| ID | Date | Author | Operational Mode | Subject | Text |  | 
| 286 | Fri Nov 12 16:54:50 2010 | Doug Preddy | Normal Running | Solenoid ramped up |  |  | 
| 285 | Thu Nov 11 15:46:40 2010 | Doug Preddy | Other | switched to parallel cold return | At 14:45. |  | 
| 284 | Mon Aug  9 14:27:12 2010 | Doug Preddy | Shut Down | Helium plant shut down |  |  | 
| 283 | Mon Jun  7 09:24:47 2010 | Doug Preddy | Quench Recovery | started flow to power leads |  |  | 
| 282 | Tue May 18 15:04:36 2010 | Kevin Trithardt | Maintenance | Solenoid ramping up |  |  | 
| 281 | Thu May 13 14:37:13 2010 | Doug Preddy | Quench Recovery | switching to series mode. |  |  | 
| 280 | Thu May 13 07:39:55 2010 | Doug Preddy | Quench Recovery | solenoid quench last night | Solenoid quenched on power lead temps last evening. I am recovering it in parallel cold
 return mode. C12~60K C1~50K.
 |  | 
| 279 | Mon May  3 07:49:54 2010 | Doug Preddy | Quench Recovery | Solenoid quench Saturday | Solenoid quenched at ~13:30 Saturday. Kevin came in later in the day to recover it. System
 was in all around mode, no flow to solenoid.
 |  | 
| 278 | Fri Apr  9 16:07:28 2010 | Doug Preddy | Start Up | flow to power leads started |  |  | 
| 277 | Fri Apr  9 07:34:55 2010 | Doug Preddy | Quench Recovery | switched to parallel cold return | Switched to parallel cold return at about 7:30 this morning
 
 |  | 
| 276 | Thu Apr  1 11:57:55 2010 | Doug Preddy | Start Up | started flow to power leads | Solenoid cold. Buffer tank pressure ~3psi, so Kevin will add a bottle of helium. Started
 flow to power leads at 11:45
 |  | 
| 275 | Mon Mar 29 09:25:52 2010 | Kevin Trithardt | Start Up | Started cool down this morning |  |  | 
| 274 | Wed Mar 24 08:30:33 2010 | Doug Preddy | Start Up | Start up | Compressor started yesterday afternoon. Pumping out solenoid overnight. Cold box
 vacuum good. circulating helium to solenoid
 |  | 
| 273 | Mon Oct  5 11:54:58 2009 | Doug Preddy | Normal Running | Switched TIC121 into auto mode |  |  | 
| 272 | Tue Jul 14 15:28:36 2009 | Doug Preddy | Quench Recovery | Solenoid ramped up | The solenoid is cold and I have started to ramp up the power supply. Some slight
 ripples in the power leads, but they are
 |  | 
| 271 | Fri Jul 10 01:10:59 2009 | Doug Preddy | Quench Recovery | switch to parallel warm return |  |  | 
| 270 | Fri Jul 10 00:34:03 2009 | Doug Preddy | Quench Recovery | solewnoid quench due to power bump | Solenoid quench due to power bump caused by an airplane crash in Richmond at ~ 22:00.
 Compressor off. Water pump off. Restarted
 |  | 
| 269 | Wed May 27 15:01:34 2009 | Doug Preddy | Quench Recovery | power leads | Power leads are cold. There is still a small ripple. The power supply can be ramped
 up as soon as the ripples die out.
 |  | 
| 268 | Wed May 27 13:37:22 2009 | Doug Preddy | Quench Recovery | switch to series mode | Switched to series mode at ~11:00. Built liquid to 55% by 13:30. Started flow to the
 power leads.
 |  | 
| 267 | Wed May 27 08:06:00 2009 | Doug Preddy | Quench Recovery | Solenoid quenched at 5:00 | The solenoid quenched at 5:00 this morning due to power lead temps. The temps started
 oscillating about 10 minutes before the trip.
 |  | 
| 266 | Wed Dec 10 15:44:58 2008 | Doug Preddy | Maintenance | Repairs to solenoid power supply | Klaus and company pulled the ramp rate card out of the solenoid power supply. Operational
 checks were done and the rate switch was
 |  | 
| 265 | Wed Dec 10 15:41:04 2008 | Doug Preddy | Shut Down | shuttting down solenoid | shuttting down solenoid for winter shutdown. |  | 
| 264 | Wed Nov  5 16:33:52 2008 | Doug Preddy | Normal Running | sbplc light out | On main menu the sbplc light was black. All vacuum systems looked good, but there was
 no control of any device. Jane reset the
 |  | 
| 263 | Sun Oct  5 18:54:27 2008 | Doug Preddy | Quench Recovery | Solenoid Quench ~16:30 | Restarted compressor and flow to coldbox by 18:30
 |  | 
| 262 | Sun Oct  5 11:12:48 2008 | Doug Preddy | Normal Running | M9B:V1 interlocks and a IOC crash | M9A:V1 was removed this shutdown. It's "in" limit is required to allow M9B:V1 to be opened.
 I have bypassed the interlocks on M9B:V1
 |  | 
| 261 | Fri Sep 26 09:50:18 2008 | Doug Preddy | Quench Recovery | started flow in parallel cold return mode |  |  | 
| 260 | Fri Sep 26 09:45:18 2008 | Doug Preddy | Quench Recovery | Solenoid quenched at 7:50 | Compressor on. Turbines started. Vacuum system recovering.
 |  | 
| 259 | Fri Sep 26 09:44:25 2008 | Doug Preddy | Maintenance | Water flow interlock restored | yesterday |  | 
| 258 | Tue Sep  9 15:04:23 2008 | Doug Preddy | Maintenance | Jumpered interlock for diff pump water flow | Jumpered interlock for diff pump water flow in Proteus box.
 |  | 
| 257 | Thu Sep  4 10:02:55 2008 | Doug Preddy | Normal Running | ramping down solenoid | Solenoid power supply will be turned off for Fall shutdown.
 |  | 
| 256 | Sun Aug 24 06:45:14 2008 | Doug Preddy | Quench Recovery | switch to series mode |  |  | 
| 255 | Sat Aug 23 17:02:38 2008 | Doug Preddy | Quench Recovery | Solenoid quenched | The solenoid quenched this afternoon at ~12:30. The compressor shut down. Fault lights were
 cold box trip.
 Restarted the compressor
 |  | 
| 254 | Sat Jul 26 01:29:21 2008 | Doug Preddy | Quench Recovery | power leads | ops will monitor liquid level. When it's at 60% they will come down and open the power
 lead valve. When the temperature ripples
 |  | 
| 253 | Sat Jul 26 01:06:04 2008 | Doug Preddy | Quench Recovery | Refrigerator in series mode | Switched to series mode. Solenoid ~8K, starting to build liquid.
 |  | 
| 252 | Fri Jul 25 21:48:15 2008 | Doug Preddy | Quench Recovery | another quench | System was stable and solenoid ramped up at 4:00 today. At about 7:30 it tripped off
 again. I came in and restarted the compressor
 |  | 
| 251 | Thu Jul 24 16:28:50 2008 | Doug Preddy | Quench Recovery | power bump | There was a power bump at ~15:00 today. All mseon hall cooling water went off. Compressor
 tripped. Main magnet went off.
 Restarted
 |  | 
| 250 | Tue Jul 15 08:47:13 2008 | Doug Preddy | Normal Running | Screen grab of normal running | Solenoid was recovered on Sunday July 13. The power leads started to ripple slightly
 this morning, so I cycled HIC142. No real
 |    | 
| 249 | Sun Jul 13 20:17:24 2008 | Doug Preddy | Other | Helium refrigerator problems | In early June the helium compressor was running with the solenoid cold. A vavlve broke off
 at the suction side of the compressor. The
 |  | 
| 248 | Mon Apr 28 09:03:22 2008 | Doug Preddy | Normal Running | power leads warmed during weekend | The power lead temperatures suddenly rose on Sunday morning. No rippling occured before
 the woarm up.
 I shut flow off this morning
 |  | 
| 247 | Thu Apr 17 13:52:25 2008 | Doug Preddy | Start Up | switched to series mode at 9:30 this morning | switched to series mode at 9:30 this morning. Added two bottles of helium gas.
 Started
 flow to power leads.
 |  | 
| 246 | Wed Apr 16 16:51:16 2008 | Doug Preddy | Start Up | switched to parallel cold return |  |  | 
| 245 | Wed Apr 16 15:15:44 2008 | Doug Preddy | Start Up | refrigerator started | Turned on compressor last Tuesday. Ran purifier until today. Final value 0.15ppm.
 Started
 cool down yesterday. Switched into parallel
 |  | 
| 244 | Wed Apr 16 15:12:01 2008 | Doug Preddy | Normal Running | graphs to load | There are three strip tools saved: coils
 powerlead
 startup
 |  | 
| 243 | Tue Apr  1 15:17:06 2008 | Doug Preddy | Start Up | Starting up vacuum systems | Monday afternoon there was no control of the vacuum system. The controls were working,
 but no output to devices. The "output" breaker
 |  | 
| 242 | Mon Dec  3 14:59:58 2007 | Doug Preddy | Quench Recovery | Not really a quench recovery, But... | Not really a quench recovery, but...I wanted to send this out via e-mail.
 The contactor for the compressor building
 |  | 
| 241 | Wed Nov 28 15:07:08 2007 | Kevin Trithardt | Other | power lead regeneration | Power leads cold again, power supply ramping up.
 |  | 
| 240 | Tue Nov 27 21:02:08 2007 | Doug Preddy | Normal Running | power lead ripples | power leads were oscillating tonight. John K. came down and cycled CV142.
 (Thanks John!) Temps settling out.
 |  | 
| 239 | Mon Nov 26 09:45:43 2007 | Doug Preddy | Normal Running | power lead problems | Power lead temperatures were rippling yesterday afternoon. I asked one of
 the operators to cycle CV142.
 |  | 
| 238 | Fri Nov 23 08:41:45 2007 | Doug Preddy | Normal Running | screen grab of powerlead temps |  |   | 
| 237 | Fri Nov 23 08:35:37 2007 | Doug Preddy | Normal Running | power lead problems | About 10 hours ago the power lead temperatures started to fluctuate and
 climb.
 |  | 
| 236 | Mon Nov 19 04:30:47 2007 | Doug Preddy | Quench Recovery | switched to series mode |  |  | 
| 235 | Mon Nov 19 02:58:44 2007 | Doug Preddy | Quench Recovery | Solenoid quenched on power lead over temperature | I received a page at 1:30 this morning. The power lead temperatures were
 climbing. Cycling CV142 did not fix the problem.
 |  | 
| 234 | Fri Nov  9 18:56:20 2007 | Doug Preddy | Maintenance | System Ready to Run | M9BMP2 was replaced. Vacuum good. Solenoid is ramped up. :)
 |  | 
| 233 | Fri Nov  9 16:24:33 2007 | Doug Preddy | Maintenance | 2APLC Died | thumb wheel panel all lite up. 2APLC rebooted by Klara (Thanks). M9BMP2 not
 pumping properly. Edi is looking at it.
 |  | 
| 232 | Fri Nov  9 14:50:03 2007 | Doug Preddy | Quench Recovery | Solenoid cold | Solenoid cold. Added two bottles of helium. Allowing to stablize. Will
 ramp power supply in half an hour.
 |  | 
| 231 | Fri Nov  9 07:37:44 2007 | Doug Preddy | Quench Recovery | switched to series mode | switched to series mode. |  | 
| 230 | Thu Nov  8 09:28:53 2007 | Doug Preddy | Quench Recovery | started compressor | And started flow to coldbox,solenoid in parallel warm return.
 
 |  | 
| 229 | Wed Oct 24 12:04:06 2007 | Doug Preddy | Maintenance | power lead and cryopump regeneration | Regenerating M9B:CP2 as temperature was up to almost 20k.
 Shut off flow to power leads to allow them
 |  | 
| 228 | Tue Oct  9 13:16:44 2007 | Doug Preddy | Quench Recovery | switched to series mode |  |  | 
| 227 | Tue Oct  9 12:58:22 2007 | Doug Preddy | Quench Recovery | Solenoid Quench 2 | Refrigeration sytem shut off at ~ 12:30 today. No apparent reason. Only
 light on was a cold box fault.
 |  | 
| 226 | Fri Oct  5 13:25:41 2007 | Doug Preddy | Quench Recovery | funny graph | There was a fluctuation of the positive lead temp with no flow going to
 the power leads.
 |   | 
| 225 | Fri Oct  5 10:25:43 2007 | Doug Preddy | Quench Recovery | Unknown quench 2 | Kevin and I were at the Sulzer panel when the system quenched 2. No
 obvious reasons. Restarted the system.
 |  | 
| 224 | Tue Oct  2 13:02:37 2007 | Doug Preddy | Normal Running | PLC problems | While downloading new code for 2A PLC the M9 and M15 vacuum systems were
 shut off. This is due to the two systems
 |  | 
| 223 | Thu Sep 27 07:41:15 2007 | Doug Preddy | Start Up | Solenoid cold | Solenoid cold and in series mode. 50% liquid level. Starting flow to
 powerleads.
 |  | 
| 222 | Wed Sep 26 08:52:46 2007 | Doug Preddy | Start Up | switched to parallel cold return |  |  | 
| 221 | Tue Sep 25 07:34:04 2007 | Doug Preddy | Start Up | PIC229 to 11 bar | adjusted turbine speeds and HIC104 to maintain their values.
 |  | 
| 220 | Mon Sep 24 15:08:41 2007 | Doug Preddy | Start Up | Started cool down this morning | The vacuum in the cold box was good on Friday 0.0 X 10-4. Let it pump over
 the weekend. Added two bottles of UHP helium
 |  | 
| 219 | Wed Sep 19 20:53:18 2007 | Doug Preddy | Maintenance | More diff pump news | Cold box pumped out to ~ 190 mTorr overnight.Leak checked the pump again.
 Still leaks at dip stick.Tried resetting
 |  | 
| 218 | Tue Sep 18 18:48:11 2007 | Curtis Ballard | Maintenance | cold box leak check | The leak check of the cold box diff pump fittings. Found a leak at the
 base of the diff pump on the tee that is
 |  | 
| 217 | Fri Sep 14 12:44:52 2007 | Doug Preddy | Start Up | started compressor | running purifier over weekend |  | 
| 216 | Wed Sep 12 12:01:02 2007 | Doug Preddy | Maintenance | regenerated adsorber | regenerated adsorber replaced MV902 "ON" switch (used nitrogen
 precooling on switch)
 |  | 
| 215 | Tue Aug 14 12:55:12 2007 | Doug Preddy | Normal Running | Strip tool saves | There are two strip tools saved: coils
 powerlead
 |  | 
| 214 | Tue Aug 14 10:15:22 2007 | Doug Preddy | Maintenance | Solenoid ramped down | Solenoid ramped down and locked out for M9A slits work.
 |  | 
| 213 | Mon Aug  6 08:41:31 2007 | Doug Preddy | Quench Recovery | Cold box trip at 6:00 | compressor off due to cold box trip at 6:00 this morning. Restarted system
 and started flow to solenoid in parallel
 |  | 
| 212 | Tue Jul 31 07:53:57 2007 | Kevin Trithardt | Normal Running | cycled power leads | got a page for the power lead temp came down and cycled the power leads.. did
 not seem to make a difference..
 |  | 
| 211 | Sun Jul 29 04:58:27 2007 | Doug Preddy | Quench Recovery | Energizing solenoid | LHe level~50%. Still have a slight ripple in the power leads, but they are
 cold enough to power the solenoid. LHe_in
 |  | 
| 210 | Sat Jul 28 22:30:13 2007 | Doug Preddy | Quench Recovery | quench again | recovered system and energized solenoid at about 6:00 this evening.
 At 9:30 power lead temps quenched solenoid
 |  | 
| 209 | Sat Jul 28 15:45:17 2007 | Doug Preddy | Quench Recovery | quench recovery | The solenoid tripped off this morning due to pwer lead over temp. The
 compressor was also off. I restarted this
 |  | 
| 208 | Thu Jul  5 11:01:47 2007 | Kevin Trithardt | Other | Current lead warm up | Due to curent lead temperature occilation Doug recommended warming up the
 current leads.
 |  | 
| 207 | Thu Jul  5 03:04:50 2007 | Doug Preddy | Normal Running | Power lead problems | The power leads temps are starting to cycle again. Earlier today M9
 experimentor cycled HIC142 to stablize the
 |  | 
| 206 | Wed Jul  4 11:13:12 2007 | Doug Preddy | Quench Recovery | test | This is a test to make sure the e-mail works. |  | 
| 205 | Sun Jul  1 16:09:37 2007 | Doug Preddy | Normal Running | M9B:DP2 tripped off | Diffusion pump tripped off and vacuum (and power lead temps) going up. I
 had an experimentor cycle HIC 142 at 4:15
 |  | 
| 204 | Tue Jun 19 21:15:41 2007 | Doug Preddy | Quench Recovery | Today's quench | power lead temps caused a quench at 8:30 this morning. Recovery of system
 and flow to solenoid was started by 11:00.
 |  | 
| 203 | Sat Jun 16 15:05:03 2007 | Doug Preddy | Normal Running | Solenoid power lead temp problem | Power lead temps rippling and increasing. I asked an operator to cycle
 HIC142. This seems to have fixed the problem
 |  | 
| 202 | Thu Jun 14 00:10:10 2007 | Doug Preddy | Quench Recovery | recovering from earlier quench | Solenoid quenched on power lead temperature last night at about 6:00. I
 restarted flow to the solenoid in parallel
 |  | 
| 201 | Tue Jun 12 12:48:25 2007 | Doug Preddy | Maintenance | added one bottle UHP helium to system |  |  | 
| 200 | Thu May 31 11:51:49 2007 | Doug Preddy | Quench Recovery | Switched to series mode |  |  | 
| 199 | Thu May 31 07:11:05 2007 | Doug Preddy | Quench Recovery | switched to parallel cold return | switched to parallel cold return earlier this morning. Vacuum is 5 x 10-5.
 |  | 
| 198 | Wed May 30 11:27:32 2007 | Doug Preddy | Quench Recovery | Problems recovering | Various start up problems : MV51 stuck open. Finally got it closed by
 hitting the valve with a hammer.
 |  | 
| 197 | Wed May 30 08:38:38 2007 | Doug Preddy | Quench Recovery | Solenoid quenched this morning | Solenoid quenched this morning on power lead temp (I believe).
 I restarted it at ~7:00, but the capacity
 |  | 
| 196 | Wed May 16 10:02:35 2007 | Doug Preddy | Maintenance | warming up power leads | Power lead temps have been slowly increasing over the last wekk. I ramped
 the solenoid down and am warming the power
 |  | 
| 195 | Wed May  2 16:28:16 2007 | Doug Preddy | Normal Running | Energizing solenoid to 500 amps | Power supply initially would not ramp up. I exercised the rate control
 selector switch and found that the power
 |  | 
| 194 | Mon Apr 30 09:30:26 2007 | Doug Preddy | Normal Running | problems over weekend | The PLC Limit switch power supply was tripped off on Saturday. I came in
 and reset it and restarted all the vacuum
 |  | 
| 193 | Fri Apr 27 10:32:07 2007 | Doug Preddy | Quench Recovery | Reset quench relay logic box | The quench relay logic controller was not functioning correctly. It showed
 a "Trip" light and no other lights. I cycled
 |  | 
| 192 | Fri Apr 27 10:16:44 2007 | Kevin Trithardt | Quench Recovery | Test | Don't Panic -- There was no quench. I am testing the e-log.
 I have added Kevin's name and e-mail to the
 |  | 
| 191 | Fri Apr 27 10:10:07 2007 | Doug Preddy | Normal Running | Solenoid Cold | Solenoid cold, power leads cold. Alarm handler started.
 |  | 
| 190 | Tue Apr 24 10:21:49 2007 | Doug Preddy | Start Up | Started flow to turbines | Started flow to turbines at about 10:00 in parallel warm mode.
 |  | 
| 189 | Mon Apr 23 16:31:02 2007 | Doug Preddy | Start Up | Started compressor | Startewd compressor and purifer today. |  | 
| 188 | Fri Apr 20 12:15:12 2007 | Doug Preddy | Start Up | remote PLC switch | I could not control the vacuum devices using EPICS. The buttons could be
 pushed, but nothing would respond. The PLC
 |  | 
| 187 | Thu Apr 19 15:18:25 2007 | Doug Preddy | Start Up | preparing to start the refrigerator | Starting the vacuum systems in the coldbox and solenoid.
 |  | 
| 186 | Mon Mar  5 14:59:41 2007 | Doug Preddy | Maintenance | Password | On the weekend someone used this log to send spam messages. I have
 installed a password for the write privilege.
 |  | 
| 185 | Fri Jan 26 15:10:37 2007 | Doug Preddy | Maintenance | new control computer installed | I have installed a new control computer at the M9 solenoid control panel
 |  | 
| 184 | Tue Nov 28 10:32:04 2006 | Doug Preddy | Quench Recovery | recovering after looong power outage | The power has been out for about 24 hours. Recovering coldbox, solenoid and
 beamline vacuum.PLC is down for the vacuum
 |  | 
| 183 | Wed Nov 15 17:05:49 2006 | Willy Andersson | Quench Recovery | powerleads cold, energizing solenoid | Powerlead temperatures have bottomed out, energizing solenoid to 300
 amps, ramp rate 60.
 |  | 
| 182 | Wed Nov 15 13:56:00 2006 | Doug Preddy | Quench Recovery | switch to series mode |  |  | 
| 181 | Wed Nov 15 11:09:28 2006 | Doug Preddy | Quench Recovery | Compressor tripped off | Compressor tripped this morning at ~8:45. I restarted the compressor, but
 turbine 1 would not spin up. I let it warm
 |  | 
| 180 | Tue Nov  7 10:35:32 2006 | Doug Preddy | Normal Running | Maintenance day | I ramped down the solenoid power suplly and shut off flow to the power
 leads. I will allow them to warm up and (hopefully)
 |  | 
| 179 | Fri Nov  3 12:09:51 2006 | Doug Preddy | Normal Running | Fine tuning | The power lead temperatures have been climbing at a rate of ~ 0.02K very 8
 hours. The negative lead started to show
 |  | 
| 178 | Tue Oct 31 08:37:34 2006 | Doug Preddy | Quench Recovery | Energizing solenoid |  |  | 
| 177 | Tue Oct 31 05:36:14 2006 | Doug Preddy | Quench Recovery | switch to series mode |  |  | 
| 176 | Tue Oct 31 03:52:29 2006 | Doug Preddy | Quench Recovery | power lead quench at 2:30 this morning | Compressor still on. Starting flow to solenoid |  | 
| 175 | Thu Oct 26 14:14:57 2006 | Doug Preddy | Quench Recovery | power lead quench | solenoid tripped on power lead temp. I was here and caught it. System
 cooling.
 |  | 
| 174 | Thu Oct 26 11:45:22 2006 | Doug Preddy | Quench Recovery | Solenoid cold | power leads cold 50% liquid. Turning on power supply
 |  | 
| 173 | Thu Oct 26 08:18:10 2006 | Doug Preddy | Quench Recovery | switched to series mode | Switched to series mode at about 7:45. |  | 
| 172 | Thu Oct 26 04:34:15 2006 | Willy Andersson | Quench Recovery | Still in // cold return mode. | Leaving system cooling in // cold return mode, cooldown rates still good.
 Had flow on to powerleads since re-starting
 |  | 
| 171 | Thu Oct 26 04:27:31 2006 | Willy Andersson | Quench Recovery | Powerlead overtemp | 02:30 this morning got page from operations. Looked at system online, appears that there
 was a powerlead overtemp
 |  | 
| 170 | Sun Oct 15 07:09:41 2006 | Other | Normal Running | cryopump bcp2 trip | Donald Arseneau reporting 
 Not really solenoid realated, but FYI...
 |  | 
| 169 | Fri Oct 13 00:13:57 2006 | Experimentor | Quench Recovery | Energized solenoid. | Hi there, 
 This is Khashayar (experimenter). Lookin
 |  | 
| 168 | Thu Oct 12 19:54:10 2006 | Willy Andersson | Quench Recovery | Energized solenoid. | Switched to series mode about 6 hours ago. Cooling continued including
 power leads until LHe level at about 50%
 |  | 
| 167 | Thu Oct 12 11:53:04 2006 | Willy Andersson | Quench Recovery | Compressor trip, 09:50 this morning. | Main compressor tripped, refrigerator shutdown. Checked with experimentors
 and main control room, no indications of
 |  | 
| 166 | Thu Oct  5 14:53:54 2006 | Doug Preddy | Start Up | Switched to series mode |  |  | 
| 165 | Thu Oct  5 11:51:53 2006 | Doug Preddy | Start Up | Adde 2 bottles UHP Helium | Valved off purifiers, but left LN2 on. |  | 
| 164 | Thu Oct  5 10:26:26 2006 | Doug Preddy | Start Up | switched to parallel cold return |  |  | 
| 163 | Wed Oct  4 18:10:18 2006 | Willy Andersson | Start Up | Quick check on system | System in good // warm return mode, temps coming down nicely. BCP2 cryopump
 does not look too healthy, temperature is
 |  | 
| 162 | Wed Oct  4 07:29:04 2006 | Doug Preddy | Start Up | Starting cool down | Started compressor yesterday with flow through the purifier overnight.
 Starting tubines this morning in parallel
 |  | 
| 161 | Thu Aug 17 19:45:27 2006 | Willy Andersson | Quench Recovery | switch to series cooling | Temps of flow returning from solenoid flat lined at about 10K.
 Gradually closed CV155.
 |  | 
| 160 | Thu Aug 17 08:04:44 2006 | Doug Preddy | Quench Recovery | Solenoid tripped at about 3:00 this morning | restarting system. Compressor would not start. Idid not see
 anything that would not let it
 |  | 
| 159 | Tue Aug 15 15:39:27 2006 | Willy Andersson | Quench Recovery | switched to parallel cold return | Return flow temps from solenoid about 40K, TIC121 at about 40K so switching
 to //cold return.
 |  | 
| 158 | Tue Aug 15 08:45:56 2006 | Doug Preddy | Quench Recovery | restarted the compressor | Solenoid tripped off Sunday. Restarted compressor and cool down of cold box.
 |  | 
| 157 | Thu Aug 10 20:26:09 2006 | Willy Andersson | Quench Recovery | solenoid energized | Powerleads seem stable. |  | 
| 156 | Thu Aug 10 16:33:11 2006 | Willy Andersson | Quench Recovery | Flow to solenoid tripped on powerlead temperature. | Power lead temps were drifting up, tried to tweak TIC121 and HIC141 to
 unstall powerlead flow. Got quench. Restarted
 |  | 
| 155 | Thu Aug 10 08:33:44 2006 | Doug Preddy | Normal Running | Running low on helium | Temperture okay, but there are fluctuation in the power leads. PIC229 is
 trying to charge. The buffer tank pressure
 |  | 
| 154 | Wed Aug  9 07:50:49 2006 | Doug Preddy | Quench Recovery | Switched to series mode. | Cool down continues. |  | 
| 153 | Wed Aug  9 06:59:43 2006 | Willy Andersson | Quench Recovery | more // cold return stuff | About 21:00 last night the solenoid temps had not changed much, pretty well
 flat lined. I guess my reducing CV155 stalled
 |  | 
| 152 | Wed Aug  9 06:44:58 2006 | Willy Andersson | Quench Recovery | // cold mode cooling continues. | After restarting flow to the solenoid yesterday in // cold return mode
 TIC121 just refused to go lower than about
 |  | 
| 151 | Tue Aug  8 07:27:49 2006 | Doug Preddy | Quench Recovery | Solenoid Quench at about 6:00 this morning | No indication of problem. Turbines still running, but no flow to solenoid.
 Started flow. Solenoid ~20K.
 |  | 
| 150 | Sat Jul 29 09:38:06 2006 | Doug Preddy | Quench Recovery | power leads cooled | opened power lead flow about 40 minutes ago. ~12K will let them stablize.
 LHE level >20%
 |  | 
| 149 | Sat Jul 29 05:30:28 2006 | Doug Preddy | Quench Recovery | switched to series mode | Temps bottomed out at ~16K. Switching to series mode.
 |  | 
| 148 | Fri Jul 28 20:39:40 2006 | Doug Preddy | Quench Recovery | Restarting refrigerator | Solenoid quenched at about 6:00 tonight. I restarted the compressor, then the cold
 box and turbines. When cold box
 |  | 
| 147 | Sat Jul 22 06:34:14 2006 | Experimentor | Quench Recovery | looking good | Donald Arseneau reporting. 
 It is looking pretty good, so I came back
 |  | 
| 146 | Fri Jul 21 23:36:21 2006 | Doug Preddy | Quench Recovery | Final cooldown problems | 80K shield is down to <60K Liquid level not building. Power leads rippling
 and warming. T1 inlet 85K with 10%heater
 |  | 
| 145 | Fri Jul 21 17:18:19 2006 | Willy Andersson | Quench Recovery | Opened HIC142 fully open | Power leads have flatlined but still building LHe level.
 |  | 
| 144 | Fri Jul 21 14:08:04 2006 | Doug Preddy | Quench Recovery | switch to series mode | Coils temps <10K.Switching to series mode. |  | 
| 143 | Fri Jul 21 07:39:37 2006 | Doug Preddy | Quench Recovery | Cool down continues | Raised PIC229 to 11Bar. Started to close HIC155. Cooling rates improve.
 |  | 
| 142 | Thu Jul 20 13:34:26 2006 | Willy Andersson | Quench Recovery | cold return mode continues... | Even though cr_c1 and cr_c12 have begun to tick down, cr_c7 and cr_c6 are
 actually ticking up. This is probably a wave
 |  | 
| 141 | Thu Jul 20 13:27:21 2006 | Willy Andersson | Quench Recovery | cold return mode continues... | cr_c1 and cr_12 are both at about 45K and heading slowly down, so the
 gamble of going to // cold return worked.
 |  | 
| 140 | Thu Jul 20 13:21:57 2006 | Willy Andersson | Quench Recovery | started // warm return flow. | TIC121 was down to 27K warm return valves were wide open
 opened supply valves CV143 and CV146.
 |  | 
| 139 | Thu Jul 20 11:01:09 2006 | Willy Andersson | Quench Recovery | Opened up warm return valves CV151 and CV153 | Mu2 pressure was just about at trip point due to trapped volume of helium
 warming. So opened up warm returns to bleed
 |  | 
| 138 | Thu Jul 20 08:18:31 2006 | Doug Preddy | Quench Recovery | T1 started | Started flow to turbines. Both started. T1 seems a bit slow. HIC110 at 0.85
 SI110 = 4100.
 |  | 
| 137 | Thu Jul 20 02:59:06 2006 | Doug Preddy | Quench Recovery | T1 still not starting | I will leave the cold box to warm overnight and try it again in the morning.
 |  | 
| 136 | Thu Jul 20 00:52:13 2006 | Doug Preddy | Quench Recovery | T1 frozen | T1 will not spin up. Warming up cold box. |  | 
| 135 | Thu Jul 20 00:25:16 2006 | Doug Preddy | Quench Recovery | Solenoid quenched @ 11:15 | Solenoid Quenched on C1 temp. It went above 5K. Shut off compressor.
 Restarting system @ 12:15
 |  | 
| 134 | Thu Jul  6 15:04:17 2006 | Doug Preddy | Normal Running | test | test |   | 
| 133 | Thu Jul  6 15:00:15 2006 | Doug Preddy | Normal Running | test |  |  | 
| 132 | Mon Jul  3 23:01:34 2006 | Doug Preddy | Quench Recovery | power lead warming/ripples caused quench | power lead warming/ripples caused quench at 18:45 today. Compressor stayed
 on, but flow to solenoid stopped.
 |  | 
| 131 | Fri Jun 30 11:17:57 2006 | Doug Preddy | Normal Running | Solenoid recovered | Tuesday night. June 27, 2006, the solenoid quenched for no apparent
 reason. It was restarted Wednesday morning.
 |  | 
| 130 | Fri Jun 30 11:13:00 2006 | Doug Preddy | Normal Running | E-log is back up and running | I would have made an entry when it died, but----- |  | 
| 129 | Fri Jun 16 08:12:04 2006 | Doug Preddy | Normal Running | LI108 indicates high on EPICS screen | LI108 indicates high on EPICS screen (150%) Controller shows ~64%
 |  | 
| 128 | Wed Jun 14 16:27:01 2006 | Willy Andersson | Quench Recovery | Energized solenoid | Powerleads appear stable. |  | 
| 127 | Wed Jun 14 10:35:30 2006 | Willy Andersson | Quench Recovery | Switched to series cooling. | Had system in //return cooling overnight with small flow to keep power
 leads cool. Switched to series mode about
 |  | 
| 126 | Tue Jun 13 16:04:33 2006 | Willy Andersson | Quench Recovery | TIC 121 set to auto mode |  |  | 
| 125 | Tue Jun 13 15:52:46 2006 | Willy Andersson | Quench Recovery | Doug has reduced and now closed CV155. | Now in full // cold return mode. Had just talked with Curtis, He said thta
 he has already informed the
 |  | 
| 124 | Tue Jun 13 10:54:27 2006 | Doug Preddy | Quench Recovery | Compressor tripped off | No indicator lights etc as to why the compressor tripped. The only  light
 was a cold box fault. I had to shut off the
 |  | 
| 123 | Tue May 30 14:36:41 2006 | Doug Preddy | Normal Running | power leads warming | power leads warming. closed HIC 142, then re-opened. temps falling
 |  | 
| 122 | Fri May 26 14:34:24 2006 | Doug Preddy | Start Up | Ready to run | Solenoid is cold, liquid level is 35%, power leads cold and seem stable.
 Added one bottle helium to buffer tank. Pressure
 |  | 
| 121 | Fri May 26 10:03:03 2006 | Doug Preddy | Start Up | Started flow to power leads | Started flow to power leads. Liquid level 15%
 |  | 
| 120 | Fri May 26 08:45:38 2006 | Doug Preddy | Start Up | Changed to series mode | Changed to series mode. Liquid level building.
 |  | 
| 119 | Fri May 26 07:29:54 2006 | Doug Preddy | Start Up | CV155 closed | CV155 closed. Temps about 15K
 |  | 
| 118 | Thu May 25 07:34:18 2006 | Doug Preddy | Start Up | switched to parallel cold return | switched to parallel cold return |  | 
| 117 | Tue May 23 08:44:28 2006 | Doug Preddy | Start Up | Started turbines | Solenoid and cold box were pumped out last Tuesday. Cold box roughing pump
 initially would not start. had to retighten
 |  | 
| 116 | Thu Dec 22 15:10:44 2005 | Doug Preddy | Shut Down | shut down system | shuttting down the system. isolated cold box by closing v135. preparing to
 pump out cold box.
 |  | 
| 115 | Mon Dec 12 09:24:29 2005 | Doug Preddy | Quench Recovery | Solenoid Quench on power lead temp | The solenoid tripped on power lead temps at about 9:15. I am recovering it
 from it's all around mode into parallel cold
 |  | 
| 114 | Tue Nov 15 07:25:10 2005 | Doug Preddy | Normal Running | power supply problem | Saturday November 12 the solenoid power supply ramped itself down to zero.
 It would not ramp up unless the ramp rate
 |  | 
| 113 | Thu Nov  3 09:27:16 2005 | Doug Preddy | Quench Recovery | started flow to solenoid | started flow to solenoid in parallel cold return.
 |  | 
| 112 | Thu Nov  3 06:19:50 2005 | Doug Preddy | Quench Recovery | restart system | Last night at about 7:00 there was a large power bump that shut the
 compressor and water pump off. I came in
 |  | 
| 111 | Wed Nov  2 10:17:27 2005 | Willy Andersson | Start Up | M9B helium refrigerator and solenoid cooldown | Started M9B MYCOM helium compressor circulating helium through system last
 Friday. Right hand purifier was regenerated
 |  | 
| 110 | Mon Oct 31 07:48:57 2005 | Willy Andersson | Quench Recovery | what a fine entry | Keep IT safe boys |  | 
| 109 | Sun Aug 14 10:03:13 2005 | Willy Andersson | Normal Running | System stable | No experiment running in M9B. 
 Solenoid current is zero and power supply
 |  | 
| 108 | Tue Jul 19 18:09:56 2005 | Willy Andersson | Quench Recovery | Quench II | Machine trip just as I was exiting the TRIUMF gate. QRL said cold box fault
 as did red lite on compressor. At cold box
 |  | 
| 107 | Tue Jul 19 16:13:32 2005 | Willy Andersson | Quench Recovery | switch to series mode | CR_6KS_M down to 11 K. Switched to series mode.
 Started small flow to power leads to get
 |  | 
| 106 | Mon Jul 18 15:51:17 2005 | Doug Preddy | Quench Recovery | Compressor restarted, flow to solenoid | Recovered the vacuum in the solenoid, then started the compressor.
 Started flow to solenoid in parallel warm
 |  | 
| 105 | Mon Jul 18 11:18:40 2005 | Doug Preddy | Quench Recovery | Solenoid Quenched over the weekend | The solenoid was found to be warm on Monday morning. M9Q3 had a bad water
 leak over the weekend. The water header serving
 |  | 
| 104 | Tue Jun 14 23:01:22 2005 | Doug Preddy | Quench Recovery | switched to parallel cold return |  |  | 
| 103 | Tue Jun 14 20:53:19 2005 | Doug Preddy | Quench Recovery | restarted refrigerator after power bump | see bitmap for temps/pressures after presurrizing coldbox. Preparing to
 start flow to turbines, then will start flow
 |   | 
| 102 | Tue Jun 14 18:48:14 2005 | Experimentor | Normal Running | Power bump at 18:40 | Solenoid tripped on a power bump at 18:40 Tuesday
 -- Donald Arseneau
 |  | 
| 101 | Thu Jun  9 21:41:25 2005 | Willy Andersson | Quench Recovery | Trying to achieve stable cooling. | From about 08:00 to 15:00 
 See Doug's previous entry for quench recovery
 |  | 
| 100 | Thu Jun  9 08:56:10 2005 | Doug Preddy | Quench Recovery | system back on | The compressor was restarted and the solenoid is cooling (sort of) in
 parallel cold return.
 |  | 
| 99 | Thu Jun  9 06:53:40 2005 | Experimentor | Other | Trip at 6:45 Today. | Solenoid quenched at 6:45.  All temperature monitors were flat until trip.
 Hearing compressor stop was first warning.
 |  | 
| 98 | Thu Jun  2 09:16:40 2005 | Doug Preddy | Quench Recovery | continuing to cool | See screen grab |   | 
| 97 | Thu Jun  2 01:38:25 2005 | Doug Preddy | Quench Recovery | strip tool plot | This is a strip tool  plot of the cooldown |   | 
| 96 | Thu Jun  2 01:26:30 2005 | Doug Preddy | Quench Recovery | started flow to solenoid | See attached file for a screen grab of temps. |   | 
| 95 | Wed Jun  1 23:48:02 2005 | Doug Preddy | Quench Recovery | the first quench | I recieved a call from ops at 22:00 tonight that the solenoid had alarmed.
 I came in and found a cold box fault, with
 |  | 
| 94 | Wed Jun  1 08:23:19 2005 | Doug Preddy | Start Up | start up notes | The solenoid was pumped out Monday and yesterday after the quads just
 downstream of the solenoid were put back
 |  | 
| 93 | Thu May 19 15:05:31 2005 | Doug Preddy | Maintenance | STA 120 replaced | The turbine 2 speed indicator control box STA 120 has been replaced with a
 newer style Jaquet controller. This is similar
 |   | 
| 92 | Mon Apr 18 09:52:11 2005 | Doug Preddy | Maintenance | Relays replaced | The following relays have been replaced in the Sulzer control panel:
 K341, K379, K132, K324A, K314, K125, K127,
 |  | 
| 91 | Fri Mar 11 09:18:30 2005 | Doug Preddy | Maintenance | services on system | The cold box was isolated and pumped out after it warmed up. It has been
 refilled with helium.
 |  | 
| 90 | Sun Dec  5 11:15:24 2004 | Doug Preddy | Normal Running | more ripples | I loked at the power lead temps this morning. LHe_in 10.4-12.0  PL-Neg 8.0-
 9.0  PL-Pos 9.8-10.4. Flows down to 160mm.
 |  | 
| 89 | Thu Dec  2 10:51:16 2004 | Doug Preddy | Normal Running | power lead ripple | Ripples got worse overnight.LHe in 10.3-11.7, PL Neg 8.0-8.7. Flows <200mm.
 cycled CV142. Flows 240mm. LHe in 10.3-11.3
 |  | 
| 88 | Wed Dec  1 12:55:56 2004 | Doug Preddy | Normal Running | added one bottle helium gas | buffer tank pressure is up to 6.5 psi. |  | 
| 87 | Mon Nov 29 09:28:16 2004 | Doug Preddy | Normal Running | Buffer  tank empty | Came down to check on system. CV211 was fully open, trying to charge the
 system. Ripples on LHe in and PL neg. CV141
 |  | 
| 86 | Sun Nov 28 10:15:18 2004 | Willy Andersson | Quench Recovery | Powerlead quench ? | Got page very early this morning. Couldn't come in til 08:30. System was
 in idle running mode and someone had reset
 |  | 
| 85 | Thu Nov 25 07:21:37 2004 | Doug Preddy | Quench Recovery | started power lead flow | Curtis switched to series mode last night at 7:00.
 Startedflow to power leads at 7:00 this morning.
 |  | 
| 84 | Wed Nov 24 14:58:42 2004 | Doug Preddy | Quench Recovery | today | Came in this morning to find the helium return line to the buffer tank
 frosted up. MV51 has been stuck open the
 |  | 
| 83 | Wed Nov 24 07:28:38 2004 | Other |  | Mission Impossible |  |  | 
| 82 | Wed Nov 24 07:28:30 2004 | Other |  | Mission Impossible |  |  | 
| 81 | Tue Nov 23 15:10:06 2004 | Doug Preddy | Quench Recovery | switched to parallel cold return mode |  |  | 
| 80 | Tue Nov 23 11:35:44 2004 | Doug Preddy | Quench Recovery | Another power bump | Compressor tripped off again due to a power bump. Meson hall water off, so
 vacuum pumps off again.
 |  | 
| 79 | Tue Nov 23 09:30:15 2004 | Doug Preddy | Quench Recovery | started flow to solenoid | Started flow to solenoid in parallel warm return.
 |  | 
| 78 | Tue Nov 23 07:47:00 2004 | Doug Preddy | Quench Recovery | Solenoid quench Nov 21 | Solenoid quenched Nov 21 due to alarge power bump. Everything tripped off
 including vacuum and water pumps. Cold box
 |   | 
| 77 | Thu Nov  4 13:33:49 2004 | Doug Preddy | Normal Running | Power supply problems | Last night the power supply was ramping up at a very slow rate. This
 morning it was energized,but with no current
 |  | 
| 76 | Mon Nov  1 07:17:44 2004 | Doug Preddy | Normal Running | power lead flow gone bad | power temps as follows: LHE_IN 23
 PL_POS 36
 |  | 
| 75 | Fri Oct 29 14:25:23 2004 | Doug Preddy | Normal Running | power lead flows dropping | Flow is down to <190 mm. temps are starting to climb. Cycled HIC142
 |  | 
| 74 | Tue Oct 26 09:45:24 2004 | Doug Preddy | Normal Running | power leads warming | Flow to power leads dropped to <80 mm. Cycled HIC142. Temps dropping, flows
 >250.
 |  | 
| 73 | Mon Oct 25 15:44:15 2004 | Doug Preddy | Normal Running | Power lead temps starting to ripple | Power lead temps starting to ripple at a 1.5min period. PL flows are
 both>250 mm.
 |  | 
| 72 | Fri Oct 22 09:15:06 2004 | Doug Preddy | Start Up | started flow to power leads | Started flow to power leads HIC142 at 0.8. 
 |  | 
| 71 | Thu Oct 21 15:41:29 2004 | Doug Preddy | Start Up | switched to series mode |  |  | 
| 70 | Thu Oct 21 07:16:35 2004 | Doug Preddy | Start Up | switched to parallel cold return |  |  | 
| 69 | Wed Oct 20 09:38:17 2004 | Doug Preddy | Start Up | restarting system | Solenoid tripped  off Monday morning due to high compressor pressure.
 Regenerated purifier on Tuesday. Started
 |  | 
| 68 | Sun Oct 17 05:03:13 2004 | Other | Start Up | Cycle valve after oscillations begin | (Sat night/Sun AM) Helium temperature started oscillating, so
 powered off magnet (precautionary) and cycled
 |  | 
| 67 | Sun Oct 17 04:26:29 2004 | Other | Start Up | cycling 142 when temp hit 10.68 | > I tried cycling the valve twice -- at about 11:30 and again around 1:00,
 > trying both the buttons and the dial --
 |  | 
| 66 | Sat Oct 16 17:18:50 2004 | Other |  | lame |  |  | 
| 65 | Sat Oct 16 17:13:19 2004 | Other |  | s | s |  | 
| 64 | Wed Oct 13 08:17:00 2004 | Experimentor | Start Up | cycling 142 when temp hit 10.68 | I tried cycling the valve twice -- at about 11:30 and again around 1:00,
 trying both the buttons and the dial -- but
 |  | 
| 63 | Tue Oct 12 13:00:08 2004 | Curtis Ballard | Start Up | cv142 valve | Curtis closes and opens slowly at about 11:00 am this morning. Wants to see
 if that works better than the sudden on/off
 |  | 
| 62 | Tue Oct 12 02:05:31 2004 | Experimentor | Start Up | cv142 valve | after only two hours since last lead cycling, they warm up quickly.
 ops cycles the valve, but more quickly this
 |  | 
| 61 | Mon Oct 11 22:32:11 2004 | Experimentor | Start Up | cv142 valve | temp of leads begins its exponential rise again (caught it at 12K), approx
 11 hours after last reset.  this time asked
 |  | 
| 60 | Mon Oct 11 11:49:34 2004 | Experimentor | Start Up | cv142 valve | noticed temp rise, which occured only about 5 hours after the
 last "reset" .... the time between resets
 |  | 
| 59 | Mon Oct 11 03:56:14 2004 | Experimentor | Start Up | cv142 valve | noticed that the lead temp rose to 24K over the past 4 hours. had control
 room personnle cycle the cv142 valve. temperature
 |  | 
| 58 | Sun Oct 10 17:36:35 2004 | Experimentor | Start Up | power leads | At 5:30 pm sundat, power lead temp is about 10.6K.  script tool still
 notworking
 |  | 
| 57 | Sat Oct  9 22:50:26 2004 | Other |  | power leads warmed | At 10:30 PM Saturday, Donald Arseneau sees power leads at
 25K, having gone up for the previous 10 hours.
 |  | 
| 56 | Thu Oct  7 13:33:20 2004 | Doug Preddy | Start Up | turned off purifier | turned off purifier |  | 
| 55 | Wed Oct  6 09:25:54 2004 | Doug Preddy | Start Up | switch to series mode | switched to series mode |  | 
| 54 | Wed Oct  6 07:57:01 2004 | Doug Preddy | Start Up | parallel cold return | Switched to parallel cold return at 4:00 yesterday afternoon.
 6:30 this morning started closing cv155 as
 |  | 
| 53 | Tue Oct  5 07:18:16 2004 | Doug Preddy | Start Up | adjust compressor discharge pressure | adjusted compressor discharge pressure to 11 Bar
 |  | 
| 52 | Mon Oct  4 09:16:29 2004 | Doug Preddy | Start Up | Started turbines | Started flow to the turbimes with the solenoid in parallel warm return.
 |  | 
| 51 | Fri Oct  1 11:39:55 2004 | Doug Preddy | Maintenance | Loading of compressor not working | When the compressor was restarted on Monday, the remote load button would
 not respond. The compressor was shut off.
 |  | 
| 50 | Mon Sep 13 13:56:06 2004 | Doug Preddy | Shut Down | compressor off | compressor off |  | 
| 49 | Mon Sep 13 10:55:25 2004 | Doug Preddy | Shut Down | Quench and shutdown | The solenoid quenched on powerlead temps this morning at 10:00. It is
 warming up and the power leads are >200K.Isolating
 |  | 
| 48 | Fri Sep  3 01:28:43 2004 | Doug Preddy | Quench Recovery | power up | started flow to the power leads at 11:30. Cold and stable at 1:00am.
 Energized solenoid at 1:30.
 |  | 
| 47 | Thu Sep  2 21:58:34 2004 | Willy Andersson | Quench Recovery | Switch to series cooling | Solenoid had tripped on powerlead overtemp at about 09:00 this morning.
 Doug recovered the system to // cooling.
 |  | 
| 46 | Sun Aug 29 14:39:19 2004 | Doug Preddy | Normal Running | Power lead ripple | The LHe in and PL-neg temps have been building up a ripple over the last
 few days. Curtis cycled HIC142 Friday evening
 |  | 
| 45 | Wed Aug 25 12:52:53 2004 | Doug Preddy | Normal Running | power lead ripple | Lhe ripple has started. 10.9-11.4K PL-Pos also has a smaller ripple. Both
 flows down to 230mm. Cycled HIC 142 and the
 |  | 
| 44 | Wed Aug 25 07:47:05 2004 | Doug Preddy | Normal Running | Power lead flow | The power lead flows had dropped to 210 mm this morning. I increased HIC142
 from 0.8 to 0.83. The flows are back to <250
 |  | 
| 43 | Tue Aug 24 12:09:50 2004 | Doug Preddy | Normal Running | The past few days | The solenoid has been re-started as of August 19. It was cooled down and
 energized over the weekend. Sunday night
 |  | 
| 42 | Thu Aug 19 10:28:32 2004 | Doug Preddy | Start Up | after M9bv5 FIX | Over the last couple of days we replaced the gate on the M9BV5 valve. The
 old one had worn the teflon sliders on the
 |  | 
| 41 | Sat Aug  7 23:40:47 2004 | Willy Andersson | Quench Recovery | Energizing solenoid | 23:39 Powerleads cold, LHe level over 40% and rising. Energizing solenoid
 to 500 Amps, ramp rate 60. Checking compressor
 |  | 
| 40 | Sat Aug  7 22:37:04 2004 | Willy Andersson | Quench Recovery | Compressor tripped off. | Got a call at about 10:00 this morning, the whole system had tripped off.
 No power bumps or other disturbances. When
 |  | 
| 39 | Thu Jul 29 13:25:29 2004 | Doug Preddy | Quench Recovery | start flow to power leads |  |  | 
| 38 | Thu Jul 29 09:01:11 2004 | Doug Preddy | Quench Recovery | switched to parallel cold | Changed to series mode. |  | 
| 37 | Wed Jul 28 23:07:01 2004 | Doug Preddy | Quench Recovery | solenoid cooling | cooling rates for the solenoid are good. In parallel cold return.
 |  | 
| 36 | Wed Jul 28 22:01:31 2004 | Doug Preddy | Quench Recovery | Starting flow to solenoid | c1 28, c12 29, c6 11.6, c7 12.5 TI121 22.5 |  | 
| 35 | Wed Jul 28 21:26:19 2004 | Doug Preddy | Quench Recovery | Double Power bump | Two power bumps tonight. Everything tripped of. Restarted cooling water
 pump, diff. pump under solenoid etc. Computer
 |  | 
| 34 | Wed Jul 28 07:00:18 2004 | Doug Preddy | Quench Recovery | switch to series mode | Switching to series mode. Jumper on vacuum gauge removed.
 |  | 
| 33 | Wed Jul 28 00:04:30 2004 | Doug Preddy | Quench Recovery | started flow to solenoid | Solenoid temps were: C1 & C12 about 30K. C6 &C7 about 11K. Started flow to
 solenoid in parallel cold return. Jumpered
 |  | 
| 32 | Tue Jul 27 23:33:27 2004 | Doug Preddy | Quench Recovery | Solenoid quench 9:30 | I was phoned at about 9:30 tonight with a solenoid alarm. After looking at
 the display I found the system to be tripped
 |  | 
| 31 | Tue Jun 29 14:59:46 2004 | Willy Andersson | Normal Running | Removed jumper on vacuum controller. | Tuesday 14:55 
 Removed jumper on back of PIC 901. This jumper
 |  | 
| 30 | Fri Jun 25 14:20:58 2004 | Willy Andersson | Quench Recovery | Switch to series mode | At about 14:00 solenoid temps dipping below 8 K. Trace of LHe level (dense
 gas) starting to show. Switching to series
 |  | 
| 29 | Fri Jun 25 12:36:04 2004 | Willy Andersson | Quench Recovery | Compressor unloaded etc. | At about 09:30 this morning I noticed the sound coming from the compressor
 building seemed louder than usual. After
 |  | 
| 28 | Fri May 21 09:27:39 2004 | Doug Preddy | Maintenance | Idle mode | Power supply was ramped down yesterday as there are no users.
 The Sundance server was rebooted yesterday.
 |  | 
| 27 | Thu May  6 09:32:13 2004 | Willy Andersson | Quench Recovery | Additional notes from last night power lead problems. | When performing last nights refrigerator adjustments I also increased the
 compressor discharge pressure to 11.25 barA.
 |  | 
| 26 | Wed May  5 23:57:38 2004 | Willy Andersson | Quench Recovery | Powerleads warming up again. | Got a page from Don Flemming at about 18:50 wondering why the power
 lead temperatures were rising. See attachment
 |   | 
| 25 | Wed May  5 12:13:51 2004 | Doug Preddy | Normal Running | power leads cold | should be ready to run. |  | 
| 24 | Wed May  5 09:59:10 2004 | Doug Preddy | Quench Recovery | started flow to power leads | Started flow to power leads. TIC121 into auto mode
 |  | 
| 23 | Wed May  5 07:55:37 2004 | Doug Preddy | Quench Recovery | Solenoid quenched on power lead temp | Solenoid quenched on power lead temp. I ran down and managed to recover it
 with minimal temp rise in the solenoid. Coil
 |   | 
| 22 | Tue May  4 11:26:38 2004 | Doug Preddy | Normal Running | power leads | Power cycled HIC 142 with solenoid power at 400 amps. Flows > 240mm and
 temps have dropped.
 |   | 
| 21 | Tue May  4 11:20:57 2004 | Doug Preddy | Normal Running | power leads going unstable | power lead ripples increased over last couple of days. They are quickly
 dieing out now and the temps are startint
 |  | 
| 20 | Fri Apr 30 01:53:34 2004 | Doug Preddy | Quench Recovery | switched to parallel cold |  |  | 
| 19 | Thu Apr 29 23:43:15 2004 | Doug Preddy | Quench Recovery | Starting flow to solenoid | Solenoid quenched due to power leads at 9:00 tonight. Restarted flow in
 parallel warm mode.
 |  | 
| 18 | Wed Apr 28 09:30:09 2004 | Doug Preddy | Normal Running | power leads again | Same story as yesterday. Ripples dieing out, flows down to <200 mm, temps
 increasing. Turned HIC 142 down to 0.4 then
 |  | 
| 17 | Tue Apr 27 09:03:18 2004 | Doug Preddy | Normal Running | Power leads | Power leads are being monitored with the strip tool graphs. I saw the temps
 go from having a ripple to almost no ripple,
 |  | 
| 16 | Mon Apr 26 08:12:30 2004 | Doug Preddy | Quench Recovery | Power leads flow | Throttled back the LHe in until the power leads started to warm. My goal is
 to flow "warm" gas (hopefully 80K) through
 |  | 
| 15 | Mon Apr 26 07:48:46 2004 | Doug Preddy | Quench Recovery | Series mode | Switching over to series mode. |  | 
| 14 | Sun Apr 25 21:41:00 2004 | Willy Andersson | Quench Recovery | Trip on PL Temp. | Got pager call at 16:20 this afternoon. Solenoid had tripped on PL
 temperature. Arrived at about 17:00. Did
 |  | 
| 13 | Fri Apr 23 09:19:30 2004 | Doug Preddy | Normal Running | CV211 | PIC 229 was down to 10.8 bar this morning. PIC 229 was trying to open CV211
 to charge the system. I realized that in
 |  | 
| 12 | Thu Apr 22 10:39:16 2004 | Doug Preddy | Normal Running | Yesterday | Added two bottles of UHP helium to buffer tank.
 Had no liquifation rate, so optimized Ti121
 |  | 
| 11 | Wed Apr 21 09:57:59 2004 | Willy Andersson | Start Up | System status | Buffer tank pressure is down to about 6 psig. Too low to keep up
 with liquefaction requirements to build LHe
 |  | 
| 10 | Wed Apr 21 08:16:18 2004 | Doug Preddy | Start Up | switch to series mode | Switched to series mode. |  | 
| 9 | Wed Apr 21 06:48:31 2004 | Doug Preddy | Start Up | JT bypass closed | Closed cv155 |  | 
| 8 | Wed Apr 21 06:39:17 2004 | Doug Preddy | Start Up | over night | Changed to parallel cold mode at 9:30. At about 4:30 the city water chiller was
 turned on. The air compressor had
 |  | 
| 7 | Tue Apr 20 07:33:03 2004 | Doug Preddy | Start Up | PIC229 at 11bar | PIC229 at 11 bar. Changed strip tool to 15 sec increments.
 |  | 
| 6 | Tue Apr 20 07:06:41 2004 | Doug Preddy | Start Up | morning day 2 | Temps in solenoid are equalling out. Increasing PIC229 to 11 bar.
 |   | 
| 5 | Mon Apr 19 15:06:49 2004 | Doug Preddy | Start Up | Started turbines | Purifier run over the weekend. Both LN2 dewars empty, but purifier cold.
 Disabled CV211 and flowed all charge gas
 |   | 
| 4 | Fri Apr 16 10:15:26 2004 | Doug Preddy | Start Up | Changes to M9sol Vacuum PC | Yesterday and today there have been a number of changes made to the M9
 Vacuum control computer. The computer now
 |  | 
| 3 | Fri Dec 19 13:59:53 2003 | Doug Preddy | Maintenance | To-Do list for M9SOL | > I had distributed this list by email earlier this month. I have added
 > a few items at the bottom. I am posting
 |  | 
| 2 | Fri Dec 19 13:07:25 2003 | Willy Andersson | Maintenance | To-Do list for M9SOL | I had distributed this list by email earlier this month. I have added
 a few items at the bottom. I am posting it
 |   | 
| 1 | Fri Dec 19 11:34:10 2003 | Doug Preddy | Quench Recovery | Final home of this E-log | http://isac.triumf.ca/m9sol/ Chris has set up the M9 solenoid e-log at