Message ID: 47     Entry time: Thu Sep 2 21:58:34 2004
Author: Willy Andersson 
Operational Mode: Quench Recovery 
Subject: Switch to series cooling 
Solenoid had tripped on powerlead overtemp at about 09:00 this morning. 
Doug recovered the system to // cooling. I switched system to series 
cooling at about 21:30. LHe level at about 10%. Buffer tank pressure is low 
at about 13 psig. Letting LHe level build. Once LHe level rises to about 
35% then I suggest starting flow to the powerleads by opening HIC142 fully 
open. LHe level should continue to build but at a much slower rate. Once 
power lead temps have bottomed out to cr_pl_neg < 9K and cr_pl_pos < 11 K 
(check these values with archived data first) and are stable then can 
energize solenoid.