246
Wed Apr 16 16:51:16 2008
Doug Preddy
Start Up
switched to parallel cold return
245
Wed Apr 16 15:15:44 2008
Doug Preddy
Start Up
refrigerator started
Turned on compressor last Tuesday. Ran purifier
until today. Final value 0.15ppm.
Started
cool down yesterday. Switched into parallel
244
Wed Apr 16 15:12:01 2008
Doug Preddy
Normal Running
graphs to load
There are three strip tools saved:
coils
powerlead
startup
243
Tue Apr 1 15:17:06 2008
Doug Preddy
Start Up
Starting up vacuum systems
Monday afternoon there was no control of the
vacuum system. The controls were working,
but no output to devices. The "output" breaker
242
Mon Dec 3 14:59:58 2007
Doug Preddy
Quench Recovery
Not really a quench recovery, But...
Not really a quench recovery, but...I wanted
to send this out via e-mail.
The contactor for the compressor building
241
Wed Nov 28 15:07:08 2007
Kevin Trithardt
Other
power lead regeneration
Power leads cold again, power supply ramping
up.
240
Tue Nov 27 21:02:08 2007
Doug Preddy
Normal Running
power lead ripples
power leads were oscillating tonight. John
K. came down and cycled CV142.
(Thanks John!) Temps settling out.
239
Mon Nov 26 09:45:43 2007
Doug Preddy
Normal Running
power lead problems
Power lead temperatures were rippling yesterday
afternoon. I asked one of
the operators to cycle CV142.
238
Fri Nov 23 08:41:45 2007
Doug Preddy
Normal Running
screen grab of powerlead temps
237
Fri Nov 23 08:35:37 2007
Doug Preddy
Normal Running
power lead problems
About 10 hours ago the power lead temperatures
started to fluctuate and
climb.
236
Mon Nov 19 04:30:47 2007
Doug Preddy
Quench Recovery
switched to series mode
235
Mon Nov 19 02:58:44 2007
Doug Preddy
Quench Recovery
Solenoid quenched on power lead over temperature
I received a page at 1:30 this morning. The
power lead temperatures were
climbing. Cycling CV142 did not fix the problem.
234
Fri Nov 9 18:56:20 2007
Doug Preddy
Maintenance
System Ready to Run
M9BMP2 was replaced. Vacuum good. Solenoid
is ramped up. :)
233
Fri Nov 9 16:24:33 2007
Doug Preddy
Maintenance
2APLC Died
thumb wheel panel all lite up. 2APLC rebooted
by Klara (Thanks). M9BMP2 not
pumping properly. Edi is looking at it.
232
Fri Nov 9 14:50:03 2007
Doug Preddy
Quench Recovery
Solenoid cold
Solenoid cold. Added two bottles of helium.
Allowing to stablize. Will
ramp power supply in half an hour.
231
Fri Nov 9 07:37:44 2007
Doug Preddy
Quench Recovery
switched to series mode
switched to series mode.
230
Thu Nov 8 09:28:53 2007
Doug Preddy
Quench Recovery
started compressor
And started flow to coldbox,solenoid in parallel
warm return.
229
Wed Oct 24 12:04:06 2007
Doug Preddy
Maintenance
power lead and cryopump regeneration
Regenerating M9B:CP2 as temperature was up
to almost 20k.
Shut off flow to power leads to allow them
228
Tue Oct 9 13:16:44 2007
Doug Preddy
Quench Recovery
switched to series mode
227
Tue Oct 9 12:58:22 2007
Doug Preddy
Quench Recovery
Solenoid Quench 2
Refrigeration sytem shut off at ~ 12:30 today.
No apparent reason. Only
light on was a cold box fault.