M9sol Logbooks Solenoid Daily log
  Solenoid logbook, all entries  ELOG logo
ID Date Author Operational Mode Subjectdown Text Attachments
  110   Mon Oct 31 07:48:57 2005 Willy AnderssonQuench Recoverywhat a fine entryKeep IT safe boys   
  196   Wed May 16 10:02:35 2007 Doug PreddyMaintenancewarming up power leadsPower lead temps have been slowly increasing
over the last wekk. I ramped 
the solenoid down and am warming the power
  
  56   Thu Oct 7 13:33:20 2004 Doug PreddyStart Upturned off purifierturned off purifier   
  84   Wed Nov 24 14:58:42 2004 Doug PreddyQuench RecoverytodayCame in this morning to find the helium return
line to the buffer tank 
frosted up. MV51 has been stuck open the
  
  95   Wed Jun 1 23:48:02 2005 Doug PreddyQuench Recoverythe first quenchI recieved a call from ops at 22:00 tonight
that the solenoid had alarmed. 
I came in and found a cold box fault, with
  
  133   Thu Jul 6 15:00:15 2006 Doug PreddyNormal Runningtest    
  134   Thu Jul 6 15:04:17 2006 Doug PreddyNormal Runningtesttest  M9daily log070306.XLS 
  206   Wed Jul 4 11:13:12 2007 Doug PreddyQuench RecoverytestThis is a test to make sure the e-mail works.   
  100   Thu Jun 9 08:56:10 2005 Doug PreddyQuench Recoverysystem back onThe compressor was restarted and the solenoid
is cooling (sort of) in 
parallel cold return.
  
  281   Thu May 13 14:37:13 2010 Doug PreddyQuench Recoveryswitching to series mode.    
  247   Thu Apr 17 13:52:25 2008 Doug PreddyStart Upswitched to series mode at 9:30 this morningswitched to series mode at 9:30 this morning.
Added two bottles of helium gas. 
Started
flow to power leads.
  
  71   Thu Oct 21 15:41:29 2004 Doug PreddyStart Upswitched to series mode    
  149   Sat Jul 29 05:30:28 2006 Doug PreddyQuench Recoveryswitched to series modeTemps bottomed out at ~16K. Switching to series
mode.
  
  173   Thu Oct 26 08:18:10 2006 Doug PreddyQuench Recoveryswitched to series modeSwitched to series mode at about 7:45.   
  228   Tue Oct 9 13:16:44 2007 Doug PreddyQuench Recoveryswitched to series mode    
  231   Fri Nov 9 07:37:44 2007 Doug PreddyQuench Recoveryswitched to series modeswitched to series mode.   
  236   Mon Nov 19 04:30:47 2007 Doug PreddyQuench Recoveryswitched to series mode    
  81   Tue Nov 23 15:10:06 2004 Doug PreddyQuench Recoveryswitched to parallel cold return mode    
  70   Thu Oct 21 07:16:35 2004 Doug PreddyStart Upswitched to parallel cold return    
  104   Tue Jun 14 23:01:22 2005 Doug PreddyQuench Recoveryswitched to parallel cold return    
  118   Thu May 25 07:34:18 2006 Doug PreddyStart Upswitched to parallel cold returnswitched to parallel cold return   
  159   Tue Aug 15 15:39:27 2006 Willy AnderssonQuench Recoveryswitched to parallel cold returnReturn flow temps from solenoid about 40K,
TIC121 at about 40K so switching 
to //cold return.
  
  164   Thu Oct 5 10:26:26 2006 Doug PreddyStart Upswitched to parallel cold return    
  199   Thu May 31 07:11:05 2007 Doug PreddyQuench Recoveryswitched to parallel cold returnswitched to parallel cold return earlier this
morning. Vacuum is 5 x 10-5.
  
  222   Wed Sep 26 08:52:46 2007 Doug PreddyStart Upswitched to parallel cold return    
  246   Wed Apr 16 16:51:16 2008 Doug PreddyStart Upswitched to parallel cold return    
  277   Fri Apr 9 07:34:55 2010 Doug PreddyQuench Recoveryswitched to parallel cold returnSwitched to parallel cold return at about
7:30 this morning
 
  
  285   Thu Nov 11 15:46:40 2010 Doug PreddyOtherswitched to parallel cold returnAt 14:45.   
  20   Fri Apr 30 01:53:34 2004 Doug PreddyQuench Recoveryswitched to parallel cold    
  38   Thu Jul 29 09:01:11 2004 Doug PreddyQuench Recoveryswitched to parallel coldChanged to series mode.   
  10   Wed Apr 21 08:16:18 2004 Doug PreddyStart Upswitch to series modeSwitched to series mode.   
  34   Wed Jul 28 07:00:18 2004 Doug PreddyQuench Recoveryswitch to series modeSwitching to series mode. Jumper on vacuum
gauge removed.
  
  55   Wed Oct 6 09:25:54 2004 Doug PreddyStart Upswitch to series modeswitched to series mode   
  107   Tue Jul 19 16:13:32 2005 Willy AnderssonQuench Recoveryswitch to series modeCR_6KS_M down to 11 K.
Switched to series mode.
Started small flow to power leads to get
  
  144   Fri Jul 21 14:08:04 2006 Doug PreddyQuench Recoveryswitch to series modeCoils temps <10K.Switching to series mode.   
  177   Tue Oct 31 05:36:14 2006 Doug PreddyQuench Recoveryswitch to series mode    
  182   Wed Nov 15 13:56:00 2006 Doug PreddyQuench Recoveryswitch to series mode    
  256   Sun Aug 24 06:45:14 2008 Doug PreddyQuench Recoveryswitch to series mode    
  268   Wed May 27 13:37:22 2009 Doug PreddyQuench Recoveryswitch to series modeSwitched to series mode at ~11:00. Built
liquid to 55% by 13:30. Started flow to the
power leads.
  
  161   Thu Aug 17 19:45:27 2006 Willy AnderssonQuench Recoveryswitch to series coolingTemps of flow returning from solenoid flat
lined at about 10K.
Gradually closed CV155.
  
  271   Fri Jul 10 01:10:59 2009 Doug PreddyQuench Recoveryswitch to parallel warm return    
  97   Thu Jun 2 01:38:25 2005 Doug PreddyQuench Recoverystrip tool plotThis is a strip tool  plot of the cooldown  02-06-05-12.57.26 AM.jpg 
  85   Thu Nov 25 07:21:37 2004 Doug PreddyQuench Recoverystarted power lead flowCurtis switched to series mode last night
at 7:00.
Startedflow to power leads at 7:00 this morning.
  
  33   Wed Jul 28 00:04:30 2004 Doug PreddyQuench Recoverystarted flow to solenoidSolenoid temps were: C1 & C12 about 30K. C6
&C7 about 11K. Started flow to 
solenoid in parallel cold return. Jumpered
  
  79   Tue Nov 23 09:30:15 2004 Doug PreddyQuench Recoverystarted flow to solenoidStarted flow to solenoid in parallel warm
return.
  
  96   Thu Jun 2 01:26:30 2005 Doug PreddyQuench Recoverystarted flow to solenoidSee attached file for a screen grab of temps.  Capture6-2-05-1.13.33 AM.jpg 
  113   Thu Nov 3 09:27:16 2005 Doug PreddyQuench Recoverystarted flow to solenoidstarted flow to solenoid in parallel cold
return.
  
  24   Wed May 5 09:59:10 2004 Doug PreddyQuench Recoverystarted flow to power leadsStarted flow to power leads.
TIC121 into auto mode
  
  72   Fri Oct 22 09:15:06 2004 Doug PreddyStart Upstarted flow to power leadsStarted flow to power leads HIC142 at 0.8.
  
  276   Thu Apr 1 11:57:55 2010 Doug PreddyStart Upstarted flow to power leadsSolenoid cold. Buffer tank pressure ~3psi,
so Kevin will add a bottle of helium. Started
flow to power leads at 11:45
  
  283   Mon Jun 7 09:24:47 2010 Doug PreddyQuench Recoverystarted flow to power leads    
  261   Fri Sep 26 09:50:18 2008 Doug PreddyQuench Recoverystarted flow in parallel cold return mode    
  217   Fri Sep 14 12:44:52 2007 Doug PreddyStart Upstarted compressorrunning purifier over weekend   
  230   Thu Nov 8 09:28:53 2007 Doug PreddyQuench Recoverystarted compressorAnd started flow to coldbox,solenoid in parallel
warm return.
  
  140   Thu Jul 20 13:21:57 2006 Willy AnderssonQuench Recoverystarted // warm return flow.TIC121 was down to 27K
warm return valves were wide open
opened supply valves CV143 and CV146.
  
  94   Wed Jun 1 08:23:19 2005 Doug PreddyStart Upstart up notesThe solenoid was pumped out Monday and yesterday
after the quads just 
downstream of the solenoid were put back
  
  39   Thu Jul 29 13:25:29 2004 Doug PreddyQuench Recoverystart flow to power leads    
  270   Fri Jul 10 00:34:03 2009 Doug PreddyQuench Recoverysolewnoid quench due to power bumpSolenoid quench due to power bump caused
by an airplane crash in Richmond at ~ 22:00.
Compressor off. Water pump off. Restarted
  
  280   Thu May 13 07:39:55 2010 Doug PreddyQuench Recoverysolenoid quench last nightSolenoid quenched on power lead temps last
evening. I am recovering it in parallel cold
return mode. C12~60K C1~50K.
  
  157   Thu Aug 10 20:26:09 2006 Willy AnderssonQuench Recoverysolenoid energizedPowerleads seem stable.   
  37   Wed Jul 28 23:07:01 2004 Doug PreddyQuench Recoverysolenoid coolingcooling rates for the solenoid are good. In
parallel cold return.
  
  265   Wed Dec 10 15:41:04 2008 Doug PreddyShut Downshuttting down solenoidshuttting down solenoid for winter shutdown.   
  116   Thu Dec 22 15:10:44 2005 Doug PreddyShut Downshut down systemshuttting down the system. isolated cold box
by closing v135. preparing to 
pump out cold box.
  
  91   Fri Mar 11 09:18:30 2005 Doug PreddyMaintenanceservices on systemThe cold box was isolated and pumped out after
it warmed up. It has been 
refilled with helium.
  
  238   Fri Nov 23 08:41:45 2007 Doug PreddyNormal Runningscreen grab of powerlead temps   Capture.jpg 
  264   Wed Nov 5 16:33:52 2008 Doug PreddyNormal Runningsbplc light outOn main menu the sbplc light was black. All
vacuum systems looked good, but there was
no control of any device. Jane reset the
  
  65   Sat Oct 16 17:13:19 2004 Other ss   
  69   Wed Oct 20 09:38:17 2004 Doug PreddyStart Uprestarting systemSolenoid tripped  off Monday morning due to
high compressor pressure. 
Regenerated purifier on Tuesday. Started
  
  158   Tue Aug 15 08:45:56 2006 Doug PreddyQuench Recoveryrestarted the compressorSolenoid tripped off Sunday. Restarted compressor
and cool down of cold box.
  
  103   Tue Jun 14 20:53:19 2005 Doug PreddyQuench Recoveryrestarted refrigerator after power bumpsee bitmap for temps/pressures after presurrizing
coldbox. Preparing to 
start flow to turbines, then will start flow
 june142045.BMP 
  112   Thu Nov 3 06:19:50 2005 Doug PreddyQuench Recoveryrestart systemLast night at about 7:00 there was a large
power bump that shut the 
compressor and water pump off. I came in
  
  188   Fri Apr 20 12:15:12 2007 Doug PreddyStart Upremote PLC switchI could not control the vacuum devices using
EPICS. The buttons could be 
pushed, but nothing would respond. The PLC
  
  216   Wed Sep 12 12:01:02 2007 Doug PreddyMaintenanceregenerated adsorberregenerated adsorber
replaced MV902 "ON" switch (used nitrogen
precooling on switch)
  
  245   Wed Apr 16 15:15:44 2008 Doug PreddyStart Uprefrigerator started Turned on compressor last Tuesday. Ran purifier
until today. Final value 0.15ppm.
Started
cool down yesterday. Switched into parallel
  
  202   Thu Jun 14 00:10:10 2007 Doug PreddyQuench Recoveryrecovering from earlier quenchSolenoid quenched on power lead temperature
last night at about 6:00. I 
restarted flow to the solenoid in parallel
  
  184   Tue Nov 28 10:32:04 2006 Doug PreddyQuench Recoveryrecovering after looong power outageThe power has been out for about 24 hours.
Recovering coldbox, solenoid and 
beamline vacuum.PLC is down for the vacuum
  
  257   Thu Sep 4 10:02:55 2008 Doug PreddyNormal Runningramping down solenoidSolenoid power supply will be turned off for
Fall shutdown.
  
  209   Sat Jul 28 15:45:17 2007 Doug PreddyQuench Recoveryquench recoveryThe solenoid tripped off this morning due
to pwer lead over temp. The 
compressor was also off. I restarted this
  
  210   Sat Jul 28 22:30:13 2007 Doug PreddyQuench Recoveryquench againrecovered system and energized solenoid at
about 6:00 this evening.
At 9:30 power lead temps quenched solenoid
  
  194   Mon Apr 30 09:30:26 2007 Doug PreddyNormal Runningproblems over weekendThe PLC Limit switch power supply was tripped
off on Saturday. I came in 
and reset it and restarted all the vacuum
  
  187   Thu Apr 19 15:18:25 2007 Doug PreddyStart Uppreparing to start the refrigeratorStarting the vacuum systems in the coldbox
and solenoid.
  
  183   Wed Nov 15 17:05:49 2006 Willy AnderssonQuench Recoverypowerleads cold, energizing solenoidPowerlead temperatures have bottomed out,
energizing solenoid to 300 
amps, ramp rate 60.
  
  48   Fri Sep 3 01:28:43 2004 Doug PreddyQuench Recoverypower upstarted flow to the power leads at 11:30.
Cold and stable at 1:00am. 
Energized solenoid at 1:30.
  
  114   Tue Nov 15 07:25:10 2005 Doug PreddyNormal Runningpower supply problemSaturday November 12 the solenoid power supply
ramped itself down to zero. 
It would not ramp up unless the ramp rate
  
  74   Tue Oct 26 09:45:24 2004 Doug PreddyNormal Runningpower leads warmingFlow to power leads dropped to <80 mm. Cycled
HIC142. Temps dropping, flows 
>250.
  
  123   Tue May 30 14:36:41 2006 Doug PreddyNormal Runningpower leads warmingpower leads warming. closed HIC 142, then
re-opened. temps falling
  
  248   Mon Apr 28 09:03:22 2008 Doug PreddyNormal Runningpower leads warmed during weekendThe power lead temperatures suddenly rose
on Sunday morning. No rippling occured before
the woarm up. 
I shut flow off this morning
  
  57   Sat Oct 9 22:50:26 2004 Other power leads warmedAt 10:30 PM Saturday, Donald Arseneau sees
power leads at
25K, having gone up for the previous 10 hours.
  
  21   Tue May 4 11:20:57 2004 Doug PreddyNormal Runningpower leads going unstablepower lead ripples increased over last couple
of days. They are quickly 
dieing out now and the temps are startint
  
  150   Sat Jul 29 09:38:06 2006 Doug PreddyQuench Recoverypower leads cooledopened power lead flow about 40 minutes ago.
~12K will let them stablize. 
LHE level >20%
  
  25   Wed May 5 12:13:51 2004 Doug PreddyNormal Runningpower leads coldshould be ready to run.   
  18   Wed Apr 28 09:30:09 2004 Doug PreddyNormal Runningpower leads againSame story as yesterday. Ripples dieing out,
flows down to <200 mm, temps 
increasing. Turned HIC 142 down to 0.4 then
  
  22   Tue May 4 11:26:38 2004 Doug PreddyNormal Runningpower leads Power cycled HIC 142 with solenoid power at
400 amps. Flows > 240mm and 
temps have dropped.
 may4_1053.BMP 
  58   Sun Oct 10 17:36:35 2004 ExperimentorStart Uppower leadsAt 5:30 pm sundat, power lead temp is about
10.6K.  script tool still 
notworking
  
  254   Sat Jul 26 01:29:21 2008 Doug PreddyQuench Recoverypower leadsops will monitor liquid level. When it's at
60% they will come down and open the power
lead valve. When the temperature ripples
  
  269   Wed May 27 15:01:34 2009 Doug PreddyQuench Recoverypower leadsPower leads are cold. There is still a
small ripple. The power supply can be ramped
up as soon as the ripples die out.
  
  132   Mon Jul 3 23:01:34 2006 Doug PreddyQuench Recoverypower lead warming/ripples caused quenchpower lead warming/ripples caused quench at
18:45 today. Compressor stayed 
on, but flow to solenoid stopped.
  
  240   Tue Nov 27 21:02:08 2007 Doug PreddyNormal Runningpower lead ripplespower leads were oscillating tonight. John
K. came down and cycled CV142. 
(Thanks John!) Temps settling out.
  
  45   Wed Aug 25 12:52:53 2004 Doug PreddyNormal Runningpower lead rippleLhe ripple has started. 10.9-11.4K PL-Pos
also has a smaller ripple. Both 
flows down to 230mm. Cycled HIC 142 and the
  
  89   Thu Dec 2 10:51:16 2004 Doug PreddyNormal Runningpower lead rippleRipples got worse overnight.LHe in 10.3-11.7,
PL Neg 8.0-8.7. Flows <200mm.
cycled CV142. Flows 240mm. LHe in 10.3-11.3
  
  241   Wed Nov 28 15:07:08 2007 Kevin TrithardtOtherpower lead regenerationPower leads cold again, power supply ramping
up.
  
  176   Tue Oct 31 03:52:29 2006 Doug PreddyQuench Recoverypower lead quench at 2:30 this morningCompressor still on. Starting flow to solenoid   
  175   Thu Oct 26 14:14:57 2006 Doug PreddyQuench Recoverypower lead quenchsolenoid tripped on power lead temp. I was
here and caught it. System 
cooling.
  
  237   Fri Nov 23 08:35:37 2007 Doug PreddyNormal Runningpower lead problemsAbout 10 hours ago the power lead temperatures
started to fluctuate and 
climb.
  
  239   Mon Nov 26 09:45:43 2007 Doug PreddyNormal Runningpower lead problemsPower lead temperatures were rippling yesterday
afternoon. I asked one of 
the operators to cycle CV142.
  
  75   Fri Oct 29 14:25:23 2004 Doug PreddyNormal Runningpower lead flows droppingFlow is down to <190 mm. temps are starting
to climb. Cycled HIC142
  
  76   Mon Nov 1 07:17:44 2004 Doug PreddyNormal Runningpower lead flow gone badpower temps as follows:
LHE_IN 23
PL_POS 36
  
  229   Wed Oct 24 12:04:06 2007 Doug PreddyMaintenancepower lead and cryopump regenerationRegenerating M9B:CP2 as temperature was up
to almost 20k.
Shut off flow to power leads to allow them
  
  251   Thu Jul 24 16:28:50 2008 Doug PreddyQuench Recoverypower bumpThere was a power bump at ~15:00 today. All
mseon hall cooling water went off. Compressor
tripped. Main magnet went off. 
Restarted
  
  54   Wed Oct 6 07:57:01 2004 Doug PreddyStart Upparallel cold returnSwitched to parallel cold return at 4:00 yesterday
afternoon.
6:30 this morning started closing cv155 as
  
  8   Wed Apr 21 06:39:17 2004 Doug PreddyStart Upover nightChanged to parallel cold mode at 9:30.
At about 4:30 the city water chiller was
turned on. The air compressor had 
  
  185   Fri Jan 26 15:10:37 2007 Doug PreddyMaintenancenew control computer installedI have installed a new control computer at
the M9 solenoid control panel
  
  6   Tue Apr 20 07:06:41 2004 Doug PreddyStart Upmorning day 2Temps in solenoid are equalling out. Increasing
PIC229 to 11 bar. 
 april200705.BMP 
  90   Sun Dec 5 11:15:24 2004 Doug PreddyNormal Runningmore ripplesI loked at the power lead temps this morning.
LHe_in 10.4-12.0  PL-Neg 8.0-
9.0  PL-Pos 9.8-10.4. Flows down to 160mm.
  
  153   Wed Aug 9 06:59:43 2006 Willy AnderssonQuench Recoverymore // cold return stuffAbout 21:00 last night the solenoid temps
had not changed much, pretty well 
flat lined. I guess my reducing CV155 stalled
  
  147   Sat Jul 22 06:34:14 2006 ExperimentorQuench Recoverylooking goodDonald Arseneau reporting.

It is looking pretty good, so I came back
  
  66   Sat Oct 16 17:18:50 2004 Other lame    
  244   Wed Apr 16 15:12:01 2008 Doug PreddyNormal Runninggraphs to loadThere are three strip tools saved:
coils
powerlead
startup
  
  226   Fri Oct 5 13:25:41 2007 Doug PreddyQuench Recoveryfunny graph There was a fluctuation of the positive lead
temp with no flow going to 
the power leads.
 Capture.jpg 
  278   Fri Apr 9 16:07:28 2010 Doug PreddyStart Upflow to power leads started    
  64   Wed Oct 13 08:17:00 2004 ExperimentorStart Upcycling 142 when temp hit 10.68I tried cycling the valve twice -- at about
11:30 and again around 1:00,
trying both the buttons and the dial -- but
  
  67   Sun Oct 17 04:26:29 2004 OtherStart Upcycling 142 when temp hit 10.68> I tried cycling the valve twice -- at about
11:30 and again around 1:00,
> trying both the buttons and the dial --
  
  212   Tue Jul 31 07:53:57 2007 Kevin TrithardtNormal Runningcycled power leadsgot a page for the power lead temp
came down and cycled the power leads.. did
not seem to make a difference..
  
  59   Mon Oct 11 03:56:14 2004 ExperimentorStart Upcv142 valvenoticed that the lead temp rose to 24K over
the past 4 hours. had control 
room personnle cycle the cv142 valve. temperature
  
  60   Mon Oct 11 11:49:34 2004 ExperimentorStart Upcv142 valvenoticed temp rise, which occured only about
5 hours after the 
last "reset" .... the time between resets
  
  61   Mon Oct 11 22:32:11 2004 ExperimentorStart Upcv142 valvetemp of leads begins its exponential rise
again (caught it at 12K), approx 
11 hours after last reset.  this time asked
  
  62   Tue Oct 12 02:05:31 2004 ExperimentorStart Upcv142 valveafter only two hours since last lead cycling,
they warm up quickly. 
ops cycles the valve, but more quickly this
  
  63   Tue Oct 12 13:00:08 2004 Curtis BallardStart Upcv142 valveCurtis closes and opens slowly at about 11:00
am this morning. Wants to see 
if that works better than the sudden on/off
  
  170   Sun Oct 15 07:09:41 2006 OtherNormal Runningcryopump bcp2 tripDonald Arseneau reporting

Not really solenoid realated, but FYI...
  
  98   Thu Jun 2 09:16:40 2005 Doug PreddyQuench Recoverycontinuing to coolSee screen grab  02-06-05-7.32.49 AM.jpg 
  50   Mon Sep 13 13:56:06 2004 Doug PreddyShut Downcompressor offcompressor off   
  141   Thu Jul 20 13:27:21 2006 Willy AnderssonQuench Recoverycold return mode continues...cr_c1 and cr_12 are both at about 45K and
heading slowly down, so the 
gamble of going to // cold return worked.
  
  142   Thu Jul 20 13:34:26 2006 Willy AnderssonQuench Recoverycold return mode continues...Even though cr_c1 and cr_c12 have begun to
tick down, cr_c7 and cr_c6 are 
actually ticking up. This is probably a wave
  
  218   Tue Sep 18 18:48:11 2007 Curtis BallardMaintenancecold box leak checkThe leak check of the cold box diff pump fittings.
 Found a leak at the 
base of the diff pump on the tee that is
  
  252   Fri Jul 25 21:48:15 2008 Doug PreddyQuench Recoveryanother quenchSystem was stable and solenoid ramped up at
4:00 today. At about 7:30 it tripped off
again. I came in and restarted the compressor
  
  42   Thu Aug 19 10:28:32 2004 Doug PreddyStart Upafter M9bv5 FIXOver the last couple of days we replaced the
gate on the M9BV5 valve. The 
old one had worn the teflon sliders on the
  
  53   Tue Oct 5 07:18:16 2004 Doug PreddyStart Upadjust compressor discharge pressureadjusted compressor discharge pressure to
11 Bar
  
  88   Wed Dec 1 12:55:56 2004 Doug PreddyNormal Runningadded one bottle helium gasbuffer tank pressure is up to 6.5 psi.   
  201   Tue Jun 12 12:48:25 2007 Doug PreddyMaintenanceadded one bottle UHP helium to system    
  12   Thu Apr 22 10:39:16 2004 Doug PreddyNormal RunningYesterdayAdded two bottles of UHP helium to buffer
tank.
Had no liquifation rate, so optimized Ti121
  
  259   Fri Sep 26 09:44:25 2008 Doug PreddyMaintenanceWater flow interlock restoredyesterday   
  225   Fri Oct 5 10:25:43 2007 Doug PreddyQuench RecoveryUnknown quench 2Kevin and I were at the Sulzer panel when
the system quenched 2. No 
obvious reasons. Restarted the system.
  
  101   Thu Jun 9 21:41:25 2005 Willy AnderssonQuench RecoveryTrying to achieve stable cooling.From about 08:00 to 15:00

See Doug's previous entry for quench recovery
  
  14   Sun Apr 25 21:41:00 2004 Willy AnderssonQuench RecoveryTrip on PL Temp.Got pager call at 16:20 this afternoon. Solenoid
had tripped on PL 
temperature. Arrived at about 17:00. Did
  
  99   Thu Jun 9 06:53:40 2005 ExperimentorOtherTrip at 6:45 Today. Solenoid quenched at 6:45.  All temperature
monitors were flat until trip.  
Hearing compressor stop was first warning.
  
  204   Tue Jun 19 21:15:41 2007 Doug PreddyQuench RecoveryToday's quenchpower lead temps caused a quench at 8:30 this
morning. Recovery of system 
and flow to solenoid was started by 11:00.
  
  2   Fri Dec 19 13:07:25 2003 Willy AnderssonMaintenanceTo-Do list for M9SOLI had distributed this list by email earlier
this month. I have added
a few items at the bottom. I am posting it
 M9Blist3.txt 
  3   Fri Dec 19 13:59:53 2003 Doug PreddyMaintenanceTo-Do list for M9SOL> I had distributed this list by email earlier
this month. I have added
> a few items at the bottom. I am posting
  
  43   Tue Aug 24 12:09:50 2004 Doug PreddyNormal RunningThe past few daysThe solenoid has been re-started as of August
19. It was cooled down and 
energized over the weekend. Sunday night
  
  192   Fri Apr 27 10:16:44 2007 Kevin TrithardtQuench RecoveryTestDon't Panic -- There was no quench. I am testing
the e-log.
I have added Kevin's name and e-mail to the
  
  126   Tue Jun 13 16:04:33 2006 Willy AnderssonQuench RecoveryTIC 121 set to auto mode    
  137   Thu Jul 20 02:59:06 2006 Doug PreddyQuench RecoveryT1 still not startingI will leave the cold box to warm overnight
and try it again in the morning.
  
  138   Thu Jul 20 08:18:31 2006 Doug PreddyQuench RecoveryT1 startedStarted flow to turbines. Both started. T1
seems a bit slow. HIC110 at 0.85 
SI110 = 4100.
  
  136   Thu Jul 20 00:52:13 2006 Doug PreddyQuench RecoveryT1 frozenT1 will not spin up. Warming up cold box.   
  11   Wed Apr 21 09:57:59 2004 Willy AnderssonStart UpSystem statusBuffer tank pressure is down to about 6 psig.
Too low to keep up
with liquefaction requirements to build LHe
  
  109   Sun Aug 14 10:03:13 2005 Willy AnderssonNormal RunningSystem stableNo experiment running in M9B.

Solenoid current is zero and power supply
  
  234   Fri Nov 9 18:56:20 2007 Doug PreddyMaintenanceSystem Ready to RunM9BMP2 was replaced. Vacuum good. Solenoid
is ramped up. :)
  
  154   Wed Aug 9 07:50:49 2006 Doug PreddyQuench RecoverySwitched to series mode.Cool down continues.   
  166   Thu Oct 5 14:53:54 2006 Doug PreddyStart UpSwitched to series mode    
  200   Thu May 31 11:51:49 2007 Doug PreddyQuench RecoverySwitched to series mode    
  127   Wed Jun 14 10:35:30 2006 Willy AnderssonQuench RecoverySwitched to series cooling.Had system in //return cooling overnight with
small flow to keep power 
leads cool. Switched to series mode about
  
  273   Mon Oct 5 11:54:58 2009 Doug PreddyNormal RunningSwitched TIC121 into auto mode    
  30   Fri Jun 25 14:20:58 2004 Willy AnderssonQuench RecoverySwitch to series modeAt about 14:00 solenoid temps dipping below
8 K. Trace of LHe level (dense 
gas) starting to show. Switching to series
  
  47   Thu Sep 2 21:58:34 2004 Willy AnderssonQuench RecoverySwitch to series coolingSolenoid had tripped on powerlead overtemp
at about 09:00 this morning. 
Doug recovered the system to // cooling.
  
  215   Tue Aug 14 12:55:12 2007 Doug PreddyNormal RunningStrip tool savesThere are two strip tools saved:
coils
powerlead
  
  172   Thu Oct 26 04:34:15 2006 Willy AnderssonQuench RecoveryStill in // cold return mode.Leaving system cooling in // cold return mode,
cooldown rates still good. 
Had flow on to powerleads since re-starting
  
  243   Tue Apr 1 15:17:06 2008 Doug PreddyStart UpStarting up vacuum systemsMonday afternoon there was no control of the
vacuum system. The controls were working,
but no output to devices. The "output" breaker
  
  19   Thu Apr 29 23:43:15 2004 Doug PreddyQuench RecoveryStarting flow to solenoidSolenoid quenched due to power leads at 9:00
tonight. Restarted flow in 
parallel warm mode. 
  
  36   Wed Jul 28 22:01:31 2004 Doug PreddyQuench RecoveryStarting flow to solenoidc1 28, c12 29, c6 11.6, c7 12.5 TI121 22.5   
  162   Wed Oct 4 07:29:04 2006 Doug PreddyStart UpStarting cool downStarted compressor yesterday with flow through
the purifier overnight.
Starting tubines this morning in parallel
  
  5   Mon Apr 19 15:06:49 2004 Doug PreddyStart UpStarted turbinesPurifier run over the weekend. Both LN2 dewars
empty, but purifier cold. 
Disabled CV211 and flowed all charge gas
 april191500.BMP 
  52   Mon Oct 4 09:16:29 2004 Doug PreddyStart UpStarted turbinesStarted flow to the turbimes with the solenoid
in parallel warm return.
  
  117   Tue May 23 08:44:28 2006 Doug PreddyStart UpStarted turbinesSolenoid and cold box were pumped out last
Tuesday. Cold box roughing pump 
initially would not start. had to retighten
  
  190   Tue Apr 24 10:21:49 2007 Doug PreddyStart UpStarted flow to turbinesStarted flow to turbines at about 10:00 in
parallel warm mode.
  
  121   Fri May 26 10:03:03 2006 Doug PreddyStart UpStarted flow to power leadsStarted flow to power leads.
Liquid level 15%
  
  220   Mon Sep 24 15:08:41 2007 Doug PreddyStart UpStarted cool down this morningThe vacuum in the cold box was good on Friday
0.0 X 10-4. Let it pump over 
the weekend. Added two bottles of UHP helium
  
  275   Mon Mar 29 09:25:52 2010 Kevin TrithardtStart UpStarted cool down this morning    
  189   Mon Apr 23 16:31:02 2007 Doug PreddyStart UpStarted compressorStartewd compressor and purifer today.   
  274   Wed Mar 24 08:30:33 2010 Doug PreddyStart UpStart upCompressor started yesterday afternoon.
Pumping out solenoid overnight. Cold box
vacuum good. circulating helium to solenoid
  
  160   Thu Aug 17 08:04:44 2006 Doug PreddyQuench RecoverySolenoid tripped at about 3:00 this morningrestarting system.
Compressor would not start. Idid not see
anything that would not let it 
  
  131   Fri Jun 30 11:17:57 2006 Doug PreddyNormal RunningSolenoid recoveredTuesday night. June 27, 2006, the solenoid
quenched for no apparent 
reason. It was restarted Wednesday morning.
  
  282   Tue May 18 15:04:36 2010 Kevin TrithardtMaintenanceSolenoid ramping up    
  272   Tue Jul 14 15:28:36 2009 Doug PreddyQuench RecoverySolenoid ramped upThe solenoid is cold and I have started
to ramp up the power supply. Some slight
ripples in the power leads, but they are
  
  286   Fri Nov 12 16:54:50 2010 Doug PreddyNormal RunningSolenoid ramped up    
  214   Tue Aug 14 10:15:22 2007 Doug PreddyMaintenanceSolenoid ramped downSolenoid ramped down and locked out for M9A
slits work.
  
  197   Wed May 30 08:38:38 2007 Doug PreddyQuench RecoverySolenoid quenched this morningSolenoid quenched this morning on power lead
temp (I believe).
I restarted it at ~7:00, but the capacity
  
  23   Wed May 5 07:55:37 2004 Doug PreddyQuench RecoverySolenoid quenched on power lead tempSolenoid quenched on power lead temp. I ran
down and managed to recover it 
with minimal temp rise in the solenoid. Coil
 may5_745.BMP 
  235   Mon Nov 19 02:58:44 2007 Doug PreddyQuench RecoverySolenoid quenched on power lead over temperatureI received a page at 1:30 this morning. The
power lead temperatures were 
climbing. Cycling CV142 did not fix the problem.
  
  260   Fri Sep 26 09:45:18 2008 Doug PreddyQuench RecoverySolenoid quenched at 7:50Compressor on. Turbines started. Vacuum system
recovering.
  
  267   Wed May 27 08:06:00 2009 Doug PreddyQuench RecoverySolenoid quenched at 5:00The solenoid quenched at 5:00 this morning
due to power lead temps. The temps started
oscillating about 10 minutes before the trip.
  
  135   Thu Jul 20 00:25:16 2006 Doug PreddyQuench RecoverySolenoid quenched @ 11:15Solenoid Quenched on C1 temp. It went above
5K. Shut off compressor. 
Restarting system @ 12:15
  
  255   Sat Aug 23 17:02:38 2008 Doug PreddyQuench RecoverySolenoid quenchedThe solenoid quenched this afternoon at ~12:30.
The compressor shut down. Fault lights were
cold box trip.
Restarted the compressor
  
  279   Mon May 3 07:49:54 2010 Doug PreddyQuench RecoverySolenoid quench SaturdaySolenoid quenched at ~13:30 Saturday. Kevin
came in later in the day to recover it. System
was in all around mode, no flow to solenoid.
  
  78   Tue Nov 23 07:47:00 2004 Doug PreddyQuench RecoverySolenoid quench Nov 21Solenoid quenched Nov 21 due to alarge power
bump. Everything tripped off 
including vacuum and water pumps. Cold box
 11_22_04_7_40.BMP 
  32   Tue Jul 27 23:33:27 2004 Doug PreddyQuench RecoverySolenoid quench 9:30I was phoned at about 9:30 tonight with a
solenoid alarm. After looking at 
the display I found the system to be tripped
  
  203   Sat Jun 16 15:05:03 2007 Doug PreddyNormal RunningSolenoid power lead temp problemPower lead temps rippling and increasing.
I asked an operator to cycle 
HIC142. This seems to have fixed the problem
  
  174   Thu Oct 26 11:45:22 2006 Doug PreddyQuench RecoverySolenoid coldpower leads cold 50% liquid.
Turning on power supply
  
  223   Thu Sep 27 07:41:15 2007 Doug PreddyStart UpSolenoid coldSolenoid cold and in series mode. 50% liquid
level. Starting flow to 
powerleads.
  
  232   Fri Nov 9 14:50:03 2007 Doug PreddyQuench RecoverySolenoid coldSolenoid cold. Added two bottles of helium.
Allowing to stablize. Will 
ramp power supply in half an hour.
  
  105   Mon Jul 18 11:18:40 2005 Doug PreddyQuench RecoverySolenoid Quenched over the weekendThe solenoid was found to be warm on Monday
morning. M9Q3 had a bad water 
leak over the weekend. The water header serving
  
  263   Sun Oct 5 18:54:27 2008 Doug PreddyQuench RecoverySolenoid Quench ~16:30Restarted compressor and flow to coldbox by
18:30
  
  115   Mon Dec 12 09:24:29 2005 Doug PreddyQuench RecoverySolenoid Quench on power lead tempThe solenoid tripped on power lead temps at
about 9:15. I am recovering it 
from it's all around mode into parallel cold
  
  151   Tue Aug 8 07:27:49 2006 Doug PreddyQuench RecoverySolenoid Quench at about 6:00 this morningNo indication of problem. Turbines still running,
but no flow to solenoid. 
Started flow. Solenoid ~20K.
  
  227   Tue Oct 9 12:58:22 2007 Doug PreddyQuench RecoverySolenoid Quench 2Refrigeration sytem shut off at ~ 12:30 today.
No apparent reason. Only 
light on was a cold box fault. 
  
  191   Fri Apr 27 10:10:07 2007 Doug PreddyNormal RunningSolenoid ColdSolenoid cold, power leads cold. Alarm handler
started.
  
  15   Mon Apr 26 07:48:46 2004 Doug PreddyQuench RecoverySeries modeSwitching over to series mode.   
  250   Tue Jul 15 08:47:13 2008 Doug PreddyNormal RunningScreen grab of normal runningSolenoid was recovered on Sunday July 13.
The power leads started to ripple slightly
this morning, so I cycled HIC142. No real
 july_15_08_0841.jpgjuly_15_08_0846.jpg 
  93   Thu May 19 15:05:31 2005 Doug PreddyMaintenanceSTA 120 replacedThe turbine 2 speed indicator control box
STA 120 has been replaced with a 
newer style Jaquet controller. This is similar
 M9 Turbine 2.doc 
  155   Thu Aug 10 08:33:44 2006 Doug PreddyNormal RunningRunning low on heliumTemperture okay, but there are fluctuation
in the power leads. PIC229 is 
trying to charge. The buffer tank pressure
  
  148   Fri Jul 28 20:39:40 2006 Doug PreddyQuench RecoveryRestarting refrigeratorSolenoid quenched at about 6:00 tonight.
I restarted the compressor, then the cold
box and turbines. When cold box 
  
  193   Fri Apr 27 10:32:07 2007 Doug PreddyQuench RecoveryReset quench relay logic boxThe quench relay logic controller was not
functioning correctly. It showed 
a "Trip" light and no other lights. I cycled
  
  266   Wed Dec 10 15:44:58 2008 Doug PreddyMaintenanceRepairs to solenoid power supplyKlaus and company pulled the ramp rate card
out of the solenoid power supply. Operational
checks were done and the rate switch was
  
  31   Tue Jun 29 14:59:46 2004 Willy AnderssonNormal RunningRemoved jumper on vacuum controller.Tuesday 14:55

Removed jumper on back of PIC 901. This jumper
  
  92   Mon Apr 18 09:52:11 2005 Doug PreddyMaintenanceRelays replacedThe following relays have been replaced in
the Sulzer control panel:
K341, K379, K132, K324A, K314, K125, K127,
  
  253   Sat Jul 26 01:06:04 2008 Doug PreddyQuench RecoveryRefrigerator in series modeSwitched to series mode. Solenoid ~8K, starting
to build liquid. 
  
  122   Fri May 26 14:34:24 2006 Doug PreddyStart UpReady to runSolenoid is cold, liquid level is 35%, power
leads cold and seem stable.
Added one bottle helium to buffer tank. Pressure
  
  163   Wed Oct 4 18:10:18 2006 Willy AnderssonStart UpQuick check on systemSystem in good // warm return mode, temps
coming down nicely. BCP2 cryopump 
does not look too healthy, temperature is
  
  49   Mon Sep 13 10:55:25 2004 Doug PreddyShut DownQuench and shutdownThe solenoid quenched on powerlead temps this
morning at 10:00. It is 
warming up and the power leads are >200K.Isolating
  
  108   Tue Jul 19 18:09:56 2005 Willy AnderssonQuench RecoveryQuench IIMachine trip just as I was exiting the TRIUMF
gate. QRL said cold box fault 
as did red lite on compressor. At cold box
  
  198   Wed May 30 11:27:32 2007 Doug PreddyQuench RecoveryProblems recoveringVarious start up problems :
MV51 stuck open. Finally got it closed by
hitting the valve with a hammer.
  
  26   Wed May 5 23:57:38 2004 Willy AnderssonQuench RecoveryPowerleads warming up again.Got a page from Don Flemming at about 18:50
wondering why the power
lead temperatures were rising. See attachment
 may5_04_1800.BMP 
  86   Sun Nov 28 10:15:18 2004 Willy AnderssonQuench RecoveryPowerlead quench ?Got page very early this morning. Couldn't
come in til 08:30. System was 
in idle running mode and someone had reset
  
  171   Thu Oct 26 04:27:31 2006 Willy AnderssonQuench RecoveryPowerlead overtemp02:30 this morning got page from operations.
Looked at system online, appears that there
was a powerlead overtemp
  
  77   Thu Nov 4 13:33:49 2004 Doug PreddyNormal RunningPower supply problemsLast night the power supply was ramping up
at a very slow rate. This 
morning it was energized,but with no current
  
  16   Mon Apr 26 08:12:30 2004 Doug PreddyQuench RecoveryPower leads flowThrottled back the LHe in until the power
leads started to warm. My goal is 
to flow "warm" gas (hopefully 80K) through
  
  17   Tue Apr 27 09:03:18 2004 Doug PreddyNormal RunningPower leadsPower leads are being monitored with the strip
tool graphs. I saw the temps 
go from having a ripple to almost no ripple,
  
  73   Mon Oct 25 15:44:15 2004 Doug PreddyNormal RunningPower lead temps starting to ripplePower lead temps starting to ripple at a 1.5min
period. PL flows are 
both>250 mm.
  
  46   Sun Aug 29 14:39:19 2004 Doug PreddyNormal RunningPower lead rippleThe LHe in and PL-neg temps have been building
up a ripple over the last 
few days. Curtis cycled HIC142 Friday evening
  
  207   Thu Jul 5 03:04:50 2007 Doug PreddyNormal RunningPower lead problemsThe power leads temps are starting to cycle
again. Earlier today M9 
experimentor cycled HIC142 to stablize the
  
  44   Wed Aug 25 07:47:05 2004 Doug PreddyNormal RunningPower lead flowThe power lead flows had dropped to 210 mm
this morning. I increased HIC142 
from 0.8 to 0.83. The flows are back to <250
  
  102   Tue Jun 14 18:48:14 2005 ExperimentorNormal RunningPower bump at 18:40Solenoid tripped on a power bump at 18:40
Tuesday
-- Donald Arseneau
  
  186   Mon Mar 5 14:59:41 2007 Doug PreddyMaintenancePasswordOn the weekend someone used this log to send
spam messages. I have 
installed a password for the write privilege.
  
  224   Tue Oct 2 13:02:37 2007 Doug PreddyNormal RunningPLC problemsWhile downloading new code for 2A PLC the
M9 and M15 vacuum systems were 
shut off. This is due to the two systems
  
  221   Tue Sep 25 07:34:04 2007 Doug PreddyStart UpPIC229 to 11 baradjusted turbine speeds and HIC104 to maintain
their values.
  
  7   Tue Apr 20 07:33:03 2004 Doug PreddyStart UpPIC229 at 11barPIC229 at 11 bar. Changed strip tool to 15
sec increments.
  
  139   Thu Jul 20 11:01:09 2006 Willy AnderssonQuench RecoveryOpened up warm return valves CV151 and CV153Mu2 pressure was just about at trip point
due to trapped volume of helium 
warming. So opened up warm returns to bleed
  
  145   Fri Jul 21 17:18:19 2006 Willy AnderssonQuench RecoveryOpened HIC142 fully openPower leads have flatlined but still building
LHe level.
  
  242   Mon Dec 3 14:59:58 2007 Doug PreddyQuench RecoveryNot really a quench recovery, But...Not really a quench recovery, but...I wanted
to send this out via e-mail. 
The contactor for the compressor building
  
  219   Wed Sep 19 20:53:18 2007 Doug PreddyMaintenanceMore diff pump newsCold box pumped out to ~ 190 mTorr overnight.Leak
checked the pump again. 
Still leaks at dip stick.Tried resetting
  
  82   Wed Nov 24 07:28:30 2004 Other Mission Impossible   
  83   Wed Nov 24 07:28:38 2004 Other Mission Impossible   
  180   Tue Nov 7 10:35:32 2006 Doug PreddyNormal RunningMaintenance dayI ramped down the solenoid power suplly and
shut off flow to the power 
leads. I will allow them to warm up and (hopefully)
  
  262   Sun Oct 5 11:12:48 2008 Doug PreddyNormal RunningM9B:V1 interlocks and a IOC crashM9A:V1 was removed this shutdown. It's "in"
limit is required to allow M9B:V1 to be opened.
I have bypassed the interlocks on M9B:V1
  
  205   Sun Jul 1 16:09:37 2007 Doug PreddyNormal RunningM9B:DP2 tripped off Diffusion pump tripped off and vacuum (and
power lead temps) going up. I 
had an experimentor cycle HIC 142 at 4:15
  
  111   Wed Nov 2 10:17:27 2005 Willy AnderssonStart UpM9B helium refrigerator and solenoid cooldownStarted M9B MYCOM helium compressor circulating
helium through system last 
Friday. Right hand purifier was regenerated
  
  51   Fri Oct 1 11:39:55 2004 Doug PreddyMaintenanceLoading of compressor not workingWhen the compressor was restarted on Monday,
the remote load button would 
not respond. The compressor was shut off.
  
  129   Fri Jun 16 08:12:04 2006 Doug PreddyNormal RunningLI108 indicates high on EPICS screenLI108 indicates high on EPICS screen (150%)
 Controller shows ~64%
  
  258   Tue Sep 9 15:04:23 2008 Doug PreddyMaintenanceJumpered interlock for diff pump water flowJumpered interlock for diff pump water flow
in Proteus box.
  
  9   Wed Apr 21 06:48:31 2004 Doug PreddyStart UpJT bypass closedClosed cv155   
  28   Fri May 21 09:27:39 2004 Doug PreddyMaintenanceIdle modePower supply was ramped down yesterday as
there are no users.
The Sundance server was rebooted yesterday.
  
  249   Sun Jul 13 20:17:24 2008 Doug PreddyOtherHelium refrigerator problemsIn early June the helium compressor was running
with the solenoid cold. A vavlve broke off
at the suction side of the compressor. The
  
  284   Mon Aug 9 14:27:12 2010 Doug PreddyShut DownHelium plant shut down    
  156   Thu Aug 10 16:33:11 2006 Willy AnderssonQuench RecoveryFlow to solenoid tripped on powerlead temperature.Power lead temps were drifting up, tried to
tweak TIC121 and HIC141 to 
unstall powerlead flow. Got quench. Restarted
  
  179   Fri Nov 3 12:09:51 2006 Doug PreddyNormal RunningFine tuningThe power lead temperatures have been climbing
at a rate of ~ 0.02K very 8 
hours. The negative lead started to show
  
  1   Fri Dec 19 11:34:10 2003 Doug PreddyQuench RecoveryFinal home of this E-loghttp://isac.triumf.ca/m9sol/
Chris has set up the M9 solenoid e-log at
the above web address. Play with 
  
  146   Fri Jul 21 23:36:21 2006 Doug PreddyQuench RecoveryFinal cooldown problems80K shield is down to <60K Liquid level not
building. Power leads rippling 
and warming. T1 inlet 85K with 10%heater
  
  195   Wed May 2 16:28:16 2007 Doug PreddyNormal RunningEnergizing solenoid to 500 ampsPower supply initially would not ramp up.
I exercised the rate control 
selector switch and found that the power
  
  41   Sat Aug 7 23:40:47 2004 Willy AnderssonQuench RecoveryEnergizing solenoid23:39 Powerleads cold, LHe level over 40%
and rising. Energizing solenoid 
to 500 Amps, ramp rate 60. Checking compressor
  
  178   Tue Oct 31 08:37:34 2006 Doug PreddyQuench RecoveryEnergizing solenoid    
  211   Sun Jul 29 04:58:27 2007 Doug PreddyQuench RecoveryEnergizing solenoidLHe level~50%. Still have a slight ripple
in the power leads, but they are 
cold enough to power the solenoid. LHe_in
  
  168   Thu Oct 12 19:54:10 2006 Willy AnderssonQuench RecoveryEnergized solenoid.Switched to series mode about 6 hours ago.
Cooling continued including 
power leads until LHe level at about 50%
  
  169   Fri Oct 13 00:13:57 2006 ExperimentorQuench RecoveryEnergized solenoid.Hi there,

This is Khashayar (experimenter). Lookin
  
  128   Wed Jun 14 16:27:01 2006 Willy AnderssonQuench RecoveryEnergized solenoidPowerleads appear stable.   
  130   Fri Jun 30 11:13:00 2006 Doug PreddyNormal RunningE-log is back up and runningI would have made an entry when it died, but-----   
  125   Tue Jun 13 15:52:46 2006 Willy AnderssonQuench RecoveryDoug has reduced and now closed CV155.Now in full // cold return mode.
Had just talked with Curtis, He said thta
he has already informed the 
  
  35   Wed Jul 28 21:26:19 2004 Doug PreddyQuench RecoveryDouble Power bumpTwo power bumps tonight. Everything tripped
of. Restarted cooling water 
pump, diff. pump under solenoid etc. Computer
  
  68   Sun Oct 17 05:03:13 2004 OtherStart UpCycle valve after oscillations begin(Sat night/Sun AM) Helium temperature started
oscillating, so
powered off magnet (precautionary) and cycled
  
  208   Thu Jul 5 11:01:47 2007 Kevin TrithardtOtherCurrent lead warm upDue to curent lead temperature occilation
Doug recommended warming up the 
current leads.
  
  143   Fri Jul 21 07:39:37 2006 Doug PreddyQuench RecoveryCool down continuesRaised PIC229 to 11Bar. Started to close HIC155.
Cooling rates improve.
  
  29   Fri Jun 25 12:36:04 2004 Willy AnderssonQuench RecoveryCompressor unloaded etc.At about 09:30 this morning I noticed the
sound coming from the compressor 
building seemed louder than usual. After
  
  40   Sat Aug 7 22:37:04 2004 Willy AnderssonQuench RecoveryCompressor tripped off.Got a call at about 10:00 this morning, the
whole system had tripped off. 
No power bumps or other disturbances. When
  
  124   Tue Jun 13 10:54:27 2006 Doug PreddyQuench RecoveryCompressor tripped offNo indicator lights etc as to why the compressor
tripped. The only  light 
was a cold box fault. I had to shut off the
  
  181   Wed Nov 15 11:09:28 2006 Doug PreddyQuench RecoveryCompressor tripped offCompressor tripped this morning at ~8:45.
I restarted the compressor, but 
turbine 1 would not spin up. I let it warm
  
  167   Thu Oct 12 11:53:04 2006 Willy AnderssonQuench RecoveryCompressor trip, 09:50 this morning.Main compressor tripped, refrigerator shutdown.
Checked with experimentors 
and main control room, no indications of
  
  106   Mon Jul 18 15:51:17 2005 Doug PreddyQuench RecoveryCompressor restarted, flow to solenoidRecovered the vacuum in the solenoid, then
started the compressor.
Started flow to solenoid in parallel warm
  
  213   Mon Aug 6 08:41:31 2007 Doug PreddyQuench RecoveryCold box trip at 6:00compressor off due to cold box trip at 6:00
this morning. Restarted system 
and started flow to solenoid in parallel
  
  4   Fri Apr 16 10:15:26 2004 Doug PreddyStart UpChanges to M9sol Vacuum PCYesterday and today there have been a number
of changes made to the M9 
Vacuum control computer. The computer now
  
  120   Fri May 26 08:45:38 2006 Doug PreddyStart UpChanged to series modeChanged to series mode.
Liquid level building.
  
  13   Fri Apr 23 09:19:30 2004 Doug PreddyNormal RunningCV211PIC 229 was down to 10.8 bar this morning.
PIC 229 was trying to open CV211 
to charge the system. I realized that in
  
  119   Fri May 26 07:29:54 2006 Doug PreddyStart UpCV155 closedCV155 closed.
Temps about 15K
  
  87   Mon Nov 29 09:28:16 2004 Doug PreddyNormal RunningBuffer tank emptyCame down to check on system. CV211 was fully
open, trying to charge the 
system. Ripples on LHe in and PL neg. CV141
  
  80   Tue Nov 23 11:35:44 2004 Doug PreddyQuench RecoveryAnother power bumpCompressor tripped off again due to a power
bump. Meson hall water off, so 
vacuum pumps off again.
  
  27   Thu May 6 09:32:13 2004 Willy AnderssonQuench RecoveryAdditional notes from last night power lead problems.When performing last nights refrigerator adjustments
I also increased the 
compressor discharge pressure to 11.25 barA.
  
  165   Thu Oct 5 11:51:53 2006 Doug PreddyStart UpAdde 2 bottles UHP HeliumValved off purifiers, but left LN2 on.   
  233   Fri Nov 9 16:24:33 2007 Doug PreddyMaintenance2APLC Diedthumb wheel panel all lite up. 2APLC rebooted
by Klara (Thanks). M9BMP2 not 
pumping properly. Edi is looking at it. 
  
  152   Wed Aug 9 06:44:58 2006 Willy AnderssonQuench Recovery// cold mode cooling continues.After restarting flow to the solenoid yesterday
in // cold return mode 
TIC121 just refused to go lower than about
  