  30   Tuesday, December 11, 2012, 15:44 Anders MjosSiCHP-FEBIADITWTM4nana (old style fittings)ReadyTest target for the Conditioning Station

 The target has been used for operator training previously at the ISAC Test Stand.

  • 9-pin connector was completed and grounded
  • RH gripper arms added to the heat shield

Further documentation on DocuShare

  31   Tuesday, March 19, 2013, 13:32 John WongSiC#28-RFQLP-SIS TM4LP-RFQ (NEW) SpentSiC#28-RFQ target description

 Tuesday, March 19, 2013, 12:28

SiC#28 - RFQ target description: 

Total SiC foils = 325 discs

Total SiC + C-grv foils = 38.48 g 

Total length = 10.9cm 

Average mass of SiC/disc (without graphite foil) = 0.083 g

Average thickness of SiC (without graphite foil) = 0.18 mm 

No Rh foil added

See attachment #1:  SiC#28 RFQ density measurements 

See link for more details:  https://documents.triumf.ca/docushare/dsweb/View/Collection-9771


 

 

Attachment 1: SiC#28_RFQ_density_measurements.xlsx
  90   Monday, July 22, 2019, 16:33 Marla CervantesUCx#27LP-SIS  49044-9 ReadyUCx#27 Target Description

 Monday, July 22, 2019, 16:31

Attached please find the UCx#27 target description. 


 

Attachment 1: UCx#27.JPG
UCx#27.JPG
  123   Tuesday, October 25, 2022, 13:10 Darwin Ortiz RosalesUCx#40LP-SIS  45550-2 ReadyUCx#40 Target Description

Tuesday, October 25, 2022, 13:10

Target in Glovebox

Attachment 1: UCx#40_Description.png
UCx#40_Description.png
  135   Thursday, August 24, 2023, 14:16 Aaron SchmidtNb#11LP-IG-LISITWTM245447-9 In progressNb#11 Target description

Nb#11 LP-IG-LIS target description:

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm (0.001")

Target density =  7.13 g/cm3

Target length = 3.53 cm

Total Mass = 25.165 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven=  434 g (w/ end caps)

Discs in target are arranged as follows: 1pack / 75 singles / 1pack / 100 singles / 2packs / 100 singles / 1pack / 75 singles / 1pack

Dimpling direction on foils/ Packs.                                                               --> Center <--

End caps installed (Not expanded). No Ta foil heat shield added

Replacement for Nb#10

Attachment 1: Nb#11_Target_Loading_Data.pdf
Nb#11_Target_Loading_Data.pdf
  150   Wednesday, April 23, 2025, 10:35 Darwin Ortiz RosalesUCx#48LP-IG-LIS  45447-8 ReadyUCx#48_Description
Attachment 1: UCx#48_Description.png
UCx#48_Description.png
  132   Friday, June 23, 2023, 09:02 Aaron SchmidtNb#10LP-IG-LISITWTM245447-7  Nb#10 target description.

 Nb#10 LP-IG-LIS target description:

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm (0.001")

Target density =  7.36 g/cm3

Target length = 3.5 cm

Total Mass = 25.7576 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven=  438 g (w/ end caps)

Discs in target are arranged as follows: 1pack / 75 singles / 1pack / 100 singles / 2packs / 100 singles / 1pack / 75 singles / 1pack

Dimpling direction on foils/ Packs.                                                               --> Center <--

End caps installed (Not expanded). No Ta foil heat shield added

  142   Friday, June 14, 2024, 14:16 Darwin Ortiz RosalesTiC#7HP-FEBIAD  45447-545235-15ReadyTiC#7_Description

Please note that this target comes in a Graphite cartridge.

Attachment 1: TiC#7.png
TiC#7.png
  145   Tuesday, September 10, 2024, 09:53 Darwin Ortiz RosalesTiC#9HP-SIS  45447-2945235-19ReadyTiC#9_Description
Attachment 1: TiC#9.png
TiC#9.png
  147   Monday, October 07, 2024, 13:16 Aaron SchmidtTa#68HP-SISITETM245447-2645235-16ReadyTa#68 Target Description

 Ta#68 HP-SIS target description: 

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.0254mm (0.0010") avg Measured. 

Target material batch: PO# TR-2023-06149

Target density = 14.204 g/cm3

Target length = 3.50 cm

Total Mass = 49.7146 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven =  1483g (w/ end caps)

Discs in target are arranged as follows: 1pack / 75 singles / 1pack / 100 singles / 2packs / 100 singles / 1pack / 75 singles / 1pack

Dimpling direction on foils/ Packs.                                                               --> Center <--

End caps installed (Expanded). No Ta foil heat shield added

Attachment 1: ITE-TM2-Ta#68-HP-SIS_Target_Loading_Data.pdf
ITE-TM2-Ta#68-HP-SIS_Target_Loading_Data.pdf
  144   Tuesday, September 03, 2024, 16:11 Darwin Ortiz RosalesTiC#8HP-FEBIAD  45447-2 ReadyTiC#8_Description

This was back-up target

Attachment 1: TiC#8_(2).png
TiC#8_(2).png
  151   Monday, May 26, 2025, 09:54 Darwin Ortiz RosalesUCx#49LP-SIS  45447-1942972-2ReadyUCx#49_Description
Attachment 1: UCx#49_Description.png
UCx#49_Description.png
  146   Friday, September 20, 2024, 14:26 Aaron SchmidtTa#LP-SISITETM245447-1845235-13ReadyTa# Target Description

Spare for Ta#68 Ta# LP-SIS target description: 

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.0254mm (0.0010") avg Measured. 

Target material batch: PO# TR-2023-06149

Target density = 14.17 g/cm3

Target length = 3.505 cm

Total Mass = 47.6553 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven =  454g (w/ end caps)

Discs in target are arranged as follows: 1pack / 75 singles / 1pack / 100 singles / 2packs / 100 singles / 1pack / 75 singles / 1pack

Dimpling direction on foils/ Packs.                                                               --> Center <--

End caps installed (Expanded). No Ta foil heat shield added

  139   Tuesday, March 05, 2024, 10:15 Aaron SchmidtTaLP-SIS  45447-15 ReadyTa# (March stock) Target Description.

Tuesday, March 05, 2024, 10:12: Ta# (Stock - Loaded today) LP-SIS target description: 

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.033mm (0.0013") avg Measured. 

Target material batch: Ta 2023-1  Mt'l PO# TR-2022-03028

Target density = 18.2 g/cm3

Target length = 3.4 cm

Total Mass = 61.9 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven =  459.6 g (w/ end caps)

Discs in target are arranged as follows: 1pack / 75 singles / 1pack / 100 singles / 2packs / 100 singles / 1pack / 75 singles / 1pack

Dimpling direction on foils/ Packs.                                                               --> Center <--

End caps installed (Not expanded). No Ta foil heat shield added.

  143   Tuesday, August 13, 2024, 10:08 Aaron SchmidtTa#67 Thin Foil TargetLP-SISITETM245447-1445235-10ReadyTa#67 Thin Foil Target description & Notes

  Ta#67 Thin foil LP-SIS target description: 

Total Ta foils = 470 loaded into 2 packs of 235 foils each.

Foil Thickness = 0.010mm (0.0004") avg Measured. 

Target material batch: See attached*

Target density = 4.498 g/cm3

Target length = 3.7465 cm

Total Mass = 16.8509 g [Incl foils + Ta wire on 2 packs]

Total mass w/ Target oven =  415g (No end caps)

Discs in target are arranged as follows: 2packs

Dimpling direction on foils/ Packs.                                                               --> Center <--

No Ta foil heat shield added


The original HS 45253-23 Had to be replaced due to a leak being discovered during a leak check. Reassembled in 45235-10

However post-TS https://elog.triumf.ca/TIS/Test-Stand/1014 & https://elog.triumf.ca/TIS/Test-Stand/1015, it was discovered there was electrical shorts, components stuck together, including the retainer to the ovens tube. This had to be repaired & rebuilt.

Repairs included opening ITD0542 inner diameter to 11/16"

increasing the slot gap on the HS for the target leg to bottom insulator clearance

Source Cup inner ID opened 0.030" as the short was found to be the tube touching the Cu wall.

Top & bottom insulators were also replaced pre-TS. Checked post TS. They held 540 OL & 1049 OL respectively afterwards.

Run in TS: https://elog.triumf.ca/TIS/Test-Stand/1018

Link to Checklist & Docushare details below.

https://documents.triumf.ca/docushare/dsweb/View/Collection-39715

Attachment 1: Certificate_of_Analysis_-_Ta_0.01mm.pdf
Certificate_of_Analysis_-_Ta_0.01mm.pdf
  149   Thursday, January 02, 2025, 12:54 Darwin Ortiz RosalesUCx#47LP-SIS  45447-12 ReadyUCx#47_Description
Attachment 1: UCx#47_Description.png
UCx#47_Description.png
  134   Wednesday, July 26, 2023, 12:03 Darwin Ortiz RosalesUCx#43LP-SIS  45447-11 ReadyUCx#43_Description
Attachment 1: UCx#43_Description.png
UCx#43_Description.png
  140   Wednesday, May 22, 2024, 22:32 Darwin Ortiz RosalesUCx#45LP-SIS  45220-9 ReadyUCx#45_Description
Attachment 1: UCx#45_Description.png
UCx#45_Description.png
  127   Monday, March 06, 2023, 13:38 Aaron SchmidtTa#65LP-SIS  45220-7 ReadyTa#65 Target Description.

 Monday, March 06, 2023, 13:38:

Ta#65 LP-SIS target description: 

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.033mm (0.0013") avg Measured. 

Target material batch: Ta 2023-1  Mt'l PO# TR-2022-03028

Target density = 18.2685 g/cm3

Target length = 3.40 cm

Total Mass = 62.113 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven= 460.8 g (w/ end caps)

Discs in target are arranged as follows: 1pack / 75 singles / 1pack / 100 singles / 2packs / 100 singles / 1pack / 75 singles / 1pack

Dimpling direction on foils/ Packs.                                                               --> Center <--

End caps installed (Not expanded). No Ta foil heat shield added

 

Tuesday, May 16, 2023, 10:44: End cap removed post conditioning to remove rhenium foil. End cap replaced, & weight updated. Length remains the same.

  126   Monday, February 06, 2023, 10:02 Darwin Ortiz RosalesTarget Reassigned (VOID ENTRY)   45220-6 ReadyTarget C#6 Description
Attachment 1: C#6_Description.png
C#6_Description.png
  152   Friday, May 30, 2025, 14:23 Darwin Ortiz RosalesC#5LP-SIS  45220-6 ReadyC#5 conditioning

Friday, May 30, 2025, 14:21

Attachment 1: C#5_Description.png
C#5_Description.png
  136   Wednesday, September 20, 2023, 14:25 Aaron SchmidtTa#66LP-SISITWTM245220-5 In progressTa#66 Target description.

Ta#66 LP-SIS target description: 

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.033mm (0.0013") avg Measured. 

Target material batch: Ta 2023-1  Mt'l PO# TR-2022-03028

Target density = 12.446 g/cm3

Target length = 5.0 cm

Total Mass = 62.230 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven=  g (w/ end caps)

Discs in target are arranged as follows: 1pack / 75 singles / 1pack / 100 singles / 2packs / 100 singles / 1pack / 75 singles / 1pack

Dimpling direction on foils/ Packs.                                                               --> Center <--

End caps installed (Not expanded). No Ta foil heat shield added

Attachment 1: Ta#66_Target_Loading_Data.pdf
Ta#66_Target_Loading_Data.pdf
  137   Thursday, October 05, 2023, 12:48 Aaron Schmidt LP-SIS  45220-3  Ta#( ) Target description. (Spare)

Ta# LP-SIS target description: 

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.033mm (0.0013") avg Measured. 

Target material batch: Ta 2023-1  Mt'l PO# TR-2022-03028

Target density = 12.515 g/cm3

Target length = 4.97 cm

Total Mass = 62.202 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven =  461g (w/ end caps)

Discs in target are arranged as follows: 1pack / 75 singles / 1pack / 100 singles / 2packs / 100 singles / 1pack / 75 singles / 1pack

Dimpling direction on foils/ Packs.                                                               --> Center <--

End caps installed (Not expanded). No Ta foil heat shield added

Attachment 1: Ta#68_Target_Loading_Data.pdf
Ta#68_Target_Loading_Data.pdf
  124   Tuesday, November 15, 2022, 08:41 Darwin Ortiz RosalesUCx#41LP-SIS  45220-13 Ready 
Attachment 1: UCx#41_Description.png
UCx#41_Description.png
  133   Wednesday, July 26, 2023, 09:42 Darwin Ortiz RosalesUCx#42LP-SIS  45220-1 ReadyUCx#42_Description
Attachment 1: UCx#42_Description.png
UCx#42_Description.png
  115   Wednesday, May 25, 2022, 10:15 Darwin Ortiz RosalesSiC#42HP-FEBIAD  44692-9 ReadyTarget SiC#42 Description

Wednesday, May 25, 2022, 10:15

Attachment 1: Description-SiC#42.png
Description-SiC#42.png
  129   Monday, May 01, 2023, 14:59 Darwin Ortiz RosalesTiC #5LP-FEBIAD  44692-5 ReadyTarget TiC #5 Description
Attachment 1: TiC#5.png
TiC#5.png
  131   Friday, May 26, 2023, 11:53 Darwin Ortiz RosalesSiC#45HP-SIS  44692-3 Ready 
Attachment 1: SiC#45.png
SiC#45.png
  141   Friday, June 14, 2024, 14:12 Darwin Ortiz RosalesSiC#46HP-SIS  44692-245235-6ReadySiC#46_description
Attachment 1: SiC#46.png
SiC#46.png
  130   Friday, May 26, 2023, 11:35 Darwin Ortiz RosalesTiC#6HP-SIS  44692-10 ReadyTarget TiC #6 Description
Attachment 1: TiC_#6.png
TiC_#6.png
  111   Tuesday, September 21, 2021, 21:08 Darwin Ortiz RosalesC#3LP-SIS  44691-7 ReadyTarget C #3 Description

 Tuesday, September 21, 2021, 21:09

See attached file for target description.

Attachment 1: C#3_Target_description.png
C#3_Target_description.png
  109   Monday, August 16, 2021, 11:02 John WongUCx#34LP-SISITWTM244691-544378-1ReadyTarget description

Monday, August 16, 2021, 11:01

See attachments for the target description and SEM images


 

Attachment 1: UCx#34_target_description.PNG
UCx#34_target_description.PNG
Attachment 2: UCx_34_SEM_Density_Measurements.pdf
UCx_34_SEM_Density_Measurements.pdf UCx_34_SEM_Density_Measurements.pdf UCx_34_SEM_Density_Measurements.pdf UCx_34_SEM_Density_Measurements.pdf UCx_34_SEM_Density_Measurements.pdf UCx_34_SEM_Density_Measurements.pdf UCx_34_SEM_Density_Measurements.pdf UCx_34_SEM_Density_Measurements.pdf
  108   Wednesday, June 23, 2021, 18:13 Aaron SchmidtTa#62LP-SISITETM444691-4 In progressTa#62 Target Description (2)

 Ta#62 LP-SIS target description: (REPLACEMENT FOR 44139-10)

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm

Target density = 14.29 g/cm3

Target length = 3.4 cm

Total Mass = 48.602 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven= 456 g (w/ end caps)

Discs in target are arranged as follows: 1pack / 75 singles / 1pack / 100 singles / 2packs / 100 singles / 1pack / 75 singles / 1pack

End caps installed (Not expanded). No Ta foil heat shield added

  122   Thursday, September 22, 2022, 13:18 Darwin Ortiz RosalesC#4LP-SIS  44691-3 ReadyTarget C#4 Description

Thursday, September 22, 2022, 13:17

Attachment 1: C#4_Target_Description.png
C#4_Target_Description.png
  125   Wednesday, December 14, 2022, 10:28 Darwin Ortiz RosalesTarget reassigned (VOID ENTRY)   44691-13 ReadyTarget C#5 Description

Graphite target

Attachment 1: C#5_Description.png
C#5_Description.png
  116   Thursday, June 02, 2022, 16:55 Aaron SchmidtTa#64LP-SISITETM444691-12 In progressTa#64 Target Description

Thursday, June 02, 2022, 16:54: 

Ta#64 LP-SIS target description: 

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.0254mm (0.0010") Measured.

Target density = 12.005 g/cm3

Target length = 3.35 cm

Total Mass = 40.2166 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven= 446 g (w/ end caps)

Discs in target are arranged as follows: 1pack / 75 singles / 1pack / 100 singles / 2packs / 100 singles / 1pack / 75 singles / 1pack

End caps installed (Not expanded). No Ta foil heat shield added

 
Attachment 1: Ta#64_Target_Loading_Data.pdf
Ta#64_Target_Loading_Data.pdf
  114   Wednesday, November 10, 2021, 18:09 Aaron SchmidtTa#63LP-SISITETM444691-11 In progressTa#63

  Ta#63 LP-SIS target description: 

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm

Target density = 15.38 g/cm3

Target length = 3.4 cm

Total Mass = 52.296 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven= 457 g (w/ end caps)

Discs in target are arranged as follows: 1pack / 75 singles / 1pack / 100 singles / 2packs / 100 singles / 1pack / 75 singles / 1pack

End caps installed (Not expanded). No Ta foil heat shield added

Attachment 1: Ta#63_Target_Loading_Data.pdf
Ta#63_Target_Loading_Data.pdf
  113   Monday, November 08, 2021, 11:30 Darwin Ortiz RosalesUCx#36LP-SIS  44691-1 ReadyUCx#36_Target_Description
Attachment 1: UCx#36_properties.png
UCx#36_properties.png
  119   Friday, July 15, 2022, 10:47 Darwin Ortiz RosalesUCx#38LP-SIS  44508-3 ReadyUCx#38_Target_Description

The bulk density was corrected.  Use attachment number 2.

 

Darwin

Attachment 1: UCx#38.png
UCx#38.png
Attachment 2: UCx#38.png
UCx#38.png
  103   Tuesday, October 27, 2020, 12:53 Aaron SchmidtNb#9LP-SISITETM444508-2 ReadyNb#9 LP-SIS target description

 Nb#9 LP-SIS target description:

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm

Target density =  7.24 g/cm3

Target length = 3.5 cm

Total Mass = 25.343 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven=  425 g (w/ end caps)

Discs in target are arranged as follows: 1pack / 75 singles / 1pack / 100 singles / 2packs / 100 singles / 1pack / 75 singles / 1pack

Attachment 1: Nb#9_Target_Loading_data.pdf
Nb#9_Target_Loading_data.pdf Nb#9_Target_Loading_data.pdf
  102   Tuesday, October 06, 2020, 14:25 Aaron SchmidtTa#61LP-SISITETM444508-1 In progressTa#61 Target Description

 Ta#61 LP-SIS target description:

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm

Target density =  12.76 g/cm3

Target length = 3.4 cm

Total Mass = 43.388 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven=  464 g (w/ end caps)

Discs in target are arranged as follows: 1pack / 100 singles / 1pack / 75 singles / 2packs / 75 singles / 1pack / 100 singles / 1pack

End caps installed (not yet expanded). No Ta foil heat shield added 

  104   Thursday, November 12, 2020, 11:32 Darwin Ortiz RosalesUCx#31HP-SISITETM44416043822ReadyUCx#31-Target_description

 Target full name ITE-TM4-UC#31-P2N-SIS

 

 

Attachment 1: UCx31-target_description.png
UCx31-target_description.png
  101   Tuesday, September 29, 2020, 11:50 Marla CervantesC#2LP-SISITWTM244139-9 In progressTarget C#2 Description

 Tuesday, September 29, 2020, 11:49

See attached picture for description 

Attachment 1: C2-Description.JPG
C2-Description.JPG
  98   Thursday, February 20, 2020, 08:11 Aaron SchmidtTa#60LP-SIS  44139-13 ReadyTa#60 Target Description

 Thursday, February 20, 2020, 08:07: 

Ta#60 LP-SIS target description:

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm

Target density =  14.081 g/cm3

Target length = 3.33 cm

Total Mass =  46.891 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven=  445g w/ end caps)

Discs in target are arranged as follows: 1pack / 75singles / 1pack / 100singles / 2packs / 100singles / 1pack / 75singles / 1pack

End caps installed (Expanded). No Ta heat shield added.

Attachment 1: Ta#60_Loading_Data.pdf
Ta#60_Loading_Data.pdf
  107   Wednesday, June 09, 2021, 11:45 Darwin Ortiz RosalesUCx#33LP-SIS  44139-12 ReadyUCx#33-Target_description

 Cartridge with discs was further heated in EVAP1 (eLog 271 Evaporator 1).

Attachment 1: UCx33-target_description.png
UCx33-target_description.png
  106   Thursday, April 08, 2021, 09:54 Aaron SchmidtTa#62LP-SIS  44139-10 ReadyTa#62 target description

Ta#62 LP-SIS target description:

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm

Target density = 13.96 g/cm3

Target length = 3.4 cm

Total Mass = 47.461 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven= 445 g (w/ end caps)

Discs in target are arranged as follows: 1pack / 75 singles / 1pack / 100 singles / 2packs / 100 singles / 1pack / 75 singles / 1pack

End caps installed (expanded). No Ta foil heat shield added

  95   Monday, October 21, 2019, 12:38 Darwin Ortiz RosalesUCx#29LP-SIS  44062-6  ReadyUCx29-target_description
Attachment 1: UCx#29.PNG
UCx#29.PNG
  92   Tuesday, September 17, 2019, 08:47 Aaron SchmidtTa#59LP-SIS TM244062-5 In progressTa#59 Target description

Tuesday, September 17, 2019, 08:44: 

Ta#59 LP-SIS target description:

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm

Target density =  13.74 g/cm3

Target length = 3.4 cm

Total Mass =  46.7314 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven=  436 g (440 g w/ end caps)

Discs in target are arranged as follows: 1pack / 75singles / 1pack / 100singles / 2packs / 100singles / 1pack / 75singles / 1pack

End caps installed (not yet expanded). No Ta heat shield added.

Attachment 1: Ta#59_Loading_Data.pdf
Ta#59_Loading_Data.pdf
  93   Wednesday, October 02, 2019, 11:18 Marla CervantesUCx#28LP-SIS  44062-4 ReadyUCx #28 Target description

 Wednesday, October 02, 2019, 11:17

Attached please find the UCx#28 target description. 

Attachment 1: UCx28-target_description.JPG
UCx28-target_description.JPG
  96   Tuesday, November 19, 2019, 12:20 Darwin Ortiz RosalesUCx#30LP-SIS  44062-3  Ready UCx30-target_description
Attachment 1: UCx#30.PNG
UCx#30.PNG
Attachment 2: UCx30-taret_description_revised.png
UCx30-taret_description_revised.png
  100   Thursday, June 04, 2020, 15:00 Marla CervantesC#1LP-SIS  43044-8 In progressTarget C#1 Description

 Thursday, June 04, 2020, 15:00

See attached picture for description 

Attachment 1: C1-Description.PNG
C1-Description.PNG
  91   Tuesday, July 30, 2019, 15:52 Darwin Ortiz RosalesZrC#8LP-SISITETM243044-5 ReadyTarget description

 

Tuesday, July 30, 2019, 15:50

See attachment for details


Thursday, August 01, 2019, 07:41

See attachment#2 for SEM - JW

 

 

 

Attachment 1: ZrC8.png
ZrC8.png
Attachment 2: ZrC_8_SEM.pdf
ZrC_8_SEM.pdf ZrC_8_SEM.pdf ZrC_8_SEM.pdf ZrC_8_SEM.pdf ZrC_8_SEM.pdf
  89   Friday, May 31, 2019, 16:11 Aaron SchmidtTa#58LP-SISITETM243044-442972-7ReadyTa#58 Target desription

 Ta#58 LP-SIS target description:

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm

Target density =  14.16 g/cm3

Target length = 3.4 cm

Total Mass =  48.138 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven=  464 g

Discs in target are arranged as follows: 1pack / 75singles / 1pack / 100singles / 2packs / 100singles / 1pack / 75singles / 1pack

End caps installed (not yet expanded). No Ta heat shield added.

Attachment 1: Ta#58_Loading_data.pdf
Ta#58_Loading_data.pdf
  85   Monday, September 17, 2018, 11:39 John Wongnp-SiC#1LP-SIS  43044-1 Readynano-SiC#1 Target description

Monday, September 17, 2018, 11:32

See attachment for the target description.

Some notes about the target conditioning in Evap#2 (see elog ID#414):

TGHT at 500A, 8.4V, temp at 1600 C (4.2 kW)

TGHT at 475/7.84V, temp at 1550 C (3.72 kW - Max operating target heating power)

TGHT at 450/7.31V, temp at 1500 C (3.28 kW)


 

Attachment 1: target_description.JPG
target_description.JPG
  94   Friday, October 11, 2019, 09:26 Darwin Ortiz RosalesSiC#38HP-FEBIAD TM242464-4 ReadyTarget SiC#38 Description

Friday, October 11, 2019, 09:25

See attached picture for description.

 

Attachment 1: SiC#38.PNG
SiC#38.PNG
  97   Friday, November 29, 2019, 10:53 Darwin Ortiz RosalesSiC#39HP-FEBIAD TM242464-3 ReadyTarget SiC#39 Description

Friday, December 06, 2019, 15:34

Added SEM images (see attachment#2) - JW


 

Attachment 1: SiC#39_description.PNG
SiC#39_description.PNG
Attachment 2: SiC39_SEM.pdf
SiC39_SEM.pdf SiC39_SEM.pdf SiC39_SEM.pdf SiC39_SEM.pdf
  117   Monday, June 20, 2022, 10:35 Darwin Ortiz RosalesSiC#43HP-FEBIAD  42464-1 ReadyTarget SiC#43 Description

Monday, June 20, 2022, 10:33

  88   Thursday, November 29, 2018, 14:23 Marla CervantesUCx#26LP-IG-LIS  42281-3 ReadyUCx#26 Target description

Thursday, November 29, 2018, 14:20

Attached please find the UCx#26 target description. 


 

Attachment 1: UCx_26-description.JPG
UCx_26-description.JPG
  82   Thursday, June 21, 2018, 11:34 Aaron SchmidtTa#56HP-SIS  42218-9 ReadyTa#56 target description

 Thursday, June 21, 2018, 11:33:

Ta#56 HP-SIS target description:

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm

Target density = 14.01 g/cm3

Target length = 3.4 cm

Total Mass = 47.374 g [Incl foils + Ta wire on 6 packs]

Total mass w/ Target oven= 1469 g

No endcaps inserted, no Rh foil inserted.


Tuesday, July 03, 2018, 08:30: Endcaps inserted & expanded. Rh foil inserted.

Attachment 1: Ta#56_Loading_Data.pdf
Ta#56_Loading_Data.pdf
  99   Wednesday, April 15, 2020, 08:10 Darwin Ortiz RosalesSiC#40HP-SIS  42218-8 ReadyTarget SiC#40 Description

Wednesday, April 15, 2020, 07:51

See attached picture for description 

Attachment 1: Annotation_2020-04-15_080836.png
Annotation_2020-04-15_080836.png
  110   Monday, September 13, 2021, 14:53 Darwin Ortiz RosalesSiC#41HP-SIS  42218-7 ReadyTarget SiC#41 Description

Monday, September 13, 2021, 14:52

Attachment 1: Screenshot_2021-09-13_144830.png
Screenshot_2021-09-13_144830.png
  76   Monday, June 26, 2017, 13:36 Laura LambertTa #53HP-SIS TM242218-6 ReadyTa#53 target description

 Monday, June 26, 2017, 13:32

Ta#53 HP-SIS target description:

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm

Target density = 14.80 g/cm3

Target length = 3.4cm

Total Mass = 50.3105g [Incl foils + Ta wire on 6 packs]


 

  112   Wednesday, September 29, 2021, 10:37 Darwin Ortiz RosalesUCx#35LP-IG-LIS  42218-5 ReadyUCx#35-Target_description

 

Target ID UCx#35
General Total No. Discs 179  
Disc area 0.95 cm2
Ave graphite foil thickness/disc  0.013 cm
Ave. graphite foil mass/disc 0.015 g
  Total mass of discs 14.97 g
Green Target Total Green mass of discs 13.89 g
Ave. mass/disc  0.084 g
Ave. mass/disc (no C-foil & org) 0.078 g
Ave. thickness/disc (no C-foil) 0.037 cm
Theoretical length (with C-foil) 9.0 cm
Total packed length (with C-foil) 10.0 cm
Final Target UC2 in Target 11.05 g
Free C in Target 0.83 g
Total massUC2+free C 11.88 g
U mass 10.04 g
U thickness 10.56 g/cm2
UC2 bulk density (measured) 2.88 g/cm3
Theoretical Density 8.24 g/cm3
% Yield Density 34.91 %
Theoretical Porosity 0.651  
Attachment 1: UCx#35_Target_Description_.png
UCx#35_Target_Description_.png
  120   Tuesday, August 09, 2022, 11:02 Darwin Ortiz RosalesUCx#39LP-IG-LIS  42218-4 ReadyUCx#39_Target_Description
Attachment 1: UCx#39_Description.png
UCx#39_Description.png
  148   Tuesday, November 12, 2024, 12:49 Darwin Ortiz RosalesUCx#46LP-IG-LIS  42218-2 ReadyUCx#46_Description
Attachment 1: UCx#46_Description.png
UCx#46_Description.png
  78   Monday, August 21, 2017, 11:13 Laura LambertUCx#21LP-SIS  42218-15 ReadyUCx#21 Target Description

 Monday, August 21, 2017, 11:09

Attached please find the UCx#21 target description.


 

Attachment 1: target_description.JPG
target_description.JPG
  75   Wednesday, May 31, 2017, 09:53 Laura LambertTa# 52LP-SIS TM442218-14 In progressTa#52 target description

 Wednesday, May 31, 2017, 09:51

Ta#52 LP-SIS target description:

Total Ta foils = 400 [4x20 packs + 320 single discs]

Foil Thickness = 0.025mm

Target density = 14.77 g/cm3

Target length = 2.9cm

Total Mass = 42.8228g [Incl foils + Ta wire on 4 packs]


 

  81   Monday, April 16, 2018, 09:55 Laura LambertUCx#22LP-SIS  42218-13 ReadyUCx#22 Target Description

 Monday, April 16, 2018, 09:53

Attached please find the UCx#22 target description.


 

Attachment 1: target_description.JPG
target_description.JPG
  71   Tuesday, March 07, 2017, 10:00 Laura LambertUCx#19LP-SIS TM442218-12 ReadyUCx#19 Target Description

 Tuesday, March 07, 2017, 09:54

Attached please find the UCx#19 target description.


 

Attachment 1: UCx#19_description.jpg
UCx#19_description.jpg
  77   Thursday, July 06, 2017, 15:36 Anders MjosTa#55LP-SIS  42218-11 ReadyTa#55 target description

Target loaded by Matt yesterday

320 foils = 33.7832 g
4 x 20 packs = 9.1170 g
Total Weight = 42.9002 g

Length of foils is 2.88 cm

Target conditioned at ISAC Test Stand (580 A) without Ta heat shields

  79   Tuesday, September 26, 2017, 12:07 Laura LambertTa#54HP-SISITWTM242218-10 ReadyTa#54 target description

Tuesday, September 26, 2017, 09:31

Ta#54 HP-SIS target description:

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm

Target density = 14.72 g/cm3

Target length = 3.4cm

Total Mass = 50.0372g [Incl foils + Ta wire on 6 packs]


 

  73   Tuesday, May 16, 2017, 13:18 Laura LambertUCx#20LP-IG-LISITETM442218-1 ReadyUCx#20 Target Description

 Tuesday, May 16, 2017, 13:13

Attached please find the UCx#20 target description.


 

Attachment 1: target_description.JPG
target_description.JPG
  138   Tuesday, October 24, 2023, 11:30 Darwin Ortiz RosalesUCx#44LP-FEBIAD  41790-7 ReadyUCx#44_Description

Three Ta heatshields left installed on container.

Attachment 1: UCx#44_Description.png
UCx#44_Description.png
  70   Friday, October 21, 2016, 13:47 Laura LambertUCx#18LP-SIS  41790-6 ReadyUCx#18 Target Description

 Friday, October 21, 2016, 13:46

Attached please find the UCx#18 target description.


 

Attachment 1: UCx#18_description.jpg
UCx#18_description.jpg
  69   Tuesday, October 18, 2016, 12:06 Laura LambertTa#50LP-SIS  41790-5 ReadyTa#50 target description

Tuesday, October 18, 2016, 10:12

Ta#50 LP-SIS target description:

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm

Target density = 14.03 g/cm3

Target length = 3.4cm

Total Mass = 47.6889g [Incl foils + Ta wire on 6 packs]

  72   Wednesday, March 08, 2017, 15:42 Laura LambertTa#51LP-SIS TM241790-4 In progressTa#51 target description

 Wednesday, March 08, 2017, 15:40

Ta#51 LP-SIS target description:

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm

Target density = 14.64 g/cm3

Target length = 3.4cm

Total Mass = 49.7916g [Incl foils + Ta wire on 6 packs]


 

  65   Tuesday, August 02, 2016, 15:28 Laura LambertUCx#17LP-SIS  41790-3 ReadyUCx#17 Target Description

 Tuesday, August 02, 2016, 15:27

Attached please find the UCx#17 target description.


 

Attachment 1: UCx#17_description.jpg
UCx#17_description.jpg
  66   Tuesday, August 02, 2016, 15:42 Laura LambertTa # 49HP-SIS  41369-6 ReadyTa#49 target description

 Tuesday, August 02, 2016, 15:30

Ta#49 HP-SIS target description:

Total Ta foils = 500 [6x20 packs + 380 single discs]

Foil Thickness = 0.025mm

Target density = 14.87 g/cm3

Target length = 3.4cm

Total Mass = 50.6142g [Incl foils + Ta wire on 6 packs]


 

  62   Tuesday, April 19, 2016, 11:33 Laura LambertTa # 48HP-SIS  41369-5 ReadyTa # 48 target description

 Tuesday, April 19, 2016, 11:24

Ta#48 HP-SIS target description:

Total Ta foils = 505

Foil Thickness = 0.025mm

Target density = 15.86 g/cm3

Target length = 3.4cm

Total Mass = 53.9381g [Incl foils + Ta wire on 6 packs]


 

  64   Wednesday, June 29, 2016, 11:49 Laura LambertSiC # 33HP-SIS  41369-4 ReadySiC#33 target description

 Wednesday, June 29, 2016, 11:47

See attachment for target description.


 

Attachment 1: elog_target_description.JPG
elog_target_description.JPG
  56   Thursday, September 10, 2015, 13:33 Laura LambertTa # 46HP-SIS  41369-3 ReadyTa#46 target description

 Thursday, September 10, 2015, 13:31

Ta#46 HP-SIS target description:

Total Ta foils = 505

Foil Thickness = 0.025mm

Target density = 15.85 g/cm3

Target length = 3.4cm

Total Mass = 75.2772g


 

  58   Wednesday, September 30, 2015, 17:08 John WongTa # 46HP-SIS  41369-3 ReadyTa#46 target description

Laura Lambert wrote:

 Thursday, September 10, 2015, 13:31

Ta#46 HP-SIS target description:

Total Ta foils = 505

Foil Thickness = 0.025mm

Target density = 16.69 g/cm3

Target length = 3.4cm

Total Mass = 75.2772g


 

 Wednesday, September 30, 2015, 17:01

The average mass per disc = 0.105g

Total mass (505 discs + some Ta wire for packing) = 53.876g  ( Not 75.277g as mentioned above.  See Target Logbook 2015 Dropbox for more details)


 

  74   Tuesday, May 30, 2017, 15:15 Laura LambertSiC#36HP-FEBIAD-CTL TM241369-10 ReadySiC#36 target description

 Tuesday, May 30, 2017, 15:12

See attachment for target description.


 

Attachment 1: SiC#36_Target_Description.JPG
SiC#36_Target_Description.JPG
  87   Monday, October 15, 2018, 07:36 Aaron SchmidtTa#57 ITETM441177-3  Ta#57 Target Description.

(Laura Lambert: Evap 2 Elog ID: 404) Tuesday, April 17, 2018, 10:20

Target consists of 470 discs.

Total packed length 3.4cm [exactly 7.8cm space on both sides]

Total mass g [Incl foils + Ta wire on 6 packs]

Target consists of 350 single discs and 6x20 packs of Ta discs

Discs in target are arranged as follows: 1pack / 75singles / 1pack / 100singles / 2packs / 100singles / 1pack / 75singles / 1pack

 

Weight not included. Should be identical to Ta#54 (Below)

 


Comparable as the foils are all 0.001" & the quantities are the same. 

(Laura Lambert: Target Production Elog ID: 79)

Ta#54 HP-SIS target description:

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm

Target density = 14.72 g/cm3

Target length = 3.4cm

Total Mass = 50.0372g [Incl foils + Ta wire on 6 packs]

  50   Friday, April 17, 2015, 16:00 Laura LambertUCx # 12LP-SISITWTM141177-2 ReadyUCx#12 Target Description

 Thursday, April 16, 2015, 14:07

See attachment for target description

Attachment 1: UCx_12_target_description.PNG
UCx_12_target_description.PNG
  54   Tuesday, July 07, 2015, 13:00 John WongZrC#7LP-SISITWTM241177-2 ReadyZrC#7 (backup1) target description

Tuesday, July 07, 2015, 12:36 

Total no. discs = 422

Total packed length = 13.2 cm (measured after the sintering process) 

Note:  The theoretical length (ZrC + graphite foils) is 10.52cm    After the sintering process, the stack of discs expanded in length.  To avoid crushing the materials while inserting the endcaps, the discs are gentally packed to 13.2cm.

(see attachments for more details)


 

Attachment 1: ZrC#7_backup.PNG
ZrC#7_backup.PNG
Attachment 2: ZrC_specification_(backup#1_WO#41177-2).xlsx
  52   Wednesday, June 03, 2015, 10:55 Laura LambertZrC#7LP-SISITWTM241177-1 ReadyZrC#7 target description

 Wednesday, June 03, 2015, 10:53

Total length packed = 12.3cm

See attachment for target description.


 

Attachment 1: target_production_elog_ZrC#7.JPG
target_production_elog_ZrC#7.JPG
Attachment 2: ZrC#7_final_density.xlsx
  53   Monday, July 06, 2015, 12:18 John WongZrC#7LP-SISITWTM241177-1 FailedZrC#7 target description

Laura Lambert wrote:

 Wednesday, June 03, 2015, 10:53

Total length packed = 12.3cm

See attachment for target description.


 

 Monday, July 06, 2015, 12:10

This target did not pass the test-stand and will not be used for online (unless if it gets approved).  The target container is deformed.  An increase of voltage was observed during the test; perhaps it's a sign of the container being cracked?!  See Test-stand e-log#450 for more details.  

Another target is already prepared and being tested.  (Target Oven WO# 41177-2). 


 

  59   Monday, October 26, 2015, 16:33 Laura LambertUCx#14LP-SIS  41077-5 ReadyUCx#14 Target Description

 Monday, October 26, 2015, 16:23

Attached please find the UCx#14 target description.

Note: This is the first UCx target using the new LP-SIS small center block design.


 

 

 

 

Attachment 1: Capture.JPG
Capture.JPG
  60   Friday, March 04, 2016, 14:57 Laura LambertUCx#15LP-SISITW 41077-1 ReadyUCx#15 Target Description

 Friday, March 04, 2016, 14:56

Attached please find the UCx#15 target description.

Note: This is the second UCx target using the new LP-SIS small center block design.


 

Attachment 1: UCx_15_description.jpg
UCx_15_description.jpg
  63   Wednesday, June 08, 2016, 15:55 Laura LambertUCx#16LP-IG-LIS  40860-2 ReadyUCx#16 Target Description

 Wednesday, June 08, 2016, 15:52

Attached please find the UCx#16 target description.


 

Attachment 1: Target_Description.JPG
Target_Description.JPG
  55   Friday, July 10, 2015, 11:23 John WongUCx#13LP-IG-LISITETM440860-1 ReadyUCx#13 Target description

 Friday, July 10, 2015, 11:15

Attached please find the UCx#13 target description.   

Note:  Target container was coated/sintered with TaC twice.  Would like to see if this will improve the life-span of the target.  


 

Attachment 1: UCx_13_target_description.PNG
UCx_13_target_description.PNG
  48   Thursday, November 06, 2014, 13:35 Laura LambertUCx # 11LP-SISITETM440282-5 ReadyUCx#11 Target Description

 Thursday, November 06, 2014, 13:29

See attachment for target description


 

Attachment 1: UCx_11_target_description.PNG
UCx_11_target_description.PNG
  45   Monday, July 28, 2014, 15:57 John WongZrC#6LP-SISITWTM140282-3 SpentZrC target description

 Monday, July 28, 2014, 15:52

See attachment#1 for ZrC target description. 

Target is currently undergoing the sintering process


 

Attachment 1: ZrC#6.JPG
ZrC#6.JPG
  43   Tuesday, July 15, 2014, 10:54 Laura LambertTa # 43LP-SISITWTM140282-2 ReadyTa # 43 Target Description

 Tuesday, July 15, 2014, 9:45

Ta#42 LP-SIS target description:

Total Ta foils = 400

Foil Thickness = 0.025mm

Target density = 16.69 g/cm3

Target length = 4.9cm

Total Mass (400 foils + Ta wire) = 45.920g

 

 

 

 

 

 

  39   Monday, March 17, 2014, 16:15 John WongTa#42LP-SISITWTM440282-139956-1SpentTa#42 Target description

Monday, March 17, 2014, 16:14

Ta#42 LP-SIS target description:

Total Ta foils = 400

Foil Thickness = 0.025mm

Target density = 16.69 g/cm3

Target length = 4.2cm

Total Mass (400 foils + Ta wire) = 45.899g


 

  67   Friday, September 02, 2016, 14:45 John WongSiC#34LP-IG-LIS  39814-V2-1 ReadySiC#34 Target Description

 Friday, September 02, 2016, 14:43

See attachment for target description 


 

Attachment 1: SiC34_target_description.JPG
SiC34_target_description.JPG
  38   Wednesday, November 27, 2013, 17:56 John WongUCx#8LP-IG-LISITETM439814-IV SpentUCx#8 target description

Wednesday, November 27, 2013, 17:49

UCx#8 target description - see attachment#1


 

Attachment 1: UCx_8_specification.PNG
UCx_8_specification.PNG
  68   Friday, October 07, 2016, 16:27 John WongSiC#35HP-FEBIAD  39530-9 ReadySiC35 Target description

Friday, October 07, 2016, 16:28

See attachment for the target description


 

Attachment 1: SiC35_target_description.JPG
SiC35_target_description.JPG
  36   Friday, August 30, 2013, 14:45 John WongNiO#2HP-FEBIAD  39530-639069-4SpentNiO target description

Friday, August 30, 2013, 14:38 

NiO/Ni#2                                             Average

NiO+ Ni (g)                                          0.114

NiO - Ni foil (g)                                    0.066

NiO + Ni thickness (mm)                  0.116

NiO - Ni thickness (mm)                   0.104

Area (cm2)                                           2.212

NiO + Ni - organics (8.95%) (g)       0.104

NiO - Ni - organics (8.95%) (g)        0.060

NiO Density (g/cm3)                          2.110

% Yield density                                   32%

# discs in target                                  530

Total target length (cm)                      9.0


 

  57   Thursday, September 24, 2015, 13:42 Laura LambertSiC#32HP-FEBIAD  39530-5 ReadySiC#32 target description

 Thursday, September 24, 2015, 11:46

See attachment for target description.


 

Attachment 1: elog_snapshot.JPG
elog_snapshot.JPG
  80   Wednesday, September 27, 2017, 09:14 Laura LambertSiC#37HP-FEBIADITETM439530-10 ReadySiC#37 target description

 Wednesday, September 27, 2017, 09:13

See attachment for target description.


 

Attachment 1: Target_Description.JPG
Target_Description.JPG
  42   Friday, June 06, 2014, 14:09 John WongUCx#9LP-SIS  39165-9 ReadyUCx#9 - Target description

 Friday, June 06, 2014, 14:08

See attachment for the target description


 

Attachment 1: UCx#9_target_description_1.JPG
UCx#9_target_description_1.JPG
  32   Thursday, April 11, 2013, 14:30 John WongTa#39LP-SISITWTM139165-7 SpentTa#39 target description

Ta#38 LP-SIS target description:

Total Ta foils = 400

Foil Thickness = 0.025mm

Target density = 16.69 g/cm3

Target length = 4.1cm

Total Mass (400 foils + Ta wire) = 43.15g

Disc area = 2.12 cm2

Ta thickness = 20.354g/cm2

See link for more details:

http://documents.triumf.ca/docushare/dsweb/View/Collection-9934


 

  28   Wednesday, September 12, 2012, 01:53 John WongNb#7LP-SIS  39165-6 SpentNb#7 target description

 Wednesday, September 12, 2012, 02:10

Nb#7 target description: 

Total Nb foils = 400 discs

Total Nb mass (Nb foils + Ta wires) = 26.529g 

Total length (after filled) = 4.6cm 

(see attachment #1) 

Attachment 1: Nb#7_target_description.jpg
Nb#7_target_description.jpg
  33   Thursday, May 30, 2013, 16:58 John WongUCx#6LP-SIS TM139165-639230-8SpentUCx#6 - target description

 Thursday, May 30, 2013, 16:57

UCx#6 target specification, see attachment #1. 

The target is currently undergoing the final conditioning in the Evaporator#1.


 Wednesday, June 12, 2013, 17:10

Target is assembled and stored in the Ar glove box.

It's ready to go into the Hotcell.


 

Attachment 1: UCx#6_specifications.PNG
UCx#6_specifications.PNG
  27   Wednesday, August 29, 2012, 14:52 John WongUCx#5LP-SISITWTM439165-5 FailedUCx#5 target description

Wednesday, August 29, 2012, 14:50

UCx#5 target description, see attachment#1

The target is currently undergoing the final conditioning in the Evaporator #1. 


 Wednesday, December 05, 2012, 15:46

This target was broken right before the end of the final run.  The target was inspected today inside the Hotcell and the pictures can be found here: http://documents.triumf.ca/docushare/dsweb/View/Collection-9357

Looks like the protective layer (TaC coating) was worn out over the course of the run time. 

We need to improve the target coating so this can be prevented in the future.

More info about this later...


 

Attachment 1: UCx#5_specifications.PNG
UCx#5_specifications.PNG
  22   Friday, June 01, 2012, 11:43 John WongTa#38LP-SISITWTM139165-238633-2SpentTa#38 target description

Ta#38 LP-SIS target description:

Total Ta foils = 400

Foil thickness = 0.025mm

Target density = 16.69 g/cm3

Target area = 2.12cm2

Target length = 4.3cm

Total Mass (400 foils + Ta wire) = 42.8013g

See the link for more details: http://documents.triumf.ca/docushare/dsweb/View/Collection-8541


Friday, June 01, 2012, 11:36

The target is now in the test-stand, and by next week Tuesday, it should go in the Hotcell.

 


 Monday, June 04, 2012, 16:56

The target was tested in the Test-stand.  The EE was inspected and a small amount of deposit was discovered after the test (the same stuff was found on Ta#36),  see attachment #1.  The samples were collected and will be analyzed through SEM/EDX/XRD.

The EE and the common plate were cleaned with methanol and were assembled back.

The target is ready to go in the Hotcell tomorrow, June 05.


 Monday, November 05, 2012, 16:42

The deposits on the EE and CC were analyzed using XRD

On the EE, it was mainly copper hydroxide nitrate, Cu2(OH)3(NO3). (see attachment #2)

On the CE, it was mainly tantalum, Ta. (see attachment #3)


 

Attachment 1: Ta_38_EE_deposit.JPG
Ta_38_EE_deposit.JPG
Attachment 2: Ta#38_EE.PNG
Ta#38_EE.PNG
Attachment 3: Ta#38_CE.PNG
Ta#38_CE.PNG
  44   Tuesday, July 15, 2014, 11:01 Laura LambertUCx # 10LP-SISITETM439165-10 ReadyUCx#10 Target Description

Tuesday, July 15, 2014, 10:58

See attachment for target description 

Attachment 1: UCx_10_target_description.PNG
UCx_10_target_description.PNG
  47   Thursday, September 25, 2014, 16:01 Laura LambertTa#44HP-SISITETM438817-940316-11ReadyTa#44 target description

Thursday, September 25, 2014, 15:59

Ta#44 HP-SIS target description:

Total Ta foils = 400

Foil Thickness = 0.025mm

Target density = 16.69 g/cm3

Target length = 4.5cm

Total Mass (400 foils + Ta wire) = 47.49g


 

  40   Monday, March 31, 2014, 15:02 John WongTiC#4HP-SIS  38817-839956-10SpentTarget Description

 Monday, March 31, 2014, 15:00

Target consists of 360 discs of TiC + graphite foils. 

Total length = 11.7cm 

Total green mass = 40.945g

(see attachment for more details) 


 

 

Attachment 1: TiC#4_description.JPG
TiC#4_description.JPG
  51   Wednesday, April 29, 2015, 15:39 Laura LambertSiC#31HP-SIS  38817-7 ReadySiC#31 Target description

 Wednesday, April 29, 2015, 15:30

SiC#31 HP-SIS target description: 

Total SiC foils = 520 discs 

Total mass SiC + C_grv foils = 60.2441 g (before sintering, expecting ~13% loss of organic mass after burnt off) 

Average mass SiC/disc (without C_grv) = 0.0853g

Average thickness SiC (without C_grv) = 0.2093mm 

Total packed length = 17cm

See attachment #1 for the density measurement.  


 

Attachment 1: final_density_measurements.JPG
final_density_measurements.JPG
Attachment 2: density_measurements.xlsx
Attachment 3: SiC_density_Jan_10_'14_cast_-_Copy.xlsx
Attachment 4: SiC_density_Oct_31_'14_cast.xlsx
  49   Wednesday, April 01, 2015, 11:27 Laura LambertTa # 45HP-SIS  38817-5 In UseTa#45 target description

 Wednesday, April 01, 2015, 11:25

Ta#45 HP-SIS target description:

Total Ta foils = 400

Foil Thickness = 0.025mm

Target density = 13.48 g/cm3

Target length = 3.4cm

Total Mass (400 foils + Ta wire) = 45.8181g


 
  21   Thursday, May 24, 2012, 11:04 John WongSiC#27HP-SISITETM438817-339069-2SpentHP-SIS SiC27 target description

 Thursday, May 24, 2012, 10:34

SiC#27- HPSIS target description: 

Total SiC foils = 240 discs

Total SiC+ C-grv = 27.79g

Total length = 8.3cm 

Average mass of SiC/disc (without C-grv) = 0.085g 

Average thickness of SiC (without C-grv) = 0.24mm  

No Rh foil added

Salt added = 10uL Li, 10uL Na 

see attachment#1: SiC#27_density measurements

see link for more details: http://documents.triumf.ca/docushare/dsweb/View/Collection-8530


 

Attachment 1: SiC#27_specification.png
SiC#27_specification.png
  11   Wednesday, July 06, 2011, 14:29 John WongNb/SpareHP-SIS  38817-2 In progressTarget ready for Assembly

Target oven w.o# 38817-2:

inner diameter: left = 18.29mm, right = 18.16mm.

Filled with 400 foils Nb (16pk, 5 foils/pk).  

Discs spec: 18.65mm/14.65mm, thickness = 0.022mm  

Total mass + Ta-wire = 22.66g

total length = 4.5cm

Rn foil inserted for ionization.  (0.1mm, 0.004", 1cm x 1cm)  

 

 

  37   Friday, October 25, 2013, 17:35 John WongTa#41HP-SISITWTM438817-10 SpentTa#41 target description

Friday, October 25, 2013, 17:35

Ta#41 HP-SIS target description:

Total Ta foils = 400

Foil Thickness = 0.025mm

Target density = 16.69g/cm3

Target length = 4.0cm

Total Mass (400 foils + Ta wire) = 42.19g


 

  8   Thursday, June 16, 2011, 17:26 John WongNbHP-SIS  38817-1 SpentTarget is ready and stored in cabinet

 Thu Jun 16 17:21:10 2011

Two targets of Nb-HP-SIS are complete and ready to be installed in the heat shields:

-----------------------------------------------------------------------------------

Target oven w.o# 38817-1:

inner diameter: left = 18.27mm, right = 18.25mm.

Filled with 400 foils Nb (16pk, 5 foils/pk).  

Discs spec: 18.65mm/14.65mm, thickness = 0.022mm  

Total mass + Ta-wire = 22.74g

Total length = 4.5cm

Rn foil inserted for ionization.  (0.1mm, 0.004", 1cm x 1cm)  

 

------------------------------------------------------------------------------------

Wednesday, July 06, 2011, 14:35: Salts were added and tested in test-stand last 2 weeks.

 

  25   Friday, July 20, 2012, 11:39 John WongUCx#4LP-SISITWTM138634-936833-4SpentUCx#4 target description

Friday, July 20, 2012, 11:35

UCx#4 target description, see attachment#1

The target is assembled and stored in the Ar-glove box.  It will be transported to the South Hotcell on July 31, 2012.


 

Attachment 1: UCx#4_specfications.PNG
UCx#4_specfications.PNG
  13   Tuesday, August 09, 2011, 14:31 John WongUCx#2LP-SISITWTM138634-838633-5SpentUCx#2 run is finished.

UCx#2 Spec:

Target wrapped with 3 Ta-heat shields

Target length = 11.2cm

Total UC2/C foils = 245 discs

Total U/cm2 = 15.27g

Standard salts = 10uL Na, 10uL Li


Tuesday, August 09, 2011, 14:27: 

Target came out from the glovebox and transported to the cold lab for assembly.  During the assembly, one of the screw-heads for the heat-shield broke (see attachment#1: broken_screwhead). 

The target was leak checked and fine. 

Target is going back to the glovebox to be kept under Ar until RH is ready (see attachment #2: UCx#2).


 Tuesday, October 04, 2011, 10:21

UCx#2 run was finished last week.  

Target was moved from ITW to south hot cell, and stored in a bucket under argon feed. 


 

Attachment 1: broke_screwhead.jpg
broke_screwhead.jpg
Attachment 2: UCx#2.jpg
UCx#2.jpg
  35   Friday, August 30, 2013, 12:08 John WongUCx#7LP-SISITWTM138634-739230-4SpentUCx#7 - target description

Friday, August 30, 2013, 12:06

UCx#7 target specification, see attachment #1.

The target is assembled and going online soon.


Attachment 1: UCx#7_specifications.PNG
UCx#7_specifications.PNG
  14   Tuesday, September 06, 2011, 13:19 John WongTaC#1LP-SISITETM138634-6 SpentTarget is done

Tuesday, September 06, 2011, 13:14: 

Average mass = 0.206g

Average Thickness = 0.0189 cm

Area of D-shaped disc = 2.21 cm2

D-shape discs diameter = 18.5mm (14.5mm cut-out length)

Oven ID, L= 18.22mm, R = 18.20mm 

Density = 4.9266 g/cm3

% Yield Density = 47.20%

Attachment#1 - TaC density measurements.

 Loaded target with 125 discs --> 9.6g Ta/cm2

Total discs mass = 30.41g

Total length = 4.5 cm


 Thursday, September 15, 2011, 10:45

Na & Li were added.

Re foil inserted.

No Ta-heat shields required.


Friday, September 30, 2011, 14:10

After the test-stand, 10uL of Na and Li were added yesterday.   

Target is now ready, and will be moved into the hotcell next week. 


Wednesday, November 16, 2011, 15:51 

After the target was finished and transferred into the hotcell, it was found that the target broke apart from the center block.  This might have happened during the end of the target cooldown. See link for more details: 

http://documents.triumf.ca/docushare/dsweb/View/Collection-7438


 

 

Attachment 1: TaC_density_measurements.xls
  16   Tuesday, September 27, 2011, 10:36 Anders MjosTaC#1LP-SISITETM138634-636833-6Problems/SolutionsMaterial deposited on Extraction Electrode after Test Stand

John Wong wrote:

Tuesday, September 06, 2011, 13:14: 

Average mass = 0.206g

Average Thickness = 0.0189 cm

Area of D-shaped disc = 2.21 cm2

D-shape discs diameter = 18.5mm (14.5mm cut-out length)

Oven ID, L= 18.22mm, R = 18.20mm 

Density = 4.9266 g/cm3

% Yield Density = 47.20%

Attachment#1 - TaC density measurements.

 Loaded target with 125 discs --> 9.6g Ta/cm2

Total discs mass = 30.41g

Total length = 4.5 cm


 Thursday, September 15, 2011, 10:45

Na & Li were added.

Re foil inserted.

No Ta-heat shields required.

 


Tuesday, September 27, 2011, 10:29

 After the test stand, deposited material was found on the Extraction Electrode. A sample of the material was taken for later analysis. The electrode was cleaned with methanol. See attachment for photos.

Attachment 1: Deposited_material.pdf
Deposited_material.pdf
  7   Friday, June 03, 2011, 09:16 John WongTa#37LP-SISITWTM138634-536833-7SpentTarget is done

 Fri Jun 3 12:40:30 2011

Ta#37-LP-SIS (w.o#38634-5)

Filled with 400 Ta foils (25/pk), 4.0 cm in length.  Total mass + Ta wire = 42.32g.  The heat shield was assembled and tested, and the standard salts were added.  Target was installed in the test stand on Jun 1, 2011.  

Fri Jun 17 09:19:46 2011: The following salts have been added: Li, Na, Rb @ 10uL, K @ 5uL, Cs @ 15uL, and Be @ 10uL.

Wednesday, July 06, 2011, 14:08:  TM1 is still under repairing.  It was tested at the test-stand. Target is on the shelve and ready to be used.

Tuesday, July 26, 2011, 15:58:  Salt added before going into the hotcell: Li,Na,Rb,K,Cs.  Target is ready to be used for TM1 test.

Wednesday, July 27, 2011, 10:48:  It was found that the extraction electrode had some arc marking on it (see attachment: EE).  Investigating... 

Wednesday, July 27, 2011, 12:51: After further investigation on the extraction electrode,  we found some black and blue(or green) deposit on the outside surface and inside of the EE.  This deposit is most likely copper oxide, which formed during the HV test.  The EE was removed, and wiped clean with methanol.  (see attachment: EE_CuO).  This target will be used in TM1. 

 

Friday, August 26, 2011, 09:48:  As per Friday Aug 26 TIS meeting, bNMR is very satisfied with the beam; the count rate is very stable.

 

Attachment 1: EE.JPG
EE.JPG
Attachment 2: EE-CuO.jpg
EE-CuO.jpg
  19   Monday, February 13, 2012, 16:00 John WongNb#6LP-SISITWTM138634-436833-1SpentNb#5 ready in the Conditioning Station

 


Monday, February 13, 2012, 16:00 

Nb#5 target description:

Total Nb foils = 400

Total Mass (400 foils + Ta wire) = 24.52g

Total Length = 4.0cm

Target thickness = 11.57 g/cm2

Target Density = 8.57g/cm3

Inserted Rh foil (1cmx1cm)

Added standard salts before test-stand:

15ml Cs

10ml Li, Na, Rb  

5ml K 


Monday, April 23, 2012, 08:58

Target was tested in the test stand, see https://elog.triumf.ca/TIS/Test-Stand/77 for more details.

No salts were added after the test.  The target is now in the conditioning station. 

 


 

  23   Monday, June 11, 2012, 13:46 John WongNb#6LP-SISITWTM138634-436833-1SpentNb#5 run finished

John Wong wrote:

 


Monday, February 13, 2012, 16:00 

Nb#5 target description:

Total Nb foils = 400

Total Mass (400 foils + Ta wire) = 24.52g

Total Length = 4.0cm

Target thickness = 11.57 g/cm2

Target Density = 8.57g/cm3

Inserted Rh foil (1cmx1cm)

Added standard salts before test-stand:

15ml Cs

10ml Li, Na, Rb  

5ml K 


Monday, April 23, 2012, 08:58

Target was tested in the test stand, see https://elog.triumf.ca/TIS/Test-Stand/77 for more details.

No salts were added after the test.  The target is now in the conditioning station. 

 


 

 Monday, June 11, 2012, 13:42

The Nb#5 run was good.  The target is now in the storage. 

see https://elog.triumf.ca/TIS/RH-ISAC/124


 

  12   Wednesday, July 06, 2011, 14:32 John WongTa#36LP-SISITETM438634-238307-10SpentTarget ready, will go online on July 27

Ta-LP-SIS (w.o#38634-2)

Target consists of 400 Ta foils, 4.4 cm in length.  Total mass + Ta wire = 40.87g.

Thursday, July 07, 2011, 15:43:  Begin target assembly. 

Friday, July 08, 2011, 14:24: assembly completed.  Std stds added for test-stand: Li, Na, Rb, K, Cs.

Friday, July 15, 2011, 09:09:  Target currently in test-stand.  Will come out on Monday next week.  Need to add std salts + Be. 

Monday, July 18, 2011, 10:58:  Target removed from test-stand.  Std salts added (Li, Na, Rb, K, Cs) & Be (as requested by Jens).  Maico D.V will do a final inspection on the target before it goes into the hotcell

 

  18   Wednesday, November 16, 2011, 14:08 John WongUCx#3LP-SISITWTM138634-1036833-3SpentUCx#3 ready to go online

 Wednesday, November 16, 2011, 13:51

UCx#3 descriptions: 

Target oven wrapped with 3-Ta heat shields, including the back block (see attachment#1)

Total UCx/C discs = 365

Total disc length = 12.9cm 

UC2/C (365 foils) = 44.76 g

UC2/C (365 foils) - organics = 39.24g

U (365 foils) = 34.05g

U (365 foils) thickness = 16.03 g  

Salts added:  Na, Li, K, Rb, & Ce. 

Attachment 1: UCx#3_assembly.jpg
UCx#3_assembly.jpg
  20   Monday, February 20, 2012, 16:39 John WongUO2/Nb #1LP-FEBIADITETM438481-9 Spenttarget ready for assembly

 Monday, February 20, 2012, 16:25

Mass of empty target oven used = 461.55g

Target oven filled with 402 discs of UO2/Nb, 5.0cm in length, and tot mass (before sintered) was 61.49g (UO2+Nb+org). 

Target finished the sintering process on Feb 20, 2011.  

U(402 foils) mass = 35.689g

U(402 foils) thickness = 16.834 g/cm2

Target is stored in the Hotlab and ready for assembly. 


 

  24   Wednesday, July 04, 2012, 11:32 Anders MjosUO2/Nb #1LP-FEBIADITETM438481-939069-3Spenttarget assembled

John Wong wrote:

 Monday, February 20, 2012, 16:25

Mass of empty target oven used = 461.55g

Target oven filled with 402 discs of UO2/Nb, 5.0cm in length, and tot mass (before sintered) was 61.49g (UO2+Nb+org). 

Target finished the sintering process on Feb 20, 2011.  

U(402 foils) mass = 35.689g

U(402 foils) thickness = 16.834 g/cm2

Target is stored in the Hotlab and ready for assembly. 


 

Target assembled with new anode end cap.

See attachment for gap size and DocuShare for further information.

Attachment 1: UO2#3_Anode_End_Cap.xlsx
  41   Wednesday, April 30, 2014, 16:36 John WongSiC#30HP-FEBIADITETM338481-8 SpentSiC#30 Target description

Wednesday, April 30, 2014, 15:16

Note:

This target was built in 2011.  The endcaps were replaced after they got vaporized in Evap#1 during the sintering process, and the target was re-conditioned.

See elogs for more details:

https://elog.triumf.ca/TIS/Evaporator_1/20

The target was inspected and was good for online use.


Total SiC foils = 555 discs

Total length (packed) = 17.5cm

Total SiC mass (plus graphite foils) = 52.902g

See attachment for density measurements


 

 

 

Attachment 1: SiC#30_density_measurements.xlsx
  29   Thursday, October 11, 2012, 11:41 John WongNiO#1HP-FEBIADITETM438481-739069-1/10(lid)SpentNiO target description

(Should have posted this before my long vacation... Anyway, it's for documentation)

Thursday, October 11, 2012, 11:38

 

NiO/Ni #1 Average
NiO +  Ni  (g) 0.109
NiO - Ni foil (g) 0.06
NiO + Ni  thickness (mm) 0.116
NiO - Ni Thickness (mm) 0.091
Area (cm2) 2.212
NiO + Ni - organics 8.95% (g) 0.099
NiO - Ni - Organics (g) 0.055
NiO Density (g/cm3) 2.145 (32%)

 

Target contains 500 NiO/Ni discs and is measured at 9.0 in length.

  34   Wednesday, July 10, 2013, 18:41 John WongSiC#29HP-FEBIAD-CTLITETM438481-639956-4SpentSiC#29 - Target description

Wednesday, July 10, 2013, 18:32 

SiC#29 HP-FEBIAD-CTL target description: 

Total SiC foils = 510 discs 

Total mass SiC + C_grv foils = 62 g (before sintering, expecting ~13% loss of organic mass after burnt off) 

Average mass SiC/disc (without C_grv) = 0.0892 g

Average thickness SiC (without C_grv) = 0.21mm 

No Rh foil and salt added for this run. 

See attachment #1 for the density measurement.  


 Wednesday, July 10, 2013, 18:40

Target is currently in the Test-Stand.  Once it's tested; it is ready to go online.  

Happy Beaming! 


 

Attachment 1: SiC#29_density_measurements_final.xlsx
  3   Tuesday, March 29, 2011, 14:14 John WongTa#35HP-SISITETM438481-338303-4SpentITE-TM4-Ta#35-Hp-SIS

ITE-TM4-Ta#35-Hp-SIS

Thu Apr 7 15:32:32 2011:

Target oven - w.o#38481-3.

Packs of Ta foils were previously prepared by Vicky H.

Marik loaded the target:  400 of 0.025mm foils (25 foils/pk) in 3.5cm length.  (Recommended length 4.0-4.5cm; If too closely packed, wires get squeezed and snapped)

Average mass per discs = 0.105g

Total mass (Ta+wire) = 45.06g

Disc area = 2.12cm2

Ta thickness = 21.25g/cm2

Thu Apr 28 :

Target heat Shield (w.o 38303-4) cleaned

Target is now assembled by John and Rick

Megger Test Performed:

TA-->HS = 540V, OL

TA-->EE = 540V, OL 

HS--> EE = 1049V, OL

Leak Check OK

*SIS restraint used in this target ion source

Mon May 9 10:32:40 2011:  Standard salts added: Li, Na, Rb @ 10uL, K @ 5uL (but added 10uL), Cs @ 15uL.  Target will be going to the test stand.

Fri May 20 16:57:22 2011: 

Target was tested on the Test Stand but problems with the software (see Test stand elog). 

Standard salts added: Li, Na, Rb @ 10uL, K @ 5uL (but added 10uL), Cs @ 15uL, and Be @ 10uL.

Monday, July 18, 2011, 11:15:  Target ended on July 16.  The target was used online for an extra 2 weeks and performed very well. 

 

 

Attachment 1: Ta-HPSIS.jpg
Ta-HPSIS.jpg
  4   Thursday, March 31, 2011, 15:38 John WongSiC#25HP-SISITETM438481-138307-?SpentITE-TM4-SiC#25-Hp-SIS

Thu Mar 31 15:19:51 2011:  SiC#25 was made sometime in December of 2010.  It is now in the target module and ready to be conditioned. 

Mon Apr 18 18:17:05 2011: It was found later that no salt was added to this target after it came out from the test stand.  However,  there were enough 40Ca and 39K  for tuning (See attachment#1 for the big mass scan).  Once the P+ is back, the yields will be measured.

Fri Apr 29 15:30:26 2011:  Low 27Al was ionized from the surface for some reason.  However, after the green laser was shone into the extraction tube, more 27Al was produced.  Then, the laser was turned off, the target continued to surface ionizing 27Al.  There could be a blockage from the rhenium (Re) ionizer foil.  Jens L. will investigate this further and do a test with the lasers and the rhenium foil.

 

 


Attachment 1: big_mass_scan.jpg
big_mass_scan.jpg
  15   Friday, September 23, 2011, 17:03 Anders MjosSiC#26HP-FEBIADITETM438295-938307-8SpentRun finished

Friday, September 23, 2011, 16:54:

  • Cu-gaskets used for leak checking target had a leak rate of about  10-8  10-9 atm cc/sec. Replaced with silver plated SS gaskets and no leak detected above 10-11 atm cc/sec.
  • The target can not hold high voltage when it is sitting on the jig. The insulator on the jig have reduced resistance over time (less than 1 G Ohm). All high voltage tests passed when target is tested not sitting on the jig.

 

  • Target is now in the test stand and pumping down.
  17   Monday, October 17, 2011, 14:45 John WongSiC#26HP-FEBIADITETM438295-938307-8SpentRun finished

Anders Mjos wrote:

Friday, September 23, 2011, 16:54:

  • Cu-gaskets used for leak checking target had a leak rate of about  10-8  10-9 atm cc/sec. Replaced with silver plated SS gaskets and no leak detected above 10-11 atm cc/sec.
  • The target can not hold high voltage when it is sitting on the jig. The insulator on the jig have reduced resistance over time (less than 1 G Ohm). All high voltage tests passed when target is tested not sitting on the jig.

 

  • Target is now in the test stand and pumping down.

 Monday, October 17, 2011, 14:41

After the test-stand, the extraction electrode showed some electric arcing mark, but no solid deposits were found.  (see attachment#1: EE_after_teststand.jpg)

See attachment for the target specification.  

Attachment 1: EE_after_teststand.jpg
EE_after_teststand.jpg
Attachment 2: SiC#26_specification.png
SiC#26_specification.png
  10   Wednesday, July 06, 2011, 14:13 John WongTa#40HP-SIS  38295-4 SpentTarget description

Details about Ta Hp surface, w.o#38295-4 (spare):

This target was built as a spare.

 

Note: Left side target arm replaced, Nov 18, 2010.  Weld prep on centre block misaligned.  (Marik has more details about this).

Packs of Ta foils were prepared by JW. 

Target was filled with 400 of 0.025mm foils (25 foils/pk) in 4.0cm length. 

Total Mass of 16 packs of Ta foils + Ta wires = 42.70g.  

See attachment#1 for the measurement.

 


Friday, April 19, 2013, 09:44 

Target was tested at the Teststand.  Result Good. 

Ready for use.


 

  121   Tuesday, August 16, 2022, 09:48 Darwin Ortiz RosalesSiC#44HP-FEBIAD  37978-1 ReadyTarget SiC#44 Description

Tuesday, August 16, 2022, 09:47

  105   Tuesday, March 16, 2021, 17:05 Luca EgoritiITE-TM4-UC#32-P2N-SISHP-SISITETM4--In progressUO2 discs cutting

This afternoon, Darwin prepared in the hot lab fume hood the two UO2 slurries to be cut into P2N target discs. Later, I took over cutting both slurries. A total of 49 annular discs were cut and loaded into the P2N graphite container.

The discs thickness was measured by stacking 5 discs on each other, and measuring their thickness in two points.

#discs = 49

Average disc thickness = 0.453mm

Empty cartridge weight = 9.42 g

Full cartridge weight = 39.86 g

UO2+C weight = 30.44 g

The unused material plus the cutting scrap were left in a zip lock bag on a shelf in the right hand side of the fume hood. The center scraps and the loaded target were placed in a plastic tray on the top shelf. The used tools and scales were left on the right of the working area, the napkins on the bottom were put into the garbage bag (inside the fume hood) and replaced, while the only napkin left is the one with the target information (recorded in attached picture).

Attachment 1: UC32-P2N_material_info.jpg
UC32-P2N_material_info.jpg
Attachment 2: Fume_hood.jpg
Fume_hood.jpg
  9   Friday, June 17, 2011, 11:58 John WongUCx (flakes)LP-SIS    Problems/SolutionsMaking UC2 flakes (stock# 2011_A & _B)

Fri Jun 17 10:28:47 2011

First batch of UO2/C mixture of carbonization (Stock# 2011_A):   

UO2 used = 50.120g (left overs + container = 411.23g)

C(grv) used = 10.653g (or 21.25% w/w)

Mass of organics = 4.842g (or 9.66% w/w) 

Mass of dried cast = 47.420g 

Waste collected in jar = 18.195g (including UO2, C and Org) 

Waste accountable for UO2 = 12.57g

The carbonization process started from May 26 to June 4 (9 days).  

See attachment #1 and #2 for the striptools during the UO2/C carbonization

Total Mass UC2 flakes = 35.310g 


 Fri Jun 17 11:56:15 2011

Attempted to make the second batch of UO2/C mixture of carbonization on June 13, but it failed.  The milling jar gasket seal partially came out during the high speed milling.  As a result, the UO2/C mixture squirted out and contaminated the planetary mill (See attachment #3 & #4).  RPG and all other groups were informed.  The planetary mill and the lab were properly cleaned up (see attachment #5), no activities were found after the cleaning.  Attachments #6, #7, #8 are some of the discussion about the UO2 spill.  

All the UO2 taken out from the container (50.23g) is accountable for waste.  

 


 June 28, 2011 - Re-making 2nd batch of UO2/C mixture was successful.  

Stock#2011_B:

UO2 used = 50.29g

C(grv) used = 10.0g 

Mass of Organics = 4.84g 

Mass of Dried green = 71.96g (the cast was let dry over night <24hrs due to time restriction.  It might not be completely dried as a result of such heavy mass remained.  However, the UO2 waste was very minimum, <5.0g) 

Total Mass UC2 flakes = 56.853g 

 


 

 

 

 

 

Attachment 1: UO2_C_IG1_CG2_2011_A.jpg
UO2_C_IG1_CG2_2011_A.jpg
Attachment 2: UO2_C_TGHT_IG1_2011_A.jpg
UO2_C_TGHT_IG1_2011_A.jpg
Attachment 3: UO2_mill_spill_01.png
UO2_mill_spill_01.png
Attachment 4: UO2_mill_spill_02.png
UO2_mill_spill_02.png
Attachment 5: UO2_spill_cleanup.png
UO2_spill_cleanup.png
Attachment 6: UO2_spill_discussion_01.png
UO2_spill_discussion_01.png
Attachment 7: UO2_spill_discussion_02.png
UO2_spill_discussion_02.png
Attachment 8: UO2_spill_discussion_03.png
UO2_spill_discussion_03.png
  26   Wednesday, August 01, 2012, 09:37 Pierre BricaultITE-TM4-UO#3LP-FEBIADITETM4  Problems/SolutionsShort has developed while lowering the cathode heater

 On tuesday July we lower the cathode heater to swap the anode-cathode bias. While lowering the cathode heater a short on the anode developed. 

The polarity swap was not possible because the target/tube assembly has also a short to common. 

We return the configuration has it was and we added two 100 Ohms: 5 W resistor between the common and the heaters to prevent current to flow back to the common. 

The anode short disappear during the cathode and target heating. 

  46   Monday, August 18, 2014, 14:39 Laura LambertThO2 #1LP-SIS     ThO2 #1 Target Description

Monday, August 18, 2014, 14:53

See attachment for target description


 

Attachment 1: ThO2#1_target_description.JPG
ThO2#1_target_description.JPG
  61   Friday, March 11, 2016, 10:18 Laura LambertTa # 47 ITETM4  SpentTa # 47 target desscription

Friday, March 11, 2016, 10:13 

Ta#47 LP-SIS [new small center block] target description:

Total Ta foils = 470

Foil Thickness = 0.025mm

Target density = 15.05 g/cm3

Target length = 3.4cm [exactly 7.8cm space on both sides]

Total Mass = 51.1533g [Incl foils + Ta wire on 6 packs]

Note: Target consists of 350 single discs and 6x20 packs of Ta discs


 
  83   Friday, July 13, 2018, 14:41 John WongUCx#23LP-FEBIAD-CTL    ReadyUCx#23 target description

Friday, July 13, 2018, 14:40

Target is ready for installation. 

See attachment for description


 

Attachment 1: UCx_23-target_description.JPG
UCx_23-target_description.JPG
  84   Monday, August 27, 2018, 17:11 John WongUCx#24LP-SIS    ReadyUCx#24 Target Description

Monday, August 27, 2018, 17:09

See attachment for UCx#24 target description


 

Attachment 1: Capture.JPG
Capture.JPG
  86   Friday, October 12, 2018, 17:14 John WongUCx#25LP-SIS    In progressUCx#25 Target Description

Friday, October 12, 2018, 17:11

See attachment for UCx#25 target description.


 

Attachment 1: UCx#25_target_description.JPG
UCx#25_target_description.JPG
  118   Tuesday, June 21, 2022, 15:34 Matthew GareauSiC#42, SiC#43  TM2  Problems/SolutionsMultipin guide pin issues

 While removing the SiC#42 it became apparent that the multipin guide pin had become loose again and was not allowing the target to dismount from the module.

The multipin shoulder bolts were removed and then the multipin was able to be pried off.

However the NHC manipulator were not able to reach the guide pin to remove it, which would be required before the SiC#43 could be installed.

The target module was then moved to the SHC where the guide pin could be removed and the installation of SiC#43 took place, along with the transfer of the parasitic attachment onto the SiC#43.

This problem seems to suggest that the multipin could use some redesign to ensure smoother target exchanges.

  128   Friday, March 10, 2023, 07:21 Aaron Schmidt      ReadyTarget Material Batch - Ta

 Tantalum - Standard foils.

MTR PO: TR-2022-03028

Average foil thickness: 0.0013"

Attachment 1: Ta_2023_-_01.pdf
Ta_2023_-_01.pdf Ta_2023_-_01.pdf