Logbooks
Lab Maintenance
Evaporator_1
Evaporator_2
Laser cutter
Target Production
Test-Stand
RH-ISAC
RH-Cyclotron
RH-Meson Hall
RH-Beamlines
RH-ARIEL
|
Target Production |
Not logged in |
 |
|
Message ID: 112
Entry time: Wednesday, September 29, 2021, 10:37
|
Author: |
Darwin Ortiz Rosales |
Target/No.#: |
UCx#35 |
Source: |
LP-IG-LIS |
Station: |
|
Module: |
|
Target Oven W.O#: |
42218-5 |
Heat Shield W.O#: |
|
Status: |
Ready |
Subject: |
UCx#35-Target_description |
|
|
Target ID |
UCx#35 |
General |
Total No. Discs |
179 |
|
Disc area |
0.95 |
cm2 |
Ave graphite foil thickness/disc |
0.013 |
cm |
Ave. graphite foil mass/disc |
0.015 |
g |
|
Total mass of discs |
14.97 |
g |
Green Target |
Total Green mass of discs |
13.89 |
g |
Ave. mass/disc |
0.084 |
g |
Ave. mass/disc (no C-foil & org) |
0.078 |
g |
Ave. thickness/disc (no C-foil) |
0.037 |
cm |
Theoretical length (with C-foil) |
9.0 |
cm |
Total packed length (with C-foil) |
10.0 |
cm |
Final Target |
UC2 in Target |
11.05 |
g |
Free C in Target |
0.83 |
g |
Total massUC2+free C |
11.88 |
g |
U mass |
10.04 |
g |
U thickness |
10.56 |
g/cm2 |
UC2 bulk density (measured) |
2.88 |
g/cm3 |
Theoretical Density |
8.24 |
g/cm3 |
% Yield Density |
34.91 |
% |
Theoretical Porosity |
0.651 |
|
|
|