Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  Test-Stand, Page 4 of 53  Not logged in ELOG logo
New entries since:Wed Dec 31 16:00:00 1969
ID Date Author Target/No.# Sourcedown Station Module Target Oven W.O# Heat Shield W.O# Status Subject
  512   Thursday, June 09, 2016, 14:56 Aaron Schmidt LP-SIS     TUBE ONLY

Thursday, June 09, 2016, 14:50: Begin roughing.

Thursday, June 09, 2016, 15:34: Turbo pumps started.

Thursday, June 09, 2016, 15:42: Ion gauge on.


Friday, June 10, 2016, 08:27: RGA on.

Friday, June 10, 2016, 08:35: Water opened. RGA ok. Vacuum stable @ 1.9E-7T

Friday, June 10, 2016, 08:50: BIAS/EZL/EE to 30/13/1.5kV.

Friday, June 10, 2016, 08:52: Auto ramping TBHT only @ 2A min to 50A.

Friday, June 10, 2016, 10:02: Auto ramping TBHT only @ 1A min to 100A.

Friday, June 10, 2016, 11:14: Continue ramping to 150 A -AM

Friday, June 10, 2016, 11:28: Stopped ramping ag 110 A -AM


Monday, June 13, 2016, 08:12: Auto ramping TBHT only @ 1A min to 150A.

Monday, June 13, 2016, 09:59: Ramping to 150 A - AM


Monday, June 13, 2016, 15:51: Ramping down.


Tuesday, June 14, 2016, 07:47: IV4 Reset. Ion gauge off. Turbos off.

Tuesday, June 14, 2016, 07:52: Power/sources off. Water purged.

Tuesday, June 14, 2016, 09:20: Venting.

Attachment 1: Tube_only_SIS_test.zip
  514   Tuesday, June 14, 2016, 10:46 Aaron Schmidt LP-SIS     TUBE ONLY - NO PLASMA ELECTRODE

Tuesday, June 14, 2016, 10:46: Begin roughing.

Tuesday, June 14, 2016, 11:16: Turbo pumps started.

Tuesday, June 14, 2016, 11:23: Ion gauge on.


Wednesday, June 15, 2016, 07:45: Water opened. RGA ok. Vacuum stable @ 1.9E-7T. Power/sources on.

Wednesday, June 15, 2016, 08:24: BIAS / EZL / EE to 30 / 13 / 1.5 kV

Wednesday, June 15, 2016, 08:24: Ramping TBHT to 120 A at 1 A/min

Wednesday, June 15, 2016, 10:23: EZL P/S was not on. Switched on and set to 13 kV

Wednesday, June 15, 2016, 15:05: Ramping down.


Thursday, June 16, 2016, 08:57: IV4 reset. RGA off. Ion gauge off. Turbo pumps off.

Thursday, June 16, 2016, 13:22: Water lines purged. Power/sources off. Venting.

Attachment 1: Tube_only_SIS_test-all-160615.zip
Attachment 2: 2016-06-13.zip
Attachment 3: 2016-06-15.zip
Attachment 4: re-em-plots.zip
  515   Thursday, June 16, 2016, 14:34 Aaron Schmidt LP-SIS     TUBE ONLY (3)

Thursday, June 16, 2016, 14:33: Roughing. Plasma Electrode re-attached. Salts replaced.

Thursday, June 16, 2016, 15:12: Turbo pumps started.

Thursday, June 16, 2016, 15:22: Ion gauge on.


Friday, June 17, 2016, 08:32: Water opened. RGA ok. Vacuum stable @ 2.4E-7T. Power/sources on.


Saturday, June 18, 2016, 09:31: Cool down started.


Monday, June 20, 2016, 09:48: Sources/power/voltages off.

Monday, June 20, 2016, 10:06: IV4 reset. Water lines purged.

 Monday, June 20, 2016, 10:11: RGA off. Ion gauge off. Turbos off.

Monday, June 20, 2016, 11:50: Venting

Attachment 1: Tube_only-full-fernando_test-160617-PE-RPE.zip
Attachment 2: Tube_only-full-fernando_test-160617-WPE.zip
  547   Tuesday, October 18, 2016, 13:29 Aaron SchmidtTa#50LP-SISITETM441790-541751-1In progressConditioning Ta#50

Tuesday, October 18, 2016, 13:22: Roughing.

Tuesday, October 18, 2016, 13:46: Turbo pumps started.

Tuesday, October 18, 2016, 13:52: Ion gauge on.

Tuesday, October 18, 2016, 13:54: Both turbo pumps at speed.

Tuesday, October 18, 2016, 15:05: RGA on.


Wednesday, October 19, 2016, 08:46: Power/Sources on

Wednesday, October 19, 2016, 08:49: Waterline opened. No response on RGA. Vacuum stable @4.40E-7T

Wednesday, October 19, 2016, 08:57: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .

Wednesday, October 19, 2016, 08:58 : IV4 opened.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 27.9 3.90
 TBHT 27.7 3.06
 COIL + EE 27.6 5.08
 TGHT  27.7 5.70

Wednesday, October 19, 2016, 09:01: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.

Wednesday, October 19, 2016, 11:50: TMS:FC0 moved out & TMS:RPM0B began scanning continuously on their own. Monitor.

Wednesday, October 19, 2016, 14:45: RGA Restarted.

Wednesday, October 19, 2016, 15:25: RGA Restarted (Unable to save 14:45 Startup)


Friday, October 21, 2016, 13:27: Auto Cooling @ 2A per min

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 37.3 3.92
 TBHT 28.2 3.07
 COIL + EE 27.9 5.07
 TGHT  29.8 5.62

Friday, October 21, 2016, 15:39: IV4 Closed


Monday, October 24, 2016, 09:10:

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 28.1 3.95
 TBHT 28.2 3.10
 COIL + EE 27.7 5.04
 TGHT  27.7 5.70

Monday, October 24, 2016, 09:10: Power/Sources off.

Monday, October 24, 2016, 09:20: Water lines purged.

Monday, October 24, 2016, 09:21: RGA off.

Monday, October 24, 2016, 09:23: Ion gauge & Turbo pumps off

Monday, October 24, 2016, 12:54: Venting.

Attachment 1: ER-580-200-30kV-m27.txt
value (Voltage offset      40.0000)
     80.0000
      0.7000
     80.0000
  29978.0220
21
21
./emit.dat
 -0.1154 -0.0916 -0.0937 -0.0952 -0.0964 -0.0955 -0.0980 -0.1019 -0.0943 -0.0964 -0.0980 -0.0940 -0.0919 -0.0919 -0.0964 -0.0946 -0.0992 -0.0949 -0.1016 -0.0992 -0.1031
 -0.0894 -0.0903 -0.0925 -0.0928 -0.0946 -0.0943 -0.0949 -0.0977 -0.0952 -0.0943 -0.0949 -0.0925 -0.0922 -0.0909 -0.0964 -0.0934 -0.0989 -0.0925 -0.1004 -0.0986 -0.1001
 -0.0909 -0.0925 -0.0928 -0.0943 -0.0949 -0.0946 -0.0946 -0.0952 -0.0943 -0.0967 -0.0961 -0.0925 -0.0934 -0.0925 -0.0937 -0.0916 -0.0916 -0.0922 -0.0970 -0.0983 -0.0928
 -0.0916 -0.0931 -0.0934 -0.0934 -0.0916 -0.0916 -0.0925 -0.0909 -0.0943 -0.0909 -0.0974 -0.0925 -0.0961 -0.0919 -0.0916 -0.0909 -0.0900 -0.0916 -0.0949 -0.0983 -0.0931
 -0.0894 -0.0903 -0.0934 -0.0934 -0.0934 -0.0882 -0.0891 -0.0903 -0.0922 -0.0919 -0.0955 -0.0916 -0.0943 -0.0900 -0.0891 -0.0876 -0.0879 -0.0879 -0.0937 -0.0955 -0.0888
 -0.0888 -0.0891 -0.0885 -0.0916 -0.0943 -0.0919 -0.0497 -0.0320 -0.0818 -0.0912 -0.0928 -0.0912 -0.0912 -0.0891 -0.0894 -0.0916 -0.0919 -0.0894 -0.0946 -0.0970 -0.0903
 -0.0876 -0.0888 -0.0894 -0.0912 -0.0961 -0.0919 -0.0354 0.0482 0.0952 -0.0259 -0.0916 -0.0897 -0.0906 -0.0919 -0.0928 -0.0903 -0.0919 -0.0912 -0.0961 -0.0974 -0.0900
 -0.0912 -0.0903 -0.0882 -0.0882 -0.0928 -0.0912 -0.0403 0.0674 0.2994 0.2567 0.0485 -0.0671 -0.0882 -0.0912 -0.0916 -0.0912 -0.0900 -0.0928 -0.0949 -0.0983 -0.0912
 -0.0922 -0.0928 -0.0912 -0.0876 -0.0848 -0.0797 -0.0470 -0.0299 0.4971 0.5460 0.3802 0.0650 -0.0308 -0.0906 -0.0891 -0.0876 -0.0900 -0.0909 -0.0943 -0.0970 -0.0900
 -0.0903 -0.0903 -0.0903 -0.0912 -0.0800 -0.0711 -0.0632 -0.0327 0.5188 0.7532 0.5743 0.1593 -0.0430 -0.0607 -0.0827 -0.0900 -0.0916 -0.0897 -0.0937 -0.0955 -0.0897
 -0.0891 -0.0900 -0.0916 -0.0900 -0.0919 -0.0723 -0.0430 -0.0201 0.2475 0.6775 0.5890 0.2744 -0.0375 -0.0583 -0.0742 -0.0818 -0.0879 -0.0897 -0.0943 -0.0952 -0.0882
 -0.0897 -0.0909 -0.0906 -0.0922 -0.0937 -0.0922 -0.0824 -0.0095 0.1239 0.4227 0.4150 0.2563 -0.0189 -0.0629 -0.0839 -0.0845 -0.0858 -0.0909 -0.0952 -0.0964 -0.0919
 -0.0925 -0.0912 -0.0925 -0.0919 -0.0949 -0.0912 -0.0897 -0.0885 -0.0565 0.1428 0.2325 0.1718 0.0174 -0.0620 -0.0885 -0.0897 -0.0897 -0.0894 -0.0943 -0.0986 -0.0903
 -0.0931 -0.0909 -0.0922 -0.0946 -0.0952 -0.0906 -0.0885 -0.0897 -0.0906 -0.0806 0.0296 0.0879 0.0458 -0.0549 -0.0876 -0.0894 -0.0900 -0.0894 -0.0922 -0.0955 -0.0906
 -0.0909 -0.0897 -0.0916 -0.0916 -0.0922 -0.0919 -0.0903 -0.0916 -0.0903 -0.0894 -0.0900 -0.0793 -0.0656 -0.0769 -0.0885 -0.0888 -0.0873 -0.0885 -0.0943 -0.0943 -0.0879
 -0.0891 -0.0873 -0.0891 -0.0906 -0.0922 -0.0903 -0.0922 -0.0931 -0.0879 -0.0882 -0.0897 -0.0885 -0.0870 -0.0894 -0.0916 -0.0906 -0.0900 -0.0894 -0.0934 -0.0961 -0.0897
 -0.0891 -0.0900 -0.0909 -0.0922 -0.0949 -0.0919 -0.0916 -0.0940 -0.0894 -0.0903 -0.0912 -0.0897 -0.0894 -0.0925 -0.0912 -0.0894 -0.0903 -0.0912 -0.0958 -0.0964 -0.0894
 -0.0906 -0.0909 -0.0922 -0.0934 -0.0940 -0.0909 -0.0888 -0.0900 -0.0891 -0.0906 -0.0937 -0.0897 -0.0891 -0.0882 -0.0903 -0.0900 -0.0912 -0.0912 -0.0940 -0.0980 -0.0925
 -0.0912 -0.0928 -0.0925 -0.0940 -0.0937 -0.0897 -0.0900 -0.0903 -0.0873 -0.0906 -0.0931 -0.0912 -0.0922 -0.0897 -0.0900 -0.0903 -0.0894 -0.0906 -0.0937 -0.0958 -0.0912
 -0.0888 -0.0891 -0.0909 -0.0919 -0.0937 -0.0912 -0.0906 -0.0879 -0.0854 -0.0876 -0.0912 -0.0900 -0.0906 -0.0888 -0.0894 -0.0897 -0.0882 -0.0885 -0.0912 -0.0940 -0.0903
 -0.0870 -0.0891 -0.0891 -0.0906 -0.0909 -0.0928 -0.0897 -0.0894 -0.0848 -0.0839 -0.0882 -0.0888 -0.0882 -0.0909 -0.0903 -0.0894 -0.0897 -0.0888 -0.0937 -0.0937 -0.0864
Attachment 2: ER-580-200-30kV-m27(2).txt
value (Voltage offset      40.0000)
     80.0000
      0.7000
     80.0000
  29978.0220
21
21
./emit.dat
 -0.1202 -0.0977 -0.0977 -0.0967 -0.0958 -0.1004 -0.0983 -0.0977 -0.1047 -0.0989 -0.0983 -0.0970 -0.0970 -0.0964 -0.0958 -0.0961 -0.0983 -0.0964 -0.0961 -0.1004 -0.0986
 -0.0916 -0.0903 -0.0919 -0.0919 -0.0903 -0.0986 -0.0937 -0.0952 -0.0946 -0.0916 -0.0928 -0.0919 -0.0919 -0.0891 -0.0922 -0.0912 -0.0912 -0.0919 -0.0955 -0.0946 -0.0925
 -0.0912 -0.0928 -0.0903 -0.0922 -0.0916 -0.0958 -0.0937 -0.0900 -0.0909 -0.0943 -0.0919 -0.0909 -0.0894 -0.0928 -0.0912 -0.0943 -0.0888 -0.0891 -0.0928 -0.0946 -0.0909
 -0.0916 -0.0906 -0.0916 -0.0967 -0.0958 -0.0912 -0.0949 -0.0922 -0.0897 -0.0882 -0.0912 -0.0916 -0.0934 -0.0909 -0.0916 -0.0906 -0.0925 -0.0916 -0.0931 -0.0961 -0.0943
 -0.0940 -0.0937 -0.0934 -0.0974 -0.0964 -0.0888 -0.0946 -0.0922 -0.0986 -0.0964 -0.0928 -0.0928 -0.0940 -0.0903 -0.0903 -0.0912 -0.0928 -0.0928 -0.0934 -0.0974 -0.0937
 -0.0925 -0.0925 -0.0931 -0.0970 -0.0958 -0.0833 -0.0595 -0.0632 -0.0943 -0.0922 -0.0940 -0.0934 -0.0943 -0.0937 -0.0922 -0.0940 -0.0925 -0.0909 -0.0952 -0.0946 -0.0922
 -0.0891 -0.0916 -0.0903 -0.0955 -0.0940 -0.0854 -0.0610 -0.0021 -0.0043 -0.0693 -0.0906 -0.0909 -0.0897 -0.0912 -0.0937 -0.0919 -0.0909 -0.0894 -0.0919 -0.0946 -0.0897
 -0.0903 -0.0891 -0.0891 -0.0925 -0.0922 -0.0916 -0.0751 0.0476 0.1105 0.0775 -0.0342 -0.0861 -0.0912 -0.0879 -0.0906 -0.0894 -0.0894 -0.0909 -0.0940 -0.0931 -0.0906
 -0.0934 -0.0912 -0.0909 -0.0916 -0.0912 -0.0836 -0.0720 0.0098 0.2060 0.1773 0.0888 -0.0381 -0.0729 -0.0922 -0.0891 -0.0909 -0.0931 -0.0922 -0.0919 -0.0958 -0.0925
 -0.0937 -0.0916 -0.0931 -0.0949 -0.0922 -0.0827 -0.0793 -0.0565 0.2127 0.2264 0.1419 -0.0339 -0.0781 -0.0821 -0.0912 -0.0937 -0.0943 -0.0925 -0.0946 -0.0952 -0.0934
 -0.0919 -0.0928 -0.0925 -0.0964 -0.0958 -0.0830 -0.0729 -0.0583 0.0943 0.1904 0.1215 -0.0253 -0.0732 -0.0830 -0.0873 -0.0897 -0.0934 -0.0922 -0.0937 -0.0949 -0.0900
 -0.0912 -0.0916 -0.0906 -0.0928 -0.0928 -0.0919 -0.0864 -0.0555 0.0027 0.1202 0.0879 -0.0012 -0.0650 -0.0861 -0.0897 -0.0891 -0.0894 -0.0919 -0.0919 -0.0925 -0.0903
 -0.0897 -0.0897 -0.0916 -0.0943 -0.0946 -0.0885 -0.0906 -0.0885 -0.0732 0.0076 0.0360 -0.0037 -0.0497 -0.0848 -0.0891 -0.0903 -0.0897 -0.0891 -0.0909 -0.0943 -0.0912
 -0.0922 -0.0906 -0.0903 -0.0961 -0.0958 -0.0897 -0.0912 -0.0922 -0.0946 -0.0836 -0.0357 -0.0269 -0.0415 -0.0839 -0.0909 -0.0897 -0.0909 -0.0919 -0.0919 -0.0961 -0.0928
 -0.0919 -0.0916 -0.0925 -0.0983 -0.0986 -0.0903 -0.0934 -0.0940 -0.0967 -0.0934 -0.0909 -0.0861 -0.0775 -0.0894 -0.0922 -0.0934 -0.0925 -0.0909 -0.0928 -0.0955 -0.0925
 -0.0925 -0.0928 -0.0909 -0.0964 -0.0961 -0.0931 -0.0912 -0.0952 -0.0940 -0.0903 -0.0903 -0.0916 -0.0906 -0.0900 -0.0922 -0.0909 -0.0903 -0.0900 -0.0922 -0.0949 -0.0903
 -0.0903 -0.0897 -0.0909 -0.0940 -0.0946 -0.0909 -0.0919 -0.0912 -0.0922 -0.0897 -0.0891 -0.0900 -0.0891 -0.0900 -0.0906 -0.0894 -0.0912 -0.0894 -0.0912 -0.0928 -0.0894
 -0.0906 -0.0888 -0.0891 -0.0934 -0.0928 -0.0897 -0.0888 -0.0906 -0.0940 -0.0882 -0.0900 -0.0900 -0.0909 -0.0885 -0.0894 -0.0885 -0.0903 -0.0903 -0.0897 -0.0946 -0.0906
 -0.0897 -0.0903 -0.0919 -0.0952 -0.0928 -0.0894 -0.0912 -0.0912 -0.0909 -0.0906 -0.0906 -0.0916 -0.0903 -0.0922 -0.0900 -0.0912 -0.0919 -0.0916 -0.0916 -0.0961 -0.0922
 -0.0925 -0.0925 -0.0916 -0.0952 -0.0949 -0.0922 -0.0919 -0.0922 -0.0940 -0.0912 -0.0922 -0.0916 -0.0940 -0.0919 -0.0912 -0.0940 -0.0925 -0.0922 -0.0943 -0.0958 -0.0922
 -0.0916 -0.0916 -0.0919 -0.0943 -0.0952 -0.0919 -0.0922 -0.0909 -0.0900 -0.0888 -0.0906 -0.0928 -0.0906 -0.0916 -0.0912 -0.0916 -0.0900 -0.0906 -0.0925 -0.0937 -0.0912
Attachment 3: ER-580-200-30kV-m27(3).txt
value (Voltage offset      40.0000)
     80.0000
      0.7000
     80.0000
  29978.0220
21
21
./emit.dat
 -0.1181 -0.0970 -0.0977 -0.0970 -0.0992 -0.0970 -0.1059 -0.0967 -0.0974 -0.0970 -0.0983 -0.0986 -0.0961 -0.0961 -0.0970 -0.0989 -0.0986 -0.0967 -0.0989 -0.0970 -0.0964
 -0.0946 -0.0928 -0.0919 -0.0937 -0.0949 -0.0943 -0.1013 -0.0906 -0.0967 -0.0943 -0.0946 -0.0934 -0.0961 -0.0937 -0.0928 -0.0928 -0.0912 -0.0922 -0.0958 -0.0925 -0.0912
 -0.0928 -0.0934 -0.0928 -0.0937 -0.0922 -0.0931 -0.0940 -0.0934 -0.0955 -0.0970 -0.0934 -0.0934 -0.0928 -0.0934 -0.0922 -0.0916 -0.0903 -0.0909 -0.0940 -0.0928 -0.0925
 -0.0916 -0.0909 -0.0912 -0.0922 -0.0943 -0.0888 -0.0946 -0.0955 -0.0906 -0.0891 -0.0916 -0.0912 -0.0909 -0.0888 -0.0934 -0.0919 -0.0925 -0.0925 -0.0952 -0.0937 -0.0931
 -0.0916 -0.0906 -0.0912 -0.0925 -0.0955 -0.0922 -0.0949 -0.0934 -0.0934 -0.0943 -0.0931 -0.0919 -0.0922 -0.0912 -0.0912 -0.0934 -0.0922 -0.0922 -0.0949 -0.0940 -0.0946
 -0.0906 -0.0916 -0.0919 -0.0937 -0.0970 -0.0922 -0.0803 -0.0644 -0.0928 -0.0922 -0.0943 -0.0934 -0.0952 -0.0949 -0.0931 -0.0946 -0.0925 -0.0928 -0.0970 -0.0940 -0.0894
 -0.0919 -0.0934 -0.0940 -0.0937 -0.0964 -0.0897 -0.0693 -0.0311 -0.0131 -0.0650 -0.0955 -0.0928 -0.0916 -0.0937 -0.0937 -0.0943 -0.0916 -0.0925 -0.0952 -0.0916 -0.0916
 -0.0919 -0.0919 -0.0922 -0.0891 -0.0937 -0.0891 -0.0714 -0.0162 0.0900 0.0693 -0.0272 -0.0790 -0.0909 -0.0894 -0.0922 -0.0903 -0.0894 -0.0906 -0.0931 -0.0931 -0.0903
 -0.0912 -0.0906 -0.0888 -0.0891 -0.0903 -0.0864 -0.0742 -0.0595 0.1898 0.2142 0.1416 -0.0146 -0.0610 -0.0891 -0.0888 -0.0912 -0.0916 -0.0916 -0.0934 -0.0934 -0.0931
 -0.0912 -0.0894 -0.0912 -0.0928 -0.0903 -0.0827 -0.0833 -0.0626 0.2075 0.3119 0.2277 0.0372 -0.0696 -0.0787 -0.0888 -0.0916 -0.0937 -0.0934 -0.0931 -0.0946 -0.0919
 -0.0916 -0.0928 -0.0931 -0.0946 -0.0967 -0.0851 -0.0729 -0.0586 0.0763 0.2628 0.2240 0.0833 -0.0693 -0.0793 -0.0851 -0.0897 -0.0931 -0.0919 -0.0946 -0.0925 -0.0916
 -0.0940 -0.0928 -0.0943 -0.0934 -0.0958 -0.0912 -0.0916 -0.0513 0.0076 0.1422 0.1401 0.0693 -0.0598 -0.0790 -0.0894 -0.0885 -0.0888 -0.0897 -0.0922 -0.0906 -0.0891
 -0.0916 -0.0925 -0.0909 -0.0912 -0.0922 -0.0900 -0.0916 -0.0897 -0.0815 0.0153 0.0528 0.0244 -0.0400 -0.0787 -0.0879 -0.0885 -0.0894 -0.0888 -0.0931 -0.0925 -0.0925
 -0.0909 -0.0885 -0.0891 -0.0916 -0.0940 -0.0903 -0.0934 -0.0906 -0.0931 -0.0851 -0.0385 -0.0131 -0.0281 -0.0760 -0.0885 -0.0909 -0.0916 -0.0909 -0.0943 -0.0940 -0.0931
 -0.0891 -0.0894 -0.0916 -0.0928 -0.0952 -0.0928 -0.0955 -0.0928 -0.0980 -0.0891 -0.0882 -0.0851 -0.0839 -0.0882 -0.0919 -0.0912 -0.0916 -0.0919 -0.0952 -0.0949 -0.0903
 -0.0912 -0.0916 -0.0931 -0.0937 -0.0958 -0.0909 -0.0961 -0.0934 -0.0937 -0.0928 -0.0928 -0.0925 -0.0928 -0.0928 -0.0928 -0.0922 -0.0925 -0.0922 -0.0940 -0.0925 -0.0906
 -0.0912 -0.0934 -0.0925 -0.0919 -0.0946 -0.0916 -0.0928 -0.0894 -0.1001 -0.0922 -0.0912 -0.0903 -0.0916 -0.0931 -0.0909 -0.0897 -0.0909 -0.0906 -0.0912 -0.0919 -0.0888
 -0.0919 -0.0912 -0.0909 -0.0916 -0.0937 -0.0897 -0.0937 -0.0903 -0.0943 -0.0916 -0.0916 -0.0916 -0.0906 -0.0909 -0.0891 -0.0897 -0.0900 -0.0906 -0.0916 -0.0903 -0.0925
 -0.0906 -0.0900 -0.0894 -0.0909 -0.0937 -0.0912 -0.0934 -0.0912 -0.0983 -0.0876 -0.0891 -0.0900 -0.0916 -0.0894 -0.0894 -0.0891 -0.0900 -0.0909 -0.0946 -0.0928 -0.0925
 -0.0912 -0.0909 -0.0891 -0.0937 -0.0946 -0.0912 -0.0940 -0.0934 -0.0955 -0.0888 -0.0912 -0.0912 -0.0912 -0.0919 -0.0928 -0.0943 -0.0925 -0.0931 -0.0943 -0.0934 -0.0928
 -0.0906 -0.0916 -0.0928 -0.0934 -0.0955 -0.0928 -0.0946 -0.0891 -0.0952 -0.0885 -0.0900 -0.0909 -0.0925 -0.0922 -0.0922 -0.0900 -0.0922 -0.0922 -0.0937 -0.0909 -0.0894
Attachment 4: ER-580-200-30kV-m39.txt
value (Voltage offset      40.0000)
     80.0000
      0.7000
     80.0000
  29978.0220
21
21
./emit.dat
 -0.1172 -0.0961 -0.0958 -0.0970 -0.0995 -0.0980 -0.0961 -0.0995 -0.0992 -0.0961 -0.0995 -0.0989 -0.0992 -0.0961 -0.0983 -0.0974 -0.0986 -0.0940 -0.1013 -0.0989 -0.0989
 -0.0906 -0.0903 -0.0906 -0.0909 -0.0922 -0.0928 -0.0931 -0.0986 -0.0946 -0.0946 -0.0943 -0.0961 -0.0937 -0.0958 -0.0922 -0.0916 -0.0922 -0.0940 -0.0986 -0.0955 -0.0955
 -0.0928 -0.0925 -0.0912 -0.0928 -0.0940 -0.0934 -0.0934 -0.0958 -0.0964 -0.0937 -0.0940 -0.0934 -0.0928 -0.0928 -0.0925 -0.0925 -0.0952 -0.0934 -0.0986 -0.0970 -0.0940
 -0.0940 -0.0931 -0.0937 -0.0970 -0.0897 -0.0925 -0.0958 -0.0931 -0.0919 -0.0909 -0.0909 -0.0931 -0.0897 -0.0916 -0.0940 -0.0940 -0.0934 -0.0928 -0.0958 -0.0967 -0.0900
 -0.0934 -0.0928 -0.0922 -0.0986 -0.0934 -0.0897 -0.0922 -0.0943 -0.0934 -0.0912 -0.0919 -0.0912 -0.0946 -0.0922 -0.0916 -0.0937 -0.0925 -0.0919 -0.0946 -0.0955 -0.0940
 -0.0912 -0.0906 -0.0925 -0.0958 -0.0952 -0.0955 -0.0961 -0.0928 -0.0952 -0.0916 -0.0928 -0.0949 -0.0937 -0.0952 -0.0937 -0.0952 -0.0912 -0.0931 -0.0946 -0.0946 -0.0894
 -0.0897 -0.0922 -0.0888 -0.0949 -0.0940 -0.0934 -0.0961 -0.0946 -0.0912 -0.0928 -0.0943 -0.0940 -0.0937 -0.0961 -0.0928 -0.0937 -0.0922 -0.0931 -0.0949 -0.0961 -0.0958
 -0.0919 -0.0906 -0.0912 -0.0970 -0.0906 -0.0912 -0.0958 -0.0925 -0.0809 -0.0858 -0.0897 -0.0937 -0.0903 -0.0919 -0.0906 -0.0934 -0.0937 -0.0928 -0.0974 -0.0964 -0.0916
 -0.0943 -0.0940 -0.0925 -0.0961 -0.0931 -0.0906 -0.0912 -0.0934 -0.0610 -0.0494 -0.0681 -0.0894 -0.0937 -0.0894 -0.0900 -0.0909 -0.0934 -0.0925 -0.0961 -0.0952 -0.0912
 -0.0928 -0.0903 -0.0934 -0.0998 -0.0937 -0.0919 -0.0940 -0.0934 -0.0641 -0.0165 -0.0314 -0.0763 -0.0903 -0.0925 -0.0928 -0.0952 -0.0925 -0.0922 -0.0955 -0.0943 -0.0909
 -0.0912 -0.0925 -0.0900 -0.0961 -0.0961 -0.0934 -0.0937 -0.0937 -0.0851 -0.0357 -0.0394 -0.0638 -0.0909 -0.0906 -0.0928 -0.0952 -0.0919 -0.0916 -0.0952 -0.0940 -0.0928
 -0.0912 -0.0900 -0.0909 -0.0974 -0.0934 -0.0925 -0.0977 -0.0949 -0.0903 -0.0745 -0.0598 -0.0702 -0.0876 -0.0900 -0.0839 -0.0900 -0.0916 -0.0925 -0.0952 -0.0961 -0.0922
 -0.0934 -0.0922 -0.0912 -0.0977 -0.0916 -0.0909 -0.0958 -0.0940 -0.0931 -0.0894 -0.0848 -0.0836 -0.0912 -0.0900 -0.0528 -0.0897 -0.0888 -0.0922 -0.0977 -0.0958 -0.0925
 -0.0931 -0.0906 -0.0940 -0.0989 -0.0909 -0.0916 -0.0952 -0.0919 -0.0925 -0.0916 -0.0888 -0.0912 -0.0885 -0.0891 -0.0864 -0.0897 -0.0897 -0.0912 -0.0922 -0.0943 -0.0900
 -0.0937 -0.0928 -0.0919 -0.0992 -0.0940 -0.0943 -0.0937 -0.0937 -0.0937 -0.0897 -0.0912 -0.0922 -0.0931 -0.0925 -0.0912 0.2444 -0.0891 -0.0900 -0.0952 -0.0925 -0.0885
 -0.0912 -0.0903 -0.0912 -0.0974 -0.0943 -0.0934 -0.0958 -0.0940 -0.0925 -0.0937 -0.0922 -0.0946 -0.0919 -0.0922 -0.0934 -0.0317 -0.0873 -0.0912 -0.0964 -0.0946 -0.0894
 -0.0922 -0.0906 -0.0903 -0.0958 -0.0952 -0.0937 -0.0955 -0.0955 -0.0949 -0.0916 -0.0931 -0.0934 -0.0937 -0.0943 -0.0919 -0.0595 -0.0928 -0.0922 -0.0967 -0.0955 -0.0952
 -0.0931 -0.0897 -0.0906 -0.0961 -0.0928 -0.0916 -0.0970 -0.0946 -0.0922 -0.0925 -0.0897 -0.0922 -0.0903 -0.0873 -0.0900 -0.0919 -0.0949 -0.0937 -0.0937 -0.0952 -0.0916
 -0.0940 -0.0928 -0.0925 -0.0974 -0.0922 -0.0912 -0.0946 -0.0937 -0.0922 -0.0891 -0.0916 -0.0916 -0.0916 -0.0900 -0.0916 -0.0922 -0.0919 -0.0922 -0.0949 -0.0943 -0.0931
 -0.0931 -0.0909 -0.0919 -0.0970 -0.0931 -0.0922 -0.0949 -0.0925 -0.0900 -0.0922 -0.0891 -0.0928 -0.0903 -0.0922 -0.0940 -0.0958 -0.0925 -0.0903 -0.0922 -0.0952 -0.0900
 -0.0919 -0.0900 -0.0888 -0.0958 -0.0940 -0.0922 -0.0949 -0.0937 -0.0934 -0.0922 -0.0928 -0.0937 -0.0943 -0.0949 -0.0928 -0.0937 -0.0903 -0.0906 -0.0955 -0.0937 -0.0925
Attachment 5: ITE-TM4-Ta#50-LP-SIS_Test_Stand.docx
Attachment 6: Ta#50.docx
Attachment 7: TS-2016-10-18-15-05.rga
Attachment 8: TS-2016-10-19-14-42.rga
Attachment 9: OMEGA_OM-DAQPRO-5300_24-Oct-16_11.03.10.dlp
  549   Thursday, October 27, 2016, 13:37 Aaron SchmidtUC#18LP-SISITWTM2 41751-3In progressConditioning Heat shield for UC#18

Thursday, October 27, 2016, 13:36: Roughing.

Thursday, October 27, 2016, 14:27: Turbo pumps started.

Thursday, October 27, 2016, 14:33: Ion gauge on.

Thursday, October 27, 2016, 14:35: Both Turbo pumps at speed.

Thursday, October 27, 2016, 16:00: RGA on.


Friday, October 28, 2016, 07:57: Water lines opened. No response on RGA. Vacuum stable @ 2.80e-07T.


Monday, October 31, 2016, 09:42: Power/sources on.

 Monday, October 31, 2016, 09:43: Auto-ramping TGHT to 500 A @ 2A per min.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 26.3 3.88
 TBHT 26.1 3.10
 COIL + EE 26.0 5.03
 TGHT 26.0 5.67

Monday, October 31, 2016, 14:46: TGHT @ 500 A, looks ok. Auto-ramping TGHT to 580 A @ 2A per min.

Monday, October 31, 2016, 15:28: TGHT @ 580 A. 


Device Temp
[C]
 Flow
[lpm]
 HS + MSP 35.3 3.96
 TBHT 26.1 3.08
 COIL + EE 26.1 5.05
 TGHT 28.2 5.67

Wednesday, November 02, 2016, 08:23: Auto-cooling TGHT @ 2A per min.

Wednesday, November 02, 2016, 13:15: Auto-cooling complete.


Thursday, November 03, 2016, 08:40: Sources/power off.

Thursday, November 03, 2016, 08:42: Water lines purged.


Monday, November 07, 2016, 08:08: RGA off. Ion gauge off. Turbo pumps off.


Tuesday, November 08, 2016, 08:07: Turbo pumps restarted. Removal not yet required.

Tuesday, November 08, 2016, 08:13: Ion gauge switched on.


Wednesday, November 16, 2016, 08:05: Ion Gauge off. Turbo pumps off.

Wednesday, November 16, 2016, 09:45: Venting.

Attachment 1: UC#18_Heat_shield_conditioning.docx
Attachment 2: TS-2016-10-27-16-00.rga
  574   Friday, March 24, 2017, 14:43 Aaron SchmidtTa#51LP-SISITWTM241790-441751-5In progressConditioning Ta#51

Friday, March 24, 2017, 15:40: Roughing.

Friday, March 24, 2017, 15:59: Turbo pumps on.

Friday, March 24, 2017, 16:04: Ion gauge on.

Friday, March 24, 2017, 16:07: RGA on. Turbo pumps @ speed.


Monday, March 27, 2017, 09:07: Water lines opened. No response on RGA. Vacuum @ 1.81e-07T

Monday, March 27, 2017, 09:10: Power on.

Monday, March 27, 2017, 09:50: Troubleshooting: Sources/Optics non-responsive. Controls group notified.

Monday, March 27, 2017, 11:16: Optics all green minus the magnet TMS : MB1. PLC issue reported.

Monday, March 27, 2017, 11:30: PLC reboot. Brief whiteout on screens. Screen function returned. TMS : MB1 still interlocked.

Monday, March 27, 2017, 14:30: TMS : MB1 no longer interlocked. Optics non-responsive to 'P/S On'. Voltages can be adjust while optics are 'in black'


Tuesday, March 28, 2017, 09:30: Optics still have error but can be run.

Tuesday, March 28, 2017, 09:43: BIAS / EZL / EE @ 30 / 13 / 1.5 kV . Spark noted while bring up BIAS.

Tuesday, March 28, 2017, 09:45: IV4 opened.

Tuesday, March 28, 2017, 11:13: RGA Fatal error. Restarted 11:15.

Tuesday, March 28, 2017, 12:57: Optics ok.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 27.5 3.89
 TBHT 27.3 3.00
 COIL + EE 27.1 4.94
 TGHT  27.1 5.59

Tuesday, March 28, 2017, 13:01: Auto ramping TGHT / TBHT to 300 / 100 A @ 2A per min.

Tuesday, March 28, 2017, 15:52: TGHT / TBHT @ 300 / 100. Vacuum @ 2.16e-07T

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 27.5 3.85
 TBHT 26.7 3.12
 COIL + EE 25.8 4.92
 TGHT  26.5 5.59

 


Wednesday, March 29, 2017, 08:25: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.

Wednesday, March 29, 2017, 11:23: TMS : MB1 set from 'Reverse' to 'Normal' manually. Interlock Reset. (Mass Scan ok).

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 36.9 3.82
 TBHT 27.2 3.07
 COIL + EE 27.4 4.97
 TGHT  29.4 5.58

Wednesday, March 29, 2017, 13:09: Auto ramping TBHT to 220 A @ 2A per min.

Wednesday, March 29, 2017, 13:20: Source page @ 218.9A - Read back on power supply @ 219A. Fatal Error on RGA.

Wednesday, March 29, 2017, 13:27: Manually ramp TBHT to 230 A @ 2A per min (approx).

Wednesday, March 29, 2017, 15:15: DAC power supply appears to have failed. Pop sound and electrical smoke/smell noticed. 

Wednesday, March 29, 2017, 15:28: Auto cooling TBHT to 220 A @ 2A per min.


Thursday, March 30, 2017, 08:42: Auto-cooling @ 2A per min.


Friday, March 31, 2017, 08:35: IV4 closed.

Friday, March 31, 2017, 08:44: Power/sources off.

Friday, March 31, 2017, 08:46: Ion gauge, Turbos off. Waterlines purged.


Thursday, April 06, 2017, 10:40: Venting

Attachment 1: ER-580-230-30kV-M22.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
     80.0000
  29978.0220
21
21
./emit.dat
 -0.1138 -0.0684 -0.0720 -0.0696 -0.0681 -0.0662 -0.0662 -0.0668 -0.0671 -0.0763 -0.0684 -0.0696 -0.0702 -0.0705 -0.0674 -0.0674 -0.0690 -0.0674 -0.0772 -0.0684 -0.0668
 -0.0650 -0.0668 -0.0659 -0.0653 -0.0620 -0.0626 -0.0653 -0.0635 -0.0653 -0.0726 -0.0665 -0.0674 -0.0644 -0.0677 -0.0635 -0.0656 -0.0644 -0.0620 -0.0717 -0.0641 -0.0684
 -0.0632 -0.0647 -0.0638 -0.0632 -0.0638 -0.0638 -0.0650 -0.0650 -0.0635 -0.0629 -0.0635 -0.0653 -0.0647 -0.0635 -0.0659 -0.0626 -0.0638 -0.0613 -0.0604 -0.0626 -0.0656
 -0.0616 -0.0647 -0.0641 -0.0656 -0.0571 -0.0641 -0.0644 -0.0650 -0.0620 -0.0610 -0.0638 -0.0626 -0.0613 -0.0635 -0.0656 -0.0629 -0.0650 -0.0620 -0.0623 -0.0644 -0.0638
 -0.0598 -0.0653 -0.0629 -0.0702 -0.0656 -0.0629 -0.0638 -0.0635 -0.0629 -0.0632 -0.0616 -0.0693 -0.0677 -0.0626 -0.0644 -0.0647 -0.0644 -0.0626 -0.0732 -0.0650 -0.0623
 -0.0616 -0.0668 -0.0644 -0.0662 -0.0665 -0.0635 -0.0626 -0.0629 -0.0626 -0.0647 -0.0656 -0.0665 -0.0644 -0.0644 -0.0632 -0.0641 -0.0629 -0.0659 -0.0613 -0.0607 -0.0595
 -0.0632 -0.0626 -0.0632 -0.0674 -0.0632 -0.0659 -0.0641 -0.0623 -0.0653 -0.0644 -0.0656 -0.0687 -0.0623 -0.0632 -0.0623 -0.0613 -0.0629 -0.0650 -0.0598 -0.0598 -0.0598
 -0.0623 -0.0592 -0.0568 -0.0705 -0.0635 -0.0613 -0.0629 -0.0626 -0.0653 -0.0650 -0.0616 -0.0674 -0.0577 -0.0635 -0.0650 -0.0607 -0.0610 -0.0604 -0.0623 -0.0623 -0.0644
 -0.0650 -0.0595 -0.0613 -0.0629 -0.0641 -0.0629 -0.0626 -0.0644 -0.0650 -0.0613 -0.0586 -0.0684 -0.0626 -0.0616 -0.0620 -0.0623 -0.0668 -0.0623 -0.0632 -0.0656 -0.0656
 -0.0650 -0.0620 -0.0638 -0.0659 -0.0677 -0.0623 -0.0632 -0.0620 -0.0632 -0.0595 -0.0616 -0.0705 -0.0693 -0.0623 -0.0616 -0.0629 -0.0665 -0.0613 -0.0610 -0.0647 -0.0638
 -0.0629 -0.0638 -0.0668 -0.0729 -0.0607 -0.0632 -0.0601 -0.0601 -0.0607 -0.0620 -0.0620 -0.0705 -0.0671 -0.0623 -0.0647 -0.0641 -0.0632 -0.0604 -0.0668 -0.0629 -0.0626
 -0.0607 -0.0653 -0.0635 -0.0635 -0.0653 -0.0629 -0.0610 -0.0613 -0.0632 -0.0638 -0.0659 -0.0702 -0.0635 -0.0626 -0.0653 -0.0623 -0.0604 -0.0598 -0.0620 -0.0623 -0.0626
 -0.0598 -0.0659 -0.0616 -0.0623 -0.0638 -0.0620 -0.0620 -0.0613 -0.0616 -0.0635 -0.0638 -0.0681 -0.0644 -0.0604 -0.0656 -0.0616 -0.0613 -0.0613 -0.0638 -0.0635 -0.0632
 -0.0629 -0.0610 -0.0598 -0.0687 -0.0635 -0.0629 -0.0638 -0.0616 -0.0610 -0.0604 -0.0623 -0.0638 -0.0653 -0.0620 -0.0632 -0.0595 -0.0632 -0.0650 -0.0659 -0.0647 -0.0641
 -0.0629 -0.0595 -0.0632 -0.0690 -0.0650 -0.0635 -0.0616 -0.0604 -0.0595 -0.0577 -0.0583 -0.0702 -0.0613 -0.0647 -0.0598 -0.0635 -0.0647 -0.0653 -0.0696 -0.0595 -0.0620
 -0.0638 -0.0613 -0.0641 -0.0684 -0.0620 -0.0616 -0.0610 -0.0604 -0.0632 -0.0583 -0.0595 -0.0720 -0.0635 -0.0638 -0.0626 -0.0604 -0.0650 -0.0632 -0.0629 -0.0583 -0.0604
 -0.0610 -0.0623 -0.0613 -0.0690 -0.0613 -0.0607 -0.0601 -0.0610 -0.0626 -0.0595 -0.0580 -0.0674 -0.0632 -0.0644 -0.0644 -0.0620 -0.0641 -0.0629 -0.0647 -0.0601 -0.0632
 -0.0635 -0.0595 -0.0638 -0.0647 -0.0644 -0.0632 -0.0623 -0.0638 -0.0632 -0.0577 -0.0549 -0.0656 -0.0641 -0.0632 -0.0638 -0.0601 -0.0635 -0.0607 -0.0623 -0.0644 -0.0671
 -0.0650 -0.0616 -0.0632 -0.0662 -0.0684 -0.0635 -0.0644 -0.0635 -0.0620 -0.0607 -0.0574 -0.0644 -0.0601 -0.0589 -0.0635 -0.0641 -0.0616 -0.0616 -0.0641 -0.0650 -0.0644
 -0.0620 -0.0659 -0.0629 -0.0674 -0.0653 -0.0644 -0.0632 -0.0620 -0.0626 -0.0607 -0.0613 -0.0674 -0.0610 -0.0598 -0.0641 -0.0635 -0.0626 -0.0623 -0.0650 -0.0613 -0.0607
 -0.0595 -0.0641 -0.0623 -0.0653 -0.0632 -0.0647 -0.0635 -0.0601 -0.0613 -0.0626 -0.0650 -0.0690 -0.0613 -0.0623 -0.0665 -0.0638 -0.0613 -0.0644 -0.0601 -0.0620 -0.0601
Attachment 2: ER-580-230-30kV-M38.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
     80.0000
  29978.0220
21
21
./emit.dat
 -0.1169 -0.0748 -0.0684 -0.0653 -0.0705 -0.0720 -0.0681 -0.0693 -0.0677 -0.0766 -0.0677 -0.0668 -0.0668 -0.0687 -0.0690 -0.0674 -0.0690 -0.0690 -0.0784 -0.0711 -0.0699
 -0.0665 -0.0693 -0.0690 -0.0626 -0.0681 -0.0662 -0.0629 -0.0638 -0.0610 -0.0717 -0.0677 -0.0650 -0.0638 -0.0616 -0.0665 -0.0656 -0.0638 -0.0659 -0.0754 -0.0674 -0.0659
 -0.0671 -0.0635 -0.0668 -0.0647 -0.0659 -0.0659 -0.0604 -0.0653 -0.0635 -0.0745 -0.0659 -0.0677 -0.0647 -0.0647 -0.0623 -0.0674 -0.0665 -0.0653 -0.0623 -0.0623 -0.0641
 -0.0699 -0.0644 -0.0656 -0.0690 -0.0623 -0.0641 -0.0656 -0.0635 -0.0656 -0.0668 -0.0650 -0.0662 -0.0650 -0.0650 -0.0632 -0.0650 -0.0629 -0.0613 -0.0677 -0.0665 -0.0653
 -0.0702 -0.0635 -0.0684 -0.0647 -0.0723 -0.0650 -0.0647 -0.0641 -0.0674 -0.0644 -0.0632 -0.0668 -0.0620 -0.0638 -0.0656 -0.0653 -0.0632 -0.0601 -0.0623 -0.0668 -0.0665
 -0.0671 -0.0629 -0.0659 -0.0677 -0.0674 -0.0656 -0.0641 -0.0638 -0.0668 -0.0613 -0.0632 -0.0693 -0.0604 -0.0647 -0.0647 -0.0656 -0.0632 -0.0626 -0.0632 -0.0638 -0.0635
 -0.0659 -0.0620 -0.0684 -0.0638 -0.0690 -0.0635 -0.0620 -0.0620 -0.0632 -0.0647 -0.0635 -0.0668 -0.0696 -0.0662 -0.0659 -0.0668 -0.0650 -0.0665 -0.0668 -0.0620 -0.0586
 -0.0644 -0.0629 -0.0650 -0.0684 -0.0629 -0.0616 -0.0613 -0.0647 -0.0620 -0.0647 -0.0644 -0.0650 -0.0751 -0.0635 -0.0659 -0.0732 -0.0595 -0.0644 -0.0650 -0.0629 -0.0623
 -0.0653 -0.0647 -0.0662 -0.0668 -0.0620 -0.0595 -0.0626 -0.0641 -0.0650 -0.0632 -0.0598 -0.0729 -0.0620 -0.0623 -0.0626 -0.0620 -0.0613 -0.0632 -0.0632 -0.0620 -0.0650
 -0.0681 -0.0659 -0.0684 -0.0687 -0.0629 -0.0632 -0.0653 -0.0620 -0.0656 -0.0613 -0.0616 -0.0702 -0.0714 -0.0610 -0.0629 -0.0638 -0.0638 -0.0626 -0.0668 -0.0638 -0.0635
 -0.0699 -0.0638 -0.0677 -0.0644 -0.0641 -0.0644 -0.0650 -0.0629 -0.0607 -0.0595 -0.0616 -0.0708 -0.0674 -0.0607 -0.0644 -0.0613 -0.0641 -0.0641 -0.0671 -0.0629 -0.0629
 -0.0687 -0.0613 -0.0638 -0.0659 -0.0635 -0.0632 -0.0626 -0.0610 -0.0613 -0.0620 -0.0650 -0.0677 -0.0681 -0.0616 -0.0653 -0.0641 -0.0635 -0.0647 -0.0620 -0.0607 -0.0616
 -0.0677 -0.0601 -0.0659 -0.0650 -0.0626 -0.0610 -0.0635 -0.0604 -0.0616 -0.0647 -0.0647 -0.0653 -0.0620 -0.0635 -0.0644 -0.0656 -0.0613 -0.0629 -0.0632 -0.0635 -0.0632
 -0.0659 -0.0623 -0.0668 -0.0653 -0.0623 -0.0610 -0.0626 -0.0623 -0.0702 -0.0623 -0.0620 -0.0702 -0.0723 -0.0616 -0.0635 -0.0638 -0.0610 -0.0604 -0.0659 -0.0653 -0.0629
 -0.0644 -0.0650 -0.0699 -0.0674 -0.0632 -0.0638 -0.0626 -0.0647 -0.0638 -0.0571 -0.0601 -0.0653 -0.0720 -0.0616 -0.0635 -0.0604 -0.0629 -0.0638 -0.0653 -0.0668 -0.0647
 -0.0662 -0.0632 -0.0665 -0.0671 -0.0641 -0.0635 -0.0620 -0.0647 -0.0626 -0.0577 -0.0562 -0.0641 -0.0662 -0.0635 -0.0647 -0.0650 -0.0616 -0.0644 -0.0671 -0.0595 -0.0607
 -0.0665 -0.0610 -0.0659 -0.0656 -0.0613 -0.0610 -0.0610 -0.0616 -0.0610 -0.0604 -0.0574 -0.0665 -0.0665 -0.0644 -0.0635 -0.0647 -0.0644 -0.0623 -0.0647 -0.0595 -0.0595
 -0.0656 -0.0623 -0.0653 -0.0684 -0.0613 -0.0592 -0.0629 -0.0635 -0.0629 -0.0620 -0.0589 -0.0659 -0.0714 -0.0595 -0.0613 -0.0629 -0.0635 -0.0626 -0.0656 -0.0613 -0.0620
 -0.0662 -0.0653 -0.0671 -0.0696 -0.0598 -0.0610 -0.0644 -0.0650 -0.0635 -0.0647 -0.0604 -0.0656 -0.0638 -0.0595 -0.0638 -0.0629 -0.0613 -0.0613 -0.0665 -0.0647 -0.0641
 -0.0696 -0.0629 -0.0644 -0.0705 -0.0629 -0.0647 -0.0635 -0.0656 -0.0626 -0.0616 -0.0620 -0.0656 -0.0601 -0.0629 -0.0604 -0.0635 -0.0613 -0.0598 -0.0644 -0.0629 -0.0635
 -0.0699 -0.0626 -0.0632 -0.0677 -0.0629 -0.0641 -0.0620 -0.0595 -0.0623 -0.0626 -0.0616 -0.0659 -0.0626 -0.0629 -0.0626 -0.0647 -0.0610 -0.0601 -0.0650 -0.0650 -0.0620
Attachment 3: ER-580-230-30kV-M133.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
     80.0000
  29978.0220
21
21
./emit.dat
 -0.0662 -0.0671 -0.0674 -0.0693 -0.0681 -0.0693 -0.0674 -0.0693 -0.0717 -0.0690 -0.0684 -0.0662 -0.0653 -0.0616 -0.0665 -0.0674 -0.0665 -0.0662 -0.0763 -0.0702 -0.0742
 -0.0665 -0.0726 -0.0665 -0.0696 -0.0677 -0.0684 -0.0656 -0.0693 -0.0681 -0.0626 -0.0677 -0.0644 -0.0653 -0.0589 -0.0671 -0.0671 -0.0635 -0.0626 -0.0739 -0.0656 -0.0650
 -0.0662 -0.0732 -0.0699 -0.0702 -0.0699 -0.0705 -0.0711 -0.0677 -0.0623 -0.0635 -0.0635 -0.0644 -0.0647 -0.0653 -0.0659 -0.0644 -0.0610 -0.0592 -0.0604 -0.0610 -0.0583
 -0.0641 -0.0671 -0.0650 -0.0702 -0.0726 -0.0601 -0.0693 -0.0662 -0.0644 -0.0610 -0.0601 -0.0659 -0.0620 -0.0674 -0.0632 -0.0613 -0.0607 -0.0623 -0.0650 -0.0604 -0.0662
 -0.0671 -0.0632 -0.0671 -0.0650 -0.0665 -0.0626 -0.0684 -0.0638 -0.0629 -0.0616 -0.0623 -0.0668 -0.0613 -0.0598 -0.0543 -0.0610 -0.0638 -0.0641 -0.0650 -0.0620 -0.0644
 -0.0684 -0.0656 -0.0693 -0.0668 -0.0684 -0.0726 -0.0702 -0.0635 -0.0659 -0.0671 -0.0671 -0.0647 -0.0635 -0.0574 -0.0604 -0.0644 -0.0644 -0.0632 -0.0684 -0.0635 -0.0650
 -0.0674 -0.0726 -0.0677 -0.0656 -0.0668 -0.0690 -0.0699 -0.0159 -0.0644 -0.0656 -0.0641 -0.0705 -0.0607 -0.0671 -0.0641 -0.0641 -0.0638 -0.0626 -0.0632 -0.0635 -0.0620
 -0.0668 -0.0699 -0.0684 -0.0671 -0.0665 -0.0659 -0.0699 -0.0641 -0.0610 -0.0607 -0.0577 -0.0665 -0.0632 -0.0638 -0.0638 -0.0638 -0.0629 -0.0604 -0.0626 -0.0623 -0.0629
 -0.0690 -0.0668 -0.0684 -0.0684 -0.0665 -0.0668 -0.0659 -0.0613 -0.0629 -0.0613 -0.0607 -0.0711 -0.0644 -0.0610 -0.0626 -0.0620 -0.0623 -0.0595 -0.0601 -0.0623 -0.0674
 -0.0674 -0.0668 -0.0659 -0.0684 -0.0696 -0.0656 -0.0677 -0.0650 -0.0644 -0.0598 -0.0635 -0.0729 -0.0616 -0.0610 -0.0620 -0.0623 -0.0607 -0.0638 -0.0626 -0.0620 -0.0626
 -0.0635 -0.0647 -0.0665 -0.0668 -0.0681 -0.0665 -0.0693 -0.0644 -0.0653 -0.0626 -0.0632 -0.0708 -0.0644 -0.0623 -0.0650 -0.0623 -0.0604 -0.0644 -0.0647 -0.0632 -0.0620
 -0.0647 -0.0659 -0.0671 -0.0696 -0.0671 -0.0690 -0.0699 -0.0623 -0.0607 -0.0616 -0.0641 -0.0693 -0.0626 -0.0647 -0.0656 -0.0635 -0.0653 -0.0650 -0.0656 -0.0620 -0.0626
 -0.0671 -0.0662 -0.0677 -0.0665 -0.0674 -0.0653 -0.0711 -0.0613 -0.0516 -0.0592 -0.0647 -0.0656 -0.0632 -0.0644 -0.0638 -0.0616 -0.0650 -0.0629 -0.0656 -0.0601 -0.0616
 -0.0656 -0.0647 -0.0656 -0.0671 -0.0671 -0.0632 -0.0668 -0.0626 -0.0568 -0.0519 -0.0586 -0.0665 -0.0604 -0.0601 -0.0592 -0.0620 -0.0598 -0.0604 -0.0592 -0.0610 -0.0620
 -0.0641 -0.0671 -0.0671 -0.0656 -0.0665 -0.0681 -0.0684 -0.0638 -0.0623 -0.0500 -0.0473 -0.0708 -0.0607 -0.0595 -0.0607 -0.0613 -0.0616 -0.0610 -0.0641 -0.0632 -0.0644
 -0.0641 -0.0684 -0.0674 -0.0671 -0.0684 -0.0702 -0.0696 -0.0650 -0.0623 -0.0552 -0.0436 -0.0647 -0.0644 -0.0595 -0.0632 -0.0641 -0.0653 -0.0616 -0.0635 -0.0653 -0.0632
 -0.0671 -0.0690 -0.0684 -0.0674 -0.0699 -0.0677 -0.0699 -0.0644 -0.0647 -0.0610 -0.0430 -0.0467 -0.0629 -0.0623 -0.0671 -0.0638 -0.0620 -0.0635 -0.0632 -0.0635 -0.0610
 -0.0696 -0.0702 -0.0681 -0.0696 -0.0647 -0.0656 -0.0659 -0.0623 -0.0635 -0.0592 -0.0446 -0.0461 -0.0522 -0.0620 -0.0650 -0.0641 -0.0586 -0.0610 -0.0647 -0.0616 -0.0616
 -0.0668 -0.0644 -0.0650 -0.0677 -0.0668 -0.0632 -0.0656 -0.0616 -0.0620 -0.0543 -0.0531 -0.0491 -0.0418 -0.0586 -0.0613 -0.0626 -0.0589 -0.0623 -0.0674 -0.0592 -0.0613
 -0.0662 -0.0623 -0.0653 -0.0656 -0.0656 -0.0650 -0.0699 -0.0641 -0.0616 -0.0607 -0.0604 -0.0488 -0.0464 -0.0528 -0.0616 -0.0595 -0.0632 -0.0626 -0.0671 -0.0613 -0.0598
 -0.0644 -0.0671 -0.0674 -0.0696 -0.0668 -0.0671 -0.0696 -0.0626 -0.0607 -0.0644 -0.0638 -0.0696 -0.0626 -0.0583 -0.0623 -0.0610 -0.0635 -0.0626 -0.0659 -0.0613 -0.0635
Attachment 4: OMEGA_OM-DAQPRO-5300_06-Apr-17_13.58.21.dlp
Attachment 5: Ta#51.docx
Attachment 6: TS-2017-03-24-16-07.rga
Attachment 7: TS-2017-03-28-11-15.rga
Attachment 8: TS-2017-03-29-13-26.rga
Attachment 9: ITW-TM2-Ta#51-LP-SIS_Test_Stand.docx
  592   Friday, June 09, 2017, 13:28 Aaron SchmidtTa#52LP-SISITETM442218-1441751-8In progressConditioning Ta#52

Friday, June 09, 2017, 13:27: Roughing.

Friday, June 09, 2017, 14:09: Turbo pumps on.

Friday, June 09, 2017, 14:16: Ion gauge on.


Monday, June 12, 2017, 07:20: Waterlines opened. No response on RGA. Vacuum @1.52e-07T.

Monday, June 12, 2017, 07:23: Power/Sources on.

Monday, June 12, 2017, 08:15: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .

Monday, June 12, 2017, 08:16: IV4 Opened.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 26.0 3.78
 TBHT 27.3 3.03
 COIL + EE 25.8 4.91
 TGHT  26.0 5.46

Monday, June 12, 2017, 08:18: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.

Monday, June 12, 2017, 11:33: Spark around 10:30 tripped RGA. Restarting RGA.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 35.9 3.78
 TBHT 26.6 2.95
 COIL + EE 26.1 4.89
 TGHT  28.4 5.46

 


Tuesday, June 13, 2017, 14:50: Auto-cooling @ 2A  per min


Wednesday, June 14, 2017, 07:41: Sources/power off. IV4 closed


Friday, June 16, 2017, 07:16: RGA off

Friday, June 16, 2017, 07:18: Ion gauge & turbo pumps off.

Friday, June 16, 2017, 07:20: Water lines purged.


Monday, June 19, 2017, 13:35: While writing report, it was noticed a step was skipped on the mass correlations. The MassCorr data was not saved. - AS

Attachment 1: ER-580-200-30kV-m7.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1483 -0.0916 -0.0842 -0.0763 -0.0848 -0.0891 -0.0793 -0.0638 -0.0726 -0.0879 -0.0949 -0.0916 -0.0839 -0.0815 -0.0836 -0.0937 -0.0824 -0.0797 -0.0909 -0.0873 -0.0891
 -0.0891 -0.0842 -0.0806 -0.0757 -0.0836 -0.0900 -0.0821 -0.0668 -0.0754 -0.0854 -0.0916 -0.0876 -0.0812 -0.0754 -0.0806 -0.0882 -0.0833 -0.0748 -0.0861 -0.0812 -0.0833
 -0.0873 -0.0833 -0.0714 -0.0745 -0.0787 -0.0916 -0.0827 -0.0699 -0.0705 -0.0775 -0.0891 -0.0769 -0.0790 -0.0732 -0.0809 -0.0891 -0.0754 -0.0714 -0.0757 -0.0720 -0.0684
 -0.0894 -0.0772 -0.0720 -0.0781 -0.0732 -0.0888 -0.0845 -0.0720 -0.0641 -0.0760 -0.0800 -0.0720 -0.0665 -0.0647 -0.0784 -0.0891 -0.0726 -0.0723 -0.0696 -0.0674 -0.0681
 -0.0906 -0.0769 -0.0705 -0.0708 -0.0705 -0.0870 -0.0836 -0.0690 -0.0681 -0.0778 -0.0766 -0.0693 -0.0696 -0.0693 -0.0760 -0.0867 -0.0705 -0.0708 -0.0714 -0.0696 -0.0693
 -0.0906 -0.0766 -0.0674 -0.0812 -0.0742 -0.0864 -0.0845 -0.0702 -0.0742 -0.0793 -0.0775 -0.0726 -0.0684 -0.0732 -0.0760 -0.0876 -0.0717 -0.0684 -0.0671 -0.0690 -0.0711
 -0.0900 -0.0754 -0.0708 -0.0299 -0.0372 -0.0882 -0.0867 -0.0726 -0.0723 -0.0806 -0.0751 -0.0739 -0.0732 -0.0711 -0.0787 -0.0885 -0.0705 -0.0708 -0.0711 -0.0699 -0.0684
 -0.0769 -0.0714 -0.0650 0.1456 0.2567 0.1089 -0.0485 -0.0705 -0.0702 -0.0769 -0.0739 -0.0723 -0.0705 -0.0665 -0.0803 -0.0885 -0.0693 -0.0705 -0.0714 -0.0690 -0.0668
 0.0165 0.0143 0.0021 0.1740 0.2505 0.4282 0.2008 -0.0229 -0.0650 -0.0742 -0.0732 -0.0711 -0.0705 -0.0684 -0.0787 -0.0891 -0.0687 -0.0705 -0.0696 -0.0690 -0.0677
 0.0232 0.0513 0.1300 0.2176 0.3403 0.6772 0.6064 0.1114 -0.0391 -0.0623 -0.0711 -0.0699 -0.0720 -0.0681 -0.0772 -0.0885 -0.0690 -0.0720 -0.0729 -0.0693 -0.0684
 -0.0848 0.1108 0.1334 0.1318 0.2350 0.5215 0.8182 0.4596 0.0214 0.0711 -0.0400 -0.0549 -0.0583 -0.0696 -0.0787 -0.0885 -0.0705 -0.0720 -0.0760 -0.0681 -0.0693
 -0.0879 -0.0702 -0.0650 0.1450 0.3421 0.3604 0.6375 0.5753 0.1495 0.0430 -0.0662 -0.0537 -0.0552 -0.0562 -0.0662 -0.0876 -0.0696 -0.0708 -0.0699 -0.0681 -0.0690
 -0.0891 -0.0757 -0.0708 -0.0742 -0.0568 0.2441 0.4233 0.4581 0.1941 0.0522 -0.0720 -0.0699 -0.0677 -0.0626 -0.0717 -0.0800 -0.0589 -0.0723 -0.0702 -0.0677 -0.0681
 -0.0903 -0.0760 -0.0684 -0.0668 -0.0687 -0.0769 -0.0671 0.0940 0.2032 0.1163 0.0055 -0.0696 -0.0693 -0.0696 -0.0772 -0.0864 -0.0659 -0.0659 -0.0641 -0.0681 -0.0684
 -0.0912 -0.0732 -0.0671 -0.0812 -0.0702 -0.0821 -0.0806 -0.0650 -0.0647 -0.0693 -0.0677 -0.0708 -0.0696 -0.0696 -0.0763 -0.0900 -0.0687 -0.0684 -0.0656 -0.0623 -0.0696
 -0.0876 -0.0732 -0.0671 -0.0726 -0.0690 -0.0809 -0.0818 -0.0681 -0.0696 -0.0766 -0.0711 -0.0729 -0.0717 -0.0693 -0.0772 -0.0909 -0.0681 -0.0708 -0.0687 -0.0696 -0.0677
 -0.0867 -0.0729 -0.0687 -0.0732 -0.0638 -0.0812 -0.0821 -0.0699 -0.0684 -0.0769 -0.0723 -0.0702 -0.0690 -0.0677 -0.0772 -0.0900 -0.0693 -0.0711 -0.0696 -0.0687 -0.0681
 -0.0861 -0.0745 -0.0690 -0.0729 -0.0632 -0.0775 -0.0815 -0.0690 -0.0674 -0.0757 -0.0729 -0.0696 -0.0699 -0.0668 -0.0775 -0.0900 -0.0677 -0.0714 -0.0677 -0.0677 -0.0668
 -0.0858 -0.0745 -0.0699 -0.0693 -0.0601 -0.0714 -0.0775 -0.0681 -0.0665 -0.0742 -0.0717 -0.0677 -0.0684 -0.0662 -0.0766 -0.0885 -0.0702 -0.0708 -0.0711 -0.0659 -0.0677
 -0.0864 -0.0720 -0.0671 -0.0711 -0.0607 -0.0659 -0.0684 -0.0650 -0.0668 -0.0754 -0.0720 -0.0720 -0.0696 -0.0690 -0.0766 -0.0873 -0.0693 -0.0705 -0.0714 -0.0690 -0.0677
 -0.0858 -0.0690 -0.0620 -0.0662 -0.0534 -0.0583 -0.0607 -0.0604 -0.0671 -0.0757 -0.0732 -0.0711 -0.0687 -0.0684 -0.0751 -0.0876 -0.0687 -0.0711 -0.0717 -0.0684 -0.0681
Attachment 2: ER-580-200-30kV-m23.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.4541 -0.0946 -0.0922 -0.0946 -0.0876 -0.0720 -0.0491 0.0244 0.1660 0.1889 0.0699 0.0421 -0.0452 -0.0714 -0.0754 -0.0748 -0.0809 -0.0827 -0.0925 -0.0912 -0.0946
 -0.0806 -0.0879 -0.0879 -0.0937 -0.0946 -0.0912 -0.0757 -0.0491 0.0555 0.1382 0.0497 0.0366 -0.0494 -0.0775 -0.0754 -0.0714 -0.0751 -0.0781 -0.0922 -0.0858 -0.0867
 -0.0769 -0.0839 -0.0769 -0.0790 -0.0970 -0.0943 -0.0900 -0.0784 -0.0418 0.0555 0.0235 0.0232 -0.0479 -0.0793 -0.0836 -0.0778 -0.0803 -0.0760 -0.0790 -0.0769 -0.0708
 -0.0754 -0.0827 -0.0735 -0.0827 -0.0745 -0.0970 -0.1038 -0.1031 -0.0830 -0.0449 -0.0208 -0.0140 -0.0449 -0.0757 -0.0839 -0.0821 -0.0800 -0.0836 -0.0711 -0.0699 -0.0696
 -0.0742 -0.0836 -0.0757 -0.0879 -0.0766 -0.0940 -0.1016 -0.1096 -0.1108 -0.1038 -0.0745 -0.0513 -0.0601 -0.0775 -0.0836 -0.0827 -0.0839 -0.0818 -0.0751 -0.0732 -0.0705
 -0.0778 -0.0845 -0.0763 -0.0793 -0.0809 -0.1038 -0.1068 -0.1062 -0.0928 -0.0952 -0.1074 -0.0909 -0.0824 -0.0793 -0.0824 -0.0797 -0.0824 -0.0833 -0.0742 -0.0739 -0.0742
 -0.0803 -0.0848 -0.0778 -0.0879 -0.0797 -0.1163 -0.1208 -0.1181 -0.1028 -0.0916 -0.0934 -0.0812 -0.0766 -0.0790 -0.0809 -0.0787 -0.0812 -0.0830 -0.0708 -0.0726 -0.0739
 -0.0513 -0.0827 -0.0732 -0.0656 0.0027 0.5679 0.0867 -0.1138 -0.1123 -0.0974 -0.0958 -0.0845 -0.0739 -0.0775 -0.0821 -0.0793 -0.0848 -0.0827 -0.0809 -0.0711 -0.0717
 1.2949 1.0226 0.2405 0.0510 -0.0192 4.0729 3.4451 1.0425 0.0763 -0.0818 -0.0916 -0.0830 -0.0745 -0.0797 -0.0821 -0.0803 -0.0833 -0.0824 -0.0711 -0.0732 -0.0723
 0.4846 1.1899 1.5659 1.6354 1.1652 4.6243 7.5842 5.8371 1.2793 0.0476 -0.0735 -0.0766 -0.0751 -0.0781 -0.0806 -0.0787 -0.0803 -0.0809 -0.0705 -0.0729 -0.0714
 -0.0562 0.5368 0.7703 1.5552 3.1232 5.3870 7.5870 9.9997 7.3779 2.2556 1.7380 1.2317 0.4492 -0.0720 -0.0793 -0.0793 -0.0806 -0.0800 -0.0775 -0.0739 -0.0739
 -0.0739 -0.0690 -0.0330 1.1514 1.3330 2.6193 5.1074 9.9997 9.9997 6.2830 2.7347 0.7892 0.9729 1.2283 0.4895 -0.0659 -0.0793 -0.0824 -0.0702 -0.0742 -0.0729
 -0.0748 -0.0800 -0.0693 -0.0500 -0.0009 2.8204 3.1949 5.0052 9.5169 9.9997 3.6981 3.7781 0.2814 0.5304 0.5841 0.7227 0.6760 -0.0620 -0.0797 -0.0705 -0.0708
 -0.0708 -0.0793 -0.0720 -0.0821 -0.0668 -0.0140 0.2133 4.2502 7.5003 9.9997 5.7306 3.4097 -0.0299 -0.0128 0.1605 0.4138 0.3284 0.2347 0.5820 -0.0699 -0.0711
 -0.0739 -0.0809 -0.0732 -0.0769 -0.0745 -0.0745 -0.0681 0.0278 0.2014 4.1779 7.2113 5.8627 2.5699 -0.0647 -0.0699 -0.0580 -0.0131 0.1538 0.1840 0.2417 0.4391
 -0.0748 -0.0809 -0.0751 -0.0836 -0.0745 -0.0781 -0.0766 -0.0641 -0.0452 0.0305 0.1208 0.3046 0.1395 -0.0665 -0.0732 -0.0717 -0.0763 -0.0720 -0.0412 0.0812 0.0699
 -0.0732 -0.0833 -0.0757 -0.0851 -0.0748 -0.0766 -0.0775 -0.0714 -0.0601 -0.0516 -0.0620 -0.0552 -0.0610 -0.0739 -0.0754 -0.0732 -0.0772 -0.0793 -0.0757 -0.0677 -0.0543
 -0.0693 -0.0800 -0.0717 -0.0864 -0.0745 -0.0607 -0.0693 -0.0708 -0.0638 -0.0616 -0.0705 -0.0635 -0.0690 -0.0757 -0.0797 -0.0769 -0.0790 -0.0790 -0.0708 -0.0693 -0.0699
 -0.0610 -0.0720 -0.0659 -0.0806 -0.0732 -0.0348 -0.0369 -0.0574 -0.0653 -0.0671 -0.0757 -0.0696 -0.0714 -0.0787 -0.0824 -0.0769 -0.0787 -0.0772 -0.0729 -0.0693 -0.0705
 -0.0595 -0.0653 -0.0601 -0.0693 -0.0653 -0.0027 0.0113 -0.0067 -0.0519 -0.0677 -0.0763 -0.0729 -0.0714 -0.0809 -0.0809 -0.0766 -0.0787 -0.0803 -0.0772 -0.0708 -0.0723
 -0.0595 -0.0641 -0.0528 -0.0607 -0.0488 0.0327 0.0610 0.0708 0.0015 -0.0488 -0.0653 -0.0616 -0.0656 -0.0748 -0.0797 -0.0757 -0.0787 -0.0793 -0.0793 -0.0729 -0.0717
Attachment 3: ER-580-200-30kV-m39.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1331 -0.0787 -0.0827 -0.0873 -0.0845 -0.0864 -0.0876 -0.0812 -0.0732 -0.0800 -0.0842 -0.0742 -0.0742 -0.0845 -0.0864 -0.0845 -0.0821 -0.0851 -0.0931 -0.0879 -0.0916
 -0.0739 -0.0742 -0.0760 -0.0848 -0.0812 -0.0861 -0.0858 -0.0821 -0.0739 -0.0797 -0.0818 -0.0711 -0.0745 -0.0827 -0.0833 -0.0809 -0.0769 -0.0830 -0.0894 -0.0833 -0.0888
 -0.0748 -0.0735 -0.0732 -0.0763 -0.0800 -0.0793 -0.0836 -0.0803 -0.0757 -0.0732 -0.0797 -0.0705 -0.0726 -0.0806 -0.0873 -0.0821 -0.0766 -0.0815 -0.0775 -0.0748 -0.0757
 -0.0754 -0.0711 -0.0702 -0.0854 -0.0803 -0.0766 -0.0818 -0.0800 -0.0803 -0.0745 -0.0809 -0.0778 -0.0684 -0.0708 -0.0851 -0.0821 -0.0723 -0.0787 -0.0708 -0.0714 -0.0745
 -0.0754 -0.0711 -0.0766 -0.0836 -0.0729 -0.0763 -0.0818 -0.0800 -0.0784 -0.0757 -0.0848 -0.0803 -0.0757 -0.0748 -0.0845 -0.0824 -0.0732 -0.0803 -0.0711 -0.0745 -0.0739
 -0.0735 -0.0726 -0.0732 -0.0766 -0.0751 -0.0781 -0.0830 -0.0818 -0.0769 -0.0787 -0.0848 -0.0781 -0.0742 -0.0745 -0.0812 -0.0818 -0.0729 -0.0778 -0.0754 -0.0720 -0.0723
 -0.0735 -0.0717 -0.0751 -0.0824 -0.0812 -0.0797 -0.0861 -0.0824 -0.0803 -0.0775 -0.0854 -0.0821 -0.0745 -0.0732 -0.0845 -0.0830 -0.0726 -0.0803 -0.0751 -0.0757 -0.0720
 -0.0696 -0.0723 -0.0757 -0.0754 -0.0815 -0.0732 -0.0812 -0.0824 -0.0775 -0.0754 -0.0854 -0.0754 -0.0745 -0.0714 -0.0867 -0.0827 -0.0732 -0.0803 -0.0757 -0.0705 -0.0739
 0.0568 0.0281 -0.0275 -0.0436 -0.0586 -0.0366 0.0165 -0.0409 -0.0717 -0.0760 -0.0845 -0.0775 -0.0742 -0.0693 -0.0858 -0.0830 -0.0739 -0.0821 -0.0784 -0.0729 -0.0732
 -0.0470 0.0848 0.1245 0.2023 0.1379 0.0052 0.2209 0.1651 0.0595 -0.0668 -0.0824 -0.0732 -0.0726 -0.0754 -0.0824 -0.0800 -0.0745 -0.0800 -0.0732 -0.0739 -0.0726
 -0.0742 -0.0616 -0.0397 0.0076 0.1636 0.1840 0.3632 0.4947 0.6839 0.1077 0.0528 0.0790 -0.0037 -0.0732 -0.0836 -0.0818 -0.0732 -0.0803 -0.0717 -0.0708 -0.0729
 -0.0745 -0.0711 -0.0732 -0.0671 -0.0406 0.0446 0.1764 0.3564 1.2970 0.5920 0.2246 0.2917 0.2277 0.2655 0.0363 -0.0699 -0.0748 -0.0781 -0.0732 -0.0726 -0.0742
 -0.0729 -0.0720 -0.0735 -0.0806 -0.0735 -0.0339 0.0107 0.1123 0.4984 0.6879 0.3940 0.1642 0.0809 0.1001 0.1968 0.2277 0.1578 -0.0424 -0.0732 -0.0748 -0.0745
 -0.0723 -0.0726 -0.0772 -0.0839 -0.0748 -0.0732 -0.0778 -0.0174 0.0809 0.3207 0.4752 0.1981 0.0888 -0.0320 0.0278 0.0708 0.0891 0.0406 0.1053 -0.0693 -0.0742
 -0.0732 -0.0729 -0.0696 -0.0824 -0.0739 -0.0748 -0.0809 -0.0766 -0.0693 -0.0085 0.1529 0.2121 0.1392 -0.0714 -0.0812 -0.0772 -0.0375 0.0128 0.0153 0.0046 0.0821
 -0.0720 -0.0708 -0.0729 -0.0745 -0.0772 -0.0748 -0.0824 -0.0790 -0.0729 -0.0760 -0.0775 -0.0613 -0.0473 -0.0714 -0.0827 -0.0818 -0.0714 -0.0754 -0.0577 -0.0079 -0.0162
 -0.0720 -0.0714 -0.0757 -0.0806 -0.0760 -0.0757 -0.0833 -0.0800 -0.0729 -0.0748 -0.0830 -0.0833 -0.0729 -0.0708 -0.0851 -0.0815 -0.0720 -0.0793 -0.0763 -0.0708 -0.0674
 -0.0720 -0.0735 -0.0763 -0.0812 -0.0742 -0.0763 -0.0833 -0.0800 -0.0732 -0.0766 -0.0842 -0.0720 -0.0708 -0.0720 -0.0833 -0.0818 -0.0751 -0.0812 -0.0726 -0.0720 -0.0726
 -0.0714 -0.0745 -0.0739 -0.0793 -0.0763 -0.0757 -0.0800 -0.0766 -0.0751 -0.0766 -0.0815 -0.0745 -0.0714 -0.0720 -0.0833 -0.0827 -0.0739 -0.0797 -0.0720 -0.0723 -0.0732
 -0.0742 -0.0699 -0.0687 -0.0845 -0.0748 -0.0745 -0.0775 -0.0751 -0.0732 -0.0754 -0.0824 -0.0757 -0.0723 -0.0745 -0.0824 -0.0803 -0.0720 -0.0778 -0.0702 -0.0702 -0.0702
 -0.0723 -0.0684 -0.0723 -0.0818 -0.0717 -0.0717 -0.0757 -0.0748 -0.0690 -0.0757 -0.0818 -0.0787 -0.0729 -0.0726 -0.0821 -0.0809 -0.0720 -0.0751 -0.0702 -0.0723 -0.0699
Attachment 4: OMEGA_OM-DAQPRO-5300_19-Jun-17_10.29.48.dlp
Attachment 5: Ta#52.docx
Attachment 6: TS-2017-06-09-14-20.rga
Attachment 7: TS-2017-06-12-11-34.rga
Attachment 8: ITE-TM4-Ta#52-LP-SIS_Test_Stand.docx
  595   Thursday, July 06, 2017, 13:16 Aaron SchmidtTa#55LP-SIS  42218-1142738-2In progressConditioning Ta spare for Ta#52

Thursday, July 06, 2017, 13:56: Roughing.

Thursday, July 06, 2017, 14:25: Turbo  pumps started.

Thursday, July 06, 2017, 14:30: Ion gauge on. RGA Started


Friday, July 07, 2017, 07:29: Waterlines opened. No response on RGA. Vacuum ok @ 1.97e-07T

Friday, July 07, 2017, 07:33: Power/sources on.

Friday, July 07, 2017, 07:39: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .

Friday, July 07, 2017, 07:40: IV4 Opened.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 26.0 3.78
 TBHT 26.5 3.02
 COIL + EE 25.8 4.86
 TGHT  26.0 5.47

Friday, July 07, 2017, 07:42: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.

Friday, July 07, 2017, 07:45: Optics reset.

Friday, July 07, 2017, 14:28: Anode brought up to 5, 10 15, and 50 V using FEBIAD controls. No Current detected.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 36.5 3.75
 TBHT 27.6 3.04
 COIL + EE 26.5 4.87
 TGHT  28.7 5.54

 


Saturday, July 08, 2017, 08:58: Ramping down heaters


Monday, July 10, 2017, 08:15: Power/Sources off. *BIAS 30 to 0 kV due to shut off. EE & EZL Brought to 0.

Monday, July 10, 2017, 08:23: Water lines purged. IV4 Closed

Monday, July 10, 2017, 08:27: IV4 reset.

Monday, July 10, 2017, 09:40: RGA off. Ion Gauge off. Turbos off.

Monday, July 10, 2017, 11:02: Venting.

Attachment 1: ER-580-200-30kV-m23.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1459 -0.0992 -0.0943 -0.0925 -0.0934 -0.1016 -0.0986 -0.0839 -0.0732 -0.0800 -0.0766 -0.0739 -0.0772 -0.0876 -0.0931 -0.0876 -0.0989 -0.0861 -0.0934 -0.0989 -0.0958
 -0.0903 -0.0943 -0.0906 -0.0906 -0.0879 -0.0986 -0.0931 -0.0797 -0.0793 -0.0821 -0.0723 -0.0693 -0.0726 -0.0848 -0.0888 -0.0818 -0.0867 -0.0803 -0.0882 -0.0946 -0.0897
 -0.0897 -0.0925 -0.0867 -0.0821 -0.0854 -0.0977 -0.0916 -0.0793 -0.0842 -0.0882 -0.0720 -0.0687 -0.0751 -0.0821 -0.0879 -0.0839 -0.0861 -0.0775 -0.0793 -0.0815 -0.0797
 -0.0882 -0.0903 -0.0793 -0.0937 -0.0769 -0.0977 -0.0949 -0.0806 -0.0778 -0.0858 -0.0751 -0.0729 -0.0867 -0.0748 -0.0864 -0.0867 -0.0760 -0.0781 -0.0800 -0.0784 -0.0772
 -0.0876 -0.0879 -0.0787 -0.0958 -0.0787 -0.0986 -0.0952 -0.0812 -0.0821 -0.0848 -0.0797 -0.0787 -0.0723 -0.0787 -0.0851 -0.0870 -0.0790 -0.0787 -0.0787 -0.0775 -0.0781
 -0.0858 -0.0888 -0.0781 -0.0748 -0.0720 -0.0970 -0.0961 -0.0809 -0.0897 -0.0845 -0.0793 -0.0787 -0.0769 -0.0778 -0.0839 -0.0836 -0.0757 -0.0775 -0.0769 -0.0778 -0.0778
 -0.0885 -0.0909 -0.0769 -0.0912 0.0586 -0.0385 -0.0851 -0.0790 -0.0778 -0.0870 -0.0790 -0.0790 -0.0864 -0.0766 -0.0851 -0.0812 -0.0757 -0.0760 -0.0787 -0.0751 -0.0781
 -0.0888 -0.0912 -0.0775 -0.0833 -0.0034 0.2264 0.0839 -0.0494 -0.0876 -0.0891 -0.0797 -0.0790 -0.0778 -0.0797 -0.0851 -0.0830 -0.0790 -0.0784 -0.0790 -0.0766 -0.0769
 -0.0717 -0.0797 -0.0766 -0.0916 -0.0726 0.3128 0.6992 0.1498 -0.0574 -0.0739 -0.0732 -0.0790 -0.0790 -0.0760 -0.0848 -0.0836 -0.0787 -0.0800 -0.0775 -0.0769 -0.0778
 -0.0720 -0.0684 -0.0549 -0.0714 -0.0342 -0.0745 1.2405 0.6815 0.1675 -0.0424 -0.0638 -0.0699 -0.0739 -0.0790 -0.0854 -0.0821 -0.0803 -0.0781 -0.0803 -0.0775 -0.0797
 -0.0623 -0.0571 -0.0558 -0.0565 -0.0299 -0.0323 0.2798 0.8789 1.0532 0.0723 -0.0464 -0.0595 -0.0674 -0.0699 -0.0769 -0.0815 -0.0793 -0.0772 -0.0800 -0.0781 -0.0775
 -0.0879 -0.0882 -0.0211 -0.0644 -0.0330 -0.0381 0.0974 0.2484 1.1102 0.7678 0.0488 -0.0729 -0.0705 -0.0690 -0.0763 -0.0720 -0.0647 -0.0754 -0.0760 -0.0751 -0.0766
 -0.0894 -0.0903 -0.0778 -0.0952 -0.0354 -0.0589 0.0488 0.0494 0.2368 1.1792 0.4474 0.0125 -0.0751 -0.0732 -0.0827 -0.0790 -0.0705 -0.0629 -0.0745 -0.0763 -0.0772
 -0.0879 -0.0903 -0.0787 -0.0970 -0.0760 -0.0934 -0.0165 0.0262 0.0156 0.3046 0.8878 0.2240 -0.0699 -0.0729 -0.0858 -0.0818 -0.0763 -0.0717 -0.0699 -0.0546 -0.0769
 -0.0870 -0.0900 -0.0766 -0.0888 -0.0787 -0.0961 -0.0925 -0.0714 -0.0507 0.0174 0.2612 0.5963 0.1871 -0.0787 -0.0854 -0.0815 -0.0778 -0.0787 -0.0742 -0.0684 -0.0671
 -0.0888 -0.0897 -0.0760 -0.0909 -0.0757 -0.0961 -0.0912 -0.0745 -0.0891 -0.0720 0.0137 0.1895 0.3024 -0.0314 -0.0827 -0.0818 -0.0766 -0.0787 -0.0781 -0.0790 -0.0693
 -0.0879 -0.0897 -0.0766 -0.0900 -0.0754 -0.0919 -0.0885 -0.0754 -0.0790 -0.0827 -0.0723 -0.0687 -0.0305 -0.0381 -0.0821 -0.0806 -0.0766 -0.0757 -0.0763 -0.0742 -0.0769
 -0.0885 -0.0888 -0.0766 -0.0897 -0.0745 -0.0879 -0.0873 -0.0757 -0.0864 -0.0827 -0.0775 -0.0769 -0.0751 -0.0766 -0.0842 -0.0836 -0.0766 -0.0757 -0.0778 -0.0763 -0.0757
 -0.0900 -0.0909 -0.0784 -0.0970 -0.0751 -0.0870 -0.0858 -0.0751 -0.0912 -0.0806 -0.0790 -0.0793 -0.0787 -0.0769 -0.0848 -0.0833 -0.0778 -0.0781 -0.0769 -0.0760 -0.0781
 -0.0864 -0.0900 -0.0757 -0.0961 -0.0757 -0.0864 -0.0809 -0.0696 -0.0751 -0.0821 -0.0772 -0.0793 -0.0787 -0.0760 -0.0836 -0.0818 -0.0790 -0.0778 -0.0787 -0.0778 -0.0772
 -0.0858 -0.0894 -0.0748 -0.0934 -0.0739 -0.0848 -0.0787 -0.0629 -0.0809 -0.0778 -0.0769 -0.0775 -0.0775 -0.0772 -0.0818 -0.0815 -0.0763 -0.0763 -0.0769 -0.0772 -0.0781
Attachment 2: ER-580-200-30kV-m38.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1575 -0.1071 -0.1016 -0.1065 -0.1062 -0.1035 -0.1004 -0.0903 -0.0760 -0.0363 -0.0461 -0.0601 -0.0778 -0.0818 -0.0949 -0.0995 -0.0931 -0.0916 -0.1071 -0.1041 -0.0958
 -0.0970 -0.1013 -0.0980 -0.1010 -0.1016 -0.1001 -0.0970 -0.0812 -0.0861 -0.0699 -0.0449 -0.0555 -0.0717 -0.0787 -0.0937 -0.0961 -0.0854 -0.0867 -0.1013 -0.0983 -0.0897
 -0.0952 -0.0986 -0.0851 -0.0897 -0.0955 -0.0977 -0.0983 -0.0803 -0.0836 -0.1077 -0.0610 -0.0546 -0.0717 -0.0772 -0.0894 -0.0958 -0.0864 -0.0821 -0.0961 -0.0861 -0.0809
 -0.0964 -0.0977 -0.0775 -0.0858 -0.0739 -0.0974 -0.0989 -0.0818 -0.0787 -0.1105 -0.0882 -0.0601 -0.0775 -0.0739 -0.0882 -0.0958 -0.0861 -0.0815 -0.0806 -0.0772 -0.0778
 -0.1007 -0.1004 -0.0800 -0.0861 -0.0845 -0.0989 -0.0992 -0.0848 -0.0861 -0.1050 -0.1031 -0.0732 -0.0757 -0.0751 -0.0867 -0.0967 -0.0854 -0.0815 -0.0821 -0.0797 -0.0787
 -0.0995 -0.0986 -0.0790 -0.0839 -0.0821 -0.0989 -0.0989 -0.0854 -0.0864 -0.1059 -0.0977 -0.0821 -0.0803 -0.0793 -0.0894 -0.0970 -0.0848 -0.0806 -0.0797 -0.0797 -0.0778
 -0.0964 -0.0974 -0.0735 -0.0873 -0.0375 -0.0601 -0.0922 -0.0870 -0.0867 -0.1074 -0.0952 -0.0809 -0.0812 -0.0803 -0.0894 -0.0992 -0.0845 -0.0809 -0.0809 -0.0757 -0.0772
 -0.0961 -0.0955 -0.0742 -0.0452 -0.0705 0.0897 0.0931 -0.0305 -0.0882 -0.1099 -0.0922 -0.0793 -0.0797 -0.0763 -0.0894 -0.0989 -0.0818 -0.0809 -0.0809 -0.0760 -0.0769
 -0.0134 -0.0549 -0.0696 -0.0800 -0.0336 0.0757 0.4651 0.6528 0.0015 -0.0769 -0.0827 -0.0784 -0.0775 -0.0742 -0.0861 -0.0989 -0.0858 -0.0845 -0.0821 -0.0821 -0.0775
 0.0378 0.0250 0.0470 0.0342 0.1025 -0.0128 0.2185 1.8475 1.6077 -0.0134 -0.0391 -0.0351 -0.0470 -0.0769 -0.0882 -0.0961 -0.0867 -0.0827 -0.0806 -0.0784 -0.0781
 -0.0510 0.0760 0.0638 0.0986 0.2078 0.3448 0.0732 0.6653 3.9651 1.0074 0.0226 -0.0113 -0.0189 -0.0146 -0.0632 -0.0958 -0.0836 -0.0821 -0.0809 -0.0766 -0.0797
 -0.0961 -0.0946 0.0452 0.0372 0.1334 0.1855 0.2502 0.3159 2.5229 6.0635 0.1724 -0.0327 -0.0235 -0.0327 -0.0494 -0.0406 -0.0385 -0.0790 -0.0748 -0.0751 -0.0775
 -0.0958 -0.0964 -0.0726 -0.0839 0.0092 0.0116 0.1544 0.2350 0.5399 4.1684 3.1567 0.1071 -0.0070 -0.0229 -0.0632 -0.0671 -0.0507 -0.0302 -0.0647 -0.0745 -0.0760
 -0.0974 -0.0980 -0.0784 -0.0873 -0.0800 -0.0888 0.0009 0.1086 0.1468 0.9015 3.8147 1.2885 0.1019 -0.0085 -0.0812 -0.0833 -0.0699 -0.0494 -0.0470 0.0104 -0.0754
 -0.0986 -0.0970 -0.0797 -0.0888 -0.0836 -0.0958 -0.0937 -0.0748 -0.0476 0.0757 0.7455 1.4111 0.2850 -0.0409 -0.0198 -0.0919 -0.0806 -0.0729 -0.0549 -0.0415 -0.0406
 -0.0980 -0.0974 -0.0778 -0.0882 -0.0827 -0.0946 -0.0943 -0.0790 -0.0751 -0.0818 0.0134 0.2280 0.3674 0.0885 -0.0677 -0.0916 -0.0812 -0.0793 -0.0766 -0.0671 -0.0412
 -0.0958 -0.0986 -0.0766 -0.0867 -0.0803 -0.0937 -0.0925 -0.0803 -0.0778 -0.0916 -0.0754 -0.0659 -0.0241 0.0232 -0.0778 -0.0897 -0.0803 -0.0787 -0.0781 -0.0751 -0.0745
 -0.0949 -0.0955 -0.0763 -0.0909 -0.0769 -0.0922 -0.0912 -0.0781 -0.0751 -0.0940 -0.0775 -0.0726 -0.0754 -0.0745 -0.0851 -0.0940 -0.0821 -0.0806 -0.0757 -0.0745 -0.0754
 -0.0946 -0.0980 -0.0793 -0.0851 -0.0803 -0.0934 -0.0894 -0.0726 -0.0732 -0.0952 -0.0778 -0.0748 -0.0754 -0.0748 -0.0851 -0.0946 -0.0836 -0.0793 -0.0787 -0.0763 -0.0769
 -0.0958 -0.0967 -0.0772 -0.0864 -0.0812 -0.0943 -0.0897 -0.0668 -0.0638 -0.0964 -0.0800 -0.0739 -0.0763 -0.0775 -0.0858 -0.0922 -0.0836 -0.0787 -0.0769 -0.0754 -0.0769
 -0.0955 -0.0937 -0.0751 -0.0864 -0.0790 -0.0912 -0.0876 -0.0601 -0.0464 -0.0861 -0.0790 -0.0763 -0.0766 -0.0757 -0.0854 -0.0906 -0.0824 -0.0790 -0.0754 -0.0760 -0.0781
Attachment 3: ER-580-200-30kV-m39.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1590 -0.1062 -0.1086 -0.0906 -0.0906 -0.1022 -0.1093 -0.1038 -0.0885 -0.0534 -0.0565 -0.0681 -0.0763 -0.0925 -0.1007 -0.1041 -0.1007 -0.0952 -0.1074 -0.1068 -0.1041
 -0.1016 -0.1031 -0.1047 -0.0867 -0.0873 -0.0998 -0.1019 -0.0995 -0.0922 -0.0665 -0.0510 -0.0604 -0.0708 -0.0894 -0.0949 -0.1001 -0.0940 -0.0928 -0.1013 -0.1004 -0.0989
 -0.1007 -0.0989 -0.0916 -0.0845 -0.0827 -0.0983 -0.1016 -0.0992 -0.0867 -0.0845 -0.0607 -0.0586 -0.0677 -0.0867 -0.0946 -0.0995 -0.0934 -0.0851 -0.0998 -0.0943 -0.0848
 -0.0986 -0.0970 -0.0763 -0.0781 -0.0751 -0.0919 -0.1038 -0.0992 -0.0717 -0.0867 -0.0726 -0.0647 -0.0650 -0.0717 -0.0928 -0.1019 -0.0870 -0.0888 -0.0800 -0.0790 -0.0842
 -0.0992 -0.0989 -0.0790 -0.0760 -0.0745 -0.0922 -0.1047 -0.1001 -0.0751 -0.0830 -0.0787 -0.0760 -0.0723 -0.0717 -0.0940 -0.1031 -0.0867 -0.0891 -0.0824 -0.0812 -0.0803
 -0.0989 -0.0989 -0.0790 -0.0867 -0.0757 -0.0946 -0.1056 -0.0964 -0.0720 -0.0870 -0.0787 -0.0775 -0.0772 -0.0772 -0.0934 -0.1028 -0.0827 -0.0903 -0.0790 -0.0787 -0.0763
 -0.1004 -0.0986 -0.0790 -0.0790 -0.0241 -0.0101 -0.0891 -0.0931 -0.0351 -0.0485 -0.0778 -0.0745 -0.0778 -0.0803 -0.0943 -0.1007 -0.0790 -0.0867 -0.0763 -0.0766 -0.0766
 -0.0809 -0.0986 -0.0763 -0.0790 -0.0714 0.3131 0.1181 -0.0345 0.0058 0.0626 -0.0830 -0.0748 -0.0748 -0.0766 -0.0903 -0.1010 -0.0809 -0.0873 -0.0806 -0.0790 -0.0772
 -0.0894 -0.0177 -0.0507 -0.0848 -0.0745 0.0546 0.5795 0.3409 0.0809 0.1352 -0.0708 -0.0778 -0.0803 -0.0760 -0.0894 -0.1028 -0.0824 -0.0900 -0.0784 -0.0790 -0.0784
 -0.0922 -0.0919 -0.0690 -0.0726 -0.0629 -0.0781 0.4315 1.1215 0.9171 0.2142 0.0015 -0.0677 -0.0769 -0.0787 -0.0900 -0.1031 -0.0815 -0.0888 -0.0812 -0.0787 -0.0781
 -0.0995 -0.0888 -0.0693 -0.0781 -0.0705 -0.0766 -0.0616 0.5667 1.8753 1.0110 0.1111 -0.0580 -0.0735 -0.0763 -0.0916 -0.1010 -0.0781 -0.0864 -0.0775 -0.0769 -0.0778
 -0.1001 -0.0964 -0.0726 -0.0742 -0.0717 -0.0769 -0.0699 0.0687 1.0928 3.1250 0.4935 -0.0278 -0.0681 -0.0711 -0.0888 -0.0983 -0.0754 -0.0845 -0.0800 -0.0754 -0.0763
 -0.1007 -0.0995 -0.0775 -0.0809 -0.0754 -0.0787 -0.0769 -0.0305 0.1749 1.6907 1.8475 0.2292 -0.0671 -0.0705 -0.0903 -0.0974 -0.0751 -0.0861 -0.0760 -0.0769 -0.0745
 -0.0986 -0.0986 -0.0769 -0.0824 -0.0775 -0.0888 -0.0916 -0.0552 0.0089 0.3650 1.8930 1.0120 0.1450 -0.0696 -0.0919 -0.0989 -0.0790 -0.0861 -0.0815 -0.0760 -0.0778
 -0.0986 -0.0980 -0.0754 -0.0809 -0.0751 -0.0916 -0.1025 -0.0925 -0.0546 0.0333 0.3699 0.9113 0.5783 -0.0540 -0.0885 -0.1007 -0.0809 -0.0864 -0.0772 -0.0739 -0.0766
 -0.0995 -0.0992 -0.0754 -0.0812 -0.0769 -0.0916 -0.1007 -0.0952 -0.0708 -0.0702 0.0208 0.2255 0.5313 0.2142 -0.0876 -0.0977 -0.0778 -0.0842 -0.0790 -0.0745 -0.0723
 -0.1004 -0.1010 -0.0766 -0.0812 -0.0775 -0.0894 -0.1010 -0.0958 -0.0745 -0.0848 -0.0656 -0.0620 -0.0168 0.0571 -0.0797 -0.0970 -0.0772 -0.0867 -0.0757 -0.0748 -0.0769
 -0.0986 -0.1001 -0.0769 -0.0891 -0.0803 -0.0876 -0.0992 -0.0952 -0.0784 -0.0815 -0.0702 -0.0726 -0.0760 -0.0732 -0.0891 -0.0992 -0.0784 -0.0851 -0.0763 -0.0769 -0.0763
 -0.0970 -0.0970 -0.0739 -0.0839 -0.0763 -0.0876 -0.0943 -0.0922 -0.0763 -0.0806 -0.0717 -0.0739 -0.0769 -0.0766 -0.0919 -0.1010 -0.0781 -0.0864 -0.0784 -0.0772 -0.0769
 -0.0974 -0.0958 -0.0751 -0.0845 -0.0769 -0.0867 -0.0919 -0.0864 -0.0705 -0.0803 -0.0726 -0.0754 -0.0763 -0.0772 -0.0897 -0.1010 -0.0790 -0.0864 -0.0766 -0.0778 -0.0772
 -0.0989 -0.0967 -0.0757 -0.0839 -0.0775 -0.0867 -0.0916 -0.0821 -0.0635 -0.0745 -0.0708 -0.0729 -0.0766 -0.0763 -0.0916 -0.0986 -0.0781 -0.0861 -0.0797 -0.0775 -0.0760
Attachment 4: ER-580-200-30kV-m133.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1505 -0.0983 -0.0916 -0.0909 -0.0943 -0.0958 -0.0989 -0.0854 -0.0821 -0.0665 -0.0467 -0.0623 -0.0763 -0.0919 -0.1028 -0.1016 -0.0998 -0.0983 -0.1074 -0.1065 -0.1059
 -0.0885 -0.0925 -0.0885 -0.0870 -0.0931 -0.0916 -0.0946 -0.0781 -0.0815 -0.0836 -0.0473 -0.0568 -0.0702 -0.0854 -0.0983 -0.0946 -0.0974 -0.0961 -0.1035 -0.1035 -0.0986
 -0.0885 -0.0903 -0.0815 -0.0830 -0.0845 -0.0903 -0.0916 -0.0757 -0.0790 -0.0909 -0.0604 -0.0549 -0.0671 -0.0793 -0.0977 -0.0946 -0.0949 -0.0912 -0.1022 -0.0989 -0.0864
 -0.0861 -0.0873 -0.0775 -0.0803 -0.0797 -0.0836 -0.0943 -0.0800 -0.0766 -0.0848 -0.0800 -0.0650 -0.0650 -0.0714 -0.0955 -0.0946 -0.0839 -0.0885 -0.0946 -0.0757 -0.0757
 -0.0864 -0.0870 -0.0751 -0.0793 -0.0742 -0.0842 -0.0946 -0.0775 -0.0797 -0.0812 -0.0839 -0.0790 -0.0699 -0.0757 -0.0952 -0.0934 -0.0876 -0.0885 -0.0812 -0.0757 -0.0769
 -0.0903 -0.0885 -0.0760 -0.0891 -0.0735 -0.0854 -0.0937 -0.0781 -0.0812 -0.0827 -0.0778 -0.0790 -0.0784 -0.0824 -0.0974 -0.0946 -0.0861 -0.0897 -0.0876 -0.0760 -0.0766
 -0.0894 -0.0882 -0.0772 -0.0851 -0.0436 -0.0208 -0.0787 -0.0797 -0.0815 -0.0827 -0.0784 -0.0800 -0.0818 -0.0842 -0.0974 -0.0946 -0.0845 -0.0885 -0.0818 -0.0784 -0.0751
 -0.0873 -0.0861 -0.0775 -0.0775 -0.0751 0.2106 0.1611 0.0058 -0.0708 -0.0824 -0.0784 -0.0784 -0.0793 -0.0839 -0.0980 -0.0949 -0.0870 -0.0897 -0.0870 -0.0787 -0.0800
 -0.0873 -0.0858 -0.0729 -0.0864 -0.0742 -0.0360 0.5655 0.5167 0.0711 -0.0668 -0.0778 -0.0781 -0.0775 -0.0812 -0.0967 -0.0940 -0.0845 -0.0882 -0.0821 -0.0790 -0.0793
 -0.0854 -0.0861 -0.0723 -0.0778 -0.0735 -0.0790 0.1804 1.1356 1.1194 0.0473 -0.0671 -0.0751 -0.0748 -0.0803 -0.0980 -0.0946 -0.0806 -0.0873 -0.0824 -0.0781 -0.0772
 -0.0867 -0.0815 -0.0748 -0.0824 -0.0702 -0.0729 -0.0751 0.3769 2.1054 1.6153 0.0116 -0.0620 -0.0739 -0.0827 -0.0977 -0.0940 -0.0845 -0.0894 -0.0830 -0.0745 -0.0751
 -0.0897 -0.0897 -0.0708 -0.0748 -0.0717 -0.0739 -0.0720 0.0027 0.6451 3.8977 0.9238 -0.0403 -0.0723 -0.0800 -0.0970 -0.0922 -0.0812 -0.0879 -0.0833 -0.0757 -0.0742
 -0.0873 -0.0885 -0.0751 -0.0897 -0.0717 -0.0742 -0.0760 -0.0336 0.1282 1.8280 3.6035 0.3409 -0.0598 -0.0797 -0.0961 -0.0928 -0.0809 -0.0867 -0.0793 -0.0766 -0.0793
 -0.0891 -0.0870 -0.0757 -0.0842 -0.0766 -0.0818 -0.0803 -0.0497 -0.0037 0.2542 2.2183 1.9345 0.1843 -0.0778 -0.0964 -0.0928 -0.0830 -0.0882 -0.0830 -0.0772 -0.0790
 -0.0891 -0.0882 -0.0748 -0.0916 -0.0778 -0.0815 -0.0906 -0.0742 -0.0574 0.0226 0.3580 1.1969 0.9808 0.0894 -0.0970 -0.0943 -0.0800 -0.0848 -0.0824 -0.0760 -0.0763
 -0.0894 -0.0858 -0.0754 -0.0769 -0.0739 -0.0803 -0.0894 -0.0772 -0.0748 -0.0653 0.0101 0.2203 0.6061 0.4141 -0.0912 -0.0922 -0.0784 -0.0861 -0.0824 -0.0763 -0.0748
 -0.0906 -0.0864 -0.0754 -0.0876 -0.0742 -0.0797 -0.0894 -0.0760 -0.0742 -0.0726 -0.0687 -0.0598 0.0128 0.1819 -0.0031 -0.0946 -0.0833 -0.0873 -0.0836 -0.0800 -0.0760
 -0.0885 -0.0882 -0.0757 -0.0879 -0.0754 -0.0800 -0.0885 -0.0772 -0.0748 -0.0742 -0.0720 -0.0751 -0.0720 -0.0745 -0.0858 -0.0952 -0.0809 -0.0891 -0.0854 -0.0769 -0.0769
 -0.0867 -0.0861 -0.0772 -0.0815 -0.0748 -0.0797 -0.0876 -0.0745 -0.0711 -0.0757 -0.0754 -0.0742 -0.0739 -0.0793 -0.0961 -0.0937 -0.0800 -0.0848 -0.0815 -0.0766 -0.0760
 -0.0854 -0.0858 -0.0751 -0.0812 -0.0735 -0.0775 -0.0864 -0.0714 -0.0671 -0.0732 -0.0766 -0.0751 -0.0766 -0.0800 -0.0967 -0.0925 -0.0815 -0.0876 -0.0848 -0.0754 -0.0739
 -0.0885 -0.0864 -0.0757 -0.0800 -0.0742 -0.0772 -0.0842 -0.0684 -0.0616 -0.0653 -0.0729 -0.0778 -0.0778 -0.0824 -0.0967 -0.0925 -0.0824 -0.0858 -0.0809 -0.0748 -0.0760
Attachment 5: TS-2017-07-06-14-30.rga
Attachment 6: TS-2017-07-07-10-10.rga
Attachment 7: TS-2017-07-07-12-50.rga
Attachment 8: Spare_Ta_LP_SIS.docx
Attachment 9: OMEGA_OM-DAQPRO-5300_10-Jul-17_10.27.49.dlp
Attachment 10: ITE-TM4-Ta#(Spare)-LP-SIS_Test_Stand.docx
  602   Tuesday, August 15, 2017, 08:41 Aaron SchmidtTa#55LP-SIS  42218-1142738-2In progressTBHT sight in test

Heat TGHT to 580 A & TGHT to  200 A for view port test.

Tuesday, August 15, 2017, 08:40: Roughing.

Tuesday, August 15, 2017, 08:59: Both Turbo pumps started.

Tuesday, August 15, 2017, 09:05: Ion gauge on. TP1B @ speed 42 KRPM.

Tuesday, August 15, 2017, 09:07: TP1A @ speed 38 KRPM.

Tuesday, August 15, 2017, 09:10: RGA on. Vacuum @ 9.14e-06T

Tuesday, August 15, 2017, 11:40: Check. Vacuum @ 6.76e-07T

Tuesday, August 15, 2017, 11:41: Waterlines opened. No response on RGA. Vacuum @ 6.49e-07T

Tuesday, August 15, 2017, 11:43: Power/sources on.

Tuesday, August 15, 2017, 13:57: Check. Vacuum @ 3.88e-07T.

Tuesday, August 15, 2017, 14:33: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.


Wednesday, August 16, 2017, 10:33: Auto-cooling TBHT to 100 A @ 4A per min.


Thursday, August 17, 2017, 07:05: TGHT / TBHT off. FC0 closed. IV4 Closed.

Thursday, August 17, 2017, 07:11: IV4 reset.

Thursday, August 17, 2017, 07:13: Power off

Thursday, August 17, 2017, 07:15 Waterlines purged


Thursday, August 24, 2017, 13:03: Waterlines ok (open). Power/source on.

Thursday, August 24, 2017, 13:10: Auto ramping TGHT to 580 A @ 2A per min.


Friday, August 25, 2017, 07:59: Open IV4. Open TMS : FC0.


Friday, September 08, 2017, 06:51: Auto ramping TGHT to 580 A @ 2A per min.


Monday, September 11, 2017, 10:55: Close TMS : FC0. Close IV4

Monday, September 11, 2017, 11:58: power/sources off

Monday, September 11, 2017, 11:00: RGA off

Monday, September 11, 2017, 11:01: Ion gauge, turbos off.

Monday, September 11, 2017, 11:59: Venting.

Monday, September 11, 2017, 13:17: Roughing.

Monday, September 11, 2017, 13:34: Turbo pumps on

Monday, September 11, 2017, 13:40: Ion gauge on.


Tuesday, September 12, 2017, 07:24: RGA, power, & sources on.

Tuesday, September 12, 2017, 07:25: Auto ramping TGHT to 580 A @ 2A per min.

Tuesday, September 12, 2017, 07:26: Open IV4. Open TMS : FC0.


Monday, September 18, 2017, 10:53: Close TMS : FC0. Close IV4

Monday, September 18, 2017, 10:57: Power/sources off.


Wednesday, September 20, 2017, 10:25: RGA, Ion gauge, Turbo pumps off. Waterlines purged.

Wednesday, September 20, 2017, 12:49: Venting.

  615   Friday, December 01, 2017, 14:31 Aaron Schmidt LP-SIS  (Blank)42752-3In progressSilver plated bolts test / Outgas TS

Chamber 'open state, no turbos' 2 weeks. Testing Silver plated 3/8"-16x1" (SCR, SHC, 18-8, AG) bolt on the TBHT.

Friday, December 01, 2017, 14:26: Roughing. BV1B open.

Friday, December 01, 2017, 14:27: BV1A open

Friday, December 01, 2017, 14:47: Turbo pumps on.

Friday, December 01, 2017, 14:53: Ion gauge on (Vacuum @ 1.44 e-04T)

Friday, December 01, 2017, 14:57: Turbos @ speed. RGA on.


Monday, December 04, 2017, 09:27: System rebooted after weekend power outage.


Tuesday, December 05, 2017, 08:10: RGA on.

Tuesday, December 05, 2017, 08:17: Waterlines opened. Vacuum ok @ 1.56e-06T

Tuesday, December 05, 2017, 08:20: Power on. (Power supplies in off position) Turn power off

Open back. switch power supplies to on.

Tuesday, December 05, 2017, 08:35: Power on

Tuesday, December 05, 2017, 08:37: Auto ramping TGHT / TBHT to 150 / 75 A @ 2A per min.

Tuesday, December 05, 2017, 09:55: Auto ramping TGHT / TBHT to 200 / 75 A @ 2A per min.

Tuesday, December 05, 2017, 10:28: Auto ramping TGHT / TBHT to 400 / 150 A @ 2A per min. Vacuum @ 1.56e-06T.

Tuesday, December 05, 2017, 12:50: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min. Vacuum @ 3.03e-06T.

Tuesday, December 05, 2017, 12:58: Auto ramping halted. Vacuum @ 3.12e-06T. Allow vacuum to improve.


Wednesday, December 06, 2017, 08:03: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min. Vacuum @ 2.65e-06T. Added note*** The Silver bolts are NOT vented ***.

Wednesday, December 06, 2017, 08:30: TGHT / TBHT @ 449 / 199 A. Vacuum @ 3.03e-06T and holding. Auto-ramping continued. No halt.

Wednesday, December 06, 2017, 10:45: Check. TGHT / TBHT @ 449 / 199 A. Vacuum @ 3.36e-06T and holding. Auto-ramping on wait.

Wednesday, December 06, 2017, 14:37: Check. TGHT / TBHT @ 449 / 199 A. Vacuum @ 3.21e-06T and holding. Auto-ramping halted. Allow Vacuum to improve.


Thursday, December 07, 2017, 07:52: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min. Vacuum @ 2.33e-06T.

Thursday, December 07, 2017, 09:00: Check. TGHT / TBHT @ 571 / 200 A. Vacuum @ 3.19e-06T. Auto-ramping halted. Allow Vacuum to improve.


Friday, December 08, 2017, 07:52: Auto ramping TGHT to 580 A from 571 A @ 2A per min. Vacuum @ 2.04e-06T.


Monday, December 11, 2017, 08:00: Auto cooling @ 2A per min.

Monday, December 11, 2017, 13:45: Sources/power off. Water cooling off.


Tuesday, December 12, 2017, 06:35: RGA, Ion gauge, turbo pumps off.

Tuesday, December 12, 2017, 07:45: Venting.

Tuesday, December 12, 2017, 08:23: Doors open.

Attachment 1: 1st_silver_bolts_test.docx
Attachment 2: TS-2017-12-05-08-10.rga
  617   Tuesday, December 12, 2017, 08:32 Aaron Schmidt LP-SIS  (Blank)42752-3DonePump down w/ possible outgas

Check achievable vacuum. Watch for spikes on RGA. Silver bolts removed from the chamber. Vented steel screws in.

Tuesday, December 12, 2017, 08:23: Doors open from previous test. (Open for 7 minutes)

Tuesday, December 12, 2017, 08:30: Doors closed. Roughing.

Tuesday, December 12, 2017, 08:50: Turbo pumps on.

Tuesday, December 12, 2017, 08:56: Ion gauge on.


Wednesday, December 13, 2017, 06:46: Check. Vacuum @ 5.44e-07T. RGA Error.

Wednesday, December 13, 2017, 06:49: RGA restarted.

Wednesday, December 13, 2017, 06:52: Water lines opened. Vacuum @ 5.44e-07T

Wednesday, December 13, 2017, 06:54: Power on.

Wednesday, December 13, 2017, 06:56: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.


Friday, December 15, 2017, 13:37: Auto-cooling @ 2A per min


Monday, December 18, 2017, 08:15: Power/source off.


Wednesday, December 20, 2017, 09:07: RGA off.

Wednesday, December 20, 2017, 09:08: Both Ion gauges and all 4 turbo pumps off.

Attachment 1: Empty_Chamber_-_Silver_bolts_removed.docx
Attachment 2: TS-2017-12-12-09-00.rga
Attachment 3: TS-2017-12-13-06-50.rga
  618   Wednesday, December 20, 2017, 09:11 Aaron Schmidt LP-SIS  (Blank)42752-3DoneHoliday Shut Down

Power/sources already off. The water cooling is currently left on.

Wednesday, December 20, 2017, 09:07: RGA off.

Wednesday, December 20, 2017, 09:08: Both Ion gauges and all 4 turbo pumps off.

Wednesday, December 20, 2017, 11:21: Valves closed. Roughing pumps off.

Wednesday, December 20, 2017, 11:22: RGA shut off completely.

  623   Friday, January 26, 2018, 13:59 Aaron Schmidt LP-SIS  (Blank)42752-3DoneHV test 1

Friday, January 26, 2018, 14:00: BIAS/EZL/EE ramp up.

Friday, January 26, 2018, 14:06: BIAS/EZL/EE @ 30/13/1.5 kV.

Friday, January 26, 2018, 14:09: BIAS/EZL/EE @ 35/13/1.5 kV.

Spark at BIAS 30 kV and 40kV

Friday, January 26, 2018, 14:12: BIAS/EZL/EE @ 40/13/1.5 kV.

Friday, January 26, 2018, 14:15: BIAS/EZL/EE @ 45/13/1.5 kV.

Audible sparks, but none showing on the strip tool.

Friday, January 26, 2018, 14:19: BIAS/EZL/EE @ 45/19/1.5 kV.

Friday, January 26, 2018, 14:22: BIAS/EZL/EE @ 47/22/1.5 kV. Sparks @ 47000 to 47250 V.

Sparks observed on EE feed through in cage.

Friday, January 26, 2018, 14:27: BIAS/EZL/EE @ 50/22/1.7 kV.

Friday, January 26, 2018, 14:39: BIAS/EZL/EE @ 55/23.5/1.8 kV.

Bias controls 55 kV Actual @ 50 kV.

Friday, January 26, 2018, 14:44: Spark free 5 min.

Friday, January 26, 2018, 14:49: Spark free 10 min.

Attachment 1: 2018_HV_test_1.docx
  624   Monday, January 29, 2018, 10:07 Aaron Schmidt LP-SIS  (Blank)42752-3DonePower outage / HV test 2

HV test 2 cancelled.  9:38: BIAS/EZL/EE @ 30/13/1.5 kV

Monday, January 29, 2018, 09:40 (Roughly) Power went out completely

Monday, January 29, 2018, 10:00: Power up TS

Monday, January 29, 2018, 10:07:Turbos on.

Monday, January 29, 2018, 10:11: Ion gauge on Vacuum @ 3.3e-06T

Monday, January 29, 2018, 11:47; Power cycled. PLC and Canbus reset/cfg.

Monday, January 29, 2018, 11:54: Optics & Sources ok/reset.

 

Monday, January 29, 2018, 13:27: BIAS/EZL/EE @ 30 / 13 / 1.5 kV.

Monday, January 29, 2018, 13:30: BIAS/EZL/EE @ 40 / 17 / 2 kV.

Monday, January 29, 2018, 13:33: BIAS/EZL/EE @ 45 / 19.5 / 2.25 kV.

Monday, January 29, 2018, 13:36: BIAS/EZL/EE @ 50 / 21.5 / 2.5 kV.

Monday, January 29, 2018, 13:40: BIAS/EZL/EE @ 55 / 23.8 / 2.75 kV.

Monday, January 29, 2018, 13:43: BIAS/EZL/EE @ 60 / 26 / 3 kV.

SPARKS*

Monday, January 29, 2018, 13:48: BIAS/EZL/EE @ 55 / 23.8 / 2.75 kV.

Monday, January 29, 2018, 13:51: BIAS/EZL/EE @ 57.5 / 26 / 2.9 kV.

Sparks

Monday, January 29, 2018, 13:57: BIAS/EZL/EE @ 58 / 27 / 2.95 kV.

Monday, January 29, 2018, 14:03: Holding.

Monday, January 29, 2018, 14:07: BIAS/EZL/EE @ 59 / 28 / 3.1 kV.

Monday, January 29, 2018, 14:14: Spark*

Monday, January 29, 2018, 14:45: Holding

Monday, January 29, 2018, 14:56: BIAS/EZL/EE @ 0. Test Complete

Attachment 1: 2018_HV_test_2.docx
  625   Tuesday, January 30, 2018, 06:51 Aaron Schmidt LP-SIS  (Blank)42752-3DoneHV test 3

Tuesday, January 30, 2018, 06:50: BIAS / EZL / EE @ 30 / 13 / 1.5

Tuesday, January 30, 2018, 06:52: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.

Tuesday, January 30, 2018, 06:58: Optics on / set

Tuesday, January 30, 2018, 06:59: IV4 opened.

RGA switched completely off and unplugged.

Tuesday, January 30, 2018, 13:15: BIAS / EZL / EE @ 40 / 17 / 2

Tuesday, January 30, 2018, 13:20: BIAS / EZL / EE @ 50 / 21.8 / 2.5

Tuesday, January 30, 2018, 13:50: Half hour wait. Ok. No sparks.

Tuesday, January 30, 2018, 13:55: BIAS / EZL / EE @ 55 / 24 / 2.8

15 minute wait. No sparks.

Tuesday, January 30, 2018, 14:10: BIAS / EZL / EE @ 57.5 / 27 / 3.1

Tuesday, January 30, 2018, 14:23: BIAS / EZL / EE @ 60 / 28 / 3.1

Tuesday, January 30, 2018, 14:35: No sparks.

Tuesday, January 30, 2018, 15:05: BIAS / EZL / EE @ 0

Tuesday, January 30, 2018, 15:06: Auto-cooling started.


Wednesday, January 31, 2018, 06:59: Power/sources off.

 

Attachment 1: 2018_HV_test_3.docx
  627   Wednesday, January 31, 2018, 12:58 Aaron Schmidt LP-SIS  (Blank)42752-3In progressHV test 4

Wednesday, January 31, 2018, 12:53: Power/sources on.

Wednesday, January 31, 2018, 12:58: BIAS / EZL / EE @ 40 / 17 / 2

Wednesday, January 31, 2018, 13:02: BIAS / EZL / EE @ 40 / 17 / 2

Wednesday, January 31, 2018, 13:05: BIAS / EZL / EE @ 50 / 22 / 2.5

Wednesday, January 31, 2018, 13:08: BIAS / EZL / EE @ 55 / 24 / 2.8

Wednesday, January 31, 2018, 13:10: BIAS / EZL / EE @ 57.5 / 27 / 3.0

Wednesday, January 31, 2018, 13:14: BIAS / EZL / EE @ 60 / 28 / 3.1

Wednesday, January 31, 2018, 13:34: No Sparks.

Wednesday, January 31, 2018, 14:14: No sparks after one hour.

Wednesday, January 31, 2018, 14:44: Spark free at 1-1/2 hours.

Wednesday, January 31, 2018, 14:45: Bringing down voltage.

Wednesday, January 31, 2018, 14:56: BIAS / EZL / EE @ 0

Wednesday, January 31, 2018, 14:57: Sources off.

Attachment 1: 2018_HV_test_4.docx
  648   Tuesday, March 13, 2018, 06:00 Aaron SchmidtUC#22LP-SISITETM4  In progressConditioning Heat Shield for UC#22

Tuesday, March 13, 2018, 08:50: Roughing.

Tuesday, March 13, 2018, 09:10: Turbo Pumps started.

Tuesday, March 13, 2018, 09:15: Ion gauge on.

Tuesday, March 13, 2018, 10:17: RGA started.

Tuesday, March 13, 2018, 13:15: Check Vacuum @ 6.56e-07T


Wednesday, March 14, 2018, 07:27: Waterlines opened. No response on RGA. Vacuum ok @ 2.69e-07T

Wednesday, March 14, 2018, 07:30: Power/sources on.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 21.7 3.61
 TBHT 21.4 2.82
 COIL + EE 21.5 4.64
 TGHT  21.7 5.23

Wednesday, March 14, 2018, 07:35: Auto ramping TGHT to 580 A @ 2A per min.

Wednesday, March 14, 2018, 13:18: Vacuum ok @ 5.48e-07T

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 32.6 3.72
 TBHT 22.5 2.86
 COIL + EE 22.6 4.75
 TGHT  24.9 5.08

 Wednesday, March 14, 2018, 13:29: BIAS/EZL/EE to 30 / 13.5 / 1.5 kV.

BIAS current: 40uA

EZL current: 49uA

EE current: 518uA


Thursday, March 15, 2018, 08:00: Check, Vacuum @ 4.76e-07T

Thursday, March 15, 2018, 13:19: Check, Vacuum @ 4.65e-07T

Thursday, March 15, 2018, 15:13: Started cool down. Vacuum at 4.76e-7 by AM


Friday, March 16, 2018, 06:24: Power/sources off.

Friday, March 16, 2018, 06:27: Waterlines purged

Friday, March 16, 2018, 06:28: RGA, ion gauge, and turbos off.

Friday, March 16, 2018, 07:57: Venting.

 

  658   Wednesday, April 25, 2018, 09:50 Aaron SchmidtTa (test)LP-SIS  41177-3 In progressTa test and window change

Window changed on view port. Ta target prepared for Aurelia's tests.

***Note*** No Ta window foils on the end plates. Initial ramping to on the TGHT to 300 A. Follow up in increments of 50 A and check the right door window temperature.

Wednesday, April 25, 2018, 09:47: Roughing both the Evap/box end and beam end.

Wednesday, April 25, 2018, 10:27: All 4 turbo pumps started.

Wednesday, April 25, 2018, 10:49: Both Ion gauges on.

Wednesday, April 25, 2018, 14:56: RGA on.


Thursday, April 26, 2018, 06:54: Water lines opened. No response on RGA. Vacuum stable @ 4.23e-07T

Thursday, April 26, 2018, 06:56: Power/sources on.

Thursday, April 26, 2018, 07:00: Auto ramping TGHT  to 300 @ 2A per min.

Thursday, April 26, 2018, 07:05: IV4 opened. 

Thursday, April 26, 2018, 08:50: Door temp @ 25 C @ left window.

Thursday, April 26, 2018, 09:30: 300A reached. Door temp @ 26 C @ left window.

Thursday, April 26, 2018, 13:01: 550A Reached. Door temp @ 29 C @ left window.


Monday, April 30, 2018, 07:05: IV4 closed, RGA off.

Monday, April 30, 2018, 07:10: Ion gauge off. Turbo pumps off.

Monday, April 30, 2018, 11:00: Venting. Purging water lines.

Attachment 1: TS-2018-04-25-14-55.rga
Attachment 2: Ta_target_heating_test_-_Test_stand.docx
  676   Friday, July 27, 2018, 14:44 Aaron SchmidtUC#24LP-SISITETM441177-342752-6In progressConditioning Heat Shield for UC#24 & voltage test prep

4 Kapton wire routed into the 9-pin connector. Unused pins are grounded on the bracket.

Friday, July 27, 2018, 14:25: Roughing

Friday, July 27, 2018, 14:41: Turbo pumps on.

Friday, July 27, 2018, 15:53: RGA on.


Saturday, July 28, 2018, 07:03: Waterlines opened. No response on RGA. Vacuum steady @ 1.34e-06T


Friday, August 03, 2018, 10:30: Exposed 9-pin wires inside the cage insulated with electrical tape.

Friday, August 03, 2018, 10:35: Auto-ramping TGHT / TBHT to 580 / 225 A @ 2A per min.


Tuesday, August 07, 2018, 07:40: Small spikes but frequent noted in the RGA on Water, Nitrogen, and Oxygen. More than usual although the vacuum in reasonable at 7.0e-07T.

Tuesday, August 07, 2018, 07:49: Auto-cooling started @ 2A per min.

  678   Tuesday, August 14, 2018, 11:03 Aaron Schmidt LP-SIS  41177-342752-6 Voltage Tests (Ta target)

Tuesday, August 14, 2018, 10:55: Auto ramping TGHT @ 6 A per min to 40 A.

Tuesday, August 14, 2018, 11:03: TGHT @ 40 A. Display @ 0.198 V. PS @ 0.18 V. Multimeter @ 033 mV.

Tuesday, August 14, 2018, 11:09: Auto ramping TGHT @ 6 A per min to 80 A.

Tuesday, August 14, 2018, 11:16: TGHT @ 80.1 A. Display @ 0.405 V. PS @ 0.38 V. Multimeter @ 084 mV.

Tuesday, August 14, 2018, 11:18: Auto ramping TGHT @ 6 A per min to 120 A.

Tuesday, August 14, 2018, 11:28: TGHT @ 120.1 A. Display @ 0.718 V. PS @ 0.69 V. Multimeter @ 187 mV.

Tuesday, August 14, 2018, 11:35: TGHT @ 120.1 A. Display @ 0.745 V. PS @ 0.72 V. Multimeter @ 201 mV.

Tuesday, August 14, 2018, 11:36: Auto ramping TGHT @ 6 A per min to 160 A.

Tuesday, August 14, 2018, 11:44: TGHT @ 160.3 A. Display @ 1.131 V. PS @ 1.11 V. Multimeter @ 340 mV.

Tuesday, August 14, 2018, 11:54 TGHT @ 160.3 A. Display @ 1.148 V. PS @ 1.12 V. Multimeter @ 347 mV.

Tuesday, August 14, 2018, 11:54: Auto ramping TGHT @ 6 A per min to 200 A.

Tuesday, August 14, 2018, 12:01: TGHT @ 200.4 A. Display @ 1.565 V. PS @ 1.54 V. Multimeter @ 502 mV.

Tuesday, August 14, 2018, 12:47 TGHT @ 200.4 A. Display @ 1.582 V. PS @ 1.56 V. Multimeter @ 508 mV.

Tuesday, August 14, 2018, 12:47: Auto ramping TGHT @ 6 A per min to 230A.

Tuesday, August 14, 2018, 12:54: TGHT @ 230.6 A. Display @ 1.924 V. PS @ 1.89 V. Multimeter @ 635 mV.

Tuesday, August 14, 2018, 13:00: TGHT @ 230.6 A. Display @ 1.929 V. PS @ 1.90 V. Multimeter @ 638 mV.

Tuesday, August 14, 2018, 13:01: Auto ramping TGHT @ 6 A per min to 240A.

Tuesday, August 14, 2018, 13:05: TGHT @ 240.6 A. Display @ 2.046 V. PS @ 2.02 V. Multimeter @ 682 mV.

Tuesday, August 14, 2018, 13:10: TGHT @ 240.6 A. Display @ 2.049 V. PS @ 2.02 V. Multimeter @ 683 mV.

Tuesday, August 14, 2018, 13:11 Auto ramping TGHT @ 6 A per min to 280A.

Tuesday, August 14, 2018, 13:19: TGHT @ 280.7 A. Display @ 2.537 V. PS @ 2.51 V. Multimeter @ 869 mV.

Tuesday, August 14, 2018, 13:24: TGHT @ 280.7 A. Display @ 2.540 V. PS @ 2.51 V. Multimeter @ 871 mV.

Tuesday, August 14, 2018, 13:25 Auto ramping TGHT @ 6 A per min to 300A.

Tuesday, August 14, 2018, 13:30: TGHT @ 300.8 A. Display @ 2.794 V. PS @ 2.76 V. Multimeter @ 968 mV.

Tuesday, August 14, 2018, 13:34: TGHT @ 300.8 A. Display @ 2.796 V. PS @ 2.77 V. Multimeter @ 970 mV.

Tuesday, August 14, 2018, 13:36: Auto ramping TGHT @ 6 A per min to 354A.

Tuesday, August 14, 2018, 13:46: TGHT @ 354.9 A. Display @ 3.509 V. PS @ 3.48 V. Multimeter @ 1248 mV.

Tuesday, August 14, 2018, 13:51: TGHT @ 354.9 A. Display @ 3.512 V. PS @ 3.48 V. Multimeter @ 1249 mV.

Tuesday, August 14, 2018, 13:52 Auto ramping TGHT @ 6 A per min to 360A.

Tuesday, August 14, 2018, 13:55: TGHT @ 360.9 A. Display @ 3.592 V. PS @ 3.56 V. Multimeter @ 1281 mV.

Tuesday, August 14, 2018, 13:56 Auto ramping TGHT @ 6 A per min to 370A.

Tuesday, August 14, 2018, 14:00 TGHT @ 370.8 A. Display @ 3.729 V. PS @ 3.70 V. Multimeter @ 1336 mV.

Tuesday, August 14, 2018, 14:03: Auto ramping TGHT @ 6 A per min to 400A.

Tuesday, August 14, 2018, 14:10 TGHT @ 401.0 A. Display @ 4.149 V. PS @ 4.12 V. Multimeter @ 1501 mV.

Tuesday, August 14, 2018, 14:15: Auto ramping TGHT @ 6 A per min to 410A.

Tuesday, August 14, 2018, 14:19: TGHT @ 411.0 A. Display @ 4.291 V. PS @ 4.26 V. Multimeter @ 1558 mV.

Tuesday, August 14, 2018, 14:23: Auto ramping TGHT @ 6 A per min to 427A.

Tuesday, August 14, 2018, 14:28: TGHT @ 428.0 A. Display @ 4.537 V. PS @ 4.51 V. Multimeter @ 1657 mV.

Tuesday, August 14, 2018, 14:32:  Auto ramping TGHT @ 6 A per min to 451A.

Tuesday, August 14, 2018, 14:38: TGHT @ 452.2 A. Display @ 4.891 V. PS @ 4.86 V. Multimeter @ 1805 mV.

Tuesday, August 14, 2018, 14:42: Auto ramping TGHT @ 6 A per min to 460A.

Tuesday, August 14, 2018, 14:47:  TGHT @ 461.2 A. Display @ 5.026 V. PS @ 4.99 V. Multimeter @ 1862 mV.

Tuesday, August 14, 2018, 14:52: Auto ramping TGHT @ 6 A per min to 475A.

Tuesday, August 14, 2018, 14:55: TGHT @ 476.3 A. Display @ 5.250 V. PS @ 5.22 V. Multimeter @ 1962 mV.

Tuesday, August 14, 2018, 14:58: Auto ramping TGHT @ 6 A per min to 500A.

Tuesday, August 14, 2018, 15:04: TGHT @ 501.4 A. Display @ 5.629 V. PS @ 5.60 V. Multimeter @ 2130 mV.

Tuesday, August 14, 2018, 15:08:  Auto ramping TGHT @ 6 A per min to 510A.

Tuesday, August 14, 2018, 15:11: TGHT @ 511.5 A. Display @ 5.783 V. PS @ 5.75 V. Multimeter @ 2190 mV.

Tuesday, August 14, 2018, 15:13:  Auto ramping TGHT @ 6 A per min to 520A.

Tuesday, August 14, 2018, 15:17: TGHT @ 521.5 A. Display @ 5.937 V. PS @ 5.91 V. Multimeter @ 2260 mV.


Wednesday, August 15, 2018, 09:29: TGHT @ 521.5 A. Display @ 5.956 V. PS @ 5.92 V. Multimeter @ 2260 mV.

Wednesday, August 15, 2018, 09:34: Auto ramping TGHT @ 6 A per min to 550A.

Wednesday, August 15, 2018, 09:39: TGHT @ 551.5 A. Display @ 6.422 V. PS @ 6.39 V. Multimeter @ 2480 mV.

Wednesday, August 15, 2018, 09:42: Auto ramping TGHT @ 6 A per min to 560A.

Wednesday, August 15, 2018, 09:44: TGHT @ 561.5 A. Display @ 6.579 V. PS @ 6.55 V. Multimeter @ 2550 mV.

Wednesday, August 15, 2018, 09:46: Auto ramping TGHT @ 6 A per min to 580A.

Wednesday, August 15, 2018, 09:53: TGHT @ 581.7 A. Display @ 6.896 V. PS @ 6.86 V. Multimeter @ 2680 mV.

Wednesday, August 15, 2018, 09:55: Auto ramping TGHT @ 6 A per min to 600A.

Wednesday, August 15, 2018, 09:59: TGHT @ 600.0 A. Display @ 7.214 V. PS @ 7.18 V. Multimeter @ 2820 mV.

Wednesday, August 15, 2018, 10:07: Auto cooling TGHT @ 6 A per min to 580A.

Wednesday, August 15, 2018, 10:10: TGHT @ 581.7 A. Display @ 6.896 V. PS @ 6.86 V. Multimeter @ 2680 mV.

Wednesday, August 15, 2018, 10:12: Auto cooling TGHT @ 6 A per min to 560A.

Wednesday, August 15, 2018, 10:17: TGHT @ 561.6 A. Display @ 6.584 V. PS @ 6.55 V. Multimeter @ 2550 mV.

Wednesday, August 15, 2018, 10:18: Auto cooling TGHT @ 6 A per min to 550A.

Wednesday, August 15, 2018, 10:23: TGHT @ 551.5 A. Display @ 6.427 V. PS @ 6.39 V. Multimeter @ 2480 mV.

Wednesday, August 15, 2018, 10:24: Auto cooling TGHT @ 6 A per min to 520A.

Wednesday, August 15, 2018, 10:33: TGHT @ 521.5 A. Display @ 5.963 V. PS @ 5.93 V. Multimeter @ 2290 mV.

Wednesday, August 15, 2018, 10:35 Auto cooling TGHT @ 6 A per min to 510A.

Wednesday, August 15, 2018, 10:39: TGHT @ 511.5 A. Display @ 5.814 V. PS @ 5.78 V. Multimeter @ 2200 mV.

Wednesday, August 15, 2018, 10:40: Auto cooling TGHT @ 6 A per min to 500A.

Wednesday, August 15, 2018, 10:45 TGHT @ 501.4 A. Display @ 5.665 V. PS @ 5.63 V. Multimeter @ 2140 mV.

Wednesday, August 15, 2018, 10:45: Auto cooling TGHT @ 6 A per min to 475A.

Wednesday, August 15, 2018, 10:52: TGHT @ 476.3 A. Display @ 5.292 V. PS @ 5.26 V. Multimeter @ 1997 mV.

Wednesday, August 15, 2018, 10:53: Auto cooling TGHT @ 6 A per min to 460A.

Wednesday, August 15, 2018, 10:59: TGHT @ 461.2 A. Display @ 5.070 V. PS @ 5.04 V. Multimeter @ 1879 mV.

Wednesday, August 15, 2018, 11:00:  Auto cooling TGHT @ 6 A per min to 451A.

Wednesday, August 15, 2018, 11:05: TGHT @ 452.2 A. Display @ 4.935 V. PS @ 4.90 V. Multimeter @ 1821 mV.

Wednesday, August 15, 2018, 11:06: Auto cooling TGHT @ 6 A per min to 427A.

Wednesday, August 15, 2018, 11:15: TGHT @ 428.0 A. Display @ 4.584 V. PS @ 4.55 V. Multimeter @ 1674 mV.

Wednesday, August 15, 2018, 11:16: Auto cooling TGHT @ 6 A per min to 410A.

Wednesday, August 15, 2018, 11:20: TGHT @ 411.0 A. Display @ 4.337 V. PS @ 4.31 V. Multimeter @ 1576 mV.

Wednesday, August 15, 2018, 11:21: Auto cooling TGHT @ 6 A per min to 400A.

Wednesday, August 15, 2018, 11:25: TGHT @ 401.0 A. Display @ 4.195 V. PS @ 4.17 V. Multimeter @ 1519 mV.

Wednesday, August 15, 2018, 13:13: TGHT @ 401.0 A. Display @ 4.193 V. PS @ 4.16 V. Multimeter @ 1516 mV.

Wednesday, August 15, 2018, 13:14: Auto cooling TGHT @ 6 A per min to 370A.

Wednesday, August 15, 2018, 13:23:  TGHT @ 370.8 A. Display @ 3.773 V. PS @ 3.74 V. Multimeter @ 1350 mV.

Wednesday, August 15, 2018, 13:24 Auto cooling TGHT @ 6 A per min to 360A.

Wednesday, August 15, 2018, 13:30: TGHT @ 360.9 A. Display @ 3.634 V. PS @ 3.61 V. Multimeter @ 1296 mV.

Wednesday, August 15, 2018, 13:31: Auto cooling TGHT @ 6 A per min to 354A.

Wednesday, August 15, 2018, 13:35: TGHT @ 354.9 A. Display @ 3.553 V. PS @ 3.52 V. Multimeter @ 1263 mV.

Wednesday, August 15, 2018, 13:36 Auto cooling TGHT @ 6 A per min to 300A.

Wednesday, August 15, 2018, 13:46: TGHT @ 300.8 A. Display @ 2.830 V. PS @ 2.80 V. Multimeter @ 0984 mV.

Wednesday, August 15, 2018, 13:51: TGHT @ 300.8 A. Display @ 2.828 V. PS @ 2.80 V. Multimeter @ 0981 mV.

Wednesday, August 15, 2018, 13:52: Auto cooling TGHT @ 6 A per min to 280A.

Wednesday, August 15, 2018, 13:58: TGHT @ 280.7 A. Display @ 2.569 V. PS @ 2.54 V. Multimeter @ 0882 mV.

Wednesday, August 15, 2018, 13:59: Auto cooling TGHT @ 6 A per min to 240A.

Wednesday, August 15, 2018, 14:07: TGHT @ 240.6 A. Display @ 2.073 V. PS @ 2.05 V. Multimeter @ 0685 mV.

Wednesday, August 15, 2018, 14:11: TGHT @ 240.6 A. Display @ 2.071 V. PS @ 2.04 V. Multimeter @ 0692 mV.

Wednesday, August 15, 2018, 14:12: Auto cooling TGHT @ 6 A per min to 230A.

Wednesday, August 15, 2018, 14:18: TGHT @ 230.6 A. Display @ 1.951 V. PS @ 1.92 V. Multimeter @ 0647 mV.

Wednesday, August 15, 2018, 14:19: Auto cooling TGHT @ 6 A per min to 200A.

Wednesday, August 15, 2018, 14:25: TGHT @ 200.4 A. Display @ 1.612 V. PS @ 1.59 V. Multimeter @ 0522 mV.

Wednesday, August 15, 2018, 14:31: TGHT @ 200.4 A. Display @ 1.604 V. PS @ 1.58 V. Multimeter @ 0517 mV.

Wednesday, August 15, 2018, 14:31: Auto cooling TGHT @ 6 A per min to 160A.

Wednesday, August 15, 2018, 14:39: TGHT @ 160.3 A. Display @ 1.179 V. PS @ 1.16 V. Multimeter @ 0364 mV.

Wednesday, August 15, 2018, 14:44: TGHT @ 160.3 A. Display @ 1.170 V. PS @ 1.14 V. Multimeter @ 0356 mV.

Wednesday, August 15, 2018, 14:44: Auto cooling TGHT @ 6 A per min to 120A.

Wednesday, August 15, 2018, 14:52: TGHT @ 120.1 A. Display @ 0.794 V. PS @ 0.77 V. Multimeter @ 0226 mV.

Wednesday, August 15, 2018, 14:56: TGHT @ 120.3 A. Display @ 0.779 V. PS @ 0.76 V. Multimeter @ 0218 mV.

Wednesday, August 15, 2018, 14:56: Auto cooling TGHT @ 6 A per min to 80A.

Wednesday, August 15, 2018, 15:04: TGHT @ 80.1 A. Display @ 0.466 V. PS @ 0.45 V. Multimeter @ 0117 mV.

Wednesday, August 15, 2018, 15:09: TGHT @ 80.1 A. Display @ 0.454 V. PS @ 0.43 V. Multimeter @ 0109 mV.

Wednesday, August 15, 2018, 15:13: TGHT @ 80.1 A. Display @ 0.447 V. PS @ 0.42 V. Multimeter @ 0105 mV.

Wednesday, August 15, 2018, 15:14: Auto cooling TGHT @ 6 A per min to 40A.

Wednesday, August 15, 2018, 15:22: TGHT @ 40.0 A. Display @ 0.217 V. PS @ 0.19 V. Multimeter @ 0041 mV.

Wednesday, August 15, 2018, 15:27: TGHT @ 40.0 A. Display @ 0.210 V. PS @ 0.19 V. Multimeter @ 0039 mV.

Wednesday, August 15, 2018, 15:28: Auto cooling TGHT @ 6 A per min to 0A.

------------------------------------

Thursday, August 16, 2018, 13:41: Auto ramping TGHT @ 6 A per min to 40A. Auto ramping TBHT @ 6 A per min to 40A.

Thursday, August 16, 2018, 13:52 TGHT @ 40.0 A. Display @ 0.200 V. PS @ 0.18 V. Multimeter @ 0116 mV. TBHT @ 38.4 A. Display @ 0.195 V. PS @ 0.18 V.

Thursday, August 16, 2018, 13:53 Auto ramping TGHT @ 6 A per min to 80A. Auto ramping TBHT @ 6 A per min to 80A.

Thursday, August 16, 2018, 14:02: TGHT @ 80.1 A. Display @ 0.405 V. PS @ 0.38 V. Multimeter @ 0279 mV. TBHT @ 78.5 A. Display @ 0.391 V. PS @ 0.38 V.

Thursday, August 16, 2018, 14:13: TGHT @ 80.1 A. Display @ 0.437 V. PS @ 0.41 V. Multimeter @ 0304 mV. TBHT @ 78.5 A. Display @ 0.410 V. PS @ 0.40 V.

Thursday, August 16, 2018, 14:22: TGHT @ 80.1 A. Display @ 0.444 V. PS @ 0.42 V. Multimeter @ 0313 mV. TBHT @ 78.5 A. Display @ 0.418 V. PS @ 0.40 V.

Thursday, August 16, 2018, 14:22: Auto ramping TGHT @ 6 A per min to 120A. Auto ramping TBHT @ 6 A per min to 120A.

Thursday, August 16, 2018, 14:31: TGHT @ 120.1 A. Display @ 0.735 V. PS @ 0.72 V. Multimeter @ 0610 mV. TBHT @ 118.5 A. Display @ 0.711 V. PS @ 0.71 V.

Thursday, August 16, 2018, 14:44: TGHT @ 120.1 A. Display @ 0.794 V. PS @ 0.77 V. Multimeter @ 0674 mV. TBHT @ 118.5 A. Display @ 0.767 V. PS @ 0.75 V.

Thursday, August 16, 2018, 14:45: Auto ramping TGHT @ 6 A per min to 160A. Auto ramping TBHT @ 6 A per min to 160A.

Thursday, August 16, 2018, 14:53: TGHT @ 160.3 A. Display @ 1.165 V. PS @ 1.15 V. Multimeter @ 0927 mV. TBHT @ 158.7 A. Display @ 1.197 V. PS @ 1.19 V.

Thursday, August 16, 2018, 15:01: TGHT @ 160.3 A. Display @ 1.197 V. PS @ 1.17 V. Multimeter @ 0946 mV. TBHT @ 158.7 A. Display @ 1.221 V. PS @ 1.21 V.

Thursday, August 16, 2018, 15:02: Auto ramping TGHT @ 6 A per min to 200A. Auto ramping TBHT @ 6 A per min to 200A.

Thursday, August 16, 2018, 15:10: TGHT @ 200.4 A. Display @ 1.617 V. PS @ 1.60 V. Multimeter @ 1382 mV. TBHT @ 198.8 A. Display @ 1.678 V. PS @ 1.68 V.

Thursday, August 16, 2018, 15:17: TGHT @ 200.4 A. Display @ 1.634 V. PS @ 1.61 V. Multimeter @ 1390 mV. TBHT @ 198.7 A. Display @ 1.690 V. PS @ 1.69 V.

Thursday, August 16, 2018, 15:18: Auto ramping TGHT @ 6 A per min to 230A. Auto ramping TBHT @ 6 A per min to 230A.

Thursday, August 16, 2018, 15:25: TGHT @ 230.5 A. Display @ 1.976 V. PS @ 1.95 V. Multimeter @ 1751 mV. TBHT @ 228.9 A. Display @ 2.071 V. PS @ 2.07 V.

Thursday, August 16, 2018, 15:34: TGHT @ 230.6 A. Display @ 1.985 V. PS @ 1.96 V. Multimeter @ 1754 mV. TBHT @ 228.9 A. Display @ 2.081 V. PS @ 2.08 V.

Thursday, August 16, 2018, 15:35: Auto ramping TGHT @ 6 A per min to 240A.

Thursday, August 16, 2018, 15:42: TGHT @ 240.6 A. Display @ 2.100 V. PS @ 2.07 V. Multimeter @ 1801 mV. TBHT @ 228.9 A. Display @ 2.083 V. PS @ 2.08 V.

Thursday, August 16, 2018, 15:44: Auto ramping TGHT @ 6 A per min to 280A.

Thursday, August 16, 2018, 15:52: TGHT @ 280.6 A. Display @ 2.581 V. PS @ 2.56 V. Multimeter @ 1996 mV. TBHT @ 228.9 A. Display @ 2.095 V. PS @ 2.09 V.

Thursday, August 16, 2018, 16:00: TGHT @ 280.7 A. Display @ 2.586 V. PS @ 2.56 V. Multimeter @ 1998 mV. TBHT @ 228.9 A. Display @ 2.098 V. PS @ 2.10 V.

Thursday, August 16, 2018, 16:01: Auto ramping TGHT @ 6 A per min to 300A.

Thursday, August 16, 2018, 16:06: TGHT @ 300.8 A. Display @ 2.835 V. PS @ 2.81 V. Multimeter @ 2090 mV. TBHT @ 228.9 A. Display @ 2.103 V. PS @ 2.10 V.

Thursday, August 16, 2018, 16:11: TGHT @ 300.8 A. Display @ 2.840 V. PS @ 2.81 V. Multimeter @ 2090 mV. TBHT @ 228.9 A. Display @ 2.105 V. PS @ 2.10 V.

Thursday, August 16, 2018, 16:11: Auto ramping TGHT @ 6 A per min to 354A.

Thursday, August 16, 2018, 16:22: TGHT @ 354.9 A. Display @ 3.546 V. PS @ 3.52 V. Multimeter @ 2390 mV. TBHT @ 228.9 A. Display @ 2.122 V. PS @ 2.12 V.

Thursday, August 16, 2018, 16:26: TGHT @ 354.9 A. Display @ 3.551 V. PS @ 3.52 V. Multimeter @ 2390 mV. TBHT @ 228.9 A. Display @ 2.125 V. PS @ 2.12 V.

Thursday, August 16, 2018, 16:26: Auto ramping TGHT @ 6 A per min to 360A.

Thursday, August 16, 2018, 16:32: TGHT @ 360.9 A. Display @ 3.634 V. PS @ 3.60 V. Multimeter @ 2420 mV. TBHT @ 228.9 A. Display @ 2.127 V. PS @ 2.12 V.

Thursday, August 16, 2018, 16:33: Auto ramping TGHT @ 6 A per min to 370A.

Thursday, August 16, 2018, 16:40: TGHT @ 370.8 A. Display @ 3.768 V. PS @ 3.74 V. Multimeter @ 2480 mV. TBHT @ 228.9 A. Display @ 2.129 V. PS @ 2.13 V.

Thursday, August 16, 2018, 16:41: Auto ramping TGHT @ 6 A per min to 400A.

Thursday, August 16, 2018, 16:47: TGHT @ 401.0 A. Display @ 4.183 V. PS @ 4.15 V. Multimeter @ 2650 mV. TBHT @ 228.9 A. Display @ 2.139 V. PS @ 2.14 V.

--------------------------------------------------------------

Friday, August 17, 2018, 09:12: TGHT @ 400.9 A. Display @ 4.190 V. PS @ 4.16 V. Multimeter @ 2650 mV. TBHT @ 228.9 A. Display @ 2.144 V. PS @ 2.14 V.

Friday, August 17, 2018, 09:13: Auto ramping TGHT @ 6 A per min to 410A.

Friday, August 17, 2018, 09:16: TGHT @ 410.8 A. Display @ 4.327 V. PS @ 4.30 V. Multimeter @ 2710 mV. TBHT @ 228.9 A. Display @ 2.147 V. PS @ 2.15 V.

Friday, August 17, 2018, 09:19: TGHT @ 410.8 A. Display @ 4.332 V. PS @ 4.30 V. Multimeter @ 2720 mV. TBHT @ 228.9 A. Display @ 2.149 V. PS @ 2.15 V.

Friday, August 17, 2018, 09:20: Auto ramping TGHT @ 6 A per min to 427A.

Friday, August 17, 2018, 09:24: TGHT @ 428.0 A. Display @ 4.571 V. PS @ 4.55 V. Multimeter @ 2820 mV. TBHT @ 228.9 A. Display @ 2.154 V. PS @ 2.15 V.

Friday, August 17, 2018, 09:28: TGHT @ 428.0 A. Display @ 4.574 V. PS @ 4.55 V. Multimeter @ 2820 mV. TBHT @ 228.9 A. Display @ 2.156 V. PS @ 2.15 V.

Friday, August 17, 2018, 09:29: Auto ramping TGHT @ 6 A per min to 451A.

Friday, August 17, 2018, 09:37: TGHT @ 452.0 A. Display @ 4.923 V. PS @ 4.89 V. Multimeter @ 2970 mV. TBHT @ 228.9 A. Display @ 2.166 V. PS @ 2.17 V.

Friday, August 17, 2018, 09:38: Auto ramping TGHT @ 6 A per min to 460A.

Friday, August 17, 2018, 09:44: TGHT @ 461.2 A. Display @ 5.057 V. PS @ 5.03 V. Multimeter @ 3020 mV. TBHT @ 228.9 A. Display @ 2.168 V. PS @ 2.17 V.

Friday, August 17, 2018, 09:45: Auto ramping TGHT @ 6 A per min to 475A.

Friday, August 17, 2018, 09:51: TGHT @ 476.2 A. Display @ 5.280 V. PS @ 5.25 V. Multimeter @ 3120 mV. TBHT @ 228.9 A. Display @ 2.176 V. PS @ 2.18 V.

Friday, August 17, 2018, 09:52: Auto ramping TGHT @ 6 A per min to 500A.

Friday, August 17, 2018, 10:01: TGHT @ 501.4 A. Display @ 5.656 V. PS @ 5.63 V. Multimeter @ 3280 mV. TBHT @ 228.9 A. Display @ 2.188 V. PS @ 2.19 V.

Friday, August 17, 2018, 10:02: Auto ramping TGHT @ 6 A per min to 510A.

Friday, August 17, 2018, 10:08: TGHT @ 511.4 A. Display @ 5.810 V. PS @ 5.78 V. Multimeter @ 3350 mV. TBHT @ 228.9 A. Display @ 2.193 V. PS @ 2.19 V.

Friday, August 17, 2018, 10:09: Auto ramping TGHT @ 6 A per min to 520A.

Friday, August 17, 2018, 10:14: TGHT @ 521.3 A. Display @ 5.961 V. PS @ 5.93 V. Multimeter @ 3420 mV. TBHT @ 228.9 A. Display @ 2.203 V. PS @ 2.20 V.

Friday, August 17, 2018, 10:16: Auto ramping TGHT @ 6 A per min to 550A.

Friday, August 17, 2018, 10:24: TGHT @ 551.5 A. Display @ 6.425 V. PS @ 6.39 V. Multimeter @ 3630 mV. TBHT @ 228.9 A. Display @ 2.217 V. PS @ 2.22 V.

Friday, August 17, 2018, 10:25: Auto ramping TGHT @ 6 A per min to 560A.

Friday, August 17, 2018, 10:32: TGHT @ 561.5 A. Display @ 6.584 V. PS @ 6.55 V. Multimeter @ 3690 mV. TBHT @ 228.9 A. Display @ 2.222 V. PS @ 2.22 V.

Friday, August 17, 2018, 10:32: Auto ramping TGHT @ 6 A per min to 580A.

Friday, August 17, 2018, 10:40: TGHT @ 581.7 A. Display @ 6.899 V. PS @ 6.87 V. Multimeter @ 3830 mV. TBHT @ 228.9 A. Display @ 2.232 V. PS @ 2.23 V.

Friday, August 17, 2018, 10:41: Auto ramping TGHT @ 6 A per min to 600A.

Friday, August 17, 2018, 10:47: TGHT @ 600.0 A. Display @ 7.214 V. PS @ 7.18 V. Multimeter @ 3970 mV. TBHT @ 228.9 A. Display @ 2.239 V. PS @ 2.24 V.

Friday, August 17, 2018, 10:56: TGHT @ 600.0 A. Display @ 7.214 V. PS @ 7.18 V. Multimeter @ 3970 mV. TBHT @ 228.9 A. Display @ 2.242 V. PS @ 2.24 V.

Friday, August 17, 2018, 10:58: Auto cooling TGHT @ 6 A per min to 580A.

Friday, August 17, 2018, 11:09: TGHT @ 581.7 A. Display @ 6.901 V. PS @ 6.87 V. Multimeter @ 3830 mV. TBHT @ 228.9 A. Display @ 2.234 V. PS @ 2.23 V.

Friday, August 17, 2018, 11:09: Auto cooling TGHT @ 6 A per min to 560A.

Friday, August 17, 2018, 11:19: TGHT @ 561.5 A. Display @ 6.586 V. PS @ 6.55 V. Multimeter @ 3700 mV. TBHT @ 228.9 A. Display @ 2.225 V. PS @ 2.22 V.

Friday, August 17, 2018, 11:20: Auto cooling TGHT @ 6 A per min to 550A.

Friday, August 17, 2018, 11:25: TGHT @ 551.5 A. Display @ 6.432 V. PS @ 6.40 V. Multimeter @ 3630 mV. TBHT @ 228.9 A. Display @ 2.220 V. PS @ 2.22 V.

Friday, August 17, 2018, 11:26: Auto cooling TGHT @ 6 A per min to 520A.

Friday, August 17, 2018, 11:35: TGHT @ 521.3 A. Display @ 5.976 V. PS @ 5.94 V. Multimeter @ 3430 mV. TBHT @ 228.9 A. Display @ 2.208 V. PS @ 2.21 V.

Friday, August 17, 2018, 11:36: Auto cooling TGHT @ 6 A per min to 510A.

Friday, August 17, 2018, 11:41: TGHT @ 511.4 A. Display @ 5.827 V. PS @ 5.80 V. Multimeter @ 3370 mV. TBHT @ 228.9 A. Display @ 2.203 V. PS @ 2.20 V.

Friday, August 17, 2018, 11:42: Auto cooling TGHT @ 6 A per min to 500A.

Friday, August 17, 2018, 11:48: TGHT @ 501.4 A. Display @ 5.678 V. PS @ 5.65 V. Multimeter @ 3300 mV. TBHT @ 228.9 A. Display @ 2.198 V. PS @ 2.20 V.

Friday, August 17, 2018, 11:49: Auto cooling TGHT @ 6 A per min to 475A.

Friday, August 17, 2018, 11:55: TGHT @ 476.2 A. Display @ 5.314 V. PS @ 5.28 V. Multimeter @ 3140 mV. TBHT @ 228.9 A. Display @ 2.190 V. PS @ 2.19 V.

Friday, August 17, 2018, 12:59: TGHT @ 476.2 A. Display @ 5.309 V. PS @ 5.28 V. Multimeter @ 3140 mV. TBHT @ 228.9 A. Display @ 2.188 V. PS @ 2.19 V.

Friday, August 17, 2018, 13:00: Auto cooling TGHT @ 6 A per min to 460A.

Friday, August 17, 2018, 13:08: TGHT @ 461.2 A. Display @ 5.092 V. PS @ 5.06 V. Multimeter @ 3050 mV. TBHT @ 228.9 A. Display @ 2.183 V. PS @ 2.18 V.

Friday, August 17, 2018, 13:09: Auto cooling TGHT @ 6 A per min to 451A.

Friday, August 17, 2018, 13:16: TGHT @ 452.2 A. Display @ 4.960 V. PS @ 4.93 V. Multimeter @ 2990 mV. TBHT @ 228.9 A. Display @ 2.173 V. PS @ 2.17 V.

Friday, August 17, 2018, 13:17: Auto cooling TGHT @ 6 A per min to 427A.

Friday, August 17, 2018, 13:26: TGHT @ 428.0 A. Display @ 4.608 V. PS @ 4.58 V. Multimeter @ 2830 mV. TBHT @ 228.9 A. Display @ 2.159 V. PS @ 2.16 V.

Friday, August 17, 2018, 13:27: Auto cooling TGHT @ 6 A per min to 410A.

Friday, August 17, 2018, 13:34: TGHT @ 411.0 A. Display @ 4.364 V. PS @ 4.33 V. Multimeter @ 2730 mV. TBHT @ 228.9 A. Display @ 2.151 V. PS @ 2.15 V.

Friday, August 17, 2018, 13:36: Auto cooling TGHT @ 6 A per min to 400A.

Friday, August 17, 2018, 13:41: TGHT @ 400.9 A. Display @ 4.220 V. PS @ 4.19 V. Multimeter @ 2670 mV. TBHT @ 228.9 A. Display @ 2.149 V. PS @ 2.15 V.

Friday, August 17, 2018, 13:42: Auto cooling TGHT @ 6 A per min to 370A.

Friday, August 17, 2018, 13:52: TGHT @ 370.8 A. Display @ 3.797 V. PS @ 3.77 V. Multimeter @ 2490 mV. TBHT @ 228.9 A. Display @ 2.137 V. PS @ 2.14 V.

Friday, August 17, 2018, 13:53: Auto cooling TGHT @ 6 A per min to 360A.

Friday, August 17, 2018, 14:01: TGHT @ 360.9 A. Display @ 3.661 V. PS @ 3.63 V. Multimeter @ 2430 mV. TBHT @ 228.9 A. Display @ 2.132 V. PS @ 2.13 V.

Friday, August 17, 2018, 14:02: Auto cooling TGHT @ 6 A per min to 354A.

Friday, August 17, 2018, 14:05: TGHT @ 354.9 A. Display @ 3.580 V. PS @ 3.55 V. Multimeter @ 2400 mV. TBHT @ 228.9 A. Display @ 2.129 V. PS @ 2.13 V.

Friday, August 17, 2018, 14:07: Auto cooling TGHT @ 6 A per min to 300A.

Friday, August 17, 2018, 14:17: TGHT @ 300.8 A. Display @ 2.869 V. PS @ 2.84 V. Multimeter @ 2110 mV. TBHT @ 228.9 A. Display @ 2.112 V. PS @ 2.11 V.

Friday, August 17, 2018, 14:22: TGHT @ 300.8 A. Display @ 2.862 V. PS @ 2.83 V. Multimeter @ 2100 mV. TBHT @ 228.9 A. Display @ 2.110 V. PS @ 2.11 V.

Friday, August 17, 2018, 14:22: Auto cooling TGHT @ 6 A per min to 280A.

Friday, August 17, 2018, 14:30: TGHT @ 280.7 A. Display @ 2.606 V. PS @ 2.58 V. Multimeter @ 2000 mV. TBHT @ 228.9 A. Display @ 2.103 V. PS @ 2.10 V

.Friday, August 17, 2018, 14:22: Auto cooling TGHT @ 6 A per min to 240A.

Friday, August 17, 2018, 14:56: TGHT @ 240.6 A. Display @ 2.081 V. PS @ 2.09 V. Multimeter @ 1798 mV. TBHT @ 228.9 A. Display @ 2.081 V. PS @ 2.08 V

Friday, August 17, 2018, 14:58: Auto cooling TGHT @ 6 A per min to 230A.

Friday, August 17, 2018, 15:05: TGHT @ 230.5 A. Display @ 1.998 V. PS @ 1.97 V. Multimeter @ 1750 mV. TBHT @ 228.9 A. Display @ 2.078 V. PS @ 2.08 V

------------------------------------------------------------------------------

Sunday, August 19, 2018, 12:44: TGHT @ 230.5 A. Display @ 1.998 V. PS @ 1.97 V. Multimeter @ 1748 mV. TBHT @ 228.9 A. Display @ 2.078 V. PS @ 2.08 V

Sunday, August 19, 2018, 12:45: Auto cooling TGHT @ 6 A per min to 200A.

Sunday, August 19, 2018, 12:51: TGHT @ 200.4 A. Display @ 1.661 V. PS @ 1.64 V. Multimeter @ 1617 mV. TBHT @ 228.9 A. Display @ 2.071 V. PS @ 2.07 V

Sunday, August 19, 2018, 12:56: TGHT @ 200.4 A. Display @ 1.658 V. PS @ 1.63 V. Multimeter @ 1614 mV. TBHT @ 228.9 A. Display @ 2.068 V. PS @ 2.07 V

Sunday, August 19, 2018, 12:59: Auto cooling TGHT @ 6 A per min to 160A.

Sunday, August 19, 2018, 13:06: TGHT @ 160.3 A. Display @ 1.245 V. PS @ 1.22 V. Multimeter @ 1452 mV. TBHT @ 228.9 A. Display @ 2.061 V. PS @ 2.06 V

Sunday, August 19, 2018, 13:11: TGHT @ 160.3 A. Display @ 1.236 V. PS @ 1.21 V. Multimeter @ 1449 mV. TBHT @ 228.9 A. Display @ 2.056 V. PS @ 2.06 V

Sunday, August 19, 2018, 13:11: Auto cooling TGHT @ 6 A per min to 120A.

Sunday, August 19, 2018, 13:20: TGHT @ 120.1 A. Display @ 0.867 V. PS @ 0.84 V. Multimeter @ 1311 mV. TBHT @ 228.9 A. Display @ 2.049 V. PS @ 2.05 V

Sunday, August 19, 2018, 13:24: TGHT @ 120.1 A. Display @ 0.857 V. PS @ 0.83 V. Multimeter @ 1306 mV. TBHT @ 228.9 A. Display @ 2.046 V. PS @ 2.05 V

Sunday, August 19, 2018, 13:25: Auto cooling TGHT @ 6 A per min to 80A.

Sunday, August 19, 2018, 13:33: TGHT @ 80.1 A. Display @ 0.540 V. PS @ 0.51 V. Multimeter @ 1193 mV. TBHT @ 228.9 A. Display @ 2.039 V. PS @ 2.04 V

Sunday, August 19, 2018, 13:37: TGHT @ 80.1 A. Display @ 0.535 V. PS @ 0.51 V. Multimeter @ 1190 mV. TBHT @ 228.9 A. Display @ 2.037 V. PS @ 2.04 V

Sunday, August 19, 2018, 13:38: Auto cooling TGHT @ 6 A per min to 40A.

Sunday, August 19, 2018, 13:53: TGHT @ 40.0 A. Display @ 0.259 V. PS @ 0.24 V. Multimeter @ 1100 mV. TBHT @ 228.9 A. Display @ 2.032 V. PS @ 2.03 V

Sunday, August 19, 2018, 13:53: Auto cooling TGHT @ 6 A per min to 0A.

Sunday, August 19, 2018, 14:02: TGHT @ 4.4 A. Display @ 0.042 V. PS @ 0.02 V. Multimeter @ 1033 mV. TBHT @ 228.9 A. Display @ 2.029 V. PS @ 2.03 V

Sunday, August 19, 2018, 14:02: Auto cooling TBHT @ 6 A per min to 200A.

Sunday, August 19, 2018, 14:12: TGHT @ 4.4 A. Display @ 0.042 V. PS @ 0.02 V. Multimeter @ 0807 mV. TBHT @ 198.8 A. Display @ 1.648 V. PS @ 1.65 V

Sunday, August 19, 2018, 14:15: Auto cooling TBHT @ 6 A per min to 160A.

Sunday, August 19, 2018, 14:27: TGHT @ 4.4 A. Display @ 0.042 V. PS @ 0.02 V. Multimeter @ 0531 mV. TBHT @ 158.7 A. Display @ 1.179 V. PS @ 1.17 V

Sunday, August 19, 2018, 14:28: Auto cooling TBHT @ 6 A per min to 120A.

Sunday, August 19, 2018, 14:50: TGHT @ 4.4 A. Display @ 0.042 V. PS @ 0.02 V. Multimeter @ 0419 mV. TBHT @ 118.5 A. Display @ 0.725 V. PS @ 0.72 V

Sunday, August 19, 2018, 14:50: Auto cooling TBHT @ 6 A per min to 80A.

Sunday, August 19, 2018, 15:02: TGHT @ 4.4 A. Display @ 0.042 V. PS @ 0.02 V. Multimeter @ 0208 mV. TBHT @ 79.0 A. Display @ 0.415 V. PS @ 0.40 V

Sunday, August 19, 2018, 15:02: Auto cooling TBHT @ 6 A per min to 40A.

Sunday, August 19, 2018, 15:20: TGHT @ 4.4 A. Display @ 0.042 V. PS @ 0.02 V. Multimeter @ 0083 mV. TBHT @ 38.4 A. Display @ 0.198 V. PS @ 0.19 V

Sunday, August 19, 2018, 15:21: Auto cooling TBHT @ 6 A per min to 0A.


Monday, August 20, 2018, 09:10: RGA off.

Monday, August 20, 2018, 09:14: Ion gauge, turbo pumps off.

Monday, August 20, 2018, 09:15: Power off.

Monday, August 20, 2018, 10:26: Venting.

ELOG V2.9.2-2455