Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  Test-Stand, Page 7 of 53  Not logged in ELOG logo
ID Date Author Target/No.# Source Station Module Target Oven W.O#down Heat Shield W.O# Status Subject
  594   Friday, June 30, 2017, 13:31 Aaron SchmidtTa#53HP-SISITWTM242218-641751-6In progressConditioning Ta#53

Friday, June 30, 2017, 13:48: Roughing.

Friday, June 30, 2017, 14:26: Turbo pumps on.

Friday, June 30, 2017, 14:31: Ion gauge on.


Tuesday, July 04, 2017, 07:47: Waterlines opened. No response on RGA. Vacuum @1.35e-07T.

Tuesday, July 04, 2017, 07:51: Power/sources on.

Tuesday, July 04, 2017, 08:01: BIAS / EZL / EE @ 20 / 13 / 1.5 kV .

Tuesday, July 04, 2017, 08:02: IV4 Opened.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 26.2 3.78
 TBHT 26.1 2.99
 COIL + EE 25.8 4.87
 TGHT  26.0 5.48

***Halted to bring in spare target for Ta#52.***


Thursday, July 06, 2017, 09:20: Power/sources off

Thursday, July 06, 2017, 09:24: IV4 closed. Timeout error. Fixed. RGA off

Thursday, July 06, 2017, 09:27: Waterlines purged.

Thursday, July 06, 2017, 09:29: Ion gauge off. Turbos off.

Thursday, July 06, 2017, 13:14: Venting.

***Halted to bring in spare target for Ta#52.***

Attachment 1: Ta#53.docx
Attachment 2: TS-2017-06-30-14-35.rga
  596   Monday, July 10, 2017, 13:27 Aaron SchmidtTa#53HP-SISITWTM242218-641751-6In progressConditioning Ta#53

Monday, July 10, 2017, 14:03: Roughing.

Monday, July 10, 2017, 14:27: Turbo pumps started.

Monday, July 10, 2017, 14:33: Ion gauge on.

Monday, July 10, 2017, 14:36: RGA on. 


Tuesday, July 11, 2017, 06:52: Waterlines opened. No response on RGA. Vacuum ok @ 2.18e-07T

Tuesday, July 11, 2017, 06:56: Power/sources on.

Tuesday, July 11, 2017, 07:03: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .

Tuesday, July 11, 2017, 07:04: IV4 Opened.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 27.1 3.78
 TBHT 27.0 2.95
 COIL + EE 26.9 4.87
 TGHT  27.1 5.48

Tuesday, July 11, 2017, 07:10: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.

Tuesday, July 11, 2017, 10:10: RGA error (Spark?). Restarted.

Tuesday, July 11, 2017, 15:08: Anode brought up to 5, 10 15, and 50 V using FEBIAD controls. No Current detected.


Wednesday, July 12, 2017, 08:23:

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 33.2 3.85
 TBHT 27.5 2.98
 COIL + EE 27.2 4.87
 TGHT  29.5 5.48

Wednesday, July 12, 2017, 14:42: Auto-cooling.


Friday, July 14, 2017, 12:13: Sources off. IV4 Closed (Reset)


Monday, July 17, 2017, 07:55: Power off.

Monday, July 17, 2017, 07:57: Waterlines purged.

Monday, July 17, 2017, 07:59: RGA off. Ion gauge & Turbos off.

Monday, July 17, 2017, 09:43: Venting.

Attachment 1: ER-580-200-30kV-m7.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1480 -0.0989 -0.0958 -0.0916 -0.0922 -0.0961 -0.1141 -0.0934 -0.0870 -0.1038 -0.1031 -0.0906 -0.0870 -0.0986 -0.1053 -0.0986 -0.1013 -0.0961 -0.1010 -0.1022 -0.1062
 -0.0861 -0.0937 -0.0912 -0.0873 -0.0894 -0.0931 -0.1102 -0.0882 -0.0854 -0.0998 -0.0967 -0.0870 -0.0858 -0.0925 -0.1001 -0.0916 -0.0946 -0.0906 -0.0955 -0.0980 -0.1004
 -0.0900 -0.0922 -0.0873 -0.0882 -0.0888 -0.0919 -0.1080 -0.0900 -0.0891 -0.0879 -0.1004 -0.0879 -0.0894 -0.0885 -0.0974 -0.0912 -0.0928 -0.0861 -0.0876 -0.0876 -0.0955
 -0.0897 -0.0888 -0.0864 -0.0931 -0.0882 -0.0888 -0.1083 -0.0897 -0.0833 -0.0787 -0.0842 -0.0903 -0.0836 -0.0839 -0.0961 -0.0912 -0.0885 -0.0861 -0.0809 -0.0845 -0.0839
 -0.0873 -0.0870 -0.0833 -0.0974 -0.0851 -0.0916 -0.1089 -0.0912 -0.0876 -0.0848 -0.0891 -0.0912 -0.0906 -0.0861 -0.0977 -0.0946 -0.0900 -0.0833 -0.0824 -0.0858 -0.0851
 -0.0861 -0.0839 -0.0751 -0.0772 -0.0815 -0.0894 -0.1096 -0.0906 -0.0854 -0.0876 -0.0873 -0.0931 -0.0873 -0.0882 -0.0983 -0.0946 -0.0925 -0.0845 -0.0879 -0.0873 -0.0851
 -0.0870 -0.0876 -0.0833 -0.0137 -0.0476 -0.0806 -0.1086 -0.0876 -0.0809 -0.0833 -0.0833 -0.0967 -0.0851 -0.0870 -0.0998 -0.0912 -0.0909 -0.0867 -0.0949 -0.0861 -0.0827
 -0.0867 -0.0894 -0.0888 -0.0568 0.0336 -0.0510 -0.1013 -0.0873 -0.0845 -0.0845 -0.0845 -0.0879 -0.0851 -0.0836 -0.0955 -0.0909 -0.0873 -0.0851 -0.0937 -0.0842 -0.0842
 -0.0867 -0.0891 -0.0858 -0.0891 -0.0403 0.0198 -0.0790 -0.0870 -0.0867 -0.0845 -0.0867 -0.0876 -0.0864 -0.0854 -0.0958 -0.0897 -0.0903 -0.0845 -0.0845 -0.0812 -0.0842
 -0.0876 -0.0876 -0.0854 -0.0964 -0.0714 0.0116 0.0070 -0.0620 -0.0772 -0.0812 -0.0845 -0.0867 -0.0909 -0.0845 -0.0977 -0.0897 -0.0931 -0.0873 -0.0861 -0.0827 -0.0858
 -0.0861 -0.0851 -0.0861 -0.1010 -0.0839 -0.0760 -0.0195 0.0287 -0.0610 -0.0797 -0.0830 -0.0854 -0.0885 -0.0845 -0.0986 -0.0909 -0.0912 -0.0848 -0.0815 -0.0858 -0.0861
 -0.0851 -0.0839 -0.0873 -0.1022 -0.0864 -0.0836 -0.0937 -0.0201 0.0531 -0.0632 -0.0833 -0.0858 -0.0848 -0.0870 -0.0961 -0.0897 -0.0888 -0.0845 -0.0815 -0.0854 -0.0845
 -0.0873 -0.0861 -0.0873 -0.0995 -0.0830 -0.0836 -0.1041 -0.0769 -0.0165 0.0104 -0.0790 -0.0873 -0.0848 -0.0839 -0.0958 -0.0876 -0.0867 -0.0858 -0.0964 -0.0854 -0.0836
 -0.0888 -0.0882 -0.0885 -0.0912 -0.0861 -0.0839 -0.1080 -0.0867 -0.0665 -0.0031 -0.0391 -0.0888 -0.0873 -0.0854 -0.0964 -0.0909 -0.0873 -0.0854 -0.0809 -0.0842 -0.0833
 -0.0848 -0.0854 -0.0885 -0.0916 -0.0861 -0.0885 -0.1071 -0.0894 -0.0845 -0.0677 -0.0476 -0.0848 -0.0858 -0.0882 -0.0986 -0.0906 -0.0879 -0.0845 -0.0821 -0.0824 -0.0842
 -0.0851 -0.0851 -0.0848 -0.0989 -0.0815 -0.0900 -0.1065 -0.0897 -0.0854 -0.0848 -0.0842 -0.0864 -0.0879 -0.0882 -0.0995 -0.0903 -0.0906 -0.0867 -0.1007 -0.0833 -0.0839
 -0.0854 -0.0839 -0.0854 -0.1004 -0.0839 -0.0861 -0.1050 -0.0873 -0.0793 -0.0836 -0.0827 -0.0848 -0.0870 -0.0851 -0.0970 -0.0891 -0.0885 -0.0873 -0.0806 -0.0836 -0.0830
 -0.0864 -0.0861 -0.0879 -0.0952 -0.0851 -0.0839 -0.1068 -0.0888 -0.0854 -0.0858 -0.0845 -0.0845 -0.0848 -0.0858 -0.0946 -0.0903 -0.0870 -0.0842 -0.0888 -0.0839 -0.0848
 -0.0873 -0.0870 -0.0867 -0.0931 -0.0836 -0.0876 -0.1068 -0.0925 -0.0900 -0.0870 -0.0851 -0.0858 -0.0885 -0.0854 -0.0955 -0.0900 -0.0864 -0.0839 -0.0827 -0.0833 -0.0827
 -0.0891 -0.0873 -0.0824 -0.1007 -0.0824 -0.0845 -0.1031 -0.0912 -0.0842 -0.0818 -0.0845 -0.0870 -0.0894 -0.0864 -0.0967 -0.0912 -0.0888 -0.0848 -0.0772 -0.0833 -0.0851
 -0.0861 -0.0830 -0.0839 -0.0970 -0.0806 -0.0815 -0.1007 -0.0864 -0.0815 -0.0833 -0.0830 -0.0870 -0.0858 -0.0845 -0.0986 -0.0922 -0.0873 -0.0824 -0.0992 -0.0833 -0.0833
Attachment 2: ER-580-200-30kV-m129.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1501 -0.1071 -0.0986 -0.0906 -0.0946 -0.0998 -0.1016 -0.1120 -0.0882 0.0452 0.0061 -0.0577 -0.0891 -0.0995 -0.1028 -0.0934 -0.1013 -0.0998 -0.1022 -0.0943 -0.1047
 -0.0931 -0.1025 -0.0919 -0.0888 -0.0894 -0.0955 -0.0974 -0.0970 -0.1263 -0.0412 -0.0046 -0.0531 -0.0839 -0.0980 -0.0977 -0.0900 -0.0961 -0.0949 -0.0967 -0.0888 -0.1022
 -0.0891 -0.0995 -0.0897 -0.0897 -0.0900 -0.0903 -0.0970 -0.0925 -0.1083 -0.1120 -0.0461 -0.0583 -0.0833 -0.0909 -0.0964 -0.0906 -0.0934 -0.0922 -0.0925 -0.0885 -0.0922
 -0.0851 -0.0885 -0.0830 -0.0934 -0.0864 -0.0885 -0.0974 -0.0940 -0.0943 -0.1089 -0.0900 -0.0769 -0.0784 -0.0848 -0.0952 -0.0897 -0.0842 -0.0949 -0.0906 -0.0888 -0.0900
 -0.0842 -0.0891 -0.0864 -0.0888 -0.0870 -0.0888 -0.0986 -0.0934 -0.0955 -0.0916 -0.1007 -0.0949 -0.0861 -0.0876 -0.0961 -0.0897 -0.0873 -0.0946 -0.0925 -0.0854 -0.0861
 -0.0858 -0.0900 -0.0864 -0.0967 -0.0882 -0.0900 -0.0986 -0.0986 -0.1065 -0.0925 -0.0912 -0.0946 -0.0861 -0.0888 -0.0961 -0.0879 -0.0891 -0.0934 -0.0916 -0.0851 -0.0867
 -0.0870 -0.0879 -0.0830 -0.0928 -0.0778 -0.0748 -0.0995 -0.1077 -0.1132 -0.1013 -0.0919 -0.0946 -0.0912 -0.0894 -0.0955 -0.0876 -0.0867 -0.0940 -0.0888 -0.0842 -0.0870
 -0.0873 -0.0858 -0.0858 -0.0815 -0.0818 0.1971 0.2328 -0.0522 -0.1306 -0.1120 -0.0955 -0.0952 -0.0909 -0.0873 -0.0952 -0.0888 -0.0882 -0.0940 -0.0909 -0.0867 -0.0876
 -0.0842 -0.0842 -0.0876 -0.0842 -0.0790 -0.0400 1.9156 2.0398 0.3156 -0.0864 -0.0922 -0.0970 -0.0888 -0.0842 -0.0937 -0.0906 -0.0891 -0.0937 -0.0876 -0.0858 -0.0882
 -0.0848 -0.0839 -0.0925 -0.0946 -0.0803 -0.0693 0.1837 5.7727 7.4100 0.4596 -0.0653 -0.0885 -0.0870 -0.0848 -0.0931 -0.0903 -0.0873 -0.0906 -0.0928 -0.0873 -0.0864
 -0.0873 -0.0864 -0.0931 -0.0922 -0.0848 -0.0726 -0.0433 2.6282 9.9997 5.0674 0.0568 -0.0699 -0.0836 -0.0845 -0.0928 -0.0894 -0.0861 -0.0909 -0.0891 -0.0873 -0.0845
 -0.0879 -0.0900 -0.0909 -0.0964 -0.0879 -0.0803 -0.0571 0.2975 6.2939 9.9997 1.7688 -0.0534 -0.0699 -0.0809 -0.0934 -0.0870 -0.0854 -0.0931 -0.0925 -0.0867 -0.0833
 -0.0870 -0.0891 -0.0864 -0.0970 -0.0900 -0.0882 -0.0848 -0.0153 0.8432 7.7576 6.0599 -0.0018 -0.0732 -0.0732 -0.0797 -0.0861 -0.0848 -0.0943 -0.0885 -0.0858 -0.0854
 -0.0867 -0.0876 -0.0882 -0.0992 -0.0848 -0.0891 -0.0912 -0.0787 0.0012 1.1005 5.1297 2.8534 -0.0516 -0.0800 -0.0873 -0.0836 -0.0818 -0.0864 -0.0934 -0.0836 -0.0870
 -0.0851 -0.0873 -0.0873 -0.0986 -0.0864 -0.0867 -0.0922 -0.0854 -0.0711 -0.0067 0.4382 0.9674 0.5585 -0.0751 -0.0864 -0.0845 -0.0851 -0.0894 -0.0870 -0.0769 -0.0842
 -0.0870 -0.0854 -0.0885 -0.0909 -0.0864 -0.0851 -0.0922 -0.0861 -0.0775 -0.0616 -0.0549 -0.0162 0.0714 -0.0476 -0.0876 -0.0851 -0.0833 -0.0903 -0.0919 -0.0830 -0.0815
 -0.0885 -0.0876 -0.0879 -0.0885 -0.0861 -0.0848 -0.0919 -0.0858 -0.0806 -0.0732 -0.0766 -0.0839 -0.0827 -0.0824 -0.0925 -0.0854 -0.0824 -0.0879 -0.0861 -0.0842 -0.0851
 -0.0879 -0.0885 -0.0879 -0.0876 -0.0864 -0.0864 -0.0891 -0.0833 -0.0876 -0.0790 -0.0812 -0.0851 -0.0897 -0.0867 -0.0925 -0.0894 -0.0839 -0.0909 -0.0842 -0.0876 -0.0854
 -0.0854 -0.0876 -0.0873 -0.0974 -0.0897 -0.0879 -0.0821 -0.0684 -0.0693 -0.0803 -0.0836 -0.0851 -0.0900 -0.0882 -0.0922 -0.0894 -0.0851 -0.0925 -0.0916 -0.0861 -0.0848
 -0.0848 -0.0845 -0.0925 -0.0937 -0.0873 -0.0864 -0.0815 -0.0433 -0.0241 -0.0681 -0.0803 -0.0848 -0.0894 -0.0882 -0.0919 -0.0876 -0.0839 -0.0912 -0.0906 -0.0839 -0.0842
 -0.0861 -0.0867 -0.0931 -0.0879 -0.0876 -0.0833 -0.0803 -0.0323 0.0284 -0.0406 -0.0784 -0.0836 -0.0879 -0.0879 -0.0925 -0.0851 -0.0854 -0.0912 -0.0925 -0.0851 -0.0830
Attachment 3: TS-2017-07-10-14-35.rga
Attachment 4: TS-2017-07-11-10-10.rga
Attachment 5: Ta#53.docx
Attachment 6: OMEGA_OM-DAQPRO-5300_18-Jul-17_08.11.44.dlp
Attachment 7: ITW-TM2-Ta#53-HP-SIS_Test_Stand.docx
  567   Wednesday, January 25, 2017, 14:56 Aaron Schmidt    42218-1441751-6DoneTarget cleaning test LP-SIS CC Clean

Target Test (Fully Cleaned vs Crystal Clean only).

For all tests:

-Start Turbo pumps & Ion gauge as soon as available on ISAC TEST STAND VACUUM Screen.

- Ensure previous test turbos are off by 8:15 and started venting by 9:15am to meet timeline.

-Start Roughing by 10:10am +/- 1 minute.

-TGHT / TBHT ramping starts @ 8:30am. 2A per min to 580A / 200A

ALL START TIMES MUST BE CONSISTENT.


***Pre-outgassed*** Heat Shield W.O# 41751-6: Includes insulators, new Al grounded 9-pin, bracket, & gripper arms. Blank/Plate. No source cup/EE.


LP-SIS Target # 42218-14: Crystal Clean only.


Tuesday, February 07, 2017, 09:48: HS removed from test stand. Exchange target.

Tuesday, February 07, 2017, 10:07: Target # 42218-14 placed into test stand. Door open for 19 minutes.

Tuesday, February 07, 2017, 10:10: Roughing.

Tuesday, February 07, 2017, 10:10: Doors sealed for roughing.

Tuesday, February 07, 2017, 10:11: BV1B OPEN. Opened @ 430T.

Tuesday, February 07, 2017, 10:29: Both Turbo pumps started.

Tuesday, February 07, 2017, 10:34: Ion gauge on. TP1B @ speed 42 KRPM.

Tuesday, February 07, 2017, 10:37: TP1A @ speed 38 KRPM.

Tuesday, February 07, 2017, 10:39: RGA on. Vacuum @ 1.49E-05T.

Tuesday, February 07, 2017, 15:30: Check. Vacuum @ 3.26E-07T.


Wednesday, February 08, 2017, 08:12: Water lines opened. No response on RGA. Vacuum stable @ 1.29E-07T.

Wednesday, February 08, 2017, 08:14 : Power/Sources on.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 20.8 4.06
 TBHT 20.5 3.17
 COIL + EE 20.6 5.27
 TGHT  20.7 5.88

Wednesday, February 08, 2017, 08:30: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.

Wednesday, February 08, 2017, 10:00: Check. Vacuum @ 3.33E-07T.

Wednesday, February 08, 2017, 11:40: Check. Vacuum @ 1.70E-07T.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 30.2 4.06
 TBHT 21.0  3.17
 COIL + EE  21.1 5.22
 TGHT

 21.3

 5.82

Wednesday, February 08, 2017, 13:30: Check. Vacuum @ 3.53E-07T.

Wednesday, February 08, 2017, 15:30: Auto-cooling started @ 2A per min. Vacuum @ 3.88E-07T.

***NOTE: Small water anomaly on RGA affecting Vacuum***


***NOTE: Small water anomaly on RGA affecting Vacuum continues***

Thursday, February 09, 2017, 08:08: Sources/power off.

Thursday, February 09, 2017, 08:10: Water lines purged.

Thursday, February 09, 2017, 08:11: RGA off.

Thursday, February 09, 2017, 09:00: Venting

Attachment 1: LP-SIS_CC_Clean.docx
Attachment 2: TS-2017-02-07-10-40.rga
Attachment 3: OMEGA_OM-DAQPRO-5300_09-Feb-17_10.49.31.dlp
  592   Friday, June 09, 2017, 13:28 Aaron SchmidtTa#52LP-SISITETM442218-1441751-8In progressConditioning Ta#52

Friday, June 09, 2017, 13:27: Roughing.

Friday, June 09, 2017, 14:09: Turbo pumps on.

Friday, June 09, 2017, 14:16: Ion gauge on.


Monday, June 12, 2017, 07:20: Waterlines opened. No response on RGA. Vacuum @1.52e-07T.

Monday, June 12, 2017, 07:23: Power/Sources on.

Monday, June 12, 2017, 08:15: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .

Monday, June 12, 2017, 08:16: IV4 Opened.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 26.0 3.78
 TBHT 27.3 3.03
 COIL + EE 25.8 4.91
 TGHT  26.0 5.46

Monday, June 12, 2017, 08:18: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.

Monday, June 12, 2017, 11:33: Spark around 10:30 tripped RGA. Restarting RGA.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 35.9 3.78
 TBHT 26.6 2.95
 COIL + EE 26.1 4.89
 TGHT  28.4 5.46

 


Tuesday, June 13, 2017, 14:50: Auto-cooling @ 2A  per min


Wednesday, June 14, 2017, 07:41: Sources/power off. IV4 closed


Friday, June 16, 2017, 07:16: RGA off

Friday, June 16, 2017, 07:18: Ion gauge & turbo pumps off.

Friday, June 16, 2017, 07:20: Water lines purged.


Monday, June 19, 2017, 13:35: While writing report, it was noticed a step was skipped on the mass correlations. The MassCorr data was not saved. - AS

Attachment 1: ER-580-200-30kV-m7.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1483 -0.0916 -0.0842 -0.0763 -0.0848 -0.0891 -0.0793 -0.0638 -0.0726 -0.0879 -0.0949 -0.0916 -0.0839 -0.0815 -0.0836 -0.0937 -0.0824 -0.0797 -0.0909 -0.0873 -0.0891
 -0.0891 -0.0842 -0.0806 -0.0757 -0.0836 -0.0900 -0.0821 -0.0668 -0.0754 -0.0854 -0.0916 -0.0876 -0.0812 -0.0754 -0.0806 -0.0882 -0.0833 -0.0748 -0.0861 -0.0812 -0.0833
 -0.0873 -0.0833 -0.0714 -0.0745 -0.0787 -0.0916 -0.0827 -0.0699 -0.0705 -0.0775 -0.0891 -0.0769 -0.0790 -0.0732 -0.0809 -0.0891 -0.0754 -0.0714 -0.0757 -0.0720 -0.0684
 -0.0894 -0.0772 -0.0720 -0.0781 -0.0732 -0.0888 -0.0845 -0.0720 -0.0641 -0.0760 -0.0800 -0.0720 -0.0665 -0.0647 -0.0784 -0.0891 -0.0726 -0.0723 -0.0696 -0.0674 -0.0681
 -0.0906 -0.0769 -0.0705 -0.0708 -0.0705 -0.0870 -0.0836 -0.0690 -0.0681 -0.0778 -0.0766 -0.0693 -0.0696 -0.0693 -0.0760 -0.0867 -0.0705 -0.0708 -0.0714 -0.0696 -0.0693
 -0.0906 -0.0766 -0.0674 -0.0812 -0.0742 -0.0864 -0.0845 -0.0702 -0.0742 -0.0793 -0.0775 -0.0726 -0.0684 -0.0732 -0.0760 -0.0876 -0.0717 -0.0684 -0.0671 -0.0690 -0.0711
 -0.0900 -0.0754 -0.0708 -0.0299 -0.0372 -0.0882 -0.0867 -0.0726 -0.0723 -0.0806 -0.0751 -0.0739 -0.0732 -0.0711 -0.0787 -0.0885 -0.0705 -0.0708 -0.0711 -0.0699 -0.0684
 -0.0769 -0.0714 -0.0650 0.1456 0.2567 0.1089 -0.0485 -0.0705 -0.0702 -0.0769 -0.0739 -0.0723 -0.0705 -0.0665 -0.0803 -0.0885 -0.0693 -0.0705 -0.0714 -0.0690 -0.0668
 0.0165 0.0143 0.0021 0.1740 0.2505 0.4282 0.2008 -0.0229 -0.0650 -0.0742 -0.0732 -0.0711 -0.0705 -0.0684 -0.0787 -0.0891 -0.0687 -0.0705 -0.0696 -0.0690 -0.0677
 0.0232 0.0513 0.1300 0.2176 0.3403 0.6772 0.6064 0.1114 -0.0391 -0.0623 -0.0711 -0.0699 -0.0720 -0.0681 -0.0772 -0.0885 -0.0690 -0.0720 -0.0729 -0.0693 -0.0684
 -0.0848 0.1108 0.1334 0.1318 0.2350 0.5215 0.8182 0.4596 0.0214 0.0711 -0.0400 -0.0549 -0.0583 -0.0696 -0.0787 -0.0885 -0.0705 -0.0720 -0.0760 -0.0681 -0.0693
 -0.0879 -0.0702 -0.0650 0.1450 0.3421 0.3604 0.6375 0.5753 0.1495 0.0430 -0.0662 -0.0537 -0.0552 -0.0562 -0.0662 -0.0876 -0.0696 -0.0708 -0.0699 -0.0681 -0.0690
 -0.0891 -0.0757 -0.0708 -0.0742 -0.0568 0.2441 0.4233 0.4581 0.1941 0.0522 -0.0720 -0.0699 -0.0677 -0.0626 -0.0717 -0.0800 -0.0589 -0.0723 -0.0702 -0.0677 -0.0681
 -0.0903 -0.0760 -0.0684 -0.0668 -0.0687 -0.0769 -0.0671 0.0940 0.2032 0.1163 0.0055 -0.0696 -0.0693 -0.0696 -0.0772 -0.0864 -0.0659 -0.0659 -0.0641 -0.0681 -0.0684
 -0.0912 -0.0732 -0.0671 -0.0812 -0.0702 -0.0821 -0.0806 -0.0650 -0.0647 -0.0693 -0.0677 -0.0708 -0.0696 -0.0696 -0.0763 -0.0900 -0.0687 -0.0684 -0.0656 -0.0623 -0.0696
 -0.0876 -0.0732 -0.0671 -0.0726 -0.0690 -0.0809 -0.0818 -0.0681 -0.0696 -0.0766 -0.0711 -0.0729 -0.0717 -0.0693 -0.0772 -0.0909 -0.0681 -0.0708 -0.0687 -0.0696 -0.0677
 -0.0867 -0.0729 -0.0687 -0.0732 -0.0638 -0.0812 -0.0821 -0.0699 -0.0684 -0.0769 -0.0723 -0.0702 -0.0690 -0.0677 -0.0772 -0.0900 -0.0693 -0.0711 -0.0696 -0.0687 -0.0681
 -0.0861 -0.0745 -0.0690 -0.0729 -0.0632 -0.0775 -0.0815 -0.0690 -0.0674 -0.0757 -0.0729 -0.0696 -0.0699 -0.0668 -0.0775 -0.0900 -0.0677 -0.0714 -0.0677 -0.0677 -0.0668
 -0.0858 -0.0745 -0.0699 -0.0693 -0.0601 -0.0714 -0.0775 -0.0681 -0.0665 -0.0742 -0.0717 -0.0677 -0.0684 -0.0662 -0.0766 -0.0885 -0.0702 -0.0708 -0.0711 -0.0659 -0.0677
 -0.0864 -0.0720 -0.0671 -0.0711 -0.0607 -0.0659 -0.0684 -0.0650 -0.0668 -0.0754 -0.0720 -0.0720 -0.0696 -0.0690 -0.0766 -0.0873 -0.0693 -0.0705 -0.0714 -0.0690 -0.0677
 -0.0858 -0.0690 -0.0620 -0.0662 -0.0534 -0.0583 -0.0607 -0.0604 -0.0671 -0.0757 -0.0732 -0.0711 -0.0687 -0.0684 -0.0751 -0.0876 -0.0687 -0.0711 -0.0717 -0.0684 -0.0681
Attachment 2: ER-580-200-30kV-m23.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.4541 -0.0946 -0.0922 -0.0946 -0.0876 -0.0720 -0.0491 0.0244 0.1660 0.1889 0.0699 0.0421 -0.0452 -0.0714 -0.0754 -0.0748 -0.0809 -0.0827 -0.0925 -0.0912 -0.0946
 -0.0806 -0.0879 -0.0879 -0.0937 -0.0946 -0.0912 -0.0757 -0.0491 0.0555 0.1382 0.0497 0.0366 -0.0494 -0.0775 -0.0754 -0.0714 -0.0751 -0.0781 -0.0922 -0.0858 -0.0867
 -0.0769 -0.0839 -0.0769 -0.0790 -0.0970 -0.0943 -0.0900 -0.0784 -0.0418 0.0555 0.0235 0.0232 -0.0479 -0.0793 -0.0836 -0.0778 -0.0803 -0.0760 -0.0790 -0.0769 -0.0708
 -0.0754 -0.0827 -0.0735 -0.0827 -0.0745 -0.0970 -0.1038 -0.1031 -0.0830 -0.0449 -0.0208 -0.0140 -0.0449 -0.0757 -0.0839 -0.0821 -0.0800 -0.0836 -0.0711 -0.0699 -0.0696
 -0.0742 -0.0836 -0.0757 -0.0879 -0.0766 -0.0940 -0.1016 -0.1096 -0.1108 -0.1038 -0.0745 -0.0513 -0.0601 -0.0775 -0.0836 -0.0827 -0.0839 -0.0818 -0.0751 -0.0732 -0.0705
 -0.0778 -0.0845 -0.0763 -0.0793 -0.0809 -0.1038 -0.1068 -0.1062 -0.0928 -0.0952 -0.1074 -0.0909 -0.0824 -0.0793 -0.0824 -0.0797 -0.0824 -0.0833 -0.0742 -0.0739 -0.0742
 -0.0803 -0.0848 -0.0778 -0.0879 -0.0797 -0.1163 -0.1208 -0.1181 -0.1028 -0.0916 -0.0934 -0.0812 -0.0766 -0.0790 -0.0809 -0.0787 -0.0812 -0.0830 -0.0708 -0.0726 -0.0739
 -0.0513 -0.0827 -0.0732 -0.0656 0.0027 0.5679 0.0867 -0.1138 -0.1123 -0.0974 -0.0958 -0.0845 -0.0739 -0.0775 -0.0821 -0.0793 -0.0848 -0.0827 -0.0809 -0.0711 -0.0717
 1.2949 1.0226 0.2405 0.0510 -0.0192 4.0729 3.4451 1.0425 0.0763 -0.0818 -0.0916 -0.0830 -0.0745 -0.0797 -0.0821 -0.0803 -0.0833 -0.0824 -0.0711 -0.0732 -0.0723
 0.4846 1.1899 1.5659 1.6354 1.1652 4.6243 7.5842 5.8371 1.2793 0.0476 -0.0735 -0.0766 -0.0751 -0.0781 -0.0806 -0.0787 -0.0803 -0.0809 -0.0705 -0.0729 -0.0714
 -0.0562 0.5368 0.7703 1.5552 3.1232 5.3870 7.5870 9.9997 7.3779 2.2556 1.7380 1.2317 0.4492 -0.0720 -0.0793 -0.0793 -0.0806 -0.0800 -0.0775 -0.0739 -0.0739
 -0.0739 -0.0690 -0.0330 1.1514 1.3330 2.6193 5.1074 9.9997 9.9997 6.2830 2.7347 0.7892 0.9729 1.2283 0.4895 -0.0659 -0.0793 -0.0824 -0.0702 -0.0742 -0.0729
 -0.0748 -0.0800 -0.0693 -0.0500 -0.0009 2.8204 3.1949 5.0052 9.5169 9.9997 3.6981 3.7781 0.2814 0.5304 0.5841 0.7227 0.6760 -0.0620 -0.0797 -0.0705 -0.0708
 -0.0708 -0.0793 -0.0720 -0.0821 -0.0668 -0.0140 0.2133 4.2502 7.5003 9.9997 5.7306 3.4097 -0.0299 -0.0128 0.1605 0.4138 0.3284 0.2347 0.5820 -0.0699 -0.0711
 -0.0739 -0.0809 -0.0732 -0.0769 -0.0745 -0.0745 -0.0681 0.0278 0.2014 4.1779 7.2113 5.8627 2.5699 -0.0647 -0.0699 -0.0580 -0.0131 0.1538 0.1840 0.2417 0.4391
 -0.0748 -0.0809 -0.0751 -0.0836 -0.0745 -0.0781 -0.0766 -0.0641 -0.0452 0.0305 0.1208 0.3046 0.1395 -0.0665 -0.0732 -0.0717 -0.0763 -0.0720 -0.0412 0.0812 0.0699
 -0.0732 -0.0833 -0.0757 -0.0851 -0.0748 -0.0766 -0.0775 -0.0714 -0.0601 -0.0516 -0.0620 -0.0552 -0.0610 -0.0739 -0.0754 -0.0732 -0.0772 -0.0793 -0.0757 -0.0677 -0.0543
 -0.0693 -0.0800 -0.0717 -0.0864 -0.0745 -0.0607 -0.0693 -0.0708 -0.0638 -0.0616 -0.0705 -0.0635 -0.0690 -0.0757 -0.0797 -0.0769 -0.0790 -0.0790 -0.0708 -0.0693 -0.0699
 -0.0610 -0.0720 -0.0659 -0.0806 -0.0732 -0.0348 -0.0369 -0.0574 -0.0653 -0.0671 -0.0757 -0.0696 -0.0714 -0.0787 -0.0824 -0.0769 -0.0787 -0.0772 -0.0729 -0.0693 -0.0705
 -0.0595 -0.0653 -0.0601 -0.0693 -0.0653 -0.0027 0.0113 -0.0067 -0.0519 -0.0677 -0.0763 -0.0729 -0.0714 -0.0809 -0.0809 -0.0766 -0.0787 -0.0803 -0.0772 -0.0708 -0.0723
 -0.0595 -0.0641 -0.0528 -0.0607 -0.0488 0.0327 0.0610 0.0708 0.0015 -0.0488 -0.0653 -0.0616 -0.0656 -0.0748 -0.0797 -0.0757 -0.0787 -0.0793 -0.0793 -0.0729 -0.0717
Attachment 3: ER-580-200-30kV-m39.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1331 -0.0787 -0.0827 -0.0873 -0.0845 -0.0864 -0.0876 -0.0812 -0.0732 -0.0800 -0.0842 -0.0742 -0.0742 -0.0845 -0.0864 -0.0845 -0.0821 -0.0851 -0.0931 -0.0879 -0.0916
 -0.0739 -0.0742 -0.0760 -0.0848 -0.0812 -0.0861 -0.0858 -0.0821 -0.0739 -0.0797 -0.0818 -0.0711 -0.0745 -0.0827 -0.0833 -0.0809 -0.0769 -0.0830 -0.0894 -0.0833 -0.0888
 -0.0748 -0.0735 -0.0732 -0.0763 -0.0800 -0.0793 -0.0836 -0.0803 -0.0757 -0.0732 -0.0797 -0.0705 -0.0726 -0.0806 -0.0873 -0.0821 -0.0766 -0.0815 -0.0775 -0.0748 -0.0757
 -0.0754 -0.0711 -0.0702 -0.0854 -0.0803 -0.0766 -0.0818 -0.0800 -0.0803 -0.0745 -0.0809 -0.0778 -0.0684 -0.0708 -0.0851 -0.0821 -0.0723 -0.0787 -0.0708 -0.0714 -0.0745
 -0.0754 -0.0711 -0.0766 -0.0836 -0.0729 -0.0763 -0.0818 -0.0800 -0.0784 -0.0757 -0.0848 -0.0803 -0.0757 -0.0748 -0.0845 -0.0824 -0.0732 -0.0803 -0.0711 -0.0745 -0.0739
 -0.0735 -0.0726 -0.0732 -0.0766 -0.0751 -0.0781 -0.0830 -0.0818 -0.0769 -0.0787 -0.0848 -0.0781 -0.0742 -0.0745 -0.0812 -0.0818 -0.0729 -0.0778 -0.0754 -0.0720 -0.0723
 -0.0735 -0.0717 -0.0751 -0.0824 -0.0812 -0.0797 -0.0861 -0.0824 -0.0803 -0.0775 -0.0854 -0.0821 -0.0745 -0.0732 -0.0845 -0.0830 -0.0726 -0.0803 -0.0751 -0.0757 -0.0720
 -0.0696 -0.0723 -0.0757 -0.0754 -0.0815 -0.0732 -0.0812 -0.0824 -0.0775 -0.0754 -0.0854 -0.0754 -0.0745 -0.0714 -0.0867 -0.0827 -0.0732 -0.0803 -0.0757 -0.0705 -0.0739
 0.0568 0.0281 -0.0275 -0.0436 -0.0586 -0.0366 0.0165 -0.0409 -0.0717 -0.0760 -0.0845 -0.0775 -0.0742 -0.0693 -0.0858 -0.0830 -0.0739 -0.0821 -0.0784 -0.0729 -0.0732
 -0.0470 0.0848 0.1245 0.2023 0.1379 0.0052 0.2209 0.1651 0.0595 -0.0668 -0.0824 -0.0732 -0.0726 -0.0754 -0.0824 -0.0800 -0.0745 -0.0800 -0.0732 -0.0739 -0.0726
 -0.0742 -0.0616 -0.0397 0.0076 0.1636 0.1840 0.3632 0.4947 0.6839 0.1077 0.0528 0.0790 -0.0037 -0.0732 -0.0836 -0.0818 -0.0732 -0.0803 -0.0717 -0.0708 -0.0729
 -0.0745 -0.0711 -0.0732 -0.0671 -0.0406 0.0446 0.1764 0.3564 1.2970 0.5920 0.2246 0.2917 0.2277 0.2655 0.0363 -0.0699 -0.0748 -0.0781 -0.0732 -0.0726 -0.0742
 -0.0729 -0.0720 -0.0735 -0.0806 -0.0735 -0.0339 0.0107 0.1123 0.4984 0.6879 0.3940 0.1642 0.0809 0.1001 0.1968 0.2277 0.1578 -0.0424 -0.0732 -0.0748 -0.0745
 -0.0723 -0.0726 -0.0772 -0.0839 -0.0748 -0.0732 -0.0778 -0.0174 0.0809 0.3207 0.4752 0.1981 0.0888 -0.0320 0.0278 0.0708 0.0891 0.0406 0.1053 -0.0693 -0.0742
 -0.0732 -0.0729 -0.0696 -0.0824 -0.0739 -0.0748 -0.0809 -0.0766 -0.0693 -0.0085 0.1529 0.2121 0.1392 -0.0714 -0.0812 -0.0772 -0.0375 0.0128 0.0153 0.0046 0.0821
 -0.0720 -0.0708 -0.0729 -0.0745 -0.0772 -0.0748 -0.0824 -0.0790 -0.0729 -0.0760 -0.0775 -0.0613 -0.0473 -0.0714 -0.0827 -0.0818 -0.0714 -0.0754 -0.0577 -0.0079 -0.0162
 -0.0720 -0.0714 -0.0757 -0.0806 -0.0760 -0.0757 -0.0833 -0.0800 -0.0729 -0.0748 -0.0830 -0.0833 -0.0729 -0.0708 -0.0851 -0.0815 -0.0720 -0.0793 -0.0763 -0.0708 -0.0674
 -0.0720 -0.0735 -0.0763 -0.0812 -0.0742 -0.0763 -0.0833 -0.0800 -0.0732 -0.0766 -0.0842 -0.0720 -0.0708 -0.0720 -0.0833 -0.0818 -0.0751 -0.0812 -0.0726 -0.0720 -0.0726
 -0.0714 -0.0745 -0.0739 -0.0793 -0.0763 -0.0757 -0.0800 -0.0766 -0.0751 -0.0766 -0.0815 -0.0745 -0.0714 -0.0720 -0.0833 -0.0827 -0.0739 -0.0797 -0.0720 -0.0723 -0.0732
 -0.0742 -0.0699 -0.0687 -0.0845 -0.0748 -0.0745 -0.0775 -0.0751 -0.0732 -0.0754 -0.0824 -0.0757 -0.0723 -0.0745 -0.0824 -0.0803 -0.0720 -0.0778 -0.0702 -0.0702 -0.0702
 -0.0723 -0.0684 -0.0723 -0.0818 -0.0717 -0.0717 -0.0757 -0.0748 -0.0690 -0.0757 -0.0818 -0.0787 -0.0729 -0.0726 -0.0821 -0.0809 -0.0720 -0.0751 -0.0702 -0.0723 -0.0699
Attachment 4: OMEGA_OM-DAQPRO-5300_19-Jun-17_10.29.48.dlp
Attachment 5: Ta#52.docx
Attachment 6: TS-2017-06-09-14-20.rga
Attachment 7: TS-2017-06-12-11-34.rga
Attachment 8: ITE-TM4-Ta#52-LP-SIS_Test_Stand.docx
  568   Wednesday, January 25, 2017, 14:57 Aaron Schmidt    42218-1141751-6DoneTarget cleaning test LP-SIS Full Clean

Target Test (Fully Cleaned vs Crystal Clean only).

For all tests:

-Start Turbo pumps & Ion gauge as soon as available on ISAC TEST STAND VACUUM Screen.

- Ensure previous test turbos are off by 8:15 and started venting by 9:15am to meet timeline.

-Start Roughing by 10:10am +/- 1 minute.

-TGHT / TBHT ramping starts @ 8:30am. 2A per min to 580A / 200A

ALL START TIMES MUST BE CONSISTENT.


***Pre-outgassed*** Heat Shield W.O# 41751-6: Includes insulators, new Al grounded 9-pin, bracket, & gripper arms. Blank/Plate. No source cup/EE.


HP-SIS Target # 42218-11: Full Clean. Crystal Clear, Acetone, & Methanol used.


Thursday, February 09, 2017, 09:44: HS removed from test stand. Exchange target.

Thursday, February 09, 2017, 10:04: Target # 42218-11 placed into test stand. Door open for 20 minutes.

Thursday, February 09, 2017, 10:10: Roughing.

 Thursday, February 09, 2017, 10:10: Doors sealed for roughing.

Thursday, February 09, 2017, 10:11: BV1B OPEN. Opened @ 428T.

Thursday, February 09, 2017, 10:29: Both Turbo pumps started.

Thursday, February 09, 2017, 10:34: Ion gauge on. TP1B @ speed 42 KRPM.

Thursday, February 09, 2017, 10:37: TP1A @ speed 38 KRPM.

Thursday, February 09, 2017, 10:40: RGA on. Vacuum @ 1.45E-05T.

Thursday, February 09, 2017, 10:47: Strip tool Vacuum.Current restarted.

Thursday, February 09, 2017, 15:34: Check. Vacuum @ 3.97E-07T.

***NOTE: Small water anomaly on RGA affecting Vacuum***


***NOTE: Small water anomaly on RGA affecting Vacuum*** Continues. Monitor through heating.

Friday, February 10, 2017, 08:06: Water lines opened. No response on RGA. Vacuum stable @ 3.24E-07T.

Friday, February 10, 2017, 08:08: Power/Sources on.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 24.4 4.02
 TBHT 24.2 3.22
 COIL + EE 24.2 5.26
 TGHT  24.4 5.90

08:30: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.

10:00: Check. Vacuum @ 3.68-4.87E-07T. (Anomaly variant)

11:40: Check. Vacuum @ XE-07T.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 34.6 4.12
 TBHT 25.5  3.21
 COIL + EE  25.5 5.29
 TGHT

 27.3

 5.88

Check. Vacuum @ 4.41E-07T.

 Auto-cooling started @ 2A per min. Vacuum @ 4.25E-07T.


Tuesday, February 14, 2017, 08:22: Power/sources off.

Tuesday, February 14, 2017, 08:23: Water lines purged.

Tuesday, February 14, 2017, 08:26: RGA off. Ion Gauge off. Turbo pumps off.

Attachment 1: LP-SIS_Full_Clean.docx
Attachment 2: OMEGA_OM-DAQPRO-5300_15-Feb-17_13.30.42.dlp
Attachment 3: TS-2017-02-09-10-40.rga
  570   Wednesday, February 15, 2017, 13:42 Aaron Schmidt    42218-1141751-6Problems/SolutionsVacuum Check (Water leak)

*Vacuum from previous test still active, minus turbo pumps. Same target, & heat shield as well.*

Wednesday, February 15, 2017, 13:38: Both Turbo pumps started.

Wednesday, February 15, 2017, 13:43: Ion gauge on. TP1B @ speed 42 KRPM.

Wednesday, February 15, 2017, 13:51: RGA on. Vacuum @4.79E-05T.

Monitor water & vacuum with the RGA.


Thursday, February 16, 2017, 09:05: Main water supply & return shut off to the test stand. Purged. Vacuum @ 7.13E-07T.

Thursday, February 16, 2017, 09:16: Vacuum @ 3.59E-07T. Confirms water leak

Thursday, February 16, 2017, 09:17: RGA stopped.

Thursday, February 16, 2017, 09:20: RGA restarted.

Thursday, February 16, 2017, 09:30: Open main supply & return water supply. Individual valves are closed.

Thursday, February 16, 2017, 09:35: Open supply and return to FS #2. Vacuum @ 3.24E-07T. No response on RGA

Thursday, February 16, 2017, 09:40: Open supply and return to FS #3. Vacuum @ 6.76E-07T. Response on RGA. (Leak detected)

Thursday, February 16, 2017, 09:42: Close main Supply & Return. Purge & close FS #2 & FS #3 . Allow vacuum to stabilize.

Thursday, February 16, 2017, 09:50: Open main supply & return water supply. Open supply and return to FS #3. Vacuum @ 6.76E-07T. Response on RGA. (Leak confirmed)

Thursday, February 16, 2017, 09:53: Close main Supply & Return. Purge & close FS #3 . Allow vacuum to stabilize.

Thursday, February 16, 2017, 09:55: Open main supply & return. All valves are closed. Vacuum @ 3.51E-07T

Thursday, February 16, 2017, 10:00: Open supply and return to FS #4. Vacuum @3.26E-07T. No response on RGA.

Water leak confirmed in FS #3

Thursday, February 16, 2017, 10:05: Close main Supply & Return. Purge & close FS #4 . Allow vacuum to stabilize.

Thursday, February 16, 2017, 10:37: RGA off.

Thursday, February 16, 2017, 10:38: Ion gauge. Turbos off

Thursday, February 16, 2017, 11:58: Venting


Friday, February 17, 2017, 14:10: Roughing.

Friday, February 17, 2017, 14:30: Both Turbo pumps started.

Friday, February 17, 2017, 14:36: Ion gauge on.

Friday, February 17, 2017, 14:44: RGA on. Vacuum @ 2.32E-05T. Note* Waterlines FS #3 currently pressurized with 38 PSI Helium.


Monday, February 20, 2017, 08:26: Vacuum @ 2.60E-07T

Leak is apparent on RGA via Helium. He @ 20 PSI

Monday, February 20, 2017, 09:55: He in line brought up to 40 PSI. Monitor on RGA.

Monday, February 20, 2017, 10:00: No response on RGA. Connectivity gaps?. 

Monday, February 20, 2017, 10:05: RGA restarted. 

Monday, February 20, 2017, 10:39: Top cage removed from test stand. Waterlines purged of Helium. Immediate drop in RGA Helium reading.

Monday, February 20, 2017, 11:23: FS #3 (K) re-routed with return from (F). EE (K) pressurized with He @ 20-30 PSI. No significant response on RGA. Purge He

Monday, February 20, 2017, 11:36: FS #3 (K) returned back to (G). Return connected through (E). Pressurized to 20-30 PSI. No significant response on RGA. Purge He

Monday, February 20, 2017, 11:46: FS #3 (K) routed through to (H). (F) return to return. Pressurized to 20-30 PSI. Left pressurized for one hour

Monday, February 20, 2017, 12:51: No response on RGA. He Purged.

Monday, February 20, 2017, 12:55: Re-pressurize all of FS #3 with He @ 30 PSI.

Monday, February 20, 2017, 13:00: Purge Helium. Switch to water.

Monday, February 20, 2017, 13:12: Water supply/return on. No response on RGA.

Monday, February 20, 2017, 13:22: Slow response noted on RGA. Slow climb on water observed.

Monday, February 20, 2017, 13:30: Slight increase from 2.52E-07T to 3.26E-07T noted on Vacuum. Purge water lines. Reconnect/route lines & Test.

Monday, February 20, 2017, 13:40: FS #3 (K) re-routed with return from (F). EE (K). Open waterlines & observe.

Monday, February 20, 2017, 13:57: No response on RGA or Vacuum. Purge water & Re-route lines.

Monday, February 20, 2017, 14:05: FS #3 (K) returned back to (G). Return connected through (E). Open waterlines & observe.

Monday, February 20, 2017, 14:10: The water leak is apparent on (G) & (E) route. Vacuum climbed from 2.80E-07T to 3.33E-07T. RGA shows increase on water.

Monday, February 20, 2017, 14:14: Purge water lines.

Monday, February 20, 2017, 14:21: FS #3 (K) routed through to (H). (F) return to return. Waterlines opened to rule out (F) & (H) route.

Monday, February 20, 2017, 14:30: No response on RGA. Vacuum Stable. Purge water & Re-connect lines back to standard.

Monday, February 20, 2017, 15:22: Intermittent purging. Flow meter paddles pick up speed with air if purged for too long.


Tuesday, February 21, 2017, 08:34: RGA off, Ion gauge off, Turbos off.

Tuesday, February 21, 2017, 09:32: Venting.

  571   Tuesday, February 21, 2017, 10:35 Aaron Schmidt    42218-1141751-6Problems/SolutionsWater leak (E)(G)

Water leak found on water block lines (E) & (G) on FS #3.

Both lines were removed at the water blocks and the coil connection at the heat shield. They ends were blanked off.

Tuesday, February 21, 2017, 10:32: Roughing

Tuesday, February 21, 2017, 10:34: BV1B opened.

Tuesday, February 21, 2017, 10:51: Turbo pumps started.

Tuesday, February 21, 2017, 11:23: Ion gauge & RGA started. Vacuum @ 7.60E-06T.


Wednesday, February 22, 2017, 10:30: The test stand was pressurized with Air (HPA). The loop was completed through the heat shield line FS #5.

Pressurized @ 80 PSI w/ HPA. Monitor. No response on RGA or Vacuum.

Wednesday, February 22, 2017, 10:42: The test stand was pressurized with Helium. The loop was completed through the heat shield line FS #5. The air was purged from the lines using the Helium.

Pressurized @ 30 PSI w/ Helium. Monitor.

Wednesday, February 22, 2017, 11:57: De-pressurize the lines. No response on RGA.

Wednesday, February 22, 2017, 13:47: Pressurize line (G) @ 40 PSI w/ Helium directly.

Wednesday, February 22, 2017, 14:00: No response on RGA or Vacuum. Disconnect.

Wednesday, February 22, 2017, 14:03: Pressurize line (E) @ 38 PSI w/ Helium directly.

Wednesday, February 22, 2017, 14:30: No response on RGA or Vacuum. Disconnect.


Thursday, February 23, 2017, 08:11: RGA, Ion gauge, & Turbo pumps off.

Thursday, February 23, 2017, 10:21: Venting.

Thursday, February 23, 2017, 11:45 Lines (G)(E) reinstalled w/ new seals

Thursday, February 23, 2017, 11:58: Roughing.

Thursday, February 23, 2017, 13:00:  Turbos on.

Thursday, February 23, 2017, 13:05: Ion Gauge on.

Thursday, February 23, 2017, 13:10: RGA on.


Friday, February 24, 2017, 08:15: Waterlines opened on FS #3.

Friday, February 24, 2017, 08:18: RGA froze.

Friday, February 24, 2017, 08:20: Restarted RGA.

Friday, February 24, 2017, 08:48: FS #2, #4, #5 all opened.

Friday, February 24, 2017, 10:20: Vacuum & Leak appear to be stable. Vacuum @ 2.54E-07T

Friday, February 24, 2017, 10:25: Cage re-installed on top of vacuum Chamber

Friday, February 24, 2017, 10:28: Power/sources on

Friday, February 24, 2017, 10:30: Auto ramping TGHT / TBHT to 580 / 200 A @ 4A per min. Heat check.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 21.4 4.06
 TBHT 22.2 3.28
 COIL + EE 21.2 5.16
 TGHT  21.4 5.82

Friday, February 24, 2017, 10:37: RGA froze (Fatal communication error message).

Friday, February 24, 2017, 10:40: RGA restarted.

Friday, February 24, 2017, 13:30: No major sign of leak detected. Standard response on Vacuum and RGA to heating.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 32.1 4.06
 TBHT 24.5 3.28
 COIL + EE 23.0 5.17
 TGHT  24.9 5.80

Friday, February 24, 2017, 13:55: Vacuum and water appear leveled off with no indication of a major leak.

Friday, February 24, 2017, 14:00: Auto-cooling started @ 4A per min.


Monday, February 27, 2017, 08:12: Power/sources off. The likely cause for the leak was one of the seals within lines (G)(E). Either one of the O-rings or VCR's. Once the seals were changed, no further leaks were detected.

Wednesday, March 08, 2017, 08:40: RGA timed out. RGA off. No leak detected.


Friday, March 10, 2017, 10:05: Water lines purged.

Friday, March 10, 2017, 10:09: Ion gauge & Turbos off.

Friday, March 10, 2017, 14:24: Venting.

  595   Thursday, July 06, 2017, 13:16 Aaron SchmidtTa#55LP-SIS  42218-1142738-2In progressConditioning Ta spare for Ta#52

Thursday, July 06, 2017, 13:56: Roughing.

Thursday, July 06, 2017, 14:25: Turbo  pumps started.

Thursday, July 06, 2017, 14:30: Ion gauge on. RGA Started


Friday, July 07, 2017, 07:29: Waterlines opened. No response on RGA. Vacuum ok @ 1.97e-07T

Friday, July 07, 2017, 07:33: Power/sources on.

Friday, July 07, 2017, 07:39: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .

Friday, July 07, 2017, 07:40: IV4 Opened.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 26.0 3.78
 TBHT 26.5 3.02
 COIL + EE 25.8 4.86
 TGHT  26.0 5.47

Friday, July 07, 2017, 07:42: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.

Friday, July 07, 2017, 07:45: Optics reset.

Friday, July 07, 2017, 14:28: Anode brought up to 5, 10 15, and 50 V using FEBIAD controls. No Current detected.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 36.5 3.75
 TBHT 27.6 3.04
 COIL + EE 26.5 4.87
 TGHT  28.7 5.54

 


Saturday, July 08, 2017, 08:58: Ramping down heaters


Monday, July 10, 2017, 08:15: Power/Sources off. *BIAS 30 to 0 kV due to shut off. EE & EZL Brought to 0.

Monday, July 10, 2017, 08:23: Water lines purged. IV4 Closed

Monday, July 10, 2017, 08:27: IV4 reset.

Monday, July 10, 2017, 09:40: RGA off. Ion Gauge off. Turbos off.

Monday, July 10, 2017, 11:02: Venting.

Attachment 1: ER-580-200-30kV-m23.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1459 -0.0992 -0.0943 -0.0925 -0.0934 -0.1016 -0.0986 -0.0839 -0.0732 -0.0800 -0.0766 -0.0739 -0.0772 -0.0876 -0.0931 -0.0876 -0.0989 -0.0861 -0.0934 -0.0989 -0.0958
 -0.0903 -0.0943 -0.0906 -0.0906 -0.0879 -0.0986 -0.0931 -0.0797 -0.0793 -0.0821 -0.0723 -0.0693 -0.0726 -0.0848 -0.0888 -0.0818 -0.0867 -0.0803 -0.0882 -0.0946 -0.0897
 -0.0897 -0.0925 -0.0867 -0.0821 -0.0854 -0.0977 -0.0916 -0.0793 -0.0842 -0.0882 -0.0720 -0.0687 -0.0751 -0.0821 -0.0879 -0.0839 -0.0861 -0.0775 -0.0793 -0.0815 -0.0797
 -0.0882 -0.0903 -0.0793 -0.0937 -0.0769 -0.0977 -0.0949 -0.0806 -0.0778 -0.0858 -0.0751 -0.0729 -0.0867 -0.0748 -0.0864 -0.0867 -0.0760 -0.0781 -0.0800 -0.0784 -0.0772
 -0.0876 -0.0879 -0.0787 -0.0958 -0.0787 -0.0986 -0.0952 -0.0812 -0.0821 -0.0848 -0.0797 -0.0787 -0.0723 -0.0787 -0.0851 -0.0870 -0.0790 -0.0787 -0.0787 -0.0775 -0.0781
 -0.0858 -0.0888 -0.0781 -0.0748 -0.0720 -0.0970 -0.0961 -0.0809 -0.0897 -0.0845 -0.0793 -0.0787 -0.0769 -0.0778 -0.0839 -0.0836 -0.0757 -0.0775 -0.0769 -0.0778 -0.0778
 -0.0885 -0.0909 -0.0769 -0.0912 0.0586 -0.0385 -0.0851 -0.0790 -0.0778 -0.0870 -0.0790 -0.0790 -0.0864 -0.0766 -0.0851 -0.0812 -0.0757 -0.0760 -0.0787 -0.0751 -0.0781
 -0.0888 -0.0912 -0.0775 -0.0833 -0.0034 0.2264 0.0839 -0.0494 -0.0876 -0.0891 -0.0797 -0.0790 -0.0778 -0.0797 -0.0851 -0.0830 -0.0790 -0.0784 -0.0790 -0.0766 -0.0769
 -0.0717 -0.0797 -0.0766 -0.0916 -0.0726 0.3128 0.6992 0.1498 -0.0574 -0.0739 -0.0732 -0.0790 -0.0790 -0.0760 -0.0848 -0.0836 -0.0787 -0.0800 -0.0775 -0.0769 -0.0778
 -0.0720 -0.0684 -0.0549 -0.0714 -0.0342 -0.0745 1.2405 0.6815 0.1675 -0.0424 -0.0638 -0.0699 -0.0739 -0.0790 -0.0854 -0.0821 -0.0803 -0.0781 -0.0803 -0.0775 -0.0797
 -0.0623 -0.0571 -0.0558 -0.0565 -0.0299 -0.0323 0.2798 0.8789 1.0532 0.0723 -0.0464 -0.0595 -0.0674 -0.0699 -0.0769 -0.0815 -0.0793 -0.0772 -0.0800 -0.0781 -0.0775
 -0.0879 -0.0882 -0.0211 -0.0644 -0.0330 -0.0381 0.0974 0.2484 1.1102 0.7678 0.0488 -0.0729 -0.0705 -0.0690 -0.0763 -0.0720 -0.0647 -0.0754 -0.0760 -0.0751 -0.0766
 -0.0894 -0.0903 -0.0778 -0.0952 -0.0354 -0.0589 0.0488 0.0494 0.2368 1.1792 0.4474 0.0125 -0.0751 -0.0732 -0.0827 -0.0790 -0.0705 -0.0629 -0.0745 -0.0763 -0.0772
 -0.0879 -0.0903 -0.0787 -0.0970 -0.0760 -0.0934 -0.0165 0.0262 0.0156 0.3046 0.8878 0.2240 -0.0699 -0.0729 -0.0858 -0.0818 -0.0763 -0.0717 -0.0699 -0.0546 -0.0769
 -0.0870 -0.0900 -0.0766 -0.0888 -0.0787 -0.0961 -0.0925 -0.0714 -0.0507 0.0174 0.2612 0.5963 0.1871 -0.0787 -0.0854 -0.0815 -0.0778 -0.0787 -0.0742 -0.0684 -0.0671
 -0.0888 -0.0897 -0.0760 -0.0909 -0.0757 -0.0961 -0.0912 -0.0745 -0.0891 -0.0720 0.0137 0.1895 0.3024 -0.0314 -0.0827 -0.0818 -0.0766 -0.0787 -0.0781 -0.0790 -0.0693
 -0.0879 -0.0897 -0.0766 -0.0900 -0.0754 -0.0919 -0.0885 -0.0754 -0.0790 -0.0827 -0.0723 -0.0687 -0.0305 -0.0381 -0.0821 -0.0806 -0.0766 -0.0757 -0.0763 -0.0742 -0.0769
 -0.0885 -0.0888 -0.0766 -0.0897 -0.0745 -0.0879 -0.0873 -0.0757 -0.0864 -0.0827 -0.0775 -0.0769 -0.0751 -0.0766 -0.0842 -0.0836 -0.0766 -0.0757 -0.0778 -0.0763 -0.0757
 -0.0900 -0.0909 -0.0784 -0.0970 -0.0751 -0.0870 -0.0858 -0.0751 -0.0912 -0.0806 -0.0790 -0.0793 -0.0787 -0.0769 -0.0848 -0.0833 -0.0778 -0.0781 -0.0769 -0.0760 -0.0781
 -0.0864 -0.0900 -0.0757 -0.0961 -0.0757 -0.0864 -0.0809 -0.0696 -0.0751 -0.0821 -0.0772 -0.0793 -0.0787 -0.0760 -0.0836 -0.0818 -0.0790 -0.0778 -0.0787 -0.0778 -0.0772
 -0.0858 -0.0894 -0.0748 -0.0934 -0.0739 -0.0848 -0.0787 -0.0629 -0.0809 -0.0778 -0.0769 -0.0775 -0.0775 -0.0772 -0.0818 -0.0815 -0.0763 -0.0763 -0.0769 -0.0772 -0.0781
Attachment 2: ER-580-200-30kV-m38.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1575 -0.1071 -0.1016 -0.1065 -0.1062 -0.1035 -0.1004 -0.0903 -0.0760 -0.0363 -0.0461 -0.0601 -0.0778 -0.0818 -0.0949 -0.0995 -0.0931 -0.0916 -0.1071 -0.1041 -0.0958
 -0.0970 -0.1013 -0.0980 -0.1010 -0.1016 -0.1001 -0.0970 -0.0812 -0.0861 -0.0699 -0.0449 -0.0555 -0.0717 -0.0787 -0.0937 -0.0961 -0.0854 -0.0867 -0.1013 -0.0983 -0.0897
 -0.0952 -0.0986 -0.0851 -0.0897 -0.0955 -0.0977 -0.0983 -0.0803 -0.0836 -0.1077 -0.0610 -0.0546 -0.0717 -0.0772 -0.0894 -0.0958 -0.0864 -0.0821 -0.0961 -0.0861 -0.0809
 -0.0964 -0.0977 -0.0775 -0.0858 -0.0739 -0.0974 -0.0989 -0.0818 -0.0787 -0.1105 -0.0882 -0.0601 -0.0775 -0.0739 -0.0882 -0.0958 -0.0861 -0.0815 -0.0806 -0.0772 -0.0778
 -0.1007 -0.1004 -0.0800 -0.0861 -0.0845 -0.0989 -0.0992 -0.0848 -0.0861 -0.1050 -0.1031 -0.0732 -0.0757 -0.0751 -0.0867 -0.0967 -0.0854 -0.0815 -0.0821 -0.0797 -0.0787
 -0.0995 -0.0986 -0.0790 -0.0839 -0.0821 -0.0989 -0.0989 -0.0854 -0.0864 -0.1059 -0.0977 -0.0821 -0.0803 -0.0793 -0.0894 -0.0970 -0.0848 -0.0806 -0.0797 -0.0797 -0.0778
 -0.0964 -0.0974 -0.0735 -0.0873 -0.0375 -0.0601 -0.0922 -0.0870 -0.0867 -0.1074 -0.0952 -0.0809 -0.0812 -0.0803 -0.0894 -0.0992 -0.0845 -0.0809 -0.0809 -0.0757 -0.0772
 -0.0961 -0.0955 -0.0742 -0.0452 -0.0705 0.0897 0.0931 -0.0305 -0.0882 -0.1099 -0.0922 -0.0793 -0.0797 -0.0763 -0.0894 -0.0989 -0.0818 -0.0809 -0.0809 -0.0760 -0.0769
 -0.0134 -0.0549 -0.0696 -0.0800 -0.0336 0.0757 0.4651 0.6528 0.0015 -0.0769 -0.0827 -0.0784 -0.0775 -0.0742 -0.0861 -0.0989 -0.0858 -0.0845 -0.0821 -0.0821 -0.0775
 0.0378 0.0250 0.0470 0.0342 0.1025 -0.0128 0.2185 1.8475 1.6077 -0.0134 -0.0391 -0.0351 -0.0470 -0.0769 -0.0882 -0.0961 -0.0867 -0.0827 -0.0806 -0.0784 -0.0781
 -0.0510 0.0760 0.0638 0.0986 0.2078 0.3448 0.0732 0.6653 3.9651 1.0074 0.0226 -0.0113 -0.0189 -0.0146 -0.0632 -0.0958 -0.0836 -0.0821 -0.0809 -0.0766 -0.0797
 -0.0961 -0.0946 0.0452 0.0372 0.1334 0.1855 0.2502 0.3159 2.5229 6.0635 0.1724 -0.0327 -0.0235 -0.0327 -0.0494 -0.0406 -0.0385 -0.0790 -0.0748 -0.0751 -0.0775
 -0.0958 -0.0964 -0.0726 -0.0839 0.0092 0.0116 0.1544 0.2350 0.5399 4.1684 3.1567 0.1071 -0.0070 -0.0229 -0.0632 -0.0671 -0.0507 -0.0302 -0.0647 -0.0745 -0.0760
 -0.0974 -0.0980 -0.0784 -0.0873 -0.0800 -0.0888 0.0009 0.1086 0.1468 0.9015 3.8147 1.2885 0.1019 -0.0085 -0.0812 -0.0833 -0.0699 -0.0494 -0.0470 0.0104 -0.0754
 -0.0986 -0.0970 -0.0797 -0.0888 -0.0836 -0.0958 -0.0937 -0.0748 -0.0476 0.0757 0.7455 1.4111 0.2850 -0.0409 -0.0198 -0.0919 -0.0806 -0.0729 -0.0549 -0.0415 -0.0406
 -0.0980 -0.0974 -0.0778 -0.0882 -0.0827 -0.0946 -0.0943 -0.0790 -0.0751 -0.0818 0.0134 0.2280 0.3674 0.0885 -0.0677 -0.0916 -0.0812 -0.0793 -0.0766 -0.0671 -0.0412
 -0.0958 -0.0986 -0.0766 -0.0867 -0.0803 -0.0937 -0.0925 -0.0803 -0.0778 -0.0916 -0.0754 -0.0659 -0.0241 0.0232 -0.0778 -0.0897 -0.0803 -0.0787 -0.0781 -0.0751 -0.0745
 -0.0949 -0.0955 -0.0763 -0.0909 -0.0769 -0.0922 -0.0912 -0.0781 -0.0751 -0.0940 -0.0775 -0.0726 -0.0754 -0.0745 -0.0851 -0.0940 -0.0821 -0.0806 -0.0757 -0.0745 -0.0754
 -0.0946 -0.0980 -0.0793 -0.0851 -0.0803 -0.0934 -0.0894 -0.0726 -0.0732 -0.0952 -0.0778 -0.0748 -0.0754 -0.0748 -0.0851 -0.0946 -0.0836 -0.0793 -0.0787 -0.0763 -0.0769
 -0.0958 -0.0967 -0.0772 -0.0864 -0.0812 -0.0943 -0.0897 -0.0668 -0.0638 -0.0964 -0.0800 -0.0739 -0.0763 -0.0775 -0.0858 -0.0922 -0.0836 -0.0787 -0.0769 -0.0754 -0.0769
 -0.0955 -0.0937 -0.0751 -0.0864 -0.0790 -0.0912 -0.0876 -0.0601 -0.0464 -0.0861 -0.0790 -0.0763 -0.0766 -0.0757 -0.0854 -0.0906 -0.0824 -0.0790 -0.0754 -0.0760 -0.0781
Attachment 3: ER-580-200-30kV-m39.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1590 -0.1062 -0.1086 -0.0906 -0.0906 -0.1022 -0.1093 -0.1038 -0.0885 -0.0534 -0.0565 -0.0681 -0.0763 -0.0925 -0.1007 -0.1041 -0.1007 -0.0952 -0.1074 -0.1068 -0.1041
 -0.1016 -0.1031 -0.1047 -0.0867 -0.0873 -0.0998 -0.1019 -0.0995 -0.0922 -0.0665 -0.0510 -0.0604 -0.0708 -0.0894 -0.0949 -0.1001 -0.0940 -0.0928 -0.1013 -0.1004 -0.0989
 -0.1007 -0.0989 -0.0916 -0.0845 -0.0827 -0.0983 -0.1016 -0.0992 -0.0867 -0.0845 -0.0607 -0.0586 -0.0677 -0.0867 -0.0946 -0.0995 -0.0934 -0.0851 -0.0998 -0.0943 -0.0848
 -0.0986 -0.0970 -0.0763 -0.0781 -0.0751 -0.0919 -0.1038 -0.0992 -0.0717 -0.0867 -0.0726 -0.0647 -0.0650 -0.0717 -0.0928 -0.1019 -0.0870 -0.0888 -0.0800 -0.0790 -0.0842
 -0.0992 -0.0989 -0.0790 -0.0760 -0.0745 -0.0922 -0.1047 -0.1001 -0.0751 -0.0830 -0.0787 -0.0760 -0.0723 -0.0717 -0.0940 -0.1031 -0.0867 -0.0891 -0.0824 -0.0812 -0.0803
 -0.0989 -0.0989 -0.0790 -0.0867 -0.0757 -0.0946 -0.1056 -0.0964 -0.0720 -0.0870 -0.0787 -0.0775 -0.0772 -0.0772 -0.0934 -0.1028 -0.0827 -0.0903 -0.0790 -0.0787 -0.0763
 -0.1004 -0.0986 -0.0790 -0.0790 -0.0241 -0.0101 -0.0891 -0.0931 -0.0351 -0.0485 -0.0778 -0.0745 -0.0778 -0.0803 -0.0943 -0.1007 -0.0790 -0.0867 -0.0763 -0.0766 -0.0766
 -0.0809 -0.0986 -0.0763 -0.0790 -0.0714 0.3131 0.1181 -0.0345 0.0058 0.0626 -0.0830 -0.0748 -0.0748 -0.0766 -0.0903 -0.1010 -0.0809 -0.0873 -0.0806 -0.0790 -0.0772
 -0.0894 -0.0177 -0.0507 -0.0848 -0.0745 0.0546 0.5795 0.3409 0.0809 0.1352 -0.0708 -0.0778 -0.0803 -0.0760 -0.0894 -0.1028 -0.0824 -0.0900 -0.0784 -0.0790 -0.0784
 -0.0922 -0.0919 -0.0690 -0.0726 -0.0629 -0.0781 0.4315 1.1215 0.9171 0.2142 0.0015 -0.0677 -0.0769 -0.0787 -0.0900 -0.1031 -0.0815 -0.0888 -0.0812 -0.0787 -0.0781
 -0.0995 -0.0888 -0.0693 -0.0781 -0.0705 -0.0766 -0.0616 0.5667 1.8753 1.0110 0.1111 -0.0580 -0.0735 -0.0763 -0.0916 -0.1010 -0.0781 -0.0864 -0.0775 -0.0769 -0.0778
 -0.1001 -0.0964 -0.0726 -0.0742 -0.0717 -0.0769 -0.0699 0.0687 1.0928 3.1250 0.4935 -0.0278 -0.0681 -0.0711 -0.0888 -0.0983 -0.0754 -0.0845 -0.0800 -0.0754 -0.0763
 -0.1007 -0.0995 -0.0775 -0.0809 -0.0754 -0.0787 -0.0769 -0.0305 0.1749 1.6907 1.8475 0.2292 -0.0671 -0.0705 -0.0903 -0.0974 -0.0751 -0.0861 -0.0760 -0.0769 -0.0745
 -0.0986 -0.0986 -0.0769 -0.0824 -0.0775 -0.0888 -0.0916 -0.0552 0.0089 0.3650 1.8930 1.0120 0.1450 -0.0696 -0.0919 -0.0989 -0.0790 -0.0861 -0.0815 -0.0760 -0.0778
 -0.0986 -0.0980 -0.0754 -0.0809 -0.0751 -0.0916 -0.1025 -0.0925 -0.0546 0.0333 0.3699 0.9113 0.5783 -0.0540 -0.0885 -0.1007 -0.0809 -0.0864 -0.0772 -0.0739 -0.0766
 -0.0995 -0.0992 -0.0754 -0.0812 -0.0769 -0.0916 -0.1007 -0.0952 -0.0708 -0.0702 0.0208 0.2255 0.5313 0.2142 -0.0876 -0.0977 -0.0778 -0.0842 -0.0790 -0.0745 -0.0723
 -0.1004 -0.1010 -0.0766 -0.0812 -0.0775 -0.0894 -0.1010 -0.0958 -0.0745 -0.0848 -0.0656 -0.0620 -0.0168 0.0571 -0.0797 -0.0970 -0.0772 -0.0867 -0.0757 -0.0748 -0.0769
 -0.0986 -0.1001 -0.0769 -0.0891 -0.0803 -0.0876 -0.0992 -0.0952 -0.0784 -0.0815 -0.0702 -0.0726 -0.0760 -0.0732 -0.0891 -0.0992 -0.0784 -0.0851 -0.0763 -0.0769 -0.0763
 -0.0970 -0.0970 -0.0739 -0.0839 -0.0763 -0.0876 -0.0943 -0.0922 -0.0763 -0.0806 -0.0717 -0.0739 -0.0769 -0.0766 -0.0919 -0.1010 -0.0781 -0.0864 -0.0784 -0.0772 -0.0769
 -0.0974 -0.0958 -0.0751 -0.0845 -0.0769 -0.0867 -0.0919 -0.0864 -0.0705 -0.0803 -0.0726 -0.0754 -0.0763 -0.0772 -0.0897 -0.1010 -0.0790 -0.0864 -0.0766 -0.0778 -0.0772
 -0.0989 -0.0967 -0.0757 -0.0839 -0.0775 -0.0867 -0.0916 -0.0821 -0.0635 -0.0745 -0.0708 -0.0729 -0.0766 -0.0763 -0.0916 -0.0986 -0.0781 -0.0861 -0.0797 -0.0775 -0.0760
Attachment 4: ER-580-200-30kV-m133.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1505 -0.0983 -0.0916 -0.0909 -0.0943 -0.0958 -0.0989 -0.0854 -0.0821 -0.0665 -0.0467 -0.0623 -0.0763 -0.0919 -0.1028 -0.1016 -0.0998 -0.0983 -0.1074 -0.1065 -0.1059
 -0.0885 -0.0925 -0.0885 -0.0870 -0.0931 -0.0916 -0.0946 -0.0781 -0.0815 -0.0836 -0.0473 -0.0568 -0.0702 -0.0854 -0.0983 -0.0946 -0.0974 -0.0961 -0.1035 -0.1035 -0.0986
 -0.0885 -0.0903 -0.0815 -0.0830 -0.0845 -0.0903 -0.0916 -0.0757 -0.0790 -0.0909 -0.0604 -0.0549 -0.0671 -0.0793 -0.0977 -0.0946 -0.0949 -0.0912 -0.1022 -0.0989 -0.0864
 -0.0861 -0.0873 -0.0775 -0.0803 -0.0797 -0.0836 -0.0943 -0.0800 -0.0766 -0.0848 -0.0800 -0.0650 -0.0650 -0.0714 -0.0955 -0.0946 -0.0839 -0.0885 -0.0946 -0.0757 -0.0757
 -0.0864 -0.0870 -0.0751 -0.0793 -0.0742 -0.0842 -0.0946 -0.0775 -0.0797 -0.0812 -0.0839 -0.0790 -0.0699 -0.0757 -0.0952 -0.0934 -0.0876 -0.0885 -0.0812 -0.0757 -0.0769
 -0.0903 -0.0885 -0.0760 -0.0891 -0.0735 -0.0854 -0.0937 -0.0781 -0.0812 -0.0827 -0.0778 -0.0790 -0.0784 -0.0824 -0.0974 -0.0946 -0.0861 -0.0897 -0.0876 -0.0760 -0.0766
 -0.0894 -0.0882 -0.0772 -0.0851 -0.0436 -0.0208 -0.0787 -0.0797 -0.0815 -0.0827 -0.0784 -0.0800 -0.0818 -0.0842 -0.0974 -0.0946 -0.0845 -0.0885 -0.0818 -0.0784 -0.0751
 -0.0873 -0.0861 -0.0775 -0.0775 -0.0751 0.2106 0.1611 0.0058 -0.0708 -0.0824 -0.0784 -0.0784 -0.0793 -0.0839 -0.0980 -0.0949 -0.0870 -0.0897 -0.0870 -0.0787 -0.0800
 -0.0873 -0.0858 -0.0729 -0.0864 -0.0742 -0.0360 0.5655 0.5167 0.0711 -0.0668 -0.0778 -0.0781 -0.0775 -0.0812 -0.0967 -0.0940 -0.0845 -0.0882 -0.0821 -0.0790 -0.0793
 -0.0854 -0.0861 -0.0723 -0.0778 -0.0735 -0.0790 0.1804 1.1356 1.1194 0.0473 -0.0671 -0.0751 -0.0748 -0.0803 -0.0980 -0.0946 -0.0806 -0.0873 -0.0824 -0.0781 -0.0772
 -0.0867 -0.0815 -0.0748 -0.0824 -0.0702 -0.0729 -0.0751 0.3769 2.1054 1.6153 0.0116 -0.0620 -0.0739 -0.0827 -0.0977 -0.0940 -0.0845 -0.0894 -0.0830 -0.0745 -0.0751
 -0.0897 -0.0897 -0.0708 -0.0748 -0.0717 -0.0739 -0.0720 0.0027 0.6451 3.8977 0.9238 -0.0403 -0.0723 -0.0800 -0.0970 -0.0922 -0.0812 -0.0879 -0.0833 -0.0757 -0.0742
 -0.0873 -0.0885 -0.0751 -0.0897 -0.0717 -0.0742 -0.0760 -0.0336 0.1282 1.8280 3.6035 0.3409 -0.0598 -0.0797 -0.0961 -0.0928 -0.0809 -0.0867 -0.0793 -0.0766 -0.0793
 -0.0891 -0.0870 -0.0757 -0.0842 -0.0766 -0.0818 -0.0803 -0.0497 -0.0037 0.2542 2.2183 1.9345 0.1843 -0.0778 -0.0964 -0.0928 -0.0830 -0.0882 -0.0830 -0.0772 -0.0790
 -0.0891 -0.0882 -0.0748 -0.0916 -0.0778 -0.0815 -0.0906 -0.0742 -0.0574 0.0226 0.3580 1.1969 0.9808 0.0894 -0.0970 -0.0943 -0.0800 -0.0848 -0.0824 -0.0760 -0.0763
 -0.0894 -0.0858 -0.0754 -0.0769 -0.0739 -0.0803 -0.0894 -0.0772 -0.0748 -0.0653 0.0101 0.2203 0.6061 0.4141 -0.0912 -0.0922 -0.0784 -0.0861 -0.0824 -0.0763 -0.0748
 -0.0906 -0.0864 -0.0754 -0.0876 -0.0742 -0.0797 -0.0894 -0.0760 -0.0742 -0.0726 -0.0687 -0.0598 0.0128 0.1819 -0.0031 -0.0946 -0.0833 -0.0873 -0.0836 -0.0800 -0.0760
 -0.0885 -0.0882 -0.0757 -0.0879 -0.0754 -0.0800 -0.0885 -0.0772 -0.0748 -0.0742 -0.0720 -0.0751 -0.0720 -0.0745 -0.0858 -0.0952 -0.0809 -0.0891 -0.0854 -0.0769 -0.0769
 -0.0867 -0.0861 -0.0772 -0.0815 -0.0748 -0.0797 -0.0876 -0.0745 -0.0711 -0.0757 -0.0754 -0.0742 -0.0739 -0.0793 -0.0961 -0.0937 -0.0800 -0.0848 -0.0815 -0.0766 -0.0760
 -0.0854 -0.0858 -0.0751 -0.0812 -0.0735 -0.0775 -0.0864 -0.0714 -0.0671 -0.0732 -0.0766 -0.0751 -0.0766 -0.0800 -0.0967 -0.0925 -0.0815 -0.0876 -0.0848 -0.0754 -0.0739
 -0.0885 -0.0864 -0.0757 -0.0800 -0.0742 -0.0772 -0.0842 -0.0684 -0.0616 -0.0653 -0.0729 -0.0778 -0.0778 -0.0824 -0.0967 -0.0925 -0.0824 -0.0858 -0.0809 -0.0748 -0.0760
Attachment 5: TS-2017-07-06-14-30.rga
Attachment 6: TS-2017-07-07-10-10.rga
Attachment 7: TS-2017-07-07-12-50.rga
Attachment 8: Spare_Ta_LP_SIS.docx
Attachment 9: OMEGA_OM-DAQPRO-5300_10-Jul-17_10.27.49.dlp
Attachment 10: ITE-TM4-Ta#(Spare)-LP-SIS_Test_Stand.docx
  602   Tuesday, August 15, 2017, 08:41 Aaron SchmidtTa#55LP-SIS  42218-1142738-2In progressTBHT sight in test

Heat TGHT to 580 A & TGHT to  200 A for view port test.

Tuesday, August 15, 2017, 08:40: Roughing.

Tuesday, August 15, 2017, 08:59: Both Turbo pumps started.

Tuesday, August 15, 2017, 09:05: Ion gauge on. TP1B @ speed 42 KRPM.

Tuesday, August 15, 2017, 09:07: TP1A @ speed 38 KRPM.

Tuesday, August 15, 2017, 09:10: RGA on. Vacuum @ 9.14e-06T

Tuesday, August 15, 2017, 11:40: Check. Vacuum @ 6.76e-07T

Tuesday, August 15, 2017, 11:41: Waterlines opened. No response on RGA. Vacuum @ 6.49e-07T

Tuesday, August 15, 2017, 11:43: Power/sources on.

Tuesday, August 15, 2017, 13:57: Check. Vacuum @ 3.88e-07T.

Tuesday, August 15, 2017, 14:33: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.


Wednesday, August 16, 2017, 10:33: Auto-cooling TBHT to 100 A @ 4A per min.


Thursday, August 17, 2017, 07:05: TGHT / TBHT off. FC0 closed. IV4 Closed.

Thursday, August 17, 2017, 07:11: IV4 reset.

Thursday, August 17, 2017, 07:13: Power off

Thursday, August 17, 2017, 07:15 Waterlines purged


Thursday, August 24, 2017, 13:03: Waterlines ok (open). Power/source on.

Thursday, August 24, 2017, 13:10: Auto ramping TGHT to 580 A @ 2A per min.


Friday, August 25, 2017, 07:59: Open IV4. Open TMS : FC0.


Friday, September 08, 2017, 06:51: Auto ramping TGHT to 580 A @ 2A per min.


Monday, September 11, 2017, 10:55: Close TMS : FC0. Close IV4

Monday, September 11, 2017, 11:58: power/sources off

Monday, September 11, 2017, 11:00: RGA off

Monday, September 11, 2017, 11:01: Ion gauge, turbos off.

Monday, September 11, 2017, 11:59: Venting.

Monday, September 11, 2017, 13:17: Roughing.

Monday, September 11, 2017, 13:34: Turbo pumps on

Monday, September 11, 2017, 13:40: Ion gauge on.


Tuesday, September 12, 2017, 07:24: RGA, power, & sources on.

Tuesday, September 12, 2017, 07:25: Auto ramping TGHT to 580 A @ 2A per min.

Tuesday, September 12, 2017, 07:26: Open IV4. Open TMS : FC0.


Monday, September 18, 2017, 10:53: Close TMS : FC0. Close IV4

Monday, September 18, 2017, 10:57: Power/sources off.


Wednesday, September 20, 2017, 10:25: RGA, Ion gauge, Turbo pumps off. Waterlines purged.

Wednesday, September 20, 2017, 12:49: Venting.

  599   Monday, July 17, 2017, 10:04 Aaron Schmidt HP-SIS  42218-1042738-1 Target cleaning test HP-SIS Full Clean

 Target Test (Fully Cleaned vs Crystal Clean only).

For all tests:

-Start Turbo pumps & Ion gauge as soon as available on ISAC TEST STAND VACUUM Screen.

- Ensure previous test turbos are off by 8:15 and started venting by 9:15am to meet timeline.

-Start Roughing by 10:10am +/- 1 minute.

-TGHT / TBHT ramping starts @ 8:30am. 2A per min to 580A / 200A

ALL START TIMES MUST BE CONSISTENT.


***Pre-outgassed*** Heat Shield W.O# 42738-1: Includes insulators, new Al grounded 9-pin, bracket, & gripper arms. Blank/Plate. No source cup/EE.


HP-SIS Target # 42218-10: Full Clean. Crystal Clear, Acetone, & Methanol used.


Thursday, July 27, 2017, 09:49: HS removed from test stand. Exchange target.

Thursday, July 27, 2017, 10:07: Target # 42218-10 placed into test stand. Door open for 20 minutes.

Thursday, July 27, 2017, 10:10: Roughing.

Thursday, July 27, 2017, 10:10: Doors sealed for roughing.

Thursday, July 27, 2017, 10:13 BV1B OPEN. Opened @ 230T.

Thursday, July 27, 2017, 10:29: Both Turbo pumps started.

Thursday, July 27, 2017, 10:34: Ion gauge on. TP1B @ speed 42 KRPM.

Thursday, July 27, 2017, 10:37: TP1A @ speed 38 KRPM.

Thursday, July 27, 2017, 10:40: RGA on. Vacuum @ 1.42T.

Thursday, July 27, 2017,15:00: Check. Vacuum @ 5.51e-07T.

Thursday, July 27, 2017, 15:02: Waterlines opened. No response on RGA. Vacuum @ 5.42e-07T

Thursday, July 27, 2017, 15:05: Power on.


Friday, July 28, 2017, 08:30: Ramping started


Monday, July 31, 2017, 07:56: Ramping down

Monday, July 31, 2017, 14:25: Power supplies off. RGA off.


Tuesday, August 15, 2017, 06:48: Ion gauge, Turbo pumps off.

Tuesday, August 15, 2017, 06:50: power off. Waterlines purged.

Tuesday, August 15, 2017, 07:55: Venting.

Attachment 1: HP-SIS_Full_Clean_(Short).docx
Attachment 2: TS-2017-07-27-10-40.rga
  605   Wednesday, September 20, 2017, 14:11 Aaron SchmidtTa#54HP-SISITWTM242218-10 In progressConditioning Ta#54

Wednesday, September 20, 2017, 14:08: Roughing.

Wednesday, September 20, 2017, 14:28: Turbos on.

Wednesday, September 20, 2017, 14:33: Ion gauge on.

Wednesday, September 20, 2017, 14:39: RGA on.


Thursday, September 21, 2017, 08:43: Waterlines opened. No response on RGA. Vacuum ok @ 1.7e-07T

Thursday, September 21, 2017, 08:49: Power/sources on.

Thursday, September 21, 2017, 09:03: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .

Thursday, September 21, 2017, 09:04: IV4 Opened.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 25.1 3.85
 TBHT 25.4 3.04
 COIL + EE 24.9 4.94
 TGHT  25.1 5.59

Thursday, September 21, 2017, 09:08: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.


Friday, September 22, 2017, 13:42: Bias / EZL increased to 40 / 17.3 kV. Tune accordingly.

Mass scan ok

Friday, September 22, 2017, 14:33: Bias / EZL decreased to 30 / 13 kV. Tune accordingly.

Friday, September 22, 2017, 14:39: Auto-cooling @ 2A / min


Monday, September 25, 2017, 10:20: BIAS / EZL / EE @ 0 V.

Monday, September 25, 2017, 10:21:Power/sources off.

Monday, September 25, 2017, 10:32: RGA off. IV4 closed

Monday, September 25, 2017, 10:33: Ion gauge, turbos off.

Monday, September 25, 2017, 10:37: Waterlines purged.

Monday, September 25, 2017, 13:19: Venting.

Attachment 1: ER-580-200-30kV-m7.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1376 -0.0848 -0.0870 -0.0836 -0.0839 -0.0894 -0.0861 -0.0635 -0.0439 -0.0433 -0.0583 -0.0772 -0.0827 -0.0809 -0.0888 -0.0916 -0.0842 -0.0839 -0.0864 -0.1010 -0.0934
 -0.0790 -0.0790 -0.0787 -0.0769 -0.0803 -0.0815 -0.0845 -0.0790 -0.0577 -0.0427 -0.0485 -0.0769 -0.0806 -0.0793 -0.0851 -0.0845 -0.0778 -0.0800 -0.0833 -0.0946 -0.0894
 -0.0790 -0.0793 -0.0797 -0.0797 -0.0797 -0.0806 -0.0806 -0.0839 -0.0797 -0.0546 -0.0458 -0.0735 -0.0809 -0.0797 -0.0851 -0.0861 -0.0793 -0.0797 -0.0797 -0.0836 -0.0812
 -0.0790 -0.0769 -0.0854 -0.0949 -0.0809 -0.0812 -0.0809 -0.0790 -0.0861 -0.0748 -0.0537 -0.0769 -0.0839 -0.0793 -0.0870 -0.0891 -0.0760 -0.0772 -0.0754 -0.0769 -0.0803
 -0.0787 -0.0797 -0.0833 -0.0873 -0.0867 -0.0818 -0.0821 -0.0821 -0.0848 -0.0803 -0.0729 -0.0735 -0.0760 -0.0793 -0.0839 -0.0891 -0.0815 -0.0809 -0.0784 -0.0784 -0.0787
 -0.0742 -0.0775 -0.0668 -0.0311 -0.0766 -0.0845 -0.0854 -0.0842 -0.0861 -0.0836 -0.0836 -0.0793 -0.0818 -0.0803 -0.0845 -0.0891 -0.0815 -0.0806 -0.0836 -0.0790 -0.0793
 -0.0787 -0.0757 -0.0781 0.3857 0.1312 -0.0653 -0.0882 -0.0839 -0.0873 -0.0842 -0.0827 -0.0815 -0.0800 -0.0800 -0.0839 -0.0888 -0.0790 -0.0763 -0.0806 -0.0800 -0.0797
 -0.0760 -0.0769 -0.0806 -0.0201 1.2869 0.1776 -0.0616 -0.0797 -0.0858 -0.0812 -0.0797 -0.0806 -0.0818 -0.0781 -0.0851 -0.0876 -0.0772 -0.0772 -0.0766 -0.0793 -0.0787
 -0.0772 -0.0781 -0.0781 -0.0925 0.7306 1.4966 0.2347 -0.0467 -0.0781 -0.0751 -0.0748 -0.0800 -0.0809 -0.0766 -0.0848 -0.0891 -0.0775 -0.0797 -0.0806 -0.0775 -0.0787
 -0.0766 -0.0775 -0.0803 -0.0958 -0.0534 1.9830 1.6452 0.2716 -0.0613 -0.0760 -0.0769 -0.0778 -0.0800 -0.0787 -0.0861 -0.0903 -0.0793 -0.0797 -0.0870 -0.0781 -0.0757
 -0.0714 -0.0763 -0.0766 -0.0940 -0.0674 0.1242 2.7417 2.0587 0.1642 -0.0760 -0.0772 -0.0778 -0.0787 -0.0803 -0.0848 -0.0903 -0.0781 -0.0797 -0.0854 -0.0793 -0.0787
 -0.0775 -0.0769 -0.0705 -0.0897 -0.0595 -0.0275 0.4175 2.2665 2.2058 0.0433 -0.0778 -0.0809 -0.0800 -0.0763 -0.0833 -0.0888 -0.0766 -0.0793 -0.0812 -0.0787 -0.0797
 -0.0793 -0.0769 -0.0797 -0.0925 -0.0635 -0.0409 0.0406 0.3757 1.6052 2.2055 -0.0446 -0.0793 -0.0830 -0.0787 -0.0839 -0.0891 -0.0778 -0.0757 -0.0806 -0.0793 -0.0806
 -0.0787 -0.0769 -0.0797 -0.0912 -0.0790 -0.0653 -0.0391 0.0531 0.4321 1.3553 1.2790 -0.0763 -0.0812 -0.0818 -0.0842 -0.0897 -0.0793 -0.0781 -0.0806 -0.0784 -0.0790
 -0.0818 -0.0787 -0.0797 -0.0974 -0.0784 -0.0778 -0.0763 -0.0555 0.0208 0.3412 1.2677 0.3552 -0.0790 -0.0781 -0.0839 -0.0894 -0.0797 -0.0787 -0.0763 -0.0769 -0.0757
 -0.0763 -0.0797 -0.0842 -0.0922 -0.0769 -0.0757 -0.0787 -0.0763 -0.0803 -0.0302 0.2393 0.3601 -0.0763 -0.0760 -0.0833 -0.0885 -0.0809 -0.0781 -0.0842 -0.0778 -0.0766
 -0.0769 -0.0766 -0.0769 -0.0894 -0.0684 -0.0769 -0.0769 -0.0769 -0.0766 -0.0745 -0.0696 -0.0696 -0.0812 -0.0772 -0.0824 -0.0851 -0.0754 -0.0772 -0.0842 -0.0781 -0.0781
 -0.0757 -0.0766 -0.0790 -0.0885 -0.0546 -0.0696 -0.0748 -0.0775 -0.0772 -0.0757 -0.0766 -0.0784 -0.0824 -0.0760 -0.0830 -0.0851 -0.0751 -0.0775 -0.0781 -0.0787 -0.0797
 -0.0778 -0.0778 -0.0797 -0.0888 -0.0494 -0.0504 -0.0690 -0.0775 -0.0815 -0.0739 -0.0754 -0.0778 -0.0827 -0.0797 -0.0867 -0.0861 -0.0778 -0.0775 -0.0815 -0.0769 -0.0778
 -0.0790 -0.0778 -0.0781 -0.0906 -0.0497 -0.0342 -0.0510 -0.0705 -0.0824 -0.0757 -0.0769 -0.0772 -0.0806 -0.0803 -0.0842 -0.0870 -0.0775 -0.0790 -0.0830 -0.0757 -0.0766
 -0.0809 -0.0793 -0.0781 -0.0900 -0.0491 -0.0293 -0.0314 -0.0531 -0.0760 -0.0778 -0.0781 -0.0787 -0.0790 -0.0784 -0.0839 -0.0858 -0.0769 -0.0763 -0.0803 -0.0775 -0.0763
Attachment 2: ER-580-200-30kV-m38.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1535 -0.1013 -0.0998 -0.0961 -0.1013 -0.0931 -0.0974 -0.0989 -0.0989 -0.0992 -0.1007 -0.1004 -0.0992 -0.0949 -0.0931 -0.0897 -0.0903 -0.0952 -0.1025 -0.0989 -0.1019
 -0.0961 -0.0961 -0.0977 -0.0919 -0.0946 -0.0888 -0.0916 -0.0928 -0.0916 -0.0946 -0.0952 -0.0980 -0.0955 -0.0909 -0.0900 -0.0839 -0.0827 -0.0909 -0.0943 -0.0940 -0.0955
 -0.0958 -0.0919 -0.0891 -0.0851 -0.0894 -0.0858 -0.0897 -0.0925 -0.0861 -0.0851 -0.0903 -0.0851 -0.0867 -0.0873 -0.0888 -0.0864 -0.0848 -0.0888 -0.0848 -0.0894 -0.0864
 -0.0949 -0.0876 -0.0775 -0.0864 -0.0775 -0.0848 -0.0894 -0.0922 -0.0842 -0.0790 -0.0784 -0.0781 -0.0745 -0.0742 -0.0897 -0.0861 -0.0784 -0.0891 -0.0861 -0.0845 -0.0800
 -0.0946 -0.0873 -0.0815 -0.0818 -0.0769 -0.0839 -0.0925 -0.0940 -0.0809 -0.0806 -0.0778 -0.0797 -0.0812 -0.0778 -0.0888 -0.0851 -0.0851 -0.0897 -0.0824 -0.0800 -0.0836
 -0.0931 -0.0845 -0.0848 -0.0821 -0.0803 -0.0848 -0.0934 -0.0928 -0.0836 -0.0812 -0.0797 -0.0784 -0.0787 -0.0809 -0.0885 -0.0848 -0.0836 -0.0873 -0.0833 -0.0769 -0.0800
 -0.0925 -0.0854 -0.0815 -0.0775 -0.0797 -0.0858 -0.0912 -0.0909 -0.0833 -0.0806 -0.0809 -0.0827 -0.0833 -0.0797 -0.0900 -0.0848 -0.0797 -0.0864 -0.0839 -0.0751 -0.0784
 -0.0949 -0.0858 -0.0787 -0.0818 -0.0775 -0.0821 -0.0894 -0.0919 -0.0815 -0.0800 -0.0763 -0.0769 -0.0818 -0.0778 -0.0891 -0.0858 -0.0800 -0.0821 -0.0809 -0.0818 -0.0778
 -0.0949 -0.0870 -0.0793 -0.0812 -0.0769 -0.0839 -0.0803 -0.0873 -0.0815 -0.0772 -0.0754 -0.0787 -0.0800 -0.0769 -0.0894 -0.0870 -0.0836 -0.0845 -0.0824 -0.0824 -0.0790
 -0.0937 -0.0864 -0.0778 -0.0900 -0.0797 -0.0824 -0.0858 -0.0693 -0.0650 -0.0760 -0.0760 -0.0772 -0.0772 -0.0766 -0.0894 -0.0870 -0.0793 -0.0861 -0.0842 -0.0797 -0.0815
 -0.0934 -0.0818 -0.0772 -0.0864 -0.0778 -0.0809 -0.0891 -0.0824 -0.0302 -0.0662 -0.0760 -0.0787 -0.0763 -0.0790 -0.0876 -0.0848 -0.0790 -0.0824 -0.0812 -0.0812 -0.0772
 -0.0958 -0.0830 -0.0812 -0.0769 -0.0769 -0.0812 -0.0867 -0.0900 -0.0488 0.0195 -0.0723 -0.0815 -0.0778 -0.0763 -0.0879 -0.0839 -0.0775 -0.0824 -0.0821 -0.0797 -0.0772
 -0.0949 -0.0824 -0.0787 -0.0821 -0.0769 -0.0806 -0.0854 -0.0903 -0.0735 -0.0204 -0.0235 -0.0766 -0.0778 -0.0778 -0.0885 -0.0842 -0.0778 -0.0821 -0.0839 -0.0754 -0.0803
 -0.0937 -0.0854 -0.0803 -0.0821 -0.0760 -0.0827 -0.0873 -0.0906 -0.0778 -0.0601 -0.0107 -0.0446 -0.0766 -0.0781 -0.0885 -0.0867 -0.0809 -0.0806 -0.0830 -0.0812 -0.0818
 -0.0922 -0.0824 -0.0790 -0.0836 -0.0781 -0.0867 -0.0903 -0.0919 -0.0818 -0.0739 -0.0528 -0.0388 -0.0665 -0.0766 -0.0861 -0.0854 -0.0815 -0.0824 -0.0812 -0.0818 -0.0815
 -0.0922 -0.0803 -0.0769 -0.0806 -0.0787 -0.0864 -0.0882 -0.0937 -0.0833 -0.0793 -0.0739 -0.0757 -0.0699 -0.0739 -0.0851 -0.0839 -0.0818 -0.0824 -0.0809 -0.0775 -0.0787
 -0.0925 -0.0815 -0.0772 -0.0806 -0.0778 -0.0839 -0.0888 -0.0928 -0.0818 -0.0797 -0.0775 -0.0797 -0.0797 -0.0769 -0.0854 -0.0845 -0.0787 -0.0781 -0.0806 -0.0818 -0.0763
 -0.0952 -0.0821 -0.0769 -0.0851 -0.0772 -0.0803 -0.0879 -0.0934 -0.0778 -0.0772 -0.0775 -0.0793 -0.0778 -0.0784 -0.0873 -0.0848 -0.0790 -0.0775 -0.0806 -0.0806 -0.0772
 -0.0940 -0.0839 -0.0769 -0.0821 -0.0775 -0.0824 -0.0882 -0.0909 -0.0809 -0.0757 -0.0751 -0.0769 -0.0803 -0.0763 -0.0876 -0.0864 -0.0775 -0.0812 -0.0827 -0.0803 -0.0775
 -0.0946 -0.0803 -0.0775 -0.0861 -0.0800 -0.0842 -0.0882 -0.0925 -0.0818 -0.0757 -0.0763 -0.0766 -0.0772 -0.0787 -0.0876 -0.0845 -0.0809 -0.0815 -0.0821 -0.0790 -0.0775
 -0.0946 -0.0790 -0.0763 -0.0787 -0.0781 -0.0812 -0.0873 -0.0931 -0.0797 -0.0763 -0.0778 -0.0778 -0.0793 -0.0769 -0.0854 -0.0833 -0.0809 -0.0809 -0.0809 -0.0751 -0.0766
Attachment 3: OMEGA_OM-DAQPRO-5300_27-Sep-17_09.18.45.dlp
Attachment 4: Ta#54.docx
Attachment 5: TS-2017-09-20-14-37.rga
Attachment 6: ITW-TM2-Ta#54-HP-SIS_Test_Stand.docx
  529   Monday, August 08, 2016, 09:48 Aaron Schmidt    41882-441751-1 HS Test Full Clean

Heat Shield Test (Fully Cleaned).

For all 6 tests:

-Start Turbo pumps & Ion gauge as soon as available on ISAC TEST STAND VACUUM Screen.

-Start Roughing by 10:10am +/- 1 minute and ensure previous HS started venting by 9:15am to meet timeline.


HS# 41751-1 . OR, Acetone, Methanol. TGHT only.

No insulators, bracket, or gripper arms. Blank/Plate. No source cup.

Monday, August 08, 2016, 09:58: Ta#49 removed from test stand.

Monday, August 08, 2016, 10:07: 41751-1 placed into test stand. Door open for 9 minutes.

Monday, August 08, 2016, 10:10: Roughing.

Monday, August 08, 2016, 10:18: BV1B OPEN. Opened @ 88T.

Monday, August 08, 2016, 10:40: Both Turbo pumps started.

Monday, August 08, 2016, 10:45: Ion gauge on. TP1B @ speed 42 KRPM.

Monday, August 08, 2016, 10:48: TP1A @ speed 38 KRPM.

Monday, August 08, 2016, 11:45: RGA on. Vacuum @ 2.02E-6T


Tuesday, August 09, 2016, 08:14: Water lines opened. No response on RGA. Vacuum stable @ 1.99E-7T

Tuesday, August 09, 2016, 08:19: Power/sources on.

Tuesday, August 09, 2016, 08:30: TGHT auto-ramping to 580A @ 2A/min. Vacuum stable @ 1.99E-7T

Tuesday, August 09, 2016, 11:15: Screen shots and check made @ 329-330A

Tuesday, August 09, 2016, 13:20: TGHT @ 580A. Vacuum @4.23E-7T

Tuesday, August 09, 2016, 16:00: Auto-cooling @ 2A/min. Vacuum @ 3.77E-7T


Wednesday, August 10, 2016, 08:00: Vacuum @ 1.70E-7T. RGA off. Ion gauge off. Turbos off.

Wednesday, August 10, 2016, 09:14: Venting.

Attachment 1: HS_41751-1_TEST.docx
Attachment 2: TS-2016-08-08-11-45.rga
Attachment 3: 20160824_144103.jpg
20160824_144103.jpg
Attachment 4: 20160824_144106.jpg
20160824_144106.jpg
  530   Wednesday, August 10, 2016, 08:09 Aaron Schmidt    41882-441751-2 HS Test Full Clean

Heat Shield Test (Fully Cleaned).

For all 6 tests:

-Start Turbo pumps & Ion gauge as soon as available on ISAC TEST STAND VACUUM Screen.

-Start Roughing by 10:10am +/- 1 minute and ensure previous HS started venting by 9:15am to meet timeline.


HS# 41751-2 . OR, Acetone, Methanol. TGHT only.

No insulators, bracket, or gripper arms. Blank/Plate. No source cup.

Wednesday, August 10, 2016, 09:47: 41751-1 removed from test stand. Exchange target & End plates to 41751-2

Wednesday, August 10, 2016, 10:07: 41751-2 placed into test stand. Door open for 20 minutes.

Wednesday, August 10, 2016, 10:10: Roughing.

Wednesday, August 10, 2016, 10:13: Doors finally sealed for roughing.

Wednesday, August 10, 2016, 10:23: BV1B OPEN. Opened @ 59T.

Wednesday, August 10, 2016, 10:45: Both Turbo pumps started.

Wednesday, August 10, 2016, 10:50: Ion gauge on. TP1B @ speed 42 KRPM.

Wednesday, August 10, 2016, 10:53: TP1A @ speed 38 KRPM

Wednesday, August 10, 2016, 11:45: RGA on. Vacuum @ 3.56E-6T


Thursday, August 11, 2016, 08:16: Water lines opened. No response on RGA. Vacuum stable @ 2.34E-7T

Thursday, August 11, 2016, 08:19: Power/sources on

Thursday, August 11, 2016, 08:30: TGHT auto-ramping to 580A @ 2A/min. Vacuum stable @ 2.36E-7T

Thursday, August 11, 2016, 11:15: Screen shots and check made @ 329-330A.

Thursday, August 11, 2016, 13:25: TGHT @ 580A. Vacuum @4.16E-7T

Thursday, August 11, 2016, 15:54: Screen shots and check made @ 580A. Vacuum @4.43E-7T and increasing. Cool in Am.


Friday, August 12, 2016, 08:30: Auto-cooling started @ 2A/min. Vacuum @ 3.88E-7T

Friday, August 12, 2016, 13:20: Cooling Complete. Vacuum @ 2.16E-7T.

Friday, August 12, 2016, 13:33: Water lines purged. Power/sources off.


Monday, August 15, 2016, 08:07: Turbo pumps, Ion gauge, and RGA off.

Monday, August 15, 2016, 09:14: Venting.

Attachment 1: HS_41751-2_TEST.docx
Attachment 2: TS-2016-08-10-11-45.rga
Attachment 3: 20160824_143539.jpg
20160824_143539.jpg
  532   Monday, August 15, 2016, 08:09 Aaron Schmidt    41882-441751-4 HS Test OR Clean

Heat Shield Test (OR Cleaned).

For all 6 tests:

-Start Turbo pumps & Ion gauge as soon as available on ISAC TEST STAND VACUUM Screen.

-Start Roughing by 10:10am +/- 1 minute and ensure previous HS started venting by 9:15am to meet timeline.


HS# 41751-4 . OR only. TGHT only.

No insulators, bracket, or gripper arms. Blank/Plate. No source cup.

Monday, August 15, 2016, 09:45: 41751-2 removed from test stand. Exchange target & End plates to 41751-2

Monday, August 15, 2016, 10:07: 41751-4 placed into test stand. Door open for 22 minutes.

Monday, August 15, 2016, 10:10: Roughing.

Monday, August 15, 2016, 10:12: Doors sealed for roughing.

Monday, August 15, 2016, 10:21: BV1B OPEN. Opened @ 104T.

Monday, August 15, 2016, 10:44: Both Turbo pumps started.

Monday, August 15, 2016, 10:50: Ion gauge on. TP1B @ speed 42 KRPM.

Monday, August 15, 2016, 10:52: TP1A @ speed 38 KRPM.

Monday, August 15, 2016, 11:45: RGA on. Vacuum @ 4.20E-6T.


Tuesday, August 16, 2016, 08:15: Water lines opened. No response on RGA. Vacuum stable @ 2.43E-7T

Tuesday, August 16, 2016, 08:20: Power/sources on

Tuesday, August 16, 2016, 08:30: TGHT auto-ramping to 580A @ 2A/min. Vacuum stable @ 2.45E-7T

Tuesday, August 16, 2016, 11:15: Screen shots and check made @ 329-330A.

Tuesday, August 16, 2016, 13:20: TGHT @ 580A. Vacuum @4.16E-7T.

Tuesday, August 16, 2016, 16:00: Auto-cooling started @ 2A/min. Vacuum @ 4.25E-7T


Wednesday, August 17, 2016, 08:11: Vacuum @ 1.90E-7T. RGA off. Ion gauge off. Turbos off.

Wednesday, August 17, 2016, 08:16: Water lines purged. Power/sources off.

Wednesday, August 17, 2016, 09:09: Venting.

Attachment 1: HS_41751-4_TEST.docx
Attachment 2: TS-2016-08-15-11-45.rga
Attachment 3: 20160824_143550.jpg
20160824_143550.jpg
  533   Wednesday, August 17, 2016, 08:19 Aaron Schmidt    41882-441751-5 HS Test OR Clean

Heat Shield Test (OR Cleaned).

For all 6 tests:

-Start Turbo pumps & Ion gauge as soon as available on ISAC TEST STAND VACUUM Screen.

-Start Roughing by 10:10am +/- 1 minute and ensure previous HS started venting by 9:15am to meet timeline.


HS# 41751-5 . OR only. TGHT only.

No insulators, bracket, or gripper arms. Blank/Plate. No source cup.

Wednesday, August 17, 2016, 09:45: 41751-4 removed from test stand. Exchange target & End plates to 41751-5

Wednesday, August 17, 2016, 10:08: 41751-5 placed into test stand. Door open for 22 minutes.

Wednesday, August 17, 2016, 10:10: Roughing.

Wednesday, August 17, 2016, 10:10: Doors sealed for roughing.

Wednesday, August 17, 2016, 10:18: BV1B OPEN. Opened @ 136T.

Wednesday, August 17, 2016, 10:41: Both Turbo pumps started.

Wednesday, August 17, 2016, 10:46: Ion gauge on. TP1B @ speed 42 KRPM.

Wednesday, August 17, 2016, 10:50: TP1A @ speed 38 KRPM.

Wednesday, August 17, 2016, 11:47: RGA on. Vacuum @ 3.65E-6T.


Thursday, August 18, 2016, 08:19: Water lines opened. No response on RGA. Vacuum stable @ 2.53E-7T.

Thursday, August 18, 2016, 08:22: Power/sources on.

Thursday, August 18, 2016, 08:30: TGHT auto-ramping to 580A @ 2A/min. Vacuum stable @ 2.45E-7T.

Thursday, August 18, 2016, 11:39: Screen shots and check made @ 380-381A.

Thursday, August 18, 2016, 13:20: TGHT @ 580A. Vacuum @4.21E-7T.


Friday, August 19, 2016, 08:30: Auto-cooling started @ 2A/min. Vacuum @ 3.88E-7T

Friday, August 19, 2016, 13:20: Cooling Complete. Vacuum @ 2.25E-7T.

Friday, August 19, 2016, 13:42: Water lines purged. Power/sources off.

Monday, August 22, 2016, 08:17: Vacuum @ 1.37E-7T. RGA off. Ion gauge off. Turbos off.

Monday, August 22, 2016, 09:16: Venting.

Flat Head screw on bottom end plate stripped. Warping found on cup Blank/plate. Discoloration noted on heat shield

Attachment 1: HS_41751-5_TEST.docx
Attachment 2: TS-2016-08-17-11-45.rga
Attachment 3: 20160824_143555.jpg
20160824_143555.jpg
Attachment 4: 20160824_143558.jpg
20160824_143558.jpg
  534   Monday, August 22, 2016, 08:20 Aaron Schmidt    41882-441751-3 HS Test Full Clean

Heat Shield Test (Fully Cleaned).

For all 6 tests:

-Start Turbo pumps & Ion gauge as soon as available on ISAC TEST STAND VACUUM Screen.

-Start Roughing by 10:10am +/- 1 minute and ensure previous HS started venting by 9:15am to meet timeline.


HS# 41751-3 . OR, Acetone, Methanol. TGHT only.

No insulators, bracket, or gripper arms. Blank/Plate (Warped from 41751-5). No source cup.

Monday, August 22, 2016, 09:50: 41751-5 removed from test stand. Exchange target & End plates to 41751-3

Monday, August 22, 2016, 10:15: 41751-3 placed into test stand. Door open for 25 minutes.

Monday, August 22, 2016, 10:18: Roughing.

Monday, August 22, 2016, 10:19: Doors sealed for roughing.

Monday, August 22, 2016, 10:25: BV1B OPEN. Opened @ 151T.

Monday, August 22, 2016, 10:50: Both Turbo pumps started.

Monday, August 22, 2016, 10:54: Ion gauge on. TP1B @ speed 42 KRPM.

Monday, August 22, 2016, 10:58: TP1A @ speed 38 KRPM.

Monday, August 22, 2016, 11:50: RGA on. Vacuum @ 3.91E-6T.


Tuesday, August 23, 2016, 08:16: Water lines opened. No response on RGA. Vacuum stable @ 2.45E-7T

Tuesday, August 23, 2016, 08:19: Power/sources on.

Tuesday, August 23, 2016, 08:30: TGHT auto-ramping to 580A @ 2A/min. Vacuum @ 2.45E-7T.

Tuesday, August 23, 2016, 13:20: TGHT @ 580A. Vacuum @3.79E-7T.

Tuesday, August 23, 2016, 16:06: Auto-cooling started @ 2A/min. Vacuum @ 3.96E-7T.


Wednesday, August 24, 2016, 08:08: RGA off. Ion gauge and Turbos off.

Wednesday, August 24, 2016, 08:14; Power/sources off.

Wednesday, August 24, 2016, 08:25: Waterlines purged.

Wednesday, August 24, 2016, 09:14: Venting.

Attachment 1: HS_41751-3_TEST.docx
Attachment 2: TS-2016-08-22-11-50.rga
Attachment 3: 20160824_143544.jpg
20160824_143544.jpg
  535   Wednesday, August 24, 2016, 09:16 Aaron Schmidt    41882-441751-6 HS Test OR Clean

Heat Shield Test (OR only).

For all 6 tests:

-Start Turbo pumps & Ion gauge as soon as available on ISAC TEST STAND VACUUM Screen.

-Start Roughing by 10:10am +/- 1 minute and ensure previous HS started venting by 9:15am to meet timeline.


HS# 41751-6 . OR only.

No insulators, bracket, or gripper arms. Blank/Plate (new one, old plate was too warped). No source cup.

Wednesday, August 24, 2016, 09:46 : 41751-3 removed from test stand. Exchange target & End plates to 41751-6

Wednesday, August 24, 2016, 10:08: 41751-6 placed into test stand. Door open for 22 minutes.

Wednesday, August 24, 2016, 10:10: Roughing.

Wednesday, August 24, 2016, 10:10: Doors sealed for roughing.

Wednesday, August 24, 2016, 10:18: BV1B OPEN. Opened @ 79T.

Wednesday, August 24, 2016, 10:41: Both Turbo pumps started.

Wednesday, August 24, 2016, 10:46: Ion gauge on. TP1B @ speed 42 KRPM.

Wednesday, August 24, 2016, 10:49: TP1A @ speed 38 KRPM.

Wednesday, August 24, 2016, 11:45: RGA on. Vacuum @ 3.38E-6T.


Thursday, August 25, 2016, 08:15: Water lines opened. No response on RGA. Vacuum stable @ 2.36E-7T

Thursday, August 25, 2016, 08:20: Power/sources on.

Thursday, August 25, 2016, 08:30: TGHT auto-ramping to 580A @ 2A/min. Vacuum @ 2.36E-7T.

Thursday, August 25, 2016, 11:21: Screen shots and check made @ 340-344A.

Thursday, August 25, 2016, 13:20: TGHT @ 580A. Vacuum @3.70E-7T.


Friday, August 26, 2016, 08:30: Auto-cooling started @ 2A/min. Vacuum @ 3.53E-7T

Friday, August 26, 2016, 13:20: Cooling Complete. Vacuum @ 1.97E-7T


Monday, August 29, 2016, 08:19: RGA off. Ion gauge and Turbos off.

Monday, August 29, 2016, 09:28: Venting.

Monday, August 29, 2016, 09:35: Power/sources off. Waterlines purged.

Attachment 1: HS_41751-6_TEST.docx
Attachment 2: TS-2016-08-24-11-45.rga
Attachment 3: OMEGA_OM-DAQPRO-5300_29-Aug-16_09.47.01.dlp
Attachment 4: 20160829_110707.jpg
20160829_110707.jpg
Attachment 5: 20160829_110713.jpg
20160829_110713.jpg
Attachment 6: 20160829_110716.jpg
20160829_110716.jpg
Attachment 7: 20160829_110719.jpg
20160829_110719.jpg
  564   Wednesday, January 25, 2017, 09:48 Aaron Schmidt    41882-441751-6DoneOut gas Heat Shield for Target cleaning test

No EE. Insulators, gripper arms, and blank EE plate included

-New grounded 9 pins included in brackets

-TGHT included to out gas Heat shield


Wednesday, January 25, 2017, 09:45: Roughing

Wednesday, January 25, 2017, 10:04: Turbo pumps started.

Wednesday, January 25, 2017, 11:26: Ion gauge & RGA on.


Thursday, January 26, 2017, 08:25: RGA restarted. (Fatal communication error)

Thursday, January 26, 2017, 08:32: Water lines opened. No response on RGA. Vacuum stable @ 1.44e-07T.

Thursday, January 26, 2017, 08:35: Power/sources on.

Thursday, January 26, 2017, 08:37: Auto-ramping TGHT to 580A @ 2A per min. Vacuum @ 1.44e-07T.

  Device Temp
[C]
 Flow
[lpm]
 HS + MSP 23.5 4.02
 TBHT 23.3 3.13
 COIL + EE 23.3 5.18
 TGHT  23.4 5.78

Thursday, January 26, 2017, 13:11: RGA restarted. (Fatal communication error)

Thursday, January 26, 2017, 15:20: RGA restarted. (Fatal communication error)

Thursday, January 26, 2017, 15:26: RGA being shut down. (Fatal communication error)

Thursday, January 26, 2017, 15:31: RGA Restarted.


Friday, January 27, 2017, 08:06: RGA running ok.

  Device Temp
[C]
 Flow
[lpm]
 HS + MSP 30.2 4.06
 TBHT 21.2 3.14
 COIL + EE 21.4 5.11
 TGHT  23.3 5.78

Friday, January 27, 2017, 15:23: Auto-cooling started @ 2A per min.


Monday, January 30, 2017, 08:02: RGA off. Ion Gauge off. Turbo pumps off.

Monday, January 30, 2017, 08:05: Power/sources off. Water lines purged.

Monday, January 30, 2017, 09:09: Venting.

Attachment 1: OMEGA_OM-DAQPRO-5300_30-Jan-17_09.03.56.dlp
Attachment 2: Target_cleaning_test_HS_out_gassing.docx
Attachment 3: TS-2017-01-25-11-25.rga
Attachment 4: TS-2017-01-26-08-25.rga
Attachment 5: TS-2017-01-26-15-20.rga
Attachment 6: TS-2017-01-26-15-30.rga
Attachment 7: TS-2017-01-26-15-50.rga
  572   Friday, March 10, 2017, 15:42 Aaron SchmidtUC#19   41882-441751-2In progressConditioning Heat shield for UC#19

Friday, March 10, 2017, 15:40: Roughing.

Friday, March 10, 2017, 15:58: Turbo pumps started.

Friday, March 10, 2017, 16:19: Ion gauge on.

Friday, March 10, 2017, 16:22: RGA on.


Monday, March 13, 2017, 09:20: Water lines opened. No response on RGA. Vacuum @ 1.99e-07T

Monday, March 13, 2017, 09:38: Power/sources on

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 24.6 3.84
 TBHT 24.4 3.00
 COIL + EE 24.4 4.89
 TGHT  24.6 5.56

Monday, March 13, 2017, 09:42: Auto ramping TGHT to 580A @ 2A per min.

Monday, March 13, 2017, 11:58: Water: 95 PSI on Supply. 10 PSI on Return.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 26.1 3.82
 TBHT 26.3 3.20
 COIL + EE 24.7 4.95
 TGHT  25.4 5.56

 


Tuesday, March 14, 2017, 08:55: Vacuum @ 3.26e-07T

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 34.0 3.85
 TBHT 25.9 3.11
 COIL + EE 24.9 4.98
 TGHT  26.8 5.57

Tuesday, March 14, 2017, 15:54:

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 34.1 3.82
 TBHT 25.2 3.07
 COIL + EE 25.0 4.94
 TGHT  27.1 5.58

Tuesday, March 14, 2017, 15:55: Auto-cooling started @ 2A per min


Tuesday, March 21, 2017, 09:30: Power/Sources off.

Tuesday, March 21, 2017, 14:11: RGA off

Tuesday, March 21, 2017, 14:13: Ion gauge & Turbo pumps off.


by Anders

Wednesday, March 22, 2017, 08:47: Venting

Wednesday, March 22, 2017, 09:3: HS removed

Attachment 1: TS-2017-03-10-16-20.rga
Attachment 2: OMEGA_OM-DAQPRO-5300_22-Mar-17_09.05.23.dlp
  597   Monday, July 17, 2017, 09:55 Aaron Schmidt    41882-442738-1In progressOut Gas HS for Target Cleaning Test

-No EE. Insulators, gripper arms, and blank EE plate included

-New grounded 9 pins included in brackets

-TGHT included to out gas Heat shield


Monday, July 17, 2017, 10:48: Roughing.

Monday, July 17, 2017, 11:08: Turbo pumps on.

Monday, July 17, 2017, 11:17: Ion gauge & RGA on.


Tuesday, July 18, 2017, 08:17: Water lines opened. No response on RGA. Vacuum stable @ 1.79e-07T.

Tuesday, July 18, 2017, 08:20: Power/sources on.

  Device Temp
[C]
 Flow
[lpm]
 HS + MSP 27.1 3.78
 TBHT 26.9 2.97
 COIL + EE 26.9 4.94
 TGHT 26.9 5.48

Tuesday, July 18, 2017, 08:25: Auto-ramping TGHT to 580A @ 2A per min. Vacuum @ 1.44e-07T.


 

  Device Temp
[C]
 Flow
[lpm]
 HS + MSP 36.4 3.78
 TBHT 27.1 2.97
 COIL + EE 27.4 4.84
 TGHT 29.5 5.48

Wednesday, July 19, 2017, 08:00: Auto cooling TGHT.


Monday, July 24, 2017, 07:38: Power/sources off.

Monday, July 24, 2017, 07:40: Waterlines purged

Monday, July 24, 2017, 07:42: RGA off. Ion gauge off. Turbos off.

Monday, July 24, 2017, 09:10: Venting.

Attachment 1: OMEGA_OM-DAQPRO-5300_24-Jul-17_10.23.02.dlp
Attachment 2: Target_Cleaning_Test_HS_Outgassing.docx
Attachment 3: TS-2017-07-17-11-17.rga
ELOG V2.9.2-2455