Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  Test-Stand, Page 8 of 53  Not logged in ELOG logo
ID Date Author Target/No.# Source Station Module Target Oven W.O#down Heat Shield W.O# Status Subject
  660   Tuesday, May 08, 2018, 10:36 Aaron Schmidt LP-FEBIAD  41790-942752-5In progressC-seal Test

2 C-Seals installed with SS springs on water blocks. The SS springs are also used in the module.

Long pkg seal (A):

T: 0.098" x 0.429" Diameter.

Installed on line FS#5 , Block L, Heat shield cooling block. (Top seal, bottom seal of 2 is still the rubber O-ring)

Short pkg seal (B):

T: 0.095"  x 0.424" Diameter.

Installed on line FS#3 , Block G, Coil cooling block.

All the waterlines are purged and shut. Open individually and monitor for leaks.


Tuesday, May 08, 2018, 10:27: Roughing

Tuesday, May 08, 2018, 10:31: Both BV1A & BV1B are open.

Tuesday, May 08, 2018, 10:57: Turbo Pumps started.

Tuesday, May 08, 2018, 11:03: Ion gauge on.

Tuesday, May 08, 2018, 11:15: RGA started.

Tuesday, May 08, 2018, 13:30: Vacuum @ 3.14e-06T


Wednesday, May 09, 2018, 06:48: Vacuum 5.13E-07T. RGA ok. 

Wednesday, May 09, 2018, 06:50: Hook up Leak Detector. Leak detector # 5 (DRY) from Vacuum Group. Startup

Baseline Reading:

TP pressure 0.0e-04

Leak rate: Below 0.0e-09 atm-cc/sec

Valve to chamber opened. LD connection clear.

LD: 1.3e-08 atm-cc/sec

He TEST:

Wednesday, May 09, 2018, 07:00: Purge both lines with He. Do not pressurize.

Vacuum @ 5.13e-07T. RGA status:

Water @ 3.22e-007

He:-1.93e-009

LD: 1.3e-08 atm-cc/sec

Wednesday, May 09, 2018, 07:15: Open line FS#5 to heat shield. Pressurize to 30 PSI.

Wednesday, May 09, 2018, 07:40: Vacuum @ 5.02e-07T. RGA status:

Water @ 3.13e-007

He:-1.67e-009

LD: 9.3e-08 atm-cc/sec. Purge and close FS#5.

Wednesday, May 09, 2018, 07:45: Open line FS#3 to Coil block. Pressurize to 30 PSI.

08:10: Vacuum @ 4.9e-07T. RGA status:

Water @ 3.08e-007

He:-1.61e-009

LD: 1.5e-07 atm-cc/sec

Wednesday, May 09, 2018, 08:15: Depressurize lines. Connect HPA. 

Wednesday, May 09, 2018, 08:20: Vacuum @ 4.76e-07T. RGA status:

Water @ 3.05e-007

He:-1.68e-009

LD: 1.7e-07 atm-cc/sec

08:25: LD: 1.7e-07 atm-cc/sec. Close Valve & disconnect LD


Water & Cycling TEST:

Wednesday, May 09, 2018, 08:45: Vacuum @ 4.67e-07T. RGA status:

Water: 3.02e-007

Open lines FS#2 and FS#4

Wednesday, May 09, 2018, 08:55: Vacuum @ 4.65e-07T. RGA status: Water @ 2.97e-007

Close lines FS#2 and FS#4. Water filled.

Open line FS#5 to heat shield.  Vacuum and RGA ok.

09:00: Begin C-Seal (A) shock cycling. 20 cycles @ 5 min apart.

Close Supply/Return, Open Supply/Return to 'Shock'

TIME Vacuum (T) RGA (water) TIME Vacuum (T) RGA (Water)
09:00  4.65e-07  2.94e-007 09:50  4.58e-07  2.87e-007
09:05  4.58e-07  2.93e-007 09:55  4.58e-07  2.88e-007
09:10  4.58e-07  2.93e-007 10:00  4.58e-07  2.87e-007
09:15  4.65e-07  2.92e-007 10:05  4.56e-07  2.87e-007
09:20  4.65e-07  2.92e-007 10:10  4.56e-07  2.85e-007
09:25  4.65e-07  2.90e-007 10:15  4.56e-07  2.86e-007
09:30  4.65e-07  2.89e-007 10:20  4.49e-07  2.83e-007
09:35  4.56e-07  2.89e-007 10:25  4.49e-07  2.84e-007
09:40  4.65e-07  2.89e-007 10:30  4.49e-07  2.84e-007
09:45  4.58e-07  2.89e-007 10:35  4.47e-07  2.83e-007

Wednesday, May 09, 2018, 10:40: Close line FS#5. Vacuum @ 4.49e-07T. RGA status: Water @ 2.84e-007

Open line FS#3 to Coil block.  Vacuum and RGA ok

10:45: Begin C-Seal (B) shock cycling. 20 cycles @ 5 min apart

Close Supply/Return, Open Supply/Return to 'Shock'

TIME Vacuum (T) RGA (water) TIME Vacuum (T) RGA (Water)
10:45  4.47e-07  2.80e-007 11:35  4.32e-07  2.77e-007
10:50  4.47e-07  2.80e-007 11:40  4.38e-07  2.75e-007
10:55  4.47e-07  2.81e-007 11:45  4.38e-07  2.75e-007
11:00  4.38e-07  2.80e-007 11:50  4.38e-07  2.75e-007
11:05  4.38e-07  2.78e-007 11:55  4.38e-07  2.75e-007
11:10  4.40-e07  2.79e-007 12:00  4.40e-07  2.75e-007
11:15  4.38e-07  2.77e-007 12:05  4.38e-07  2.73e-007
11:20  4.47e-07  2.77e-007 12:10  4.40e-07  2.73e-007
11:25  4.40e-07  2.78e-007 12:15  4.30e-07  2.71e-007
11:30  4.38e-07  2.76e-007 12:20  4.32e-07  2.73e-007

12:25: Open FS#5 and leave FS#3 open. Status Vacuum @ 4.32e-07T . RGA status: Water @ 2.70e-007

13:00: Status Vacuum @ 4.30e-07T . RGA status: Water @ 2.67e-007

Wednesday, May 09, 2018, 14:24: Status Vacuum @ 4.21e-07T . RGA status: Water @ 2.52e-007


Thursday, May 10, 2018, 07:41: Status Vacuum @ 3.24e-07T . RGA status: Water @ 1.76e-007.

Purge all water from lines. Prep for He.

Hook up Leak Detector. Leak detector # 5 (DRY) from Vacuum Group. Startup

Thursday, May 10, 2018, 08:00: Baseline Reading:

TP pressure 0.0e-04

Leak rate: Below 0.0e-09 atm-cc/sec

Vacuum @ 3.15e-07T . RGA status: Water @ 1.65e-007.

Valve to chamber opened. LD connection clear.

Leak rate:  1.6e-08 atm-cc/sec

He TEST #2:

Thursday, May 10, 2018, 08:13: Purge both lines with He. Do not pressurize.

Vacuum @ 3.15e-07T. RGA status:

Water @ 1.76e-007

He: -1.51e-009

LD:  1.5e-08 atm-cc/sec

Thursday, May 10, 2018, 08:15: Open line FS#3 to Coil Block. Pressurize to 30 PSI.

Thursday, May 10, 2018, 08:40: Vacuum @ 3.22e-07T . RGA status:

Water @ 1.80e-007

He:-1.98e-009

LD: 7.1e-08 atm-cc/sec. Purge and close FS#3.

Thursday, May 10, 2018, 08:45: Open line FS#5 to Heat Shield. Pressurize to 30 PSI.

Thursday, May 10, 2018, 09:10: Vacuum @ 3.22e-07T. RGA status:

Water @ 1.79e-007

He:-1.39e-009

LD:  1.5e-07 atm-cc/sec

09:15: Depressurize lines. Connect HPA. 

09:30: Vacuum @ 3.22e-07T . RGA status:

Water @ 1.78e-007

He:-1.56-009

LD:  1.7e-07 atm-cc/sec

Thursday, May 10, 2018, 09:32: Close Valve & disconnect LD


Thursday, May 10, 2018, 09:35: Opened all waterlines completely. Vacuum stable @3.22e-07T and no response on the RGA.


Device Temp
[C]
 Flow
[lpm]
 HS + MSP 24.8 2.83
 TBHT 24.7 3.09
 COIL + EE 24.7 5.08
 TGHT 24.9 5.77

Tuesday, May 15, 2018, 06:06: Auto ramping TGHT / TBHT to 580 / 270 A @ 2A per min.

Tuesday, May 15, 2018, 06:13: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .

Tuesday, May 15, 2018, 06:17: Optics set. IV4 opened.

Tuesday, May 15, 2018, 08:35: RGA Restarted. Spikes noticed on BIAS current. Vacuum @ 2.57E-06T


Wednesday, May 16, 2018, 06:40: TMS:IG6 (between 3-4pm yesterday) and the RGA were both knocked out. Restarting both.

Wednesday, May 16, 2018, 06:48: Repeated attempts to reset/restart IG6 have failed. Closing IV4


Friday, May 18, 2018, 10:55: Auto-cooling @ 1A per min. IG6 still non-responsive.

  664   Monday, May 28, 2018, 08:46 Aaron Schmidt LP-FEBIAD TM241790-942752-5In progressNew TM source tray swap out.

Monday, May 28, 2018, 08:45: Venting. Prep For removal.

Close FS#8. Clean up water

Adjust


Tuesday, May 29, 2018, 08:30: Test Stand source tray & Optics tray removed

Clean EZL on optics tray.

Tuesday, May 29, 2018, 11:30: New Source tray installed w/ Original TS optics tray. Connect.

Tuesday, May 29, 2018, 13:30: Roughing.

Tuesday, May 29, 2018, 14:38: Turbos on

Tuesday, May 29, 2018, 14:47: Ion gauge on.

Tuesday, May 29, 2018, 14:49: Exterior line FS# 8 opened.


Wednesday, May 30, 2018, 06:29: RGA on. Vacuum @ 3.38e-06T


He Leak TEST #1:

Vacuum Group Leak detector LD#2.

Friday, June 01, 2018, 08:30: Vacuum @ 9.15e-07T

LD Base line before opening valve to turbo pumps: 3.5e-10 atm-cc/sec

Friday, June 01, 2018, 08:40: LD baseline after opening valve to turbo pumps: 3.7e-10 atm-cc/sec

Full open base:

Vacuum @ 3.24e-07T. RGA status:

Water @ 7.76e-007

He: -2.31e-009

LD:  2.6e-09 atm-cc/sec

Friday, June 01, 2018, 10:30: Open line FS#2 . Pressurize to 30 PSI. Record after 10 minutes

Vacuum @ 9.24e-07T . RGA status:

Water @ 7.91e-007

He:-1.87e-009

LD: 2.8e-09 atm-cc/sec. Purge and close FS#2.

Friday, June 01, 2018, 10:45: Open line FS#3 . Pressurize to 30 PSI. Record after 10 minutes

Vacuum @ 9.24e-07T. RGA status:

Water @ 7.97e-007

He:-1.92e-009

LD:  3.0e-09 atm-cc/sec. Purge and close FS#3.

Friday, June 01, 2018, 11:00: Open line FS#4 . Pressurize to 30 PSI. Record after 10 minutes

Vacuum @ 9.21e-07T. RGA status:

Water @ 7.92e-007

He:-1.85e-009

LD:  4.7e-09 atm-cc/sec. Purge and close FS#4.

Friday, June 01, 2018, 11:15: Open line FS#5 . Pressurize to 30 PSI. Record after 10 minutes

Vacuum @ 9.24e-07T. RGA status:

Water @ 7.919e-007

He:-1.81e-009

LD:  4.0e-09 atm-cc/sec. Purge and close FS#5

Friday, June 01, 2018, 11:30: Depressurize lines. Connect HPA & Purge all lines. 

After 5 minutes: Vacuum @ 9.24e-07T . RGA status:

Water @ 7.87e-007

He:-1.86-009

LD:  4.6e-09 atm-cc/sec

Close Valve & disconnect LD. HE LEAK TEST COMPLETE


Friday, June 01, 2018, 11:41: Open water supply and return mainlines. Vacuum @ 9.15e-07T.

Friday, June 01, 2018, 11:45: Open line FS#2. Vacuum ok. RGA ok

Friday, June 01, 2018, 11:46: Open line FS#3. Vacuum ok. RGA ok

Friday, June 01, 2018, 11:47: Open line FS#4. Vacuum ok. RGA ok

Friday, June 01, 2018, 11:48: Open line FS#5. Vacuum ok. RGA ok.

Friday, June 01, 2018, 12:05: Power/sources switched on.

Friday, June 01, 2018, 12:11: Auto ramping TGHT / TBHT to 300 / 150 A @ 2A per min.

Friday, June 01, 2018, 12:53: Power bump. Lights flickered, however TS ok.

Friday, June 01, 2018, 12:58: Monitored for 5 minutes. ok.

Friday, June 01, 2018, 15:08: Auto ramping TGHT to 350 A @ 2A per min.


Monday, June 04, 2018, 07:27: BIAS / EZL / EE @ 30 / 13 / 1.5 kV.

Monday, June 04, 2018, 07:28: Auto ramping TGHT / TBHT to 580 / 290 A @ 2A per min.

Monday, June 04, 2018, 07:30: Optics reset / Switched on. IV4 opened.

Monday, June 04, 2018, 10:00: FC0 GAIN on 1 uA = 1.65e-09. Wire scans no reading. Adjust steering.

Monday, June 04, 2018, 10:34: Anode @ 200V. Coil @ 10A.

Monday, June 04, 2018, 13:30: Optics tune changed with EZL off. Brad. S.

Monday, June 04, 2018, 13:45: Resetting tunes. COIL @ 50 A. Auto ramping TBHT to 300 A @ 2A per min. (310 A on read back)

Monday, June 04, 2018, 13:51: EZL / EE @ 13 / -1.5 kV.

Monday, June 04, 2018, 13:54: Optics reset to EZL on.


Tuesday, June 05, 2018, 08:00: RGA off.

Tuesday, June 05, 2018, 08:10: Auto ramping TBHT to 310 A @ 2A per min. (320 A on read back)

E-rig issue. Not scanning. Proceed with HV testing.

Tuesday, June 05, 2018, 09:00: BIAS / EZL / EE @ 35 / 15 / 1.75 kV.

Tuesday, June 05, 2018, 09:06: BIAS / EZL / EE @ 40 / 17.5 / 2 kV.

Tuesday, June 05, 2018, 09:13: Small spark knocked out optics & TMS:TP6

 Tuesday, June 05, 2018, 09:21: IV4 closed/opened/reset. TP6 restarted.

Tuesday, June 05, 2018, 09:30: Spark*. Knocked out optics. Reset.

Tuesday, June 05, 2018, 10:37: BIAS / EZL / EE @ 45 / 19.5 / 2.25 kV.

Tuesday, June 05, 2018, 10:44: BIAS / EZL / EE @ 50 / 22 / 2.5 kV.

Tuesday, June 05, 2018, 10:47: Spark. Optics & CBC reset.

Tuesday, June 05, 2018, 11:25: BIAS / EZL / EE @ 52 / 24 / 2.8 kV. SPARK *knocked out optics, magnet & Turbo TMS:TP6. IV4 closed/opened/reset. TP6/optics restarted. Magnet still off.

Tuesday, June 05, 2018, 11:35: BIAS / EZL / EE @ 55 / 24 / 2.8 kV. SPARK* knocked out optics, magnet & Turbos TMS:TP6 & TP1B. IV4 closed/opened/reset. Turbos/optics restarted. Magnet still off.

Tuesday, June 05, 2018, 12:46: BIAS / EZL / EE @ 55 / 24 / 2.8 kV. SPARK* knocked out optics, magnet & Turbos TMS:TP6 & TP1B. IV4 closed/opened/reset. Turbos/optics restarted. Magnet still off.

Tuesday, June 05, 2018, 13:06: BIAS / EZL / EE @ 55 / 24 / 2.8 kV. SPARK* knocked out optics, magnet & Turbos TMS:TP6. IV4 closed/opened/reset. Turbo/optics restarted. Magnet still off. Anode @ 200V (reads 9V, shorted)

Tuesday, June 05, 2018, 13:16: TBHT brought down to 300A. Anode brought to 0V.

Tuesday, June 05, 2018, 13:20: Attempted to bring up to 30V. (30 input) Actual: 8V & 50.061 mA. Possible short, post spark*

Tuesday, June 05, 2018, 13:47: Auto-cooling @ 2A/min

Tuesday, June 05, 2018, 13:56: BIAS / EZL / EE @ 50 / 22 / 2.5 kV. Continue down to 30 kV settings.

Tuesday, June 05, 2018, 14:04: BIAS / EZL / EE @ 30 / 13 / 1.5 kV.


Wednesday, June 06, 2018, 06:49: Anode still unresponsive cold. Short*

Wednesday, June 06, 2018, 11:09: Vacuum @ 4.30e-07T ok. Test EZL

Wednesday, June 06, 2018, 11:16: BIAS / EZL / EE @ 30 / 20 / 1.5 kV.

Wednesday, June 06, 2018, 11:20: BIAS / EZL / EE @ 30 / 25 / 1.5 kV.

Wednesday, June 06, 2018, 11:24: BIAS / EZL / EE @ 30 / 30 / 1.5 kV.

Wednesday, June 06, 2018, 11:30: Coil @ 0A.

Wednesday, June 06, 2018, 12:51: No sparks* BIAS / EZL / EE @ 30 / 30 / 1.5 kV.

Wednesday, June 06, 2018, 12:53: Move Steerers to zero.

TSG:CB1: 0V

TSG:VCB1: 0V

TSG:CBC(2): 0V

Increase each one to 1000V individually. Hold 5 minutes and bump back to zero on each.

Wednesday, June 06, 2018, 12:55: TSG:CB1: 0 - 1000V. Hold for 5 minutes. Zero Sparks. Back to 0V.

Wednesday, June 06, 2018, 13:01: TSG:VCB1: 0 - 1000V. Goes red @ 1000V, stays blue @ 999V Hold for 5 minutes. Zero Sparks. Back to 0V.

Wednesday, June 06, 2018, 13:06: TSG:CBC(2): 0 - 1000V. Hold for 5 minutes. Zero Sparks. Back to 0V.

Wednesday, June 06, 2018, 13:40: Canbus contoller was swapped. The magnet is back.

Wednesday, June 06, 2018, 13:50: Sources to 0V.

Wednesday, June 06, 2018, 14:05: IV4 closed. Turbo pumps off. Ion gauge off.

Wednesday, June 06, 2018, 14:08: Power to the sources off.


Thursday, June 07, 2018, 06:35: Venting & waterlines purged.

Attachment 1: New_TM_Source_Tray_2018.docx
Attachment 2: TS-2018-05-30-06-30.rga
Attachment 3: ER-580-310-30kV-m28.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
     80.0000
  29978.0220
21
21
./emit.dat
 -0.0977 -0.1007 -0.0986 -0.0970 -0.1013 -0.0989 -0.0998 -0.0989 -0.1004 -0.0970 -0.0998 -0.0958 -0.1028 -0.1001 -0.1022 -0.0998 -0.1007 -0.1016 -0.1050 -0.1065 -0.1056
 -0.0983 -0.0998 -0.1004 -0.1019 -0.1007 -0.1031 -0.0995 -0.0995 -0.1035 -0.0977 -0.1001 -0.0998 -0.1010 -0.0992 -0.1022 -0.0992 -0.1010 -0.0992 -0.1047 -0.1080 -0.1050
 -0.0967 -0.0989 -0.0998 -0.1004 -0.1025 -0.1022 -0.1007 -0.1016 -0.1028 -0.0980 -0.1016 -0.1016 -0.1016 -0.0995 -0.0998 -0.1004 -0.0983 -0.0986 -0.1041 -0.1068 -0.0974
 -0.0995 -0.0995 -0.1001 -0.1001 -0.0974 -0.0995 -0.1019 -0.1019 -0.0998 -0.1016 -0.0986 -0.1019 -0.1038 -0.0986 -0.1001 -0.0983 -0.1001 -0.0980 -0.1031 -0.1056 -0.1016
 -0.0983 -0.1016 -0.0977 -0.1004 -0.0967 -0.0974 -0.1007 -0.1004 -0.1038 -0.1004 -0.1013 -0.1016 -0.1016 -0.0986 -0.1019 -0.1013 -0.1010 -0.1019 -0.1062 -0.1041 -0.1053
 -0.0946 -0.0995 -0.0970 -0.1004 -0.1001 -0.1001 -0.1010 -0.0992 -0.1022 -0.0980 -0.1013 -0.1007 -0.1047 -0.1035 -0.1022 -0.1044 -0.1016 -0.1022 -0.1080 -0.1062 -0.1016
 -0.0961 -0.0992 -0.0995 -0.0998 -0.1007 -0.1019 -0.0992 -0.1022 -0.1035 -0.1010 -0.0977 -0.1010 -0.0989 -0.0998 -0.1004 -0.1016 -0.1013 -0.1004 -0.1102 -0.1086 -0.0980
 -0.0977 -0.0986 -0.1007 -0.0980 -0.0989 -0.1010 -0.1016 -0.1016 -0.1038 -0.1004 -0.1001 -0.0992 -0.0983 -0.1007 -0.0980 -0.1001 -0.0995 -0.0983 -0.1083 -0.1074 -0.0998
 -0.0970 -0.1013 -0.0995 -0.0995 -0.1022 -0.0986 -0.1041 -0.0998 -0.1028 -0.1010 -0.1013 -0.0992 -0.1028 -0.1010 -0.1001 -0.0998 -0.1010 -0.0977 -0.1111 -0.1059 -0.1019
 -0.0989 -0.0998 -0.0998 -0.1010 -0.1001 -0.0992 -0.1035 -0.1007 -0.1010 -0.0998 -0.1010 -0.1010 -0.1031 -0.1035 -0.1031 -0.1028 -0.1007 -0.1013 -0.1062 -0.1062 -0.0992
 -0.0970 -0.0986 -0.0964 -0.1007 -0.0989 -0.1004 -0.1001 -0.1007 -0.0992 -0.0998 -0.1010 -0.0980 -0.1007 -0.0983 -0.1016 -0.1031 -0.0995 -0.0995 -0.1035 -0.1074 -0.1010
 -0.0961 -0.0989 -0.0967 -0.1010 -0.0992 -0.1001 -0.1016 -0.1019 -0.1004 -0.1019 -0.0989 -0.0992 -0.1010 -0.0989 -0.1004 -0.1007 -0.0983 -0.1013 -0.1080 -0.1056 -0.1068
 -0.0986 -0.0998 -0.0983 -0.1019 -0.1010 -0.0977 -0.1022 -0.0998 -0.1019 -0.1007 -0.0977 -0.0977 -0.0998 -0.0967 -0.0983 -0.1007 -0.0998 -0.0989 -0.1111 -0.1041 -0.0989
 -0.0998 -0.0998 -0.0980 -0.0983 -0.0986 -0.1001 -0.1035 -0.0977 -0.1016 -0.1013 -0.1019 -0.1013 -0.1025 -0.0998 -0.1013 -0.1007 -0.1019 -0.0986 -0.1117 -0.1053 -0.0983
 -0.0967 -0.0992 -0.0998 -0.0970 -0.0986 -0.1004 -0.1022 -0.1007 -0.1028 -0.0992 -0.1004 -0.1025 -0.1010 -0.1028 -0.0998 -0.1044 -0.1019 -0.1022 -0.1071 -0.1086 -0.1038
 -0.0989 -0.0989 -0.1019 -0.0977 -0.0992 -0.1013 -0.1044 -0.1019 -0.1007 -0.0977 -0.0992 -0.1013 -0.1016 -0.1050 -0.1010 -0.1010 -0.1007 -0.1019 -0.1062 -0.1077 -0.1031
 -0.0955 -0.0970 -0.0995 -0.1019 -0.1022 -0.1004 -0.1044 -0.1022 -0.1013 -0.0998 -0.1004 -0.1004 -0.0995 -0.1007 -0.0995 -0.0992 -0.0995 -0.1013 -0.1035 -0.1062 -0.1007
 -0.0943 -0.0989 -0.0980 -0.1025 -0.1028 -0.1013 -0.1013 -0.1013 -0.1007 -0.0995 -0.1016 -0.0992 -0.0998 -0.0983 -0.0998 -0.1007 -0.0995 -0.1001 -0.1065 -0.1047 -0.1025
 -0.0983 -0.0998 -0.0989 -0.1025 -0.0998 -0.1004 -0.1022 -0.1035 -0.1050 -0.1010 -0.1028 -0.1013 -0.1004 -0.0986 -0.0995 -0.1013 -0.0995 -0.0983 -0.1068 -0.1068 -0.0995
 -0.0961 -0.1001 -0.0974 -0.0995 -0.0986 -0.0980 -0.1001 -0.1016 -0.1004 -0.1004 -0.1007 -0.0995 -0.1013 -0.1010 -0.0998 -0.1031 -0.1022 -0.0998 -0.1120 -0.1065 -0.0983
 -0.0958 -0.0995 -0.0983 -0.1001 -0.0995 -0.0961 -0.1019 -0.0995 -0.1013 -0.0995 -0.0967 -0.1013 -0.1010 -0.0998 -0.1013 -0.1022 -0.1010 -0.1035 -0.1123 -0.1071 -0.1010
Attachment 4: ER-580-320-30kV-M39.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    160.0000
  29978.0220
21
21
./emit.dat
 -0.0845 -0.0845 -0.0906 -0.0940 -0.0659 -0.0629 -0.0525 -0.0476 -0.0516 -0.0879 -0.0839 -0.0858 -0.0864 -0.0821 -0.0854 -0.0848 -0.0848 -0.0839 -0.0943 -0.0888 -0.0882
 -0.0842 -0.0830 -0.0864 -0.0922 -0.0635 -0.0632 -0.0497 -0.0482 -0.0510 -0.0894 -0.0833 -0.0830 -0.0790 -0.0839 -0.0851 -0.0821 -0.0827 -0.0870 -0.0943 -0.0897 -0.0867
 -0.0851 -0.0833 -0.0845 -0.0867 -0.0671 -0.0595 -0.0519 -0.0507 -0.0522 -0.0909 -0.0851 -0.0851 -0.0870 -0.0885 -0.0842 -0.0851 -0.0839 -0.0842 -0.0864 -0.0873 -0.0854
 -0.0851 -0.0854 -0.0845 -0.0861 -0.0699 -0.0626 -0.0531 -0.0473 -0.0522 -0.0925 -0.0854 -0.0864 -0.0821 -0.0867 -0.0854 -0.0842 -0.0821 -0.0836 -0.0876 -0.0891 -0.0864
 -0.0842 -0.0854 -0.0851 -0.0854 -0.0693 -0.0610 -0.0534 -0.0519 -0.0549 -0.0925 -0.0861 -0.0867 -0.0851 -0.0848 -0.0851 -0.0854 -0.0827 -0.0824 -0.0937 -0.0900 -0.0854
 -0.0836 -0.0842 -0.0861 -0.0870 -0.0699 -0.0607 -0.0528 -0.0476 -0.0577 -0.0931 -0.0861 -0.0854 -0.0861 -0.0842 -0.0845 -0.0854 -0.0848 -0.0854 -0.0922 -0.0885 -0.0848
 -0.0839 -0.0836 -0.0836 -0.0815 -0.0681 -0.0601 -0.0497 -0.0482 -0.0531 -0.0909 -0.0827 -0.0836 -0.0778 -0.0836 -0.0836 -0.0830 -0.0842 -0.0833 -0.0912 -0.0861 -0.0836
 -0.0867 -0.0836 -0.0848 -0.0836 -0.0653 -0.0580 -0.0510 -0.0488 -0.0513 -0.0900 -0.0821 -0.0818 -0.0861 -0.0845 -0.0824 -0.0833 -0.0836 -0.0827 -0.0919 -0.0867 -0.0854
 -0.0851 -0.0864 -0.0836 -0.0867 -0.0671 -0.0583 -0.0543 -0.0476 -0.0504 -0.0873 -0.0839 -0.0858 -0.0864 -0.0848 -0.0836 -0.0854 -0.0842 -0.0854 -0.0922 -0.0879 -0.0879
 -0.0803 -0.0845 -0.0827 -0.0854 -0.0717 -0.0607 -0.0513 -0.0494 -0.0522 -0.0919 -0.0861 -0.0870 -0.0845 -0.0851 -0.0851 -0.0851 -0.0861 -0.0873 -0.0977 -0.0894 -0.0867
 -0.0845 -0.0827 -0.0842 -0.0839 -0.0705 -0.0607 -0.0522 -0.0470 -0.0531 -0.0894 -0.0858 -0.0864 -0.0848 -0.0851 -0.0854 -0.0833 -0.0842 -0.0848 -0.0946 -0.0876 -0.0851
 -0.0824 -0.0867 -0.0845 -0.0839 -0.0696 -0.0586 -0.0513 -0.0461 -0.0543 -0.0916 -0.0839 -0.0827 -0.0842 -0.0845 -0.0830 -0.0827 -0.0815 -0.0836 -0.0851 -0.0864 -0.0818
 -0.0830 -0.0861 -0.0848 -0.0861 -0.0702 -0.0604 -0.0488 -0.0519 -0.0552 -0.0912 -0.0836 -0.0842 -0.0824 -0.0836 -0.0830 -0.0848 -0.0824 -0.0827 -0.0858 -0.0864 -0.0842
 -0.0867 -0.0867 -0.0864 -0.0854 -0.0671 -0.0595 -0.0543 -0.0507 -0.0528 -0.0925 -0.0845 -0.0842 -0.0848 -0.0848 -0.0845 -0.0848 -0.0836 -0.0836 -0.0934 -0.0876 -0.0867
 -0.0867 -0.0870 -0.0864 -0.0879 -0.0681 -0.0601 -0.0519 -0.0522 -0.0546 -0.0922 -0.0851 -0.0867 -0.0851 -0.0861 -0.0858 -0.0870 -0.0845 -0.0845 -0.0970 -0.0888 -0.0882
 -0.0848 -0.0848 -0.0854 -0.0867 -0.0711 -0.0583 -0.0522 -0.0479 -0.0537 -0.0882 -0.0839 -0.0858 -0.0833 -0.0851 -0.0848 -0.0836 -0.0839 -0.0830 -0.0961 -0.0870 -0.0876
 -0.0858 -0.0839 -0.0836 -0.0812 -0.0684 -0.0568 -0.0519 -0.0479 -0.0531 -0.0873 -0.0836 -0.0836 -0.0848 -0.0818 -0.0827 -0.0812 -0.0845 -0.0848 -0.0922 -0.0882 -0.0854
 -0.0830 -0.0827 -0.0824 -0.0836 -0.0656 -0.0574 -0.0528 -0.0467 -0.0540 -0.0919 -0.0836 -0.0842 -0.0812 -0.0818 -0.0839 -0.0830 -0.0842 -0.0836 -0.0912 -0.0867 -0.0858
 -0.0842 -0.0845 -0.0821 -0.0876 -0.0690 -0.0607 -0.0510 -0.0500 -0.0531 -0.0928 -0.0842 -0.0833 -0.0885 -0.0854 -0.0851 -0.0858 -0.0848 -0.0858 -0.0937 -0.0888 -0.0827
 -0.0867 -0.0848 -0.0854 -0.0854 -0.0720 -0.0620 -0.0534 -0.0476 -0.0534 -0.0934 -0.0848 -0.0870 -0.0836 -0.0864 -0.0854 -0.0870 -0.0848 -0.0864 -0.0952 -0.0888 -0.0879
 -0.0848 -0.0833 -0.0851 -0.0864 -0.0687 -0.0592 -0.0552 -0.0507 -0.0522 -0.0873 -0.0851 -0.0839 -0.0848 -0.0839 -0.0845 -0.0845 -0.0833 -0.0836 -0.0955 -0.0882 -0.0827
Attachment 5: 2018_TM2_Source_Tray_in_Test_Stand_-_Test_Results.docx
  670   Thursday, June 21, 2018, 16:32 Anders Mjos LP-FEBIAD  41790-942752-5In progressHeating target

Thursday, June 21, 2018, 16:30: Short between ANODE and EE. Do not use until repaired.

Thursday, June 21, 2018, 16:32: Ramping TBHT/TGHT to 290 / 580 Aat 2 A/min


Friday, June 22, 2018, 07:49: Target at temperature. Vacuum at 1.0E-7 T


Monday, June 25, 2018, 09:10: Auto-cooling at 2A/min - AS.


Friday, June 29, 2018, 08:52: Turbos/PS/Sources/RGA/Ion gauge off.

Friday, June 29, 2018, 10:03: Venting.

Friday, June 29, 2018, 14:32: Roughing. Anode to EE short fixed.

Attachment 1: TS-2018-06-19-10-30.rga
Attachment 2: TS_Source_tray_2018_-_Check_2.docx
  672   Monday, July 09, 2018, 10:44 Aaron Schmidt LP-FEBIAD  41790-942752-5In progressCoil Test July

Monday, July 09, 2018, 10:43: Rouging.

Monday, July 09, 2018, 13:14: Turbo pumps on.


Tuesday, July 10, 2018, 07:45: Waterlines opened. Vacuum ok @ 6.93e-07T. No response on RGA.

Tuesday, July 10, 2018, 08:05: Opened gas lines TSH:SV3 & TSH:SV4. Vacuum spiked to 1.08e-06T. Response registered on RGA. Closed valves.

Tuesday, July 10, 2018, 08:22: Vacuum Holding at 1.08e-06T.

Tuesday, July 10, 2018, 08:24: Close valve to small gas line pump. Pump is running.

Tuesday, July 10, 2018, 09:02: Path to TSH:SV1 through SV2 & SV5 ok. Vacuum @ 5.55e-07T. Close and begin SV3 & SV4 purging.

Tuesday, July 10, 2018, 09:12: All gas lines opened.

Tuesday, July 10, 2018, 09:16: Power/sources on. Vacuum @ 7.26e-07T.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 24.4 3.85
 TBHT 24.2 2.99
 COIL + EE 24.2 4.97
 TGHT  24.6 5.61

Tuesday, July 10, 2018, 09:18: Auto ramping TGHT / TBHT to 580 / 270 A @ 2A per min.

Tuesday, July 10, 2018, 09:20: IV4 opened. Optics on (already set).

Tuesday, July 10, 2018, 14:11: Abort auto-ramping. Increase TBHT ramping to 300A. Continue auto ramping TGHT / TBHT to 580 / 270 A @ 2A per min.

Tuesday, July 10, 2018, 13:56: BIAS/EZL/EE @ 30/13/1.5 kV. Anode set to 200 V

Tuesday, July 10, 2018, 14:12: Auto ramping complete.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 34.6 3.85
 TBHT 25.0 2.97
 COIL + EE 24.9 5.00
 TGHT  26.8 5.60

 Tuesday, July 10, 2018, 14:17: TBHT set to 300 A. Read back reads 312 A


Wednesday, July 11, 2018, 06:52: Coil to 10A.

Wednesday, July 11, 2018, 07:00: Coil to 20A.

Wednesday, July 11, 2018, 07:05: Coil to 30A.

Wednesday, July 11, 2018, 07:10: Coil to 40A.

Wednesday, July 11, 2018, 07:20: Coil to 50A.


Fernando Maldonado

Wednesday, July 11, 2018, 09:59 Anode current=4.884 mA, Anode Voltage=200V. FC0=1.59e-9 . Source seems to be too cold. ramping up TBHT to 320 A (for a 330A readback). Vaccum=9.06e-7

 Wednesday, July 11, 2018, 10:18 Ramp up compleete TBHT=320A Read Back=33.8A. Anode current=9.768mA. Anode voltage=200V. FC0=1.77e-8A. Vaccum=9.13e-7

Wednesday, July 11, 2018, 11:52 Mass scan performed for nominal values

Lunch

Steering

Adjusting Emittance parameter for more precise scan

Wednesday, July 11, 2018, 15:37 Cathode is still "cold" , at 325A and 250V the anode current is 15 mA. The way to obtained more current and to have better emittances readings is to increase the tbht current

Wednesday, July 11, 2018, 15:39 TBHT=325, TBHT RB =339.1A, TBHT PS=330A. Increasing TBHT to 335A. Where TBHT RB is the read back value. TBHT PS is the value display on the Power Supply

Wednesday, July 11, 2018, 15:46 tbht=335, tbht rb=349.7, tbht ps=341. With this the anode current is 24.42 at anode voltage=250V. *THIS WILL BE THE MAXIMUM VALUE"

Wednesday, July 11, 2018, 17:15 Mass Scan measurement seem to not vary with anode voltage

Wednesday, July 11, 2018, 17:45 Coil to 60 A to let out gass a bit more overnight. Vacuum ok at 9.6e-7


Thursday, July 12, 2018, 07:25: Power outage. (07/11/2018 @ 21:53 approx) The test stand was completely knocked out. Power & pumps were all shut down.

Thursday, July 12, 2018, 07:31: Power supplies turned off and all settings manually re-zeroed.

  673   Thursday, July 12, 2018, 08:24 Aaron Schmidt LP-FEBIAD  41790-942752-5In progressPower outage / restart

Power outage (07/11/2018 @ approx 21:53) shut down the test stand.

 

Thursday, July 12, 2018, 08:17: Reset vacuum controls & IV4

Thursday, July 12, 2018, 08:19: Restart all the turbo pumps.

Thursday, July 12, 2018, 08:24: IG6 & IG1 restarted.

Thursday, July 12, 2018, 08:33: Power on.

Thursday, July 12, 2018, 08:36: Opened gas lines SV2 & SV5 for FEBIAD to pump down lines with the chamber.

Thursday, July 12, 2018, 08:56: Reset optics. Operation appears ok.

Thursday, July 12, 2018, 09:02: Sources except EE & BIAS are reset. BIAS requires EE and the EE won't restart / respond.

Thursday, July 12, 2018, 09:31: Cycle EE PS? No response.

Thursday, July 12, 2018, 10:08: Cycle PLC. EE is responsive / back. BIAS operates as well.

Thursday, July 12, 2018, 10:10: Opened IV4.

  674   Thursday, July 12, 2018, 10:11 Aaron Schmidt LP-FEBIAD  41790-942752-5In progressCoil Test #5

 *** REVIEW Elog entry #673 Power outage & restart ***

Thursday, July 12, 2018, 10:10: Opened IV4

Thursday, July 12, 2018, 10:20: Attempt auto-ramp cancelled. Max increment values reached. Wait on Vacuum.

Thursday, July 12, 2018, 13:10: Auto-ramping TGHT / TBHT to 500 / 200 A @ 2A per min.

Thursday, July 12, 2018, 13:16: BIAS/EZL/EE @ 30/13/1.5 kV. 

Thursday, July 12, 2018, 13:19: Anode set to 200 V

Thursday, July 12, 2018, 13:23: Gas lines SV1, SV3, SV4 opened without issue.


Friday, July 13, 2018, 07:02: Auto-ramping TGHT / TBHT to 580 / 300 A @ 2A per min.


FM:

Friday, July 13, 2018, 09:26 Ramping up to TBHT=335A to get at least 20 mA on the anode current.

Friday, July 13, 2018, 09:52 Ramp up finished: TBHT=335A, TBHT RB=349.6A, TBHT PS=341A. Anode current=20.757mA@200V. Vaccum ok@8.86e-7Torr

Friday, July 13, 2018, 09:56 Closing Neon valve in order to only deliver Argon.


Friday, July 13, 2018, 13:32   When anode voltage=0V, there are four peaks on the mass scan on masses  23,38,41,83 u. This indicate surface ionized particles. Most likely Sodium (23) and Potassium (38,41) and not sure about 83.

Friday, July 13, 2018, 16:58  Cooling it down for the weekend: TBHT=250A, TGHT=450

Friday, July 13, 2018, 17:01 wow! accidentally, changed to those values immediatly. I hope this does not compromise the cathode - Fernando


Monday, July 16, 2018, 06:57: Odd RGA values. Restarting RGA.

Monday, July 16, 2018, 07:14: RGA only displays Water, Nitrogen, and Oxygen. Investigate.

Monday, July 16, 2018, 09:57 Auto-ramping TGHT / TBHT to 580 / 300 A @ 2A per min from 449 / 209 A.

Monday, July 16, 2018, 10:00: Restarted the RGA with an older file loaded. RGA still only displays lines for Water, Nitrogen, and Oxygen. A table of values is still given for the other masses.

Monday, July 16, 2018, 11:08: TGHT / TBHT to 580 / 320 A. Anode to 200V

Monday, July 16, 2018, 11:19: Neon valve carefully opened. Vacuum stable @ 6.56e-07T. Neon response also showed up on the RGA.

Monday, July 16, 2018, 11:25: TBHT @ 340.0A (332A on PS)


FM

Monday, July 16, 2018, 11:55  TBHT@340A @ 354.7A (RB)@346A PS

Monday, July 16, 2018, 15:02 TBHT@345A @ 359.8A (RB)@351A PS

Monday, July 16, 2018, 15:18 Investigation the possible location of the surface ionized ions by sweeping the voltage in lower values <10V

Monday, July 16, 2018, 15:29 Values swept: 10,8,6,4,2,1 V. the same peak was observed at all cases

Monday, July 16, 2018, 15:52 Closing Neon Valve for efficiency measurement.

Monday, July 16, 2018, 15:58 Steady  ion source current. closing Argon valve.

Monday, July 16, 2018, 17:00 Opening Argon Valve

Monday, July 16, 2018, 17:51 Argon valve closed. Neon Valve open. Measuring Mass=20

Monday, July 16, 2018, 18:13 Manually tapped TBHT down to 325A.


FM

Tuesday, July 17, 2018, 09:53  Ramping up TBHT to 345A

Tuesday, July 17, 2018, 10:13 Ramp up finish. TBHT@345A @ 359.7A (RB)@351A PS.  Vacuum ok @ 7.43e-7 T

Tuesday, July 17, 2018, 10:14 Restarting  quads that were off.

Tuesday, July 17, 2018, 10:18 Tuning

Tuesday, July 17, 2018, 11:01 Measurements will start for mass 20.

Tuesday, July 17, 2018, 11:59 Lunch break

Tuesday, July 17, 2018, 12:41 Coming back. TMSQ2 & TMSQ5 off. Turning them on.

Tuesday, July 17, 2018, 13:54 Retuning emittance scanner to default parameters. Cycle the emittance scanner PS.

Tuesday, July 17, 2018, 14:18 Increasing current limit for anode. From 51.282 mA to  51.400 mA

 Tuesday, July 17, 2018, 14:34 Increasing current limin on anode to 52.00 mA

Tuesday, July 17, 2018, 14:35 Closing Neon Valve

Tuesday, July 17, 2018, 15:02 FC6 stable

Tuesday, July 17, 2018, 15:02 Opening "Argon" bottle

Tuesday, July 17, 2018, 15:21 Fc6 stable. The present current lies between the on/off Neon Valve.

Tuesday, July 17, 2018, 15:30 Opening Neon Valve

Tuesday, July 17, 2018, 15:51 Re-tuning. Taking a mass scan.

Tuesday, July 17, 2018, 15:55 Checking the RGA, I noticed that the maximum reading for Neon was 7.05e-8T, and for Argon, 6.15e-7T

Tuesday, July 17, 2018, 16:45 Increasing Anode current limit to 100mA. Increasing TBHT to 347A, 361.8 RB, 353 PS. Vaccum ok@7.87e-7T

Tuesday, July 17, 2018, 16:54 Finish for today.


FM

Wednesday, July 18, 2018, 09:27  Starting new round of measurements.

Wednesday, July 18, 2018, 09:31 TBHT@350A@364.5A (RB) @ 356 (PS).  Vacuum ok at 6.71e-7T

Wednesday, July 18, 2018, 11:48 Finish mass scan sweeps. Leaving for lunch. Coming back at around 3pm due to science week

Wednesday, July 18, 2018, 15:29 Starting again.

Wednesday, July 18, 2018, 15:30 Mass scan at AB=0V

Wednesday, July 18, 2018, 15:56 Mass scan set to 38.5 (K).

Wednesday, July 18, 2018, 15:56 Measuring emittance.

Wednesday, July 18, 2018, 16:02 Closing Argon bottle.

Wednesday, July 18, 2018, 16:08 Closing Neon bottle.

Wednesday, July 18, 2018, 16:13 Measuring what it looks to be a potassium signal. Let's see if the signal decreases overnight.


Friday, July 20, 2018, 11:48: FC0 put in place & the gas lines closed. - Aaron S.

Friday, July 20, 2018, 11:50  : Auto cooling @ 2A per min.


Monday, July 23, 2018, 07:05: Sources off. Optics checked.

Monday, July 23, 2018, 07:13 : RGA error. Restart for proper shut off. IV4 closed.

Monday, July 23, 2018, 07:14 : Ion gauge, turbos off.

  687   Tuesday, October 02, 2018, 13:39 Aaron Schmidt LP-FEBIAD  41790-942752-5In progressTest FEBIAD w/ Silver bolts & sealed end caps w/ stoppage

Silver bolts used on Coil, TGHT, & TBHT.

End caps in target

Tuesday, October 02, 2018, 13:36: Roughing.

Tuesday, October 02, 2018, 13:37: Doors sealed.

Tuesday, October 02, 2018, 14:10: Turbos on.


Due to potential power & cooling loss. Shutdown vac.

Wednesday, October 03, 2018, 07:34: Chamber Turbos off.

Wednesday, October 03, 2018, 10:58: Beam line Turbos off. Closed both PV1 & PV2 & turned off both roughing pumps.


Thursday, October 11, 2018, 10:37: RGA on. Strip tools re-opened.

Thursday, October 11, 2018, 10:43: Waterlines opened. No response on RGA. Vacuum stable @ 3.40e-07T.

Thursday, October 11, 2018, 10:48: Power on.

Thursday, October 11, 2018, 10:49: Opened manual valve in gas control to operate small pump. Vacuum spike into Xe-05T range.

Thursday, October 11, 2018, 10:51: Very gradual opening on source gas control. Pump ineffective.

Thursday, October 11, 2018, 10:52: SV5 not opened yet. Open SV2. Vacuum spike noted.

Thursday, October 11, 2018, 10:53: SV5 opened/closed very quickly. Allow stabilization. Alternate between SV5 & SV2 until stable.

Thursday, October 11, 2018, 10:58: SV5 closed. Open SV4. Vacuum spike noted up to 3.3e-05T.

Thursday, October 11, 2018, 11:30: All Gas lines opened.

Thursday, October 11, 2018, 11:36: Sources on.

Thursday, October 11, 2018, 11:43: BIAS / EZL / EE @ 30 / 13 / 1.5 kV.

Thursday, October 11, 2018, 11:46: IV4 opened. Error / closed / red. Fix. IV4 re-opened.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 21.9 3.96
 TBHT 21.7 3.09
 COIL + EE 22.0 5.15
 TGHT  22.2 5.83

 Thursday, October 11, 2018, 12:08: Auto-ramping TGHT / TBHT to 300 / 200 A @ 2A per min


Friday, October 12, 2018, 08:07: Auto-ramping TGHT / TBHT to 580 / 290 A @ 2A per min.

Friday, October 12, 2018, 08:13: Optics reset.Can Diagnostics re-cfg.

Friday, October 12, 2018, 10:49: Anode @ 200 V.

Friday, October 12, 2018, 10:50: Coil @ 10 A. Vacuum @ 6.25e-07T

Friday, October 12, 2018, 10:53: Coil @ 20 A. Vacuum @ 6.27e-07T

Friday, October 12, 2018, 10:56: Coil @ 30 A. Vacuum @ 6.27e-07T

Friday, October 12, 2018, 10:58: Coil @ 40 A. Vacuum @ 6.34e-07T

Friday, October 12, 2018, 11:00: Coil @ 50 A. Vacuum @ 6.34e-07T

Friday, October 12, 2018, 11:01: Auto-ramping TBHT to 300 A @ 2A per min.

Friday, October 12, 2018, 11:06: TBHT @ 300 A.

Friday, October 12, 2018, 11:51: Nothing found on Mass or wire scans?

Friday, October 12, 2018, 11:52: Anode set to 0 V.

Friday, October 12, 2018, 11:54: Still nothing found on wire scans.


Fernando:

Friday, October 12, 2018, 12:56: Anode leakage current is 6.105 mA on the readback with AV@200V Coil@0A TGHT@580A TBHT@310

Friday, October 12, 2018, 12:59 : This is very low anode lakage current. It should be on the bulk of 100 mA.

Friday, October 12, 2018, 12:59: Ramping up TBHT to 320 A @ 2A per min while keeping the AV@200 V. STOP

 


Aaron: Anode / Cathode BIAS PS polarity requires a reversal. John Chow notified. To be reversed Monday after 11am.

Friday, October 12, 2018, 13:24: Continue auto-ramping TBHT to 320 A @ 2A per min.

Friday, October 12, 2018, 13:30: TBHT 320 A / Read back @ 313.1 A.

Friday, October 12, 2018, 13:34: Wire & mass scans

Friday, October 12, 2018, 13:54: Anode to 0 V. Coil to 50 A.


Monday, October 15, 2018, 07:14: Auto-cooling. Coil to 0 A.

Monday, October 15, 2018, 11:54: Sources/power off. IV4 closed

Attachment 1: FEBIAD_Coil_Test_6_-_POS_Cathode.docx
Attachment 2: TS-2018-10-11-10-37.rga
  688   Monday, October 15, 2018, 14:23 Aaron Schmidt LP-FEBIAD  41790-942752-5In progressTest FEBIAD w/ Silver bolts & sealed end caps

Anode / Cathode BIAS PS polarity reversed by John Chow. PS: Negative

Monday, October 15, 2018, 14:12: Power/sources on

Monday, October 15, 2018, 14:17: RGA on.

Monday, October 15, 2018, 14:22: Gas-lines opened. Vacuum ok @ 4.38e-07T

Monday, October 15, 2018, 14:29: Auto-ramping TGHT / TBHT to 580 / 300 A @ 2A per min.

Monday, October 15, 2018, 14:30: Optics on

Monday, October 15, 2018, 14:33: BIAS / EZL / EE @ 30 / 13 / 1.5 kV.

Monday, October 15, 2018, 14:35: Anode @ 200 V.

Monday, October 15, 2018, 14:36: IV4 Open.


Tuesday, October 16, 2018, 06:33: Auto-ramping TBHT to 320 A @ 2A per min. 

Tuesday, October 16, 2018, 07:15: Coil @ 50 A. 

Tuesday, October 16, 2018, 14:40: FC6 Auto range & Polarity set. Mass scans work now.


FM

Wednesday, October 17, 2018, 10:02 : Auto-ramping TBHT to 330 A @ 2A per min. 

Wednesday, October 17, 2018, 13:21: Auto-ramping TBHT to 340 A @ 2A per min.

Wednesday, October 17, 2018, 13:45: Increasing TBHT voltage from 3.001 to 3.5 to  avoid voltage limitation when increasing the current

 Wednesday, October 17, 2018, 13:48: Auto-ramping TBHT to 345 A @ 2A per min.

Wednesday, October 17, 2018, 13:54: FC0 start decreasing once the ramp up finished. Vacuum ok @ 5.17e-7T.

Wednesday, October 17, 2018, 17:16: FC0 stable. Probably outgassing. Vacuum ok @ 5.37e-T

Wednesday, October 17, 2018, 17:16: Auto-ramping TBHT to 350 A @ 2A per min.

Wednesday, October 17, 2018, 17:30: Ramping finished. FC0 decreasing again. Vacuum ok @5.57e-7T

Wednesday, October 17, 2018, 17:51: Potassium signal seems to be decreasing and now the Argon is slightly higher than Potassium.

Wednesday, October 17, 2018, 17:51: Mass optimized for 40Ar, which is shown in the mass scan as ~39.52. Dipole current=102.2736 A

Wednesday, October 17, 2018, 17:52: Starting steerers an einzel lense tuning to increase FC6 signal. Will continue tomorrow


FM

Thursday, October 18, 2018, 08:03: FC0 signal is stable now. Auto-ramping TBHT to 355 A @ 2A per min.

Thursday, October 18, 2018, 08:08 Ramping finished. Vacuum ok @5.26e-7T. Anode current=65 mA.

Thursday, October 18, 2018, 09:55: FC0 looks stable now. Manually tapping 2 A to have TBHT=357A.

 Thursday, October 18, 2018, 11:40: Vaccum ok @ 4.56e-7T

Thursday, October 18, 2018, 16:25: Vacuum ok @ 5.28e-7T

Thursday, October 18, 2018, 16:26: Manually tapping 2 A to have TBHT=359A.

Thursday, October 18, 2018, 17:29: Steerers, einzel lens and dipole tune to increase signal on FC6. Argon current=4 nA. Total source current=54 nA.

Thursday, October 18, 2018, 17:37: Coil to 50A.

Thursday, October 18, 2018, 17:47: Closing Neon valve.

Thursday, October 18, 2018, 17:57: Closing Argon valve.

Thursday, October 18, 2018, 19:51: Assuming 720 nA entering the source (from Brad). Efficiency  measurement = 0.44%

Thursday, October 18, 2018, 19:52: Opening valves.

Thursday, October 18, 2018, 19:54: Vaccum ok @ 5.46e-7T


 FM

Friday, October 19, 2018, 08:27: Manually tapping 1 A to have TBHT=360.

Friday, October 19, 2018, 08:39: FC6 is stable. Closing Neon Valve.

Friday, October 19, 2018, 08:44: FC6 is stable. Closing Argon Valve.

 Friday, October 19, 2018, 09:13: TBHT@360A, AV=150V. Efficiency measurement = 0.41%

Friday, October 19, 2018, 09:16: Opening the valves.

Friday, October 19, 2018, 09:40: FC6 stable. Carefully increasing AV (anode voltage) to 200V keeping an eye on the AC (anode current).

Attachment 1: FEBIAD_Coil_Test_6-2_-_Neg_Cathode.docx
Attachment 2: TS-2018-10-15-14-15.rga
  691   Monday, October 22, 2018, 11:07 Aaron Schmidt LP-FEBIAD  41790-942752-5In progressTest FEBIAD w/ Silver bolts & sealed end caps

Restarted post power outage.

Water & Turbos on.

Monday, October 22, 2018, 10:55: Power / sources on. MB1?

Monday, October 22, 2018, 11:07: Open gas lines slowly.

Monday, October 22, 2018, 11:11: Gas lines all opened. Vacuum @ 2.37e-06T & decreasing.


Tuesday, October 23, 2018, 08:00: MB1 PS on.

Tuesday, October 23, 2018, 08:03: Auto-ramping TGHT / TBHT to 580 / 300 A @ 2A per min.

Tuesday, October 23, 2018, 08:08: BIAS / EZL / EE @ 30 / 13.5 / 1.5 kV.

Tuesday, October 23, 2018, 08:10: Epics lost. Communication error. Internet loss.

Tuesday, October 23, 2018, 08:21: Auto ramp found halted in communication loss. Auto-ramping TGHT / TBHT to 580 / 300 A @ 2A per min restarted from 14.0A.

Tuesday, October 23, 2018, 08:23: Anode @ 200 V.

Tuesday, October 23, 2018, 08:28: Optics set. IV4 opened.

Tuesday, October 23, 2018, 13:03: Auto-ramping TBHT / 320 A @ 2A per min.

Tuesday, October 23, 2018, 14:00: Coil @ 50 A. TBHT @ 330 A.

Tuesday, October 23, 2018, 14:30: Mass scans ok. TBHT back @ 320.

Emittance rig not calibrating?


Wednesday, October 24, 2018, 08:17 : E-rig drive motor PS re-plugged in. Switched on. E-rig calibrated & ready.

 FM:

Wednesday, October 24, 2018, 12:40 :  Vacuum 6.52e-7T. Auto-ramping TBHT  to 320 A @ 2A per min.

Wednesday, October 24, 2018, 13:02 : Auto-ramping finished. Anode Current@8.54mA. Vaccum ok @ 6.86E-7T

Wednesday, October 24, 2018, 13:40 : Vacuum 6.58e-7T. Auto-ramping TBHT  to 340 A @ 2A per min.

Wednesday, October 24, 2018, 13:55 : Auto-ramping finished. Anode Current@19.53mA. Vaccum ok @ 6.67E-7T

Wednesday, October 24, 2018, 15:18 : Vacuum 6.68e-7T. Auto-ramping TBHT  to 350 A @ 2A per min.

Wednesday, October 24, 2018, 15:27: Auto-ramping finished. Anode Current@41.51mA. Vaccum ok @ 6.82E-7T


Thursday, October 25, 2018, 08:27: IG6 Tripped off around 3am this morning. Attempts to restart, unsuccessful.

Thursday, October 25, 2018, 08:29: IV4 Closed. Vacuum group notified..

Thursday, October 25, 2018, 08:40: Filament burnt out. Vac group switched on IG6 secondary filament.

Thursday, October 25, 2018, 08:44: IV4 re-opened. Optics reset.


 Friday, October 26, 2018, 13:49: Auto-ramping TBHT / 360 A @ 1A per min.

Friday, October 26, 2018, 14:08:

TBHT @ 359.0 A / 3.074 V.

Anode @ 200 V / 94.017 - 95.238 mA.

Friday, October 26, 2018, 14:34: Mass & wire scans complete w/ screen shots on the word doc (FEBIAD COIL TEST 6-3 Caps in).


Sunday, October 28, 2018, 13:19: Remotely connect to check the status. Something happened around midnight.  See attached screenshots.

 It seems there is no bias. EPICS control whiteout.  Not sure how to proceed. Leaving evreything as is. 

Attachment 1: epics-plot-whiteout-181028_1322.PNG
epics-plot-whiteout-181028_1322.PNG
Attachment 2: epics-whiteout-181028_1322.PNG
epics-whiteout-181028_1322.PNG
Attachment 3: FEBIAD_COIL_Test_6-3_Caps_in.docx
Attachment 4: TS-2018-10-22-09-47.rga
  692   Monday, October 29, 2018, 10:56 Aaron Schmidt LP-FEBIAD  41790-942752-5Problems/SolutionsTS communications issues.

Sometime around 00:30 Sunday, October 28, 2018, the ISAC test stand lost communication to the PLC and shut down abruptly.

Monday, October 29, 2018, 09:00: PLC communications issues.

Rod N. was contacted this morning& the issues fixed by controls group.

  693   Monday, October 29, 2018, 11:03 Aaron Schmidt LP-FEBIAD  41790-942752-5DoneTest FEBIAD w/ Silver bolts & sealed end caps

Restarting FEBIAD

Monday, October 29, 2018, 10:56: Restart roughing & Turbo pumps.

Monday, October 29, 2018, 11:03: Ion gauges on.

Monday, October 29, 2018, 11:10: Power on

Monday, October 29, 2018, 11:12: Gas lines opened to pump down with TS. Vacuum @ 1.20e-05T.

Monday, October 29, 2018, 11:20: RGA restarted.


Tuesday, October 30, 2018, 07:24: BIAS / EZL / EE @ 30 / 13.5 / 1.5 kV. Anode @ 200 V.

Tuesday, October 30, 2018, 07:26: Auto-ramping TGHT / TBHT to 580 / 320 A @ 2A per min.

Tuesday, October 30, 2018, 07:31: Optics set.

Tuesday, October 30, 2018, 07:32 : IV4 Opened.

Tuesday, October 30, 2018, 08:50: Auto ramping aborted. While the source does increment, the actual power supply does not. It's not actually ramping.

Tuesday, October 30, 2018, 08:51: Unable to bump up the coil either. Tapping down the all the voltage to 0.

Tuesday, October 30, 2018, 08:55: IV4 closed. Gases closed.

Tuesday, October 30, 2018, 08:57: Power shut off & restarted.

Tuesday, October 30, 2018, 09:01: Gases re-opened.

Tuesday, October 30, 2018, 09:03: Sources switched on. Manually bumped up TGHT & TBHT. No response.

Tuesday, October 30, 2018, 09:04: Gases closed. Sources off.

Tuesday, October 30, 2018, 10:41: IV4 Opened.

Tuesday, October 30, 2018, 10:47: BIAS / EZL / EE @ 30 / 13.5 / 1.5 kV. Anode @ 200 V. Voltage limits reset for the TGHT, TBHT, & Coil. Responses given

Tuesday, October 30, 2018, 10:49: Auto-ramping TGHT / TBHT to 580 / 320 A @ 2A per min.

Tuesday, October 30, 2018, 10:55: Optics reset.

Tuesday, October 30, 2018, 11:01: Gases reopened. Vacuum @ 4.52e-07T.


Wednesday, October 31, 2018, 06:59: Auto-ramping TBHT to 330 A @ 2A per min.

Wednesday, October 31, 2018, 07:00: Coil @ 50 A.

Wednesday, October 31, 2018, 10:57 : Mass & wire scans appear ok.


FM:

Wednesday, October 31, 2018, 12:10 : Auto-ramping TBHT to 340 A @ 1A per min.

Wednesday, October 31, 2018, 12:49 : Vacuum @ 5.90e-07T.

Wednesday, October 31, 2018, 12:54 : Auto-ramping TBHT to 350 A @ 1A per min.

Wednesday, October 31, 2018, 15:25 : Vacuum @ 6.03E-7T.

Wednesday, October 31, 2018, 15:28 : Auto-ramping TBHT to 360 A @ 1A per min.

Wednesday, October 31, 2018, 15:34 : Changing Gain in FC0 from 33 nA to 100 nA

Wednesday, October 31, 2018, 15:42 : Increasing Gain in FC0 from 100 nA to 1 uA.

Voltage drop from 200 to 196 V.  The anode is current limited. Increasing the current limit on the anode from 100 mA to 110 mA .

Voltage back to 200 V 


FM

Thursday, November 01, 2018, 11:35 : Optics set to the previous tune. Check entry 688.

Anode set to 150 V. FC6 stable

Thursday, November 01, 2018, 11:37 Closing Neon Valve

Thursday, November 01, 2018, 12:38 : FC6 stable. Closing Argon valve.

Thursday, November 01, 2018, 13:05 : Assuming 720 nA entering the source (from Brad). Efficiency  measurement = 0.96%. Almost a factor of two larger than previous measurement with the same conditions!

Thursday, November 01, 2018, 13:08 : Opening the valves

Thursday, November 01, 2018, 15:41 : FC6 stable 5.38e-9A. Closing neon valve.

Thursday, November 01, 2018, 16:50 : FC6 stable overall 5.33e-9A. Closing argon Valve

Thursday, November 01, 2018, 18:42 : FC6 stable 1.44e-10A. efficiency= 0.72%

Opening gas valves


FM

Friday, November 02, 2018, 09:39 : Increasing AV to 200V.

Friday, November 02, 2018, 09:47 : Increasing anode current to 150 mA.

Friday, November 02, 2018, 18:01 : Coil sweep from 0-80 A for AV=25,50 completed. Efficiencies of 0.035 % and 0.34% respectively.

Leaving the cathode hot to continue test tomorrow morning.


Saturday, November 03, 2018, 18:01: Coil sweep from 0-80 A for AV=75,100 completed. Efficiencies of 0.56 % and 0.77% respectively.


Sunday, November 04, 2018, 09:01 : Anode current increased from 56 to around 95 mA overnight under the same conditions. This could indicate a cleaner surface on the cathode and hence a better emission or, the anode cathode distance is reducing fue to the thermal expansion.

Cooling down manually  TBHT to 300 A @ 2A per min.


Monday, November 05, 2018, 09:00: Short on TBHT appeared around 8 am. Investigate.

Monday, November 05, 2018, 09:15: Auto-cooling TGHT / TBHT @ 2 A per min.

Monday, November 05, 2018, 14:22: Gas lines closed.

Monday, November 05, 2018, 14:32: Sources/power off.

Monday, November 05, 2018, 14:40: Short confirmed up top. RGA off.

Monday, November 05, 2018, 14:43: IV4 closed/reset, Ion gauge off, Turbos off.

  760   Friday, September 27, 2019, 11:54 Aaron Schmidt LP-SIS  41790-943515-6 Conditioning UC#(29) HS

Using broken LP-FEBIAD oven. TBHT off.

Friday, September 27, 2019, 11:53: Roughing.

Friday, September 27, 2019, 13:14: Turbo pumps on.

Friday, September 27, 2019, 13:20: Ion gauge on.

Friday, September 27, 2019, 13:24: RGA on.


Monday, September 30, 2019, 07:30: Waterlines Opened. No response on RGA. Vacuum ok @ 5.39e-07T.

Monday, September 30, 2019, 07:34: Power / Sources on.

Monday, September 30, 2019, 07:35: Auto-ramping TGHT to 580 A @ 2 A per min.


Tuesday, October 01, 2019, 14:29: Auto-Cooling TGHT @ 2A per min.


Wednesday, October 02, 2019, 08:24: Power / sources off.

Wednesday, October 02, 2019, 08:25: RGA off.

Wednesday, October 02, 2019, 08:26: Ion gauge & Turbo pumps off.

Wednesday, October 02, 2019, 12:50: Venting & Waterlines purged.

  761   Wednesday, October 02, 2019, 13:42 Aaron Schmidt LP-SIS  41790-943515-9In progressConditioning UC#(30) HS

Wednesday, October 02, 2019, 13:42: Roughing.

Wednesday, October 02, 2019, 14:13: Turbo Pumps started.

Wednesday, October 02, 2019, 14:19: Ion gauge on.

Wednesday, October 02, 2019, 14:24: RGA on.


Thursday, October 03, 2019, 07:35: Waterlines opened.  No response on RGA. Vacuum ok @ 7.53e-07T

Thursday, October 03, 2019, 07:37: Power / Sources on.

Thursday, October 03, 2019, 07:38: Auto-ramping TGHT to 580 A @ 2 A per min.


Friday, October 04, 2019, 14:53: Auto-cooling @ 2 A per min. 


Wednesday, October 16, 2019, 10:17: RGA off.

Wednesday, October 16, 2019, 10:19: Power / Sources off. IG1 off. Turbo pumps off.

Attachment 1: UC#30.docx
  769   Friday, November 01, 2019, 16:01 Fernando Maldonado LP-SIS  41790-9 43515-9 In progressPeek seal tests on UCx#30 HS

Friday, November 01, 2019, 15:47: A symbiotic  peek seal assembly  was installed on the manipulator grip to verify the temperature at high voltage and high current.

Friday, November 01, 2019, 15:48: Three  thermocouples added for temperature readings.

Friday, November 01, 2019, 15:47: Roughing

Friday, November 01, 2019, 16:19: Turbos on.

Friday, November 01, 2019, 16:27: IG1 on.

Friday, November 01, 2019, 16:29: RGA on.


AS

Monday, November 04, 2019, 07:41: Waterlines opened. No response on RGA. Vacuum ok @ 3.97e-07T.

Monday, November 04, 2019, 07:44: Power / sources on.

Monday, November 04, 2019, 07:47: Auto-ramping TGHT to 300 A @ 2 A per min.

Monday, November 04, 2019, 10:57: Auto-ramping TGHT to 400 A @ 2 A per min.

FM

Monday, November 04, 2019, 12:11: Auto-ramping TGHT to 500 A @ 2 A per min.

Monday, November 04, 2019, 13:35: Auto-ramping TGHT to 600 A @ 2 A per min.

Monday, November 04, 2019, 14:48: Bias/EE/EZL @ 30/1.5/13 kV.

Monday, November 04, 2019, 15:03: Bias/EE/EZL @ 40/2.0/17 kV.

Monday, November 04, 2019, 15:04: Sparks observed after reaching 40 kV. TMS:TP6 down. Back down to 32 kV.

Monday, November 04, 2019, 15:04:  Bias to 40 kV. One spark only. Seems stable now.

Sparks occurred sparsely. Max time without a spark was 26 minutes.

Monday, November 04, 2019, 16:09: Bias/EE/EZL @ 30/1.5/13 kV.

Monday, November 04, 2019, 16:16: RGA restart.


AS

Tuesday, November 05, 2019, 08:30: RGA Restarted.

Tuesday, November 05, 2019, 08:32: Begin bringing up HV to 35 kV on BIAS.

Tuesday, November 05, 2019, 08:35: BIAS / EE / EZL @ 35 / 1.8 / 15 kV.

Tuesday, November 05, 2019, 08:55: No sparks after 20 minutes. Move up to 40 kV.

Tuesday, November 05, 2019, 08:59: BIAS / EE / EZL @ 40 / 2.0 / 20 kV.

Tuesday, November 05, 2019, 09:34: Audible Spark*. Restart TSG: CB1, VCB1, CBCx2. Reset & restart turbo pump. TMS:TP6. Appears on EZl cur.

Tuesday, November 05, 2019, 09:37: RGA off. Leave it off.

Tuesday, November 05, 2019, 09:38: Audible Spark* but not loud at all. Restart TSG: CB1, VCB1, CBCx2. Reset & restart turbo pump. TMS:TP6.

Tuesday, November 05, 2019, 10:00: BIAS / EE / EZL @ 35 / 1.8 / 17.5 kV.

Tuesday, November 05, 2019, 10:06: BIAS / EE / EZL @ 0 / 0 / 0 kV.

Tuesday, November 05, 2019, 10:07: Auto-cooling TGHT to 500 A @ 4 A per min.


Wednesday, November 13, 2019, 07:29: Turbo pumps & ion gauge off. RGA was already off.


Friday, November 15, 2019, 08:43: Venting.

 

 

Attachment 1: UC#30-plus-peak-seal.docx
Attachment 2: TS-2019-11-01-16-22.rga
Attachment 3: TS-2019-11-04-16-12.rga
Attachment 4: TS-2019-11-05-08-30.rga
  770   Friday, November 08, 2019, 14:21 Aaron Schmidt LP-SIS  41790-943515-9In progressNitrogen Bottle leak test

Friday, November 01, 2019, 14:00: Nitrogen bottle installed. Purged waterlines HS / FS#5. Purge done in conjunction with SiC#38

4.8 Purity

35-40 PSI pressure.

Successfully purged the line after 5 minutes.


Friday, November 08, 2019, 14:16: Begin FS#5 waterline purge w/ Nitrogen. Purge done in conjunction with UC#30 HS & assy.

4.8 Purity

12-14 PSI line pressure.

2500 PSI bottle pressure.

Full stream for a minute or so, then purged lightly. Drops / mist outgoing.

Friday, November 08, 2019, 14:24: Line Pressure at 10-11 PSI, brought back up to 14 PSI.

2400 PSI bottle pressure. Water / mist outgoing still.

Friday, November 08, 2019, 14:31: No water appears on line with purge.

Approx 2350 PSI bottle pressure.

Line pressure increased to 40 PSI. No water appeared on the purge line.

Friday, November 08, 2019, 14:34: Purge complete. 14 PSI appears successful.

  784   Wednesday, February 12, 2020, 11:04 Aaron Schmidt LP-SIS  41790-943515-1In progressConditioning HS

Using a broken LP-FEBIAD oven for out gassing. Tentative UC#31*

Wednesday, February 12, 2020, 10:57: Roughing.

Wednesday, February 12, 2020, 11:30: Turbo pumps on.


Thursday, February 13, 2020, 07:19: Waterlines opened. Vacuum ok @ 6.91e-07T.

Thursday, February 13, 2020, 07:22: Power / Sources on. TGHT only.

Thursday, February 13, 2020, 07:25: Auto-ramping TGHT @ 2A per min to 580 A.


Friday, February 14, 2020, 14:03: Auto-cooling @ 1 A per min.


Thursday, February 20, 2020, 07:33: PS, sources off, locked out.

Thursday, February 20, 2020, 07:34: Turbo pumps & IG1 off.

Thursday, February 20, 2020, 07:35: Waterlines purged. Nitrogen Bottle around 2050 PSI.

Thursday, February 20, 2020, 09:22: Venting.

 

Attachment 1: UC#31.docx
  800   Monday, July 13, 2020, 14:52 Aaron Schmidt LP-FEBIAD  41790-943515-2 UCx HS conditioning.

Broken LP-FEBIAD used to out gas.

Monday, July 13, 2020, 14:52: Roughing.

Monday, July 13, 2020, 15:29: Turbo pumps on. 

Monday, July 13, 2020, 15:34: Ion gauge on.


Tuesday, July 14, 2020, 06:44: Waterlines opened.

Tuesday, July 14, 2020, 06:47: Power / sources on.

Tuesday, July 14, 2020, 06:49: Auto-ramping TGHT to 580 A @ 2 A per min.

Tuesday, July 14, 2020, 14:15: Vacuum @ 2.59e-06T 


Friday, July 17, 2020, 10:54: Auto-cooling @ 1 A per min. 


Monday, July 20, 2020, 06:51: Power /  sources off.

Monday, July 20, 2020, 06:52: Ion gauge / Turbo pumps off.

Monday, July 20, 2020, 06:53: Re-attached N bottle / regulator for purging.

Monday, July 20, 2020, 06:57: Waterlines Purged. N bottle @ 1300 PSI.

  547   Tuesday, October 18, 2016, 13:29 Aaron SchmidtTa#50LP-SISITETM441790-541751-1In progressConditioning Ta#50

Tuesday, October 18, 2016, 13:22: Roughing.

Tuesday, October 18, 2016, 13:46: Turbo pumps started.

Tuesday, October 18, 2016, 13:52: Ion gauge on.

Tuesday, October 18, 2016, 13:54: Both turbo pumps at speed.

Tuesday, October 18, 2016, 15:05: RGA on.


Wednesday, October 19, 2016, 08:46: Power/Sources on

Wednesday, October 19, 2016, 08:49: Waterline opened. No response on RGA. Vacuum stable @4.40E-7T

Wednesday, October 19, 2016, 08:57: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .

Wednesday, October 19, 2016, 08:58 : IV4 opened.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 27.9 3.90
 TBHT 27.7 3.06
 COIL + EE 27.6 5.08
 TGHT  27.7 5.70

Wednesday, October 19, 2016, 09:01: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.

Wednesday, October 19, 2016, 11:50: TMS:FC0 moved out & TMS:RPM0B began scanning continuously on their own. Monitor.

Wednesday, October 19, 2016, 14:45: RGA Restarted.

Wednesday, October 19, 2016, 15:25: RGA Restarted (Unable to save 14:45 Startup)


Friday, October 21, 2016, 13:27: Auto Cooling @ 2A per min

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 37.3 3.92
 TBHT 28.2 3.07
 COIL + EE 27.9 5.07
 TGHT  29.8 5.62

Friday, October 21, 2016, 15:39: IV4 Closed


Monday, October 24, 2016, 09:10:

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 28.1 3.95
 TBHT 28.2 3.10
 COIL + EE 27.7 5.04
 TGHT  27.7 5.70

Monday, October 24, 2016, 09:10: Power/Sources off.

Monday, October 24, 2016, 09:20: Water lines purged.

Monday, October 24, 2016, 09:21: RGA off.

Monday, October 24, 2016, 09:23: Ion gauge & Turbo pumps off

Monday, October 24, 2016, 12:54: Venting.

Attachment 1: ER-580-200-30kV-m27.txt
value (Voltage offset      40.0000)
     80.0000
      0.7000
     80.0000
  29978.0220
21
21
./emit.dat
 -0.1154 -0.0916 -0.0937 -0.0952 -0.0964 -0.0955 -0.0980 -0.1019 -0.0943 -0.0964 -0.0980 -0.0940 -0.0919 -0.0919 -0.0964 -0.0946 -0.0992 -0.0949 -0.1016 -0.0992 -0.1031
 -0.0894 -0.0903 -0.0925 -0.0928 -0.0946 -0.0943 -0.0949 -0.0977 -0.0952 -0.0943 -0.0949 -0.0925 -0.0922 -0.0909 -0.0964 -0.0934 -0.0989 -0.0925 -0.1004 -0.0986 -0.1001
 -0.0909 -0.0925 -0.0928 -0.0943 -0.0949 -0.0946 -0.0946 -0.0952 -0.0943 -0.0967 -0.0961 -0.0925 -0.0934 -0.0925 -0.0937 -0.0916 -0.0916 -0.0922 -0.0970 -0.0983 -0.0928
 -0.0916 -0.0931 -0.0934 -0.0934 -0.0916 -0.0916 -0.0925 -0.0909 -0.0943 -0.0909 -0.0974 -0.0925 -0.0961 -0.0919 -0.0916 -0.0909 -0.0900 -0.0916 -0.0949 -0.0983 -0.0931
 -0.0894 -0.0903 -0.0934 -0.0934 -0.0934 -0.0882 -0.0891 -0.0903 -0.0922 -0.0919 -0.0955 -0.0916 -0.0943 -0.0900 -0.0891 -0.0876 -0.0879 -0.0879 -0.0937 -0.0955 -0.0888
 -0.0888 -0.0891 -0.0885 -0.0916 -0.0943 -0.0919 -0.0497 -0.0320 -0.0818 -0.0912 -0.0928 -0.0912 -0.0912 -0.0891 -0.0894 -0.0916 -0.0919 -0.0894 -0.0946 -0.0970 -0.0903
 -0.0876 -0.0888 -0.0894 -0.0912 -0.0961 -0.0919 -0.0354 0.0482 0.0952 -0.0259 -0.0916 -0.0897 -0.0906 -0.0919 -0.0928 -0.0903 -0.0919 -0.0912 -0.0961 -0.0974 -0.0900
 -0.0912 -0.0903 -0.0882 -0.0882 -0.0928 -0.0912 -0.0403 0.0674 0.2994 0.2567 0.0485 -0.0671 -0.0882 -0.0912 -0.0916 -0.0912 -0.0900 -0.0928 -0.0949 -0.0983 -0.0912
 -0.0922 -0.0928 -0.0912 -0.0876 -0.0848 -0.0797 -0.0470 -0.0299 0.4971 0.5460 0.3802 0.0650 -0.0308 -0.0906 -0.0891 -0.0876 -0.0900 -0.0909 -0.0943 -0.0970 -0.0900
 -0.0903 -0.0903 -0.0903 -0.0912 -0.0800 -0.0711 -0.0632 -0.0327 0.5188 0.7532 0.5743 0.1593 -0.0430 -0.0607 -0.0827 -0.0900 -0.0916 -0.0897 -0.0937 -0.0955 -0.0897
 -0.0891 -0.0900 -0.0916 -0.0900 -0.0919 -0.0723 -0.0430 -0.0201 0.2475 0.6775 0.5890 0.2744 -0.0375 -0.0583 -0.0742 -0.0818 -0.0879 -0.0897 -0.0943 -0.0952 -0.0882
 -0.0897 -0.0909 -0.0906 -0.0922 -0.0937 -0.0922 -0.0824 -0.0095 0.1239 0.4227 0.4150 0.2563 -0.0189 -0.0629 -0.0839 -0.0845 -0.0858 -0.0909 -0.0952 -0.0964 -0.0919
 -0.0925 -0.0912 -0.0925 -0.0919 -0.0949 -0.0912 -0.0897 -0.0885 -0.0565 0.1428 0.2325 0.1718 0.0174 -0.0620 -0.0885 -0.0897 -0.0897 -0.0894 -0.0943 -0.0986 -0.0903
 -0.0931 -0.0909 -0.0922 -0.0946 -0.0952 -0.0906 -0.0885 -0.0897 -0.0906 -0.0806 0.0296 0.0879 0.0458 -0.0549 -0.0876 -0.0894 -0.0900 -0.0894 -0.0922 -0.0955 -0.0906
 -0.0909 -0.0897 -0.0916 -0.0916 -0.0922 -0.0919 -0.0903 -0.0916 -0.0903 -0.0894 -0.0900 -0.0793 -0.0656 -0.0769 -0.0885 -0.0888 -0.0873 -0.0885 -0.0943 -0.0943 -0.0879
 -0.0891 -0.0873 -0.0891 -0.0906 -0.0922 -0.0903 -0.0922 -0.0931 -0.0879 -0.0882 -0.0897 -0.0885 -0.0870 -0.0894 -0.0916 -0.0906 -0.0900 -0.0894 -0.0934 -0.0961 -0.0897
 -0.0891 -0.0900 -0.0909 -0.0922 -0.0949 -0.0919 -0.0916 -0.0940 -0.0894 -0.0903 -0.0912 -0.0897 -0.0894 -0.0925 -0.0912 -0.0894 -0.0903 -0.0912 -0.0958 -0.0964 -0.0894
 -0.0906 -0.0909 -0.0922 -0.0934 -0.0940 -0.0909 -0.0888 -0.0900 -0.0891 -0.0906 -0.0937 -0.0897 -0.0891 -0.0882 -0.0903 -0.0900 -0.0912 -0.0912 -0.0940 -0.0980 -0.0925
 -0.0912 -0.0928 -0.0925 -0.0940 -0.0937 -0.0897 -0.0900 -0.0903 -0.0873 -0.0906 -0.0931 -0.0912 -0.0922 -0.0897 -0.0900 -0.0903 -0.0894 -0.0906 -0.0937 -0.0958 -0.0912
 -0.0888 -0.0891 -0.0909 -0.0919 -0.0937 -0.0912 -0.0906 -0.0879 -0.0854 -0.0876 -0.0912 -0.0900 -0.0906 -0.0888 -0.0894 -0.0897 -0.0882 -0.0885 -0.0912 -0.0940 -0.0903
 -0.0870 -0.0891 -0.0891 -0.0906 -0.0909 -0.0928 -0.0897 -0.0894 -0.0848 -0.0839 -0.0882 -0.0888 -0.0882 -0.0909 -0.0903 -0.0894 -0.0897 -0.0888 -0.0937 -0.0937 -0.0864
Attachment 2: ER-580-200-30kV-m27(2).txt
value (Voltage offset      40.0000)
     80.0000
      0.7000
     80.0000
  29978.0220
21
21
./emit.dat
 -0.1202 -0.0977 -0.0977 -0.0967 -0.0958 -0.1004 -0.0983 -0.0977 -0.1047 -0.0989 -0.0983 -0.0970 -0.0970 -0.0964 -0.0958 -0.0961 -0.0983 -0.0964 -0.0961 -0.1004 -0.0986
 -0.0916 -0.0903 -0.0919 -0.0919 -0.0903 -0.0986 -0.0937 -0.0952 -0.0946 -0.0916 -0.0928 -0.0919 -0.0919 -0.0891 -0.0922 -0.0912 -0.0912 -0.0919 -0.0955 -0.0946 -0.0925
 -0.0912 -0.0928 -0.0903 -0.0922 -0.0916 -0.0958 -0.0937 -0.0900 -0.0909 -0.0943 -0.0919 -0.0909 -0.0894 -0.0928 -0.0912 -0.0943 -0.0888 -0.0891 -0.0928 -0.0946 -0.0909
 -0.0916 -0.0906 -0.0916 -0.0967 -0.0958 -0.0912 -0.0949 -0.0922 -0.0897 -0.0882 -0.0912 -0.0916 -0.0934 -0.0909 -0.0916 -0.0906 -0.0925 -0.0916 -0.0931 -0.0961 -0.0943
 -0.0940 -0.0937 -0.0934 -0.0974 -0.0964 -0.0888 -0.0946 -0.0922 -0.0986 -0.0964 -0.0928 -0.0928 -0.0940 -0.0903 -0.0903 -0.0912 -0.0928 -0.0928 -0.0934 -0.0974 -0.0937
 -0.0925 -0.0925 -0.0931 -0.0970 -0.0958 -0.0833 -0.0595 -0.0632 -0.0943 -0.0922 -0.0940 -0.0934 -0.0943 -0.0937 -0.0922 -0.0940 -0.0925 -0.0909 -0.0952 -0.0946 -0.0922
 -0.0891 -0.0916 -0.0903 -0.0955 -0.0940 -0.0854 -0.0610 -0.0021 -0.0043 -0.0693 -0.0906 -0.0909 -0.0897 -0.0912 -0.0937 -0.0919 -0.0909 -0.0894 -0.0919 -0.0946 -0.0897
 -0.0903 -0.0891 -0.0891 -0.0925 -0.0922 -0.0916 -0.0751 0.0476 0.1105 0.0775 -0.0342 -0.0861 -0.0912 -0.0879 -0.0906 -0.0894 -0.0894 -0.0909 -0.0940 -0.0931 -0.0906
 -0.0934 -0.0912 -0.0909 -0.0916 -0.0912 -0.0836 -0.0720 0.0098 0.2060 0.1773 0.0888 -0.0381 -0.0729 -0.0922 -0.0891 -0.0909 -0.0931 -0.0922 -0.0919 -0.0958 -0.0925
 -0.0937 -0.0916 -0.0931 -0.0949 -0.0922 -0.0827 -0.0793 -0.0565 0.2127 0.2264 0.1419 -0.0339 -0.0781 -0.0821 -0.0912 -0.0937 -0.0943 -0.0925 -0.0946 -0.0952 -0.0934
 -0.0919 -0.0928 -0.0925 -0.0964 -0.0958 -0.0830 -0.0729 -0.0583 0.0943 0.1904 0.1215 -0.0253 -0.0732 -0.0830 -0.0873 -0.0897 -0.0934 -0.0922 -0.0937 -0.0949 -0.0900
 -0.0912 -0.0916 -0.0906 -0.0928 -0.0928 -0.0919 -0.0864 -0.0555 0.0027 0.1202 0.0879 -0.0012 -0.0650 -0.0861 -0.0897 -0.0891 -0.0894 -0.0919 -0.0919 -0.0925 -0.0903
 -0.0897 -0.0897 -0.0916 -0.0943 -0.0946 -0.0885 -0.0906 -0.0885 -0.0732 0.0076 0.0360 -0.0037 -0.0497 -0.0848 -0.0891 -0.0903 -0.0897 -0.0891 -0.0909 -0.0943 -0.0912
 -0.0922 -0.0906 -0.0903 -0.0961 -0.0958 -0.0897 -0.0912 -0.0922 -0.0946 -0.0836 -0.0357 -0.0269 -0.0415 -0.0839 -0.0909 -0.0897 -0.0909 -0.0919 -0.0919 -0.0961 -0.0928
 -0.0919 -0.0916 -0.0925 -0.0983 -0.0986 -0.0903 -0.0934 -0.0940 -0.0967 -0.0934 -0.0909 -0.0861 -0.0775 -0.0894 -0.0922 -0.0934 -0.0925 -0.0909 -0.0928 -0.0955 -0.0925
 -0.0925 -0.0928 -0.0909 -0.0964 -0.0961 -0.0931 -0.0912 -0.0952 -0.0940 -0.0903 -0.0903 -0.0916 -0.0906 -0.0900 -0.0922 -0.0909 -0.0903 -0.0900 -0.0922 -0.0949 -0.0903
 -0.0903 -0.0897 -0.0909 -0.0940 -0.0946 -0.0909 -0.0919 -0.0912 -0.0922 -0.0897 -0.0891 -0.0900 -0.0891 -0.0900 -0.0906 -0.0894 -0.0912 -0.0894 -0.0912 -0.0928 -0.0894
 -0.0906 -0.0888 -0.0891 -0.0934 -0.0928 -0.0897 -0.0888 -0.0906 -0.0940 -0.0882 -0.0900 -0.0900 -0.0909 -0.0885 -0.0894 -0.0885 -0.0903 -0.0903 -0.0897 -0.0946 -0.0906
 -0.0897 -0.0903 -0.0919 -0.0952 -0.0928 -0.0894 -0.0912 -0.0912 -0.0909 -0.0906 -0.0906 -0.0916 -0.0903 -0.0922 -0.0900 -0.0912 -0.0919 -0.0916 -0.0916 -0.0961 -0.0922
 -0.0925 -0.0925 -0.0916 -0.0952 -0.0949 -0.0922 -0.0919 -0.0922 -0.0940 -0.0912 -0.0922 -0.0916 -0.0940 -0.0919 -0.0912 -0.0940 -0.0925 -0.0922 -0.0943 -0.0958 -0.0922
 -0.0916 -0.0916 -0.0919 -0.0943 -0.0952 -0.0919 -0.0922 -0.0909 -0.0900 -0.0888 -0.0906 -0.0928 -0.0906 -0.0916 -0.0912 -0.0916 -0.0900 -0.0906 -0.0925 -0.0937 -0.0912
Attachment 3: ER-580-200-30kV-m27(3).txt
value (Voltage offset      40.0000)
     80.0000
      0.7000
     80.0000
  29978.0220
21
21
./emit.dat
 -0.1181 -0.0970 -0.0977 -0.0970 -0.0992 -0.0970 -0.1059 -0.0967 -0.0974 -0.0970 -0.0983 -0.0986 -0.0961 -0.0961 -0.0970 -0.0989 -0.0986 -0.0967 -0.0989 -0.0970 -0.0964
 -0.0946 -0.0928 -0.0919 -0.0937 -0.0949 -0.0943 -0.1013 -0.0906 -0.0967 -0.0943 -0.0946 -0.0934 -0.0961 -0.0937 -0.0928 -0.0928 -0.0912 -0.0922 -0.0958 -0.0925 -0.0912
 -0.0928 -0.0934 -0.0928 -0.0937 -0.0922 -0.0931 -0.0940 -0.0934 -0.0955 -0.0970 -0.0934 -0.0934 -0.0928 -0.0934 -0.0922 -0.0916 -0.0903 -0.0909 -0.0940 -0.0928 -0.0925
 -0.0916 -0.0909 -0.0912 -0.0922 -0.0943 -0.0888 -0.0946 -0.0955 -0.0906 -0.0891 -0.0916 -0.0912 -0.0909 -0.0888 -0.0934 -0.0919 -0.0925 -0.0925 -0.0952 -0.0937 -0.0931
 -0.0916 -0.0906 -0.0912 -0.0925 -0.0955 -0.0922 -0.0949 -0.0934 -0.0934 -0.0943 -0.0931 -0.0919 -0.0922 -0.0912 -0.0912 -0.0934 -0.0922 -0.0922 -0.0949 -0.0940 -0.0946
 -0.0906 -0.0916 -0.0919 -0.0937 -0.0970 -0.0922 -0.0803 -0.0644 -0.0928 -0.0922 -0.0943 -0.0934 -0.0952 -0.0949 -0.0931 -0.0946 -0.0925 -0.0928 -0.0970 -0.0940 -0.0894
 -0.0919 -0.0934 -0.0940 -0.0937 -0.0964 -0.0897 -0.0693 -0.0311 -0.0131 -0.0650 -0.0955 -0.0928 -0.0916 -0.0937 -0.0937 -0.0943 -0.0916 -0.0925 -0.0952 -0.0916 -0.0916
 -0.0919 -0.0919 -0.0922 -0.0891 -0.0937 -0.0891 -0.0714 -0.0162 0.0900 0.0693 -0.0272 -0.0790 -0.0909 -0.0894 -0.0922 -0.0903 -0.0894 -0.0906 -0.0931 -0.0931 -0.0903
 -0.0912 -0.0906 -0.0888 -0.0891 -0.0903 -0.0864 -0.0742 -0.0595 0.1898 0.2142 0.1416 -0.0146 -0.0610 -0.0891 -0.0888 -0.0912 -0.0916 -0.0916 -0.0934 -0.0934 -0.0931
 -0.0912 -0.0894 -0.0912 -0.0928 -0.0903 -0.0827 -0.0833 -0.0626 0.2075 0.3119 0.2277 0.0372 -0.0696 -0.0787 -0.0888 -0.0916 -0.0937 -0.0934 -0.0931 -0.0946 -0.0919
 -0.0916 -0.0928 -0.0931 -0.0946 -0.0967 -0.0851 -0.0729 -0.0586 0.0763 0.2628 0.2240 0.0833 -0.0693 -0.0793 -0.0851 -0.0897 -0.0931 -0.0919 -0.0946 -0.0925 -0.0916
 -0.0940 -0.0928 -0.0943 -0.0934 -0.0958 -0.0912 -0.0916 -0.0513 0.0076 0.1422 0.1401 0.0693 -0.0598 -0.0790 -0.0894 -0.0885 -0.0888 -0.0897 -0.0922 -0.0906 -0.0891
 -0.0916 -0.0925 -0.0909 -0.0912 -0.0922 -0.0900 -0.0916 -0.0897 -0.0815 0.0153 0.0528 0.0244 -0.0400 -0.0787 -0.0879 -0.0885 -0.0894 -0.0888 -0.0931 -0.0925 -0.0925
 -0.0909 -0.0885 -0.0891 -0.0916 -0.0940 -0.0903 -0.0934 -0.0906 -0.0931 -0.0851 -0.0385 -0.0131 -0.0281 -0.0760 -0.0885 -0.0909 -0.0916 -0.0909 -0.0943 -0.0940 -0.0931
 -0.0891 -0.0894 -0.0916 -0.0928 -0.0952 -0.0928 -0.0955 -0.0928 -0.0980 -0.0891 -0.0882 -0.0851 -0.0839 -0.0882 -0.0919 -0.0912 -0.0916 -0.0919 -0.0952 -0.0949 -0.0903
 -0.0912 -0.0916 -0.0931 -0.0937 -0.0958 -0.0909 -0.0961 -0.0934 -0.0937 -0.0928 -0.0928 -0.0925 -0.0928 -0.0928 -0.0928 -0.0922 -0.0925 -0.0922 -0.0940 -0.0925 -0.0906
 -0.0912 -0.0934 -0.0925 -0.0919 -0.0946 -0.0916 -0.0928 -0.0894 -0.1001 -0.0922 -0.0912 -0.0903 -0.0916 -0.0931 -0.0909 -0.0897 -0.0909 -0.0906 -0.0912 -0.0919 -0.0888
 -0.0919 -0.0912 -0.0909 -0.0916 -0.0937 -0.0897 -0.0937 -0.0903 -0.0943 -0.0916 -0.0916 -0.0916 -0.0906 -0.0909 -0.0891 -0.0897 -0.0900 -0.0906 -0.0916 -0.0903 -0.0925
 -0.0906 -0.0900 -0.0894 -0.0909 -0.0937 -0.0912 -0.0934 -0.0912 -0.0983 -0.0876 -0.0891 -0.0900 -0.0916 -0.0894 -0.0894 -0.0891 -0.0900 -0.0909 -0.0946 -0.0928 -0.0925
 -0.0912 -0.0909 -0.0891 -0.0937 -0.0946 -0.0912 -0.0940 -0.0934 -0.0955 -0.0888 -0.0912 -0.0912 -0.0912 -0.0919 -0.0928 -0.0943 -0.0925 -0.0931 -0.0943 -0.0934 -0.0928
 -0.0906 -0.0916 -0.0928 -0.0934 -0.0955 -0.0928 -0.0946 -0.0891 -0.0952 -0.0885 -0.0900 -0.0909 -0.0925 -0.0922 -0.0922 -0.0900 -0.0922 -0.0922 -0.0937 -0.0909 -0.0894
Attachment 4: ER-580-200-30kV-m39.txt
value (Voltage offset      40.0000)
     80.0000
      0.7000
     80.0000
  29978.0220
21
21
./emit.dat
 -0.1172 -0.0961 -0.0958 -0.0970 -0.0995 -0.0980 -0.0961 -0.0995 -0.0992 -0.0961 -0.0995 -0.0989 -0.0992 -0.0961 -0.0983 -0.0974 -0.0986 -0.0940 -0.1013 -0.0989 -0.0989
 -0.0906 -0.0903 -0.0906 -0.0909 -0.0922 -0.0928 -0.0931 -0.0986 -0.0946 -0.0946 -0.0943 -0.0961 -0.0937 -0.0958 -0.0922 -0.0916 -0.0922 -0.0940 -0.0986 -0.0955 -0.0955
 -0.0928 -0.0925 -0.0912 -0.0928 -0.0940 -0.0934 -0.0934 -0.0958 -0.0964 -0.0937 -0.0940 -0.0934 -0.0928 -0.0928 -0.0925 -0.0925 -0.0952 -0.0934 -0.0986 -0.0970 -0.0940
 -0.0940 -0.0931 -0.0937 -0.0970 -0.0897 -0.0925 -0.0958 -0.0931 -0.0919 -0.0909 -0.0909 -0.0931 -0.0897 -0.0916 -0.0940 -0.0940 -0.0934 -0.0928 -0.0958 -0.0967 -0.0900
 -0.0934 -0.0928 -0.0922 -0.0986 -0.0934 -0.0897 -0.0922 -0.0943 -0.0934 -0.0912 -0.0919 -0.0912 -0.0946 -0.0922 -0.0916 -0.0937 -0.0925 -0.0919 -0.0946 -0.0955 -0.0940
 -0.0912 -0.0906 -0.0925 -0.0958 -0.0952 -0.0955 -0.0961 -0.0928 -0.0952 -0.0916 -0.0928 -0.0949 -0.0937 -0.0952 -0.0937 -0.0952 -0.0912 -0.0931 -0.0946 -0.0946 -0.0894
 -0.0897 -0.0922 -0.0888 -0.0949 -0.0940 -0.0934 -0.0961 -0.0946 -0.0912 -0.0928 -0.0943 -0.0940 -0.0937 -0.0961 -0.0928 -0.0937 -0.0922 -0.0931 -0.0949 -0.0961 -0.0958
 -0.0919 -0.0906 -0.0912 -0.0970 -0.0906 -0.0912 -0.0958 -0.0925 -0.0809 -0.0858 -0.0897 -0.0937 -0.0903 -0.0919 -0.0906 -0.0934 -0.0937 -0.0928 -0.0974 -0.0964 -0.0916
 -0.0943 -0.0940 -0.0925 -0.0961 -0.0931 -0.0906 -0.0912 -0.0934 -0.0610 -0.0494 -0.0681 -0.0894 -0.0937 -0.0894 -0.0900 -0.0909 -0.0934 -0.0925 -0.0961 -0.0952 -0.0912
 -0.0928 -0.0903 -0.0934 -0.0998 -0.0937 -0.0919 -0.0940 -0.0934 -0.0641 -0.0165 -0.0314 -0.0763 -0.0903 -0.0925 -0.0928 -0.0952 -0.0925 -0.0922 -0.0955 -0.0943 -0.0909
 -0.0912 -0.0925 -0.0900 -0.0961 -0.0961 -0.0934 -0.0937 -0.0937 -0.0851 -0.0357 -0.0394 -0.0638 -0.0909 -0.0906 -0.0928 -0.0952 -0.0919 -0.0916 -0.0952 -0.0940 -0.0928
 -0.0912 -0.0900 -0.0909 -0.0974 -0.0934 -0.0925 -0.0977 -0.0949 -0.0903 -0.0745 -0.0598 -0.0702 -0.0876 -0.0900 -0.0839 -0.0900 -0.0916 -0.0925 -0.0952 -0.0961 -0.0922
 -0.0934 -0.0922 -0.0912 -0.0977 -0.0916 -0.0909 -0.0958 -0.0940 -0.0931 -0.0894 -0.0848 -0.0836 -0.0912 -0.0900 -0.0528 -0.0897 -0.0888 -0.0922 -0.0977 -0.0958 -0.0925
 -0.0931 -0.0906 -0.0940 -0.0989 -0.0909 -0.0916 -0.0952 -0.0919 -0.0925 -0.0916 -0.0888 -0.0912 -0.0885 -0.0891 -0.0864 -0.0897 -0.0897 -0.0912 -0.0922 -0.0943 -0.0900
 -0.0937 -0.0928 -0.0919 -0.0992 -0.0940 -0.0943 -0.0937 -0.0937 -0.0937 -0.0897 -0.0912 -0.0922 -0.0931 -0.0925 -0.0912 0.2444 -0.0891 -0.0900 -0.0952 -0.0925 -0.0885
 -0.0912 -0.0903 -0.0912 -0.0974 -0.0943 -0.0934 -0.0958 -0.0940 -0.0925 -0.0937 -0.0922 -0.0946 -0.0919 -0.0922 -0.0934 -0.0317 -0.0873 -0.0912 -0.0964 -0.0946 -0.0894
 -0.0922 -0.0906 -0.0903 -0.0958 -0.0952 -0.0937 -0.0955 -0.0955 -0.0949 -0.0916 -0.0931 -0.0934 -0.0937 -0.0943 -0.0919 -0.0595 -0.0928 -0.0922 -0.0967 -0.0955 -0.0952
 -0.0931 -0.0897 -0.0906 -0.0961 -0.0928 -0.0916 -0.0970 -0.0946 -0.0922 -0.0925 -0.0897 -0.0922 -0.0903 -0.0873 -0.0900 -0.0919 -0.0949 -0.0937 -0.0937 -0.0952 -0.0916
 -0.0940 -0.0928 -0.0925 -0.0974 -0.0922 -0.0912 -0.0946 -0.0937 -0.0922 -0.0891 -0.0916 -0.0916 -0.0916 -0.0900 -0.0916 -0.0922 -0.0919 -0.0922 -0.0949 -0.0943 -0.0931
 -0.0931 -0.0909 -0.0919 -0.0970 -0.0931 -0.0922 -0.0949 -0.0925 -0.0900 -0.0922 -0.0891 -0.0928 -0.0903 -0.0922 -0.0940 -0.0958 -0.0925 -0.0903 -0.0922 -0.0952 -0.0900
 -0.0919 -0.0900 -0.0888 -0.0958 -0.0940 -0.0922 -0.0949 -0.0937 -0.0934 -0.0922 -0.0928 -0.0937 -0.0943 -0.0949 -0.0928 -0.0937 -0.0903 -0.0906 -0.0955 -0.0937 -0.0925
Attachment 5: ITE-TM4-Ta#50-LP-SIS_Test_Stand.docx
Attachment 6: Ta#50.docx
Attachment 7: TS-2016-10-18-15-05.rga
Attachment 8: TS-2016-10-19-14-42.rga
Attachment 9: OMEGA_OM-DAQPRO-5300_24-Oct-16_11.03.10.dlp
  574   Friday, March 24, 2017, 14:43 Aaron SchmidtTa#51LP-SISITWTM241790-441751-5In progressConditioning Ta#51

Friday, March 24, 2017, 15:40: Roughing.

Friday, March 24, 2017, 15:59: Turbo pumps on.

Friday, March 24, 2017, 16:04: Ion gauge on.

Friday, March 24, 2017, 16:07: RGA on. Turbo pumps @ speed.


Monday, March 27, 2017, 09:07: Water lines opened. No response on RGA. Vacuum @ 1.81e-07T

Monday, March 27, 2017, 09:10: Power on.

Monday, March 27, 2017, 09:50: Troubleshooting: Sources/Optics non-responsive. Controls group notified.

Monday, March 27, 2017, 11:16: Optics all green minus the magnet TMS : MB1. PLC issue reported.

Monday, March 27, 2017, 11:30: PLC reboot. Brief whiteout on screens. Screen function returned. TMS : MB1 still interlocked.

Monday, March 27, 2017, 14:30: TMS : MB1 no longer interlocked. Optics non-responsive to 'P/S On'. Voltages can be adjust while optics are 'in black'


Tuesday, March 28, 2017, 09:30: Optics still have error but can be run.

Tuesday, March 28, 2017, 09:43: BIAS / EZL / EE @ 30 / 13 / 1.5 kV . Spark noted while bring up BIAS.

Tuesday, March 28, 2017, 09:45: IV4 opened.

Tuesday, March 28, 2017, 11:13: RGA Fatal error. Restarted 11:15.

Tuesday, March 28, 2017, 12:57: Optics ok.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 27.5 3.89
 TBHT 27.3 3.00
 COIL + EE 27.1 4.94
 TGHT  27.1 5.59

Tuesday, March 28, 2017, 13:01: Auto ramping TGHT / TBHT to 300 / 100 A @ 2A per min.

Tuesday, March 28, 2017, 15:52: TGHT / TBHT @ 300 / 100. Vacuum @ 2.16e-07T

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 27.5 3.85
 TBHT 26.7 3.12
 COIL + EE 25.8 4.92
 TGHT  26.5 5.59

 


Wednesday, March 29, 2017, 08:25: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.

Wednesday, March 29, 2017, 11:23: TMS : MB1 set from 'Reverse' to 'Normal' manually. Interlock Reset. (Mass Scan ok).

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 36.9 3.82
 TBHT 27.2 3.07
 COIL + EE 27.4 4.97
 TGHT  29.4 5.58

Wednesday, March 29, 2017, 13:09: Auto ramping TBHT to 220 A @ 2A per min.

Wednesday, March 29, 2017, 13:20: Source page @ 218.9A - Read back on power supply @ 219A. Fatal Error on RGA.

Wednesday, March 29, 2017, 13:27: Manually ramp TBHT to 230 A @ 2A per min (approx).

Wednesday, March 29, 2017, 15:15: DAC power supply appears to have failed. Pop sound and electrical smoke/smell noticed. 

Wednesday, March 29, 2017, 15:28: Auto cooling TBHT to 220 A @ 2A per min.


Thursday, March 30, 2017, 08:42: Auto-cooling @ 2A per min.


Friday, March 31, 2017, 08:35: IV4 closed.

Friday, March 31, 2017, 08:44: Power/sources off.

Friday, March 31, 2017, 08:46: Ion gauge, Turbos off. Waterlines purged.


Thursday, April 06, 2017, 10:40: Venting

Attachment 1: ER-580-230-30kV-M22.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
     80.0000
  29978.0220
21
21
./emit.dat
 -0.1138 -0.0684 -0.0720 -0.0696 -0.0681 -0.0662 -0.0662 -0.0668 -0.0671 -0.0763 -0.0684 -0.0696 -0.0702 -0.0705 -0.0674 -0.0674 -0.0690 -0.0674 -0.0772 -0.0684 -0.0668
 -0.0650 -0.0668 -0.0659 -0.0653 -0.0620 -0.0626 -0.0653 -0.0635 -0.0653 -0.0726 -0.0665 -0.0674 -0.0644 -0.0677 -0.0635 -0.0656 -0.0644 -0.0620 -0.0717 -0.0641 -0.0684
 -0.0632 -0.0647 -0.0638 -0.0632 -0.0638 -0.0638 -0.0650 -0.0650 -0.0635 -0.0629 -0.0635 -0.0653 -0.0647 -0.0635 -0.0659 -0.0626 -0.0638 -0.0613 -0.0604 -0.0626 -0.0656
 -0.0616 -0.0647 -0.0641 -0.0656 -0.0571 -0.0641 -0.0644 -0.0650 -0.0620 -0.0610 -0.0638 -0.0626 -0.0613 -0.0635 -0.0656 -0.0629 -0.0650 -0.0620 -0.0623 -0.0644 -0.0638
 -0.0598 -0.0653 -0.0629 -0.0702 -0.0656 -0.0629 -0.0638 -0.0635 -0.0629 -0.0632 -0.0616 -0.0693 -0.0677 -0.0626 -0.0644 -0.0647 -0.0644 -0.0626 -0.0732 -0.0650 -0.0623
 -0.0616 -0.0668 -0.0644 -0.0662 -0.0665 -0.0635 -0.0626 -0.0629 -0.0626 -0.0647 -0.0656 -0.0665 -0.0644 -0.0644 -0.0632 -0.0641 -0.0629 -0.0659 -0.0613 -0.0607 -0.0595
 -0.0632 -0.0626 -0.0632 -0.0674 -0.0632 -0.0659 -0.0641 -0.0623 -0.0653 -0.0644 -0.0656 -0.0687 -0.0623 -0.0632 -0.0623 -0.0613 -0.0629 -0.0650 -0.0598 -0.0598 -0.0598
 -0.0623 -0.0592 -0.0568 -0.0705 -0.0635 -0.0613 -0.0629 -0.0626 -0.0653 -0.0650 -0.0616 -0.0674 -0.0577 -0.0635 -0.0650 -0.0607 -0.0610 -0.0604 -0.0623 -0.0623 -0.0644
 -0.0650 -0.0595 -0.0613 -0.0629 -0.0641 -0.0629 -0.0626 -0.0644 -0.0650 -0.0613 -0.0586 -0.0684 -0.0626 -0.0616 -0.0620 -0.0623 -0.0668 -0.0623 -0.0632 -0.0656 -0.0656
 -0.0650 -0.0620 -0.0638 -0.0659 -0.0677 -0.0623 -0.0632 -0.0620 -0.0632 -0.0595 -0.0616 -0.0705 -0.0693 -0.0623 -0.0616 -0.0629 -0.0665 -0.0613 -0.0610 -0.0647 -0.0638
 -0.0629 -0.0638 -0.0668 -0.0729 -0.0607 -0.0632 -0.0601 -0.0601 -0.0607 -0.0620 -0.0620 -0.0705 -0.0671 -0.0623 -0.0647 -0.0641 -0.0632 -0.0604 -0.0668 -0.0629 -0.0626
 -0.0607 -0.0653 -0.0635 -0.0635 -0.0653 -0.0629 -0.0610 -0.0613 -0.0632 -0.0638 -0.0659 -0.0702 -0.0635 -0.0626 -0.0653 -0.0623 -0.0604 -0.0598 -0.0620 -0.0623 -0.0626
 -0.0598 -0.0659 -0.0616 -0.0623 -0.0638 -0.0620 -0.0620 -0.0613 -0.0616 -0.0635 -0.0638 -0.0681 -0.0644 -0.0604 -0.0656 -0.0616 -0.0613 -0.0613 -0.0638 -0.0635 -0.0632
 -0.0629 -0.0610 -0.0598 -0.0687 -0.0635 -0.0629 -0.0638 -0.0616 -0.0610 -0.0604 -0.0623 -0.0638 -0.0653 -0.0620 -0.0632 -0.0595 -0.0632 -0.0650 -0.0659 -0.0647 -0.0641
 -0.0629 -0.0595 -0.0632 -0.0690 -0.0650 -0.0635 -0.0616 -0.0604 -0.0595 -0.0577 -0.0583 -0.0702 -0.0613 -0.0647 -0.0598 -0.0635 -0.0647 -0.0653 -0.0696 -0.0595 -0.0620
 -0.0638 -0.0613 -0.0641 -0.0684 -0.0620 -0.0616 -0.0610 -0.0604 -0.0632 -0.0583 -0.0595 -0.0720 -0.0635 -0.0638 -0.0626 -0.0604 -0.0650 -0.0632 -0.0629 -0.0583 -0.0604
 -0.0610 -0.0623 -0.0613 -0.0690 -0.0613 -0.0607 -0.0601 -0.0610 -0.0626 -0.0595 -0.0580 -0.0674 -0.0632 -0.0644 -0.0644 -0.0620 -0.0641 -0.0629 -0.0647 -0.0601 -0.0632
 -0.0635 -0.0595 -0.0638 -0.0647 -0.0644 -0.0632 -0.0623 -0.0638 -0.0632 -0.0577 -0.0549 -0.0656 -0.0641 -0.0632 -0.0638 -0.0601 -0.0635 -0.0607 -0.0623 -0.0644 -0.0671
 -0.0650 -0.0616 -0.0632 -0.0662 -0.0684 -0.0635 -0.0644 -0.0635 -0.0620 -0.0607 -0.0574 -0.0644 -0.0601 -0.0589 -0.0635 -0.0641 -0.0616 -0.0616 -0.0641 -0.0650 -0.0644
 -0.0620 -0.0659 -0.0629 -0.0674 -0.0653 -0.0644 -0.0632 -0.0620 -0.0626 -0.0607 -0.0613 -0.0674 -0.0610 -0.0598 -0.0641 -0.0635 -0.0626 -0.0623 -0.0650 -0.0613 -0.0607
 -0.0595 -0.0641 -0.0623 -0.0653 -0.0632 -0.0647 -0.0635 -0.0601 -0.0613 -0.0626 -0.0650 -0.0690 -0.0613 -0.0623 -0.0665 -0.0638 -0.0613 -0.0644 -0.0601 -0.0620 -0.0601
Attachment 2: ER-580-230-30kV-M38.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
     80.0000
  29978.0220
21
21
./emit.dat
 -0.1169 -0.0748 -0.0684 -0.0653 -0.0705 -0.0720 -0.0681 -0.0693 -0.0677 -0.0766 -0.0677 -0.0668 -0.0668 -0.0687 -0.0690 -0.0674 -0.0690 -0.0690 -0.0784 -0.0711 -0.0699
 -0.0665 -0.0693 -0.0690 -0.0626 -0.0681 -0.0662 -0.0629 -0.0638 -0.0610 -0.0717 -0.0677 -0.0650 -0.0638 -0.0616 -0.0665 -0.0656 -0.0638 -0.0659 -0.0754 -0.0674 -0.0659
 -0.0671 -0.0635 -0.0668 -0.0647 -0.0659 -0.0659 -0.0604 -0.0653 -0.0635 -0.0745 -0.0659 -0.0677 -0.0647 -0.0647 -0.0623 -0.0674 -0.0665 -0.0653 -0.0623 -0.0623 -0.0641
 -0.0699 -0.0644 -0.0656 -0.0690 -0.0623 -0.0641 -0.0656 -0.0635 -0.0656 -0.0668 -0.0650 -0.0662 -0.0650 -0.0650 -0.0632 -0.0650 -0.0629 -0.0613 -0.0677 -0.0665 -0.0653
 -0.0702 -0.0635 -0.0684 -0.0647 -0.0723 -0.0650 -0.0647 -0.0641 -0.0674 -0.0644 -0.0632 -0.0668 -0.0620 -0.0638 -0.0656 -0.0653 -0.0632 -0.0601 -0.0623 -0.0668 -0.0665
 -0.0671 -0.0629 -0.0659 -0.0677 -0.0674 -0.0656 -0.0641 -0.0638 -0.0668 -0.0613 -0.0632 -0.0693 -0.0604 -0.0647 -0.0647 -0.0656 -0.0632 -0.0626 -0.0632 -0.0638 -0.0635
 -0.0659 -0.0620 -0.0684 -0.0638 -0.0690 -0.0635 -0.0620 -0.0620 -0.0632 -0.0647 -0.0635 -0.0668 -0.0696 -0.0662 -0.0659 -0.0668 -0.0650 -0.0665 -0.0668 -0.0620 -0.0586
 -0.0644 -0.0629 -0.0650 -0.0684 -0.0629 -0.0616 -0.0613 -0.0647 -0.0620 -0.0647 -0.0644 -0.0650 -0.0751 -0.0635 -0.0659 -0.0732 -0.0595 -0.0644 -0.0650 -0.0629 -0.0623
 -0.0653 -0.0647 -0.0662 -0.0668 -0.0620 -0.0595 -0.0626 -0.0641 -0.0650 -0.0632 -0.0598 -0.0729 -0.0620 -0.0623 -0.0626 -0.0620 -0.0613 -0.0632 -0.0632 -0.0620 -0.0650
 -0.0681 -0.0659 -0.0684 -0.0687 -0.0629 -0.0632 -0.0653 -0.0620 -0.0656 -0.0613 -0.0616 -0.0702 -0.0714 -0.0610 -0.0629 -0.0638 -0.0638 -0.0626 -0.0668 -0.0638 -0.0635
 -0.0699 -0.0638 -0.0677 -0.0644 -0.0641 -0.0644 -0.0650 -0.0629 -0.0607 -0.0595 -0.0616 -0.0708 -0.0674 -0.0607 -0.0644 -0.0613 -0.0641 -0.0641 -0.0671 -0.0629 -0.0629
 -0.0687 -0.0613 -0.0638 -0.0659 -0.0635 -0.0632 -0.0626 -0.0610 -0.0613 -0.0620 -0.0650 -0.0677 -0.0681 -0.0616 -0.0653 -0.0641 -0.0635 -0.0647 -0.0620 -0.0607 -0.0616
 -0.0677 -0.0601 -0.0659 -0.0650 -0.0626 -0.0610 -0.0635 -0.0604 -0.0616 -0.0647 -0.0647 -0.0653 -0.0620 -0.0635 -0.0644 -0.0656 -0.0613 -0.0629 -0.0632 -0.0635 -0.0632
 -0.0659 -0.0623 -0.0668 -0.0653 -0.0623 -0.0610 -0.0626 -0.0623 -0.0702 -0.0623 -0.0620 -0.0702 -0.0723 -0.0616 -0.0635 -0.0638 -0.0610 -0.0604 -0.0659 -0.0653 -0.0629
 -0.0644 -0.0650 -0.0699 -0.0674 -0.0632 -0.0638 -0.0626 -0.0647 -0.0638 -0.0571 -0.0601 -0.0653 -0.0720 -0.0616 -0.0635 -0.0604 -0.0629 -0.0638 -0.0653 -0.0668 -0.0647
 -0.0662 -0.0632 -0.0665 -0.0671 -0.0641 -0.0635 -0.0620 -0.0647 -0.0626 -0.0577 -0.0562 -0.0641 -0.0662 -0.0635 -0.0647 -0.0650 -0.0616 -0.0644 -0.0671 -0.0595 -0.0607
 -0.0665 -0.0610 -0.0659 -0.0656 -0.0613 -0.0610 -0.0610 -0.0616 -0.0610 -0.0604 -0.0574 -0.0665 -0.0665 -0.0644 -0.0635 -0.0647 -0.0644 -0.0623 -0.0647 -0.0595 -0.0595
 -0.0656 -0.0623 -0.0653 -0.0684 -0.0613 -0.0592 -0.0629 -0.0635 -0.0629 -0.0620 -0.0589 -0.0659 -0.0714 -0.0595 -0.0613 -0.0629 -0.0635 -0.0626 -0.0656 -0.0613 -0.0620
 -0.0662 -0.0653 -0.0671 -0.0696 -0.0598 -0.0610 -0.0644 -0.0650 -0.0635 -0.0647 -0.0604 -0.0656 -0.0638 -0.0595 -0.0638 -0.0629 -0.0613 -0.0613 -0.0665 -0.0647 -0.0641
 -0.0696 -0.0629 -0.0644 -0.0705 -0.0629 -0.0647 -0.0635 -0.0656 -0.0626 -0.0616 -0.0620 -0.0656 -0.0601 -0.0629 -0.0604 -0.0635 -0.0613 -0.0598 -0.0644 -0.0629 -0.0635
 -0.0699 -0.0626 -0.0632 -0.0677 -0.0629 -0.0641 -0.0620 -0.0595 -0.0623 -0.0626 -0.0616 -0.0659 -0.0626 -0.0629 -0.0626 -0.0647 -0.0610 -0.0601 -0.0650 -0.0650 -0.0620
Attachment 3: ER-580-230-30kV-M133.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
     80.0000
  29978.0220
21
21
./emit.dat
 -0.0662 -0.0671 -0.0674 -0.0693 -0.0681 -0.0693 -0.0674 -0.0693 -0.0717 -0.0690 -0.0684 -0.0662 -0.0653 -0.0616 -0.0665 -0.0674 -0.0665 -0.0662 -0.0763 -0.0702 -0.0742
 -0.0665 -0.0726 -0.0665 -0.0696 -0.0677 -0.0684 -0.0656 -0.0693 -0.0681 -0.0626 -0.0677 -0.0644 -0.0653 -0.0589 -0.0671 -0.0671 -0.0635 -0.0626 -0.0739 -0.0656 -0.0650
 -0.0662 -0.0732 -0.0699 -0.0702 -0.0699 -0.0705 -0.0711 -0.0677 -0.0623 -0.0635 -0.0635 -0.0644 -0.0647 -0.0653 -0.0659 -0.0644 -0.0610 -0.0592 -0.0604 -0.0610 -0.0583
 -0.0641 -0.0671 -0.0650 -0.0702 -0.0726 -0.0601 -0.0693 -0.0662 -0.0644 -0.0610 -0.0601 -0.0659 -0.0620 -0.0674 -0.0632 -0.0613 -0.0607 -0.0623 -0.0650 -0.0604 -0.0662
 -0.0671 -0.0632 -0.0671 -0.0650 -0.0665 -0.0626 -0.0684 -0.0638 -0.0629 -0.0616 -0.0623 -0.0668 -0.0613 -0.0598 -0.0543 -0.0610 -0.0638 -0.0641 -0.0650 -0.0620 -0.0644
 -0.0684 -0.0656 -0.0693 -0.0668 -0.0684 -0.0726 -0.0702 -0.0635 -0.0659 -0.0671 -0.0671 -0.0647 -0.0635 -0.0574 -0.0604 -0.0644 -0.0644 -0.0632 -0.0684 -0.0635 -0.0650
 -0.0674 -0.0726 -0.0677 -0.0656 -0.0668 -0.0690 -0.0699 -0.0159 -0.0644 -0.0656 -0.0641 -0.0705 -0.0607 -0.0671 -0.0641 -0.0641 -0.0638 -0.0626 -0.0632 -0.0635 -0.0620
 -0.0668 -0.0699 -0.0684 -0.0671 -0.0665 -0.0659 -0.0699 -0.0641 -0.0610 -0.0607 -0.0577 -0.0665 -0.0632 -0.0638 -0.0638 -0.0638 -0.0629 -0.0604 -0.0626 -0.0623 -0.0629
 -0.0690 -0.0668 -0.0684 -0.0684 -0.0665 -0.0668 -0.0659 -0.0613 -0.0629 -0.0613 -0.0607 -0.0711 -0.0644 -0.0610 -0.0626 -0.0620 -0.0623 -0.0595 -0.0601 -0.0623 -0.0674
 -0.0674 -0.0668 -0.0659 -0.0684 -0.0696 -0.0656 -0.0677 -0.0650 -0.0644 -0.0598 -0.0635 -0.0729 -0.0616 -0.0610 -0.0620 -0.0623 -0.0607 -0.0638 -0.0626 -0.0620 -0.0626
 -0.0635 -0.0647 -0.0665 -0.0668 -0.0681 -0.0665 -0.0693 -0.0644 -0.0653 -0.0626 -0.0632 -0.0708 -0.0644 -0.0623 -0.0650 -0.0623 -0.0604 -0.0644 -0.0647 -0.0632 -0.0620
 -0.0647 -0.0659 -0.0671 -0.0696 -0.0671 -0.0690 -0.0699 -0.0623 -0.0607 -0.0616 -0.0641 -0.0693 -0.0626 -0.0647 -0.0656 -0.0635 -0.0653 -0.0650 -0.0656 -0.0620 -0.0626
 -0.0671 -0.0662 -0.0677 -0.0665 -0.0674 -0.0653 -0.0711 -0.0613 -0.0516 -0.0592 -0.0647 -0.0656 -0.0632 -0.0644 -0.0638 -0.0616 -0.0650 -0.0629 -0.0656 -0.0601 -0.0616
 -0.0656 -0.0647 -0.0656 -0.0671 -0.0671 -0.0632 -0.0668 -0.0626 -0.0568 -0.0519 -0.0586 -0.0665 -0.0604 -0.0601 -0.0592 -0.0620 -0.0598 -0.0604 -0.0592 -0.0610 -0.0620
 -0.0641 -0.0671 -0.0671 -0.0656 -0.0665 -0.0681 -0.0684 -0.0638 -0.0623 -0.0500 -0.0473 -0.0708 -0.0607 -0.0595 -0.0607 -0.0613 -0.0616 -0.0610 -0.0641 -0.0632 -0.0644
 -0.0641 -0.0684 -0.0674 -0.0671 -0.0684 -0.0702 -0.0696 -0.0650 -0.0623 -0.0552 -0.0436 -0.0647 -0.0644 -0.0595 -0.0632 -0.0641 -0.0653 -0.0616 -0.0635 -0.0653 -0.0632
 -0.0671 -0.0690 -0.0684 -0.0674 -0.0699 -0.0677 -0.0699 -0.0644 -0.0647 -0.0610 -0.0430 -0.0467 -0.0629 -0.0623 -0.0671 -0.0638 -0.0620 -0.0635 -0.0632 -0.0635 -0.0610
 -0.0696 -0.0702 -0.0681 -0.0696 -0.0647 -0.0656 -0.0659 -0.0623 -0.0635 -0.0592 -0.0446 -0.0461 -0.0522 -0.0620 -0.0650 -0.0641 -0.0586 -0.0610 -0.0647 -0.0616 -0.0616
 -0.0668 -0.0644 -0.0650 -0.0677 -0.0668 -0.0632 -0.0656 -0.0616 -0.0620 -0.0543 -0.0531 -0.0491 -0.0418 -0.0586 -0.0613 -0.0626 -0.0589 -0.0623 -0.0674 -0.0592 -0.0613
 -0.0662 -0.0623 -0.0653 -0.0656 -0.0656 -0.0650 -0.0699 -0.0641 -0.0616 -0.0607 -0.0604 -0.0488 -0.0464 -0.0528 -0.0616 -0.0595 -0.0632 -0.0626 -0.0671 -0.0613 -0.0598
 -0.0644 -0.0671 -0.0674 -0.0696 -0.0668 -0.0671 -0.0696 -0.0626 -0.0607 -0.0644 -0.0638 -0.0696 -0.0626 -0.0583 -0.0623 -0.0610 -0.0635 -0.0626 -0.0659 -0.0613 -0.0635
Attachment 4: OMEGA_OM-DAQPRO-5300_06-Apr-17_13.58.21.dlp
Attachment 5: Ta#51.docx
Attachment 6: TS-2017-03-24-16-07.rga
Attachment 7: TS-2017-03-28-11-15.rga
Attachment 8: TS-2017-03-29-13-26.rga
Attachment 9: ITW-TM2-Ta#51-LP-SIS_Test_Stand.docx
  502   Monday, April 04, 2016, 11:33 Aaron Schmidt LP-SIS  41790-239956-8In progressTesting restraint insulator prototype

Monday, April 04, 2016, 11:31: Begin roughing. (No plasma electrode)

Monday, April 04, 2016, 12:47: Started Turbos and Ion gauge


Tuesday, April 05, 2016, 09:09: Open water, Vacuum increased from 3.7E-07 to 2.2E-06 (roughly). Leak detected, Confirmed via RGA, Water increased from 7.0E-7T to 5.8E-6T.

Tuesday, April 05, 2016, 09:17: Turbo Pumps stopped. Ion gauge shut off. RGA off.

Tuesday, April 05, 2016, 10:28: Venting for inspection.

Tuesday, April 05, 2016, 11:31: Inspect, re-tighten with new seals.. Begin roughing.

Tuesday, April 05, 2016, 11:54: Turbo pumps started.

Tuesday, April 05, 2016, 11:59: Ion gauge turned on.

Tuesday, April 05, 2016, 15:45: Opened water for heat shield. Vacuum stable @ 6.4E-7T - 6.8E-7T.

 Tuesday, April 05, 2016, 16:07: Vacuum stable @6.4E-7T.


Wednesday, April 06, 2016, 09:09: Started auto ramping up to 500A and 200A at 2A/min BIAS/EZL/EE to 30/13/1.5 KV

Wednesday, April 06, 2016, 09:22: Opened IV4. Switched on optics.

Wednesday, April 06, 2016, 14:29: Continue auto ramp up to 580A and 230A at 2A/min.


Thursday, April 07, 2016, 13:30: Most of the salts have evaporated FC0: at 3E-10A

Thursday, April 07, 2016, 14:10: Ramp down 2A/min.


Friday, April 08, 2016, 08:57: Ramp up 2A/min to 100A and 50A. ANODE to 200V.

Friday, April 08, 2016, 13:50: Ramp up 2A/min to 500A and 200A.


Monday, April 11, 2016, 08:29: Ramp up 2A/min to 580A and 230A. Vacuum stable @ 2.9E-7T

Monday, April 11, 2016, 09:44: ANODE current increased from 0 to 16 mA during the ramp up. -AM

Monday, April 11, 2016, 11:36: Increased TGHT voltage limit to 7.35 V. ANODE current now at 18 mA -AM

Monday, April 11, 2016, 16:00: BIAS / EZL / EE to 30 / 13 / 1.5 kV -AM


Tuesday, April 12, 2016, 08:47: Ramp Down 2A/min 230A to 200A TBHT only.

Tuesday, April 12, 2016, 09:31: Ramp Down 2A/min 580A to 500A TGHT only.

Tuesday, April 12, 2016, 13:55: Ramp up 2A/min 500A to 580A TGHT only.

 Tuesday, April 12, 2016, 15:43: Ramp up 2A/min 580A to 590A, Voltage limit increased to 7.50 TGHT only.

Tuesday, April 12, 2016, 15:59: Ramp down 2A/min and 1A/min.


Wednesday, April 13, 2016, 08:04 : RGA off.

Wednesday, April 13, 2016, 08:20: Source off. IV4 closed. Turbo pumps off.

Wednesday, April 13, 2016, 09:38: Venting.

Wednesday, April 13, 2016, 10:59: Venting paused.

Wednesday, April 13, 2016, 11:03: Venting Continued.

Wednesday, April 13, 2016, 11:24: Venting complete. RP#1 removed. (No plasma electrode).

Attachment 1: Anode_Voltage_VS_Current.xlsx
Attachment 2: TS-2016-04-05_Restraint_Test.rga
Attachment 3: TS-2016-04-13_Restraint_Test.rga
Attachment 4: RP#1.docx
ELOG V2.9.2-2455