ID |
Date |
Author |
Target/No.# |
Source |
Station |
Module |
Target Oven W.O# |
Heat Shield W.O# |
Status |
Subject |
725
|
Tuesday, May 28, 2019, 15:05 |
Aaron Schmidt | Test FEBIAD / TM4 | LP-FEBIAD | | TM4 | 395530-3 | 42752-5 | In progress | TM4 Source Tray - Episode II | SEE ELOG 724
Tuesday, May 28, 2019, 15:02: Roughing
Tuesday, May 28, 2019, 15:32: Turbos on.
Wednesday, May 29, 2019, 07:08: RGA on.
Wednesday, May 29, 2019, 07:10: Waterlines opened on FS#5. No response on RGA. Vacuum ok @ 1.32e-06T.
Power on.
Wednesday, May 29, 2019, 07:13: Mini-pump & manual valve opened. Prep / rough gas lines.
Alternate SV1, SV3, & SV4.
Wednesday, May 29, 2019, 07:27: Shut off mini-pump, manual valve, & SV1. Open / close SV5, SV3, & SV4.
Wednesday, May 29, 2019, 07:32: Gas lines fully opened.
Wednesday, May 29, 2019, 07:37: Sources on.
Wednesday, May 29, 2019, 07:39: Anode to 50 V. No response on EE. EE to -100 V. Current response normal @ 34.188 uA.
Wednesday, May 29, 2019, 07:52: BIAS / EE / EZL @ 30 / -1.5 / 13 kV. Anode to 100 V. Coil to 10 A. Small sparks heard around 23 kV.
Wednesday, May 29, 2019, 07:54: IV4 opened.
Wednesday, May 29, 2019, 07:57: Auto ramping TGHT / TBHT to 350 / 150 A @ 2A per min. Vac limit increased to 3.000e-06T for out gassing.
Wednesday, May 29, 2019, 07:57: Spark* Computer screens flickered. RGA knocked out.
Wednesday, May 29, 2019, 08:00: Shut off & unplug the RGA.
Wednesday, May 29, 2019, 08:04: Optics set.
Wednesday, May 29, 2019, 08:23: Small spark* Nothing affected.
Wednesday, May 29, 2019, 08:35: Small spark* TSG:CBC restarted.
Wednesday, May 29, 2019, 10:26: Small spark* TSG:CBC boosted from 500 V to 2501 V. Reset to 500 V. TPS:Q1 boosted from 790 / 675 V to 1500 / 1230 V. Reset.
Wednesday, May 29, 2019, 12:04: Auto ramping TGHT / TBHT to 580 / 270 A @ 2A per min. Over 100 minutes spark free.
Wednesday, May 29, 2019, 12:14: Spark* Reset optics & TSG:CBC
Wednesday, May 29, 2019, 12:35: Note*: Gradual climb on EZL current noticed as ramping current climbs.
Wednesday, May 29, 2019, 13:01: Spark* Reset optics & TSG:CBC
Wednesday, May 29, 2019, 13:50: Spark* Reset optics & TSG:CBC. TSG:CBC was boosted from 500 V to 2501 V.
Wednesday, May 29, 2019, 13:54: A small but notable out gassing event, without an audible spark occurred. Brief instability in BIAS voltage & current. Vacuum spike to over 3.0e-06T.
Wednesday, May 29, 2019, 14:01: Auto ramping TBHT to 290 A @ 2A per min. TGHT at 577 (580.4 A)
Wednesday, May 29, 2019, 14:15: Wire Scans taken. TMS:RPM0A (X Axis) revealed no beam / peak. TMS:RPM0B (Y Axis) revealed a beam peak. Will attempt to steer TMS:RPM0A (X Axis) until beam peak is revealed.
Wednesday, May 29, 2019, 14:49: Further tuning required. Optics (TPS) Q1, Q2, Q3, Q4 all adjusted. FEBIAD source side (TSG) CB1, VCB1, & CBC all adjusted. Small peak found on RPM0A (X Axis). TMS:RPM0B (Y Axis) revealed a beam peak
Mass scans have not been successful. Yet.
Wednesday, May 29, 2019, 15:00: BIAS / EE / EZL brought to 0 kV. Allow out gassing of the source tray & protect equipment overnight.
Thursday, May 30, 2019, 07:43: BIAS / EE / EZL @ 30 / -1.5 / 13 kV
Thursday, May 30, 2019, 07:55: Spark* Reset optics & TSG:CBC.
Thursday, May 30, 2019, 08:37: Spark* Tripped TMS:TP6 turbo pump. Closed IV4. Tripped & Boosted TPS:Q1. Reset optics & TSG:CBC. Open IV4.
Thursday, May 30, 2019, 09:00: Spark* Visible inside TS on the back of the optics tray off one of the EZL wires. EZL current spiked 1440 uA. Voltage dropped and became unresponsive over 8.5 kV.
Thursday, May 30, 2019, 09:13: BIAS / EE / EZL brought to 0 kV. Anode to 0 V. Coil to 0 A.
EZL to 7000V - Current spikes to over 500 uA.
Thursday, May 30, 2019, 09:32: Auto-cooling @ 2 A per min.
Thursday, May 30, 2019, 09:41: Auto-cooling @ 3 A per min.
Thursday, May 30, 2019, 09:49: IV4 closed. Gas lines closed. All sources expect TGHT / TBHT off.
Thursday, May 30, 2019, 12:24: Halt auto-cooling. Sources on.
Thursday, May 30, 2019, 12:32: BIAS @ 20 kV.
Thursday, May 30, 2019, 12:34: BIAS @ 30 kV.
Thursday, May 30, 2019, 12:38: BIAS / EE @ 30 / -1.5 kV
Thursday, May 30, 2019, 12:42: BIAS / EE @ 39.5 / -2 kV. Two sparks* Shut off turbos
Thursday, May 30, 2019, 12:45: Turbo pumps TP6 & TP1B restarted. Optics reset.
Thursday, May 30, 2019, 12:51: BIAS / EE / EZL @ 40 / -2 / 5.25 kV. Spark* Shut off turbos / reset.
Thursday, May 30, 2019, 12:56: BIAS / EE / EZL @ 41.5 / -2.5 / 7.5 kV. Spark* Shut off turbos / Optics. Reset.
Thursday, May 30, 2019, 13:01: BIAS / EE / EZL @ 45 / -2.5 / 7.5 kV. Visible spark inside the vacuum* Shut off turbos / Optics. Reset.
Thursday, May 30, 2019, 13:26: BIAS / EE / EZL @ 46.25 / -2.5 / 7.5 kV. Visible spark inside the vacuum, off multi-pin corner? * Shut off turbos / Optics. Reset. FC0 affected. Reset.
Thursday, May 30, 2019, 13:38: Go pro setup
Thursday, May 30, 2019, 13:41: BIAS / EE / EZL @ 48 / -2.5 / 7.5 kV. Visible spark inside the vacuum, Did the go pro catch it? * Shut off turbos / Optics. Reset. FC0 affected. Reset.
*Spark seen by Carla on the back of multi-pin (Anode wire perhaps?).
Thursday, May 30, 2019, 13:52: BIAS / EE / EZL @ 48.9 / -2.5 / 7.5 kV. Visible spark inside the vacuum, Did the go pro catch it? * Shut off turbos / Optics. Reset. FC0 affected. Reset.
*Spark seen on the base of the EZL.
Thursday, May 30, 2019, 14:00: BIAS / EE / EZL @ 50 / -2.5 / 7.5 kV.
Thursday, May 30, 2019, 14:07: Spark* No trips
Thursday, May 30, 2019, 14:15: BIAS / EE / EZL @ 50 / -2.5 / 7.5 kV. * Shut off turbos / Optics. Reset. FC0 affected. Reset.
Thursday, May 30, 2019, 14:24 BIAS / EE / EZL @ 51.3 / -2.5 / 7.5 kV. * Shut off TP6 / Optics. Reset. RPM6B affected, ran up & down on its own. Reset.
Thursday, May 30, 2019, 14:25: BIAS back to 50 kV.
Thursday, May 30, 2019, 14:31: BIAS / EE / EZL @ 51.9 / -2.5 / 7.5 kV. Visible spark inside the vacuum, from source tray to the rear ground electrode plate * Shut off turbos / Optics. Reset. FC0 affected. Reset.
IG6 ion gauge & TMS TP6 tripped. Fix. Switched to filament 2.
Thursday, May 30, 2019, 14:43: BIAS / EE / EZL @ 52.8 / -2.5 / 7.5 kV. Visible spark inside the vacuum, Bottom corner of multi-pin bracket * Shut off turbos / Optics. Reset. FC0 affected. Reset. RPM6B Reset.
Thursday, May 30, 2019, 14:54: BIAS / EE / EZL @ 53.5 / -2.5 / 7.5 kV. Sparks*
Thursday, May 30, 2019, 14:59: BIAS / EE / EZL @ 54 / -2.5 / 7.5 kV. Visible spark inside the vacuum, Source tray EE over insulators to ground (Matt got a video, really nice)* Shut off turbos / Optics. Reset. FC0 affected. Reset. RPM6B Reset.
Thursday, May 30, 2019, 15:06: BIAS / EE / EZL @ 54.7 / -2.5 / 7.5 kV. Spark. Reset a bunch.
Thursday, May 30, 2019, 15:15: BIAS @ 55 kV.
Thursday, May 30, 2019, 15:24: BIAS / EE / EZL @ 0 kV.
Thursday, May 30, 2019, 15:24: Auto ramping TGHT / TBHT to 580 / 290 A @ 2A per min.
Friday, May 31, 2019, 07:22: BIAS @ 10 kV. IV4 opened.
Friday, May 31, 2019, 07:33: BIAS / EE / EZL @ 30 / -1.5 / 7.5 kV. Anode @ 200 V. Coil @ 20 A. FC0? No response.
Friday, May 31, 2019, 07:48: Coil @ 50 A. FC0 BIAS off. Troubleshoot FC0.
Friday, May 31, 2019, 08:01: Nothing on wire scans. FC0 randomly gaining, but no BIAS response. Rod N contacted on FC0.
Friday, May 31, 2019, 09:56: FC0 Fixed by Rod. N. MB1 acting up. Fixed itself through power / Canbus cycling?. Wire scans: RPM0A (X) non-responsive, RPM0B (Y) responds. Turn down HV to prep for gases.
Friday, May 31, 2019, 10:14: BIAS / EE / EZL / Anode @ 0 kV.
Friday, May 31, 2019, 10:18: Mini pump on. Manual valve opened. Prep / rough gas lines.
Alternate SV1, SV3, & SV4.
Friday, May 31, 2019, 10:31: Shut off mini-pump, manual valve, & SV1. Open / close SV5, SV3, & SV4.
Friday, May 31, 2019, 10:36: Gas lines fully opened.
Friday, May 31, 2019, 10:48: Vacuum stable @ 1.25e-06T. HV sources back on.
Friday, May 31, 2019, 11:04: BIAS / EE / EZL @ 30 / -1.5 / 8.25 kV. Anode @ 200 V. Coil @ 50 A.
Friday, May 31, 2019, 11:32: Spark knocked out Optics & Steering. Wire scanners to zero. Unable to pick up wire scan for X or Y.
Friday, May 31, 2019, 11:35: BIAS / EE / EZL @ 30 / -2.0 / 7.5 kV.
Friday, May 31, 2019, 12:05: Tuning adjusted. RPM0A (X) no response. RPM0B (Y) Response (It's back!!!).
Friday, May 31, 2019, 12:13: But it sparked...* Reset Optics. Re-scan to recover. Wire scanners to zero. Unable to pick up wire scan for X.
Carla, Fernando, Matt K.
Friday, May 31, 2019, 12:33 Using Brad's tuning values we see a nice peak at the first set of wire scanners on the y axis but nothing on the x axis. Using CB2, went from 200V to 900V in steps of 100V and saw nothing on the x axis. This seems to indicate x axis scanner is broken. Proceeding without x axis readout. Also tried with CB1 steerers, same effect.
Friday, May 31, 2019, 13:30: Auto-ramping TBHT to 305 A @ 2 A per min.
Friday, May 31, 2019, 13:31: EZL @ 8 kV.
Friday, May 31, 2019, 13:36: Coil to 0 A.
Friday, May 31, 2019, 15:32: Successful wire scans across RPM0A & RPM0B (X,Y) and RPM6B & RPM6A (Y,X).
Friday, May 31, 2019, 16:27 : Wire scanners after the magnet look ok, but not enough beam to do a good emittance scan. Raised the tube heater to 310 A set point, got about 0.5nA at FC6. Probably could be solved with tuning, to be done Monday. Turning everything off but leaving target heating on. See pasted screenshots for current tune on mass 28.
Monday, June 03, 2019, 07:35: BIAS / EE / EZL @ 30 / -1.5 / 7 kV. Anode @ 200 V. Coil @ 10 A.
Monday, June 03, 2019, 07:38: Coil @ 30 A.
Monday, June 03, 2019, 07:41: IV4 opened.
Monday, June 03, 2019, 07:48: BIAS / EE / EZL @ 0 kV. A named Hossein Rafighi with IT began poking around the HV rack and running cable w/ permission reference to Dale Nagra. Locked it out. No work permit was issued. Main Control Room Contacted.
Monday, June 03, 2019, 08:03: Cable run complete. Lock off. Sources on again.
Monday, June 03, 2019, 08:23: BIAS / EE / EZL @ 30 / -1.5 / 7 kV. Anode @ 200 V. Coil @ 30 A.
Monday, June 03, 2019, 08:35: BIAS / EE / EZL @ 40 / -2 / 8 kV
Monday, June 03, 2019, 08:40: Spark* Optics tripped. Turbo pump TP6 tripped. Close / re-open IV4. Reset.
Monday, June 03, 2019, 08:41: BIAS @ 45 kV.
Monday, June 03, 2019, 08:45: Spark* Optics tripped. Turbo pump TP6 tripped. Close / re-open IV4. Reset. Mass Magnet MB1 left off.
Monday, June 03, 2019, 08:55: Spark* Optics tripped. Turbo pump TP6 tripped. Close / re-open IV4. Reset.
Monday, June 03, 2019, 08:58: BIAS / EE / EZL @ 50 / -2 / 8 kV.
Monday, June 03, 2019, 09:07 : Spark* Optics tripped. Turbo pump TP6 tripped. Close / re-open IV4. Reset.
Monday, June 03, 2019, 09:21: 2 Sparks* Optics tripped. Turbo pump TP6 tripped. Close / re-open IV4. Reset.
Monday, June 03, 2019, 09:29: Spark* Optics tripped. Turbo pump TP6 tripped. Close / re-open IV4. Reset. BIAS @ 54 kV.
Monday, June 03, 2019, 09:31: BIAS / EE / EZL @ 55 / -2.5 / 8 kV.
Monday, June 03, 2019, 09:35: Spark* Optics tripped. Turbo pump TP6 tripped. Close / re-open IV4. Reset
Monday, June 03, 2019, 09:44: Spark* Optics tripped. Turbo pumps TP6 & TP1A Tripped. Close / re-open IV4. Reset
Monday, June 03, 2019, 09:49: Near BIAS 57.5 kV. Multiple sparks. BIAS unable to hold. Optics tripped, wire scanners & FC0 began running independently. Turbo pumps tripped.
Monday, June 03, 2019, 09:51: BIAS down to 50 kV. Burnt plastic smell lingering. Wait*
Monday, June 03, 2019, 09:57: Spark* Reset Optics. Turn off Optics for now.
Monday, June 03, 2019, 10:19: Spark* Reset Optics & Wire scanners. Anode current dropped from 10.989 mA to 7.768 mA
Monday, June 03, 2019, 10:36: BIAS Near 55 kV, brought back down to 50 kV. Multiple sparks. Optics tripped, wire scanners & FC0 began running independently. Anode current dropped from 7.328 mA to 6.105 mA. Knocked out the TGHT
The target heater dropped from 580A to 3.1 A. Attempted reset. Non-responsive.
Monday, June 03, 2019, 10:46: TGHT Still non-responsive. Bump down HV to open the cage & cycle the PS.
Monday, June 03, 2019, 11:02: BIAS / EE / EZL / Anode @ 0 kV.
Monday, June 03, 2019, 11:36: TGHT PS determined to have failed. Run w/ TBHT and test up to 50 kV.
Monday, June 03, 2019, 11:41: Sources on.
Monday, June 03, 2019, 13:17: BIAS / EE / EZL @ 30 / -1.5 / 7.5 kV. Anode @ 200 V.
Monday, June 03, 2019, 13:24: BIAS / EE / EZL @ 40 / -2.0 / 7.5 kV.
Monday, June 03, 2019, 13:37: BIAS / EE / EZL @ 50 / -2.5 / 7.5 kV.
Monday, June 03, 2019, 13:49: Odd sound from Turbo pump TP6?
Vacuum @ 1.13e-06T. Was in the 5e-07 range earlier. Strip tool revealing gradual increase.
Monday, June 03, 2019, 13:52: Spark* Optics tripped, wire scanners & FC0 running independently. Reset
Monday, June 03, 2019, 14:13: Spark* Optics tripped. Loud, outside vacuum chamber.
Monday, June 03, 2019, 14:55: BIAS / EE / EZL @ 0 kV. Anode to 0 V.
Tuesday, June 04, 2019, 08:40: BIAS / EE / EZL @ 30 / -1.5 / 7.5 kV. Anode @ 200 V.
Tuesday, June 04, 2019, 08:54: BIAS / EE / EZL @ 40 / -2.0 / 7.5 kV.
Tuesday, June 04, 2019, 09:09: BIAS / EE / EZL @ 45 / -2.5 / 7.5 kV.
Tuesday, June 04, 2019, 09:23: BIAS / EE / EZL @ 50 / -2.5 / 7.5 kV.
Tuesday, June 04, 2019, 10:21: Spark* Optics & Turbo pump TP6 tripped. Restarted. Quiet spark* Inside Vacuum chamber.
Tuesday, June 04, 2019, 10:23: BIAS / EE / EZL @ 50 / -2.5 / 7.5 kV. Almost 1 hour of no sparks.
Tuesday, June 04, 2019, 10:29: Note* Minute fluctuations of Anode current have ended after the spark. Also slight 'Burnt' smell noticed around the test stand area.
Tuesday, June 04, 2019, 10:32: BIAS / EE / EZL @ 51 / -2.5 / 7.5 kV.
Tuesday, June 04, 2019, 10:39: BIAS / EE / EZL @ 52.5 / -2.7 / 7.5 kV.
Tuesday, June 04, 2019, 11:02: BIAS / EE / EZL @ 53.5 / -2.8 / 7.5 kV.
Tuesday, June 04, 2019, 11:07: Spark* Outside TS. IG6 tripped / reset. BIAS current spike. EZL current dropped afterwards near 460 uA from 550 uA.
Tuesday, June 04, 2019, 11:22: BIAS / EE / EZL @ 54.5 / -2.9 / 7.5 kV.
Tuesday, June 04, 2019, 11:27: BIAS / EE / EZL @ 55 / -3.0 / 7.5 kV.
Tuesday, June 04, 2019, 11:35: Spark* Optics & Turbo pump - TP6 tripped. Restarted. Quiet spark* Inside Vacuum chamber.
Tuesday, June 04, 2019, 11:39: Spark* Optics tripped. Loud*, outside the vacuum chamber, above TS. EZL current jumped back up to 550 uA
BIAS current gradually increasing. from 50 uA to 55 uA since 11:35 spark.
Tuesday, June 04, 2019, 12:01: Runaway sparks* Turbo pumps & ion gauges tripped. Optics tripped. Wire scanners running independently. Both quiet & loud sparks. 3 Consistent marks in the strip tool. Reset.
Brought BIAS down to 55 kV. EE to -2.5 kV. BIAS & EZL currents have a small drop.
Tuesday, June 04, 2019, 13:21: BIAS / EE / EZL @ 50 / -2.5 / 7.5 kV. Over 1 hr no sparks.
Tuesday, June 04, 2019, 13:34: BIAS / EE / EZL @ 52.5 / -2.7 / 7.5 kV.
Tuesday, June 04, 2019, 13:40: BIAS / EE / EZL @ 53.5 / -2.8 / 7.5 kV.
Tuesday, June 04, 2019, 13:41: Spark* Optics & Turbo pump - TP6 tripped. Wire scanners and FC0 operated independently. Restarted. Quiet spark* Inside Vacuum chamber.
Tuesday, June 04, 2019, 13:46: Spark* Loud*, outside the vacuum chamber, above TS. EZL current jumped from 580 uA to 610 uA approx & is gradually increasing. BIAS current slowly decreasing.
Tuesday, June 04, 2019, 14:02: BIAS / EE / EZL @ 54.5 / -2.9 / 7.5 kV. Variation in EZL & BIAS currents noted.
Tuesday, June 04, 2019, 14:08: BIAS / EE / EZL @ 55 / -3.0 / 7.5 kV.
Tuesday, June 04, 2019, 14:28: 20 minutes. No Spark.
Tuesday, June 04, 2019, 14:47: BIAS / EE / EZL @ 0 kV. Anode to 0 V.
Wednesday, June 05, 2019, 08:34: BIAS / EE / EZL @ 30 / -1.5 / 7.5 kV. Anode @ 200 V.
Wednesday, June 05, 2019, 08:35: Coil down from 30 to 0 A.
Wednesday, June 05, 2019, 08:36: Open IV4.
Wednesday, June 05, 2019, 08:37: Optics on / reset.
Wednesday, June 05, 2019, 08:38: Anode to 400 V.
Wednesday, June 05, 2019, 08:44: Anode to 450 V. TBHT @ 305 A.
Wednesday, June 05, 2019, 11:15: BIAS / EE / EZL @ 40 / -2.0 / 7.5 kV. Anode @ 200 V.
Wednesday, June 05, 2019, 11:30: BIAS / EE / EZL @ 50 / -2.5 / 7.5 kV.
Wednesday, June 05, 2019, 11:40: BIAS / EE / EZL @ 52.5 / -2.7 / 7.5 kV.
Wednesday, June 05, 2019, 11:53: BIAS / EE / EZL @ 55 / -2.8 / 7.5 kV.
Wednesday, June 05, 2019, 12:04: Spark* Optics & Turbo pump - TP6 tripped. Restarted. Quiet spark* Inside Vacuum chamber. bias to 54.9 kV.
Wednesday, June 05, 2019, 12:04: IV4 closed.
Wednesday, June 05, 2019, 12:15: BIAS / EE / EZL @ 55 / -2.8 / 7.5 kV.
Wednesday, June 05, 2019, 12:16: Spark* Loud*, outside the vacuum chamber, above TS. EZL current came down from 601 uA to 537uA . IG6 off. Restarted. Bias to 54.9 kV.
Wednesday, June 05, 2019, 12:24: BIAS / EE / EZL @ 55 / -2.8 / 7.5 kV.
Wednesday, June 05, 2019, 12:35: Spark* inside vacuum chamber. Optics & turbo pumps tripped
Wednesday, June 05, 2019, 13:35: Note* 1 hour @ 55 kV no sparks.
Wednesday, June 05, 2019, 13:42: BIAS / EE / EZL @ 56 / -2.8 / 7.5 kV.
Wednesday, June 05, 2019, 13:49: BIAS / EE / EZL @ 57 / -2.8 / 7.5 kV.
Wednesday, June 05, 2019, 13:52: TSG:CB1 went red. Communication lost it.
Wednesday, June 05, 2019, 13:54: BIAS / EE / EZL @ 57.5 / -2.8 / 7.5 kV.
Wednesday, June 05, 2019, 14:01: Spark* inside vacuum chamber. Optics & turbo pumps tripped. Appears to be on the Anode. The test stand phone also redialed the last number contacted on its own.
Wednesday, June 05, 2019, 14:25: BIAS / EE / EZL @ 0 kV. Anode to 0 V. Auto-cooling @ 2 A per min.
Thursday, June 06, 2019, 09:10: Gas lines closed. Sources off. Optics off.
Thursday, June 06, 2019, 09:13: EE / BIAS on.
Thursday, June 06, 2019, 09:20: BIAS @ 30 kV.
Thursday, June 06, 2019, 09:30: BIAS @ 40 kV.
Thursday, June 06, 2019, 09:40: BIAS @ 45 kV. EZL slowly charging (Not on) from 12 to 420 V. EE with -15 V.
Thursday, June 06, 2019, 09:48: BIAS @ 50 kV. EZL / EE @ 606 / -17 V.
Thursday, June 06, 2019, 10:06: BIAS @ 52.2 back to 50 kV. Loud buzzing heard up top. Charge on EZL from 600 V to 1200 - 2000 V fluctuations. EZL & EE to be brought up before we break something.
Thursday, June 06, 2019, 10:16: BIAS / EE / EZL @ 50 / -2.5 / 7.5 kV.
Thursday, June 06, 2019, 10:27: BIAS / EE / EZL @ 55 / -2.7 / 7.5 kV.
Thursday, June 06, 2019, 10:39: BIAS / EE / EZL @ 57.5 / -2.8 / 7.5 kV.
Thursday, June 06, 2019, 11:04: BIAS / EE / EZL @ 58 / -2.9 / 7.5 kV.
Thursday, June 06, 2019, 11:06: Buzzing / crackling up top in the cage again. Unstable voltage drops. BIAS to 57 kV.
Thursday, June 06, 2019, 11:11: BIAS / EE / EZL @ 58 / -2.9 / 8.25 kV.
Thursday, June 06, 2019, 11:27: BIAS was up to 58.5 & EZL to 9 kV. Buzzing / crackling. Unstable voltage drops / Current increase. BIAS / EZL to 57.5 / 8.5 kV.
Thursday, June 06, 2019, 11:36: BIAS was up to 58.0 & EZL to 9 kV. Buzzing / crackling. More unstable voltage drops / Current increases. BIAS dropped on it's own to 56.1 kV. BIAS / EZL to 56 / 7.5 kV.
Thursday, June 06, 2019, 11:36: Spark*. Inside chamber. Turbo pumps tripped. Buzzing stopped.
Thursday, June 06, 2019, 11:51: Unable to surpass 57 kV before voltage breakdown. Buzzing comes back. Breakdown dropped to 51 kV on BIAS.
Thursday, June 06, 2019, 11:53: BIAS @ 50 kV. EZL to 0, PS off, in attempt to relieve EZL current.
Thursday, June 06, 2019, 11:56: BIAS @ 50 kV. EZL to 7.5 kV w/ 1245 uA. Typically at 550-600 uA on EZL. Unable to relieve the current.
Thursday, June 06, 2019, 12:18: BIAS / EE / EZL @ 55 / -2.7 / 7.5 kV. Current: BIAS / EZL @ 65 / 1306 uA.
Thursday, June 06, 2019, 13:14: BIAS / EE / EZL @ 56 / -2.7 / 7.25 kV. Current: BIAS / EZL @ 70 / 1450 uA.
Thursday, June 06, 2019, 13:15: Voltage breakdown. Buzzing / crackling back. Breakdown dropped to 56.2 kV on BIAS. BIAS set to 56.1 kV to stabilize.
Thursday, June 06, 2019, 13:19: EZL current limit increased.
Thursday, June 06, 2019, 13:30: BIAS / EE / EZL @ 56.5 / -2.7 / 7.00 kV. Current: BIAS / EZL @ 70 / 1856 uA.
Thursday, June 06, 2019, 13:31: Passed to BIAS 56.6 kV. Voltage breakdown. Buzzing / crackling. Breakdown dropped to 55.6 kV on BIAS. BIAS set to 55.5 kV to stabilize.
Thursday, June 06, 2019, 13:51: BIAS / EE / EZL @ 56.2 / -2.7 / 7.00 kV. Current: BIAS / EZL @ 70 / 1950 uA.
Thursday, June 06, 2019, 13:52: BIAS 56.3 kV. Voltage breakdown. Buzzing / crackling. Breakdown dropped to 55.1 kV on BIAS. BIAS set to 55.0 kV to stabilize. Unable to regain HV. Decreases each time we go up.
Thursday, June 06, 2019, 14:00: Limited via the test stand voltage breakdown up in the cage. Source tray held @ 58 kV.
Bring it down.
Thursday, June 06, 2019, 14:31: BIAS / EE / EZL @ 0 kV.
Friday, June 07, 2019, 06:41: Power / Sources off.
Friday, June 07, 2019, 06:43: IG1 & Turbo pumps off.
|
Attachment 1: TM4_Source_Tray_-_2019_-_2.docx
|
Attachment 2: TS-2019-05-29-07-08.rga
|
Attachment 3: TM4_Source_Tray_-_2019_-_3_-_HV_-_1.docx
|
Attachment 4: TM4_Source_Tray_-_2019_-_4_-_HV_-_2_-_Heat_on.docx
|
Attachment 5: TM4_Source_Tray_-_2019_-_5_-_HV_over_50_-Heat_on.docx
|
726
|
Monday, June 03, 2019, 11:35 |
Aaron Schmidt | | LP-SIS | | TM4 | 395530-3 | 42752-5 | Problems/Solutions | TGHT PS Failed. | Monday, June 03, 2019, 10:36: BIAS Near 55 kV, brought back down to 50 kV. Multiple sparks. Optics tripped, wire scanners & FC0 began running independently. Anode current dropped from 7.328 mA to 6.105 mA. Knocked out the TGHT
The target heater dropped from 580A to 3.1 A. Attempted reset. Non-responsive.
Monday, June 03, 2019, 10:46: TGHT Still non-responsive. Bump down HV to open the cage & cycle the PS.
PS model: Xantrex XPR 10-600
Message lights on the front of the power supply:
OTP - INT - OVP - SNS Triggered.
Front Panel Display, Status Annunciators
1. AC: The AC input is out of range and has
caused the output of
the supply to be
shut down (red)
2. OTP: The internal Over Temperature
Protection threshold has been exceeded,
disabling the supply output until the sensor
s cool to within the normal operating
temperature range (red)
3. INT: The external shutdo
wn line has been asserted, disabling the output of the
supply (green)
4. OVP: The output Over Voltage Prot
ection threshold have been exceeded,
latching the supply output off (red)
5. SNS: The remote sense line protection
circuit has been activated, latching the
supply output off (red)
6. ON: The supply output
is enabled (green)
7. OFF: The supply output
has been disabled (green)
8. LCL: The supply is in
Local control mode (green)
9. RMT: The supply is in Remote program mode (green)
10. CV: The supply is operating in
Constant Voltage mode (green)
11. CC: The supply is operating in
Constant Current mode (green)
Thursday, June 13, 2019, 14:05: ELOG 732 - PS replaced. |
727
|
Monday, June 03, 2019, 14:13 |
Aaron Schmidt | | | | TM4 | 395530-3 | 42752-5 | Done | TMS:TP6 Turbo pump issue. (Ok now) | Monday, June 03, 2019, 14:09:
Turbo pump TMS:TP6 was making a loud hum / buzz sound. The vacuum pressure is steadily increasing on IG6. Edi Contacted. Turbo pump likely failing.
The sound tapered off.

Monday, June 03, 2019, 14:32: Sound was the turbo pump spinning down. Edi restarted & reset the main controller. It was knocked out by one of the morning sparks.

|
731
|
Tuesday, June 11, 2019, 14:33 |
Aaron Schmidt | Test FEBIAD | LP-FEBIAD | | | 395530-3 | 42752-5 | In progress | ISAC Source tray re-installed w/ optics tray. | Tuesday, June 11, 2019, 14:33: Roughing.
Wednesday, June 12, 2019, 08:18: Venting to install FEBIAD.
Wednesday, June 12, 2019, 09:13: LP-FEBIAD installed. Roughing.
Wednesday, June 12, 2019, 11:13: Turbo pumps on.
Wednesday, June 12, 2019, 11:21: Ion gauge on.
Wednesday, June 12, 2019, 11:48: RGA re-connected & switched on.
Thursday, June 13, 2019, 10:27:
He Leak TEST:
Vacuum Group Leak detector LD#4.
Thursday, June 13, 2019, 10:28: Vacuum @ 1.60e-06T
LD Base line before opening valve to turbo pumps: 0.0e-09 atm-cc/sec
Thursday, June 13, 2019, 10:35: LD baseline after opening valve to turbo pumps: 1.2e-09 atm-cc/sec
Full open base:
Vacuum @ 1.60e-06T. RGA status:
Water: 2.20e-006
He: 4.55e-010
LD: 1.2e-09 atm-cc/sec
Thursday, June 13, 2019, 10:35:
Open line FS#2 . Pressurize to 30 PSI. Record after 5 minutes
Thursday, June 13, 2019, 10:40 Vacuum @ 1.54e-06T . RGA status:
Water: 2.20e-006
He: 4.20e-010
LD: 0.1e-09 atm-cc/sec. Purge and close FS#2.
Thursday, June 13, 2019, 10:45: Open line FS#3 . Pressurize to 30 PSI. Record after 5 minutes
Thursday, June 13, 2019, 10:50: Vacuum @ 1.52e-06T. RGA status:
Water: 2.18e-006
He: 3.68e-010
LD: 0.7e-09 atm-cc/sec. Purge and close FS#3.
Thursday, June 13, 2019, 10:50: Open line FS#4 . Pressurize to 30 PSI. Record after 5 minutes
Thursday, June 13, 2019, 10:55: Vacuum @ 1.54e-06T. RGA status:
Water: 2.16e-006
He: 9.24e-011
*LD: 3.7e-09 atm-cc/sec. Run for another 5 minutes.
Thursday, June 13, 2019, 11:00: Vacuum @ 1.52e-06T. RGA status:
Water: 2.15e-006
He: 2.06e-010
*LD: 0.7e-08 atm-cc/sec. Potential He escaping through o-rings*
Purge and close FS#4.
Thursday, June 13, 2019, 11:01: Open line FS#5 . Pressurize to 30 PSI. Record after 5 minutes
Thursday, June 13, 2019, 11:06: Vacuum @ 1.52e-06T. RGA status:
Water: 2.13e-006
He: 2.75e-010
LD: 1.7e-08 atm-cc/sec. Purge and close FS#5
Thursday, June 13, 2019, 11:07: Depressurize lines. Connect HPA & Purge all lines.
Thursday, June 13, 2019, 11:12: Vacuum @ 1.54e-06T . RGA status:
Water: 2.12e-006
He: 6.21e-010
LD: 3.3e-08 atm-cc/sec
Close Valve & disconnect LD. HE LEAK TEST COMPLETE
Thursday, June 13, 2019, 11:20: Vacuum @1.52e-06T :Begin gradually opening the waterlines.
Thursday, June 13, 2019, 11:22: Open FS#2. Vacuum @ 1.52e-06T. No Response on RGA.
Thursday, June 13, 2019, 11:23: Open FS#3. Vacuum @ 1.52e-06T. No Response on RGA.
Thursday, June 13, 2019, 11:23: Open FS#4. Vacuum @ 1.52e-06T. No Response on RGA.
Thursday, June 13, 2019, 11:24: Open FS#5. Vacuum @ 1.52e-06T. No Response on RGA.
Thursday, June 13, 2019, 11:28: Power sources: Tested. TGHT PS is still a no go. Power switched off, lock out.
Thursday, June 13, 2019, 14:03: New TGHT PS installed.
Thursday, June 13, 2019, 14:06: Power / sources on.
Thursday, June 13, 2019, 14:20: BIAS / EE / EZL @ 30 / -1.5 / 13 kV. EZL current @ 61 uA. EZL appears to be fixed. :)
Thursday, June 13, 2019, 14:21: Spark*
Thursday, June 13, 2019, 14:29: Sources to 0 kV. FS#8 needs re-connection.
Thursday, June 13, 2019, 14:43: FS#8 connected. Vacuum ok @ 1.52e-06T. No response on RGA.
Thursday, June 13, 2019, 14:45: Auto-ramping TGHT / TBHT to 300 / 150 A @ 2 A per min.
Friday, June 14, 2019, 06:52: Restart RGA.
Friday, June 14, 2019, 08:37: Run mini pump to prep gas-lines.
Friday, June 14, 2019, 08:44: Close SV1, manual valve & turn off mini pump. Gradually open gas lines.
Friday, June 14, 2019, 08:48: Gas lines fully opened.
Friday, June 14, 2019, 08:49: Auto-ramping TGHT / TBHT to 570 / 270 A @ 2 A per min.
Friday, June 14, 2019, 08:52: HV sources ready.
FM
Friday, June 14, 2019, 11:11: Anode current @ 0V 7.326mA. Setting Anode to 150 V.
AS
Friday, June 14, 2019, 11:16: Auto-ramping TGHT / TBHT to 580 / 290 A @ 2 A per min.
Friday, June 14, 2019, 11:18: Optics set.
Friday, June 14, 2019, 11:29: BIAS / EE / EZL @ 30 / -1.5 / 13 kV. One spark while bring the BIAS to 30 kv around 27.5 kV. IV4 opened.
Friday, June 14, 2019, 14:39: Multiple wire scans attempted since 1pm. Controls & Diagnostics contacted. Ongoing troubleshooting.
Monday, June 17, 2019, 08:34: TMS:MB1 - Reset.
Monday, June 17, 2019, 09:18: Auto-ramping TBHT to 310 A @ 2 A per min.
FM
Monday, June 17, 2019, 10:30: TBHT manually tapped to 320 A ~2A/min
Tuesday, June 18, 2019, 18:42 : BIAS / EZL / EE @ 0 kV. Anode @ 0 V.
Tuesday, June 18, 2019, 18:43: Begin auto-cooling @ 2A per min.
AS
Wednesday, June 19, 2019, 07:10: IV4 Closed.
Wednesday, June 19, 2019, 07:12: Gas lines & Sources / power off.
Wednesday, June 19, 2019, 07:15: RGA off.
Wednesday, June 19, 2019, 07:16: IG1 & Turbo pumps off.
Wednesday, June 19, 2019, 08:35: Purging / venting. |
543
|
Friday, September 30, 2016, 11:11 |
Aaron Schmidt | SiC#35 | HP-FEBIAD | ITW | TM2 | 39530-9 | 41120-8 | Done | Conditioning SiC#35 | Friday, September 30, 2016, 10:00: Roughing.
Friday, September 30, 2016, 10:01: Roughing stopped. RP1 failure. Breaker tripped.
Friday, September 30, 2016, 11:49: Roughing continued.
Friday, September 30, 2016, 12:37: Turbos started -AM
Monday, October 03, 2016, 09:00: RGA on.
Monday, October 03, 2016, 09:05: Waterline opened. No response on RGA. Vacuum stable @1.70E-7T.
Monday, October 03, 2016, 09:20: Power/Sources on. Gas sources opened/closed.
Monday, October 03, 2016, 09:39: Optics switched on. Monitor TMS:SX2
Monday, October 03, 2016, 09:47: Anode on @ 200 V
Monday, October 03, 2016, 09:49: Coil on @ 50A
Monday, October 03, 2016, 09:50: IV4 opened.
Monday, October 03, 2016, 09:58: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .
Monday, October 03, 2016, 10:08: Auto ramping TGHT / TBHT to 580 / 290 A @ 2A per min.
Monday, October 03, 2016, 10:12
Device
|
Temp
[C] |
Flow
[lpm] |
HS + MSP |
27.0 |
3.99 |
TBHT |
27.5 |
3.21 |
COIL + EE |
26.8 |
5.26 |
TGHT |
26.9 |
5.78 |
Monday, October 03, 2016, 11:09: TMS:SX2 Fixed. No optics or sources affected.
Monday, October 03, 2016, 11:38: BIAS @ 30 kV. (Was actually set to 20 kV before). Ramping progress ok.
Monday, October 03, 2016, 15:56: TGHT / TBHT @ 580 / 290 A. RGA restarted.
Wednesday, October 05, 2016, 08:29: Optics found to be tripped : TMS:OCT4 , TMS:Q4 ,TMS:OCT3 ,TMS:SX3 ,TMS:Q3 ,TMS:SX2. Restarted
|
Temp
[C] |
Flow
[lpm] |
HS + MSP |
35.3 |
3.99 |
TBHT |
29.9 |
3.14 |
COIL + EE |
28.8 |
5.07 |
TGHT |
30.8 |
5.72 |
Wednesday, October 05, 2016, 09:20: Wire scanners reset. Emittance rig not reading properly.
Wednesday, October 05, 2016, 09:22: Auto ramping TBHT to 300 A @ 2A per min.
Thursday, October 06, 2016, 15:58: Auto-cooling TGHT / TBHT @ 2A per min.
|
Temp
[C] |
Flow
[lpm] |
HS + MSP |
35.4 |
3.95 |
TBHT |
29.4 |
3.09 |
COIL + EE |
28.8 |
5.05 |
TGHT |
30.8 |
5.70 |
Friday, October 07, 2016, 08:12: Source off - AM
Tuesday, October 11, 2016, 08:28: Turbos off. RGA off -AM
Tuesday, October 11, 2016, 13:38: Venting -AM |
Attachment 1: ER-580-290-M20-2016-10-05-09-10.txt
|
value (Voltage offset 0.0000)
75.0000
0.6333
75.0000
29978.0220
31
31
./emit.dat
-0.0006 -0.0012 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006 -0.0006 -0.0009 -0.0012 -0.0012 -0.0009 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006 -0.0006 -0.0009 -0.0009 -0.0009
-0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0003 -0.0009 -0.0003 -0.0009 -0.0009 -0.0006 -0.0006
-0.0009 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006 -0.0012 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0003 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006
-0.0009 -0.0009 -0.0006 -0.0006 -0.0009 -0.0009 -0.0003 -0.0009 -0.0009 -0.0006 -0.0012 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006 -0.0012 -0.0003 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0006
-0.0009 -0.0012 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0012 -0.0009 -0.0009 -0.0012 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006
-0.0009 -0.0009 -0.0006 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0006 0.0000 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0006 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0006
-0.0009 -0.0006 -0.0009 -0.0006 -0.0006 -0.0006 -0.0006 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0006 -0.0009 -0.0009 -0.0009
-0.0012 -0.0006 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0012 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0003 -0.0006 -0.0006 -0.0012 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0012 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009
-0.0009 -0.0009 -0.0006 -0.0003 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0003 -0.0012 -0.0009 -0.0003
-0.0006 -0.0009 -0.0012 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0003 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006 -0.0006 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006
-0.0006 -0.0009 -0.0012 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0006 -0.0009 -0.0006 -0.0006 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0003
-0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0006 -0.0009 -0.0009 -0.0006 -0.0012 -0.0009 -0.0009 -0.0009 -0.0009 -0.0012 -0.0006 -0.0009 -0.0012 -0.0006 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006
-0.0009 -0.0012 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0003 -0.0009 -0.0009 -0.0009 -0.0006 -0.0006 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006 -0.0009 -0.0012 -0.0009 -0.0009
-0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0006 -0.0009 -0.0012 -0.0009 -0.0012 -0.0009 -0.0006 -0.0009 -0.0006 -0.0006 -0.0006 -0.0006 -0.0006 -0.0006 -0.0003 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009
-0.0006 -0.0006 -0.0006 -0.0006 -0.0009 -0.0009 -0.0009 -0.0012 -0.0006 -0.0006 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0012 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0006 -0.0009 -0.0006 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006
-0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0006 -0.0006 -0.0003 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0012 -0.0009
-0.0009 -0.0009 -0.0009 -0.0006 -0.0006 -0.0009 -0.0006 -0.0012 -0.0009 -0.0012 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009
-0.0006 -0.0009 -0.0006 -0.0006 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006 -0.0006 -0.0009 -0.0012 -0.0006 -0.0009 -0.0009 -0.0003 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0003 -0.0006 -0.0009 -0.0006 -0.0003 -0.0009 -0.0012
-0.0006 -0.0006 -0.0003 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0003 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0006 -0.0009 -0.0006 -0.0006 -0.0006
-0.0009 -0.0006 -0.0009 -0.0003 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0012 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0012
-0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0003 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006
-0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0012 -0.0006 -0.0009 -0.0006 -0.0012 -0.0012 -0.0009 -0.0009 -0.0006 -0.0012 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0012 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006
-0.0012 -0.0006 -0.0009 -0.0006 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006 -0.0009 -0.0009 -0.0012 -0.0009
-0.0009 -0.0006 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0003 -0.0006 -0.0009 -0.0009 -0.0012 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0012 -0.0006 -0.0006 -0.0009
-0.0009 -0.0006 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0012 -0.0012 -0.0012 -0.0003 -0.0009 -0.0012 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0012 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006
-0.0006 -0.0009 -0.0009 -0.0006 -0.0003 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0003 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006 -0.0009
-0.0009 -0.0006 -0.0012 -0.0006 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0012 -0.0012 -0.0009 -0.0012 -0.0006 -0.0009 -0.0009 -0.0012 -0.0009 -0.0012 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0003 -0.0006 -0.0006 -0.0006
-0.0009 -0.0006 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006 -0.0006 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0012 -0.0009 -0.0006 -0.0003 -0.0009 -0.0009 -0.0009 -0.0009 -0.0003 -0.0009 -0.0006 -0.0012
-0.0006 -0.0006 -0.0009 -0.0009 -0.0009 -0.0012 -0.0003 -0.0006 -0.0006 -0.0006 -0.0006 -0.0012 -0.0009 -0.0009 -0.0006 -0.0009 -0.0012 -0.0006 -0.0006 -0.0009 -0.0009 -0.0006 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006 -0.0009 -0.0009 -0.0003 -0.0006
-0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0012 -0.0009 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0012 -0.0009 -0.0009 -0.0006 -0.0006 -0.0012 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0003 -0.0009 -0.0012 -0.0012 -0.0009 -0.0009
-0.0012 -0.0009 -0.0009 -0.0009 -0.0006 -0.0009 -0.0006 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0012 -0.0009 -0.0009 -0.0012 -0.0006 -0.0006 -0.0009 -0.0006 -0.0009 -0.0009 -0.0009 -0.0006 -0.0006 -0.0006 -0.0009 -0.0009 -0.0009 -0.0009 -0.0006
|
Attachment 2: SiC#35.docx
|
Attachment 3: TS-2016-10-03-09-00.rga
|
Attachment 4: TS-2016-10-03-16-00.rga
|
Attachment 5: TS-2016-10-05-09-43.rga
|
Attachment 6: ITW-TM2-SiC#35-HP-FEBIAD_Test_Stand.docx
|
330
|
Friday, August 30, 2013, 15:09 |
Anders Mjos | NiO#2 | HP-FEBIAD | ITE | TM3 | 39530-6 | 39069-4 | Done | Conditioning of NiO#2 | Friday, August 30, 2013, 15:09 : Turbos stopped in preparation for installation.
Friday, August 30, 2013, 16:22: Venting
Friday, August 30, 2013, 17:08: Roughing with NiO#2
Friday, August 30, 2013, 17:27: Starting turbos
Friday, August 30, 2013, 17:35: Turbos at speed
Friday, August 30, 2013, 17:37: Vacuum at 1.9 E-7 T
Monday, September 09, 2013, 14:04: Opened up water. Vacuum stable at 1.5E-7 T
Tuesday, September 10, 2013, 08:59: Pumping out the gas lines.
Tuesday, September 10, 2013, 09:10: Gas lines evacuated.
Tuesday, September 10, 2013, 09:23: TPS:IV4 opened
Tuesday, September 10, 2013, 10:06: TGHT PS was off. Debiased to access el cage.
Tuesday, September 10, 2013, 10:07: EZL/EE/BIAS: 15/2/30 kV 61/674/55 uA
Tuesday, September 10, 2013, 10:08: COIL/ANODE: 50A/200V
Tuesday, September 10, 2013, 10:09: Started target heat to TGHT/TBHT 250/230A
Tuesday, September 10, 2013, 15:12: Started target heat to TGHT 350A (TBHT remains at 230A)
Wednesday, September 11, 2013, 13:06: Heating TGHT to 500A
Thursday, September 12, 2013, 08:44: Heating TGHT to 580A
Thursday, September 12, 2013, 09:39: Heating TBHT to 260A
Thursday, September 12, 2013, 10:06: BIAS sparking occurred at 9:48 TBHT:RDCUR at 256A. Gradually reduced TBHT:RDCUR to 254A. Sparking stopped after ~20 min. ANODE:RDCUR, EE:RDCUR and IG1:RDVAC showed correlated sparking responses. ANODE:RDCUR went to 0 mA and back up after sparking had stopped.
Thursday, September 12, 2013, 10:30: Manually increased TBHT:RDCUR to 260A.
Thursday, September 12, 2013, 11:06: A jump in TBHT:RDVOLT and TBHT:RDCUR at 10:44. Reduced TBHT:RDCUR to 255A
Thursday, September 12, 2013, 12:10: TPS:Q4 was taken out by a spark. Restarted.
Thursday, September 12, 2013, 12:19: EZL to 13 kV
Thursday, September 12, 2013, 13:14: TBHT:RDCUR to 260A
Thursday, September 12, 2013, 13:31: TBHT:RDCUR to 265A. Sparking returns. Lower to 260A
Friday, September 13, 2013, 10:51: Increased TBHT:RDCUR to 265A
Friday, September 13, 2013, 10:58 : Increased TBHT:RDCUR to 270A
Friday, September 13, 2013, 11:05: Some sparking. Lowered EE to 1kV
Friday, September 13, 2013, 11:07 : Increased TBHT:RDCUR to 275A
Friday, September 13, 2013, 11:15 : Increased TBHT:RDCUR to 280A
Friday, September 13, 2013, 11:24 : Increased TBHT:RDCUR to 290A. ANODE:RDCUR 8.5mA. FC0 1.3E-6 A
Friday, September 13, 2013, 11:53: SV5 and SV2 gas lines opened.
Friday, September 13, 2013, 12:05: A spark took out Q4. Restarted.
Friday, September 13, 2013, 12:07: Lowered TBHT:RDCUR to 285A
Friday, September 13, 2013, 13:03: Started mass scan.
Friday, September 13, 2013, 14:27: Lowered TBHT:RDCUR to 280A
Friday, September 13, 2013, 15:01 : Lowered TBHT:RDCUR to 270A
Monday, September 16, 2013, 08:00: Mass scan started.
Monday, September 16, 2013, 09:00: Starting heater ramp down. 2A/min. Gas lines closed.
Monday, September 16, 2013, 13:30: Source off. Target debiased.
Monday, September 16, 2013, 13:56: Turbos stopped
Monday, September 16, 2013, 14:50: Venting. |
Attachment 1: 2013-09-12_NiO#2_TBHT_jump.PNG
|
|
333
|
Monday, September 16, 2013, 17:07 |
Anders Mjos | NiO#2 | HP-FEBIAD | ITE | TM3 | 39530-6 | 39069-4 | Done | Conditioning NiO#2 take 2 | The GE on the Source tray and EE on the target was cleaned. The target will be conditioned again to verify if sparking has been reduced. Torque 120 in lbs
Monday, September 16, 2013, 17:05: Roughing. The roughing pump is somewhat noisy.
Monday, September 16, 2013, 17:26: Starting turbos.
Monday, September 16, 2013, 17:34: Turbos at speed
Monday, September 16, 2013, 17:36: Vacuum at 3.8E-5 T.
Tuesday, September 17, 2013, 08:10: Opened cooling water. Vacuum stable at 6.5E-7. EE flow 4.8 l/min 27.9 C
Tuesday, September 17, 2013, 08:20: Target biased: 30kV, 15kV, -1kV.
Tuesday, September 17, 2013, 08:25: Opening gas line SV5 and Ne SV3 evacuated
Tuesday, September 17, 2013, 08:32: Started auto-ramping to 580A/230A
Tuesday, September 17, 2013, 08:34: Coil to 50A, Anode to 200V (0mA)
Tuesday, September 17, 2013, 08:36: Opened IV4
Tuesday, September 17, 2013, 10:39: EE flow 4.9 l/min 28.5 C
Tuesday, September 17, 2013, 13:05: EE flow 4.9 l/min 28.8 C
Tuesday, September 17, 2013, 15:03: EE flow 4.9 l/min 28.8 C
Tuesday, September 17, 2013, 15:04 :TGHT at 580A, TBHT at 280A.
Tuesday, September 17, 2013, 15:30: TBHT:RDCUR to 290A
Wednesday, September 18, 2013, 08:07: Stared mass scan.
Wednesday, September 18, 2013, 08:08:EE flow 4.9 l/min 28.8 C
Wednesday, September 18, 2013, 09:14: EE to 2kV. Current at 676 uA
Wednesday, September 18, 2013, 09:23: Opened He gas to 6.0. It was later found the SV2 was closed. I.e. no He was supplied to the target.
Wednesday, September 18, 2013, 09:26: Tuning source. EE 0.8kV, EZL 1.3kV
Wednesday, September 18, 2013, 09:32: Started mass scan.
Wednesday, September 18, 2013, 09:55: A jump in BIAS current from 42 to 61 uA. Came down towards 42 after 30 min
Wednesday, September 18, 2013, 10:22: Ramping down heaters at 2A and 2.5A /min
Wednesday, September 18, 2013, 10:23: He and Ne off.
Wednesday, September 18, 2013, 10:48: Gas lines closed.
Wednesday, September 18, 2013, 10:53: EE flow 4.9 l/min 28.5 C
Wednesday, September 18, 2013, 10:55: IV4 closed.
Wednesday, September 18, 2013, 10:59: Target de-biased.
Wednesday, September 18, 2013, 12:18: Heaters off.
Wednesday, September 18, 2013, 12:18: Stopped turbos.
Wednesday, September 18, 2013, 12:56: Started wenting.
Wednesday, September 18, 2013, 13:02: EE flow 4.9 l/min 28.2 C
Wednesday, September 18, 2013, 13:20 : Chamber vented. |
559
|
Wednesday, January 18, 2017, 09:23 |
Aaron Schmidt | Ta Test | LP-FEBIAD | | | 39530-3 | | In progress | Tri-Coil FEBIAD for testing | Wednesday, January 18, 2017, 09:22: Roughing
Wednesday, January 18, 2017, 09:41: Turbo pumps started.
Wednesday, January 18, 2017, 11:01: Ion gauge on.
Thursday, January 19, 2017, 08:15: Water lines opened. No response on RGA. Vacuum stable @ 1.70e-07T.
Thursday, January 19, 2017, 08:55: Power/Sources on. BIAS/EZL/EE @ 30/15/1.5 kV . Coil @ 50 A . Anode @ 200 V.
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
24.9 |
3.99 |
TBHT |
24.8 |
3.15 |
COIL + EE |
24.7 |
5.16 |
TGHT |
24.7 |
5.05 |
Thursday, January 19, 2017, 08:59: IV4 opened. Auto ramping TGHT / TBHT to 580 / 290 A @ 2A per min.
Friday, January 20, 2017, 08:25: TSG : TP1A and TSG : BV1A were noticed to be tripped around 6am.
Friday, January 20, 2017, 08:42: Bypass to restart TSG : TP1A and TSG : BV1A. TP1A is at speed.
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
34.1 |
3.99 |
TBHT |
25.0 |
3.19 |
COIL + EE |
24.7 |
5.19 |
TGHT |
26.5 |
5.05 |
Friday, January 20, 2017, 11:48: Auto ramping TBHT to 320 A @ 2A per min.
Friday, January 20, 2017, 15:51: Coil @ 100A
Saturday, January 21, 2017, 11:00: Increased TB HT voltage limit to 2.75 V. TBHT current went back up from 313 to 319 A. -AM
Monday, January 23, 2017, 10:08: Argon Valve closed.
Monday, January 23, 2017, 10:33: Neon Valve closed.
Monday, January 23, 2017, 11:00: Auto-cooling @ 4A per min. Coil reduced from 100 A to 50 A
Monday, January 23, 2017, 13:32: Power/sources off.
Monday, January 23, 2017, 13:35: RGA off. IV4 closed. Ion gauge off. Turbo pumps off.
Monday, January 23, 2017, 14:45: Venting |
Attachment 1: ER-580-290-30kV-m27.txt
|
value (Voltage offset 40.0000)
80.0000
0.7000
80.0000
29978.0220
21
21
./emit.dat
-0.0037 -0.0031 -0.0009 -0.0018 -0.0024 -0.0040 -0.0043 -0.0043 -0.0043 -0.0037 -0.0040 -0.0037 -0.0040 -0.0043 -0.0040 -0.0046 -0.0043 -0.0046 -0.0043 -0.0046 -0.0040
-0.0040 -0.0031 -0.0012 -0.0018 -0.0034 -0.0043 -0.0043 -0.0037 -0.0040 -0.0040 -0.0037 -0.0037 -0.0040 -0.0043 -0.0037 -0.0040 -0.0046 -0.0040 -0.0043 -0.0040 -0.0043
-0.0043 -0.0031 -0.0009 -0.0021 -0.0034 -0.0034 -0.0034 -0.0040 -0.0040 -0.0040 -0.0043 -0.0040 -0.0037 -0.0037 -0.0043 -0.0037 -0.0040 -0.0037 -0.0040 -0.0040 -0.0037
-0.0040 -0.0034 -0.0012 -0.0021 -0.0037 -0.0040 -0.0043 -0.0040 -0.0043 -0.0040 -0.0043 -0.0040 -0.0040 -0.0034 -0.0040 -0.0037 -0.0040 -0.0037 -0.0037 -0.0037 -0.0037
-0.0043 -0.0027 -0.0006 -0.0021 -0.0031 -0.0037 -0.0043 -0.0043 -0.0043 -0.0040 -0.0043 -0.0043 -0.0043 -0.0040 -0.0043 -0.0043 -0.0043 -0.0040 -0.0037 -0.0040 -0.0040
-0.0037 -0.0027 -0.0009 -0.0018 -0.0031 -0.0040 -0.0040 -0.0040 -0.0043 -0.0043 -0.0040 -0.0040 -0.0037 -0.0040 -0.0043 -0.0046 -0.0043 -0.0040 -0.0043 -0.0046 -0.0043
-0.0040 -0.0031 -0.0009 -0.0018 -0.0027 -0.0043 -0.0034 -0.0040 -0.0040 -0.0043 -0.0040 -0.0043 -0.0037 -0.0037 -0.0040 -0.0037 -0.0040 -0.0043 -0.0043 -0.0043 -0.0040
-0.0040 -0.0034 -0.0012 -0.0015 -0.0031 -0.0043 -0.0003 -0.0018 -0.0034 -0.0034 -0.0040 -0.0037 -0.0040 -0.0040 -0.0037 -0.0040 -0.0037 -0.0043 -0.0040 -0.0037 -0.0034
-0.0043 -0.0034 -0.0012 -0.0021 -0.0034 -0.0043 -0.0031 -0.0009 -0.0006 -0.0031 -0.0040 -0.0043 -0.0043 -0.0040 -0.0040 -0.0037 -0.0043 -0.0037 -0.0040 -0.0040 -0.0037
-0.0040 -0.0031 -0.0012 -0.0021 -0.0031 -0.0037 -0.0043 -0.0043 -0.0027 -0.0012 -0.0021 -0.0034 -0.0043 -0.0043 -0.0040 -0.0040 -0.0043 -0.0040 -0.0043 -0.0040 -0.0040
-0.0037 -0.0024 -0.0009 -0.0015 -0.0027 -0.0037 -0.0040 -0.0040 -0.0037 -0.0034 -0.0024 -0.0012 -0.0024 -0.0040 -0.0043 -0.0040 -0.0043 -0.0043 -0.0049 -0.0043 -0.0043
-0.0037 -0.0031 -0.0012 -0.0021 -0.0031 -0.0037 -0.0040 -0.0037 -0.0037 -0.0037 -0.0037 -0.0034 -0.0027 -0.0015 -0.0031 -0.0040 -0.0043 -0.0037 -0.0040 -0.0043 -0.0043
-0.0043 -0.0034 -0.0012 -0.0018 -0.0034 -0.0040 -0.0040 -0.0043 -0.0043 -0.0037 -0.0040 -0.0040 -0.0034 -0.0031 -0.0018 -0.0027 -0.0037 -0.0037 -0.0040 -0.0040 -0.0037
-0.0043 -0.0034 -0.0015 -0.0024 -0.0037 -0.0043 -0.0043 -0.0043 -0.0043 -0.0040 -0.0043 -0.0043 -0.0043 -0.0040 -0.0037 -0.0034 -0.0024 -0.0031 -0.0037 -0.0040 -0.0037
-0.0040 -0.0034 -0.0009 -0.0021 -0.0031 -0.0040 -0.0043 -0.0043 -0.0049 -0.0043 -0.0046 -0.0043 -0.0043 -0.0040 -0.0043 -0.0037 -0.0037 -0.0031 -0.0034 -0.0043 -0.0040
-0.0037 -0.0027 -0.0006 -0.0018 -0.0031 -0.0037 -0.0037 -0.0037 -0.0043 -0.0040 -0.0037 -0.0043 -0.0040 -0.0043 -0.0043 -0.0037 -0.0037 -0.0040 -0.0031 -0.0031 -0.0043
-0.0040 -0.0031 -0.0012 -0.0021 -0.0027 -0.0037 -0.0037 -0.0040 -0.0040 -0.0040 -0.0034 -0.0037 -0.0040 -0.0040 -0.0037 -0.0040 -0.0040 -0.0037 -0.0043 -0.0040 -0.0031
-0.0043 -0.0031 -0.0009 -0.0021 -0.0034 -0.0040 -0.0037 -0.0040 -0.0043 -0.0037 -0.0043 -0.0037 -0.0040 -0.0037 -0.0037 -0.0037 -0.0040 -0.0037 -0.0037 -0.0043 -0.0040
-0.0040 -0.0031 -0.0012 -0.0021 -0.0031 -0.0040 -0.0040 -0.0043 -0.0040 -0.0040 -0.0043 -0.0037 -0.0040 -0.0040 -0.0040 -0.0037 -0.0037 -0.0043 -0.0040 -0.0037 -0.0040
-0.0040 -0.0031 -0.0012 -0.0021 -0.0031 -0.0043 -0.0046 -0.0043 -0.0046 -0.0040 -0.0043 -0.0043 -0.0040 -0.0040 -0.0040 -0.0043 -0.0040 -0.0037 -0.0037 -0.0043 -0.0043
-0.0043 -0.0031 -0.0009 -0.0015 -0.0024 -0.0040 -0.0043 -0.0037 -0.0043 -0.0043 -0.0040 -0.0043 -0.0040 -0.0043 -0.0043 -0.0046 -0.0043 -0.0037 -0.0043 -0.0043 -0.0043
|
Attachment 2: ER-580-290-30kV-m28(2).txt
|
value (Voltage offset 40.0000)
80.0000
0.7000
80.0000
29978.0220
21
21
./emit.dat
-0.1248 -0.1074 -0.1031 -0.1035 -0.1016 -0.1047 -0.1035 -0.1041 -0.1086 -0.1050 -0.1047 -0.1041 -0.1038 -0.1016 -0.1022 -0.1025 -0.1031 -0.1065 -0.1074 -0.1071 -0.1038
-0.1001 -0.1038 -0.0992 -0.0980 -0.0961 -0.1016 -0.1044 -0.1013 -0.1050 -0.1028 -0.1031 -0.1001 -0.0992 -0.0970 -0.1016 -0.1010 -0.1025 -0.1010 -0.1038 -0.1025 -0.0989
-0.0980 -0.1025 -0.0989 -0.0998 -0.0980 -0.0977 -0.0998 -0.1031 -0.1068 -0.1028 -0.1016 -0.1041 -0.0995 -0.1010 -0.1025 -0.1004 -0.1016 -0.0977 -0.1019 -0.0995 -0.0970
-0.0977 -0.1013 -0.0995 -0.1013 -0.0940 -0.0946 -0.0961 -0.1019 -0.1013 -0.0967 -0.0995 -0.0986 -0.0986 -0.0998 -0.0995 -0.0998 -0.0970 -0.0974 -0.1004 -0.0998 -0.1001
-0.0998 -0.1044 -0.0998 -0.1038 -0.0974 -0.0943 -0.0949 -0.0919 -0.0958 -0.0946 -0.0998 -0.0989 -0.0980 -0.0967 -0.0952 -0.0995 -0.0983 -0.0958 -0.1010 -0.1013 -0.1010
-0.1007 -0.1041 -0.0986 -0.1053 -0.0970 -0.0894 -0.0922 -0.0867 -0.0928 -0.0912 -0.0998 -0.0955 -0.0989 -0.0992 -0.0983 -0.1007 -0.0983 -0.0986 -0.1031 -0.1035 -0.0986
-0.0986 -0.1013 -0.0949 -0.1041 -0.0958 -0.0974 -0.0922 -0.0842 -0.0897 -0.0937 -0.0964 -0.0934 -0.0995 -0.1010 -0.0995 -0.1016 -0.0995 -0.1001 -0.1038 -0.1013 -0.0983
-0.0946 -0.0980 -0.0937 -0.1013 -0.0961 -0.0964 -0.0949 -0.0833 -0.0858 -0.0906 -0.0946 -0.0946 -0.0967 -0.1010 -0.1001 -0.0983 -0.0974 -0.0986 -0.1007 -0.0989 -0.0967
-0.0986 -0.0986 -0.0946 -0.1010 -0.0980 -0.0928 -0.0961 -0.0873 -0.0812 -0.0806 -0.0900 -0.0952 -0.0989 -0.0983 -0.0952 -0.0958 -0.0967 -0.0970 -0.1010 -0.0995 -0.0967
-0.0992 -0.1025 -0.0995 -0.1016 -0.0931 -0.0916 -0.0980 -0.0897 -0.0818 -0.0726 -0.0851 -0.0949 -0.0970 -0.0940 -0.0955 -0.0980 -0.0980 -0.0989 -0.1025 -0.1001 -0.0986
-0.0995 -0.1041 -0.0983 -0.1056 -0.0958 -0.0903 -0.0970 -0.0885 -0.0830 -0.0739 -0.0815 -0.0891 -0.0946 -0.0940 -0.0977 -0.1019 -0.1004 -0.0989 -0.1019 -0.1059 -0.0989
-0.0995 -0.1025 -0.0955 -0.1019 -0.0967 -0.0955 -0.0989 -0.0894 -0.0864 -0.0778 -0.0781 -0.0845 -0.0912 -0.0958 -0.0983 -0.1016 -0.1001 -0.0970 -0.1007 -0.1001 -0.0995
-0.0980 -0.1016 -0.0970 -0.1035 -0.0974 -0.0974 -0.1044 -0.0967 -0.0928 -0.0827 -0.0812 -0.0839 -0.0894 -0.0949 -0.0995 -0.1004 -0.0970 -0.0958 -0.0989 -0.0986 -0.0958
-0.0986 -0.1022 -0.0983 -0.1065 -0.0986 -0.0998 -0.1053 -0.1010 -0.1016 -0.0897 -0.0888 -0.0861 -0.0894 -0.0955 -0.0992 -0.1001 -0.0961 -0.0967 -0.1004 -0.1010 -0.0977
-0.0998 -0.1028 -0.0989 -0.1071 -0.0983 -0.1007 -0.1044 -0.0986 -0.1007 -0.0967 -0.0989 -0.0919 -0.0928 -0.0934 -0.0949 -0.1001 -0.0967 -0.0983 -0.1022 -0.1025 -0.0977
-0.0998 -0.1041 -0.0977 -0.1056 -0.0977 -0.0977 -0.1038 -0.0986 -0.1019 -0.0995 -0.1007 -0.0964 -0.0964 -0.0967 -0.0940 -0.0986 -0.0989 -0.0995 -0.1035 -0.1016 -0.0964
-0.0983 -0.1025 -0.0961 -0.1044 -0.0983 -0.0970 -0.1031 -0.0995 -0.1028 -0.0998 -0.1001 -0.0980 -0.0986 -0.0995 -0.0964 -0.1013 -0.0992 -0.0998 -0.1013 -0.1001 -0.0952
-0.0980 -0.0998 -0.0977 -0.1050 -0.0974 -0.0961 -0.1053 -0.1028 -0.1041 -0.0983 -0.1007 -0.0989 -0.0995 -0.0995 -0.0983 -0.1013 -0.0980 -0.0961 -0.0983 -0.0983 -0.0961
-0.0964 -0.1013 -0.0977 -0.1065 -0.0995 -0.0980 -0.1074 -0.1001 -0.1041 -0.0983 -0.0989 -0.0995 -0.0974 -0.0958 -0.0977 -0.1007 -0.0961 -0.0952 -0.0995 -0.0980 -0.0983
-0.0992 -0.1031 -0.0995 -0.1053 -0.0967 -0.0986 -0.1086 -0.0986 -0.1007 -0.0964 -0.0974 -0.0974 -0.0961 -0.0964 -0.0980 -0.0989 -0.0989 -0.0967 -0.1016 -0.1013 -0.1007
-0.0998 -0.1028 -0.0970 -0.1044 -0.0961 -0.0958 -0.1041 -0.0970 -0.0992 -0.0980 -0.0974 -0.0974 -0.0970 -0.0958 -0.0995 -0.1019 -0.0986 -0.0964 -0.1016 -0.1019 -0.0983
|
Attachment 3: ER-580-290-30kV-m28.txt
|
value (Voltage offset 40.0000)
80.0000
0.7000
80.0000
29978.0220
21
21
./emit.dat
-0.1248 -0.1175 -0.1169 -0.1175 -0.1184 -0.1144 -0.1187 -0.1187 -0.1154 -0.1178 -0.1163 -0.1187 -0.1105 -0.1099 -0.1096 -0.1117 -0.1230 -0.1071 -0.1129 -0.1154 -0.1184
-0.1022 -0.1117 -0.1114 -0.1108 -0.1111 -0.1117 -0.1169 -0.1154 -0.1117 -0.1135 -0.1141 -0.1154 -0.1025 -0.1068 -0.1093 -0.1108 -0.1135 -0.1016 -0.1099 -0.1138 -0.1135
-0.0992 -0.1123 -0.1105 -0.1028 -0.0952 -0.1025 -0.1038 -0.1053 -0.1056 -0.1117 -0.1154 -0.1160 -0.1022 -0.1050 -0.1074 -0.1044 -0.1053 -0.1010 -0.1108 -0.1126 -0.1028
-0.1004 -0.1123 -0.1105 -0.1028 -0.0977 -0.0842 -0.0858 -0.0922 -0.1013 -0.1038 -0.1154 -0.1135 -0.0974 -0.1001 -0.1035 -0.1007 -0.1028 -0.1016 -0.1074 -0.1102 -0.0995
-0.0998 -0.1117 -0.1126 -0.1086 -0.1074 -0.0754 -0.0592 -0.0571 -0.0763 -0.1001 -0.1135 -0.1120 -0.0992 -0.0998 -0.0961 -0.1001 -0.0989 -0.0998 -0.1062 -0.1105 -0.0998
-0.0909 -0.1099 -0.1129 -0.1068 -0.1053 -0.0900 -0.0525 -0.0357 -0.0497 -0.0778 -0.1028 -0.1117 -0.1004 -0.1007 -0.0967 -0.0998 -0.1001 -0.1019 -0.1074 -0.1111 -0.1010
-0.0879 -0.0992 -0.1019 -0.0946 -0.1016 -0.1025 -0.0711 -0.0122 -0.0278 -0.0580 -0.0842 -0.0937 -0.0980 -0.1007 -0.1016 -0.0992 -0.1007 -0.1025 -0.1089 -0.1129 -0.1019
-0.0906 -0.1013 -0.0949 -0.0812 -0.0845 -0.0897 -0.0842 -0.0336 0.0085 -0.0372 -0.0671 -0.0925 -0.0848 -0.0937 -0.1007 -0.1044 -0.1016 -0.1010 -0.1071 -0.1123 -0.1022
-0.0986 -0.1071 -0.0943 -0.0891 -0.0812 -0.0717 -0.0732 -0.0504 0.0134 0.0082 -0.0507 -0.0742 -0.0864 -0.0897 -0.0940 -0.0992 -0.1016 -0.0983 -0.1050 -0.1108 -0.1013
-0.0974 -0.1120 -0.1105 -0.0928 -0.0864 -0.0735 -0.0714 -0.0601 -0.0131 0.0418 -0.0113 -0.0635 -0.0754 -0.0891 -0.0949 -0.0937 -0.0955 -0.0989 -0.1056 -0.1108 -0.1007
-0.1007 -0.1123 -0.1108 -0.1028 -0.1001 -0.0754 -0.0784 -0.0674 -0.0421 0.0128 0.0302 -0.0391 -0.0693 -0.0891 -0.0958 -0.0986 -0.0952 -0.0961 -0.1062 -0.1111 -0.0998
-0.1028 -0.1120 -0.1123 -0.1050 -0.1041 -0.0967 -0.0815 -0.0702 -0.0589 -0.0150 0.0162 -0.0098 -0.0604 -0.0839 -0.0955 -0.0986 -0.0974 -0.0998 -0.1059 -0.1105 -0.1007
-0.0995 -0.1108 -0.1117 -0.1074 -0.1025 -0.1004 -0.1007 -0.0903 -0.0665 -0.0464 -0.0153 -0.0027 -0.0378 -0.0827 -0.0934 -0.0964 -0.0995 -0.0986 -0.1059 -0.1099 -0.0995
-0.0986 -0.1105 -0.1108 -0.1041 -0.1041 -0.1001 -0.1019 -0.0989 -0.0992 -0.0784 -0.0555 -0.0360 -0.0363 -0.0732 -0.0876 -0.0983 -0.1007 -0.0986 -0.1044 -0.1086 -0.0992
-0.0977 -0.1114 -0.1120 -0.1022 -0.1041 -0.0989 -0.1013 -0.1004 -0.1062 -0.1083 -0.0992 -0.0784 -0.0583 -0.0656 -0.0845 -0.0974 -0.0998 -0.0992 -0.1035 -0.1080 -0.1004
-0.0974 -0.1129 -0.1129 -0.1025 -0.1031 -0.1016 -0.1028 -0.1025 -0.1074 -0.1080 -0.1099 -0.1111 -0.0891 -0.0818 -0.0851 -0.0931 -0.1010 -0.0995 -0.1044 -0.1093 -0.0998
-0.0998 -0.1135 -0.1111 -0.1010 -0.1022 -0.1004 -0.1031 -0.1022 -0.1059 -0.1086 -0.1129 -0.1135 -0.0992 -0.0970 -0.0919 -0.0925 -0.0995 -0.1004 -0.1062 -0.1117 -0.1010
-0.1016 -0.1120 -0.1089 -0.1010 -0.1019 -0.0977 -0.1013 -0.1013 -0.1028 -0.1053 -0.1144 -0.1135 -0.1010 -0.0992 -0.1013 -0.1007 -0.0974 -0.0974 -0.1059 -0.1114 -0.1013
-0.0998 -0.1102 -0.1102 -0.1022 -0.1041 -0.0964 -0.1016 -0.0989 -0.0995 -0.1047 -0.1135 -0.1120 -0.0995 -0.0998 -0.0998 -0.1025 -0.0980 -0.0974 -0.1038 -0.1120 -0.0992
-0.0992 -0.1108 -0.1117 -0.1062 -0.1053 -0.0986 -0.1025 -0.0995 -0.1035 -0.1062 -0.1105 -0.1108 -0.0977 -0.1001 -0.0983 -0.0983 -0.0970 -0.0992 -0.1050 -0.1093 -0.0974
-0.1022 -0.1123 -0.1114 -0.1025 -0.1041 -0.1001 -0.1035 -0.1016 -0.1035 -0.1083 -0.1089 -0.1111 -0.0992 -0.0977 -0.0992 -0.0970 -0.0992 -0.1022 -0.1053 -0.1102 -0.0977
|
Attachment 4: ER-580-320-30kV-m20.txt
|
value (Voltage offset 40.0000)
80.0000
0.7000
80.0000
29978.0220
21
21
./emit.dat
-0.1202 -0.1038 -0.1041 -0.1031 -0.1041 -0.1047 -0.1016 -0.1041 -0.1105 -0.1059 -0.1053 -0.1022 -0.1044 -0.1025 -0.1025 -0.1016 -0.1004 -0.1022 -0.1044 -0.1016 -0.1035
-0.0980 -0.0986 -0.1004 -0.0980 -0.0983 -0.1001 -0.0980 -0.1004 -0.1041 -0.0983 -0.1001 -0.0974 -0.0986 -0.0977 -0.0961 -0.0974 -0.0992 -0.0983 -0.1016 -0.1007 -0.0989
-0.0974 -0.1031 -0.1019 -0.1007 -0.0983 -0.0992 -0.1013 -0.1001 -0.1019 -0.0992 -0.0986 -0.0983 -0.0977 -0.0986 -0.0980 -0.1001 -0.0995 -0.0989 -0.1022 -0.1019 -0.0980
-0.0998 -0.1050 -0.1028 -0.0995 -0.0989 -0.0983 -0.0989 -0.1019 -0.1007 -0.0970 -0.0998 -0.0998 -0.1013 -0.0952 -0.1028 -0.1004 -0.1013 -0.0998 -0.1025 -0.1022 -0.0986
-0.1001 -0.1016 -0.1010 -0.1013 -0.0983 -0.0974 -0.1004 -0.0952 -0.1007 -0.0998 -0.1071 -0.0998 -0.0989 -0.1001 -0.0995 -0.1004 -0.0974 -0.0977 -0.1019 -0.1010 -0.0995
-0.0986 -0.0998 -0.0992 -0.1010 -0.0980 -0.0955 -0.0946 -0.0958 -0.1007 -0.0998 -0.1019 -0.0974 -0.0967 -0.1001 -0.0970 -0.0992 -0.0983 -0.0964 -0.1025 -0.0998 -0.0992
-0.0974 -0.1007 -0.0986 -0.0992 -0.0970 -0.0949 -0.0943 -0.0818 -0.0949 -0.0958 -0.0995 -0.0964 -0.0958 -0.1004 -0.0983 -0.0995 -0.0955 -0.0970 -0.1035 -0.1007 -0.0992
-0.0986 -0.1001 -0.0995 -0.0980 -0.0964 -0.0967 -0.0967 -0.0839 -0.0836 -0.0912 -0.0961 -0.0967 -0.0995 -0.0998 -0.0986 -0.1007 -0.0989 -0.0995 -0.1031 -0.1007 -0.0998
-0.1004 -0.1010 -0.0995 -0.0998 -0.0977 -0.0964 -0.0998 -0.0909 -0.0769 -0.0845 -0.0952 -0.0977 -0.0998 -0.0998 -0.0983 -0.0998 -0.0998 -0.1013 -0.1035 -0.1007 -0.0989
-0.0961 -0.1013 -0.1013 -0.0995 -0.0952 -0.0946 -0.1013 -0.0952 -0.0854 -0.0797 -0.0940 -0.0983 -0.1004 -0.1001 -0.0967 -0.1004 -0.1004 -0.0977 -0.1022 -0.0992 -0.0980
-0.0961 -0.0998 -0.1010 -0.0995 -0.0970 -0.0934 -0.1004 -0.0928 -0.0894 -0.0769 -0.0879 -0.0958 -0.0964 -0.0970 -0.0961 -0.1001 -0.0983 -0.0946 -0.1010 -0.0986 -0.0970
-0.0977 -0.1007 -0.1007 -0.0992 -0.0967 -0.0977 -0.0983 -0.0903 -0.0900 -0.0806 -0.0815 -0.0916 -0.0946 -0.0943 -0.0967 -0.1010 -0.0949 -0.0961 -0.0998 -0.1004 -0.0958
-0.1004 -0.1028 -0.1004 -0.1013 -0.0986 -0.0983 -0.1022 -0.0964 -0.0946 -0.0854 -0.0861 -0.0882 -0.0946 -0.0958 -0.0998 -0.1004 -0.0977 -0.0967 -0.1028 -0.1010 -0.0967
-0.0992 -0.1010 -0.1007 -0.1035 -0.1004 -0.0974 -0.1038 -0.0992 -0.1047 -0.0958 -0.0925 -0.0922 -0.0952 -0.0983 -0.0998 -0.0977 -0.0986 -0.0986 -0.1022 -0.1007 -0.1004
-0.0983 -0.1019 -0.0995 -0.1022 -0.0989 -0.0989 -0.1047 -0.1007 -0.1044 -0.0995 -0.0998 -0.0961 -0.0967 -0.0977 -0.0998 -0.0998 -0.0974 -0.0961 -0.1025 -0.1001 -0.1007
-0.0977 -0.1007 -0.0989 -0.1013 -0.0977 -0.0964 -0.1041 -0.1019 -0.1050 -0.0992 -0.1004 -0.0961 -0.0970 -0.0992 -0.0970 -0.0995 -0.0943 -0.0970 -0.1013 -0.0986 -0.0986
-0.0967 -0.0989 -0.0983 -0.0998 -0.0967 -0.0946 -0.1035 -0.0986 -0.1004 -0.0961 -0.0974 -0.0949 -0.0980 -0.0983 -0.0958 -0.0980 -0.0961 -0.0983 -0.1019 -0.0989 -0.0964
-0.0970 -0.1007 -0.1004 -0.1019 -0.0970 -0.0964 -0.1025 -0.0977 -0.0989 -0.0946 -0.0974 -0.0989 -0.0986 -0.0967 -0.0961 -0.1004 -0.0992 -0.0995 -0.1028 -0.1010 -0.0955
-0.0980 -0.1022 -0.1010 -0.1013 -0.0983 -0.0992 -0.1022 -0.1004 -0.0998 -0.0970 -0.1004 -0.1004 -0.1004 -0.0980 -0.0989 -0.1050 -0.0998 -0.0995 -0.1025 -0.1022 -0.0977
-0.0964 -0.1025 -0.1004 -0.1013 -0.0992 -0.0989 -0.1056 -0.1007 -0.1035 -0.0998 -0.1022 -0.0998 -0.0986 -0.0992 -0.0977 -0.1038 -0.0989 -0.0964 -0.1016 -0.1022 -0.0970
-0.0964 -0.1007 -0.0989 -0.1004 -0.0970 -0.0967 -0.1022 -0.0967 -0.1044 -0.1004 -0.1010 -0.0986 -0.0949 -0.0964 -0.0983 -0.0977 -0.0961 -0.0946 -0.0998 -0.0989 -0.0970
|
Attachment 5: ER-580-320-30kV-m22.txt
|
value (Voltage offset 40.0000)
80.0000
0.7000
80.0000
29978.0220
21
21
./emit.dat
-0.1166 -0.0989 -0.0992 -0.0989 -0.0970 -0.0986 -0.0983 -0.1007 -0.1050 -0.0964 -0.0986 -0.0977 -0.0977 -0.0974 -0.0970 -0.0970 -0.0974 -0.0995 -0.1044 -0.1016 -0.1016
-0.0916 -0.0964 -0.0974 -0.0961 -0.0934 -0.0967 -0.0958 -0.0946 -0.1007 -0.0931 -0.0943 -0.0949 -0.0946 -0.0958 -0.0952 -0.0940 -0.0949 -0.0967 -0.1010 -0.0946 -0.0964
-0.0928 -0.0998 -0.0967 -0.0958 -0.0952 -0.0940 -0.0943 -0.0952 -0.0970 -0.0949 -0.0949 -0.0958 -0.0952 -0.0970 -0.0937 -0.0949 -0.0958 -0.0940 -0.1007 -0.0964 -0.0934
-0.0967 -0.0977 -0.0967 -0.0964 -0.0912 -0.0949 -0.0928 -0.0946 -0.0946 -0.0928 -0.0967 -0.0949 -0.0937 -0.0848 -0.0940 -0.0958 -0.0946 -0.0928 -0.0983 -0.0961 -0.0916
-0.0964 -0.0970 -0.0925 -0.0977 -0.0922 -0.0952 -0.0946 -0.0961 -0.0974 -0.0931 -0.0961 -0.0946 -0.0922 -0.0931 -0.0906 -0.0946 -0.0952 -0.0928 -0.0986 -0.0970 -0.0928
-0.0943 -0.0955 -0.0937 -0.0998 -0.0940 -0.0934 -0.0964 -0.0977 -0.0970 -0.0931 -0.0949 -0.0925 -0.0928 -0.0955 -0.0940 -0.0967 -0.0961 -0.0925 -0.0998 -0.0992 -0.0946
-0.0928 -0.0955 -0.0949 -0.1047 -0.0955 -0.0964 -0.0940 -0.0949 -0.0952 -0.0937 -0.0958 -0.0928 -0.0937 -0.0946 -0.0967 -0.0958 -0.0946 -0.0934 -0.1031 -0.0992 -0.0934
-0.0919 -0.0955 -0.0970 -0.1031 -0.0925 -0.0946 -0.0925 -0.0955 -0.0934 -0.0922 -0.0952 -0.0934 -0.0943 -0.0958 -0.0946 -0.0961 -0.0928 -0.0928 -0.0995 -0.0943 -0.0912
-0.0931 -0.0964 -0.0958 -0.1010 -0.0912 -0.0900 -0.0937 -0.0983 -0.0967 -0.0903 -0.0946 -0.0937 -0.0925 -0.0949 -0.0906 -0.0937 -0.0900 -0.0937 -0.0977 -0.0937 -0.0943
-0.0931 -0.0964 -0.0937 -0.0986 -0.0922 -0.0925 -0.0980 -0.1001 -0.0977 -0.0906 -0.0906 -0.0928 -0.0943 -0.0952 -0.0912 -0.0946 -0.0946 -0.0934 -0.0992 -0.0934 -0.0955
-0.0937 -0.0958 -0.0946 -0.0992 -0.0943 -0.0949 -0.0970 -0.0983 -0.0955 -0.0909 -0.0876 -0.0922 -0.0943 -0.0934 -0.0922 -0.0967 -0.0946 -0.0949 -0.1001 -0.0974 -0.0961
-0.0940 -0.0974 -0.0964 -0.1010 -0.0922 -0.0961 -0.0943 -0.0946 -0.0937 -0.0922 -0.0946 -0.0800 -0.0931 -0.0937 -0.0940 -0.0970 -0.0949 -0.0943 -0.0992 -0.0967 -0.0937
-0.0946 -0.0974 -0.0955 -0.1013 -0.0931 -0.0922 -0.0922 -0.0943 -0.0946 -0.0928 -0.0952 -0.0925 -0.0903 -0.0931 -0.0943 -0.0955 -0.0922 -0.0909 -0.0998 -0.0955 -0.0903
-0.0937 -0.0952 -0.0964 -0.1016 -0.0925 -0.0925 -0.0937 -0.0970 -0.0977 -0.0940 -0.0961 -0.0934 -0.0888 -0.0906 -0.0937 -0.0937 -0.0919 -0.0912 -0.0983 -0.0949 -0.0903
-0.0943 -0.0958 -0.0949 -0.1013 -0.0928 -0.0931 -0.0955 -0.0989 -0.0992 -0.0946 -0.0958 -0.0903 -0.0894 -0.0916 -0.0928 -0.0940 -0.0922 -0.0940 -0.0989 -0.0955 -0.0925
-0.0912 -0.0955 -0.0952 -0.1019 -0.0940 -0.0946 -0.0955 -0.0974 -0.0974 -0.0928 -0.0958 -0.0919 -0.0919 -0.0931 -0.0934 -0.0967 -0.0940 -0.0943 -0.1004 -0.0955 -0.0943
-0.0946 -0.0958 -0.0961 -0.1013 -0.0934 -0.0925 -0.0955 -0.0983 -0.0967 -0.0912 -0.0925 -0.0940 -0.0922 -0.0949 -0.0928 -0.0946 -0.0925 -0.0928 -0.0995 -0.0937 -0.0928
-0.0922 -0.0961 -0.0974 -0.1010 -0.0934 -0.0925 -0.0955 -0.0970 -0.0949 -0.0919 -0.0940 -0.0949 -0.0961 -0.0931 -0.0925 -0.0955 -0.0919 -0.0922 -0.0977 -0.0943 -0.0925
-0.0931 -0.0974 -0.0964 -0.1001 -0.0897 -0.0922 -0.0967 -0.0989 -0.0970 -0.0919 -0.0961 -0.0943 -0.0934 -0.0928 -0.0903 -0.0949 -0.0974 -0.0934 -0.0961 -0.0952 -0.0937
-0.0922 -0.0970 -0.0940 -0.0961 -0.0900 -0.0940 -0.0970 -0.0992 -0.0964 -0.0906 -0.0952 -0.0934 -0.0909 -0.0912 -0.0912 -0.0980 -0.0946 -0.0943 -0.0977 -0.0964 -0.0937
-0.0925 -0.0961 -0.0946 -0.1001 -0.0940 -0.0961 -0.0952 -0.0958 -0.0949 -0.0909 -0.0940 -0.0928 -0.0922 -0.0922 -0.0940 -0.0989 -0.0958 -0.0928 -0.1010 -0.0967 -0.0925
|
Attachment 6: ER-580-320-30kV-m40.txt
|
value (Voltage offset 40.0000)
80.0000
0.7000
80.0000
29978.0220
21
21
./emit.dat
-0.1270 -0.1038 -0.1025 -0.1031 -0.1031 -0.1056 -0.1031 -0.1068 -0.1019 -0.1047 -0.1044 -0.1016 -0.1010 -0.0998 -0.1022 -0.1013 -0.1016 -0.1041 -0.1135 -0.1053 -0.1050
-0.0970 -0.1013 -0.0986 -0.0943 -0.0946 -0.1013 -0.0998 -0.0952 -0.1004 -0.1041 -0.1004 -0.0983 -0.0980 -0.0974 -0.1016 -0.1019 -0.0974 -0.0983 -0.1065 -0.0998 -0.0983
-0.0983 -0.1016 -0.0967 -0.0943 -0.0919 -0.0928 -0.0952 -0.0992 -0.1010 -0.1013 -0.1004 -0.0989 -0.1004 -0.1031 -0.1007 -0.1004 -0.0992 -0.0964 -0.1028 -0.1004 -0.0986
-0.0986 -0.0992 -0.0974 -0.0946 -0.0900 -0.0861 -0.0879 -0.0952 -0.0964 -0.0970 -0.1013 -0.1004 -0.1007 -0.0922 -0.0992 -0.0992 -0.1041 -0.0986 -0.1041 -0.1013 -0.1004
-0.0980 -0.0974 -0.0989 -0.1022 -0.0922 -0.0854 -0.0821 -0.0848 -0.0851 -0.0922 -0.1004 -0.0986 -0.0992 -0.0986 -0.0940 -0.0964 -0.0977 -0.0986 -0.1053 -0.1031 -0.1016
-0.0946 -0.0980 -0.0961 -0.1007 -0.0928 -0.0821 -0.0760 -0.0720 -0.0650 -0.0775 -0.0912 -0.0958 -0.0952 -0.0949 -0.0964 -0.0961 -0.0970 -0.0986 -0.1038 -0.1038 -0.0980
-0.0916 -0.0961 -0.0967 -0.0958 -0.0928 -0.0879 -0.0751 -0.0592 -0.0430 -0.0607 -0.0766 -0.0839 -0.0937 -0.0958 -0.1007 -0.1010 -0.0974 -0.0974 -0.1016 -0.1010 -0.0955
-0.0934 -0.0949 -0.0928 -0.0928 -0.0891 -0.0842 -0.0745 -0.0583 -0.0256 -0.0363 -0.0644 -0.0754 -0.0815 -0.0919 -0.1007 -0.1019 -0.0974 -0.0992 -0.1019 -0.0998 -0.0967
-0.0983 -0.0940 -0.0916 -0.0949 -0.0876 -0.0803 -0.0711 -0.0565 -0.0339 -0.0024 -0.0488 -0.0717 -0.0821 -0.0903 -0.0909 -0.0949 -0.1004 -0.1004 -0.1031 -0.1013 -0.0983
-0.1007 -0.0961 -0.0967 -0.0986 -0.0916 -0.0821 -0.0726 -0.0571 -0.0397 0.0064 -0.0238 -0.0632 -0.0787 -0.0879 -0.0922 -0.0925 -0.0967 -0.0983 -0.1047 -0.1028 -0.0986
-0.1004 -0.0995 -0.0986 -0.1028 -0.0949 -0.0919 -0.0797 -0.0668 -0.0488 -0.0214 -0.0195 -0.0546 -0.0757 -0.0909 -0.0925 -0.0934 -0.0961 -0.0928 -0.1031 -0.1025 -0.1044
-0.0967 -0.1001 -0.0995 -0.0995 -0.0955 -0.0940 -0.0928 -0.0842 -0.0681 -0.0473 -0.0351 -0.0449 -0.0705 -0.0879 -0.0925 -0.0964 -0.0952 -0.0940 -0.0995 -0.1001 -0.0961
-0.0958 -0.1010 -0.0992 -0.0983 -0.0955 -0.0949 -0.0964 -0.0922 -0.0900 -0.0748 -0.0604 -0.0574 -0.0708 -0.0861 -0.0937 -0.0974 -0.0943 -0.0949 -0.1025 -0.1004 -0.0949
-0.0952 -0.0992 -0.0970 -0.1001 -0.0992 -0.0970 -0.0995 -0.0986 -0.0964 -0.0919 -0.0879 -0.0815 -0.0812 -0.0812 -0.0922 -0.0998 -0.0995 -0.0970 -0.1007 -0.1025 -0.0961
-0.0983 -0.0986 -0.0974 -0.1019 -0.1001 -0.1010 -0.1001 -0.0992 -0.1016 -0.0980 -0.0970 -0.0916 -0.0909 -0.0830 -0.0900 -0.0967 -0.0986 -0.1001 -0.1038 -0.1035 -0.0977
-0.0974 -0.0992 -0.0989 -0.1010 -0.0980 -0.0995 -0.0970 -0.0974 -0.1019 -0.0967 -0.0992 -0.0964 -0.0912 -0.0903 -0.0900 -0.0940 -0.0970 -0.0995 -0.1013 -0.1013 -0.0961
-0.0967 -0.1010 -0.0974 -0.0995 -0.0967 -0.0974 -0.0958 -0.0958 -0.1013 -0.0983 -0.0986 -0.0970 -0.0955 -0.0970 -0.0949 -0.0958 -0.0943 -0.0970 -0.1013 -0.0998 -0.0946
-0.0980 -0.0998 -0.0983 -0.1001 -0.0974 -0.0961 -0.0952 -0.0986 -0.0992 -0.0983 -0.0983 -0.0986 -0.0989 -0.0989 -0.1001 -0.0995 -0.0940 -0.0961 -0.1010 -0.1004 -0.0961
-0.0977 -0.0964 -0.0952 -0.1013 -0.0986 -0.0977 -0.0983 -0.0995 -0.1001 -0.0995 -0.1007 -0.0989 -0.0986 -0.1001 -0.0974 -0.0992 -0.0992 -0.0977 -0.1041 -0.1013 -0.0992
-0.0989 -0.0958 -0.0967 -0.1019 -0.0980 -0.0989 -0.0998 -0.0983 -0.0998 -0.0986 -0.0992 -0.0967 -0.0980 -0.0980 -0.0946 -0.0964 -0.0989 -0.0986 -0.1050 -0.1031 -0.1007
-0.0967 -0.0998 -0.0986 -0.0995 -0.0986 -0.0970 -0.0995 -0.0970 -0.0983 -0.0949 -0.0974 -0.0961 -0.0952 -0.0961 -0.0946 -0.0977 -0.0967 -0.0964 -0.1044 -0.1007 -0.0964
|
Attachment 7: TS-2017-01-18-11-02.rga
|
Attachment 8: TS-2017-01-18-11-40.rga
|
Attachment 9: TS-2017-01-20-08-28.rga
|
Attachment 10: TS-2017-01-21-11-01.rga
|
Attachment 11: Ta_Tri-coil_FEBIAD_Test_stand.docx
|
Attachment 12: ITE-Test_Ta-LP-FEBIAD-_TRI_COIL-_Test_Stand.docx
|
607
|
Friday, October 06, 2017, 12:14 |
Aaron Schmidt | SiC#37 | HP-FEBIAD | ITE | TM4 | 39530-10 | 41751-4 | In progress | Conditioning SiC#37 | NOTE* : The right coil terminal terminates into the heat shield. DO NOT TURN ON COIL.
Friday, October 06, 2017, 12:12: Roughing. * Note on installation. Middle TBHT prong 'stuck' to insulator slightly and was moved to fit. Fit ok in Jig.
Friday, October 06, 2017, 13:53: Turbo pumps on.
Friday, October 06, 2017, 14:00: Ion gauge, RGA on.
Tuesday, October 10, 2017, 07:45: Waterlines opened. No response on RGA. Vacuum ok @ 1.35e-07T.
Tuesday, October 10, 2017, 07:49: RGA Fatal communications error. Restart RGA.
Tuesday, October 10, 2017, 07:57: Power/sources on.
Tuesday, October 10, 2017, 08:17: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .
Tuesday, October 10, 2017, 08:57: BIAS / EZL / EE @ 0 kV. Power/sources off. Locked out*
Tuesday, October 10, 2017, 09:30: Install/swap gas bottles. Elog.
Tuesday, October 10, 2017, 10:36: Power on. Gas bottles opened. Await vacuum stabilization.
Tuesday, October 10, 2017, 10:40: Sources on.
Tuesday, October 10, 2017, 10:50: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .
Unable to bring Anode to voltage. Power supply was changed and does not 'Float'. Test TGHT & TBHT. Then switch from Cathode Negative Bias to Anode positive*.
Tuesday, October 10, 2017, 11:37: Auto ramping TGHT / TBHT to 20 / 20 A @ 4A per min. (TEST then cool)
Tuesday, October 10, 2017, 11:45: Auto-cooling
Tuesday, October 10, 2017, 11:57: Sources off / BIAS / EZL / EE @ 0 kV.
Wednesday, October 11, 2017, 10:25: Anode & PS switched*. Power/Sources on.
Wednesday, October 11, 2017, 10:38: BIAS / EZL / EE @ 30 / 13 / 1.5 kV.
Wednesday, October 11, 2017, 10:40: Coil @ 0 A. Anode 200 V / 7.326mA.
Wednesday, October 11, 2017, 10:45: Gas bottles opened. Await vacuum stabilization.
Wednesday, October 11, 2017, 10:47: IV4 Opened. Vacuum @ 2.43e-07T
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
24.6 |
3.78 |
TBHT |
24.6 |
2.75 |
COIL + EE |
24.4 |
4.84 |
TGHT |
24.6 |
5.44 |
Wednesday, October 11, 2017, 10:50: Auto ramping TGHT / TBHT to 580 / 290 A @ 2A per min.
Wednesday, October 11, 2017, 14:21: Check. Vacuum @ 2.99e-06T. Outgassing.
Wednesday, October 11, 2017, 14:42: Note* Slight discrepancy on the TBHT. Direct read out off the PS @ 296A. Source page: TBHT set to 290.0A, shows 302.2A
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
31.4 |
3.84 |
TBHT |
25.4 |
2.98 |
COIL + EE |
25.0 |
4.86 |
TGHT |
27.1 |
5.45 |
Thursday, October 12, 2017, 09:00. Mass scans ok. Spark @ 8:57.
Thursday, October 12, 2017, 10:48: Auto ramping TBHT to 310 A @ 2A per min.
Friday, October 13, 2017, 07:15: Anode short determined. Plotted. Occurred around 18:00 (6pm) yesterday.
Friday, October 13, 2017, 07:20: Auto-cooling @ 4A per min
Friday, October 13, 2017, 10:03: Sources off. Gas lines closed.
Friday, October 13, 2017, 10:06: RGA off. IV4 Closed. Reset IV4
Friday, October 13, 2017, 10:08: Ion gauge off. Turbos off.
Friday, October 13, 2017, 11:10: Venting.
Friday, October 13, 2017, 11:45: power off. Waterlines purged.
Friday, October 13, 2017, 11:49: Doors open, |
Attachment 1: TS-2017-10-10-07-55.rga
|
Attachment 2: TS-2017-10-06-14-57.rga
|
Attachment 3: SiC#37.docx
|
Attachment 4: ER-580-290-30kV-m23.txt
|
value (Voltage offset 0.0000)
80.0000
0.7000
100.0000
29978.0220
21
21
./emit.dat
-0.4483 -0.0751 -0.0757 -0.0751 -0.0729 -0.0732 -0.0696 -0.0650 -0.0528 0.0006 0.0024 -0.0165 -0.0446 -0.0583 -0.0677 -0.0693 -0.0671 -0.0677 -0.0699 -0.0726 -0.0732
-0.0720 -0.0696 -0.0702 -0.0708 -0.0720 -0.0763 -0.0742 -0.0699 -0.0708 -0.0452 -0.0064 -0.0085 -0.0354 -0.0525 -0.0620 -0.0626 -0.0626 -0.0656 -0.0671 -0.0677 -0.0665
0.0259 -0.0623 -0.0696 -0.0696 -0.0705 -0.0757 -0.0757 -0.0748 -0.0754 -0.0754 -0.0351 -0.0156 -0.0366 -0.0516 -0.0592 -0.0632 -0.0662 -0.0650 -0.0647 -0.0653 -0.0610
0.1025 0.1196 0.0458 -0.0378 -0.0757 -0.0766 -0.0787 -0.0790 -0.0772 -0.0739 -0.0693 -0.0476 -0.0485 -0.0494 -0.0616 -0.0616 -0.0635 -0.0665 -0.0629 -0.0623 -0.0632
-0.0568 0.2243 0.3433 0.1666 0.0699 -0.0650 -0.0848 -0.0879 -0.0861 -0.0778 -0.0760 -0.0726 -0.0610 -0.0525 -0.0607 -0.0647 -0.0647 -0.0659 -0.0641 -0.0644 -0.0647
-0.0653 -0.0513 0.4471 0.6042 0.4266 0.6778 0.3799 -0.0507 -0.0992 -0.0864 -0.0830 -0.0858 -0.0754 -0.0623 -0.0613 -0.0638 -0.0653 -0.0629 -0.0632 -0.0659 -0.0656
-0.0665 -0.0671 -0.0388 0.6653 1.1581 1.1627 1.5646 1.1548 0.7242 -0.0723 -0.0833 -0.0775 -0.0699 -0.0662 -0.0644 -0.0632 -0.0659 -0.0629 -0.0702 -0.0647 -0.0650
-0.0003 -0.0552 -0.0647 -0.0327 1.0379 1.8552 2.8146 2.4734 1.7310 1.2500 0.4538 -0.0674 -0.0671 -0.0623 -0.0665 -0.0674 -0.0662 -0.0623 -0.0647 -0.0659 -0.0638
0.4077 0.3668 0.3448 0.4446 0.5948 2.3346 3.7173 6.2762 4.0143 2.7844 1.7270 0.8627 0.2435 -0.0632 -0.0656 -0.0684 -0.0644 -0.0629 -0.0650 -0.0665 -0.0650
0.2783 0.6293 1.0846 1.4621 2.1027 3.2513 4.7601 6.8808 8.5223 4.3607 3.5660 1.6492 0.8087 0.2847 0.0327 -0.0662 -0.0626 -0.0616 -0.0656 -0.0662 -0.0647
-0.0458 0.1624 0.5487 1.0056 1.6989 2.4228 3.2785 4.7852 7.3224 5.2383 3.2251 1.1319 0.5197 0.3995 0.2014 0.0772 -0.0061 -0.0641 -0.0696 -0.0650 -0.0650
-0.0623 -0.0571 -0.0314 0.1779 0.6763 1.1481 1.9574 2.9968 4.6860 7.1210 2.5647 0.5713 0.0821 0.0027 0.0122 0.0421 0.0479 0.0089 -0.0516 -0.0638 -0.0659
-0.0601 -0.0607 -0.0641 -0.0739 -0.0226 0.4214 0.8472 1.2827 1.8497 3.7027 5.4932 1.3995 0.2875 0.0146 -0.0488 -0.0571 -0.0562 -0.0510 -0.0327 -0.0110 -0.0558
-0.0604 -0.0592 -0.0632 -0.0797 -0.0592 -0.0485 0.0153 0.4959 0.8826 1.3910 2.9550 3.6655 0.9778 0.1694 -0.0165 -0.0629 -0.0650 -0.0613 -0.0632 -0.0562 -0.0558
-0.0613 -0.0595 -0.0620 -0.0742 -0.0574 -0.0571 -0.0531 -0.0439 -0.0140 0.5255 1.1789 2.1106 1.8564 0.6610 0.0726 -0.0317 -0.0629 -0.0604 -0.0653 -0.0632 -0.0650
-0.0635 -0.0632 -0.0601 -0.0604 -0.0519 -0.0525 -0.0510 -0.0531 -0.0558 -0.0449 -0.0113 0.2933 0.6607 0.7449 0.3354 0.0275 -0.0537 -0.0616 -0.0644 -0.0635 -0.0659
-0.0647 -0.0604 -0.0638 -0.0632 -0.0443 -0.0439 -0.0375 -0.0430 -0.0479 -0.0546 -0.0543 -0.0555 -0.0504 0.0449 0.3394 0.1416 -0.0079 -0.0604 -0.0641 -0.0641 -0.0653
-0.0610 -0.0571 -0.0592 -0.0641 -0.0424 -0.0323 -0.0204 -0.0217 -0.0345 -0.0433 -0.0525 -0.0580 -0.0647 -0.0607 -0.0571 -0.0317 0.0150 -0.0391 -0.0644 -0.0635 -0.0635
-0.0555 -0.0516 -0.0525 -0.0568 -0.0372 -0.0143 0.0043 0.0156 -0.0061 -0.0272 -0.0394 -0.0580 -0.0595 -0.0662 -0.0653 -0.0623 -0.0604 -0.0626 -0.0659 -0.0638 -0.0613
-0.0540 -0.0513 -0.0491 -0.0488 -0.0241 0.0049 0.0296 0.0525 0.0427 -0.0040 -0.0211 -0.0494 -0.0607 -0.0623 -0.0629 -0.0607 -0.0616 -0.0635 -0.0732 -0.0616 -0.0647
-0.0571 -0.0519 -0.0427 -0.0510 -0.0137 0.0159 0.0476 0.0787 0.0891 0.0320 -0.0058 -0.0403 -0.0531 -0.0580 -0.0604 -0.0595 -0.0613 -0.0656 -0.0723 -0.0620 -0.0684
|
Attachment 5: ER-580-290-30kV-m28.txt
|
value (Voltage offset 0.0000)
80.0000
0.7000
100.0000
29978.0220
21
21
./emit.dat
-0.1282 -0.0800 -0.0717 -0.0702 -0.0705 -0.0699 -0.0690 -0.0702 -0.0699 -0.0351 -0.0140 -0.0296 -0.0507 -0.0610 -0.0668 -0.0677 -0.0684 -0.0681 -0.0745 -0.0726 -0.0717
-0.0699 -0.0748 -0.0656 -0.0641 -0.0671 -0.0668 -0.0644 -0.0638 -0.0699 -0.0656 -0.0214 -0.0247 -0.0476 -0.0589 -0.0610 -0.0623 -0.0647 -0.0653 -0.0702 -0.0687 -0.0690
-0.0357 -0.0717 -0.0674 -0.0684 -0.0711 -0.0677 -0.0656 -0.0656 -0.0684 -0.0708 -0.0439 -0.0330 -0.0516 -0.0595 -0.0598 -0.0623 -0.0647 -0.0665 -0.0647 -0.0659 -0.0644
-0.0412 0.0220 0.0235 -0.0403 -0.0690 -0.0687 -0.0717 -0.0726 -0.0711 -0.0650 -0.0677 -0.0546 -0.0531 -0.0549 -0.0604 -0.0613 -0.0650 -0.0632 -0.0620 -0.0632 -0.0696
-0.0687 -0.0037 0.1111 0.1181 0.1062 -0.0452 -0.0793 -0.0809 -0.0781 -0.0735 -0.0708 -0.0690 -0.0635 -0.0595 -0.0610 -0.0601 -0.0644 -0.0620 -0.0595 -0.0644 -0.0674
-0.0659 -0.0665 0.0455 0.2451 0.3592 0.4355 0.2582 -0.0067 -0.0882 -0.0809 -0.0751 -0.0705 -0.0677 -0.0653 -0.0626 -0.0616 -0.0647 -0.0601 -0.0616 -0.0650 -0.0684
-0.0653 -0.0674 -0.0616 0.0934 0.4333 0.8362 1.0184 0.7513 0.4086 -0.0378 -0.0739 -0.0684 -0.0674 -0.0644 -0.0665 -0.0635 -0.0629 -0.0632 -0.0641 -0.0656 -0.0653
-0.0586 -0.0656 -0.0638 -0.0665 0.2002 0.7745 1.6849 1.4682 0.9906 0.5087 0.2045 -0.0400 -0.0708 -0.0659 -0.0681 -0.0638 -0.0638 -0.0656 -0.0647 -0.0647 -0.0659
-0.0195 -0.0208 -0.0259 -0.0400 -0.0351 0.4218 1.4456 2.9980 2.0459 1.0242 0.3888 0.1514 0.0729 -0.0595 -0.0699 -0.0632 -0.0653 -0.0653 -0.0662 -0.0635 -0.0650
0.0217 0.0214 0.0552 0.0967 0.1715 0.2869 0.9854 3.1738 5.3613 2.0422 0.7935 0.2029 0.0650 0.0204 0.0153 -0.0613 -0.0629 -0.0653 -0.0729 -0.0632 -0.0665
-0.0531 0.0513 0.0540 0.0793 0.1791 0.3333 0.7407 1.7371 5.2942 3.7866 1.0818 0.2380 0.0311 0.0104 0.0009 -0.0113 -0.0110 -0.0659 -0.0641 -0.0635 -0.0662
-0.0690 -0.0616 -0.0299 0.0650 0.1306 0.1947 0.3656 0.7361 1.8848 5.6970 1.7746 0.5069 0.0656 -0.0485 -0.0485 -0.0412 -0.0375 -0.0284 -0.0421 -0.0641 -0.0681
-0.0659 -0.0638 -0.0629 -0.0610 -0.0171 0.1617 0.2716 0.4413 0.8774 2.1158 4.6158 1.0266 0.3339 0.0119 -0.0644 -0.0638 -0.0632 -0.0549 -0.0473 -0.0375 -0.0510
-0.0671 -0.0650 -0.0653 -0.0717 -0.0595 -0.0552 0.0220 0.2460 0.5835 0.9680 1.9342 2.9599 0.6287 0.2295 -0.0299 -0.0650 -0.0644 -0.0632 -0.0598 -0.0580 -0.0543
-0.0668 -0.0644 -0.0653 -0.0656 -0.0616 -0.0626 -0.0626 -0.0513 -0.0192 0.3293 0.8240 1.3589 1.2018 0.3604 0.1443 -0.0436 -0.0623 -0.0623 -0.0693 -0.0659 -0.0671
-0.0638 -0.0638 -0.0656 -0.0607 -0.0604 -0.0589 -0.0589 -0.0568 -0.0543 -0.0497 -0.0223 0.1840 0.3668 0.3961 0.1917 0.0891 -0.0522 -0.0641 -0.0635 -0.0629 -0.0656
-0.0668 -0.0641 -0.0613 -0.0693 -0.0558 -0.0522 -0.0519 -0.0510 -0.0543 -0.0558 -0.0555 -0.0562 -0.0571 0.0098 0.1544 0.0937 0.0403 -0.0644 -0.0729 -0.0632 -0.0690
-0.0699 -0.0653 -0.0595 -0.0653 -0.0568 -0.0482 -0.0433 -0.0430 -0.0494 -0.0534 -0.0549 -0.0677 -0.0644 -0.0647 -0.0601 -0.0476 0.0348 -0.0266 -0.0638 -0.0647 -0.0684
-0.0674 -0.0659 -0.0684 -0.0674 -0.0571 -0.0488 -0.0391 -0.0314 -0.0409 -0.0491 -0.0522 -0.0641 -0.0610 -0.0653 -0.0632 -0.0629 -0.0616 -0.0632 -0.0677 -0.0632 -0.0677
-0.0665 -0.0647 -0.0702 -0.0693 -0.0568 -0.0479 -0.0354 -0.0201 -0.0180 -0.0406 -0.0543 -0.0583 -0.0635 -0.0604 -0.0629 -0.0613 -0.0629 -0.0647 -0.0632 -0.0644 -0.0674
-0.0635 -0.0623 -0.0595 -0.0647 -0.0537 -0.0458 -0.0323 -0.0116 0.0067 -0.0241 -0.0494 -0.0580 -0.0659 -0.0604 -0.0626 -0.0616 -0.0626 -0.0616 -0.0665 -0.0632 -0.0677
|
Attachment 6: ER-580-290-30kV-m38.txt
|
value (Voltage offset 0.0000)
80.0000
0.7000
100.0000
29978.0220
21
21
./emit.dat
-0.1303 -0.0723 -0.0732 -0.0699 -0.0705 -0.0677 -0.0717 -0.0702 -0.0714 -0.0812 -0.0714 -0.0708 -0.0687 -0.0708 -0.0723 -0.0726 -0.0714 -0.0705 -0.0702 -0.0824 -0.0769
-0.0708 -0.0693 -0.0674 -0.0656 -0.0641 -0.0662 -0.0674 -0.0644 -0.0647 -0.0766 -0.0659 -0.0644 -0.0635 -0.0671 -0.0665 -0.0677 -0.0650 -0.0650 -0.0659 -0.0781 -0.0739
-0.0650 -0.0699 -0.0668 -0.0671 -0.0656 -0.0659 -0.0662 -0.0647 -0.0659 -0.0613 -0.0632 -0.0647 -0.0650 -0.0674 -0.0650 -0.0681 -0.0653 -0.0653 -0.0641 -0.0693 -0.0644
-0.0623 -0.0598 -0.0586 -0.0656 -0.0659 -0.0681 -0.0653 -0.0644 -0.0650 -0.0662 -0.0632 -0.0677 -0.0751 -0.0620 -0.0659 -0.0693 -0.0662 -0.0668 -0.0748 -0.0656 -0.0665
-0.0684 -0.0574 -0.0516 -0.0534 -0.0510 -0.0632 -0.0659 -0.0674 -0.0677 -0.0674 -0.0656 -0.0729 -0.0687 -0.0626 -0.0671 -0.0684 -0.0656 -0.0647 -0.0702 -0.0659 -0.0677
-0.0684 -0.0671 -0.0507 -0.0449 -0.0357 -0.0314 -0.0433 -0.0629 -0.0684 -0.0653 -0.0659 -0.0705 -0.0693 -0.0629 -0.0662 -0.0684 -0.0641 -0.0662 -0.0693 -0.0635 -0.0626
-0.0681 -0.0662 -0.0616 -0.0473 -0.0330 -0.0006 0.0153 0.0040 -0.0269 -0.0620 -0.0665 -0.0702 -0.0635 -0.0629 -0.0653 -0.0668 -0.0623 -0.0653 -0.0684 -0.0644 -0.0623
-0.0677 -0.0644 -0.0644 -0.0662 -0.0351 0.0052 0.0427 0.0522 0.0061 -0.0073 -0.0336 -0.0647 -0.0656 -0.0641 -0.0650 -0.0635 -0.0656 -0.0659 -0.0778 -0.0653 -0.0662
-0.0632 -0.0586 -0.0604 -0.0601 -0.0525 -0.0101 0.0418 0.1007 0.0568 0.0046 -0.0189 -0.0372 -0.0455 -0.0644 -0.0659 -0.0644 -0.0665 -0.0668 -0.0742 -0.0653 -0.0681
-0.0580 -0.0510 -0.0513 -0.0476 -0.0375 -0.0101 0.0281 0.0787 0.1413 0.0485 0.0183 -0.0204 -0.0436 -0.0522 -0.0537 -0.0662 -0.0668 -0.0671 -0.0696 -0.0662 -0.0671
-0.0671 -0.0436 -0.0491 -0.0461 -0.0360 -0.0281 0.0006 0.0244 0.0888 0.1120 0.0461 -0.0046 -0.0485 -0.0543 -0.0546 -0.0577 -0.0586 -0.0629 -0.0711 -0.0650 -0.0671
-0.0671 -0.0665 -0.0604 -0.0381 -0.0385 -0.0400 -0.0269 0.0040 0.0412 0.1096 0.0479 -0.0089 -0.0372 -0.0623 -0.0629 -0.0586 -0.0562 -0.0568 -0.0690 -0.0641 -0.0659
-0.0677 -0.0659 -0.0641 -0.0684 -0.0610 -0.0296 -0.0299 -0.0229 -0.0061 0.0369 0.1013 -0.0040 -0.0150 -0.0507 -0.0641 -0.0620 -0.0626 -0.0623 -0.0613 -0.0601 -0.0629
-0.0674 -0.0659 -0.0659 -0.0681 -0.0656 -0.0647 -0.0580 -0.0299 -0.0064 0.0012 0.0311 0.0531 -0.0153 -0.0424 -0.0598 -0.0638 -0.0644 -0.0656 -0.0720 -0.0644 -0.0616
-0.0677 -0.0674 -0.0647 -0.0705 -0.0662 -0.0653 -0.0656 -0.0656 -0.0610 -0.0287 -0.0037 0.0067 0.0049 -0.0296 -0.0537 -0.0604 -0.0641 -0.0647 -0.0742 -0.0641 -0.0644
-0.0677 -0.0641 -0.0632 -0.0665 -0.0644 -0.0644 -0.0662 -0.0662 -0.0629 -0.0656 -0.0620 -0.0464 -0.0320 -0.0311 -0.0485 -0.0546 -0.0616 -0.0650 -0.0693 -0.0620 -0.0638
-0.0684 -0.0620 -0.0650 -0.0690 -0.0623 -0.0626 -0.0653 -0.0644 -0.0610 -0.0647 -0.0641 -0.0616 -0.0653 -0.0580 -0.0500 -0.0519 -0.0577 -0.0629 -0.0748 -0.0626 -0.0665
-0.0687 -0.0653 -0.0668 -0.0656 -0.0641 -0.0626 -0.0623 -0.0623 -0.0653 -0.0613 -0.0623 -0.0665 -0.0662 -0.0662 -0.0629 -0.0626 -0.0601 -0.0586 -0.0717 -0.0650 -0.0681
-0.0690 -0.0656 -0.0674 -0.0681 -0.0632 -0.0635 -0.0626 -0.0635 -0.0647 -0.0623 -0.0610 -0.0668 -0.0690 -0.0659 -0.0647 -0.0671 -0.0653 -0.0650 -0.0696 -0.0644 -0.0696
-0.0684 -0.0665 -0.0656 -0.0647 -0.0635 -0.0653 -0.0635 -0.0623 -0.0638 -0.0641 -0.0629 -0.0717 -0.0650 -0.0653 -0.0668 -0.0671 -0.0653 -0.0647 -0.0726 -0.0644 -0.0650
-0.0653 -0.0653 -0.0632 -0.0644 -0.0629 -0.0626 -0.0644 -0.0635 -0.0641 -0.0623 -0.0629 -0.0693 -0.0644 -0.0632 -0.0662 -0.0656 -0.0638 -0.0632 -0.0690 -0.0635 -0.0629
|
Attachment 7: ER-580-310-30kV-m23.txt
|
value (Voltage offset 0.0000)
80.0000
0.7000
100.0000
29978.0220
21
21
./emit.dat
-0.0967 -0.1022 -0.0934 -0.0641 -0.0513 -0.0598 -0.0418 -0.0223 -0.0104 0.0793 0.1715 0.1703 0.1410 0.0769 0.0122 -0.0241 -0.0247 -0.0317 -0.0497 -0.0620 -0.0629
-0.0800 -0.1086 -0.1105 -0.0851 -0.0684 -0.0739 -0.0562 -0.0308 -0.0253 -0.0073 0.1077 0.1431 0.1120 0.0626 0.0122 -0.0287 -0.0311 -0.0375 -0.0528 -0.0592 -0.0583
1.0831 -0.0116 -0.1114 -0.1022 -0.1007 -0.1099 -0.0873 -0.0616 -0.0482 -0.0412 0.0174 0.0882 0.0912 0.0381 0.0089 -0.0308 -0.0354 -0.0436 -0.0607 -0.0586 -0.0586
1.0553 2.6859 2.1231 0.4813 -0.1233 -0.1447 -0.1276 -0.1050 -0.0854 -0.0644 -0.0415 -0.0024 0.0558 0.0214 -0.0021 -0.0241 -0.0299 -0.0479 -0.0723 -0.0650 -0.0623
-0.0415 2.5452 3.1885 3.1613 2.6624 0.1715 -0.1715 -0.1477 -0.1291 -0.1089 -0.0784 -0.0601 -0.0159 -0.0037 -0.0247 -0.0327 -0.0262 -0.0427 -0.0668 -0.0662 -0.0668
-0.0793 0.1428 3.9502 3.8171 4.2780 5.6683 4.5184 1.0385 -0.1633 -0.1303 -0.1108 -0.0952 -0.0851 -0.0476 -0.0443 -0.0555 -0.0278 -0.0372 -0.0668 -0.0677 -0.0671
-0.0580 -0.0757 0.3665 4.5935 4.5792 7.0300 6.0605 5.5484 4.7726 0.4169 -0.1202 -0.1141 -0.1080 -0.0873 -0.0726 -0.0729 -0.0479 -0.0400 -0.0626 -0.0632 -0.0659
0.7700 0.4062 0.3680 1.1285 4.3320 5.5103 9.8682 7.5186 6.6586 4.8758 4.5834 0.3418 -0.0940 -0.0833 -0.0925 -0.0943 -0.0696 -0.0571 -0.0647 -0.0656 -0.0616
1.2888 1.2250 1.2610 1.3541 2.4429 4.5264 7.2159 9.9997 7.9214 5.3458 4.8840 3.8849 3.7399 0.0433 -0.0739 -0.0845 -0.0644 -0.0632 -0.0610 -0.0668 -0.0638
2.2534 1.4795 1.5872 1.6635 1.9397 4.9500 6.4972 9.9997 9.9997 6.9727 5.4489 4.9753 2.8958 2.3810 2.0789 -0.0809 -0.0613 -0.0659 -0.0693 -0.0647 -0.0656
0.7089 3.3536 2.4023 1.7398 1.9894 2.5311 4.1031 6.3916 9.9997 9.9997 7.4921 5.8585 5.2655 2.5928 1.9415 1.4978 1.5005 -0.0638 -0.0711 -0.0623 -0.0687
-0.0433 0.0089 0.9427 3.0133 3.0850 2.3279 2.7740 3.8568 5.9601 9.9997 9.8004 6.1935 5.5389 2.8302 2.0892 1.8539 1.4917 1.2170 0.9894 -0.0613 -0.0659
-0.0467 -0.0479 -0.0284 0.0064 1.0059 3.3408 3.8303 3.7759 4.1678 6.2854 9.9997 8.2861 3.9346 3.9020 0.4007 0.5817 0.9482 1.1398 1.1365 1.1145 0.5487
-0.0458 -0.0363 -0.0278 -0.0470 -0.0284 0.0668 1.2939 3.3481 4.2123 5.6619 7.0963 9.9997 6.1972 3.0945 3.0719 -0.0452 -0.0195 0.0439 0.4980 0.7767 0.7108
-0.0464 -0.0342 -0.0070 -0.0223 -0.0250 -0.0363 -0.0192 0.0479 0.4440 3.6618 4.7522 6.2686 9.0561 4.1541 2.6917 2.5610 -0.0510 -0.0568 -0.0540 -0.0058 0.2258
-0.0504 -0.0357 0.0018 0.0116 -0.0024 0.0031 -0.0116 -0.0134 -0.0177 0.0510 0.2686 3.2034 5.1953 5.5859 3.4576 2.3410 2.1951 -0.0565 -0.0546 -0.0592 -0.0568
-0.0497 -0.0339 0.0058 0.0293 0.0253 0.0375 0.0330 0.0189 0.0064 -0.0165 -0.0284 0.0198 0.0809 0.9250 2.6086 2.2415 1.9977 0.8820 -0.0623 -0.0607 -0.0577
-0.0427 -0.0317 0.0037 0.0281 0.0461 0.0629 0.0845 0.0763 0.0415 0.0061 -0.0119 -0.0317 -0.0360 -0.0354 -0.0021 0.1132 1.5292 1.7456 0.0690 -0.0610 -0.0650
-0.0339 -0.0275 0.0085 0.0323 0.0607 0.0861 0.1212 0.1465 0.0946 0.0360 0.0140 -0.0134 -0.0333 -0.0467 -0.0598 -0.0720 -0.0339 -0.0067 0.3729 -0.0500 -0.0641
-0.0201 -0.0177 0.0168 0.0345 0.0662 0.1044 0.1520 0.1990 0.1794 0.0812 0.0510 0.0168 -0.0143 -0.0354 -0.0543 -0.0711 -0.0601 -0.0626 -0.0641 -0.0635 -0.0690
-0.0150 -0.0049 0.0311 0.0461 0.0714 0.1181 0.1700 0.2310 0.2481 0.1517 0.0903 0.0519 0.0192 -0.0165 -0.0412 -0.0616 -0.0543 -0.0601 -0.0662 -0.0620 -0.0629
|
Attachment 8: ER-580-310-30kV-m38.txt
|
value (Voltage offset 0.0000)
80.0000
0.7000
100.0000
29978.0220
21
21
./emit.dat
-0.1071 -0.0974 -0.0732 -0.0662 -0.0607 -0.0650 -0.0604 -0.0455 -0.0439 -0.0482 -0.0308 -0.0125 -0.0076 -0.0146 -0.0360 -0.0699 -0.0647 -0.0595 -0.0665 -0.0729 -0.0806
-0.0806 -0.0967 -0.0803 -0.0729 -0.0665 -0.0693 -0.0620 -0.0473 -0.0482 -0.0601 -0.0458 -0.0208 -0.0146 -0.0125 -0.0281 -0.0577 -0.0629 -0.0601 -0.0665 -0.0677 -0.0778
0.9167 -0.0537 -0.0800 -0.0867 -0.0815 -0.0784 -0.0729 -0.0601 -0.0558 -0.0531 -0.0571 -0.0302 -0.0177 -0.0259 -0.0269 -0.0531 -0.0607 -0.0555 -0.0668 -0.0635 -0.0635
1.1377 0.8154 0.9314 0.4590 -0.0925 -0.0952 -0.0894 -0.0760 -0.0638 -0.0629 -0.0595 -0.0516 -0.0320 -0.0259 -0.0320 -0.0519 -0.0577 -0.0510 -0.0644 -0.0626 -0.0662
-0.0558 1.1316 1.2906 2.2522 2.0361 0.1923 -0.1093 -0.0928 -0.0836 -0.0818 -0.0714 -0.0632 -0.0549 -0.0363 -0.0430 -0.0558 -0.0562 -0.0510 -0.0681 -0.0632 -0.0650
-0.0806 -0.0467 1.2347 2.5235 2.4680 2.3276 1.8118 0.5157 -0.0937 -0.0928 -0.0897 -0.0769 -0.0757 -0.0607 -0.0565 -0.0668 -0.0598 -0.0525 -0.0601 -0.0626 -0.0653
-0.0650 -0.0821 -0.0211 1.2958 2.1375 2.5375 2.3993 2.0953 1.8317 0.2002 -0.0946 -0.0867 -0.0906 -0.0769 -0.0729 -0.0784 -0.0702 -0.0555 -0.0681 -0.0604 -0.0677
0.0867 0.0488 0.0455 0.1535 1.4014 2.3450 2.9492 2.4246 2.0340 1.7654 1.5717 0.1254 -0.0809 -0.0775 -0.0806 -0.0903 -0.0812 -0.0644 -0.0687 -0.0623 -0.0653
0.2988 0.2176 0.1981 0.1779 0.3204 1.3181 2.6715 3.0154 2.7887 2.1344 1.9815 1.9897 1.7218 0.0278 -0.0696 -0.0845 -0.0797 -0.0656 -0.0641 -0.0610 -0.0681
1.0544 0.5304 0.4434 0.4306 0.3915 0.5493 1.4310 3.1207 3.3359 2.9495 2.5803 2.3004 1.2375 1.1011 1.3828 -0.0803 -0.0790 -0.0665 -0.0763 -0.0635 -0.0653
0.5692 1.4536 1.1288 0.6842 0.6494 0.6287 1.1530 1.6870 3.0573 3.7460 3.4006 2.7585 1.6827 0.8313 0.8008 0.8655 1.1084 -0.0671 -0.0732 -0.0653 -0.0665
-0.0558 -0.0330 1.0129 1.4017 1.4478 0.8286 0.9335 1.7447 1.8158 3.1241 3.6096 2.5269 1.8466 0.9354 0.4218 0.4636 0.4962 0.6219 0.7980 -0.0650 -0.0653
-0.0598 -0.0687 -0.0433 -0.0259 1.1697 1.3602 1.5585 1.4749 1.5070 1.7593 2.8372 3.2993 2.0493 1.8457 0.3799 0.1254 0.1379 0.2139 0.4285 0.6036 0.5618
-0.0595 -0.0665 -0.0507 -0.0473 -0.0342 0.0137 1.4301 1.4197 1.4612 1.4874 1.8011 2.5586 2.9877 1.7834 1.6428 0.1089 -0.0424 0.0119 0.0824 0.1895 0.3391
-0.0610 -0.0693 -0.0455 -0.0427 -0.0458 -0.0565 -0.0391 -0.0131 0.3320 1.4764 1.5033 1.9058 2.7496 2.6636 1.5988 1.3913 0.0638 -0.0620 -0.0601 -0.0159 0.0433
-0.0638 -0.0668 -0.0427 -0.0357 -0.0351 -0.0470 -0.0525 -0.0485 -0.0494 -0.0244 0.0995 1.4993 1.9073 2.4872 2.1170 1.4285 1.0736 0.0995 -0.0745 -0.0610 -0.0620
-0.0632 -0.0681 -0.0449 -0.0409 -0.0287 -0.0357 -0.0406 -0.0397 -0.0452 -0.0513 -0.0555 -0.0317 -0.0156 0.6174 1.7032 1.5125 1.1856 0.8423 -0.0668 -0.0623 -0.0681
-0.0592 -0.0638 -0.0439 -0.0430 -0.0238 -0.0284 -0.0281 -0.0256 -0.0360 -0.0436 -0.0513 -0.0528 -0.0656 -0.0479 -0.0269 0.0604 1.1856 1.0800 0.6958 -0.0568 -0.0653
-0.0589 -0.0653 -0.0464 -0.0317 -0.0204 -0.0204 -0.0198 -0.0165 -0.0238 -0.0348 -0.0424 -0.0443 -0.0500 -0.0528 -0.0635 -0.0775 -0.0531 -0.0299 0.4526 0.3867 -0.0638
-0.0546 -0.0623 -0.0430 -0.0385 -0.0223 -0.0189 -0.0153 -0.0073 -0.0098 -0.0211 -0.0293 -0.0311 -0.0595 -0.0507 -0.0607 -0.0772 -0.0723 -0.0647 -0.0653 -0.0500 -0.0494
-0.0504 -0.0574 -0.0378 -0.0226 -0.0201 -0.0192 -0.0128 -0.0009 0.0000 -0.0061 -0.0171 -0.0269 -0.0424 -0.0482 -0.0574 -0.0748 -0.0693 -0.0632 -0.0751 -0.0610 -0.0647
|
Attachment 9: ER-580-310-30kV-m202.txt
|
value (Voltage offset 0.0000)
80.0000
0.7000
100.0000
29978.0220
21
21
./emit.dat
-0.1477 -0.0919 -0.0912 -0.0732 -0.0742 -0.0897 -0.0873 -0.0751 -0.0708 -0.0821 -0.0836 -0.0742 -0.0772 -0.0760 -0.0876 -0.0943 -0.0851 -0.0748 -0.0854 -0.0824 -0.0876
-0.0864 -0.0851 -0.0848 -0.0677 -0.0681 -0.0836 -0.0803 -0.0671 -0.0653 -0.0803 -0.0809 -0.0684 -0.0726 -0.0708 -0.0818 -0.0827 -0.0793 -0.0702 -0.0830 -0.0784 -0.0824
-0.0854 -0.0867 -0.0757 -0.0677 -0.0659 -0.0806 -0.0793 -0.0650 -0.0671 -0.0674 -0.0735 -0.0656 -0.0696 -0.0684 -0.0803 -0.0842 -0.0790 -0.0656 -0.0754 -0.0677 -0.0644
-0.0858 -0.0854 -0.0684 -0.0653 -0.0626 -0.0793 -0.0800 -0.0674 -0.0674 -0.0674 -0.0668 -0.0665 -0.0729 -0.0653 -0.0821 -0.0848 -0.0784 -0.0702 -0.0739 -0.0650 -0.0647
-0.0879 -0.0879 -0.0656 -0.0723 -0.0626 -0.0793 -0.0815 -0.0656 -0.0638 -0.0681 -0.0684 -0.0671 -0.0674 -0.0650 -0.0803 -0.0858 -0.0784 -0.0693 -0.0769 -0.0632 -0.0647
-0.0845 -0.0842 -0.0647 -0.0650 -0.0659 -0.0812 -0.0812 -0.0659 -0.0641 -0.0653 -0.0629 -0.0653 -0.0644 -0.0647 -0.0787 -0.0845 -0.0754 -0.0674 -0.0653 -0.0644 -0.0677
-0.0827 -0.0861 -0.0641 -0.0659 -0.0677 -0.0809 -0.0778 -0.0644 -0.0653 -0.0662 -0.0638 -0.0641 -0.0641 -0.0641 -0.0793 -0.0836 -0.0751 -0.0681 -0.0690 -0.0647 -0.0684
-0.0854 -0.0879 -0.0665 -0.0729 -0.0650 -0.0775 -0.0769 -0.0623 -0.0616 -0.0671 -0.0659 -0.0653 -0.0644 -0.0635 -0.0790 -0.0858 -0.0742 -0.0671 -0.0754 -0.0665 -0.0653
-0.0848 -0.0858 -0.0668 -0.0681 -0.0635 -0.0772 -0.0769 -0.0632 -0.0586 -0.0653 -0.0665 -0.0693 -0.0659 -0.0644 -0.0812 -0.0854 -0.0714 -0.0687 -0.0674 -0.0644 -0.0665
-0.0845 -0.0851 -0.0647 -0.0656 -0.0656 -0.0793 -0.0800 -0.0644 -0.0577 -0.0641 -0.0665 -0.0647 -0.0656 -0.0674 -0.0806 -0.0827 -0.0702 -0.0684 -0.0772 -0.0647 -0.0665
-0.0833 -0.0861 -0.0650 -0.0662 -0.0659 -0.0781 -0.0781 -0.0641 -0.0583 -0.0613 -0.0610 -0.0644 -0.0641 -0.0659 -0.0793 -0.0836 -0.0705 -0.0665 -0.0763 -0.0632 -0.0650
-0.0842 -0.0845 -0.0659 -0.0723 -0.0650 -0.0800 -0.0806 -0.0659 -0.0641 -0.0616 -0.0577 -0.0607 -0.0650 -0.0644 -0.0775 -0.0845 -0.0699 -0.0653 -0.0726 -0.0632 -0.0641
-0.0851 -0.0858 -0.0656 -0.0671 -0.0635 -0.0772 -0.0815 -0.0656 -0.0650 -0.0632 -0.0595 -0.0616 -0.0656 -0.0647 -0.0787 -0.0851 -0.0717 -0.0668 -0.0806 -0.0644 -0.0665
-0.0858 -0.0848 -0.0656 -0.0629 -0.0632 -0.0787 -0.0800 -0.0629 -0.0629 -0.0677 -0.0620 -0.0632 -0.0613 -0.0638 -0.0793 -0.0848 -0.0739 -0.0690 -0.0681 -0.0653 -0.0647
-0.0848 -0.0833 -0.0629 -0.0720 -0.0650 -0.0815 -0.0790 -0.0632 -0.0632 -0.0656 -0.0632 -0.0635 -0.0647 -0.0601 -0.0772 -0.0851 -0.0711 -0.0668 -0.0653 -0.0641 -0.0632
-0.0839 -0.0861 -0.0623 -0.0726 -0.0662 -0.0800 -0.0793 -0.0650 -0.0635 -0.0650 -0.0632 -0.0644 -0.0665 -0.0607 -0.0748 -0.0842 -0.0690 -0.0671 -0.0729 -0.0644 -0.0677
-0.0830 -0.0870 -0.0638 -0.0693 -0.0656 -0.0784 -0.0803 -0.0659 -0.0620 -0.0656 -0.0644 -0.0644 -0.0644 -0.0620 -0.0772 -0.0854 -0.0705 -0.0659 -0.0684 -0.0653 -0.0681
-0.0836 -0.0864 -0.0659 -0.0717 -0.0638 -0.0775 -0.0818 -0.0659 -0.0641 -0.0653 -0.0671 -0.0656 -0.0644 -0.0641 -0.0784 -0.0861 -0.0677 -0.0647 -0.0793 -0.0662 -0.0620
-0.0824 -0.0864 -0.0668 -0.0650 -0.0653 -0.0793 -0.0815 -0.0632 -0.0653 -0.0684 -0.0656 -0.0647 -0.0656 -0.0662 -0.0797 -0.0848 -0.0693 -0.0668 -0.0739 -0.0659 -0.0659
-0.0836 -0.0848 -0.0644 -0.0668 -0.0647 -0.0790 -0.0809 -0.0626 -0.0641 -0.0690 -0.0635 -0.0644 -0.0662 -0.0647 -0.0760 -0.0833 -0.0720 -0.0693 -0.0754 -0.0635 -0.0644
-0.0824 -0.0824 -0.0647 -0.0720 -0.0656 -0.0800 -0.0800 -0.0650 -0.0653 -0.0635 -0.0626 -0.0659 -0.0644 -0.0641 -0.0754 -0.0827 -0.0681 -0.0668 -0.0720 -0.0613 -0.0659
|
610
|
Friday, October 13, 2017, 15:53 |
Aaron Schmidt | SiC#37 | HP-FEBIAD | ITE | TM4 | 39530-10 | 41751-4 | In progress | Conditioning SiC#37 (ii) | Previous short fixed. Grid clamp removed. Grid tack welded.
Friday, October 13, 2017, 15:49: Roughing.
By Anders
Friday, October 13, 2017, 16:10: TPs started
Friday, October 13, 2017, 16:20: Vacuum at 2.7E-7 T
Saturday, October 14, 2017, 06:37: Opened water supply to HS. 95/10 psi. Vacuum stable at 5.7E-7 T
Saturday, October 14, 2017, 06:42: Opened gas lines.
Saturday, October 14, 2017, 06:43: Started heater ramping to 580 / 290 A
Saturday, October 14, 2017, 06:45: BIAS/EZL/EE to 30 / 13.5 / 1.5 kV. EE current somewhat unstable around 500 uA - seems related to elevated BIAS current. Improving as BIAS current is gradually coming down.
Saturday, October 14, 2017, 15:19: TBHT to 310 A. Local readback 316 A
Sunday, October 15, 2017, 17:36 : Ramping down heaters.
Monday, October 16, 2017, 09:10: Sources off.
Monday, October 16, 2017, 09:13; IV4 Closed & Reset. FEBIAD Gas lines closed.
Monday, October 16, 2017, 09:15: RGA, Ion gauge, turbos off.
Monday, October 16, 2017, 10:40: Power off. Waterlines purged.
Monday, October 16, 2017, 10:43: Venting. |
Attachment 1: SiC#37_(iii).docx
|
Attachment 2: TS-2017-10-13-17-44.rga
|
Attachment 3: TS-2017-10-14-06-35.rga
|
Attachment 4: ER-580-310-30kV-m20.txt
|
value (Voltage offset 0.0000)
76.0000
0.9000
100.0000
29978.0220
21
21
./emit.dat
-0.1251 -0.0882 -0.0931 -0.0888 -0.0848 -0.0879 -0.0848 -0.0702 -0.0754 -0.0833 -0.0699 -0.0717 -0.0760 -0.0717 -0.0763 -0.0766 -0.0870 -0.0797 -0.0772 -0.0873 -0.0861
-0.0650 -0.0851 -0.0873 -0.0824 -0.0797 -0.0830 -0.0793 -0.0644 -0.0723 -0.0766 -0.0653 -0.0674 -0.0739 -0.0650 -0.0720 -0.0754 -0.0812 -0.0751 -0.0739 -0.0818 -0.0800
-0.0693 -0.0766 -0.0864 -0.0766 -0.0638 -0.0671 -0.0586 -0.0702 -0.0674 -0.0732 -0.0641 -0.0629 -0.0665 -0.0613 -0.0613 -0.0699 -0.0790 -0.0702 -0.0714 -0.0787 -0.0659
-0.0693 -0.0757 -0.0873 -0.0787 -0.0629 -0.0662 -0.0647 -0.0659 -0.0650 -0.0748 -0.0638 -0.0629 -0.0659 -0.0623 -0.0656 -0.0671 -0.0806 -0.0690 -0.0702 -0.0784 -0.0641
-0.0705 -0.0732 -0.0885 -0.0769 -0.0659 -0.0635 -0.0644 -0.0711 -0.0635 -0.0769 -0.0659 -0.0604 -0.0644 -0.0632 -0.0629 -0.0671 -0.0818 -0.0702 -0.0662 -0.0790 -0.0659
-0.0726 -0.0757 -0.0864 -0.0772 -0.0656 -0.0653 -0.0665 -0.0665 -0.0626 -0.0748 -0.0653 -0.0650 -0.0653 -0.0659 -0.0638 -0.0702 -0.0824 -0.0708 -0.0690 -0.0793 -0.0705
-0.0653 -0.0775 -0.0870 -0.0757 -0.0638 -0.0690 -0.0638 -0.0705 -0.0607 -0.0720 -0.0626 -0.0653 -0.0668 -0.0638 -0.0647 -0.0693 -0.0809 -0.0693 -0.0699 -0.0784 -0.0696
-0.0668 -0.0775 -0.0861 -0.0748 -0.0644 -0.0699 -0.0610 -0.0674 -0.0616 -0.0723 -0.0598 -0.0623 -0.0635 -0.0638 -0.0638 -0.0705 -0.0806 -0.0702 -0.0708 -0.0797 -0.0671
-0.0632 -0.0760 -0.0864 -0.0745 -0.0638 -0.0668 -0.0616 -0.0650 -0.0601 -0.0739 -0.0571 -0.0595 -0.0616 -0.0638 -0.0613 -0.0681 -0.0806 -0.0705 -0.0708 -0.0800 -0.0671
-0.0677 -0.0772 -0.0861 -0.0769 -0.0616 -0.0668 -0.0647 -0.0681 -0.0620 -0.0754 -0.0558 -0.0546 -0.0613 -0.0641 -0.0635 -0.0687 -0.0815 -0.0714 -0.0677 -0.0812 -0.0690
-0.0681 -0.0763 -0.0845 -0.0751 -0.0635 -0.0671 -0.0644 -0.0653 -0.0647 -0.0739 -0.0552 -0.0510 -0.0595 -0.0601 -0.0653 -0.0717 -0.0821 -0.0711 -0.0681 -0.0790 -0.0714
-0.0650 -0.0775 -0.0839 -0.0745 -0.0653 -0.0665 -0.0644 -0.0693 -0.0635 -0.0745 -0.0613 -0.0485 -0.0568 -0.0583 -0.0623 -0.0681 -0.0821 -0.0711 -0.0693 -0.0775 -0.0668
-0.0632 -0.0766 -0.0845 -0.0742 -0.0647 -0.0656 -0.0629 -0.0671 -0.0644 -0.0751 -0.0626 -0.0537 -0.0537 -0.0580 -0.0620 -0.0668 -0.0836 -0.0687 -0.0720 -0.0775 -0.0616
-0.0717 -0.0742 -0.0873 -0.0717 -0.0632 -0.0653 -0.0668 -0.0778 -0.0620 -0.0775 -0.0613 -0.0580 -0.0546 -0.0568 -0.0604 -0.0671 -0.0824 -0.0702 -0.0708 -0.0790 -0.0620
-0.0699 -0.0754 -0.0842 -0.0745 -0.0632 -0.0656 -0.0665 -0.0748 -0.0623 -0.0772 -0.0632 -0.0650 -0.0577 -0.0592 -0.0595 -0.0690 -0.0824 -0.0702 -0.0677 -0.0800 -0.0620
-0.0726 -0.0787 -0.0851 -0.0751 -0.0638 -0.0668 -0.0623 -0.0620 -0.0629 -0.0778 -0.0641 -0.0671 -0.0626 -0.0616 -0.0601 -0.0668 -0.0815 -0.0720 -0.0681 -0.0797 -0.0650
-0.0690 -0.0763 -0.0845 -0.0732 -0.0632 -0.0656 -0.0601 -0.0626 -0.0635 -0.0778 -0.0632 -0.0638 -0.0616 -0.0610 -0.0629 -0.0677 -0.0803 -0.0705 -0.0681 -0.0787 -0.0616
-0.0662 -0.0766 -0.0851 -0.0726 -0.0635 -0.0659 -0.0626 -0.0623 -0.0629 -0.0760 -0.0623 -0.0616 -0.0638 -0.0641 -0.0620 -0.0693 -0.0815 -0.0696 -0.0696 -0.0803 -0.0659
-0.0671 -0.0760 -0.0864 -0.0742 -0.0638 -0.0662 -0.0629 -0.0711 -0.0656 -0.0775 -0.0620 -0.0607 -0.0641 -0.0650 -0.0635 -0.0674 -0.0812 -0.0693 -0.0684 -0.0800 -0.0662
-0.0662 -0.0763 -0.0858 -0.0723 -0.0635 -0.0653 -0.0653 -0.0754 -0.0632 -0.0760 -0.0635 -0.0620 -0.0650 -0.0623 -0.0647 -0.0674 -0.0833 -0.0690 -0.0681 -0.0790 -0.0662
-0.0690 -0.0748 -0.0854 -0.0742 -0.0638 -0.0641 -0.0641 -0.0623 -0.0620 -0.0763 -0.0644 -0.0635 -0.0635 -0.0604 -0.0644 -0.0684 -0.0848 -0.0674 -0.0690 -0.0745 -0.0641
|
Attachment 5: ER-580-310-30kV-m23.txt
|
value (Voltage offset 0.0000)
76.0000
0.9000
100.0000
29978.0220
21
21
./emit.dat
-0.1132 -0.0763 -0.0882 -0.0815 -0.0909 -0.1007 -0.0751 -0.0522 -0.0201 0.0049 0.0455 0.2124 0.4019 0.3796 0.3268 0.2060 0.0980 0.0424 -0.0186 -0.0430 -0.0586
-0.0589 -0.0696 -0.0772 -0.0632 -0.0848 -0.1013 -0.0940 -0.0839 -0.0598 -0.0345 -0.0131 0.0604 0.2576 0.2960 0.2536 0.1974 0.0928 0.0262 -0.0266 -0.0397 -0.0546
-0.0610 -0.0650 -0.0772 0.2383 1.4291 0.0836 -0.0998 -0.1157 -0.1053 -0.0772 -0.0540 -0.0183 0.1093 0.2029 0.1865 0.1715 0.0995 0.0174 -0.0412 -0.0479 -0.0443
-0.0647 -0.0653 -0.0790 -0.0491 1.4752 2.6566 1.9754 -0.1193 -0.1337 -0.1251 -0.1062 -0.0751 -0.0204 0.0833 0.1221 0.1111 0.1025 0.0226 -0.0433 -0.0616 -0.0626
-0.0558 -0.0665 -0.0815 -0.0601 0.1520 3.0429 6.1838 5.6906 1.1609 -0.1587 -0.1459 -0.1297 -0.0934 -0.0409 0.0317 0.0513 0.0708 0.0305 -0.0415 -0.0638 -0.0607
0.2582 0.0153 -0.0800 -0.0623 -0.0583 1.0925 6.0068 8.4924 6.7996 4.0253 -0.0656 -0.1505 -0.1321 -0.1123 -0.0684 -0.0128 0.0092 0.0244 -0.0391 -0.0644 -0.0687
2.1964 0.7330 0.1846 0.0098 -0.0473 -0.0342 3.0319 9.7589 9.9997 9.9997 8.0103 2.4527 -0.1282 -0.1208 -0.1114 -0.0763 -0.0461 -0.0171 -0.0385 -0.0620 -0.0681
0.0967 2.7423 1.5198 0.9164 0.5261 0.0790 0.1855 6.0474 9.9997 9.9997 9.9997 9.9997 4.9326 0.0308 -0.0900 -0.1013 -0.1013 -0.0711 -0.0504 -0.0607 -0.0696
-0.0598 0.0070 2.4112 2.3657 1.9696 1.4609 1.2375 1.6498 9.9881 9.9997 9.9997 9.9997 9.9997 8.4024 2.9581 -0.0681 -0.0748 -0.0665 -0.0635 -0.0629 -0.0693
-0.0616 -0.0638 -0.0314 1.1356 3.4067 2.5894 2.9434 3.2678 7.1732 9.9997 9.9997 9.9997 9.9997 9.9997 6.3333 4.5761 -0.0378 -0.0626 -0.0671 -0.0714 -0.0674
-0.0626 -0.0671 -0.0781 -0.0436 0.0507 3.0930 3.5120 4.0921 5.3421 9.9997 9.9997 9.9997 9.9997 9.9997 7.5992 4.5221 3.8205 1.7584 -0.0641 -0.0729 -0.0665
-0.0620 -0.0665 -0.0800 -0.0589 -0.0464 0.0168 2.0221 4.3130 4.9597 6.9238 9.9997 9.9997 9.9997 9.9997 9.9997 3.8733 3.1485 2.4753 2.4164 -0.0705 -0.0656
-0.0638 -0.0647 -0.0821 -0.0610 -0.0397 -0.0360 -0.0076 0.1013 3.3884 5.9052 7.3169 9.9997 9.9997 9.9997 9.9997 6.1499 1.1453 1.3699 1.5634 1.9119 0.5576
-0.0650 -0.0650 -0.0842 -0.0632 -0.0348 -0.0180 -0.0064 -0.0299 0.0577 2.4121 4.5850 7.7075 9.9997 9.9997 9.9997 8.0157 0.5927 0.1312 0.5273 0.9311 0.9802
-0.0626 -0.0677 -0.0824 -0.0638 -0.0366 -0.0082 0.0284 0.0201 0.0009 0.0238 0.1215 2.7127 7.3132 9.9997 9.9997 8.0194 6.8076 -0.0146 -0.0311 0.0830 0.4599
-0.0507 -0.0668 -0.0800 -0.0635 -0.0403 -0.0092 0.0543 0.0830 0.0620 0.0211 0.0006 0.0375 0.2414 4.2218 7.6529 7.8983 6.4822 3.3484 -0.0430 -0.0577 -0.0464
-0.0409 -0.0549 -0.0745 -0.0607 -0.0397 -0.0110 0.0613 0.1324 0.1398 0.0998 0.0613 0.0189 -0.0055 0.0623 0.1926 5.6717 6.6541 5.2524 0.6009 -0.0565 -0.0543
-0.0400 -0.0430 -0.0623 -0.0507 -0.0330 -0.0095 0.0565 0.1620 0.2112 0.2039 0.1672 0.0922 0.0238 -0.0055 -0.0266 0.0601 0.1840 3.4863 2.0746 0.7181 -0.0562
-0.0415 -0.0421 -0.0525 -0.0348 -0.0192 0.0024 0.0592 0.1627 0.2631 0.3098 0.3015 0.2017 0.0980 0.0317 -0.0119 -0.0381 -0.0562 0.0058 0.0195 0.9073 -0.0534
-0.0415 -0.0427 -0.0479 -0.0259 -0.0031 0.0195 0.0763 0.1651 0.2972 0.3961 0.4553 0.3519 0.2002 0.1019 0.0223 -0.0259 -0.0571 -0.0516 -0.0635 -0.0531 -0.0604
-0.0433 -0.0424 -0.0491 -0.0250 0.0018 0.0330 0.0964 0.1950 0.3262 0.4553 0.5765 0.5301 0.3271 0.1938 0.0745 0.0012 -0.0415 -0.0476 -0.0626 -0.0699 -0.0662
|
Attachment 6: ER-580-310-30kV-m28.txt
|
value (Voltage offset 0.0000)
76.0000
0.9000
100.0000
29978.0220
21
21
./emit.dat
-0.1187 -0.0696 -0.0842 -0.0742 -0.0809 -0.0940 -0.0714 -0.0360 0.0162 0.0665 0.1373 0.5869 1.1185 0.8875 0.5414 0.1840 0.0412 -0.0162 -0.0540 -0.0662 -0.0754
-0.0595 -0.0696 -0.0818 -0.0671 -0.0717 -0.0870 -0.0842 -0.0754 -0.0403 0.0131 0.0674 0.2820 0.8087 0.7736 0.5081 0.1941 0.0494 -0.0217 -0.0558 -0.0607 -0.0705
-0.0598 -0.0729 -0.0797 -0.0629 -0.0592 -0.0653 -0.0629 -0.0919 -0.0983 -0.0595 0.0137 0.0867 0.4562 0.5594 0.4379 0.1944 0.0574 -0.0235 -0.0562 -0.0644 -0.0589
-0.0647 -0.0735 -0.0775 -0.0659 -0.0461 0.1990 0.1093 -0.0995 -0.1205 -0.1315 -0.0946 -0.0339 0.0912 0.2820 0.3189 0.1581 0.0604 -0.0229 -0.0626 -0.0671 -0.0632
-0.0677 -0.0653 -0.0790 -0.0671 -0.0525 0.3851 2.2314 2.3416 -0.0034 -0.1788 -0.1575 -0.1410 -0.0906 0.0049 0.1236 0.0906 0.0482 -0.0162 -0.0613 -0.0681 -0.0677
0.0015 -0.0677 -0.0839 -0.0659 -0.0623 -0.0375 2.6950 5.9198 4.4141 1.5039 -0.2081 -0.1733 -0.1437 -0.1343 -0.0751 -0.0055 0.0079 -0.0189 -0.0604 -0.0671 -0.0653
1.3000 0.5222 -0.0537 -0.0626 -0.0696 -0.0665 0.3500 6.6342 9.9997 9.9997 6.5463 0.6955 -0.1389 -0.1205 -0.1266 -0.0842 -0.0458 -0.0369 -0.0616 -0.0681 -0.0665
-0.0034 1.7517 1.2799 0.7275 0.1010 -0.0507 -0.0269 1.8335 9.9997 9.9997 9.9997 9.9997 3.0206 -0.0671 -0.0848 -0.0958 -0.0940 -0.0616 -0.0626 -0.0665 -0.0653
-0.0598 -0.0211 1.6721 2.0697 1.9296 1.1948 0.5298 0.4449 7.5043 9.9997 9.9997 9.9997 9.9997 6.2766 1.6965 -0.0626 -0.0717 -0.0607 -0.0656 -0.0653 -0.0620
-0.0601 -0.0604 -0.0446 0.3900 3.1848 2.8616 2.8793 3.1970 3.4647 9.9997 9.9997 9.9997 9.9997 8.4708 3.4933 2.9553 -0.0632 -0.0583 -0.0653 -0.0659 -0.0641
-0.0571 -0.0690 -0.0735 -0.0558 0.0223 3.2553 4.3109 5.0281 6.7178 9.9997 9.9997 9.9997 9.9997 9.9997 5.2866 2.2714 1.7648 0.7895 -0.0616 -0.0681 -0.0647
-0.0552 -0.0714 -0.0766 -0.0616 -0.0546 0.0031 1.1572 6.2912 6.8265 9.1522 9.9997 9.9997 9.9997 9.9997 9.4885 1.9513 1.3593 1.0706 0.9070 -0.0662 -0.0647
-0.0583 -0.0641 -0.0766 -0.0616 -0.0528 -0.0525 -0.0238 0.1154 4.3103 9.9997 9.9997 9.9997 9.9997 9.9997 9.9997 2.5046 0.6808 0.9274 0.7187 0.4718 0.0467
-0.0616 -0.0589 -0.0775 -0.0601 -0.0522 -0.0519 -0.0360 -0.0238 0.0760 1.6986 9.2844 9.9997 9.9997 9.9997 9.9997 6.6360 0.0238 0.0186 0.2628 0.3699 0.2271
-0.0632 -0.0604 -0.0793 -0.0598 -0.0565 -0.0549 -0.0290 -0.0119 -0.0073 0.0375 0.2820 2.9593 9.9997 9.9997 9.9997 9.9997 3.8776 -0.0449 -0.0552 -0.0046 0.1202
-0.0586 -0.0693 -0.0772 -0.0629 -0.0613 -0.0571 -0.0192 0.0146 0.0284 0.0238 0.0653 0.1486 0.5829 5.3879 9.9997 9.9997 6.1002 1.5350 -0.0574 -0.0616 -0.0595
-0.0552 -0.0641 -0.0745 -0.0665 -0.0592 -0.0586 -0.0189 0.0372 0.0952 0.0946 0.0949 0.0949 0.1031 0.1544 0.3656 5.3406 6.2766 2.5500 -0.0174 -0.0601 -0.0607
-0.0565 -0.0565 -0.0677 -0.0610 -0.0516 -0.0601 -0.0195 0.0439 0.1477 0.2090 0.1990 0.1373 0.0867 0.0418 0.0128 0.0638 0.1416 1.5460 0.3824 -0.0607 -0.0623
-0.0580 -0.0522 -0.0635 -0.0522 -0.0433 -0.0516 -0.0177 0.0443 0.1752 0.3250 0.3522 0.2570 0.1184 0.0406 -0.0015 -0.0357 -0.0537 -0.0323 -0.0510 -0.0592 -0.0620
-0.0580 -0.0531 -0.0650 -0.0449 -0.0363 -0.0345 -0.0098 0.0555 0.1871 0.3961 0.5533 0.4825 0.1907 0.0653 0.0006 -0.0372 -0.0580 -0.0558 -0.0623 -0.0708 -0.0659
-0.0552 -0.0562 -0.0626 -0.0427 -0.0360 -0.0241 0.0079 0.0748 0.2063 0.4385 0.7339 0.7773 0.3360 0.1132 0.0156 -0.0320 -0.0555 -0.0558 -0.0607 -0.0690 -0.0632
|
Attachment 7: ER-580-310-30kV-m40.txt
|
value (Voltage offset 0.0000)
76.0000
0.9000
100.0000
29978.0220
21
21
./emit.dat
-0.1373 -0.0854 -0.0903 -0.0897 -0.0861 -0.0925 -0.0922 -0.0696 -0.0858 -0.0909 -0.0739 -0.0726 -0.0809 -0.0699 -0.0690 -0.0790 -0.0833 -0.0854 -0.0861 -0.0851 -0.0848
-0.0699 -0.0800 -0.0858 -0.0836 -0.0821 -0.0882 -0.0864 -0.0638 -0.0812 -0.0842 -0.0720 -0.0671 -0.0769 -0.0641 -0.0653 -0.0748 -0.0778 -0.0818 -0.0781 -0.0830 -0.0793
-0.0720 -0.0726 -0.0873 -0.0766 -0.0626 -0.0827 -0.0577 -0.0708 -0.0613 -0.0839 -0.0629 -0.0620 -0.0681 -0.0626 -0.0613 -0.0720 -0.0769 -0.0787 -0.0723 -0.0809 -0.0665
-0.0717 -0.0735 -0.0867 -0.0748 -0.0620 -0.0571 -0.0439 -0.0620 -0.0656 -0.0833 -0.0635 -0.0659 -0.0671 -0.0638 -0.0626 -0.0723 -0.0778 -0.0775 -0.0745 -0.0784 -0.0705
-0.0748 -0.0732 -0.0851 -0.0787 -0.0690 -0.0638 -0.0287 -0.0165 -0.0482 -0.0858 -0.0632 -0.0641 -0.0677 -0.0656 -0.0635 -0.0696 -0.0806 -0.0760 -0.0766 -0.0760 -0.0681
-0.0705 -0.0732 -0.0854 -0.0769 -0.0659 -0.0818 -0.0394 0.0076 0.0073 -0.0424 -0.0571 -0.0620 -0.0732 -0.0647 -0.0656 -0.0702 -0.0797 -0.0769 -0.0766 -0.0772 -0.0647
-0.0525 -0.0681 -0.0830 -0.0742 -0.0632 -0.0848 -0.0488 -0.0153 0.0351 -0.0046 -0.0055 -0.0302 -0.0751 -0.0647 -0.0653 -0.0717 -0.0787 -0.0778 -0.0742 -0.0803 -0.0601
-0.0705 -0.0519 -0.0717 -0.0687 -0.0592 -0.0824 -0.0604 -0.0388 0.0327 0.0385 0.0308 -0.0012 -0.0299 -0.0607 -0.0629 -0.0726 -0.0784 -0.0800 -0.0729 -0.0803 -0.0620
-0.0711 -0.0732 -0.0604 -0.0537 -0.0488 -0.0702 -0.0571 -0.0592 -0.0092 0.0513 0.0702 0.0562 -0.0006 -0.0174 -0.0323 -0.0723 -0.0797 -0.0800 -0.0745 -0.0778 -0.0632
-0.0745 -0.0714 -0.0824 -0.0708 -0.0378 -0.0574 -0.0491 -0.0568 -0.0281 0.0040 0.0873 0.1031 0.0580 0.0113 -0.0110 -0.0378 -0.0797 -0.0781 -0.0769 -0.0766 -0.0644
-0.0739 -0.0729 -0.0830 -0.0748 -0.0659 -0.0525 -0.0369 -0.0446 -0.0317 -0.0232 0.0644 0.1093 0.1202 0.0546 -0.0067 -0.0360 -0.0461 -0.0589 -0.0760 -0.0781 -0.0690
-0.0714 -0.0729 -0.0839 -0.0748 -0.0638 -0.0824 -0.0427 -0.0375 -0.0269 -0.0366 0.0098 0.0888 0.0977 0.0821 0.0278 -0.0372 -0.0537 -0.0558 -0.0464 -0.0803 -0.0620
-0.0739 -0.0732 -0.0845 -0.0748 -0.0629 -0.0842 -0.0586 -0.0708 -0.0281 -0.0473 -0.0128 0.0143 0.0793 0.0607 0.0598 -0.0336 -0.0690 -0.0656 -0.0574 -0.0534 -0.0616
-0.0742 -0.0742 -0.0836 -0.0775 -0.0610 -0.0824 -0.0632 -0.0626 -0.0629 -0.0641 -0.0256 -0.0165 0.0177 0.0476 0.0339 0.0076 -0.0668 -0.0745 -0.0705 -0.0656 -0.0476
-0.0717 -0.0745 -0.0836 -0.0790 -0.0632 -0.0812 -0.0638 -0.0620 -0.0647 -0.0842 -0.0632 -0.0287 -0.0168 0.0098 0.0229 0.0076 -0.0266 -0.0766 -0.0745 -0.0757 -0.0565
-0.0684 -0.0745 -0.0833 -0.0778 -0.0641 -0.0818 -0.0629 -0.0662 -0.0623 -0.0851 -0.0635 -0.0607 -0.0702 -0.0195 -0.0034 -0.0024 -0.0247 -0.0592 -0.0729 -0.0800 -0.0613
-0.0693 -0.0754 -0.0833 -0.0760 -0.0638 -0.0812 -0.0601 -0.0644 -0.0613 -0.0845 -0.0656 -0.0604 -0.0717 -0.0641 -0.0604 -0.0143 -0.0177 -0.0421 -0.0656 -0.0812 -0.0598
-0.0714 -0.0723 -0.0854 -0.0729 -0.0650 -0.0797 -0.0632 -0.0775 -0.0613 -0.0827 -0.0629 -0.0604 -0.0708 -0.0650 -0.0626 -0.0708 -0.0760 -0.0507 -0.0571 -0.0769 -0.0674
-0.0702 -0.0714 -0.0858 -0.0739 -0.0641 -0.0787 -0.0641 -0.0641 -0.0653 -0.0809 -0.0598 -0.0638 -0.0681 -0.0644 -0.0626 -0.0739 -0.0787 -0.0772 -0.0717 -0.0748 -0.0662
-0.0720 -0.0723 -0.0842 -0.0751 -0.0632 -0.0815 -0.0632 -0.0592 -0.0613 -0.0824 -0.0583 -0.0626 -0.0690 -0.0629 -0.0604 -0.0705 -0.0769 -0.0763 -0.0757 -0.0784 -0.0641
-0.0726 -0.0735 -0.0830 -0.0732 -0.0632 -0.0827 -0.0616 -0.0613 -0.0571 -0.0833 -0.0586 -0.0616 -0.0668 -0.0626 -0.0620 -0.0696 -0.0778 -0.0745 -0.0751 -0.0790 -0.0616
|
Attachment 8: SiC#37_(ii).docx
|
Attachment 9: OMEGA_OM-DAQPRO-5300_19-Oct-17_11.17.50.dlp
|
Attachment 10: ITE-TM4-SiC#37-HP-FEBIAD_Test_Stand.docx
|
497
|
Friday, March 04, 2016, 14:23 |
Anders Mjos | Target tube only | LP-SIS | ITW | TM2 | 39459-2 | 41120-10 | In progress | Conditiong Heat Shield for UC#15 | Friday, March 04, 2016, 14:20: Roughing
Friday, March 04, 2016, 15:05: Turbos started
Friday, March 04, 2016, 15:11: Turbos at speed. Vacuum at 3.5E-5 T.
Monday, March 07, 2016, 08:34: Opened water. Vacuum stable at 6.7E-7 T. No response on RGA
Monday, March 07, 2016, 08:46: Started auto-ramping TGHT to 420 A at 2 A/min. EE to 1500 V
Tuesday, March 08, 2016, 08:26: Continued auto-ramping TGHT to 500 A.
Wednesday, March 09, 2016, 09:25: TGHT voltage limited at 488 A. Increased voltage limit from 6 to 6.5 V. Resumed auto-ramping to 500 A
Thursday, March 10, 2016, 08:19: Begin cool down @ 2 A/min ; Vacuum @ 7.1E-7 T
Thursday, March 10, 2016, 13:20: TGHT to 0 A. Switched off RGA.
Thursday, March 10, 2016, 14:04 : Source off. Stopped Turbo pumps.
Friday, March 11, 2016, 08:15: Begin venting.
Friday, March 11, 2016, 08:28: Heat shield water circuit has been purged.
|
499
|
Wednesday, March 16, 2016, 09:59 |
Anders Mjos | Target tube only | | | TM1 | 39459-2 | 40898-2 | In progress | Conditiong Heat Shield for UC#15 | Back up heat shield for UC#15
Wednesday, March 16, 2016, 09:58: Roughing
Wednesday, March 16, 2016, 12:54: Started Turbo pumps
Wednesday, March 16, 2016, 13:02: Turbos at speed, Vacuum at 7.16E-06T
Thursday, March 17, 2016, 08:30: Opened water on heat shield circuit. Vacuum stable at 3.0E-07T. No response on RGA.
Thursday, March 17, 2016, 08:37: Auto-Ramping started. TGHT to 500A @ 2A/min. EE to 1500V.
Thursday, March 17, 2016, 15:40 @ 500A and 6.76E-07T
Friday, March 18, 2016, 08:35: Auto-cooling. Vacuum @5.87E-07T.
Friday, March 18, 2016, 14:13: Source off. Stopped turbo pumps.
Tuesday, March 29, 2016, 13:22: venting |
337
|
Friday, September 27, 2013, 11:41 |
Anders Mjos | Ta#41 | HP-SIS | ITW | TM4 | 39069-5 | 38817-10 | Done | Ta#41 Conditioning | Friday, September 27, 2013, 11:41: Roughing
Friday, September 27, 2013, 12:08: Turbos started
Friday, September 27, 2013, 12:16: Turbos at speed.
Friday, September 27, 2013, 12:18: Vacuum at 1.5E-5 T
Monday, September 30, 2013, 08:41: Opened water supply. Vacuum stable at 1.5E-7 T
TGHT Q=5.5 T=28.0
TBHT Q=5.8 T=28.0
Monday, September 30, 2013, 15:19: Flow sensor moved from TBHT to HS
TGHT Q=5.5 T=27.9
HS Q=3.6 T=27.8
Tuesday, October 01, 2013, 12:08: Taget Biased to 30kV.
Wednesday, October 02, 2013, 08:50: Started autoramping to 580A/230A
Wednesday, October 02, 2013, 08:50: Opened IV4
Wednesday, October 02, 2013, 10:51 TGHT/TBHT 250A / 239A
TGHT Q=5.5 T=28.5
HS Q=3.6 T=29.2
SUPPLY (ext) 27.3
RETURN (ext) 27.8
Wednesday, October 02, 2013, 13:28 TGHT/TBHT 250A / 570A
TGHT Q=5.5 T=30.4
HS Q=3.6 T=33.4
SUPPLY (ext) 27.4
RETURN (ext) 29.3
Wednesday, October 02, 2013, 13:30: TBHT was starting to limit on voltage. Increased voltage limit to 2.4V.
Thursday, October 03, 2013, 08:21: TMS:TP2 timed out over night.
Magnet PS breaker tripped during a mass scan yesterday. Reset breaker and cleared interlocks.
TGHT Q=5.5 T=30.4
HS Q=3.6 T=34.0
SUPPLY (ext) 27.3
RETURN (ext) 29.3
Thursday, October 03, 2013, 08:28: Started mass scan 200 -> 0 m/q. FC0 7.5E-8 A
Thursday, October 03, 2013, 14:29
TGHT Q=5.5 T=30.6
HS Q=3.6 T=33.9
SUPPLY (ext) 27.4
RETURN (ext) 29.6
Thursday, October 03, 2013, 15:14: Lowered TBHT:RDCUR to 230A.
Thursday, October 03, 2013, 15:37: Increased EE to -2000 V
Thursday, October 03, 2013, 16:22
TGHT Q=5.5 T=30.6
HS Q=3.6 T=33.9
SUPPLY (ext) 27.5
RETURN (ext) 29.6
Friday, October 04, 2013, 08:08
TGHT Q=5.5 T=30.6
HS Q=3.5 T=33.9
SUPPLY (ext) 27.5
RETURN (ext) 29.3
Mass scan started.
Friday, October 04, 2013, 08:46: Cool down started.
Friday, October 04, 2013, 13:35: TPS:IV4 closed
Friday, October 04, 2013, 13:41: Source debiased.
Tuesday, October 15, 2013, 09:56: Turbos stopped. Vacuum at 1.15 E-7 T
Tuesday, October 15, 2013, 12:44: Venting
After the conditioning, some oxidation (?) from the standards (?) were found on the Plasma Electrode and around the Target Tube. See attachment |
Attachment 1: Ta#41_EE_After_Test_Stand.pdf
|
|
402
|
Friday, August 29, 2014, 14:52 |
Anders Mjos | Ta#44 | HP-SIS | ITE | TM4 | 38817-9 | 40316-11 | Done | Conditioning Ta#44 | Friday, August 29, 2014, 14:51: Roughing
Friday, August 29, 2014, 15:11: Starting turbos.
Friday, August 29, 2014, 15:19: Turbos at speed
Friday, August 29, 2014, 15:21: Vacuum at 1.6E-5 T
Tuesday, September 02, 2014, 08:36: HS cooling water supply opened. Vacuum stable at 1.4E-7 T
Tuesday, September 02, 2014, 08:50 : BIAS/ EZL / EE to 30 / 14 / -1.5 kV. Started auto ramping to 500 / 150 (180) A
Tuesday, September 02, 2014, 09:23
Supply ~97psi, Return ~7psi
Device
|
Flow
[lpm] |
Temp
[C] |
HS + MSP |
3.9 |
28.0 |
TBHT |
6.5 |
27.6 |
COIL + EE |
5.4 |
28.6 |
TGHT |
5.8 |
27.8 |
SUPPLY |
|
28.2 |
Thursday, September 04, 2014, 12:38 : Starting cool-down
Thursday, September 04, 2014, 15:20: Closed IV4
Thursday, September 04, 2014, 15:37 : BIAS / EZL / EE to 0
Friday, September 05, 2014
Target cooled down. Power supplies switched off.
Monday, September 08, 2014, 14:07: Turbos stopped.
Monday, September 08, 2014, 15:15: Venting. HS cooling water lines purged.
Monday, September 08, 2014, 16:00: Target removed. |
446
|
Monday, May 25, 2015, 14:40 |
Anders Mjos | SiC#31 | HP-SIS | ITW | TM4 | 38817-7 | | In progress | SiC#31 Conditioning | Monday, May 25, 2015, 14:35: Rouging
Monday, May 25, 2015, 15:05: Turbos started.
Monday, May 25, 2015, 15:15: Turbos at speed. Vacuum at 2.4E-5 T
Tuesday, May 26, 2015, 07:45: Opened water circuit for HS line. Vacuum stable at 6.3E-7 T. No response on RGA
Tuesday, May 26, 2015, 07:50: Autoramping to 500/200 A at 2/1 A/min. Good vacuum 2E-6, bad vacuum 4E-6 T
Tuesday, May 26, 2015, 07:52: IV4 open
Tuesday, May 26, 2015, 13:15: Continue auto-ramping to 580 A
Tuesday, May 26, 2015, 13:48: BIAS/EZL/EE 30 / 15 / 1.5 kV
Tuesday, May 26, 2015, 17:29: ~6 sparks/hour
Wednesday, May 27, 2015, 08:14: A couple of sparks over night.
Wednesday, May 27, 2015, 09:23: Autoramping TBHT to 220 A
Wednesday, May 27, 2015, 09:41: Found FC0 and FC6 bias to be off
Wednesday, May 27, 2015, 09:42: Autoramping TBHT to 230 A
Wednesday, May 27, 2015, 10:30: EE to 2 kV
Wednesday, May 27, 2015, 10:37: Starting cool down at 4/2 A/min
Wednesday, May 27, 2015, 10:40: EE to 1.5 kV
Wednesday, May 27, 2015, 12:37 : BIAS/EZL/EE OFF
Wednesday, May 27, 2015, 12:37 : IV4 closed
Wednesday, May 27, 2015, 13:02: Stopped turbos
Wednesday, May 27, 2015, 13:41: Venting |
Attachment 1: SiC#31.docx
|
Attachment 2: ER30kV580-200.txt
|
value (Voltage offset 0.0000)
75.0000
0.8000
50.0000
29978.0220
21
21
./emit.dat
0.0131 -0.0009 0.0000 0.0003 0.0000 -0.0006 -0.0015 0.0015 -0.0043 -0.0024 -0.0034 -0.0031 -0.0018 -0.0012 -0.0031 -0.0018 0.0092 0.0006 0.0024 0.0034 0.0055
0.0107 -0.0037 -0.0018 0.0015 -0.0006 -0.0003 -0.0037 -0.0027 -0.0024 -0.0015 -0.0024 -0.0015 0.0058 -0.0015 0.0009 -0.0012 0.0073 0.0003 -0.0006 0.0034 0.0055
0.0137 -0.0018 -0.0021 0.0015 -0.0009 0.0006 0.0006 -0.0024 -0.0006 -0.0009 -0.0024 -0.0018 -0.0009 -0.0018 -0.0015 0.0000 0.0089 0.0027 0.0000 0.0003 0.0003
0.0082 -0.0009 -0.0024 -0.0006 -0.0027 0.0012 0.0000 -0.0012 -0.0021 0.0012 -0.0015 -0.0018 -0.0049 -0.0015 0.0009 -0.0043 0.0107 0.0031 0.0000 0.0015 0.0003
0.0122 0.0003 -0.0027 -0.0018 -0.0009 -0.0006 0.0015 0.0040 0.0009 -0.0006 -0.0009 -0.0015 0.0000 -0.0012 -0.0006 -0.0015 0.0058 0.0024 0.0000 0.0040 -0.0015
0.0128 -0.0024 -0.0034 0.0006 -0.0012 0.0006 -0.0003 0.0049 -0.0003 0.0040 0.0140 0.0012 -0.0009 0.0000 0.0000 -0.0006 0.0061 -0.0003 0.0015 0.0000 0.0027
0.0119 -0.0027 -0.0015 -0.0003 0.0006 -0.0012 -0.0009 0.0006 0.0027 0.0391 0.1184 0.1059 0.0482 0.0055 -0.0024 -0.0018 0.0098 0.0021 0.0027 0.0024 0.0024
0.0140 -0.0034 -0.0018 -0.0009 -0.0031 -0.0012 0.0000 0.0009 0.0015 0.0778 0.1892 0.2051 0.1620 0.0763 0.0027 -0.0018 0.0052 0.0018 0.0009 0.0003 0.0049
0.0119 -0.0006 -0.0018 -0.0021 -0.0018 -0.0006 0.0015 0.0012 0.0027 0.0885 0.2640 0.3473 0.2960 0.1498 0.0403 -0.0006 0.0092 0.0021 0.0000 0.0021 0.0027
0.0119 -0.0018 -0.0021 -0.0018 -0.0012 0.0021 0.0021 0.0021 0.0003 0.1001 0.4187 0.6180 0.5743 0.3156 0.0836 0.0116 0.0183 0.0018 -0.0027 0.0046 0.0015
0.0079 0.0043 -0.0009 -0.0015 0.0015 0.0125 0.0052 0.0009 -0.0003 0.0952 0.5145 0.9247 0.9613 0.5710 0.1602 0.0247 0.0220 0.0015 -0.0018 0.0024 0.0021
0.0134 -0.0021 0.0000 -0.0009 0.0061 0.0058 0.0043 0.0040 0.0021 0.0671 0.5103 1.1981 1.4487 0.9302 0.2875 0.0439 0.0116 0.0027 0.0006 0.0034 0.0018
0.0095 -0.0015 0.0000 0.0055 0.0049 0.0037 0.0000 0.0012 0.0024 0.0418 0.3949 1.6235 2.2305 1.6080 0.4742 0.0952 0.0504 0.0348 0.0009 0.0012 0.0027
0.0107 -0.0018 0.0003 0.0003 -0.0018 0.0095 0.0076 -0.0021 0.0018 0.0177 0.2280 1.1844 1.8930 1.7804 0.8380 0.1935 0.0409 0.0446 0.1483 0.0034 0.0018
0.0101 -0.0009 -0.0003 -0.0034 -0.0027 0.0034 0.0027 -0.0021 -0.0012 0.0049 0.1041 0.5759 1.1823 1.2555 0.9869 0.3024 0.0571 0.0208 0.2191 0.0772 0.0018
0.0076 -0.0009 -0.0009 -0.0027 -0.0018 0.0006 -0.0021 0.0018 -0.0015 0.0015 0.0381 0.2502 0.6845 1.0934 0.9415 0.4208 0.1108 0.0168 0.0348 0.0507 0.0009
0.0101 -0.0031 -0.0018 -0.0006 -0.0006 0.0000 -0.0021 -0.0009 0.0034 0.0009 0.0098 0.0992 0.3369 0.6430 0.7822 0.4846 0.1813 0.0165 0.0110 0.0067 0.0027
0.0073 -0.0006 -0.0012 0.0000 -0.0006 -0.0009 0.0012 -0.0009 -0.0006 0.0134 0.0021 0.0342 0.1541 0.3702 0.4874 0.4376 0.2237 0.0369 0.0119 0.0015 0.0031
0.0101 -0.0024 0.0006 0.0000 -0.0009 -0.0027 0.0009 0.0021 0.0003 -0.0003 0.0082 0.0180 0.0650 0.1892 0.3043 0.2966 0.2097 0.0723 0.0067 0.0040 0.0003
0.0067 0.0000 -0.0003 -0.0012 -0.0018 0.0027 0.0000 -0.0012 0.0009 0.0000 0.0009 0.0058 0.0375 0.0885 0.1822 0.1828 0.1566 0.0882 0.0125 0.0037 0.0012
0.0082 -0.0006 0.0009 -0.0024 -0.0021 0.0024 0.0000 0.0027 -0.0009 -0.0003 0.0015 -0.0006 0.0089 0.0400 0.0885 0.1178 0.1004 0.0714 0.0107 0.0195 0.0024
|
Attachment 3: ER30kV580-230.txt
|
value (Voltage offset 0.0000)
75.0000
0.8000
50.0000
29978.0220
21
21
./emit.dat
-0.0009 -0.0009 0.0024 -0.0006 -0.0031 0.0064 0.0027 -0.0009 0.0043 0.0018 0.0000 -0.0024 -0.0006 -0.0037 -0.0006 -0.0024 -0.0012 0.0012 -0.0024 0.0006 0.0027
0.0003 -0.0031 0.0000 0.0037 0.0012 0.0061 0.0031 0.0006 -0.0021 0.0037 -0.0012 -0.0024 -0.0040 -0.0021 -0.0012 -0.0043 -0.0018 0.0006 -0.0009 0.0009 0.0092
-0.0027 -0.0009 0.0031 -0.0021 0.0024 0.0037 0.0049 -0.0015 -0.0031 0.0021 0.0000 -0.0049 -0.0015 -0.0052 0.0009 -0.0015 -0.0034 0.0000 -0.0003 0.0018 0.0092
-0.0027 -0.0037 0.0031 -0.0049 -0.0027 0.0049 0.0043 0.0006 -0.0040 0.0015 -0.0009 -0.0027 -0.0024 -0.0024 0.0018 -0.0024 -0.0037 0.0027 -0.0009 0.0018 0.0067
-0.0015 -0.0027 0.0034 -0.0031 -0.0024 0.0052 0.0021 -0.0021 -0.0031 0.0034 -0.0015 -0.0018 -0.0021 -0.0037 0.0031 -0.0015 0.0006 0.0040 -0.0006 -0.0006 0.0098
-0.0021 -0.0024 -0.0006 -0.0009 -0.0009 0.0064 0.0024 -0.0024 -0.0018 0.0061 0.0006 -0.0021 -0.0021 -0.0018 0.0034 -0.0027 -0.0012 0.0015 0.0012 -0.0006 0.0027
-0.0034 -0.0003 0.0012 0.0003 -0.0043 0.0058 0.0058 -0.0009 0.0003 0.0095 0.0089 0.0034 0.0000 0.0018 0.0043 -0.0021 0.0006 0.0003 -0.0009 0.0012 0.0040
-0.0046 0.0012 0.0015 0.0012 -0.0012 0.0040 0.0049 -0.0006 0.0006 0.0482 0.1517 0.2005 0.0616 0.0122 0.0015 0.0006 -0.0027 0.0015 -0.0012 0.0024 0.0098
-0.0027 0.0012 0.0000 -0.0003 0.0006 0.0027 0.0015 -0.0009 0.0110 0.1740 0.4355 0.7028 0.3497 0.0717 0.0168 -0.0006 -0.0015 0.0027 0.0003 -0.0012 0.0104
-0.0031 -0.0021 0.0009 -0.0027 -0.0003 0.0058 0.0018 -0.0009 0.0052 0.2228 0.6641 0.9564 0.6808 0.1855 0.0323 0.0043 -0.0024 0.0003 -0.0006 -0.0006 0.0131
-0.0043 -0.0003 0.0024 -0.0018 -0.0040 0.0058 0.0031 -0.0006 0.0040 0.1422 0.7123 1.0974 1.1118 0.4178 0.0848 0.0052 0.0000 -0.0009 0.0000 -0.0018 0.0052
-0.0027 -0.0012 0.0000 -0.0024 -0.0024 0.0049 0.0027 -0.0009 0.0027 0.0806 0.5936 1.1768 1.3040 0.8737 0.2533 0.0107 -0.0018 0.0009 -0.0031 0.0024 0.0027
-0.0031 -0.0003 0.0012 -0.0031 0.0000 0.0034 0.0037 -0.0034 0.0018 0.0403 0.3281 1.1108 1.3599 1.2180 0.5121 0.0201 0.0034 0.0467 -0.0015 0.0015 0.0064
-0.0024 0.0012 0.0012 -0.0024 0.0012 0.0034 0.0018 -0.0012 0.0079 0.0156 0.1288 0.7346 1.2213 1.1874 0.6903 0.0458 0.0012 0.0418 0.1047 0.0015 0.0064
-0.0040 -0.0012 0.0021 -0.0043 -0.0015 0.0055 0.0027 0.0006 -0.0021 0.0082 0.0403 0.3973 0.7742 0.8746 0.7599 0.0897 0.0034 0.0101 0.0275 0.0122 0.0076
-0.0031 -0.0043 0.0040 -0.0018 0.0089 0.0064 0.0031 0.0003 -0.0027 0.0058 0.0134 0.1047 0.3873 0.6091 0.6433 0.1285 0.0052 0.0015 0.0052 0.0015 0.0049
-0.0021 -0.0018 0.0024 -0.0031 -0.0021 0.0049 0.0034 -0.0012 -0.0018 0.0043 0.0012 0.0217 0.1511 0.3198 0.3851 0.1474 0.0076 0.0003 0.0137 0.0024 0.0040
-0.0009 -0.0024 0.0024 -0.0015 0.0000 0.0052 0.0058 -0.0024 0.0003 0.0024 0.0015 0.0037 0.0253 0.1089 0.1926 0.1376 0.0079 0.0061 0.0153 0.0027 0.0055
-0.0015 -0.0015 0.0018 -0.0024 0.0021 0.0040 0.0027 -0.0012 -0.0003 0.0021 0.0006 0.0027 0.0089 0.0195 0.0735 0.0671 0.0101 0.0082 0.0104 0.0012 0.0134
-0.0027 -0.0034 0.0024 -0.0012 -0.0018 0.0049 0.0024 0.0000 -0.0006 0.0034 -0.0024 -0.0009 -0.0006 0.0049 0.0186 0.0323 0.0043 0.0067 0.0027 0.0015 0.0046
-0.0037 -0.0021 0.0006 0.0006 0.0323 0.0040 0.0027 0.0003 -0.0043 0.0052 -0.0024 -0.0021 -0.0046 -0.0009 0.0076 0.0049 0.0055 0.0031 0.0009 0.0015 0.0107
|
431
|
Monday, March 16, 2015, 18:53 |
Anders Mjos | Ta#45 | HP-SIS | ITE | TM4 | 38817-5 | | In progress | Conditioning Ta#45 | Monday, March 16, 2015, 18:53: Started turbos
Monday, March 16, 2015, 19:00: Turbos at speed. Vacuum at 1.3E-5 T
Tuesday, March 17, 2015, 09:03: Opened cooling water. Vacuum stable at 1.9E-7 T
Tuesday, March 17, 2015, 09:12: Opened IV4. BIAS/EZL/EE to 30/14/1.5 kV
Tuesday, March 17, 2015, 09:20: Started heating TGHT/TBHT to 400/150 A at 2/1 A/min
Wednesday, March 18, 2015, 08:41: Continued heating TGHT/TBHT to 580/170 A at 2/1 A/min
Wednesday, March 18, 2015, 09:33: MB1 found tripped on water flow. Water pressure is low this morning as cyclotron main magnet was switched on, but not all water pumps. Expected to be resolved soon..
Some emittance scans at 170, 180, 190, 200 A TBHT were done
Thursday, March 19, 2015, 08:47: Heating TBHT to 220 A
Thursday, March 19, 2015, 08:56: Heating TBHT to 230 A
Thursday, March 19, 2015, 10:50: EE at -2000 V for 10 min
Thursday, March 19, 2015, 10:52: EE back at -1.5 kV
Thursday, March 19, 2015, 17:54: Started cool down of target
Friday, March 20, 2015, 08:47: Closed IV4
Friday, March 20, 2015, 08:47: RGA off.
Friday, March 20, 2015, 08:50: Stopped turbos
Friday, March 20, 2015, 10:24: Venting
Friday, March 20, 2015, 11:45: Target removed. |
Attachment 1: Ta#45_Test_Stand.docx
|
Attachment 2: Ta#45_2015-03-17_Mar_17_2015_09-05-31_AM.ana
|
98
|
Thursday, May 17, 2012, 13:24 |
John Wong | SiC#27 | HP-SIS | ITE | TM4 | 38817-3 | | Problems/Solutions | Solid residues found on EE | Thursday, May 17, 2012, 13:15
After the test-stand, the EE was removed for inspection. It was discovered that there were some solid residues coated on the extraction electrode and the common plate, but the ionizer tube was clean (see attachment#1). Some samples were collected and will try to do some analysis. This problem has been seen before but it's still not clear why the solid residues form after the test-stand.
The EE & the common plate were cleaned with acetone and re-assembled back to the target.
|
Attachment 1: SiC#27_after_teststand.pdf
|
|
Attachment 2: SiC#27_clean_EE.pdf
|
|
649
|
Friday, March 16, 2018, 08:15 |
Aaron Schmidt | Nb#8 | HP-SIS | ITW | TM2 | 38817-2 | 42738-12 | In progress | Conditioning Nb#08 | Friday, March 16, 2018, 08:42: Roughing.
Friday, March 16, 2018, 10:34: Turbo pumps started.
Friday, March 16, 2018, 10:40: Ion gauge on.
Friday, March 16, 2018, 10:45: RGA on.
Friday, March 16, 2018, 13:10: Check. Vacuum @ 7.37e-7T
Monday, March 19, 2018, 06:44: Waterlines opened: Vacuum stable @ 2.16e-07T
Monday, March 19, 2018, 06:46: Power/sources on.
Monday, March 19, 2018, 06:49: IV4 Opened.
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
21.6 |
3.68 |
TBHT |
22.2 |
2.97 |
COIL + EE |
21.4 |
4.73 |
TGHT |
21.5 |
5.39 |
Monday, March 19, 2018, 06:51: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.
Monday, March 19, 2018, 06:57: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .
Monday, March 19, 2018, 07:02: Optics on & reset.
Monday, March 19, 2018, 10:53: Drops in BIAS voltage from 30kV to below 22.5 kV and spikes to the BIAS current (282 uA) observed. Unstable.
Monday, March 19, 2018, 12:56: Communications issue noted with the mass separator magnet. Increase in Amp achieves no change in Gauss. Kyle L notified.
TMS:HALLMB1:RDFIELD Communication error.
Monday, March 19, 2018, 14:00: MB1 Communication error fixed. However the negative voltage on Q2 & Q5 in unavailable.
Monday, March 19, 2018, 15:07: Auto cooling TGHT / TBHT to 500 / 150 A @ 6A/4A per min.
Monday, March 19, 2018, 15:17: Auto cooling halted. TGHT / TBHT @ 501 / 152 A. Manually put to 500 / 150 A.
Controller for TMS:Q5 & TMS:Q2 off for repair.
Tuesday, March 20, 2018, 09:09: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.
Tuesday, March 20, 2018, 12:58: The controller was fixed. Nb#08 conditioning continues with tests.
Wednesday, March 21, 2018, 06:24: Auto-cooling @ 2A per min.
Wednesday, March 21, 2018, 11:23: Sources at 0. Off.
Wednesday, March 21, 2018, 11:29: Power off. Waterlines purged.
Wednesday, March 21, 2018, 11:32: RGA off. IV4 closed/reset.
Wednesday, March 21, 2018, 11:33: Ion gauge off. Turbos off.
Wednesday, March 21, 2018, 13:15: Venting. |
Attachment 1: Nb#08.docx
|
Attachment 2: TS-2018-03-16-10-45.rga
|
Attachment 3: OMEGA_OM-DAQPRO-5300_21-Mar-18_14.40.06.dlp
|
Attachment 4: ER-580-200-30kV-m7.txt
|
value (Voltage offset 0.0000)
80.0000
0.7000
80.0000
29978.0220
21
21
./emit.dat
-0.0854 -0.0821 -0.0848 -0.0864 -0.0867 -0.0943 -0.0598 0.6357 0.8398 0.3336 0.0101 -0.0858 -0.0882 -0.0851 -0.0885 -0.0885 -0.0867 -0.0861 -0.0977 -0.0876 -0.0873
-0.0836 -0.0821 -0.0845 -0.0870 -0.0861 -0.0967 -0.0610 0.6345 0.8401 0.3336 0.0131 -0.0858 -0.0885 -0.0888 -0.0903 -0.0897 -0.0922 -0.0900 -0.1007 -0.0909 -0.0894
-0.0848 -0.0803 -0.0842 -0.0873 -0.0867 -0.0919 -0.0555 0.6326 0.8380 0.3336 0.0122 -0.0851 -0.0854 -0.0903 -0.0873 -0.0912 -0.0873 -0.0903 -0.0983 -0.0891 -0.0888
-0.0854 -0.0839 -0.0848 -0.0888 -0.0879 -0.0891 -0.0519 0.6268 0.8475 0.3339 0.0119 -0.0864 -0.0912 -0.0897 -0.0879 -0.0879 -0.0885 -0.0870 -0.0980 -0.0894 -0.0882
-0.0854 -0.0842 -0.0858 -0.0845 -0.0851 -0.0870 -0.0528 0.6296 0.8478 0.3354 0.0128 -0.0867 -0.0800 -0.0882 -0.0870 -0.0873 -0.0879 -0.0876 -0.0977 -0.0867 -0.0885
-0.0839 -0.0806 -0.0833 -0.0870 -0.0873 -0.0882 -0.0555 0.6247 0.8484 0.3363 0.0119 -0.0882 -0.0928 -0.0879 -0.0885 -0.0891 -0.0864 -0.0888 -0.0970 -0.0873 -0.0854
-0.0842 -0.0790 -0.0833 -0.0845 -0.0873 -0.0888 -0.0571 0.6268 0.8481 0.3342 0.0128 -0.0861 -0.0891 -0.0885 -0.0870 -0.0903 -0.0858 -0.0885 -0.0970 -0.0873 -0.0848
-0.0842 -0.0812 -0.0851 -0.0836 -0.0882 -0.0873 -0.0543 0.6274 0.8508 0.3348 0.0122 -0.0836 -0.0961 -0.0854 -0.0851 -0.0882 -0.0858 -0.0894 -0.0961 -0.0876 -0.0876
-0.0848 -0.0833 -0.0848 -0.0876 -0.0879 -0.0851 -0.0537 0.6268 0.8493 0.3363 0.0116 -0.0851 -0.0809 -0.0845 -0.0842 -0.0864 -0.0845 -0.0882 -0.0970 -0.0885 -0.0867
-0.0851 -0.0842 -0.0848 -0.0888 -0.0879 -0.0858 -0.0540 0.6274 0.8508 0.3360 0.0122 -0.0885 -0.0873 -0.0876 -0.0870 -0.0864 -0.0851 -0.0867 -0.0977 -0.0864 -0.0894
-0.0861 -0.0821 -0.0851 -0.0867 -0.0885 -0.0888 -0.0546 0.6262 0.8475 0.3360 0.0116 -0.0864 -0.0891 -0.0897 -0.0888 -0.0861 -0.0885 -0.0867 -0.0943 -0.0873 -0.0867
-0.0827 -0.0821 -0.0824 -0.0879 -0.0882 -0.0888 -0.0549 0.6250 0.8472 0.3348 0.0116 -0.0848 -0.0879 -0.0900 -0.0891 -0.0891 -0.0897 -0.0885 -0.0940 -0.0894 -0.0885
-0.0836 -0.0800 -0.0845 -0.0876 -0.0888 -0.0882 -0.0546 0.6281 0.8475 0.3320 0.0134 -0.0851 -0.0888 -0.0851 -0.0888 -0.0870 -0.0879 -0.0882 -0.0958 -0.0897 -0.0894
-0.0858 -0.0812 -0.0861 -0.0864 -0.0854 -0.0854 -0.0546 0.6259 0.8481 0.3351 0.0125 -0.0876 -0.0888 -0.0848 -0.0867 -0.0870 -0.0851 -0.0885 -0.0980 -0.0882 -0.0888
-0.0839 -0.0833 -0.0851 -0.0888 -0.0876 -0.0861 -0.0552 0.6290 0.8466 0.3366 0.0104 -0.0879 -0.0861 -0.0876 -0.0851 -0.0873 -0.0851 -0.0876 -0.0946 -0.0851 -0.0854
-0.0845 -0.0815 -0.0858 -0.0879 -0.0876 -0.0867 -0.0565 0.6287 0.8505 0.3348 0.0116 -0.0870 -0.0876 -0.0888 -0.0839 -0.0885 -0.0864 -0.0861 -0.0922 -0.0854 -0.0867
-0.0830 -0.0815 -0.0842 -0.0870 -0.0885 -0.0876 -0.0552 0.6226 0.8481 0.3354 0.0128 -0.0848 -0.0870 -0.0882 -0.0879 -0.0864 -0.0879 -0.0867 -0.0943 -0.0867 -0.0873
-0.0836 -0.0818 -0.0848 -0.0876 -0.0906 -0.0867 -0.0558 0.6250 0.8466 0.3326 0.0159 -0.0854 -0.0870 -0.0861 -0.0879 -0.0897 -0.0885 -0.0870 -0.0992 -0.0876 -0.0888
-0.0851 -0.0839 -0.0848 -0.0879 -0.0870 -0.0861 -0.0562 0.6274 0.8481 0.3326 0.0125 -0.0885 -0.0839 -0.0864 -0.0870 -0.0873 -0.0906 -0.0879 -0.0983 -0.0897 -0.0897
-0.0854 -0.0827 -0.0842 -0.0900 -0.0858 -0.0876 -0.0546 0.6265 0.8447 0.3345 0.0119 -0.0867 -0.0867 -0.0873 -0.0864 -0.0854 -0.0864 -0.0867 -0.0931 -0.0873 -0.0888
-0.0848 -0.0824 -0.0839 -0.0873 -0.0888 -0.0864 -0.0531 0.6284 0.8475 0.3339 0.0116 -0.0836 -0.0906 -0.0879 -0.0876 -0.0876 -0.0851 -0.0876 -0.0916 -0.0873 -0.0882
|
Attachment 5: ER-580-200-30kV-m23.txt
|
value (Voltage offset 0.0000)
80.0000
0.7000
80.0000
29978.0220
21
21
./emit.dat
-0.0937 -0.0977 -0.0861 -0.0876 -0.0943 -0.0958 -0.0946 0.0430 0.1172 0.0085 -0.0690 -0.0882 -0.0879 -0.0888 -0.0891 -0.0876 -0.0858 -0.0906 -0.0992 -0.0922 -0.0925
-0.0943 -0.0949 -0.0861 -0.0851 -0.0940 -0.0967 -0.0934 0.0433 0.1163 0.0049 -0.0681 -0.0897 -0.0879 -0.0903 -0.0912 -0.0891 -0.0848 -0.0891 -0.1010 -0.0912 -0.0928
-0.0949 -0.0977 -0.0854 -0.0867 -0.0879 -0.0891 -0.0873 0.0446 0.1178 0.0137 -0.0641 -0.0879 -0.0900 -0.0909 -0.0922 -0.0937 -0.0867 -0.0882 -0.0964 -0.0864 -0.0876
-0.0928 -0.0989 -0.0888 -0.0830 -0.0876 -0.0885 -0.0848 0.0397 0.1233 0.0137 -0.0653 -0.0885 -0.0900 -0.0894 -0.0888 -0.0888 -0.0888 -0.0882 -0.0958 -0.0858 -0.0891
-0.0912 -0.0952 -0.0891 -0.0861 -0.0864 -0.0891 -0.0851 0.0406 0.1218 0.0134 -0.0671 -0.0873 -0.0864 -0.0867 -0.0873 -0.0888 -0.0854 -0.0894 -0.0970 -0.0864 -0.0897
-0.0946 -0.0955 -0.0854 -0.0879 -0.0888 -0.0888 -0.0864 0.0369 0.1190 0.0104 -0.0662 -0.0888 -0.0891 -0.0891 -0.0870 -0.0879 -0.0882 -0.0867 -0.0989 -0.0867 -0.0873
-0.0940 -0.0958 -0.0851 -0.0867 -0.0873 -0.0864 -0.0845 0.0400 0.1181 0.0098 -0.0662 -0.0894 -0.0879 -0.0900 -0.0876 -0.0900 -0.0864 -0.0851 -0.1001 -0.0891 -0.0900
-0.0955 -0.0977 -0.0885 -0.0873 -0.0873 -0.0845 -0.0836 0.0427 0.1202 0.0128 -0.0650 -0.0888 -0.0882 -0.0885 -0.0888 -0.0888 -0.0854 -0.0848 -0.0977 -0.0864 -0.0900
-0.0946 -0.0995 -0.0873 -0.0934 -0.0894 -0.0848 -0.0854 0.0427 0.1190 0.0150 -0.0665 -0.0888 -0.0900 -0.0858 -0.0867 -0.0867 -0.0861 -0.0870 -0.0961 -0.0858 -0.0888
-0.0925 -0.0983 -0.0879 -0.0891 -0.0891 -0.0864 -0.0854 0.0391 0.1175 0.0119 -0.0665 -0.0882 -0.0916 -0.0854 -0.0848 -0.0861 -0.0870 -0.0897 -0.0964 -0.0879 -0.0858
-0.0912 -0.0949 -0.0870 -0.0952 -0.0879 -0.0879 -0.0864 0.0357 0.1169 0.0095 -0.0674 -0.0894 -0.0897 -0.0888 -0.0851 -0.0867 -0.0891 -0.0891 -0.0986 -0.0891 -0.0897
-0.0943 -0.0946 -0.0864 -0.0876 -0.0867 -0.0864 -0.0861 0.0391 0.1205 0.0113 -0.0665 -0.0885 -0.0882 -0.0861 -0.0885 -0.0845 -0.0870 -0.0891 -0.0970 -0.0876 -0.0873
-0.0952 -0.0961 -0.0864 -0.0946 -0.0845 -0.0854 -0.0830 0.0381 0.1224 0.0116 -0.0644 -0.0885 -0.0827 -0.0873 -0.0897 -0.0873 -0.0851 -0.0873 -0.0949 -0.0876 -0.0894
-0.0934 -0.0983 -0.0888 -0.0958 -0.0867 -0.0867 -0.0839 0.0372 0.1233 0.0104 -0.0659 -0.0888 -0.0900 -0.0864 -0.0885 -0.0897 -0.0870 -0.0867 -0.0958 -0.0861 -0.0867
-0.0922 -0.0983 -0.0885 -0.0925 -0.0900 -0.0891 -0.0848 0.0391 0.1212 0.0104 -0.0653 -0.0891 -0.0900 -0.0867 -0.0888 -0.0885 -0.0894 -0.0854 -0.0977 -0.0861 -0.0873
-0.0922 -0.0955 -0.0870 -0.0876 -0.0885 -0.0888 -0.0848 0.0345 0.1199 0.0079 -0.0656 -0.0909 -0.0876 -0.0873 -0.0867 -0.0885 -0.0888 -0.0873 -0.0970 -0.0870 -0.0873
-0.0937 -0.0952 -0.0864 -0.0906 -0.0848 -0.0861 -0.0861 0.0351 0.1221 0.0073 -0.0656 -0.0903 -0.0861 -0.0867 -0.0861 -0.0876 -0.0861 -0.0873 -0.0964 -0.0879 -0.0891
-0.0946 -0.0964 -0.0867 -0.0885 -0.0885 -0.0864 -0.0830 0.0391 0.1230 0.0085 -0.0662 -0.0885 -0.0906 -0.0879 -0.0882 -0.0879 -0.0854 -0.0876 -0.0949 -0.0861 -0.0903
-0.0946 -0.0980 -0.0894 -0.0937 -0.0873 -0.0867 -0.0815 0.0385 0.1224 0.0101 -0.0665 -0.0873 -0.0897 -0.0861 -0.0867 -0.0885 -0.0888 -0.0882 -0.0937 -0.0864 -0.0876
-0.0928 -0.0964 -0.0900 -0.0970 -0.0854 -0.0885 -0.0842 0.0372 0.1224 0.0085 -0.0665 -0.0888 -0.0894 -0.0867 -0.0858 -0.0873 -0.0897 -0.0912 -0.0961 -0.0885 -0.0870
-0.0922 -0.0967 -0.0873 -0.0876 -0.0876 -0.0882 -0.0864 0.0366 0.1212 0.0082 -0.0674 -0.0928 -0.0854 -0.0858 -0.0864 -0.0873 -0.0867 -0.0897 -0.0974 -0.0897 -0.0867
|
Attachment 6: ITW-TM2-Nb#08-HP-SIS_Test_Stand.docx
|
3
|
Wednesday, September 14, 2011, 15:40 |
Anders Mjos | TaC#1 | LP-SIS | ITE | TM1 | 38634-6 | 36833-6 | In progress | TaC#1 installed and pumping down. | Wednesday, September 14, 2011, 15:39: Vacuum at 1.3e-5 torr
Thursday, September 15, 2011, 08:59: Vacuum at 2.9e-7 torr. Water flowing and vacuum stable. |
6
|
Friday, September 16, 2011, 12:39 |
Aurelia Laxdal | TaC#1 | LP-SIS | ITE | TM1 | 38634-6 | 36833-6 | In progress | HV conditioning | Started the HV conditioning: TSG:BIAS=30kV and TSH:EE=-700V. |
|