1044
|
Tuesday, January 07, 2025, 17:06 |
Dave Neilson | | LP-FEBIAD | | | 44692-4 | 42972-10 | In progress | FEBIAD development | Tuesday, January 07, 2025, 16:00 - Target assembly installed in test stand. electrical connections assessed:
Prior to installation: oven and cathode isolated from heatshield. no shorts
After installation: Oven and Cathode 35kΩ from HVC.
Tuesday, January 07, 2025, 16:05 - pumping down to rough vacuum
Tuesday, January 07, 2025, 16:45 TP1B on. TP1A on. Pumping down
Prior to proceeding, ensure that ALL water cooling is turned on, as it was all drained and empty for the past 2 weeks.
Tuesday, January 07, 2025, 17:10 Beam line TP1 and TP2 on.
Wednesday, January 08, 2025, 09:01 chamber pressure at 1.0E-6 T, beamline pressure at 3.7E-7 T
Wednesday, January 08, 2025, 10:32 - water cooling all on. no effect on vacuum pressure.
Wednesday, January 08, 2025, 10:33 - Increased argon pressure in the gas line has no effect on vacuum pressure. it is registered on the pressure gauge at feedthrough.-
- However, roughing pump doesn't seem to be venting the argon gas line. I disconnected the copper line to the venting valve to confirm pump activity. - no activity detectable.
- removed pump from assembly. switching on by itself shows no evidence of pump functionality. When on, pump is heating to very hot temperatures. I've emailed Edi at vacuum group to investigate further.
Wednesday, January 08, 2025, 15:02 - Vacuum group says that the pump is dead. We can replace it if we'd like, but i've created a temporary solution by connecting the argon venting valve to the backing pumps. For the sake of being careful (although this may be overly cautious), we must avoid opening this venting valve without first isolating the connection from the turbo pumps via closing TSG:PV1
Wednesday, January 08, 2025, 15:04 ramping up COIL to ensure conditioning against outgassing later. vacuum appears to be stabilizing below 2e-6 T
Wednesday, January 08, 2025, 16:32 vacuum recovered, beginning to heat at 1A/30s up to 580A/230A TGHT/TBHT
Wednesday, January 08, 2025, 22:32 with COIL on, the vacuum pressure increased to ~2.2E-6 T before beginning to recover, leaving on to encourage vacuum recovery.
Thursday, January 09, 2025, 08:27 Vacuum at 1.0E-6 T. COIL off.
Thursday, January 09, 2025, 10:15 Argon gas line at 0.5 bar.a
Thursday, January 09, 2025, 10:56 increasing BIAS/EE/EZL to 30/1.5/13 kV
Thursday, January 09, 2025, 11:19 turning on optics elements. TMS:Q2 is timed out and possibly unalive. - must be investigated
Thursday, January 09, 2025, 12:38 HV at 30kV. monitoring for spark activity
Thursday, January 09, 2025, 12:53 over 15 minutes, there was just one possible spark. I see also that TPS:Q2 (quadrupole prior to FC0) seems to be half-down. i can't seem to get it to stop timing out and recharge. must investigate further
after investigating further and attempting to reset/reconfigure. I tried power cycling the canbus controller as well. In the end, controls group determined that this controller needed to be replaced.
Thursday, January 09, 2025, 14:34 all optics set. setting COIL CATHODE to 40A 120V. CATHODE is leaking 0 current. (0.000 mA readback).
Thursday, January 09, 2025, 14:37 reading 2.91e-8A at FC0 . found Argon peak at 40 exactly. MB1 current 96.4561 A
Thursday, January 09, 2025, 14:42 using COIL CATHODE scan program to optimize. (data file measure_id=9106) A local maximum was identified at 140V with 40A coil current. However, a short seems to show up when the cathode voltage reaches ~130V. it lets through only ~10mA at 130 and then will release up to at least 100mA at 140V.
As time passes, the sustained voltage drops over time, given a current limit of 100mA. it dropped to ~50V. On subsequent increases of the power supply, current begins leaking through at ~65V.
Given this, I will continue (while monitoring the cathode power supply) with a cathode coil of 50V 90A.
Thursday, January 09, 2025, 15:28 FC0 = 2.21 e-8 A. FC6 at 3.21e-10A RPM taken for all 4.
Thursday, January 09, 2025, 16:04 mass scan complete - 2025_01_09-16_01 FEBIAD no retainer 1.txt CATHODE voltage was stable at 50V throughout.
Thursday, January 09, 2025, 16:07 decreasing BIAS/EE/EZL to 0, 0, 0
Friday, January 10, 2025, 09:09 ramping BIAS/EE/EZL to 30/1.5/13 kV
Friday, January 10, 2025, 11:51 Increasing COIL and CATHODE to 90A and 50V (was able to ramp cathode up to 100V before seeing current loss)
Friday, January 10, 2025, 11:53 FC0 = 1.7e-8 A. FC6 = 4.9e-10 A wire scans taken.
Friday, January 10, 2025, 12:00 Beginning mass scan with COIL CATHODE at 90A 50V - 2025_01_10-12_28 FEBIAD no retainer 2.txt
Friday, January 10, 2025, 12:28 decreasing BIAS/EE/EZL to 0, 0, 0
Friday, January 10, 2025, 13:15 gas inflow OFF (at rough vacuum pressure). Ramping up again BIAS/EE/EZL to 30/1.5/13 kV
Friday, January 10, 2025, 14:27 FC0 = 1.22e-8A. FC6 = 2.93e-12A (at 40) . mass scan taken = 2025_01_10-14_18 FEBIAD no retainer 3 argon off.txt wire scans taken.
Friday, January 10, 2025, 14:28 TGHT TBHT cooling to 0A 0A at 1A/30s
Monday, January 13, 2025, 11:00 testing cathode isolation on cold target by ramping up cathode voltage until breakdown. At 120V, CATHODE current is 2.4mA. (125V, 8.5mA), at 130V, current limiting and dropping voltage.
Monday, January 13, 2025, 11:10 power supplies off. Winding down turbo pumps for chamber evac
Monday, January 13, 2025, 13:07 venting
|
|
717
|
Monday, April 15, 2019, 14:42 |
Aaron Schmidt | | LP-SIS | | | | 42972-1 | Done | Conditioning UC#27 HS & test FS#3 | New coil lines installed for P2N testing later.
Monday, April 15, 2019, 14:15: Roughing.
Monday, April 15, 2019, 14:49: Turbo pumps on.
Tuesday, April 16, 2019, 08:20: RGA on.
Gradually open each line. Wait 5 minutes & monitor the response.
Tuesday, April 16, 2019, 08:30: Open Return & Supply main valves.
Tuesday, April 16, 2019, 08:32: Open FS#5. Vacuum @ 4.29e-06T. No response on RGA (water).
Tuesday, April 16, 2019, 08:37: Vacuum @ 4.37e-06T
Tuesday, April 16, 2019, 08:38: Open FS#4. Vacuum @ 4.37e-06T. No response on RGA (water).
Tuesday, April 16, 2019, 08:43 Vacuum @ 4.29e-06T
Tuesday, April 16, 2019, 08:44 Open FS#3. Vacuum @ 4.29e-06T. No response on RGA (water).
Tuesday, April 16, 2019, 08:49: Vacuum @ 4.26e-06T.
Tuesday, April 16, 2019, 08:50: Open FS#2. Vacuum @ 4.29e-06T. No response on RGA (water).
Tuesday, April 16, 2019, 08:57: Vacuum @ 4.29e-06T.
Wednesday, April 17, 2019, 08:30: Power/sources on. Vacuum @ 2.46e-06T.
Wednesday, April 17, 2019, 10:30: Auto-ramping TGHT to 150 @ 2A per min. Vacuum limit increased to 3.000e-06T.
Wednesday, April 17, 2019, 11:53: Auto-ramping TGHT to 300 @ 2A per min. Vacuum @ 2.50e-06T.
Wednesday, April 17, 2019, 13:10: Auto-ramping TGHT to 350 @ 2A per min. Vacuum @ 2.95e-06T.
Thursday, April 18, 2019, 07:23: Auto-ramping TGHT to 575 @ 2A per min. Vacuum @ 2.40e-06T.
Tuesday, April 23, 2019, 06:45: Auto-Cooling TGHT @ 2A per min. Vacuum @ 9.22e-07T.
Tuesday, April 23, 2019, 08:29: RGA error. Shut off RGA.
Tuesday, April 23, 2019, 11:34: Power/sources off.
Tuesday, April 23, 2019, 11:37: IG1 off. Turbos off.
Tuesday, April 23, 2019, 12:53: Venting. |
Attachment 1: UC#27_&_FS#3_check.docx
|
Attachment 2: TS-2019-04-16-08-20.rga
|
|
999
|
Thursday, March 14, 2024, 17:43 |
Dave Neilson | | LP-FEBIAD | | | 44692-4 | 42972 | | Calibrated Leak Development | Thursday, March 14, 2024, 15:39: Roughing
Thursday, March 14, 2024, 16:20:TP1B on
Thursday, March 14, 2024, 16:40: TP1A on
Thursday, March 14, 2024, 16:51: IG1 on
Thursday, March 14, 2024, 17:28: Waterline FS#5 open
Friday, March 15, 2024, 13:19: Gradually increased BIAS to 25kV, then up to 50kV more slowly. Accompanied by increase in EZL voltage.
Friday, March 15, 2024, 13:50: Periodically increased BIAS voltage in 100V increments watching for sparks. Decrease BIAS whenever sparks. settled at 51.5 kV (cannot increase further without sparking), left at high voltage.
Friday, March 15, 2024, 15:48: Decreasing BIAS to 10kV
Friday, March 15, 2024, 16:28: ran coil at 80A until vacuum recovered.
Friday, March 15, 2024, 19:29: decrease coil to 0A, decrease BIAS to 0V
Monday, March 18, 2024, 10:08 increasing TGHT/TBHT to 440A/290A at 1A/30s
Monday, March 18, 2024, 16:00 increasing COIL to 40A, waiting for vacuum to return to normal
Monday, March 18, 2024, 21:05: decreasing COIL to 0A
Tuesday, March 19, 2024, 09:35: increased COIL to 40A.
Tuesday, March 19, 2024, 10:17: increased TDH:ANODE (cathode mode) to (-)200V
Tuesday, March 19, 2024, 10:20: increasing BIAS to 30kV
Tuesday, March 19, 2024, 13:12: turned on gas. vacuum jumped up and tripped everything off. gas was immediately turned back off. TGHT and TBHT were off for approximately 1 minute before being turned back on.
Tuesday, March 19, 2024, 13:46: cooling. need to investigate possible leaks.
Tuesday, March 19, 2024, 15:58: TGHT and TBHT at 0A. Turbo pumps turned off. waiting to slow before venting.
Tuesday, March 19, 2024, 16:20 venting
Wednesday, March 20, 2024, 14:47: Target assembly was removed today and the tube from the inlet to the calibrated leak was replaced. Now reinstalled, roughing. Gas was turned on at ~0.2 bar during roughing. No effect on vacuum.
water cooling also turned on.
Wednesday, March 20, 2024, 14:53 Turbo pumps turned on.
Wednesday, March 20, 2024, 20:24: begin ramping TGHT TBHT to 440/290A at 1A/30s
Thursday, March 21, 2024, 09:54: Target heated, vacuum good. Tested gas to ensure there is no noticeable effect on vacuum. Lines subsequently vented through the line roughing pump.
Thursday, March 21, 2024, 09:55 Optics switched on. increasing cathode to 200V and coil to 40A. Watching vacuum, checking beam tune
Thursday, March 21, 2024, 10:20 ramping BIAS to 30kV
Thursday, March 21, 2024, 11:14 : scanned, data recorded
Thursday, March 21, 2024, 11:14 cathode reduced to 100V. scanned
Thursday, March 21, 2024, 11:32: coil reduced to 20A. scanned.
Thursday, March 21, 2024, 11:50 cathode increased to 200V. scanned
Thursday, March 21, 2024, 12:07 coil increased to 40A. Argon gas on (-0.8bar at regulator). scanned
Thursday, March 21, 2024, 12:31 reducing bias to 0kV
Thursday, March 21, 2024, 13:02 argon regulator pressure increased to -.64bar
Thursday, March 21, 2024, 13:05 increasing bias to 30kV
Thursday, March 21, 2024, 13:15: vented lines, scanning
Thursday, March 21, 2024, 13:35 gas on. scanning
Thursday, March 21, 2024, 15:05 cathode at 100V, scanning
Thursday, March 21, 2024, 15:20 coil at 20A, scanning
Thursday, March 21, 2024, 15:55 coil off. scanning
Thursday, March 21, 2024, 16:10 cathode off, scanning
Thursday, March 21, 2024, 17:23 - bias down. changing gas pressure
Friday, March 22, 2024, 09:08 increasing bias, checking tune, increasing cathode to 200V, coil to 40A, EZL to 13kV
Friday, March 22, 2024, 09:48 gas vented, scanned, gas on, scanned.
Friday, March 22, 2024, 09:48 decreasing bias to 0V, changing gas pressure.
Friday, March 22, 2024, 10:12 pressure at 1bar.abs. increased bias to 30kV, checked beam tune, cathode to 200V, coil to 40A, EZL to 13kV
Friday, March 22, 2024, 10:13 gas on. scanning
Friday, March 22, 2024, 10:33 decrease bias, changing pressure.
Friday, March 22, 2024, 10:57 gas line pressure at 1.5bar.a. increasing bias to 30kV, checking beam tune, cathode to 200V, coil to 40A, EZL to 13kV
Friday, March 22, 2024, 11:42 decreased bias, changing pressure.
Friday, March 22, 2024, 11:52 gas line pressure at ~2bar.a increasing bias to 30kV, checking beam tune, cathode to 200V, coil to 40A, EZL to 13kV
Friday, March 22, 2024, 15:40 bias down to 0V. reducing TGHT TBHT to 0/0
Monday, March 25, 2024, 09:30 Turbo pumps turned off
Monday, March 25, 2024, 10:00 venting chamber, purged water lines.
Monday, March 25, 2024,11:00 removed target assembly
Monday, March 25, 2024,13:00 replaced target assembly, began roughing
Monday, March 25, 2024,13:50: turned on water cooling. Vacuum stable. Turned on turbo pumps.
Monday, March 25, 2024, 15:16 turned on optics elements, cathode to 200V, coil to 80A, EZL to 13kV. monitoring vacuum. Gas leak test shows no effect on vacuum (5e-6T).
Monday, March 25, 2024, 15:22: Ramping TGHT TBHT to 440/290A at 1A/30s
Tuesday, March 26, 2024, 10:32 Ran mass scans with gas off and on.
Tuesday, March 26, 2024, 10:32 scanning parameter space for optimal beam output - cathode voltage and coil current.
Tuesday, March 26, 2024, 15:29 ramping TBHT to 300A, scanning
Tuesday, March 26, 2024, 16:10 ramping TBHT to 310A, scanning
Tuesday, March 26, 2024, 16:40 ramping TBHT to 320A, scanning
Wednesday, March 27, 2024, 09:25 - ramping bias to 30kV
Wednesday, March 27, 2024, 09:40 scanning cathode / coil parameter space to optimize output --- highest output at 200V 60A - 200V 62A - magnet drifted and is not right at Ar peak. corrected. scan repeated
Wednesday, March 27, 2024, 10:31 scanning cathode / coil parameter space to optimize output
Wednesday, March 27, 2024, 11:02 mass scan with 200V 62A cathode coil.
Wednesday, March 27, 2024, 11:17 - lines evacuated. mass scan.
Wednesday, March 27, 2024, 11:40 bias to 0V. change gas pressure
Wednesday, March 27, 2024, 11:50 bias to 30kV. mass scan
Wednesday, March 27, 2024, 12:10 bias to 0V. change gas pressure
Wednesday, March 27, 2024, 12:21 bias to 30kV. mass scan
Wednesday, March 27, 2024, 12:31 bias to 0V. change gas pressure
Wednesday, March 27, 2024, 12:43 Ramping bias to 30kV.
Wednesday, March 27, 2024, 15:00 bias to 0V. changing pressure to .5 bar.
Wednesday, March 27, 2024, 15:13 bias at 30kV.
Wednesday, March 27, 2024, 15:13 increasing TGHT to 468A (~2250W) - mass scan - monitoring vacuum. stable at 6e-7T
Wednesday, March 27, 2024, 15:39 increasing TGHT to 490A (~2500W) - mass scan - monitoring vacuum. stable at 6e-7T
Wednesday, March 27, 2024, 16:23 reduced cathode voltage to 50V to record the electron current (17.09mA). mass scan taken with coil at 35A
Wednesday, March 27, 2024, 16:35 ramping down BIAS to 0V.
Wednesday, March 27, 2024, 16:38 decreasing TGHT TBHT to 0A 0A at 1A/30s in preparation to switch to anode mode in the morning.
Thursday, March 28, 2024, 13:59 - Switched to anode mode
Thursday, March 28, 2024, 13:59 - ramping TGHT and TBHT to 100A at 1A/30s
Thursday, March 28, 2024, 14:37 increasing BIAS to 30kV
Thursday, March 28, 2024, 14:47 - ramping TGHT/TBHT to 440A/290A at 1A/7s
Thursday, March 28, 2024, 15:56 mass scan with 200V 62A cathode coil. gas on at 0.5 bar. (bias at 29.8kV)
Thursday, March 28, 2024, 15:56 mass scan with 200V 62A cathode coil. gas vented. (bias at 29.8kV)
Thursday, March 28, 2024, 16:29 ramping TGHT/TBHT to 490A/320A at 1.4A/10s
Thursday, March 28, 2024, 16:36 mass scan with 200V 62A cathode coil. gas on at 0.5 bar. (bias at 29.8kV)
Thursday, March 28, 2024, 16:56 mass scan with 200V 62A cathode coil. gas vented. (bias at 29.8kV)
Thursday, March 28, 2024, 17:13 dropping bias to 0
Thursday, March 28, 2024, 17:14 cooling tght and tbht at 1A/30s |
|
685
|
Monday, September 17, 2018, 11:52 |
Aaron Schmidt | UC#25 | LP-SIS | ITW | TM2 | | 42752-9 | In progress | Conditioning Heat Shield for UC#25 | Monday, September 17, 2018, 11:47: Roughing*. Doors not sealing, help acq'd
Monday, September 17, 2018, 11:51: Roughing. Doors sealed
Monday, September 17, 2018, 12:30: Turbo pumps started.
Monday, September 17, 2018, 14:09: RGA on.
Tuesday, September 18, 2018, 09:01: Waterlines opened. No response on RGA. Vacuum ok @ 5.92e-07T.
Tuesday, September 18, 2018, 09:02: Power on
Tuesday, September 18, 2018, 09:05: Sources on.
Tuesday, September 18, 2018, 09:08: Auto-ramping TGHT / TBHT to 580 / 230 A @ 2A per min
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
24.8 |
4.02 |
TBHT |
24.4 |
3.11 |
COIL + EE |
24.6 |
5.15 |
TGHT |
24.6 |
5.79 |
Wednesday, September 19, 2018, 08:14: Vacuum @ 7.73e-07T.
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
34.8 |
4.02 |
TBHT |
24.7 |
3.10 |
COIL + EE |
24.9 |
5.13 |
TGHT |
26.8 |
5.77 |
Wednesday, September 19, 2018, 14:30: Auto-cooling @ 2A per min.
Friday, September 21, 2018, 07:52: Power / Sources off.
Friday, September 21, 2018, 07:53: RGA off. Ion Gauge off. Turbo pumps off.
Friday, September 21, 2018, 07:56: Waterlines purged. TS locked out.
|
Attachment 1: OMEGA_OM-DAQPRO-5300_24-Sep-18_08.08.26.dlp
|
Attachment 2: TS-2018-09-17-14-10.rga
|
Attachment 3: UC#25.docx
|
|
718
|
Tuesday, April 23, 2019, 14:39 |
Aaron Schmidt | | LP-SIS | | | | 42752-8 | Done | Conditioning UC#28 HS | Tuesday, April 23, 2019, 14:39: Roughing.
Tuesday, April 23, 2019, 15:09: Turbo pumps started.
Tuesday, April 23, 2019, 15:18: Ion gauge on.
Wednesday, April 24, 2019, 07:33: RGA on.
Wednesday, April 24, 2019, 07:37: Waterlines opened. Vacuum ok @ 7.99e-07T. No response on RGA
Wednesday, April 24, 2019, 07:41: Power / sources on.
Wednesday, April 24, 2019, 07:43: Auto-ramping TGHT to 575 @ 2A per min. Vacuum limit remains at 3.000e-06T.
Thursday, April 25, 2019, 14:07: Auto-cooling @ 2 A per min.
Friday, April 26, 2019, 07:53: sources off.
Friday, April 26, 2019, 07:55: RGA off. Power off.
Friday, April 26, 2019, 07:57: Turbos off.
Friday, April 26, 2019, 08:00: HS waterlines purged. |
Attachment 1: TS-2019-04-24-07-30.rga
|
Attachment 2: UC#28_HS.docx
|
|
753
|
Thursday, August 08, 2019, 09:12 |
Aaron Schmidt | | LP-FEBIAD | | | | 42752-7 | In progress | P2N Setup - LP-FEBIAD target | Inserted a used (heating) LP-FEBIAD target with a basic heat shield assembly with no end plate foil on the window side to observe heating. The P2N Coil / TGHT connections were added.
Test whether the TS connections are affecting the P2N heating or if it's the P2N target itself.
-Heat the TGHT only, observe, then ramp down.
-Heat Coil only, observe, then ramp down.
-Heat TGHT w/ Coil, observe, then ramp down.
Note* Heating should be even across the target. Elog & take pictures.
NO HV ALLOWED!!!!
Thursday, August 08, 2019, 09:05: Roughing.
Thursday, August 08, 2019, 09:14: BV1A closed to allow pressure stabilization. Opened about within 2 seconds later with BV1B.
Thursday, August 08, 2019, 09:41: Turbo pumps on.
Thursday, August 08, 2019, 09:47: Ion gauge on.
Thursday, August 08, 2019, 10:37: RGA on.
TSG:CG1A & TSG:CG1B do not appear to be pumping down efficiently - May need to swap out TSG:RP1 after this test. Last done April 18, 2018. elog: 655.
Friday, August 09, 2019, 08:49: Power on.
Friday, August 09, 2019, 08:51: Waterlines opened. No response on RGA. Vacuum ok @ 3.03e-06T
Friday, August 09, 2019, 08:56: Auto-ramping TGHT @ 4 @ per min to 150 A.
Friday, August 09, 2019, 10:10: Ramping the TGHT down from 150 A to 0A.
Friday, August 09, 2019, 10:39: Anode Current @ 200 mA w/ 19V? Anode PS on. Voltage & current dropped off.
Friday, August 09, 2019, 11:18: TGHT and TBHT power supplies ramped to 150A each, total of 300A in the target
Friday, August 09, 2019, 13:12: Target glows symmetrically. TGHT and TBHT ramped from 150A to 200A each for a total of 400A on the target
Friday, August 09, 2019, 14:05: RGA error.
Friday, August 09, 2019, 14:08: RGA Restarted.
Friday, August 09, 2019, 14:08: Target heating evenly on total 400 A.
Friday, August 09, 2019, 15:57: Auto-ramping TGHT/TBHT @ 4 @ per min to 250/250 A.
Friday, August 09, 2019, 16:28: Vacuum limit reach at 241 and 238 A for TGHT and TBHT, respectively.
Friday, August 09, 2019, 16:28: Auto-cooling started at 4A/min
Monday, August 12, 2019, 09:22: Sources off. RGA off.
Monday, August 12, 2019, 10:04: Power off. Ion gauge off. Turbos off. |
Attachment 1: 20190808_085752.jpg
|
|
Attachment 2: 20190808_085758.jpg
|
|
Attachment 3: 20190809_135736.jpg
|
|
Attachment 4: 20190809_135747.jpg
|
|
Attachment 5: 20190809_135820.jpg
|
|
Attachment 6: Aug_2019_test_target_-_P2N_-_LP-FEBIAD_setup.docx
|
Attachment 7: TS-2019-08-08-10-37.rga
|
Attachment 8: TS-2019-08-09-14-07.rga
|
|
1016
|
Wednesday, July 31, 2024, 18:50 |
Aaron Schmidt | TiC Spare | HP-FEBIAD | ITW | TM4 | 45447-2 | 42752-7 | In progress | Conditioning Spare for TiC#7 | Wednesday, July 31, 2024, 18:47: Roughing.
Wednesday, July 31, 2024, 19:24: TP1B on.
Wednesday, July 31, 2024, 19:30: TP1B at speed. Backing pressure CG1A @ 0.212T.
Wednesday, July 31, 2024, 19:33: TP1A Started. CG1A @ 0.164T.
Wednesday, July 31, 2024, 19:41: TP1A at speed. Whining a bit, but it's alive.
Wednesday, July 31, 2024, 19:42: IG1 on.
Thursday, August 01, 2024, 08:00: Vacuum @ 3.14e-0T. Waterline FS#5 open. Vacuum ok. CG1A @ 0.028T.
Thursday, August 01, 2024, 08:04: IV4 open.
Thursday, August 01, 2024, 08:09: Cage closed up. PS / Sources on. Note* TGHT/TBHT breakers were on. Anode was off. All "On" now.
Thursday, August 01, 2024, 08:13: Manual tap up checks. TGHT / TBHT @ 5 / 5 A. Anode @ 100 V / 15.873 mA. Coil PS acting up. Set to 7 A, displays 9.87A & fluctuates.
Thursday, August 01, 2024, 08:19: Anode to 0 V. Coil reduced to 0 A, however still reads back 4.15 A on epics. PS @ 0A.
Thursday, August 01, 2024, 08:24: Cage open. HV sources off. Coil back to 7A. Epics shows 9.82-10.11A (Blue). Fluctuates, PS direct reading @ 7A - Steady.
Thursday, August 01, 2024, 08:39: Cage closed up. HV sources back on.
Thursday, August 01, 2024, 08:40: BIAS / EE / EZL set to 1 / -0.5 / 1 kV. Anode to 100 V / 15.873 mA. Coil remains @ 7A. Epics 10.35A (Blue) Fluctuates still. PS direct reading @ 7A.
Thursday, August 01, 2024, 08:49: BIAS / EE / EZL @ 30 / -1.5 / 13 kV.
Thursday, August 01, 2024, 08:54: BIAS / EE / EZL @ 35 / -1.8 / 15 kV.
Thursday, August 01, 2024, 08:56: BIAS / EE / EZL @ 40 / -2.0 / 17 kV.
Thursday, August 01, 2024, 08:57: 1 Spark. External - Audible.
Thursday, August 01, 2024, 09:01: Anode @ 200 V / 17.094 mA. Vacuum @ 2.75e-06T.
Thursday, August 01, 2024, 09:04: Anode @ 300 V / 18.315 mA.
Thursday, August 01, 2024, 09:05: Anode @ 400 V / 19.536 mA.
Thursday, August 01, 2024, 09:05: Anode @ 500 V / 20.757 mA. Interesting, the coil epics read back (Blue) moved to 6.98 (7) A.The coil read back stability appears to have occurred in conjunction with the 8:57am spark*.
Thursday, August 01, 2024, 09:17: 20 min stable. Can assume the spark was the coil instability discharging. Coil remain stable reading @ 7A. BIAS / EE / EZL @ 40 / -2.0 / 17 kV (20 min). Anode @ 500 V / 19.536 mA. (10 min+).
Thursday, August 01, 2024, 09:29: BIAS / EE / EZL to 0 kV. Anode to 0.
Thursday, August 01, 2024, 09:36: Auto-ramping TGHT / TBHT to 350 / 150 A @ 2 @ per min.
Thursday, August 01, 2024, 09:37: Tap up coil to 10 A. Stable. Read back on all ends match up.
Thursday, August 01, 2024, 09:39: Coil @ 20 A. Stable. Hold. Vacuum @ 2.66e-06T.
Thursday, August 01, 2024, 10:05: Coil @ 30 A. Vacuum @ 2.66e-06T.
Thursday, August 01, 2024, 10:13: Coil @ 40 A. Vacuum @ 2.68e-06T.
Thursday, August 01, 2024, 10:18: Coil @ 50 A. Vacuum @ 2.66e-06T.
Thursday, August 01, 2024, 10:24: Coil @ 60 A. Vacuum @ 2.66e-06T.
Thursday, August 01, 2024, 10:27: Coil @ 70 A. Vacuum @ 2.77e-06T.
Thursday, August 01, 2024, 10:32: Coil @ 80 A. Vacuum @ 3.03e-06T. Note* Coil was cleaned prior to installation & acrylic within dissolved. Successful run on coil.
Thursday, August 01, 2024, 10:43: Coil to 0 A. However in the last 20 A, the read back & epics did not match again. Slowly discharging on epics, although the PS read back was accurate.
Thursday, August 01, 2024, 13:29 FBP: Auto-ramping TGHT / TBHT to 580 / 290 A @ 2 A per min . IG1:3.56e-6.
Thursday, August 01, 2024, 16:00: AS - Aborted target ramping. TGHT / TBHT @ 448 / 248 A. Vacuum limited at 5.0e-06T. Allow to out gas. Continue tomorrow am.
Friday, August 02, 2024, 06:01: Auto-ramping TGHT / TBHT to 580 / 290 A @ 2 A per min. Vacuum: 1.60e-06T.
Friday, August 02, 2024, 08:28 turning on HV
Friday, August 02, 2024, 08:35 BIAS / EE / EZL @ 30 / -1.5 / 13 kV. No issues.
Friday, August 02, 2024, 08:40 Same kind of sparks as the previous thin foil Ta observed

Friday, August 02, 2024, 08:49 Bias lowered to 20kV
Friday, August 02, 2024, 10:39 HV issues could not be identified, we recorded some sparks from within the target near the multipin and TBHT. Outgassing coil and running mass scan. Source tray to be thoroughly checked after retrieving the target.
Friday, August 02, 2024, 15:04 gradually increased COIL current to 100A over time while monitoring vacuum to ensure no heater trips. Vacuum now at 2.4e-6 T
Friday, August 02, 2024, 15:05 increasing BIAS / EE / EZL to 30 / -1.5 / 13 kV
Friday, August 02, 2024, 15:26 now at BIAS / EE / EZL to 30 / -1.5 / 13 kV --- first spark at 30kV.
Friday, August 02, 2024, 15:37 sparking occurring at 4-5min intervals. set anode to 120V, coil to 40A. finding Ar peak via scan. Argon operated at 0.5 bar.
Friday, August 02, 2024, 15:57 Ar peak located at 95.5332A. on magnet. tested gas on / off to confirm - current at FC6 = 1.37 nA
Friday, August 02, 2024, 16:02 ran febiad coil/anode scan.
Friday, August 02, 2024, 16:33 data at test number = 7462 can be further visualized using the shared jupyter notebook code. data taken at anode voltage 260V was disrupted due to a spark derailing a number of optics elements in the beamline. they were soon corrected, but this data was missed.

Friday, August 02, 2024, 16:34 - mass scan with 180V 70A (Argon ON) ---- doesn't seem to be recording background noise well. ie the common background recording artifact is present. records constant numbers instead of numbers in range (1e-13,5e-12). often appears in mass scan image at slants
Friday, August 02, 2024, 17:07 - mass scan rerun after switching off argon and pumping down the lines. however, only run from mass 49 down, as the signal above was completely blank
Friday, August 02, 2024, 17:25 ramping down bias ee and ezl
Friday, August 02, 2024, 17:45 decreasing TGHT and TBHT to 0a at 1A/30s
|
|
676
|
Friday, July 27, 2018, 14:44 |
Aaron Schmidt | UC#24 | LP-SIS | ITE | TM4 | 41177-3 | 42752-6 | In progress | Conditioning Heat Shield for UC#24 & voltage test prep | 4 Kapton wire routed into the 9-pin connector. Unused pins are grounded on the bracket.
Friday, July 27, 2018, 14:25: Roughing
Friday, July 27, 2018, 14:41: Turbo pumps on.
Friday, July 27, 2018, 15:53: RGA on.
Saturday, July 28, 2018, 07:03: Waterlines opened. No response on RGA. Vacuum steady @ 1.34e-06T
Friday, August 03, 2018, 10:30: Exposed 9-pin wires inside the cage insulated with electrical tape.
Friday, August 03, 2018, 10:35: Auto-ramping TGHT / TBHT to 580 / 225 A @ 2A per min.
Tuesday, August 07, 2018, 07:40: Small spikes but frequent noted in the RGA on Water, Nitrogen, and Oxygen. More than usual although the vacuum in reasonable at 7.0e-07T.
Tuesday, August 07, 2018, 07:49: Auto-cooling started @ 2A per min. |
|
678
|
Tuesday, August 14, 2018, 11:03 |
Aaron Schmidt | | LP-SIS | | | 41177-3 | 42752-6 | | Voltage Tests (Ta target) | Tuesday, August 14, 2018, 10:55: Auto ramping TGHT @ 6 A per min to 40 A.
Tuesday, August 14, 2018, 11:03: TGHT @ 40 A. Display @ 0.198 V. PS @ 0.18 V. Multimeter @ 033 mV.
Tuesday, August 14, 2018, 11:09: Auto ramping TGHT @ 6 A per min to 80 A.
Tuesday, August 14, 2018, 11:16: TGHT @ 80.1 A. Display @ 0.405 V. PS @ 0.38 V. Multimeter @ 084 mV.
Tuesday, August 14, 2018, 11:18: Auto ramping TGHT @ 6 A per min to 120 A.
Tuesday, August 14, 2018, 11:28: TGHT @ 120.1 A. Display @ 0.718 V. PS @ 0.69 V. Multimeter @ 187 mV.
Tuesday, August 14, 2018, 11:35: TGHT @ 120.1 A. Display @ 0.745 V. PS @ 0.72 V. Multimeter @ 201 mV.
Tuesday, August 14, 2018, 11:36: Auto ramping TGHT @ 6 A per min to 160 A.
Tuesday, August 14, 2018, 11:44: TGHT @ 160.3 A. Display @ 1.131 V. PS @ 1.11 V. Multimeter @ 340 mV.
Tuesday, August 14, 2018, 11:54 TGHT @ 160.3 A. Display @ 1.148 V. PS @ 1.12 V. Multimeter @ 347 mV.
Tuesday, August 14, 2018, 11:54: Auto ramping TGHT @ 6 A per min to 200 A.
Tuesday, August 14, 2018, 12:01: TGHT @ 200.4 A. Display @ 1.565 V. PS @ 1.54 V. Multimeter @ 502 mV.
Tuesday, August 14, 2018, 12:47 TGHT @ 200.4 A. Display @ 1.582 V. PS @ 1.56 V. Multimeter @ 508 mV.
Tuesday, August 14, 2018, 12:47: Auto ramping TGHT @ 6 A per min to 230A.
Tuesday, August 14, 2018, 12:54: TGHT @ 230.6 A. Display @ 1.924 V. PS @ 1.89 V. Multimeter @ 635 mV.
Tuesday, August 14, 2018, 13:00: TGHT @ 230.6 A. Display @ 1.929 V. PS @ 1.90 V. Multimeter @ 638 mV.
Tuesday, August 14, 2018, 13:01: Auto ramping TGHT @ 6 A per min to 240A.
Tuesday, August 14, 2018, 13:05: TGHT @ 240.6 A. Display @ 2.046 V. PS @ 2.02 V. Multimeter @ 682 mV.
Tuesday, August 14, 2018, 13:10: TGHT @ 240.6 A. Display @ 2.049 V. PS @ 2.02 V. Multimeter @ 683 mV.
Tuesday, August 14, 2018, 13:11 Auto ramping TGHT @ 6 A per min to 280A.
Tuesday, August 14, 2018, 13:19: TGHT @ 280.7 A. Display @ 2.537 V. PS @ 2.51 V. Multimeter @ 869 mV.
Tuesday, August 14, 2018, 13:24: TGHT @ 280.7 A. Display @ 2.540 V. PS @ 2.51 V. Multimeter @ 871 mV.
Tuesday, August 14, 2018, 13:25 Auto ramping TGHT @ 6 A per min to 300A.
Tuesday, August 14, 2018, 13:30: TGHT @ 300.8 A. Display @ 2.794 V. PS @ 2.76 V. Multimeter @ 968 mV.
Tuesday, August 14, 2018, 13:34: TGHT @ 300.8 A. Display @ 2.796 V. PS @ 2.77 V. Multimeter @ 970 mV.
Tuesday, August 14, 2018, 13:36: Auto ramping TGHT @ 6 A per min to 354A.
Tuesday, August 14, 2018, 13:46: TGHT @ 354.9 A. Display @ 3.509 V. PS @ 3.48 V. Multimeter @ 1248 mV.
Tuesday, August 14, 2018, 13:51: TGHT @ 354.9 A. Display @ 3.512 V. PS @ 3.48 V. Multimeter @ 1249 mV.
Tuesday, August 14, 2018, 13:52 Auto ramping TGHT @ 6 A per min to 360A.
Tuesday, August 14, 2018, 13:55: TGHT @ 360.9 A. Display @ 3.592 V. PS @ 3.56 V. Multimeter @ 1281 mV.
Tuesday, August 14, 2018, 13:56 Auto ramping TGHT @ 6 A per min to 370A.
Tuesday, August 14, 2018, 14:00 TGHT @ 370.8 A. Display @ 3.729 V. PS @ 3.70 V. Multimeter @ 1336 mV.
Tuesday, August 14, 2018, 14:03: Auto ramping TGHT @ 6 A per min to 400A.
Tuesday, August 14, 2018, 14:10 TGHT @ 401.0 A. Display @ 4.149 V. PS @ 4.12 V. Multimeter @ 1501 mV.
Tuesday, August 14, 2018, 14:15: Auto ramping TGHT @ 6 A per min to 410A.
Tuesday, August 14, 2018, 14:19: TGHT @ 411.0 A. Display @ 4.291 V. PS @ 4.26 V. Multimeter @ 1558 mV.
Tuesday, August 14, 2018, 14:23: Auto ramping TGHT @ 6 A per min to 427A.
Tuesday, August 14, 2018, 14:28: TGHT @ 428.0 A. Display @ 4.537 V. PS @ 4.51 V. Multimeter @ 1657 mV.
Tuesday, August 14, 2018, 14:32: Auto ramping TGHT @ 6 A per min to 451A.
Tuesday, August 14, 2018, 14:38: TGHT @ 452.2 A. Display @ 4.891 V. PS @ 4.86 V. Multimeter @ 1805 mV.
Tuesday, August 14, 2018, 14:42: Auto ramping TGHT @ 6 A per min to 460A.
Tuesday, August 14, 2018, 14:47: TGHT @ 461.2 A. Display @ 5.026 V. PS @ 4.99 V. Multimeter @ 1862 mV.
Tuesday, August 14, 2018, 14:52: Auto ramping TGHT @ 6 A per min to 475A.
Tuesday, August 14, 2018, 14:55: TGHT @ 476.3 A. Display @ 5.250 V. PS @ 5.22 V. Multimeter @ 1962 mV.
Tuesday, August 14, 2018, 14:58: Auto ramping TGHT @ 6 A per min to 500A.
Tuesday, August 14, 2018, 15:04: TGHT @ 501.4 A. Display @ 5.629 V. PS @ 5.60 V. Multimeter @ 2130 mV.
Tuesday, August 14, 2018, 15:08: Auto ramping TGHT @ 6 A per min to 510A.
Tuesday, August 14, 2018, 15:11: TGHT @ 511.5 A. Display @ 5.783 V. PS @ 5.75 V. Multimeter @ 2190 mV.
Tuesday, August 14, 2018, 15:13: Auto ramping TGHT @ 6 A per min to 520A.
Tuesday, August 14, 2018, 15:17: TGHT @ 521.5 A. Display @ 5.937 V. PS @ 5.91 V. Multimeter @ 2260 mV.
Wednesday, August 15, 2018, 09:29: TGHT @ 521.5 A. Display @ 5.956 V. PS @ 5.92 V. Multimeter @ 2260 mV.
Wednesday, August 15, 2018, 09:34: Auto ramping TGHT @ 6 A per min to 550A.
Wednesday, August 15, 2018, 09:39: TGHT @ 551.5 A. Display @ 6.422 V. PS @ 6.39 V. Multimeter @ 2480 mV.
Wednesday, August 15, 2018, 09:42: Auto ramping TGHT @ 6 A per min to 560A.
Wednesday, August 15, 2018, 09:44: TGHT @ 561.5 A. Display @ 6.579 V. PS @ 6.55 V. Multimeter @ 2550 mV.
Wednesday, August 15, 2018, 09:46: Auto ramping TGHT @ 6 A per min to 580A.
Wednesday, August 15, 2018, 09:53: TGHT @ 581.7 A. Display @ 6.896 V. PS @ 6.86 V. Multimeter @ 2680 mV.
Wednesday, August 15, 2018, 09:55: Auto ramping TGHT @ 6 A per min to 600A.
Wednesday, August 15, 2018, 09:59: TGHT @ 600.0 A. Display @ 7.214 V. PS @ 7.18 V. Multimeter @ 2820 mV.
Wednesday, August 15, 2018, 10:07: Auto cooling TGHT @ 6 A per min to 580A.
Wednesday, August 15, 2018, 10:10: TGHT @ 581.7 A. Display @ 6.896 V. PS @ 6.86 V. Multimeter @ 2680 mV.
Wednesday, August 15, 2018, 10:12: Auto cooling TGHT @ 6 A per min to 560A.
Wednesday, August 15, 2018, 10:17: TGHT @ 561.6 A. Display @ 6.584 V. PS @ 6.55 V. Multimeter @ 2550 mV.
Wednesday, August 15, 2018, 10:18: Auto cooling TGHT @ 6 A per min to 550A.
Wednesday, August 15, 2018, 10:23: TGHT @ 551.5 A. Display @ 6.427 V. PS @ 6.39 V. Multimeter @ 2480 mV.
Wednesday, August 15, 2018, 10:24: Auto cooling TGHT @ 6 A per min to 520A.
Wednesday, August 15, 2018, 10:33: TGHT @ 521.5 A. Display @ 5.963 V. PS @ 5.93 V. Multimeter @ 2290 mV.
Wednesday, August 15, 2018, 10:35 Auto cooling TGHT @ 6 A per min to 510A.
Wednesday, August 15, 2018, 10:39: TGHT @ 511.5 A. Display @ 5.814 V. PS @ 5.78 V. Multimeter @ 2200 mV.
Wednesday, August 15, 2018, 10:40: Auto cooling TGHT @ 6 A per min to 500A.
Wednesday, August 15, 2018, 10:45 TGHT @ 501.4 A. Display @ 5.665 V. PS @ 5.63 V. Multimeter @ 2140 mV.
Wednesday, August 15, 2018, 10:45: Auto cooling TGHT @ 6 A per min to 475A.
Wednesday, August 15, 2018, 10:52: TGHT @ 476.3 A. Display @ 5.292 V. PS @ 5.26 V. Multimeter @ 1997 mV.
Wednesday, August 15, 2018, 10:53: Auto cooling TGHT @ 6 A per min to 460A.
Wednesday, August 15, 2018, 10:59: TGHT @ 461.2 A. Display @ 5.070 V. PS @ 5.04 V. Multimeter @ 1879 mV.
Wednesday, August 15, 2018, 11:00: Auto cooling TGHT @ 6 A per min to 451A.
Wednesday, August 15, 2018, 11:05: TGHT @ 452.2 A. Display @ 4.935 V. PS @ 4.90 V. Multimeter @ 1821 mV.
Wednesday, August 15, 2018, 11:06: Auto cooling TGHT @ 6 A per min to 427A.
Wednesday, August 15, 2018, 11:15: TGHT @ 428.0 A. Display @ 4.584 V. PS @ 4.55 V. Multimeter @ 1674 mV.
Wednesday, August 15, 2018, 11:16: Auto cooling TGHT @ 6 A per min to 410A.
Wednesday, August 15, 2018, 11:20: TGHT @ 411.0 A. Display @ 4.337 V. PS @ 4.31 V. Multimeter @ 1576 mV.
Wednesday, August 15, 2018, 11:21: Auto cooling TGHT @ 6 A per min to 400A.
Wednesday, August 15, 2018, 11:25: TGHT @ 401.0 A. Display @ 4.195 V. PS @ 4.17 V. Multimeter @ 1519 mV.
Wednesday, August 15, 2018, 13:13: TGHT @ 401.0 A. Display @ 4.193 V. PS @ 4.16 V. Multimeter @ 1516 mV.
Wednesday, August 15, 2018, 13:14: Auto cooling TGHT @ 6 A per min to 370A.
Wednesday, August 15, 2018, 13:23: TGHT @ 370.8 A. Display @ 3.773 V. PS @ 3.74 V. Multimeter @ 1350 mV.
Wednesday, August 15, 2018, 13:24 Auto cooling TGHT @ 6 A per min to 360A.
Wednesday, August 15, 2018, 13:30: TGHT @ 360.9 A. Display @ 3.634 V. PS @ 3.61 V. Multimeter @ 1296 mV.
Wednesday, August 15, 2018, 13:31: Auto cooling TGHT @ 6 A per min to 354A.
Wednesday, August 15, 2018, 13:35: TGHT @ 354.9 A. Display @ 3.553 V. PS @ 3.52 V. Multimeter @ 1263 mV.
Wednesday, August 15, 2018, 13:36 Auto cooling TGHT @ 6 A per min to 300A.
Wednesday, August 15, 2018, 13:46: TGHT @ 300.8 A. Display @ 2.830 V. PS @ 2.80 V. Multimeter @ 0984 mV.
Wednesday, August 15, 2018, 13:51: TGHT @ 300.8 A. Display @ 2.828 V. PS @ 2.80 V. Multimeter @ 0981 mV.
Wednesday, August 15, 2018, 13:52: Auto cooling TGHT @ 6 A per min to 280A.
Wednesday, August 15, 2018, 13:58: TGHT @ 280.7 A. Display @ 2.569 V. PS @ 2.54 V. Multimeter @ 0882 mV.
Wednesday, August 15, 2018, 13:59: Auto cooling TGHT @ 6 A per min to 240A.
Wednesday, August 15, 2018, 14:07: TGHT @ 240.6 A. Display @ 2.073 V. PS @ 2.05 V. Multimeter @ 0685 mV.
Wednesday, August 15, 2018, 14:11: TGHT @ 240.6 A. Display @ 2.071 V. PS @ 2.04 V. Multimeter @ 0692 mV.
Wednesday, August 15, 2018, 14:12: Auto cooling TGHT @ 6 A per min to 230A.
Wednesday, August 15, 2018, 14:18: TGHT @ 230.6 A. Display @ 1.951 V. PS @ 1.92 V. Multimeter @ 0647 mV.
Wednesday, August 15, 2018, 14:19: Auto cooling TGHT @ 6 A per min to 200A.
Wednesday, August 15, 2018, 14:25: TGHT @ 200.4 A. Display @ 1.612 V. PS @ 1.59 V. Multimeter @ 0522 mV.
Wednesday, August 15, 2018, 14:31: TGHT @ 200.4 A. Display @ 1.604 V. PS @ 1.58 V. Multimeter @ 0517 mV.
Wednesday, August 15, 2018, 14:31: Auto cooling TGHT @ 6 A per min to 160A.
Wednesday, August 15, 2018, 14:39: TGHT @ 160.3 A. Display @ 1.179 V. PS @ 1.16 V. Multimeter @ 0364 mV.
Wednesday, August 15, 2018, 14:44: TGHT @ 160.3 A. Display @ 1.170 V. PS @ 1.14 V. Multimeter @ 0356 mV.
Wednesday, August 15, 2018, 14:44: Auto cooling TGHT @ 6 A per min to 120A.
Wednesday, August 15, 2018, 14:52: TGHT @ 120.1 A. Display @ 0.794 V. PS @ 0.77 V. Multimeter @ 0226 mV.
Wednesday, August 15, 2018, 14:56: TGHT @ 120.3 A. Display @ 0.779 V. PS @ 0.76 V. Multimeter @ 0218 mV.
Wednesday, August 15, 2018, 14:56: Auto cooling TGHT @ 6 A per min to 80A.
Wednesday, August 15, 2018, 15:04: TGHT @ 80.1 A. Display @ 0.466 V. PS @ 0.45 V. Multimeter @ 0117 mV.
Wednesday, August 15, 2018, 15:09: TGHT @ 80.1 A. Display @ 0.454 V. PS @ 0.43 V. Multimeter @ 0109 mV.
Wednesday, August 15, 2018, 15:13: TGHT @ 80.1 A. Display @ 0.447 V. PS @ 0.42 V. Multimeter @ 0105 mV.
Wednesday, August 15, 2018, 15:14: Auto cooling TGHT @ 6 A per min to 40A.
Wednesday, August 15, 2018, 15:22: TGHT @ 40.0 A. Display @ 0.217 V. PS @ 0.19 V. Multimeter @ 0041 mV.
Wednesday, August 15, 2018, 15:27: TGHT @ 40.0 A. Display @ 0.210 V. PS @ 0.19 V. Multimeter @ 0039 mV.
Wednesday, August 15, 2018, 15:28: Auto cooling TGHT @ 6 A per min to 0A.
------------------------------------
Thursday, August 16, 2018, 13:41: Auto ramping TGHT @ 6 A per min to 40A. Auto ramping TBHT @ 6 A per min to 40A.
Thursday, August 16, 2018, 13:52 TGHT @ 40.0 A. Display @ 0.200 V. PS @ 0.18 V. Multimeter @ 0116 mV. TBHT @ 38.4 A. Display @ 0.195 V. PS @ 0.18 V.
Thursday, August 16, 2018, 13:53 Auto ramping TGHT @ 6 A per min to 80A. Auto ramping TBHT @ 6 A per min to 80A.
Thursday, August 16, 2018, 14:02: TGHT @ 80.1 A. Display @ 0.405 V. PS @ 0.38 V. Multimeter @ 0279 mV. TBHT @ 78.5 A. Display @ 0.391 V. PS @ 0.38 V.
Thursday, August 16, 2018, 14:13: TGHT @ 80.1 A. Display @ 0.437 V. PS @ 0.41 V. Multimeter @ 0304 mV. TBHT @ 78.5 A. Display @ 0.410 V. PS @ 0.40 V.
Thursday, August 16, 2018, 14:22: TGHT @ 80.1 A. Display @ 0.444 V. PS @ 0.42 V. Multimeter @ 0313 mV. TBHT @ 78.5 A. Display @ 0.418 V. PS @ 0.40 V.
Thursday, August 16, 2018, 14:22: Auto ramping TGHT @ 6 A per min to 120A. Auto ramping TBHT @ 6 A per min to 120A.
Thursday, August 16, 2018, 14:31: TGHT @ 120.1 A. Display @ 0.735 V. PS @ 0.72 V. Multimeter @ 0610 mV. TBHT @ 118.5 A. Display @ 0.711 V. PS @ 0.71 V.
Thursday, August 16, 2018, 14:44: TGHT @ 120.1 A. Display @ 0.794 V. PS @ 0.77 V. Multimeter @ 0674 mV. TBHT @ 118.5 A. Display @ 0.767 V. PS @ 0.75 V.
Thursday, August 16, 2018, 14:45: Auto ramping TGHT @ 6 A per min to 160A. Auto ramping TBHT @ 6 A per min to 160A.
Thursday, August 16, 2018, 14:53: TGHT @ 160.3 A. Display @ 1.165 V. PS @ 1.15 V. Multimeter @ 0927 mV. TBHT @ 158.7 A. Display @ 1.197 V. PS @ 1.19 V.
Thursday, August 16, 2018, 15:01: TGHT @ 160.3 A. Display @ 1.197 V. PS @ 1.17 V. Multimeter @ 0946 mV. TBHT @ 158.7 A. Display @ 1.221 V. PS @ 1.21 V.
Thursday, August 16, 2018, 15:02: Auto ramping TGHT @ 6 A per min to 200A. Auto ramping TBHT @ 6 A per min to 200A.
Thursday, August 16, 2018, 15:10: TGHT @ 200.4 A. Display @ 1.617 V. PS @ 1.60 V. Multimeter @ 1382 mV. TBHT @ 198.8 A. Display @ 1.678 V. PS @ 1.68 V.
Thursday, August 16, 2018, 15:17: TGHT @ 200.4 A. Display @ 1.634 V. PS @ 1.61 V. Multimeter @ 1390 mV. TBHT @ 198.7 A. Display @ 1.690 V. PS @ 1.69 V.
Thursday, August 16, 2018, 15:18: Auto ramping TGHT @ 6 A per min to 230A. Auto ramping TBHT @ 6 A per min to 230A.
Thursday, August 16, 2018, 15:25: TGHT @ 230.5 A. Display @ 1.976 V. PS @ 1.95 V. Multimeter @ 1751 mV. TBHT @ 228.9 A. Display @ 2.071 V. PS @ 2.07 V.
Thursday, August 16, 2018, 15:34: TGHT @ 230.6 A. Display @ 1.985 V. PS @ 1.96 V. Multimeter @ 1754 mV. TBHT @ 228.9 A. Display @ 2.081 V. PS @ 2.08 V.
Thursday, August 16, 2018, 15:35: Auto ramping TGHT @ 6 A per min to 240A.
Thursday, August 16, 2018, 15:42: TGHT @ 240.6 A. Display @ 2.100 V. PS @ 2.07 V. Multimeter @ 1801 mV. TBHT @ 228.9 A. Display @ 2.083 V. PS @ 2.08 V.
Thursday, August 16, 2018, 15:44: Auto ramping TGHT @ 6 A per min to 280A.
Thursday, August 16, 2018, 15:52: TGHT @ 280.6 A. Display @ 2.581 V. PS @ 2.56 V. Multimeter @ 1996 mV. TBHT @ 228.9 A. Display @ 2.095 V. PS @ 2.09 V.
Thursday, August 16, 2018, 16:00: TGHT @ 280.7 A. Display @ 2.586 V. PS @ 2.56 V. Multimeter @ 1998 mV. TBHT @ 228.9 A. Display @ 2.098 V. PS @ 2.10 V.
Thursday, August 16, 2018, 16:01: Auto ramping TGHT @ 6 A per min to 300A.
Thursday, August 16, 2018, 16:06: TGHT @ 300.8 A. Display @ 2.835 V. PS @ 2.81 V. Multimeter @ 2090 mV. TBHT @ 228.9 A. Display @ 2.103 V. PS @ 2.10 V.
Thursday, August 16, 2018, 16:11: TGHT @ 300.8 A. Display @ 2.840 V. PS @ 2.81 V. Multimeter @ 2090 mV. TBHT @ 228.9 A. Display @ 2.105 V. PS @ 2.10 V.
Thursday, August 16, 2018, 16:11: Auto ramping TGHT @ 6 A per min to 354A.
Thursday, August 16, 2018, 16:22: TGHT @ 354.9 A. Display @ 3.546 V. PS @ 3.52 V. Multimeter @ 2390 mV. TBHT @ 228.9 A. Display @ 2.122 V. PS @ 2.12 V.
Thursday, August 16, 2018, 16:26: TGHT @ 354.9 A. Display @ 3.551 V. PS @ 3.52 V. Multimeter @ 2390 mV. TBHT @ 228.9 A. Display @ 2.125 V. PS @ 2.12 V.
Thursday, August 16, 2018, 16:26: Auto ramping TGHT @ 6 A per min to 360A.
Thursday, August 16, 2018, 16:32: TGHT @ 360.9 A. Display @ 3.634 V. PS @ 3.60 V. Multimeter @ 2420 mV. TBHT @ 228.9 A. Display @ 2.127 V. PS @ 2.12 V.
Thursday, August 16, 2018, 16:33: Auto ramping TGHT @ 6 A per min to 370A.
Thursday, August 16, 2018, 16:40: TGHT @ 370.8 A. Display @ 3.768 V. PS @ 3.74 V. Multimeter @ 2480 mV. TBHT @ 228.9 A. Display @ 2.129 V. PS @ 2.13 V.
Thursday, August 16, 2018, 16:41: Auto ramping TGHT @ 6 A per min to 400A.
Thursday, August 16, 2018, 16:47: TGHT @ 401.0 A. Display @ 4.183 V. PS @ 4.15 V. Multimeter @ 2650 mV. TBHT @ 228.9 A. Display @ 2.139 V. PS @ 2.14 V.
--------------------------------------------------------------
Friday, August 17, 2018, 09:12: TGHT @ 400.9 A. Display @ 4.190 V. PS @ 4.16 V. Multimeter @ 2650 mV. TBHT @ 228.9 A. Display @ 2.144 V. PS @ 2.14 V.
Friday, August 17, 2018, 09:13: Auto ramping TGHT @ 6 A per min to 410A.
Friday, August 17, 2018, 09:16: TGHT @ 410.8 A. Display @ 4.327 V. PS @ 4.30 V. Multimeter @ 2710 mV. TBHT @ 228.9 A. Display @ 2.147 V. PS @ 2.15 V.
Friday, August 17, 2018, 09:19: TGHT @ 410.8 A. Display @ 4.332 V. PS @ 4.30 V. Multimeter @ 2720 mV. TBHT @ 228.9 A. Display @ 2.149 V. PS @ 2.15 V.
Friday, August 17, 2018, 09:20: Auto ramping TGHT @ 6 A per min to 427A.
Friday, August 17, 2018, 09:24: TGHT @ 428.0 A. Display @ 4.571 V. PS @ 4.55 V. Multimeter @ 2820 mV. TBHT @ 228.9 A. Display @ 2.154 V. PS @ 2.15 V.
Friday, August 17, 2018, 09:28: TGHT @ 428.0 A. Display @ 4.574 V. PS @ 4.55 V. Multimeter @ 2820 mV. TBHT @ 228.9 A. Display @ 2.156 V. PS @ 2.15 V.
Friday, August 17, 2018, 09:29: Auto ramping TGHT @ 6 A per min to 451A.
Friday, August 17, 2018, 09:37: TGHT @ 452.0 A. Display @ 4.923 V. PS @ 4.89 V. Multimeter @ 2970 mV. TBHT @ 228.9 A. Display @ 2.166 V. PS @ 2.17 V.
Friday, August 17, 2018, 09:38: Auto ramping TGHT @ 6 A per min to 460A.
Friday, August 17, 2018, 09:44: TGHT @ 461.2 A. Display @ 5.057 V. PS @ 5.03 V. Multimeter @ 3020 mV. TBHT @ 228.9 A. Display @ 2.168 V. PS @ 2.17 V.
Friday, August 17, 2018, 09:45: Auto ramping TGHT @ 6 A per min to 475A.
Friday, August 17, 2018, 09:51: TGHT @ 476.2 A. Display @ 5.280 V. PS @ 5.25 V. Multimeter @ 3120 mV. TBHT @ 228.9 A. Display @ 2.176 V. PS @ 2.18 V.
Friday, August 17, 2018, 09:52: Auto ramping TGHT @ 6 A per min to 500A.
Friday, August 17, 2018, 10:01: TGHT @ 501.4 A. Display @ 5.656 V. PS @ 5.63 V. Multimeter @ 3280 mV. TBHT @ 228.9 A. Display @ 2.188 V. PS @ 2.19 V.
Friday, August 17, 2018, 10:02: Auto ramping TGHT @ 6 A per min to 510A.
Friday, August 17, 2018, 10:08: TGHT @ 511.4 A. Display @ 5.810 V. PS @ 5.78 V. Multimeter @ 3350 mV. TBHT @ 228.9 A. Display @ 2.193 V. PS @ 2.19 V.
Friday, August 17, 2018, 10:09: Auto ramping TGHT @ 6 A per min to 520A.
Friday, August 17, 2018, 10:14: TGHT @ 521.3 A. Display @ 5.961 V. PS @ 5.93 V. Multimeter @ 3420 mV. TBHT @ 228.9 A. Display @ 2.203 V. PS @ 2.20 V.
Friday, August 17, 2018, 10:16: Auto ramping TGHT @ 6 A per min to 550A.
Friday, August 17, 2018, 10:24: TGHT @ 551.5 A. Display @ 6.425 V. PS @ 6.39 V. Multimeter @ 3630 mV. TBHT @ 228.9 A. Display @ 2.217 V. PS @ 2.22 V.
Friday, August 17, 2018, 10:25: Auto ramping TGHT @ 6 A per min to 560A.
Friday, August 17, 2018, 10:32: TGHT @ 561.5 A. Display @ 6.584 V. PS @ 6.55 V. Multimeter @ 3690 mV. TBHT @ 228.9 A. Display @ 2.222 V. PS @ 2.22 V.
Friday, August 17, 2018, 10:32: Auto ramping TGHT @ 6 A per min to 580A.
Friday, August 17, 2018, 10:40: TGHT @ 581.7 A. Display @ 6.899 V. PS @ 6.87 V. Multimeter @ 3830 mV. TBHT @ 228.9 A. Display @ 2.232 V. PS @ 2.23 V.
Friday, August 17, 2018, 10:41: Auto ramping TGHT @ 6 A per min to 600A.
Friday, August 17, 2018, 10:47: TGHT @ 600.0 A. Display @ 7.214 V. PS @ 7.18 V. Multimeter @ 3970 mV. TBHT @ 228.9 A. Display @ 2.239 V. PS @ 2.24 V.
Friday, August 17, 2018, 10:56: TGHT @ 600.0 A. Display @ 7.214 V. PS @ 7.18 V. Multimeter @ 3970 mV. TBHT @ 228.9 A. Display @ 2.242 V. PS @ 2.24 V.
Friday, August 17, 2018, 10:58: Auto cooling TGHT @ 6 A per min to 580A.
Friday, August 17, 2018, 11:09: TGHT @ 581.7 A. Display @ 6.901 V. PS @ 6.87 V. Multimeter @ 3830 mV. TBHT @ 228.9 A. Display @ 2.234 V. PS @ 2.23 V.
Friday, August 17, 2018, 11:09: Auto cooling TGHT @ 6 A per min to 560A.
Friday, August 17, 2018, 11:19: TGHT @ 561.5 A. Display @ 6.586 V. PS @ 6.55 V. Multimeter @ 3700 mV. TBHT @ 228.9 A. Display @ 2.225 V. PS @ 2.22 V.
Friday, August 17, 2018, 11:20: Auto cooling TGHT @ 6 A per min to 550A.
Friday, August 17, 2018, 11:25: TGHT @ 551.5 A. Display @ 6.432 V. PS @ 6.40 V. Multimeter @ 3630 mV. TBHT @ 228.9 A. Display @ 2.220 V. PS @ 2.22 V.
Friday, August 17, 2018, 11:26: Auto cooling TGHT @ 6 A per min to 520A.
Friday, August 17, 2018, 11:35: TGHT @ 521.3 A. Display @ 5.976 V. PS @ 5.94 V. Multimeter @ 3430 mV. TBHT @ 228.9 A. Display @ 2.208 V. PS @ 2.21 V.
Friday, August 17, 2018, 11:36: Auto cooling TGHT @ 6 A per min to 510A.
Friday, August 17, 2018, 11:41: TGHT @ 511.4 A. Display @ 5.827 V. PS @ 5.80 V. Multimeter @ 3370 mV. TBHT @ 228.9 A. Display @ 2.203 V. PS @ 2.20 V.
Friday, August 17, 2018, 11:42: Auto cooling TGHT @ 6 A per min to 500A.
Friday, August 17, 2018, 11:48: TGHT @ 501.4 A. Display @ 5.678 V. PS @ 5.65 V. Multimeter @ 3300 mV. TBHT @ 228.9 A. Display @ 2.198 V. PS @ 2.20 V.
Friday, August 17, 2018, 11:49: Auto cooling TGHT @ 6 A per min to 475A.
Friday, August 17, 2018, 11:55: TGHT @ 476.2 A. Display @ 5.314 V. PS @ 5.28 V. Multimeter @ 3140 mV. TBHT @ 228.9 A. Display @ 2.190 V. PS @ 2.19 V.
Friday, August 17, 2018, 12:59: TGHT @ 476.2 A. Display @ 5.309 V. PS @ 5.28 V. Multimeter @ 3140 mV. TBHT @ 228.9 A. Display @ 2.188 V. PS @ 2.19 V.
Friday, August 17, 2018, 13:00: Auto cooling TGHT @ 6 A per min to 460A.
Friday, August 17, 2018, 13:08: TGHT @ 461.2 A. Display @ 5.092 V. PS @ 5.06 V. Multimeter @ 3050 mV. TBHT @ 228.9 A. Display @ 2.183 V. PS @ 2.18 V.
Friday, August 17, 2018, 13:09: Auto cooling TGHT @ 6 A per min to 451A.
Friday, August 17, 2018, 13:16: TGHT @ 452.2 A. Display @ 4.960 V. PS @ 4.93 V. Multimeter @ 2990 mV. TBHT @ 228.9 A. Display @ 2.173 V. PS @ 2.17 V.
Friday, August 17, 2018, 13:17: Auto cooling TGHT @ 6 A per min to 427A.
Friday, August 17, 2018, 13:26: TGHT @ 428.0 A. Display @ 4.608 V. PS @ 4.58 V. Multimeter @ 2830 mV. TBHT @ 228.9 A. Display @ 2.159 V. PS @ 2.16 V.
Friday, August 17, 2018, 13:27: Auto cooling TGHT @ 6 A per min to 410A.
Friday, August 17, 2018, 13:34: TGHT @ 411.0 A. Display @ 4.364 V. PS @ 4.33 V. Multimeter @ 2730 mV. TBHT @ 228.9 A. Display @ 2.151 V. PS @ 2.15 V.
Friday, August 17, 2018, 13:36: Auto cooling TGHT @ 6 A per min to 400A.
Friday, August 17, 2018, 13:41: TGHT @ 400.9 A. Display @ 4.220 V. PS @ 4.19 V. Multimeter @ 2670 mV. TBHT @ 228.9 A. Display @ 2.149 V. PS @ 2.15 V.
Friday, August 17, 2018, 13:42: Auto cooling TGHT @ 6 A per min to 370A.
Friday, August 17, 2018, 13:52: TGHT @ 370.8 A. Display @ 3.797 V. PS @ 3.77 V. Multimeter @ 2490 mV. TBHT @ 228.9 A. Display @ 2.137 V. PS @ 2.14 V.
Friday, August 17, 2018, 13:53: Auto cooling TGHT @ 6 A per min to 360A.
Friday, August 17, 2018, 14:01: TGHT @ 360.9 A. Display @ 3.661 V. PS @ 3.63 V. Multimeter @ 2430 mV. TBHT @ 228.9 A. Display @ 2.132 V. PS @ 2.13 V.
Friday, August 17, 2018, 14:02: Auto cooling TGHT @ 6 A per min to 354A.
Friday, August 17, 2018, 14:05: TGHT @ 354.9 A. Display @ 3.580 V. PS @ 3.55 V. Multimeter @ 2400 mV. TBHT @ 228.9 A. Display @ 2.129 V. PS @ 2.13 V.
Friday, August 17, 2018, 14:07: Auto cooling TGHT @ 6 A per min to 300A.
Friday, August 17, 2018, 14:17: TGHT @ 300.8 A. Display @ 2.869 V. PS @ 2.84 V. Multimeter @ 2110 mV. TBHT @ 228.9 A. Display @ 2.112 V. PS @ 2.11 V.
Friday, August 17, 2018, 14:22: TGHT @ 300.8 A. Display @ 2.862 V. PS @ 2.83 V. Multimeter @ 2100 mV. TBHT @ 228.9 A. Display @ 2.110 V. PS @ 2.11 V.
Friday, August 17, 2018, 14:22: Auto cooling TGHT @ 6 A per min to 280A.
Friday, August 17, 2018, 14:30: TGHT @ 280.7 A. Display @ 2.606 V. PS @ 2.58 V. Multimeter @ 2000 mV. TBHT @ 228.9 A. Display @ 2.103 V. PS @ 2.10 V
.Friday, August 17, 2018, 14:22: Auto cooling TGHT @ 6 A per min to 240A.
Friday, August 17, 2018, 14:56: TGHT @ 240.6 A. Display @ 2.081 V. PS @ 2.09 V. Multimeter @ 1798 mV. TBHT @ 228.9 A. Display @ 2.081 V. PS @ 2.08 V
Friday, August 17, 2018, 14:58: Auto cooling TGHT @ 6 A per min to 230A.
Friday, August 17, 2018, 15:05: TGHT @ 230.5 A. Display @ 1.998 V. PS @ 1.97 V. Multimeter @ 1750 mV. TBHT @ 228.9 A. Display @ 2.078 V. PS @ 2.08 V
------------------------------------------------------------------------------
Sunday, August 19, 2018, 12:44: TGHT @ 230.5 A. Display @ 1.998 V. PS @ 1.97 V. Multimeter @ 1748 mV. TBHT @ 228.9 A. Display @ 2.078 V. PS @ 2.08 V
Sunday, August 19, 2018, 12:45: Auto cooling TGHT @ 6 A per min to 200A.
Sunday, August 19, 2018, 12:51: TGHT @ 200.4 A. Display @ 1.661 V. PS @ 1.64 V. Multimeter @ 1617 mV. TBHT @ 228.9 A. Display @ 2.071 V. PS @ 2.07 V
Sunday, August 19, 2018, 12:56: TGHT @ 200.4 A. Display @ 1.658 V. PS @ 1.63 V. Multimeter @ 1614 mV. TBHT @ 228.9 A. Display @ 2.068 V. PS @ 2.07 V
Sunday, August 19, 2018, 12:59: Auto cooling TGHT @ 6 A per min to 160A.
Sunday, August 19, 2018, 13:06: TGHT @ 160.3 A. Display @ 1.245 V. PS @ 1.22 V. Multimeter @ 1452 mV. TBHT @ 228.9 A. Display @ 2.061 V. PS @ 2.06 V
Sunday, August 19, 2018, 13:11: TGHT @ 160.3 A. Display @ 1.236 V. PS @ 1.21 V. Multimeter @ 1449 mV. TBHT @ 228.9 A. Display @ 2.056 V. PS @ 2.06 V
Sunday, August 19, 2018, 13:11: Auto cooling TGHT @ 6 A per min to 120A.
Sunday, August 19, 2018, 13:20: TGHT @ 120.1 A. Display @ 0.867 V. PS @ 0.84 V. Multimeter @ 1311 mV. TBHT @ 228.9 A. Display @ 2.049 V. PS @ 2.05 V
Sunday, August 19, 2018, 13:24: TGHT @ 120.1 A. Display @ 0.857 V. PS @ 0.83 V. Multimeter @ 1306 mV. TBHT @ 228.9 A. Display @ 2.046 V. PS @ 2.05 V
Sunday, August 19, 2018, 13:25: Auto cooling TGHT @ 6 A per min to 80A.
Sunday, August 19, 2018, 13:33: TGHT @ 80.1 A. Display @ 0.540 V. PS @ 0.51 V. Multimeter @ 1193 mV. TBHT @ 228.9 A. Display @ 2.039 V. PS @ 2.04 V
Sunday, August 19, 2018, 13:37: TGHT @ 80.1 A. Display @ 0.535 V. PS @ 0.51 V. Multimeter @ 1190 mV. TBHT @ 228.9 A. Display @ 2.037 V. PS @ 2.04 V
Sunday, August 19, 2018, 13:38: Auto cooling TGHT @ 6 A per min to 40A.
Sunday, August 19, 2018, 13:53: TGHT @ 40.0 A. Display @ 0.259 V. PS @ 0.24 V. Multimeter @ 1100 mV. TBHT @ 228.9 A. Display @ 2.032 V. PS @ 2.03 V
Sunday, August 19, 2018, 13:53: Auto cooling TGHT @ 6 A per min to 0A.
Sunday, August 19, 2018, 14:02: TGHT @ 4.4 A. Display @ 0.042 V. PS @ 0.02 V. Multimeter @ 1033 mV. TBHT @ 228.9 A. Display @ 2.029 V. PS @ 2.03 V
Sunday, August 19, 2018, 14:02: Auto cooling TBHT @ 6 A per min to 200A.
Sunday, August 19, 2018, 14:12: TGHT @ 4.4 A. Display @ 0.042 V. PS @ 0.02 V. Multimeter @ 0807 mV. TBHT @ 198.8 A. Display @ 1.648 V. PS @ 1.65 V
Sunday, August 19, 2018, 14:15: Auto cooling TBHT @ 6 A per min to 160A.
Sunday, August 19, 2018, 14:27: TGHT @ 4.4 A. Display @ 0.042 V. PS @ 0.02 V. Multimeter @ 0531 mV. TBHT @ 158.7 A. Display @ 1.179 V. PS @ 1.17 V
Sunday, August 19, 2018, 14:28: Auto cooling TBHT @ 6 A per min to 120A.
Sunday, August 19, 2018, 14:50: TGHT @ 4.4 A. Display @ 0.042 V. PS @ 0.02 V. Multimeter @ 0419 mV. TBHT @ 118.5 A. Display @ 0.725 V. PS @ 0.72 V
Sunday, August 19, 2018, 14:50: Auto cooling TBHT @ 6 A per min to 80A.
Sunday, August 19, 2018, 15:02: TGHT @ 4.4 A. Display @ 0.042 V. PS @ 0.02 V. Multimeter @ 0208 mV. TBHT @ 79.0 A. Display @ 0.415 V. PS @ 0.40 V
Sunday, August 19, 2018, 15:02: Auto cooling TBHT @ 6 A per min to 40A.
Sunday, August 19, 2018, 15:20: TGHT @ 4.4 A. Display @ 0.042 V. PS @ 0.02 V. Multimeter @ 0083 mV. TBHT @ 38.4 A. Display @ 0.198 V. PS @ 0.19 V
Sunday, August 19, 2018, 15:21: Auto cooling TBHT @ 6 A per min to 0A.
Monday, August 20, 2018, 09:10: RGA off.
Monday, August 20, 2018, 09:14: Ion gauge, turbo pumps off.
Monday, August 20, 2018, 09:15: Power off.
Monday, August 20, 2018, 10:26: Venting. |
|
684
|
Wednesday, September 12, 2018, 11:19 |
Aaron Schmidt | Ta#57 | LP-SIS | ITE | TM4 | 41177-3 | 42752-6 | In progress | Conditioning Ta#57 | Wednesday, September 12, 2018, 11:06: Roughing. Help req'd. Doors wouldn't seal.
Wednesday, September 12, 2018, 11:52: Turbo pumps started.
Wednesday, September 12, 2018, 12:40: Ion gauge on.
Thursday, September 13, 2018, 07:30: Waterlines opened. No response on RGA. Vacuum ok @ 5.5e-07T
Thursday, September 13, 2018, 07:33: Power/sources on.
Thursday, September 13, 2018, 07:37: Auto-ramping TGHT / TBHT to 580 / 230 A @ 2A per min
Thursday, September 13, 2018, 07:42: BIAS / EZL / EE @ 30 / 13 / 1.5 kV.
Thursday, September 13, 2018, 07:43: IV4 opened.
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
24.6 |
4.02 |
TBHT |
24.4 |
3.10 |
COIL + EE |
24.4 |
5.13 |
TGHT |
24.6 |
5.77 |
Thursday, September 13, 2018, 07:44: Optics set.
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
33.5 |
4.02 |
TBHT |
24.6 |
3.13 |
COIL + EE |
24.4 |
5.16 |
TGHT |
26.5 |
5.82 |
Thursday, September 13, 2018, 13:56: Vacuum @ 9.97e-07T.
Friday, September 14, 2018, 09:55:
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
33.5 |
4.02 |
TBHT |
24.6 |
3.13 |
COIL + EE |
24.4 |
5.16 |
TGHT |
26.5 |
5.82 |
Friday, September 14, 2018, 09:55:
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
33.7 |
3.99 |
TBHT |
24.6 |
3.11 |
COIL + EE |
24.6 |
5.12 |
TGHT |
26.5 |
5.83 |
Friday, September 14, 2018, 14:41: Auto-cooling @ 2 A per min.
Monday, September 17, 2018, 09:15: HV to 0. RGA off, IV4 closed & reset
Monday, September 17, 2018, 09:20: Sources off. Turbo pumps & Ion gauge off. Power off.
Monday, September 17, 2018, 09:23: Waterlines purged.
|
Attachment 1: ER-580-230-30kV-M7.txt
|
value (Voltage offset 40.0000)
80.0000
0.7000
80.0000
29978.0220
21
21
./emit.dat
-0.1041 -0.0919 -0.0903 -0.0830 -0.0800 -0.0824 -0.0729 -0.0745 -0.0851 -0.0824 -0.0836 -0.0858 -0.0861 -0.0842 -0.0836 -0.0833 -0.0815 -0.0803 -0.0873 -0.0833 -0.0836
-0.0812 -0.0897 -0.0867 -0.0726 -0.0003 0.0238 0.0455 0.0076 0.0107 -0.0644 -0.0812 -0.0809 -0.0827 -0.0790 -0.0800 -0.0815 -0.0806 -0.0809 -0.0803 -0.0812 -0.0803
-0.0363 -0.0693 -0.0787 -0.0348 0.0748 0.0604 0.1294 0.0235 0.0601 -0.0052 -0.0497 -0.0766 -0.0830 -0.0797 -0.0839 -0.0815 -0.0824 -0.0848 -0.0891 -0.0809 -0.0806
0.0046 -0.0131 -0.0156 0.1093 0.1465 0.0888 0.1617 0.0455 0.1926 0.0470 -0.0171 -0.0577 -0.0763 -0.0760 -0.0809 -0.0790 -0.0818 -0.0806 -0.0797 -0.0812 -0.0800
0.0208 0.0235 0.0778 0.1755 0.2148 0.1544 0.2103 0.1010 0.2051 0.0211 -0.0339 -0.0598 -0.0644 -0.0766 -0.0790 -0.0775 -0.0787 -0.0751 -0.0797 -0.0800 -0.0778
-0.0479 0.1074 0.1547 0.1547 0.1947 0.1337 0.1697 0.0522 0.1599 -0.0217 -0.0513 -0.0656 -0.0671 -0.0568 -0.0677 -0.0800 -0.0778 -0.0769 -0.0928 -0.0787 -0.0766
-0.0797 -0.0858 0.0458 0.0531 0.0851 0.0638 0.0507 0.0146 0.1797 -0.0516 -0.0757 -0.0705 -0.0720 -0.0708 -0.0702 -0.0592 -0.0751 -0.0812 -0.0815 -0.0806 -0.0793
-0.0797 -0.0885 -0.0839 -0.0803 0.0049 0.0058 0.0031 -0.0162 0.0607 -0.0754 -0.0790 -0.0778 -0.0790 -0.0766 -0.0778 -0.0717 -0.0699 -0.0641 -0.0845 -0.0812 -0.0812
-0.0778 -0.0870 -0.0839 -0.0803 -0.0757 -0.0711 -0.0446 -0.0647 -0.0806 -0.0781 -0.0778 -0.0790 -0.0803 -0.0790 -0.0781 -0.0772 -0.0748 -0.0696 -0.0482 -0.0809 -0.0806
-0.0781 -0.0879 -0.0839 -0.0793 -0.0781 -0.0797 -0.0778 -0.0797 -0.0803 -0.0781 -0.0778 -0.0778 -0.0790 -0.0769 -0.0803 -0.0803 -0.0757 -0.0751 -0.0696 -0.0455 -0.0787
-0.0797 -0.0894 -0.0842 -0.0800 -0.0790 -0.0815 -0.0790 -0.0830 -0.0790 -0.0800 -0.0797 -0.0784 -0.0824 -0.0787 -0.0803 -0.0812 -0.0836 -0.0809 -0.0858 -0.0726 -0.0595
-0.0803 -0.0903 -0.0858 -0.0830 -0.0797 -0.0803 -0.0787 -0.0775 -0.0793 -0.0815 -0.0812 -0.0800 -0.0797 -0.0781 -0.0797 -0.0809 -0.0821 -0.0818 -0.0864 -0.0797 -0.0754
-0.0800 -0.0882 -0.0858 -0.0845 -0.0784 -0.0775 -0.0784 -0.0787 -0.0821 -0.0809 -0.0809 -0.0818 -0.0815 -0.0812 -0.0793 -0.0809 -0.0797 -0.0778 -0.0867 -0.0815 -0.0781
-0.0781 -0.0894 -0.0848 -0.0870 -0.0775 -0.0778 -0.0781 -0.0787 -0.0769 -0.0800 -0.0806 -0.0772 -0.0790 -0.0800 -0.0793 -0.0787 -0.0781 -0.0751 -0.0876 -0.0803 -0.0772
-0.0766 -0.0882 -0.0824 -0.0848 -0.0781 -0.0797 -0.0787 -0.0751 -0.0858 -0.0775 -0.0797 -0.0778 -0.0793 -0.0778 -0.0790 -0.0793 -0.0766 -0.0797 -0.0894 -0.0797 -0.0775
-0.0778 -0.0885 -0.0827 -0.0833 -0.0769 -0.0769 -0.0751 -0.0766 -0.0751 -0.0763 -0.0800 -0.0778 -0.0784 -0.0754 -0.0778 -0.0787 -0.0772 -0.0809 -0.0854 -0.0766 -0.0778
-0.0793 -0.0876 -0.0851 -0.0781 -0.0745 -0.0742 -0.0754 -0.0735 -0.0781 -0.0763 -0.0793 -0.0793 -0.0815 -0.0781 -0.0784 -0.0766 -0.0790 -0.0775 -0.0867 -0.0800 -0.0797
-0.0784 -0.0873 -0.0836 -0.0775 -0.0726 -0.0714 -0.0729 -0.0705 -0.0739 -0.0745 -0.0787 -0.0790 -0.0818 -0.0809 -0.0797 -0.0809 -0.0800 -0.0778 -0.0867 -0.0787 -0.0803
-0.0781 -0.0848 -0.0800 -0.0742 -0.0693 -0.0711 -0.0720 -0.0674 -0.0732 -0.0745 -0.0781 -0.0769 -0.0797 -0.0793 -0.0793 -0.0815 -0.0797 -0.0800 -0.0906 -0.0778 -0.0800
-0.0748 -0.0858 -0.0793 -0.0772 -0.0705 -0.0726 -0.0717 -0.0723 -0.0723 -0.0751 -0.0790 -0.0769 -0.0778 -0.0763 -0.0800 -0.0790 -0.0772 -0.0809 -0.0891 -0.0769 -0.0778
-0.0757 -0.0854 -0.0797 -0.0754 -0.0729 -0.0711 -0.0723 -0.0681 -0.0674 -0.0729 -0.0787 -0.0790 -0.0775 -0.0787 -0.0775 -0.0769 -0.0772 -0.0763 -0.0891 -0.0784 -0.0775
|
Attachment 2: ER-580-230-30kV-M38.txt
|
value (Voltage offset 40.0000)
80.0000
0.7000
80.0000
29978.0220
21
21
./emit.dat
-0.1047 -0.0854 -0.0870 -0.0839 -0.0812 -0.0833 -0.0714 -0.0671 -0.0836 -0.0858 -0.0873 -0.0906 -0.0845 -0.0864 -0.0845 -0.0864 -0.0879 -0.0867 -0.0851 -0.0864 -0.0848
-0.0784 -0.0806 -0.0806 -0.0754 -0.0723 0.0015 0.1178 0.1584 0.1105 -0.0311 -0.0803 -0.0818 -0.0830 -0.0827 -0.0824 -0.0812 -0.0806 -0.0839 -0.0864 -0.0821 -0.0803
-0.0034 -0.0601 -0.0732 -0.0656 -0.0476 0.1443 0.2267 0.1889 0.1532 0.0931 0.0171 -0.0620 -0.0809 -0.0800 -0.0818 -0.0809 -0.0806 -0.0812 -0.0876 -0.0827 -0.0815
0.0577 0.0604 0.0372 -0.0018 0.0333 0.2884 0.2066 0.1212 0.1318 0.0674 0.0772 -0.0049 -0.0635 -0.0787 -0.0806 -0.0800 -0.0800 -0.0772 -0.0888 -0.0806 -0.0806
0.0528 0.0870 0.1120 0.1562 0.3827 0.5875 0.3125 0.2417 0.2646 0.1102 0.0943 0.0024 -0.0381 -0.0620 -0.0797 -0.0800 -0.0800 -0.0790 -0.0879 -0.0806 -0.0821
-0.0412 0.0250 0.0858 0.1660 0.4089 0.5222 0.3882 0.3125 0.2863 0.1181 0.0641 -0.0226 -0.0513 -0.0592 -0.0613 -0.0824 -0.0803 -0.0824 -0.0891 -0.0787 -0.0806
-0.0803 -0.0775 0.0793 0.1678 0.2972 0.3137 0.2509 0.2155 0.1559 0.1059 0.0543 -0.0290 -0.0439 -0.0540 -0.0620 -0.0662 -0.0690 -0.0818 -0.0912 -0.0803 -0.0821
-0.0800 -0.0797 -0.0797 -0.0714 0.1044 0.1242 0.1099 0.0998 0.0598 0.0760 0.0031 -0.0754 -0.0726 -0.0690 -0.0571 -0.0702 -0.0711 -0.0696 -0.0903 -0.0830 -0.0812
-0.0790 -0.0790 -0.0800 -0.0797 -0.0754 -0.0568 0.0415 0.0439 0.0403 0.0494 -0.0241 -0.0800 -0.0800 -0.0787 -0.0757 -0.0726 -0.0711 -0.0677 -0.0775 -0.0797 -0.0812
-0.0806 -0.0793 -0.0800 -0.0818 -0.0769 -0.0772 -0.0775 -0.0708 -0.0146 -0.0171 -0.0769 -0.0812 -0.0818 -0.0815 -0.0806 -0.0824 -0.0787 -0.0751 -0.0656 -0.0714 -0.0665
-0.0793 -0.0812 -0.0800 -0.0879 -0.0800 -0.0797 -0.0787 -0.0790 -0.0809 -0.0797 -0.0821 -0.0803 -0.0824 -0.0809 -0.0818 -0.0815 -0.0809 -0.0815 -0.0800 -0.0739 0.0928
-0.0793 -0.0806 -0.0803 -0.0827 -0.0797 -0.0806 -0.0797 -0.0778 -0.0797 -0.0793 -0.0809 -0.0809 -0.0797 -0.0778 -0.0797 -0.0803 -0.0793 -0.0793 -0.0906 -0.0797 -0.0610
-0.0784 -0.0790 -0.0806 -0.0809 -0.0806 -0.0812 -0.0793 -0.0809 -0.0778 -0.0778 -0.0790 -0.0800 -0.0790 -0.0772 -0.0772 -0.0784 -0.0812 -0.0800 -0.0900 -0.0824 -0.0772
-0.0778 -0.0793 -0.0790 -0.0833 -0.0803 -0.0790 -0.0775 -0.0800 -0.0821 -0.0800 -0.0793 -0.0781 -0.0778 -0.0775 -0.0769 -0.0787 -0.0809 -0.0797 -0.0833 -0.0824 -0.0818
-0.0806 -0.0812 -0.0784 -0.0806 -0.0790 -0.0781 -0.0778 -0.0775 -0.0873 -0.0797 -0.0818 -0.0803 -0.0809 -0.0809 -0.0800 -0.0797 -0.0781 -0.0800 -0.0797 -0.0815 -0.0812
-0.0797 -0.0793 -0.0793 -0.0793 -0.0781 -0.0766 -0.0760 -0.0757 -0.0769 -0.0790 -0.0806 -0.0809 -0.0800 -0.0815 -0.0815 -0.0793 -0.0797 -0.0790 -0.0937 -0.0812 -0.0793
-0.0769 -0.0790 -0.0818 -0.0827 -0.0775 -0.0739 -0.0739 -0.0769 -0.0760 -0.0760 -0.0781 -0.0778 -0.0800 -0.0797 -0.0803 -0.0815 -0.0824 -0.0803 -0.0876 -0.0787 -0.0800
-0.0769 -0.0784 -0.0790 -0.0858 -0.0757 -0.0726 -0.0745 -0.0760 -0.0879 -0.0772 -0.0784 -0.0769 -0.0775 -0.0790 -0.0790 -0.0818 -0.0815 -0.0827 -0.0909 -0.0797 -0.0803
-0.0763 -0.0766 -0.0751 -0.0766 -0.0735 -0.0702 -0.0739 -0.0775 -0.0757 -0.0757 -0.0778 -0.0766 -0.0800 -0.0800 -0.0790 -0.0778 -0.0793 -0.0809 -0.0894 -0.0793 -0.0803
-0.0775 -0.0769 -0.0769 -0.0769 -0.0739 -0.0693 -0.0726 -0.0763 -0.0763 -0.0769 -0.0790 -0.0772 -0.0790 -0.0800 -0.0793 -0.0778 -0.0784 -0.0787 -0.0903 -0.0790 -0.0778
-0.0781 -0.0793 -0.0769 -0.0800 -0.0726 -0.0696 -0.0699 -0.0729 -0.0766 -0.0769 -0.0769 -0.0772 -0.0790 -0.0781 -0.0806 -0.0797 -0.0793 -0.0775 -0.0894 -0.0772 -0.0790
|
Attachment 3: ER-580-230-30kV-M83.txt
|
value (Voltage offset 40.0000)
80.0000
0.7000
80.0000
29978.0220
21
21
./emit.dat
-0.1050 -0.0864 -0.0854 -0.0848 -0.0812 -0.0824 -0.0812 -0.0778 -0.0821 -0.0842 -0.0861 -0.0839 -0.0851 -0.0854 -0.0873 -0.0851 -0.0879 -0.0854 -0.0848 -0.0854 -0.0864
-0.0778 -0.0809 -0.0818 -0.0803 -0.0726 -0.0464 -0.0089 0.0061 0.0009 -0.0531 -0.0812 -0.0800 -0.0815 -0.0797 -0.0827 -0.0806 -0.0824 -0.0821 -0.0818 -0.0809 -0.0827
-0.0381 -0.0699 -0.0797 -0.0714 -0.0681 0.0180 0.0522 0.0269 0.0272 0.0269 -0.0165 -0.0677 -0.0803 -0.0812 -0.0809 -0.0830 -0.0787 -0.0797 -0.0815 -0.0797 -0.0793
-0.0192 -0.0235 -0.0229 -0.0406 -0.0354 0.0851 0.0604 0.0085 0.0052 0.0104 0.0299 -0.0204 -0.0497 -0.0787 -0.0818 -0.0806 -0.0803 -0.0793 -0.0824 -0.0797 -0.0797
-0.0302 -0.0269 -0.0064 0.0580 0.0906 0.2505 0.0992 0.0574 0.0830 0.0482 0.0488 -0.0003 0.1349 0.1010 -0.0763 -0.0818 -0.0809 -0.0800 -0.0803 -0.0824 -0.0803
-0.0659 -0.0510 -0.0140 0.1251 0.1233 0.2231 0.1401 0.0897 0.0992 0.0427 0.0308 -0.0098 0.0592 -0.0250 -0.0266 -0.0833 -0.0815 -0.0803 -0.0797 -0.0824 -0.0897
-0.0812 -0.0797 -0.0177 0.1657 0.0900 0.1364 0.0745 0.0385 0.0409 0.0351 0.0406 -0.0128 0.0232 -0.0409 -0.0549 -0.0491 -0.0595 -0.0790 -0.0833 -0.0827 -0.0800
-0.0793 -0.0778 -0.0790 -0.0735 0.0079 0.0250 0.0076 -0.0082 0.0018 0.0092 -0.0064 -0.0748 -0.0330 -0.0641 -0.0293 -0.0580 -0.0607 -0.0546 -0.0812 -0.0827 -0.0793
-0.0781 -0.0793 -0.0800 -0.0824 -0.0790 -0.0705 -0.0262 -0.0299 -0.0122 -0.0031 -0.0177 -0.0809 -0.0800 -0.0781 -0.0540 -0.0644 -0.0543 -0.0653 -0.0513 -0.0800 -0.0790
-0.0812 -0.0818 -0.0800 -0.0824 -0.0806 -0.0800 -0.0797 -0.0775 -0.0433 -0.0345 -0.0751 -0.0806 -0.0812 -0.0818 -0.0812 -0.0797 -0.0745 -0.0641 -0.0677 -0.0613 -0.0748
-0.0790 -0.0818 -0.0803 -0.0848 -0.0787 -0.0800 -0.0812 -0.0797 -0.0940 -0.0800 -0.0800 -0.0803 -0.0797 -0.0809 -0.0812 -0.0821 -0.0797 -0.0790 -0.0732 -0.0702 -0.0705
-0.0797 -0.0812 -0.0803 -0.0815 -0.0806 -0.0797 -0.0784 -0.0790 -0.0797 -0.0809 -0.0806 -0.0778 -0.0772 -0.0790 -0.0800 -0.0790 -0.0793 -0.0803 -0.0827 -0.0739 -0.0696
-0.0784 -0.0775 -0.0812 -0.0815 -0.0806 -0.0797 -0.0812 -0.0793 -0.0778 -0.0793 -0.0818 -0.0784 -0.0803 -0.0763 -0.0784 -0.0790 -0.0787 -0.0797 -0.0821 -0.0815 -0.0739
-0.0812 -0.0815 -0.0806 -0.0818 -0.0797 -0.0793 -0.0803 -0.0821 -0.0891 -0.0806 -0.0830 -0.0821 -0.0803 -0.0806 -0.0790 -0.0797 -0.0793 -0.0793 -0.0803 -0.0803 -0.0793
-0.0824 -0.0833 -0.0790 -0.0803 -0.0778 -0.0797 -0.0800 -0.0803 -0.0836 -0.0787 -0.0815 -0.0821 -0.0800 -0.0824 -0.0806 -0.0797 -0.0797 -0.0784 -0.0806 -0.0812 -0.0787
-0.0806 -0.0803 -0.0784 -0.0806 -0.0772 -0.0778 -0.0760 -0.0781 -0.0775 -0.0766 -0.0800 -0.0793 -0.0787 -0.0809 -0.0793 -0.0830 -0.0806 -0.0809 -0.0809 -0.0833 -0.0797
-0.0781 -0.0781 -0.0797 -0.0827 -0.0784 -0.0766 -0.0775 -0.0797 -0.0778 -0.0763 -0.0809 -0.0781 -0.0781 -0.0781 -0.0800 -0.0797 -0.0809 -0.0806 -0.0830 -0.0821 -0.0812
-0.0766 -0.0769 -0.0803 -0.0818 -0.0784 -0.0763 -0.0775 -0.0790 -0.0760 -0.0800 -0.0818 -0.0803 -0.0803 -0.0787 -0.0790 -0.0793 -0.0790 -0.0793 -0.0803 -0.0806 -0.0806
-0.0775 -0.0797 -0.0787 -0.0809 -0.0769 -0.0775 -0.0809 -0.0790 -0.0812 -0.0784 -0.0824 -0.0800 -0.0800 -0.0803 -0.0815 -0.0784 -0.0781 -0.0775 -0.0803 -0.0793 -0.0797
-0.0800 -0.0803 -0.0775 -0.0824 -0.0772 -0.0763 -0.0732 -0.0766 -0.0790 -0.0766 -0.0809 -0.0809 -0.0778 -0.0793 -0.0800 -0.0766 -0.0772 -0.0766 -0.0797 -0.0800 -0.0793
-0.0797 -0.0787 -0.0784 -0.0790 -0.0760 -0.0739 -0.0739 -0.0757 -0.0751 -0.0760 -0.0778 -0.0787 -0.0778 -0.0800 -0.0781 -0.0775 -0.0787 -0.0787 -0.0854 -0.0824 -0.0787
|
Attachment 4: ER-580-230-30kV-M129.txt
|
value (Voltage offset 40.0000)
80.0000
0.7000
80.0000
29978.0220
21
21
./emit.dat
-0.1038 -0.0870 -0.0873 -0.0861 -0.0827 -0.0879 -0.0793 -0.0711 -0.0842 -0.0873 -0.0867 -0.0848 -0.0836 -0.0839 -0.0827 -0.0839 -0.0845 -0.0845 -0.0858 -0.0891 -0.0861
-0.0775 -0.0821 -0.0815 -0.0751 -0.0745 -0.0305 0.0513 0.0909 0.0665 -0.0195 -0.0821 -0.0812 -0.0806 -0.0815 -0.0806 -0.0824 -0.0793 -0.0815 -0.0812 -0.0836 -0.0821
-0.0125 -0.0610 -0.0735 -0.0650 -0.0580 0.0922 0.1590 0.1352 0.1193 0.1236 0.0455 -0.0510 -0.0815 -0.0827 -0.0824 -0.0821 -0.0803 -0.0806 -0.0818 -0.0827 -0.0809
0.0232 0.0366 0.0226 -0.0119 0.0018 0.2161 0.1642 0.0845 0.1028 0.0977 0.1288 0.0266 -0.0519 -0.0781 -0.0833 -0.0842 -0.0839 -0.0812 -0.0827 -0.0848 -0.0815
0.0104 0.0336 0.0537 0.0922 0.2652 0.4791 0.2448 0.1974 0.2075 0.1489 0.1703 0.0433 0.0076 -0.0308 -0.0806 -0.0824 -0.0815 -0.0827 -0.0842 -0.0885 -0.0809
-0.0500 -0.0150 0.0397 0.1108 0.3088 0.4303 0.2905 0.3024 0.2249 0.1633 0.1065 0.0165 -0.0156 -0.0308 -0.0403 -0.0760 -0.0809 -0.0818 -0.0845 -0.0885 -0.0821
-0.0797 -0.0790 0.0281 0.1035 0.1779 0.2435 0.1834 0.1431 0.1355 0.1328 0.1297 0.0211 0.0253 0.0177 -0.0339 -0.0336 -0.0381 -0.0797 -0.0812 -0.0858 -0.0803
-0.0815 -0.0806 -0.0781 -0.0732 0.0314 0.0638 0.0558 0.0552 0.0690 0.0946 0.0500 -0.0739 -0.0537 -0.0470 -0.0073 -0.0378 -0.0351 -0.0311 -0.0809 -0.0864 -0.0803
-0.0809 -0.0797 -0.0806 -0.0803 -0.0757 -0.0632 0.0116 0.0140 0.0418 0.0577 0.0174 -0.0800 -0.0769 -0.0775 -0.0684 -0.0497 -0.0378 -0.0391 -0.0418 -0.0842 -0.0781
-0.0809 -0.0824 -0.0824 -0.0824 -0.0797 -0.0793 -0.0778 -0.0778 -0.0327 -0.0137 -0.0732 -0.0790 -0.0784 -0.0797 -0.0806 -0.0809 -0.0687 -0.0473 -0.0494 -0.0513 -0.0684
-0.0806 -0.0824 -0.0815 -0.0815 -0.0800 -0.0806 -0.0812 -0.0864 -0.0769 -0.0781 -0.0790 -0.0781 -0.0797 -0.0790 -0.0797 -0.0800 -0.0800 -0.0800 -0.0525 -0.0558 -0.0549
-0.0790 -0.0818 -0.0803 -0.0806 -0.0793 -0.0800 -0.0797 -0.0803 -0.0824 -0.0784 -0.0830 -0.0797 -0.0793 -0.0778 -0.0790 -0.0784 -0.0793 -0.0790 -0.0815 -0.0681 -0.0543
-0.0800 -0.0781 -0.0797 -0.0793 -0.0784 -0.0784 -0.0800 -0.0851 -0.0818 -0.0803 -0.0827 -0.0809 -0.0797 -0.0806 -0.0803 -0.0809 -0.0815 -0.0775 -0.0815 -0.0864 -0.0681
-0.0806 -0.0806 -0.0809 -0.0800 -0.0778 -0.0781 -0.0775 -0.0836 -0.0797 -0.0775 -0.0815 -0.0797 -0.0821 -0.0806 -0.0824 -0.0839 -0.0809 -0.0815 -0.0827 -0.0870 -0.0803
-0.0809 -0.0815 -0.0815 -0.0806 -0.0797 -0.0790 -0.0769 -0.0800 -0.0873 -0.0784 -0.0815 -0.0797 -0.0797 -0.0803 -0.0824 -0.0821 -0.0800 -0.0806 -0.0842 -0.0876 -0.0803
-0.0797 -0.0812 -0.0815 -0.0809 -0.0806 -0.0793 -0.0778 -0.0803 -0.0784 -0.0809 -0.0809 -0.0784 -0.0787 -0.0812 -0.0781 -0.0790 -0.0806 -0.0803 -0.0833 -0.0873 -0.0812
-0.0787 -0.0784 -0.0787 -0.0803 -0.0793 -0.0763 -0.0790 -0.0800 -0.0797 -0.0790 -0.0800 -0.0784 -0.0809 -0.0787 -0.0772 -0.0793 -0.0781 -0.0769 -0.0812 -0.0867 -0.0793
-0.0787 -0.0775 -0.0797 -0.0790 -0.0754 -0.0732 -0.0751 -0.0797 -0.0766 -0.0763 -0.0809 -0.0784 -0.0787 -0.0772 -0.0781 -0.0800 -0.0787 -0.0793 -0.0839 -0.0864 -0.0797
-0.0793 -0.0790 -0.0790 -0.0797 -0.0763 -0.0726 -0.0742 -0.0787 -0.0739 -0.0754 -0.0793 -0.0766 -0.0787 -0.0806 -0.0812 -0.0812 -0.0797 -0.0809 -0.0833 -0.0873 -0.0793
-0.0778 -0.0809 -0.0806 -0.0800 -0.0751 -0.0726 -0.0714 -0.0800 -0.0729 -0.0729 -0.0763 -0.0751 -0.0784 -0.0793 -0.0797 -0.0784 -0.0806 -0.0803 -0.0833 -0.0885 -0.0784
-0.0787 -0.0787 -0.0781 -0.0775 -0.0757 -0.0732 -0.0723 -0.0751 -0.0702 -0.0742 -0.0769 -0.0763 -0.0763 -0.0772 -0.0775 -0.0778 -0.0809 -0.0787 -0.0824 -0.0891 -0.0806
|
Attachment 5: ER-580-230-30kV-M130.txt
|
value (Voltage offset 40.0000)
80.0000
0.7000
80.0000
29978.0220
21
21
./emit.dat
-0.1044 -0.0864 -0.0876 -0.0854 -0.0845 -0.0873 -0.0891 -0.0815 -0.0870 -0.0888 -0.0894 -0.0854 -0.0864 -0.0824 -0.0870 -0.0873 -0.0870 -0.0833 -0.0861 -0.0858 -0.0873
-0.0754 -0.0803 -0.0803 -0.0763 -0.0705 -0.0415 0.0507 0.1022 0.0833 -0.0085 -0.0781 -0.0836 -0.0818 -0.0778 -0.0803 -0.0836 -0.0818 -0.0806 -0.0806 -0.0818 -0.0815
0.0287 -0.0427 -0.0674 -0.0558 -0.0571 0.1105 0.2243 0.1810 0.1608 0.1733 0.0864 -0.0299 -0.0812 -0.0821 -0.0815 -0.0821 -0.0790 -0.0821 -0.0809 -0.0800 -0.0803
0.0613 0.0775 0.0690 0.0293 0.0397 0.3207 0.2563 0.1126 0.0876 0.0714 0.1642 0.0873 -0.0281 -0.0745 -0.0821 -0.0815 -0.0809 -0.0824 -0.0812 -0.0827 -0.0800
0.0259 0.0800 0.1157 0.1895 0.4358 0.6754 0.3445 0.2350 0.2557 0.1257 0.2408 0.1294 0.0357 -0.0089 -0.0760 -0.0818 -0.0818 -0.0793 -0.0818 -0.0830 -0.0815
-0.0577 0.0162 0.0787 0.1935 0.4044 0.5386 0.3766 0.3082 0.2881 0.1538 0.2075 0.0644 0.0040 -0.0119 -0.0317 -0.0677 -0.0815 -0.0797 -0.0824 -0.0830 -0.0818
-0.0806 -0.0787 0.0537 0.1517 0.2155 0.2774 0.2200 0.1868 0.1724 0.1419 0.2045 0.0543 0.0272 0.0391 -0.0266 -0.0232 -0.0372 -0.0812 -0.0815 -0.0815 -0.0818
-0.0809 -0.0797 -0.0781 -0.0748 0.0427 0.0751 0.0684 0.0784 0.0851 0.0882 0.0766 -0.0711 -0.0543 -0.0424 -0.0131 -0.0302 -0.0400 -0.0308 -0.0797 -0.0806 -0.0793
-0.0833 -0.0797 -0.0821 -0.0818 -0.0769 -0.0671 0.0403 0.0351 0.0797 0.0452 -0.0024 -0.0803 -0.0806 -0.0797 -0.0711 -0.0488 -0.0363 -0.0449 -0.0415 -0.0720 -0.0790
-0.0821 -0.0815 -0.0827 -0.0824 -0.0815 -0.0809 -0.0797 -0.0754 -0.0491 -0.0528 -0.0775 -0.0790 -0.0821 -0.0800 -0.0793 -0.0806 -0.0705 -0.0458 -0.0592 -0.0424 -0.0525
-0.0806 -0.0815 -0.0797 -0.0845 -0.0806 -0.0830 -0.0800 -0.0818 -0.0800 -0.0812 -0.0809 -0.0806 -0.0793 -0.0797 -0.0809 -0.0815 -0.0800 -0.0790 -0.0650 -0.0534 -0.0568
-0.0800 -0.0809 -0.0800 -0.0836 -0.0800 -0.0793 -0.0800 -0.0815 -0.0793 -0.0818 -0.0827 -0.0827 -0.0815 -0.0803 -0.0818 -0.0806 -0.0806 -0.0806 -0.0812 -0.0665 -0.0574
-0.0821 -0.0797 -0.0803 -0.0839 -0.0797 -0.0790 -0.0793 -0.0848 -0.0848 -0.0803 -0.0830 -0.0809 -0.0815 -0.0803 -0.0803 -0.0806 -0.0824 -0.0787 -0.0809 -0.0797 -0.0729
-0.0821 -0.0821 -0.0827 -0.0821 -0.0803 -0.0803 -0.0793 -0.0879 -0.0781 -0.0769 -0.0818 -0.0793 -0.0812 -0.0793 -0.0800 -0.0803 -0.0803 -0.0790 -0.0809 -0.0812 -0.0778
-0.0821 -0.0830 -0.0812 -0.0818 -0.0815 -0.0812 -0.0806 -0.0842 -0.0824 -0.0800 -0.0806 -0.0790 -0.0803 -0.0815 -0.0824 -0.0824 -0.0800 -0.0818 -0.0836 -0.0815 -0.0800
-0.0775 -0.0797 -0.0800 -0.0815 -0.0793 -0.0790 -0.0793 -0.0812 -0.0809 -0.0790 -0.0818 -0.0793 -0.0854 -0.0781 -0.0833 -0.0818 -0.0800 -0.0800 -0.0821 -0.0812 -0.0812
-0.0760 -0.0790 -0.0781 -0.0806 -0.0766 -0.0757 -0.0766 -0.0790 -0.0800 -0.0790 -0.0821 -0.0800 -0.0790 -0.0787 -0.0803 -0.0818 -0.0803 -0.0793 -0.0800 -0.0806 -0.0815
-0.0784 -0.0787 -0.0769 -0.0784 -0.0757 -0.0735 -0.0754 -0.0827 -0.0760 -0.0772 -0.0793 -0.0793 -0.0827 -0.0818 -0.0778 -0.0824 -0.0818 -0.0781 -0.0803 -0.0803 -0.0797
-0.0803 -0.0809 -0.0787 -0.0812 -0.0763 -0.0742 -0.0729 -0.0772 -0.0745 -0.0766 -0.0769 -0.0763 -0.0806 -0.0806 -0.0784 -0.0787 -0.0800 -0.0809 -0.0824 -0.0812 -0.0784
-0.0797 -0.0793 -0.0797 -0.0824 -0.0772 -0.0757 -0.0702 -0.0784 -0.0769 -0.0757 -0.0763 -0.0760 -0.0781 -0.0797 -0.0800 -0.0781 -0.0784 -0.0815 -0.0839 -0.0818 -0.0787
-0.0787 -0.0769 -0.0769 -0.0800 -0.0775 -0.0757 -0.0729 -0.0827 -0.0754 -0.0757 -0.0754 -0.0775 -0.0790 -0.0790 -0.0803 -0.0797 -0.0793 -0.0784 -0.0824 -0.0839 -0.0803
|
Attachment 6: OMEGA_OM-DAQPRO-5300_17-Sep-18_14.13.36.dlp
|
Attachment 7: TS-2018-09-12-12-50.rga
|
Attachment 8: Ta#57.docx
|
Attachment 9: ITE-TM4-Ta#57-LP-SIS_Test_Stand.docx
|
|
799
|
Monday, July 06, 2020, 05:28 |
Fernando Maldonado | Test FEBIAD | HP-FEBIAD | | | 38295-8 | 42752-5 | Done | FEBIAD development part 3/3 | Reminder. Mini pump still linked to SV1. DO NOT OPEN SV2!!!
The plan for today Monday, July 06, 2020 is:
- Ramp up TGHT/TBH to 580/315 A.
- Anode sweep
- Mass scan without gas.
- Coil sweep without gas
- Gases in.
Mass scan with gases
Coil sweep with gasses
Efficiency measurements at highest and lowest FC6 current
COVID-19 protocol: buddy is Marla Cervantes
New idea: Screenshot e-log on the ISAC TS team notebook: 20-07-06 FEBIAD development part 3 (Web view)
Monday, July 06, 2020, 05:27: Auto-ramping TGHT/TBHT to 580/315 @ 2 A/min.
Monday, July 06, 2020, 14:2: IV4 open.
Monday, July 06, 2020, 14:41: Anode sweep completed. No anode current registered.
Monday, July 06, 2020, 14:42: HV spark. Anode current now show something. 13.431 mA. The source current however did not change after the spark.
Monday, July 06, 2020, 15:08: Anode sweep completed.
Monday, July 06, 2020, 15:19: IG1=2.57e-6T.
Monday, July 06, 2020, 15:43: Mass scan completed.
Monday, July 06, 2020, 16:02: Coil sweep completed.
Monday, July 06, 2020, 16:04: FC6=8.40e-11A. IG1=2.59e-6T.
Monday, July 06, 2020, 16:07: SV1 closed. SV3 and SV4 open/closed x5.
Monday, July 06, 2020, 16:10: SV3 and SV4 open.
Monday, July 06, 2020, 16:12: SV1 closed. SV5 open/closed x5.
Monday, July 06, 2020, 16:13: SV5 open. No trip!!!
Monday, July 06, 2020, 16:25: FC6=1.16e-10A. IG1=2.74e-6T.
Monday, July 06, 2020, 16:33: Closing gas lines to prevent more leakage.
Tomorrow I will move to mass 40, to see if the ionization changes with the small gas load.
Monday, July 06, 2020, 16:36: Done for the day.
The plan for today Tuesday, July 07, 2020 is:
- Gases in.
- Mass scan with gases
- Coil sweep with gasses
- Efficiency measurements at highest and lowest FC6 current
Tuesday, July 07, 2020, 13:38:FC6=4.46e-11A. IG1=2.41e-6T.
SV3 and SV4 open/closed x5. SV1 open.
Tuesday, July 07, 2020, 13:40: SV3 and SV4 open.
Tuesday, July 07, 2020, 13:41: SV1 closed. SV5 open/closed x5.
SV5 open. No trip!!!
Tuesday, July 07, 2020, 13:57: FC6=8.46e-11A. IG1=2.59e-6T.
Tuesday, July 07, 2020, 14:15: Mass scan completed.
Tuesday, July 07, 2020, 14:22: mass=39.82. A slight bump between the potassiums 38 and 41.
Tuesday, July 07, 2020, 14:38: Coil sweep completed.
Coil=0A. FC6=1.18e-11A. IG1=2.59e-6T.
Tuesday, July 07, 2020, 14:39: SV3 closed.
Tuesday, July 07, 2020, 14:47: FC6=1.26e-11A. IG1=2.59e-6T.
Tuesday, July 07, 2020, 14:48: SV4 closed.
Tuesday, July 07, 2020, 14:59: 1.15e-11A. IG1=2.41e-6T.
Tuesday, July 07, 2020, 15:00: SV5 closed.
Tuesday, July 07, 2020, 15:07: FC6=1.15e-11A. IG1=2.41e-6T.
Tuesday, July 07, 2020, 15:08: SV3 and SV4 open/closed x5. SV1 open.
Tuesday, July 07, 2020, 15:09: SV3 and SV4 open. SV1 closed.
Tuesday, July 07, 2020, 15:10: SV5 open/closed x5. SV5 open. No trip!!!
Tuesday, July 07, 2020, 15:2: FC6=1.26e-11A. IG1=2.59e-6T.
Tuesday, July 07, 2020, 15:32: SV3 closed.
Tuesday, July 07, 2020, 15:44: FC6=1.4e-11A. IG1=2.57e-6T
Tuesday, July 07, 2020, 15:44: SV4 closed.
Tuesday, July 07, 2020, 16:01: FC6=1.15e-11A. IG1=2.39e-6T.
Tuesday, July 07, 2020, 16:01: SV5 closed.
Tuesday, July 07, 2020, 16:31: Coil sweep with no gas completed.
Tuesday, July 07, 2020, 16:52: Mass scan completed.
Tuesday, July 07, 2020, 16:52: Done for the day.
The plan for today Wednesday, July 08, 2020 is:
- Ramp up TBHT to 580/320 A.
- Anode sweep
- Mass scan without gas.
- Coil sweep without gas
- Gases in.
- Mass scan with gases
- Coil sweep with gasses
- Efficiency measurements at highest FC6 current
- Coil sweep without gas
Wednesday, July 08, 2020, 15:00: IG1=2.41e-6T.
Wednesday, July 08, 2020, 15:29: Auto-ramping TBHT to 320A @ 2A/min.
Wednesday, July 08, 2020, 15:54: Anode sweep completed.
Wednesday, July 08, 2020, 16:33: Mass scan completed
Wednesday, July 08, 2020, 16:56: Coil completed.
Wednesday, July 08, 2020, 16:57. IG1=2.59e-6T. FC6=8.63e-11A.
Wednesday, July 08, 2020, 16:58: SV3 and SV4 open/closed x5. SV1 open.
Wednesday, July 08, 2020, 16:59: SV3 and SV4 open. SV1 closed.
Wednesday, July 08, 2020, 17:01: SV5 open/closed x5. SV5 open. No trip!!!
Wednesday, July 08, 2020, 17:12: FC6=9.20e-12A. IG1=2.65e-6T.
Wednesday, July 08, 2020, 17:44: Mass scan completed.
FC6=8.61e-12A. IG1=2.65e-6T.
Wednesday, July 08, 2020, 18:14: Coil sweep completed.
FC6=5.52e-11A. IG1=2.68e-6T.
Wednesday, July 08, 2020, 18:19: SV3 closed.
Wednesday, July 08, 2020, 18:25: FC6=6.8e-11A. IG1=2.65e-6T
Wednesday, July 08, 2020, 18:25: SV4 closed.
Wednesday, July 08, 2020, 18:39 : FC6=1.70e-11A. IG1=2.59e-6T.
Wednesday, July 08, 2020, 18:40 : SV5 closed.
Wednesday, July 08, 2020, 19:08: FC6=1.32e-11A. IG1=2.59e-6T.
Wednesday, July 08, 2020, 19:12: Sudden TBHT current increase of approx 2A. TBHT voltage seem to have droped a little 0.02V.
Wednesday, July 08, 2020, 19:14: Coil=0A. I'll wait until IG1 reading stabilizes.
Wednesday, July 08, 2020, 19:31: FC6=8.09e-12. IG1=2.48e-6T.
Wednesday, July 08, 2020, 19:31: TBHT returned to values pior to sudden jump.
Wednesday, July 08, 2020, 19:58: Coil sweep with no gas completed.
Wednesday, July 08, 2020, 19:58: Done for the day.
The plan for today Thursday, July 09,2020 is:
- Ramp down TBHT to 580/290 A.
- Anode sweep
- Coil sweep without gas
- Gases in.
- Coil sweep with gases
- Ramp down test stand
Thursday, July 09, 2020, 09:53: Manually tap down TBHT to 290A @ ~~2A/min
Thursday, July 09, 2020, 11:02: Anode sweep completed.
Thursday, July 09, 2020, 13:15: Coil=0A. IG1=2.24e-6T. FC0=2.48e-10A. FC6=7.43e-12A.
Thursday, July 09, 2020, 14:15: SV3 and SV4 open/closed x5. SV1 open.
Thursday, July 09, 2020, 14:16: SV3 and SV4 open. SV1 closed.
Thursday, July 09, 2020, 14:17 : SV5 open/closed x5. SV5 open. No trip!!!
Thursday, July 09, 2020, 14:31: FC6=3.03e-11A. IG1=2.50e-6T.
Thursday, July 09, 2020, 14:54: Coil sweep completed.
Thursday, July 09, 2020, 15:16: Anode=0 V.
Thursday, July 09, 2020, 15:16: SV3, SV4, SV5 closed
Thursday, July 09, 2020, 15:18: IV4 closed. timeout, reset it
Thursday, July 09, 2020, 15:20: BIAS/EZL/EE 0 kV.
Thursday, July 09, 2020, 15:21: Auto-cooling 2 A/min
Thursday, July 09, 2020, 15:25: Done for the day.
Monday, July 13, 2020, 07:01: Ion gauge & Turbo pumps off. Noted HV cage is already removed & PS locked out. |
|
798
|
Tuesday, June 30, 2020, 21:59 |
Fernando Maldonado | Test FEBIAD | HP-FEBIAD | | | 38295-8 | 42752-5 | Done | FEBIAD development part 2/3 | Reminder. Mini pump still linked to SV1. DO NOT OPEN SV2!!!
The plan for today Wednesday, July 01, 2020 (  ) is:
- Coil sweep with gas lines opened.
- Coil sweep with gas lines closed
- Coil sweep with gas lines opened.
- Efficiency measurements at highest and lowest FC6 signal
- Mass scan.
COVID-19 protocol: buddy is Marla Cervantes
New idea: Screenshot e-log on the ISAC TS team notebook: 20-07-01 FEBIAD development part 2 (Web view)
Wednesday, July 01, 2020, 11:50: FC6=6.89e-12A. IG1=2.76e-6T.
FC6 has dropped 60% and the vacuum 15% with the respect to the last set of measurements.
Maybe the mass is not Ar+ but K? Or the decrease in pressure has this large correlation to FC6 magnitude.
Wednesday, July 01, 2020, 12:41: Coil sweep up and down completed.
Wednesday, July 01, 2020, 12:47: SV3, SV4, and SV5 close
IG1=2.59e-6T. The gas bottles are definitely running out, as there is no much change in pressure whether they are opened or closed
Wednesday, July 01, 2020, 13:15: Coil sweep completed.
Wednesday, July 01, 2020, 13:22: FC6 and IG1 increasing. Probably due to the coil still at 90A. reducing coil current.
Wednesday, July 01, 2020, 13:29: SV1 open. SV3 open/closed x5. SV4 open/closed x5.
Wednesday, July 01, 2020, 13:31: SV3 and SV4 open.
Wednesday, July 01, 2020, 13:32: SV1 closed. SV5 open/closed x5.
Wednesday, July 01, 2020, 13:33 SV5 open. No trip!!!
Wednesday, July 01, 2020, 14:30: Coil=0A.
Wednesday, July 01, 2020, 14:34 : SV3 closed.
Wednesday, July 01, 2020, 14:37:SV4 closed.
Coffee break.
Wednesday, July 01, 2020, 14:54: FC6~~9.76E-13A. IG1= 2.44e-6T
Wednesday, July 01, 2020, 15:11:SV1 open. SV3 open/closed x5. SV4 open/closed x5.
Wednesday, July 01, 2020, 15:13: SV3 and SV4 open.
Wednesday, July 01, 2020, 15:30: SV1 closed. SV5 open/closed x5.
Wednesday, July 01, 2020, 15:30: SV5 open. No trip!!!
Wednesday, July 01, 2020, 15:52: Coil=90A. FC6=1.41e-11A . IG1=2.68e-6T.
Wednesday, July 01, 2020, 15:57: FC6=1.16e-11A. IG1=2.68E-6T.
Wednesday, July 01, 2020, 15:58: SV3 closed.
Wednesday, July 01, 2020, 16:03: SV4 closed.
Wednesday, July 01, 2020, 16:07: 4.95e-12A. IG1=2.77e-6T.
Pressure increasing.
Wednesday, July 01, 2020, 16:14:SV1 open. FC6 decreasing momentarily, now increasing again due to the pressure increase.
Wednesday, July 01, 2020, 16:15: SV3 open/closed x5. SV4 open/closed x5.
IG1=3.1e-6T. FC6=6.00e-12A. Seems table now.
Wednesday, July 01, 2020, 16:17: SV3 and SV4 open.
Wednesday, July 01, 2020, 16:20: SV1 closed. SV5 open/closed x5.
Wednesday, July 01, 2020, 16:20: SV5 open. No trip!!!
Wednesday, July 01, 2020, 16:44: FC6=1.12e-11A. IG1=3.14e-6T.
Wednesday, July 01, 2020, 17:40: Mass scan completed. Peaks are really bad resolved.
Wednesday, July 01, 2020, 17:43;FC0 in. Coil=0A.
Wednesday, July 01, 2020, 17:43: Done for the day
The plan for today Thursday, July 02, 2020 is:
- Increase cathode current to 310A.
- Anode voltage sweep.
- Mass scan at nominal values.
- Coil sweep with gases
- Efficiency measurement at highest coil current.
- Coil sweep without gas.
- Mass scan without gas.
Thursday, July 02, 2020, 08:03: IG1=2.68e-6T. Vacuum decreased overnight, reading on FC0 decreased as well.
Thursday, July 02, 2020, 08:11: Auto-ramp TBHT to 310A @ 2A/min
Thursday, July 02, 2020, 09:15: FC0 increasing
Thursday, July 02, 2020, 10:30: FC0 seems to be stable now on 5.51e-9A.
Thursday, July 02, 2020, 11:10: Mass scan completed.
Thursday, July 02, 2020, 11:26: Mass=40.
Thursday, July 02, 2020, 13:47: Coil sweep completed.
Thursday, July 02, 2020, 13:48: Coil=90A. FC6=1.35e-10A. IG1=2.76e-6T.
Thursday, July 02, 2020, 13:49: SV3 closed.
Thursday, July 02, 2020, 13:52: SV4 closed.
Thursday, July 02, 2020, 14:00: SV5 closed.
Thursday, July 02, 2020, 14:02: FC6=5.80e-11A. IG1=2.59e-6T.
Thursday, July 02, 2020, 14:08 : SV1 open. SV3 open/closed x5. SV4 open/closed x5.
Thursday, July 02, 2020, 14:09: SV3 and SV4 open.
Thursday, July 02, 2020, 14:11: SV1 closed. SV5 open/closed x5.
Thursday, July 02, 2020, 14:11: SV5 open. No trip!!!
Thursday, July 02, 2020, 14:30: FC6=1.40e-10A. IG1=2.96e-6T.
Thursday, July 02, 2020, 14:32: SV3 closed.
Thursday, July 02, 2020, 14:35 : SV4 closed.
Thursday, July 02, 2020, 14:50 : SV5 closed.
Thursday, July 02, 2020, 15:28: FC6=3.34e-11A. IG1=2.59e-6T.
Thursday, July 02, 2020, 16:05: Coil sweep without gases finished.
Thursday, July 02, 2020, 16:35: Mass scan finished.
Thursday, July 02, 2020, 16:51: Coil=90A.
Thursday, July 02, 2020, 17:07: Mass scan completed.
Thursday, July 02, 2020, 17:07 : SV1 open. SV3 open/closed x5. SV4 open/closed x5.
Thursday, July 02, 2020, 17:09: SV3 and SV4 open.
Thursday, July 02, 2020, 17:10: SV1 closed. SV5 open/closed x5.
The mini-pump line was open for a moment while SV5 was opened, there was no trip but this might have added air into the system.
Thursday, July 02, 2020, 17:13: SV5 open. No trip!!!
Thursday, July 02, 2020, 17:21: FC6=4.10e-11A. IG1=2.76e-6T.
Thursday, July 02, 2020, 17:38: Mass scan completed.
Thursday, July 02, 2020, 17:40: Coil=0A.
Thursday, July 02, 2020, 17:56: Mass scan finished.
Thursday, July 02, 2020, 17:57: Done for the day.
The plan for today Friday, July 03, 2020 is:
- Emittance scans at difference coil values.
Friday, July 03, 2020, 11:45: Coil=0. Mass=40u
Friday, July 03, 2020, 11:45: FC6=4.12e-11A. IG1=2.72e-6T.
Friday, July 03, 2020, 12:13: Coil=0. Mass=28u. FC6=7.88e-10A. IG1=2.74e-6T.
Friday, July 03, 2020, 12:25: Test Stand tripped and the lights around flicker. EPICS whiteout. Even turbos shut off.
Friday, July 03, 2020, 12:27: Called control room and they told that there were a couple of power bumps.
Friday, July 03, 2020, 12:41: EPICS restarted but still whiteout problem similar to e-log 702. Likely to submit a faulty report
Friday, July 03, 2020, 12:57: PLC restarted, EPICS workings.
Friday, July 03, 2020, 12:58: Turbos at speed.
Friday, July 03, 2020, 13:25: Source/Optics on.
Friday, July 03, 2020, 13:36: BIAS/EZL/EE @ 30/13/1.5 kV.
Friday, July 03, 2020, 13:39: IG1=2.15e-6T.
Friday, July 03, 2020, 13:41: Auto-ramping TGHT/TBHT to 400/200A @ 2A/min for the weekend.
Friday, July 03, 2020, 13:44: Done for the day. To continue on Monday.
|
|
659
|
Monday, April 30, 2018, 12:07 |
Aaron Schmidt | | LP-FEBIAD | | | 41790-9 | 42752-5 | In progress | Test ready | Ready for C-seals or FEBIAD work
Monday, April 30, 2018, 12:05: Roughing.
Monday, April 30, 2018, 13:11: Turbo pumps on.
Monday, May 07, 2018, 15:02: RGA, ion gauge, and turbo pumps off. |
|
660
|
Tuesday, May 08, 2018, 10:36 |
Aaron Schmidt | | LP-FEBIAD | | | 41790-9 | 42752-5 | In progress | C-seal Test | 2 C-Seals installed with SS springs on water blocks. The SS springs are also used in the module.
Long pkg seal (A):
T: 0.098" x 0.429" Diameter.
Installed on line FS#5 , Block L, Heat shield cooling block. (Top seal, bottom seal of 2 is still the rubber O-ring)
Short pkg seal (B):
T: 0.095" x 0.424" Diameter.
Installed on line FS#3 , Block G, Coil cooling block.
All the waterlines are purged and shut. Open individually and monitor for leaks.
Tuesday, May 08, 2018, 10:27: Roughing
Tuesday, May 08, 2018, 10:31: Both BV1A & BV1B are open.
Tuesday, May 08, 2018, 10:57: Turbo Pumps started.
Tuesday, May 08, 2018, 11:03: Ion gauge on.
Tuesday, May 08, 2018, 11:15: RGA started.
Tuesday, May 08, 2018, 13:30: Vacuum @ 3.14e-06T
Wednesday, May 09, 2018, 06:48: Vacuum 5.13E-07T. RGA ok.
Wednesday, May 09, 2018, 06:50: Hook up Leak Detector. Leak detector # 5 (DRY) from Vacuum Group. Startup
Baseline Reading:
TP pressure 0.0e-04
Leak rate: Below 0.0e-09 atm-cc/sec
Valve to chamber opened. LD connection clear.
LD: 1.3e-08 atm-cc/sec
He TEST:
Wednesday, May 09, 2018, 07:00: Purge both lines with He. Do not pressurize.
Vacuum @ 5.13e-07T. RGA status:
Water @ 3.22e-007
He:-1.93e-009
LD: 1.3e-08 atm-cc/sec
Wednesday, May 09, 2018, 07:15: Open line FS#5 to heat shield. Pressurize to 30 PSI.
Wednesday, May 09, 2018, 07:40: Vacuum @ 5.02e-07T. RGA status:
Water @ 3.13e-007
He:-1.67e-009
LD: 9.3e-08 atm-cc/sec. Purge and close FS#5.
Wednesday, May 09, 2018, 07:45: Open line FS#3 to Coil block. Pressurize to 30 PSI.
08:10: Vacuum @ 4.9e-07T. RGA status:
Water @ 3.08e-007
He:-1.61e-009
LD: 1.5e-07 atm-cc/sec
Wednesday, May 09, 2018, 08:15: Depressurize lines. Connect HPA.
Wednesday, May 09, 2018, 08:20: Vacuum @ 4.76e-07T. RGA status:
Water @ 3.05e-007
He:-1.68e-009
LD: 1.7e-07 atm-cc/sec
08:25: LD: 1.7e-07 atm-cc/sec. Close Valve & disconnect LD
Water & Cycling TEST:
Wednesday, May 09, 2018, 08:45: Vacuum @ 4.67e-07T. RGA status:
Water: 3.02e-007
Open lines FS#2 and FS#4
Wednesday, May 09, 2018, 08:55: Vacuum @ 4.65e-07T. RGA status: Water @ 2.97e-007
Close lines FS#2 and FS#4. Water filled.
Open line FS#5 to heat shield. Vacuum and RGA ok.
09:00: Begin C-Seal (A) shock cycling. 20 cycles @ 5 min apart.
Close Supply/Return, Open Supply/Return to 'Shock'
TIME |
Vacuum (T) |
RGA (water) |
TIME |
Vacuum (T) |
RGA (Water) |
09:00 |
4.65e-07 |
2.94e-007 |
09:50 |
4.58e-07 |
2.87e-007 |
09:05 |
4.58e-07 |
2.93e-007 |
09:55 |
4.58e-07 |
2.88e-007 |
09:10 |
4.58e-07 |
2.93e-007 |
10:00 |
4.58e-07 |
2.87e-007 |
09:15 |
4.65e-07 |
2.92e-007 |
10:05 |
4.56e-07 |
2.87e-007 |
09:20 |
4.65e-07 |
2.92e-007 |
10:10 |
4.56e-07 |
2.85e-007 |
09:25 |
4.65e-07 |
2.90e-007 |
10:15 |
4.56e-07 |
2.86e-007 |
09:30 |
4.65e-07 |
2.89e-007 |
10:20 |
4.49e-07 |
2.83e-007 |
09:35 |
4.56e-07 |
2.89e-007 |
10:25 |
4.49e-07 |
2.84e-007 |
09:40 |
4.65e-07 |
2.89e-007 |
10:30 |
4.49e-07 |
2.84e-007 |
09:45 |
4.58e-07 |
2.89e-007 |
10:35 |
4.47e-07 |
2.83e-007 |
Wednesday, May 09, 2018, 10:40: Close line FS#5. Vacuum @ 4.49e-07T. RGA status: Water @ 2.84e-007
Open line FS#3 to Coil block. Vacuum and RGA ok
10:45: Begin C-Seal (B) shock cycling. 20 cycles @ 5 min apart
Close Supply/Return, Open Supply/Return to 'Shock'
TIME |
Vacuum (T) |
RGA (water) |
TIME |
Vacuum (T) |
RGA (Water) |
10:45 |
4.47e-07 |
2.80e-007 |
11:35 |
4.32e-07 |
2.77e-007 |
10:50 |
4.47e-07 |
2.80e-007 |
11:40 |
4.38e-07 |
2.75e-007 |
10:55 |
4.47e-07 |
2.81e-007 |
11:45 |
4.38e-07 |
2.75e-007 |
11:00 |
4.38e-07 |
2.80e-007 |
11:50 |
4.38e-07 |
2.75e-007 |
11:05 |
4.38e-07 |
2.78e-007 |
11:55 |
4.38e-07 |
2.75e-007 |
11:10 |
4.40-e07 |
2.79e-007 |
12:00 |
4.40e-07 |
2.75e-007 |
11:15 |
4.38e-07 |
2.77e-007 |
12:05 |
4.38e-07 |
2.73e-007 |
11:20 |
4.47e-07 |
2.77e-007 |
12:10 |
4.40e-07 |
2.73e-007 |
11:25 |
4.40e-07 |
2.78e-007 |
12:15 |
4.30e-07 |
2.71e-007 |
11:30 |
4.38e-07 |
2.76e-007 |
12:20 |
4.32e-07 |
2.73e-007 |
12:25: Open FS#5 and leave FS#3 open. Status Vacuum @ 4.32e-07T . RGA status: Water @ 2.70e-007
13:00: Status Vacuum @ 4.30e-07T . RGA status: Water @ 2.67e-007
Wednesday, May 09, 2018, 14:24: Status Vacuum @ 4.21e-07T . RGA status: Water @ 2.52e-007
Thursday, May 10, 2018, 07:41: Status Vacuum @ 3.24e-07T . RGA status: Water @ 1.76e-007.
Purge all water from lines. Prep for He.
Hook up Leak Detector. Leak detector # 5 (DRY) from Vacuum Group. Startup
Thursday, May 10, 2018, 08:00: Baseline Reading:
TP pressure 0.0e-04
Leak rate: Below 0.0e-09 atm-cc/sec
Vacuum @ 3.15e-07T . RGA status: Water @ 1.65e-007.
Valve to chamber opened. LD connection clear.
Leak rate: 1.6e-08 atm-cc/sec
He TEST #2:
Thursday, May 10, 2018, 08:13: Purge both lines with He. Do not pressurize.
Vacuum @ 3.15e-07T. RGA status:
Water @ 1.76e-007
He: -1.51e-009
LD: 1.5e-08 atm-cc/sec
Thursday, May 10, 2018, 08:15: Open line FS#3 to Coil Block. Pressurize to 30 PSI.
Thursday, May 10, 2018, 08:40: Vacuum @ 3.22e-07T . RGA status:
Water @ 1.80e-007
He:-1.98e-009
LD: 7.1e-08 atm-cc/sec. Purge and close FS#3.
Thursday, May 10, 2018, 08:45: Open line FS#5 to Heat Shield. Pressurize to 30 PSI.
Thursday, May 10, 2018, 09:10: Vacuum @ 3.22e-07T. RGA status:
Water @ 1.79e-007
He:-1.39e-009
LD: 1.5e-07 atm-cc/sec
09:15: Depressurize lines. Connect HPA.
09:30: Vacuum @ 3.22e-07T . RGA status:
Water @ 1.78e-007
He:-1.56-009
LD: 1.7e-07 atm-cc/sec
Thursday, May 10, 2018, 09:32: Close Valve & disconnect LD
Thursday, May 10, 2018, 09:35: Opened all waterlines completely. Vacuum stable @3.22e-07T and no response on the RGA.
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
24.8 |
2.83 |
TBHT |
24.7 |
3.09 |
COIL + EE |
24.7 |
5.08 |
TGHT |
24.9 |
5.77 |
Tuesday, May 15, 2018, 06:06: Auto ramping TGHT / TBHT to 580 / 270 A @ 2A per min.
Tuesday, May 15, 2018, 06:13: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .
Tuesday, May 15, 2018, 06:17: Optics set. IV4 opened.
Tuesday, May 15, 2018, 08:35: RGA Restarted. Spikes noticed on BIAS current. Vacuum @ 2.57E-06T
Wednesday, May 16, 2018, 06:40: TMS:IG6 (between 3-4pm yesterday) and the RGA were both knocked out. Restarting both.
Wednesday, May 16, 2018, 06:48: Repeated attempts to reset/restart IG6 have failed. Closing IV4
Friday, May 18, 2018, 10:55: Auto-cooling @ 1A per min. IG6 still non-responsive. |
|
664
|
Monday, May 28, 2018, 08:46 |
Aaron Schmidt | | LP-FEBIAD | | TM2 | 41790-9 | 42752-5 | In progress | New TM source tray swap out. | Monday, May 28, 2018, 08:45: Venting. Prep For removal.
Close FS#8. Clean up water
Adjust
Tuesday, May 29, 2018, 08:30: Test Stand source tray & Optics tray removed
Clean EZL on optics tray.
Tuesday, May 29, 2018, 11:30: New Source tray installed w/ Original TS optics tray. Connect.
Tuesday, May 29, 2018, 13:30: Roughing.
Tuesday, May 29, 2018, 14:38: Turbos on
Tuesday, May 29, 2018, 14:47: Ion gauge on.
Tuesday, May 29, 2018, 14:49: Exterior line FS# 8 opened.
Wednesday, May 30, 2018, 06:29: RGA on. Vacuum @ 3.38e-06T
He Leak TEST #1:
Vacuum Group Leak detector LD#2.
Friday, June 01, 2018, 08:30: Vacuum @ 9.15e-07T
LD Base line before opening valve to turbo pumps: 3.5e-10 atm-cc/sec
Friday, June 01, 2018, 08:40: LD baseline after opening valve to turbo pumps: 3.7e-10 atm-cc/sec
Full open base:
Vacuum @ 3.24e-07T. RGA status:
Water @ 7.76e-007
He: -2.31e-009
LD: 2.6e-09 atm-cc/sec
Friday, June 01, 2018, 10:30: Open line FS#2 . Pressurize to 30 PSI. Record after 10 minutes
Vacuum @ 9.24e-07T . RGA status:
Water @ 7.91e-007
He:-1.87e-009
LD: 2.8e-09 atm-cc/sec. Purge and close FS#2.
Friday, June 01, 2018, 10:45: Open line FS#3 . Pressurize to 30 PSI. Record after 10 minutes
Vacuum @ 9.24e-07T. RGA status:
Water @ 7.97e-007
He:-1.92e-009
LD: 3.0e-09 atm-cc/sec. Purge and close FS#3.
Friday, June 01, 2018, 11:00: Open line FS#4 . Pressurize to 30 PSI. Record after 10 minutes
Vacuum @ 9.21e-07T. RGA status:
Water @ 7.92e-007
He:-1.85e-009
LD: 4.7e-09 atm-cc/sec. Purge and close FS#4.
Friday, June 01, 2018, 11:15: Open line FS#5 . Pressurize to 30 PSI. Record after 10 minutes
Vacuum @ 9.24e-07T. RGA status:
Water @ 7.919e-007
He:-1.81e-009
LD: 4.0e-09 atm-cc/sec. Purge and close FS#5
Friday, June 01, 2018, 11:30: Depressurize lines. Connect HPA & Purge all lines.
After 5 minutes: Vacuum @ 9.24e-07T . RGA status:
Water @ 7.87e-007
He:-1.86-009
LD: 4.6e-09 atm-cc/sec
Close Valve & disconnect LD. HE LEAK TEST COMPLETE
Friday, June 01, 2018, 11:41: Open water supply and return mainlines. Vacuum @ 9.15e-07T.
Friday, June 01, 2018, 11:45: Open line FS#2. Vacuum ok. RGA ok
Friday, June 01, 2018, 11:46: Open line FS#3. Vacuum ok. RGA ok
Friday, June 01, 2018, 11:47: Open line FS#4. Vacuum ok. RGA ok
Friday, June 01, 2018, 11:48: Open line FS#5. Vacuum ok. RGA ok.
Friday, June 01, 2018, 12:05: Power/sources switched on.
Friday, June 01, 2018, 12:11: Auto ramping TGHT / TBHT to 300 / 150 A @ 2A per min.
Friday, June 01, 2018, 12:53: Power bump. Lights flickered, however TS ok.
Friday, June 01, 2018, 12:58: Monitored for 5 minutes. ok.
Friday, June 01, 2018, 15:08: Auto ramping TGHT to 350 A @ 2A per min.
Monday, June 04, 2018, 07:27: BIAS / EZL / EE @ 30 / 13 / 1.5 kV.
Monday, June 04, 2018, 07:28: Auto ramping TGHT / TBHT to 580 / 290 A @ 2A per min.
Monday, June 04, 2018, 07:30: Optics reset / Switched on. IV4 opened.
Monday, June 04, 2018, 10:00: FC0 GAIN on 1 uA = 1.65e-09. Wire scans no reading. Adjust steering.
Monday, June 04, 2018, 10:34: Anode @ 200V. Coil @ 10A.
Monday, June 04, 2018, 13:30: Optics tune changed with EZL off. Brad. S.
Monday, June 04, 2018, 13:45: Resetting tunes. COIL @ 50 A. Auto ramping TBHT to 300 A @ 2A per min. (310 A on read back)
Monday, June 04, 2018, 13:51: EZL / EE @ 13 / -1.5 kV.
Monday, June 04, 2018, 13:54: Optics reset to EZL on.
Tuesday, June 05, 2018, 08:00: RGA off.
Tuesday, June 05, 2018, 08:10: Auto ramping TBHT to 310 A @ 2A per min. (320 A on read back)
E-rig issue. Not scanning. Proceed with HV testing.
Tuesday, June 05, 2018, 09:00: BIAS / EZL / EE @ 35 / 15 / 1.75 kV.
Tuesday, June 05, 2018, 09:06: BIAS / EZL / EE @ 40 / 17.5 / 2 kV.
Tuesday, June 05, 2018, 09:13: Small spark knocked out optics & TMS:TP6
Tuesday, June 05, 2018, 09:21: IV4 closed/opened/reset. TP6 restarted.
Tuesday, June 05, 2018, 09:30: Spark*. Knocked out optics. Reset.
Tuesday, June 05, 2018, 10:37: BIAS / EZL / EE @ 45 / 19.5 / 2.25 kV.
Tuesday, June 05, 2018, 10:44: BIAS / EZL / EE @ 50 / 22 / 2.5 kV.
Tuesday, June 05, 2018, 10:47: Spark. Optics & CBC reset.
Tuesday, June 05, 2018, 11:25: BIAS / EZL / EE @ 52 / 24 / 2.8 kV. SPARK *knocked out optics, magnet & Turbo TMS:TP6. IV4 closed/opened/reset. TP6/optics restarted. Magnet still off.
Tuesday, June 05, 2018, 11:35: BIAS / EZL / EE @ 55 / 24 / 2.8 kV. SPARK* knocked out optics, magnet & Turbos TMS:TP6 & TP1B. IV4 closed/opened/reset. Turbos/optics restarted. Magnet still off.
Tuesday, June 05, 2018, 12:46: BIAS / EZL / EE @ 55 / 24 / 2.8 kV. SPARK* knocked out optics, magnet & Turbos TMS:TP6 & TP1B. IV4 closed/opened/reset. Turbos/optics restarted. Magnet still off.
Tuesday, June 05, 2018, 13:06: BIAS / EZL / EE @ 55 / 24 / 2.8 kV. SPARK* knocked out optics, magnet & Turbos TMS:TP6. IV4 closed/opened/reset. Turbo/optics restarted. Magnet still off. Anode @ 200V (reads 9V, shorted)
Tuesday, June 05, 2018, 13:16: TBHT brought down to 300A. Anode brought to 0V.
Tuesday, June 05, 2018, 13:20: Attempted to bring up to 30V. (30 input) Actual: 8V & 50.061 mA. Possible short, post spark*
Tuesday, June 05, 2018, 13:47: Auto-cooling @ 2A/min
Tuesday, June 05, 2018, 13:56: BIAS / EZL / EE @ 50 / 22 / 2.5 kV. Continue down to 30 kV settings.
Tuesday, June 05, 2018, 14:04: BIAS / EZL / EE @ 30 / 13 / 1.5 kV.
Wednesday, June 06, 2018, 06:49: Anode still unresponsive cold. Short*
Wednesday, June 06, 2018, 11:09: Vacuum @ 4.30e-07T ok. Test EZL
Wednesday, June 06, 2018, 11:16: BIAS / EZL / EE @ 30 / 20 / 1.5 kV.
Wednesday, June 06, 2018, 11:20: BIAS / EZL / EE @ 30 / 25 / 1.5 kV.
Wednesday, June 06, 2018, 11:24: BIAS / EZL / EE @ 30 / 30 / 1.5 kV.
Wednesday, June 06, 2018, 11:30: Coil @ 0A.
Wednesday, June 06, 2018, 12:51: No sparks* BIAS / EZL / EE @ 30 / 30 / 1.5 kV.
Wednesday, June 06, 2018, 12:53: Move Steerers to zero.
TSG:CB1: 0V
TSG:VCB1: 0V
TSG:CBC(2): 0V
Increase each one to 1000V individually. Hold 5 minutes and bump back to zero on each.
Wednesday, June 06, 2018, 12:55: TSG:CB1: 0 - 1000V. Hold for 5 minutes. Zero Sparks. Back to 0V.
Wednesday, June 06, 2018, 13:01: TSG:VCB1: 0 - 1000V. Goes red @ 1000V, stays blue @ 999V Hold for 5 minutes. Zero Sparks. Back to 0V.
Wednesday, June 06, 2018, 13:06: TSG:CBC(2): 0 - 1000V. Hold for 5 minutes. Zero Sparks. Back to 0V.
Wednesday, June 06, 2018, 13:40: Canbus contoller was swapped. The magnet is back.
Wednesday, June 06, 2018, 13:50: Sources to 0V.
Wednesday, June 06, 2018, 14:05: IV4 closed. Turbo pumps off. Ion gauge off.
Wednesday, June 06, 2018, 14:08: Power to the sources off.
Thursday, June 07, 2018, 06:35: Venting & waterlines purged. |
Attachment 1: New_TM_Source_Tray_2018.docx
|
Attachment 2: TS-2018-05-30-06-30.rga
|
Attachment 3: ER-580-310-30kV-m28.txt
|
value (Voltage offset 0.0000)
80.0000
0.7000
80.0000
29978.0220
21
21
./emit.dat
-0.0977 -0.1007 -0.0986 -0.0970 -0.1013 -0.0989 -0.0998 -0.0989 -0.1004 -0.0970 -0.0998 -0.0958 -0.1028 -0.1001 -0.1022 -0.0998 -0.1007 -0.1016 -0.1050 -0.1065 -0.1056
-0.0983 -0.0998 -0.1004 -0.1019 -0.1007 -0.1031 -0.0995 -0.0995 -0.1035 -0.0977 -0.1001 -0.0998 -0.1010 -0.0992 -0.1022 -0.0992 -0.1010 -0.0992 -0.1047 -0.1080 -0.1050
-0.0967 -0.0989 -0.0998 -0.1004 -0.1025 -0.1022 -0.1007 -0.1016 -0.1028 -0.0980 -0.1016 -0.1016 -0.1016 -0.0995 -0.0998 -0.1004 -0.0983 -0.0986 -0.1041 -0.1068 -0.0974
-0.0995 -0.0995 -0.1001 -0.1001 -0.0974 -0.0995 -0.1019 -0.1019 -0.0998 -0.1016 -0.0986 -0.1019 -0.1038 -0.0986 -0.1001 -0.0983 -0.1001 -0.0980 -0.1031 -0.1056 -0.1016
-0.0983 -0.1016 -0.0977 -0.1004 -0.0967 -0.0974 -0.1007 -0.1004 -0.1038 -0.1004 -0.1013 -0.1016 -0.1016 -0.0986 -0.1019 -0.1013 -0.1010 -0.1019 -0.1062 -0.1041 -0.1053
-0.0946 -0.0995 -0.0970 -0.1004 -0.1001 -0.1001 -0.1010 -0.0992 -0.1022 -0.0980 -0.1013 -0.1007 -0.1047 -0.1035 -0.1022 -0.1044 -0.1016 -0.1022 -0.1080 -0.1062 -0.1016
-0.0961 -0.0992 -0.0995 -0.0998 -0.1007 -0.1019 -0.0992 -0.1022 -0.1035 -0.1010 -0.0977 -0.1010 -0.0989 -0.0998 -0.1004 -0.1016 -0.1013 -0.1004 -0.1102 -0.1086 -0.0980
-0.0977 -0.0986 -0.1007 -0.0980 -0.0989 -0.1010 -0.1016 -0.1016 -0.1038 -0.1004 -0.1001 -0.0992 -0.0983 -0.1007 -0.0980 -0.1001 -0.0995 -0.0983 -0.1083 -0.1074 -0.0998
-0.0970 -0.1013 -0.0995 -0.0995 -0.1022 -0.0986 -0.1041 -0.0998 -0.1028 -0.1010 -0.1013 -0.0992 -0.1028 -0.1010 -0.1001 -0.0998 -0.1010 -0.0977 -0.1111 -0.1059 -0.1019
-0.0989 -0.0998 -0.0998 -0.1010 -0.1001 -0.0992 -0.1035 -0.1007 -0.1010 -0.0998 -0.1010 -0.1010 -0.1031 -0.1035 -0.1031 -0.1028 -0.1007 -0.1013 -0.1062 -0.1062 -0.0992
-0.0970 -0.0986 -0.0964 -0.1007 -0.0989 -0.1004 -0.1001 -0.1007 -0.0992 -0.0998 -0.1010 -0.0980 -0.1007 -0.0983 -0.1016 -0.1031 -0.0995 -0.0995 -0.1035 -0.1074 -0.1010
-0.0961 -0.0989 -0.0967 -0.1010 -0.0992 -0.1001 -0.1016 -0.1019 -0.1004 -0.1019 -0.0989 -0.0992 -0.1010 -0.0989 -0.1004 -0.1007 -0.0983 -0.1013 -0.1080 -0.1056 -0.1068
-0.0986 -0.0998 -0.0983 -0.1019 -0.1010 -0.0977 -0.1022 -0.0998 -0.1019 -0.1007 -0.0977 -0.0977 -0.0998 -0.0967 -0.0983 -0.1007 -0.0998 -0.0989 -0.1111 -0.1041 -0.0989
-0.0998 -0.0998 -0.0980 -0.0983 -0.0986 -0.1001 -0.1035 -0.0977 -0.1016 -0.1013 -0.1019 -0.1013 -0.1025 -0.0998 -0.1013 -0.1007 -0.1019 -0.0986 -0.1117 -0.1053 -0.0983
-0.0967 -0.0992 -0.0998 -0.0970 -0.0986 -0.1004 -0.1022 -0.1007 -0.1028 -0.0992 -0.1004 -0.1025 -0.1010 -0.1028 -0.0998 -0.1044 -0.1019 -0.1022 -0.1071 -0.1086 -0.1038
-0.0989 -0.0989 -0.1019 -0.0977 -0.0992 -0.1013 -0.1044 -0.1019 -0.1007 -0.0977 -0.0992 -0.1013 -0.1016 -0.1050 -0.1010 -0.1010 -0.1007 -0.1019 -0.1062 -0.1077 -0.1031
-0.0955 -0.0970 -0.0995 -0.1019 -0.1022 -0.1004 -0.1044 -0.1022 -0.1013 -0.0998 -0.1004 -0.1004 -0.0995 -0.1007 -0.0995 -0.0992 -0.0995 -0.1013 -0.1035 -0.1062 -0.1007
-0.0943 -0.0989 -0.0980 -0.1025 -0.1028 -0.1013 -0.1013 -0.1013 -0.1007 -0.0995 -0.1016 -0.0992 -0.0998 -0.0983 -0.0998 -0.1007 -0.0995 -0.1001 -0.1065 -0.1047 -0.1025
-0.0983 -0.0998 -0.0989 -0.1025 -0.0998 -0.1004 -0.1022 -0.1035 -0.1050 -0.1010 -0.1028 -0.1013 -0.1004 -0.0986 -0.0995 -0.1013 -0.0995 -0.0983 -0.1068 -0.1068 -0.0995
-0.0961 -0.1001 -0.0974 -0.0995 -0.0986 -0.0980 -0.1001 -0.1016 -0.1004 -0.1004 -0.1007 -0.0995 -0.1013 -0.1010 -0.0998 -0.1031 -0.1022 -0.0998 -0.1120 -0.1065 -0.0983
-0.0958 -0.0995 -0.0983 -0.1001 -0.0995 -0.0961 -0.1019 -0.0995 -0.1013 -0.0995 -0.0967 -0.1013 -0.1010 -0.0998 -0.1013 -0.1022 -0.1010 -0.1035 -0.1123 -0.1071 -0.1010
|
Attachment 4: ER-580-320-30kV-M39.txt
|
value (Voltage offset 0.0000)
80.0000
0.7000
160.0000
29978.0220
21
21
./emit.dat
-0.0845 -0.0845 -0.0906 -0.0940 -0.0659 -0.0629 -0.0525 -0.0476 -0.0516 -0.0879 -0.0839 -0.0858 -0.0864 -0.0821 -0.0854 -0.0848 -0.0848 -0.0839 -0.0943 -0.0888 -0.0882
-0.0842 -0.0830 -0.0864 -0.0922 -0.0635 -0.0632 -0.0497 -0.0482 -0.0510 -0.0894 -0.0833 -0.0830 -0.0790 -0.0839 -0.0851 -0.0821 -0.0827 -0.0870 -0.0943 -0.0897 -0.0867
-0.0851 -0.0833 -0.0845 -0.0867 -0.0671 -0.0595 -0.0519 -0.0507 -0.0522 -0.0909 -0.0851 -0.0851 -0.0870 -0.0885 -0.0842 -0.0851 -0.0839 -0.0842 -0.0864 -0.0873 -0.0854
-0.0851 -0.0854 -0.0845 -0.0861 -0.0699 -0.0626 -0.0531 -0.0473 -0.0522 -0.0925 -0.0854 -0.0864 -0.0821 -0.0867 -0.0854 -0.0842 -0.0821 -0.0836 -0.0876 -0.0891 -0.0864
-0.0842 -0.0854 -0.0851 -0.0854 -0.0693 -0.0610 -0.0534 -0.0519 -0.0549 -0.0925 -0.0861 -0.0867 -0.0851 -0.0848 -0.0851 -0.0854 -0.0827 -0.0824 -0.0937 -0.0900 -0.0854
-0.0836 -0.0842 -0.0861 -0.0870 -0.0699 -0.0607 -0.0528 -0.0476 -0.0577 -0.0931 -0.0861 -0.0854 -0.0861 -0.0842 -0.0845 -0.0854 -0.0848 -0.0854 -0.0922 -0.0885 -0.0848
-0.0839 -0.0836 -0.0836 -0.0815 -0.0681 -0.0601 -0.0497 -0.0482 -0.0531 -0.0909 -0.0827 -0.0836 -0.0778 -0.0836 -0.0836 -0.0830 -0.0842 -0.0833 -0.0912 -0.0861 -0.0836
-0.0867 -0.0836 -0.0848 -0.0836 -0.0653 -0.0580 -0.0510 -0.0488 -0.0513 -0.0900 -0.0821 -0.0818 -0.0861 -0.0845 -0.0824 -0.0833 -0.0836 -0.0827 -0.0919 -0.0867 -0.0854
-0.0851 -0.0864 -0.0836 -0.0867 -0.0671 -0.0583 -0.0543 -0.0476 -0.0504 -0.0873 -0.0839 -0.0858 -0.0864 -0.0848 -0.0836 -0.0854 -0.0842 -0.0854 -0.0922 -0.0879 -0.0879
-0.0803 -0.0845 -0.0827 -0.0854 -0.0717 -0.0607 -0.0513 -0.0494 -0.0522 -0.0919 -0.0861 -0.0870 -0.0845 -0.0851 -0.0851 -0.0851 -0.0861 -0.0873 -0.0977 -0.0894 -0.0867
-0.0845 -0.0827 -0.0842 -0.0839 -0.0705 -0.0607 -0.0522 -0.0470 -0.0531 -0.0894 -0.0858 -0.0864 -0.0848 -0.0851 -0.0854 -0.0833 -0.0842 -0.0848 -0.0946 -0.0876 -0.0851
-0.0824 -0.0867 -0.0845 -0.0839 -0.0696 -0.0586 -0.0513 -0.0461 -0.0543 -0.0916 -0.0839 -0.0827 -0.0842 -0.0845 -0.0830 -0.0827 -0.0815 -0.0836 -0.0851 -0.0864 -0.0818
-0.0830 -0.0861 -0.0848 -0.0861 -0.0702 -0.0604 -0.0488 -0.0519 -0.0552 -0.0912 -0.0836 -0.0842 -0.0824 -0.0836 -0.0830 -0.0848 -0.0824 -0.0827 -0.0858 -0.0864 -0.0842
-0.0867 -0.0867 -0.0864 -0.0854 -0.0671 -0.0595 -0.0543 -0.0507 -0.0528 -0.0925 -0.0845 -0.0842 -0.0848 -0.0848 -0.0845 -0.0848 -0.0836 -0.0836 -0.0934 -0.0876 -0.0867
-0.0867 -0.0870 -0.0864 -0.0879 -0.0681 -0.0601 -0.0519 -0.0522 -0.0546 -0.0922 -0.0851 -0.0867 -0.0851 -0.0861 -0.0858 -0.0870 -0.0845 -0.0845 -0.0970 -0.0888 -0.0882
-0.0848 -0.0848 -0.0854 -0.0867 -0.0711 -0.0583 -0.0522 -0.0479 -0.0537 -0.0882 -0.0839 -0.0858 -0.0833 -0.0851 -0.0848 -0.0836 -0.0839 -0.0830 -0.0961 -0.0870 -0.0876
-0.0858 -0.0839 -0.0836 -0.0812 -0.0684 -0.0568 -0.0519 -0.0479 -0.0531 -0.0873 -0.0836 -0.0836 -0.0848 -0.0818 -0.0827 -0.0812 -0.0845 -0.0848 -0.0922 -0.0882 -0.0854
-0.0830 -0.0827 -0.0824 -0.0836 -0.0656 -0.0574 -0.0528 -0.0467 -0.0540 -0.0919 -0.0836 -0.0842 -0.0812 -0.0818 -0.0839 -0.0830 -0.0842 -0.0836 -0.0912 -0.0867 -0.0858
-0.0842 -0.0845 -0.0821 -0.0876 -0.0690 -0.0607 -0.0510 -0.0500 -0.0531 -0.0928 -0.0842 -0.0833 -0.0885 -0.0854 -0.0851 -0.0858 -0.0848 -0.0858 -0.0937 -0.0888 -0.0827
-0.0867 -0.0848 -0.0854 -0.0854 -0.0720 -0.0620 -0.0534 -0.0476 -0.0534 -0.0934 -0.0848 -0.0870 -0.0836 -0.0864 -0.0854 -0.0870 -0.0848 -0.0864 -0.0952 -0.0888 -0.0879
-0.0848 -0.0833 -0.0851 -0.0864 -0.0687 -0.0592 -0.0552 -0.0507 -0.0522 -0.0873 -0.0851 -0.0839 -0.0848 -0.0839 -0.0845 -0.0845 -0.0833 -0.0836 -0.0955 -0.0882 -0.0827
|
Attachment 5: 2018_TM2_Source_Tray_in_Test_Stand_-_Test_Results.docx
|
|
670
|
Thursday, June 21, 2018, 16:32 |
Anders Mjos | | LP-FEBIAD | | | 41790-9 | 42752-5 | In progress | Heating target | Thursday, June 21, 2018, 16:30: Short between ANODE and EE. Do not use until repaired.
Thursday, June 21, 2018, 16:32: Ramping TBHT/TGHT to 290 / 580 Aat 2 A/min
Friday, June 22, 2018, 07:49: Target at temperature. Vacuum at 1.0E-7 T
Monday, June 25, 2018, 09:10: Auto-cooling at 2A/min - AS.
Friday, June 29, 2018, 08:52: Turbos/PS/Sources/RGA/Ion gauge off.
Friday, June 29, 2018, 10:03: Venting.
Friday, June 29, 2018, 14:32: Roughing. Anode to EE short fixed. |
Attachment 1: TS-2018-06-19-10-30.rga
|
Attachment 2: TS_Source_tray_2018_-_Check_2.docx
|
|
672
|
Monday, July 09, 2018, 10:44 |
Aaron Schmidt | | LP-FEBIAD | | | 41790-9 | 42752-5 | In progress | Coil Test July | Monday, July 09, 2018, 10:43: Rouging.
Monday, July 09, 2018, 13:14: Turbo pumps on.
Tuesday, July 10, 2018, 07:45: Waterlines opened. Vacuum ok @ 6.93e-07T. No response on RGA.
Tuesday, July 10, 2018, 08:05: Opened gas lines TSH:SV3 & TSH:SV4. Vacuum spiked to 1.08e-06T. Response registered on RGA. Closed valves.
Tuesday, July 10, 2018, 08:22: Vacuum Holding at 1.08e-06T.
Tuesday, July 10, 2018, 08:24: Close valve to small gas line pump. Pump is running.
Tuesday, July 10, 2018, 09:02: Path to TSH:SV1 through SV2 & SV5 ok. Vacuum @ 5.55e-07T. Close and begin SV3 & SV4 purging.
Tuesday, July 10, 2018, 09:12: All gas lines opened.
Tuesday, July 10, 2018, 09:16: Power/sources on. Vacuum @ 7.26e-07T.
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
24.4 |
3.85 |
TBHT |
24.2 |
2.99 |
COIL + EE |
24.2 |
4.97 |
TGHT |
24.6 |
5.61 |
Tuesday, July 10, 2018, 09:18: Auto ramping TGHT / TBHT to 580 / 270 A @ 2A per min.
Tuesday, July 10, 2018, 09:20: IV4 opened. Optics on (already set).
Tuesday, July 10, 2018, 14:11: Abort auto-ramping. Increase TBHT ramping to 300A. Continue auto ramping TGHT / TBHT to 580 / 270 A @ 2A per min.
Tuesday, July 10, 2018, 13:56: BIAS/EZL/EE @ 30/13/1.5 kV. Anode set to 200 V
Tuesday, July 10, 2018, 14:12: Auto ramping complete.
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
34.6 |
3.85 |
TBHT |
25.0 |
2.97 |
COIL + EE |
24.9 |
5.00 |
TGHT |
26.8 |
5.60 |
Tuesday, July 10, 2018, 14:17: TBHT set to 300 A. Read back reads 312 A
Wednesday, July 11, 2018, 06:52: Coil to 10A.
Wednesday, July 11, 2018, 07:00: Coil to 20A.
Wednesday, July 11, 2018, 07:05: Coil to 30A.
Wednesday, July 11, 2018, 07:10: Coil to 40A.
Wednesday, July 11, 2018, 07:20: Coil to 50A.
Fernando Maldonado
Wednesday, July 11, 2018, 09:59 Anode current=4.884 mA, Anode Voltage=200V. FC0=1.59e-9 . Source seems to be too cold. ramping up TBHT to 320 A (for a 330A readback). Vaccum=9.06e-7
Wednesday, July 11, 2018, 10:18 Ramp up compleete TBHT=320A Read Back=33.8A. Anode current=9.768mA. Anode voltage=200V. FC0=1.77e-8A. Vaccum=9.13e-7
Wednesday, July 11, 2018, 11:52 Mass scan performed for nominal values
Lunch
Steering
Adjusting Emittance parameter for more precise scan
Wednesday, July 11, 2018, 15:37 Cathode is still "cold" , at 325A and 250V the anode current is 15 mA. The way to obtained more current and to have better emittances readings is to increase the tbht current
Wednesday, July 11, 2018, 15:39 TBHT=325, TBHT RB =339.1A, TBHT PS=330A. Increasing TBHT to 335A. Where TBHT RB is the read back value. TBHT PS is the value display on the Power Supply
Wednesday, July 11, 2018, 15:46 tbht=335, tbht rb=349.7, tbht ps=341. With this the anode current is 24.42 at anode voltage=250V. *THIS WILL BE THE MAXIMUM VALUE"
Wednesday, July 11, 2018, 17:15 Mass Scan measurement seem to not vary with anode voltage
Wednesday, July 11, 2018, 17:45 Coil to 60 A to let out gass a bit more overnight. Vacuum ok at 9.6e-7
Thursday, July 12, 2018, 07:25: Power outage. (07/11/2018 @ 21:53 approx) The test stand was completely knocked out. Power & pumps were all shut down.
Thursday, July 12, 2018, 07:31: Power supplies turned off and all settings manually re-zeroed. |
|
673
|
Thursday, July 12, 2018, 08:24 |
Aaron Schmidt | | LP-FEBIAD | | | 41790-9 | 42752-5 | In progress | Power outage / restart | Power outage (07/11/2018 @ approx 21:53) shut down the test stand.

Thursday, July 12, 2018, 08:17: Reset vacuum controls & IV4
Thursday, July 12, 2018, 08:19: Restart all the turbo pumps.
Thursday, July 12, 2018, 08:24: IG6 & IG1 restarted.
Thursday, July 12, 2018, 08:33: Power on.
Thursday, July 12, 2018, 08:36: Opened gas lines SV2 & SV5 for FEBIAD to pump down lines with the chamber.
Thursday, July 12, 2018, 08:56: Reset optics. Operation appears ok.
Thursday, July 12, 2018, 09:02: Sources except EE & BIAS are reset. BIAS requires EE and the EE won't restart / respond.
Thursday, July 12, 2018, 09:31: Cycle EE PS? No response.
Thursday, July 12, 2018, 10:08: Cycle PLC. EE is responsive / back. BIAS operates as well.
Thursday, July 12, 2018, 10:10: Opened IV4. |
|
674
|
Thursday, July 12, 2018, 10:11 |
Aaron Schmidt | | LP-FEBIAD | | | 41790-9 | 42752-5 | In progress | Coil Test #5 | *** REVIEW Elog entry #673 Power outage & restart ***
Thursday, July 12, 2018, 10:10: Opened IV4
Thursday, July 12, 2018, 10:20: Attempt auto-ramp cancelled. Max increment values reached. Wait on Vacuum.
Thursday, July 12, 2018, 13:10: Auto-ramping TGHT / TBHT to 500 / 200 A @ 2A per min.
Thursday, July 12, 2018, 13:16: BIAS/EZL/EE @ 30/13/1.5 kV.
Thursday, July 12, 2018, 13:19: Anode set to 200 V
Thursday, July 12, 2018, 13:23: Gas lines SV1, SV3, SV4 opened without issue.
Friday, July 13, 2018, 07:02: Auto-ramping TGHT / TBHT to 580 / 300 A @ 2A per min.
FM:
Friday, July 13, 2018, 09:26 Ramping up to TBHT=335A to get at least 20 mA on the anode current.
Friday, July 13, 2018, 09:52 Ramp up finished: TBHT=335A, TBHT RB=349.6A, TBHT PS=341A. Anode current=20.757mA@200V. Vaccum ok@8.86e-7Torr
Friday, July 13, 2018, 09:56 Closing Neon valve in order to only deliver Argon.
Friday, July 13, 2018, 13:32 When anode voltage=0V, there are four peaks on the mass scan on masses 23,38,41,83 u. This indicate surface ionized particles. Most likely Sodium (23) and Potassium (38,41) and not sure about 83.
Friday, July 13, 2018, 16:58 Cooling it down for the weekend: TBHT=250A, TGHT=450
Friday, July 13, 2018, 17:01 wow! accidentally, changed to those values immediatly. I hope this does not compromise the cathode - Fernando
Monday, July 16, 2018, 06:57: Odd RGA values. Restarting RGA.
Monday, July 16, 2018, 07:14: RGA only displays Water, Nitrogen, and Oxygen. Investigate.
Monday, July 16, 2018, 09:57 Auto-ramping TGHT / TBHT to 580 / 300 A @ 2A per min from 449 / 209 A.
Monday, July 16, 2018, 10:00: Restarted the RGA with an older file loaded. RGA still only displays lines for Water, Nitrogen, and Oxygen. A table of values is still given for the other masses.
Monday, July 16, 2018, 11:08: TGHT / TBHT to 580 / 320 A. Anode to 200V
Monday, July 16, 2018, 11:19: Neon valve carefully opened. Vacuum stable @ 6.56e-07T. Neon response also showed up on the RGA.
Monday, July 16, 2018, 11:25: TBHT @ 340.0A (332A on PS)
FM
Monday, July 16, 2018, 11:55 TBHT@340A @ 354.7A (RB)@346A PS
Monday, July 16, 2018, 15:02 TBHT@345A @ 359.8A (RB)@351A PS
Monday, July 16, 2018, 15:18 Investigation the possible location of the surface ionized ions by sweeping the voltage in lower values <10V
Monday, July 16, 2018, 15:29 Values swept: 10,8,6,4,2,1 V. the same peak was observed at all cases
Monday, July 16, 2018, 15:52 Closing Neon Valve for efficiency measurement.
Monday, July 16, 2018, 15:58 Steady ion source current. closing Argon valve.
Monday, July 16, 2018, 17:00 Opening Argon Valve
Monday, July 16, 2018, 17:51 Argon valve closed. Neon Valve open. Measuring Mass=20
Monday, July 16, 2018, 18:13 Manually tapped TBHT down to 325A.
FM
Tuesday, July 17, 2018, 09:53 Ramping up TBHT to 345A
Tuesday, July 17, 2018, 10:13 Ramp up finish. TBHT@345A @ 359.7A (RB)@351A PS. Vacuum ok @ 7.43e-7 T
Tuesday, July 17, 2018, 10:14 Restarting quads that were off.
Tuesday, July 17, 2018, 10:18 Tuning
Tuesday, July 17, 2018, 11:01 Measurements will start for mass 20.
Tuesday, July 17, 2018, 11:59 Lunch break
Tuesday, July 17, 2018, 12:41 Coming back. TMSQ2 & TMSQ5 off. Turning them on.
Tuesday, July 17, 2018, 13:54 Retuning emittance scanner to default parameters. Cycle the emittance scanner PS.
Tuesday, July 17, 2018, 14:18 Increasing current limit for anode. From 51.282 mA to 51.400 mA
Tuesday, July 17, 2018, 14:34 Increasing current limin on anode to 52.00 mA
Tuesday, July 17, 2018, 14:35 Closing Neon Valve
Tuesday, July 17, 2018, 15:02 FC6 stable
Tuesday, July 17, 2018, 15:02 Opening "Argon" bottle
Tuesday, July 17, 2018, 15:21 Fc6 stable. The present current lies between the on/off Neon Valve.
Tuesday, July 17, 2018, 15:30 Opening Neon Valve
Tuesday, July 17, 2018, 15:51 Re-tuning. Taking a mass scan.
Tuesday, July 17, 2018, 15:55 Checking the RGA, I noticed that the maximum reading for Neon was 7.05e-8T, and for Argon, 6.15e-7T
Tuesday, July 17, 2018, 16:45 Increasing Anode current limit to 100mA. Increasing TBHT to 347A, 361.8 RB, 353 PS. Vaccum ok@7.87e-7T
Tuesday, July 17, 2018, 16:54 Finish for today.
FM
Wednesday, July 18, 2018, 09:27 Starting new round of measurements.
Wednesday, July 18, 2018, 09:31 TBHT@350A@364.5A (RB) @ 356 (PS). Vacuum ok at 6.71e-7T
Wednesday, July 18, 2018, 11:48 Finish mass scan sweeps. Leaving for lunch. Coming back at around 3pm due to science week
Wednesday, July 18, 2018, 15:29 Starting again.
Wednesday, July 18, 2018, 15:30 Mass scan at AB=0V
Wednesday, July 18, 2018, 15:56 Mass scan set to 38.5 (K).
Wednesday, July 18, 2018, 15:56 Measuring emittance.
Wednesday, July 18, 2018, 16:02 Closing Argon bottle.
Wednesday, July 18, 2018, 16:08 Closing Neon bottle.
Wednesday, July 18, 2018, 16:13 Measuring what it looks to be a potassium signal. Let's see if the signal decreases overnight.
Friday, July 20, 2018, 11:48: FC0 put in place & the gas lines closed. - Aaron S.
Friday, July 20, 2018, 11:50 : Auto cooling @ 2A per min.
Monday, July 23, 2018, 07:05: Sources off. Optics checked.
Monday, July 23, 2018, 07:13 : RGA error. Restart for proper shut off. IV4 closed.
Monday, July 23, 2018, 07:14 : Ion gauge, turbos off.
|
|
687
|
Tuesday, October 02, 2018, 13:39 |
Aaron Schmidt | | LP-FEBIAD | | | 41790-9 | 42752-5 | In progress | Test FEBIAD w/ Silver bolts & sealed end caps w/ stoppage | Silver bolts used on Coil, TGHT, & TBHT.
End caps in target
Tuesday, October 02, 2018, 13:36: Roughing.
Tuesday, October 02, 2018, 13:37: Doors sealed.
Tuesday, October 02, 2018, 14:10: Turbos on.
Due to potential power & cooling loss. Shutdown vac.
Wednesday, October 03, 2018, 07:34: Chamber Turbos off.
Wednesday, October 03, 2018, 10:58: Beam line Turbos off. Closed both PV1 & PV2 & turned off both roughing pumps.
Thursday, October 11, 2018, 10:37: RGA on. Strip tools re-opened.
Thursday, October 11, 2018, 10:43: Waterlines opened. No response on RGA. Vacuum stable @ 3.40e-07T.
Thursday, October 11, 2018, 10:48: Power on.
Thursday, October 11, 2018, 10:49: Opened manual valve in gas control to operate small pump. Vacuum spike into Xe-05T range.
Thursday, October 11, 2018, 10:51: Very gradual opening on source gas control. Pump ineffective.
Thursday, October 11, 2018, 10:52: SV5 not opened yet. Open SV2. Vacuum spike noted.
Thursday, October 11, 2018, 10:53: SV5 opened/closed very quickly. Allow stabilization. Alternate between SV5 & SV2 until stable.
Thursday, October 11, 2018, 10:58: SV5 closed. Open SV4. Vacuum spike noted up to 3.3e-05T.
Thursday, October 11, 2018, 11:30: All Gas lines opened.
Thursday, October 11, 2018, 11:36: Sources on.
Thursday, October 11, 2018, 11:43: BIAS / EZL / EE @ 30 / 13 / 1.5 kV.
Thursday, October 11, 2018, 11:46: IV4 opened. Error / closed / red. Fix. IV4 re-opened.
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
21.9 |
3.96 |
TBHT |
21.7 |
3.09 |
COIL + EE |
22.0 |
5.15 |
TGHT |
22.2 |
5.83 |
Thursday, October 11, 2018, 12:08: Auto-ramping TGHT / TBHT to 300 / 200 A @ 2A per min
Friday, October 12, 2018, 08:07: Auto-ramping TGHT / TBHT to 580 / 290 A @ 2A per min.
Friday, October 12, 2018, 08:13: Optics reset.Can Diagnostics re-cfg.
Friday, October 12, 2018, 10:49: Anode @ 200 V.
Friday, October 12, 2018, 10:50: Coil @ 10 A. Vacuum @ 6.25e-07T
Friday, October 12, 2018, 10:53: Coil @ 20 A. Vacuum @ 6.27e-07T
Friday, October 12, 2018, 10:56: Coil @ 30 A. Vacuum @ 6.27e-07T
Friday, October 12, 2018, 10:58: Coil @ 40 A. Vacuum @ 6.34e-07T
Friday, October 12, 2018, 11:00: Coil @ 50 A. Vacuum @ 6.34e-07T
Friday, October 12, 2018, 11:01: Auto-ramping TBHT to 300 A @ 2A per min.
Friday, October 12, 2018, 11:06: TBHT @ 300 A.
Friday, October 12, 2018, 11:51: Nothing found on Mass or wire scans?
Friday, October 12, 2018, 11:52: Anode set to 0 V.
Friday, October 12, 2018, 11:54: Still nothing found on wire scans.
Fernando:
Friday, October 12, 2018, 12:56: Anode leakage current is 6.105 mA on the readback with AV@200V Coil@0A TGHT@580A TBHT@310
Friday, October 12, 2018, 12:59 : This is very low anode lakage current. It should be on the bulk of 100 mA.
Friday, October 12, 2018, 12:59: Ramping up TBHT to 320 A @ 2A per min while keeping the AV@200 V. STOP
Aaron: Anode / Cathode BIAS PS polarity requires a reversal. John Chow notified. To be reversed Monday after 11am.
Friday, October 12, 2018, 13:24: Continue auto-ramping TBHT to 320 A @ 2A per min.
Friday, October 12, 2018, 13:30: TBHT 320 A / Read back @ 313.1 A.
Friday, October 12, 2018, 13:34: Wire & mass scans
Friday, October 12, 2018, 13:54: Anode to 0 V. Coil to 50 A.
Monday, October 15, 2018, 07:14: Auto-cooling. Coil to 0 A.
Monday, October 15, 2018, 11:54: Sources/power off. IV4 closed |
Attachment 1: FEBIAD_Coil_Test_6_-_POS_Cathode.docx
|
Attachment 2: TS-2018-10-11-10-37.rga
|
|