Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  RH-ISAC, Page 115 of 141  Not logged in ELOG logo
  ID Date Author Category Type Module Target/Number Subjectdown
  1291   Monday, July 25, 2016, 08:15 Anders MjosConditioning StationRepairTM3 HV Test

 Monday, July 25, 2016, 08:14: Enabled staging for chiller

Monday, July 25, 2016, 08:33: Disabled staging. Fault 2. Staging does not work at this point.

Monday, July 25, 2016, 08:34: Running CP1 manually to cool down water.

Monday, July 25, 2016, 09:15: Bias current limiting at 150 V. Investigating short.

Monday, July 25, 2016, 09:51: EE cable was shorting the BIAS on the TCS side.

Monday, July 25, 2016, 09:57: BIAS/ / EL at 20 / 10 kV

Monday, July 25, 2016, 10:08: Sparking at 24 kV. Lowered to 22 kV

Monday, July 25, 2016, 10:44: BIAS at 25 kV

 Monday, July 25, 2016, 11:00: Sparking at 25.5 kV. Lowered to 25 kV

Monday, July 25, 2016, 12:03: 25.5 kV. No sparking for 10 min. Lower to 20 kV for lunch break.

Monday, July 25, 2016, 12:51: 26 kV

Monday, July 25, 2016, 13:00: Sparking at 27.3 kV. Lowered to 27 kV. EL at 14 kV

Monday, July 25, 2016, 13:04: More sparking. Lowered to 25 kV

Monday, July 25, 2016, 13:39: 25 kV

Monday, July 25, 2016, 14:21: A spark tripped the PS on water resistivity

Monday, July 25, 2016, 14:31: 25 kV

Monday, July 25, 2016, 15:34: ~20 sparks for the last 60 min

Monday, July 25, 2016, 16:28: ~10 sparks for the last 60 min. BIAS an EL off

Monday, July 25, 2016, 16:36: Stopped water cooling system.


Tuesday, July 26, 2016, 08:20: Started water cooling system.

Tuesday, July 26, 2016, 08:24: Ran CP1 manually for 3 min. Enabled staging.

Tuesday, July 26, 2016, 08:31: BIAS / EL at 25 / 0 kV

Tuesday, July 26, 2016, 08:39: EL float to 300 V. Set to 500 V

Tuesday, July 26, 2016, 08:45: BIAS to 25.5 kV

Tuesday, July 26, 2016, 09:07: 26 kV

Tuesday, July 26, 2016, 09:12: Lost contact with EPICS. Scheduled maintenance

Tuesday, July 26, 2016, 10:26: 6 sparks for the last 60 min

Tuesday, July 26, 2016, 10:33: BIAS to 27 kV

Tuesday, July 26, 2016, 10:43: Sparking at 27.8 kV. Lowered to 23 kV

Tuesday, July 26, 2016, 10:51: BIAS back at 27 kV. A spark tripped the BIAS on water resistivity

Tuesday, July 26, 2016, 11:08: BIAS at 27 kV

Tuesday, July 26, 2016, 11:22: BIAS at 28 kV

Tuesday, July 26, 2016, 11:42: BIAS at 27 kV after trip due to water resistivity (spark)

Tuesday, July 26, 2016, 12:00: BIAS / EL off for lunch break

Tuesday, July 26, 2016, 14:10: EL to 10 kV

Tuesday, July 26, 2016, 14:14: El to 20 kV

Tuesday, July 26, 2016, 14:20: EL to 25 kV

Tuesday, July 26, 2016, 14:27: First spark at 27.6 kV

Tuesday, July 26, 2016, 14:29: EL to 28 kV (BIAS at 0 kV)

Tuesday, July 26, 2016, 14:43: RGA on

Tuesday, July 26, 2016, 14:49: No sparking for 20 min

Tuesday, July 26, 2016, 16:02: No sparks for 90 min.

Tuesday, July 26, 2016, 16:06: PS and water off.


Wednesday, July 27, 2016, 09:14: BIAS at 24 kV

Wednesday, July 27, 2016, 09:19: Lowered to 20 kV due to sparking

Wednesday, July 27, 2016, 09:28: BIAS at 22 kV

Wednesday, July 27, 2016, 15:30: BIAS and water off


Thursday, July 28, 2016, 08:43: RGA off

Attachment 1: TM3_HV_conditioning_2016-07-25.docx
Attachment 2: TCS-2016-07-25_TM3_no_target.rga
Attachment 3: TCS-2016-07-26_TM3_no_target.rga
  1637   Friday, October 06, 2017, 13:25 Anders MjosConditioning StationRepair  HV Test

 HV was brought up to 70 kV for 1 h without sparks to test the new water installation.

Attachment 1: 2017-10-06_HV_Water_Line_Test.xlsx
  578   Thursday, December 12, 2013, 11:14 Bevan MossConditioning StationStandard OperationTM2 HV Chase Electrical Test

While under vacuum Travis and I completed diagnostics of the HV chase. Majority of the lines showed OL at 1kV but there were the following "failures"

A to P @ 250V 716kOhm
A to Q @ 1kV 25.18MOhm
B to C with ohm meter 8Mohm @250V 830kOhm
B to D @ 1kV 15.57MOhm
C to D @1kV 16.18 MOhm
P to Q @1kV 1.069 GOhm
G to J with ohm meter dead short.

Ideally all of these values would be OL and they may improve or get worse when the source tray is installed depending on where the electrical contact is. I need to discuss with Friedhelm but I believe that as these values the module can still operate. Without changing the HV chase there is very little chance of improving these values.

  2385   Thursday, November 03, 2022, 13:15 David WangConditioning StationStandard OperationTM2UCx#40HS line helium pressure test. TM4 with UCx#40at TCS.

Helium 70 psi. tested 5minutes at base leak rate 0.9 xE09 atm.cc/sec on HS line.  No leak.

  2374   Wednesday, October 05, 2022, 09:31 David WangConditioning StationStandard OperationTM4C#4HS line helium pressure test. TM4 with C#4 at TCS.

TM4 HS line helium pressure test is done. 70psi helium pressure@ 3XE-8 atm.cc/sec . No leak.

  2356   Thursday, August 11, 2022, 09:10 David WangConditioning StationStandard OperationTM4SiC #44HS line helium pressure test. Gas line fitting on module top and gas line leak check.

HS line helium pressure test is done at TCS. No leak. I also leak checked gasline adapter and fittings on TM4 top. No leak.  1/4 copper gas line between module and ITE was leak checked yesterday. It is good.

  2364   Thursday, September 08, 2022, 09:06 David WangConditioning StationStandard OperationTM2UCx#39HS line helium pressure test.

TM2 with UCX#39 HS line helium pressure test is done at TCS. 70 psi helium 3minutes. No leak @0.0XE-9 atm.cc/sec.

  1980   Monday, September 09, 2019, 06:38 David WangConditioning StationStandard OperationTM4UCx27HS and window line leak check.

TM4 HS and window line are helium pressure checked at TCS. 70 psi helium 5 minutes. No response @1.2XE-8 atm.cc/sec. They are leak free.

  485   Wednesday, October 02, 2013, 09:40 Bevan MossConditioning StationStandard OperationTM3NiO#2HP conditioning

Yesterday the TBHT was ramped to 250 amp and the TGHT was ramped to 440 amps. Attached is the data from that run the strip tool that had this screen shot crashed but it can be recreated with the data. The TBHT ps was not reading back accurate results as such there are large steps in current or voltage or some instability for no apparent reason. In addition to this the ramping program is still not working so the ramping was a very manual process. The strip tool was left running over night, I have attached a screen shot of the the run. You will notice there are changes in vacuum that corresponds to changes in the TGHT and the TBHT voltage. Anders and myself used the archiver to plot the data with TS6 (TS3 would be better and I will have it added to the achiver), the changes in voltage and vacuum match perfectly with the compressor coming on and off. This should be investigated and possibly the temperature range of the cycling of the compressor reduced.

Attachment 1: 01102013_TM3_NiO2_TGHT_and_TBHT_ramp_up_eod.txt
Time	TCS:IG1:RDVAC [T]	TCS:IG2:RDVAC [T]	TCS:IG1S:RDVAC [T]	TCS:TGHT:RDVOL [V]	TCS:TGHT:RDCUR [A]	TCS:TBHT:RDVOL [V]	TCS:TBHT:RDCUR [A]	
10/01/2013 12:26:24.968541	                      0.000003494                      0.000008483                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:25.968221	                      0.000003516                      0.000008483                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:26.968076	                      0.000003516                      0.000008483                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:27.967994	                      0.000003494                      0.000008483                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:28.967909	                      0.000003516                      0.000008483                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:29.967723	                      0.000003472                      0.000008461                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:30.967596	                      0.000003494                      0.000008461                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:31.967466	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:32.992188	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:33.992187	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:34.992103	                      0.000003516                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:35.991944	                      0.000003516                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:36.991845	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:37.991694	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:38.992589	                      0.000003494                      0.000008505                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:39.992441	                      0.000003494                      0.000008483                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:40.992335	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:42.60462	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:43.60139	                      0.000003516                      0.000008461                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:44.61035	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:45.60806	                      0.000003494                      0.000008505                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:46.60702	                      0.000003494                      0.000008505                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:47.60766	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:48.61462	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:49.62355	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:50.62179	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:51.62047	                      0.000003494                      0.000008483                      0.000007428                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:52.61945	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:53.61799	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:54.62693	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:55.62785	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:56.62449	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:57.62550	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:58.63169	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:59.64295	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:00.63960	                      0.000003472                      0.000008461                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:01.63807	                      0.000003472                      0.000008461                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:02.63666	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:03.63581	                      0.000003494                      0.000008505                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:04.64414	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:05.64537	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:06.64172	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:07.64079	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:08.63970	                      0.000003516                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:09.64820	                      0.000003516                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:10.64701	                      0.000003494                      0.000008505                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:11.64536	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:12.64452	                      0.000003516                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:13.64287	                      0.000003516                      0.000008505                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:14.65215	                      0.000003494                      0.000008505                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:15.65043	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:16.64919	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:17.64797	                      0.000003494                      0.000008505                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:18.64668	                      0.000003494                      0.000008505                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:19.65546	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:20.65649	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:21.66282	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:22.66192	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:23.66088	                      0.000003494                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:24.66997	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:25.67780	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:26.67657	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:27.68761	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:28.69545	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:29.173136	                      0.000003516                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:30.173380	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:31.174041	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:32.173956	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:33.175017	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:34.175756	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:35.175519	                      0.000003472                      0.000008483                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:36.175439	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:37.175269	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:38.177533	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:39.178038	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:40.177940	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:41.177878	                      0.000003494                      0.000008483                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:42.177674	                      0.000003450                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:43.177534	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:44.178494	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:45.178420	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:47.556742	                      0.000003472                      0.000008483                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:48.556877	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:49.556716	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:50.556647	                      0.000003450                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:51.556470	                      0.000003450                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:52.555722	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:53.556262	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:54.556119	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:55.555966	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:56.557075	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:57.556794	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:58.556582	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:59.556637	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:00.556339	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:01.556215	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:02.556121	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:03.555991	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:04.556884	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:05.557704	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:06.557663	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:07.557508	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:08.557363	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:09.557201	                      0.000003494                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:10.557206	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:11.556952	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:12.603135	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:13.603729	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:14.603627	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:15.604448	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:16.604357	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:17.604557	                      0.000003494                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:18.604130	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:19.604964	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:20.604817	                      0.000003472                      0.000008461                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:21.605840	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:22.605655	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:23.605561	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:24.606470	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:25.606226	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:26.606108	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:27.607225	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:28.606838	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:29.607689	                      0.000003472                      0.000008461                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:30.607592	                      0.000003450                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:31.607731	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:32.607319	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:33.607229	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:34.608083	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:35.607942	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:36.607853	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:37.639580	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:38.639598	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:39.639435	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:40.639309	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:41.639237	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:42.639092	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:43.642376	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:44.642881	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:45.643933	                      0.000003450                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:46.669142	                      0.000003450                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:47.669528	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:48.670348	                      0.000003450                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:49.671254	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:50.692124	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:51.692466	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:52.691865	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:53.691703	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:54.692702	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:55.692422	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:56.692317	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:57.692221	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:58.692092	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:59.692955	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:00.692831	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:01.692670	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:02.692596	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:03.692421	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:04.693353	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:05.694193	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:06.773807	                      0.000003450                      0.000008439                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:07.773908	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:08.773785	                      0.000003450                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:09.773707	                      0.000003472                      0.000008461                      0.000007340                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:10.773565	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:11.773425	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:12.773282	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:13.774375	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:14.775067	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:15.774905	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:16.774981	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:17.775909	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:18.775561	                      0.000003450                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:19.775556	                      0.000003450                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:20.775386	                      0.000003450                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:21.775154	                      0.000003450                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:22.775046	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:23.774939	                      0.000003450                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:24.775322	                      0.000003428                      0.000008439                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:25.775712	                      0.000003450                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:26.775518	                      0.000003450                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:27.776544	                      0.000003450                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:28.776275	                      0.000003450                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:29.776180	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:30.777108	                      0.000003450                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:31.789402	                      0.000003450                      0.000008483                      0.000007362                              0.0                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:32.789771	                      0.000003450                      0.000008483                      0.000007362                              0.0                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:33.789680	                      0.000003472                      0.000008483                      0.000007384                              0.0                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:34.789545	                      0.000003450                      0.000008483                      0.000007384                              0.0                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:35.789418	                      0.000003428                      0.000008461                      0.000007384                              0.0                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:37.442708	                      0.000003472                      0.000008461                      0.000007362                              0.0                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:38.442085	                      0.000003450                      0.000008461                      0.000007362                              0.0                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:39.442135	                      0.000003450                      0.000008483                      0.000007362                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:40.441811	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:41.442707	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:42.442561	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:43.442435	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:44.443379	                      0.000003472                      0.000008461                      0.000007362                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:45.443198	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:46.444259	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:47.444924	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:48.444821	                      0.000003450                      0.000008483                      0.000007384                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:49.444709	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:50.444638	                      0.000003472                      0.000008461                      0.000007362                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:51.549284	                      0.000003472                      0.000008461                      0.000007384                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:52.552322	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:29:53.552168	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:29:54.553247	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:29:55.553967	                      0.000003472                      0.000008461                      0.000007340                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:29:56.553827	                      0.000003450                      0.000008461                      0.000007406                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:29:57.553658	                      0.000003450                      0.000008417                      0.000007384                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:29:58.553540	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:29:59.554438	                      0.000003472                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:30:00.554431	                      0.000003472                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:30:01.555204	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:30:02.555036	                      0.000003450                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:30:03.554940	                      0.000003472                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:04.555874	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:05.555662	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:06.555571	                      0.000003472                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:07.555438	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:08.555439	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:09.556229	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:10.557048	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:11.556908	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:12.557182	                      0.000003428                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:13.557690	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:14.558563	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:15.558521	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:16.558282	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:17.558166	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:18.558067	                      0.000003428                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:19.557905	                      0.000003428                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:20.557772	                      0.000003428                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:21.557692	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:22.557585	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:23.662422	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:24.792773	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:25.793155	                      0.000003428                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:26.793121	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:27.814109	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:28.814848	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:29.816080	                      0.000003428                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:30.936335	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:31.936479	                      0.000003450                      0.000008439                      0.000007340                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:32.936542	                      0.000003428                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:33.937159	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:34.936998	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:36.28953	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:37.29782	                      0.000003450                      0.000008461                      0.000007340                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:38.101823	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:39.102451	                      0.000003450                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:40.102434	                      0.000003428                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:41.102235	                      0.000003428                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:42.102174	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:43.178823	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:44.178888	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:45.178756	                      0.000003472                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:46.178627	                      0.000003406                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:47.178529	                      0.000003428                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:48.178427	                      0.000003450                      0.000008439                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:49.178213	                      0.000003450                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:50.179230	                      0.000003450                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:51.179939	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:52.561149	                      0.000003428                      0.000008461                      0.000007340                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:53.563445	                      0.000003450                      0.000008439                      0.000007340                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:54.563591	                      0.000003450                      0.000008439                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:55.707336	                      0.000003406                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:56.732944	                      0.000003428                      0.000008439                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:57.846017	                      0.000003406                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:58.988827	                      0.000003428                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:00.90800	                      0.000003428                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:01.124740	                      0.000003450                      0.000008439                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:02.124607	                      0.000003428                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:03.124442	                      0.000003428                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:04.124375	                      0.000003406                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:05.124268	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:06.125092	                      0.000003450                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:07.124979	                      0.000003428                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:08.124845	                      0.000003450                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:09.124778	                      0.000003428                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:10.125618	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:11.125456	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:12.125314	                      0.000003450                      0.000008439                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:13.125188	                      0.000003450                      0.000008439                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:14.126061	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:15.126120	                      0.000003406                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:16.125873	                      0.000003450                      0.000008439                      0.000007340                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:17.125797	                      0.000003450                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:18.125584	                      0.000003450                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:19.125575	                      0.000003428                      0.000008439                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:20.126343	                      0.000003406                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:21.126184	                      0.000003406                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:22.126058	                      0.000003406                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:23.125975	                      0.000003428                      0.000008439                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:24.125801	                      0.000003428                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:25.125684	                      0.000003428                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:26.125551	                      0.000003428                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:27.126146	                      0.000003428                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
... 14745 more lines ...
Attachment 2: 02102013_TM3_NiO#2_HP_testing_overnight_run.JPG
02102013_TM3_NiO#2_HP_testing_overnight_run.JPG
Attachment 3: 02102013_TM3_NiO#2_HP_testing_overnight_run_archiver.JPG
02102013_TM3_NiO#2_HP_testing_overnight_run_archiver.JPG
  476   Friday, September 27, 2013, 12:53 Bevan MossConditioning StationStandard OperationTM3NiO#2HP TGHT and TBHT conditioning

 Yesterday TM3 had the TGHT current raised to 160 amps and the TBHT raised to 200 amps. At this power the TGHT showed typical results in that vacuum was affected and the increased voltage as the target increased in temperature. The tube heater on the other hand did not change voltage or show a change in vacuum. As a result it is believe there is a short somewhere in the module. I have attached a screen shot of the strip tool and the raw data.

Attachment 1: 26092013_TM3_NiO#2_HP_testing.JPG
26092013_TM3_NiO#2_HP_testing.JPG
Attachment 2: 26092013_TM3_NiO2_HP_testing.txt
Time	TCS:IG2:RDVAC [T]	TCS:IG1S:RDVAC [T]	TCS:TGHT:RDVOL [V]	TCS:TGHT:RDCUR [A]	TCS:TBHT:RDVOL [V]	TCS:TBHT:RDCUR [A]	
09/26/2013 13:56:46.578505	                      0.000008241                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:47.577921	                      0.000008263                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:48.577796	                      0.000008241                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:49.578660	                      0.000008241                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:50.578515	                      0.000008241                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:51.579406	                      0.000008241                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:52.579315	                      0.000008241                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:53.634912	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:54.635006	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:55.634932	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:56.635812	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:57.636661	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:58.636560	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:59.636386	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:00.636250	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:01.637154	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:02.637232	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:03.636895	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:04.636752	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:05.636626	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:06.636503	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:07.636378	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:08.636282	                      0.000008263                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:09.636155	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:10.635997	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:11.635868	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:12.635748	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:13.635872	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:14.636574	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:15.636412	                      0.000008263                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:16.636294	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:17.636120	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:18.635997	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:19.636923	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:20.636746	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:21.668993	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:22.714260	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:23.714359	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:24.714296	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:25.714103	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:26.714040	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:27.713860	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:28.714757	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:29.714771	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:30.714484	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:31.715358	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:32.716275	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:33.716123	                      0.000008263                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:34.717000	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:35.716854	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:36.717731	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:37.717602	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:38.718299	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:39.718395	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:40.718234	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:41.718109	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:42.718043	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:43.717868	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:44.717782	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:45.717687	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:46.718498	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:47.718410	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:48.717910	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:49.717196	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:50.716999	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:51.717976	                      0.000008263                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:52.717775	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:53.717737	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:54.718503	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:55.718379	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:56.718284	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:57.718371	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:58.718558	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:59.718850	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:00.718981	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:01.719618	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:02.719497	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:03.719359	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:04.719222	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:05.719118	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:06.719000	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:07.718868	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:08.719780	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:09.719649	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:10.719473	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:11.720399	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:12.720219	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:13.720537	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:14.720105	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:15.719968	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:16.719928	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:17.719609	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:18.719466	                      0.000008241                      0.000006857                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:19.719357	                      0.000008263                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:20.719250	                      0.000008263                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:21.720139	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:22.719967	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:23.719864	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:24.719740	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:25.719641	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:26.720504	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:27.720334	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:28.808036	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:29.808223	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:30.807982	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:31.807990	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:32.807723	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:33.807576	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:34.808451	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:35.808364	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:36.808305	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:37.808311	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:38.807950	                      0.000008241                      0.000006879                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:39.807853	                      0.000008263                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:40.807789	                      0.000008263                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:41.808598	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:42.808442	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:43.807745	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:44.807218	                      0.000008241                      0.000006879                             0.01                             3.36                           BadVal                           BadVal
09/26/2013 13:58:45.807185	                      0.000008241                      0.000006901                             0.01                             3.51                           BadVal                           BadVal
09/26/2013 13:58:46.807942	                      0.000008241                      0.000006901                             0.01                             3.51                           BadVal                           BadVal
09/26/2013 13:58:47.807835	                      0.000008241                      0.000006901                             0.01                             3.51                           BadVal                           BadVal
09/26/2013 13:58:48.807734	                      0.000008241                      0.000006901                             0.01                             3.51                           BadVal                           BadVal
09/26/2013 13:58:49.807600	                      0.000008241                      0.000006901                             0.01                             3.51                           BadVal                           BadVal
09/26/2013 13:58:50.807446	                      0.000008241                      0.000006901                             0.01                             3.51                           BadVal                           BadVal
09/26/2013 13:58:51.808335	                      0.000008263                      0.000006901                             0.01                             3.36                           BadVal                           BadVal
09/26/2013 13:58:52.808409	                      0.000008241                      0.000006901                             0.01                             3.36                           BadVal                           BadVal
09/26/2013 13:58:53.808264	                      0.000008263                      0.000006901                             0.01                             3.51                           BadVal                           BadVal
09/26/2013 13:58:54.809105	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                           BadVal
09/26/2013 13:58:55.808874	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                           BadVal
09/26/2013 13:58:56.809685	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                           BadVal
09/26/2013 13:58:57.809598	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                           BadVal
09/26/2013 13:58:58.809481	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                           BadVal
09/26/2013 13:58:59.809374	                      0.000008219                      0.000006901                             0.01                             3.36                             0.01                           BadVal
09/26/2013 13:59:00.809218	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:01.809059	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:02.808942	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:03.808893	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:04.808683	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:05.808558	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:06.809426	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:07.809384	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:08.809184	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:09.809055	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:10.808936	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:11.808804	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:12.808117	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:13.807562	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:14.807437	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:15.807335	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:16.807178	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:17.807061	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:18.806936	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:19.806850	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:20.806758	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:21.807575	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:22.807434	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:23.807343	                      0.000008263                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:24.807219	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:25.808051	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:26.807928	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:27.807840	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:28.807693	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:29.807579	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:30.807486	                      0.000008263                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:31.808408	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:32.808183	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:33.808055	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:34.807923	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:35.814752	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:36.814785	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:37.857316	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:38.906460	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:39.906791	                      0.000008263                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:40.907200	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:41.907040	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:43.54114	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:44.54762	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:45.54866	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:46.55513	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:47.55541	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:48.56263	                      0.000008241                      0.000006857                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:49.56250	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:50.56068	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:51.56089	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:52.56015	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:53.56641	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:54.56536	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:55.56409	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:56.57287	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:57.57141	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:58.57008	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:59.56962	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:00.56794	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:01.57631	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:02.57507	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:03.57404	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:04.57264	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:05.57140	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:06.57187	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:07.56888	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:08.57796	                      0.000008263                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:09.57676	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:10.57505	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:11.58416	                      0.000008263                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:12.58402	                      0.000008263                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:13.58264	                      0.000008263                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:14.59001	                      0.000008263                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:15.58444	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:16.58774	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:17.58778	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:18.58509	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:19.58385	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:20.58280	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:21.59151	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:22.59007	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:23.59082	                      0.000008263                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:24.59789	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:25.60718	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:26.61742	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:27.61382	                      0.000008241                      0.000006857                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:28.61834	                      0.000008241                      0.000006857                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:29.61207	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:30.61995	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:31.62910	                      0.000008263                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:33.133357	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:34.133490	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:36.60493	                      0.000008263                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:37.60166	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:38.60002	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:39.59878	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:40.60744	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:41.60649	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:42.60501	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:43.60366	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:44.60273	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:45.60145	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:46.60000	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:47.59995	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:48.59750	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:49.59805	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:50.59523	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:51.59371	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:52.59286	                      0.000008263                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:53.59148	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:54.59093	                      0.000008263                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:55.58909	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:56.58815	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:57.59642	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:58.59530	                      0.000008263                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:59.59655	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:00.60281	                      0.000008263                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:01.60159	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:02.59986	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:03.59868	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:04.59736	                      0.000008241                      0.000006857                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:05.59636	                      0.000008241                      0.000006857                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:06.60509	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:07.60389	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:08.60242	                      0.000008241                      0.000006923                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:09.60171	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:10.61014	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:11.60867	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:12.60792	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:13.60652	                      0.000008263                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:14.60564	                      0.000008241                      0.000006857                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:15.60381	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:16.60265	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:17.59410	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:18.58988	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:19.58888	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:20.58998	                      0.000008263                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:21.58677	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:22.59507	                      0.000008241                      0.000006879                             0.01                             3.36                             0.08                             8.21
09/26/2013 14:01:23.59361	                      0.000008241                      0.000006879                             0.01                             3.36                             0.08                             8.54
09/26/2013 14:01:24.59276	                      0.000008263                      0.000006901                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:25.60197	                      0.000008241                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:26.60323	                      0.000008263                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:27.60873	                      0.000008263                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:28.60753	                      0.000008263                      0.000006879                             0.01                             3.36                             0.07                             8.54
09/26/2013 14:01:29.60689	                      0.000008241                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:30.60484	                      0.000008241                      0.000006879                             0.01                             3.36                             0.07                             8.54
09/26/2013 14:01:31.61383	                      0.000008241                      0.000006879                             0.01                             3.36                             0.07                             8.54
09/26/2013 14:01:32.61284	                      0.000008263                      0.000006879                             0.01                             3.36                             0.07                             8.54
09/26/2013 14:01:33.61108	                      0.000008263                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:34.61997	                      0.000008263                      0.000006901                             0.01                             3.36                             0.08                             8.54
09/26/2013 14:01:35.62848	                      0.000008263                      0.000006901                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:36.62787	                      0.000008241                      0.000006901                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:37.62638	                      0.000008219                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:38.62557	                      0.000008263                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:39.62356	                      0.000008263                      0.000006901                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:40.62272	                      0.000008241                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:41.62109	                      0.000008241                      0.000006879                             0.01                             3.36                             0.07                             8.54
09/26/2013 14:01:42.61973	                      0.000008241                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:43.61895	                      0.000008263                      0.000006879                             0.01                             3.36                             0.08                             8.54
09/26/2013 14:01:44.61768	                      0.000008241                      0.000006901                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:45.62636	                      0.000008219                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:46.62506	                      0.000008263                      0.000006945                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:47.62451	                      0.000008241                      0.000006879                             0.01                             3.36                             0.09                             8.54
... 7809 more lines ...
  489   Friday, October 04, 2013, 19:05 Grant MinorConditioning StationStandard OperationTM3NiO#2HP Conditioning - e-mail notes from Oct 2nd to Oct 4th

I have compiled the e-mail comments from people involved in High-Power Conditioning of TM3 w/ the new Rev 3 source tray and NiO#2 since Bevan Moss' last log on October 2nd below:

On October 2nd at 6:35pm Friedhelm Ames wrote:
TM3 heaters have been tested already yesterday in the conditioning station. Current and voltage readings were in the normal operating range.
The heaters were kept on over the night and no changes except a slow improvement in the vacuum has been seen. Today the high voltage was started. A high bias current of up to 150 micro Amps at 10 kV and an oscillation of the current with a frequency of about 3/minutes has been seen. Going higher in voltage resulted in an erratic behavior. After a visual inspection in the target hall some water leakage on top of the module was detected. it could be traced to 2 of the water connectors. They have been replaced. After this the high voltage current was down to normal values and the oscillation disappeared. With the heaters off the voltage could be raised to 22 kV, after which sparking occurred.
Earlier with the heaters on the functioning of the source has been tested. Extracted beam could be seen on the Faraday cup and the dependance on both the anode voltage and the coils could be verified.
One of the steerer plates (bottom) appears to have a short. As there is some redundancy in the steering elements in the target station optics, I don't believe this will effect the beam delivery very much.
A full test of all the source systems is planned for tomorrow including a careful attempt to rise the high voltage to a higher value. For the acceleration of 10C we will need 20.4 kV.

On October 3rd at 9:32am Bevan Moss wrote:

FYI Module is holding 24kV with zero sparks as of yet  I will continue with updates as I go

On October 3rd at 10:18am Bevan Moss wrote:

Sparking started at ~28kV I have had to dial it back to 22kV to hold stable voltage. I will continue with updates.

On October 3rd at 11:17am Bevan Moss wrote:

Sparking has forced the voltage down to 20kV. Friedhelm has asked me to turn everything on and start ramping the heaters.

On October 3rd at 4:15pm Bevan Moss wrote:

The heaters have been brought online and the Bias raised to 25kV. It has been sitting there for the last 10 minutes with 1 spark during this time. All other PS are on and beam is being extracted.

On October 3rd at 4:37pm Bevan Moss wrote:

Sparking has just occurred and the Vacuum is now getting worse significantly faster. Voltage was lowered to 21KV where it has been stable for the last 5 minutes. I am going to turn off the HV leave the module overnight with the heaters on. Hopefully we will have good Vacuum again in the morning.

On October 4th at 2:20pm Bevan Moss wrote:

I have finished completing another complete leak test on the module and it is leak tight. Once the heaters were turned off the vacuum did improve it is currently sitting at 9.3 x 10^-7 Torr in the service cap, 5.71 x 10^-6 Torr in the diagnostics box and 4.0 x 10^-6 in in the secondary.  I have raise the bias voltage and it can hold 20kV for 5 minutes but did have 1 spark. At 22.5kV it appears to be doing some conditioning as the current is unstable. I would not recommend moving TM3 to ITE on Monday until we can establish that the module can hold 25kV for at least 1 hour preferably with the heaters on. I have left the turbo pumps running in case there is more testing you would like to do. If you choose to move to ITE on Monday please have someone stop the turbo pumps otherwise I must wait an hour before venting on Monday. The TCS laptop on my desk should not have the screen saver come on so TCS can be operated through it.

  493   Sunday, October 06, 2013, 15:39 Friedhelm AmesConditioning StationStandard OperationTM3NiO#2HEATERS STARTED

The high voltage was sparking several times during the night. It may be correlated with the cycling of the cooling water temperature and conductance.

vacuum levels at 11:00

IG1 7.5e-7 T

IG2 3.9e-6 T

IG1s 2.0 e-6 T

heaters have been ramped up to TGH 580 A, TBHT 250 A,

anode 100 V 5.3 mA, coil 50 A Faraday cup 50 nA bias 20 kV EL 6 kV

vacuum

IG1 7.9e-7

IG2 4.5e-6

IG1s 2.2e-6

  2560   Monday, April 08, 2024, 09:29 David WangCoolingStandard Operation  HALCW system general inspection.

water are supplied to both ITE and ITW. I did a general inspection on HALCW system. Two small leaks on 2" shut off valves are fixed. No other leak has been found so far. all water flow meters work fine. MP2 pump discharging pressure is ok which is just above 100 psi on local gauge reading. pump noise are low. pump body temperature is less than 45C which is good. 

  2495   Friday, October 13, 2023, 11:46 David WangCoolingMaintenance  HALCW system O2 sensor service and calibration.

HALCW system O2 sensor service and calibration were done in Sep 26th. electrolyte and membrane on rose mount sensor 499ADO were refilled and replaced. Sensor was zero and full scale calibrated. The sensor works fine after the job.

  1444   Monday, March 06, 2017, 08:01 David WangCoolingRepair  HALCW resin tank replacement.

HALCW resin tank has been replaced. The old tank is stored in middle-west silo now. The field on old resin tank bottom is about 3.5 msv/ hr(on contact). By shielding it into silo and putting two lead blankets on the top cover of the silo,The general field 1 feet above this silo is 25usv/hr. And, The field around the silo area are less affected. This old tank will stay there as long as possible.  For the new resin tank during preparation, It was found that on return line side, the 1.25" pipe thread plug with return line tube  was very difficult to be installed onto the female thread adapter which is welded onto to tank body. It was jammed in half way of threading onto tank at manufacturer. This plug was removed out first and inspected. The first and second male threads were galled.  Maico repaired the thread on a lathe. The female thread on the tank was tapped. A tight spot was found during the tap. After that, the return line plug could be installed onto the tank properly.  The new tank was installed into the vessel and connected back to HALCW system. Water are refilled into new tank. During refill air were trapped into the system. This caused  pump restarting problem. By Several times of starting up the pump, stopping the pump and venting,  Most air were breathed out of  system. Finally,The system was restarted by supply water to ITW( without module) only. ITE and resin can circuits were brought back to system after that without problem.

 Before removing the old tank, We drained water out of old tank to decay tank. Since the resin tank is largely filled with  resin, and water are soaked into resin, we only can drain out less than 10 liters water out of tank. Also, The drain caused some part of return line on pumps have to been drained. This is the main reason air was trapped in system during the new resin tank water refilling. For the future replacement, we should not drain water in old tank. It does not worth to do it. 

Attach TH HALCW resin tank replacement procedure here.

2022-02-17 reload replacement procedure and draining schematic.

Attachment 1: new_resin_tank_installation_004.JPG
new_resin_tank_installation_004.JPG
Attachment 2: new_resin_tank_installation_005.JPG
new_resin_tank_installation_005.JPG
Attachment 3: new_resin_tank_installation_006.JPG
new_resin_tank_installation_006.JPG
Attachment 4: new_resin_tank_installation_008.JPG
new_resin_tank_installation_008.JPG
Attachment 5: new_resin_tank_installation_010.JPG
new_resin_tank_installation_010.JPG
Attachment 6: new_resin_tank_installation_012.JPG
new_resin_tank_installation_012.JPG
Attachment 7: new_resin_tank_installation_013.JPG
new_resin_tank_installation_013.JPG
Attachment 8: HALCW_Resin_Tank_Replacement_Procedure..docx
Attachment 9: HALCW_Resin_Tank_Replacement_Procedure..pdf
HALCW_Resin_Tank_Replacement_Procedure..pdf HALCW_Resin_Tank_Replacement_Procedure..pdf HALCW_Resin_Tank_Replacement_Procedure..pdf
Attachment 10: HALCW_resin_can_replacement__procedure..pdf
HALCW_resin_can_replacement__procedure..pdf
Attachment 11: resin_tank_draining.pdf
resin_tank_draining.pdf
  1164   Monday, March 14, 2016, 11:24 David WangITE/ITW CoolingMaintenance  HALCW resin tank filter cartridges are changed today

Two of resin tank filter cartridges(supply and return) are changed today. Old cartridges and bagged and tagged. The contact field on them are 650usv/hr. The new cartridges are fiber-glass, Triumf store item. After replacement, the field on contact with the filter case are about 10 usv/hr. But the field on contact with top resin tank  has been increased from 45usv/hr  to 225 usv/hr

 The flow rate is relatively stable before and after change. See attachment

Attachment 1: 002.JPG
002.JPG
Attachment 2: 003.JPG
003.JPG
Attachment 3: 006.JPG
006.JPG
  2781   Monday, March 03, 2025, 15:33 David WangCoolingMaintenance  HALCW resin can loop two filters have been replaced.

HALCW resin can loop two filters have been replaced. no water leak on after job check.

  1693   Wednesday, March 07, 2018, 10:36 David WangITE/ITW CoolingDevelopment  HALCW resin can flow meter replacment

HALCW resin can flow meter replacement is done. Less that 100 mL water was drained to a small tray during the job. RPG checked the water. Less than 20 counts. Water are dumped the high active sump.   The HALCW MP1 is started. No leak on new gauges and fittings. Old gauge is labeled , bagged, and stored in TH.

  2763   Monday, February 03, 2025, 13:10 David WangCoolingStandard Operation  HALCW pump ITC:MP1 has been started.

HALCW pump ITC:MP1 has been started. water are supplied to both ITE and ITW system. all water signals for individual circuits work fine. no major leak on system. Some small water seeping from small ball valve stems are found in ITW. they will be fixed soon.

  2776   Friday, February 21, 2025, 09:18 David WangCoolingStandard Operation  HALCW pump ITC:MP1 and MP2 phase current draw measurement.

HALCW pump ITC:MP1 and MP2 phase current draw measurement result: water supplied to both station. MP1: Black 17.0A, blue 16.9A, red 16.7A.   MP2: black 17.3A, blue17.8A, red 17.3A.    The results are good.  No concern on both motors.

ELOG V2.9.2-2455