  481   Wed Aug 13 15:04:05 2014 Raymond Dubé  OLIS Bender Magnet PS (0-500 amps)Arthur
and Derek had a look at the OLIS Bender Magnet
P/S and found a problem with the Polarity
  
  482   Wed Aug 13 15:09:42 2014 Raymond DubéRoutineGeneralOLIS Attenuators #2 fixed, #1 gone!Today
Probe group (Scott K.) installed the OLIS
Attenuator, they fixed the vacuum leak on
  
  483   Tue Sep 16 11:55:28 2014 Raymond DubéRoutineGeneralMCIS infoI found
these files about the MCIS in Geoff folders
in M:\ISIS\Geoff-110715\OLIS\SuperNanogan
 Cart-Power1.pdfSuperNanogan_Control-R1.pdfSupernanogan_Interlocks-R1.pdf 
  484   Fri Oct 10 14:02:56 2014 Raymond DubeRoutineGeneralTest with new Alga, ALPA 50W RF AMPWe did a test yesterday with the new Alga
Microwave RF AMP, model # ALPA 50 Watt, serial
# 130684 to be used for the MCIS.  We
 ALPA-RF_AMP-50W.pdf 
  485   Tue Oct 14 08:31:53 2014 Brian MinatoOtherGeneralWater Flow Switch #3This entry was incorrectly added to the
OLIS logbook.  it has been moved to
the I3 logbook.
  
  486   Tue Oct 21 07:14:26 2014 Brian MinatoOtherGeneralOLIS WorkMCIS-
the cart was moved into the HV cage. 
The facilities will be connected and an oven
  
  487   Wed Oct 22 07:58:32 2014 Brian MinatoOtherGeneralMCIS WorkThe remaining connections were done on
the MCIS. 
A piece of 1/8" teflon was installed
  
  488   Thu Oct 23 10:52:21 2014 Brian MinatoRoutineGeneralALGA AmplifierThe ALGA amplifier and its signal generator
have been turned off at the request of Controls. 
They have had communication issues and are
  
  489   Fri Oct 24 10:50:20 2014 Brian MinatoOtherOtherOLIS UpdateThe nitrogen bottle needs to be replaced.
 
MCIS-
  
  490   Mon Oct 27 15:13:18 2014 Brian Minato   MCIS-
cleaning out source to remove potassium
and calcium.
  
  491   Tue Oct 28 13:30:47 2014 Brian MinatoOtherOtherMCIS UpdateMCIS-
the cleaning has been completed. 
The oven assembly is removed and the gate
  
  492   Thu Oct 30 13:14:51 2014 Brian MinatoOtherOtherOLIS UpdateMCIS-
a new oven assembly was installed
on the MCIS yesterday.  The cart has
  
  493   Thu Oct 30 22:09:02 2014 Keerthi Jayamanna  Suoernanogan steeringIOS mass scan does not work.
A/Q        
   FC6     
  
  494   Thu Oct 30 22:44:45 2014 Keerthi JayamannaRoutineGeneralMass scan 3 to 4IOS mass scan if faulty. These is a manual
scan from mass 3 to 4
Const22 = 0.04187  adjusted
  
  495   Thu Oct 30 23:37:58 2014 Keerthi JayamannaRoutineGeneralMass scan 4 to 6A/Q           
FC6
4               
  
  496   Thu Oct 30 23:44:57 2014 Keerthi JayamannaRoutineGeneralOLIS Supernanogan mass scan 6 to 7.5A/q          
FC6  (only current over 100pA)
6.0          
  
  497   Fri Oct 31 20:39:56 2014 Brian MinatoOtherOtherOLIS UpdateMCIS-
the MWS Bias and PE have been moved
on to the MCIS (the MCIS connections have
  
  498   Sat Nov 1 13:31:17 2014 Keerthi Jayamanna  MCIS readyness for 76SeSupernanogan is running with oven no heat,
without 76selenium. MCIS:IG1=3e-7Torr.
Tune saved as 141101_1318_76Se+11tune.snap
  
  499   Thu Nov 6 08:37:52 2014 Brian MinatoOtherOtherOLIS UpdateMCIS-
the source is off.  It is under
vacuum and in the HV cage.
  
  500   Wed Nov 12 13:32:47 2014 Raymond DubeRoutineGeneralNew in-line resistors for IOS:PEToday we replace the in-line resistors
for the Plasma Electrode (PE) it was a 25
ohm 200W (2 x 50 Ohm 100W) with a 50 Ohm
 ISK4583D-rev10.pdf 