  582   Friday, May 05, 2017, 14:28 Aaron SchmidtUC#20LP-IG-LIS     Conditioning Heat shield for UC#20 (2)

New O-rings installed on FS#3.

Friday, May 05, 2017, 14:24: Roughing.

Friday, May 05, 2017, 14:34: Waterlines reconnected to air lines. He likely still in lines. May show in RGA if there's an issue prior to opening the waterlines.

Friday, May 05, 2017, 14:47: Turbos started.

Friday, May 05, 2017, 17:59: Water leak still present on FS#3


Monday, May 08, 2017, 08:54: Line re-route on FS#3. Return line on (F) re-routed to Return line on (K). (EE)

Monday, May 08, 2017, 09:00: Open water into line FS#3 wait 10 minutes. Vacuum @ 4.52e-07T.

Monday, May 08, 2017, 09:12: Vacuum @ 4.52e-07T. Lines purged. No response on Vacuum or RGA.

Monday, May 08, 2017, 09:20: Line re-route on FS#3 through (K) supply to (G) & return on (E)

Monday, May 08, 2017, 09:28: Vacuum increased to 6.41e-07T. Response noticed on RGA. FS#3 purged & closed.

Monday, May 08, 2017, 09:30: Slow Vacuum increase to 6.67e-07T & climbing.

Monday, May 08, 2017, 10:13: Vacuum has decreased.

Monday, May 08, 2017, 10:26: Line re-route on FS#3 through (K) supply to (H) & return on (F).

Monday, May 08, 2017, 10:20: Open water into line FS#3 wait 10 minutes. Vacuum @ 4.69e-07T.

Monday, May 08, 2017, 10:30: Vacuum @ 4.60e-07T. Lines purged. No response on Vacuum or RGA.

Monday, May 08, 2017, 10:39: Spike on RGA while tightening (G)

Monday, May 08, 2017, 12:02: RGA off. Ion gauge & turbos off

Monday, May 08, 2017, 14:14: Vented. Lines adapted to leak detector on line (G) of FS#3. Line (E) blanked off. Check for leaks.

Monday, May 08, 2017, 14:49: Roughing.

Further purging of the lines required. Leak detector unable to achieve vacuum in the lines. Adapters removed. Line re-route on FS#3 through (K) supply to (G) & return on (E) for purge.


Tuesday, May 09, 2017, 07:30: Venting & Purging. Purge for a minimum of 20 mins.

Attach adapters. Hook up leak detector

Tuesday, May 09, 2017, 09:24: Leak found @ (G) line block

The (G) block line was disconnected and 'Blanked' off. The (G) line, block to feed through, was re-tested. No leak was found on blank off.

O-ring area cleaned. O-ring replaced. Blank removed. Block to Coil end reconnected. Leak highest when He sprayed directly on block.

Masking tape and/or 'putty' added block to block & confirm solder joint ok.

(E)(G) line removed. Block to block leak/sealing issue confirmed. (G) block leaks on the seat. (E) block ok. Tested externally. Achieves 1.0e-09 leak rate.


Wednesday, May 10, 2017, 09:22: Test stand (G) line from block to feed through has a separate leak. Cannot achieve below e-04 leak rate. Does not fully seal.


Removed heat shield to condition in evaporator 2.

Attachment 1: UC#20_Heat_shield_conditioning-2.docx
Attachment 2: TS-2017-05-05-17-11.rga
Attachment 3: TS-2017-05-11-07-50.rga
  583   Wednesday, May 10, 2017, 15:07 Aaron Schmidt       Pump down empty chamber. (FS#3 out)

FS#3 lines blanked off.

Wednesday, May 10, 2017, 14:40: Roughing

Wednesday, May 10, 2017, 15:05: Turbos on.

Wednesday, May 10, 2017, 15:10: Ion gauge on.


Disconnect Leak detector. Connect He to FS#3 (G) line.

Thursday, May 11, 2017, 07:44: RGA on.

Thursday, May 11, 2017, 08:23: Vacuum @ 1.85e-06T. He line pressurized to 50 PSI. Monitor RGA.

Thursday, May 11, 2017, 09:07: No response on RGA. Vacuum @ 1.85e-06T. Purge Helium.

Hook up dry system leak detector to leak check valve behind turbo.

Leak rate: 0.0e-09 atm-cc/sec on detector. T.P: 0.0e-04T.

Thursday, May 11, 2017, 09:18: Open check valve. Vacuum ok @ 1.76e-06T

Base leak rate: 2.5e-09 atm-cc/sec.

Thursday, May 11, 2017, 09:25: Open He & Pressurize to 50 PSI.

Thursday, May 11, 2017, 09:30: Base leak rate dropping slowly: 2.0e-09 atm-cc/sec. No leak detected. No response on RGA.

Thursday, May 11, 2017, 09:35: Base leak rate dropping slowly: 1.7e-09 atm-cc/sec. No leak detected. No response on RGA.

Thursday, May 11, 2017, 09:45: Base leak rate Ok: 1.7e-09 atm-cc/sec. No leak detected. No response on RGA. Close check valve. Purge line. Vent detector.

Leak detector on. Leak rate: 0.0e-09 atm-cc/sec on detector. T.P: 0.0e-04T.

Thursday, May 11, 2017, 13:24: Open check valve. Vacuum @ 1.41e-06T

Base leak rate: @.2.0e-09 atm-cc/sec.

Thursday, May 11, 2017, 13:30: Open He & Pressurize to 50 PSI.

Thursday, May 11, 2017, 13:37: Base leak rate dropping slowly: 1.0e-09 atm-cc/sec. No leak detected. No response on RGA.

Thursday, May 11, 2017, 13:40: Base leak rate Ok: 0.1e-09 atm-cc/sec. No leak detected. No response on RGA. Vent He.

Connect spray nozzle to He. Vacuum issue noticed on RGA: N & O. Check for leak on Chamber


Tuesday, May 16, 2017, 08:45: RGA off. Ion gauge off. Turbos off

Tuesday, May 16, 2017, 09:50: Venting

Attachment 1: 05-11-2017_Water_leak.docx
  584   Tuesday, May 16, 2017, 14:25 Aaron Schmidt HP-FEBIAD-CTL     Tri-Coil CTL-FEBIAD test III

Coil loose (Remove lid, Tighten and place lid again)

Tuesday, May 16, 2017, 14:24: Roughing.

Tuesday, May 16, 2017, 14:45: Turbo pumps on.


Wednesday, May 17, 2017, 06:53: RGA on. Vacuum @ 7.40e-07T

Wednesday, May 17, 2017, 08:04: Open waterline FS#5. OK. Vacuum @ 6.93e-07T

Wednesday, May 17, 2017, 08:05: Open waterline FS#4. OK.

Wednesday, May 17, 2017, 08:06: Open waterline FS#2. OK.

Wednesday, May 17, 2017, 08:07: Open waterline FS#3. OK.

Wednesday, May 17, 2017, 08:10: Vacuum @ 6.93e-07T. No response on RGA

Wednesday, May 17, 2017, 08:26: RGA Error out. Restart RGA.

Wednesday, May 17, 2017, 08:35: Vacuum ok @ 6.87e-07T. RGA ok.


Sparking issues


Wednesday, May 17, 2017, 13:00:

BIAS / EZL / EE @ 35 / 13 / 1.5 kV .

Coil @ 50 A . Anode @ 200 V

RGA Fatal errors. unable to restart. Reboot in morning.

Wednesday, May 17, 2017, 14:25: Auto ramping TGHT / TBHT to 300 / 150 A @ 2/1 A per min.

Wednesday, May 17, 2017, 15:05: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .


Thursday, May 18, 2017, 08:27: Auto ramping TGHT / TBHT to 400 / 200 A @ 2/1 A per min.

Thursday, May 18, 2017, 08:33: RGA on.

Thursday, May 18, 2017, 08:37: IV4 opened

Thursday, May 18, 2017, 08:42: Optics on/set


Monday, May 22, 2017, 07:55: RGA, Power/Sources off.

Monday, May 22, 2017, 07:59: Turbo pumps, Ion gauge off.

Monday, May 22, 2017, 09:03: Venting

Monday, May 22, 2017, 09:40: 6 block screws tightened.

Attachment 1: CTL_FEBIAD_Test_target_III.docx
Attachment 2: TS-2017-05-17-13-00.rga
Attachment 3: TS-2017-05-17-13-30.rga
Attachment 4: TS-2017-05-18-08-30.rga
Attachment 5: CTL_FEBIAD_Test_target_II_(Spark_out).docx
Attachment 6: TS-2017-05-17-06-55.rga
Attachment 7: TS-2017-05-17-08-30.rga
Attachment 8: TS-2017-05-17-12-33.rga
  585   Wednesday, May 17, 2017, 10:17 Anders Mjos HP-FEBIAD-CTL    Problems/SolutionsSparking between new coil line and ground electrode

Wednesday, May 17, 2017, 10:17: Sparking around 20 kV. Will need to vent to increase gap between water lines.

Turbos stopped.

Wednesday, May 17, 2017, 10:50: Reset IV4.

Wednesday, May 17, 2017, 10:59: Venting.

Wednesday, May 17, 2017, 11:42: Increased clearance to GND electrode.

Wednesday, May 17, 2017, 11:42: Roughing

Wednesday, May 17, 2017, 12:01: Turbos started

 

  586   Monday, May 22, 2017, 09:43 Aaron Schmidt HP-FEBIAD-CTL     Tri-Coil CTL-FEBIAD test IV

Monday, May 22, 2017, 09:41: Roughing.

Monday, May 22, 2017, 10:00: Turbo pumps started.


Tuesday, May 23, 2017, 07:45: Power/sources on.

Tuesday, May 23, 2017, 07:58: BIAS / EZL / EE @ 35 / 13 / 1.5 kV .

Tuesday, May 23, 2017, 08:19: Coil @ 50 A . Anode @ 200 V

Tuesday, May 23, 2017, 08:20: Auto ramping TGHT / TBHT to 580 / 270 A @ 2 A per min.

Tuesday, May 23, 2017, 10:22: IV4 opened.

Tuesday, May 23, 2017, 10:57: Increased ramping rate for TGHT to 4 A/min - AM

Tuesday, May 23, 2017, 19:05: Cooling down target for change to Anode bias tomorrow. - AM

Tuesday, May 23, 2017, 21:04: BIAS/EZL/EE off


Wednesday, May 24, 2017, 07:16: IV4 Closed.

Wednesday, May 24, 2017, 07:18: Power/Sources off.

Changed to Anode Bias.

Wednesday, May 24, 2017, 11:00: Coil @ 50 A . Anode @ 200 V

Wednesday, May 24, 2017, 11:02: Auto ramping TGHT / TBHT to 580 / 270 A @ 4 / 2 A per min.

Wednesday, May 24, 2017, 11:09: BIAS / EZL / EE @ 35 / 13 / 1.5 kV .


Tuesday, May 30, 2017, 07:02: Auto Cooling @ 2A/min


Thursday, June 01, 2017, 14:08: IV4 Closed. RGA off.

Thursday, June 01, 2017, 14:10: Ion gauge off. Turbos off.

Thursday, June 01, 2017, 16:51: Venting

Thursday, June 01, 2017, 21:16: Roughing.

Attachment 1: CTL_FEBIAD_Test_IV.docx
Attachment 2: TS-2017-05-22-10-07.rga
Attachment 3: TS-2017-05-25-15-42.rga
  587   Wednesday, May 24, 2017, 11:12 Anders Mjos      Problems/SolutionsFEBIAD Anode bias configuration

 The Test Stand has been configured for Anode bias. John Chow changed the power supply and the electrical connections have been wired according to ISK4523D REV 6.

 

Wednesday, May 24, 2017, 17:37: Anode current reduce to 285 A. Anode drain current limited  to a maximum of 300 A. This will be left overnight to see if there is an increase in anode drain current. as was seen today.

Attachment 1: 20170524-ST-currentvscoil-CHT300A.PNG
20170524-ST-currentvscoil-CHT300A.PNG
  589   Saturday, June 03, 2017, 09:34 Anders Mjos      Problems/SolutionsFEBIAD Cathode bias configuration

Anders Mjos wrote:

 The Test Stand has been configured for Anode bias. John Chow changed the power supply and the electrical connections have been wired according to ISK4523D REV 6.

 

Wednesday, May 24, 2017, 17:37: Anode current reduce to 285 A. Anode drain current limited  to a maximum of 300 A. This will be left overnight to see if there is an increase in anode drain current. as was seen today.

 The Test Stand has been  configured for Cathode bias. John Chow changed the power supply and the electrical connections have been wired according to ISK4563 REV 5.

  590   Tuesday, June 06, 2017, 09:22 Aaron Schmidt       Pump down empty chamber

Tuesday, June 06, 2017, 09:21: roughing.


Friday, June 09, 2017, 08:50: Ion gauge off. Turbo pumps off.

Friday, June 09, 2017, 10:46: Venting.

  591   Wednesday, June 07, 2017, 12:49 Aaron Schmidt       Gas bottles

 Gas bottles for FEBIAD testing were exchanged yesterday around 1:30. One Neon & one Argon bottle were removed.

Two Neon bottles were added.

  593   Friday, June 16, 2017, 11:23 Aaron Schmidt       Pump down empty chamber.

Friday, June 16, 2017, 11:22: roughing.

Friday, June 16, 2017, 14:28: Turbo pumps on


Friday, June 30, 2017, 10:37: Ion gauge, turbos off.

Friday, June 30, 2017, 13:01: Venting.

  595   Thursday, July 06, 2017, 13:16 Aaron SchmidtTa#55LP-SIS  42218-1142738-2In progressConditioning Ta spare for Ta#52

Thursday, July 06, 2017, 13:56: Roughing.

Thursday, July 06, 2017, 14:25: Turbo  pumps started.

Thursday, July 06, 2017, 14:30: Ion gauge on. RGA Started


Friday, July 07, 2017, 07:29: Waterlines opened. No response on RGA. Vacuum ok @ 1.97e-07T

Friday, July 07, 2017, 07:33: Power/sources on.

Friday, July 07, 2017, 07:39: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .

Friday, July 07, 2017, 07:40: IV4 Opened.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 26.0 3.78
 TBHT 26.5 3.02
 COIL + EE 25.8 4.86
 TGHT  26.0 5.47

Friday, July 07, 2017, 07:42: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.

Friday, July 07, 2017, 07:45: Optics reset.

Friday, July 07, 2017, 14:28: Anode brought up to 5, 10 15, and 50 V using FEBIAD controls. No Current detected.

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 36.5 3.75
 TBHT 27.6 3.04
 COIL + EE 26.5 4.87
 TGHT  28.7 5.54

 


Saturday, July 08, 2017, 08:58: Ramping down heaters


Monday, July 10, 2017, 08:15: Power/Sources off. *BIAS 30 to 0 kV due to shut off. EE & EZL Brought to 0.

Monday, July 10, 2017, 08:23: Water lines purged. IV4 Closed

Monday, July 10, 2017, 08:27: IV4 reset.

Monday, July 10, 2017, 09:40: RGA off. Ion Gauge off. Turbos off.

Monday, July 10, 2017, 11:02: Venting.

Attachment 1: ER-580-200-30kV-m23.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1459 -0.0992 -0.0943 -0.0925 -0.0934 -0.1016 -0.0986 -0.0839 -0.0732 -0.0800 -0.0766 -0.0739 -0.0772 -0.0876 -0.0931 -0.0876 -0.0989 -0.0861 -0.0934 -0.0989 -0.0958
 -0.0903 -0.0943 -0.0906 -0.0906 -0.0879 -0.0986 -0.0931 -0.0797 -0.0793 -0.0821 -0.0723 -0.0693 -0.0726 -0.0848 -0.0888 -0.0818 -0.0867 -0.0803 -0.0882 -0.0946 -0.0897
 -0.0897 -0.0925 -0.0867 -0.0821 -0.0854 -0.0977 -0.0916 -0.0793 -0.0842 -0.0882 -0.0720 -0.0687 -0.0751 -0.0821 -0.0879 -0.0839 -0.0861 -0.0775 -0.0793 -0.0815 -0.0797
 -0.0882 -0.0903 -0.0793 -0.0937 -0.0769 -0.0977 -0.0949 -0.0806 -0.0778 -0.0858 -0.0751 -0.0729 -0.0867 -0.0748 -0.0864 -0.0867 -0.0760 -0.0781 -0.0800 -0.0784 -0.0772
 -0.0876 -0.0879 -0.0787 -0.0958 -0.0787 -0.0986 -0.0952 -0.0812 -0.0821 -0.0848 -0.0797 -0.0787 -0.0723 -0.0787 -0.0851 -0.0870 -0.0790 -0.0787 -0.0787 -0.0775 -0.0781
 -0.0858 -0.0888 -0.0781 -0.0748 -0.0720 -0.0970 -0.0961 -0.0809 -0.0897 -0.0845 -0.0793 -0.0787 -0.0769 -0.0778 -0.0839 -0.0836 -0.0757 -0.0775 -0.0769 -0.0778 -0.0778
 -0.0885 -0.0909 -0.0769 -0.0912 0.0586 -0.0385 -0.0851 -0.0790 -0.0778 -0.0870 -0.0790 -0.0790 -0.0864 -0.0766 -0.0851 -0.0812 -0.0757 -0.0760 -0.0787 -0.0751 -0.0781
 -0.0888 -0.0912 -0.0775 -0.0833 -0.0034 0.2264 0.0839 -0.0494 -0.0876 -0.0891 -0.0797 -0.0790 -0.0778 -0.0797 -0.0851 -0.0830 -0.0790 -0.0784 -0.0790 -0.0766 -0.0769
 -0.0717 -0.0797 -0.0766 -0.0916 -0.0726 0.3128 0.6992 0.1498 -0.0574 -0.0739 -0.0732 -0.0790 -0.0790 -0.0760 -0.0848 -0.0836 -0.0787 -0.0800 -0.0775 -0.0769 -0.0778
 -0.0720 -0.0684 -0.0549 -0.0714 -0.0342 -0.0745 1.2405 0.6815 0.1675 -0.0424 -0.0638 -0.0699 -0.0739 -0.0790 -0.0854 -0.0821 -0.0803 -0.0781 -0.0803 -0.0775 -0.0797
 -0.0623 -0.0571 -0.0558 -0.0565 -0.0299 -0.0323 0.2798 0.8789 1.0532 0.0723 -0.0464 -0.0595 -0.0674 -0.0699 -0.0769 -0.0815 -0.0793 -0.0772 -0.0800 -0.0781 -0.0775
 -0.0879 -0.0882 -0.0211 -0.0644 -0.0330 -0.0381 0.0974 0.2484 1.1102 0.7678 0.0488 -0.0729 -0.0705 -0.0690 -0.0763 -0.0720 -0.0647 -0.0754 -0.0760 -0.0751 -0.0766
 -0.0894 -0.0903 -0.0778 -0.0952 -0.0354 -0.0589 0.0488 0.0494 0.2368 1.1792 0.4474 0.0125 -0.0751 -0.0732 -0.0827 -0.0790 -0.0705 -0.0629 -0.0745 -0.0763 -0.0772
 -0.0879 -0.0903 -0.0787 -0.0970 -0.0760 -0.0934 -0.0165 0.0262 0.0156 0.3046 0.8878 0.2240 -0.0699 -0.0729 -0.0858 -0.0818 -0.0763 -0.0717 -0.0699 -0.0546 -0.0769
 -0.0870 -0.0900 -0.0766 -0.0888 -0.0787 -0.0961 -0.0925 -0.0714 -0.0507 0.0174 0.2612 0.5963 0.1871 -0.0787 -0.0854 -0.0815 -0.0778 -0.0787 -0.0742 -0.0684 -0.0671
 -0.0888 -0.0897 -0.0760 -0.0909 -0.0757 -0.0961 -0.0912 -0.0745 -0.0891 -0.0720 0.0137 0.1895 0.3024 -0.0314 -0.0827 -0.0818 -0.0766 -0.0787 -0.0781 -0.0790 -0.0693
 -0.0879 -0.0897 -0.0766 -0.0900 -0.0754 -0.0919 -0.0885 -0.0754 -0.0790 -0.0827 -0.0723 -0.0687 -0.0305 -0.0381 -0.0821 -0.0806 -0.0766 -0.0757 -0.0763 -0.0742 -0.0769
 -0.0885 -0.0888 -0.0766 -0.0897 -0.0745 -0.0879 -0.0873 -0.0757 -0.0864 -0.0827 -0.0775 -0.0769 -0.0751 -0.0766 -0.0842 -0.0836 -0.0766 -0.0757 -0.0778 -0.0763 -0.0757
 -0.0900 -0.0909 -0.0784 -0.0970 -0.0751 -0.0870 -0.0858 -0.0751 -0.0912 -0.0806 -0.0790 -0.0793 -0.0787 -0.0769 -0.0848 -0.0833 -0.0778 -0.0781 -0.0769 -0.0760 -0.0781
 -0.0864 -0.0900 -0.0757 -0.0961 -0.0757 -0.0864 -0.0809 -0.0696 -0.0751 -0.0821 -0.0772 -0.0793 -0.0787 -0.0760 -0.0836 -0.0818 -0.0790 -0.0778 -0.0787 -0.0778 -0.0772
 -0.0858 -0.0894 -0.0748 -0.0934 -0.0739 -0.0848 -0.0787 -0.0629 -0.0809 -0.0778 -0.0769 -0.0775 -0.0775 -0.0772 -0.0818 -0.0815 -0.0763 -0.0763 -0.0769 -0.0772 -0.0781
Attachment 2: ER-580-200-30kV-m38.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1575 -0.1071 -0.1016 -0.1065 -0.1062 -0.1035 -0.1004 -0.0903 -0.0760 -0.0363 -0.0461 -0.0601 -0.0778 -0.0818 -0.0949 -0.0995 -0.0931 -0.0916 -0.1071 -0.1041 -0.0958
 -0.0970 -0.1013 -0.0980 -0.1010 -0.1016 -0.1001 -0.0970 -0.0812 -0.0861 -0.0699 -0.0449 -0.0555 -0.0717 -0.0787 -0.0937 -0.0961 -0.0854 -0.0867 -0.1013 -0.0983 -0.0897
 -0.0952 -0.0986 -0.0851 -0.0897 -0.0955 -0.0977 -0.0983 -0.0803 -0.0836 -0.1077 -0.0610 -0.0546 -0.0717 -0.0772 -0.0894 -0.0958 -0.0864 -0.0821 -0.0961 -0.0861 -0.0809
 -0.0964 -0.0977 -0.0775 -0.0858 -0.0739 -0.0974 -0.0989 -0.0818 -0.0787 -0.1105 -0.0882 -0.0601 -0.0775 -0.0739 -0.0882 -0.0958 -0.0861 -0.0815 -0.0806 -0.0772 -0.0778
 -0.1007 -0.1004 -0.0800 -0.0861 -0.0845 -0.0989 -0.0992 -0.0848 -0.0861 -0.1050 -0.1031 -0.0732 -0.0757 -0.0751 -0.0867 -0.0967 -0.0854 -0.0815 -0.0821 -0.0797 -0.0787
 -0.0995 -0.0986 -0.0790 -0.0839 -0.0821 -0.0989 -0.0989 -0.0854 -0.0864 -0.1059 -0.0977 -0.0821 -0.0803 -0.0793 -0.0894 -0.0970 -0.0848 -0.0806 -0.0797 -0.0797 -0.0778
 -0.0964 -0.0974 -0.0735 -0.0873 -0.0375 -0.0601 -0.0922 -0.0870 -0.0867 -0.1074 -0.0952 -0.0809 -0.0812 -0.0803 -0.0894 -0.0992 -0.0845 -0.0809 -0.0809 -0.0757 -0.0772
 -0.0961 -0.0955 -0.0742 -0.0452 -0.0705 0.0897 0.0931 -0.0305 -0.0882 -0.1099 -0.0922 -0.0793 -0.0797 -0.0763 -0.0894 -0.0989 -0.0818 -0.0809 -0.0809 -0.0760 -0.0769
 -0.0134 -0.0549 -0.0696 -0.0800 -0.0336 0.0757 0.4651 0.6528 0.0015 -0.0769 -0.0827 -0.0784 -0.0775 -0.0742 -0.0861 -0.0989 -0.0858 -0.0845 -0.0821 -0.0821 -0.0775
 0.0378 0.0250 0.0470 0.0342 0.1025 -0.0128 0.2185 1.8475 1.6077 -0.0134 -0.0391 -0.0351 -0.0470 -0.0769 -0.0882 -0.0961 -0.0867 -0.0827 -0.0806 -0.0784 -0.0781
 -0.0510 0.0760 0.0638 0.0986 0.2078 0.3448 0.0732 0.6653 3.9651 1.0074 0.0226 -0.0113 -0.0189 -0.0146 -0.0632 -0.0958 -0.0836 -0.0821 -0.0809 -0.0766 -0.0797
 -0.0961 -0.0946 0.0452 0.0372 0.1334 0.1855 0.2502 0.3159 2.5229 6.0635 0.1724 -0.0327 -0.0235 -0.0327 -0.0494 -0.0406 -0.0385 -0.0790 -0.0748 -0.0751 -0.0775
 -0.0958 -0.0964 -0.0726 -0.0839 0.0092 0.0116 0.1544 0.2350 0.5399 4.1684 3.1567 0.1071 -0.0070 -0.0229 -0.0632 -0.0671 -0.0507 -0.0302 -0.0647 -0.0745 -0.0760
 -0.0974 -0.0980 -0.0784 -0.0873 -0.0800 -0.0888 0.0009 0.1086 0.1468 0.9015 3.8147 1.2885 0.1019 -0.0085 -0.0812 -0.0833 -0.0699 -0.0494 -0.0470 0.0104 -0.0754
 -0.0986 -0.0970 -0.0797 -0.0888 -0.0836 -0.0958 -0.0937 -0.0748 -0.0476 0.0757 0.7455 1.4111 0.2850 -0.0409 -0.0198 -0.0919 -0.0806 -0.0729 -0.0549 -0.0415 -0.0406
 -0.0980 -0.0974 -0.0778 -0.0882 -0.0827 -0.0946 -0.0943 -0.0790 -0.0751 -0.0818 0.0134 0.2280 0.3674 0.0885 -0.0677 -0.0916 -0.0812 -0.0793 -0.0766 -0.0671 -0.0412
 -0.0958 -0.0986 -0.0766 -0.0867 -0.0803 -0.0937 -0.0925 -0.0803 -0.0778 -0.0916 -0.0754 -0.0659 -0.0241 0.0232 -0.0778 -0.0897 -0.0803 -0.0787 -0.0781 -0.0751 -0.0745
 -0.0949 -0.0955 -0.0763 -0.0909 -0.0769 -0.0922 -0.0912 -0.0781 -0.0751 -0.0940 -0.0775 -0.0726 -0.0754 -0.0745 -0.0851 -0.0940 -0.0821 -0.0806 -0.0757 -0.0745 -0.0754
 -0.0946 -0.0980 -0.0793 -0.0851 -0.0803 -0.0934 -0.0894 -0.0726 -0.0732 -0.0952 -0.0778 -0.0748 -0.0754 -0.0748 -0.0851 -0.0946 -0.0836 -0.0793 -0.0787 -0.0763 -0.0769
 -0.0958 -0.0967 -0.0772 -0.0864 -0.0812 -0.0943 -0.0897 -0.0668 -0.0638 -0.0964 -0.0800 -0.0739 -0.0763 -0.0775 -0.0858 -0.0922 -0.0836 -0.0787 -0.0769 -0.0754 -0.0769
 -0.0955 -0.0937 -0.0751 -0.0864 -0.0790 -0.0912 -0.0876 -0.0601 -0.0464 -0.0861 -0.0790 -0.0763 -0.0766 -0.0757 -0.0854 -0.0906 -0.0824 -0.0790 -0.0754 -0.0760 -0.0781
Attachment 3: ER-580-200-30kV-m39.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1590 -0.1062 -0.1086 -0.0906 -0.0906 -0.1022 -0.1093 -0.1038 -0.0885 -0.0534 -0.0565 -0.0681 -0.0763 -0.0925 -0.1007 -0.1041 -0.1007 -0.0952 -0.1074 -0.1068 -0.1041
 -0.1016 -0.1031 -0.1047 -0.0867 -0.0873 -0.0998 -0.1019 -0.0995 -0.0922 -0.0665 -0.0510 -0.0604 -0.0708 -0.0894 -0.0949 -0.1001 -0.0940 -0.0928 -0.1013 -0.1004 -0.0989
 -0.1007 -0.0989 -0.0916 -0.0845 -0.0827 -0.0983 -0.1016 -0.0992 -0.0867 -0.0845 -0.0607 -0.0586 -0.0677 -0.0867 -0.0946 -0.0995 -0.0934 -0.0851 -0.0998 -0.0943 -0.0848
 -0.0986 -0.0970 -0.0763 -0.0781 -0.0751 -0.0919 -0.1038 -0.0992 -0.0717 -0.0867 -0.0726 -0.0647 -0.0650 -0.0717 -0.0928 -0.1019 -0.0870 -0.0888 -0.0800 -0.0790 -0.0842
 -0.0992 -0.0989 -0.0790 -0.0760 -0.0745 -0.0922 -0.1047 -0.1001 -0.0751 -0.0830 -0.0787 -0.0760 -0.0723 -0.0717 -0.0940 -0.1031 -0.0867 -0.0891 -0.0824 -0.0812 -0.0803
 -0.0989 -0.0989 -0.0790 -0.0867 -0.0757 -0.0946 -0.1056 -0.0964 -0.0720 -0.0870 -0.0787 -0.0775 -0.0772 -0.0772 -0.0934 -0.1028 -0.0827 -0.0903 -0.0790 -0.0787 -0.0763
 -0.1004 -0.0986 -0.0790 -0.0790 -0.0241 -0.0101 -0.0891 -0.0931 -0.0351 -0.0485 -0.0778 -0.0745 -0.0778 -0.0803 -0.0943 -0.1007 -0.0790 -0.0867 -0.0763 -0.0766 -0.0766
 -0.0809 -0.0986 -0.0763 -0.0790 -0.0714 0.3131 0.1181 -0.0345 0.0058 0.0626 -0.0830 -0.0748 -0.0748 -0.0766 -0.0903 -0.1010 -0.0809 -0.0873 -0.0806 -0.0790 -0.0772
 -0.0894 -0.0177 -0.0507 -0.0848 -0.0745 0.0546 0.5795 0.3409 0.0809 0.1352 -0.0708 -0.0778 -0.0803 -0.0760 -0.0894 -0.1028 -0.0824 -0.0900 -0.0784 -0.0790 -0.0784
 -0.0922 -0.0919 -0.0690 -0.0726 -0.0629 -0.0781 0.4315 1.1215 0.9171 0.2142 0.0015 -0.0677 -0.0769 -0.0787 -0.0900 -0.1031 -0.0815 -0.0888 -0.0812 -0.0787 -0.0781
 -0.0995 -0.0888 -0.0693 -0.0781 -0.0705 -0.0766 -0.0616 0.5667 1.8753 1.0110 0.1111 -0.0580 -0.0735 -0.0763 -0.0916 -0.1010 -0.0781 -0.0864 -0.0775 -0.0769 -0.0778
 -0.1001 -0.0964 -0.0726 -0.0742 -0.0717 -0.0769 -0.0699 0.0687 1.0928 3.1250 0.4935 -0.0278 -0.0681 -0.0711 -0.0888 -0.0983 -0.0754 -0.0845 -0.0800 -0.0754 -0.0763
 -0.1007 -0.0995 -0.0775 -0.0809 -0.0754 -0.0787 -0.0769 -0.0305 0.1749 1.6907 1.8475 0.2292 -0.0671 -0.0705 -0.0903 -0.0974 -0.0751 -0.0861 -0.0760 -0.0769 -0.0745
 -0.0986 -0.0986 -0.0769 -0.0824 -0.0775 -0.0888 -0.0916 -0.0552 0.0089 0.3650 1.8930 1.0120 0.1450 -0.0696 -0.0919 -0.0989 -0.0790 -0.0861 -0.0815 -0.0760 -0.0778
 -0.0986 -0.0980 -0.0754 -0.0809 -0.0751 -0.0916 -0.1025 -0.0925 -0.0546 0.0333 0.3699 0.9113 0.5783 -0.0540 -0.0885 -0.1007 -0.0809 -0.0864 -0.0772 -0.0739 -0.0766
 -0.0995 -0.0992 -0.0754 -0.0812 -0.0769 -0.0916 -0.1007 -0.0952 -0.0708 -0.0702 0.0208 0.2255 0.5313 0.2142 -0.0876 -0.0977 -0.0778 -0.0842 -0.0790 -0.0745 -0.0723
 -0.1004 -0.1010 -0.0766 -0.0812 -0.0775 -0.0894 -0.1010 -0.0958 -0.0745 -0.0848 -0.0656 -0.0620 -0.0168 0.0571 -0.0797 -0.0970 -0.0772 -0.0867 -0.0757 -0.0748 -0.0769
 -0.0986 -0.1001 -0.0769 -0.0891 -0.0803 -0.0876 -0.0992 -0.0952 -0.0784 -0.0815 -0.0702 -0.0726 -0.0760 -0.0732 -0.0891 -0.0992 -0.0784 -0.0851 -0.0763 -0.0769 -0.0763
 -0.0970 -0.0970 -0.0739 -0.0839 -0.0763 -0.0876 -0.0943 -0.0922 -0.0763 -0.0806 -0.0717 -0.0739 -0.0769 -0.0766 -0.0919 -0.1010 -0.0781 -0.0864 -0.0784 -0.0772 -0.0769
 -0.0974 -0.0958 -0.0751 -0.0845 -0.0769 -0.0867 -0.0919 -0.0864 -0.0705 -0.0803 -0.0726 -0.0754 -0.0763 -0.0772 -0.0897 -0.1010 -0.0790 -0.0864 -0.0766 -0.0778 -0.0772
 -0.0989 -0.0967 -0.0757 -0.0839 -0.0775 -0.0867 -0.0916 -0.0821 -0.0635 -0.0745 -0.0708 -0.0729 -0.0766 -0.0763 -0.0916 -0.0986 -0.0781 -0.0861 -0.0797 -0.0775 -0.0760
Attachment 4: ER-580-200-30kV-m133.txt
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1505 -0.0983 -0.0916 -0.0909 -0.0943 -0.0958 -0.0989 -0.0854 -0.0821 -0.0665 -0.0467 -0.0623 -0.0763 -0.0919 -0.1028 -0.1016 -0.0998 -0.0983 -0.1074 -0.1065 -0.1059
 -0.0885 -0.0925 -0.0885 -0.0870 -0.0931 -0.0916 -0.0946 -0.0781 -0.0815 -0.0836 -0.0473 -0.0568 -0.0702 -0.0854 -0.0983 -0.0946 -0.0974 -0.0961 -0.1035 -0.1035 -0.0986
 -0.0885 -0.0903 -0.0815 -0.0830 -0.0845 -0.0903 -0.0916 -0.0757 -0.0790 -0.0909 -0.0604 -0.0549 -0.0671 -0.0793 -0.0977 -0.0946 -0.0949 -0.0912 -0.1022 -0.0989 -0.0864
 -0.0861 -0.0873 -0.0775 -0.0803 -0.0797 -0.0836 -0.0943 -0.0800 -0.0766 -0.0848 -0.0800 -0.0650 -0.0650 -0.0714 -0.0955 -0.0946 -0.0839 -0.0885 -0.0946 -0.0757 -0.0757
 -0.0864 -0.0870 -0.0751 -0.0793 -0.0742 -0.0842 -0.0946 -0.0775 -0.0797 -0.0812 -0.0839 -0.0790 -0.0699 -0.0757 -0.0952 -0.0934 -0.0876 -0.0885 -0.0812 -0.0757 -0.0769
 -0.0903 -0.0885 -0.0760 -0.0891 -0.0735 -0.0854 -0.0937 -0.0781 -0.0812 -0.0827 -0.0778 -0.0790 -0.0784 -0.0824 -0.0974 -0.0946 -0.0861 -0.0897 -0.0876 -0.0760 -0.0766
 -0.0894 -0.0882 -0.0772 -0.0851 -0.0436 -0.0208 -0.0787 -0.0797 -0.0815 -0.0827 -0.0784 -0.0800 -0.0818 -0.0842 -0.0974 -0.0946 -0.0845 -0.0885 -0.0818 -0.0784 -0.0751
 -0.0873 -0.0861 -0.0775 -0.0775 -0.0751 0.2106 0.1611 0.0058 -0.0708 -0.0824 -0.0784 -0.0784 -0.0793 -0.0839 -0.0980 -0.0949 -0.0870 -0.0897 -0.0870 -0.0787 -0.0800
 -0.0873 -0.0858 -0.0729 -0.0864 -0.0742 -0.0360 0.5655 0.5167 0.0711 -0.0668 -0.0778 -0.0781 -0.0775 -0.0812 -0.0967 -0.0940 -0.0845 -0.0882 -0.0821 -0.0790 -0.0793
 -0.0854 -0.0861 -0.0723 -0.0778 -0.0735 -0.0790 0.1804 1.1356 1.1194 0.0473 -0.0671 -0.0751 -0.0748 -0.0803 -0.0980 -0.0946 -0.0806 -0.0873 -0.0824 -0.0781 -0.0772
 -0.0867 -0.0815 -0.0748 -0.0824 -0.0702 -0.0729 -0.0751 0.3769 2.1054 1.6153 0.0116 -0.0620 -0.0739 -0.0827 -0.0977 -0.0940 -0.0845 -0.0894 -0.0830 -0.0745 -0.0751
 -0.0897 -0.0897 -0.0708 -0.0748 -0.0717 -0.0739 -0.0720 0.0027 0.6451 3.8977 0.9238 -0.0403 -0.0723 -0.0800 -0.0970 -0.0922 -0.0812 -0.0879 -0.0833 -0.0757 -0.0742
 -0.0873 -0.0885 -0.0751 -0.0897 -0.0717 -0.0742 -0.0760 -0.0336 0.1282 1.8280 3.6035 0.3409 -0.0598 -0.0797 -0.0961 -0.0928 -0.0809 -0.0867 -0.0793 -0.0766 -0.0793
 -0.0891 -0.0870 -0.0757 -0.0842 -0.0766 -0.0818 -0.0803 -0.0497 -0.0037 0.2542 2.2183 1.9345 0.1843 -0.0778 -0.0964 -0.0928 -0.0830 -0.0882 -0.0830 -0.0772 -0.0790
 -0.0891 -0.0882 -0.0748 -0.0916 -0.0778 -0.0815 -0.0906 -0.0742 -0.0574 0.0226 0.3580 1.1969 0.9808 0.0894 -0.0970 -0.0943 -0.0800 -0.0848 -0.0824 -0.0760 -0.0763
 -0.0894 -0.0858 -0.0754 -0.0769 -0.0739 -0.0803 -0.0894 -0.0772 -0.0748 -0.0653 0.0101 0.2203 0.6061 0.4141 -0.0912 -0.0922 -0.0784 -0.0861 -0.0824 -0.0763 -0.0748
 -0.0906 -0.0864 -0.0754 -0.0876 -0.0742 -0.0797 -0.0894 -0.0760 -0.0742 -0.0726 -0.0687 -0.0598 0.0128 0.1819 -0.0031 -0.0946 -0.0833 -0.0873 -0.0836 -0.0800 -0.0760
 -0.0885 -0.0882 -0.0757 -0.0879 -0.0754 -0.0800 -0.0885 -0.0772 -0.0748 -0.0742 -0.0720 -0.0751 -0.0720 -0.0745 -0.0858 -0.0952 -0.0809 -0.0891 -0.0854 -0.0769 -0.0769
 -0.0867 -0.0861 -0.0772 -0.0815 -0.0748 -0.0797 -0.0876 -0.0745 -0.0711 -0.0757 -0.0754 -0.0742 -0.0739 -0.0793 -0.0961 -0.0937 -0.0800 -0.0848 -0.0815 -0.0766 -0.0760
 -0.0854 -0.0858 -0.0751 -0.0812 -0.0735 -0.0775 -0.0864 -0.0714 -0.0671 -0.0732 -0.0766 -0.0751 -0.0766 -0.0800 -0.0967 -0.0925 -0.0815 -0.0876 -0.0848 -0.0754 -0.0739
 -0.0885 -0.0864 -0.0757 -0.0800 -0.0742 -0.0772 -0.0842 -0.0684 -0.0616 -0.0653 -0.0729 -0.0778 -0.0778 -0.0824 -0.0967 -0.0925 -0.0824 -0.0858 -0.0809 -0.0748 -0.0760
Attachment 5: TS-2017-07-06-14-30.rga
Attachment 6: TS-2017-07-07-10-10.rga
Attachment 7: TS-2017-07-07-12-50.rga
Attachment 8: Spare_Ta_LP_SIS.docx
Attachment 9: OMEGA_OM-DAQPRO-5300_10-Jul-17_10.27.49.dlp
Attachment 10: ITE-TM4-Ta#(Spare)-LP-SIS_Test_Stand.docx
  597   Monday, July 17, 2017, 09:55 Aaron Schmidt    41882-442738-1In progressOut Gas HS for Target Cleaning Test

-No EE. Insulators, gripper arms, and blank EE plate included

-New grounded 9 pins included in brackets

-TGHT included to out gas Heat shield


Monday, July 17, 2017, 10:48: Roughing.

Monday, July 17, 2017, 11:08: Turbo pumps on.

Monday, July 17, 2017, 11:17: Ion gauge & RGA on.


Tuesday, July 18, 2017, 08:17: Water lines opened. No response on RGA. Vacuum stable @ 1.79e-07T.

Tuesday, July 18, 2017, 08:20: Power/sources on.

  Device Temp
[C]
 Flow
[lpm]
 HS + MSP 27.1 3.78
 TBHT 26.9 2.97
 COIL + EE 26.9 4.94
 TGHT 26.9 5.48

Tuesday, July 18, 2017, 08:25: Auto-ramping TGHT to 580A @ 2A per min. Vacuum @ 1.44e-07T.


 

  Device Temp
[C]
 Flow
[lpm]
 HS + MSP 36.4 3.78
 TBHT 27.1 2.97
 COIL + EE 27.4 4.84
 TGHT 29.5 5.48

Wednesday, July 19, 2017, 08:00: Auto cooling TGHT.


Monday, July 24, 2017, 07:38: Power/sources off.

Monday, July 24, 2017, 07:40: Waterlines purged

Monday, July 24, 2017, 07:42: RGA off. Ion gauge off. Turbos off.

Monday, July 24, 2017, 09:10: Venting.

Attachment 1: OMEGA_OM-DAQPRO-5300_24-Jul-17_10.23.02.dlp
Attachment 2: Target_Cleaning_Test_HS_Outgassing.docx
Attachment 3: TS-2017-07-17-11-17.rga
  598   Monday, July 17, 2017, 09:58 Aaron Schmidt HP-SIS  42218-842738-1 Target cleaning test (Long) HP-SIS CC Clean

 Target Test (Fully Cleaned vs Crystal Clean only).

For all tests:

-Start Turbo pumps & Ion gauge as soon as available on ISAC TEST STAND VACUUM Screen.

- Ensure previous test turbos are off by 8:15 and started venting by 9:15am to meet timeline.

-Start Roughing by 10:10am +/- 1 minute.

-TGHT / TBHT ramping starts @ 8:30am. 2A per min to 580A / 200A

ALL START TIMES MUST BE CONSISTENT.


***Pre-outgassed*** Heat Shield W.O# 42738-1: Includes insulators, new Al grounded 9-pin, bracket, & gripper arms. Blank/Plate. No source cup/EE.


HP-SIS Target # 42218-8: Crystal Clean only.


Monday, July 24, 2017, 09:46: HS removed from test stand. Exchange target.

Monday, July 24, 2017, 10:08: Target # 42218-8 placed into test stand. Door open for 22 minutes.

Monday, July 24, 2017, 10:10: Roughing.

Monday, July 24, 2017, 10:10: Doors sealed for roughing.

Monday, July 24, 2017, 10:11: BV1B OPEN. Opened @ 306T.

Monday, July 24, 2017, 10:29: Both Turbo pumps started.

Monday, July 24, 2017, 10:34: Ion gauge on. TP1B @ speed 42 KRPM.

Monday, July 24, 2017, 10:38: TP1A @ speed 38 KRPM.

Monday, July 24, 2017, 10:40: RGA on. Vacuum @ 1.51E-05T.

Monday, July 24, 2017, 15:03: Vacuum @ 6.14e-07T


Tuesday, July 25, 2017, 08:02: Water lines opened. No response on RGA. Vacuum stable @ 1.97E-07T.

Tuesday, July 25, 2017, 08:05: Power/Sources on.

Tuesday, July 25, 2017, 08:23: Anode increased to 5, 10, 25, 50 V. Ok @ 0.000mA

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 27.3 3.77
 TBHT 27.3 3.02
 COIL + EE 27.1 4.91
 TGHT  27.3 5.53

Tuesday, July 25, 2017, 08:30: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.

Tuesday, July 25, 2017, 09:00: Open back power supply panel. Using multimeter, 9-pin ground connection verified. Panel closed.

10:45: Check. Vacuum @ 2E-06T.

11:45: Check. Vacuum @ 3.33E-07T.

13:30: Check. Vacuum @ 2.78E-07T.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP  33.5 3.78
 TBHT  28.1 3.03
 COIL + EE 27.2  4.92
 TGHT

29.4

4.48

Tuesday, July 25, 2017, 14:27: Open back power supply panel. Using multimeter, 9-pin ground connection verified. Panel closed


Wednesday, July 26, 2017, 08:15: Check. Vacuum @ 2.60E-07T.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP  33.5 3.78
 TBHT  27.6 2.99
 COIL + EE 27.2  4.06
 TGHT

29.4

 5.52

 Wednesday, July 26, 2017, 15:00: Auto-Cooling @ 2A per min. Check. Vacuum @ @2.43e-07T


Thursday, July 27, 2017, 07:20: Check. Vacuum @ 1.44e-07T

Thursday, July 27, 2017, 07:22: Power/sources off.

Thursday, July 27, 2017, 07:24: Waterlines purged.

Thursday, July 27, 2017, 07:26: RGA off.

Thursday, July 27, 2017, 07:27: Ion gauge, Turbos off

Thursday, July 27, 2017, 09:13: Venting.

Attachment 1: HP-SIS_CC_Clean_(Long).docx
Attachment 2: OMEGA_OM-DAQPRO-5300_27-Jul-17_10.25.04.dlp
Attachment 3: TS-2017-07-24-10-40.rga
  599   Monday, July 17, 2017, 10:04 Aaron Schmidt HP-SIS  42218-1042738-1 Target cleaning test HP-SIS Full Clean

 Target Test (Fully Cleaned vs Crystal Clean only).

For all tests:

-Start Turbo pumps & Ion gauge as soon as available on ISAC TEST STAND VACUUM Screen.

- Ensure previous test turbos are off by 8:15 and started venting by 9:15am to meet timeline.

-Start Roughing by 10:10am +/- 1 minute.

-TGHT / TBHT ramping starts @ 8:30am. 2A per min to 580A / 200A

ALL START TIMES MUST BE CONSISTENT.


***Pre-outgassed*** Heat Shield W.O# 42738-1: Includes insulators, new Al grounded 9-pin, bracket, & gripper arms. Blank/Plate. No source cup/EE.


HP-SIS Target # 42218-10: Full Clean. Crystal Clear, Acetone, & Methanol used.


Thursday, July 27, 2017, 09:49: HS removed from test stand. Exchange target.

Thursday, July 27, 2017, 10:07: Target # 42218-10 placed into test stand. Door open for 20 minutes.

Thursday, July 27, 2017, 10:10: Roughing.

Thursday, July 27, 2017, 10:10: Doors sealed for roughing.

Thursday, July 27, 2017, 10:13 BV1B OPEN. Opened @ 230T.

Thursday, July 27, 2017, 10:29: Both Turbo pumps started.

Thursday, July 27, 2017, 10:34: Ion gauge on. TP1B @ speed 42 KRPM.

Thursday, July 27, 2017, 10:37: TP1A @ speed 38 KRPM.

Thursday, July 27, 2017, 10:40: RGA on. Vacuum @ 1.42T.

Thursday, July 27, 2017,15:00: Check. Vacuum @ 5.51e-07T.

Thursday, July 27, 2017, 15:02: Waterlines opened. No response on RGA. Vacuum @ 5.42e-07T

Thursday, July 27, 2017, 15:05: Power on.


Friday, July 28, 2017, 08:30: Ramping started


Monday, July 31, 2017, 07:56: Ramping down

Monday, July 31, 2017, 14:25: Power supplies off. RGA off.


Tuesday, August 15, 2017, 06:48: Ion gauge, Turbo pumps off.

Tuesday, August 15, 2017, 06:50: power off. Waterlines purged.

Tuesday, August 15, 2017, 07:55: Venting.

Attachment 1: HP-SIS_Full_Clean_(Short).docx
Attachment 2: TS-2017-07-27-10-40.rga
  600   Wednesday, July 19, 2017, 09:13 Aaron SchmidtTa Targets     DoneIon Peaks

Attached is a recent list of Ta targets with their extracted ions peaks.

Attachment 1: Magnet_peaks.xlsx
  601   Thursday, August 03, 2017, 17:00 Anders Mjos      Problems/SolutionsVacuum system tripped on power interruption

 Thursday, August 03, 2017, 17:00: Restarted TPs

  602   Tuesday, August 15, 2017, 08:41 Aaron SchmidtTa#55LP-SIS  42218-1142738-2In progressTBHT sight in test

Heat TGHT to 580 A & TGHT to  200 A for view port test.

Tuesday, August 15, 2017, 08:40: Roughing.

Tuesday, August 15, 2017, 08:59: Both Turbo pumps started.

Tuesday, August 15, 2017, 09:05: Ion gauge on. TP1B @ speed 42 KRPM.

Tuesday, August 15, 2017, 09:07: TP1A @ speed 38 KRPM.

Tuesday, August 15, 2017, 09:10: RGA on. Vacuum @ 9.14e-06T

Tuesday, August 15, 2017, 11:40: Check. Vacuum @ 6.76e-07T

Tuesday, August 15, 2017, 11:41: Waterlines opened. No response on RGA. Vacuum @ 6.49e-07T

Tuesday, August 15, 2017, 11:43: Power/sources on.

Tuesday, August 15, 2017, 13:57: Check. Vacuum @ 3.88e-07T.

Tuesday, August 15, 2017, 14:33: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.


Wednesday, August 16, 2017, 10:33: Auto-cooling TBHT to 100 A @ 4A per min.


Thursday, August 17, 2017, 07:05: TGHT / TBHT off. FC0 closed. IV4 Closed.

Thursday, August 17, 2017, 07:11: IV4 reset.

Thursday, August 17, 2017, 07:13: Power off

Thursday, August 17, 2017, 07:15 Waterlines purged


Thursday, August 24, 2017, 13:03: Waterlines ok (open). Power/source on.

Thursday, August 24, 2017, 13:10: Auto ramping TGHT to 580 A @ 2A per min.


Friday, August 25, 2017, 07:59: Open IV4. Open TMS : FC0.


Friday, September 08, 2017, 06:51: Auto ramping TGHT to 580 A @ 2A per min.


Monday, September 11, 2017, 10:55: Close TMS : FC0. Close IV4

Monday, September 11, 2017, 11:58: power/sources off

Monday, September 11, 2017, 11:00: RGA off

Monday, September 11, 2017, 11:01: Ion gauge, turbos off.

Monday, September 11, 2017, 11:59: Venting.

Monday, September 11, 2017, 13:17: Roughing.

Monday, September 11, 2017, 13:34: Turbo pumps on

Monday, September 11, 2017, 13:40: Ion gauge on.


Tuesday, September 12, 2017, 07:24: RGA, power, & sources on.

Tuesday, September 12, 2017, 07:25: Auto ramping TGHT to 580 A @ 2A per min.

Tuesday, September 12, 2017, 07:26: Open IV4. Open TMS : FC0.


Monday, September 18, 2017, 10:53: Close TMS : FC0. Close IV4

Monday, September 18, 2017, 10:57: Power/sources off.


Wednesday, September 20, 2017, 10:25: RGA, Ion gauge, Turbo pumps off. Waterlines purged.

Wednesday, September 20, 2017, 12:49: Venting.

  603   Thursday, August 17, 2017, 12:45 Anders Mjos      Problems/SolutionsTBHT power supply removed

 The Xantrex 10V 600 A power supply for the TBHT has been moved to ITE to replace the failed TGHT power supply

  604   Monday, August 21, 2017, 16:20 Anders Mjos      Problems/SolutionsTBHT power supply removed

Anders Mjos wrote:

 The Xantrex 10V 600 A power supply for the TBHT has been moved to ITE to replace the failed TGHT power supply

 A spare EMI-EMS 10 V - 500 A power supply was installed last Friday by John Chow. Dave Morris has connected the controls today and updated EPICS. The local voltage readback on the power supply does not work.

  606   Monday, September 25, 2017, 14:10 Aaron Schmidt       Pump down empty chamber

Monday, September 25, 2017, 14:10: Roughing.


Wednesday, September 27, 2017, 08:23: Turbo pumps on


Friday, October 06, 2017, 09:04: Ion gauge off. Turbos off.

Friday, October 06, 2017, 11:17: Venting.