Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  Test-Stand, Page 22 of 53  Not logged in ELOG logo
ID Datedown Author Target/No.# Source Station Module Target Oven W.O# Heat Shield W.O# Status Subject
  629   Wednesday, February 07, 2018, 08:46 Aaron Schmidt       BIAS Power supply installation

Wednesday, February 07, 2018, 08:40: A 75 kV Glassman power supply was installed in the rack for use in further HV testing.

-The 75kV power supply is linked into the keyed interlock system. The key & switch labelled BIAS/EZL will unlock it.

-The PS is currently not connected to the epics controls and can only be adjusted manually.

-The old Glassman supply is disconnected and removed from the cage.

Attachment 1: BIAS_PS_off_top_switch.jpg
BIAS_PS_off_top_switch.jpg
Attachment 2: BIAS_PS_70kV_rear_bottom.jpg
BIAS_PS_70kV_rear_bottom.jpg
Attachment 3: BIAS_PS_70kV_Front_bottom.jpg
BIAS_PS_70kV_Front_bottom.jpg
  628   Monday, February 05, 2018, 09:17 Aaron Schmidt      Problems/SolutionsEpics loss

Monday, February 05, 2018, 09:10: the epics program was in 'White-out'. Turbo pumps tripped.

Unplug, re-plug PLC.

Monday, February 05, 2018, 09:16: Restart Turbo pumps on chamber & beamline.

Monday, February 05, 2018, 09:21: Ion gauges on.


Tuesday, February 06, 2018, 09:00: TSG:CB1 noted to be red. Attempts to restart/reconfigure failed. Alive displays red MAJOR. Fault submitted.

  627   Wednesday, January 31, 2018, 12:58 Aaron Schmidt LP-SIS  (Blank)42752-3In progressHV test 4

Wednesday, January 31, 2018, 12:53: Power/sources on.

Wednesday, January 31, 2018, 12:58: BIAS / EZL / EE @ 40 / 17 / 2

Wednesday, January 31, 2018, 13:02: BIAS / EZL / EE @ 40 / 17 / 2

Wednesday, January 31, 2018, 13:05: BIAS / EZL / EE @ 50 / 22 / 2.5

Wednesday, January 31, 2018, 13:08: BIAS / EZL / EE @ 55 / 24 / 2.8

Wednesday, January 31, 2018, 13:10: BIAS / EZL / EE @ 57.5 / 27 / 3.0

Wednesday, January 31, 2018, 13:14: BIAS / EZL / EE @ 60 / 28 / 3.1

Wednesday, January 31, 2018, 13:34: No Sparks.

Wednesday, January 31, 2018, 14:14: No sparks after one hour.

Wednesday, January 31, 2018, 14:44: Spark free at 1-1/2 hours.

Wednesday, January 31, 2018, 14:45: Bringing down voltage.

Wednesday, January 31, 2018, 14:56: BIAS / EZL / EE @ 0

Wednesday, January 31, 2018, 14:57: Sources off.

Attachment 1: 2018_HV_test_4.docx
  626   Wednesday, January 31, 2018, 09:53 Aaron Schmidt    (Blank)42752-3DoneWire moved.

Wednesday, January 31, 2018, 09:50: A wire (B - F) was too close (Almost touching) to the EZL and had visible sparks during HV tests 1 & 2. The wire was adjusted to sit higher. and away from the EZL

  625   Tuesday, January 30, 2018, 06:51 Aaron Schmidt LP-SIS  (Blank)42752-3DoneHV test 3

Tuesday, January 30, 2018, 06:50: BIAS / EZL / EE @ 30 / 13 / 1.5

Tuesday, January 30, 2018, 06:52: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.

Tuesday, January 30, 2018, 06:58: Optics on / set

Tuesday, January 30, 2018, 06:59: IV4 opened.

RGA switched completely off and unplugged.

Tuesday, January 30, 2018, 13:15: BIAS / EZL / EE @ 40 / 17 / 2

Tuesday, January 30, 2018, 13:20: BIAS / EZL / EE @ 50 / 21.8 / 2.5

Tuesday, January 30, 2018, 13:50: Half hour wait. Ok. No sparks.

Tuesday, January 30, 2018, 13:55: BIAS / EZL / EE @ 55 / 24 / 2.8

15 minute wait. No sparks.

Tuesday, January 30, 2018, 14:10: BIAS / EZL / EE @ 57.5 / 27 / 3.1

Tuesday, January 30, 2018, 14:23: BIAS / EZL / EE @ 60 / 28 / 3.1

Tuesday, January 30, 2018, 14:35: No sparks.

Tuesday, January 30, 2018, 15:05: BIAS / EZL / EE @ 0

Tuesday, January 30, 2018, 15:06: Auto-cooling started.


Wednesday, January 31, 2018, 06:59: Power/sources off.

 

Attachment 1: 2018_HV_test_3.docx
  624   Monday, January 29, 2018, 10:07 Aaron Schmidt LP-SIS  (Blank)42752-3DonePower outage / HV test 2

HV test 2 cancelled.  9:38: BIAS/EZL/EE @ 30/13/1.5 kV

Monday, January 29, 2018, 09:40 (Roughly) Power went out completely

Monday, January 29, 2018, 10:00: Power up TS

Monday, January 29, 2018, 10:07:Turbos on.

Monday, January 29, 2018, 10:11: Ion gauge on Vacuum @ 3.3e-06T

Monday, January 29, 2018, 11:47; Power cycled. PLC and Canbus reset/cfg.

Monday, January 29, 2018, 11:54: Optics & Sources ok/reset.

 

Monday, January 29, 2018, 13:27: BIAS/EZL/EE @ 30 / 13 / 1.5 kV.

Monday, January 29, 2018, 13:30: BIAS/EZL/EE @ 40 / 17 / 2 kV.

Monday, January 29, 2018, 13:33: BIAS/EZL/EE @ 45 / 19.5 / 2.25 kV.

Monday, January 29, 2018, 13:36: BIAS/EZL/EE @ 50 / 21.5 / 2.5 kV.

Monday, January 29, 2018, 13:40: BIAS/EZL/EE @ 55 / 23.8 / 2.75 kV.

Monday, January 29, 2018, 13:43: BIAS/EZL/EE @ 60 / 26 / 3 kV.

SPARKS*

Monday, January 29, 2018, 13:48: BIAS/EZL/EE @ 55 / 23.8 / 2.75 kV.

Monday, January 29, 2018, 13:51: BIAS/EZL/EE @ 57.5 / 26 / 2.9 kV.

Sparks

Monday, January 29, 2018, 13:57: BIAS/EZL/EE @ 58 / 27 / 2.95 kV.

Monday, January 29, 2018, 14:03: Holding.

Monday, January 29, 2018, 14:07: BIAS/EZL/EE @ 59 / 28 / 3.1 kV.

Monday, January 29, 2018, 14:14: Spark*

Monday, January 29, 2018, 14:45: Holding

Monday, January 29, 2018, 14:56: BIAS/EZL/EE @ 0. Test Complete

Attachment 1: 2018_HV_test_2.docx
  623   Friday, January 26, 2018, 13:59 Aaron Schmidt LP-SIS  (Blank)42752-3DoneHV test 1

Friday, January 26, 2018, 14:00: BIAS/EZL/EE ramp up.

Friday, January 26, 2018, 14:06: BIAS/EZL/EE @ 30/13/1.5 kV.

Friday, January 26, 2018, 14:09: BIAS/EZL/EE @ 35/13/1.5 kV.

Spark at BIAS 30 kV and 40kV

Friday, January 26, 2018, 14:12: BIAS/EZL/EE @ 40/13/1.5 kV.

Friday, January 26, 2018, 14:15: BIAS/EZL/EE @ 45/13/1.5 kV.

Audible sparks, but none showing on the strip tool.

Friday, January 26, 2018, 14:19: BIAS/EZL/EE @ 45/19/1.5 kV.

Friday, January 26, 2018, 14:22: BIAS/EZL/EE @ 47/22/1.5 kV. Sparks @ 47000 to 47250 V.

Sparks observed on EE feed through in cage.

Friday, January 26, 2018, 14:27: BIAS/EZL/EE @ 50/22/1.7 kV.

Friday, January 26, 2018, 14:39: BIAS/EZL/EE @ 55/23.5/1.8 kV.

Bias controls 55 kV Actual @ 50 kV.

Friday, January 26, 2018, 14:44: Spark free 5 min.

Friday, January 26, 2018, 14:49: Spark free 10 min.

Attachment 1: 2018_HV_test_1.docx
  622   Friday, January 26, 2018, 11:10 Aaron Schmidt      DoneHelium tank replacement

Friday, January 26, 2018, 11:08: New Helium tank installed inside the Test Stand.

-The 'Full' tag remains. The He tank was not opened.

-The old tank was placed beside the Test Stand beneath/beside the power switch.

  621   Monday, January 22, 2018, 09:04 Aaron Schmidt      DonePower outage & Controller replacement.

Computer, TS, epics, restarted. Power checks

Monday, January 22, 2018, 09:00: ISAC PLC TLCIOC Restarted. 

Turbo pumps restarted.

Monday, January 22, 2018, 09:26: New controller for TSG:TP1B brought in.

(Connect) Global reset. Resart Turbo pumps.

Monday, January 22, 2018, 09:40: Both turbo pumps @ speed.

Vacuum controls ok

Sources ok

Optics -  Issues. Some restart, others do not.

 

  620   Tuesday, January 02, 2018, 09:18 Anders Mjos      Problems/SolutionsStarting system

Tuesday, January 02, 2018

Cycled power for TCSIOC and PLCIOC

TP2 and TP6 started

TP1B does not start. TP1B controller does not start up correctly - controller will have to be replaced. E-fault submitted (10739).

Controller for PS for Q1,6 was switching constantly. Unplugged and plugged back in resolved the problem

TP1A started


(Aaron Schmidt)

Wednesday, January 03, 2018, 08:15:  TSG:TP1B Still non-responsive.

Wednesday, January 03, 2018, 08:20: Power/sources on. Add Voltage/Amp and check power supplies.

BIAS - 300V Reads 290V - OK

EE - 300V Reads 298V - OK

EZL - 300V Reads 293 V - OK

CB1 - 10V Reads 10V - OK

VCB1 - 10V Reads 10V - OK

CBC (both)  - 10V Reads 9V - OK

TBHT - 10A Reads 9.9A (009 direct off PS) - OK

TGHT - 10A Reads 9.8A (007 direct off PS) - OK

Optics Check:

Magnet ok - Some optics start, some do not respond. Submit fault report for communications issues.

Fault Report submitted - 10742

 


Thursday, January 04, 2018, 14:38:

Kyle and Dave were able to get all the canbus optics for ISAC test stand re-configured except one device: TMS:Q2

"We investigated and it looks like the controller board for the TMS:Q2 supply is toast. Unfortunately these boards are a very old style so we don't have a surplus of replacements on hand. Our electronics shop did have some spares but the spares are not ready for use and need to have a few components soldered on. " - Kyle

The repaired controller board should be ready for next week.


Wednesday, January 10, 2018, 13:22: TS optics all fixed. Turbo pump TP1B is still out.

  619   Thursday, December 21, 2017, 15:12 Anders Mjos       Powered down devices

Powered down vacuum gauge and turbo pump controllers

  618   Wednesday, December 20, 2017, 09:11 Aaron Schmidt LP-SIS  (Blank)42752-3DoneHoliday Shut Down

Power/sources already off. The water cooling is currently left on.

Wednesday, December 20, 2017, 09:07: RGA off.

Wednesday, December 20, 2017, 09:08: Both Ion gauges and all 4 turbo pumps off.

Wednesday, December 20, 2017, 11:21: Valves closed. Roughing pumps off.

Wednesday, December 20, 2017, 11:22: RGA shut off completely.

  617   Tuesday, December 12, 2017, 08:32 Aaron Schmidt LP-SIS  (Blank)42752-3DonePump down w/ possible outgas

Check achievable vacuum. Watch for spikes on RGA. Silver bolts removed from the chamber. Vented steel screws in.

Tuesday, December 12, 2017, 08:23: Doors open from previous test. (Open for 7 minutes)

Tuesday, December 12, 2017, 08:30: Doors closed. Roughing.

Tuesday, December 12, 2017, 08:50: Turbo pumps on.

Tuesday, December 12, 2017, 08:56: Ion gauge on.


Wednesday, December 13, 2017, 06:46: Check. Vacuum @ 5.44e-07T. RGA Error.

Wednesday, December 13, 2017, 06:49: RGA restarted.

Wednesday, December 13, 2017, 06:52: Water lines opened. Vacuum @ 5.44e-07T

Wednesday, December 13, 2017, 06:54: Power on.

Wednesday, December 13, 2017, 06:56: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.


Friday, December 15, 2017, 13:37: Auto-cooling @ 2A per min


Monday, December 18, 2017, 08:15: Power/source off.


Wednesday, December 20, 2017, 09:07: RGA off.

Wednesday, December 20, 2017, 09:08: Both Ion gauges and all 4 turbo pumps off.

Attachment 1: Empty_Chamber_-_Silver_bolts_removed.docx
Attachment 2: TS-2017-12-12-09-00.rga
Attachment 3: TS-2017-12-13-06-50.rga
  616   Monday, December 04, 2017, 08:29 Aaron Schmidt      Problems/SolutionsPower outage

Power outage over the weekend. Reboot PLC and systems.

Monday, December 04, 2017, 08:22: Turbo pumps TMS:TP6 & TP2 Restarted.

Monday, December 04, 2017, 08:28: Time out on turbo pump TP1B.

Monday, December 04, 2017, 09:25: Turn off/restart TSG:TP1B Maincontrol.

Monday, December 04, 2017, 09:26: TMS:IG6 Restarted.

Monday, December 04, 2017, 14:24: Multi gauge controller to TSG:IG1 burned out. Replaced. Vacuum @ 4.45e-06T. Monitor.

Attachment 1: 20171205_103324.jpg
20171205_103324.jpg
  615   Friday, December 01, 2017, 14:31 Aaron Schmidt LP-SIS  (Blank)42752-3In progressSilver plated bolts test / Outgas TS

Chamber 'open state, no turbos' 2 weeks. Testing Silver plated 3/8"-16x1" (SCR, SHC, 18-8, AG) bolt on the TBHT.

Friday, December 01, 2017, 14:26: Roughing. BV1B open.

Friday, December 01, 2017, 14:27: BV1A open

Friday, December 01, 2017, 14:47: Turbo pumps on.

Friday, December 01, 2017, 14:53: Ion gauge on (Vacuum @ 1.44 e-04T)

Friday, December 01, 2017, 14:57: Turbos @ speed. RGA on.


Monday, December 04, 2017, 09:27: System rebooted after weekend power outage.


Tuesday, December 05, 2017, 08:10: RGA on.

Tuesday, December 05, 2017, 08:17: Waterlines opened. Vacuum ok @ 1.56e-06T

Tuesday, December 05, 2017, 08:20: Power on. (Power supplies in off position) Turn power off

Open back. switch power supplies to on.

Tuesday, December 05, 2017, 08:35: Power on

Tuesday, December 05, 2017, 08:37: Auto ramping TGHT / TBHT to 150 / 75 A @ 2A per min.

Tuesday, December 05, 2017, 09:55: Auto ramping TGHT / TBHT to 200 / 75 A @ 2A per min.

Tuesday, December 05, 2017, 10:28: Auto ramping TGHT / TBHT to 400 / 150 A @ 2A per min. Vacuum @ 1.56e-06T.

Tuesday, December 05, 2017, 12:50: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min. Vacuum @ 3.03e-06T.

Tuesday, December 05, 2017, 12:58: Auto ramping halted. Vacuum @ 3.12e-06T. Allow vacuum to improve.


Wednesday, December 06, 2017, 08:03: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min. Vacuum @ 2.65e-06T. Added note*** The Silver bolts are NOT vented ***.

Wednesday, December 06, 2017, 08:30: TGHT / TBHT @ 449 / 199 A. Vacuum @ 3.03e-06T and holding. Auto-ramping continued. No halt.

Wednesday, December 06, 2017, 10:45: Check. TGHT / TBHT @ 449 / 199 A. Vacuum @ 3.36e-06T and holding. Auto-ramping on wait.

Wednesday, December 06, 2017, 14:37: Check. TGHT / TBHT @ 449 / 199 A. Vacuum @ 3.21e-06T and holding. Auto-ramping halted. Allow Vacuum to improve.


Thursday, December 07, 2017, 07:52: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min. Vacuum @ 2.33e-06T.

Thursday, December 07, 2017, 09:00: Check. TGHT / TBHT @ 571 / 200 A. Vacuum @ 3.19e-06T. Auto-ramping halted. Allow Vacuum to improve.


Friday, December 08, 2017, 07:52: Auto ramping TGHT to 580 A from 571 A @ 2A per min. Vacuum @ 2.04e-06T.


Monday, December 11, 2017, 08:00: Auto cooling @ 2A per min.

Monday, December 11, 2017, 13:45: Sources/power off. Water cooling off.


Tuesday, December 12, 2017, 06:35: RGA, Ion gauge, turbo pumps off.

Tuesday, December 12, 2017, 07:45: Venting.

Tuesday, December 12, 2017, 08:23: Doors open.

Attachment 1: 1st_silver_bolts_test.docx
Attachment 2: TS-2017-12-05-08-10.rga
  614   Wednesday, November 15, 2017, 10:56 Aaron Schmidt      Problems/SolutionsTMS:TP6 replacement

Turbo pump TP6 requires replacement.

Wednesday, November 15, 2017, 09:55: Turbo (TP2) off. Ion gauge off.

Wednesday, November 15, 2017, 10:55: Venting

  613   Monday, October 30, 2017, 08:54 Anders Mjos      Problems/SolutionsTMS:TP6 shut down due to high noise

The turbo pump will need replacement.

  612   Tuesday, October 24, 2017, 11:38 Anders Mjos       FEBIAD coil magnetic field measurment

with Fernando

Tuesday, October 24, 2017: Testing setup with triple coil

Wednesday, October 25, 2017: Field mapping at 70 A and current/field measurements on triple coil.

Attachment 1: FEBIAD_Coil_Field_Measurement_2017-10-25.xlsx
Attachment 2: 171025-First-Run.xlsx
  611   Monday, October 16, 2017, 11:46 Aaron Schmidt      In progressPump down empty chamber Test EE

Monday, October 16, 2017, 11:45: Roughing.

Monday, October 16, 2017, 12:05: Turbo pumps started.

Monday, October 16, 2017, 12:10; Ion gauge on.


Tuesday, October 17, 2017, 10:43: Waterlines opened. RGA not on. Vacuum ok @ 1.53e-07T. Power/sources on.

Tuesday, October 17, 2017, 10:45: RGA on.

Tuesday, October 17, 2017, 10:50: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .

   

      

Tuesday, October 17, 2017, 14:44: EE Current Stable.

Tuesday, October 17, 2017, 14:52: Power/sources off.

Tuesday, October 17, 2017, 14:54: Waterlines purged.

 

 

  610   Friday, October 13, 2017, 15:53 Aaron SchmidtSiC#37HP-FEBIADITETM439530-1041751-4In progressConditioning SiC#37 (ii)

Previous short fixed. Grid clamp removed. Grid tack welded.

Friday, October 13, 2017, 15:49: Roughing.


By Anders

Friday, October 13, 2017, 16:10: TPs started

Friday, October 13, 2017, 16:20: Vacuum at 2.7E-7 T


Saturday, October 14, 2017, 06:37: Opened water supply to HS. 95/10 psi. Vacuum stable at 5.7E-7 T

Saturday, October 14, 2017, 06:42: Opened gas lines.

Saturday, October 14, 2017, 06:43: Started heater ramping to 580 / 290 A

Saturday, October 14, 2017, 06:45: BIAS/EZL/EE to 30 / 13.5 / 1.5 kV. EE current somewhat unstable around 500 uA - seems related to elevated BIAS current. Improving as BIAS current is gradually coming down.

Saturday, October 14, 2017, 15:19: TBHT to 310 A. Local readback 316 A


Sunday, October 15, 2017, 17:36 : Ramping down heaters.


Monday, October 16, 2017, 09:10: Sources off.

Monday, October 16, 2017, 09:13; IV4 Closed & Reset. FEBIAD Gas lines closed.

Monday, October 16, 2017, 09:15: RGA, Ion gauge, turbos off.

Monday, October 16, 2017, 10:40: Power off. Waterlines purged.

Monday, October 16, 2017, 10:43: Venting.

Attachment 1: SiC#37_(iii).docx
Attachment 2: TS-2017-10-13-17-44.rga
Attachment 3: TS-2017-10-14-06-35.rga
Attachment 4: ER-580-310-30kV-m20.txt
value (Voltage offset       0.0000)
     76.0000
      0.9000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1251 -0.0882 -0.0931 -0.0888 -0.0848 -0.0879 -0.0848 -0.0702 -0.0754 -0.0833 -0.0699 -0.0717 -0.0760 -0.0717 -0.0763 -0.0766 -0.0870 -0.0797 -0.0772 -0.0873 -0.0861
 -0.0650 -0.0851 -0.0873 -0.0824 -0.0797 -0.0830 -0.0793 -0.0644 -0.0723 -0.0766 -0.0653 -0.0674 -0.0739 -0.0650 -0.0720 -0.0754 -0.0812 -0.0751 -0.0739 -0.0818 -0.0800
 -0.0693 -0.0766 -0.0864 -0.0766 -0.0638 -0.0671 -0.0586 -0.0702 -0.0674 -0.0732 -0.0641 -0.0629 -0.0665 -0.0613 -0.0613 -0.0699 -0.0790 -0.0702 -0.0714 -0.0787 -0.0659
 -0.0693 -0.0757 -0.0873 -0.0787 -0.0629 -0.0662 -0.0647 -0.0659 -0.0650 -0.0748 -0.0638 -0.0629 -0.0659 -0.0623 -0.0656 -0.0671 -0.0806 -0.0690 -0.0702 -0.0784 -0.0641
 -0.0705 -0.0732 -0.0885 -0.0769 -0.0659 -0.0635 -0.0644 -0.0711 -0.0635 -0.0769 -0.0659 -0.0604 -0.0644 -0.0632 -0.0629 -0.0671 -0.0818 -0.0702 -0.0662 -0.0790 -0.0659
 -0.0726 -0.0757 -0.0864 -0.0772 -0.0656 -0.0653 -0.0665 -0.0665 -0.0626 -0.0748 -0.0653 -0.0650 -0.0653 -0.0659 -0.0638 -0.0702 -0.0824 -0.0708 -0.0690 -0.0793 -0.0705
 -0.0653 -0.0775 -0.0870 -0.0757 -0.0638 -0.0690 -0.0638 -0.0705 -0.0607 -0.0720 -0.0626 -0.0653 -0.0668 -0.0638 -0.0647 -0.0693 -0.0809 -0.0693 -0.0699 -0.0784 -0.0696
 -0.0668 -0.0775 -0.0861 -0.0748 -0.0644 -0.0699 -0.0610 -0.0674 -0.0616 -0.0723 -0.0598 -0.0623 -0.0635 -0.0638 -0.0638 -0.0705 -0.0806 -0.0702 -0.0708 -0.0797 -0.0671
 -0.0632 -0.0760 -0.0864 -0.0745 -0.0638 -0.0668 -0.0616 -0.0650 -0.0601 -0.0739 -0.0571 -0.0595 -0.0616 -0.0638 -0.0613 -0.0681 -0.0806 -0.0705 -0.0708 -0.0800 -0.0671
 -0.0677 -0.0772 -0.0861 -0.0769 -0.0616 -0.0668 -0.0647 -0.0681 -0.0620 -0.0754 -0.0558 -0.0546 -0.0613 -0.0641 -0.0635 -0.0687 -0.0815 -0.0714 -0.0677 -0.0812 -0.0690
 -0.0681 -0.0763 -0.0845 -0.0751 -0.0635 -0.0671 -0.0644 -0.0653 -0.0647 -0.0739 -0.0552 -0.0510 -0.0595 -0.0601 -0.0653 -0.0717 -0.0821 -0.0711 -0.0681 -0.0790 -0.0714
 -0.0650 -0.0775 -0.0839 -0.0745 -0.0653 -0.0665 -0.0644 -0.0693 -0.0635 -0.0745 -0.0613 -0.0485 -0.0568 -0.0583 -0.0623 -0.0681 -0.0821 -0.0711 -0.0693 -0.0775 -0.0668
 -0.0632 -0.0766 -0.0845 -0.0742 -0.0647 -0.0656 -0.0629 -0.0671 -0.0644 -0.0751 -0.0626 -0.0537 -0.0537 -0.0580 -0.0620 -0.0668 -0.0836 -0.0687 -0.0720 -0.0775 -0.0616
 -0.0717 -0.0742 -0.0873 -0.0717 -0.0632 -0.0653 -0.0668 -0.0778 -0.0620 -0.0775 -0.0613 -0.0580 -0.0546 -0.0568 -0.0604 -0.0671 -0.0824 -0.0702 -0.0708 -0.0790 -0.0620
 -0.0699 -0.0754 -0.0842 -0.0745 -0.0632 -0.0656 -0.0665 -0.0748 -0.0623 -0.0772 -0.0632 -0.0650 -0.0577 -0.0592 -0.0595 -0.0690 -0.0824 -0.0702 -0.0677 -0.0800 -0.0620
 -0.0726 -0.0787 -0.0851 -0.0751 -0.0638 -0.0668 -0.0623 -0.0620 -0.0629 -0.0778 -0.0641 -0.0671 -0.0626 -0.0616 -0.0601 -0.0668 -0.0815 -0.0720 -0.0681 -0.0797 -0.0650
 -0.0690 -0.0763 -0.0845 -0.0732 -0.0632 -0.0656 -0.0601 -0.0626 -0.0635 -0.0778 -0.0632 -0.0638 -0.0616 -0.0610 -0.0629 -0.0677 -0.0803 -0.0705 -0.0681 -0.0787 -0.0616
 -0.0662 -0.0766 -0.0851 -0.0726 -0.0635 -0.0659 -0.0626 -0.0623 -0.0629 -0.0760 -0.0623 -0.0616 -0.0638 -0.0641 -0.0620 -0.0693 -0.0815 -0.0696 -0.0696 -0.0803 -0.0659
 -0.0671 -0.0760 -0.0864 -0.0742 -0.0638 -0.0662 -0.0629 -0.0711 -0.0656 -0.0775 -0.0620 -0.0607 -0.0641 -0.0650 -0.0635 -0.0674 -0.0812 -0.0693 -0.0684 -0.0800 -0.0662
 -0.0662 -0.0763 -0.0858 -0.0723 -0.0635 -0.0653 -0.0653 -0.0754 -0.0632 -0.0760 -0.0635 -0.0620 -0.0650 -0.0623 -0.0647 -0.0674 -0.0833 -0.0690 -0.0681 -0.0790 -0.0662
 -0.0690 -0.0748 -0.0854 -0.0742 -0.0638 -0.0641 -0.0641 -0.0623 -0.0620 -0.0763 -0.0644 -0.0635 -0.0635 -0.0604 -0.0644 -0.0684 -0.0848 -0.0674 -0.0690 -0.0745 -0.0641
Attachment 5: ER-580-310-30kV-m23.txt
value (Voltage offset       0.0000)
     76.0000
      0.9000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1132 -0.0763 -0.0882 -0.0815 -0.0909 -0.1007 -0.0751 -0.0522 -0.0201 0.0049 0.0455 0.2124 0.4019 0.3796 0.3268 0.2060 0.0980 0.0424 -0.0186 -0.0430 -0.0586
 -0.0589 -0.0696 -0.0772 -0.0632 -0.0848 -0.1013 -0.0940 -0.0839 -0.0598 -0.0345 -0.0131 0.0604 0.2576 0.2960 0.2536 0.1974 0.0928 0.0262 -0.0266 -0.0397 -0.0546
 -0.0610 -0.0650 -0.0772 0.2383 1.4291 0.0836 -0.0998 -0.1157 -0.1053 -0.0772 -0.0540 -0.0183 0.1093 0.2029 0.1865 0.1715 0.0995 0.0174 -0.0412 -0.0479 -0.0443
 -0.0647 -0.0653 -0.0790 -0.0491 1.4752 2.6566 1.9754 -0.1193 -0.1337 -0.1251 -0.1062 -0.0751 -0.0204 0.0833 0.1221 0.1111 0.1025 0.0226 -0.0433 -0.0616 -0.0626
 -0.0558 -0.0665 -0.0815 -0.0601 0.1520 3.0429 6.1838 5.6906 1.1609 -0.1587 -0.1459 -0.1297 -0.0934 -0.0409 0.0317 0.0513 0.0708 0.0305 -0.0415 -0.0638 -0.0607
 0.2582 0.0153 -0.0800 -0.0623 -0.0583 1.0925 6.0068 8.4924 6.7996 4.0253 -0.0656 -0.1505 -0.1321 -0.1123 -0.0684 -0.0128 0.0092 0.0244 -0.0391 -0.0644 -0.0687
 2.1964 0.7330 0.1846 0.0098 -0.0473 -0.0342 3.0319 9.7589 9.9997 9.9997 8.0103 2.4527 -0.1282 -0.1208 -0.1114 -0.0763 -0.0461 -0.0171 -0.0385 -0.0620 -0.0681
 0.0967 2.7423 1.5198 0.9164 0.5261 0.0790 0.1855 6.0474 9.9997 9.9997 9.9997 9.9997 4.9326 0.0308 -0.0900 -0.1013 -0.1013 -0.0711 -0.0504 -0.0607 -0.0696
 -0.0598 0.0070 2.4112 2.3657 1.9696 1.4609 1.2375 1.6498 9.9881 9.9997 9.9997 9.9997 9.9997 8.4024 2.9581 -0.0681 -0.0748 -0.0665 -0.0635 -0.0629 -0.0693
 -0.0616 -0.0638 -0.0314 1.1356 3.4067 2.5894 2.9434 3.2678 7.1732 9.9997 9.9997 9.9997 9.9997 9.9997 6.3333 4.5761 -0.0378 -0.0626 -0.0671 -0.0714 -0.0674
 -0.0626 -0.0671 -0.0781 -0.0436 0.0507 3.0930 3.5120 4.0921 5.3421 9.9997 9.9997 9.9997 9.9997 9.9997 7.5992 4.5221 3.8205 1.7584 -0.0641 -0.0729 -0.0665
 -0.0620 -0.0665 -0.0800 -0.0589 -0.0464 0.0168 2.0221 4.3130 4.9597 6.9238 9.9997 9.9997 9.9997 9.9997 9.9997 3.8733 3.1485 2.4753 2.4164 -0.0705 -0.0656
 -0.0638 -0.0647 -0.0821 -0.0610 -0.0397 -0.0360 -0.0076 0.1013 3.3884 5.9052 7.3169 9.9997 9.9997 9.9997 9.9997 6.1499 1.1453 1.3699 1.5634 1.9119 0.5576
 -0.0650 -0.0650 -0.0842 -0.0632 -0.0348 -0.0180 -0.0064 -0.0299 0.0577 2.4121 4.5850 7.7075 9.9997 9.9997 9.9997 8.0157 0.5927 0.1312 0.5273 0.9311 0.9802
 -0.0626 -0.0677 -0.0824 -0.0638 -0.0366 -0.0082 0.0284 0.0201 0.0009 0.0238 0.1215 2.7127 7.3132 9.9997 9.9997 8.0194 6.8076 -0.0146 -0.0311 0.0830 0.4599
 -0.0507 -0.0668 -0.0800 -0.0635 -0.0403 -0.0092 0.0543 0.0830 0.0620 0.0211 0.0006 0.0375 0.2414 4.2218 7.6529 7.8983 6.4822 3.3484 -0.0430 -0.0577 -0.0464
 -0.0409 -0.0549 -0.0745 -0.0607 -0.0397 -0.0110 0.0613 0.1324 0.1398 0.0998 0.0613 0.0189 -0.0055 0.0623 0.1926 5.6717 6.6541 5.2524 0.6009 -0.0565 -0.0543
 -0.0400 -0.0430 -0.0623 -0.0507 -0.0330 -0.0095 0.0565 0.1620 0.2112 0.2039 0.1672 0.0922 0.0238 -0.0055 -0.0266 0.0601 0.1840 3.4863 2.0746 0.7181 -0.0562
 -0.0415 -0.0421 -0.0525 -0.0348 -0.0192 0.0024 0.0592 0.1627 0.2631 0.3098 0.3015 0.2017 0.0980 0.0317 -0.0119 -0.0381 -0.0562 0.0058 0.0195 0.9073 -0.0534
 -0.0415 -0.0427 -0.0479 -0.0259 -0.0031 0.0195 0.0763 0.1651 0.2972 0.3961 0.4553 0.3519 0.2002 0.1019 0.0223 -0.0259 -0.0571 -0.0516 -0.0635 -0.0531 -0.0604
 -0.0433 -0.0424 -0.0491 -0.0250 0.0018 0.0330 0.0964 0.1950 0.3262 0.4553 0.5765 0.5301 0.3271 0.1938 0.0745 0.0012 -0.0415 -0.0476 -0.0626 -0.0699 -0.0662
Attachment 6: ER-580-310-30kV-m28.txt
value (Voltage offset       0.0000)
     76.0000
      0.9000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1187 -0.0696 -0.0842 -0.0742 -0.0809 -0.0940 -0.0714 -0.0360 0.0162 0.0665 0.1373 0.5869 1.1185 0.8875 0.5414 0.1840 0.0412 -0.0162 -0.0540 -0.0662 -0.0754
 -0.0595 -0.0696 -0.0818 -0.0671 -0.0717 -0.0870 -0.0842 -0.0754 -0.0403 0.0131 0.0674 0.2820 0.8087 0.7736 0.5081 0.1941 0.0494 -0.0217 -0.0558 -0.0607 -0.0705
 -0.0598 -0.0729 -0.0797 -0.0629 -0.0592 -0.0653 -0.0629 -0.0919 -0.0983 -0.0595 0.0137 0.0867 0.4562 0.5594 0.4379 0.1944 0.0574 -0.0235 -0.0562 -0.0644 -0.0589
 -0.0647 -0.0735 -0.0775 -0.0659 -0.0461 0.1990 0.1093 -0.0995 -0.1205 -0.1315 -0.0946 -0.0339 0.0912 0.2820 0.3189 0.1581 0.0604 -0.0229 -0.0626 -0.0671 -0.0632
 -0.0677 -0.0653 -0.0790 -0.0671 -0.0525 0.3851 2.2314 2.3416 -0.0034 -0.1788 -0.1575 -0.1410 -0.0906 0.0049 0.1236 0.0906 0.0482 -0.0162 -0.0613 -0.0681 -0.0677
 0.0015 -0.0677 -0.0839 -0.0659 -0.0623 -0.0375 2.6950 5.9198 4.4141 1.5039 -0.2081 -0.1733 -0.1437 -0.1343 -0.0751 -0.0055 0.0079 -0.0189 -0.0604 -0.0671 -0.0653
 1.3000 0.5222 -0.0537 -0.0626 -0.0696 -0.0665 0.3500 6.6342 9.9997 9.9997 6.5463 0.6955 -0.1389 -0.1205 -0.1266 -0.0842 -0.0458 -0.0369 -0.0616 -0.0681 -0.0665
 -0.0034 1.7517 1.2799 0.7275 0.1010 -0.0507 -0.0269 1.8335 9.9997 9.9997 9.9997 9.9997 3.0206 -0.0671 -0.0848 -0.0958 -0.0940 -0.0616 -0.0626 -0.0665 -0.0653
 -0.0598 -0.0211 1.6721 2.0697 1.9296 1.1948 0.5298 0.4449 7.5043 9.9997 9.9997 9.9997 9.9997 6.2766 1.6965 -0.0626 -0.0717 -0.0607 -0.0656 -0.0653 -0.0620
 -0.0601 -0.0604 -0.0446 0.3900 3.1848 2.8616 2.8793 3.1970 3.4647 9.9997 9.9997 9.9997 9.9997 8.4708 3.4933 2.9553 -0.0632 -0.0583 -0.0653 -0.0659 -0.0641
 -0.0571 -0.0690 -0.0735 -0.0558 0.0223 3.2553 4.3109 5.0281 6.7178 9.9997 9.9997 9.9997 9.9997 9.9997 5.2866 2.2714 1.7648 0.7895 -0.0616 -0.0681 -0.0647
 -0.0552 -0.0714 -0.0766 -0.0616 -0.0546 0.0031 1.1572 6.2912 6.8265 9.1522 9.9997 9.9997 9.9997 9.9997 9.4885 1.9513 1.3593 1.0706 0.9070 -0.0662 -0.0647
 -0.0583 -0.0641 -0.0766 -0.0616 -0.0528 -0.0525 -0.0238 0.1154 4.3103 9.9997 9.9997 9.9997 9.9997 9.9997 9.9997 2.5046 0.6808 0.9274 0.7187 0.4718 0.0467
 -0.0616 -0.0589 -0.0775 -0.0601 -0.0522 -0.0519 -0.0360 -0.0238 0.0760 1.6986 9.2844 9.9997 9.9997 9.9997 9.9997 6.6360 0.0238 0.0186 0.2628 0.3699 0.2271
 -0.0632 -0.0604 -0.0793 -0.0598 -0.0565 -0.0549 -0.0290 -0.0119 -0.0073 0.0375 0.2820 2.9593 9.9997 9.9997 9.9997 9.9997 3.8776 -0.0449 -0.0552 -0.0046 0.1202
 -0.0586 -0.0693 -0.0772 -0.0629 -0.0613 -0.0571 -0.0192 0.0146 0.0284 0.0238 0.0653 0.1486 0.5829 5.3879 9.9997 9.9997 6.1002 1.5350 -0.0574 -0.0616 -0.0595
 -0.0552 -0.0641 -0.0745 -0.0665 -0.0592 -0.0586 -0.0189 0.0372 0.0952 0.0946 0.0949 0.0949 0.1031 0.1544 0.3656 5.3406 6.2766 2.5500 -0.0174 -0.0601 -0.0607
 -0.0565 -0.0565 -0.0677 -0.0610 -0.0516 -0.0601 -0.0195 0.0439 0.1477 0.2090 0.1990 0.1373 0.0867 0.0418 0.0128 0.0638 0.1416 1.5460 0.3824 -0.0607 -0.0623
 -0.0580 -0.0522 -0.0635 -0.0522 -0.0433 -0.0516 -0.0177 0.0443 0.1752 0.3250 0.3522 0.2570 0.1184 0.0406 -0.0015 -0.0357 -0.0537 -0.0323 -0.0510 -0.0592 -0.0620
 -0.0580 -0.0531 -0.0650 -0.0449 -0.0363 -0.0345 -0.0098 0.0555 0.1871 0.3961 0.5533 0.4825 0.1907 0.0653 0.0006 -0.0372 -0.0580 -0.0558 -0.0623 -0.0708 -0.0659
 -0.0552 -0.0562 -0.0626 -0.0427 -0.0360 -0.0241 0.0079 0.0748 0.2063 0.4385 0.7339 0.7773 0.3360 0.1132 0.0156 -0.0320 -0.0555 -0.0558 -0.0607 -0.0690 -0.0632
Attachment 7: ER-580-310-30kV-m40.txt
value (Voltage offset       0.0000)
     76.0000
      0.9000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1373 -0.0854 -0.0903 -0.0897 -0.0861 -0.0925 -0.0922 -0.0696 -0.0858 -0.0909 -0.0739 -0.0726 -0.0809 -0.0699 -0.0690 -0.0790 -0.0833 -0.0854 -0.0861 -0.0851 -0.0848
 -0.0699 -0.0800 -0.0858 -0.0836 -0.0821 -0.0882 -0.0864 -0.0638 -0.0812 -0.0842 -0.0720 -0.0671 -0.0769 -0.0641 -0.0653 -0.0748 -0.0778 -0.0818 -0.0781 -0.0830 -0.0793
 -0.0720 -0.0726 -0.0873 -0.0766 -0.0626 -0.0827 -0.0577 -0.0708 -0.0613 -0.0839 -0.0629 -0.0620 -0.0681 -0.0626 -0.0613 -0.0720 -0.0769 -0.0787 -0.0723 -0.0809 -0.0665
 -0.0717 -0.0735 -0.0867 -0.0748 -0.0620 -0.0571 -0.0439 -0.0620 -0.0656 -0.0833 -0.0635 -0.0659 -0.0671 -0.0638 -0.0626 -0.0723 -0.0778 -0.0775 -0.0745 -0.0784 -0.0705
 -0.0748 -0.0732 -0.0851 -0.0787 -0.0690 -0.0638 -0.0287 -0.0165 -0.0482 -0.0858 -0.0632 -0.0641 -0.0677 -0.0656 -0.0635 -0.0696 -0.0806 -0.0760 -0.0766 -0.0760 -0.0681
 -0.0705 -0.0732 -0.0854 -0.0769 -0.0659 -0.0818 -0.0394 0.0076 0.0073 -0.0424 -0.0571 -0.0620 -0.0732 -0.0647 -0.0656 -0.0702 -0.0797 -0.0769 -0.0766 -0.0772 -0.0647
 -0.0525 -0.0681 -0.0830 -0.0742 -0.0632 -0.0848 -0.0488 -0.0153 0.0351 -0.0046 -0.0055 -0.0302 -0.0751 -0.0647 -0.0653 -0.0717 -0.0787 -0.0778 -0.0742 -0.0803 -0.0601
 -0.0705 -0.0519 -0.0717 -0.0687 -0.0592 -0.0824 -0.0604 -0.0388 0.0327 0.0385 0.0308 -0.0012 -0.0299 -0.0607 -0.0629 -0.0726 -0.0784 -0.0800 -0.0729 -0.0803 -0.0620
 -0.0711 -0.0732 -0.0604 -0.0537 -0.0488 -0.0702 -0.0571 -0.0592 -0.0092 0.0513 0.0702 0.0562 -0.0006 -0.0174 -0.0323 -0.0723 -0.0797 -0.0800 -0.0745 -0.0778 -0.0632
 -0.0745 -0.0714 -0.0824 -0.0708 -0.0378 -0.0574 -0.0491 -0.0568 -0.0281 0.0040 0.0873 0.1031 0.0580 0.0113 -0.0110 -0.0378 -0.0797 -0.0781 -0.0769 -0.0766 -0.0644
 -0.0739 -0.0729 -0.0830 -0.0748 -0.0659 -0.0525 -0.0369 -0.0446 -0.0317 -0.0232 0.0644 0.1093 0.1202 0.0546 -0.0067 -0.0360 -0.0461 -0.0589 -0.0760 -0.0781 -0.0690
 -0.0714 -0.0729 -0.0839 -0.0748 -0.0638 -0.0824 -0.0427 -0.0375 -0.0269 -0.0366 0.0098 0.0888 0.0977 0.0821 0.0278 -0.0372 -0.0537 -0.0558 -0.0464 -0.0803 -0.0620
 -0.0739 -0.0732 -0.0845 -0.0748 -0.0629 -0.0842 -0.0586 -0.0708 -0.0281 -0.0473 -0.0128 0.0143 0.0793 0.0607 0.0598 -0.0336 -0.0690 -0.0656 -0.0574 -0.0534 -0.0616
 -0.0742 -0.0742 -0.0836 -0.0775 -0.0610 -0.0824 -0.0632 -0.0626 -0.0629 -0.0641 -0.0256 -0.0165 0.0177 0.0476 0.0339 0.0076 -0.0668 -0.0745 -0.0705 -0.0656 -0.0476
 -0.0717 -0.0745 -0.0836 -0.0790 -0.0632 -0.0812 -0.0638 -0.0620 -0.0647 -0.0842 -0.0632 -0.0287 -0.0168 0.0098 0.0229 0.0076 -0.0266 -0.0766 -0.0745 -0.0757 -0.0565
 -0.0684 -0.0745 -0.0833 -0.0778 -0.0641 -0.0818 -0.0629 -0.0662 -0.0623 -0.0851 -0.0635 -0.0607 -0.0702 -0.0195 -0.0034 -0.0024 -0.0247 -0.0592 -0.0729 -0.0800 -0.0613
 -0.0693 -0.0754 -0.0833 -0.0760 -0.0638 -0.0812 -0.0601 -0.0644 -0.0613 -0.0845 -0.0656 -0.0604 -0.0717 -0.0641 -0.0604 -0.0143 -0.0177 -0.0421 -0.0656 -0.0812 -0.0598
 -0.0714 -0.0723 -0.0854 -0.0729 -0.0650 -0.0797 -0.0632 -0.0775 -0.0613 -0.0827 -0.0629 -0.0604 -0.0708 -0.0650 -0.0626 -0.0708 -0.0760 -0.0507 -0.0571 -0.0769 -0.0674
 -0.0702 -0.0714 -0.0858 -0.0739 -0.0641 -0.0787 -0.0641 -0.0641 -0.0653 -0.0809 -0.0598 -0.0638 -0.0681 -0.0644 -0.0626 -0.0739 -0.0787 -0.0772 -0.0717 -0.0748 -0.0662
 -0.0720 -0.0723 -0.0842 -0.0751 -0.0632 -0.0815 -0.0632 -0.0592 -0.0613 -0.0824 -0.0583 -0.0626 -0.0690 -0.0629 -0.0604 -0.0705 -0.0769 -0.0763 -0.0757 -0.0784 -0.0641
 -0.0726 -0.0735 -0.0830 -0.0732 -0.0632 -0.0827 -0.0616 -0.0613 -0.0571 -0.0833 -0.0586 -0.0616 -0.0668 -0.0626 -0.0620 -0.0696 -0.0778 -0.0745 -0.0751 -0.0790 -0.0616
Attachment 8: SiC#37_(ii).docx
Attachment 9: OMEGA_OM-DAQPRO-5300_19-Oct-17_11.17.50.dlp
Attachment 10: ITE-TM4-SiC#37-HP-FEBIAD_Test_Stand.docx
ELOG V2.9.2-2455