ID |
Date |
Author |
Target Material |
Target Oven W.O# |
Source |
Status |
Subject |
583
|
Monday, June 20, 2022, 14:52 |
Darwin Ortiz Rosales | Empty | | | Done | softstart of new TP | Monday, June 20, 2022, 14:50
A new Turbo pump was installed this morning and it was soft started. It has operated correctly and it is ready for use. Derek O. installed it.
Previous pump had issues at start-up and was diagnosed as finished by Edi/
|
584
|
Tuesday, June 21, 2022, 09:48 |
Darwin Ortiz Rosales | SiC | 37978-1 | HP-FEBIAD | Done | T-220617 sintering | Tuesday, June 21, 2022, 09:43
Total number of foils: 377 ( including 6 analysis discs on the right side of the container)
Total weight = 1454 g (includes heatshields, spring and T bar)
Total length packed 15.5 cm (excluding 6 analysis discs that will be removed)
container has 3 Ta heat shields and is coated
No endcaps inserted to facilitate out-gassing
Ta coated spring at front (right) side
Initially to ramp TGHT and TBHT to 80A at 2 A/min
then to 120A and slow down ramp when pressure begins to increase
max approx 675A on TGHT and 230A on TBHT (Omega slope set = 0.990)
Then temperature = 1800C
TBHT will be set fixed at 230 A (max). TBHT heated manually and automatically to max.
Good vac at 3.00 e -4 Torr
Bad vac at 5.00 e-2 Torr
IGP1=3.6 e-07 Torr at start of initial ramp.
Tuesday, June 21, 2022, 13:54
IGP1= 7.1 e-05 and dropping. Start ramp to 150A for both TGHT and TBHT @1A/min
Tuesday, June 21, 2022, 21:27
Increased ramp rate to 3A/min for TGHT to 675A
Tuesday, June 21, 2022, 22:38
Ramp up to 685 A on TGHT and leave for 4 hours.
OMEGA = 1645 C but suspect reading not on Ta surface.
Wednesday, June 22, 2022, 02:28
IGP1=7.7e-6 Torr
OMEGA =1610C
TGHT = 685A, 10.5V
TBHT = 230A, 3V
initiate rampdown at -2A/min and -1Amin TGHT and TBHT |
586
|
Wednesday, June 29, 2022, 10:12 |
Darwin Ortiz Rosales | TaC | 44692-10 | HP-FEBIAD | Done | TaC coating sintering | Wednesday, June 29, 2022, 10:07
Ta HP-FEBIAD was coated with TaC and wrapped with Ta heat-shields (3 layers)
Desired temperature = 1800 C , TBHT will be set fixed at 230A (max).
Evap 2 will be pumped down over previous night
OMEGA slope= 0.990
Good vacuum = 3.00 e-4 Torr
Bad vacuum = 5.00 e-2 Torr
IGP1= 1.2 e-06 Torr
Start autoramp @2A/min to 80 A then autoramp heat @1A/min to 150 A. Then autoramp TGHT to 675A and TBHT to 230A.1
Wednesday, June 29, 2022, 10:13
Chamber isolation performed and no change of pressure observed.
Water leak test performed and no pressure change detected.
Starting ramp up
Wednesday, June 29, 2022, 13:23
Ramping to 675A on TGHT and 230A on TBHT @ 2A/min
Ramped TGHT to 680A, 10.6V and held for 4 hours
Wednesday, June 29, 2022, 21:52
Started rampdown @-2A/min TGHT and -1A/min TBHT |
589
|
Thursday, August 25, 2022, 10:16 |
Darwin Ortiz Rosales | TiC | 44692-10 | HP-FEBIAD | Done | TiC 220818 Sintering (TiC #6) | Thursday, August 25, 2022, 10:06
Target container is loaded with 379 Discs of TiC for Target and 5 extra discs for analysis on left end
Target length = 12.5
3 Ta foil heatshields and on TaC coated Ta spring on right side
Total weight including heatshields and spring = 1551.4 g Final total weight = 1545.5 g
Final power should be ~10988 Watt
TGHT ramp up to 820A, and TGHT to 300A at 0.1A/min.
IGP1= 7.7e-7 Torr
Good vacuum =3.0e-05 Torr
Bad vacuum = 5e-04 Torr
Initial TGHT and TBHT autoramped @1A/min to 50A
TBHT mas 230A
TGHT max depends on final power
Thursday, August 25, 2022, 11:17
Start autoramp at 1A/min to 80A on both TGHT and TBHT
Thursday, August 25, 2022, 11:54
leaving at 80A until pressure stops rising
Thursday, August 25, 2022, 22:29
left to autoramp to 230A TBHT and 750 A TGHT since approximately 2:00 pm
Will leave ramping @1A/min to 750 A on TGHT
TBHT currently at 230A
Friday, August 26, 2022, 07:27
TGHT = 850A, 9.3V
Started autoramp to 850A at 1A/min for TGHT
TBHT held at 230A
OMEGA = 1584C
Friday, August 26, 2022, 09:34
Target running at ~820A, 13.1 V since 9
Cooling to start between 1:15 to 1:30 pm
OMEGA reads 1668 C at top current
current reading oscillates between 812A and 828A and 13.1V to 13.2V.
Monday, August 29, 2022, 12:21
Cooling at -2A/min for TGHT and -1A/min TBHT started on time on Friday August 26th.
Target remained under vacuum after cooling, over the weekend.
Target removed this morning and looks good. 5 discs removed for analysis
|
590
|
Thursday, September 01, 2022, 12:32 |
Darwin Ortiz Rosales | TaC | 44691-13 | LP-SIS | Done | TaC coating sintering | Thursday, September 01, 2022, 10:34
The coated container was wrapped in three Ta foil heatshields.
Target temperature on OMEGA= 1850-1900 C
OMEGA slope:= 0.990
Target time at max temperature = 4 hours
Final Current TBHT= 230A
Final TGHT current= ~420A (adjustable)
Heating autoramp starged at 2A/min on both TGHT and TBHT to 40A
Then at 1A/min on both heaters to 120A.
Good vacuum = 5.0 e-05 Torr
Bad vacuum = 5.0E-04 Torr
Thursday, September 01, 2022, 12:39
Has been stabilizing at 40A on both heaters.
Starting 1A/min autoramp to 80A
A discrepancy between setpoint and readback currents was observed for both TGHT and TBHT. The difference is substancial and warrants investigation. This process will be paused until further notice
Tuesday, September 06, 2022, 14:45
Controls has been troubleshooting EVAP2. In the process TBHT has been damaged. Arthur will take a look at it ASAP and his group will try and fix it tomorrow.
Regarding our initial problem, it appears that when an electrical safety box was installed, the copper water lines connected to output terminals of the power supply, were moved so that they were in contact with the evaporator cage. See the figure below.
Thursday, September 08, 2022, 12:47
Target has been removed and a new test target (HP-FEBIAD) has been placed inside. System left to generate enough vacuum and then testing will be performed.
Friday, September 09, 2022, 15:58
System was tested and all seems good. This coat/sintering will continue after conditioning of graphite material
Saturday, September 24, 2022, 09:13
Continuation of process.
IGP1= 4.3e-07 Torr
ramping @ 1A/min to 230 A on both TGHt and TBHT
Saturday, September 24, 2022, 16:28
ramped to 430A, 7.2V on TGHT and 233A, 4.2V on TBHT. Will leave until 7 pm.
Saturday, September 24, 2022, 19:44
Starting rampdown @-2A/min TGHT -1A/min TBHT
camera offline
IGP1= 1.2e-6 Torr
Monday, September 26, 2022, 09:41
IGP1=2.2e-07 Torr. Shutting off TBHT and TGHT.
Thursday, September 29, 2022, 09:53
Final notes: coating looks good . Probably 3 hours at final temperature is better to prevent material degradation. |
Attachment 1: MicrosoftTeams-image_(13).png
|
|
591
|
Wednesday, September 14, 2022, 11:48 |
Darwin Ortiz Rosales | | | TARGET BOAT | Done | Grahite disc conditioning 220907G | Wednesday, September 14, 2022, 9:30
A Ta container tube was wrapped with 3 Ta heatshields.
322 graphite discs were loaded.
one Tantalum cup (modified to fit losely) was inserted on the front end along with a Ta spring.
A Ta spring was also installed on the back end to maintain the discs inside the tube.
The system was left to establish vacuum over night.
Good vacuum set to5e-03Torr
Bad vacuum was set to 5e-2 Torr
Total weight of system including springs = 249.63 g
Two Ta foils were installed on the copper heatshield facing the ends of the container.
The initial weight of the front Ta foil = 22.5 g
The initial weight of the back Ta foil will be determined by difference after cleaning =
OMEGA slope = 0.99
The target will be heated to 2100 C, held there for 10 hours. The target will be removed, the end Ta foils will be removed to analyses, new Ta end foils will be installed and the Target will be heated to 2100 plus 10 A higher than previous and left at that temperature for 2 hours.
Wednesday, September 14, 2022, 9:47
IGP1= 4.57e--7 Torr
initial Utoramp at 1A/10s to 80A and held at 80 amps until pressure good.
Wednesday, September 14, 2022, 11:54
Start autoramp at 1A/20s to 120A
Wednesday, September 14, 2022, 12:22
Autoramp at 1A/20s to 200A
Wednesday, September 14, 2022, 12:34
Autoramp increased to 2A/20s when TGHT = 154A
Wednesday, September 14, 2022, 13:46
Autoramping to 512A as projected current for 2100 C
Wednesday, September 14, 2022, 14:24
OMEGA = 1945 C @ 512A
Tapped to 540A, 10.5V and OMEGA = 2075. Will leave at this current and see Temperature behaeviour.
After 10 min OMEGA reads 2080C. Will start cooling at 12:20 am
Wednesday, September 14, 2022, 14:29
OMEGA = 2104C will ramp down slightly to keep 2100C
Wednesday, September 14, 2022, 14:31
OMEGA = 2098 @ 530A, 10.4V
Wednesday, September 14, 2022, 14:33
OMEGA = 2100 @ 410A,9.8V
Wednesday, September 14, 2022, 14:45
OMEGA = 2093 at 500A, 9.75V
Wednesday, September 14, 2022, 15:07
497A, 9.6V, OMEGA 2100
Wednesday, September 14, 2022, 22:09
@ approximately 515 pm noticed that OMEGA was reading 1600 C while TGHT was at 497A and 10.0V
Suspect damage to the container and possible damage
Left at the same current.
Just now the OMEGA is reading 2200C at 497A and 9.7V. Out of precaution, lowered the current to 475A and the volatage is 9.2V
OMEGA now reads 2152 C. Will leave at this current until 12:30 am when the 10 hour heating will end and will start rampdown at -2A/min
Thursday, September 15, 2022, 00:05
TGHT= 475A, 9.2V
OMEGA= 2126 C
IGP1= 1.33e-6 Torr
lowered current to 470A, 9.1 V, IGP1= 1.3e-6 Torr, OMEGA stabilized at 2120C
Thursday, September 15, 2022, 00:09
IGP1= 1.3e-6 Torr
will increase 10A to see pressure response. OMEGA= 2130C
Increase in IGP1 to 1.4e-6 Torr observed but immediately started to trend down
Thursday, September 15, 2022, 00:16
IGP1= 1.36e-6 Torr
OMEGA = 2131A
TGHT = 480 A
Thursday, September 15, 2022, 00:18
Start ramp down at -2A/min
Thursday, September 15, 2022, 09:45
IGP1=4.35e-07 Torr
TGHT turned off
Starting dismounting of target tube
Thursday, September 22, 2022, 07:04
Container tube was crystalized and was easily cut with a small pair of nail scissors. All but 18 discs were transferred to a new container. WO# 44691-3 |
592
|
Thursday, September 22, 2022, 07:15 |
Darwin Ortiz Rosales | TaC | | TARGET BOAT | Done | TaC coated endcaps and spring sintering | Thursday, September 22, 2022, 07:13
Each endcap is separated with a disc of graphite foil and rolled into 3 packs of 4 within a foil envelope in a target boat.
There are also 2 coated springs
The boat is wrapped with 3-layers of Ta heatshields.
Chamber pumped down over previous night
Good vacuum = 5.0e-04 Torr
bad vacuum = 5.e-02 Torr
OMEGA slope = 0.99
Will ramp TGHT up to 40 A then autoramp to 400A at 1A/min.
Final desired temperature on OMEGA (tap)= 1800C
IGP1= 3.8e-7 Torr
Thursday, September 22, 2022, 10:08
TGHT= 168A,
IGP1= 4.7E-07=6Torr
Increased ramp to 2A/min
Thursday, September 22, 2022, 12:08
TGHT= 400A, 6.7V
OMEGA= 1604C
Continuing autoramp to 450A @2A/min
Thursday, September 22, 2022, 12:36
TGHT=450A, 7.7V
OMEGA=1700C
Tapping to 1800C manually
Thursday, September 22, 2022, 12:54
OMEGA = 1750C at 505A, 8.8 V
Tapped to 540A, 9.5V
Will leave there for two more hours
OMEGA= 1800+/- 10 C
Thursday, September 22, 2022, 15:00
TGHT = 540A, 9.3V
OMEGA= 1713
start rampdown at-2A/min
Friday, September 23, 2022, 09:49
IGP1= 4.0e-7 Torr
Will dismount container
Saturday, September 24, 2022, 09:10
Endcaps looked good. The front spring sintered onto container and not usable. signs of overheating.
|
593
|
Thursday, September 29, 2022, 10:02 |
Darwin Ortiz Rosales | | 44139-3 | TARGET BOAT | Done | 220926-Graphite conditioning in target tube | Thursday, September 29, 2022, 09:57
A Ta container tube was wrapped with 3 Ta heatshields.
338 graphite discs were loaded.
one Tantalum cup (modified to fit losely) was inserted on the front end along with a Ta spring.
on the backside there is no spring but four discs glued together with PVA
The system was left to establish vacuum over night.
Good vacuum set to5e-04Torr
Bad vacuum was set to 5e-2 Torr
Total weight of system including springs = 249.07 g
OMEGA slope = 0.99
The target will be heated to 2100 C, held there for 10 hours. The target Temperature vs Power will be tabulated and the current required to reach 2100 will be extrapolated to avoid using OMEGA at higher temperatures.
a) The target will be removed, weighed and inspected
b) If the target boat is still good, run again at 10A higher than previous for 2 hours.
c) The target will be removed, weighed and inspected
d) i) If the weight is the same or negligible, experiment ends
ii) if the weight has decreased and target still intact, repeat from step b and document all measurements and settings
Thursday, September 29, 2022, 10:04
IGP1= 1.4e-04 Torr
TGHT= 80.0 A, 1.0V
Will allow pressure to drop and then continue heating
Thursday, September 29, 2022, 10:20
IGP1= 7.9e-05 Torr pressure on downward trend
Start heating at 2A/min to 200A
The target started dropping in temperature as current increased. Suspected OMEGA issue.
Started cooling at -2A/min
Will continue next week.
Monday, October 03, 2022, 11:05
Upon inspection, the window though which the OMEGA laser travels was coated with dark precipitate (mostly a spot in the direct path of the laser).
Cleaned the window and started pumpdown.
Monday, October 03, 2022, 12:57
start heating @2A/20s = 6A/min because the target had previously been heated past additive evaporation point.
IGP1= 3.8e-06 Torr
Monday, October 03, 2022, 18:41
Will heat to 500A and leave for 10 hours. The regression shows too high current to reach 2100. Not reasonable. So will work with 500A as a baseline.
Tuesday, October 04, 2022, 06:52
Temperature @500 A read 1742C
maintained for 10 hours
start rampdown at -2A/min
OMEGA currently 1099C
Current = 500 A
Voltage =9.9 V
IGP1= 5.9e-7 Torr
Tuesday, October 04, 2022, 21:32
Quartz window was completely covered with graphite like dust. final weight after heating = 244.57 g (a couple of discs came loose and they were weighed separately and not included in the next heating).
weighted target and added one extra heatshield because it looked like a lot of heat was escaping. Will treat at 500A again for 2 hours. The extra heat-shield should provide hotter environment.
new initial weight = 254.24 g (don't have a balance with greater precision .
Wednesday, October 05, 2022, 07:42
IGP1= 3.8e-07 Torr
ramping to 280A @ 3A/min
Wednesday, October 05, 2022, 11:40
ramping to 350 @ 3A/min
Then ramp to 500A
500A at 12:50
Wednesday, October 05, 2022, 15:52
ramping down at -2A/min
Thursday, October 06, 2022, 09:27
IGP1= 1.8e-07 Torr
Will dismount and weigh, then decide how to proceed
according to second regression, second heating was ~2000 C. OMEGA not reliable after 1700 C in this case.
Thursday, October 06, 2022, 12:34
The final weight after the second heating was = 254.19 g
reinstalled target and cleaned quartz window.
will heat up to 510 A
Friday, October 07, 2022, 07:46
IGP1= 3.0 e-07 Torr
ramp at 2A/min to 500A. will pause to take data points for regression.
Friday, October 07, 2022, 13:05
Start rampdown after two hours at 500A, 10.3V
-2A/min
|
594
|
Monday, October 17, 2022, 11:32 |
Darwin Ortiz Rosales | TaC | 45220-2 | LP-SIS | Done | TaC coating sintering | Monday, October 17, 2022, 10:03
The coated container was wrapped in 4 Ta foil heatshields. (The 4 shields appear to work well)
Target temperature on OMEGA= 1850-1900 C
OMEGA slope:= 0.990
Target time at max temperature = 4 hours
Final Current TBHT= 230A
Final TGHT current= ~420A (adjustable)
Heating autoramp started at 2A/min on both TGHT and TBHT to40A
Then at 1A/min on both heaters to 120A.
Good vacuum = 5.0 e-04 Torr
Bad vacuum = 5.0E-02 Torr
Monday, October 17, 2022, 12:45
Autoramping @4A/min on both TGHT and TBHT to 400 and 30A respectively
Monday, October 17, 2022, 12:56
slowed ramp to 2A/min on both heaters as pressure was increasing too fast. (1A/30s).
Monday, October 17, 2022, 13:30
increased ramp rate to 4A/min on TGHT as TBHT is already at 230A
Monday, October 17, 2022, 14:16
leaving at TGHT : 393A, 6.5V ; TBHT 233A, 4.2V
OMEGA reads 1864C. Will start rampdown at 6:17 pm
Monday, October 17, 2022, 18:16
OMEGA read 1875C at max and now reads 1856C
starting ramp down at -2A/min TGHT and -1A/min TBHT
Final IGP1= 1.4e-06 Torr
Tuesday, October 18, 2022, 09:37
IGP1= 2.72e-7 Torr
Turning heaters off on epics and turning vacuums off. |
597
|
Tuesday, October 25, 2022, 12:25 |
Darwin Ortiz Rosales | TaC | 44139-15-SK | LP-SIS | Done | TaC coating sintering | Tuesday, October 25, 2022, 12:18
Container wrapped in 4 Ta foil heatshields
Target temperature on OMEGA= 1850-1900 C
OMEGA slope:= 0.990
Target time at max temperature = 4 hours
Final Current TBHT= 230A
Final TGHT current= ~420A (adjustable)
Heating autoramp started at 2A/min on both TGHT and TBHT to40A
Then at 1A/min on both heaters to 120A.
Good vacuum = 5.0 e-04 Torr
Bad vacuum = 5.0E-02 Torr
Wednesday, October 26, 2022, 07:07
IGP1= 7.5e-7 Torr
start ramp at 2A/min to 40A
Ramped at 1A/min to 120A then held until pressure drop. Then ramped @ 4A/min to 400A on TGHT.
Wednesday, October 26, 2022, 14:09
Manually rampted TGHT to 415A, 6.9V will hold for 4 hours
OMEGA reads 1870 C initially
Thursday, October 27, 2022, 09:44
was held for 5 hours due to technical issues but should be ok.
OMEGA @ time of rampdown = 1830C
Rampdown @-2A/min TGHT and -1A/min TBHT
IGP1= 3.5e-07 Torr
|
598
|
Friday, October 28, 2022, 07:59 |
Darwin Ortiz Rosales | TaC | 45220-6 | LP-SIS | Done | TaC coating sintering | Friday, October 28, 2022, 07:58
The coated container was wrapped in four Ta foil heatshields.
Target temperature on OMEGA= 1850-1900 C
OMEGA slope:= 0.990
Target time at max temperature = 4 hours
Final Current TBHT= 230A
Final TGHT current= ~420A (adjustable)
Heating autoramp starged at 4A/min on both TGHT and TBHT to 40A
Then at 1A/min on both heaters to 120A.
Good vacuum = 5.0 e-02 Torr
Bad vacuum = 5.0E-04 Torr
IGP1 initial = 4.5e-07 Torr
Friday, October 28, 2022, 10:50
after reaching 120A and leaving until pressure drop, autoramp at 2.5A/30s (5A/min) towards 400A TGHT and 230A TBHT.
Friday, October 28, 2022, 11:22
OMEGA= 1890 @ 412A, 7.1 V on TGHT
Leaving for 4 hours
|
600
|
Monday, November 07, 2022, 14:26 |
Darwin Ortiz Rosales | TaC | 44139-4 | TARGET BOAT | Done | TaC coating sintering | Monday, November 07, 2022, 14:24
The coated container tube was wrapped in four Ta foil heatshields.
Target temperature on OMEGA= 1850-1900 C
OMEGA slope:= 0.990
Target time at max temperature = 4 hours
Final TGHT current= ~400A (adjustable)
Heating autoramp starged at 4A/min to 40A
Then at 1A/min on both heaters to 120A.
Good vacuum = 5.0 e-02 Torr
Bad vacuum = 5.0E-04 Torr
IGP1 initial = |
601
|
Thursday, November 10, 2022, 07:40 |
Darwin Ortiz Rosales | TaC | 45220-13 | LP-SIS | Done | TaC coating sintering | Thursday, November 10, 2022, 07:40
The coated container was wrapped in four Ta foil heatshields.
Target temperature on OMEGA= 1850-1900 C
OMEGA slope:= 0.990
Target time at max temperature = 4 hours
Final Current TBHT= 230A
Final TGHT current= ~400A (adjustable)
Heating autoramp starged at 4A/min on both TGHT and TBHT to 40A
Then at 1A/min on both heaters to 120A.
Good vacuum = 5.0 e-02 Torr
Bad vacuum = 5.0E-04 Torr
IGP1 initial = 2.8e-07 Torr
Thursday, November 10, 2022, 10:31
TGHT 230A 3.4V and TBHT at 230A , 3.9V
Continue ramp up for TGHT at 4A/min (2A/30s) to 380A
Thursday, November 10, 2022, 11:30
OMEGA =1886 C @ 430A, 7.1V TGHT; will leave for 4 hours
Thursday, November 10, 2022, 15:43
rampdown at -2A/min TGHT and -1A/min TBHT |
602
|
Thursday, November 17, 2022, 08:14 |
Darwin Ortiz Rosales | TaC | 44692-12 | TARGET BOAT | Done | TaC coating sintering | Thursday, November 17, 2022, 08:11
The coated container was wrapped in four Ta foil heatshields.
Target temperature on OMEGA= 1850-1900 C
OMEGA slope:= 0.990
Target time at max temperature = 4 hours
Final Current TBHT= N/A
Final TGHT current= ~400A (adjustable)
Heating autoramp starged at 4A/min on TGHT 40A
Then at 2A/min on both heaters to 120A.
Good vacuum = 5.0 e-02 Torr
Bad vacuum = 5.0E-04 Torr
IGP1 initial = 2.7e-07 Torr
Thursday, November 17, 2022, 10:41
OMEGA=1897C at 358A, 6.0V
Will ramp down at 230 pm
Thursday, November 17, 2022, 14:34
starting rampdown at -2A/min
OMEGA reads 1865C and IGP1=7.6e-07 Torr
|
603
|
Friday, November 18, 2022, 15:13 |
Darwin Ortiz Rosales | TaC | 45220-1 | LP-SIS | Done | TaC coating sintering | Friday, November 18, 2022, 15:08
The coated container was wrapped in four Ta foil heatshields.
Target temperature on OMEGA= 1860-1900 C
OMEGA slope:= 0.990
Target time at max temperature = 4 hours
Final Current TBHT= 230A
Final TGHT current= ~400A (adjustable)
Heating autoramp starged at 4A/min on both TGHT and TBHT to 40A
Then at 2A/min on both heaters to 120A.
Good vacuum = 5.0 e-02 Torr
Bad vacuum = 5.0E-04 Torr
leaving to pressure down over weekend
Sunday, November 20, 2022, 12:11
start heatup IGP1=1.63e-07 Torr
Sunday, November 20, 2022, 14:21
OMEGA= 1900 at 395A, 6.6V on TGHT
Will leave until 6;20 pm |
604
|
Tuesday, November 22, 2022, 14:59 |
Darwin Ortiz Rosales | TaC | 45220-12 | LP-SIS | Done | TaC coating sintering | Tuesday, November 22, 2022, 14:58
The coated container was wrapped in four Ta foil heatshields.
Target temperature on OMEGA= 1860-1900 C
OMEGA slope:= 0.990
Target time at max temperature = 4 hours
Final Current TBHT= 230A
Final TGHT current= ~400A (adjustable)
Heating autoramp starged at 4A/min on both TGHT and TBHT to 40A
Then at 2A/min on both heaters to 120A.
Good vacuum = 5.0 e-02 Torr
Bad vacuum = 5.0E-04 Torr
leaving to pressure down.
Wednesday, November 22, 2022, 19:10
OMEGA = 1942 C @ 396A on TGHT
lowered to 386A to 1900 and left for 4 hours.
After four hours, started rampdown at-2A/min TGHT and -1A/min TBHT
|
606
|
Friday, December 02, 2022, 09:50 |
Darwin Ortiz Rosales | | 45220-12 | TARGET BOAT | Done | Grahite disc conditioning 221130-Graphite | Friday, December 02, 2022, 09:46
297 graphite discs packed into a Ta tube wrapped in four heatshields.
Good vacuum 5.0e-04 Torr
Bad vacuum 5.0 e-02 Torr
Will be conditioned @ 2000 C for 5 hours and then for one hour at 10A greater than final current. Final current will be calculated from
regression.
IGP1= 5.7 e-07 Torr
ramping to 40A @ 2A/min TGHT
Friday, December 02, 2022, 20:59
Was ramped to 320A but the window must be coated because Temperature readings are drastically off. Will cool down and repeat on Monday after cleaning window. Will ramp down at -2A/min
OMEGA= 1360 C, expect much higher.
Tuesday, December 06, 2022, 10:58
ramp started at 4A/min . Initial pressure was 2.6 e-07 Torr
Tuesday, December 06, 2022, 12:45
OMEGA readings low (1290 @ 331A, 5.6V )
Will tap up until pressure bump observed.
Tuesday, December 06, 2022, 13:04
Pressure bump observed around 440 A, 7.9 V
IGP1= 8.4 e-06 Torr.
Current stopped and will keep at this current for 10 hours (11 pm rampdown)
Tuesday, December 06, 2022, 13:24
Ramped to 500A, Will leave at 500A for the rest of the conditioning.
Tuesday, December 06, 2022, 23:05
Starting rampdown at -2A/min
IGP1= 4.9e-07 Torr
TGHT= 500A, 9.3V
Wednesday, December 07, 2022, 09:43
Dismounting today.
IGP1 final = 9.6e-08 Torr
Wednesday, December 07, 2022, 13:29
Reinstalled in the evaporator. Window was not shaded. Will try heating to 500A once more.
ramp at 6A/min |
607
|
Friday, December 16, 2022, 10:03 |
Darwin Ortiz Rosales | TaC | | TARGET BOAT | Done | TaC coated endcaps and spring sintering | 4 TaC coated springs and 4 Tac coated caps loaded into a target boat.
Target boat wrapped in 4 Ta heatshields
Target boat type was with endcaps at both ends.
Left in Evap2 over previous night to depressurize.
Target temperature= 1850-1900C for 4 hours
Good vacuum = 5e-04 Torr
Bad vacuum = 5e-02 Torr
Friday, December 16, 2022, 10:01
Heated @1A/min to 100A on TGHT
Then heated 2A/min to 120A
Friday, December 16, 2022, 10:08
started autoramp @4A/min to 300A
Friday, December 16, 2022, 11:15
TGHT=450 A, 7.6 V
OMEGA= 1863 C
IGP1= 7.2e-06 Torr
Will keep for 4 hours
Monday, December 19, 2022, 10:30
Sintering was perfect. Repeat same configuration next time. |
608
|
Thursday, February 09, 2023, 09:49 |
Darwin Ortiz Rosales | Nb | | TARGET BOAT | Done | condition Nb samples (Phillip) | Thursday, February 09, 2023, 09:47
Target boat with a layer of niobium foil on the bottom is used to heat up Nb samples to 1400 C for 4 hrs.
The niobium samples are wrapped in a Nb foil.
Boat is wrapped with 2 layers of Ta foil, careful to avoid touching samples.
Samples left in Evap2 overnight (pumping down).
Good vacuum = 5.0e-04 Torr
Bad vacuum = 5.0e-02 Torr
Thursday, February 09, 2023, 09:48
IGP1= 6.19e-07 Torr
Autoramping to 100A @1A/min
Thursday, February 09, 2023, 09:57
Vacuum stable. ramp at 6A/min to 50 A instead
Thursday, February 09, 2023, 10:09
ramping at 2A/min to 100A
Thursday, February 09, 2023, 11:29
1400C on OMEGA reached @ 341A, 5.5V Will stop and watch until stable
Thursday, February 09, 2023, 15:29
OMEGA= 1408 C
IGP1= 7.97e-07 Torr
Start rampdown @ -1A/min |
609
|
Thursday, February 23, 2023, 18:24 |
Darwin Ortiz Rosales | TaC | 45441-2 | HP-FEBIAD | Done | TaC coating sintering | Thursday, February 23, 2023, 18:20
Ta HP-FEBIAD was coated with TaC and wrapped with Ta heat-shields (3 layers) and 1 under layer that doesn't cover the middle section.
Desired temperature = 1800 C , TBHT will be set fixed at 230A (max).
Evap 2 will be pumped down to 5e-06 Torr range
OMEGA slope= 0.990
Good vacuum = 5.00 e-4 Torr
Bad vacuum = 5.00 e-2 Torr
IGP1=2.5 e-06 Torr
Start autoramp @1A/min to 80 A then autoramp heat @1A/min to 150 A. Then autoramp TGHT to approx 675A and TBHT to 230A.1
Friday, February 24, 2023, 09:10
heated overnight @0.5 A/min to 400 A. Then started heating at 1A/min to 600 A while taking readings for current/Temp regression.
regression so far points to 682A to get to 1800C. Will continue taking readings and possible heat to 1850 C.
Friday, February 24, 2023, 09:43
autoramping to 690A at 1A/min
Friday, February 24, 2023, 11:24
ramping manually to 690A from 650A
Friday, February 24, 2023, 11:32
4 hours at 330pm
TGHT 492A, 10.6V
TBHT = 230A, 3.2V
IGP1=5.11e-06 Torr
OMEGA =1840 C
Friday, February 24, 2023, 15:36
TGHT = 692A, 10.6V
TBHT = 230 A, 3.2V
OMEGA = 1835 C
IGP1= 2.4e-06 Torr
Starting rampdown at -2A/min TGHT and -1A/min TBHT |
Attachment 1: WO_45441-2.png
|
|
|