Friday, December 02, 2022, 09:46
297 graphite discs packed into a Ta tube wrapped in four heatshields.
Good vacuum 5.0e-04 Torr
Bad vacuum 5.0 e-02 Torr
Will be conditioned @ 2000 C for 5 hours and then for one hour at 10A greater than final current. Final current will be calculated from
regression.
IGP1= 5.7 e-07 Torr
ramping to 40A @ 2A/min TGHT
Friday, December 02, 2022, 20:59
Was ramped to 320A but the window must be coated because Temperature readings are drastically off. Will cool down and repeat on Monday after cleaning window. Will ramp down at -2A/min
OMEGA= 1360 C, expect much higher.
Tuesday, December 06, 2022, 10:58
ramp started at 4A/min . Initial pressure was 2.6 e-07 Torr
Tuesday, December 06, 2022, 12:45
OMEGA readings low (1290 @ 331A, 5.6V )
Will tap up until pressure bump observed.
Tuesday, December 06, 2022, 13:04
Pressure bump observed around 440 A, 7.9 V
IGP1= 8.4 e-06 Torr.
Current stopped and will keep at this current for 10 hours (11 pm rampdown)
Tuesday, December 06, 2022, 13:24
Ramped to 500A, Will leave at 500A for the rest of the conditioning.
Tuesday, December 06, 2022, 23:05
Starting rampdown at -2A/min
IGP1= 4.9e-07 Torr
TGHT= 500A, 9.3V
Wednesday, December 07, 2022, 09:43
Dismounting today.
IGP1 final = 9.6e-08 Torr
Wednesday, December 07, 2022, 13:29
Reinstalled in the evaporator. Window was not shaded. Will try heating to 500A once more.
ramp at 6A/min |