Entry  Tue Nov 25 07:49:50 2014, Brian Minato, Other, General, MCIS and SIS Update 
MCIS-
the Se was loaded into the oven last friday (21st).  During the weekend the presence of 76Se +11 to +15 was verified.
NOTE- the oven p/s on the cart is broken.  An attempt was made to remove it on friday.  It may be possible to remove, if the rack it
Entry  Tue Nov 25 12:23:28 2014, Keerthi Jayamanna, Routine, General, 76Se+13 tune saved 
141125_1159_OLIS_76Se+13.snap is saved in reference directory.
El=3480/4303V
MCIS:YCB1=110
Entry  Mon Apr 20 08:09:03 2015, Brian Minato, Routine, General, OLIS Shutdown Update 
The OLIS 1 week shutdown has been completed (WP 2015-04-13-1).
General-
Y-Box insulators were inspected and look good.  Beamline optics were inspected.
Entry  Mon Sep 14 11:21:16 2015, Keerthi Jayamanna, Routine, General, 20Ne+2 beam saved 
20Ne+2 beam optics saved as 150914_1117_ios0iosfc6.snap
Entry  Fri Oct 24 10:50:20 2014, Brian Minato, Other, Other, OLIS Update 
The nitrogen bottle needs to be replaced.
 
MCIS-
Entry  Tue Oct 28 13:30:47 2014, Brian Minato, Other, Other, MCIS Update 
MCIS-
the cleaning has been completed.  The oven assembly is removed and the gate valve closed.
The quartz tubes on the ground electrode have been installed with the larger diameter tube pushed as far downstream (toward the optics box) as
Entry  Thu Oct 30 13:14:51 2014, Brian Minato, Other, Other, OLIS Update 
MCIS-
a new oven assembly was installed on the MCIS yesterday.  The cart has been installed in the cage and is connected.
The power is connected to the cart oven as is the MCIS bias.  it is currently on and being tested for Se
Entry  Fri Oct 31 20:39:56 2014, Brian Minato, Other, Other, OLIS Update 
MCIS-
the MWS Bias and PE have been moved on to the MCIS (the MCIS connections have been removed).
The plug in power to the HP Signal generator has been moved to a continuous source (Removed from the source controlled by the SSW).
Entry  Thu Nov 6 08:37:52 2014, Brian Minato, Other, Other, OLIS Update 
MCIS-
the source is off.  It is under vacuum and in the HV cage.
SIS-
Entry  Thu Nov 13 16:31:54 2014, Brian Minato, Other, Other, SIS 
A new SIS source has been installed.  The oven has been loaded with Sodium Hydrogen Carbonate (Alfa Aesar 10863) and Rubidium Nitrate.
Pictures are located in 2014-11-13 Source and PS Pics
Entry  Thu Nov 13 16:34:29 2014, Brian Minato, Other, Other, MCIS Update 
The ethernet cable from the CPI TWT amplifier has been temporarily moved to the Alga 50W amplifier.  The Alga had been experiencing loss of communication. 
This has not been an issue with the CPI so this cable will be tried with the Alga.  A signal generator is not connected to the Alga.
Entry  Wed Dec 17 13:17:13 2014, Brian Minato, Other, Other, MCIS Update 
The MCIS was vented with nitrogen yesterday.  The Se oven was drawn back into the tuner.  It was removed with the tuner and blanked.
The standard quartz window was removed and replaced with a ceramic, test window supplied by CPI. (p/n VWU2054).  The system is currently pumping
down.
Entry  Wed Apr 29 07:48:22 2015, Brian Minato, Other, Other, Repair- Water leak inside HV cage 
A water leak was reported inside the HV cage (FR#8079).  The cage was accessed to determine the source of the leak (WP#I2015-04-27-1).
The WMS cusp cooling line was leaking.  It was close to the MCIS cart (at ground) and had sparked through.  The hose was cut at the puncture
location and re-connected. to the fitting.  It was re-routed to be further away from the MCIS cart.