MCIS-
the Se was loaded into the oven last friday (21st). During the weekend the presence of 76Se +11 to +15 was verified.
NOTE- the oven p/s on the cart is broken. An attempt was made to remove it on friday. It may be possible to remove, if the rack it is mounted on is loosened and turned.
SIS-
the Ionizer p/s control cable was removed to allow for local operation. Keerthi tested the power supply in local mode on the weekend and found that the voltage did fluctuate. This was confirmed by changes in the beam current.
Set-up pictures are located in -
M:\OLIS\OFF LINE ION SOURCE\SURFACE SOURCE\Single Source - Purchased Box\PICTURES\2014-11-21 Voltage Fluctuation setup
Ionizer Power supply + is connected to the common on the SIS source
Ionizer Power supply - is connected to the Ionizer on the SIS source
a cable is connected from the SIS main, bias plate to the negative terminal on the p/s. |