Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  RH-ISAC, Page 114 of 138  Not logged in ELOG logo
  ID Date Author Category Type Module Target/Number Subject
  488   Thursday, October 03, 2013, 10:18 Travis CaveSouth Hot-CellStandard OperationTM4no targetmodule move

TM#4 with no target has been moved from the south hot cell to the south west silo.

  487   Wednesday, October 02, 2013, 14:47 Travis CaveSouth Hot-CellStandard OperationTM4no targetElectrical check

Electrical check done on TM#4 with no target, see PDF.

Attachment 1: TM#4_elec_chk_No_Tgt_SHC.PDF
TM#4_elec_chk_No_Tgt_SHC.PDF
  486   Wednesday, October 02, 2013, 13:44 Bevan MossSouth Hot-CellStandard OperationTM4 Right Coil is leaking

 Today Maico and Travis completed and electrical test to determine the right and left coil on TM4. The left coil is the negative side. From there Maico pumped on both lines for a leak test. The leak cart pumped down to its lower limits on the left line and there was no response when sprayed with helium. The right line would not pump below 2.5 x 10^-2 Torr and had a base leak rate of 3.3 x 10^-7 atm*cc/sec (this indicates a massive leak). Without spraying helium the leak cart responded and peaked at 1.9 x 10^-6 atm*cc/sec. Attached is a picture of the module indicating the line.

 

Attachment 1: tm4leak.jpg
tm4leak.jpg
  485   Wednesday, October 02, 2013, 09:40 Bevan MossConditioning StationStandard OperationTM3NiO#2HP conditioning

Yesterday the TBHT was ramped to 250 amp and the TGHT was ramped to 440 amps. Attached is the data from that run the strip tool that had this screen shot crashed but it can be recreated with the data. The TBHT ps was not reading back accurate results as such there are large steps in current or voltage or some instability for no apparent reason. In addition to this the ramping program is still not working so the ramping was a very manual process. The strip tool was left running over night, I have attached a screen shot of the the run. You will notice there are changes in vacuum that corresponds to changes in the TGHT and the TBHT voltage. Anders and myself used the archiver to plot the data with TS6 (TS3 would be better and I will have it added to the achiver), the changes in voltage and vacuum match perfectly with the compressor coming on and off. This should be investigated and possibly the temperature range of the cycling of the compressor reduced.

Attachment 1: 01102013_TM3_NiO2_TGHT_and_TBHT_ramp_up_eod.txt
Time	TCS:IG1:RDVAC [T]	TCS:IG2:RDVAC [T]	TCS:IG1S:RDVAC [T]	TCS:TGHT:RDVOL [V]	TCS:TGHT:RDCUR [A]	TCS:TBHT:RDVOL [V]	TCS:TBHT:RDCUR [A]	
10/01/2013 12:26:24.968541	                      0.000003494                      0.000008483                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:25.968221	                      0.000003516                      0.000008483                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:26.968076	                      0.000003516                      0.000008483                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:27.967994	                      0.000003494                      0.000008483                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:28.967909	                      0.000003516                      0.000008483                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:29.967723	                      0.000003472                      0.000008461                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:30.967596	                      0.000003494                      0.000008461                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:31.967466	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:32.992188	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:33.992187	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:34.992103	                      0.000003516                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:35.991944	                      0.000003516                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:36.991845	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:37.991694	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:38.992589	                      0.000003494                      0.000008505                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:39.992441	                      0.000003494                      0.000008483                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:40.992335	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:42.60462	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:43.60139	                      0.000003516                      0.000008461                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:44.61035	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:45.60806	                      0.000003494                      0.000008505                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:46.60702	                      0.000003494                      0.000008505                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:47.60766	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:48.61462	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:49.62355	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:50.62179	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:51.62047	                      0.000003494                      0.000008483                      0.000007428                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:52.61945	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:53.61799	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:54.62693	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:55.62785	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:56.62449	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:57.62550	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:58.63169	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:26:59.64295	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:00.63960	                      0.000003472                      0.000008461                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:01.63807	                      0.000003472                      0.000008461                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:02.63666	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:03.63581	                      0.000003494                      0.000008505                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:04.64414	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:05.64537	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:06.64172	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:07.64079	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:08.63970	                      0.000003516                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:09.64820	                      0.000003516                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:10.64701	                      0.000003494                      0.000008505                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:11.64536	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:12.64452	                      0.000003516                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:13.64287	                      0.000003516                      0.000008505                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:14.65215	                      0.000003494                      0.000008505                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:15.65043	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:16.64919	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:17.64797	                      0.000003494                      0.000008505                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:18.64668	                      0.000003494                      0.000008505                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:19.65546	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:20.65649	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:21.66282	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:22.66192	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:23.66088	                      0.000003494                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:24.66997	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:25.67780	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:26.67657	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:27.68761	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:28.69545	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:29.173136	                      0.000003516                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:30.173380	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:31.174041	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:32.173956	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:33.175017	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:34.175756	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:35.175519	                      0.000003472                      0.000008483                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:36.175439	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:37.175269	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:38.177533	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:39.178038	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:40.177940	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:41.177878	                      0.000003494                      0.000008483                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:42.177674	                      0.000003450                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:43.177534	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:44.178494	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:45.178420	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:47.556742	                      0.000003472                      0.000008483                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:48.556877	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:49.556716	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:50.556647	                      0.000003450                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:51.556470	                      0.000003450                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:52.555722	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:53.556262	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:54.556119	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:55.555966	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:56.557075	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:57.556794	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:58.556582	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:27:59.556637	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:00.556339	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:01.556215	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:02.556121	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:03.555991	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:04.556884	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:05.557704	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:06.557663	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:07.557508	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:08.557363	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:09.557201	                      0.000003494                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:10.557206	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:11.556952	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:12.603135	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:13.603729	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:14.603627	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:15.604448	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:16.604357	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:17.604557	                      0.000003494                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:18.604130	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:19.604964	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:20.604817	                      0.000003472                      0.000008461                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:21.605840	                      0.000003494                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:22.605655	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:23.605561	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:24.606470	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:25.606226	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:26.606108	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:27.607225	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:28.606838	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:29.607689	                      0.000003472                      0.000008461                      0.000007406                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:30.607592	                      0.000003450                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:31.607731	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:32.607319	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:33.607229	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:34.608083	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:35.607942	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:36.607853	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:37.639580	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:38.639598	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:39.639435	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:40.639309	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:41.639237	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:42.639092	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:43.642376	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:44.642881	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:45.643933	                      0.000003450                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:46.669142	                      0.000003450                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:47.669528	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:48.670348	                      0.000003450                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:49.671254	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:50.692124	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:51.692466	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:52.691865	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:53.691703	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:54.692702	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:55.692422	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:56.692317	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:57.692221	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:58.692092	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:28:59.692955	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:00.692831	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:01.692670	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:02.692596	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:03.692421	                      0.000003472                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:04.693353	                      0.000003494                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:05.694193	                      0.000003494                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:06.773807	                      0.000003450                      0.000008439                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:07.773908	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:08.773785	                      0.000003450                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:09.773707	                      0.000003472                      0.000008461                      0.000007340                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:10.773565	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:11.773425	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:12.773282	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:13.774375	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:14.775067	                      0.000003472                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:15.774905	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:16.774981	                      0.000003450                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:17.775909	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:18.775561	                      0.000003450                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:19.775556	                      0.000003450                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:20.775386	                      0.000003450                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:21.775154	                      0.000003450                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:22.775046	                      0.000003472                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:23.774939	                      0.000003450                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:24.775322	                      0.000003428                      0.000008439                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:25.775712	                      0.000003450                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:26.775518	                      0.000003450                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:27.776544	                      0.000003450                      0.000008461                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:28.776275	                      0.000003450                      0.000008483                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:29.776180	                      0.000003472                      0.000008461                      0.000007384                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:30.777108	                      0.000003450                      0.000008483                      0.000007362                           BadVal                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:31.789402	                      0.000003450                      0.000008483                      0.000007362                              0.0                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:32.789771	                      0.000003450                      0.000008483                      0.000007362                              0.0                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:33.789680	                      0.000003472                      0.000008483                      0.000007384                              0.0                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:34.789545	                      0.000003450                      0.000008483                      0.000007384                              0.0                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:35.789418	                      0.000003428                      0.000008461                      0.000007384                              0.0                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:37.442708	                      0.000003472                      0.000008461                      0.000007362                              0.0                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:38.442085	                      0.000003450                      0.000008461                      0.000007362                              0.0                           BadVal                           BadVal                           BadVal
10/01/2013 12:29:39.442135	                      0.000003450                      0.000008483                      0.000007362                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:40.441811	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:41.442707	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:42.442561	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:43.442435	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:44.443379	                      0.000003472                      0.000008461                      0.000007362                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:45.443198	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:46.444259	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:47.444924	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:48.444821	                      0.000003450                      0.000008483                      0.000007384                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:49.444709	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:50.444638	                      0.000003472                      0.000008461                      0.000007362                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:51.549284	                      0.000003472                      0.000008461                      0.000007384                              0.0                             3.36                           BadVal                           BadVal
10/01/2013 12:29:52.552322	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:29:53.552168	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:29:54.553247	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:29:55.553967	                      0.000003472                      0.000008461                      0.000007340                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:29:56.553827	                      0.000003450                      0.000008461                      0.000007406                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:29:57.553658	                      0.000003450                      0.000008417                      0.000007384                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:29:58.553540	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:29:59.554438	                      0.000003472                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:30:00.554431	                      0.000003472                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:30:01.555204	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:30:02.555036	                      0.000003450                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           BadVal
10/01/2013 12:30:03.554940	                      0.000003472                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:04.555874	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:05.555662	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:06.555571	                      0.000003472                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:07.555438	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:08.555439	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:09.556229	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:10.557048	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:11.556908	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:12.557182	                      0.000003428                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:13.557690	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:14.558563	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:15.558521	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:16.558282	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:17.558166	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:18.558067	                      0.000003428                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:19.557905	                      0.000003428                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:20.557772	                      0.000003428                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:21.557692	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:22.557585	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:23.662422	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:24.792773	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:25.793155	                      0.000003428                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:26.793121	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:27.814109	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:28.814848	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:29.816080	                      0.000003428                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:30.936335	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:31.936479	                      0.000003450                      0.000008439                      0.000007340                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:32.936542	                      0.000003428                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:33.937159	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:34.936998	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:36.28953	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:37.29782	                      0.000003450                      0.000008461                      0.000007340                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:38.101823	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:39.102451	                      0.000003450                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:40.102434	                      0.000003428                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:41.102235	                      0.000003428                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:42.102174	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:43.178823	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:44.178888	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:45.178756	                      0.000003472                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:46.178627	                      0.000003406                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:47.178529	                      0.000003428                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:48.178427	                      0.000003450                      0.000008439                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:49.178213	                      0.000003450                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:50.179230	                      0.000003450                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:51.179939	                      0.000003450                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:52.561149	                      0.000003428                      0.000008461                      0.000007340                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:53.563445	                      0.000003450                      0.000008439                      0.000007340                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:54.563591	                      0.000003450                      0.000008439                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:55.707336	                      0.000003406                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:56.732944	                      0.000003428                      0.000008439                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:57.846017	                      0.000003406                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:30:58.988827	                      0.000003428                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:00.90800	                      0.000003428                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:01.124740	                      0.000003450                      0.000008439                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:02.124607	                      0.000003428                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:03.124442	                      0.000003428                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:04.124375	                      0.000003406                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:05.124268	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:06.125092	                      0.000003450                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:07.124979	                      0.000003428                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:08.124845	                      0.000003450                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:09.124778	                      0.000003428                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:10.125618	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:11.125456	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:12.125314	                      0.000003450                      0.000008439                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:13.125188	                      0.000003450                      0.000008439                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:14.126061	                      0.000003450                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:15.126120	                      0.000003406                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:16.125873	                      0.000003450                      0.000008439                      0.000007340                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:17.125797	                      0.000003450                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:18.125584	                      0.000003450                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:19.125575	                      0.000003428                      0.000008439                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:20.126343	                      0.000003406                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:21.126184	                      0.000003406                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:22.126058	                      0.000003406                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:23.125975	                      0.000003428                      0.000008439                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:24.125801	                      0.000003428                      0.000008461                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:25.125684	                      0.000003428                      0.000008439                      0.000007384                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:26.125551	                      0.000003428                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
10/01/2013 12:31:27.126146	                      0.000003428                      0.000008461                      0.000007362                              0.0                             3.36                           0.0146                           0.8058
... 14745 more lines ...
Attachment 2: 02102013_TM3_NiO#2_HP_testing_overnight_run.JPG
02102013_TM3_NiO#2_HP_testing_overnight_run.JPG
Attachment 3: 02102013_TM3_NiO#2_HP_testing_overnight_run_archiver.JPG
02102013_TM3_NiO#2_HP_testing_overnight_run_archiver.JPG
  484   Tuesday, October 01, 2013, 11:00 Travis CaveConditioning StationStandard OperationTM3NiO#2Electrical check

TM#3 with NiO#2 electrical check at conditioning station under vacuum, see pdf for details.

Attachment 1: TM#3_elec_chk_NiO#2_CS2.PDF
TM#3_elec_chk_NiO#2_CS2.PDF
  483   Tuesday, October 01, 2013, 10:59 Travis CaveSouth Hot-CellStandard OperationTM4SiC#29Electrical check

TM#4 with SiC#29 electrical check post beam see pdf for details.

Attachment 1: TM#4_elec_chk_SiC#29_SHC_post2.PDF
TM#4_elec_chk_SiC#29_SHC_post2.PDF
  482   Monday, September 30, 2013, 16:04 Bevan MossConditioning StationRepairTM3NiO#2Ion gauge filament replacement

 The filament on IG1 for TM3 was replaced by myself and Edi. The old filament did not appear to be damaged but the controller claimed it was. The current gauge on TM3 is a single filament, ideally it would be upgraded to a double.

  481   Monday, September 30, 2013, 16:02 Bevan MossConditioning StationStandard OperationTM3NiO#2Vacuum

 The roughing pump for TCS was started at ~3:00 and the turbo's turned on at ~4:00. This is the minimum time for the module to pump down.

  480   Monday, September 30, 2013, 16:00 Bevan MossSouth Hot-CellStandard OperationTM3NiO#2Electrical check

Travis Cave wrote:

Results of the various electrical checks done on TM#3 while at the hot cell since the 27th of September. View PDF for the details.

 There was also a leak check. There was some water remaining in the HS line but once removed the leak cart pumped down to 0.0 x 10-4 Torr and the leak tester was at the bottom limit. There was no response when sprayed with helium

  479   Monday, September 30, 2013, 15:59 Bevan MossSouth Hot-CellStandard OperationTM4SiC#29Module move from SW Silo to SHC

 TM4 was move by myself and Grant. Move went smoothly, no hang ups.

  478   Monday, September 30, 2013, 14:00 Travis CaveConditioning StationStandard OperationTM3NiO#2module move

TM#3 with NiO#2 has been move from south hot cell to the conditioning station.

  477   Monday, September 30, 2013, 12:23 Travis CaveSouth Hot-CellStandard OperationTM3NiO#2Electrical check

Results of the various electrical checks done on TM#3 while at the hot cell since the 27th of September. View PDF for the details.

Attachment 1: TM#3_elec_chk_NiO#2_SHC2.PDF
TM#3_elec_chk_NiO#2_SHC2.PDF TM#3_elec_chk_NiO#2_SHC2.PDF TM#3_elec_chk_NiO#2_SHC2.PDF
  476   Friday, September 27, 2013, 12:53 Bevan MossConditioning StationStandard OperationTM3NiO#2HP TGHT and TBHT conditioning

 Yesterday TM3 had the TGHT current raised to 160 amps and the TBHT raised to 200 amps. At this power the TGHT showed typical results in that vacuum was affected and the increased voltage as the target increased in temperature. The tube heater on the other hand did not change voltage or show a change in vacuum. As a result it is believe there is a short somewhere in the module. I have attached a screen shot of the strip tool and the raw data.

Attachment 1: 26092013_TM3_NiO#2_HP_testing.JPG
26092013_TM3_NiO#2_HP_testing.JPG
Attachment 2: 26092013_TM3_NiO2_HP_testing.txt
Time	TCS:IG2:RDVAC [T]	TCS:IG1S:RDVAC [T]	TCS:TGHT:RDVOL [V]	TCS:TGHT:RDCUR [A]	TCS:TBHT:RDVOL [V]	TCS:TBHT:RDCUR [A]	
09/26/2013 13:56:46.578505	                      0.000008241                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:47.577921	                      0.000008263                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:48.577796	                      0.000008241                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:49.578660	                      0.000008241                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:50.578515	                      0.000008241                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:51.579406	                      0.000008241                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:52.579315	                      0.000008241                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:53.634912	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:54.635006	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:55.634932	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:56.635812	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:57.636661	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:58.636560	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:56:59.636386	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:00.636250	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:01.637154	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:02.637232	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:03.636895	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:04.636752	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:05.636626	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:06.636503	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:07.636378	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:08.636282	                      0.000008263                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:09.636155	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:10.635997	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:11.635868	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:12.635748	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:13.635872	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:14.636574	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:15.636412	                      0.000008263                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:16.636294	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:17.636120	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:18.635997	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:19.636923	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:20.636746	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:21.668993	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:22.714260	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:23.714359	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:24.714296	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:25.714103	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:26.714040	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:27.713860	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:28.714757	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:29.714771	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:30.714484	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:31.715358	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:32.716275	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:33.716123	                      0.000008263                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:34.717000	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:35.716854	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:36.717731	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:37.717602	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:38.718299	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:39.718395	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:40.718234	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:41.718109	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:42.718043	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:43.717868	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:44.717782	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:45.717687	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:46.718498	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:47.718410	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:48.717910	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:49.717196	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:50.716999	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:51.717976	                      0.000008263                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:52.717775	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:53.717737	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:54.718503	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:55.718379	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:56.718284	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:57.718371	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:58.718558	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:57:59.718850	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:00.718981	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:01.719618	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:02.719497	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:03.719359	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:04.719222	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:05.719118	                      0.000008241                      0.000006923                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:06.719000	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:07.718868	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:08.719780	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:09.719649	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:10.719473	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:11.720399	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:12.720219	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:13.720537	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:14.720105	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:15.719968	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:16.719928	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:17.719609	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:18.719466	                      0.000008241                      0.000006857                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:19.719357	                      0.000008263                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:20.719250	                      0.000008263                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:21.720139	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:22.719967	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:23.719864	                      0.000008263                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:24.719740	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:25.719641	                      0.000008241                      0.000006901                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:26.720504	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:27.720334	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:28.808036	                      0.000008241                      0.000006879                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:29.808223	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:30.807982	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:31.807990	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:32.807723	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:33.807576	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:34.808451	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:35.808364	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:36.808305	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:37.808311	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:38.807950	                      0.000008241                      0.000006879                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:39.807853	                      0.000008263                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:40.807789	                      0.000008263                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:41.808598	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:42.808442	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:43.807745	                      0.000008241                      0.000006901                             0.01                           BadVal                           BadVal                           BadVal
09/26/2013 13:58:44.807218	                      0.000008241                      0.000006879                             0.01                             3.36                           BadVal                           BadVal
09/26/2013 13:58:45.807185	                      0.000008241                      0.000006901                             0.01                             3.51                           BadVal                           BadVal
09/26/2013 13:58:46.807942	                      0.000008241                      0.000006901                             0.01                             3.51                           BadVal                           BadVal
09/26/2013 13:58:47.807835	                      0.000008241                      0.000006901                             0.01                             3.51                           BadVal                           BadVal
09/26/2013 13:58:48.807734	                      0.000008241                      0.000006901                             0.01                             3.51                           BadVal                           BadVal
09/26/2013 13:58:49.807600	                      0.000008241                      0.000006901                             0.01                             3.51                           BadVal                           BadVal
09/26/2013 13:58:50.807446	                      0.000008241                      0.000006901                             0.01                             3.51                           BadVal                           BadVal
09/26/2013 13:58:51.808335	                      0.000008263                      0.000006901                             0.01                             3.36                           BadVal                           BadVal
09/26/2013 13:58:52.808409	                      0.000008241                      0.000006901                             0.01                             3.36                           BadVal                           BadVal
09/26/2013 13:58:53.808264	                      0.000008263                      0.000006901                             0.01                             3.51                           BadVal                           BadVal
09/26/2013 13:58:54.809105	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                           BadVal
09/26/2013 13:58:55.808874	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                           BadVal
09/26/2013 13:58:56.809685	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                           BadVal
09/26/2013 13:58:57.809598	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                           BadVal
09/26/2013 13:58:58.809481	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                           BadVal
09/26/2013 13:58:59.809374	                      0.000008219                      0.000006901                             0.01                             3.36                             0.01                           BadVal
09/26/2013 13:59:00.809218	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:01.809059	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:02.808942	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:03.808893	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:04.808683	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:05.808558	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:06.809426	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:07.809384	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:08.809184	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:09.809055	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:10.808936	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:11.808804	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:12.808117	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:13.807562	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:14.807437	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:15.807335	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:16.807178	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:17.807061	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:18.806936	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:19.806850	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:20.806758	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:21.807575	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:22.807434	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:23.807343	                      0.000008263                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:24.807219	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:25.808051	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:26.807928	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:27.807840	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:28.807693	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:29.807579	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:30.807486	                      0.000008263                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:31.808408	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:32.808183	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:33.808055	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:34.807923	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:35.814752	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:36.814785	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:37.857316	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:38.906460	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:39.906791	                      0.000008263                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:40.907200	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:41.907040	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:43.54114	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:44.54762	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:45.54866	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:46.55513	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:47.55541	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:48.56263	                      0.000008241                      0.000006857                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:49.56250	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:50.56068	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:51.56089	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:52.56015	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:53.56641	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:54.56536	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:55.56409	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:56.57287	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:57.57141	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 13:59:58.57008	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 13:59:59.56962	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:00.56794	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:01.57631	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:02.57507	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:03.57404	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:04.57264	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:05.57140	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:06.57187	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:07.56888	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:08.57796	                      0.000008263                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:09.57676	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:10.57505	                      0.000008241                      0.000006879                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:11.58416	                      0.000008263                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:12.58402	                      0.000008263                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:13.58264	                      0.000008263                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:14.59001	                      0.000008263                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:15.58444	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:16.58774	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:17.58778	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:18.58509	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:19.58385	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:20.58280	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:21.59151	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:22.59007	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:23.59082	                      0.000008263                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:24.59789	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:25.60718	                      0.000008241                      0.000006901                             0.01                             3.51                             0.01                             0.80
09/26/2013 14:00:26.61742	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:27.61382	                      0.000008241                      0.000006857                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:28.61834	                      0.000008241                      0.000006857                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:29.61207	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:30.61995	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:31.62910	                      0.000008263                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:33.133357	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:34.133490	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:36.60493	                      0.000008263                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:37.60166	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:38.60002	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:39.59878	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:40.60744	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:41.60649	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:42.60501	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:43.60366	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:44.60273	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:45.60145	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:46.60000	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:47.59995	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:48.59750	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:49.59805	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:50.59523	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:51.59371	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:52.59286	                      0.000008263                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:53.59148	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:54.59093	                      0.000008263                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:55.58909	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:56.58815	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:57.59642	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:58.59530	                      0.000008263                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:00:59.59655	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:00.60281	                      0.000008263                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:01.60159	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:02.59986	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:03.59868	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:04.59736	                      0.000008241                      0.000006857                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:05.59636	                      0.000008241                      0.000006857                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:06.60509	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:07.60389	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:08.60242	                      0.000008241                      0.000006923                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:09.60171	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:10.61014	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:11.60867	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:12.60792	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:13.60652	                      0.000008263                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:14.60564	                      0.000008241                      0.000006857                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:15.60381	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:16.60265	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:17.59410	                      0.000008241                      0.000006879                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:18.58988	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:19.58888	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:20.58998	                      0.000008263                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:21.58677	                      0.000008241                      0.000006901                             0.01                             3.36                             0.01                             0.80
09/26/2013 14:01:22.59507	                      0.000008241                      0.000006879                             0.01                             3.36                             0.08                             8.21
09/26/2013 14:01:23.59361	                      0.000008241                      0.000006879                             0.01                             3.36                             0.08                             8.54
09/26/2013 14:01:24.59276	                      0.000008263                      0.000006901                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:25.60197	                      0.000008241                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:26.60323	                      0.000008263                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:27.60873	                      0.000008263                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:28.60753	                      0.000008263                      0.000006879                             0.01                             3.36                             0.07                             8.54
09/26/2013 14:01:29.60689	                      0.000008241                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:30.60484	                      0.000008241                      0.000006879                             0.01                             3.36                             0.07                             8.54
09/26/2013 14:01:31.61383	                      0.000008241                      0.000006879                             0.01                             3.36                             0.07                             8.54
09/26/2013 14:01:32.61284	                      0.000008263                      0.000006879                             0.01                             3.36                             0.07                             8.54
09/26/2013 14:01:33.61108	                      0.000008263                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:34.61997	                      0.000008263                      0.000006901                             0.01                             3.36                             0.08                             8.54
09/26/2013 14:01:35.62848	                      0.000008263                      0.000006901                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:36.62787	                      0.000008241                      0.000006901                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:37.62638	                      0.000008219                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:38.62557	                      0.000008263                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:39.62356	                      0.000008263                      0.000006901                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:40.62272	                      0.000008241                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:41.62109	                      0.000008241                      0.000006879                             0.01                             3.36                             0.07                             8.54
09/26/2013 14:01:42.61973	                      0.000008241                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:43.61895	                      0.000008263                      0.000006879                             0.01                             3.36                             0.08                             8.54
09/26/2013 14:01:44.61768	                      0.000008241                      0.000006901                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:45.62636	                      0.000008219                      0.000006879                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:46.62506	                      0.000008263                      0.000006945                             0.01                             3.36                             0.09                             8.54
09/26/2013 14:01:47.62451	                      0.000008241                      0.000006879                             0.01                             3.36                             0.09                             8.54
... 7809 more lines ...
  475   Friday, September 27, 2013, 10:54 Travis CaveConditioning StationStandard OperationTM3NiO#2module move

TM#3 with NiO#2 has been moved to the south hot cell from the conditioning station.

  474   Thursday, September 26, 2013, 15:25 Bevan MossConditioning StationStandard OperationTM3NiO#2HV conditioning

Bevan Moss wrote:

 The first attempt at HV conditioning of TM3 with the new source tray and NiO#2 target installed. Conditioning so far has been very slow, with the current increasing quickly to a maximum of 400uA. Voltage in these cases was lowered until the current was below 200uA. Attached is the raw data. It appears the source tray and target are conditioning just there is either something burning off or there is localized poor vacuum.

 Attached screen shot. Vacuum was around 8.3 X 10^-6 on IG2 (which is in the diagnostics box)

Attachment 1: TM3_NiO#2_Frist_Attempt_HV_conditioning_Bias_and_El_V_and_I.JPG
TM3_NiO#2_Frist_Attempt_HV_conditioning_Bias_and_El_V_and_I.JPG
  473   Thursday, September 26, 2013, 15:00 Bevan MossConditioning StationStandard OperationTM3NiO#2RGA not connecting

 Following the helium pressure test Don attempted to connect to the RGA via the network. He was unable to do so. This scan was desired so a pre and post water scan could be compared to indicate if there was any leaking. It has been decided to carry on as the HV and HP tests are of greater importance. Don will continue to investigate the RGA. Travis inspected from the conditioning station roof and determined that the RGA is getting power.

  472   Thursday, September 26, 2013, 14:55 Bevan MossConditioning StationStandard OperationTM3NiO#2HV conditioning

 The first attempt at HV conditioning of TM3 with the new source tray and NiO#2 target installed. Conditioning so far has been very slow, with the current increasing quickly to a maximum of 400uA. Voltage in these cases was lowered until the current was below 200uA. Attached is the raw data. It appears the source tray and target are conditioning just there is either something burning off or there is localized poor vacuum.

Attachment 1: 26092013_TM3_NiO2_HV_conditioning_first_attempt.txt
Time	TCS:DIC1:RDCOND [Mohm]	TCS:IG2:RDVAC [T]	TCS:IG1S:RDVAC [T]	TCS:BIAS:RDVOL [V]	TCS:BIAS:RDCUR [uA]	TCS:EL:RDVOL [V]	TCS:EL:RDCUR [mA]	
09/26/2013 12:33:22.798401	                              5.0                      0.000008351                      0.000007054                         74042.11                             0.00                           BadVal                           BadVal
09/26/2013 12:33:23.798571	                              5.0                      0.000008351                      0.000007054                         74042.11                             0.00                           BadVal                           BadVal
09/26/2013 12:33:24.798174	                              5.0                      0.000008351                      0.000007054                         74042.11                             0.00                           BadVal                           BadVal
09/26/2013 12:33:25.798026	                              5.0                      0.000008351                      0.000007076                         74042.11                             0.00                           BadVal                           BadVal
09/26/2013 12:33:26.797921	                              5.0                      0.000008329                      0.000007054                         74042.11                             0.00                           BadVal                           BadVal
09/26/2013 12:33:27.797998	                              5.0                      0.000008351                      0.000007076                         74042.11                             0.00                           BadVal                           BadVal
09/26/2013 12:33:28.798672	                              5.0                      0.000008351                      0.000007076                         74042.11                             0.00                           BadVal                           BadVal
09/26/2013 12:33:29.799554	                              5.0                      0.000008351                      0.000007076                         74042.11                             0.00                           BadVal                           BadVal
09/26/2013 12:33:30.799423	                              5.0                      0.000008373                      0.000007076                         74042.11                             0.00                           BadVal                           BadVal
09/26/2013 12:33:31.799291	                              5.0                      0.000008351                      0.000007098                         74042.11                             0.00                             6.10                           BadVal
09/26/2013 12:33:32.799147	                              5.0                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                           BadVal
09/26/2013 12:33:33.799077	                              5.0                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                           BadVal
09/26/2013 12:33:34.798923	                              5.0                      0.000008373                      0.000007076                         74042.11                             0.00                             6.10                           BadVal
09/26/2013 12:33:35.855344	                              5.0                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                           BadVal
09/26/2013 12:33:36.855698	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:37.855558	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:38.856379	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:39.856500	                              5.1                      0.000008373                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:40.856215	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:41.856012	                              5.1                      0.000008373                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:42.855878	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:43.855856	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:44.855636	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:45.855511	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:46.855439	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:47.856491	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:48.856270	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:49.856034	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:50.856953	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:51.856801	                              5.1                      0.000008329                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:52.857714	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:53.857523	                              5.1                      0.000008373                      0.000007098                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:54.857498	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:55.857555	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:56.857130	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:57.857091	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:58.856941	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:33:59.856797	                              5.1                      0.000008329                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:00.856620	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:01.856523	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:02.856463	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:03.857270	                              5.1                      0.000008373                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:04.857123	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:05.857083	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:06.856878	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:07.857738	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:08.857668	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:09.857674	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:10.857375	                              5.1                      0.000008373                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:11.857251	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:12.858195	                              5.1                      0.000008373                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:13.858066	                              5.1                      0.000008373                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:14.857889	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:15.857952	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:16.857924	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:17.858515	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:18.858375	                              5.1                      0.000008373                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:19.858249	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:20.858134	                              5.1                      0.000008373                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:21.858037	                              5.1                      0.000008373                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:22.858889	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:23.859048	                              5.1                      0.000008329                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:24.859688	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:25.859494	                              5.1                      0.000008373                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:26.859368	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:27.860282	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:28.860158	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:29.860203	                              5.1                      0.000008329                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:30.859858	                              5.1                      0.000008329                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:31.859761	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:32.860674	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:33.860533	                              5.1                      0.000008373                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:34.860365	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:35.861264	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:36.861210	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:37.861126	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:38.861879	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:39.861739	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:40.861700	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:41.861528	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:42.862380	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:43.862277	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:44.862112	                              5.1                      0.000008373                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:45.862983	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:46.863148	                              5.2                      0.000008373                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:47.862822	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:48.862650	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:49.862510	                              5.1                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:50.862441	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:51.862255	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:52.862185	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:53.861987	                              5.1                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:54.861930	                              5.2                      0.000008373                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:55.861816	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:56.861631	                              5.2                      0.000008351                      0.000007032                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:57.862574	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:58.862592	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:34:59.863332	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:00.863126	                              5.2                      0.000008373                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:01.862988	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:02.864119	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:03.863759	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:04.863623	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:05.863474	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:06.863355	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:07.864247	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:08.865174	                              5.2                      0.000008329                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:09.864977	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:10.864890	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:11.864811	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:12.865594	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:13.865522	                              5.2                      0.000008373                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:14.865352	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:15.865394	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:16.865099	                              5.2                      0.000008329                      0.000007032                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:17.865215	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:18.864848	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:19.864765	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:20.865638	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:21.865502	                              5.2                      0.000008373                      0.000007032                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:22.866509	                              5.2                      0.000008373                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:23.866222	                              5.2                      0.000008373                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:24.866090	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:25.866012	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:26.865870	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:27.866737	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:28.866607	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:29.866473	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:30.866467	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:31.866478	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:32.867142	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:33.866959	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:34.867842	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:35.867961	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:36.868791	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:37.869662	                              5.2                      0.000008373                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:38.869339	                              5.2                      0.000008351                      0.000007098                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:39.869215	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:40.869132	                              5.2                      0.000008329                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:41.869018	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:42.868886	                              5.2                      0.000008329                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:43.868751	                              5.2                      0.000008329                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:44.868686	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:45.868470	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:46.868337	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:47.869278	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:48.869086	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:49.869006	                              5.2                      0.000008351                      0.000007098                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:50.868862	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:51.868752	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:52.869658	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:53.869537	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:54.869355	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:55.869223	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:56.869139	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:57.869989	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:58.869918	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:35:59.869723	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:00.869582	                              5.2                      0.000008373                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:01.869474	                              5.2                      0.000008329                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:02.869434	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:03.869240	                              5.2                      0.000008373                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:04.869079	                              5.2                      0.000008329                      0.000007098                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:05.868954	                              5.2                      0.000008329                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:06.868870	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:07.869737	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:08.869680	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:09.870454	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:10.871370	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:11.871230	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:12.872120	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:13.872047	                              5.2                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:14.871843	                              5.2                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:15.871761	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:16.872567	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:17.872536	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:18.872446	                              5.3                      0.000008351                      0.000007098                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:19.872215	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:20.872112	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:21.871983	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:22.872841	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:23.872736	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:24.873591	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:25.873620	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:26.874339	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:27.875187	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:28.875227	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:29.875940	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:30.875881	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:31.875697	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:32.876587	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:33.876467	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:34.876309	                              5.3                      0.000008329                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:35.876204	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:36.876081	                              5.3                      0.000008351                      0.000007098                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:37.876934	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:38.876816	                              5.3                      0.000008351                      0.000007032                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:39.876759	                              5.3                      0.000008351                      0.000007032                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:40.877645	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:41.877482	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:42.877457	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:43.878271	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:44.878065	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:45.877956	                              5.3                      0.000008329                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:46.877868	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:47.878910	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:48.878625	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:49.878437	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:50.878412	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:51.878179	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:52.879080	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:53.878974	                              5.3                      0.000008329                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:54.878810	                              5.3                      0.000008329                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:55.879704	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:56.879708	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:57.880427	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:58.880416	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:36:59.880178	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:00.880064	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:01.879972	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:02.879840	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:03.879740	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:04.879558	                              5.3                      0.000008329                      0.000007032                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:05.879432	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:06.879299	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:07.880472	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:08.880082	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:09.879993	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:10.879795	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:11.879733	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:12.880595	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:13.880554	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:14.880302	                              5.3                      0.000008329                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:15.880194	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:16.880051	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:17.880059	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:18.879840	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:19.879756	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:20.879609	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:21.880474	                              5.3                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:22.881316	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:23.881346	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:24.882077	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:25.881953	                              5.3                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:26.881837	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:27.881700	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:28.881548	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:29.881443	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:30.881370	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:31.881170	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:32.882063	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:33.881942	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:34.881794	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:35.881669	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:36.881536	                              5.4                      0.000008373                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:37.881446	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:38.882284	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:39.882168	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:40.882060	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:41.881991	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:42.882848	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:43.882706	                              5.4                      0.000008329                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:44.882540	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:45.882415	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:46.882435	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:47.883305	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:48.883152	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:49.883946	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:50.883788	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:51.883868	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:52.884570	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:53.884434	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:54.884309	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:55.885284	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:56.885053	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:57.885972	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:58.885785	                              5.4                      0.000008351                      0.000007076                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:37:59.885690	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:38:00.885587	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:38:01.885460	                              5.4                      0.000008351                      0.000007032                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:38:02.885458	                              5.4                      0.000008351                      0.000007032                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:38:03.886263	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:38:04.886033	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:38:05.886081	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:38:06.886768	                              5.4                      0.000008351                      0.000007054                         74042.11                             0.00                             6.10                             0.01
09/26/2013 12:38:07.886717	                              5.4                      0.000008351                      0.000007054                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 12:38:08.887522	                              5.4                      0.000008351                      0.000007076                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 12:38:09.887405	                              5.4                      0.000008329                      0.000007076                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 12:38:10.887279	                              5.4                      0.000008329                      0.000007054                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 12:38:11.887175	                              5.4                      0.000008351                      0.000007076                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 12:38:12.887075	                              5.4                      0.000008351                      0.000007076                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 12:38:13.887092	                              5.4                      0.000008329                      0.000007076                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 12:38:14.886777	                              5.4                      0.000008351                      0.000007076                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 12:38:15.886703	                              5.4                      0.000008329                      0.000007076                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 12:38:16.886519	                              5.4                      0.000008329                      0.000007076                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 12:38:17.887434	                              5.4                      0.000008329                      0.000007076                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 12:38:18.887300	                              5.4                      0.000008351                      0.000007054                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 12:38:19.887151	                              5.4                      0.000008351                      0.000007076                           BadVal                           BadVal                           BadVal                           BadVal
09/26/2013 12:38:20.887176	                              5.4                      0.000008351                      0.000007076                           BadVal                           BadVal                           BadVal                           BadVal
... 4493 more lines ...
  471   Thursday, September 26, 2013, 14:47 Bevan MossConditioning StationMaintenanceTM3NiO#2Water pumps turned on

 The water pumps for the conditioning station were turned on at ~11:45. It took longer to start the conditioning station as the air from the leak test had to be bled off and the make up water tank refilled. The valves on the water supply line were closed slightly to make up for the fact the water cooled power supply has been removed. DIF2 was installed by Maico so we can measure the flow in the primary side.

  470   Thursday, September 26, 2013, 14:42 Bevan MossSouth Hot-CellRepairTM3NiO#2Activities September 24 - 25

Bevan Moss wrote:

September 24

 After the manipulator repair Maico installed the NiO#2 onto the target module. Once connected I completed a leak test on the HS line with Isaac. The leak testing cart would not pump down originally (something wrong with the cart) but eventually it did start to work. The cart pumped to its lower limits and did not respond even with the module was flooded with helium. A electrical test was then conducted by myself and and Don Jackson (see attached). During the standard test the Target to Anode and to 60kV common read lower than expected at ~47Momh @ 1kV. Also of note during the conductance check the measurement from c-d was low at 1.82 mohm.

3  additional tests were completed 2 of which were coil to common (both sides). The resisance was measure at 47 ohm with the megger and 1.32 Kohm with the uohmmeter. The coil lines lines were also measured to get and indication of the resistance of a copper line. These lines were measured while they were not jumpered both the coil + and coil - measured 5.73 mohm.

A meeting with Lia and Friedhelm was held and and the decision was to move the module and test it in the conditioning station. Fingers crossed that the module will still work.

Sept 25th

Maico finished installing the cover today for the module to be moved. Other elogs cover the other activities for today.

 Also of note the turbo pumps were started on the conditioning station around 1:30. Don and Travis completed and electrical test, see other elogs. When attempting to start IG1 the controller indicated there was no filament, this gauge only has 1 to start. It will have to be replaced when the module is next vented. In addition to this it appear the PNG1 controller has failed again. I have emailed Dimo to see if we can get one to test. The img100 is new and different cables have been used. All 4 options for connections were also tested in case of a wiring mistake.

  469   Thursday, September 26, 2013, 14:37 Bevan MossConditioning StationStandard OperationTM3NiO#2Leak check / Pressure Test

maicodallavalle wrote:

TM3 has had all of the water lines pressurized with Helium and the Helium Leak detector was connected to the turbo pumps. There were no leaks detected.

 While pressurizing the lines it was found some of the lines had water in them. It is unknown where this water came from. Perhaps David connected the conditioning station last time it was in? Lines that were free of water were A, B, C, and window The leak detector pumped down to 4.0 x 10^-3 Torr and the leak rate was 1.4 x 10^-9 atm-cc/sec, there was no response during the leak test and the leak rate continued to decrease during the test.

ELOG V2.9.2-2455