|
ID |
Date |
Author |
Target Material |
Target Oven W.O# |
Source |
Status |
Subject |
|
270
|
Monday, January 11, 2016, 15:47 |
Laura Lambert | SiC | 41077-3 | LP-SIS | Done | Conditioning SiC test target for test stand measurements | Monday, January 11, 2016, 15:25
Small center block LP-SIS container previously sintered with TaC coating
3x Ta heat shields
Container is loaded with 260 discs of SiC with a total length of 8.5cm
Loaded into evap2 and pumped down this morning
HS water valves opened up, no leaks detected
Tapped up manually to 30A
Vacuum stable
IGP1 = 1e-6 Torr
Initiating auto-ramp at 0.2A/min and 0.1A/min for the TGHT and TBHT respectively
Final setpoints: 400A for the TGHT and 230A for the TBHT
Final desired temperature: 1700C
Tuesday, January 12, 2016, 12:55
TGHT = 138A, TBHT = 80A
IGP1 = 7.7e-6 Torr.
Vacuum stable and coming down.
Increasing ramping rate to 0.5A/min and 0.2A/min for the TGHT and TBHT respectively.
Wednesday, January 13, 2016, 09:32
TGHT reached 400A ~11.5hrs ago
TBHT reached 230A ~8hrs ago
OMEGA = 1400C
IGP1 = 1.75e-7 Torr
Tapping up manually to 1700C...
Wednesday, January 13, 2016, 10:14
Previous temperature measurement of OMEGA = 1400C is wrong due to window port being coated with SiC [see attached photos]
Will remain at TGHT = 420A for 4hrs then cool down this afternoon
Wednesday, January 13, 2016, 14:29
Remained at max setpoints for 4.5hrs.
Initiating auto-cool at 2.0A/min and 1.0A/min for the TGHT and TBHT respectively.
Thursday, January 14, 2016, 09:47
Vented in air and opened up this morning
Target looks good
Will insert and expand endcaps this afternoon
|
Attachment 1: Photo_2016-0-13.jpg
|
|
Attachment 2: Photo_2016-01-13_#2.jpg
|
|
Attachment 3: Photo_2016-01-13_#3.jpg
|
|
Attachment 4: Photo_2016-01-13_#4.jpg
|
|
|
269
|
Friday, January 08, 2016, 11:15 |
Laura Lambert | Ta | 41369-5 | HP-SIS | Done | Temperature measurements |
Aurelia Laxdal wrote: |
Temperature measurements in progress this week for Ta HP-SIS.
|
Friday, January 08, 2016, 10:49
Rh foil added.
endcaps inserted and expanded.
total length of target with end caps = 3.45cm
spot weld one thermocouple onto center block and one thermocouple onto target leg [left side wrt ionizer tube into page].
see elog id# 268 for more info regarding measurements taken and state of TBHT power supply...
|
|
268
|
Thursday, January 07, 2016, 09:19 |
Aurelia Laxdal | Ta | 41369-5 | HP-SIS | In progress | Temperature measurements | Temperature measurements in progress this week for Ta HP-SIS.
See attachment for the current settings and rampings of the TGHT/TBHT for the temperature measurements taken on Jan. 9-10, 2016 .
Note: TBHT PS tripped off on Jan. 9, 2016 (see first page of the attached file) and couldn't turn it ON again over the week-end. |
Attachment 1: Jan_9_and_10_2016_measurements.docx
|
|
267
|
Wednesday, December 16, 2015, 10:22 |
Laura Lambert | Ta | 41369-5 | HP-SIS | Done | Conditioning back-up Ta Target & emissivity measurements | Wednesday, December 16, 2015, 10:09
Target consists of 505 discs.
Total packed length 3.4cm.
*same parameters at Ta#46*
Total mass of target materials 53.9381g [4 packs of 20 with Ta wire + 425 single discs].
3 x Ta Heat shields.
No Re foil inserted.
No endcaps inserted.
Chamber pumped down overnight.
IGP1 = 5.1e-7 Torr.
Tapped up manually to 35A.
IGP1 = 5.2e-7 Torr.
Vacuum stable, initiating auto-ramp.
TBHT will be ramped up to 230A at 1.0A/min.
TGHT will be ramped up to 1800C, auto ramp to 600A at 2.0A/min [estimated final value of 720A according to elog ID 239]
Wednesday, December 16, 2015, 15:08
TBHT reached max setpoint of 230A, 3.3V
TGHT = 600A, 8.9V
IGP1 = 1.1e-6 Torr.
OMEGA = 1543C
Increasing final setpoint of TGHT to 700A and re-initiating auto-ramp.
Wednesday, December 16, 2015, 16:02
TGHT = 700A, 10.6V
IGP1 = 2.1e-6 Torr.
OMEGA = 1740C
Tapping up manually to 1800C.
Wednesday, December 16, 2015, 16:28
TGHT = 730A, 11.2V
Vacuum is coming down from IGP1 = 2.6e-6 Torr.
OMEGA = 1800C
Will condition the target at 1800C overnight and ramp down in the morning.
Thursday, December 17, 2015, 09:31
OMEGA = 1802C on right side
Temp on leg of left side = 2150C, the hottest point - measured with the Mikron pyrometer.
Friday, December 18, 2015, 08:53
Evap2 remained at 1800C for ~26.5hrs
Auto-ramped down to TGHT = 300A and TBHT = 90A at a rate of 2.0A/min and 1.0A/min respectively.
Conditioning complete.
TGHT will be ramped up later today for emissivity measurements.
Monday, January 04, 2016, 11:21
the following was performed by aurelia.
emissivity measurements were taken late december.
evap2 was cooled down and left under vacuum over the xmas break.
currently ramping back up for more emissivity measurements.
Friday, January 08, 2016, 10:47
Aurelia done measurements.
vented and opened this morning.
|
|
266
|
Monday, December 07, 2015, 09:06 |
Laura Lambert | Empty | 39530-9 | HP-FEBIAD | Done | emissivity measurements with no h.s on HP-FEBIAD container - round2 | Monday, December 07, 2015, 09:04
No heat shields.
Container empty.
IGP1 = 1.1e-7 Torr.
Slope of OMEGA = 0.98
Tapped up manually to 35A.
Vacuum stable, initiating auto-ramp at 2.0A/min and 1.0A/min for the TGHT and TBHT respectively.
Final setpoints: TBHT = 90A, TGHT = 360A
Will tap up with the same parameters as last round to confirm results.
Friday, December 11, 2015, 11:44
Took out of evap2 yesterday.
Aurelia took many emissivity measurements to compare with previous data - which matched up.
The dip in emissivity before going up again around ~560A could be due to oxidation of the materials.
|
|
265
|
Tuesday, December 01, 2015, 11:35 |
Anders Mjos | Empty | 41077-5 | LP-SIS | In progress | Test#2: Measuring temp with thermocouple (away from centre) & 2C-IR | Data from thermocouple attached. Raw data and Excel sheet with plot. |
Attachment 1: 2015-10-08_Small_Centre_Block.xlsx
|
Attachment 2: 2015-10-08.4ch
|
°C Date Time T1 - C T2 - C T3 - C T4 - K
1 2015-10-08 14:35:51 25.0
2 2015-10-08 14:36:21 25.7
3 2015-10-08 14:36:51 26.1
4 2015-10-08 14:37:21 26.9
5 2015-10-08 14:37:51 28.0
6 2015-10-08 14:38:21 28.6
7 2015-10-08 14:38:51 29.7
8 2015-10-08 14:39:21 30.8
9 2015-10-08 14:39:51 32.5
10 2015-10-08 14:40:21 34.4
11 2015-10-08 14:40:51 36.3
12 2015-10-08 14:41:21 39.0
13 2015-10-08 14:41:51 42.5
14 2015-10-08 14:42:21 46.4
15 2015-10-08 14:42:51 51.3
16 2015-10-08 14:43:21 57.3
17 2015-10-08 14:43:51 64.7
18 2015-10-08 14:44:21 72.1
19 2015-10-08 14:44:51 82.1
20 2015-10-08 14:45:21 91.5
21 2015-10-08 14:45:51 104.3
22 2015-10-08 14:46:21 116.5
23 2015-10-08 14:46:51 133.1
24 2015-10-08 14:47:21 151.2
25 2015-10-08 14:47:51 167.8
26 2015-10-08 14:48:21 188.5
27 2015-10-08 14:48:51 207.7
28 2015-10-08 14:49:21 228.1
29 2015-10-08 14:49:51 251.5
30 2015-10-08 14:50:21 275.9
31 2015-10-08 14:50:51 297.5
32 2015-10-08 14:51:21 319.8
33 2015-10-08 14:51:51 343.1
34 2015-10-08 14:52:21 366.5
35 2015-10-08 14:52:51 388.6
36 2015-10-08 14:53:21 410.5
37 2015-10-08 14:53:51 432.4
38 2015-10-08 14:54:21 452.0
39 2015-10-08 14:54:51 471.5
40 2015-10-08 14:55:21 491.6
41 2015-10-08 14:55:51 507.5
42 2015-10-08 14:56:21 523.7
43 2015-10-08 14:56:51 539.1
44 2015-10-08 14:57:21 554.6
45 2015-10-08 14:57:51 567.7
46 2015-10-08 14:58:21 581.1
47 2015-10-08 14:58:51 593.0
48 2015-10-08 14:59:21 607.7
49 2015-10-08 14:59:51 623.0
50 2015-10-08 15:00:21 638.0
51 2015-10-08 15:00:51 650.2
52 2015-10-08 15:01:21 661.4
53 2015-10-08 15:01:51 670.3
54 2015-10-08 15:02:21 678.0
55 2015-10-08 15:02:51 683.8
56 2015-10-08 15:03:21 688.8
57 2015-10-08 15:03:51 693.0
58 2015-10-08 15:04:21 697.2
59 2015-10-08 15:04:51 700.7
60 2015-10-08 15:05:21 703.5
61 2015-10-08 15:05:51 706.3
62 2015-10-08 15:06:21 709.1
63 2015-10-08 15:06:51 711.5
64 2015-10-08 15:07:21 713.7
65 2015-10-08 15:07:51 716.1
66 2015-10-08 15:08:21 718.3
67 2015-10-08 15:08:51 720.5
68 2015-10-08 15:09:21 722.5
69 2015-10-08 15:09:51 724.7
70 2015-10-08 15:10:21 726.8
71 2015-10-08 15:10:51 728.8
72 2015-10-08 15:11:21 730.8
73 2015-10-08 15:11:51 733.0
74 2015-10-08 15:12:21 734.7
75 2015-10-08 15:12:51 736.9
76 2015-10-08 15:13:21 738.8
77 2015-10-08 15:13:51 740.8
78 2015-10-08 15:14:21 742.7
79 2015-10-08 15:14:51 744.7
80 2015-10-08 15:15:21 746.6
81 2015-10-08 15:15:51 748.6
82 2015-10-08 15:16:21 750.6
83 2015-10-08 15:16:51 752.5
84 2015-10-08 15:17:21 754.4
85 2015-10-08 15:17:51 756.3
86 2015-10-08 15:18:21 758.3
87 2015-10-08 15:18:51 760.2
88 2015-10-08 15:19:21 762.2
89 2015-10-08 15:19:51 764.1
90 2015-10-08 15:20:21 765.8
91 2015-10-08 15:20:51 767.8
92 2015-10-08 15:21:21 769.9
93 2015-10-08 15:21:51 771.9
94 2015-10-08 15:22:21 773.8
95 2015-10-08 15:22:51 775.5
96 2015-10-08 15:23:21 777.3
97 2015-10-08 15:23:51 779.3
98 2015-10-08 15:24:21 781.3
99 2015-10-08 15:24:51 783.1
100 2015-10-08 15:25:21 785.0
101 2015-10-08 15:25:51 786.9
102 2015-10-08 15:26:21 788.8
103 2015-10-08 15:26:51 790.6
104 2015-10-08 15:27:21 792.6
105 2015-10-08 15:27:51 794.5
106 2015-10-08 15:28:21 796.3
107 2015-10-08 15:28:51 798.3
108 2015-10-08 15:29:21 800.1
109 2015-10-08 15:29:51 802.0
110 2015-10-08 15:30:21 803.8
111 2015-10-08 15:30:51 805.8
112 2015-10-08 15:31:21 807.8
113 2015-10-08 15:31:51 809.6
114 2015-10-08 15:32:21 811.5
115 2015-10-08 15:32:51 813.3
116 2015-10-08 15:33:21 815.3
117 2015-10-08 15:33:51 817.2
118 2015-10-08 15:34:21 819.1
119 2015-10-08 15:34:51 820.8
120 2015-10-08 15:35:21 822.8
121 2015-10-08 15:35:51 824.6
122 2015-10-08 15:36:21 826.5
123 2015-10-08 15:36:51 828.3
124 2015-10-08 15:37:21 830.3
125 2015-10-08 15:37:51 832.1
126 2015-10-08 15:38:21 834.0
127 2015-10-08 15:38:51 835.8
128 2015-10-08 15:39:21 837.7
129 2015-10-08 15:39:51 840.4
130 2015-10-08 15:40:21 848.5
131 2015-10-08 15:40:51 860.4
132 2015-10-08 15:41:21 873.7
133 2015-10-08 15:41:51 884.4
134 2015-10-08 15:42:21 891.8
135 2015-10-08 15:42:51 896.6
136 2015-10-08 15:43:21 899.4
137 2015-10-08 15:43:51 901.0
138 2015-10-08 15:44:21 902.1
139 2015-10-08 15:44:51 902.8
140 2015-10-08 15:45:21 903.0
141 2015-10-08 15:45:51 903.2
142 2015-10-08 15:46:21 903.2
143 2015-10-08 15:46:51 903.2
144 2015-10-08 15:47:21 903.5
145 2015-10-08 15:47:51 903.5
146 2015-10-08 15:48:21 903.5
147 2015-10-08 15:48:51 903.5
148 2015-10-08 15:49:21 903.5
149 2015-10-08 15:49:51 903.5
150 2015-10-08 15:50:21 903.5
151 2015-10-08 15:50:51 903.3
152 2015-10-08 15:51:21 903.5
153 2015-10-08 15:51:51 903.5
154 2015-10-08 15:52:21 903.2
155 2015-10-08 15:52:51 903.5
156 2015-10-08 15:53:21 903.3
157 2015-10-08 15:53:51 903.0
158 2015-10-08 15:54:21 903.3
159 2015-10-08 15:54:51 903.3
160 2015-10-08 15:55:21 903.2
161 2015-10-08 15:55:51 903.0
162 2015-10-08 15:56:21 903.0
163 2015-10-08 15:56:51 906.3
164 2015-10-08 15:57:21 926.1
165 2015-10-08 15:57:51 945.0
166 2015-10-08 15:58:21 956.8
167 2015-10-08 15:58:51 964.6
168 2015-10-08 15:59:21 968.6
169 2015-10-08 15:59:51 970.8
170 2015-10-08 16:00:21 972.1
171 2015-10-08 16:00:51 972.7
172 2015-10-08 16:01:21 973.0
173 2015-10-08 16:01:51 973.5
174 2015-10-08 16:02:21 973.5
175 2015-10-08 16:02:51 973.5
176 2015-10-08 16:03:21 973.5
177 2015-10-08 16:03:51 973.5
178 2015-10-08 16:04:21 973.6
179 2015-10-08 16:04:51 973.5
180 2015-10-08 16:05:21 973.5
181 2015-10-08 16:05:51 973.5
182 2015-10-08 16:06:21 973.5
183 2015-10-08 16:06:51 973.5
184 2015-10-08 16:07:21 973.5
185 2015-10-08 16:07:51 973.5
186 2015-10-08 16:08:21 992.7
187 2015-10-08 16:08:51 1018.0
188 2015-10-08 16:09:21 1032.8
189 2015-10-08 16:09:51 1041.1
190 2015-10-08 16:10:21 1045.0
191 2015-10-08 16:10:51 1047.1
192 2015-10-08 16:11:21 1047.7
193 2015-10-08 16:11:51 1053.5
194 2015-10-08 16:12:21 1082.7
195 2015-10-08 16:12:51 1100.0
196 2015-10-08 16:13:21 1108.8
197 2015-10-08 16:13:51 1112.5
198 2015-10-08 16:14:21 1114.2
199 2015-10-08 16:14:51 1116.9
200 2015-10-08 16:15:21 1148.3
201 2015-10-08 16:15:51 1168.6
202 2015-10-08 16:16:21 1176.6
203 2015-10-08 16:16:51 1180.1
204 2015-10-08 16:17:21 1181.3
205 2015-10-08 16:17:51 1182.0
206 2015-10-08 16:18:21 1217.1
207 2015-10-08 16:18:51 1239.5
208 2015-10-08 16:19:21 1247.7
209 2015-10-08 16:19:51 1250.8
210 2015-10-08 16:20:21 1251.6
211 2015-10-08 16:20:51 1252.2
212 2015-10-08 16:21:21 1252.0
213 2015-10-08 16:21:51 1252.1
214 2015-10-08 16:22:21 1252.2
215 2015-10-08 16:22:51 1254.5
216 2015-10-08 16:23:21 1289.5
217 2015-10-08 16:23:51 1305.0
218 2015-10-08 16:24:21 1310.1
219 2015-10-08 16:24:51 1311.7
220 2015-10-08 16:25:21 1312.1
221 2015-10-08 16:25:51 1312.1
222 2015-10-08 16:26:21 1312.1
223 2015-10-08 16:26:51 1312.6
224 2015-10-08 16:27:21 1312.3
225 2015-10-08 16:27:51 1312.5
226 2015-10-08 16:28:21 1312.3
227 2015-10-08 16:28:51 1312.1
228 2015-10-08 16:29:21 1312.1
229 2015-10-08 16:29:51 1312.8
230 2015-10-08 16:30:21 1312.5
231 2015-10-08 16:30:51 1312.1
232 2015-10-08 16:31:21 1311.8
233 2015-10-08 16:31:51 1311.7
234 2015-10-08 16:32:21 1311.7
235 2015-10-08 16:32:51 1312.6
236 2015-10-08 16:33:21 1354.6
237 2015-10-08 16:33:51 1370.7
238 2015-10-08 16:34:21 1375.4
239 2015-10-08 16:34:51 1376.4
240 2015-10-08 16:35:21 1376.9
241 2015-10-08 16:35:51 1376.7
242 2015-10-08 16:36:21 1377.0
243 2015-10-08 16:36:51 1377.0
244 2015-10-08 16:37:21 1384.5
245 2015-10-08 16:37:51 1426.4
246 2015-10-08 16:38:21 1436.9
247 2015-10-08 16:38:51 1439.0
248 2015-10-08 16:39:21 1440.4
249 2015-10-08 16:39:51 1440.4
250 2015-10-08 16:40:21 1440.0
251 2015-10-08 16:40:51 1440.3
252 2015-10-08 16:41:21 1469.0
253 2015-10-08 16:41:51 1489.8
254 2015-10-08 16:42:21 1494.0
255 2015-10-08 16:42:51 1495.0
256 2015-10-08 16:43:21 1495.1
257 2015-10-08 16:43:51 1495.3
258 2015-10-08 16:44:21 1495.1
259 2015-10-08 16:44:51 1543.6
260 2015-10-08 16:45:21 1555.0
261 2015-10-08 16:45:51 1557.6
262 2015-10-08 16:46:21 1557.6
263 2015-10-08 16:46:51 1557.7
264 2015-10-08 16:47:21 1558.1
265 2015-10-08 16:47:51 1571.9
266 2015-10-08 16:48:21 1607.9
267 2015-10-08 16:48:51 1612.9
268 2015-10-08 16:49:21 1613.5
269 2015-10-08 16:49:51 1613.3
270 2015-10-08 16:50:21 1613.5
271 2015-10-08 16:50:51 1612.5
272 2015-10-08 16:51:21 1612.1
273 2015-10-08 16:51:51 1646.2
274 2015-10-08 16:52:21 1664.2
275 2015-10-08 16:52:51 1665.2
276 2015-10-08 16:53:21 1665.3
277 2015-10-08 16:53:51 1665.7
278 2015-10-08 16:54:21 1688.6
279 2015-10-08 16:54:51 1717.2
280 2015-10-08 16:55:21 1720.7
281 2015-10-08 16:55:51 1720.6
282 2015-10-08 16:56:21 1720.1
283 2015-10-08 16:56:51 1722.2
284 2015-10-08 16:57:21 1757.5
285 2015-10-08 16:57:51 1769.5
286 2015-10-08 16:58:21 1771.0
287 2015-10-08 16:58:51 1770.4
288 2015-10-08 16:59:21 1771.0
289 2015-10-08 16:59:51 1769.0
290 2015-10-08 17:00:21 1769.5
291 2015-10-08 17:00:51 1769.5
292 2015-10-08 17:01:21 1768.9
293 2015-10-08 17:01:51 1769.0
294 2015-10-08 17:02:21 1769.5
295 2015-10-08 17:02:51 1768.9
296 2015-10-08 17:03:21 1769.4
297 2015-10-08 17:03:51 1769.5
298 2015-10-08 17:04:21 1769.5
299 2015-10-08 17:04:51 1768.9
... 95919 more lines ...
|
Attachment 3: 2015-10-06.4ch
|
°C Date Time T1 - C T2 - C T3 - C T4 - K
1 2015-10-06 11:06:14 21.7
2 2015-10-06 11:06:44 21.3
3 2015-10-06 11:07:14 21.6
4 2015-10-06 11:07:44 22.1
5 2015-10-06 11:08:14 21.7
6 2015-10-06 11:08:44 21.9
7 2015-10-06 11:09:14 22.3
8 2015-10-06 11:09:44 22.7
9 2015-10-06 11:10:14 21.8
10 2015-10-06 11:10:44 22.5
11 2015-10-06 11:11:14 23.5
12 2015-10-06 11:11:44 22.7
13 2015-10-06 11:12:14 22.9
14 2015-10-06 11:12:44 23.3
15 2015-10-06 11:13:14 22.5
16 2015-10-06 11:13:44 21.7
17 2015-10-06 11:14:14 22.2
18 2015-10-06 11:14:44 22.2
19 2015-10-06 11:15:14 22.1
20 2015-10-06 11:15:44 22.5
21 2015-10-06 11:16:14 22.5
22 2015-10-06 11:16:44 22.7
23 2015-10-06 11:17:14 22.7
24 2015-10-06 11:17:44 22.1
25 2015-10-06 11:18:14 22.5
26 2015-10-06 11:18:44 21.9
27 2015-10-06 11:19:14 22.2
28 2015-10-06 11:19:44 22.7
29 2015-10-06 11:20:14 22.7
30 2015-10-06 11:20:44 22.5
31 2015-10-06 11:21:14 22.5
32 2015-10-06 11:21:44 22.5
33 2015-10-06 11:22:14 22.5
34 2015-10-06 11:22:44 22.2
35 2015-10-06 11:23:14 21.3
36 2015-10-06 11:23:44 21.3
37 2015-10-06 11:24:14 21.8
38 2015-10-06 11:24:44 21.4
39 2015-10-06 11:25:14 21.5
40 2015-10-06 11:25:44 21.6
41 2015-10-06 11:26:14 21.6
42 2015-10-06 11:26:44 21.6
43 2015-10-06 11:27:14 21.7
44 2015-10-06 11:27:44 21.9
45 2015-10-06 11:28:14 21.8
46 2015-10-06 11:28:44 21.8
47 2015-10-06 11:29:14 21.9
48 2015-10-06 11:29:44 21.9
49 2015-10-06 11:30:14 21.9
50 2015-10-06 11:30:44 22.7
51 2015-10-06 11:31:14 22.2
52 2015-10-06 11:31:44 22.1
53 2015-10-06 11:32:14 22.2
54 2015-10-06 11:32:44 21.8
55 2015-10-06 11:33:14 22.2
56 2015-10-06 11:33:44 22.3
57 2015-10-06 11:34:14 21.9
58 2015-10-06 11:34:44 21.9
59 2015-10-06 11:35:14 21.4
60 2015-10-06 11:35:44 21.9
61 2015-10-06 11:36:14 22.5
62 2015-10-06 11:36:44 21.6
63 2015-10-06 11:37:14 22.2
64 2015-10-06 11:37:44 21.6
65 2015-10-06 11:38:14 21.1
66 2015-10-06 11:38:44 21.2
67 2015-10-06 11:39:14 21.7
68 2015-10-06 11:39:44 21.2
69 2015-10-06 11:40:14 22.2
70 2015-10-06 11:40:44 21.8
71 2015-10-06 11:41:14 21.9
72 2015-10-06 11:41:44 21.3
73 2015-10-06 11:42:14 21.4
74 2015-10-06 11:42:44 21.9
75 2015-10-06 11:43:14 21.4
76 2015-10-06 11:43:44 21.5
77 2015-10-06 11:44:14 21.6
78 2015-10-06 11:44:44 21.6
79 2015-10-06 11:45:14 21.6
80 2015-10-06 11:45:44 22.1
81 2015-10-06 11:46:14 21.6
82 2015-10-06 11:46:44 21.6
83 2015-10-06 11:47:14 21.6
84 2015-10-06 11:47:44 21.6
85 2015-10-06 11:48:14 21.6
86 2015-10-06 11:48:44 21.7
87 2015-10-06 11:49:14 21.7
88 2015-10-06 11:49:44 21.7
89 2015-10-06 11:50:14 21.7
90 2015-10-06 11:50:44 21.8
91 2015-10-06 11:51:14 21.7
92 2015-10-06 11:51:44 21.8
93 2015-10-06 11:52:14 21.9
94 2015-10-06 11:52:44 21.8
95 2015-10-06 11:53:14 21.9
96 2015-10-06 11:53:44 21.8
97 2015-10-06 11:54:14 21.9
98 2015-10-06 11:54:44 21.9
99 2015-10-06 11:55:14 21.9
100 2015-10-06 11:55:44 21.9
101 2015-10-06 11:56:14 21.9
102 2015-10-06 11:56:44 21.9
103 2015-10-06 11:57:14 21.9
104 2015-10-06 11:57:44 21.9
105 2015-10-06 11:58:14 21.9
106 2015-10-06 11:58:44 22.1
107 2015-10-06 11:59:14 21.9
108 2015-10-06 11:59:44 21.9
109 2015-10-06 12:00:14 22.1
110 2015-10-06 12:00:44 21.9
111 2015-10-06 12:01:14 21.9
112 2015-10-06 12:01:44 21.9
113 2015-10-06 12:02:14 22.1
114 2015-10-06 12:02:44 22.2
115 2015-10-06 12:03:14 22.1
116 2015-10-06 12:03:44 22.1
117 2015-10-06 12:04:14 22.1
118 2015-10-06 12:04:44 22.1
119 2015-10-06 12:05:14 22.1
120 2015-10-06 12:05:44 22.1
121 2015-10-06 12:06:14 22.1
122 2015-10-06 12:06:44 22.2
123 2015-10-06 12:07:14 22.2
124 2015-10-06 12:07:44 22.2
125 2015-10-06 12:08:14 22.2
126 2015-10-06 12:08:44 22.2
127 2015-10-06 12:09:14 22.2
128 2015-10-06 12:09:44 22.2
129 2015-10-06 12:10:14 22.2
130 2015-10-06 12:10:44 22.2
131 2015-10-06 12:11:14 22.2
132 2015-10-06 12:11:44 22.2
133 2015-10-06 12:12:14 22.2
134 2015-10-06 12:12:44 22.2
135 2015-10-06 12:13:14 22.2
136 2015-10-06 12:13:44 22.2
137 2015-10-06 12:14:14 22.2
138 2015-10-06 12:14:44 22.2
139 2015-10-06 12:15:14 22.2
140 2015-10-06 12:15:44 22.2
141 2015-10-06 12:16:14 22.2
142 2015-10-06 12:16:44 22.2
143 2015-10-06 12:17:14 22.2
144 2015-10-06 12:17:44 22.2
145 2015-10-06 12:18:14 22.2
146 2015-10-06 12:18:44 22.2
147 2015-10-06 12:19:14 22.3
148 2015-10-06 12:19:44 22.3
149 2015-10-06 12:20:14 22.4
150 2015-10-06 12:20:44 22.3
151 2015-10-06 12:21:14 22.3
152 2015-10-06 12:21:44 22.4
153 2015-10-06 12:22:14 22.3
154 2015-10-06 12:22:44 22.3
155 2015-10-06 12:23:14 22.3
156 2015-10-06 12:23:44 22.4
157 2015-10-06 12:24:14 22.3
158 2015-10-06 12:24:44 22.4
159 2015-10-06 12:25:14 22.5
160 2015-10-06 12:25:44 22.4
161 2015-10-06 12:26:14 22.4
162 2015-10-06 12:26:44 22.4
163 2015-10-06 12:27:14 22.4
164 2015-10-06 12:27:44 22.4
165 2015-10-06 12:28:14 22.5
166 2015-10-06 12:28:44 22.4
167 2015-10-06 12:29:14 22.4
168 2015-10-06 12:29:44 22.5
169 2015-10-06 12:30:14 22.5
170 2015-10-06 12:30:44 22.5
171 2015-10-06 12:31:14 22.5
172 2015-10-06 12:31:44 22.4
173 2015-10-06 12:32:14 22.5
174 2015-10-06 12:32:44 22.5
175 2015-10-06 12:33:14 22.5
176 2015-10-06 12:33:44 22.5
177 2015-10-06 12:34:14 22.5
178 2015-10-06 12:34:44 22.5
179 2015-10-06 12:35:14 22.5
180 2015-10-06 12:35:44 22.5
181 2015-10-06 12:36:14 22.5
182 2015-10-06 12:36:44 22.5
183 2015-10-06 12:37:14 22.5
184 2015-10-06 12:37:44 22.5
185 2015-10-06 12:38:14 22.5
186 2015-10-06 12:38:44 22.5
187 2015-10-06 12:39:14 22.5
188 2015-10-06 12:39:44 22.5
189 2015-10-06 12:40:14 22.5
190 2015-10-06 12:40:44 22.6
191 2015-10-06 12:41:14 22.5
192 2015-10-06 12:41:44 22.5
193 2015-10-06 12:42:14 22.5
194 2015-10-06 12:42:44 22.5
195 2015-10-06 12:43:14 22.6
196 2015-10-06 12:43:44 22.5
197 2015-10-06 12:44:14 22.6
198 2015-10-06 12:44:44 22.5
199 2015-10-06 12:45:14 22.5
200 2015-10-06 12:45:44 22.5
201 2015-10-06 12:46:14 22.5
202 2015-10-06 12:46:44 22.7
203 2015-10-06 12:47:14 22.6
204 2015-10-06 12:47:44 22.6
205 2015-10-06 12:48:14 22.7
206 2015-10-06 12:48:44 22.7
207 2015-10-06 12:49:14 22.6
208 2015-10-06 12:49:44 22.6
209 2015-10-06 12:50:14 22.7
210 2015-10-06 12:50:44 22.6
211 2015-10-06 12:51:14 22.6
212 2015-10-06 12:51:44 22.6
213 2015-10-06 12:52:14 22.6
214 2015-10-06 12:52:44 22.7
215 2015-10-06 12:53:14 22.7
216 2015-10-06 12:53:44 22.6
217 2015-10-06 12:54:14 22.7
218 2015-10-06 12:54:44 22.6
219 2015-10-06 12:55:14 22.6
220 2015-10-06 12:55:44 22.5
221 2015-10-06 12:56:14 22.6
222 2015-10-06 12:56:44 22.6
223 2015-10-06 12:57:14 22.5
224 2015-10-06 12:57:44 22.6
225 2015-10-06 12:58:14 22.6
226 2015-10-06 12:58:44 22.6
227 2015-10-06 12:59:14 22.7
228 2015-10-06 12:59:44 22.6
229 2015-10-06 13:00:14 22.6
230 2015-10-06 13:00:44 22.6
231 2015-10-06 13:01:14 22.6
232 2015-10-06 13:01:44 22.6
233 2015-10-06 13:02:14 22.6
234 2015-10-06 13:02:44 22.6
235 2015-10-06 13:03:14 22.6
236 2015-10-06 13:03:44 22.7
237 2015-10-06 13:04:14 22.7
238 2015-10-06 13:04:44 22.7
239 2015-10-06 13:05:14 22.6
240 2015-10-06 13:05:44 22.6
241 2015-10-06 13:06:14 22.7
242 2015-10-06 13:06:44 22.7
243 2015-10-06 13:07:14 22.7
244 2015-10-06 13:07:44 22.7
245 2015-10-06 13:08:14 22.7
246 2015-10-06 13:08:44 22.7
247 2015-10-06 13:09:14 22.7
248 2015-10-06 13:09:44 22.7
249 2015-10-06 13:10:14 22.7
250 2015-10-06 13:10:44 22.7
251 2015-10-06 13:11:14 22.7
252 2015-10-06 13:11:44 22.7
253 2015-10-06 13:12:14 22.7
254 2015-10-06 13:12:44 22.7
255 2015-10-06 13:13:14 22.7
256 2015-10-06 13:13:44 22.7
257 2015-10-06 13:14:14 22.7
258 2015-10-06 13:14:44 22.7
259 2015-10-06 13:15:14 22.7
260 2015-10-06 13:15:44 22.7
261 2015-10-06 13:16:14 22.7
262 2015-10-06 13:16:44 22.6
263 2015-10-06 13:17:14 22.7
264 2015-10-06 13:17:44 22.7
265 2015-10-06 13:18:14 22.7
266 2015-10-06 13:18:44 22.6
267 2015-10-06 13:19:14 22.7
268 2015-10-06 13:19:44 22.7
269 2015-10-06 13:20:14 22.7
270 2015-10-06 13:20:44 23.2
271 2015-10-06 13:21:14 22.7
272 2015-10-06 13:21:44 22.7
273 2015-10-06 13:22:14 22.7
274 2015-10-06 13:22:44 23.2
275 2015-10-06 13:23:14 22.7
276 2015-10-06 13:23:44 22.7
277 2015-10-06 13:24:14 22.7
278 2015-10-06 13:24:44 22.7
279 2015-10-06 13:25:14 22.7
280 2015-10-06 13:25:44 22.7
281 2015-10-06 13:26:14 22.7
282 2015-10-06 13:26:44 22.7
283 2015-10-06 13:27:14 22.7
284 2015-10-06 13:27:44 22.7
285 2015-10-06 13:28:14 22.7
286 2015-10-06 13:28:44 22.7
287 2015-10-06 13:29:14 22.7
288 2015-10-06 13:29:44 22.7
289 2015-10-06 13:30:14 22.7
290 2015-10-06 13:30:44 22.6
291 2015-10-06 13:31:14 22.6
292 2015-10-06 13:31:44 22.6
293 2015-10-06 13:32:14 23.1
294 2015-10-06 13:32:44 22.6
295 2015-10-06 13:33:14 22.6
296 2015-10-06 13:33:44 22.6
297 2015-10-06 13:34:14 22.6
298 2015-10-06 13:34:44 23.2
299 2015-10-06 13:35:14 22.5
... 5876 more lines ...
|
Attachment 4: 2015-10-06_Small_Centre_Block.xlsx
|
|
264
|
Monday, November 30, 2015, 15:38 |
Laura Lambert | Empty | 39530-9 | HP-FEBIAD | Done | emissivity measurements with no h.s on HP-FEBIAD container | Monday, November 30, 2015, 15:37
No heat shields.
Container empty.
Loaded into evap2 and currently pumping down.
Tuesday, December 01, 2015, 10:07
HS water valves opened this morning, no leaks detected.
IGP1 = 4.3e-7 Torr.
Slope of OMEGA = 0.98
Tapped up manually to 35A.
Vacuum stable, initiating auto-ramp at 2.0A/min and 1.0A/min for the TGHT and TBHT respectively.
Final setpoints: TBHT = 90A, TGHT = 360A
Once final setpoints reached, will tap up manually to max 1800C [legs = 2000C]
Monday, December 07, 2015, 09:06
Container was tapped up manually with emissivity measurements taken every 50A starting at 360A.
The final measurement was taken at 640A.
Container was ramped down for the weekend.
See attachment for raw data.
|
Attachment 1: emissivity_measurements.xlsx
|
|
263
|
Monday, November 23, 2015, 14:36 |
Laura Lambert | Empty | 39530-9 | HP-FEBIAD | Done | temp measurements with no h.s on HP-FEBIAD container | Monday, November 23, 2015, 14:29
No heat shields.
Container empty.
Loaded into evap2 and currently pumping down.
Tuesday, November 24, 2015, 09:36
HS water valves opened this morning, no leaks detected.
IGP1 = 4.5e-7 Torr.
Changed slope of OMEGA IR to 0.98 for empty container.
Tapped up manually to 35A.
IGP1 = 4.7e-7 Torr.
Vacuum stable, initiating auto-ramp at 2.0A/min and 1.0A/min for the TGHT and TBHT respectively.
Final setpoints: TBHT = 230A, TGHT = 300A
Will tap up manually to 2000C
Tuesday, November 24, 2015, 14:00
TGHT = 300A, 3.75V
TBHT = 230A, 3.08V
IGP1 = 4.7e-7 Torr
No reading yet on the OMEGA.
Re-initiating auto-ramp to final setpoint TGHT = 450A
Wednesday, November 25, 2015, 09:11
IGP1 = 3.0e-7Torr.
TGHT = 450A, 6.37V
Tapping up manually to 2000C
Wednesday, November 25, 2015, 15:31
container was tapped up manually to 650A, slope/emissivity settings were checked.
ramped back down to 450A and tapped up manually while recording data up to 1800C [legs = 2000C]
once reached 1800C, tapped TBHT down manually to see relation of TBHT to temperature/ TGHT
remained at 1800C for ~1hr [while tapping down TBHT] before initiating auto-cool at 2.0A/min for the TGHT
see attachment for raw data.
Thursday, November 26, 2015, 10:08
Vented in air and took out this morning.
|
Attachment 1: temp_measurements.xlsx
|
|
262
|
Tuesday, November 10, 2015, 14:34 |
Laura Lambert | SiC | 41077-3 | LP-SIS | Done | temp measurements with 3xTa h.s on LPSIS (small block) container with SiC | Tuesday, November 10, 2015, 14:20
Container previously sintered with TaC coating and temperature measurements taken [see elog ID# 256, 257, 258]
3x Ta heat shields
2 packs of 5 SiC discs ~2inches from the end of the right side wrt ionizer tube into page.
Loaded into evap2 and currently pumping down.
see photos attached.
Monday, November 16, 2015, 11:24
HS water valves opened up this morning, no leaks detected.
Tapped up manually to 35A.
Vacuum stable.
IGP1 = 3.4e-7 Torr.
Initiating auto-ramp at 1.0A/min and 0.5A/min for the TGHT and TBHT respectively.
Final setpoints: 250A for the TGHT and 230A for the TBHT
Monday, November 16, 2015, 13:29
Vacuum too high.
IGP1 = 2.9e-5 Torr.
Decreasing ramping rate down to 0.5A/min and 0.2A/min for the TGHT and TBHT respectively.
Tuesday, November 17, 2015, 09:12
IGP1 = 1.45e-7 Torr.
TGHT = 250A, 3.28V
TBHT = 230A, 3.33V
OMEGA = 1194C
Tapping up manually to 1900C
Tuesday, November 17, 2015, 11:22
Tapped up manually to 1800C, realized SiC should only be tapped up to 1700C, backed off immediately down to 415A from 460A/ 1800C.
See attachment for photo of striptool/ time spent above 1700C [~20min].
Tuesday, November 17, 2015, 15:46
TGHT remained at 1700C for ~4hrs.
TGHT = 420A, 6.44V.
IGP1 = 2.3e-7 Torr.
Initiating cool-down at 2.0A/min and 1.0A/min for the TGHT and TBHT respectively.
See attachment for raw data of temperature measurements.
Wednesday, November 18, 2015, 13:32
Vented in air and took out this morning.
Looks good, no visual changes with the container.
|
Attachment 1: Evernote_Camera_Roll_20151110_110128.jpg
|
|
Attachment 2: Evernote_Camera_Roll_20151110_110129.jpg
|
|
Attachment 3: striptool_screenshot_above_1700C.JPG
|
|
Attachment 4: temp_measurements_on_LPSIS_(small_block)_container.xlsx
|
|
261
|
Monday, November 09, 2015, 14:22 |
Laura Lambert | | 41077-3 | LP-SIS | Done | Sintering TaC coating on 'temporary endcaps' springs (Slope changed to 1.04) |
John Wong wrote: |
Laura Lambert wrote: |
Thursday, November 05, 2015, 16:19
Springs coated with TaC and loaded into container for sintering.
3x Ta heat shield.
EVAP2 currently pumping down.
Friday, November 06, 2015, 08:39
Opened HS water valves, no leaks detected.
Tapped up manually to 35A.
IGP1 = 5.5e-7 Torr.
Friday, November 06, 2015, 10:08
IGP1 = 5.8e-7 Torr.
Vacuum stable.
Initiating auto-ramp at 0.2A/min and 0.1A/min with final setpoints of 450A and 230A for the TGHT and TBHT respectively.
See photos attached to view how springs were coated and loaded into container.
Monday, November 09, 2015, 09:07
TGHT reached 450A, 7.2V ~8pm sat, nov 7th.
OMEGA = 1775C
IGP1 = 3.2e-7 Torr.
Tapping up manually to 1900C.
|
Monday, November 09, 2015, 12:30
Checked the temp on the inner surface (not at the spring) of the container was around 2050C at 500A.
The TaC coated spring is reading at 1900C; that's what the OMEGA is reading with the setting for 2-colors and slope at 0.98.
Lower TGHT to 450A/ 7.12V.
By comparing the data from the previous temp curve (OMEGA slope at 0.98) with no material inside the container, the voltage and current plot is aligned with the curve.
Changed OMEGA setting slope to 1.04. Current temp is reading at 1917C at 450A.
For 2-Color setting, the emissivity should be negligible. That makes me believe that different materials has different slope setting for the temp reading.
One thing to note is that, the sintering temp of TaC is around 1900-2000C. It was clear that the container was fully sintered with TGHT at 450A.
Will need to gather more data with different material to understand the slope setting better.
Try:
SiC at 1700C, slope at 1.04
ZrC at 1900C, slope at 104
Ta at 1900C, slope at 0.98.
Compare the data...
|
Monday, November 09, 2015, 14:20
IGP1 = 3.3e-7 Torr.
TGHT = 450A, 7.12V
OMEGA = 1915C
Initiating auto-cool at 2.0A/min and 1.0A/min for the TGHT and TBHT respectively.
Tuesday, November 10, 2015, 13:23
Vented and opened this morning.
springs look [mostly] sintered.
see photo attached.
|
Attachment 1: Evernote_Camera_Roll_20151110_113123.jpg
|
|
|
260
|
Monday, November 09, 2015, 12:56 |
John Wong | | 41077-3 | LP-SIS | In progress | Sintering TaC coating on 'temporary endcaps' springs (Slope changed to 1.04) |
Laura Lambert wrote: |
Thursday, November 05, 2015, 16:19
Springs coated with TaC and loaded into container for sintering.
3x Ta heat shield.
EVAP2 currently pumping down.
Friday, November 06, 2015, 08:39
Opened HS water valves, no leaks detected.
Tapped up manually to 35A.
IGP1 = 5.5e-7 Torr.
Friday, November 06, 2015, 10:08
IGP1 = 5.8e-7 Torr.
Vacuum stable.
Initiating auto-ramp at 0.2A/min and 0.1A/min with final setpoints of 450A and 230A for the TGHT and TBHT respectively.
See photos attached to view how springs were coated and loaded into container.
Monday, November 09, 2015, 09:07
TGHT reached 450A, 7.2V ~8pm sat, nov 7th.
OMEGA = 1775C
IGP1 = 3.2e-7 Torr.
Tapping up manually to 1900C.
|
Monday, November 09, 2015, 12:30
Checked the temp on the inner surface (not at the spring) of the container was around 2050C at 500A.
The TaC coated spring is reading at 1900C; that's what the OMEGA is reading with the setting for 2-colors and slope at 0.98.
Lower TGHT to 450A/ 7.12V.
By comparing the data from the previous temp curve (OMEGA slope at 0.98) with no material inside the container, the voltage and current plot is aligned with the curve.
Changed OMEGA setting slope to 1.04. Current temp is reading at 1917C at 450A.
For 2-Color setting, the emissivity should be negligible. That makes me believe that different materials has different slope setting for the temp reading.
One thing to note is that, the sintering temp of TaC is around 1900-2000C. It was clear that the container was fully sintered with TGHT at 450A.
Will need to gather more data with different material to understand the slope setting better.
Try:
SiC at 1700C, slope at 1.04
ZrC at 1900C, slope at 104
Ta at 1900C, slope at 0.98.
Compare the data...
|
|
259
|
Thursday, November 05, 2015, 16:20 |
Laura Lambert | | 41077-3 | LP-SIS | In progress | Sintering TaC coating on 'temporary endcaps' springs | Thursday, November 05, 2015, 16:19
Springs coated with TaC and loaded into container for sintering.
3x Ta heat shield.
EVAP2 currently pumping down.
Friday, November 06, 2015, 08:39
Opened HS water valves, no leaks detected.
Tapped up manually to 35A.
IGP1 = 5.5e-7 Torr.
Friday, November 06, 2015, 10:08
IGP1 = 5.8e-7 Torr.
Vacuum stable.
Initiating auto-ramp at 0.2A/min and 0.1A/min with final setpoints of 450A and 230A for the TGHT and TBHT respectively.
See photos attached to view how springs were coated and loaded into container.
Monday, November 09, 2015, 09:07
TGHT reached 450A, 7.2V ~8pm sat, nov 7th.
OMEGA = 1775C
IGP1 = 3.2e-7 Torr.
Tapping up manually to 1900C.
|
Attachment 1: 2015-11-02_16.03.10.jpg
|
|
Attachment 2: 'temporary_endcap'_springs_nov.3_2015.jpg
|
|
Attachment 3: IMG_20151105_141239627_HDR.jpg
|
|
Attachment 4: IMG_20151105_141248901_HDR.jpg
|
|
|
258
|
Monday, November 02, 2015, 16:29 |
Laura Lambert | Empty | 41077-3 | LP-SIS | Done | temp measurements with 3xTa h.s on LPSIS (small block) container | Monday, November 02, 2015, 16:28
Container previously sintered with TaC coating and temperature measurements taken with 1x Ta heat shield [see elog ID# 256] and 2x Ta heat shield [see elog ID#257]
3x Ta heat shields on this container.
Loaded into evap2 and currently pumping down.
Tuesday, November 03, 2015, 10:37
HS water valves opened up this morning, no leaks detected.
Tapped up manually to 35A.
Vacuum stable.
IGP1 = 6.8e-7 Torr.
Initiating auto-ramp at 2.0A/min and 1.0A/min for the TGHT and TBHT respectively.
Final setpoints: 250A for the TGHT and 230A for the TBHT
Tuesday, November 03, 2015, 13:09
IGP1 = 7.7e-7 Torr
TGHT = 250A, 3.42V
TBHT = 185A, 3.0V
OMEGA = 1237C
Increasing ramping rate of TBHT to 2.0A/min.
Tuesday, November 03, 2015, 13:47
Reached final setpoints.
IGP1 = 8.7e-7 Torr.
Tapping up manually to 1900C
Tuesday, November 03, 2015, 15:45
Tapped up manually to 1900C
TGHT = 450A, 7.26V
IGP1 = 1.6e-6 Torr
Will remain at 450A overnight and ramp-down in the morning.
Wednesday, November 04, 2015, 10:00
TGHT remained at 450A for ~18hrs.
IGP1 = 6e-7 Torr.
OMEGA = 1908C
See attachment for raw data of temperature measurements and picture showing voltage movement at max setpoint.
Initiating auto-cool at 2.0A/min and 1.0A/min for the TGHT and TBHT respectively.
Wednesday, November 04, 2015, 16:43
Vented and took out this afternoon.
Container looks good.
No visible changes to container compared to 2xTa heat shield temp measurements.
|
Attachment 1: temp_measurements_on_LPSIS_(small_block)_container.xlsx
|
Attachment 2: 3xTa_h.s_-_striptool_voltage.JPG
|
|
|
257
|
Monday, October 26, 2015, 16:17 |
Laura Lambert | Empty | 41077-3 | LP-SIS | Done | temp measurements with 2xTa h.s on LPSIS (small block) container | Monday, October 26, 2015, 16:16
Container previously sintered with TaC coating and temperature measurements taken with 1x Ta heat shield [see elog ID# 256]
2x Ta heat shields on this container.
Loaded into evap2 and currently pumping down.
Tuesday, October 27, 2015, 10:07
HS water valves opened up this morning, no leaks detected.
Tapped up manually to 35A.
Waited ~10min, vacuum stable
Initiating auto-ramp at 2.0A/min and 1.0A/min for the TGHT and TBHT respectively.
Final setpoints: 300A for the TGHT and 230A for the TBHT
Tuesday, October 27, 2015, 13:54
Reached final setpoints.
IGP1 = 5.3e-7 Torr.
Tapping up manually to 1900C
Tuesday, October 27, 2015, 16:42
Tapped up manually to 1900C
TGHT = 455A, 7.66V
IGP1 = 9e-7 Torr
Will remain at 455A overnight and ramp-down in the morning.
Wednesday, October 28, 2015, 09:32
TGHT remained at 455A for ~17hrs.
IGP1 = 4.3e-7 Torr.
OMEGA = 1888C
See attachment for raw data of temperature measurements.
Initiating auto-cool at 2.0A/min and 1.0A/min for the TGHT and TBHT respectively.
Wednesday, October 28, 2015, 16:32
Vented and took out this afternoon.
Container looks good.
See photos attached.
|
Attachment 1: temp_measurements_on_LPSIS_(small_block)_container.xlsx
|
Attachment 2: 20151102_145643.jpg
|
|
Attachment 3: 20151102_145753.jpg
|
|
Attachment 4: 20151102_145825.jpg
|
|
Attachment 5: 20151102_145837_1.jpg
|
|
Attachment 6: 20151102_145906.jpg
|
|
|
256
|
Monday, October 26, 2015, 16:14 |
Laura Lambert | Empty | 41077-3 | LP-SIS | Done | Sintering TaC on LPSIS (small block) container & temp measurements with 1xTa h.s |
John Wong wrote: |
Laura Lambert wrote: |
John Wong wrote: |
Wednesday, October 21, 2015, 17:35
This container is prepared for a backup UCx target (otherwise it will be used for development purposes).
The container is coated with TaC; one layer of Ta-heat shield is used.
TGHT will be ramped up to reach 1900C (previous container had TGHT ramped up to 440A).
TBHT will be ramped up to max 230A.
Chamber is currently being pumped down.
Heating can begin once the vacuum is good, currently at 4.8e-5 Torr.
|
Thursday, October 22, 2015, 11:36
HS water valves opened, no leaks detected.
Tapped up manually to 35A.
Vacuum stabilized, initiated auto-ramp at 0.2A/min and 0.1A/min for the TGHT and TBHT respectively.
Final setpoints TGHT = 300A, TBHT = 230A
IGP1 = 1.0e-6 Torr; TGHT = 45.18A, 0.17V; TBHT = 38.4A, 0.41V
|
Thursday, October 22, 2015, 17:09
Pressure coming down to 1.5e-6 Torr.
Increased ramping rate to 2A/min for both TGHT and TBHT.
|
Monday, October 26, 2015, 16:10
Target container was ramped down over the weekend and taken out today.
Coating looks good.
Will add on another Ta layer and continue with temperature measurements.
See attachment for raw data.
|
Attachment 1: temp_measurements_on_sintering_TaC_on_LPSIS_(small_block)_container.xlsx
|
|
255
|
Thursday, October 22, 2015, 17:10 |
John Wong | Empty | 41077-3 | LP-SIS | In progress | Sintering TaC on LPSIS (small block) container |
Laura Lambert wrote: |
John Wong wrote: |
Wednesday, October 21, 2015, 17:35
This container is prepared for a backup UCx target (otherwise it will be used for development purposes).
The container is coated with TaC; one layer of Ta-heat shield is used.
TGHT will be ramped up to reach 1900C (previous container had TGHT ramped up to 440A).
TBHT will be ramped up to max 230A.
Chamber is currently being pumped down.
Heating can begin once the vacuum is good, currently at 4.8e-5 Torr.
|
Thursday, October 22, 2015, 11:36
HS water valves opened, no leaks detected.
Tapped up manually to 35A.
Vacuum stabilized, initiated auto-ramp at 0.2A/min and 0.1A/min for the TGHT and TBHT respectively.
Final setpoints TGHT = 300A, TBHT = 230A
IGP1 = 1.0e-6 Torr; TGHT = 45.18A, 0.17V; TBHT = 38.4A, 0.41V
|
Thursday, October 22, 2015, 17:09
Pressure coming down to 1.5e-6 Torr.
Increased ramping rate to 2A/min for both TGHT and TBHT.
|
|
254
|
Thursday, October 22, 2015, 11:41 |
Laura Lambert | Empty | 41077-3 | LP-SIS | In progress | Sintering TaC on LPSIS (small block) container |
John Wong wrote: |
Wednesday, October 21, 2015, 17:35
This container is prepared for a backup UCx target (otherwise it will be used for development purposes).
The container is coated with TaC; one layer of Ta-heat shield is used.
TGHT will be ramped up to reach 1900C (previous container had TGHT ramped up to 440A).
TBHT will be ramped up to max 230A.
Chamber is currently being pumped down.
Heating can begin once the vacuum is good, currently at 4.8e-5 Torr.
|
Thursday, October 22, 2015, 11:36
HS water valves opened, no leaks detected.
Tapped up manually to 35A.
Vacuum stabilized, initiated auto-ramp at 0.2A/min and 0.1A/min for the TGHT and TBHT respectively.
Final setpoints TGHT = 300A, TBHT = 230A
IGP1 = 1.0e-6 Torr; TGHT = 45.18A, 0.17V; TBHT = 38.4A, 0.41V
|
|
253
|
Wednesday, October 21, 2015, 17:39 |
John Wong | Empty | 41077-3 | LP-SIS | In progress | Sintering TaC on LPSIS (small block) container | Wednesday, October 21, 2015, 17:35
This container is prepared for a backup UCx target (otherwise it will be used for development purposes).
The container is coated with TaC; one layer of Ta-heat shield is used.
TGHT will be ramped up to reach 1900C (previous container had TGHT ramped up to 440A).
TBHT will be ramped up to max 230A.
Chamber is currently being pumped down.
Heating can begin once the vacuum is good, currently at 4.8e-5 Torr.
Friday, October 23, 2015, 16:15
TGHT reached to 500A and temp measured at 1940C. More about the temp plot can be found in the dropbox.
Initiated cooling down at 2A/min.
Target will be inspected next Monday.
|
|
252
|
Thursday, October 08, 2015, 12:26 |
John Wong | Empty | 41077-5 | LP-SIS | In progress | Test#2: Measuring temp with thermocouple (away from centre) & 2C-IR |
John Wong wrote: |
John Wong wrote: |
Laura Lambert wrote: |
Thursday, September 24, 2015, 08:39
IGP1 = 6.5e-7Torr
HS water valves opened, no leaks detected.
Tapped up manually to 35A.
Waited ~20min.
Vacuum stable, IGP1 = 1.5e-6 Torr.
Initiated auto-ramp at 0.2A/min and 0.1A/min for the TGHT and TBHT respectively with final setpoints of 250A and 230A.
Final temp of 1200C on auto-ramp in order to initiate temperature measurements starting at ~1200C.
Thursday, September 24, 2015, 10:24
IGP1 = 8.7e-7 Torr
TGHT = 44A, TBHT = 38A
Increasing auto-ramp to 0.5A/min and 0.2A/min for the TGHT and TBHT respectively.
|
Thursday, September 24, 2015, 12:04
A few notes about the target container:The Ta-foil is cut in a full length of the target container, and that also covers up the back block, and the container is wrapped with 3 sheets (see attachment)
The inner tube is coated with TaC, using the "flushing technique" - the coating looks much better; will see how it turns out after the sintering process
The target will be heated up to 1900C . Due to the new wrapping of the Ta-heat shield, the heat distribution may be different. So the target heater will be ramped up to achieve the desired temperature.
Thursday, September 24, 2015, 18:45
TGHT reached to 250A3.6V, TGHT at 230A and stayed for 1.5 hours. Temp-L = 1280C, Temp-R = 1360C.
Auto-ramp TGHT to 500A at 1A/min and heat at that current for overnight. Will take temp reading tomorrow.
Thursday, September 24, 2015, 23:08
TGHT has reached to 500A/8.57V. Temp-L = 1895C, Temp-R = 2073C. (Not sure why the temperature is real, will need to investigate more to get an absolute temp)
Just to be safe… Ramping down TGHT to 425A (according to e-log id 216, Temp-R reached to 1800C with TGHT at 425A/6.84V, but no temp measured for the left at that time)
Will check again later.
Friday, September 25, 2015, 11:52
The target has been heated at 420A for ~12hours; temp measured ~1900C, slightly higher than the previous target container (W.O#40860-1); the Ta-wraping could affect the heat distribution.
The temperature is also confirmed by using the 2-C digital pyrometer. The tube inner surface is measured about 1900C, and the ionizer tube is about 2400C. (see attachment #2).
Initiating auto-cooling at 2A/min for TGHT and 1A/min for TBHT.
The target will be inspected next Monday
Wednesday, September 30, 2015, 16:01
The target container is fully sintered (see attachment#3)
It is still in the discussion where this container will be used online for the UCx. Nevertheless, the target contain is ready and it looks pretty good.
|
Tuesday, October 06, 2015, 15:09
After further discussion, we are going to reheat this container to 420A. The temperature will be measured with the thermocouple - probe inside the centre of the tube - and the 2C-IR for comparison.
This test is to ensure that the container is actually heated to 1900C before the UCx material can be loaded
The chamber has been pumped down to 1.92e-6 Torr. Heat shield valve is open and no leaks are detected.
Ramping up TBHT to 230A max and TGHT to achieve ~1900C
Wednesday, October 07, 2015, 11:12
Yesterday...
TBHT ramped up to 230A and TGHT was at 160A before taking temperature readings.
TGHT from 160A to 423A, IR2C temp reading increases linearly from 1138C to 1900C, whereas the thermocouple readouts are not quite linear, 1040C to 1748C (see attachment)
After 18 hours, both temp reader dropped about 45C.
Increased TGHT to 440C, IR2C & thermocouple currently read 1912C & 1762C, respectively.
Wednesday, October 07, 2015, 12:28
Lowered TBHT to 200A and TGHT remained at 440A. IR2C = 1900C, TypeC = 1745C. Not much affect from TBHT.
Will increase TBHT back to 230A, then start ramping down TGHT and taking temp readings.
Wednesday, October 07, 2015, 17:49
Stepped down TGHT to 160A and TBHT stayed at 230A. Cooling down temp is generated. See attachment.
Notice the overall temp during cooling is lower than that during heating.
TGHT and TBHT will be ramped down to 0A.
Chamber will be vented tomorrow.
More details later...
|
Thursday, October 08, 2015, 12:01
Test 2: To heat the target container back to ~440A and measure the temperature with the type C thermocouple (inside the tube, away from the centre block) and compared the temp reading with the 2C-IR.
Chamber is currently being pump down, vac at 9e-6 Torr.
Will ramp up TGHT to 160A and TBHT to 230A at 2A/min.
Thursday, October 08, 2015, 17:01
TGHT is now at 440A, TBHT is constant at 230A - IR2C is reading the same as before, 1910C, but Type C is reading lower at 1770C. (see attachment)
Will keep the current at 440A overnight.
Friday, October 09, 2015, 10:06
TGHT has stayed at 440A, TBHT at 230A for ~18hours. R-temp dropped slightly from 1910C to 1890C and typeC dropped from 1770 to 1756; otherwise the measurements are reproducible.
After the group discussion, the online heating curve will be re-calculated based on the temp reading from the IR2C curve.
Next, the target container will be filled with UCx material and the conditioning current for Evap#1 will only go up to 440A (or at least the required current to reach the end of the re-carbonization step).
So pay attention to the current when heating the target up in EVAP#1.
Initiate cooling at 2A/min for TGHT and 1A/min TBHT
Friday, October 09, 2015, 11:35
A note about the voltage that may explain the change in the temperature:
See attachment #2:
Left plot = Test #1:
Initially TGHT ramped up to 423A and temp measured at 1900C. Notice, about an hour later, there was a sudden drop in the voltage, and from thereon, the temp was measured at 1860C. About 18 hours later, TGHT was increased to 440A to achieve 1910C.
Right plot = Test #2
Ramped up TGHT to 440A, temp measured 1910C. At that constant current over the 6 hours period, the voltage slowly increased, and the temperature slightly dropped to 1895C.
|
Attachment 1: Test_2_tempcurves.PNG
|
|
Attachment 2: Test1&2_IV_temp_.png
|
|
|
251
|
Tuesday, October 06, 2015, 15:18 |
John Wong | Empty | 41077-5 | LP-SIS | In progress | Test#1: Measuring temp with thermocouple (centre) & 2C-IR |
John Wong wrote: |
Laura Lambert wrote: |
Thursday, September 24, 2015, 08:39
IGP1 = 6.5e-7Torr
HS water valves opened, no leaks detected.
Tapped up manually to 35A.
Waited ~20min.
Vacuum stable, IGP1 = 1.5e-6 Torr.
Initiated auto-ramp at 0.2A/min and 0.1A/min for the TGHT and TBHT respectively with final setpoints of 250A and 230A.
Final temp of 1200C on auto-ramp in order to initiate temperature measurements starting at ~1200C.
Thursday, September 24, 2015, 10:24
IGP1 = 8.7e-7 Torr
TGHT = 44A, TBHT = 38A
Increasing auto-ramp to 0.5A/min and 0.2A/min for the TGHT and TBHT respectively.
|
Thursday, September 24, 2015, 12:04
A few notes about the target container:The Ta-foil is cut in a full length of the target container, and that also covers up the back block, and the container is wrapped with 3 sheets (see attachment)
The inner tube is coated with TaC, using the "flushing technique" - the coating looks much better; will see how it turns out after the sintering process
The target will be heated up to 1900C . Due to the new wrapping of the Ta-heat shield, the heat distribution may be different. So the target heater will be ramped up to achieve the desired temperature.
Thursday, September 24, 2015, 18:45
TGHT reached to 250A3.6V, TGHT at 230A and stayed for 1.5 hours. Temp-L = 1280C, Temp-R = 1360C.
Auto-ramp TGHT to 500A at 1A/min and heat at that current for overnight. Will take temp reading tomorrow.
Thursday, September 24, 2015, 23:08
TGHT has reached to 500A/8.57V. Temp-L = 1895C, Temp-R = 2073C. (Not sure why the temperature is real, will need to investigate more to get an absolute temp)
Just to be safe… Ramping down TGHT to 425A (according to e-log id 216, Temp-R reached to 1800C with TGHT at 425A/6.84V, but no temp measured for the left at that time)
Will check again later.
Friday, September 25, 2015, 11:52
The target has been heated at 420A for ~12hours; temp measured ~1900C, slightly higher than the previous target container (W.O#40860-1); the Ta-wraping could affect the heat distribution.
The temperature is also confirmed by using the 2-C digital pyrometer. The tube inner surface is measured about 1900C, and the ionizer tube is about 2400C. (see attachment #2).
Initiating auto-cooling at 2A/min for TGHT and 1A/min for TBHT.
The target will be inspected next Monday
Wednesday, September 30, 2015, 16:01
The target container is fully sintered (see attachment#3)
It is still in the discussion where this container will be used online for the UCx. Nevertheless, the target contain is ready and it looks pretty good.
|
Tuesday, October 06, 2015, 15:09
After further discussion, we are going to reheat this container to 420A. The temperature will be measured with the thermocouple - probe inside the centre of the tube - and the 2C-IR for comparison.
This test is to ensure that the container is actually heated to 1900C before the UCx material can be loaded
The chamber has been pumped down to 1.92e-6 Torr. Heat shield valve is open and no leaks are detected.
Ramping up TBHT to 230A max and TGHT to achieve ~1900C
Wednesday, October 07, 2015, 11:12
Yesterday...
TBHT ramped up to 230A and TGHT was at 160A before taking temperature readings.
TGHT from 160A to 423A, IR2C temp reading increases linearly from 1138C to 1900C, whereas the thermocouple readouts are not quite linear, 1040C to 1748C (see attachment)
After 18 hours, both temp reader dropped about 45C.
Increased TGHT to 440C, IR2C & thermocouple currently read 1912C & 1762C, respectively.
Wednesday, October 07, 2015, 12:28
Lowered TBHT to 200A and TGHT remained at 440A. IR2C = 1900C, TypeC = 1745C. Not much affect from TBHT.
Will increase TBHT back to 230A, then start ramping down TGHT and taking temp readings.
Wednesday, October 07, 2015, 17:49
Stepped down TGHT to 160A and TBHT stayed at 230A. Cooling down temp is generated. See attachment.
Notice the overall temp during cooling is lower than that during heating.
TGHT and TBHT will be ramped down to 0A.
Chamber will be vented tomorrow.
More details later...
|
Attachment 1: Temp_LPSIS_IR2C_vs_typeC.PNG
|
|
Attachment 2: Temp_LPSIS_IR2C_vs_typeC_heating_cooling.PNG
|
|
|