546
|
Monday, November 18, 2013, 16:42 |
Bevan Moss | Conditioning Station | Standard Operation | TM2 | | HV Testing |
The Glory of TM2 continues 100V at a time. Today I brought the module to 46kV and left it there for 1 hour. During this time there was only 1 spark. I then continues at up to 53kV. At this voltage the canbus continues to be temperamental when the module sparks. The module is still conditioning and I will continue more tomorrow. Attached is the dumped data and a screen shot of the conditioning from 46kV to 53kV. |
Attachment 1: the_glory_of_TM2.JPG
|
|
Attachment 2: 18112013_TM2_No_source_tray_HV_Test.txt
|
Time TCS:IG1:RDVAC [T] TCS:IG2:RDVAC [T] TCS:IG1S:RDVAC [T] TCS:BIAS:RDVOL [V] TCS:BIAS:RDCUR [uA] TCS:EL:RDVOL [V] TCS:EL:RDCUR [mA]
11/18/2013 11:56:33.67635 0.000000350 BadVal BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:34.66886 0.000000350 BadVal BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:35.66751 0.000000350 BadVal BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:36.67660 0.000000352 BadVal BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:37.67534 0.000000352 BadVal BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:38.68405 0.000000352 BadVal BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:39.68261 0.000000352 0.000003758 BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:40.68163 0.000000350 0.000003780 BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:41.68032 0.000000350 0.000003758 BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:42.67903 0.000000354 0.000003758 BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:43.68805 0.000000350 0.000003780 BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:44.68623 0.000000350 0.000003758 BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:45.68500 0.000000350 0.000003780 BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:46.69372 0.000000350 0.000003780 BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:47.69322 0.000000352 0.000003780 BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:48.69129 0.000000352 0.000003758 BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:49.69014 0.000000352 0.000003780 BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:50.68910 0.000000350 0.000003758 BadVal BadVal BadVal BadVal BadVal
11/18/2013 11:56:51.68802 0.000000352 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:56:52.68661 0.000000350 0.000003780 0.000002000 BadVal BadVal BadVal BadVal
11/18/2013 11:56:53.68525 0.000000352 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:56:54.68443 0.000000354 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:56:55.68299 0.000000354 0.000003780 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:56:56.69356 0.000000350 0.000003780 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:56:57.69988 0.000000350 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:56:58.69998 0.000000352 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:56:59.70818 0.000000352 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:57:00.70672 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:57:01.71518 0.000000354 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:02.71393 0.000000352 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:03.72305 0.000000352 0.000003780 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:57:04.72168 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:05.71991 0.000000350 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:06.72879 0.000000350 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:07.72774 0.000000352 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:08.72786 0.000000350 0.000003780 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:57:09.73521 0.000000350 0.000003780 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:57:10.73360 0.000000350 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:11.73271 0.000000348 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:12.73140 0.000000352 0.000003802 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:57:13.72987 0.000000352 0.000003802 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:14.72862 0.000000352 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:15.73998 0.000000350 0.000003780 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:57:16.74679 0.000000352 0.000003780 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:57:17.74555 0.000000352 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:18.74357 0.000000352 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:19.74415 0.000000352 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:20.75099 0.000000352 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:21.76009 0.000000350 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:22.75889 0.000000352 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:23.75773 0.000000352 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:24.167647 0.000000352 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:25.167497 0.000000352 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:26.168548 0.000000350 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:27.168249 0.000000354 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:28.168091 0.000000354 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:57:29.168174 0.000000348 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:30.167873 0.000000357 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:31.168824 0.000000352 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:57:32.168589 0.000000352 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:57:33.168490 0.000000352 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:57:34.168368 0.000000348 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:35.169260 0.000000350 0.000003780 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:57:36.170079 0.000000350 0.000003758 0.000002000 BadVal BadVal BadVal BadVal
11/18/2013 11:57:37.169962 0.000000350 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:57:38.169833 0.000000350 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:39.169741 0.000000348 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:40.169577 0.000000348 0.000003780 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:57:41.170683 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:57:42.171478 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:43.171240 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:44.171080 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:45.171080 0.000000352 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:57:46.171861 0.000000354 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:47.171901 0.000000350 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:48.173683 0.000000350 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:57:49.173532 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:50.173355 0.000000352 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:51.174221 0.000000350 0.000003780 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:57:52.174097 0.000000350 0.000003780 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:57:53.173980 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:54.173846 0.000000352 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:55.173701 0.000000352 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:56.174624 0.000000350 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:57.174550 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:58.175484 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:57:59.175364 0.000000352 0.000003780 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:00.176142 0.000000350 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:01.176991 0.000000348 0.000003780 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:02.176907 0.000000350 0.000003780 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:03.176736 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:04.176560 0.000000350 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:05.176444 0.000000350 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:06.176439 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:07.176466 0.000000352 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:08.176115 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:09.175970 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:10.175854 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:11.176710 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:12.176587 0.000000348 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:13.176518 0.000000352 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:14.177305 0.000000352 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:15.178256 0.000000352 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:16.178161 0.000000346 0.000003780 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:17.177934 0.000000346 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:18.177862 0.000000348 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:19.177767 0.000000350 0.000003780 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:20.178771 0.000000352 0.000003780 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:58:21.179458 0.000000348 0.000003780 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:58:22.180299 0.000000348 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:23.180173 0.000000348 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:24.180086 0.000000352 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:25.180011 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:26.180879 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:27.180711 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:28.180607 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:29.180448 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:58:30.180303 0.000000350 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:31.181202 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:32.181076 0.000000346 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:33.180951 0.000000348 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:58:34.180824 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:35.180722 0.000000352 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:58:36.180621 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:58:37.180619 0.000000350 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:38.181373 0.000000352 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:39.181195 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:40.181042 0.000000348 0.000003736 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:41.181989 0.000000348 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:42.181796 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:43.181704 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:44.182677 0.000000350 0.000003758 0.000002087 BadVal BadVal BadVal BadVal
11/18/2013 11:58:45.182451 0.000000346 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:58:46.183316 0.000000348 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:58:48.803335 0.000000352 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:58:49.803216 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:58:50.802868 0.000000350 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:51.802603 0.000000350 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:52.803455 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:53.803339 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:58:54.803283 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:58:55.804102 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:58:56.803969 0.000000350 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:58:57.803933 0.000000348 0.000003758 0.000002000 BadVal BadVal BadVal BadVal
11/18/2013 11:58:58.803707 0.000000348 0.000003758 0.000002000 BadVal BadVal BadVal BadVal
11/18/2013 11:58:59.803667 0.000000350 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:00.803690 0.000000348 0.000003758 0.000002000 BadVal BadVal BadVal BadVal
11/18/2013 11:59:01.803354 0.000000350 0.000003758 0.000002000 BadVal BadVal BadVal BadVal
11/18/2013 11:59:02.803204 0.000000350 0.000003780 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:03.803102 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:59:04.802982 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:05.803883 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:06.803746 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:07.803569 0.000000348 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:59:08.803499 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:59:09.803404 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:59:10.803237 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:59:11.803090 0.000000348 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:59:12.802976 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:59:13.802824 0.000000350 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:59:14.802723 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:15.802654 0.000000348 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:59:16.802447 0.000000348 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:59:17.802315 0.000000348 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:59:18.802355 0.000000348 0.000003736 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:59:19.802084 0.000000348 0.000003736 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:20.802934 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:21.802850 0.000000350 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:59:22.802712 0.000000348 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:59:23.802672 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:24.802472 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:25.802310 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:26.802213 0.000000352 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:28.87230 0.000000350 0.000003736 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:59:29.86904 0.000000350 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:59:30.86799 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:59:31.87780 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:32.87546 0.000000348 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:59:33.87451 0.000000350 0.000003736 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:34.87271 0.000000350 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:59:35.87150 0.000000350 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:59:36.88072 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:59:37.87900 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:38.87899 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:39.87677 0.000000348 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:59:40.88567 0.000000348 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:59:41.89421 0.000000346 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:42.89294 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:43.89145 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:44.89053 0.000000348 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:59:45.88930 0.000000350 0.000003780 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:59:46.89854 0.000000352 0.000003736 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:47.89750 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:48.89545 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:49.90415 0.000000350 0.000003736 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:50.90300 0.000000348 0.000003736 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:51.91305 0.000000350 0.000003736 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 11:59:52.91017 0.000000350 0.000003736 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 11:59:53.90976 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:54.91865 0.000000350 0.000003758 0.000002087 BadVal BadVal BadVal BadVal
11/18/2013 11:59:55.92687 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:56.93563 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:57.93420 0.000000346 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:58.93286 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 11:59:59.93135 0.000000348 0.000003736 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 12:00:00.93026 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 12:00:01.93957 0.000000348 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 12:00:02.93759 0.000000348 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 12:00:03.93679 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 12:00:04.93605 0.000000352 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 12:00:05.93414 0.000000348 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 12:00:06.94302 0.000000348 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 12:00:07.94146 0.000000348 0.000003736 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 12:00:08.94029 0.000000348 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 12:00:09.93914 0.000000350 0.000003780 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 12:00:10.93783 0.000000348 0.000003736 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 12:00:11.93656 0.000000348 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 12:00:12.93579 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 12:00:13.93443 0.000000352 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 12:00:14.93255 0.000000352 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 12:00:15.93164 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 12:00:16.94024 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 12:00:17.93919 0.000000348 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 12:00:18.93782 0.000000350 0.000003758 0.000002065 BadVal BadVal BadVal BadVal
11/18/2013 12:00:19.93656 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 12:00:20.93610 0.000000348 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 12:00:21.94365 0.000000348 0.000003736 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 12:00:22.94271 0.000000350 0.000003758 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 12:00:23.94121 0.000000350 0.000003758 0.000002043 BadVal BadVal BadVal BadVal
11/18/2013 12:00:24.94077 0.000000348 0.000003802 0.000002021 BadVal BadVal BadVal BadVal
11/18/2013 12:00:25.101301 0.000000348 0.000003736 0.000002000 29.75 BadVal BadVal BadVal
11/18/2013 12:00:26.101770 0.000000352 0.000003758 0.000002043 28.61 BadVal BadVal BadVal
11/18/2013 12:00:27.628314 0.000000348 0.000003758 0.000002043 28.61 BadVal BadVal BadVal
11/18/2013 12:00:28.628508 0.000000352 0.000003758 0.000002021 28.61 BadVal BadVal BadVal
11/18/2013 12:00:29.629329 0.000000350 0.000003758 0.000002043 28.61 0.48 BadVal BadVal
11/18/2013 12:00:30.629181 0.000000350 0.000003758 0.000002043 29.75 0.48 BadVal BadVal
11/18/2013 12:00:31.630128 0.000000348 0.000003736 0.000002043 29.75 0.48 BadVal BadVal
11/18/2013 12:00:32.629945 0.000000348 0.000003758 0.000002043 28.61 0.48 BadVal BadVal
11/18/2013 12:00:33.629793 0.000000348 0.000003736 0.000002043 28.61 0.48 BadVal BadVal
11/18/2013 12:00:34.658234 0.000000348 0.000003758 0.000002043 28.61 0.48 BadVal BadVal
11/18/2013 12:00:35.657546 0.000000348 0.000003758 0.000002043 28.61 0.48 BadVal BadVal
11/18/2013 12:00:36.657419 0.000000348 0.000003758 0.000002087 27.46 0.48 6.10 BadVal
11/18/2013 12:00:37.657311 0.000000348 0.000003736 0.000002043 27.46 0.48 6.10 BadVal
11/18/2013 12:00:38.657194 0.000000348 0.000003736 0.000002043 29.75 0.48 6.10 BadVal
11/18/2013 12:00:39.658139 0.000000348 0.000003736 0.000002043 29.75 0.48 6.10 BadVal
11/18/2013 12:00:42.44788 0.000000348 0.000003758 0.000002021 28.61 0.48 6.10 BadVal
11/18/2013 12:00:43.44671 0.000000348 0.000003758 0.000002021 28.61 0.48 6.10 0.01
11/18/2013 12:00:44.44499 0.000000348 0.000003758 0.000002065 28.61 0.48 5.49 0.01
11/18/2013 12:00:45.44528 0.000000350 0.000003758 0.000002065 28.61 0.48 5.49 0.01
11/18/2013 12:00:46.44300 0.000000348 0.000003758 0.000002065 28.61 0.48 6.10 0.01
11/18/2013 12:00:47.44150 0.000000346 0.000003758 0.000002065 28.61 0.48 6.10 0.01
11/18/2013 12:00:48.43366 0.000000346 0.000003758 0.000002065 29.75 0.48 6.10 0.01
11/18/2013 12:00:49.44059 0.000000348 0.000003758 0.000002065 29.75 0.48 6.10 0.01
11/18/2013 12:00:50.43894 0.000000350 0.000003758 0.000002043 28.61 0.48 6.10 0.01
11/18/2013 12:00:51.43674 0.000000348 0.000003758 0.000002065 28.61 0.48 6.10 0.01
11/18/2013 12:00:52.44636 0.000000348 0.000003758 0.000002065 28.61 0.48 6.10 0.01
11/18/2013 12:00:53.44585 0.000000348 0.000003758 0.000002065 28.61 0.48 6.10 0.01
11/18/2013 12:00:54.45302 0.000000352 0.000003758 0.000002065 28.61 0.48 6.10 0.01
11/18/2013 12:00:55.44355 0.000000352 0.000003758 0.000002043 28.61 0.48 6.10 0.01
11/18/2013 12:00:56.44002 0.000000352 0.000003780 0.000002065 29.75 0.48 6.10 0.01
11/18/2013 12:00:57.44147 0.000000352 0.000003736 0.000002065 29.75 0.48 6.10 0.01
11/18/2013 12:00:58.44820 0.000000348 0.000003758 0.000002065 29.75 0.48 6.10 0.01
11/18/2013 12:00:59.44598 0.000000348 0.000003758 0.000002021 29.75 0.48 6.10 0.01
11/18/2013 12:01:00.44519 0.000000350 0.000003758 0.000002021 29.75 0.48 6.10 0.01
11/18/2013 12:01:01.44421 0.000000348 0.000003758 0.000002000 29.75 0.48 6.10 0.01
11/18/2013 12:01:02.44217 0.000000348 0.000003758 0.000002000 29.75 0.48 6.10 0.01
11/18/2013 12:01:03.44128 0.000000348 0.000003736 0.000002043 29.75 0.48 6.10 0.01
11/18/2013 12:01:04.43981 0.000000348 0.000003736 0.000002021 27.46 0.48 6.10 0.01
11/18/2013 12:01:05.43875 0.000000348 0.000003736 0.000002043 27.46 0.48 6.10 0.01
11/18/2013 12:01:06.43730 0.000000348 0.000003758 0.000002065 28.61 0.48 6.10 0.01
11/18/2013 12:01:07.43676 0.000000348 0.000003758 0.000002065 28.61 0.48 6.10 0.01
11/18/2013 12:01:08.43483 0.000000348 0.000003758 0.000002065 28.61 0.48 6.10 0.01
11/18/2013 12:01:10.527029 0.000000350 0.000003758 0.000002043 27.46 0.48 6.10 0.01
11/18/2013 12:01:11.526244 0.000000348 0.000003758 0.000002043 29.75 0.48 6.10 0.01
11/18/2013 12:01:12.526954 0.000000348 0.000003736 0.000002065 29.75 0.48 6.10 0.01
11/18/2013 12:01:13.526789 0.000000348 0.000003736 0.000002065 28.61 0.48 6.10 0.01
11/18/2013 12:01:14.526783 0.000000348 0.000003758 0.000002043 28.61 0.48 6.10 0.01
11/18/2013 12:01:15.526585 0.000000350 0.000003736 0.000002065 29.75 0.48 6.10 0.01
11/18/2013 12:01:17.268629 0.000000350 0.000003758 0.000002043 27.46 0.48 6.10 0.01
11/18/2013 12:01:18.269323 0.000000348 0.000003758 0.000002065 27.46 0.48 6.10 0.01
11/18/2013 12:01:19.269066 0.000000348 0.000003758 0.000002043 29.75 0.48 6.10 0.01
11/18/2013 12:01:20.269024 0.000000348 0.000003758 0.000002043 29.75 0.48 6.10 0.01
11/18/2013 12:01:21.268809 0.000000348 0.000003758 0.000002065 27.46 0.48 6.10 0.01
11/18/2013 12:01:22.268839 0.000000350 0.000003736 0.000002043 8.01 0.36 6.10 0.01
11/18/2013 12:01:23.269588 0.000000350 0.000003758 0.000002065 8.01 0.36 6.10 0.01
11/18/2013 12:01:24.269513 0.000000348 0.000003758 0.000002043 8.01 0.73 6.10 0.01
11/18/2013 12:01:25.269349 0.000000346 0.000003736 0.000002021 8.01 0.73 6.10 0.01
11/18/2013 12:01:26.269482 0.000000346 0.000003758 0.000002043 5.72 0.61 6.10 0.01
11/18/2013 12:01:27.270296 0.000000350 0.000003758 0.000002043 5.72 0.61 6.10 0.01
11/18/2013 12:01:28.269961 0.000000350 0.000003736 0.000002043 10.29 0.85 6.10 0.01
11/18/2013 12:01:29.269871 0.000000348 0.000003758 0.000002043 10.29 0.85 6.10 0.01
11/18/2013 12:01:30.269726 0.000000350 0.000003736 0.000002043 8.01 0.85 6.10 0.01
11/18/2013 12:01:31.369724 0.000000348 0.000003780 0.000002043 8.01 0.85 6.10 0.01
11/18/2013 12:01:32.369423 0.000000348 0.000003758 0.000002065 10.29 0.73 6.10 0.01
11/18/2013 12:01:33.370375 0.000000350 0.000003758 0.000002065 10.29 0.73 6.10 0.01
11/18/2013 12:01:34.370213 0.000000348 0.000003780 0.000002043 11.44 0.73 6.10 0.01
11/18/2013 12:01:35.370088 0.000000350 0.000003780 0.000002043 11.44 0.73 6.10 0.01
11/18/2013 12:01:36.369921 0.000000350 0.000003758 0.000002043 4.57 0.85 6.10 0.01
11/18/2013 12:01:37.585188 0.000000348 0.000003758 0.000002043 4.57 0.85 6.10 0.01
... 15829 more lines ...
|
547
|
Monday, November 18, 2013, 16:44 |
Bevan Moss | Conditioning Station | Standard Operation | | | Testing of cooling package |
I have tested all of the compressors for TCS. All are operational and cycled as they should. |
550
|
Tuesday, November 19, 2013, 16:55 |
Bevan Moss | Conditioning Station | Standard Operation | TM2 | | HV conditioning |
HV conditioning of TM2 continued today. At ~53.5kV the module started to spark externally to the HV cover. This is the first time the cover has been tested to this high of a voltage due to the limitations of the modules. As such for the conditioning station to achieve 60kV while keeping the target hall open of personnel the cover and cage will need to be modified. To continue with HV testing of TM2 the cover was removed and the target hall locked out as per the procedure covered in Document 70039 "ISAC Conditioning Station: First Lock-Out and High-Voltage "ON". With the cover off the module was able to achieve ~56.7kV without the Einzel lens on. Eliminate the Einzel lens as a potential source of sparking it was brought to 25kV while the module continued conditioning. Maximum voltage achieve under this configuration was 57.1kV. Increments are now 50V steps. The Canbus continues to be very sensitive to any sparking and is slowing the process down considerably. It does however appear that the module is continuing to condition. I will continue tomorrow after the module moves to attempt to reach 60kV and hold it for 1 hour. |
Attachment 1: external_sparking.JPG
|
|
Attachment 2: 19112013_HV_conditioning_TM2.JPG
|
|
Attachment 3: 19112013_TM2_No_source_tray_HV_Test_no_cover.txt
|
Time TCS:IG1:RDVAC [T] TCS:IG2:RDVAC [T] TCS:IG1S:RDVAC [T] TCS:BIAS:RDVOL [V] TCS:BIAS:RDCUR [uA] TCS:EL:RDVOL [V] TCS:EL:RDCUR [mA]
11/19/2013 14:05:53.354722 0.000000302 0.000003362 0.000001516 11.44 0.18 4.88 0.03
11/19/2013 14:05:54.354616 0.000000300 0.000003362 0.000001516 18.31 0.18 4.88 0.03
11/19/2013 14:05:55.354491 0.000000304 0.000003340 0.000001516 18.31 0.18 4.88 0.03
11/19/2013 14:05:56.355471 0.000000304 0.000003340 0.000001538 18.31 0.18 4.88 0.03
11/19/2013 14:05:57.355260 0.000000304 0.000003362 0.000001538 21.74 0.18 4.88 0.03
11/19/2013 14:05:58.356309 0.000000304 0.000003362 0.000001516 21.74 0.18 4.88 0.03
11/19/2013 14:05:59.355939 0.000000304 0.000003362 0.000001516 25.17 0.30 4.88 0.03
11/19/2013 14:06:00.356012 0.000000302 0.000003362 0.000001538 25.17 0.30 4.88 0.03
11/19/2013 14:06:01.356728 0.000000300 0.000003362 0.000001538 26.32 0.30 4.88 0.03
11/19/2013 14:06:02.356720 0.000000304 0.000003362 0.000001516 26.32 0.30 4.88 0.03
11/19/2013 14:06:03.356444 0.000000304 0.000003362 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:06:04.357380 0.000000302 0.000003362 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:06:05.357218 0.000000302 0.000003362 0.000001538 26.32 0.30 5.49 0.03
11/19/2013 14:06:06.358329 0.000000306 0.000003362 0.000001538 26.32 0.30 5.49 0.03
11/19/2013 14:06:07.359140 0.000000302 0.000003384 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:08.358814 0.000000302 0.000003340 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:06:09.358730 0.000000306 0.000003362 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:06:10.358605 0.000000304 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:11.359545 0.000000302 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:12.359312 0.000000302 0.000003362 0.000001516 27.46 0.30 5.49 0.03
11/19/2013 14:06:13.359243 0.000000304 0.000003384 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:06:14.359103 0.000000302 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:15.359040 0.000000304 0.000003362 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:06:16.359868 0.000000302 0.000003362 0.000001516 27.46 0.30 5.49 0.03
11/19/2013 14:06:17.359724 0.000000304 0.000003362 0.000001516 27.46 0.30 5.49 0.03
11/19/2013 14:06:18.359599 0.000000302 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:19.359585 0.000000304 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:20.359298 0.000000304 0.000003384 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:06:21.360304 0.000000304 0.000003384 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:06:22.360120 0.000000304 0.000003362 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:06:23.359987 0.000000302 0.000003340 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:24.359901 0.000000302 0.000003362 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:06:25.359840 0.000000304 0.000003362 0.000001516 27.46 0.30 5.49 0.03
11/19/2013 14:06:26.360642 0.000000300 0.000003340 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:06:27.360471 0.000000300 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:28.359710 0.000000306 0.000003340 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:29.360203 0.000000304 0.000003362 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:06:30.360083 0.000000302 0.000003362 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:06:31.360917 0.000000302 0.000003340 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:32.360825 0.000000302 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:33.360728 0.000000304 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:34.360551 0.000000304 0.000003362 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:06:35.360556 0.000000304 0.000003340 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:06:36.361435 0.000000302 0.000003340 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:06:37.361170 0.000000302 0.000003384 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:06:38.361078 0.000000306 0.000003340 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:06:39.360960 0.000000304 0.000003362 0.000001560 28.61 0.30 5.49 0.03
11/19/2013 14:06:40.360827 0.000000302 0.000003362 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:06:41.361763 0.000000302 0.000003362 0.000001560 27.46 0.30 5.49 0.03
11/19/2013 14:06:42.362590 0.000000302 0.000003362 0.000001560 28.61 0.30 5.49 0.03
11/19/2013 14:06:43.362450 0.000000304 0.000003362 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:06:44.362283 0.000000304 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:45.363190 0.000000302 0.000003340 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:06:46.364029 0.000000302 0.000003362 0.000001494 28.61 0.30 5.49 0.03
11/19/2013 14:06:47.363963 0.000000302 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:48.363929 0.000000304 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:49.364676 0.000000304 0.000003340 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:50.364535 0.000000304 0.000003340 0.000001560 27.46 0.30 5.49 0.03
11/19/2013 14:06:51.365463 0.000000300 0.000003362 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:06:52.366307 0.000000300 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:06:53.366193 0.000000308 0.000003362 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:06:54.366069 0.000000304 0.000003340 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:06:55.365938 0.000000306 0.000003362 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:06:56.366809 0.000000304 0.000003362 0.000001516 29.75 0.30 5.49 0.03
11/19/2013 14:06:57.366654 0.000000304 0.000003362 0.000001538 29.75 0.30 5.49 0.03
11/19/2013 14:06:58.367523 0.000000304 0.000003384 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:06:59.367430 0.000000304 0.000003362 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:07:00.367278 0.000000304 0.000003340 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:07:01.367220 0.000000306 0.000003340 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:07:02.367054 0.000000302 0.000003340 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:07:03.366900 0.000000302 0.000003340 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:07:04.366837 0.000000304 0.000003340 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:07:05.366830 0.000000304 0.000003362 0.000001560 27.46 0.30 5.49 0.03
11/19/2013 14:07:06.367680 0.000000300 0.000003340 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:07:07.367422 0.000000302 0.000003362 0.000001494 28.61 0.30 5.49 0.03
11/19/2013 14:07:08.367307 0.000000308 0.000003362 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:07:09.367180 0.000000302 0.000003362 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:07:10.367078 0.000000304 0.000003362 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:07:11.368047 0.000000302 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:07:12.367770 0.000000302 0.000003362 0.000001560 28.61 0.30 5.49 0.03
11/19/2013 14:07:13.367637 0.000000302 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:07:14.367544 0.000000302 0.000003362 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:07:15.367429 0.000000304 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:07:16.368326 0.000000302 0.000003362 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:07:17.368176 0.000000306 0.000003362 0.000001516 27.46 0.30 5.49 0.03
11/19/2013 14:07:18.368039 0.000000302 0.000003362 0.000001538 27.46 0.30 5.49 0.03
11/19/2013 14:07:19.367933 0.000000302 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:07:20.367835 0.000000302 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:07:21.368721 0.000000300 0.000003340 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:07:22.368549 0.000000302 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:07:23.368413 0.000000306 0.000003384 0.000001516 28.61 0.30 5.49 0.03
11/19/2013 14:07:24.368292 0.000000302 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:07:25.368197 0.000000304 0.000003362 0.000001538 29.75 0.30 5.49 0.03
11/19/2013 14:07:26.369153 0.000000304 0.000003362 0.000001538 29.75 0.30 5.49 0.03
11/19/2013 14:07:27.369877 0.000000304 0.000003362 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:07:28.369786 0.000000302 0.000003340 0.000001538 28.61 0.30 5.49 0.03
11/19/2013 14:07:29.369767 0.000000306 0.000003340 0.000001538 29.75 0.30 5.49 0.03
11/19/2013 14:07:30.369531 0.000000302 0.000003362 0.000001516 29.75 0.30 5.49 0.03
11/19/2013 14:07:31.370442 0.000000304 0.000003362 0.000001560 29.75 0.30 5.49 0.03
11/19/2013 14:07:32.370291 0.000000302 0.000003362 0.000001538 29.75 0.30 5.49 0.03
11/19/2013 14:07:33.370168 0.000000304 0.000003362 0.000001560 28.61 0.30 5.49 0.03
11/19/2013 14:07:34.369996 0.000000304 0.000003362 0.000001560 28.61 0.30 5.49 0.03
11/19/2013 14:07:35.369937 0.000000306 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:07:36.370838 0.000000306 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:07:37.370668 0.000000302 0.000003362 0.000001560 29.75 0.42 5.49 0.03
11/19/2013 14:07:38.370749 0.000000304 0.000003362 0.000001582 29.75 0.42 5.49 0.03
11/19/2013 14:07:39.371415 0.000000302 0.000003384 0.000001560 29.75 0.42 5.49 0.03
11/19/2013 14:07:40.371286 0.000000304 0.000003384 0.000001516 29.75 0.42 5.49 0.03
11/19/2013 14:07:41.372115 0.000000304 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:07:42.371999 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:07:43.371914 0.000000306 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:07:44.371842 0.000000304 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:07:45.371657 0.000000302 0.000003362 0.000001560 28.61 0.42 5.49 0.03
11/19/2013 14:07:46.372526 0.000000302 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:07:47.372407 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:07:48.372325 0.000000306 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:07:49.372151 0.000000306 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:07:50.373021 0.000000302 0.000003362 0.000001560 28.61 0.42 5.49 0.03
11/19/2013 14:07:51.373956 0.000000304 0.000003362 0.000001560 27.46 0.42 5.49 0.03
11/19/2013 14:07:52.373746 0.000000304 0.000003384 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:07:53.373706 0.000000304 0.000003362 0.000001538 29.75 0.42 5.49 0.03
11/19/2013 14:07:54.373541 0.000000302 0.000003362 0.000001538 29.75 0.42 5.49 0.03
11/19/2013 14:07:55.373384 0.000000308 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:07:56.373439 0.000000304 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:07:57.373107 0.000000304 0.000003362 0.000001538 29.75 0.42 5.49 0.03
11/19/2013 14:07:58.373036 0.000000306 0.000003384 0.000001538 29.75 0.42 5.49 0.03
11/19/2013 14:07:59.372863 0.000000306 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:00.372736 0.000000306 0.000003384 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:01.373696 0.000000306 0.000003362 0.000001538 27.46 0.42 5.49 0.03
11/19/2013 14:08:02.373648 0.000000302 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:03.374534 0.000000300 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:04.375263 0.000000308 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:05.376101 0.000000308 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:06.377066 0.000000297 0.000003362 0.000001494 28.61 0.42 5.49 0.03
11/19/2013 14:08:07.376910 0.000000302 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:08.376768 0.000000304 0.000003318 0.000001560 29.75 0.42 5.49 0.03
11/19/2013 14:08:09.376642 0.000000306 0.000003362 0.000001538 29.75 0.42 5.49 0.03
11/19/2013 14:08:10.376520 0.000000302 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:11.376582 0.000000300 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:12.376266 0.000000300 0.000003340 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:13.376238 0.000000302 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:14.375983 0.000000302 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:15.375886 0.000000304 0.000003384 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:16.375820 0.000000306 0.000003362 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:08:17.375626 0.000000304 0.000003340 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:08:18.375507 0.000000302 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:19.375369 0.000000302 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:20.376380 0.000000304 0.000003362 0.000001538 27.46 0.42 5.49 0.03
11/19/2013 14:08:21.377197 0.000000304 0.000003340 0.000001538 27.46 0.42 5.49 0.03
11/19/2013 14:08:22.377109 0.000000304 0.000003384 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:23.376905 0.000000304 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:24.376781 0.000000304 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:25.377666 0.000000304 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:26.377670 0.000000302 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:27.377464 0.000000302 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:28.378245 0.000000300 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:29.378131 0.000000304 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:30.378152 0.000000304 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:31.377926 0.000000306 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:32.377773 0.000000304 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:33.377608 0.000000304 0.000003340 0.000001560 28.61 0.42 5.49 0.03
11/19/2013 14:08:34.378551 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:35.378336 0.000000304 0.000003340 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:36.379229 0.000000304 0.000003340 0.000001516 29.75 0.42 5.49 0.03
11/19/2013 14:08:37.379143 0.000000306 0.000003340 0.000001516 29.75 0.42 5.49 0.03
11/19/2013 14:08:38.380035 0.000000304 0.000003362 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:08:39.379852 0.000000304 0.000003362 0.000001538 27.46 0.42 5.49 0.03
11/19/2013 14:08:40.379853 0.000000304 0.000003340 0.000001538 29.75 0.42 5.49 0.03
11/19/2013 14:08:41.380667 0.000000300 0.000003362 0.000001516 29.75 0.42 5.49 0.03
11/19/2013 14:08:42.380496 0.000000302 0.000003340 0.000001582 28.61 0.42 5.49 0.03
11/19/2013 14:08:43.381497 0.000000306 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:44.382284 0.000000306 0.000003340 0.000001494 29.75 0.42 5.49 0.03
11/19/2013 14:08:45.382081 0.000000306 0.000003362 0.000001494 29.75 0.42 5.49 0.03
11/19/2013 14:08:46.382969 0.000000306 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:47.382888 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:48.382765 0.000000302 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:49.382582 0.000000302 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:50.382483 0.000000306 0.000003362 0.000001494 28.61 0.42 5.49 0.03
11/19/2013 14:08:51.383390 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:52.383236 0.000000306 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:53.383142 0.000000306 0.000003384 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:08:54.382977 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:55.383033 0.000000302 0.000003362 0.000001560 28.61 0.42 5.49 0.03
11/19/2013 14:08:56.383746 0.000000302 0.000003384 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:08:57.383610 0.000000304 0.000003362 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:08:58.384517 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:08:59.384360 0.000000304 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:00.384233 0.000000306 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:01.384098 0.000000306 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:02.384008 0.000000302 0.000003340 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:03.383841 0.000000302 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:04.384838 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:05.384849 0.000000302 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:06.385552 0.000000306 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:07.385371 0.000000306 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:08.385188 0.000000306 0.000003384 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:09.385098 0.000000302 0.000003362 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:09:10.384964 0.000000302 0.000003362 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:09:11.385810 0.000000302 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:12.385870 0.000000306 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:13.386593 0.000000304 0.000003384 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:14.386563 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:15.386310 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:16.387331 0.000000304 0.000003340 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:17.387108 0.000000300 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:18.386975 0.000000302 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:19.386848 0.000000304 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:20.386684 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:21.387583 0.000000304 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:22.387469 0.000000302 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:23.387339 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:24.387222 0.000000304 0.000003362 0.000001560 28.61 0.42 5.49 0.03
11/19/2013 14:09:25.387167 0.000000306 0.000003362 0.000001538 27.46 0.42 5.49 0.03
11/19/2013 14:09:26.388077 0.000000302 0.000003362 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:09:27.387859 0.000000304 0.000003362 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:09:28.387670 0.000000304 0.000003384 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:09:29.387547 0.000000304 0.000003384 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:30.387462 0.000000302 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:31.388291 0.000000302 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:32.388300 0.000000302 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:33.388040 0.000000302 0.000003384 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:34.387967 0.000000308 0.000003384 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:35.387846 0.000000304 0.000003318 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:36.388795 0.000000306 0.000003384 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:37.389717 0.000000302 0.000003384 0.000001538 27.46 0.42 5.49 0.03
11/19/2013 14:09:38.389408 0.000000302 0.000003340 0.000001538 27.46 0.42 5.49 0.03
11/19/2013 14:09:39.389291 0.000000302 0.000003384 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:40.389397 0.000000304 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:41.390033 0.000000304 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:42.389915 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:43.389825 0.000000306 0.000003384 0.000001516 29.75 0.42 5.49 0.03
11/19/2013 14:09:44.389697 0.000000304 0.000003384 0.000001516 29.75 0.42 5.49 0.03
11/19/2013 14:09:45.389571 0.000000302 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:46.390485 0.000000302 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:47.391343 0.000000304 0.000003340 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:48.391159 0.000000304 0.000003340 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:49.391072 0.000000302 0.000003384 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:50.390947 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:51.391775 0.000000304 0.000003362 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:09:52.391202 0.000000304 0.000003362 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:09:53.391826 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:54.392406 0.000000304 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:55.392281 0.000000304 0.000003384 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:09:56.393178 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:57.393028 0.000000304 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:09:58.392904 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:09:59.392947 0.000000302 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:10:00.392649 0.000000302 0.000003384 0.000001516 29.75 0.42 5.49 0.03
11/19/2013 14:10:01.393587 0.000000302 0.000003340 0.000001538 29.75 0.42 5.49 0.03
11/19/2013 14:10:02.393437 0.000000300 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:10:03.393427 0.000000304 0.000003384 0.000001494 28.61 0.42 5.49 0.03
11/19/2013 14:10:04.393216 0.000000302 0.000003362 0.000001494 29.75 0.42 5.49 0.03
11/19/2013 14:10:05.393073 0.000000302 0.000003362 0.000001538 29.75 0.42 5.49 0.03
11/19/2013 14:10:06.392979 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:10:07.392808 0.000000302 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:10:08.392747 0.000000302 0.000003362 0.000001538 27.46 0.42 5.49 0.03
11/19/2013 14:10:09.393600 0.000000302 0.000003340 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:10:10.393431 0.000000304 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:10:11.394321 0.000000304 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:10:12.394255 0.000000302 0.000003362 0.000001582 27.46 0.42 5.49 0.03
11/19/2013 14:10:13.394061 0.000000306 0.000003340 0.000001538 27.46 0.42 5.49 0.03
11/19/2013 14:10:14.393935 0.000000304 0.000003340 0.000001538 27.46 0.42 5.49 0.03
11/19/2013 14:10:15.394815 0.000000304 0.000003340 0.000001538 27.46 0.42 5.49 0.03
11/19/2013 14:10:16.395754 0.000000304 0.000003362 0.000001494 28.61 0.42 5.49 0.03
11/19/2013 14:10:17.395575 0.000000302 0.000003318 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:10:18.395428 0.000000304 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:10:19.395326 0.000000302 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:10:20.396197 0.000000306 0.000003384 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:10:21.397060 0.000000304 0.000003384 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:10:22.397959 0.000000304 0.000003362 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:10:23.397760 0.000000304 0.000003362 0.000001538 27.46 0.42 5.49 0.03
11/19/2013 14:10:24.397660 0.000000304 0.000003362 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:10:25.397555 0.000000306 0.000003362 0.000001538 27.46 0.42 5.49 0.03
11/19/2013 14:10:26.397473 0.000000304 0.000003362 0.000001516 29.75 0.42 5.49 0.03
11/19/2013 14:10:27.397292 0.000000304 0.000003362 0.000001516 29.75 0.42 5.49 0.03
11/19/2013 14:10:28.397275 0.000000304 0.000003362 0.000001538 27.46 0.42 5.49 0.03
11/19/2013 14:10:29.397064 0.000000302 0.000003362 0.000001516 27.46 0.42 5.49 0.03
11/19/2013 14:10:30.397950 0.000000302 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:10:31.398916 0.000000302 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:10:32.398700 0.000000304 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:10:33.398565 0.000000302 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:10:34.398475 0.000000302 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:10:35.399271 0.000000302 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:10:36.400292 0.000000300 0.000003384 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:10:37.400283 0.000000300 0.000003384 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:10:38.399877 0.000000300 0.000003362 0.000001516 29.75 0.42 5.49 0.03
11/19/2013 14:10:39.400813 0.000000302 0.000003362 0.000001516 29.75 0.42 5.49 0.03
11/19/2013 14:10:40.401650 0.000000302 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:10:41.402517 0.000000302 0.000003362 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:10:42.403427 0.000000302 0.000003384 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:10:43.403303 0.000000304 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:10:44.403147 0.000000302 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:10:45.403017 0.000000302 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:10:46.403927 0.000000304 0.000003362 0.000001538 27.46 0.42 5.49 0.03
11/19/2013 14:10:47.403779 0.000000306 0.000003340 0.000001538 27.46 0.42 5.49 0.03
11/19/2013 14:10:48.404762 0.000000304 0.000003340 0.000001516 28.61 0.42 5.49 0.03
11/19/2013 14:10:49.405527 0.000000302 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:10:50.405406 0.000000300 0.000003362 0.000001538 28.61 0.42 5.49 0.03
11/19/2013 14:10:51.405266 0.000000304 0.000003340 0.000001538 28.61 0.42 5.49 0.03
... 9169 more lines ...
|
Attachment 4: 19112013_TM2_No_source_tray_HV_Test.txt
|
Time TCS:IG1:RDVAC [T] TCS:IG2:RDVAC [T] TCS:IG1S:RDVAC [T] TCS:BIAS:RDVOL [V] TCS:BIAS:RDCUR [uA] TCS:EL:RDVOL [V] TCS:EL:RDCUR [mA]
11/19/2013 10:41:01.318511 0.000000302 BadVal BadVal BadVal BadVal BadVal BadVal
11/19/2013 10:41:02.317614 0.000000304 BadVal BadVal BadVal BadVal BadVal BadVal
11/19/2013 10:41:03.317465 0.000000304 BadVal BadVal BadVal BadVal BadVal BadVal
11/19/2013 10:41:04.317355 0.000000304 BadVal BadVal BadVal BadVal BadVal BadVal
11/19/2013 10:41:05.402487 0.000000304 BadVal BadVal BadVal BadVal BadVal BadVal
11/19/2013 10:41:06.403328 0.000000302 BadVal BadVal BadVal BadVal BadVal BadVal
11/19/2013 10:41:07.403004 0.000000302 BadVal BadVal BadVal BadVal BadVal BadVal
11/19/2013 10:41:08.402906 0.000000302 BadVal BadVal BadVal BadVal BadVal BadVal
11/19/2013 10:41:09.402698 0.000000302 BadVal BadVal BadVal BadVal BadVal BadVal
11/19/2013 10:41:10.402606 0.000000304 0.000003384 BadVal BadVal BadVal BadVal BadVal
11/19/2013 10:41:11.403668 0.000000304 0.000003384 BadVal BadVal BadVal BadVal BadVal
11/19/2013 10:41:12.404343 0.000000302 0.000003384 BadVal BadVal BadVal BadVal BadVal
11/19/2013 10:41:13.404485 0.000000306 0.000003384 BadVal BadVal BadVal BadVal BadVal
11/19/2013 10:41:14.404091 0.000000300 0.000003384 BadVal BadVal BadVal BadVal BadVal
11/19/2013 10:41:15.404261 0.000000302 0.000003384 BadVal BadVal BadVal BadVal BadVal
11/19/2013 10:41:16.404970 0.000000300 0.000003384 BadVal BadVal BadVal BadVal BadVal
11/19/2013 10:41:17.405939 0.000000302 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:18.405596 0.000000304 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:19.405470 0.000000304 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:41:20.405591 0.000000304 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:21.405361 0.000000302 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:22.405110 0.000000304 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:41:23.404945 0.000000304 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:24.405004 0.000000302 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:41:25.404869 0.000000304 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:41:26.405664 0.000000304 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:27.405452 0.000000304 0.000003384 0.000001538 BadVal BadVal BadVal BadVal
11/19/2013 10:41:28.405690 0.000000302 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:29.406257 0.000000302 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:30.406107 0.000000302 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:41:31.407176 0.000000304 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:41:32.406962 0.000000302 0.000003362 0.000001538 BadVal BadVal BadVal BadVal
11/19/2013 10:41:33.407750 0.000000302 0.000003384 0.000001538 BadVal BadVal BadVal BadVal
11/19/2013 10:41:34.407588 0.000000304 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:35.407461 0.000000306 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:41:36.407365 0.000000304 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:37.407206 0.000000302 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:41:38.407144 0.000000302 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:41:39.406973 0.000000304 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:40.406863 0.000000302 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:41.407851 0.000000302 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:41:42.408741 0.000000304 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:41:43.408455 0.000000302 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:44.408363 0.000000300 0.000003340 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:45.408507 0.000000302 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:41:46.409102 0.000000302 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:47.408996 0.000000304 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:48.409016 0.000000304 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:49.408903 0.000000304 0.000003384 0.000001516 BadVal BadVal BadVal BadVal
11/19/2013 10:41:50.408651 0.000000302 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:51.409561 0.000000302 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:52.409366 0.000000302 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:53.671050 0.000000302 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:54.675348 0.000000306 0.000003384 0.000001538 BadVal BadVal BadVal BadVal
11/19/2013 10:41:55.675949 0.000000300 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:41:56.676845 0.000000304 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:41:57.676692 0.000000300 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:58.676230 0.000000302 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:41:59.677764 0.000000302 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:00.677259 0.000000302 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:01.678167 0.000000304 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:02.679007 0.000000304 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:03.678882 0.000000304 0.000003384 0.000001538 BadVal BadVal BadVal BadVal
11/19/2013 10:42:04.679802 0.000000302 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:05.679736 0.000000300 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:06.680510 0.000000300 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:07.680434 0.000000302 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:08.680253 0.000000302 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:09.680303 0.000000302 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:10.680002 0.000000302 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:11.706336 0.000000304 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:12.706772 0.000000302 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:13.706749 0.000000302 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:14.707652 0.000000302 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:15.708502 0.000000302 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:16.709291 0.000000306 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:17.709139 0.000000304 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:18.709013 0.000000302 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:19.708875 0.000000302 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:20.708773 0.000000300 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:21.708676 0.000000300 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:22.708491 0.000000300 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:23.708396 0.000000304 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:24.708237 0.000000304 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:25.708129 0.000000304 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:26.709068 0.000000306 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:27.709892 0.000000304 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:28.709771 0.000000302 0.000003362 0.000001538 BadVal BadVal BadVal BadVal
11/19/2013 10:42:29.709605 0.000000302 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:30.709518 0.000000300 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:31.710437 0.000000302 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:32.710268 0.000000304 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:33.710139 0.000000304 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:34.710975 0.000000304 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:35.710972 0.000000308 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:36.711931 0.000000302 0.000003340 0.000001604 BadVal BadVal BadVal BadVal
11/19/2013 10:42:37.769372 0.000000304 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:38.769499 0.000000302 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:39.769426 0.000000304 0.000003384 0.000001538 BadVal BadVal BadVal BadVal
11/19/2013 10:42:40.768586 0.000000302 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:41.769111 0.000000302 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:42.769018 0.000000304 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:43.768903 0.000000302 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:44.768752 0.000000300 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:45.768651 0.000000302 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:46.769518 0.000000304 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:47.769367 0.000000304 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:48.769247 0.000000306 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:49.769168 0.000000306 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:50.769023 0.000000302 0.000003384 0.000001538 BadVal BadVal BadVal BadVal
11/19/2013 10:42:51.769882 0.000000302 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:52.769737 0.000000302 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:53.851442 0.000000304 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:54.851454 0.000000304 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:42:55.851327 0.000000304 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:56.852247 0.000000304 0.000003340 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:57.852146 0.000000302 0.000003340 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:58.851996 0.000000300 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:42:59.851955 0.000000300 0.000003362 0.000001538 BadVal BadVal BadVal BadVal
11/19/2013 10:43:00.851730 0.000000304 0.000003362 0.000001538 BadVal BadVal BadVal BadVal
11/19/2013 10:43:01.851594 0.000000304 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:43:02.851462 0.000000304 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:43:03.851516 0.000000300 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:43:04.851223 0.000000302 0.000003384 0.000001538 BadVal BadVal BadVal BadVal
11/19/2013 10:43:05.851091 0.000000300 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:43:06.850946 0.000000300 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:43:07.850844 0.000000304 0.000003384 0.000001538 BadVal BadVal BadVal BadVal
11/19/2013 10:43:08.851722 0.000000304 0.000003384 0.000001538 BadVal BadVal BadVal BadVal
11/19/2013 10:43:10.796013 0.000000304 0.000003384 0.000001538 BadVal BadVal BadVal BadVal
11/19/2013 10:43:11.796351 0.000000302 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:43:12.796332 0.000000302 0.000003362 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:43:13.796099 0.000000302 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:43:14.795971 0.000000302 0.000003384 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:43:15.795875 0.000000304 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:43:16.795690 0.000000304 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:43:17.795627 0.000000302 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:43:18.796444 0.000000302 0.000003362 0.000001560 BadVal BadVal BadVal BadVal
11/19/2013 10:43:19.796344 0.000000300 0.000003384 0.000001582 BadVal BadVal BadVal BadVal
11/19/2013 10:43:20.796217 0.000000300 0.000003384 0.000001560 26.32 BadVal BadVal BadVal
11/19/2013 10:43:21.797113 0.000000302 0.000003362 0.000001560 26.32 BadVal BadVal BadVal
11/19/2013 10:43:22.796966 0.000000304 0.000003362 0.000001560 26.32 BadVal BadVal BadVal
11/19/2013 10:43:23.796853 0.000000302 0.000003384 0.000001560 26.32 BadVal BadVal BadVal
11/19/2013 10:43:24.797719 0.000000304 0.000003384 0.000001560 25.17 0.48 BadVal BadVal
11/19/2013 10:43:25.797602 0.000000302 0.000003362 0.000001538 25.17 0.48 BadVal BadVal
11/19/2013 10:43:26.797843 0.000000302 0.000003384 0.000001538 25.17 0.48 BadVal BadVal
11/19/2013 10:43:27.797315 0.000000302 0.000003384 0.000001582 25.17 0.48 BadVal BadVal
11/19/2013 10:43:28.797410 0.000000304 0.000003362 0.000001560 25.17 0.48 BadVal BadVal
11/19/2013 10:43:29.797095 0.000000302 0.000003362 0.000001538 25.17 0.48 BadVal BadVal
11/19/2013 10:43:31.261550 0.000000304 0.000003384 0.000001560 26.32 0.48 BadVal BadVal
11/19/2013 10:43:32.261753 0.000000304 0.000003362 0.000001560 26.32 0.48 6.10 BadVal
11/19/2013 10:43:33.261681 0.000000304 0.000003362 0.000001582 26.32 0.48 6.10 BadVal
11/19/2013 10:43:34.261527 0.000000302 0.000003362 0.000001560 26.32 0.48 6.10 BadVal
11/19/2013 10:43:35.261401 0.000000302 0.000003384 0.000001560 25.17 0.48 6.10 BadVal
11/19/2013 10:43:36.261291 0.000000304 0.000003384 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:43:37.261143 0.000000304 0.000003384 0.000001560 26.32 0.48 6.10 0.01
11/19/2013 10:43:38.357263 0.000000306 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:43:39.357863 0.000000306 0.000003384 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:43:40.357775 0.000000302 0.000003384 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:43:41.358683 0.000000302 0.000003384 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:43:42.358279 0.000000304 0.000003384 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:43:43.358382 0.000000304 0.000003384 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:43:44.358287 0.000000302 0.000003362 0.000001582 26.32 0.48 6.10 0.01
11/19/2013 10:43:45.358241 0.000000302 0.000003362 0.000001538 26.32 0.48 6.10 0.01
11/19/2013 10:43:46.359014 0.000000304 0.000003362 0.000001560 26.32 0.48 6.10 0.01
11/19/2013 10:43:47.358900 0.000000302 0.000003362 0.000001560 26.32 0.48 6.10 0.01
11/19/2013 10:43:48.553700 0.000000302 0.000003362 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:43:49.617444 0.000000302 0.000003362 0.000001538 25.17 0.48 6.10 0.01
11/19/2013 10:43:50.617485 0.000000304 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:43:51.617351 0.000000302 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:43:52.617213 0.000000302 0.000003362 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:43:53.618083 0.000000302 0.000003362 0.000001604 25.17 0.48 6.10 0.01
11/19/2013 10:43:54.693871 0.000000302 0.000003362 0.000001604 26.32 0.48 6.10 0.01
11/19/2013 10:43:55.693812 0.000000302 0.000003362 0.000001582 26.32 0.48 6.10 0.01
11/19/2013 10:43:56.693704 0.000000302 0.000003384 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:43:57.693643 0.000000304 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:43:58.693461 0.000000304 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:43:59.693350 0.000000302 0.000003384 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:00.693181 0.000000302 0.000003384 0.000001604 26.32 0.48 6.10 0.01
11/19/2013 10:44:01.694099 0.000000300 0.000003362 0.000001604 26.32 0.48 6.10 0.01
11/19/2013 10:44:02.694989 0.000000302 0.000003362 0.000001604 26.32 0.48 6.10 0.01
11/19/2013 10:44:03.694798 0.000000302 0.000003384 0.000001560 26.32 0.48 6.10 0.01
11/19/2013 10:44:04.695712 0.000000304 0.000003384 0.000001582 26.32 0.48 6.10 0.01
11/19/2013 10:44:05.695611 0.000000306 0.000003362 0.000001538 26.32 0.48 6.10 0.01
11/19/2013 10:44:06.696507 0.000000304 0.000003384 0.000001582 26.32 0.48 6.10 0.01
11/19/2013 10:44:07.697388 0.000000306 0.000003362 0.000001538 26.32 0.48 6.10 0.01
11/19/2013 10:44:08.697233 0.000000300 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:09.697084 0.000000302 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:10.697082 0.000000302 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:11.697791 0.000000302 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:12.698710 0.000000304 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:13.698611 0.000000306 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:14.698625 0.000000302 0.000003384 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:15.698313 0.000000302 0.000003362 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:44:16.699228 0.000000304 0.000003384 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:17.699088 0.000000302 0.000003384 0.000001560 26.32 0.48 6.10 0.01
11/19/2013 10:44:18.700241 0.000000302 0.000003362 0.000001582 26.32 0.48 6.10 0.01
11/19/2013 10:44:19.699985 0.000000304 0.000003362 0.000001516 26.32 0.48 6.10 0.01
11/19/2013 10:44:20.699919 0.000000302 0.000003362 0.000001582 26.32 0.48 6.10 0.01
11/19/2013 10:44:21.700749 0.000000304 0.000003362 0.000001560 26.32 0.48 6.10 0.01
11/19/2013 10:44:22.700485 0.000000302 0.000003362 0.000001560 26.32 0.48 6.10 0.01
11/19/2013 10:44:23.700313 0.000000302 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:24.700161 0.000000302 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:25.700124 0.000000302 0.000003362 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:44:26.826729 0.000000304 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:27.826774 0.000000304 0.000003362 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:44:28.826673 0.000000304 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:29.826713 0.000000302 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:30.826398 0.000000302 0.000003362 0.000001538 25.17 0.48 6.10 0.01
11/19/2013 10:44:31.826314 0.000000304 0.000003362 0.000001538 25.17 0.48 6.10 0.01
11/19/2013 10:44:32.826263 0.000000304 0.000003362 0.000001538 25.17 0.48 6.10 0.01
11/19/2013 10:44:33.826027 0.000000304 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:34.825896 0.000000304 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:35.825862 0.000000304 0.000003384 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:44:36.825771 0.000000300 0.000003384 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:37.826602 0.000000302 0.000003384 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:44:38.826434 0.000000300 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:39.826342 0.000000304 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:40.826153 0.000000304 0.000003362 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:44:41.826047 0.000000304 0.000003362 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:44:42.825889 0.000000304 0.000003384 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:43.825806 0.000000300 0.000003362 0.000001560 26.32 0.48 6.10 0.01
11/19/2013 10:44:44.825678 0.000000302 0.000003362 0.000001560 26.32 0.48 6.10 0.01
11/19/2013 10:44:45.825560 0.000000300 0.000003362 0.000001582 26.32 0.48 6.10 0.01
11/19/2013 10:44:46.825450 0.000000304 0.000003362 0.000001560 26.32 0.48 6.10 0.01
11/19/2013 10:44:47.825323 0.000000302 0.000003362 0.000001560 26.32 0.48 6.10 0.01
11/19/2013 10:44:48.825136 0.000000302 0.000003362 0.000001538 26.32 0.48 6.10 0.01
11/19/2013 10:44:49.825003 0.000000302 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:44:50.825965 0.000000304 0.000003384 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:44:51.825800 0.000000302 0.000003362 0.000001538 26.32 0.48 6.10 0.01
11/19/2013 10:44:52.825783 0.000000300 0.000003340 0.000001560 26.32 0.48 6.10 0.01
11/19/2013 10:44:53.826612 0.000000304 0.000003362 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:44:54.826401 0.000000304 0.000003362 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:44:57.57542 0.000000304 0.000003362 0.000001582 26.32 0.48 6.10 0.01
11/19/2013 10:44:58.57013 0.000000306 0.000003362 0.000001560 26.32 0.48 6.10 0.01
11/19/2013 10:44:59.56879 0.000000302 0.000003362 0.000001538 25.17 0.48 6.10 0.01
11/19/2013 10:45:00.146237 0.000000300 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:45:01.146631 0.000000302 0.000003362 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:45:02.146917 0.000000304 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:45:03.146342 0.000000304 0.000003384 0.000001560 26.32 0.48 6.10 0.01
11/19/2013 10:45:04.146247 0.000000304 0.000003362 0.000001560 26.32 0.48 6.10 0.01
11/19/2013 10:45:05.146129 0.000000302 0.000003362 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:45:06.146037 0.000000300 0.000003362 0.000001560 25.17 0.48 6.10 0.01
11/19/2013 10:45:07.146918 0.000000302 0.000003362 0.000001538 26.32 0.48 6.10 0.01
11/19/2013 10:45:08.146753 0.000000304 0.000003384 0.000001582 26.32 0.48 6.10 0.01
11/19/2013 10:45:09.146640 0.000000302 0.000003362 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:45:10.146497 0.000000304 0.000003362 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:45:11.146466 0.000000304 0.000003362 0.000001582 25.17 0.48 6.10 0.01
11/19/2013 10:45:12.147338 0.000000300 0.000003362 0.000001582 4.57 0.85 6.10 0.01
11/19/2013 10:45:13.147084 0.000000300 0.000003384 0.000001560 4.57 0.85 6.10 0.01
11/19/2013 10:45:14.147016 0.000000302 0.000003362 0.000001582 12.58 0.85 6.10 0.01
11/19/2013 10:45:15.146891 0.000000304 0.000003362 0.000001538 12.58 0.61 6.10 0.01
11/19/2013 10:45:16.146881 0.000000304 0.000003362 0.000001560 16.02 0.42 6.10 0.01
11/19/2013 10:45:17.147759 0.000000302 0.000003362 0.000001560 16.02 0.42 6.10 0.01
11/19/2013 10:45:18.147575 0.000000302 0.000003384 0.000001560 3.43 0.42 6.10 0.01
11/19/2013 10:45:19.147328 0.000000302 0.000003362 0.000001560 3.43 0.42 6.10 0.01
11/19/2013 10:45:20.147232 0.000000300 0.000003384 0.000001582 11.44 0.42 6.10 0.01
11/19/2013 10:45:21.148144 0.000000302 0.000003384 0.000001560 14.87 0.48 6.10 0.01
11/19/2013 10:45:22.149018 0.000000304 0.000003362 0.000001560 18.31 0.48 6.10 0.01
11/19/2013 10:45:23.150194 0.000000306 0.000003384 0.000001560 18.31 0.48 6.10 0.01
11/19/2013 10:45:24.150804 0.000000302 0.000003362 0.000001582 21.74 0.48 6.10 0.01
11/19/2013 10:45:25.150602 0.000000300 0.000003362 0.000001582 21.74 0.48 6.10 0.01
11/19/2013 10:45:26.150449 0.000000300 0.000003384 0.000001538 21.74 0.48 6.10 0.01
11/19/2013 10:45:27.151409 0.000000304 0.000003384 0.000001560 21.74 0.48 6.10 0.01
11/19/2013 10:45:28.151267 0.000000304 0.000003362 0.000001560 22.88 0.48 6.10 0.01
11/19/2013 10:45:29.151108 0.000000302 0.000003362 0.000001582 22.88 0.48 6.10 0.01
11/19/2013 10:45:30.151084 0.000000300 0.000003362 0.000001560 24.03 0.24 6.10 0.01
11/19/2013 10:45:31.150818 0.000000300 0.000003362 0.000001538 11.44 0.24 6.10 0.01
11/19/2013 10:45:32.151726 0.000000302 0.000003362 0.000001538 510.41 2.92 7.32 0.01
11/19/2013 10:45:33.151733 0.000000302 0.000003362 0.000001538 511.55 0.12 7.32 0.01
11/19/2013 10:45:34.151441 0.000000302 0.000003362 0.000001560 1511.78 3.84 7.93 0.01
11/19/2013 10:45:35.151315 0.000000302 0.000003362 0.000001582 2011.90 2.86 9.15 0.01
11/19/2013 10:45:36.151190 0.000000302 0.000003362 0.000001560 3512.24 12.14 9.76 0.01
11/19/2013 10:45:37.151117 0.000000302 0.000003362 0.000001538 5517.28 11.16 9.76 0.01
11/19/2013 10:45:38.150938 0.000000302 0.000003362 0.000001560 7514.30 17.15 15.25 0.01
11/19/2013 10:45:39.150851 0.000000306 0.000003384 0.000001582 9015.79 12.32 15.25 0.01
11/19/2013 10:45:40.150756 0.000000302 0.000003384 0.000001582 11016.25 15.25 15.25 0.01
11/19/2013 10:45:41.150571 0.000000302 0.000003384 0.000001560 13016.70 17.94 15.25 0.01
11/19/2013 10:45:42.151464 0.000000304 0.000003362 0.000001560 14518.19 19.22 16.47 0.01
11/19/2013 10:45:43.151354 0.000000302 0.000003362 0.000001582 16522.08 22.33 16.47 0.01
11/19/2013 10:45:44.151303 0.000000302 0.000003384 0.000001582 18519.11 23.49 17.09 0.01
11/19/2013 10:45:45.151188 0.000000302 0.000003362 0.000001582 20585.94 28.56 17.09 0.01
11/19/2013 10:45:46.151036 0.000000302 0.000003362 0.000001582 22523.46 30.76 17.70 0.01
11/19/2013 10:45:47.152035 0.000000302 0.000003362 0.000001582 24521.62 28.44 17.70 0.01
11/19/2013 10:45:48.151742 0.000000302 0.000003384 0.000001582 26520.94 31.37 17.09 0.01
11/19/2013 10:45:49.151657 0.000000304 0.000003362 0.000001560 28524.83 31.73 17.70 0.01
11/19/2013 10:45:50.151484 0.000000306 0.000003362 0.000001582 30025.17 35.09 17.70 0.01
11/19/2013 10:45:51.151476 0.000000304 0.000003362 0.000001582 32027.92 40.10 17.70 0.01
11/19/2013 10:45:52.151329 0.000000302 0.000003384 0.000001560 34024.94 42.54 17.70 0.01
11/19/2013 10:45:53.151052 0.000000302 0.000003362 0.000001538 36027.69 38.87 17.70 0.01
11/19/2013 10:45:54.150967 0.000000304 0.000003362 0.000001560 37528.03 43.15 17.70 0.01
11/19/2013 10:45:55.150804 0.000000304 0.000003362 0.000001560 39026.09 43.09 17.09 0.01
11/19/2013 10:45:56.150672 0.000000304 0.000003384 0.000001560 40528.72 43.88 17.09 0.01
11/19/2013 10:45:57.151583 0.000000306 0.000003384 0.000001538 42027.92 45.96 15.25 0.01
11/19/2013 10:45:58.151490 0.000000302 0.000003362 0.000001560 43094.52 46.50 15.25 0.01
11/19/2013 10:45:59.151326 0.000000300 0.000003362 0.000001538 44528.49 48.64 31.12 0.01
11/19/2013 10:46:00.151211 0.000000302 0.000003362 0.000001560 45529.86 49.74 50.65 0.01
11/19/2013 10:46:01.151080 0.000000304 0.000003384 0.000001538 47031.35 51.33 68.36 0.01
11/19/2013 10:46:02.151382 0.000000304 0.000003384 0.000001538 48028.15 54.07 78.73 0.01
11/19/2013 10:46:03.151823 0.000000304 0.000003384 0.000001582 49029.52 54.44 76.90 0.01
... 2782 more lines ...
|
553
|
Wednesday, November 20, 2013, 13:52 |
Bevan Moss | Conditioning Station | Standard Operation | TM2 | | HV testing of TM2 |
HV testing without the cover continued today. With the EINZEL lens at 30kV the target module was able to achieve 58.7kV. At this voltage any spark was reeking havoc on the electronics of the TCS. Vacuum gauges signals were occasionally lost, the thermocouple reading would occasionally be lost, turbo pump would turn off. All of these events caused the interlocks to trip resulting in the power supplies to be turned off. As a result I bypassed majority of the interlocks to prevent this from happening. Ultimately as spark took out the CANBUS controller first for the EINZEL lens and then for the BIAS. David Morris has been contacted to have them replaced. The damage of the CANBUS controller for these power supplies is becoming a common problem and should be investigated for if there is a better solution. During testing I experimented with raising the BIAS to near its limit followed by the EINZEL lens and vise versa. From this it was determined that the EINZEL must be raised to high voltage before the BIAS otherwise sparking occurs. Even a step from 0 to 500 volts on the ENIZEL lens with the BIAS at 55kV caused sparking. There are a few combinations of testing along this line I would like to continue with but initially my though is that the BIAS and ENIZEL do not have sufficient clearance for the BIAS to operate above 55kV. Attached is a screen shot of the strip tool and the data. I have not heard from David yet but I do not expect the TCS to be operational today. |
Attachment 1: 20112013_HV_conditioning_TM2.JPG
|
|
Attachment 2: 20112013_TM2_No_source_tray_HV_Test_no_cover.txt
|
Time TCS:IG1:RDVAC [T] TCS:IG2:RDVAC [T] TCS:IG1S:RDVAC [T] TCS:BIAS:RDVOL [V] TCS:BIAS:RDCUR [uA] TCS:EL:RDVOL [V] TCS:EL:RDCUR [mA]
11/20/2013 10:34:48.414676 0.000000253 0.000003120 BadVal BadVal BadVal BadVal BadVal
11/20/2013 10:34:49.413724 0.000000253 0.000003098 BadVal BadVal BadVal BadVal BadVal
11/20/2013 10:34:50.414780 0.000000256 0.000003120 BadVal BadVal BadVal BadVal BadVal
11/20/2013 10:34:51.414583 0.000000258 0.000003120 BadVal BadVal BadVal BadVal BadVal
11/20/2013 10:34:52.414423 0.000000260 0.000003120 BadVal BadVal BadVal BadVal BadVal
11/20/2013 10:34:53.415237 0.000000258 0.000003120 BadVal BadVal BadVal BadVal BadVal
11/20/2013 10:34:54.416059 0.000000258 0.000003098 BadVal BadVal BadVal BadVal BadVal
11/20/2013 10:34:55.506062 0.000000258 0.000003098 BadVal BadVal BadVal BadVal BadVal
11/20/2013 10:34:56.506814 0.000000258 0.000003120 0.000001230 BadVal BadVal BadVal BadVal
11/20/2013 10:34:57.506697 0.000000256 0.000003098 0.000001230 BadVal BadVal BadVal BadVal
11/20/2013 10:34:58.506780 0.000000258 0.000003120 0.000001230 BadVal BadVal BadVal BadVal
11/20/2013 10:34:59.506511 0.000000260 0.000003120 0.000001252 BadVal BadVal BadVal BadVal
11/20/2013 10:35:00.507321 0.000000258 0.000003120 0.000001230 BadVal BadVal BadVal BadVal
11/20/2013 10:35:01.507271 0.000000258 0.000003120 0.000001230 BadVal BadVal BadVal BadVal
11/20/2013 10:35:02.507196 0.000000256 0.000003098 0.000001230 BadVal BadVal BadVal BadVal
11/20/2013 10:35:03.507937 0.000000256 0.000003098 0.000001230 BadVal BadVal BadVal BadVal
11/20/2013 10:35:04.508830 0.000000256 0.000003098 0.000001230 BadVal BadVal BadVal BadVal
11/20/2013 10:35:05.509815 0.000000253 0.000003120 0.000001230 BadVal BadVal BadVal BadVal
11/20/2013 10:35:06.510560 0.000000258 0.000003120 0.000001208 BadVal BadVal BadVal BadVal
11/20/2013 10:35:07.511442 0.000000260 0.000003120 0.000001208 BadVal BadVal BadVal BadVal
11/20/2013 10:35:08.511382 0.000000258 0.000003098 0.000001208 27.46 BadVal BadVal BadVal
11/20/2013 10:35:09.511380 0.000000260 0.000003120 0.000001252 26.32 BadVal BadVal BadVal
11/20/2013 10:35:10.511063 0.000000258 0.000003120 0.000001252 26.32 BadVal BadVal BadVal
11/20/2013 10:35:11.510938 0.000000256 0.000003098 0.000001252 27.46 BadVal BadVal BadVal
11/20/2013 10:35:12.510832 0.000000256 0.000003120 0.000001208 27.46 BadVal BadVal BadVal
11/20/2013 10:35:13.510688 0.000000258 0.000003120 0.000001208 27.46 0.42 BadVal BadVal
11/20/2013 10:35:14.510527 0.000000258 0.000003098 0.000001230 27.46 0.42 BadVal BadVal
11/20/2013 10:35:15.510410 0.000000258 0.000003120 0.000001230 27.46 0.42 BadVal BadVal
11/20/2013 10:35:16.510321 0.000000256 0.000003098 0.000001230 27.46 0.42 BadVal BadVal
11/20/2013 10:35:17.510214 0.000000256 0.000003098 0.000001230 26.32 0.42 BadVal BadVal
11/20/2013 10:35:18.510025 0.000000256 0.000003120 0.000001230 26.32 0.42 5.49 BadVal
11/20/2013 10:35:19.509934 0.000000256 0.000003120 0.000001230 27.46 0.42 5.49 BadVal
11/20/2013 10:35:20.509860 0.000000256 0.000003120 0.000001230 27.46 0.42 5.49 BadVal
11/20/2013 10:35:21.511728 0.000000258 0.000003120 0.000001230 27.46 0.42 5.49 BadVal
11/20/2013 10:35:22.511632 0.000000260 0.000003120 0.000001230 27.46 0.42 5.49 BadVal
11/20/2013 10:35:23.511584 0.000000256 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:35:24.591659 0.000000258 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:35:25.729112 0.000000253 0.000003098 0.000001208 26.32 0.42 5.49 0.01
11/20/2013 10:35:26.729088 0.000000258 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:35:27.729906 0.000000253 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:35:28.729930 0.000000260 0.000003120 0.000001208 26.32 0.42 5.49 0.01
11/20/2013 10:35:29.729660 0.000000258 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:35:30.729489 0.000000258 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:35:31.729372 0.000000258 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:35:32.730305 0.000000253 0.000003098 0.000001208 26.32 0.42 5.49 0.01
11/20/2013 10:35:33.853965 0.000000256 0.000003120 0.000001252 26.32 0.42 5.49 0.01
11/20/2013 10:35:34.854006 0.000000258 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:35:35.949159 0.000000258 0.000003098 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:35:36.949749 0.000000256 0.000003098 0.000001252 27.46 0.42 5.49 0.01
11/20/2013 10:35:37.949644 0.000000256 0.000003120 0.000001252 27.46 0.42 5.49 0.01
11/20/2013 10:35:38.949535 0.000000256 0.000003120 0.000001208 26.32 0.42 5.49 0.01
11/20/2013 10:35:39.950466 0.000000256 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:35:41.28198 0.000000256 0.000003098 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:35:42.34724 0.000000256 0.000003098 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:35:43.34971 0.000000256 0.000003098 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:35:44.34865 0.000000260 0.000003098 0.000001252 27.46 0.42 5.49 0.01
11/20/2013 10:35:45.34739 0.000000258 0.000003120 0.000001252 26.32 0.42 5.49 0.01
11/20/2013 10:35:46.34576 0.000000258 0.000003098 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:35:47.34484 0.000000256 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:35:48.35380 0.000000258 0.000003120 0.000001208 26.32 0.42 5.49 0.01
11/20/2013 10:35:49.35302 0.000000256 0.000003120 0.000001208 26.32 0.42 5.49 0.01
11/20/2013 10:35:50.36136 0.000000253 0.000003120 0.000001208 26.32 0.42 5.49 0.01
11/20/2013 10:35:51.35938 0.000000253 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:35:52.35873 0.000000260 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:35:53.35802 0.000000260 0.000003098 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:35:54.35700 0.000000260 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:35:55.35442 0.000000260 0.000003098 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:35:56.35354 0.000000256 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:35:57.35238 0.000000258 0.000003098 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:35:58.36111 0.000000253 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:35:59.35984 0.000000260 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:36:00.35832 0.000000260 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:36:01.35689 0.000000260 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:36:02.144399 0.000000256 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:03.144478 0.000000258 0.000003098 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:04.144502 0.000000256 0.000003098 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:36:05.145332 0.000000258 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:36:06.145089 0.000000253 0.000003098 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:07.145209 0.000000260 0.000003098 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:08.145838 0.000000260 0.000003120 0.000001208 26.32 0.42 5.49 0.01
11/20/2013 10:36:09.146802 0.000000260 0.000003098 0.000001208 26.32 0.42 5.49 0.01
11/20/2013 10:36:10.146584 0.000000258 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:36:11.146460 0.000000256 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:36:12.146294 0.000000256 0.000003098 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:36:13.146305 0.000000258 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:36:14.146042 0.000000258 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:36:15.145915 0.000000258 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:36:16.145794 0.000000258 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:17.145705 0.000000256 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:18.145906 0.000000258 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:36:19.145434 0.000000258 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:36:20.145323 0.000000253 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:21.145202 0.000000256 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:22.145076 0.000000260 0.000003098 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:36:23.146118 0.000000260 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:36:24.145825 0.000000258 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:25.155723 0.000000258 0.000003120 0.000001208 26.32 0.42 5.49 0.01
11/20/2013 10:36:26.155570 0.000000258 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:27.155437 0.000000256 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:28.155279 0.000000256 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:36:29.155155 0.000000256 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:36:30.155132 0.000000256 0.000003098 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:36:31.155083 0.000000258 0.000003098 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:36:32.154768 0.000000258 0.000003098 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:36:33.154675 0.000000258 0.000003098 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:36:34.154594 0.000000253 0.000003120 0.000001208 26.32 0.42 5.49 0.01
11/20/2013 10:36:35.155716 0.000000256 0.000003120 0.000001208 26.32 0.42 5.49 0.01
11/20/2013 10:36:36.156306 0.000000258 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:37.156322 0.000000256 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:38.156054 0.000000256 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:36:39.155899 0.000000256 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:36:40.313843 0.000000258 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:41.313756 0.000000253 0.000003120 0.000001208 26.32 0.42 5.49 0.01
11/20/2013 10:36:42.313502 0.000000253 0.000003098 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:43.313407 0.000000256 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:44.313288 0.000000258 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:45.313123 0.000000258 0.000003098 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:46.313028 0.000000256 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:47.312863 0.000000256 0.000003098 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:48.313869 0.000000258 0.000003120 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:49.314706 0.000000256 0.000003098 0.000001230 26.32 0.42 5.49 0.01
11/20/2013 10:36:50.315547 0.000000253 0.000003120 0.000001186 26.32 0.42 5.49 0.01
11/20/2013 10:36:51.315404 0.000000260 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:36:52.315285 0.000000256 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:36:53.316161 0.000000256 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:36:54.316185 0.000000260 0.000003098 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:36:55.317074 0.000000256 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:36:56.316803 0.000000258 0.000003120 0.000001208 27.46 0.42 5.49 0.01
11/20/2013 10:36:57.316655 0.000000256 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:36:58.316082 0.000000256 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:36:59.316373 0.000000256 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:37:00.316469 0.000000256 0.000003120 0.000001230 27.46 0.42 5.49 0.01
11/20/2013 10:37:01.316114 0.000000256 0.000003120 0.000001230 27.46 0.42 0.61 0.01
11/20/2013 10:37:02.316025 0.000000260 0.000003098 0.000001230 26.32 0.42 0.61 0.01
11/20/2013 10:37:03.316970 0.000000260 0.000003120 0.000001230 8.01 0.42 0.61 0.01
11/20/2013 10:37:04.316771 0.000000256 0.000003120 0.000001230 8.01 0.42 0.61 0.01
11/20/2013 10:37:05.316651 0.000000253 0.000003098 0.000001230 9.15 0.79 4.27 0.01
11/20/2013 10:37:06.316545 0.000000256 0.000003098 0.000001230 9.15 0.79 4.27 0.01
11/20/2013 10:37:07.316358 0.000000256 0.000003120 0.000001230 12.58 0.79 6.10 0.01
11/20/2013 10:37:08.317387 0.000000256 0.000003120 0.000001208 12.58 0.79 5.49 0.01
11/20/2013 10:37:09.318100 0.000000256 0.000003076 0.000001208 10.29 0.91 5.49 0.01
11/20/2013 10:37:10.317973 0.000000256 0.000003098 0.000001230 12.58 0.73 5.49 0.01
11/20/2013 10:37:11.317854 0.000000256 0.000003120 0.000001208 12098.87 6.77 30.51 0.01
11/20/2013 10:37:12.317785 0.000000253 0.000003120 0.000001208 9037.53 5.37 8.54 0.01
11/20/2013 10:37:13.318646 0.000000256 0.000003120 0.000001208 6778.43 4.08 8.54 0.01
11/20/2013 10:37:14.318517 0.000000258 0.000003120 0.000001208 5105.28 3.41 3.05 0.01
11/20/2013 10:37:15.318456 0.000000258 0.000003120 0.000001208 3855.57 2.86 3.05 0.01
11/20/2013 10:37:16.319259 0.000000256 0.000003120 0.000001230 2921.72 1.95 0.61 0.01
11/20/2013 10:37:17.319100 0.000000256 0.000003120 0.000001208 2237.35 1.58 0.61 0.01
11/20/2013 10:37:18.320004 0.000000253 0.000003120 0.000001230 1826.50 1.15 1.83 0.01
11/20/2013 10:37:19.320985 0.000000256 0.000003120 0.000001208 1478.59 1.15 1.83 0.01
11/20/2013 10:37:20.320764 0.000000258 0.000003120 0.000001208 1197.07 1.03 3.66 0.01
11/20/2013 10:37:21.320666 0.000000253 0.000003120 0.000001230 969.32 1.03 3.66 0.01
11/20/2013 10:37:22.321547 0.000000258 0.000003120 0.000001230 785.07 1.03 4.27 0.01
11/20/2013 10:37:23.322366 0.000000258 0.000003120 0.000001230 635.15 1.03 4.27 0.01
11/20/2013 10:37:24.322396 0.000000258 0.000003120 0.000001230 513.84 0.91 4.27 0.01
11/20/2013 10:37:25.322137 0.000000256 0.000003120 0.000001230 415.42 0.91 4.27 0.01
11/20/2013 10:37:26.321987 0.000000256 0.000003142 0.000001230 335.31 0.91 4.88 0.01
11/20/2013 10:37:27.321928 0.000000253 0.000003120 0.000001208 270.08 0.91 4.88 0.01
11/20/2013 10:37:28.322956 0.000000256 0.000003098 0.000001230 216.29 0.91 4.88 0.01
11/20/2013 10:37:29.323592 0.000000256 0.000003098 0.000001230 179.67 0.91 4.88 0.01
11/20/2013 10:37:30.323486 0.000000256 0.000003142 0.000001230 141.90 0.91 4.88 0.01
11/20/2013 10:37:31.323450 0.000000258 0.000003120 0.000001230 108.72 0.91 4.88 0.01
11/20/2013 10:37:32.323271 0.000000258 0.000003120 0.000001208 89.26 0.91 4.88 0.01
11/20/2013 10:37:33.324362 0.000000258 0.000003098 0.000001208 70.95 0.91 4.88 0.01
11/20/2013 10:37:34.324005 0.000000256 0.000003120 0.000001208 53.78 0.91 5.49 0.01
11/20/2013 10:37:35.323838 0.000000256 0.000003120 0.000001208 53.78 0.91 1.83 0.01
11/20/2013 10:37:36.324765 0.000000258 0.000003120 0.000001230 33.18 0.91 1.83 0.01
11/20/2013 10:37:37.324726 0.000000258 0.000003120 0.000001230 17.16 0.91 1.22 0.01
11/20/2013 10:37:38.325505 0.000000258 0.000003120 0.000001186 17.16 0.91 1.22 0.01
11/20/2013 10:37:39.325336 0.000000258 0.000003098 0.000001208 4.57 0.36 1.22 0.01
11/20/2013 10:37:40.325348 0.000000260 0.000003120 0.000001230 4.57 0.36 1.22 0.01
11/20/2013 10:37:41.325141 0.000000258 0.000003120 0.000001230 12.58 0.79 1.22 0.01
11/20/2013 10:37:42.325021 0.000000256 0.000003120 0.000001230 12.58 0.79 1.22 0.01
11/20/2013 10:37:43.325865 0.000000253 0.000003120 0.000001230 9.15 0.91 0.61 0.01
11/20/2013 10:37:44.325763 0.000000258 0.000003120 0.000001230 9.15 0.91 0.61 0.01
11/20/2013 10:37:45.325605 0.000000258 0.000003098 0.000001230 10.29 0.91 0.61 0.01
11/20/2013 10:37:46.325508 0.000000253 0.000003120 0.000001230 10.29 0.91 0.61 0.01
11/20/2013 10:37:47.325388 0.000000256 0.000003098 0.000001230 5.72 0.91 998.55 0.04
11/20/2013 10:37:48.326442 0.000000258 0.000003120 0.000001230 5.72 0.91 3000.53 0.09
11/20/2013 10:37:49.326181 0.000000256 0.000003098 0.000001252 8.01 0.91 4500.80 0.23
11/20/2013 10:37:50.325945 0.000000256 0.000003120 0.000001230 8.01 0.91 5001.90 0.23
11/20/2013 10:37:51.326819 0.000000256 0.000003120 0.000001230 5.72 0.91 5568.93 0.23
11/20/2013 10:37:52.326754 0.000000256 0.000003120 0.000001230 5.72 0.91 7003.28 0.23
11/20/2013 10:37:53.327593 0.000000256 0.000003142 0.000001230 5.72 0.91 8003.66 0.23
11/20/2013 10:37:54.327448 0.000000256 0.000003120 0.000001208 13.73 0.91 9502.70 0.53
11/20/2013 10:37:55.328351 0.000000260 0.000003120 0.000001230 13.73 0.91 10506.14 0.35
11/20/2013 10:37:56.328292 0.000000258 0.000003120 0.000001208 17.16 0.91 11781.79 0.38
11/20/2013 10:37:57.328116 0.000000256 0.000003120 0.000001208 17.16 0.91 13006.17 0.44
11/20/2013 10:37:58.328287 0.000000256 0.000003120 0.000001208 9.15 0.91 14505.83 0.44
11/20/2013 10:37:59.328830 0.000000256 0.000003120 0.000001208 8.01 0.91 16503.54 0.58
11/20/2013 10:38:00.328696 0.000000256 0.000003120 0.000001230 8.01 0.91 18006.86 0.58
11/20/2013 10:38:01.328586 0.000000256 0.000003098 0.000001230 12.58 0.91 19506.52 0.69
11/20/2013 10:38:02.329506 0.000000258 0.000003098 0.000001208 12.58 0.91 21007.40 0.79
11/20/2013 10:38:03.330336 0.000000258 0.000003120 0.000001208 12.58 0.91 22507.66 0.79
11/20/2013 10:38:04.330331 0.000000258 0.000003098 0.000001208 4.57 0.91 23011.21 0.80
11/20/2013 10:38:05.330125 0.000000256 0.000003120 0.000001208 4.57 0.91 24009.15 0.81
11/20/2013 10:38:06.329942 0.000000256 0.000003120 0.000001208 12.58 0.79 25008.31 0.81
11/20/2013 10:38:07.329857 0.000000258 0.000003098 0.000001208 12.58 0.79 25010.14 0.81
11/20/2013 10:38:08.330714 0.000000258 0.000003098 0.000001208 5.72 0.91 25010.14 0.81
11/20/2013 10:38:09.330602 0.000000260 0.000003098 0.000001230 5.72 0.91 25011.97 0.81
11/20/2013 10:38:10.330506 0.000000253 0.000003098 0.000001230 5.72 0.73 25011.97 0.81
11/20/2013 10:38:11.331338 0.000000258 0.000003098 0.000001230 5.72 0.73 25011.97 0.81
11/20/2013 10:38:12.331188 0.000000258 0.000003120 0.000001252 6.86 0.73 25011.97 0.81
11/20/2013 10:38:13.332192 0.000000258 0.000003120 0.000001252 6.86 0.73 25011.97 0.81
11/20/2013 10:38:14.331985 0.000000256 0.000003120 0.000001252 10.29 0.85 25011.97 0.81
11/20/2013 10:38:15.331812 0.000000256 0.000003120 0.000001208 10.29 0.85 25012.58 0.81
11/20/2013 10:38:16.331676 0.000000256 0.000003120 0.000001208 13.73 0.85 25012.58 0.81
11/20/2013 10:38:17.331599 0.000000258 0.000003120 0.000001208 13.73 0.85 25012.58 0.81
11/20/2013 10:38:18.331562 0.000000260 0.000003098 0.000001208 10.29 0.85 25012.58 0.81
11/20/2013 10:38:19.331349 0.000000256 0.000003120 0.000001230 10.29 0.85 25012.58 0.81
11/20/2013 10:38:20.331173 0.000000256 0.000003120 0.000001230 8.01 0.85 25012.58 0.81
11/20/2013 10:38:21.331084 0.000000258 0.000003120 0.000001230 10.29 0.85 25012.58 0.81
11/20/2013 10:38:22.331021 0.000000256 0.000003120 0.000001230 10.29 0.85 25012.58 0.81
11/20/2013 10:38:23.330936 0.000000256 0.000003120 0.000001230 9.15 0.85 25012.58 0.81
11/20/2013 10:38:24.330710 0.000000260 0.000003120 0.000001230 9.15 0.85 25012.58 0.81
11/20/2013 10:38:25.330593 0.000000256 0.000003120 0.000001208 8.01 0.73 25012.58 0.81
11/20/2013 10:38:26.330462 0.000000258 0.000003120 0.000001208 8.01 0.73 25012.58 0.81
11/20/2013 10:38:27.330338 0.000000256 0.000003120 0.000001208 10.29 0.91 25012.58 0.81
11/20/2013 10:38:28.331226 0.000000253 0.000003120 0.000001230 10.29 0.91 25012.58 0.81
11/20/2013 10:38:29.331092 0.000000258 0.000003120 0.000001230 13.73 0.91 25012.58 0.81
11/20/2013 10:38:30.331004 0.000000258 0.000003098 0.000001230 13.73 0.91 25012.58 0.81
11/20/2013 10:38:31.330795 0.000000258 0.000003098 0.000001208 6.86 0.91 25012.58 0.81
11/20/2013 10:38:32.330772 0.000000260 0.000003120 0.000001208 6.86 0.91 25012.58 0.81
11/20/2013 10:38:33.331565 0.000000260 0.000003098 0.000001208 9.15 0.79 25012.58 0.81
11/20/2013 10:38:34.331463 0.000000258 0.000003098 0.000001230 9.15 0.79 25012.58 0.81
11/20/2013 10:38:35.331347 0.000000256 0.000003120 0.000001208 14.87 0.61 25012.58 0.81
11/20/2013 10:38:36.331211 0.000000258 0.000003120 0.000001208 14.87 0.61 25012.58 0.81
11/20/2013 10:38:37.331045 0.000000258 0.000003098 0.000001230 13.73 0.85 25012.58 0.81
11/20/2013 10:38:38.330950 0.000000256 0.000003098 0.000001230 13.73 0.85 25012.58 0.81
11/20/2013 10:38:39.330833 0.000000258 0.000003120 0.000001186 6.86 0.73 25012.58 0.81
11/20/2013 10:38:40.330708 0.000000256 0.000003120 0.000001208 6.86 0.73 25012.58 0.81
11/20/2013 10:38:41.330581 0.000000258 0.000003098 0.000001230 9.15 0.85 25012.58 0.81
11/20/2013 10:38:42.330469 0.000000253 0.000003098 0.000001186 6.86 0.73 25012.58 0.81
11/20/2013 10:38:43.331408 0.000000256 0.000003120 0.000001230 6.86 0.85 25012.58 0.81
11/20/2013 10:38:44.331413 0.000000256 0.000003120 0.000001230 1010.52 2.99 25012.58 0.81
11/20/2013 10:38:45.332046 0.000000256 0.000003120 0.000001208 1791.02 5.61 25012.58 0.81
11/20/2013 10:38:46.331944 0.000000258 0.000003098 0.000001230 2512.01 3.90 25012.58 0.81
11/20/2013 10:38:47.331830 0.000000258 0.000003120 0.000001208 3010.98 9.94 25012.58 0.81
11/20/2013 10:38:48.332698 0.000000256 0.000003120 0.000001230 4012.35 4.51 25012.58 0.81
11/20/2013 10:38:49.332575 0.000000258 0.000003120 0.000001230 5012.58 7.75 25012.58 0.81
11/20/2013 10:38:50.332461 0.000000253 0.000003098 0.000001230 6013.96 6.65 25012.58 0.81
11/20/2013 10:38:51.332339 0.000000256 0.000003098 0.000001230 7017.62 11.41 25012.58 0.81
11/20/2013 10:38:52.332236 0.000000256 0.000003098 0.000001230 8514.53 10.43 25012.58 0.81
11/20/2013 10:38:53.333086 0.000000256 0.000003098 0.000001230 9515.90 17.09 25012.58 0.81
11/20/2013 10:38:54.332965 0.000000256 0.000003098 0.000001230 11016.25 14.95 25012.58 0.81
11/20/2013 10:38:55.332817 0.000000256 0.000003120 0.000001230 12019.91 16.35 25012.58 0.81
11/20/2013 10:38:56.332686 0.000000253 0.000003098 0.000001186 13292.51 16.41 25012.58 0.81
11/20/2013 10:38:57.332571 0.000000256 0.000003120 0.000001186 14518.19 15.93 25012.58 0.81
11/20/2013 10:38:58.333428 0.000000256 0.000003120 0.000001230 15519.56 15.68 25012.58 0.81
11/20/2013 10:38:59.333281 0.000000256 0.000003120 0.000001230 16520.94 16.11 25012.58 0.81
11/20/2013 10:39:00.333153 0.000000256 0.000003098 0.000001208 18018.99 24.23 25012.58 0.81
11/20/2013 10:39:01.333093 0.000000253 0.000003142 0.000001208 19022.65 22.70 25012.58 0.81
11/20/2013 10:39:02.333017 0.000000260 0.000003120 0.000001230 20521.85 21.30 25012.58 0.81
11/20/2013 10:39:03.333874 0.000000256 0.000003098 0.000001230 22022.20 23.25 25012.58 0.81
11/20/2013 10:39:04.333737 0.000000258 0.000003098 0.000001208 23469.90 26.06 25012.58 0.81
11/20/2013 10:39:05.333680 0.000000256 0.000003120 0.000001230 24522.77 24.71 25012.58 0.81
11/20/2013 10:39:06.333393 0.000000256 0.000003120 0.000001230 26023.11 28.99 25012.58 0.81
11/20/2013 10:39:07.333417 0.000000258 0.000003120 0.000001208 27523.46 29.35 25012.58 0.81
11/20/2013 10:39:08.334148 0.000000258 0.000003120 0.000001230 29026.09 31.31 25012.58 0.81
11/20/2013 10:39:09.334056 0.000000256 0.000003120 0.000001230 30523.00 32.53 25012.58 0.81
11/20/2013 10:39:10.333927 0.000000253 0.000003120 0.000001230 31525.52 31.67 25012.58 0.81
11/20/2013 10:39:11.333973 0.000000256 0.000003120 0.000001230 32528.03 30.15 25012.58 0.81
11/20/2013 10:39:12.334739 0.000000256 0.000003120 0.000001230 33801.78 36.19 25012.58 0.81
11/20/2013 10:39:13.335695 0.000000253 0.000003120 0.000001230 35026.32 35.33 25012.58 0.81
11/20/2013 10:39:14.335425 0.000000253 0.000003120 0.000001230 36474.02 39.06 25012.58 0.81
11/20/2013 10:39:15.336686 0.000000253 0.000003098 0.000001230 37526.89 38.20 25012.58 0.81
11/20/2013 10:39:16.337275 0.000000258 0.000003098 0.000001230 38527.12 38.20 25012.58 0.81
11/20/2013 10:39:17.338221 0.000000253 0.000003098 0.000001208 39974.82 39.61 25012.58 0.81
11/20/2013 10:39:18.338918 0.000000253 0.000003098 0.000001186 41027.69 41.44 25012.58 0.81
11/20/2013 10:39:19.338838 0.000000256 0.000003098 0.000001230 42029.06 41.74 25012.58 0.81
11/20/2013 10:39:20.338621 0.000000256 0.000003098 0.000001208 43028.15 49.80 25012.58 0.81
11/20/2013 10:39:21.338512 0.000000256 0.000003098 0.000001208 44526.20 53.40 25012.58 0.81
11/20/2013 10:39:22.338428 0.000000258 0.000003098 0.000001230 45528.72 48.34 25012.58 0.81
11/20/2013 10:39:23.339288 0.000000258 0.000003098 0.000001186 46528.95 48.46 25012.58 0.81
11/20/2013 10:39:24.339252 0.000000253 0.000003098 0.000001230 47529.18 52.36 25012.58 0.81
11/20/2013 10:39:25.339125 0.000000256 0.000003098 0.000001230 48975.73 52.67 25012.58 0.81
11/20/2013 10:39:26.338885 0.000000256 0.000003120 0.000001208 50026.32 52.85 25012.58 0.81
11/20/2013 10:39:27.338831 0.000000256 0.000003098 0.000001208 51027.69 55.35 25012.58 0.81
11/20/2013 10:39:28.339727 0.000000258 0.000003120 0.000001230 52025.63 56.09 25012.58 0.81
11/20/2013 10:39:29.339894 0.000000256 0.000003120 0.000001230 53027.00 56.45 25012.58 0.81
11/20/2013 10:39:30.340396 0.000000256 0.000003120 0.000001230 54030.67 62.74 25012.58 0.81
11/20/2013 10:39:31.340369 0.000000260 0.000003120 0.000001230 54594.87 66.28 25012.58 0.81
11/20/2013 10:39:32.340350 0.000000253 0.000003120 0.000001230 55528.72 64.69 25012.58 0.81
11/20/2013 10:39:33.341106 0.000000260 0.000003120 0.000001230 56029.98 69.58 25012.58 0.81
11/20/2013 10:39:34.340924 0.000000260 0.000003098 0.000001252 56031.12 69.09 25012.58 0.81
11/20/2013 10:39:35.340837 0.000000319 0.000003758 0.000001604 56032.27 68.90 25012.58 0.81
11/20/2013 10:39:36.340732 0.000000306 0.000003714 0.000001626 56032.27 70.61 25012.58 0.81
11/20/2013 10:39:37.341651 0.000000300 0.000003582 0.000001604 56032.27 69.88 25012.58 0.81
11/20/2013 10:39:38.342419 0.000000291 0.000003538 0.000001560 56032.27 67.68 25012.58 0.81
11/20/2013 10:39:39.342271 0.000000291 0.000003494 0.000001560 56032.27 68.54 25012.58 0.81
11/20/2013 10:39:40.342168 0.000000289 0.000003494 0.000001560 56032.27 68.17 25012.58 0.81
11/20/2013 10:39:41.341996 0.000000289 0.000003472 0.000001494 56032.27 66.77 25012.58 0.81
11/20/2013 10:39:42.341916 0.000000284 0.000003450 0.000001494 56032.27 67.99 25012.58 0.81
11/20/2013 10:39:43.342793 0.000000284 0.000003450 0.000001472 56032.27 66.10 25012.58 0.81
11/20/2013 10:39:44.342661 0.000000282 0.000003384 0.000001450 56032.27 67.01 25012.58 0.81
11/20/2013 10:39:45.342577 0.000000282 0.000003384 0.000001450 56032.27 67.32 25012.58 0.81
11/20/2013 10:39:46.343125 0.000000282 0.000003362 0.000001450 56032.27 67.13 25012.58 0.81
11/20/2013 10:39:47.343318 0.000000282 0.000003384 0.000001450 56032.27 65.24 25012.58 0.81
... 10427 more lines ...
|
558
|
Wednesday, November 27, 2013, 12:50 |
Bevan Moss | South Hot-Cell | Standard Operation | TM3 | | Ground devices wiring check |
Travis and I completed an electrical check on the wiring harness for the devices on the ground plane. All lines passed 500V with the megger except for PNG and the BTM steerer. Those devices had a short to ground. of ~1.4ohm. |
559
|
Wednesday, November 27, 2013, 15:24 |
Bevan Moss | South Hot-Cell | Maintenance | TM3 | | Wiring Harness Electrical Test |
Maico removed the containment box today allow Travis and I to further test the wiring harness. I disconnected the junction block and Travis and I repeated the test from yesterday. The all lines except the PNG tested OL. The module was then lowered and the twisted pair connected to the pin that corresponds to the steerer labeled BTM (it is likely this is not labeled correctly). This testing showed showed that the source of the short found yesterday is in the wiring harness or the source tray. The PNG gauge was then tested again. When beeping out the harness to find the location of the pin it was found that the short had gone away. This circuit was then tested with the megger at 500V and showed OL. Don Jackson has suggested looking at the SHV feed through. It is difficult to determine where the source of the problem is as it is intermittent. The source tray side wiring harness for the steerers was then disconnected from the steerers and tested to OL. Finally the connections on the source tray were tested. It was determined that the third socket down (see attached picture) has a short somewhere to ground. When Chad returns on Friday I will have him complete a visual inspection and if the source can be found we may be able to deflect the line. Failing that to repair this connection a new steerer cone assembly will need to be installed. |
Attachment 1: IMG_1269.JPG
|
|
561
|
Friday, November 29, 2013, 11:57 |
Bevan Moss | South Hot-Cell | Repair | TM3 | | Steerer short investegation |
chad fisher wrote: |
Using in cell camera and transit the suspect line on the steerers was inspected as best as possible. An existing photo of the complete steerers before installation was used for comparison. The wiring and beading inspected today showed no obvious problems, breaks or spots where there may be a short. In comparison to the photo from construction, what we could see of the steerers looked identical to the original installation.
|
The test was repeated to confirm that there was a short. |
565
|
Tuesday, December 03, 2013, 14:14 |
Bevan Moss | Conditioning Station | Standard Operation | TM2 | | HV conditioning |
It was attempted to HV condition TM2 to 55kV and leave it operating there for an hour to demonstrate stability. Unfortunately at 52.65kV a spark destroyed the canbus controller. This controller had shown issues in the past but seemed to magically fixed itself. I will contact Hubert for another card. Attached is is a screen shot of the stip tools. |
Attachment 1: TM2_conditioning_03122013.JPG
|
|
575
|
Friday, December 06, 2013, 15:09 |
Bevan Moss | Conditioning Station | Standard Operation | TM2 | | TM2 no source tray 55kV 1 hour |
Today 55kV was applied to TM2 and held for over 1 hour with the EINZEL at 27.5kV. There were 2 sparks and the current was unstable at this voltage but this test should be considered a success. Once again while bringing the voltage up the canbus continued to be a problem. After successfully passing the test I continued to raise the BIAS and EINZEL. A spark at 57kV tripped a controller for the vacuum and the canbus. As stated before I believe the module will reach 60kV and that there is an interaction between the EINZEL and the BIAS. I am not certain if that interaction is internal or external to the module. Attached is a screen shot and the dumped data. |
Attachment 1: 06122013_TM2_No_source_Tray_55kV_1_hour.JPG
|
|
Attachment 2: 06122013_TM2_No_Source_Tray_55kV_1_Hour.txt
|
Time TCS:IG1:RDVAC [T] TCS:IG2:RDVAC [T] TCS:IG1S:RDVAC [T] TCS:DIC1:RDCOND [Mohm] TCS:BIAS:RDVOL [V] TCS:BIAS:RDCUR [uA] TCS:EL:RDVOL [V] TCS:EL:RDCUR [mA]
12/06/2013 10:34:30.138858 0.000000093 0.000000998 BadVal BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:31.138402 0.000000093 0.000000998 BadVal BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:32.138289 0.000000093 0.000001010 BadVal BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:33.138127 0.000000093 0.000000998 BadVal BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:34.138033 0.000000093 0.000000998 BadVal BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:35.137874 0.000000093 0.000000998 BadVal BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:36.137749 0.000000093 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:37.137657 0.000000093 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:38.137550 0.000000093 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:39.138452 0.000000093 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:40.138322 0.000000093 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:41.138150 0.000000094 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:42.139015 0.000000094 0.000000996 0.000000768 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:43.138906 0.000000093 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:44.138797 0.000000093 0.000000996 0.000000765 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:45.138654 0.000000094 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:46.144533 0.000000094 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:47.144548 0.000000094 0.000000998 0.000000770 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:48.144363 0.000000093 0.000000998 0.000000768 BadVal BadVal BadVal BadVal BadVal
12/06/2013 10:34:49.144158 0.000000093 0.000000996 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:50.144148 0.000000093 0.000000996 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:51.144894 0.000000094 0.000000998 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:52.144785 0.000000093 0.000000998 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:53.144636 0.000000094 0.000000998 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:54.144483 0.000000094 0.000000998 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:55.157531 0.000000094 0.000000998 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:56.158254 0.000000094 0.000000998 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:57.158119 0.000000094 0.000001032 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:58.157969 0.000000094 0.000001010 0.000000768 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:34:59.158877 0.000000093 0.000000998 0.000000768 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:35:00.159762 0.000000094 0.000000998 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:35:01.159615 0.000000094 0.000000998 0.000000770 16.8 BadVal BadVal BadVal BadVal
12/06/2013 10:35:03.306946 0.000000093 0.000000996 0.000000768 16.9 BadVal BadVal BadVal BadVal
12/06/2013 10:35:04.307252 0.000000094 0.000000998 0.000000768 16.9 27.46 BadVal BadVal BadVal
12/06/2013 10:35:05.308070 0.000000094 0.000000998 0.000000768 16.9 27.46 BadVal BadVal BadVal
12/06/2013 10:35:06.308956 0.000000094 0.000000994 0.000000770 16.9 27.46 BadVal BadVal BadVal
12/06/2013 10:35:07.308826 0.000000094 0.000000994 0.000000770 16.9 27.46 BadVal BadVal BadVal
12/06/2013 10:35:08.308694 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 BadVal BadVal
12/06/2013 10:35:09.308573 0.000000094 0.000000996 0.000000770 16.9 27.46 0.06 BadVal BadVal
12/06/2013 10:35:10.309442 0.000000094 0.000001010 0.000000770 16.9 27.46 0.06 BadVal BadVal
12/06/2013 10:35:11.309317 0.000000094 0.000000996 0.000000770 16.9 27.46 0.06 BadVal BadVal
12/06/2013 10:35:12.309197 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 BadVal BadVal
12/06/2013 10:35:13.309123 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 BadVal
12/06/2013 10:35:14.308964 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 BadVal
12/06/2013 10:35:15.309987 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 BadVal
12/06/2013 10:35:16.309785 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 BadVal
12/06/2013 10:35:17.309567 0.000000094 0.000000996 0.000000768 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:18.422297 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:19.422322 0.000000094 0.000000994 0.000000768 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:20.422292 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:21.422265 0.000000094 0.000001010 0.000000772 16.9 26.32 0.06 9.15 0.01
12/06/2013 10:35:22.422985 0.000000094 0.000001010 0.000000772 16.9 26.32 0.06 9.15 0.01
12/06/2013 10:35:23.532598 0.000000094 0.000000996 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:24.532696 0.000000094 0.000000996 0.000000772 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:25.700311 0.000000094 0.000000996 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:26.754388 0.000000094 0.000000996 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:27.754303 0.000000094 0.000000996 0.000000768 16.9 26.32 0.06 9.15 0.01
12/06/2013 10:35:28.754164 0.000000094 0.000000996 0.000000768 16.9 26.32 0.06 9.15 0.01
12/06/2013 10:35:29.808485 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:30.808903 0.000000094 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:31.892662 0.000000093 0.000000996 0.000000772 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:32.892649 0.000000094 0.000000994 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:33.892524 0.000000094 0.000000994 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:34.892399 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:36.59457 0.000000094 0.000000998 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:37.59452 0.000000094 0.000000998 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:38.58995 0.000000094 0.000000998 0.000000768 16.8 26.32 0.06 9.15 0.01
12/06/2013 10:35:39.58868 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:40.58702 0.000000094 0.000000998 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:41.59611 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:42.59451 0.000000093 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:43.59465 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:44.69481 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 8.54 0.01
12/06/2013 10:35:45.70113 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 8.54 0.01
12/06/2013 10:35:46.69968 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:47.69843 0.000000094 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:48.69730 0.000000093 0.000000996 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:49.69609 0.000000093 0.000000998 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:50.69501 0.000000093 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:51.70349 0.000000094 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:35:52.70327 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:53.71064 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:54.71157 0.000000094 0.000000996 0.000000768 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:55.70854 0.000000093 0.000000998 0.000000768 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:56.70728 0.000000093 0.000000996 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:57.70696 0.000000093 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:58.71465 0.000000094 0.000000998 0.000000772 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:35:59.71304 0.000000094 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:36:00.72214 0.000000093 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:36:01.72058 0.000000093 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:02.72037 0.000000094 0.000000996 0.000000765 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:03.71845 0.000000093 0.000000996 0.000000768 16.8 26.32 0.06 9.15 0.01
12/06/2013 10:36:04.71718 0.000000093 0.000000996 0.000000770 16.8 26.32 0.06 9.15 0.01
12/06/2013 10:36:05.71547 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:06.72545 0.000000094 0.000000996 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:07.72467 0.000000093 0.000000996 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:08.73208 0.000000094 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:09.73087 0.000000093 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:10.73946 0.000000093 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:11.74864 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:12.74666 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:13.75645 0.000000093 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:14.75641 0.000000093 0.000000998 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:15.75369 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:16.136319 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:17.136267 0.000000093 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:18.230037 0.000000094 0.000000998 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:19.229817 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:20.229681 0.000000093 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:21.229580 0.000000093 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:22.229431 0.000000093 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:23.229311 0.000000093 0.000000998 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:24.229201 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:25.229010 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:26.228930 0.000000094 0.000000994 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:27.273591 0.000000094 0.000000998 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:28.273699 0.000000093 0.000000998 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:29.274782 0.000000094 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:30.274160 0.000000094 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:31.274279 0.000000094 0.000000998 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:32.274186 0.000000093 0.000000998 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:33.276910 0.000000093 0.000000998 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:34.276923 0.000000094 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:35.277751 0.000000094 0.000000998 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:36.277627 0.000000094 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:37.277496 0.000000094 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:38.277402 0.000000093 0.000000998 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:39.277265 0.000000093 0.000000998 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:40.277272 0.000000093 0.000001010 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:41.278116 0.000000093 0.000000998 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:42.277876 0.000000093 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:43.277804 0.000000094 0.000000998 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:44.277667 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:45.277528 0.000000094 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:46.279324 0.000000094 0.000000998 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:47.279479 0.000000093 0.000000998 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:48.280171 0.000000094 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:49.280032 0.000000094 0.000000998 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:50.279918 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:51.280727 0.000000093 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:52.280665 0.000000093 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:53.280477 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:54.280361 0.000000093 0.000000996 0.000000770 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:55.280370 0.000000094 0.000000994 0.000000770 16.7 27.46 0.06 9.76 0.01
12/06/2013 10:36:56.281267 0.000000094 0.000000996 0.000000768 16.7 27.46 0.06 9.15 0.01
12/06/2013 10:36:57.281021 0.000000093 0.000000996 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:58.280911 0.000000094 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:36:59.280785 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:00.280647 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:01.281537 0.000000094 0.000000996 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:02.281408 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:03.281254 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:04.281106 0.000000094 0.000000998 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:05.281008 0.000000093 0.000000998 0.000000768 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:06.281889 0.000000094 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:07.281764 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:08.281734 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:09.281500 0.000000093 0.000000998 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:10.281385 0.000000094 0.000000994 0.000000770 16.8 27.46 0.06 9.15 0.01
12/06/2013 10:37:11.282277 0.000000094 0.000000996 0.000000770 16.9 27.46 0.06 9.76 0.01
12/06/2013 10:37:12.282243 0.000000093 0.000000998 0.000000768 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:37:13.282030 0.000000094 0.000000996 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:37:14.281905 0.000000093 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:37:15.281781 0.000000093 0.000000998 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:37:16.282607 0.000000093 0.000000996 0.000000770 16.9 27.46 0.06 9.15 0.01
12/06/2013 10:37:17.282622 0.000000093 0.000000998 0.000000768 16.9 6.86 1.03 9.15 0.01
12/06/2013 10:37:18.282434 0.000000094 0.000000998 0.000000770 16.9 6.86 1.03 1.22 0.01
12/06/2013 10:37:19.282257 0.000000094 0.000000998 0.000000770 16.9 3.43 1.03 1.22 0.01
12/06/2013 10:37:20.282499 0.000000094 0.000000998 0.000000770 16.9 4.57 1.03 1.83 0.01
12/06/2013 10:37:21.283005 0.000000094 0.000000998 0.000000770 16.9 4.57 1.03 493.17 0.01
12/06/2013 10:37:22.283870 0.000000093 0.000000998 0.000000770 16.9 4.57 1.03 993.05 0.01
12/06/2013 10:37:23.283923 0.000000094 0.000000998 0.000000770 16.9 6.86 0.91 1494.16 0.01
12/06/2013 10:37:24.283648 0.000000093 0.000000996 0.000000770 16.9 3.43 1.15 1494.16 0.01
12/06/2013 10:37:25.283492 0.000000094 0.000000998 0.000000770 16.9 13.73 1.15 1994.04 0.28
12/06/2013 10:37:26.283509 0.000000094 0.000000998 0.000000770 16.9 13.73 1.15 1994.04 0.08
12/06/2013 10:37:27.283262 0.000000093 0.000000998 0.000000770 16.8 5.72 0.85 2494.54 0.10
12/06/2013 10:37:28.283071 0.000000093 0.000000998 0.000000770 16.8 2.28 0.97 2995.04 0.10
12/06/2013 10:37:29.282990 0.000000094 0.000000998 0.000000770 16.8 2.28 0.97 3495.53 0.10
12/06/2013 10:37:30.282877 0.000000093 0.000000998 0.000000770 16.8 10.29 0.97 3994.81 0.15
12/06/2013 10:37:31.283728 0.000000094 0.000000996 0.000000768 16.8 8.01 0.91 3994.81 0.15
12/06/2013 10:37:32.283568 0.000000094 0.000000998 0.000000768 16.8 4.57 1.15 3996.03 0.15
12/06/2013 10:37:33.283508 0.000000093 0.000000998 0.000000770 16.8 6.86 1.15 3996.03 0.15
12/06/2013 10:37:34.283108 0.000000094 0.000000998 0.000000768 16.8 4.57 1.03 3996.03 0.15
12/06/2013 10:37:35.283283 0.000000094 0.000000998 0.000000768 16.8 5.72 1.03 3996.03 0.15
12/06/2013 10:37:36.284116 0.000000093 0.000000994 0.000000768 16.8 5.72 1.09 3996.03 0.15
12/06/2013 10:37:37.283983 0.000000094 0.000000998 0.000000770 16.8 5.72 1.09 3996.64 0.15
12/06/2013 10:37:38.283864 0.000000094 0.000000998 0.000000770 16.7 3.43 1.09 3996.03 0.15
12/06/2013 10:37:39.472917 0.000000094 0.000000998 0.000000765 16.7 3.43 1.09 3996.64 0.15
12/06/2013 10:37:40.472581 0.000000093 0.000000996 0.000000768 16.7 2.28 1.15 3996.64 0.15
12/06/2013 10:37:41.472472 0.000000094 0.000000998 0.000000770 16.7 2.28 1.15 3996.64 0.15
12/06/2013 10:37:42.472333 0.000000094 0.000000998 0.000000768 16.8 8.01 1.03 3996.64 0.15
12/06/2013 10:37:43.472212 0.000000094 0.000000998 0.000000770 16.8 8.01 1.03 3996.64 0.15
12/06/2013 10:37:44.738171 0.000000094 0.000000998 0.000000770 16.8 9.15 1.03 3996.64 0.15
12/06/2013 10:37:45.737929 0.000000093 0.000000998 0.000000770 16.8 9.15 1.03 3996.03 0.15
12/06/2013 10:37:46.737758 0.000000093 0.000000996 0.000000768 16.8 10.29 1.03 3996.64 0.15
12/06/2013 10:37:47.737679 0.000000094 0.000000996 0.000000770 16.8 8.01 0.91 3996.64 0.15
12/06/2013 10:37:48.737550 0.000000094 0.000000998 0.000000768 16.8 5.72 1.15 3996.64 0.15
12/06/2013 10:37:49.737448 0.000000094 0.000000998 0.000000768 16.8 5.72 1.15 4276.79 0.15
12/06/2013 10:37:50.737254 0.000000094 0.000000998 0.000000770 16.8 3.43 0.97 4997.02 0.15
12/06/2013 10:37:51.738371 0.000000093 0.000000996 0.000000770 16.8 12.58 0.91 5997.40 0.44
12/06/2013 10:37:52.739018 0.000000094 0.000000996 0.000000770 16.8 12.58 1.15 7996.33 0.49
12/06/2013 10:37:53.739981 0.000000094 0.000000998 0.000000770 16.9 1.14 1.15 9499.65 0.29
12/06/2013 10:37:54.739743 0.000000094 0.000000998 0.000000770 16.9 1.14 1.15 10997.48 0.59
12/06/2013 10:37:55.739858 0.000000094 0.000000996 0.000000770 16.9 14.87 1.15 11999.69 0.42
12/06/2013 10:37:56.740534 0.000000094 0.000001010 0.000000770 16.9 14.87 1.15 13499.35 0.47
12/06/2013 10:37:57.740392 0.000000094 0.000000998 0.000000765 16.9 3.43 1.15 15000.83 0.49
12/06/2013 10:37:58.740247 0.000000094 0.000000998 0.000000768 16.8 3.43 1.15 16499.88 0.57
12/06/2013 10:37:59.740213 0.000000094 0.000000998 0.000000768 16.8 12.58 1.15 18001.98 0.63
12/06/2013 10:38:00.740039 0.000000093 0.000000998 0.000000768 16.8 12.58 1.15 19068.89 0.65
12/06/2013 10:38:01.739890 0.000000093 0.000000998 0.000000774 16.8 10.29 0.97 20003.35 0.65
12/06/2013 10:38:02.739822 0.000000093 0.000000998 0.000000770 16.8 10.29 0.97 20003.35 0.65
12/06/2013 10:38:03.740608 0.000000093 0.000000998 0.000000770 16.8 1007.09 4.08 20005.18 0.65
12/06/2013 10:38:04.740535 0.000000094 0.000000998 0.000000768 16.8 2009.61 2.38 20005.18 0.65
12/06/2013 10:38:05.740421 0.000000094 0.000000998 0.000000768 16.8 3292.51 4.33 20005.18 0.65
12/06/2013 10:38:06.741299 0.000000093 0.000000998 0.000000770 16.8 4510.18 7.87 20005.18 0.65
12/06/2013 10:38:07.741154 0.000000094 0.000000998 0.000000770 16.8 5794.23 7.01 20005.79 0.65
12/06/2013 10:38:08.740978 0.000000094 0.000000994 0.000000770 16.8 7014.19 10.80 20005.79 0.65
12/06/2013 10:38:09.740907 0.000000093 0.000000998 0.000000768 16.8 8294.80 13.91 20005.79 0.65
12/06/2013 10:38:10.741753 0.000000093 0.000000998 0.000000768 16.8 10017.16 14.58 20005.79 0.65
12/06/2013 10:38:11.742650 0.000000094 0.000000996 0.000000770 16.8 11519.79 19.28 20005.79 0.65
12/06/2013 10:38:12.742528 0.000000093 0.000000996 0.000000770 16.8 12021.05 12.81 20005.79 0.65
12/06/2013 10:38:13.742395 0.000000093 0.000000998 0.000000770 16.8 12522.31 12.08 20005.79 0.65
12/06/2013 10:38:14.742226 0.000000093 0.000000998 0.000000770 16.8 13021.28 12.26 20005.79 0.65
12/06/2013 10:38:15.742200 0.000000094 0.000000998 0.000000768 16.8 13801.78 12.57 20005.79 0.65
12/06/2013 10:38:16.742099 0.000000094 0.000000998 0.000000770 16.8 14522.77 17.39 20005.79 0.65
12/06/2013 10:38:17.742891 0.000000093 0.000000996 0.000000768 16.8 15024.03 15.80 20005.79 0.65
12/06/2013 10:38:18.742772 0.000000093 0.000000998 0.000000768 16.8 15524.14 14.89 20005.79 0.65
12/06/2013 10:38:19.742602 0.000000093 0.000000998 0.000000768 16.8 16524.37 18.73 20005.79 0.65
12/06/2013 10:38:20.742477 0.000000093 0.000000998 0.000000768 16.8 17026.77 17.27 20005.79 0.65
12/06/2013 10:38:21.743352 0.000000093 0.000000996 0.000000770 16.8 18030.44 20.81 20005.79 0.65
12/06/2013 10:38:22.743286 0.000000093 0.000000996 0.000000770 16.8 18527.12 18.67 20005.79 0.65
12/06/2013 10:38:23.743116 0.000000093 0.000000998 0.000000770 16.8 19528.49 22.64 20005.79 0.65
12/06/2013 10:38:24.743988 0.000000093 0.000000996 0.000000765 16.8 20029.75 21.17 20005.79 0.65
12/06/2013 10:38:25.743892 0.000000093 0.000000996 0.000000768 16.8 20529.86 20.38 20005.79 0.65
12/06/2013 10:38:26.744754 0.000000093 0.000000998 0.000000770 16.8 21530.09 25.63 20005.79 0.65
12/06/2013 10:38:27.744640 0.000000094 0.000000998 0.000000768 16.8 22031.35 23.49 20005.79 0.65
12/06/2013 10:38:28.744504 0.000000093 0.000000998 0.000000768 16.8 22532.61 22.94 20005.79 0.65
12/06/2013 10:38:29.745410 0.000000094 0.000000998 0.000000768 16.8 23032.73 22.70 20005.79 0.65
12/06/2013 10:38:30.745262 0.000000094 0.000000998 0.000000770 16.8 23533.98 27.52 20005.79 0.65
12/06/2013 10:38:31.745131 0.000000094 0.000000996 0.000000772 16.8 24533.07 26.36 20005.79 0.65
12/06/2013 10:38:32.745008 0.000000094 0.000000998 0.000000768 16.7 25034.33 25.63 20005.79 0.65
12/06/2013 10:38:33.745050 0.000000093 0.000000998 0.000000770 16.7 26034.56 29.90 20005.79 0.65
12/06/2013 10:38:34.744755 0.000000094 0.000001010 0.000000768 16.7 26535.82 28.25 20005.79 0.65
12/06/2013 10:38:35.744622 0.000000093 0.000000998 0.000000770 16.7 27035.93 27.34 20005.79 0.65
12/06/2013 10:38:36.745477 0.000000093 0.000000998 0.000000768 16.7 27989.24 31.73 20005.79 0.65
12/06/2013 10:38:37.746375 0.000000093 0.000000998 0.000000768 16.7 28536.27 30.94 20005.79 0.65
12/06/2013 10:38:38.746259 0.000000094 0.000000998 0.000000770 16.7 29037.53 30.88 20005.79 0.65
12/06/2013 10:38:39.746094 0.000000094 0.000000996 0.000000770 16.7 29538.79 30.51 20005.79 0.65
12/06/2013 10:38:40.746006 0.000000093 0.000000998 0.000000768 16.7 30038.91 30.45 20005.79 0.65
12/06/2013 10:38:41.745828 0.000000093 0.000000998 0.000000770 16.7 30539.02 30.70 20005.79 0.65
12/06/2013 10:38:42.745708 0.000000093 0.000000994 0.000000768 16.7 31040.28 31.55 20005.79 0.65
12/06/2013 10:38:43.746645 0.000000094 0.000000994 0.000000770 16.7 31993.59 36.07 20005.79 0.65
12/06/2013 10:38:44.746538 0.000000093 0.000000998 0.000000765 16.7 32540.62 35.82 20005.79 0.65
12/06/2013 10:38:45.746327 0.000000093 0.000000998 0.000000765 16.7 33040.74 35.15 20005.79 0.65
12/06/2013 10:38:46.747194 0.000000093 0.000001010 0.000000770 16.7 33540.85 34.79 20005.79 0.65
12/06/2013 10:38:47.747076 0.000000093 0.000000996 0.000000770 16.7 34545.66 38.81 20005.79 0.65
12/06/2013 10:38:48.746949 0.000000093 0.000000998 0.000000768 16.7 35042.34 37.17 20005.79 0.65
12/06/2013 10:38:49.747854 0.000000094 0.000000998 0.000000770 16.8 36041.42 46.81 20005.79 0.65
12/06/2013 10:38:50.747794 0.000000093 0.000000996 0.000000768 16.8 36543.83 40.03 20005.79 0.65
12/06/2013 10:38:51.748607 0.000000094 0.000000998 0.000000770 16.8 37043.94 39.61 20005.79 0.65
12/06/2013 10:38:52.748492 0.000000093 0.000000998 0.000000770 16.8 37545.20 39.73 20005.79 0.65
12/06/2013 10:38:53.748395 0.000000093 0.000000998 0.000000768 16.8 38046.46 44.06 20005.79 0.65
12/06/2013 10:38:54.748195 0.000000093 0.000000996 0.000000768 16.8 39044.40 43.03 20005.79 0.65
12/06/2013 10:38:55.748153 0.000000094 0.000000998 0.000000770 16.8 39545.66 43.45 20005.79 0.65
12/06/2013 10:38:56.749112 0.000000094 0.000000998 0.000000770 16.8 39545.66 40.89 20005.79 0.65
12/06/2013 10:38:57.748818 0.000000094 0.000000998 0.000000770 16.8 40046.92 41.50 20005.79 0.65
12/06/2013 10:38:58.748731 0.000000093 0.000000998 0.000000768 16.8 40046.92 41.07 20005.79 0.65
12/06/2013 10:38:59.748612 0.000000094 0.000000996 0.000000768 16.8 40046.92 40.89 20005.79 0.65
12/06/2013 10:39:00.748440 0.000000094 0.000000996 0.000000768 16.8 40046.92 40.89 20005.79 0.65
12/06/2013 10:39:01.749351 0.000000094 0.000001010 0.000000770 16.8 40046.92 40.89 20005.79 0.65
12/06/2013 10:39:02.749233 0.000000094 0.000000996 0.000000770 16.8 40048.06 40.83 20006.40 0.65
12/06/2013 10:39:03.749082 0.000000094 0.000000998 0.000000770 16.8 40048.06 40.83 20006.40 0.65
12/06/2013 10:39:04.749026 0.000000093 0.000000998 0.000000765 16.8 40046.92 40.83 20005.79 0.65
12/06/2013 10:39:05.749899 0.000000094 0.000000998 0.000000768 16.8 40046.92 40.83 20005.79 0.65
12/06/2013 10:39:06.750723 0.000000093 0.000000998 0.000000770 16.8 40046.92 40.83 20005.79 0.65
12/06/2013 10:39:07.750716 0.000000093 0.000000998 0.000000770 16.8 40046.92 40.83 20005.79 0.65
12/06/2013 10:39:08.750436 0.000000093 0.000000998 0.000000770 16.8 40048.06 40.83 20005.79 0.65
12/06/2013 10:39:09.750420 0.000000093 0.000000998 0.000000770 16.8 40046.92 41.32 20005.79 0.65
12/06/2013 10:39:10.750265 0.000000093 0.000000998 0.000000770 16.8 40048.06 40.71 20005.79 0.65
12/06/2013 10:39:11.751047 0.000000093 0.000000996 0.000000768 16.8 40048.06 40.71 20005.79 0.65
12/06/2013 10:39:12.750973 0.000000094 0.000000996 0.000000770 16.8 40048.06 40.71 20005.79 0.65
12/06/2013 10:39:13.751821 0.000000094 0.000000998 0.000000770 16.8 40048.06 40.71 20005.79 0.65
12/06/2013 10:39:14.751678 0.000000094 0.000000996 0.000000770 16.8 40048.06 40.71 20005.79 0.65
12/06/2013 10:39:15.751659 0.000000094 0.000000996 0.000000770 16.8 40048.06 40.71 20005.79 0.65
12/06/2013 10:39:16.752539 0.000000094 0.000000998 0.000000772 16.8 40048.06 40.71 20503.85 0.71
12/06/2013 10:39:17.753413 0.000000094 0.000000996 0.000000770 16.8 40048.06 40.71 21004.34 0.71
12/06/2013 10:39:18.753168 0.000000094 0.000000998 0.000000770 16.8 40048.06 40.71 21504.23 0.71
12/06/2013 10:39:19.753173 0.000000093 0.000000998 0.000000772 16.7 40048.06 40.58 22070.64 0.76
12/06/2013 10:39:20.752945 0.000000093 0.000000994 0.000000768 16.7 40048.06 40.58 23000.83 0.99
12/06/2013 10:39:21.753831 0.000000094 0.000000996 0.000000772 16.7 40048.06 40.58 23503.16 0.80
12/06/2013 10:39:22.753731 0.000000093 0.000000996 0.000000772 16.7 40048.06 40.58 24003.66 0.80
12/06/2013 10:39:23.753588 0.000000093 0.000000996 0.000000770 16.7 40048.06 40.58 24501.71 0.85
12/06/2013 10:39:24.754474 0.000000094 0.000000998 0.000000765 16.7 40048.06 40.58 25003.43 0.87
12/06/2013 10:39:25.754339 0.000000094 0.000000998 0.000000770 16.7 40048.06 40.71 25503.31 0.87
12/06/2013 10:39:26.754168 0.000000094 0.000000998 0.000000772 16.7 40048.06 40.71 25953.15 0.87
12/06/2013 10:39:27.754087 0.000000093 0.000000998 0.000000772 16.8 40048.06 40.71 26003.20 0.87
12/06/2013 10:39:28.754026 0.000000094 0.000000998 0.000000770 16.8 40048.06 40.71 26503.70 0.99
12/06/2013 10:39:29.753829 0.000000093 0.000000998 0.000000770 16.8 40048.06 40.71 26572.67 0.99
12/06/2013 10:39:30.753655 0.000000093 0.000000998 0.000000770 16.8 40048.06 40.71 27503.47 0.95
... 14427 more lines ...
|
578
|
Thursday, December 12, 2013, 11:14 |
Bevan Moss | Conditioning Station | Standard Operation | TM2 | | HV Chase Electrical Test |
While under vacuum Travis and I completed diagnostics of the HV chase. Majority of the lines showed OL at 1kV but there were the following "failures"
A to P @ 250V 716kOhm
A to Q @ 1kV 25.18MOhm
B to C with ohm meter 8Mohm @250V 830kOhm
B to D @ 1kV 15.57MOhm
C to D @1kV 16.18 MOhm
P to Q @1kV 1.069 GOhm
G to J with ohm meter dead short.
Ideally all of these values would be OL and they may improve or get worse when the source tray is installed depending on where the electrical contact is. I need to discuss with Friedhelm but I believe that as these values the module can still operate. Without changing the HV chase there is very little chance of improving these values. |
635
|
Monday, March 31, 2014, 13:22 |
Bevan Moss | Conditioning Station | Standard Operation | TM4 | Ta #42 | HV conditioning |
TM4 with Ta#42 was HV conditioned to 35kV with the optics tray at 17.5kV. There was some vacuum response while ramping but well within normal. Attached is a screen shot and the associated data. |
Attachment 1: 31032014_TM4_Ta#42_HV_Conditioning.txt
|
Time TCS:BIAS:RDVOL [V] TCS:BIAS:RDCUR [uA] TCS:EL:RDVOL [V] TCS:EL:RDCUR [mA] TCS:IG1:RDVAC [T] TCS:IG1S:RDVAC [T] TCS:IG2:RDVAC [T] TCS:DIC1:RDCOND [Mohm]
03/31/2014 10:01:03.733174 11.44 BadVal BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:04.732844 11.44 BadVal BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:05.732769 12.58 BadVal BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:06.732712 12.58 BadVal BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:07.732465 12.58 BadVal BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:08.733393 12.58 BadVal BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:09.733515 12.58 BadVal BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:10.734113 12.58 0.18 BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:11.733990 12.58 0.18 BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:12.733856 12.58 0.18 BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:13.734748 12.58 0.18 BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:14.734630 12.58 0.18 BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:15.734460 12.58 0.18 BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:16.734336 12.58 0.18 BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:17.734235 12.58 0.18 BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:18.735107 12.58 0.18 BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:19.734959 12.58 0.18 BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:20.734835 12.58 0.18 BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:21.734750 12.58 0.18 BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:22.844126 12.58 0.18 BadVal BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:23.844486 12.58 0.18 6.71 BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:24.844315 12.58 0.18 6.71 BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:25.846185 12.58 0.18 6.71 BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:26.846055 12.58 0.18 6.71 BadVal BadVal BadVal BadVal BadVal
03/31/2014 10:01:27.845938 12.58 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:28.846887 11.44 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:29.846712 11.44 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:30.846596 12.58 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:31.846538 12.58 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:32.846330 10.29 0.30 7.32 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:33.847382 12.58 0.18 7.32 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:34.847056 12.58 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:35.846932 12.58 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:36.846800 12.58 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:37.846673 12.58 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:38.847547 12.58 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:40.104731 12.58 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:41.104567 12.58 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:42.104172 12.58 0.18 7.32 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:43.104164 12.58 0.18 7.32 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:44.104890 12.58 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:45.104770 12.58 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:46.104674 11.44 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:47.105620 11.44 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:48.106416 12.58 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:49.107534 12.58 0.18 6.71 0.01 BadVal BadVal BadVal BadVal
03/31/2014 10:01:50.107142 12.58 0.18 6.71 0.01 0.000000539 BadVal BadVal BadVal
03/31/2014 10:01:51.107016 12.58 0.18 6.71 0.01 0.000000539 BadVal BadVal BadVal
03/31/2014 10:01:52.106907 11.44 0.18 6.71 0.01 0.000000541 BadVal BadVal BadVal
03/31/2014 10:01:53.106824 11.44 0.18 6.71 0.01 0.000000537 BadVal BadVal BadVal
03/31/2014 10:01:54.106672 12.58 0.18 6.71 0.01 0.000000535 BadVal BadVal BadVal
03/31/2014 10:01:55.106512 12.58 0.18 6.71 0.01 0.000000535 BadVal BadVal BadVal
03/31/2014 10:01:56.106402 12.58 0.18 6.71 0.01 0.000000535 BadVal BadVal BadVal
03/31/2014 10:01:57.106261 12.58 0.18 6.71 0.01 0.000000537 BadVal BadVal BadVal
03/31/2014 10:01:58.106140 12.58 0.18 6.71 0.01 0.000000539 BadVal BadVal BadVal
03/31/2014 10:01:59.106045 12.58 0.18 6.71 0.01 0.000000537 BadVal BadVal BadVal
03/31/2014 10:02:00.105877 12.58 0.18 6.71 0.01 0.000000535 BadVal BadVal BadVal
03/31/2014 10:02:01.106805 12.58 0.18 6.71 0.01 0.000000537 BadVal BadVal BadVal
03/31/2014 10:02:02.106632 12.58 0.18 6.71 0.01 0.000000537 BadVal 0.000002571 BadVal
03/31/2014 10:02:03.106782 12.58 0.18 6.71 0.01 0.000000539 BadVal 0.000002571 BadVal
03/31/2014 10:02:04.107410 12.58 0.18 6.71 0.01 0.000000537 BadVal 0.000002549 BadVal
03/31/2014 10:02:05.107394 12.58 0.18 6.71 0.01 0.000000539 BadVal 0.000002549 BadVal
03/31/2014 10:02:06.108182 12.58 0.18 6.71 0.01 0.000000537 BadVal 0.000002571 BadVal
03/31/2014 10:02:07.108063 12.58 0.18 6.71 0.01 0.000000537 BadVal 0.000002593 BadVal
03/31/2014 10:02:08.107895 12.58 0.18 6.71 0.01 0.000000537 BadVal 0.000002593 BadVal
03/31/2014 10:02:09.145507 12.58 0.18 6.71 0.01 0.000000537 BadVal 0.000002571 BadVal
03/31/2014 10:02:10.195928 12.58 0.18 6.71 0.01 0.000000539 BadVal 0.000002571 BadVal
03/31/2014 10:02:11.195508 12.58 0.18 6.71 0.01 0.000000537 BadVal 0.000002571 BadVal
03/31/2014 10:02:12.195358 12.58 0.18 6.71 0.01 0.000000535 BadVal 0.000002593 BadVal
03/31/2014 10:02:13.195230 12.58 0.18 6.71 0.01 0.000000535 BadVal 0.000002549 BadVal
03/31/2014 10:02:14.196116 12.58 0.18 6.71 0.01 0.000000537 BadVal 0.000002593 BadVal
03/31/2014 10:02:15.195989 12.58 0.18 6.71 0.01 0.000000539 BadVal 0.000002593 BadVal
03/31/2014 10:02:16.195944 12.58 0.18 6.71 0.01 0.000000541 BadVal 0.000002571 BadVal
03/31/2014 10:02:17.195738 12.58 0.18 6.71 0.01 0.000000541 BadVal 0.000002593 BadVal
03/31/2014 10:02:18.195693 12.58 0.18 6.71 0.01 0.000000537 BadVal 0.000002571 BadVal
03/31/2014 10:02:19.196508 12.58 0.18 6.71 0.01 0.000000535 BadVal 0.000002571 BadVal
03/31/2014 10:02:20.196434 12.58 0.18 7.32 0.01 0.000000535 BadVal 0.000002571 BadVal
03/31/2014 10:02:21.196308 12.58 0.18 7.32 0.01 0.000000535 BadVal 0.000002549 BadVal
03/31/2014 10:02:22.196100 12.58 0.18 6.71 0.01 0.000000535 BadVal 0.000002593 BadVal
03/31/2014 10:02:23.196019 12.58 0.18 6.71 0.01 0.000000537 BadVal 0.000002593 BadVal
03/31/2014 10:02:24.283661 12.58 0.18 7.32 0.01 0.000000537 BadVal 0.000002571 BadVal
03/31/2014 10:02:25.283865 12.58 0.18 7.32 0.01 0.000000535 0.000001340 0.000002571 BadVal
03/31/2014 10:02:26.283648 12.58 0.18 6.71 0.01 0.000000535 0.000001362 0.000002571 BadVal
03/31/2014 10:02:27.284460 12.58 0.18 6.71 0.01 0.000000535 0.000001362 0.000002571 BadVal
03/31/2014 10:02:28.284342 12.58 0.18 6.71 0.01 0.000000535 0.000001340 0.000002593 BadVal
03/31/2014 10:02:29.285208 12.58 0.18 6.71 0.01 0.000000537 0.000001384 0.000002571 BadVal
03/31/2014 10:02:30.285140 12.58 0.18 6.71 0.01 0.000000539 0.000001362 0.000002571 BadVal
03/31/2014 10:02:31.285014 12.58 0.18 6.71 0.01 0.000000535 0.000001340 0.000002571 BadVal
03/31/2014 10:02:32.285861 12.58 0.18 6.71 0.01 0.000000537 0.000001318 0.000002571 BadVal
03/31/2014 10:02:33.285706 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 BadVal
03/31/2014 10:02:34.284790 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 BadVal
03/31/2014 10:02:35.285462 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002571 BadVal
03/31/2014 10:02:36.285355 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 BadVal
03/31/2014 10:02:37.286202 12.58 0.18 7.32 0.01 0.000000537 0.000001318 0.000002549 BadVal
03/31/2014 10:02:38.286087 12.58 0.18 7.32 0.01 0.000000535 0.000001318 0.000002549 BadVal
03/31/2014 10:02:39.287229 12.58 0.18 7.32 0.01 0.000000537 0.000001318 0.000002571 BadVal
03/31/2014 10:02:40.286833 12.58 0.18 7.32 0.01 0.000000537 0.000001340 0.000002571 BadVal
03/31/2014 10:02:41.286722 12.58 0.18 6.71 0.01 0.000000537 0.000001318 0.000002549 BadVal
03/31/2014 10:02:42.286663 12.58 0.18 6.71 0.01 0.000000539 0.000001362 0.000002571 BadVal
03/31/2014 10:02:43.286636 12.58 0.18 6.71 0.01 0.000000539 0.000001362 0.000002571 BadVal
03/31/2014 10:02:44.286345 12.58 0.18 6.71 0.01 0.000000535 0.000001340 0.000002571 BadVal
03/31/2014 10:02:45.287322 12.58 0.18 7.32 0.01 0.000000537 0.000001340 0.000002571 BadVal
03/31/2014 10:02:46.287072 12.58 0.18 7.32 0.01 0.000000535 0.000001340 0.000002571 BadVal
03/31/2014 10:02:47.286970 12.58 0.18 6.71 0.01 0.000000535 0.000001318 0.000002571 BadVal
03/31/2014 10:02:48.286830 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002593 BadVal
03/31/2014 10:02:49.286701 12.58 0.18 6.71 0.01 0.000000539 0.000001318 0.000002571 BadVal
03/31/2014 10:02:52.177616 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 BadVal
03/31/2014 10:02:53.178236 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 BadVal
03/31/2014 10:02:54.179135 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 BadVal
03/31/2014 10:02:55.178964 12.58 0.18 6.71 0.01 0.000000539 0.000001362 0.000002571 BadVal
03/31/2014 10:02:56.182883 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002593 BadVal
03/31/2014 10:02:57.183765 12.58 0.18 6.71 0.01 0.000000532 0.000001362 0.000002571 BadVal
03/31/2014 10:02:58.183585 12.58 0.18 6.71 0.01 0.000000535 0.000001362 0.000002571 10.5
03/31/2014 10:02:59.183583 12.58 0.18 6.71 0.01 0.000000539 0.000001362 0.000002571 10.5
03/31/2014 10:03:00.183372 11.44 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:03:01.183341 11.44 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:03:02.183147 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:03.182950 12.58 0.18 6.71 0.01 0.000000535 0.000001318 0.000002571 10.5
03/31/2014 10:03:04.183824 12.58 0.18 6.71 0.01 0.000000537 0.000001318 0.000002571 10.5
03/31/2014 10:03:05.183703 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:06.183578 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002593 10.5
03/31/2014 10:03:07.184451 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:08.184426 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:09.185200 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:10.185094 12.58 0.18 6.71 0.01 0.000000535 0.000001362 0.000002571 10.5
03/31/2014 10:03:11.184972 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:12.184928 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:13.214901 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:03:14.215641 11.44 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:15.215439 11.44 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:16.215466 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:17.216192 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:18.216066 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:19.216939 12.58 0.18 6.71 0.01 0.000000535 0.000001362 0.000002571 10.5
03/31/2014 10:03:20.216993 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:03:21.216687 12.58 0.18 6.71 0.01 0.000000532 0.000001340 0.000002593 10.5
03/31/2014 10:03:22.216602 11.44 0.18 6.71 0.01 0.000000537 0.000001318 0.000002571 10.5
03/31/2014 10:03:23.216432 11.44 0.18 6.71 0.01 0.000000537 0.000001318 0.000002571 10.5
03/31/2014 10:03:24.217344 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:25.218233 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:26.218061 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002549 10.5
03/31/2014 10:03:27.217964 12.58 0.18 6.71 0.01 0.000000535 0.000001362 0.000002571 10.5
03/31/2014 10:03:28.217956 11.44 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:29.217839 11.44 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:03:30.217602 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:03:31.217432 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:03:32.217311 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:03:33.217306 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:03:34.218096 11.44 0.18 6.71 0.01 0.000000535 0.000001362 0.000002571 10.5
03/31/2014 10:03:35.217929 11.44 0.18 6.71 0.01 0.000000535 0.000001340 0.000002571 10.5
03/31/2014 10:03:36.217843 12.58 0.18 6.71 0.01 0.000000535 0.000001318 0.000002571 10.5
03/31/2014 10:03:37.279743 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:38.374264 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:39.455103 12.58 0.18 6.71 0.01 0.000000535 0.000001318 0.000002549 10.5
03/31/2014 10:03:40.455295 12.58 0.18 6.71 0.01 0.000000535 0.000001318 0.000002549 10.5
03/31/2014 10:03:41.455156 12.58 0.18 6.71 0.01 0.000000535 0.000001340 0.000002571 10.5
03/31/2014 10:03:42.455022 12.58 0.18 6.71 0.01 0.000000535 0.000001340 0.000002549 10.5
03/31/2014 10:03:43.517929 12.58 0.18 6.71 0.01 0.000000539 0.000001318 0.000002549 10.5
03/31/2014 10:03:44.605664 12.58 0.18 6.71 0.01 0.000000539 0.000001318 0.000002549 10.5
03/31/2014 10:03:45.605769 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:03:46.605499 12.58 0.18 6.71 0.01 0.000000535 0.000001340 0.000002549 10.5
03/31/2014 10:03:47.605372 12.58 0.18 6.71 0.01 0.000000535 0.000001340 0.000002571 10.5
03/31/2014 10:03:48.605280 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:03:49.606125 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:03:50.606000 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002593 10.5
03/31/2014 10:03:51.606922 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:03:52.606757 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:53.606653 12.58 0.18 6.71 0.01 0.000000535 0.000001340 0.000002571 10.5
03/31/2014 10:03:54.606492 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:03:55.606422 12.58 0.18 6.71 0.01 0.000000552 0.000001340 0.000002571 10.5
03/31/2014 10:03:56.606253 12.58 0.18 6.71 0.01 0.000000546 0.000001340 0.000002571 10.5
03/31/2014 10:03:57.606119 12.58 0.18 6.71 0.01 0.000000546 0.000001362 0.000002615 10.5
03/31/2014 10:03:58.606091 12.58 0.18 6.71 0.01 0.000000546 0.000001340 0.000002615 10.5
03/31/2014 10:03:59.605889 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002593 10.5
03/31/2014 10:04:00.605762 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002571 10.5
03/31/2014 10:04:01.605614 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002571 10.5
03/31/2014 10:04:02.605686 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002593 10.5
03/31/2014 10:04:03.605398 12.58 0.18 6.71 0.01 0.000000539 0.000001318 0.000002593 10.5
03/31/2014 10:04:04.605227 12.58 0.18 6.71 0.01 0.000000539 0.000001318 0.000002593 10.5
03/31/2014 10:04:05.605108 12.58 0.18 6.71 0.01 0.000000537 0.000001318 0.000002593 10.5
03/31/2014 10:04:06.604986 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:04:07.604864 12.58 0.18 6.71 0.01 0.000000539 0.000001362 0.000002593 10.5
03/31/2014 10:04:08.604798 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002593 10.5
03/31/2014 10:04:09.604612 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002593 10.5
03/31/2014 10:04:10.604484 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:04:11.604377 12.58 0.18 7.32 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:04:12.604230 12.58 0.18 7.32 0.01 0.000000539 0.000001362 0.000002571 10.5
03/31/2014 10:04:13.605097 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:04:14.604980 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002593 10.5
03/31/2014 10:04:15.604853 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002593 10.5
03/31/2014 10:04:16.604721 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:04:17.604603 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:04:18.604509 12.58 0.18 6.71 0.01 0.000000539 0.000001318 0.000002571 10.5
03/31/2014 10:04:19.604351 12.58 0.18 6.71 0.01 0.000000539 0.000001318 0.000002593 10.5
03/31/2014 10:04:20.604276 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:04:21.604146 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002571 10.5
03/31/2014 10:04:22.605062 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002593 10.6
03/31/2014 10:04:23.605929 12.58 0.18 6.71 0.01 0.000000546 0.000001340 0.000002593 10.6
03/31/2014 10:04:24.605722 12.58 0.18 6.71 0.01 0.000000546 0.000001318 0.000002593 10.6
03/31/2014 10:04:25.605584 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002615 10.5
03/31/2014 10:04:26.605468 12.58 0.18 6.71 0.01 0.000000543 0.000001340 0.000002615 10.5
03/31/2014 10:04:27.616984 12.58 0.18 6.71 0.01 0.000000543 0.000001340 0.000002615 10.5
03/31/2014 10:04:28.687685 12.58 0.18 6.71 0.01 0.000000543 0.000001362 0.000002593 10.5
03/31/2014 10:04:29.688084 12.58 0.18 6.71 0.01 0.000000543 0.000001362 0.000002593 10.5
03/31/2014 10:04:30.687958 12.58 0.18 6.71 0.01 0.000000543 0.000001340 0.000002593 10.5
03/31/2014 10:04:31.809569 12.58 0.18 7.32 0.01 0.000000539 0.000001384 0.000002593 10.5
03/31/2014 10:04:32.809729 12.58 0.18 7.32 0.01 0.000000539 0.000001362 0.000002593 10.5
03/31/2014 10:04:33.810786 12.58 0.18 6.71 0.01 0.000000539 0.000001362 0.000002593 10.6
03/31/2014 10:04:34.810438 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.6
03/31/2014 10:04:35.810414 12.58 0.18 7.32 0.01 0.000000539 0.000001362 0.000002571 10.6
03/31/2014 10:04:36.810188 12.58 0.18 7.32 0.01 0.000000543 0.000001362 0.000002593 10.6
03/31/2014 10:04:37.810076 12.58 0.18 6.71 0.01 0.000000541 0.000001362 0.000002593 10.6
03/31/2014 10:04:38.867466 12.58 0.18 6.71 0.01 0.000000541 0.000001318 0.000002593 10.5
03/31/2014 10:04:39.867804 12.58 0.18 6.71 0.01 0.000000543 0.000001318 0.000002593 10.5
03/31/2014 10:04:40.886721 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002593 10.5
03/31/2014 10:04:41.886686 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002571 10.5
03/31/2014 10:04:42.886424 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002615 10.5
03/31/2014 10:04:43.886328 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002593 10.5
03/31/2014 10:04:44.886195 12.58 0.18 6.71 0.01 0.000000541 0.000001318 0.000002593 10.5
03/31/2014 10:04:45.886102 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002593 10.5
03/31/2014 10:04:46.885922 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002615 10.5
03/31/2014 10:04:47.885880 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002593 10.6
03/31/2014 10:04:48.886695 12.58 0.18 6.71 0.01 0.000000539 0.000001318 0.000002593 10.6
03/31/2014 10:04:49.886538 12.58 0.18 6.71 0.01 0.000000541 0.000001318 0.000002615 10.5
03/31/2014 10:04:50.886467 12.58 0.18 6.71 0.01 0.000000541 0.000001362 0.000002593 10.5
03/31/2014 10:04:51.886293 12.58 0.18 6.71 0.01 0.000000543 0.000001362 0.000002571 10.5
03/31/2014 10:04:52.886184 12.58 0.18 6.71 0.01 0.000000539 0.000001362 0.000002615 10.5
03/31/2014 10:04:53.886085 12.58 0.18 6.71 0.01 0.000000537 0.000001318 0.000002593 10.5
03/31/2014 10:04:54.885916 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002593 10.5
03/31/2014 10:04:55.980009 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002593 10.5
03/31/2014 10:04:56.979253 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002593 10.6
03/31/2014 10:04:57.978525 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002593 10.6
03/31/2014 10:04:58.979433 12.58 0.18 6.71 0.01 0.000000539 0.000001362 0.000002593 10.6
03/31/2014 10:04:59.979273 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.6
03/31/2014 10:05:02.606530 12.58 0.18 6.71 0.01 0.000000539 0.000001318 0.000002571 10.6
03/31/2014 10:05:03.605880 12.58 0.18 6.71 0.01 0.000000537 0.000001318 0.000002593 10.6
03/31/2014 10:05:04.606708 12.58 0.18 6.71 0.01 0.000000541 0.000001318 0.000002571 10.5
03/31/2014 10:05:05.606566 12.58 0.18 6.71 0.01 0.000000539 0.000001318 0.000002571 10.5
03/31/2014 10:05:06.606480 12.58 0.18 6.71 0.01 0.000000539 0.000001318 0.000002593 10.5
03/31/2014 10:05:07.606317 12.58 0.18 6.71 0.01 0.000000541 0.000001318 0.000002593 10.5
03/31/2014 10:05:08.606192 12.58 0.18 6.71 0.01 0.000000541 0.000001318 0.000002593 10.5
03/31/2014 10:05:09.889941 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002593 10.5
03/31/2014 10:05:11.695372 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002593 10.6
03/31/2014 10:05:12.695678 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.6
03/31/2014 10:05:13.743535 12.58 0.18 6.71 0.01 0.000000539 0.000001362 0.000002571 10.5
03/31/2014 10:05:14.743435 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:05:15.743296 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:05:16.743524 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002549 10.5
03/31/2014 10:05:17.743098 12.58 0.18 6.71 0.01 0.000000541 0.000001362 0.000002571 10.5
03/31/2014 10:05:18.742924 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:05:19.742047 11.44 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:05:20.741695 11.44 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:05:21.741602 12.58 0.18 6.71 0.01 0.000000537 0.000001318 0.000002571 10.5
03/31/2014 10:05:22.741410 12.58 0.18 6.71 0.01 0.000000539 0.000001318 0.000002549 10.5
03/31/2014 10:05:23.741280 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:05:24.918398 12.58 0.18 6.71 0.01 0.000000539 0.000001318 0.000002571 10.5
03/31/2014 10:05:25.918067 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:05:26.918881 12.58 0.18 6.71 0.01 0.000000535 0.000001340 0.000002571 10.5
03/31/2014 10:05:27.918765 12.58 0.18 6.71 0.01 0.000000535 0.000001318 0.000002571 10.5
03/31/2014 10:05:28.919631 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:05:29.919633 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.6
03/31/2014 10:05:30.920406 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002571 10.6
03/31/2014 10:05:31.920261 12.58 0.18 6.71 0.01 0.000000539 0.000001362 0.000002571 10.6
03/31/2014 10:05:34.936859 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.6
03/31/2014 10:05:35.936943 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002593 10.6
03/31/2014 10:05:36.937848 12.58 0.18 6.71 0.01 0.000000539 0.000001362 0.000002571 10.6
03/31/2014 10:05:37.938715 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:05:38.938410 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:05:39.939247 12.58 0.18 6.71 0.01 0.000000539 0.000001362 0.000002571 10.5
03/31/2014 10:05:40.939284 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:05:41.939003 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002593 10.5
03/31/2014 10:05:42.938877 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.6
03/31/2014 10:05:43.938828 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.6
03/31/2014 10:05:44.938728 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.6
03/31/2014 10:05:45.938504 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:05:46.938451 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:05:47.938375 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:05:48.938143 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:05:49.939010 12.58 0.18 6.71 0.01 0.000000539 0.000001318 0.000002615 10.5
03/31/2014 10:05:50.938898 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:05:51.938765 12.58 0.18 6.71 0.01 0.000000537 0.000001318 0.000002571 10.5
03/31/2014 10:05:52.938622 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002593 10.5
03/31/2014 10:05:53.938500 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:05:54.939556 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:05:55.939244 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:05:56.939118 12.58 0.18 6.71 0.01 0.000000541 0.000001340 0.000002571 10.5
03/31/2014 10:05:57.939019 12.58 0.18 6.71 0.01 0.000000541 0.000001362 0.000002571 10.5
03/31/2014 10:05:58.938910 12.58 0.18 6.71 0.01 0.000000535 0.000001362 0.000002549 10.5
03/31/2014 10:05:59.938928 12.58 0.18 6.71 0.01 0.000000537 0.000001362 0.000002571 10.5
03/31/2014 10:06:00.938613 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:06:01.938487 12.58 0.18 6.71 0.01 0.000000539 0.000001362 0.000002571 10.5
03/31/2014 10:06:02.938362 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:06:03.938273 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:06:04.938225 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.5
03/31/2014 10:06:05.937985 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:06:06.937870 12.58 0.18 6.71 0.01 0.000000537 0.000001340 0.000002571 10.5
03/31/2014 10:06:07.937756 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002549 10.5
03/31/2014 10:06:08.937630 12.58 0.18 6.71 0.01 0.000000539 0.000001340 0.000002571 10.6
03/31/2014 10:06:09.937472 12.58 0.18 6.71 0.01 0.000000539 0.000001362 0.000002571 10.6
... 11454 more lines ...
|
Attachment 2: 31032014_Ta#42_HV_conditioning.JPG
|
|
645
|
Thursday, April 10, 2014, 14:03 |
Bevan Moss | Conditioning Station | Standard Operation | TM3 | TiC#4 | HV conditioning |
The TiC#4 target that is on TM3 was conditioned to 25kV today with the EINZEL lens at 12.5kV. There was one spark shortly after reaching 25kV that reset the bias PS and another one after the module sat at 25kV for ~30 minutes. Attached is the strip tool data and screen shot. Prior to starting the conditioning the power on the chiller needed to by cycled as the VFD on the fans had tripped due to a recurring inrush problem. |
Attachment 1: 10042014_Tm3_TiC#4_HV_conditioning.JPG
|
|
Attachment 2: 10042014_TM3_TiC#4_HV_conditioning.txt
|
Time TCS:IG1:RDVAC [T] TCS:IG2:RDVAC [T] TCS:IG1S:RDVAC [T] TCS:EL:RDVOL [V] TCS:EL:RDCUR [mA] TCS:BIAS:RDVOL [V] TCS:BIAS:RDCUR [uA] TCS:DIC1:RDCOND [Mohm]
04/10/2014 10:51:30.980156 0.000000961 0.000005912 BadVal BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:31.979337 0.000000959 0.000005934 BadVal BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:32.979181 0.000000959 0.000005912 BadVal BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:33.979125 0.000000959 0.000005912 BadVal BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:34.979530 0.000000957 0.000005912 BadVal BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:35.979903 0.000000957 0.000005912 BadVal BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:36.979928 0.000000959 0.000005912 BadVal BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:37.979724 0.000000959 0.000005912 BadVal BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:38.980475 0.000000957 0.000005890 BadVal BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:39.980322 0.000000959 0.000005912 BadVal BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:41.721703 0.000000957 0.000005912 BadVal BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:42.720980 0.000000959 0.000005912 BadVal BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:43.721829 0.000000959 0.000005912 BadVal BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:44.721709 0.000000959 0.000005934 0.000003912 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:45.721877 0.000000959 0.000005890 0.000003912 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:46.721604 0.000000959 0.000005912 0.000003934 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:47.721360 0.000000957 0.000005912 0.000003934 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:48.722248 0.000000959 0.000005890 0.000003912 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:49.722132 0.000000954 0.000005890 0.000003912 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:50.722206 0.000000954 0.000005912 0.000003934 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:51.722849 0.000000959 0.000005912 0.000003934 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:52.722691 0.000000959 0.000005934 0.000003912 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:53.723567 0.000000959 0.000005934 0.000003934 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:54.723552 0.000000957 0.000005912 0.000003934 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:55.723346 0.000000957 0.000005934 0.000003934 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:56.723198 0.000000957 0.000005934 0.000003934 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:57.723128 0.000000961 0.000005934 0.000003934 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:58.723980 0.000000961 0.000005912 0.000003912 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:51:59.723852 0.000000957 0.000005912 0.000003890 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:52:00.723765 0.000000959 0.000005912 0.000003934 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:52:01.723774 0.000000959 0.000005890 0.000003912 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:52:02.723712 0.000000959 0.000005912 0.000003934 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:52:03.723618 0.000000957 0.000005890 0.000003912 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:52:04.724220 0.000000957 0.000005890 0.000003912 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:52:05.724345 0.000000959 0.000005890 0.000003912 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:52:06.724243 0.000000957 0.000005890 0.000003934 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:52:07.723847 0.000000959 0.000005912 0.000003912 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:52:08.723737 0.000000957 0.000005912 0.000003912 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:52:09.734748 0.000000957 0.000005890 0.000003934 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:52:10.734468 0.000000961 0.000005912 0.000003934 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:52:11.734336 0.000000961 0.000005890 0.000003934 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:52:12.734202 0.000000957 0.000005934 0.000003912 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:52:13.735274 0.000000959 0.000005934 0.000003934 BadVal BadVal BadVal BadVal BadVal
04/10/2014 10:52:14.736049 0.000000959 0.000005890 0.000003934 6.10 BadVal BadVal BadVal BadVal
04/10/2014 10:52:15.735844 0.000000961 0.000005890 0.000003934 6.10 BadVal BadVal BadVal BadVal
04/10/2014 10:52:16.735707 0.000000959 0.000005912 0.000003934 6.71 BadVal BadVal BadVal BadVal
04/10/2014 10:52:17.735550 0.000000959 0.000005912 0.000003934 6.71 BadVal BadVal BadVal BadVal
04/10/2014 10:52:18.735469 0.000000957 0.000005912 0.000003890 6.10 0.01 BadVal BadVal BadVal
04/10/2014 10:52:19.735397 0.000000959 0.000005890 0.000003934 6.10 0.01 BadVal BadVal BadVal
04/10/2014 10:52:20.736220 0.000000959 0.000005912 0.000003912 6.10 0.01 BadVal BadVal BadVal
04/10/2014 10:52:21.737118 0.000000961 0.000005912 0.000003934 6.10 0.01 BadVal BadVal BadVal
04/10/2014 10:52:22.736946 0.000000961 0.000005912 0.000003912 6.10 0.01 BadVal BadVal BadVal
04/10/2014 10:52:23.736842 0.000000959 0.000005912 0.000003912 6.10 0.01 17.16 BadVal BadVal
04/10/2014 10:52:24.736718 0.000000957 0.000005912 0.000003912 6.10 0.01 17.16 BadVal BadVal
04/10/2014 10:52:25.736582 0.000000957 0.000005912 0.000003912 6.10 0.01 17.16 BadVal BadVal
04/10/2014 10:52:26.735970 0.000000959 0.000005912 0.000003912 6.71 0.01 18.31 BadVal BadVal
04/10/2014 10:52:27.764214 0.000000959 0.000005912 0.000003934 6.71 0.01 18.31 BadVal BadVal
04/10/2014 10:52:28.764296 0.000000959 0.000005912 0.000003912 6.10 0.01 17.16 0.18 BadVal
04/10/2014 10:52:29.764046 0.000000959 0.000005912 0.000003912 6.10 0.01 18.31 0.18 BadVal
04/10/2014 10:52:30.763942 0.000000959 0.000005912 0.000003912 6.10 0.01 17.16 0.18 BadVal
04/10/2014 10:52:31.764595 0.000000959 0.000005890 0.000003934 6.10 0.01 17.16 0.18 BadVal
04/10/2014 10:52:32.764669 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 BadVal
04/10/2014 10:52:33.765611 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 BadVal
04/10/2014 10:52:34.765460 0.000000959 0.000005912 0.000003912 6.10 0.01 17.16 0.18 BadVal
04/10/2014 10:52:35.765345 0.000000961 0.000005912 0.000003934 6.10 0.01 17.16 0.18 BadVal
04/10/2014 10:52:36.765196 0.000000959 0.000005912 0.000003912 6.71 0.01 17.16 0.18 BadVal
04/10/2014 10:52:37.765304 0.000000959 0.000005890 0.000003934 6.71 0.01 17.16 0.18 BadVal
04/10/2014 10:52:38.765944 0.000000959 0.000005890 0.000003934 6.10 0.01 17.16 0.18 BadVal
04/10/2014 10:52:39.765800 0.000000959 0.000005890 0.000003956 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:52:40.765665 0.000000959 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:52:41.766003 0.000000954 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:52:42.766445 0.000000959 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:52:43.767346 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:52:44.767203 0.000000961 0.000005890 0.000003890 6.10 0.01 18.31 0.18 13.1
04/10/2014 10:52:45.767105 0.000000959 0.000005890 0.000003890 6.10 0.01 18.31 0.18 13.1
04/10/2014 10:52:46.766994 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:52:47.766939 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:52:48.851745 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:52:49.877773 0.000000959 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:52:50.877440 0.000000959 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:52:51.877267 0.000000959 0.000005912 0.000003934 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:52:52.877165 0.000000959 0.000005890 0.000003934 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:52:53.877016 0.000000959 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:52:54.877020 0.000000959 0.000005890 0.000003912 6.10 0.01 18.31 0.18 13.1
04/10/2014 10:52:55.876788 0.000000959 0.000005890 0.000003934 6.10 0.01 18.31 0.18 13.1
04/10/2014 10:52:56.876639 0.000000959 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:52:57.876545 0.000000961 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:52:58.876493 0.000000961 0.000005912 0.000003912 6.10 0.01 18.31 0.18 13.1
04/10/2014 10:52:59.876303 0.000000959 0.000005890 0.000003934 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:53:00.876207 0.000000954 0.000005890 0.000003934 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:53:01.876057 0.000000957 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:02.973167 0.000000959 0.000005912 0.000003912 6.10 0.01 18.31 0.18 13.1
04/10/2014 10:53:04.127504 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:05.245403 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:06.245504 0.000000959 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:07.246554 0.000000959 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:08.246272 0.000000961 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:09.246145 0.000000959 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:10.245957 0.000000959 0.000005890 0.000003868 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:11.245855 0.000000959 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:12.246803 0.000000959 0.000005912 0.000003934 6.10 0.01 18.31 0.18 13.1
04/10/2014 10:53:13.246602 0.000000959 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:14.246883 0.000000957 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:15.247430 0.000000957 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:16.248242 0.000000959 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:17.248123 0.000000957 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:18.248018 0.000000959 0.000005912 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:19.247819 0.000000959 0.000005912 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:20.248760 0.000000957 0.000005912 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:21.248576 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:22.248523 0.000000959 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:23.248379 0.000000959 0.000005890 0.000003868 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:24.248306 0.000000961 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:25.248120 0.000000959 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:26.247992 0.000000959 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:27.247865 0.000000957 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:28.247725 0.000000957 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:29.248587 0.000000957 0.000005912 0.000003868 6.10 0.01 18.31 0.18 13.1
04/10/2014 10:53:30.248506 0.000000959 0.000005912 0.000003912 6.10 0.01 18.31 0.18 13.1
04/10/2014 10:53:31.477726 0.000000961 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:32.510989 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:33.570288 0.000000961 0.000005912 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:34.569946 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:35.569810 0.000000954 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:36.569693 0.000000959 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:37.569517 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:38.570683 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:39.570275 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:40.570189 0.000000957 0.000005890 0.000003912 6.10 0.01 18.31 0.18 13.1
04/10/2014 10:53:41.663248 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:42.796751 0.000000961 0.000005912 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:43.796790 0.000000959 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:53:44.796660 0.000000961 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:53:45.797531 0.000000959 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:53:46.797366 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:47.963516 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:48.963095 0.000000959 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:49.964000 0.000000957 0.000005890 0.000003890 6.10 0.01 18.31 0.18 13.1
04/10/2014 10:53:50.963885 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:51.963082 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:52.962598 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:53.963488 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:54.964455 0.000000959 0.000005890 0.000003934 6.10 0.01 18.31 0.18 13.1
04/10/2014 10:53:55.964257 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:56.964088 0.000000957 0.000005890 0.000003934 6.10 0.01 18.31 0.18 13.1
04/10/2014 10:53:57.964030 0.000000959 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:58.964837 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:53:59.964711 0.000000959 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:54:00.964585 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:54:01.964460 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:54:02.964438 0.000000959 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:03.965278 0.000000959 0.000005890 0.000003912 6.10 0.01 18.31 0.18 13.2
04/10/2014 10:54:04.965134 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:06.114629 0.000000959 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:07.114842 0.000000959 0.000005890 0.000003934 6.71 0.01 17.16 0.18 13.2
04/10/2014 10:54:08.114859 0.000000959 0.000005890 0.000003912 6.71 0.01 17.16 0.18 13.2
04/10/2014 10:54:09.115601 0.000000957 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:11.95567 0.000000959 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:12.95203 0.000000959 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:13.95062 0.000000959 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:14.94930 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:15.95034 0.000000954 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:16.94756 0.000000957 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:17.94640 0.000000959 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:18.94482 0.000000959 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:19.94499 0.000000959 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:20.94191 0.000000959 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:21.94057 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:22.94970 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:23.94867 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:24.95670 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:25.95555 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:26.95473 0.000000959 0.000005934 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:27.95345 0.000000957 0.000005890 0.000003868 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:28.95171 0.000000957 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:29.156966 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:30.219271 0.000000954 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:31.218823 0.000000959 0.000005868 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:32.218700 0.000000954 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:33.218572 0.000000959 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:34.219484 0.000000957 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:35.219324 0.000000954 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:36.219155 0.000000954 0.000005890 0.000003912 6.71 0.01 17.16 0.18 13.2
04/10/2014 10:54:37.219030 0.000000954 0.000005890 0.000003934 6.71 0.01 17.16 0.18 13.2
04/10/2014 10:54:38.218939 0.000000957 0.000005868 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:39.219905 0.000000957 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:40.219644 0.000000957 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:54:41.219569 0.000000957 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:54:42.219471 0.000000954 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:54:43.219276 0.000000954 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:54:44.220207 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:54:45.226449 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:54:46.225901 0.000000957 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:54:47.225793 0.000000957 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:54:48.225688 0.000000959 0.000005890 0.000003912 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:54:49.237062 0.000000954 0.000005890 0.000003912 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:54:50.237414 0.000000954 0.000005868 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:54:51.237318 0.000000957 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:54:52.237170 0.000000957 0.000005868 0.000003868 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:54:53.237085 0.000000959 0.000005890 0.000003890 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:54:54.237035 0.000000957 0.000005890 0.000003912 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:54:55.236818 0.000000959 0.000005890 0.000003912 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:54:56.236684 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:54:57.236508 0.000000957 0.000005868 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:54:58.236539 0.000000957 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:54:59.237346 0.000000954 0.000005934 0.000003934 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:00.238278 0.000000957 0.000005912 0.000003934 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:01.238061 0.000000957 0.000005868 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:02.237902 0.000000957 0.000005868 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:03.237965 0.000000957 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:04.238630 0.000000957 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:05.239504 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:06.239438 0.000000954 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:07.239297 0.000000957 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:08.240205 0.000000959 0.000005890 0.000003890 6.10 0.01 18.31 0.18 13.1
04/10/2014 10:55:09.240127 0.000000959 0.000005890 0.000003890 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:55:10.240229 0.000000954 0.000005890 0.000003912 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:55:11.240963 0.000000957 0.000005890 0.000003868 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:12.242313 0.000000959 0.000005868 0.000003868 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:13.242643 0.000000959 0.000005890 0.000003868 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:14.243588 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:15.243253 0.000000957 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:16.243399 0.000000957 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:17.243056 0.000000957 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:18.243878 0.000000957 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:19.244834 0.000000954 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:20.245696 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:21.245500 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:22.245413 0.000000957 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:23.245446 0.000000954 0.000005912 0.000003912 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:55:24.246172 0.000000954 0.000005912 0.000003912 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:55:25.247101 0.000000954 0.000005912 0.000003912 6.71 0.01 18.31 0.18 13.1
04/10/2014 10:55:26.246944 0.000000957 0.000005912 0.000003890 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:55:27.246745 0.000000957 0.000005890 0.000003956 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:28.246707 0.000000957 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:29.247622 0.000000957 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:30.247664 0.000000957 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:31.247309 0.000000957 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:32.247107 0.000000957 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:33.247041 0.000000959 0.000005890 0.000003868 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:34.247898 0.000000957 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:35.248757 0.000000957 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:36.248622 0.000000957 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:37.248512 0.000000957 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:38.248407 0.000000954 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:39.248278 0.000000954 0.000005868 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:40.248215 0.000000954 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:41.248104 0.000000959 0.000005912 0.000003912 6.71 0.01 17.16 0.18 13.2
04/10/2014 10:55:42.247859 0.000000959 0.000005868 0.000003890 6.71 0.01 17.16 0.18 13.2
04/10/2014 10:55:43.247775 0.000000957 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:44.248699 0.000000959 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:55:45.248591 0.000000954 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:46.248414 0.000000954 0.000005890 0.000003934 6.10 0.01 18.31 0.18 13.1
04/10/2014 10:55:47.248231 0.000000954 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:48.248200 0.000000957 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:50.356306 0.000000959 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:51.356770 0.000000954 0.000005934 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:52.356658 0.000000957 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:53.356479 0.000000957 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:55.692384 0.000000959 0.000005912 0.000003868 6.71 0.01 18.31 0.18 13.1
04/10/2014 10:55:56.693040 0.000000959 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:57.693001 0.000000959 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:58.692827 0.000000959 0.000005912 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:55:59.692705 0.000000954 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:00.693597 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:01.693469 0.000000959 0.000005912 0.000003934 6.10 0.01 18.31 0.18 13.1
04/10/2014 10:56:02.693310 0.000000959 0.000005890 0.000003934 6.71 0.01 17.16 0.18 13.2
04/10/2014 10:56:03.693156 0.000000959 0.000005890 0.000003934 6.71 0.01 17.16 0.18 13.2
04/10/2014 10:56:04.693036 0.000000959 0.000005890 0.000003934 6.10 0.01 18.31 0.18 13.2
04/10/2014 10:56:05.692994 0.000000957 0.000005912 0.000003934 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:56:06.692862 0.000000957 0.000005890 0.000003934 6.71 0.01 17.16 0.18 13.2
04/10/2014 10:56:07.693748 0.000000954 0.000005912 0.000003912 6.71 0.01 17.16 0.18 13.2
04/10/2014 10:56:08.693600 0.000000954 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:56:09.694444 0.000000957 0.000005868 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:56:10.695298 0.000000957 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:56:11.695160 0.000000957 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:56:12.695091 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:56:13.694987 0.000000954 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:56:14.694827 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.2
04/10/2014 10:56:15.694742 0.000000959 0.000005912 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:16.694531 0.000000959 0.000005934 0.000003890 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:56:17.694420 0.000000959 0.000005890 0.000003890 6.71 0.01 17.16 0.18 13.1
04/10/2014 10:56:18.694323 0.000000957 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:19.694196 0.000000954 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:20.694200 0.000000954 0.000005912 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:21.693936 0.000000957 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:22.693796 0.000000957 0.000005890 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:23.693687 0.000000959 0.000005890 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:24.693613 0.000000957 0.000005912 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:25.693505 0.000000959 0.000005868 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:26.693314 0.000000957 0.000005868 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:27.693140 0.000000957 0.000005890 0.000003890 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:28.693023 0.000000957 0.000005912 0.000003934 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:29.692952 0.000000959 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:30.693831 0.000000954 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:31.693711 0.000000957 0.000005934 0.000003956 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:32.694613 0.000000959 0.000005912 0.000003912 6.10 0.01 17.16 0.18 13.1
04/10/2014 10:56:33.694463 0.000000957 0.000005912 0.000003890 6.71 0.01 17.16 0.18 13.2
04/10/2014 10:56:34.694275 0.000000957 0.000005868 0.000003890 6.71 0.01 17.16 0.18 13.2
... 7596 more lines ...
|
273
|
Friday, March 22, 2013, 18:49 |
Aurelia Laxdal | Conditioning Station | Development | TM4 | FEBIAD HP with Dummy Target | Commissioning TCS - Part 1 |
Today, the commissioning of the TCS went well. The water/vacuum interlocks were checked OK. Then, the TCS was HV conditioned up to 33kV.
Tomorrow, the heaters will be ramped up. |
274
|
Saturday, March 23, 2013, 18:14 |
Aurelia Laxdal | Conditioning Station | Development | TM4 | FEBIAD HP with Dummy Target | Commissioning TCS - Part 2 Completed |
After few hickups the TCS was successfuly commissioned.
In the morning, in the middle of the heaters conditioning: 360A/250A for the TGHT/TBHT, the target heater PS tripped and it didn't recover immediately. I cannot tell if the card is still OK. Before this even, there were few bias current sparks created when the ionized tube was raised above 250A; this also tripped the Einzel lens. Re-started the Heaters, and did the conditioning with the rest (HV, Optics) Off. After the heaters were at nominal values, raised the rest: HV to 30kV and the Optics. And everything held fine. The current on the FC was 3micro-Amperes. The water temperature levels stayed below 25 deg Celsius.
The problems that need to be fixed:
1. The biggest problem was created by the target power supply: the new card is not calibrated, the voltage is not regulating, but the main problem is the grounding of the output of power supply. Friedhelm measured 0.2V across the return and the cage (200mili-ohms resistance) . This was probably overlooked.
2. The Auto-ramping routine didn't work. Controls would need to fix this.
3. The RGA didn't want to connect from the account Don created for me. So, we have only a RGA scan taken on Friday by Don, when the target was cold.
|
1095
|
Tuesday, October 06, 2015, 21:12 |
Aurelia Laxdal | South Hot-Cell | Repair | TM3 | | TM3 service cap inspection in target hall |
David Wang and I did the service cap inspection in target hall: see WP I2015-10-06-10.<br>
Signs of sparking were seen underneath 3 insulators: E ("Cavity"), A ("Oven+") and B ("Oven -") at the pair water lines coming out of the feed-through and going to the HV chase. Specifically one water line of E is very close to the feed-through insulator (2-3mm gap) and here the sparking sings are more prominent: see picture. |
Attachment 1: Pic1.jpg
|
|
1250
|
Thursday, June 02, 2016, 18:04 |
Aurelia Laxdal | South Hot-Cell | Repair | TM3 | | 5K mega test on TM3 |
David Wang wrote: |
I did a 5KV mega test on TM3. Only5Mohm @5kV between coli line PQ and chassis(ground). It confirmed the close gap between P,Q and GE. Other lines are ok. See attachment.
|
See attachment for sparking signs of the water lines mentioned above (Coil and Ground Electrode lines) in TM3 Source Tray. |
Attachment 1: DSC_0821.JPG
|
|
Attachment 2: DSC_0822.JPG
|
|
1252
|
Friday, June 03, 2016, 16:04 |
Aurelia Laxdal | South Hot-Cell | Repair | TM3 | | 5K mega test on TM3 |
Aurelia Laxdal wrote: |
David Wang wrote: |
I did a 5KV mega test on TM3. Only5Mohm @5kV between coli line PQ and chassis(ground). It confirmed the close gap between P,Q and GE. Other lines are ok. See attachment.
|
See attachment for sparking signs of the water lines mentioned above (Coil and Ground Electrode lines) in TM3 Source Tray.
|
PLEASE NOTE: the resistance values for A and B on David's attachment are not correct since these measurements were done after Chad removed the target and the heat-shield bracket was moved, so consequently the water lines were not at their nominal locations. Once they were back in plac,e David took another measurement and the resistance values for A and B showed also 3,000 MOhm (as it was for most of the lines). The only lines that still showed a low resistance were the Coil lines. |
1289
|
Tuesday, July 19, 2016, 14:35 |
Aurelia Laxdal | Conditioning Station | Repair | TM3 | | TM3 leak checked at TCS |
David Wang wrote: |
Last Friday July-15, TM3 was leak checked at TCS. All air to vacuum seals on service cap are good. Helium pressure test on TM3 water lines(except GE, HS, and Window) were done. Leak detector has no response at 0.0xE-9 atm.cc/sec. They are leak free.
|
Friday July 15: did HV tests (with no target installed and with the anode wire removed ). <br>
TM3 went up to 22.5kV. At this voltage it started sparking. The EL was kept at 1/2 with respect to the HV BIAS. <br>
Also HV tests were done while the EL was kept at zero V. In this case, the EL voltage was once in a while spiking up (up to 4kV). <br>
During sparking we saw responses in the vacuum. Checked the Service Cap (with the HV Cover) in the Target Hall: we couldn't hear/see sparks there. <br>
The HV was turned off in the same day, after 5:30pm. |
1414
|
Monday, December 05, 2016, 10:04 |
Aurelia Laxdal | Conditioning Station | Repair | TM3 | | TM3 vacuum system have been connected at TCS. |
David Wang wrote: |
All water circuits are by passed on the top of TM3. Water pumps and vacuum turbo- pumps are started. Everything is ready for TM3 camera sparking check.
|
Wednesday November 30st: <br>
- 3 cameras (1 UV and 2 regular cameras) were installed in the Service cap and the module was pumped down. 1 UV and 1 regular were placed on the East side of the module, aiming to water lines entering the Service Tray; 1 regular camera was placed on the South side of the module, aiming at the bottom of the blue insulators.
- started pumping down. <br>
Thursday, December 1st: <br>
- 1st HV test done. TM3 started sparking above 20kV. The cameras responses were consistent with the sparks. However, the screens were flickering and it was not clear if the responses were due to electromagnetic interferences; or due to cameras' gain change caused by high intensity light due to sparking.<br>
- added helium using the regulating valve. same results were found; <br>
Friday, December 2nd: <br>
- vented the module. <br>
- shielded the cameras' cables with aluminum tape. <br>
- removed 2 blue insulator (EE and HS) and shortened them by 1". They were found too long: HS blue insulator touching the horizontal HV water line; and the EE blue insulator at 2-3mm far from the HV water line; re-installed both shortened blue insulators.<br>
- changes the location of the cameras: 1 regular camera taped to the UV camera were placed South, aiming at the bottom of the blue insulators; 1 regular camera was kept on the East side of the module, also aiming at the bottom of the blue insulators. <br>
- started pumping down. <br>
Saturday, December 3rd: <br>
- 2nd HV test was done, with cameras' installed as mentioned above. results were consistent with the first test results. <br>
- added helium. results were the same. <br>
- vented and changed the location of the cameras: 1 regular camera taped to the UV camera were placed at the top of the Service Tray (South-West corner) aiming down the Service Tray; 1 regular camera was kept (its setup was not touched at all) on the East side of the module, aiming at the bottom of the blue insulators. <br>
- attempted to pump down, but the module was leaking. reopened the flanges, cleaned them, attempted again to pump down, but it showed a bigger leak. <br>
Sunday, December 4th: <br>
- the third attempt to pump down was successful (without reopening the module!). <br>
Monday, December 5th: <br>
- did the 3rd HV test with the cameras installed as described on December 3rd. same results as before were seen. <br>
- again helium was added and same results were noticed. <br>
- stopped pumps and prepared for venting to relocate the cameras lower in the Service Tray.
|
1417
|
Friday, December 09, 2016, 18:05 |
Aurelia Laxdal | Conditioning Station | Repair | TM3 | | TM3 vacuum system have been connected at TCS. |
Aurelia Laxdal wrote: |
David Wang wrote: |
All water circuits are by passed on the top of TM3. Water pumps and vacuum turbo- pumps are started. Everything is ready for TM3 camera sparking check.
|
Wednesday November 30st: <br>
- 3 cameras (1 UV and 2 regular cameras) were installed in the Service cap and the module was pumped down. 1 UV and 1 regular were placed on the East side of the module, aiming to water lines entering the Service Tray; 1 regular camera was placed on the South side of the module, aiming at the bottom of the blue insulators.
- started pumping down. <br>
Thursday, December 1st: <br>
- 1st HV test done. TM3 started sparking above 20kV. The cameras responses were consistent with the sparks. However, the screens were flickering and it was not clear if the responses were due to electromagnetic interferences; or due to cameras' gain change caused by high intensity light due to sparking.<br>
- added helium using the regulating valve. same results were found; <br>
Friday, December 2nd: <br>
- vented the module. <br>
- shielded the cameras' cables with aluminum tape. <br>
- removed 2 blue insulator (EE and HS) and shortened them by 1". They were found too long: HS blue insulator touching the horizontal HV water line; and the EE blue insulator at 2-3mm far from the HV water line; re-installed both shortened blue insulators.<br>
- changes the location of the cameras: 1 regular camera taped to the UV camera were placed South, aiming at the bottom of the blue insulators; 1 regular camera was kept on the East side of the module, also aiming at the bottom of the blue insulators. <br>
- started pumping down. <br>
Saturday, December 3rd: <br>
- 2nd HV test was done, with cameras' installed as mentioned above. results were consistent with the first test results. <br>
- added helium. results were the same. <br>
- vented and changed the location of the cameras: 1 regular camera taped to the UV camera were placed at the top of the Service Tray (South-West corner) aiming down the Service Tray; 1 regular camera was kept (its setup was not touched at all) on the East side of the module, aiming at the bottom of the blue insulators. <br>
- attempted to pump down, but the module was leaking. reopened the flanges, cleaned them, attempted again to pump down, but it showed a bigger leak. <br>
Sunday, December 4th: <br>
- the third attempt to pump down was successful (without reopening the module!). <br>
Monday, December 5th: <br>
- did the 3rd HV test with the cameras installed as described on December 3rd. same results as before were seen. <br>
- again helium was added and same results were noticed. <br>
- stopped pumps and prepared for venting to relocate the cameras lower in the Service Tray.
|
Monday, December 5th (Con't): <br>
- vented and changed the location of the cameras;<br>
- the UV camera was placed (tapped on an aluminum spacer and on the grounded water lines) at the middle of the Service Tray (right below the first Mykroy insulator-plate) aiming down the Service Tray (at the second Mykroy insulator-plate and water lines) ; <br>
- 1 regular camera was placed right below the UV camera, also using a spacer (taped with the camera on grounded water lines) and also aiming down the Service Tray (at the second Mykroy insulator-plate and water lines); The spacer for this camera had 300counts on it after it was removed from TM3, after the test;<br>
- NOTE: when installing the 1st regular camera: dropped a small allen key in the Service Tray. It landed on last insulator with the epoxi and water lines, between the first HV line and the wall of the grounded Service Tray. This insulator is down below in the Service tray. Used a magnet stick to recuperate the allen key. <br>
- the 2nd regular camera was placed at the top of the Service Tray (tapped on the grounded water lines) looking down at the first Mykroy insulator-plate and water lines. NOTE: this camera had a filter applied to its lens. <br>
- pumped down <br>
Tuesday, December 6th:<br>
- did the 4th HV test with the cameras installed as described on December 5rd Tuesday morning very early . Same results as before were seen, except that the top camera (2nd regular camera) with the filter applied to its lens was flickering less than the other 2 cameras, and for small sparks this camera was not reacting at all. <br>
- vented shortly after the 4th test and relocated the cameras for the 5th HV test (last test); <br>
- placed only 2 regular cameras in the Service Tray;<br>
- 1 regular camera was placed right below the 2nd Mykroy insulator-plate, aiming down the Service Tray at the last insulator with the epoxi and water lines (this insulator is down below in the Service tray); <br>
- the 1st regular camera was placed a level up in the Service Tray, right below the 1st Mykroy insulator-plate, aiming down the Service Tray at the 2nd Mykroy insulator-plate; <br>
- NOTE: no spacers were used<br>
- pumped down and did the 5th HV test with the cameras installed as described above in the evening; <br>
- again same results as before were seen: the camera with the filter applied to its lens was flickering less than the other camera, and for small sparks this camera was not reacting at all. <br>
- NOTE: for this last test heard sparks outside (at atmosphere) when the HV was pushed at 26kV; <br>
Wednesday, December 7th:<br>
- vented and removed the cameras from TM3. No contamination was found on the cameras and wires. <br>
- TM3 was moved to the silo.<br> |