Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  Evaporator_2, Page 13 of 35  Not logged in ELOG logo
New entries since:Wed Dec 31 16:00:00 1969
ID Date Author Target Material Target Oven W.O# Sourcedown Status Subject
  261   Monday, November 09, 2015, 14:22 Laura Lambert 41077-3LP-SISDoneSintering TaC coating on 'temporary endcaps' springs (Slope changed to 1.04)

John Wong wrote:

Laura Lambert wrote:

 Thursday, November 05, 2015, 16:19

Springs coated with TaC and loaded into container for sintering.

3x Ta heat shield.

EVAP2 currently pumping down.


 Friday, November 06, 2015, 08:39

Opened HS water valves, no leaks detected.

Tapped up manually to 35A.

IGP1 = 5.5e-7 Torr.


 Friday, November 06, 2015, 10:08

IGP1 = 5.8e-7 Torr.

Vacuum stable.

Initiating auto-ramp at 0.2A/min and 0.1A/min with final setpoints of 450A and 230A for the TGHT and TBHT respectively.

See photos attached to view how springs were coated and loaded into container.


 Monday, November 09, 2015, 09:07

TGHT reached 450A, 7.2V ~8pm sat, nov 7th.

OMEGA = 1775C

IGP1 = 3.2e-7 Torr.

Tapping up manually to 1900C.


 

 Monday, November 09, 2015, 12:30

Checked the temp on the inner surface (not at the spring) of the container was around 2050C at 500A. 

The TaC coated spring is reading at 1900C; that's what the OMEGA is reading with the setting for 2-colors and slope at 0.98.

Lower TGHT to 450A/ 7.12V.

By comparing the data from the previous temp curve (OMEGA slope at 0.98) with no material inside the container,  the voltage and current plot is aligned with the curve.

Changed OMEGA setting slope to 1.04.  Current temp is reading at 1917C at 450A.

For 2-Color setting, the emissivity should be negligible.  That makes me believe that different materials has different slope setting for the temp reading.  

One thing to note is that, the sintering temp of TaC is around 1900-2000C.  It was clear that the container was fully sintered with TGHT at 450A.

Will need to gather more data with different material to understand the slope setting better. 

Try: 

SiC at 1700C, slope at 1.04

ZrC at 1900C, slope at 104

Ta at 1900C, slope at 0.98.

Compare the data...


 

 Monday, November 09, 2015, 14:20

IGP1 = 3.3e-7 Torr.

TGHT = 450A, 7.12V

OMEGA = 1915C

Initiating auto-cool at 2.0A/min and 1.0A/min for the TGHT and TBHT respectively.


 Tuesday, November 10, 2015, 13:23

Vented and opened this morning.

springs look [mostly] sintered.

see photo attached.

 

Attachment 1: Evernote_Camera_Roll_20151110_113123.jpg
Evernote_Camera_Roll_20151110_113123.jpg
  262   Tuesday, November 10, 2015, 14:34 Laura LambertSiC41077-3LP-SISDonetemp measurements with 3xTa h.s on LPSIS (small block) container with SiC

 Tuesday, November 10, 2015, 14:20

Container previously sintered with TaC coating and temperature measurements taken [see elog ID# 256, 257, 258]

3x Ta heat shields

2 packs of 5 SiC discs ~2inches from the end of the right side wrt ionizer tube into page.

Loaded into evap2 and currently pumping down.

see photos attached.


 Monday, November 16, 2015, 11:24

HS water valves opened up this morning, no leaks detected.

Tapped up manually to 35A.

Vacuum stable.

IGP1 = 3.4e-7 Torr.

Initiating auto-ramp at 1.0A/min and 0.5A/min for the TGHT and TBHT respectively.

Final setpoints: 250A for the TGHT and 230A for the TBHT


Monday, November 16, 2015, 13:29

Vacuum too high.

IGP1 = 2.9e-5 Torr.

Decreasing ramping rate down to 0.5A/min and 0.2A/min for the TGHT and TBHT respectively.


 Tuesday, November 17, 2015, 09:12

IGP1 = 1.45e-7 Torr.

TGHT = 250A, 3.28V

TBHT = 230A, 3.33V

OMEGA = 1194C

Tapping up manually to 1900C


 Tuesday, November 17, 2015, 11:22

Tapped up manually to 1800C, realized SiC should only be tapped up to 1700C, backed off immediately down to 415A from 460A/ 1800C.

See attachment for photo of striptool/ time spent above 1700C [~20min].


 Tuesday, November 17, 2015, 15:46

TGHT remained at 1700C for ~4hrs.

TGHT = 420A, 6.44V.

IGP1 = 2.3e-7 Torr.

Initiating cool-down at 2.0A/min and 1.0A/min for the TGHT and TBHT respectively.

See attachment for raw data of temperature measurements.


 Wednesday, November 18, 2015, 13:32

Vented in air and took out this morning.

Looks good, no visual changes with the container.


 

Attachment 1: Evernote_Camera_Roll_20151110_110128.jpg
Evernote_Camera_Roll_20151110_110128.jpg
Attachment 2: Evernote_Camera_Roll_20151110_110129.jpg
Evernote_Camera_Roll_20151110_110129.jpg
Attachment 3: striptool_screenshot_above_1700C.JPG
striptool_screenshot_above_1700C.JPG
Attachment 4: temp_measurements_on_LPSIS_(small_block)_container.xlsx
  265   Tuesday, December 01, 2015, 11:35 Anders MjosEmpty41077-5LP-SISIn progressTest#2: Measuring temp with thermocouple (away from centre) & 2C-IR

 Data from thermocouple attached. Raw data and Excel sheet with plot.

Attachment 1: 2015-10-08_Small_Centre_Block.xlsx
Attachment 2: 2015-10-08.4ch
°C	Date Time	T1 - C	T2 - C	T3 - C	T4 - K	
1	2015-10-08 14:35:51	25.0	 	 	 	
2	2015-10-08 14:36:21	25.7	 	 	 	
3	2015-10-08 14:36:51	26.1	 	 	 	
4	2015-10-08 14:37:21	26.9	 	 	 	
5	2015-10-08 14:37:51	28.0	 	 	 	
6	2015-10-08 14:38:21	28.6	 	 	 	
7	2015-10-08 14:38:51	29.7	 	 	 	
8	2015-10-08 14:39:21	30.8	 	 	 	
9	2015-10-08 14:39:51	32.5	 	 	 	
10	2015-10-08 14:40:21	34.4	 	 	 	
11	2015-10-08 14:40:51	36.3	 	 	 	
12	2015-10-08 14:41:21	39.0	 	 	 	
13	2015-10-08 14:41:51	42.5	 	 	 	
14	2015-10-08 14:42:21	46.4	 	 	 	
15	2015-10-08 14:42:51	51.3	 	 	 	
16	2015-10-08 14:43:21	57.3	 	 	 	
17	2015-10-08 14:43:51	64.7	 	 	 	
18	2015-10-08 14:44:21	72.1	 	 	 	
19	2015-10-08 14:44:51	82.1	 	 	 	
20	2015-10-08 14:45:21	91.5	 	 	 	
21	2015-10-08 14:45:51	104.3	 	 	 	
22	2015-10-08 14:46:21	116.5	 	 	 	
23	2015-10-08 14:46:51	133.1	 	 	 	
24	2015-10-08 14:47:21	151.2	 	 	 	
25	2015-10-08 14:47:51	167.8	 	 	 	
26	2015-10-08 14:48:21	188.5	 	 	 	
27	2015-10-08 14:48:51	207.7	 	 	 	
28	2015-10-08 14:49:21	228.1	 	 	 	
29	2015-10-08 14:49:51	251.5	 	 	 	
30	2015-10-08 14:50:21	275.9	 	 	 	
31	2015-10-08 14:50:51	297.5	 	 	 	
32	2015-10-08 14:51:21	319.8	 	 	 	
33	2015-10-08 14:51:51	343.1	 	 	 	
34	2015-10-08 14:52:21	366.5	 	 	 	
35	2015-10-08 14:52:51	388.6	 	 	 	
36	2015-10-08 14:53:21	410.5	 	 	 	
37	2015-10-08 14:53:51	432.4	 	 	 	
38	2015-10-08 14:54:21	452.0	 	 	 	
39	2015-10-08 14:54:51	471.5	 	 	 	
40	2015-10-08 14:55:21	491.6	 	 	 	
41	2015-10-08 14:55:51	507.5	 	 	 	
42	2015-10-08 14:56:21	523.7	 	 	 	
43	2015-10-08 14:56:51	539.1	 	 	 	
44	2015-10-08 14:57:21	554.6	 	 	 	
45	2015-10-08 14:57:51	567.7	 	 	 	
46	2015-10-08 14:58:21	581.1	 	 	 	
47	2015-10-08 14:58:51	593.0	 	 	 	
48	2015-10-08 14:59:21	607.7	 	 	 	
49	2015-10-08 14:59:51	623.0	 	 	 	
50	2015-10-08 15:00:21	638.0	 	 	 	
51	2015-10-08 15:00:51	650.2	 	 	 	
52	2015-10-08 15:01:21	661.4	 	 	 	
53	2015-10-08 15:01:51	670.3	 	 	 	
54	2015-10-08 15:02:21	678.0	 	 	 	
55	2015-10-08 15:02:51	683.8	 	 	 	
56	2015-10-08 15:03:21	688.8	 	 	 	
57	2015-10-08 15:03:51	693.0	 	 	 	
58	2015-10-08 15:04:21	697.2	 	 	 	
59	2015-10-08 15:04:51	700.7	 	 	 	
60	2015-10-08 15:05:21	703.5	 	 	 	
61	2015-10-08 15:05:51	706.3	 	 	 	
62	2015-10-08 15:06:21	709.1	 	 	 	
63	2015-10-08 15:06:51	711.5	 	 	 	
64	2015-10-08 15:07:21	713.7	 	 	 	
65	2015-10-08 15:07:51	716.1	 	 	 	
66	2015-10-08 15:08:21	718.3	 	 	 	
67	2015-10-08 15:08:51	720.5	 	 	 	
68	2015-10-08 15:09:21	722.5	 	 	 	
69	2015-10-08 15:09:51	724.7	 	 	 	
70	2015-10-08 15:10:21	726.8	 	 	 	
71	2015-10-08 15:10:51	728.8	 	 	 	
72	2015-10-08 15:11:21	730.8	 	 	 	
73	2015-10-08 15:11:51	733.0	 	 	 	
74	2015-10-08 15:12:21	734.7	 	 	 	
75	2015-10-08 15:12:51	736.9	 	 	 	
76	2015-10-08 15:13:21	738.8	 	 	 	
77	2015-10-08 15:13:51	740.8	 	 	 	
78	2015-10-08 15:14:21	742.7	 	 	 	
79	2015-10-08 15:14:51	744.7	 	 	 	
80	2015-10-08 15:15:21	746.6	 	 	 	
81	2015-10-08 15:15:51	748.6	 	 	 	
82	2015-10-08 15:16:21	750.6	 	 	 	
83	2015-10-08 15:16:51	752.5	 	 	 	
84	2015-10-08 15:17:21	754.4	 	 	 	
85	2015-10-08 15:17:51	756.3	 	 	 	
86	2015-10-08 15:18:21	758.3	 	 	 	
87	2015-10-08 15:18:51	760.2	 	 	 	
88	2015-10-08 15:19:21	762.2	 	 	 	
89	2015-10-08 15:19:51	764.1	 	 	 	
90	2015-10-08 15:20:21	765.8	 	 	 	
91	2015-10-08 15:20:51	767.8	 	 	 	
92	2015-10-08 15:21:21	769.9	 	 	 	
93	2015-10-08 15:21:51	771.9	 	 	 	
94	2015-10-08 15:22:21	773.8	 	 	 	
95	2015-10-08 15:22:51	775.5	 	 	 	
96	2015-10-08 15:23:21	777.3	 	 	 	
97	2015-10-08 15:23:51	779.3	 	 	 	
98	2015-10-08 15:24:21	781.3	 	 	 	
99	2015-10-08 15:24:51	783.1	 	 	 	
100	2015-10-08 15:25:21	785.0	 	 	 	
101	2015-10-08 15:25:51	786.9	 	 	 	
102	2015-10-08 15:26:21	788.8	 	 	 	
103	2015-10-08 15:26:51	790.6	 	 	 	
104	2015-10-08 15:27:21	792.6	 	 	 	
105	2015-10-08 15:27:51	794.5	 	 	 	
106	2015-10-08 15:28:21	796.3	 	 	 	
107	2015-10-08 15:28:51	798.3	 	 	 	
108	2015-10-08 15:29:21	800.1	 	 	 	
109	2015-10-08 15:29:51	802.0	 	 	 	
110	2015-10-08 15:30:21	803.8	 	 	 	
111	2015-10-08 15:30:51	805.8	 	 	 	
112	2015-10-08 15:31:21	807.8	 	 	 	
113	2015-10-08 15:31:51	809.6	 	 	 	
114	2015-10-08 15:32:21	811.5	 	 	 	
115	2015-10-08 15:32:51	813.3	 	 	 	
116	2015-10-08 15:33:21	815.3	 	 	 	
117	2015-10-08 15:33:51	817.2	 	 	 	
118	2015-10-08 15:34:21	819.1	 	 	 	
119	2015-10-08 15:34:51	820.8	 	 	 	
120	2015-10-08 15:35:21	822.8	 	 	 	
121	2015-10-08 15:35:51	824.6	 	 	 	
122	2015-10-08 15:36:21	826.5	 	 	 	
123	2015-10-08 15:36:51	828.3	 	 	 	
124	2015-10-08 15:37:21	830.3	 	 	 	
125	2015-10-08 15:37:51	832.1	 	 	 	
126	2015-10-08 15:38:21	834.0	 	 	 	
127	2015-10-08 15:38:51	835.8	 	 	 	
128	2015-10-08 15:39:21	837.7	 	 	 	
129	2015-10-08 15:39:51	840.4	 	 	 	
130	2015-10-08 15:40:21	848.5	 	 	 	
131	2015-10-08 15:40:51	860.4	 	 	 	
132	2015-10-08 15:41:21	873.7	 	 	 	
133	2015-10-08 15:41:51	884.4	 	 	 	
134	2015-10-08 15:42:21	891.8	 	 	 	
135	2015-10-08 15:42:51	896.6	 	 	 	
136	2015-10-08 15:43:21	899.4	 	 	 	
137	2015-10-08 15:43:51	901.0	 	 	 	
138	2015-10-08 15:44:21	902.1	 	 	 	
139	2015-10-08 15:44:51	902.8	 	 	 	
140	2015-10-08 15:45:21	903.0	 	 	 	
141	2015-10-08 15:45:51	903.2	 	 	 	
142	2015-10-08 15:46:21	903.2	 	 	 	
143	2015-10-08 15:46:51	903.2	 	 	 	
144	2015-10-08 15:47:21	903.5	 	 	 	
145	2015-10-08 15:47:51	903.5	 	 	 	
146	2015-10-08 15:48:21	903.5	 	 	 	
147	2015-10-08 15:48:51	903.5	 	 	 	
148	2015-10-08 15:49:21	903.5	 	 	 	
149	2015-10-08 15:49:51	903.5	 	 	 	
150	2015-10-08 15:50:21	903.5	 	 	 	
151	2015-10-08 15:50:51	903.3	 	 	 	
152	2015-10-08 15:51:21	903.5	 	 	 	
153	2015-10-08 15:51:51	903.5	 	 	 	
154	2015-10-08 15:52:21	903.2	 	 	 	
155	2015-10-08 15:52:51	903.5	 	 	 	
156	2015-10-08 15:53:21	903.3	 	 	 	
157	2015-10-08 15:53:51	903.0	 	 	 	
158	2015-10-08 15:54:21	903.3	 	 	 	
159	2015-10-08 15:54:51	903.3	 	 	 	
160	2015-10-08 15:55:21	903.2	 	 	 	
161	2015-10-08 15:55:51	903.0	 	 	 	
162	2015-10-08 15:56:21	903.0	 	 	 	
163	2015-10-08 15:56:51	906.3	 	 	 	
164	2015-10-08 15:57:21	926.1	 	 	 	
165	2015-10-08 15:57:51	945.0	 	 	 	
166	2015-10-08 15:58:21	956.8	 	 	 	
167	2015-10-08 15:58:51	964.6	 	 	 	
168	2015-10-08 15:59:21	968.6	 	 	 	
169	2015-10-08 15:59:51	970.8	 	 	 	
170	2015-10-08 16:00:21	972.1	 	 	 	
171	2015-10-08 16:00:51	972.7	 	 	 	
172	2015-10-08 16:01:21	973.0	 	 	 	
173	2015-10-08 16:01:51	973.5	 	 	 	
174	2015-10-08 16:02:21	973.5	 	 	 	
175	2015-10-08 16:02:51	973.5	 	 	 	
176	2015-10-08 16:03:21	973.5	 	 	 	
177	2015-10-08 16:03:51	973.5	 	 	 	
178	2015-10-08 16:04:21	973.6	 	 	 	
179	2015-10-08 16:04:51	973.5	 	 	 	
180	2015-10-08 16:05:21	973.5	 	 	 	
181	2015-10-08 16:05:51	973.5	 	 	 	
182	2015-10-08 16:06:21	973.5	 	 	 	
183	2015-10-08 16:06:51	973.5	 	 	 	
184	2015-10-08 16:07:21	973.5	 	 	 	
185	2015-10-08 16:07:51	973.5	 	 	 	
186	2015-10-08 16:08:21	992.7	 	 	 	
187	2015-10-08 16:08:51	1018.0	 	 	 	
188	2015-10-08 16:09:21	1032.8	 	 	 	
189	2015-10-08 16:09:51	1041.1	 	 	 	
190	2015-10-08 16:10:21	1045.0	 	 	 	
191	2015-10-08 16:10:51	1047.1	 	 	 	
192	2015-10-08 16:11:21	1047.7	 	 	 	
193	2015-10-08 16:11:51	1053.5	 	 	 	
194	2015-10-08 16:12:21	1082.7	 	 	 	
195	2015-10-08 16:12:51	1100.0	 	 	 	
196	2015-10-08 16:13:21	1108.8	 	 	 	
197	2015-10-08 16:13:51	1112.5	 	 	 	
198	2015-10-08 16:14:21	1114.2	 	 	 	
199	2015-10-08 16:14:51	1116.9	 	 	 	
200	2015-10-08 16:15:21	1148.3	 	 	 	
201	2015-10-08 16:15:51	1168.6	 	 	 	
202	2015-10-08 16:16:21	1176.6	 	 	 	
203	2015-10-08 16:16:51	1180.1	 	 	 	
204	2015-10-08 16:17:21	1181.3	 	 	 	
205	2015-10-08 16:17:51	1182.0	 	 	 	
206	2015-10-08 16:18:21	1217.1	 	 	 	
207	2015-10-08 16:18:51	1239.5	 	 	 	
208	2015-10-08 16:19:21	1247.7	 	 	 	
209	2015-10-08 16:19:51	1250.8	 	 	 	
210	2015-10-08 16:20:21	1251.6	 	 	 	
211	2015-10-08 16:20:51	1252.2	 	 	 	
212	2015-10-08 16:21:21	1252.0	 	 	 	
213	2015-10-08 16:21:51	1252.1	 	 	 	
214	2015-10-08 16:22:21	1252.2	 	 	 	
215	2015-10-08 16:22:51	1254.5	 	 	 	
216	2015-10-08 16:23:21	1289.5	 	 	 	
217	2015-10-08 16:23:51	1305.0	 	 	 	
218	2015-10-08 16:24:21	1310.1	 	 	 	
219	2015-10-08 16:24:51	1311.7	 	 	 	
220	2015-10-08 16:25:21	1312.1	 	 	 	
221	2015-10-08 16:25:51	1312.1	 	 	 	
222	2015-10-08 16:26:21	1312.1	 	 	 	
223	2015-10-08 16:26:51	1312.6	 	 	 	
224	2015-10-08 16:27:21	1312.3	 	 	 	
225	2015-10-08 16:27:51	1312.5	 	 	 	
226	2015-10-08 16:28:21	1312.3	 	 	 	
227	2015-10-08 16:28:51	1312.1	 	 	 	
228	2015-10-08 16:29:21	1312.1	 	 	 	
229	2015-10-08 16:29:51	1312.8	 	 	 	
230	2015-10-08 16:30:21	1312.5	 	 	 	
231	2015-10-08 16:30:51	1312.1	 	 	 	
232	2015-10-08 16:31:21	1311.8	 	 	 	
233	2015-10-08 16:31:51	1311.7	 	 	 	
234	2015-10-08 16:32:21	1311.7	 	 	 	
235	2015-10-08 16:32:51	1312.6	 	 	 	
236	2015-10-08 16:33:21	1354.6	 	 	 	
237	2015-10-08 16:33:51	1370.7	 	 	 	
238	2015-10-08 16:34:21	1375.4	 	 	 	
239	2015-10-08 16:34:51	1376.4	 	 	 	
240	2015-10-08 16:35:21	1376.9	 	 	 	
241	2015-10-08 16:35:51	1376.7	 	 	 	
242	2015-10-08 16:36:21	1377.0	 	 	 	
243	2015-10-08 16:36:51	1377.0	 	 	 	
244	2015-10-08 16:37:21	1384.5	 	 	 	
245	2015-10-08 16:37:51	1426.4	 	 	 	
246	2015-10-08 16:38:21	1436.9	 	 	 	
247	2015-10-08 16:38:51	1439.0	 	 	 	
248	2015-10-08 16:39:21	1440.4	 	 	 	
249	2015-10-08 16:39:51	1440.4	 	 	 	
250	2015-10-08 16:40:21	1440.0	 	 	 	
251	2015-10-08 16:40:51	1440.3	 	 	 	
252	2015-10-08 16:41:21	1469.0	 	 	 	
253	2015-10-08 16:41:51	1489.8	 	 	 	
254	2015-10-08 16:42:21	1494.0	 	 	 	
255	2015-10-08 16:42:51	1495.0	 	 	 	
256	2015-10-08 16:43:21	1495.1	 	 	 	
257	2015-10-08 16:43:51	1495.3	 	 	 	
258	2015-10-08 16:44:21	1495.1	 	 	 	
259	2015-10-08 16:44:51	1543.6	 	 	 	
260	2015-10-08 16:45:21	1555.0	 	 	 	
261	2015-10-08 16:45:51	1557.6	 	 	 	
262	2015-10-08 16:46:21	1557.6	 	 	 	
263	2015-10-08 16:46:51	1557.7	 	 	 	
264	2015-10-08 16:47:21	1558.1	 	 	 	
265	2015-10-08 16:47:51	1571.9	 	 	 	
266	2015-10-08 16:48:21	1607.9	 	 	 	
267	2015-10-08 16:48:51	1612.9	 	 	 	
268	2015-10-08 16:49:21	1613.5	 	 	 	
269	2015-10-08 16:49:51	1613.3	 	 	 	
270	2015-10-08 16:50:21	1613.5	 	 	 	
271	2015-10-08 16:50:51	1612.5	 	 	 	
272	2015-10-08 16:51:21	1612.1	 	 	 	
273	2015-10-08 16:51:51	1646.2	 	 	 	
274	2015-10-08 16:52:21	1664.2	 	 	 	
275	2015-10-08 16:52:51	1665.2	 	 	 	
276	2015-10-08 16:53:21	1665.3	 	 	 	
277	2015-10-08 16:53:51	1665.7	 	 	 	
278	2015-10-08 16:54:21	1688.6	 	 	 	
279	2015-10-08 16:54:51	1717.2	 	 	 	
280	2015-10-08 16:55:21	1720.7	 	 	 	
281	2015-10-08 16:55:51	1720.6	 	 	 	
282	2015-10-08 16:56:21	1720.1	 	 	 	
283	2015-10-08 16:56:51	1722.2	 	 	 	
284	2015-10-08 16:57:21	1757.5	 	 	 	
285	2015-10-08 16:57:51	1769.5	 	 	 	
286	2015-10-08 16:58:21	1771.0	 	 	 	
287	2015-10-08 16:58:51	1770.4	 	 	 	
288	2015-10-08 16:59:21	1771.0	 	 	 	
289	2015-10-08 16:59:51	1769.0	 	 	 	
290	2015-10-08 17:00:21	1769.5	 	 	 	
291	2015-10-08 17:00:51	1769.5	 	 	 	
292	2015-10-08 17:01:21	1768.9	 	 	 	
293	2015-10-08 17:01:51	1769.0	 	 	 	
294	2015-10-08 17:02:21	1769.5	 	 	 	
295	2015-10-08 17:02:51	1768.9	 	 	 	
296	2015-10-08 17:03:21	1769.4	 	 	 	
297	2015-10-08 17:03:51	1769.5	 	 	 	
298	2015-10-08 17:04:21	1769.5	 	 	 	
299	2015-10-08 17:04:51	1768.9	 	 	 	
... 95919 more lines ...
Attachment 3: 2015-10-06.4ch
°C	Date Time	T1 - C	T2 - C	T3 - C	T4 - K	
1	2015-10-06 11:06:14	21.7	 	 	 	
2	2015-10-06 11:06:44	21.3	 	 	 	
3	2015-10-06 11:07:14	21.6	 	 	 	
4	2015-10-06 11:07:44	22.1	 	 	 	
5	2015-10-06 11:08:14	21.7	 	 	 	
6	2015-10-06 11:08:44	21.9	 	 	 	
7	2015-10-06 11:09:14	22.3	 	 	 	
8	2015-10-06 11:09:44	22.7	 	 	 	
9	2015-10-06 11:10:14	21.8	 	 	 	
10	2015-10-06 11:10:44	22.5	 	 	 	
11	2015-10-06 11:11:14	23.5	 	 	 	
12	2015-10-06 11:11:44	22.7	 	 	 	
13	2015-10-06 11:12:14	22.9	 	 	 	
14	2015-10-06 11:12:44	23.3	 	 	 	
15	2015-10-06 11:13:14	22.5	 	 	 	
16	2015-10-06 11:13:44	21.7	 	 	 	
17	2015-10-06 11:14:14	22.2	 	 	 	
18	2015-10-06 11:14:44	22.2	 	 	 	
19	2015-10-06 11:15:14	22.1	 	 	 	
20	2015-10-06 11:15:44	22.5	 	 	 	
21	2015-10-06 11:16:14	22.5	 	 	 	
22	2015-10-06 11:16:44	22.7	 	 	 	
23	2015-10-06 11:17:14	22.7	 	 	 	
24	2015-10-06 11:17:44	22.1	 	 	 	
25	2015-10-06 11:18:14	22.5	 	 	 	
26	2015-10-06 11:18:44	21.9	 	 	 	
27	2015-10-06 11:19:14	22.2	 	 	 	
28	2015-10-06 11:19:44	22.7	 	 	 	
29	2015-10-06 11:20:14	22.7	 	 	 	
30	2015-10-06 11:20:44	22.5	 	 	 	
31	2015-10-06 11:21:14	22.5	 	 	 	
32	2015-10-06 11:21:44	22.5	 	 	 	
33	2015-10-06 11:22:14	22.5	 	 	 	
34	2015-10-06 11:22:44	22.2	 	 	 	
35	2015-10-06 11:23:14	21.3	 	 	 	
36	2015-10-06 11:23:44	21.3	 	 	 	
37	2015-10-06 11:24:14	21.8	 	 	 	
38	2015-10-06 11:24:44	21.4	 	 	 	
39	2015-10-06 11:25:14	21.5	 	 	 	
40	2015-10-06 11:25:44	21.6	 	 	 	
41	2015-10-06 11:26:14	21.6	 	 	 	
42	2015-10-06 11:26:44	21.6	 	 	 	
43	2015-10-06 11:27:14	21.7	 	 	 	
44	2015-10-06 11:27:44	21.9	 	 	 	
45	2015-10-06 11:28:14	21.8	 	 	 	
46	2015-10-06 11:28:44	21.8	 	 	 	
47	2015-10-06 11:29:14	21.9	 	 	 	
48	2015-10-06 11:29:44	21.9	 	 	 	
49	2015-10-06 11:30:14	21.9	 	 	 	
50	2015-10-06 11:30:44	22.7	 	 	 	
51	2015-10-06 11:31:14	22.2	 	 	 	
52	2015-10-06 11:31:44	22.1	 	 	 	
53	2015-10-06 11:32:14	22.2	 	 	 	
54	2015-10-06 11:32:44	21.8	 	 	 	
55	2015-10-06 11:33:14	22.2	 	 	 	
56	2015-10-06 11:33:44	22.3	 	 	 	
57	2015-10-06 11:34:14	21.9	 	 	 	
58	2015-10-06 11:34:44	21.9	 	 	 	
59	2015-10-06 11:35:14	21.4	 	 	 	
60	2015-10-06 11:35:44	21.9	 	 	 	
61	2015-10-06 11:36:14	22.5	 	 	 	
62	2015-10-06 11:36:44	21.6	 	 	 	
63	2015-10-06 11:37:14	22.2	 	 	 	
64	2015-10-06 11:37:44	21.6	 	 	 	
65	2015-10-06 11:38:14	21.1	 	 	 	
66	2015-10-06 11:38:44	21.2	 	 	 	
67	2015-10-06 11:39:14	21.7	 	 	 	
68	2015-10-06 11:39:44	21.2	 	 	 	
69	2015-10-06 11:40:14	22.2	 	 	 	
70	2015-10-06 11:40:44	21.8	 	 	 	
71	2015-10-06 11:41:14	21.9	 	 	 	
72	2015-10-06 11:41:44	21.3	 	 	 	
73	2015-10-06 11:42:14	21.4	 	 	 	
74	2015-10-06 11:42:44	21.9	 	 	 	
75	2015-10-06 11:43:14	21.4	 	 	 	
76	2015-10-06 11:43:44	21.5	 	 	 	
77	2015-10-06 11:44:14	21.6	 	 	 	
78	2015-10-06 11:44:44	21.6	 	 	 	
79	2015-10-06 11:45:14	21.6	 	 	 	
80	2015-10-06 11:45:44	22.1	 	 	 	
81	2015-10-06 11:46:14	21.6	 	 	 	
82	2015-10-06 11:46:44	21.6	 	 	 	
83	2015-10-06 11:47:14	21.6	 	 	 	
84	2015-10-06 11:47:44	21.6	 	 	 	
85	2015-10-06 11:48:14	21.6	 	 	 	
86	2015-10-06 11:48:44	21.7	 	 	 	
87	2015-10-06 11:49:14	21.7	 	 	 	
88	2015-10-06 11:49:44	21.7	 	 	 	
89	2015-10-06 11:50:14	21.7	 	 	 	
90	2015-10-06 11:50:44	21.8	 	 	 	
91	2015-10-06 11:51:14	21.7	 	 	 	
92	2015-10-06 11:51:44	21.8	 	 	 	
93	2015-10-06 11:52:14	21.9	 	 	 	
94	2015-10-06 11:52:44	21.8	 	 	 	
95	2015-10-06 11:53:14	21.9	 	 	 	
96	2015-10-06 11:53:44	21.8	 	 	 	
97	2015-10-06 11:54:14	21.9	 	 	 	
98	2015-10-06 11:54:44	21.9	 	 	 	
99	2015-10-06 11:55:14	21.9	 	 	 	
100	2015-10-06 11:55:44	21.9	 	 	 	
101	2015-10-06 11:56:14	21.9	 	 	 	
102	2015-10-06 11:56:44	21.9	 	 	 	
103	2015-10-06 11:57:14	21.9	 	 	 	
104	2015-10-06 11:57:44	21.9	 	 	 	
105	2015-10-06 11:58:14	21.9	 	 	 	
106	2015-10-06 11:58:44	22.1	 	 	 	
107	2015-10-06 11:59:14	21.9	 	 	 	
108	2015-10-06 11:59:44	21.9	 	 	 	
109	2015-10-06 12:00:14	22.1	 	 	 	
110	2015-10-06 12:00:44	21.9	 	 	 	
111	2015-10-06 12:01:14	21.9	 	 	 	
112	2015-10-06 12:01:44	21.9	 	 	 	
113	2015-10-06 12:02:14	22.1	 	 	 	
114	2015-10-06 12:02:44	22.2	 	 	 	
115	2015-10-06 12:03:14	22.1	 	 	 	
116	2015-10-06 12:03:44	22.1	 	 	 	
117	2015-10-06 12:04:14	22.1	 	 	 	
118	2015-10-06 12:04:44	22.1	 	 	 	
119	2015-10-06 12:05:14	22.1	 	 	 	
120	2015-10-06 12:05:44	22.1	 	 	 	
121	2015-10-06 12:06:14	22.1	 	 	 	
122	2015-10-06 12:06:44	22.2	 	 	 	
123	2015-10-06 12:07:14	22.2	 	 	 	
124	2015-10-06 12:07:44	22.2	 	 	 	
125	2015-10-06 12:08:14	22.2	 	 	 	
126	2015-10-06 12:08:44	22.2	 	 	 	
127	2015-10-06 12:09:14	22.2	 	 	 	
128	2015-10-06 12:09:44	22.2	 	 	 	
129	2015-10-06 12:10:14	22.2	 	 	 	
130	2015-10-06 12:10:44	22.2	 	 	 	
131	2015-10-06 12:11:14	22.2	 	 	 	
132	2015-10-06 12:11:44	22.2	 	 	 	
133	2015-10-06 12:12:14	22.2	 	 	 	
134	2015-10-06 12:12:44	22.2	 	 	 	
135	2015-10-06 12:13:14	22.2	 	 	 	
136	2015-10-06 12:13:44	22.2	 	 	 	
137	2015-10-06 12:14:14	22.2	 	 	 	
138	2015-10-06 12:14:44	22.2	 	 	 	
139	2015-10-06 12:15:14	22.2	 	 	 	
140	2015-10-06 12:15:44	22.2	 	 	 	
141	2015-10-06 12:16:14	22.2	 	 	 	
142	2015-10-06 12:16:44	22.2	 	 	 	
143	2015-10-06 12:17:14	22.2	 	 	 	
144	2015-10-06 12:17:44	22.2	 	 	 	
145	2015-10-06 12:18:14	22.2	 	 	 	
146	2015-10-06 12:18:44	22.2	 	 	 	
147	2015-10-06 12:19:14	22.3	 	 	 	
148	2015-10-06 12:19:44	22.3	 	 	 	
149	2015-10-06 12:20:14	22.4	 	 	 	
150	2015-10-06 12:20:44	22.3	 	 	 	
151	2015-10-06 12:21:14	22.3	 	 	 	
152	2015-10-06 12:21:44	22.4	 	 	 	
153	2015-10-06 12:22:14	22.3	 	 	 	
154	2015-10-06 12:22:44	22.3	 	 	 	
155	2015-10-06 12:23:14	22.3	 	 	 	
156	2015-10-06 12:23:44	22.4	 	 	 	
157	2015-10-06 12:24:14	22.3	 	 	 	
158	2015-10-06 12:24:44	22.4	 	 	 	
159	2015-10-06 12:25:14	22.5	 	 	 	
160	2015-10-06 12:25:44	22.4	 	 	 	
161	2015-10-06 12:26:14	22.4	 	 	 	
162	2015-10-06 12:26:44	22.4	 	 	 	
163	2015-10-06 12:27:14	22.4	 	 	 	
164	2015-10-06 12:27:44	22.4	 	 	 	
165	2015-10-06 12:28:14	22.5	 	 	 	
166	2015-10-06 12:28:44	22.4	 	 	 	
167	2015-10-06 12:29:14	22.4	 	 	 	
168	2015-10-06 12:29:44	22.5	 	 	 	
169	2015-10-06 12:30:14	22.5	 	 	 	
170	2015-10-06 12:30:44	22.5	 	 	 	
171	2015-10-06 12:31:14	22.5	 	 	 	
172	2015-10-06 12:31:44	22.4	 	 	 	
173	2015-10-06 12:32:14	22.5	 	 	 	
174	2015-10-06 12:32:44	22.5	 	 	 	
175	2015-10-06 12:33:14	22.5	 	 	 	
176	2015-10-06 12:33:44	22.5	 	 	 	
177	2015-10-06 12:34:14	22.5	 	 	 	
178	2015-10-06 12:34:44	22.5	 	 	 	
179	2015-10-06 12:35:14	22.5	 	 	 	
180	2015-10-06 12:35:44	22.5	 	 	 	
181	2015-10-06 12:36:14	22.5	 	 	 	
182	2015-10-06 12:36:44	22.5	 	 	 	
183	2015-10-06 12:37:14	22.5	 	 	 	
184	2015-10-06 12:37:44	22.5	 	 	 	
185	2015-10-06 12:38:14	22.5	 	 	 	
186	2015-10-06 12:38:44	22.5	 	 	 	
187	2015-10-06 12:39:14	22.5	 	 	 	
188	2015-10-06 12:39:44	22.5	 	 	 	
189	2015-10-06 12:40:14	22.5	 	 	 	
190	2015-10-06 12:40:44	22.6	 	 	 	
191	2015-10-06 12:41:14	22.5	 	 	 	
192	2015-10-06 12:41:44	22.5	 	 	 	
193	2015-10-06 12:42:14	22.5	 	 	 	
194	2015-10-06 12:42:44	22.5	 	 	 	
195	2015-10-06 12:43:14	22.6	 	 	 	
196	2015-10-06 12:43:44	22.5	 	 	 	
197	2015-10-06 12:44:14	22.6	 	 	 	
198	2015-10-06 12:44:44	22.5	 	 	 	
199	2015-10-06 12:45:14	22.5	 	 	 	
200	2015-10-06 12:45:44	22.5	 	 	 	
201	2015-10-06 12:46:14	22.5	 	 	 	
202	2015-10-06 12:46:44	22.7	 	 	 	
203	2015-10-06 12:47:14	22.6	 	 	 	
204	2015-10-06 12:47:44	22.6	 	 	 	
205	2015-10-06 12:48:14	22.7	 	 	 	
206	2015-10-06 12:48:44	22.7	 	 	 	
207	2015-10-06 12:49:14	22.6	 	 	 	
208	2015-10-06 12:49:44	22.6	 	 	 	
209	2015-10-06 12:50:14	22.7	 	 	 	
210	2015-10-06 12:50:44	22.6	 	 	 	
211	2015-10-06 12:51:14	22.6	 	 	 	
212	2015-10-06 12:51:44	22.6	 	 	 	
213	2015-10-06 12:52:14	22.6	 	 	 	
214	2015-10-06 12:52:44	22.7	 	 	 	
215	2015-10-06 12:53:14	22.7	 	 	 	
216	2015-10-06 12:53:44	22.6	 	 	 	
217	2015-10-06 12:54:14	22.7	 	 	 	
218	2015-10-06 12:54:44	22.6	 	 	 	
219	2015-10-06 12:55:14	22.6	 	 	 	
220	2015-10-06 12:55:44	22.5	 	 	 	
221	2015-10-06 12:56:14	22.6	 	 	 	
222	2015-10-06 12:56:44	22.6	 	 	 	
223	2015-10-06 12:57:14	22.5	 	 	 	
224	2015-10-06 12:57:44	22.6	 	 	 	
225	2015-10-06 12:58:14	22.6	 	 	 	
226	2015-10-06 12:58:44	22.6	 	 	 	
227	2015-10-06 12:59:14	22.7	 	 	 	
228	2015-10-06 12:59:44	22.6	 	 	 	
229	2015-10-06 13:00:14	22.6	 	 	 	
230	2015-10-06 13:00:44	22.6	 	 	 	
231	2015-10-06 13:01:14	22.6	 	 	 	
232	2015-10-06 13:01:44	22.6	 	 	 	
233	2015-10-06 13:02:14	22.6	 	 	 	
234	2015-10-06 13:02:44	22.6	 	 	 	
235	2015-10-06 13:03:14	22.6	 	 	 	
236	2015-10-06 13:03:44	22.7	 	 	 	
237	2015-10-06 13:04:14	22.7	 	 	 	
238	2015-10-06 13:04:44	22.7	 	 	 	
239	2015-10-06 13:05:14	22.6	 	 	 	
240	2015-10-06 13:05:44	22.6	 	 	 	
241	2015-10-06 13:06:14	22.7	 	 	 	
242	2015-10-06 13:06:44	22.7	 	 	 	
243	2015-10-06 13:07:14	22.7	 	 	 	
244	2015-10-06 13:07:44	22.7	 	 	 	
245	2015-10-06 13:08:14	22.7	 	 	 	
246	2015-10-06 13:08:44	22.7	 	 	 	
247	2015-10-06 13:09:14	22.7	 	 	 	
248	2015-10-06 13:09:44	22.7	 	 	 	
249	2015-10-06 13:10:14	22.7	 	 	 	
250	2015-10-06 13:10:44	22.7	 	 	 	
251	2015-10-06 13:11:14	22.7	 	 	 	
252	2015-10-06 13:11:44	22.7	 	 	 	
253	2015-10-06 13:12:14	22.7	 	 	 	
254	2015-10-06 13:12:44	22.7	 	 	 	
255	2015-10-06 13:13:14	22.7	 	 	 	
256	2015-10-06 13:13:44	22.7	 	 	 	
257	2015-10-06 13:14:14	22.7	 	 	 	
258	2015-10-06 13:14:44	22.7	 	 	 	
259	2015-10-06 13:15:14	22.7	 	 	 	
260	2015-10-06 13:15:44	22.7	 	 	 	
261	2015-10-06 13:16:14	22.7	 	 	 	
262	2015-10-06 13:16:44	22.6	 	 	 	
263	2015-10-06 13:17:14	22.7	 	 	 	
264	2015-10-06 13:17:44	22.7	 	 	 	
265	2015-10-06 13:18:14	22.7	 	 	 	
266	2015-10-06 13:18:44	22.6	 	 	 	
267	2015-10-06 13:19:14	22.7	 	 	 	
268	2015-10-06 13:19:44	22.7	 	 	 	
269	2015-10-06 13:20:14	22.7	 	 	 	
270	2015-10-06 13:20:44	23.2	 	 	 	
271	2015-10-06 13:21:14	22.7	 	 	 	
272	2015-10-06 13:21:44	22.7	 	 	 	
273	2015-10-06 13:22:14	22.7	 	 	 	
274	2015-10-06 13:22:44	23.2	 	 	 	
275	2015-10-06 13:23:14	22.7	 	 	 	
276	2015-10-06 13:23:44	22.7	 	 	 	
277	2015-10-06 13:24:14	22.7	 	 	 	
278	2015-10-06 13:24:44	22.7	 	 	 	
279	2015-10-06 13:25:14	22.7	 	 	 	
280	2015-10-06 13:25:44	22.7	 	 	 	
281	2015-10-06 13:26:14	22.7	 	 	 	
282	2015-10-06 13:26:44	22.7	 	 	 	
283	2015-10-06 13:27:14	22.7	 	 	 	
284	2015-10-06 13:27:44	22.7	 	 	 	
285	2015-10-06 13:28:14	22.7	 	 	 	
286	2015-10-06 13:28:44	22.7	 	 	 	
287	2015-10-06 13:29:14	22.7	 	 	 	
288	2015-10-06 13:29:44	22.7	 	 	 	
289	2015-10-06 13:30:14	22.7	 	 	 	
290	2015-10-06 13:30:44	22.6	 	 	 	
291	2015-10-06 13:31:14	22.6	 	 	 	
292	2015-10-06 13:31:44	22.6	 	 	 	
293	2015-10-06 13:32:14	23.1	 	 	 	
294	2015-10-06 13:32:44	22.6	 	 	 	
295	2015-10-06 13:33:14	22.6	 	 	 	
296	2015-10-06 13:33:44	22.6	 	 	 	
297	2015-10-06 13:34:14	22.6	 	 	 	
298	2015-10-06 13:34:44	23.2	 	 	 	
299	2015-10-06 13:35:14	22.5	 	 	 	
... 5876 more lines ...
Attachment 4: 2015-10-06_Small_Centre_Block.xlsx
  270   Monday, January 11, 2016, 15:47 Laura LambertSiC41077-3LP-SISDoneConditioning SiC test target for test stand measurements

Monday, January 11, 2016, 15:25

Small center block LP-SIS container previously sintered with TaC coating

3x Ta heat shields

Container is loaded with 260 discs of SiC with a total length of 8.5cm

Loaded into evap2 and pumped down this morning

HS water valves opened up, no leaks detected

Tapped up manually to 30A

Vacuum stable

IGP1 = 1e-6 Torr

Initiating auto-ramp at 0.2A/min and 0.1A/min for the TGHT and TBHT respectively

Final setpoints: 400A for the TGHT and 230A for the TBHT

Final desired temperature: 1700C


 Tuesday, January 12, 2016, 12:55

TGHT = 138A, TBHT = 80A

IGP1 =  7.7e-6 Torr.


Vacuum stable and coming down.

Increasing ramping rate to 0.5A/min and 0.2A/min for the TGHT and TBHT respectively.


 Wednesday, January 13, 2016, 09:32

TGHT reached 400A ~11.5hrs ago

TBHT reached 230A ~8hrs ago

OMEGA = 1400C

IGP1 = 1.75e-7 Torr

Tapping up manually to 1700C...


 Wednesday, January 13, 2016, 10:14

Previous temperature measurement of OMEGA = 1400C is wrong due to window port being coated with SiC [see attached photos]

Will remain at TGHT = 420A for 4hrs then cool down this afternoon


Wednesday, January 13, 2016, 14:29

Remained at max setpoints for 4.5hrs.

Initiating auto-cool at 2.0A/min and 1.0A/min for the TGHT and TBHT respectively.


 Thursday, January 14, 2016, 09:47

Vented in air and opened up this morning

Target looks good

Will insert and expand endcaps this afternoon


 

Attachment 1: Photo_2016-0-13.jpg
Photo_2016-0-13.jpg
Attachment 2: Photo_2016-01-13_#2.jpg
Photo_2016-01-13_#2.jpg
Attachment 3: Photo_2016-01-13_#3.jpg
Photo_2016-01-13_#3.jpg
Attachment 4: Photo_2016-01-13_#4.jpg
Photo_2016-01-13_#4.jpg
  273   Friday, February 12, 2016, 11:09 Laura LambertEmpty41077-1LP-SISDonesintering TaC coated small center block container

 Friday, February 12, 2016, 11:01

Container coated with TaC with new method - ends tapped off,  liquid poured/swirled around container, then poured back out - 'flushing technique'.

Container wrapped with 3 x Ta heat shields, incl center block - new method of each layer being one long single Ta foil covering entire container incl center block

IGP1 = 6.3e-7Torr

HS water valves opened, no leaks detected.

Tapped up manually to 35A.

Waited ~20min.

Vacuum stable, IGP1 = 1.3e-6 Torr.

Initiated auto-ramp at 0.2A/min and 0.1A/min for the TGHT and TBHT respectively with final setpoints of 400A and 230A.


 Monday, February 15, 2016, 08:56

IGP1 = 3.9e-7 Torr.

TBHT = 230A, 4.0V

TGHT = 400A, 6.58V

OMEGA = 1766C

Tapping up manually to 1900C...


 Monday, February 15, 2016, 10:18

IGP1 = 5e-7 Torr

TGHT = 450A, 7.55V

OMEGA = 1913C


 Monday, February 15, 2016, 14:30

IGP1 = 4.9e-7 Torr.

TGHT = 450A, 7.55V.

Material remained at ~1910C for ~4hrs.

Initiating auto-cool at 2.0A/min and 1.0A/min for the TGHT and TBHT respectively.


 Tuesday, February 16, 2016, 14:17

vented in air and opened up this morning.

target container sintered nicely.


 

  277   Wednesday, February 24, 2016, 17:40 John WongNb LP-SISDoneBob/Tobi Nb samples

Thursday, February 25, 2016, 18:04 Wednesday, February 24, 2016, 17:29

Heat treatment of 2 bullet-shaped Nb samples in a LP-SIS container; it is just that we don't have a target boat. 

Container is wrapped with 12" in total length.  A strip of Nb foil is used as an underlayer for the samples to sit on top. 

Chamber has been pumped down and current vacuum at 1e-6 Torr. 

HS water check - good.

Target temperature 1400C

Ramping up to 250A; this should get to ~1300C.  Will let it heat for over night and step up to 1400C tomorrow. 


 Wednesday, February 24, 2016, 17:59

I'm stopping the heating process because the camera is not working.  I can't monitor the temperature.  Tried to fit the camera; don't now why it fails to connect to the server. 

Will look in to it tomorrow...


 Thursday, February 25, 2016, 08:17

Initiated auto ramping TGHT at 2A/min.


Thursday, February 25, 2016, 17:33  

Finally the camera is fixed.  

Current temperature is at 1126C, TGHT at 250A.  Will step up to reach 1400C. 


Thursday, February 25, 2016, 18:04  

TGHT stepped up to 335A to reach 1400C.  (see attachment) 

Will heat this for few hours and cool it at around 11pm tonight.


 Thursday, February 25, 2016, 22:19

Samples have been heated over 4.5 hours. 

Initiated cool down at 1A/min.


 Friday, February 26, 2016, 15:43

done! all good.


 

Attachment 1: Screen_Shot_2016-02-25_at_6.04.20_PM.png
Screen_Shot_2016-02-25_at_6.04.20_PM.png
  279   Sunday, March 13, 2016, 14:58 Aurelia LaxdalTa41790-1LP-SISIn progressTa#47 LP-SIS (with new center block) conditioning

Started Ta#47 heating at Evap2 using the autoramp program. The TGHT current is setup to go to 300A (for now) and the TBHT to 230A (final value).

  280   Monday, March 14, 2016, 09:22 Laura LambertTa41790-1LP-SISIn progressTa#47 LP-SIS (with new center block) conditioning

Aurelia Laxdal wrote:

Started Ta#47 heating at Evap2 using the autoramp program. The TGHT current is setup to go to 300A (for now) and the TBHT to 230A (final value).

 Monday, March 14, 2016, 09:20

One thermocouple was not welded on correctly and is not working. 

Cooling evap2 back down, to remove target and re-weld thermocouple.

Once thermocouple replaced, will ramp back up and continue with tests/ conditioning of target.


 

  281   Monday, March 14, 2016, 13:12 Anders MjosTa41790-1LP-SISIn progressTa#47 LP-SIS (with new center block) conditioning

Laura Lambert wrote:

Aurelia Laxdal wrote:

Started Ta#47 heating at Evap2 using the autoramp program. The TGHT current is setup to go to 300A (for now) and the TBHT to 230A (final value).

 Monday, March 14, 2016, 09:20

One thermocouple was not welded on correctly and is not working. 

Cooling evap2 back down, to remove target and re-weld thermocouple.

Once thermocouple replaced, will ramp back up and continue with tests/ conditioning of target.


 

 Monday, March 14, 2016, 13:12: Stopped turbo pump

  282   Monday, March 14, 2016, 19:22 Aurelia LaxdalTa41790-1LP-SISIn progressTa#47 LP-SIS (with new center block) conditioning

Anders Mjos wrote:

Laura Lambert wrote:

Aurelia Laxdal wrote:

Started Ta#47 heating at Evap2 using the autoramp program. The TGHT current is setup to go to 300A (for now) and the TBHT to 230A (final value).

 Monday, March 14, 2016, 09:20

One thermocouple was not welded on correctly and is not working. 

Cooling evap2 back down, to remove target and re-weld thermocouple.

Once thermocouple replaced, will ramp back up and continue with tests/ conditioning of target.


 

 Monday, March 14, 2016, 13:12: Stopped turbo pump

 After EVAP2 was vented the thermocouple problem was fixed: it was the wire at on connector inside the box not making proper contact. The lid of the heat shield was also replaced since the one initially mounted had no holes, so one couldn't see the center block. Restarted pumping down at 4:30pm. Just started again the autoramp program: TGHT set to 300A (for the time being) and TBHT to 230A. Temperature measurements will resume tomorrow morning.

  283   Friday, March 18, 2016, 17:13 Aurelia LaxdalTa41790-1LP-SISIn progressTa#47 LP-SIS (with new center block) conditioning

Aurelia Laxdal wrote:

Anders Mjos wrote:

Laura Lambert wrote:

Aurelia Laxdal wrote:

Started Ta#47 heating at Evap2 using the autoramp program. The TGHT current is setup to go to 300A (for now) and the TBHT to 230A (final value).

 Monday, March 14, 2016, 09:20

One thermocouple was not welded on correctly and is not working. 

Cooling evap2 back down, to remove target and re-weld thermocouple.

Once thermocouple replaced, will ramp back up and continue with tests/ conditioning of target.


 

 Monday, March 14, 2016, 13:12: Stopped turbo pump

 After EVAP2 was vented the thermocouple problem was fixed: it was the wire at on connector inside the box not making proper contact. The lid of the heat shield was also replaced since the one initially mounted had no holes, so one couldn't see the center block. Restarted pumping down at 4:30pm. Just started again the autoramp program: TGHT set to 300A (for the time being) and TBHT to 230A. Temperature measurements will resume tomorrow morning.

 Raised the TGHT at 620A. The TBHT was already set at 230A. The temperature is at 1902 deg C. As per John's instructions the target will condition at these setpoints over the week-end.

  284   Monday, March 21, 2016, 18:57 Aurelia LaxdalTa41790-1LP-SISIn progressTa#47 LP-SIS (with new center block) conditioning

Aurelia Laxdal wrote:

Aurelia Laxdal wrote:

Anders Mjos wrote:

Laura Lambert wrote:

Aurelia Laxdal wrote:

Started Ta#47 heating at Evap2 using the autoramp program. The TGHT current is setup to go to 300A (for now) and the TBHT to 230A (final value).

 Monday, March 14, 2016, 09:20

One thermocouple was not welded on correctly and is not working. 

Cooling evap2 back down, to remove target and re-weld thermocouple.

Once thermocouple replaced, will ramp back up and continue with tests/ conditioning of target.


 

 Monday, March 14, 2016, 13:12: Stopped turbo pump

 After EVAP2 was vented the thermocouple problem was fixed: it was the wire at on connector inside the box not making proper contact. The lid of the heat shield was also replaced since the one initially mounted had no holes, so one couldn't see the center block. Restarted pumping down at 4:30pm. Just started again the autoramp program: TGHT set to 300A (for the time being) and TBHT to 230A. Temperature measurements will resume tomorrow morning.

 Raised the TGHT at 620A. The TBHT was already set at 230A. The temperature is at 1902 deg C. As per John's instructions the target will condition at these setpoints over the week-end.

Today, lowered the TGHT from 620A (at 1894 deg C) to 300A.  TBHT is still at 230A.

  285   Tuesday, March 22, 2016, 11:31 Aurelia LaxdalTa41790-1LP-SISIn progressTa#47 LP-SIS (with new center block) conditioning

Aurelia Laxdal wrote:

Aurelia Laxdal wrote:

Aurelia Laxdal wrote:

Anders Mjos wrote:

Laura Lambert wrote:

Aurelia Laxdal wrote:

Started Ta#47 heating at Evap2 using the autoramp program. The TGHT current is setup to go to 300A (for now) and the TBHT to 230A (final value).

 Monday, March 14, 2016, 09:20

One thermocouple was not welded on correctly and is not working. 

Cooling evap2 back down, to remove target and re-weld thermocouple.

Once thermocouple replaced, will ramp back up and continue with tests/ conditioning of target.


 

 Monday, March 14, 2016, 13:12: Stopped turbo pump

 After EVAP2 was vented the thermocouple problem was fixed: it was the wire at on connector inside the box not making proper contact. The lid of the heat shield was also replaced since the one initially mounted had no holes, so one couldn't see the center block. Restarted pumping down at 4:30pm. Just started again the autoramp program: TGHT set to 300A (for the time being) and TBHT to 230A. Temperature measurements will resume tomorrow morning.

 Raised the TGHT at 620A. The TBHT was already set at 230A. The temperature is at 1902 deg C. As per John's instructions the target will condition at these setpoints over the week-end.

Today, lowered the TGHT from 620A (at 1894 deg C) to 300A.  TBHT is still at 230A.

 Cooling down the target.

 

  286   Thursday, March 24, 2016, 11:32 John WongTa41790-1LP-SISDoneTa#47 LP-SIS (with new center block) conditioning

Aurelia Laxdal wrote:

Aurelia Laxdal wrote:

Aurelia Laxdal wrote:

Aurelia Laxdal wrote:

Anders Mjos wrote:

Laura Lambert wrote:

Aurelia Laxdal wrote:

Started Ta#47 heating at Evap2 using the autoramp program. The TGHT current is setup to go to 300A (for now) and the TBHT to 230A (final value).

 Monday, March 14, 2016, 09:20

One thermocouple was not welded on correctly and is not working. 

Cooling evap2 back down, to remove target and re-weld thermocouple.

Once thermocouple replaced, will ramp back up and continue with tests/ conditioning of target.


 

 Monday, March 14, 2016, 13:12: Stopped turbo pump

 After EVAP2 was vented the thermocouple problem was fixed: it was the wire at on connector inside the box not making proper contact. The lid of the heat shield was also replaced since the one initially mounted had no holes, so one couldn't see the center block. Restarted pumping down at 4:30pm. Just started again the autoramp program: TGHT set to 300A (for the time being) and TBHT to 230A. Temperature measurements will resume tomorrow morning.

 Raised the TGHT at 620A. The TBHT was already set at 230A. The temperature is at 1902 deg C. As per John's instructions the target will condition at these setpoints over the week-end.

Today, lowered the TGHT from 620A (at 1894 deg C) to 300A.  TBHT is still at 230A.

 Cooling down the target.

 

Thursday, March 24, 2016, 11:29 

Target was removed from the Evaporator.  

The thermocouples broke off.  

Need to insert Re foil prior to installing


 

 

  290   Tuesday, April 19, 2016, 09:28 Anders MjosTa41790-1LP-SISDoneTa#47 LP-SIS (with new center block) conditioning

John Wong wrote:

Aurelia Laxdal wrote:

Aurelia Laxdal wrote:

Aurelia Laxdal wrote:

Aurelia Laxdal wrote:

Anders Mjos wrote:

Laura Lambert wrote:

Aurelia Laxdal wrote:

Started Ta#47 heating at Evap2 using the autoramp program. The TGHT current is setup to go to 300A (for now) and the TBHT to 230A (final value).

 Monday, March 14, 2016, 09:20

One thermocouple was not welded on correctly and is not working. 

Cooling evap2 back down, to remove target and re-weld thermocouple.

Once thermocouple replaced, will ramp back up and continue with tests/ conditioning of target.


 

 Monday, March 14, 2016, 13:12: Stopped turbo pump

 After EVAP2 was vented the thermocouple problem was fixed: it was the wire at on connector inside the box not making proper contact. The lid of the heat shield was also replaced since the one initially mounted had no holes, so one couldn't see the center block. Restarted pumping down at 4:30pm. Just started again the autoramp program: TGHT set to 300A (for the time being) and TBHT to 230A. Temperature measurements will resume tomorrow morning.

 Raised the TGHT at 620A. The TBHT was already set at 230A. The temperature is at 1902 deg C. As per John's instructions the target will condition at these setpoints over the week-end.

Today, lowered the TGHT from 620A (at 1894 deg C) to 300A.  TBHT is still at 230A.

 Cooling down the target.

 

Thursday, March 24, 2016, 11:29 

Target was removed from the Evaporator.  

The thermocouples broke off.  

Need to insert Re foil prior to installing


 

 

 Tuesday, April 19, 2016, 09:28: Uploaded thermoucouple data.

Attachment 1: 2016-03-11_Evaporator_2.4ch
°C	Date Time	T1 - C	T2 - C	T3 - C	T4 - C	
1	2016/03/11 14:25:25	21.1	24.7	 	 	
2	2016/03/11 14:25:55	25.0	24.8	 	 	
3	2016/03/11 14:26:25	21.6	- - -	 	 	
4	2016/03/11 14:26:55	21.0	- - -	 	 	
5	2016/03/11 14:27:25	21.3	- - -	 	 	
6	2016/03/11 14:27:55	21.1	- - -	 	 	
7	2016/03/11 14:28:25	21.1	- - -	 	 	
8	2016/03/11 14:28:55	21.0	- - -	 	 	
9	2016/03/11 14:29:25	21.1	- - -	 	 	
10	2016/03/11 14:29:55	21.1	- - -	 	 	
11	2016/03/11 14:30:25	21.1	- - -	 	 	
12	2016/03/11 14:30:55	21.1	- - -	 	 	
13	2016/03/11 14:31:25	21.1	- - -	 	 	
14	2016/03/11 14:31:55	21.1	- - -	 	 	
15	2016/03/11 14:32:25	21.1	- - -	 	 	
16	2016/03/11 14:32:55	21.1	- - -	 	 	
17	2016/03/11 14:33:25	20.0	- - -	 	 	
18	2016/03/11 14:33:55	20.2	- - -	 	 	
19	2016/03/11 14:34:25	20.8	- - -	 	 	
20	2016/03/11 14:34:55	20.8	- - -	 	 	
21	2016/03/11 14:35:25	21.1	- - -	 	 	
22	2016/03/11 14:35:55	19.7	- - -	 	 	
23	2016/03/11 14:36:25	20.8	- - -	 	 	
24	2016/03/11 14:36:55	20.7	- - -	 	 	
25	2016/03/11 14:37:25	21.3	- - -	 	 	
26	2016/03/11 14:37:55	21.1	- - -	 	 	
27	2016/03/11 14:38:25	21.2	- - -	 	 	
28	2016/03/11 14:38:55	20.2	- - -	 	 	
29	2016/03/11 14:39:25	20.7	- - -	 	 	
30	2016/03/11 14:39:55	20.8	- - -	 	 	
31	2016/03/11 14:40:25	21.3	- - -	 	 	
32	2016/03/11 14:40:55	19.7	- - -	 	 	
33	2016/03/11 14:41:25	20.8	- - -	 	 	
34	2016/03/11 14:41:55	20.8	- - -	 	 	
35	2016/03/11 14:42:25	20.8	- - -	 	 	
36	2016/03/11 14:42:55	20.8	- - -	 	 	
37	2016/03/11 14:43:25	20.8	- - -	 	 	
38	2016/03/11 14:43:55	20.8	- - -	 	 	
39	2016/03/11 14:44:25	20.8	- - -	 	 	
40	2016/03/11 14:44:55	20.2	- - -	 	 	
41	2016/03/11 14:45:25	20.8	- - -	 	 	
42	2016/03/11 14:45:55	21.3	- - -	 	 	
43	2016/03/11 14:46:25	20.8	- - -	 	 	
44	2016/03/11 14:46:55	20.4	- - -	 	 	
45	2016/03/11 14:47:25	20.8	- - -	 	 	
46	2016/03/11 14:47:55	20.8	- - -	 	 	
47	2016/03/11 14:48:25	20.8	- - -	 	 	
48	2016/03/11 14:48:55	20.3	- - -	 	 	
49	2016/03/11 14:49:25	20.4	- - -	 	 	
50	2016/03/11 14:49:55	19.9	- - -	 	 	
51	2016/03/11 14:50:25	20.4	- - -	 	 	
52	2016/03/11 14:50:55	20.8	- - -	 	 	
53	2016/03/11 14:51:25	20.5	- - -	 	 	
54	2016/03/11 14:51:55	20.4	- - -	 	 	
55	2016/03/11 14:52:25	20.5	- - -	 	 	
56	2016/03/11 14:52:55	20.4	- - -	 	 	
57	2016/03/11 14:53:25	20.8	- - -	 	 	
58	2016/03/11 14:53:55	20.8	- - -	 	 	
59	2016/03/11 14:54:25	20.3	- - -	 	 	
60	2016/03/11 14:54:55	19.8	- - -	 	 	
61	2016/03/11 14:55:25	20.3	- - -	 	 	
62	2016/03/11 14:55:55	20.4	- - -	 	 	
63	2016/03/11 14:56:25	21.3	- - -	 	 	
64	2016/03/11 14:56:55	21.3	- - -	 	 	
65	2016/03/11 14:57:25	20.5	- - -	 	 	
66	2016/03/11 14:57:55	21.3	- - -	 	 	
67	2016/03/11 14:58:25	21.3	- - -	 	 	
68	2016/03/11 14:58:55	20.4	- - -	 	 	
69	2016/03/11 14:59:25	20.0	- - -	 	 	
70	2016/03/11 14:59:55	20.4	- - -	 	 	
71	2016/03/11 15:00:25	19.8	- - -	 	 	
72	2016/03/11 15:00:55	19.8	- - -	 	 	
73	2016/03/11 15:01:25	19.8	- - -	 	 	
74	2016/03/11 15:01:55	20.4	- - -	 	 	
75	2016/03/11 15:02:25	20.4	- - -	 	 	
76	2016/03/11 15:02:55	20.3	- - -	 	 	
77	2016/03/11 15:03:25	21.2	- - -	 	 	
78	2016/03/11 15:03:55	20.0	- - -	 	 	
79	2016/03/11 15:04:25	21.0	- - -	 	 	
80	2016/03/11 15:04:55	20.5	- - -	 	 	
81	2016/03/11 15:05:25	20.0	- - -	 	 	
82	2016/03/11 15:05:55	20.5	- - -	 	 	
83	2016/03/11 15:06:25	20.5	- - -	 	 	
84	2016/03/11 15:06:55	21.4	- - -	 	 	
85	2016/03/11 15:07:25	21.4	- - -	 	 	
86	2016/03/11 15:07:55	20.5	- - -	 	 	
87	2016/03/11 15:08:25	20.5	- - -	 	 	
88	2016/03/11 15:08:55	20.5	- - -	 	 	
89	2016/03/11 15:09:25	21.4	- - -	 	 	
90	2016/03/11 15:09:55	20.5	- - -	 	 	
91	2016/03/11 15:10:25	21.0	- - -	 	 	
92	2016/03/11 15:10:55	21.0	- - -	 	 	
93	2016/03/11 15:11:25	20.0	- - -	 	 	
94	2016/03/11 15:11:55	21.0	- - -	 	 	
95	2016/03/11 15:12:25	21.0	- - -	 	 	
96	2016/03/11 15:12:55	20.5	- - -	 	 	
97	2016/03/11 15:13:25	21.0	- - -	 	 	
98	2016/03/11 15:13:55	20.5	- - -	 	 	
99	2016/03/11 15:14:25	20.5	- - -	 	 	
100	2016/03/11 15:14:55	20.5	- - -	 	 	
101	2016/03/11 15:15:25	20.0	- - -	 	 	
102	2016/03/11 15:15:55	21.0	- - -	 	 	
103	2016/03/11 15:16:25	21.0	- - -	 	 	
104	2016/03/11 15:16:55	19.9	- - -	 	 	
105	2016/03/11 15:17:25	21.0	- - -	 	 	
106	2016/03/11 15:17:55	21.4	- - -	 	 	
107	2016/03/11 15:18:25	20.8	- - -	 	 	
108	2016/03/11 15:18:55	20.5	- - -	 	 	
109	2016/03/11 15:19:25	21.0	- - -	 	 	
110	2016/03/11 15:19:55	20.0	- - -	 	 	
111	2016/03/11 15:20:25	19.9	- - -	 	 	
112	2016/03/11 15:20:55	21.4	- - -	 	 	
113	2016/03/11 15:21:25	20.8	- - -	 	 	
114	2016/03/11 15:21:55	20.8	- - -	 	 	
115	2016/03/11 15:22:25	20.8	- - -	 	 	
116	2016/03/11 15:22:55	20.3	- - -	 	 	
117	2016/03/11 15:23:25	21.3	- - -	 	 	
118	2016/03/11 15:23:55	20.8	- - -	 	 	
119	2016/03/11 15:24:25	20.8	- - -	 	 	
120	2016/03/11 15:24:55	20.4	- - -	 	 	
121	2016/03/11 15:25:25	21.0	- - -	 	 	
122	2016/03/11 15:25:55	20.5	- - -	 	 	
123	2016/03/11 15:26:25	20.4	- - -	 	 	
124	2016/03/11 15:26:55	20.8	- - -	 	 	
125	2016/03/11 15:27:25	21.3	- - -	 	 	
126	2016/03/11 15:27:55	21.3	- - -	 	 	
127	2016/03/11 15:28:25	20.8	- - -	 	 	
128	2016/03/11 15:28:55	21.3	- - -	 	 	
129	2016/03/11 15:29:25	21.3	- - -	 	 	
130	2016/03/11 15:29:55	20.7	- - -	 	 	
131	2016/03/11 15:30:25	21.3	- - -	 	 	
132	2016/03/11 15:30:55	21.3	- - -	 	 	
133	2016/03/11 15:31:25	20.8	- - -	 	 	
134	2016/03/11 15:31:55	20.7	- - -	 	 	
135	2016/03/11 15:32:25	21.3	- - -	 	 	
136	2016/03/11 15:32:55	21.3	- - -	 	 	
137	2016/03/11 15:33:25	20.8	- - -	 	 	
138	2016/03/11 15:33:55	20.8	- - -	 	 	
139	2016/03/11 15:34:25	20.8	- - -	 	 	
140	2016/03/11 15:34:55	19.9	- - -	 	 	
141	2016/03/11 15:35:25	21.3	- - -	 	 	
142	2016/03/11 15:35:55	20.8	- - -	 	 	
143	2016/03/11 15:36:25	20.8	- - -	 	 	
144	2016/03/11 15:36:55	21.3	- - -	 	 	
145	2016/03/11 15:37:25	20.4	- - -	 	 	
146	2016/03/11 15:37:55	20.2	- - -	 	 	
147	2016/03/11 15:38:25	21.3	- - -	 	 	
148	2016/03/11 15:38:55	20.7	- - -	 	 	
149	2016/03/11 15:39:25	20.8	- - -	 	 	
150	2016/03/11 15:39:55	20.8	- - -	 	 	
151	2016/03/11 15:40:25	20.7	- - -	 	 	
152	2016/03/11 15:40:55	20.3	- - -	 	 	
153	2016/03/11 15:41:25	20.8	- - -	 	 	
154	2016/03/11 15:41:55	19.8	- - -	 	 	
155	2016/03/11 15:42:25	19.7	- - -	 	 	
156	2016/03/11 15:42:55	21.2	- - -	 	 	
157	2016/03/11 15:43:25	20.7	- - -	 	 	
158	2016/03/11 15:43:55	20.2	- - -	 	 	
159	2016/03/11 15:44:25	20.8	- - -	 	 	
160	2016/03/11 15:44:55	20.7	- - -	 	 	
161	2016/03/11 15:45:25	20.8	- - -	 	 	
162	2016/03/11 15:45:55	21.3	- - -	 	 	
163	2016/03/11 15:46:25	20.8	- - -	 	 	
164	2016/03/11 15:46:55	20.2	- - -	 	 	
165	2016/03/11 15:47:25	21.3	- - -	 	 	
166	2016/03/11 15:47:55	20.8	- - -	 	 	
167	2016/03/11 15:48:25	21.3	- - -	 	 	
168	2016/03/11 15:48:55	21.3	- - -	 	 	
169	2016/03/11 15:49:25	21.3	- - -	 	 	
170	2016/03/11 15:49:55	20.8	- - -	 	 	
171	2016/03/11 15:50:25	19.8	- - -	 	 	
172	2016/03/11 15:50:55	20.8	- - -	 	 	
173	2016/03/11 15:51:25	21.3	- - -	 	 	
174	2016/03/11 15:51:55	19.7	- - -	 	 	
175	2016/03/11 15:52:25	20.8	- - -	 	 	
176	2016/03/11 15:52:55	20.8	- - -	 	 	
177	2016/03/11 15:53:25	20.4	- - -	 	 	
178	2016/03/11 15:53:55	20.8	- - -	 	 	
179	2016/03/11 15:54:25	21.3	- - -	 	 	
180	2016/03/11 15:54:55	21.3	- - -	 	 	
181	2016/03/11 15:55:25	20.8	- - -	 	 	
182	2016/03/11 15:55:55	20.8	- - -	 	 	
183	2016/03/11 15:56:25	20.4	- - -	 	 	
184	2016/03/11 15:56:55	19.8	- - -	 	 	
185	2016/03/11 15:57:25	21.3	- - -	 	 	
186	2016/03/11 15:57:55	21.3	- - -	 	 	
187	2016/03/11 15:58:25	21.3	- - -	 	 	
188	2016/03/11 15:58:55	20.4	- - -	 	 	
189	2016/03/11 15:59:25	21.3	- - -	 	 	
190	2016/03/11 15:59:55	20.8	- - -	 	 	
191	2016/03/11 16:00:25	20.8	- - -	 	 	
192	2016/03/11 16:00:55	19.7	- - -	 	 	
193	2016/03/11 16:01:25	20.3	- - -	 	 	
194	2016/03/11 16:01:55	21.3	- - -	 	 	
195	2016/03/11 16:02:25	20.8	- - -	 	 	
196	2016/03/11 16:02:55	21.3	- - -	 	 	
197	2016/03/11 16:03:25	21.3	- - -	 	 	
198	2016/03/11 16:03:55	20.8	- - -	 	 	
199	2016/03/11 16:04:25	20.2	- - -	 	 	
200	2016/03/11 16:04:55	21.3	- - -	 	 	
201	2016/03/11 16:05:25	20.2	- - -	 	 	
202	2016/03/11 16:05:55	20.2	- - -	 	 	
203	2016/03/11 16:06:25	20.3	- - -	 	 	
204	2016/03/11 16:06:55	20.2	- - -	 	 	
205	2016/03/11 16:07:25	19.7	- - -	 	 	
206	2016/03/11 16:07:55	21.3	- - -	 	 	
207	2016/03/11 16:08:25	19.7	- - -	 	 	
208	2016/03/11 16:08:55	21.3	- - -	 	 	
209	2016/03/11 16:09:25	20.8	- - -	 	 	
210	2016/03/11 16:09:55	20.8	- - -	 	 	
211	2016/03/11 16:10:25	20.8	- - -	 	 	
212	2016/03/11 16:10:55	21.3	- - -	 	 	
213	2016/03/11 16:11:25	21.3	- - -	 	 	
214	2016/03/11 16:11:55	21.3	- - -	 	 	
215	2016/03/11 16:12:25	20.8	- - -	 	 	
216	2016/03/11 16:12:55	20.8	- - -	 	 	
217	2016/03/11 16:13:25	20.1	- - -	 	 	
218	2016/03/11 16:13:55	20.2	- - -	 	 	
219	2016/03/11 16:14:25	21.3	- - -	 	 	
220	2016/03/11 16:14:55	21.3	- - -	 	 	
221	2016/03/11 16:15:25	20.8	- - -	 	 	
222	2016/03/11 16:15:55	20.8	- - -	 	 	
223	2016/03/11 16:16:25	20.3	- - -	 	 	
224	2016/03/11 16:16:55	21.1	- - -	 	 	
225	2016/03/11 16:17:25	20.8	- - -	 	 	
226	2016/03/11 16:17:55	20.3	- - -	 	 	
227	2016/03/11 16:18:25	20.3	- - -	 	 	
228	2016/03/11 16:18:55	20.6	- - -	 	 	
229	2016/03/11 16:19:25	19.8	- - -	 	 	
230	2016/03/11 16:19:55	20.2	- - -	 	 	
231	2016/03/11 16:20:25	20.8	- - -	 	 	
232	2016/03/11 16:20:55	20.6	- - -	 	 	
233	2016/03/11 16:21:25	21.3	- - -	 	 	
234	2016/03/11 16:21:55	20.6	- - -	 	 	
235	2016/03/11 16:22:25	20.8	- - -	 	 	
236	2016/03/11 16:22:55	20.8	- - -	 	 	
237	2016/03/11 16:23:25	21.1	- - -	 	 	
238	2016/03/11 16:23:55	20.8	- - -	 	 	
239	2016/03/11 16:24:25	20.4	- - -	 	 	
240	2016/03/11 16:24:55	20.2	- - -	 	 	
241	2016/03/11 16:25:25	21.3	- - -	 	 	
242	2016/03/11 16:25:55	20.8	- - -	 	 	
243	2016/03/11 16:26:25	20.4	- - -	 	 	
244	2016/03/11 16:26:55	20.8	- - -	 	 	
245	2016/03/11 16:27:25	19.8	- - -	 	 	
246	2016/03/11 16:27:55	19.9	- - -	 	 	
247	2016/03/11 16:28:25	20.2	- - -	 	 	
248	2016/03/11 16:28:55	20.8	- - -	 	 	
249	2016/03/11 16:29:25	20.8	- - -	 	 	
250	2016/03/11 16:29:55	21.3	- - -	 	 	
251	2016/03/11 16:30:25	21.1	- - -	 	 	
252	2016/03/11 16:30:55	20.8	- - -	 	 	
253	2016/03/11 16:31:25	21.3	- - -	 	 	
254	2016/03/11 16:31:55	21.3	- - -	 	 	
255	2016/03/11 16:32:25	21.3	- - -	 	 	
256	2016/03/11 16:32:55	20.8	- - -	 	 	
257	2016/03/11 16:33:25	21.1	- - -	 	 	
258	2016/03/11 16:33:55	20.8	- - -	 	 	
259	2016/03/11 16:34:25	20.3	- - -	 	 	
260	2016/03/11 16:34:55	21.2	- - -	 	 	
261	2016/03/11 16:35:25	20.8	- - -	 	 	
262	2016/03/11 16:35:55	20.8	- - -	 	 	
263	2016/03/11 16:36:25	20.8	- - -	 	 	
264	2016/03/11 16:36:55	21.3	- - -	 	 	
265	2016/03/11 16:37:25	21.3	- - -	 	 	
266	2016/03/11 16:37:55	21.3	- - -	 	 	
267	2016/03/11 16:38:25	21.3	- - -	 	 	
268	2016/03/11 16:38:55	19.8	- - -	 	 	
269	2016/03/11 16:39:25	20.2	- - -	 	 	
270	2016/03/11 16:39:55	20.8	- - -	 	 	
271	2016/03/11 16:40:25	20.8	- - -	 	 	
272	2016/03/11 16:40:55	20.8	- - -	 	 	
273	2016/03/11 16:41:25	20.8	- - -	 	 	
274	2016/03/11 16:41:55	20.8	- - -	 	 	
275	2016/03/11 16:42:25	21.3	- - -	 	 	
276	2016/03/11 16:42:55	21.3	- - -	 	 	
277	2016/03/11 16:43:25	20.7	- - -	 	 	
278	2016/03/11 16:43:55	20.2	- - -	 	 	
279	2016/03/11 16:44:25	20.3	- - -	 	 	
280	2016/03/11 16:44:55	20.4	- - -	 	 	
281	2016/03/11 16:45:25	20.2	- - -	 	 	
282	2016/03/11 16:45:55	20.2	- - -	 	 	
283	2016/03/11 16:46:25	21.3	- - -	 	 	
284	2016/03/11 16:46:55	20.8	- - -	 	 	
285	2016/03/11 16:47:25	20.8	- - -	 	 	
286	2016/03/11 16:47:55	20.8	- - -	 	 	
287	2016/03/11 16:48:25	20.7	- - -	 	 	
288	2016/03/11 16:48:55	20.8	- - -	 	 	
289	2016/03/11 16:49:25	19.9	- - -	 	 	
290	2016/03/11 16:49:55	20.2	- - -	 	 	
291	2016/03/11 16:50:25	21.3	- - -	 	 	
292	2016/03/11 16:50:55	21.3	- - -	 	 	
293	2016/03/11 16:51:25	20.5	- - -	 	 	
294	2016/03/11 16:51:55	21.3	- - -	 	 	
295	2016/03/11 16:52:25	21.3	- - -	 	 	
296	2016/03/11 16:52:55	21.3	- - -	 	 	
297	2016/03/11 16:53:25	20.8	- - -	 	 	
298	2016/03/11 16:53:55	20.5	- - -	 	 	
299	2016/03/11 16:54:25	20.5	- - -	 	 	
... 53646 more lines ...
Attachment 2: 2016-03-11_Evaporator_2.xlsx
  308   Monday, June 27, 2016, 11:08 Laura LambertEmpty41790-3LP-SISDonesintering TaC coating on UCx#17 container

 Monday, June 27, 2016, 11:07

Container coated with TaC using 'flushing technique'.

Wrapped with 3 x Ta heat shields, incl center block - new method of each layer being one long single Ta foil covering entire container incl center block

Currently pumping down in evap2


 Monday, June 27, 2016, 13:23

HS water valves opened

Tapped up manually to 35A

Waited ~1hr, vacuum stable

IGP1 = 1.8e-6 Torr

Initiating auto ramp for TGHT @ 0.2A/min to a max setpoint of 400A

Initiating auto ramp for TBHT @ 0.1A/min to a max setpoint of 230A

Once max setpoints achieved, will tap up manually to final desired temperature of 1900C


 Wednesday, June 29, 2016, 11:33

IGP1 = 1.9e-7

TGHT = 400A, 6.0V

TBHT = 223A, 3.9V

OMEGA = 1885C

Remained at these settings for ~13hrs

Increased TBHT max voltage to 4.3V to achieve 230A

TBHT = 230A, 4.1V

OMEGA = 1890C

Tapped up to TGHT = 405A, 6.17V to achieve 1900C...OMEGA reached 1915C

Lowered TGHT to = 403A, 6.12V

OMEGA = 1905C

Measured ~1870C with the mikron, same window the OMEGA measurements are taken out of.

IGP1 = 2.1e-7


 Thursday, June 30, 2016, 08:44

TGHT = 403A, 6.07V

OMEGA = 1886C

Tapped back up to 1900C

TGHT = 410A, 6.22V

OMEGA = 1911C


Thursday, June 30, 2016, 16:01

OMEGA = 1911C

Container has remained at ~1900C for ~1.5 days

Initiating auto-cool at 2.0A/min and 1.0A/min for the TGHT and TBHT respectively.


 Monday, July 04, 2016, 09:59

Vented in air and opened up this morning.

Container looks good.

Confirmed coating fully sintered with J.Wong.

Photos attached.


 

Attachment 1: LP-SIS_post_sintering_TaC(3).jpg
LP-SIS_post_sintering_TaC(3).jpg
Attachment 2: LP-SIS_post_sintering_TaC(4).jpg
LP-SIS_post_sintering_TaC(4).jpg
  327   Wednesday, September 28, 2016, 17:09 Laura LambertEmpty41790-6LP-SISDonesintering TaC coating on UCx container

 Wednesday, September 28, 2016, 17:08

Container coated with TaC using 'flushing technique' today, cleaned ionizer tube with acetone, no coating inside ionizer tube

Dried by blowing air into tube

Wrapped with 3 x Ta heat shields, incl center block - new method of each layer being one long single Ta foil covering entire container incl center block

Currently pumping down in evap2


 Thursday, September 29, 2016, 11:11

HS water valves opened

Tapped up to 35A

Waited ~45min, vacuum stable

IGP1 = 1.4e-6 Torr

Initiating auto ramp at 0.2A/min and 0.1A/min for the TGHT and TBHT respectively.

Max setpoints for TGHT = 350A and TBHT  = 230A

Once max setpoints achieved, will tap up manually to final desired temperature of 1900C


 Friday, September 30, 2016, 10:51

IGP1 = 7.8e-7 Torr

TGHT = 315A, 4.57V

TBHT = 174A, 2.8V

Increasing ramping rate to 1.0A/min and 0.5A/min for the TGHT and TBHT respectively.


 Friday, September 30, 2016, 11:44

Changed OMEGA slope to 0.98

IGP1 = 1e-6 Torr

TGHT = 350A, 5.2V

TBHT = 201A, 3.4V

Increasing ramping rate of TBHT to 1.0A/min

OMEGA = 1625C / MIKRON = 1625C


 Friday, September 30, 2016, 15:45

IGP1 = 1.1e-6 Torr

TGHT = 430A, 6.79V

TBHT = 230A, 4.07V

OMEGA = 1903C

Will remain at max setpoints over weekend.

See attached link for raw data of manual tap up.

Also see attached photos of screenshots showing oval - not circular - tube of target container at two different temperatures...


 Monday, October 03, 2016, 13:28

IGP1 = 1.79e-7 Torr

TGHT = 430A, 6.87V

TBHT = 230A, 4.06V

OMEGA = 1892C

MIKRON = 1847C

Attached is a picture taken through the MIKRON, the slight oval shape appears to be a minor bend in the target container:

With the ionizer tube facing into the page, the bend appears to be in the center with both ends of the container bent slightly towards the page

This possible bend in the container gives the appearance of the oval shape through the camera while it is being heated, due to the angle of the camera

Remained at max setpoints over weekend.

Initiating auto-cool at 2.0A/min and 1.0A/min for the TGHT and TBHT respectively.


 Tuesday, October 04, 2016, 13:40

Vented in air and opened up this morning.

Container done, TaC coating looks good.

Waiting further inspection...


 Tuesday, October 04, 2016, 14:48

Container placed on jig and inspected.

Approved to go ahead for use online by Anders and Friedhelm.


 

Attachment 1: sintering_TaC_coating.xlsx
Attachment 2: sintering_TaC_coating.JPG
sintering_TaC_coating.JPG
Attachment 3: sintering_TaC_coating2.JPG
sintering_TaC_coating2.JPG
Attachment 4: sintering_TaC_coating-combo_of_both_pics.JPG
sintering_TaC_coating-combo_of_both_pics.JPG
  328   Friday, October 07, 2016, 16:42 Laura LambertTa41790-5LP-SISIn progressConditioning Ta#50

Friday, October 07, 2016, 16:40

Target consists of 470 discs.

Total packed length 3.4cm [exactly 7.8cm space on both sides]

Total mass 47.6889g [Incl foils + Ta wire on 6 packs]

Note: Target consists of 350 single discs and 6x20 packs of Ta discs

3x Ta Heat shields.

No endcaps inserted, no Rh foil inserted.

Chamber currently pumping down.

TBHT will be ramped up to 230A at 1.0A/min.

TGHT will be ramped up to 300A at 2.0A/min.

Final desired temperature = 1900C


 Tuesday, October 11, 2016, 10:02
 
HS water valves opened, no leaks detected
 
Tapped up manually to 35A
 
Vacuum stable, IGP1 = 2e-7 Torr
 
Initiating auto-ramp.
 
OMEGA slope = 0.98

 Wednesday, October 12, 2016, 16:15

Tapped up to 1900C? [see attached link for data concerning temperature readings]

TGHT = 450A, 7.16V

TBHT = 230A, 4.1V

OMEGA = 1748C

IGP1 = 6.1e-7 Torr


 

Attachment 1: Ta#50_no_endcaps_3xh.s..xlsx
  329   Wednesday, October 12, 2016, 21:18 John WongTa41790-5LP-SISIn progressConditioning Ta#50

Laura Lambert wrote:

Friday, October 07, 2016, 16:40

Target consists of 470 discs.

Total packed length 3.4cm [exactly 7.8cm space on both sides]

Total mass g [Incl foils + Ta wire on 6 packs]

Note: Target consists of 350 single discs and 6x20 packs of Ta discs

3x Ta Heat shields.

No endcaps inserted, no Rh foil inserted.

Chamber currently pumping down.

TBHT will be ramped up to 230A at 1.0A/min.

TGHT will be ramped up to 300A at 2.0A/min.

Final desired temperature = 1900C


 Tuesday, October 11, 2016, 10:02
 
HS water valves opened, no leaks detected
 
Tapped up manually to 35A
 
Vacuum stable, IGP1 = 2e-7 Torr
 
Initiating auto-ramp.
 
OMEGA slope = 0.98

 Wednesday, October 12, 2016, 16:15

Tapped up to 1900C? [see attached link for data concerning temperature readings]

TGHT = 450A, 7.16V

TBHT = 230A, 4.1V

OMEGA = 1748C

IGP1 = 6.1e-7 Torr


 

 Wednesday, October 12, 2016, 21:14

Current status before cooling down:

TGHT at 450A/7.32V; TBHT at 230A/4.10V; OMEGA at 1747C

Vacuum at 4.05 e-7 Torr.

Initiate auto-rampping at 2A/min for TGHT and 1A/min for TBHT


 

  330   Thursday, October 13, 2016, 17:32 Laura LambertTa41790-5LP-SISDoneConditioning Ta#50

John Wong wrote:

Laura Lambert wrote:

Friday, October 07, 2016, 16:40

Target consists of 470 discs.

Total packed length 3.4cm [exactly 7.8cm space on both sides]

Total mass g [Incl foils + Ta wire on 6 packs]

Note: Target consists of 350 single discs and 6x20 packs of Ta discs

3x Ta Heat shields.

No endcaps inserted, no Rh foil inserted.

Chamber currently pumping down.

TBHT will be ramped up to 230A at 1.0A/min.

TGHT will be ramped up to 300A at 2.0A/min.

Final desired temperature = 1900C


 Tuesday, October 11, 2016, 10:02
 
HS water valves opened, no leaks detected
 
Tapped up manually to 35A
 
Vacuum stable, IGP1 = 2e-7 Torr
 
Initiating auto-ramp.
 
OMEGA slope = 0.98

 Wednesday, October 12, 2016, 16:15

Tapped up to 1900C? [see attached link for data concerning temperature readings]

TGHT = 450A, 7.16V

TBHT = 230A, 4.1V

OMEGA = 1748C

IGP1 = 6.1e-7 Torr


 

 Wednesday, October 12, 2016, 21:14

Current status before cooling down:

TGHT at 450A/7.32V; TBHT at 230A/4.10V; OMEGA at 1747C

Vacuum at 4.05 e-7 Torr.

Initiate auto-rampping at 2A/min for TGHT and 1A/min for TBHT


 

 Thursday, October 13, 2016, 17:32

vented in air and opened up this morning.

target looks good.

ELOG V2.9.2-2455