Entry  Wed Aug 13 15:00:27 2014, Raymond Dubé, Routine, General, OLIS Surface Source broken Ionizer Coil ISK4386D-rev5.pdfISK4583D-rev7.pdfISK4584D-rev1.pdfPE-R5-FAN_ON.jpg
We spent most of the day looking at OLIS to find out why the "Ionizer Coil" keeps breaking down, we found
the following:
1.  The FAN to cool R5 (ref: ISK4583D rev.7) a 25 Ohm 200W Resistor (2x50 ohm/100W ea.) was not ON, the nylon
Entry  Wed Aug 13 15:09:42 2014, Raymond Dubé, Routine, General, OLIS Attenuators #2 fixed, #1 gone! 
Today Probe group (Scott K.) installed the OLIS Attenuator, they fixed the vacuum leak on Attenuator #2,  Marco
M. asked them to remove Attenuator #1 because it is not needed and confusing for Ops;
 
Entry  Tue Sep 16 11:55:28 2014, Raymond Dubé, Routine, General, MCIS info Cart-Power1.pdfSuperNanogan_Control-R1.pdfSupernanogan_Interlocks-R1.pdf
I found these files about the MCIS in Geoff folders in M:\ISIS\Geoff-110715\OLIS\SuperNanogan
Entry  Fri Oct 10 14:02:56 2014, Raymond Dube, Routine, General, Test with new Alga, ALPA 50W RF AMP ALPA-RF_AMP-50W.pdf
We did a test yesterday with the new Alga Microwave RF AMP, model # ALPA 50 Watt, serial # 130684 to be used for the MCIS.  We used the RF Generator
from MISTIC and set the frequency to 13.80 Ghz, note that the RF Generator was/is  limited to 16DB maxiumum RF output.
The RF cable that we used was from the 200W RF AMP, it has a 20DB SMA attenuator at the Generator end, so we could only get about 25 Watts maximum
Entry  Thu Oct 23 10:52:21 2014, Brian Minato, Routine, General, ALGA Amplifier 
The ALGA amplifier and its signal generator have been turned off at the request of Controls.  They have had communication issues and are looking
for any recent changes.
Entry  Thu Oct 30 22:44:45 2014, Keerthi Jayamanna, Routine, General, Mass scan 3 to 4 
IOS mass scan if faulty. These is a manual scan from mass 3 to 4
Const22 = 0.04187  adjusted to mass = 4
A/q               FC6           X         
Entry  Thu Oct 30 23:37:58 2014, Keerthi Jayamanna, Routine, General, Mass scan 4 to 6 
A/Q            FC6
4                1.1e-6
4.25          8e-10
Entry  Thu Oct 30 23:44:57 2014, Keerthi Jayamanna, Routine, General, OLIS Supernanogan mass scan 6 to 7.5 
A/q           FC6  (only current over 100pA)
6.0           1.85e-7
6.15         4.96e-10
Entry  Wed Nov 12 13:32:47 2014, Raymond Dube, Routine, General, New in-line resistors for IOS:PE ISK4583D-rev10.pdf
Today we replace the in-line resistors for the Plasma Electrode (PE) it was a 25 ohm 200W (2 x 50 Ohm 100W) with a 50 Ohm 200W (2 x 100 Ohm / 100W) also
added the inductance coil.
This is R5 on drawing: ISK4583D rev. 10
Entry  Wed Nov 19 11:34:55 2014, Raymond Dube, Routine, General, OLIS Electrical diagram with 3 I.S. ISK4586D-rev2.pdf
I updated the electrical Diagram  for OLIS Faraday Cage with 3 Ion Source, ref: ISK4586D rev. 2
PS:  Make sure that you do not connect both Bias P/S on same Ion Source ... 
Entry  Tue Nov 25 12:23:28 2014, Keerthi Jayamanna, Routine, General, 76Se+13 tune saved 
141125_1159_OLIS_76Se+13.snap is saved in reference directory.
El=3480/4303V
MCIS:YCB1=110
Entry  Mon Apr 20 08:09:03 2015, Brian Minato, Routine, General, OLIS Shutdown Update 
The OLIS 1 week shutdown has been completed (WP 2015-04-13-1).
General-
Y-Box insulators were inspected and look good.  Beamline optics were inspected.
Entry  Mon Sep 14 11:21:16 2015, Keerthi Jayamanna, Routine, General, 20Ne+2 beam saved 
20Ne+2 beam optics saved as 150914_1117_ios0iosfc6.snap