ID Date Authordown Operational Mode Subject Text Attachments
  127   Wed Jun 14 10:35:30 2006 Willy AnderssonQuench RecoverySwitched to series cooling.Had system in //return cooling overnight with
small flow to keep power 
leads cool. Switched to series mode about
  
  128   Wed Jun 14 16:27:01 2006 Willy AnderssonQuench RecoveryEnergized solenoidPowerleads appear stable.   
  139   Thu Jul 20 11:01:09 2006 Willy AnderssonQuench RecoveryOpened up warm return valves CV151 and CV153Mu2 pressure was just about at trip point
due to trapped volume of helium 
warming. So opened up warm returns to bleed
  
  140   Thu Jul 20 13:21:57 2006 Willy AnderssonQuench Recoverystarted // warm return flow.TIC121 was down to 27K
warm return valves were wide open
opened supply valves CV143 and CV146.
  
  141   Thu Jul 20 13:27:21 2006 Willy AnderssonQuench Recoverycold return mode continues...cr_c1 and cr_12 are both at about 45K and
heading slowly down, so the 
gamble of going to // cold return worked.
  
  142   Thu Jul 20 13:34:26 2006 Willy AnderssonQuench Recoverycold return mode continues...Even though cr_c1 and cr_c12 have begun to
tick down, cr_c7 and cr_c6 are 
actually ticking up. This is probably a wave
  
  145   Fri Jul 21 17:18:19 2006 Willy AnderssonQuench RecoveryOpened HIC142 fully openPower leads have flatlined but still building
LHe level.
  
  152   Wed Aug 9 06:44:58 2006 Willy AnderssonQuench Recovery// cold mode cooling continues.After restarting flow to the solenoid yesterday
in // cold return mode 
TIC121 just refused to go lower than about
  
  153   Wed Aug 9 06:59:43 2006 Willy AnderssonQuench Recoverymore // cold return stuffAbout 21:00 last night the solenoid temps
had not changed much, pretty well 
flat lined. I guess my reducing CV155 stalled
  
  156   Thu Aug 10 16:33:11 2006 Willy AnderssonQuench RecoveryFlow to solenoid tripped on powerlead temperature.Power lead temps were drifting up, tried to
tweak TIC121 and HIC141 to 
unstall powerlead flow. Got quench. Restarted
  
  157   Thu Aug 10 20:26:09 2006 Willy AnderssonQuench Recoverysolenoid energizedPowerleads seem stable.   
  159   Tue Aug 15 15:39:27 2006 Willy AnderssonQuench Recoveryswitched to parallel cold returnReturn flow temps from solenoid about 40K,
TIC121 at about 40K so switching 
to //cold return.
  
  161   Thu Aug 17 19:45:27 2006 Willy AnderssonQuench Recoveryswitch to series coolingTemps of flow returning from solenoid flat
lined at about 10K.
Gradually closed CV155.
  
  163   Wed Oct 4 18:10:18 2006 Willy AnderssonStart UpQuick check on systemSystem in good // warm return mode, temps
coming down nicely. BCP2 cryopump 
does not look too healthy, temperature is
  
  167   Thu Oct 12 11:53:04 2006 Willy AnderssonQuench RecoveryCompressor trip, 09:50 this morning.Main compressor tripped, refrigerator shutdown.
Checked with experimentors 
and main control room, no indications of
  
  168   Thu Oct 12 19:54:10 2006 Willy AnderssonQuench RecoveryEnergized solenoid.Switched to series mode about 6 hours ago.
Cooling continued including 
power leads until LHe level at about 50%
  
  171   Thu Oct 26 04:27:31 2006 Willy AnderssonQuench RecoveryPowerlead overtemp02:30 this morning got page from operations.
Looked at system online, appears that there
was a powerlead overtemp
  
  172   Thu Oct 26 04:34:15 2006 Willy AnderssonQuench RecoveryStill in // cold return mode.Leaving system cooling in // cold return mode,
cooldown rates still good. 
Had flow on to powerleads since re-starting
  
  183   Wed Nov 15 17:05:49 2006 Willy AnderssonQuench Recoverypowerleads cold, energizing solenoidPowerlead temperatures have bottomed out,
energizing solenoid to 300 
amps, ramp rate 60.
  
  57   Sat Oct 9 22:50:26 2004 Other power leads warmedAt 10:30 PM Saturday, Donald Arseneau sees
power leads at
25K, having gone up for the previous 10 hours.
  