ID |
Date |
Author |
Target/No.# |
Source |
Station |
Module |
Target Oven W.O# |
Heat Shield W.O# |
Status |
Subject |
Text |
|
|
388
|
Thursday, June 19, 2014, 10:48 |
Anders Mjos | | | | | | | Problems/Solutions | FC0 did not retract from beam line | Upon inspection it was found that the bolt
attaching the FC to the actuator had fallen
off. The bolt was reattached and FC0 can |
|
|
389
|
Friday, June 20, 2014, 11:44 |
Anders Mjos | | | | | | | In progress | Pumping down empty chamber | Friday, June 20, 2014, 11:44: Roughing
Friday, June 20, 2014, 13:00: Strarting
turbos |
|
|
391
|
Tuesday, July 08, 2014, 09:05 |
Anders Mjos | | | | | | | In progress | Empty chamber | Tuesday, July 08, 2014, 09:05: Roughing
Tuesday, July 08, 2014, 13:23: Starting
turbos |
|
|
392
|
Wednesday, July 09, 2014, 14:08 |
Anders Mjos | | | | | | | Problems/Solutions | FEBIAD ANODE current readback | The read back of the ANODE current was
confirmed to be working today with help of
Ray D. using a 1kOhm resistor. There is an |
|
|
393
|
Friday, July 18, 2014, 08:19 |
Anders Mjos | | | | | | | | Venting | Friday, July 18, 2014, 08:19: Stopped turbos
Friday, July 18, 2014, 10:58: Venting |
|
|
395
|
Wednesday, July 23, 2014, 14:34 |
Anders Mjos | | | | | | | In progress | Empty chamber | Wednesday, July 23, 2014, 14:34: Roughing
Wednesday, July 23, 2014, 15:03: Starting
turbos. |
|
|
396
|
Friday, August 01, 2014, 08:39 |
Anders Mjos | | | | | | | Problems/Solutions | TMS:RP2 noisy | Friday, August 01, 2014, 08:39: Backing
pressure 0 / 11 mT |
|
|
398
|
Tuesday, August 12, 2014, 09:52 |
Anders Mjos | | | | | | | Problems/Solutions | RTD installed on Supply water | Travis assisted in installing a temperature
reading RTD from Omega (SA1-RTD-MTP) on the
water supply. The RTD has been installed |
|
|
399
|
Thursday, August 14, 2014, 08:45 |
Anders Mjos | | | | | | | Problems/Solutions | FC0 reading full current | since yesterday around 19h. See attached
screen shot of archived data
Thursday, August 14, 2014, 08:40: |
|
|
400
|
Wednesday, August 20, 2014, 16:25 |
Anders Mjos | | | | | | | Problems/Solutions | Lost power on TGHT, TBHT and EE simultaniously | Possible controls problem. Breaker 18 found
latched. Can not be re-energized. Burnt smell
in PLC / HV area. e-faut 7561 |
|
|
401
|
Wednesday, August 27, 2014, 12:55 |
Anders Mjos | | | | | | | Problems/Solutions | Water leak on MB1 | e-faulted
by operations Aug 26th |
|
|
403
|
Monday, September 08, 2014, 16:23 |
Anders Mjos | UC#11 | | | TM4 | 40282-5 | | Done | UC#11 Heat Shield Conditioning | T/C installed
CH 1: Attached to TGHT outside middle
CH 2: Through HS outside TGHT right |
|
|
404
|
Thursday, September 25, 2014, 15:44 |
Anders Mjos | ThO#1 | | ITW | TM1 | 40282-5 | | Done | ThO#1 Heat Shield Conditioning | Thursday, September 25, 2014, 15:44: Roughing
Thursday, September 25, 2014, 16:06:
Starting turbos |
|
|
405
|
Tuesday, October 07, 2014, 09:32 |
Anders Mjos | | | | | | | In progress | Pumping down empty chamber | Tuesday, October 07, 2014, 09:30: Roughing
Tuesday, October 07, 2014, 10:24:
Starting turbos |
|
|
406
|
Monday, October 20, 2014, 09:17 |
Anders Mjos | | | | | | | Problems/Solutions | Power bump this weekend tripped off all turbo pumps | Monday, October 20, 2014, 09:17:
Roughing
Monday, October 20, 2014, 09:18 : |
|
|
407
|
Monday, October 27, 2014, 09:26 |
Anders Mjos | | | | | | | Problems/Solutions | TSG:TP1B noisy | e-fault 7748
Monday, October 27, 2014, 09:21:
Switched off turbos |
|
|
409
|
Monday, November 03, 2014, 14:50 |
Anders Mjos | | | | | | | Problems/Solutions | IV4 did not close until 4th attempt | Indicated with red valve on the EPICS vacuum
page |
|
|
410
|
Monday, November 24, 2014, 08:41 |
Anders Mjos | | | | | | | Problems/Solutions | TSG:RP1 replaced | by Edi on Friday Nov 11th |
|
|
411
|
Friday, December 12, 2014, 13:11 |
Anders Mjos | | | | | | | Problems/Solutions | HV inspection and cleaning | by Ray D and Anders in preparation for
60kV test. Rack insulators and f/t insulators
cleaned with windex |
|
|
412
|
Friday, December 12, 2014, 13:11 |
Anders Mjos | | | | | | | Problems/Solutions | HV inspection and cleaning | by Ray D and Anders in preparation for
60kV test. Rack insulators and f/t insulators
cleaned with windex |
|