  Test-Stand  Not logged in ELOG logo
Message ID: 398     Entry time: Tuesday, August 12, 2014, 09:52
Author: Anders Mjos 
Target/No.#:  
Source:  
Station:  
Module:  
Target Oven W.O#:  
Heat Shield W.O#:  
Status: Problems/Solutions 
Subject: RTD installed on Supply water 

Travis assisted in installing a temperature reading RTD from Omega (SA1-RTD-MTP) on the water supply. The RTD has been installed with a 503 Ohm resistor => 0V = -175.75 C, 10V = 205.71 C. The RTD has been connected to Ch 9 on the data logger.

Offset calibration data towards the SUPPLY temperature reading is attached.

Attachment 1: 2014-08-12_Dataq_Screen_Shot.png  46 kB  | Show | Hide all
Attachment 2: Dataq_Supply_Temperature_calibration_2014-08-20.docx  814 kB  Uploaded Wednesday, August 20, 2014, 11:07