  396   Friday, August 01, 2014, 08:39 Anders Mjos      Problems/SolutionsTMS:RP2 noisyFriday, August 01, 2014, 08:39: Backing
pressure 0 / 11 mT
  
  398   Tuesday, August 12, 2014, 09:52 Anders Mjos      Problems/SolutionsRTD installed on Supply waterTravis assisted in installing a temperature
reading RTD from Omega (SA1-RTD-MTP) on the
water supply. The RTD has been installed
 2014-08-12_Dataq_Screen_Shot.pngDataq_Supply_Temperature_calibration_2014-08-20.docx 
  399   Thursday, August 14, 2014, 08:45 Anders Mjos      Problems/SolutionsFC0 reading full currentsince yesterday around 19h. See attached
screen shot of archived data
Thursday, August 14, 2014, 08:40:
 FC0FullScaleReading_2014-08-14.pngFC0FullScaleReading_EPICS_2014-08-14.png 
  400   Wednesday, August 20, 2014, 16:25 Anders Mjos      Problems/SolutionsLost power on TGHT, TBHT and EE simultaniously Possible controls problem. Breaker 18 found
latched. Can not be re-energized. Burnt smell
in PLC / HV area. e-faut 7561
  
  401   Wednesday, August 27, 2014, 12:55 Anders Mjos      Problems/SolutionsWater leak on MB1e-faulted
by operations Aug 26th
  
  403   Monday, September 08, 2014, 16:23 Anders MjosUC#11  TM440282-5 DoneUC#11 Heat Shield ConditioningT/C installed
CH 1: Attached to TGHT outside middle
CH 2: Through HS outside TGHT right
 Thermocouples.xlsx 
  404   Thursday, September 25, 2014, 15:44 Anders MjosThO#1 ITWTM140282-5 DoneThO#1 Heat Shield ConditioningThursday, September 25, 2014, 15:44: Roughing
Thursday, September 25, 2014, 16:06:
Starting turbos
  
  405   Tuesday, October 07, 2014, 09:32 Anders Mjos      In progressPumping down empty chamberTuesday, October 07, 2014, 09:30: Roughing
Tuesday, October 07, 2014, 10:24:
Starting turbos
  
  406   Monday, October 20, 2014, 09:17 Anders Mjos      Problems/SolutionsPower bump this weekend tripped off all turbo pumps Monday, October 20, 2014, 09:17:
Roughing
Monday, October 20, 2014, 09:18 :
  
  407   Monday, October 27, 2014, 09:26 Anders Mjos      Problems/SolutionsTSG:TP1B noisye-fault 7748
Monday, October 27, 2014, 09:21:
Switched off turbos
  
  409   Monday, November 03, 2014, 14:50 Anders Mjos      Problems/SolutionsIV4 did not close until 4th attemptIndicated with red valve on the EPICS vacuum
page
  
  410   Monday, November 24, 2014, 08:41 Anders Mjos      Problems/SolutionsTSG:RP1 replaced by Edi on Friday Nov 11th   
  411   Friday, December 12, 2014, 13:11 Anders Mjos      Problems/SolutionsHV inspection and cleaningby Ray D and Anders in preparation for
60kV test. Rack insulators and f/t insulators
cleaned with windex
  
  412   Friday, December 12, 2014, 13:11 Anders Mjos      Problems/SolutionsHV inspection and cleaningby Ray D and Anders in preparation for
60kV test. Rack insulators and f/t insulators
cleaned with windex
  
  413   Friday, December 12, 2014, 14:09 Anders Mjos LP-FEBIAD  Tube only In progressHV BIAS TestBIAS current limit pot changed from 2.40
to 0.25
EZL current limit pot changed from
 2014-12-17_60_kV_3h.png60_kV_Test.docx 
  414   Tuesday, December 16, 2014, 10:18 Anders Mjos      Problems/SolutionsTP1B and TP6 stopped/faultpossibly due to HV spark
Tuesday, December 16, 2014, 10:24:
Starting TP1B and TP6
  
  415   Thursday, December 18, 2014, 14:34 Anders Mjos      Problems/SolutionsGE cooling water lines disconnected In preparation for leak check
Thursday, December 18, 2014, 16:37:
Reconnected after leak check
  
  416   Thursday, December 18, 2014, 16:38 Anders Mjos      Problems/SolutionsLeak found on service cap F/T Q 9 pin connectorLargest response on the bottom of the insulator.
Thursday, December 18, 2014, 16:39:
Closing pumping valves. Will allow system
 Spark_marks_on_insulator.jpgSpark_marks_on_insulator2.jpg 
  417   Monday, December 22, 2014, 11:06 Anders Mjos      Problems/SolutionsCopper rod removedA copper rod - not used was removed from
one of the cooling lines.
  
  418   Tuesday, December 23, 2014, 09:08 Anders Mjos       Water closed for ChristmasShutting of water supply to the source
box. Water not needed.
  