Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  RH-ISAC, Page 113 of 141  Not logged in ELOG logo
ID Date Author Categorydown Type Module Target/Number Subject
  645   Thursday, April 10, 2014, 14:03 Bevan MossConditioning StationStandard OperationTM3TiC#4HV conditioning

 The TiC#4 target that is on TM3 was conditioned to 25kV today with the EINZEL lens at 12.5kV. There was one spark shortly after reaching 25kV that reset the bias PS and another one  after the module sat at 25kV for ~30 minutes. Attached is the strip tool data and screen shot. Prior to starting the conditioning the power on the chiller needed to by cycled as the VFD on the fans had tripped due to a recurring inrush problem.

Attachment 1: 10042014_Tm3_TiC#4_HV_conditioning.JPG
10042014_Tm3_TiC#4_HV_conditioning.JPG
Attachment 2: 10042014_TM3_TiC#4_HV_conditioning.txt
Time	TCS:IG1:RDVAC [T]	TCS:IG2:RDVAC [T]	TCS:IG1S:RDVAC [T]	TCS:EL:RDVOL [V]	TCS:EL:RDCUR [mA]	TCS:BIAS:RDVOL [V]	TCS:BIAS:RDCUR [uA]	TCS:DIC1:RDCOND [Mohm]	
04/10/2014 10:51:30.980156	                      0.000000961                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:31.979337	                      0.000000959                      0.000005934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:32.979181	                      0.000000959                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:33.979125	                      0.000000959                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:34.979530	                      0.000000957                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:35.979903	                      0.000000957                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:36.979928	                      0.000000959                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:37.979724	                      0.000000959                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:38.980475	                      0.000000957                      0.000005890                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:39.980322	                      0.000000959                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:41.721703	                      0.000000957                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:42.720980	                      0.000000959                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:43.721829	                      0.000000959                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:44.721709	                      0.000000959                      0.000005934                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:45.721877	                      0.000000959                      0.000005890                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:46.721604	                      0.000000959                      0.000005912                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:47.721360	                      0.000000957                      0.000005912                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:48.722248	                      0.000000959                      0.000005890                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:49.722132	                      0.000000954                      0.000005890                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:50.722206	                      0.000000954                      0.000005912                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:51.722849	                      0.000000959                      0.000005912                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:52.722691	                      0.000000959                      0.000005934                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:53.723567	                      0.000000959                      0.000005934                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:54.723552	                      0.000000957                      0.000005912                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:55.723346	                      0.000000957                      0.000005934                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:56.723198	                      0.000000957                      0.000005934                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:57.723128	                      0.000000961                      0.000005934                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:58.723980	                      0.000000961                      0.000005912                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:59.723852	                      0.000000957                      0.000005912                      0.000003890                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:00.723765	                      0.000000959                      0.000005912                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:01.723774	                      0.000000959                      0.000005890                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:02.723712	                      0.000000959                      0.000005912                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:03.723618	                      0.000000957                      0.000005890                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:04.724220	                      0.000000957                      0.000005890                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:05.724345	                      0.000000959                      0.000005890                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:06.724243	                      0.000000957                      0.000005890                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:07.723847	                      0.000000959                      0.000005912                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:08.723737	                      0.000000957                      0.000005912                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:09.734748	                      0.000000957                      0.000005890                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:10.734468	                      0.000000961                      0.000005912                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:11.734336	                      0.000000961                      0.000005890                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:12.734202	                      0.000000957                      0.000005934                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:13.735274	                      0.000000959                      0.000005934                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:14.736049	                      0.000000959                      0.000005890                      0.000003934                             6.10                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:15.735844	                      0.000000961                      0.000005890                      0.000003934                             6.10                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:16.735707	                      0.000000959                      0.000005912                      0.000003934                             6.71                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:17.735550	                      0.000000959                      0.000005912                      0.000003934                             6.71                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:18.735469	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:19.735397	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:20.736220	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:21.737118	                      0.000000961                      0.000005912                      0.000003934                             6.10                             0.01                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:22.736946	                      0.000000961                      0.000005912                      0.000003912                             6.10                             0.01                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:23.736842	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                           BadVal                           BadVal
04/10/2014 10:52:24.736718	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                           BadVal                           BadVal
04/10/2014 10:52:25.736582	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                           BadVal                           BadVal
04/10/2014 10:52:26.735970	                      0.000000959                      0.000005912                      0.000003912                             6.71                             0.01                            18.31                           BadVal                           BadVal
04/10/2014 10:52:27.764214	                      0.000000959                      0.000005912                      0.000003934                             6.71                             0.01                            18.31                           BadVal                           BadVal
04/10/2014 10:52:28.764296	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:29.764046	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            18.31                             0.18                           BadVal
04/10/2014 10:52:30.763942	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:31.764595	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:32.764669	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:33.765611	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:34.765460	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:35.765345	                      0.000000961                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:36.765196	                      0.000000959                      0.000005912                      0.000003912                             6.71                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:37.765304	                      0.000000959                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:38.765944	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:39.765800	                      0.000000959                      0.000005890                      0.000003956                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:40.765665	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:41.766003	                      0.000000954                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:42.766445	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:43.767346	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:44.767203	                      0.000000961                      0.000005890                      0.000003890                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:52:45.767105	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:52:46.766994	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:47.766939	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:48.851745	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:49.877773	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:50.877440	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:51.877267	                      0.000000959                      0.000005912                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:52.877165	                      0.000000959                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:53.877016	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:54.877020	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:52:55.876788	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:52:56.876639	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:57.876545	                      0.000000961                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:58.876493	                      0.000000961                      0.000005912                      0.000003912                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:52:59.876303	                      0.000000959                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:00.876207	                      0.000000954                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:01.876057	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:02.973167	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:53:04.127504	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:05.245403	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:06.245504	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:07.246554	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:08.246272	                      0.000000961                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:09.246145	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:10.245957	                      0.000000959                      0.000005890                      0.000003868                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:11.245855	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:12.246803	                      0.000000959                      0.000005912                      0.000003934                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:53:13.246602	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:14.246883	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:15.247430	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:16.248242	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:17.248123	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:18.248018	                      0.000000959                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:19.247819	                      0.000000959                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:20.248760	                      0.000000957                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:21.248576	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:22.248523	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:23.248379	                      0.000000959                      0.000005890                      0.000003868                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:24.248306	                      0.000000961                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:25.248120	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:26.247992	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:27.247865	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:28.247725	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:29.248587	                      0.000000957                      0.000005912                      0.000003868                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:53:30.248506	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:53:31.477726	                      0.000000961                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:32.510989	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:33.570288	                      0.000000961                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:34.569946	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:35.569810	                      0.000000954                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:36.569693	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:37.569517	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:38.570683	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:39.570275	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:40.570189	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:53:41.663248	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:42.796751	                      0.000000961                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:43.796790	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:53:44.796660	                      0.000000961                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:53:45.797531	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:53:46.797366	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:47.963516	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:48.963095	                      0.000000959                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:49.964000	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:53:50.963885	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:51.963082	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:52.962598	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:53.963488	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:54.964455	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:53:55.964257	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:56.964088	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:53:57.964030	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:58.964837	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:59.964711	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:00.964585	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:01.964460	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:02.964438	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:03.965278	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            18.31                             0.18                             13.2
04/10/2014 10:54:04.965134	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:06.114629	                      0.000000959                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:07.114842	                      0.000000959                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:08.114859	                      0.000000959                      0.000005890                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:09.115601	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:11.95567	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:12.95203	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:13.95062	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:14.94930	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:15.95034	                      0.000000954                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:16.94756	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:17.94640	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:18.94482	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:19.94499	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:20.94191	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:21.94057	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:22.94970	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:23.94867	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:24.95670	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:25.95555	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:26.95473	                      0.000000959                      0.000005934                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:27.95345	                      0.000000957                      0.000005890                      0.000003868                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:28.95171	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:29.156966	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:30.219271	                      0.000000954                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:31.218823	                      0.000000959                      0.000005868                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:32.218700	                      0.000000954                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:33.218572	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:34.219484	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:35.219324	                      0.000000954                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:36.219155	                      0.000000954                      0.000005890                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:37.219030	                      0.000000954                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:38.218939	                      0.000000957                      0.000005868                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:39.219905	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:40.219644	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:41.219569	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:42.219471	                      0.000000954                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:43.219276	                      0.000000954                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:44.220207	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:45.226449	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:46.225901	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:47.225793	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:48.225688	                      0.000000959                      0.000005890                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:49.237062	                      0.000000954                      0.000005890                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:50.237414	                      0.000000954                      0.000005868                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:51.237318	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:52.237170	                      0.000000957                      0.000005868                      0.000003868                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:53.237085	                      0.000000959                      0.000005890                      0.000003890                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:54.237035	                      0.000000957                      0.000005890                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:55.236818	                      0.000000959                      0.000005890                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:56.236684	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:57.236508	                      0.000000957                      0.000005868                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:58.236539	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:59.237346	                      0.000000954                      0.000005934                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:00.238278	                      0.000000957                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:01.238061	                      0.000000957                      0.000005868                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:02.237902	                      0.000000957                      0.000005868                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:03.237965	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:04.238630	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:05.239504	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:06.239438	                      0.000000954                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:07.239297	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:08.240205	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:55:09.240127	                      0.000000959                      0.000005890                      0.000003890                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:10.240229	                      0.000000954                      0.000005890                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:11.240963	                      0.000000957                      0.000005890                      0.000003868                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:12.242313	                      0.000000959                      0.000005868                      0.000003868                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:13.242643	                      0.000000959                      0.000005890                      0.000003868                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:14.243588	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:15.243253	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:16.243399	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:17.243056	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:18.243878	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:19.244834	                      0.000000954                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:20.245696	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:21.245500	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:22.245413	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:23.245446	                      0.000000954                      0.000005912                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:24.246172	                      0.000000954                      0.000005912                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:25.247101	                      0.000000954                      0.000005912                      0.000003912                             6.71                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:55:26.246944	                      0.000000957                      0.000005912                      0.000003890                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:27.246745	                      0.000000957                      0.000005890                      0.000003956                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:28.246707	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:29.247622	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:30.247664	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:31.247309	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:32.247107	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:33.247041	                      0.000000959                      0.000005890                      0.000003868                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:34.247898	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:35.248757	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:36.248622	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:37.248512	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:38.248407	                      0.000000954                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:39.248278	                      0.000000954                      0.000005868                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:40.248215	                      0.000000954                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:41.248104	                      0.000000959                      0.000005912                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:42.247859	                      0.000000959                      0.000005868                      0.000003890                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:43.247775	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:44.248699	                      0.000000959                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:45.248591	                      0.000000954                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:46.248414	                      0.000000954                      0.000005890                      0.000003934                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:55:47.248231	                      0.000000954                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:48.248200	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:50.356306	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:51.356770	                      0.000000954                      0.000005934                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:52.356658	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:53.356479	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:55.692384	                      0.000000959                      0.000005912                      0.000003868                             6.71                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:55:56.693040	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:57.693001	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:58.692827	                      0.000000959                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:59.692705	                      0.000000954                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:00.693597	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:01.693469	                      0.000000959                      0.000005912                      0.000003934                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:56:02.693310	                      0.000000959                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:03.693156	                      0.000000959                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:04.693036	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            18.31                             0.18                             13.2
04/10/2014 10:56:05.692994	                      0.000000957                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:06.692862	                      0.000000957                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:07.693748	                      0.000000954                      0.000005912                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:08.693600	                      0.000000954                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:09.694444	                      0.000000957                      0.000005868                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:10.695298	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:11.695160	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:12.695091	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:13.694987	                      0.000000954                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:14.694827	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:15.694742	                      0.000000959                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:16.694531	                      0.000000959                      0.000005934                      0.000003890                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:17.694420	                      0.000000959                      0.000005890                      0.000003890                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:18.694323	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:19.694196	                      0.000000954                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:20.694200	                      0.000000954                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:21.693936	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:22.693796	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:23.693687	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:24.693613	                      0.000000957                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:25.693505	                      0.000000959                      0.000005868                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:26.693314	                      0.000000957                      0.000005868                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:27.693140	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:28.693023	                      0.000000957                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:29.692952	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:30.693831	                      0.000000954                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:31.693711	                      0.000000957                      0.000005934                      0.000003956                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:32.694613	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:33.694463	                      0.000000957                      0.000005912                      0.000003890                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:34.694275	                      0.000000957                      0.000005868                      0.000003890                             6.71                             0.01                            17.16                             0.18                             13.2
... 7596 more lines ...
  646   Friday, April 11, 2014, 11:35 David WangConditioning StationStandard OperationTM3 Helium pressure test on TM3 HS water line with new Ta target. at CS

The helium pressure test on TM3 HS water line with new Ta target is finished at CS. The circuit is leak tight.. No any response on LD. See pictures.

Attachment 1: 027.JPG
027.JPG
Attachment 2: 026.JPG
026.JPG
Attachment 3: 025.JPG
025.JPG
Attachment 4: 022.JPG
022.JPG
Attachment 5: 024.JPG
024.JPG
Attachment 6: 023.JPG
023.JPG
  674   Monday, June 02, 2014, 13:53 Travis CaveConditioning StationStandard OperationTM1UC#9module move

TM#1 with UC#9 has been moved from the south hot cell to the conditioning station.

  675   Monday, June 02, 2014, 15:40 David WangConditioning StationStandard OperationTM1 TM1 and CS turbo pumps have been started

TM1 and CS turbo pumps have been started. The vacuum readings are fine on all gauges.

  676   Tuesday, June 03, 2014, 09:12 David WangConditioning StationStandard OperationTM1UC9#TM1 HS line helium pressure test and TM1 electrical check at CS

TM1 HS line helium pressure test and TM1 electrical check at CS under vacuum are finished. Everything is good. See attachment

Attachment 1: 001.JPG
001.JPG
Attachment 2: 002.JPG
002.JPG
Attachment 3: 003.JPG
003.JPG
Attachment 4: 004.JPG
004.JPG
Attachment 5: 005.JPG
005.JPG
Attachment 6: 006.JPG
006.JPG
Attachment 7: TM1_UC9#_E-check_at_CS.pdf
TM1_UC9#_E-check_at_CS.pdf
  677   Monday, June 09, 2014, 10:33 David WangConditioning StationStandard OperationTM1UC9#Stop TM1 and CS trubo pumps and ruff pump.

The CS and TM1 is still under vacuum. They will be vented with argon tomorrow mornig

  678   Tuesday, June 10, 2014, 07:57 David WangConditioning StationStandard OperationTM1UC#9 TM1 are vented with argon and disconnected at CS.

TM1 will be moved to ITW this morning.

  694   Thursday, June 26, 2014, 14:18 Travis CaveConditioning StationStandard OperationTM3SiC#30module move

TM#3 with SiC#30 has been moved from the south hot cell to the conditioning station.

  695   Friday, June 27, 2014, 10:16 David WangConditioning StationStandard OperationTM3  TM3 HS water lines helium pressure test. at CS

TM3 HS water line helium pressure test is done at CS. 70 psi helium , 3 minutes.Base LR 0.0xE-9 atm.cc/sec, base pressure 0.0xE-4 torr.  No response. The line is leak free.

  708   Thursday, July 17, 2014, 10:13 David WangConditioning StationStandard OperationTM1 Leak check on TM1 HS water line with new target installed

Chad and I leak checked TM1 HS waterline with new target installed at SHC. The LD base LR:0.0xE-9atm.cc/sec, base pressure: 0.0xE-4 torr. Chad sprayed helium on both VCR fittings and flushed helium around the target. No response has been found on leak detector. The target HS line is leak tight so far. I will do helium pressure test on this line at CS on next Monday.

  709   Thursday, July 17, 2014, 11:49 David WangConditioning StationStandard OperationTM1 TM1 and CS turbo pumps have been started

I connected TM1 at CS. All turbo pumps have been started. The pumping down is  good so far.

  710   Monday, July 21, 2014, 13:11 David WangConditioning StationStandard OperationTM1 helium pressure test on TM1 HS water line at CS.

The TM1 HS line is  pressure tested this morning with 70 psi helium  at CS. No response has been found on leak detector.

  711   Tuesday, July 22, 2014, 07:32 David WangConditioning StationStandard OperationTM1 Stop TM1 and CS turbo pumps and ruff pump.

Stop TM1 and CS turbo pumps and ruff pump. TM1 is disconnect at CS. It will be vented tomorrow morning and moved to ITW.

  715   Monday, July 28, 2014, 09:29 Travis CaveConditioning StationStandard OperationTM2no targetmodule move

TM#2 has been moved from the south hot cell to the conditioning station.

  716   Monday, July 28, 2014, 14:37 David WangConditioning StationDevelopmentTM2 helium pressure test on TM2 tube heater +(D)and -(C)water lines.
  • TM2 installed in Conditioning Station
  • Varian Leak detector connected directly to Target Module 2 Service cap.
  • Vacuum draw in Target Station with both module turbo pumps
  • Leak detector baseline normal (10^-4 Torr, 10^-9 Std cc/s He)
  • Since no Source tray is installed, and tube heater supply conductor is blanked off in target enclosure, it was necessary to test both supply and return points.
  • For each line tested, 80 psi He was connected, and waited ~2 min to for a response on the leak detector.
  • Lines tested:
    • conductor D (tube heater), Supply -> no He response
    • conductor D (tube heater), Return -> no He response
    • conductor C (tube heater), Supply -> no He response
    • conductor C (tube heater), Return -> no He response
  • Based on this leak testing, both supply and return lines of conductors C & D are leak tight under 95 psi He (80 psig He supply to vacuum)

 

Attachment 1: 011.JPG
011.JPG
Attachment 2: 012.JPG
012.JPG
Attachment 3: 013.JPG
013.JPG
Attachment 4: 014.JPG
014.JPG
Attachment 5: 015.JPG
015.JPG
Attachment 6: 017.JPG
017.JPG
Attachment 7: 018.JPG
018.JPG
  717   Monday, July 28, 2014, 14:45 David WangConditioning StationDevelopmentTM2 turned off TM2 and CS turbo pumps.

I turned off TM2 and CS turbo-pumps. TM2 will be moved from CS to silo tomorrow morning.

  719   Wednesday, July 30, 2014, 11:51 Travis CaveConditioning StationMaintenance  entrance module move

ITE Entrance module has been moved from the east station to the conditioning station.

  727   Thursday, August 07, 2014, 17:07 David WangConditioning StationStandard OperationTM4UCx #10TM4 is connected in CS.

TM4 is connected at CS. All turbo pumps are started. The pumping down is good so far.

  728   Friday, August 08, 2014, 08:47 David WangConditioning StationStandard OperationTM4UCx #10helium pressure test on TM4 HS water line at CS.

Helium pressure test on TM4 HS water line at CS is finished. 80psi helium no leak. 

Attachment 1: 015.JPG
015.JPG
Attachment 2: 012.JPG
012.JPG
Attachment 3: 016.JPG
016.JPG
Attachment 4: 014.JPG
014.JPG
  729   Monday, August 11, 2014, 10:34 Travis CaveConditioning StationStandard OperationTM4UC#10module move

TM#4 with UC#10 has been moved from the conditioning station to the east target station.

ELOG V2.9.2-2455