M9sol Logbooks Solenoid Daily log
  Solenoid logbook, Page 8 of 15  ELOG logo
New entries since:Wed Dec 31 16:00:00 1969
ID Datedown Author Operational Mode Subject Text Attachments
  146   Fri Jul 21 23:36:21 2006 Doug PreddyQuench RecoveryFinal cooldown problems80K shield is down to <60K Liquid level not
building. Power leads rippling 
and warming. T1 inlet 85K with 10%heater
  
  145   Fri Jul 21 17:18:19 2006 Willy AnderssonQuench RecoveryOpened HIC142 fully openPower leads have flatlined but still building
LHe level.
  
  144   Fri Jul 21 14:08:04 2006 Doug PreddyQuench Recoveryswitch to series modeCoils temps <10K.Switching to series mode.   
  143   Fri Jul 21 07:39:37 2006 Doug PreddyQuench RecoveryCool down continuesRaised PIC229 to 11Bar. Started to close HIC155.
Cooling rates improve.
  
  142   Thu Jul 20 13:34:26 2006 Willy AnderssonQuench Recoverycold return mode continues...Even though cr_c1 and cr_c12 have begun to
tick down, cr_c7 and cr_c6 are 
actually ticking up. This is probably a wave
  
  141   Thu Jul 20 13:27:21 2006 Willy AnderssonQuench Recoverycold return mode continues...cr_c1 and cr_12 are both at about 45K and
heading slowly down, so the 
gamble of going to // cold return worked.
  
  140   Thu Jul 20 13:21:57 2006 Willy AnderssonQuench Recoverystarted // warm return flow.TIC121 was down to 27K
warm return valves were wide open
opened supply valves CV143 and CV146.
  
  139   Thu Jul 20 11:01:09 2006 Willy AnderssonQuench RecoveryOpened up warm return valves CV151 and CV153Mu2 pressure was just about at trip point
due to trapped volume of helium 
warming. So opened up warm returns to bleed
  
  138   Thu Jul 20 08:18:31 2006 Doug PreddyQuench RecoveryT1 startedStarted flow to turbines. Both started. T1
seems a bit slow. HIC110 at 0.85 
SI110 = 4100.
  
  137   Thu Jul 20 02:59:06 2006 Doug PreddyQuench RecoveryT1 still not startingI will leave the cold box to warm overnight
and try it again in the morning.
  
  136   Thu Jul 20 00:52:13 2006 Doug PreddyQuench RecoveryT1 frozenT1 will not spin up. Warming up cold box.   
  135   Thu Jul 20 00:25:16 2006 Doug PreddyQuench RecoverySolenoid quenched @ 11:15Solenoid Quenched on C1 temp. It went above
5K. Shut off compressor. 
Restarting system @ 12:15
  
  134   Thu Jul 6 15:04:17 2006 Doug PreddyNormal Runningtesttest  M9daily log070306.XLS 
  133   Thu Jul 6 15:00:15 2006 Doug PreddyNormal Runningtest    
  132   Mon Jul 3 23:01:34 2006 Doug PreddyQuench Recoverypower lead warming/ripples caused quenchpower lead warming/ripples caused quench at
18:45 today. Compressor stayed 
on, but flow to solenoid stopped.
  
  131   Fri Jun 30 11:17:57 2006 Doug PreddyNormal RunningSolenoid recoveredTuesday night. June 27, 2006, the solenoid
quenched for no apparent 
reason. It was restarted Wednesday morning.
  
  130   Fri Jun 30 11:13:00 2006 Doug PreddyNormal RunningE-log is back up and runningI would have made an entry when it died, but-----   
  129   Fri Jun 16 08:12:04 2006 Doug PreddyNormal RunningLI108 indicates high on EPICS screenLI108 indicates high on EPICS screen (150%)
 Controller shows ~64%
  
  128   Wed Jun 14 16:27:01 2006 Willy AnderssonQuench RecoveryEnergized solenoidPowerleads appear stable.   
  127   Wed Jun 14 10:35:30 2006 Willy AnderssonQuench RecoverySwitched to series cooling.Had system in //return cooling overnight with
small flow to keep power 
leads cool. Switched to series mode about
  