Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  Test-Stand, Page 26 of 53  Not logged in ELOG logo
ID Date Author Target/No.# Source Station Moduledown Target Oven W.O# Heat Shield W.O# Status Subject Text Attachments
  380   Tuesday, May 20, 2014, 12:39 Anders Mjos      DoneFEBIAD Cathode Negative BIAS configurationAs per ISK4562 Rev 3 and ISK4563 Rev 3
By Ray D. and co.
  
  383   Thursday, June 12, 2014, 09:28 Anders Mjos      Problems/SolutionsElevated backing pressure on ISAC Test Stand TMS:RP2

    

       
            
  
  385   Monday, June 16, 2014, 09:12 Anders Mjos      In progressEmpty chamberMonday, June 16, 2014, 09:12: Roughing

Tuesday, June 17, 2014, 14:38:
Venting 
  
  387   Wednesday, June 18, 2014, 09:23 Anders Mjos      Problems/SolutionsTSH:MB1 Trip & MB1 June 10th water leakA water leak on MB1 was found and reported
by operations on June 10th e fault 7429
MB1 can not be switched on due to
  
  388   Thursday, June 19, 2014, 10:48 Anders Mjos      Problems/SolutionsFC0 did not retract from beam lineUpon inspection it was found that the bolt
attaching the FC to the actuator had fallen
off. The bolt was reattached and FC0 can
  
  389   Friday, June 20, 2014, 11:44 Anders Mjos      In progressPumping down empty chamberFriday, June 20, 2014, 11:44: Roughing
Friday, June 20, 2014, 13:00: Strarting
turbos 
  
  391   Tuesday, July 08, 2014, 09:05 Anders Mjos      In progressEmpty chamberTuesday, July 08, 2014, 09:05: Roughing
Tuesday, July 08, 2014, 13:23: Starting
turbos
  
  392   Wednesday, July 09, 2014, 14:08 Anders Mjos      Problems/SolutionsFEBIAD ANODE current readbackThe read back of the ANODE current was
confirmed to be working today with help of
Ray D. using a 1kOhm resistor. There is an
  
  393   Friday, July 18, 2014, 08:19 Anders Mjos       VentingFriday, July 18, 2014, 08:19: Stopped turbos
Friday, July 18, 2014, 10:58: Venting
  
  395   Wednesday, July 23, 2014, 14:34 Anders Mjos      In progressEmpty chamberWednesday, July 23, 2014, 14:34: Roughing
Wednesday, July 23, 2014, 15:03: Starting
turbos.
  
  396   Friday, August 01, 2014, 08:39 Anders Mjos      Problems/SolutionsTMS:RP2 noisyFriday, August 01, 2014, 08:39: Backing
pressure 0 / 11 mT
  
  398   Tuesday, August 12, 2014, 09:52 Anders Mjos      Problems/SolutionsRTD installed on Supply waterTravis assisted in installing a temperature
reading RTD from Omega (SA1-RTD-MTP) on the
water supply. The RTD has been installed
 2014-08-12_Dataq_Screen_Shot.pngDataq_Supply_Temperature_calibration_2014-08-20.docx 
  399   Thursday, August 14, 2014, 08:45 Anders Mjos      Problems/SolutionsFC0 reading full currentsince yesterday around 19h. See attached
screen shot of archived data
Thursday, August 14, 2014, 08:40:
 FC0FullScaleReading_2014-08-14.pngFC0FullScaleReading_EPICS_2014-08-14.png 
  400   Wednesday, August 20, 2014, 16:25 Anders Mjos      Problems/SolutionsLost power on TGHT, TBHT and EE simultaniously Possible controls problem. Breaker 18 found
latched. Can not be re-energized. Burnt smell
in PLC / HV area. e-faut 7561
  
  401   Wednesday, August 27, 2014, 12:55 Anders Mjos      Problems/SolutionsWater leak on MB1e-faulted
by operations Aug 26th
  
  405   Tuesday, October 07, 2014, 09:32 Anders Mjos      In progressPumping down empty chamberTuesday, October 07, 2014, 09:30: Roughing
Tuesday, October 07, 2014, 10:24:
Starting turbos
  
  406   Monday, October 20, 2014, 09:17 Anders Mjos      Problems/SolutionsPower bump this weekend tripped off all turbo pumps Monday, October 20, 2014, 09:17:
Roughing
Monday, October 20, 2014, 09:18 :
  
  407   Monday, October 27, 2014, 09:26 Anders Mjos      Problems/SolutionsTSG:TP1B noisye-fault 7748
Monday, October 27, 2014, 09:21:
Switched off turbos
  
  408   Friday, October 31, 2014, 14:57 Anders Mjos LP-FEBIAD  Tube only38307-9In progressTestFriday, October 31, 2014, 14:47: Starting
turbos
Friday, October 31, 2014, 14:57:
 test20141103.pngtest20141103-2.png 
  409   Monday, November 03, 2014, 14:50 Anders Mjos      Problems/SolutionsIV4 did not close until 4th attemptIndicated with red valve on the EPICS vacuum
page
  