Friday, June 02, 2017, 15:20: Coil loose (Remove lid, Tighten and place lid again)
Friday, June 02, 2017, 15:52: Roughing
Friday, June 02, 2017, 16:13: Turbo pumps on.
Friday, June 02, 2017, 16:18: Ion gauge on
Friday, June 02, 2017, 16:23: RGA on
By Anders
Saturday, June 03, 2017, 09:07: Opened HS water supply. In/out 95 / 10 psi. 3.8 lpm. Vacuum stable at 3.3E-7 Torr
Saturday, June 03, 2017, 09:08: Open IV4
Saturday, June 03, 2017, 09:32: Opening gas bottles. Source on. Reconfigured in Cathode bias
Saturday, June 03, 2017, 09:41: Started autoramping to 580 / 250 A at 2 A/min
Saturday, June 03, 2017, 09:42: COIL at 70 A, ANODE at 200 V
Saturday, June 03, 2017, 09:42: BIAS / EL / EE at 30 / 13 / 1.5 kV
Saturday, June 03, 2017, 09:47
Daqpro time 9:33
Device |
Flow [lpm] |
Temp [C] |
HS + MSP |
3.8 |
26 |
TBHT |
3.1 |
27 |
COIL + EE |
4.9 |
26 |
TGHT |
5.5 |
26 |
Sunday, June 04, 2017, 06:52 : Source at TGHT / TBHT 580 / 290 A over night and outgassing while measuring mass 20. Both anode current andFC6 current increased whie vacuum decreased. Ramped TBHT up to 300 A. Anode current now 68 mA, FC6 0.2 nA, FC0 3 uA
Monday, June 05, 2017, 16:04: Starting cool down
Monday, June 05, 2017, 22:04: Source off. Stopped turbo pumps.
Tuesday, June 06, 2017, 07:28: RGA off.
Tuesday, June 06, 2017, 07:30: Venting |