Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  RH-ISAC, Page 108 of 140  Not logged in ELOG logo
ID Datedown Author Category Type Module Target/Number Subject
  656   Monday, May 05, 2014, 13:11 David WangITERepairTM3  ITE Entrance module has been disconnected.

ITE entrance module has been disconnected and the lifting yoke is installed for module move. See attached pictures.

Attachment 1: 011.JPG
011.JPG
Attachment 2: 025.JPG
025.JPG
Attachment 3: 033.JPG
033.JPG
Attachment 4: 035.JPG
035.JPG
Attachment 5: 032.JPG
032.JPG
  655   Monday, May 05, 2014, 13:05 David WangITERepairTM3 ITE target module flange seal has been replaced.

ITE target module flange seal has been replaced.The flan surface and the o-ring groove have been cleaned with Acetone before the new seal is installed. See attached picture.

Attachment 1: 004.JPG
004.JPG
Attachment 2: 008.JPG
008.JPG
Attachment 3: 008.JPG
008.JPG
Attachment 4: 007.JPG
007.JPG
  654   Monday, May 05, 2014, 10:40 Travis CaveITEStandard OperationTM3TiC#4module move

TM#3 with TiC#4 has been moved from the east station to the south west silo.

  653   Wednesday, April 30, 2014, 12:03 Travis CaveSpent Target VaultStandard Operation 2A3 windowWindow now in vault

The 2A3 window now has a lid, and lift yolk on it and it has been installed in the spent target vault in spot 2A. The lid is a normal unmodified and has no sealant on it.

Attachment 1: Vault_Storage_April_30_2014.pdf
Vault_Storage_April_30_2014.pdf Vault_Storage_April_30_2014.pdf Vault_Storage_April_30_2014.pdf Vault_Storage_April_30_2014.pdf
  652   Wednesday, April 16, 2014, 17:08 Keith C NgITERepair  2a3 window - He leak check

Helium leak checked 2a3 window after installing using remote line:

- There was a leak found between the target and entrance module at the target pit, rate of 9E-6 atm-cc/sec.

- One test performed with the remote helium line to the window. There was a response but the leak detector was located above the window access channel, drift cannot be ruled out.

- 3 more tests performed with vacuum group, with leak detector relocated farther away from ITE and isolated with plastic. 2 tests using remote line had no response. One test done with a wand through 2a3 wall penetration had a response of 3.2E-9 atm-cc/sec (not using remote line).

Vacuum group (D. Yosifov) comments that due to the much larger leak at the top of the tank he cannot say with confidence the response seen is a leak at the window (due to leak at top) but if there is, a tolerable leak rate.

  651   Wednesday, April 16, 2014, 10:16 Anders MjosITWDevelopmentTM3Ta#42Flow meters installed on Heat Shield and Coil water return

 

Anders Mjos wrote:

David W. has installed two SST Proteus flow meters, model OV7006FTA32, on ITW Heat Shield and Coil. The flow meters are connected to a power supply and a data logger in the electrical room.

The Heat Shield flow meter has SN 00278112 and is on the Black wire pairs and connected to CH3 & CH4 on the logger.

The Coil flow meter has SN 00278113 and is on the Green wire pairs and connected to CH1 & CH2 on the logger.

The first channel (CH1, CH3) is temperature in degrees Celcius and the second channel (CH2, CH4) is flow rate in liters per minute.

A camera has been placed on the logger.

Flow rates today (2014-04-04):

HS        8.2 - 8.3 lpm
COIL    8.0 - 8.1 lpm

Calibration sheets


UPDATE 2014-04-11: Corrected HS to CH3 & CH4.

 

PDFs of temperature and flow traces are available on DocuShare

Monday, April 28, 2014, 09:27. p+ off since 8:20

  Temp [C] Flow [lpm]
HS 28.3 8.38
COIL 26.3 8.01

Tuesday, April 29, 2014, 09:28: p+ 30 uA

  Temp [C] Flow [lpm]
HS 31.7 8.40
COIL 27.6 8.13

Monday, May 05, 2014, 10:49: p+ off since 8:39
(Downloaded data and updated traces on DocuShare)

  Temp [C] Flow [lpm]
HS 27.1 8.47
COIL 26.6 8.11

 Wednesday, May 14, 2014
(Downloaded data and updated traces on DocuShare)

  Temp [C] Flow [lpm]
HS 33.5 7.7
COIL 29.0 7.5

 

  650   Tuesday, April 15, 2014, 08:53 Travis CaveITEStandard OperationTM3TiC#4module move

TM#3 with TiC#4 has been moved from the conditioning station to the east target station.

  649   Monday, April 14, 2014, 13:58 David WangITERepair  Reconnected ITE entrance module

I reconnected the entrance module back to the ITE station. All diagnostic cables, turbo- pump, air lines, water lines on ITE entrance module and BL2A3 window are connected back.

  648   Monday, April 14, 2014, 11:57 Travis CaveITEStandard Operation  entrance module move

The East Entrance module has been moved from the south west silo to the east target station. The move went well.

  647   Friday, April 11, 2014, 18:17 Keith C NgITERepair  2a3 window - install

Today:

1) Installed new cradle and shortened beam pipe section in wall penetration.

- Removed tank dust cover from flange and cleaned

- Held flange faces together from 2A tunnel side to check alignment, slightly biased to the East axially and a small gap present in the top when resting

 

2) Prepared new 2A window

- Connected preswaged  (1/2 turn) swagelok nuts to the existing copper lines in the target hall, 1/4 turn from finger tight, 3/4 turns total

- Leak checked cooling lines with pressurized air at 75psi and snoop, no leaks were found at the joint

- Grooves were cleaned and helicoflex delta seals were installed with vacuum grease used in the periphery to hold ring in window, applied at 6 points every 60deg

 

3) Window installation

- Lowered window into position by cooling lines

- Window located into tank flange first then slid beam pipe towards it, capturing the window with seals

- Lowered clamp into place in hole

 

4) Clamp installation

- Lowered clamp into access hole and started closing clamp with steel RH drive bar

- torqued to 50 ft-lbs collar gap was still ~8"

- Increased torque to 60 ft-lbs then 70 ft-lbs stopped closing of clamp when the displacement did not change between the stop ring and clamp yoke

- It was decided not to continue to close this clamp, more assessment may be necessary

 

6) We will attempt to pump on the seal in its current configuration, tools have been removed.

  646   Friday, April 11, 2014, 11:35 David WangConditioning StationStandard OperationTM3 Helium pressure test on TM3 HS water line with new Ta target. at CS

The helium pressure test on TM3 HS water line with new Ta target is finished at CS. The circuit is leak tight.. No any response on LD. See pictures.

Attachment 1: 027.JPG
027.JPG
Attachment 2: 026.JPG
026.JPG
Attachment 3: 025.JPG
025.JPG
Attachment 4: 022.JPG
022.JPG
Attachment 5: 024.JPG
024.JPG
Attachment 6: 023.JPG
023.JPG
  645   Thursday, April 10, 2014, 14:03 Bevan MossConditioning StationStandard OperationTM3TiC#4HV conditioning

 The TiC#4 target that is on TM3 was conditioned to 25kV today with the EINZEL lens at 12.5kV. There was one spark shortly after reaching 25kV that reset the bias PS and another one  after the module sat at 25kV for ~30 minutes. Attached is the strip tool data and screen shot. Prior to starting the conditioning the power on the chiller needed to by cycled as the VFD on the fans had tripped due to a recurring inrush problem.

Attachment 1: 10042014_Tm3_TiC#4_HV_conditioning.JPG
10042014_Tm3_TiC#4_HV_conditioning.JPG
Attachment 2: 10042014_TM3_TiC#4_HV_conditioning.txt
Time	TCS:IG1:RDVAC [T]	TCS:IG2:RDVAC [T]	TCS:IG1S:RDVAC [T]	TCS:EL:RDVOL [V]	TCS:EL:RDCUR [mA]	TCS:BIAS:RDVOL [V]	TCS:BIAS:RDCUR [uA]	TCS:DIC1:RDCOND [Mohm]	
04/10/2014 10:51:30.980156	                      0.000000961                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:31.979337	                      0.000000959                      0.000005934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:32.979181	                      0.000000959                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:33.979125	                      0.000000959                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:34.979530	                      0.000000957                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:35.979903	                      0.000000957                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:36.979928	                      0.000000959                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:37.979724	                      0.000000959                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:38.980475	                      0.000000957                      0.000005890                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:39.980322	                      0.000000959                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:41.721703	                      0.000000957                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:42.720980	                      0.000000959                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:43.721829	                      0.000000959                      0.000005912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:44.721709	                      0.000000959                      0.000005934                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:45.721877	                      0.000000959                      0.000005890                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:46.721604	                      0.000000959                      0.000005912                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:47.721360	                      0.000000957                      0.000005912                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:48.722248	                      0.000000959                      0.000005890                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:49.722132	                      0.000000954                      0.000005890                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:50.722206	                      0.000000954                      0.000005912                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:51.722849	                      0.000000959                      0.000005912                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:52.722691	                      0.000000959                      0.000005934                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:53.723567	                      0.000000959                      0.000005934                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:54.723552	                      0.000000957                      0.000005912                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:55.723346	                      0.000000957                      0.000005934                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:56.723198	                      0.000000957                      0.000005934                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:57.723128	                      0.000000961                      0.000005934                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:58.723980	                      0.000000961                      0.000005912                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:51:59.723852	                      0.000000957                      0.000005912                      0.000003890                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:00.723765	                      0.000000959                      0.000005912                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:01.723774	                      0.000000959                      0.000005890                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:02.723712	                      0.000000959                      0.000005912                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:03.723618	                      0.000000957                      0.000005890                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:04.724220	                      0.000000957                      0.000005890                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:05.724345	                      0.000000959                      0.000005890                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:06.724243	                      0.000000957                      0.000005890                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:07.723847	                      0.000000959                      0.000005912                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:08.723737	                      0.000000957                      0.000005912                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:09.734748	                      0.000000957                      0.000005890                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:10.734468	                      0.000000961                      0.000005912                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:11.734336	                      0.000000961                      0.000005890                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:12.734202	                      0.000000957                      0.000005934                      0.000003912                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:13.735274	                      0.000000959                      0.000005934                      0.000003934                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:14.736049	                      0.000000959                      0.000005890                      0.000003934                             6.10                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:15.735844	                      0.000000961                      0.000005890                      0.000003934                             6.10                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:16.735707	                      0.000000959                      0.000005912                      0.000003934                             6.71                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:17.735550	                      0.000000959                      0.000005912                      0.000003934                             6.71                           BadVal                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:18.735469	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:19.735397	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:20.736220	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:21.737118	                      0.000000961                      0.000005912                      0.000003934                             6.10                             0.01                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:22.736946	                      0.000000961                      0.000005912                      0.000003912                             6.10                             0.01                           BadVal                           BadVal                           BadVal
04/10/2014 10:52:23.736842	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                           BadVal                           BadVal
04/10/2014 10:52:24.736718	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                           BadVal                           BadVal
04/10/2014 10:52:25.736582	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                           BadVal                           BadVal
04/10/2014 10:52:26.735970	                      0.000000959                      0.000005912                      0.000003912                             6.71                             0.01                            18.31                           BadVal                           BadVal
04/10/2014 10:52:27.764214	                      0.000000959                      0.000005912                      0.000003934                             6.71                             0.01                            18.31                           BadVal                           BadVal
04/10/2014 10:52:28.764296	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:29.764046	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            18.31                             0.18                           BadVal
04/10/2014 10:52:30.763942	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:31.764595	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:32.764669	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:33.765611	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:34.765460	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:35.765345	                      0.000000961                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:36.765196	                      0.000000959                      0.000005912                      0.000003912                             6.71                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:37.765304	                      0.000000959                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:38.765944	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                           BadVal
04/10/2014 10:52:39.765800	                      0.000000959                      0.000005890                      0.000003956                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:40.765665	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:41.766003	                      0.000000954                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:42.766445	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:43.767346	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:44.767203	                      0.000000961                      0.000005890                      0.000003890                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:52:45.767105	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:52:46.766994	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:47.766939	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:48.851745	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:49.877773	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:50.877440	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:51.877267	                      0.000000959                      0.000005912                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:52.877165	                      0.000000959                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:53.877016	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:54.877020	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:52:55.876788	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:52:56.876639	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:57.876545	                      0.000000961                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:52:58.876493	                      0.000000961                      0.000005912                      0.000003912                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:52:59.876303	                      0.000000959                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:00.876207	                      0.000000954                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:01.876057	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:02.973167	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:53:04.127504	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:05.245403	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:06.245504	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:07.246554	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:08.246272	                      0.000000961                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:09.246145	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:10.245957	                      0.000000959                      0.000005890                      0.000003868                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:11.245855	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:12.246803	                      0.000000959                      0.000005912                      0.000003934                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:53:13.246602	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:14.246883	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:15.247430	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:16.248242	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:17.248123	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:18.248018	                      0.000000959                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:19.247819	                      0.000000959                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:20.248760	                      0.000000957                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:21.248576	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:22.248523	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:23.248379	                      0.000000959                      0.000005890                      0.000003868                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:24.248306	                      0.000000961                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:25.248120	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:26.247992	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:27.247865	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:28.247725	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:29.248587	                      0.000000957                      0.000005912                      0.000003868                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:53:30.248506	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:53:31.477726	                      0.000000961                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:32.510989	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:33.570288	                      0.000000961                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:34.569946	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:35.569810	                      0.000000954                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:36.569693	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:37.569517	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:38.570683	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:39.570275	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:40.570189	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:53:41.663248	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:42.796751	                      0.000000961                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:43.796790	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:53:44.796660	                      0.000000961                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:53:45.797531	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:53:46.797366	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:47.963516	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:48.963095	                      0.000000959                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:49.964000	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:53:50.963885	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:51.963082	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:52.962598	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:53.963488	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:54.964455	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:53:55.964257	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:56.964088	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:53:57.964030	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:58.964837	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:53:59.964711	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:00.964585	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:01.964460	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:02.964438	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:03.965278	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            18.31                             0.18                             13.2
04/10/2014 10:54:04.965134	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:06.114629	                      0.000000959                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:07.114842	                      0.000000959                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:08.114859	                      0.000000959                      0.000005890                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:09.115601	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:11.95567	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:12.95203	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:13.95062	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:14.94930	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:15.95034	                      0.000000954                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:16.94756	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:17.94640	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:18.94482	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:19.94499	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:20.94191	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:21.94057	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:22.94970	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:23.94867	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:24.95670	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:25.95555	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:26.95473	                      0.000000959                      0.000005934                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:27.95345	                      0.000000957                      0.000005890                      0.000003868                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:28.95171	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:29.156966	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:30.219271	                      0.000000954                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:31.218823	                      0.000000959                      0.000005868                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:32.218700	                      0.000000954                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:33.218572	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:34.219484	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:35.219324	                      0.000000954                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:36.219155	                      0.000000954                      0.000005890                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:37.219030	                      0.000000954                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:38.218939	                      0.000000957                      0.000005868                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:39.219905	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:40.219644	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:41.219569	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:42.219471	                      0.000000954                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:43.219276	                      0.000000954                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:44.220207	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:45.226449	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:46.225901	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:47.225793	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:48.225688	                      0.000000959                      0.000005890                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:49.237062	                      0.000000954                      0.000005890                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:50.237414	                      0.000000954                      0.000005868                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:51.237318	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:52.237170	                      0.000000957                      0.000005868                      0.000003868                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:53.237085	                      0.000000959                      0.000005890                      0.000003890                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:54.237035	                      0.000000957                      0.000005890                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:55.236818	                      0.000000959                      0.000005890                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:56.236684	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:57.236508	                      0.000000957                      0.000005868                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:54:58.236539	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:54:59.237346	                      0.000000954                      0.000005934                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:00.238278	                      0.000000957                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:01.238061	                      0.000000957                      0.000005868                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:02.237902	                      0.000000957                      0.000005868                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:03.237965	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:04.238630	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:05.239504	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:06.239438	                      0.000000954                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:07.239297	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:08.240205	                      0.000000959                      0.000005890                      0.000003890                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:55:09.240127	                      0.000000959                      0.000005890                      0.000003890                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:10.240229	                      0.000000954                      0.000005890                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:11.240963	                      0.000000957                      0.000005890                      0.000003868                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:12.242313	                      0.000000959                      0.000005868                      0.000003868                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:13.242643	                      0.000000959                      0.000005890                      0.000003868                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:14.243588	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:15.243253	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:16.243399	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:17.243056	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:18.243878	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:19.244834	                      0.000000954                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:20.245696	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:21.245500	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:22.245413	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:23.245446	                      0.000000954                      0.000005912                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:24.246172	                      0.000000954                      0.000005912                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:25.247101	                      0.000000954                      0.000005912                      0.000003912                             6.71                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:55:26.246944	                      0.000000957                      0.000005912                      0.000003890                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:27.246745	                      0.000000957                      0.000005890                      0.000003956                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:28.246707	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:29.247622	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:30.247664	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:31.247309	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:32.247107	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:33.247041	                      0.000000959                      0.000005890                      0.000003868                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:34.247898	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:35.248757	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:36.248622	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:37.248512	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:38.248407	                      0.000000954                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:39.248278	                      0.000000954                      0.000005868                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:40.248215	                      0.000000954                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:41.248104	                      0.000000959                      0.000005912                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:42.247859	                      0.000000959                      0.000005868                      0.000003890                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:43.247775	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:44.248699	                      0.000000959                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:55:45.248591	                      0.000000954                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:46.248414	                      0.000000954                      0.000005890                      0.000003934                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:55:47.248231	                      0.000000954                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:48.248200	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:50.356306	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:51.356770	                      0.000000954                      0.000005934                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:52.356658	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:53.356479	                      0.000000957                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:55.692384	                      0.000000959                      0.000005912                      0.000003868                             6.71                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:55:56.693040	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:57.693001	                      0.000000959                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:58.692827	                      0.000000959                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:55:59.692705	                      0.000000954                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:00.693597	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:01.693469	                      0.000000959                      0.000005912                      0.000003934                             6.10                             0.01                            18.31                             0.18                             13.1
04/10/2014 10:56:02.693310	                      0.000000959                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:03.693156	                      0.000000959                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:04.693036	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            18.31                             0.18                             13.2
04/10/2014 10:56:05.692994	                      0.000000957                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:06.692862	                      0.000000957                      0.000005890                      0.000003934                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:07.693748	                      0.000000954                      0.000005912                      0.000003912                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:08.693600	                      0.000000954                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:09.694444	                      0.000000957                      0.000005868                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:10.695298	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:11.695160	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:12.695091	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:13.694987	                      0.000000954                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:14.694827	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:15.694742	                      0.000000959                      0.000005912                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:16.694531	                      0.000000959                      0.000005934                      0.000003890                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:17.694420	                      0.000000959                      0.000005890                      0.000003890                             6.71                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:18.694323	                      0.000000957                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:19.694196	                      0.000000954                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:20.694200	                      0.000000954                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:21.693936	                      0.000000957                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:22.693796	                      0.000000957                      0.000005890                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:23.693687	                      0.000000959                      0.000005890                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:24.693613	                      0.000000957                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:25.693505	                      0.000000959                      0.000005868                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:26.693314	                      0.000000957                      0.000005868                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:27.693140	                      0.000000957                      0.000005890                      0.000003890                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:28.693023	                      0.000000957                      0.000005912                      0.000003934                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:29.692952	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:30.693831	                      0.000000954                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:31.693711	                      0.000000957                      0.000005934                      0.000003956                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:32.694613	                      0.000000959                      0.000005912                      0.000003912                             6.10                             0.01                            17.16                             0.18                             13.1
04/10/2014 10:56:33.694463	                      0.000000957                      0.000005912                      0.000003890                             6.71                             0.01                            17.16                             0.18                             13.2
04/10/2014 10:56:34.694275	                      0.000000957                      0.000005868                      0.000003890                             6.71                             0.01                            17.16                             0.18                             13.2
... 7596 more lines ...
  644   Wednesday, April 09, 2014, 11:10 David WangSouth Hot-CellStandard OperationTM3 TM3 heat shield circuit leack check at SHC

This Monday morning, Chad and I leak checked heat shielding circuit on TM3 with new target installed. The base LR on LD :0.0xE-9 atm.cc/sec, base pressure 0.0xE-4 torr. Chad sprayed helium twice on the VCR joints. No leak has been found. The circuit will be pressured with helium at CS.

  643   Tuesday, April 08, 2014, 09:05 Travis CaveSouth Hot-CellStandard OperationTM3TiC#4module move

TM#3 with TiC#4 has been moved from the south hot cell to the conditioning station.

  642   Monday, April 07, 2014, 14:13 chad fisherSouth Hot-CellStandard OperationTM3TiC#4TiC#4 installation

TiC#4 has been installed onto TM3

  641   Monday, April 07, 2014, 11:49 Travis CaveSouth Hot-CellStandard OperationTM3TiC#4Electrical check

Electrical test of TM#3 with TiC#4 attached. See PDF.

Attachment 1: TM#3_elec_test_TiC#4_surf.PDF
TM#3_elec_test_TiC#4_surf.PDF
  640   Monday, April 07, 2014, 10:48 Travis CaveSouth Hot-CellStandard OperationTM3no targetElectrical check

TM#3 electrical check at the south hot  cell with no target to give us a base line and show that things are ok with the module for the start of the running year.

Attachment 1: TM#3_elec_test_no_target.PDF
TM#3_elec_test_no_target.PDF
  639   Friday, April 04, 2014, 16:59 Anders MjosITWDevelopment  Flow meters installed on Heat Shield and Coil water return

David W. has installed two SST Proteus flow meters, model OV7006FTA32, on ITW Heat Shield and Coil. The flow meters are connected to a power supply and a data logger in the electrical room.

The Heat Shield flow meter has SN 00278112 and is on the Black wire pairs and connected to CH3 & CH4 on the logger.

The Coil flow meter has SN 00278113 and is on the Green wire pairs and connected to CH1 & CH2 on the logger.

The first channel (CH1, CH3) is temperature in degrees Celcius and the second channel (CH2, CH4) is flow rate in liters per minute.

A camera has been placed on the logger.

Flow rates today (2014-04-04):

HS        8.2 - 8.3 lpm
COIL    8.0 - 8.1 lpm

Calibration sheets


UPDATE 2014-04-11: Corrected HS to CH3 & CH4.

Attachment 1: 20140404_165829_1.jpg
20140404_165829_1.jpg
  638   Friday, April 04, 2014, 14:59 Keith C NgITEMaintenance  measurement of new beam tube section

R. Kuramoto and A. Laxdal measured the distance from the 2A tunnel to the tank flange of ITE. The new fabricated beam pipe with bellows was used to gauge the distance to the tank flange; The flange face of the beam tube was butted up to the tank flange and the beam tube was marked at the plane of the wall. A metal rod was used to gauge the distance to the tank face using the same method.

From the inside of the tunnel wall:

- To tank flange: 45 5/16"

- To ITE vacuum tank face: 49"

See attached PDF.

Attachment 1: wall_to_flange_length.pdf
wall_to_flange_length.pdf
  637   Friday, April 04, 2014, 14:16 Travis CaveSouth Hot-CellStandard OperationTM3no targetmodule move

TM#3 with no target has been moved from the south west silo to the south hot cell.

ELOG V2.9.2-2455