Entry  Wed Jun 14 10:35:30 2006, Willy Andersson, Quench Recovery, Switched to series cooling. 
Had system in //return cooling overnight with small flow to keep power 
leads cool. Switched to series mode about 30 minutes ago. Building LHe 
level.
Entry  Mon Oct 5 11:54:58 2009, Doug Preddy, Normal Running, Switched TIC121 into auto mode 
 
Entry  Fri Jun 25 14:20:58 2004, Willy Andersson, Quench Recovery, Switch to series mode 
At about 14:00 solenoid temps dipping below 8 K. Trace of LHe level (dense 
gas) starting to show. Switching to series cooling mode.
Entry  Thu Sep 2 21:58:34 2004, Willy Andersson, Quench Recovery, Switch to series cooling 
Solenoid had tripped on powerlead overtemp at about 09:00 this morning. 
Doug recovered the system to // cooling. I switched system to series 
cooling at about 21:30. LHe level at about 10%. Buffer tank pressure is low 
at about 13 psig. Letting LHe level build. Once LHe level rises to about 
35% then I suggest starting flow to the powerleads by opening HIC142 fully 
open. LHe level should continue to build but at a much slower rate. Once 
power lead temps have bottomed out to cr_pl_neg < 9K and cr_pl_pos < 11 K 
(check these values with archived data first) and are stable then can 
energize solenoid.
Entry  Tue Aug 14 12:55:12 2007, Doug Preddy, Normal Running, Strip tool saves 
There are two strip tools saved:
coils
powerlead
Entry  Thu Oct 26 04:34:15 2006, Willy Andersson, Quench Recovery, Still in // cold return mode. 
Leaving system cooling in // cold return mode, cooldown rates still good. 
Had flow on to powerleads since re-starting cooling flow to solenoid.
Either I (when I return to work later this morning) or Doug will do the 
switch to series mode.
Entry  Tue Apr 1 15:17:06 2008, Doug Preddy, Start Up, Starting up vacuum systems 
Monday afternoon there was no control of the vacuum system. The controls were working, but no output to devices. The "output" breaker in the PLC box was found to be tripped. Reseting it allowed control to the vacuum systems. I started the regeneration of BCP2 this morning, but hte output breaker tripped again. reset it and continued to pump. Started to pump out the solenoid also. Removed jumper on the BDP2 temperature read-back, bypassed BV2, and BCG2. This allows pumping through the diffussion pump. Cold box has been pumping for a few weeks. It's at 2.0 x 10-3
Entry  Thu Apr 29 23:43:15 2004, Doug Preddy, Quench Recovery, Starting flow to solenoid 
Solenoid quenched due to power leads at 9:00 tonight. Restarted flow in 
parallel warm mode. 
Entry  Wed Jul 28 22:01:31 2004, Doug Preddy, Quench Recovery, Starting flow to solenoid 
c1 28, c12 29, c6 11.6, c7 12.5 TI121 22.5
Entry  Wed Oct 4 07:29:04 2006, Doug Preddy, Start Up, Starting cool down 
Started compressor yesterday with flow through the purifier overnight.
Starting tubines this morning in parallel warm return
Entry  Mon Apr 19 15:06:49 2004, Doug Preddy, Start Up, Started turbines april191500.BMP
Purifier run over the weekend. Both LN2 dewars empty, but purifier cold. 
Disabled CV211 and flowed all charge gas through purifier. Started turbine 
at 11:00am. set strip tool to monitor temps. Started M9BCP2. Pumped out 
beamline between BV% and BV7. Opened BV6.
Entry  Mon Oct 4 09:16:29 2004, Doug Preddy, Start Up, Started turbines 
Started flow to the turbimes with the solenoid in parallel warm return.
Entry  Tue May 23 08:44:28 2006, Doug Preddy, Start Up, Started turbines 
Solenoid and cold box were pumped out last Tuesday. Cold box roughing pump 
initially would not start. had to retighten all wires on the contactor in 
the Sulzer control panel (RP900).
Started compressor on Friday. ran purifier over the weekend.
Started flow in parallel warm return. Started flow to turbines.
Replacing dead monitor on other control computer.
Entry  Tue Apr 24 10:21:49 2007, Doug Preddy, Start Up, Started flow to turbines 
Started flow to turbines at about 10:00 in parallel warm mode.
Entry  Fri May 26 10:03:03 2006, Doug Preddy, Start Up, Started flow to power leads 
Started flow to power leads.
Liquid level 15%
Entry  Mon Sep 24 15:08:41 2007, Doug Preddy, Start Up, Started cool down this morning 
The vacuum in the cold box was good on Friday 0.0 X 10-4. Let it pump over 
the weekend. Added two bottles of UHP helium to system Friday.

Started the turbines this morning. Cooling in paralell warm return. 

Solenoid vacuum is 3.9 x 10-6
Entry  Mon Mar 29 09:25:52 2010, Kevin Trithardt, Start Up, Started cool down this morning 
 
Entry  Mon Apr 23 16:31:02 2007, Doug Preddy, Start Up, Started compressor 
Startewd compressor and purifer today.
Entry  Wed Mar 24 08:30:33 2010, Doug Preddy, Start Up, Start up 

Compressor started yesterday afternoon. Pumping out solenoid overnight. Cold box vacuum good. circulating helium to solenoid in Parallel warm mode, CV142 open. Purifier running.

Entry  Thu Aug 17 08:04:44 2006, Doug Preddy, Quench Recovery, Solenoid tripped at about 3:00 this morning 
restarting system.
Compressor would not start. Idid not see anything that would not let it 
start. Pushed rest button upstairs, and reset contactor. Compressor started.
preparing to start flow to solenoid in // cold return.