Wed Jun 14 10:35:30 2006, Willy Andersson, Quench Recovery, Switched to series cooling.
|
Had system in //return cooling overnight with small flow to keep power
leads cool. Switched to series mode about 30 minutes ago. Building LHe
level. |
Mon Oct 5 11:54:58 2009, Doug Preddy, Normal Running, Switched TIC121 into auto mode
|
|
Fri Jun 25 14:20:58 2004, Willy Andersson, Quench Recovery, Switch to series mode
|
At about 14:00 solenoid temps dipping below 8 K. Trace of LHe level (dense
gas) starting to show. Switching to series cooling mode. |
Thu Sep 2 21:58:34 2004, Willy Andersson, Quench Recovery, Switch to series cooling
|
Solenoid had tripped on powerlead overtemp at about 09:00 this morning.
Doug recovered the system to // cooling. I switched system to series
cooling at about 21:30. LHe level at about 10%. Buffer tank pressure is low
|
Tue Aug 14 12:55:12 2007, Doug Preddy, Normal Running, Strip tool saves
|
There are two strip tools saved:
coils
powerlead |
Thu Oct 26 04:34:15 2006, Willy Andersson, Quench Recovery, Still in // cold return mode.
|
Leaving system cooling in // cold return mode, cooldown rates still good.
Had flow on to powerleads since re-starting cooling flow to solenoid.
Either I (when I return to work later this morning) or Doug will do the
|
Tue Apr 1 15:17:06 2008, Doug Preddy, Start Up, Starting up vacuum systems
|
Monday afternoon there was no control of the vacuum system. The controls were working, but no output to devices. The "output" breaker in the PLC box was
found to be tripped. Reseting it allowed control to the vacuum systems.
I started the regeneration of BCP2 this morning, but hte output breaker tripped
again. reset it and continued to pump.
Started to pump out the solenoid also. Removed jumper on the BDP2 temperature read-back, bypassed BV2, and BCG2. |
Thu Apr 29 23:43:15 2004, Doug Preddy, Quench Recovery, Starting flow to solenoid
|
Solenoid quenched due to power leads at 9:00 tonight. Restarted flow in
parallel warm mode. |
Wed Jul 28 22:01:31 2004, Doug Preddy, Quench Recovery, Starting flow to solenoid
|
c1 28, c12 29, c6 11.6, c7 12.5 TI121 22.5 |
Wed Oct 4 07:29:04 2006, Doug Preddy, Start Up, Starting cool down
|
Started compressor yesterday with flow through the purifier overnight.
Starting tubines this morning in parallel warm return |
Mon Apr 19 15:06:49 2004, Doug Preddy, Start Up, Started turbines
|
Purifier run over the weekend. Both LN2 dewars empty, but purifier cold.
Disabled CV211 and flowed all charge gas through purifier. Started turbine
at 11:00am. set strip tool to monitor temps. Started M9BCP2. Pumped out
|
Mon Oct 4 09:16:29 2004, Doug Preddy, Start Up, Started turbines
|
Started flow to the turbimes with the solenoid in parallel warm return. |
Tue May 23 08:44:28 2006, Doug Preddy, Start Up, Started turbines
|
Solenoid and cold box were pumped out last Tuesday. Cold box roughing pump
initially would not start. had to retighten all wires on the contactor in
the Sulzer control panel (RP900).
|
Tue Apr 24 10:21:49 2007, Doug Preddy, Start Up, Started flow to turbines
|
Started flow to turbines at about 10:00 in parallel warm mode. |
Fri May 26 10:03:03 2006, Doug Preddy, Start Up, Started flow to power leads
|
Started flow to power leads.
Liquid level 15% |
Mon Sep 24 15:08:41 2007, Doug Preddy, Start Up, Started cool down this morning
|
The vacuum in the cold box was good on Friday 0.0 X 10-4. Let it pump over
the weekend. Added two bottles of UHP helium to system Friday.
|
Mon Mar 29 09:25:52 2010, Kevin Trithardt, Start Up, Started cool down this morning
|
|
Mon Apr 23 16:31:02 2007, Doug Preddy, Start Up, Started compressor
|
Startewd compressor and purifer today. |
Wed Mar 24 08:30:33 2010, Doug Preddy, Start Up, Start up
|
Compressor started yesterday afternoon. Pumping out solenoid overnight. Cold box vacuum good. circulating helium to solenoid in Parallel warm mode, CV142
open. Purifier running. |
Thu Aug 17 08:04:44 2006, Doug Preddy, Quench Recovery, Solenoid tripped at about 3:00 this morning
|
restarting system.
Compressor would not start. Idid not see anything that would not let it
start. Pushed rest button upstairs, and reset contactor. Compressor started.
|