ID Date Author Operational Mode Subjectdown
  127   Wed Jun 14 10:35:30 2006 Willy AnderssonQuench RecoverySwitched to series cooling.
Had system in //return cooling overnight with small flow to keep power 
leads cool. Switched to series mode about 30 minutes ago. Building LHe 
level.
  273   Mon Oct 5 11:54:58 2009 Doug PreddyNormal RunningSwitched TIC121 into auto mode
  30   Fri Jun 25 14:20:58 2004 Willy AnderssonQuench RecoverySwitch to series mode
At about 14:00 solenoid temps dipping below 8 K. Trace of LHe level (dense 
gas) starting to show. Switching to series cooling mode.
  47   Thu Sep 2 21:58:34 2004 Willy AnderssonQuench RecoverySwitch to series cooling
Solenoid had tripped on powerlead overtemp at about 09:00 this morning. 
Doug recovered the system to // cooling. I switched system to series 
cooling at about 21:30. LHe level at about 10%. Buffer tank pressure is low 
at about 13 psig. Letting LHe level build. Once LHe level rises to about 
35% then I suggest starting flow to the powerleads by opening HIC142 fully 
open. LHe level should continue to build but at a much slower rate. Once 
power lead temps have bottomed out to cr_pl_neg < 9K and cr_pl_pos < 11 K 
(check these values with archived data first) and are stable then can 
energize solenoid.
  215   Tue Aug 14 12:55:12 2007 Doug PreddyNormal RunningStrip tool saves
There are two strip tools saved:
coils
powerlead
  172   Thu Oct 26 04:34:15 2006 Willy AnderssonQuench RecoveryStill in // cold return mode.
Leaving system cooling in // cold return mode, cooldown rates still good. 
Had flow on to powerleads since re-starting cooling flow to solenoid.
Either I (when I return to work later this morning) or Doug will do the 
switch to series mode.
  243   Tue Apr 1 15:17:06 2008 Doug PreddyStart UpStarting up vacuum systems
Monday afternoon there was no control of the vacuum system. The controls were working, but no output to devices. The "output" breaker in the PLC box was found to be tripped. Reseting it allowed control to the vacuum systems. I started the regeneration of BCP2 this morning, but hte output breaker tripped again. reset it and continued to pump. Started to pump out the solenoid also. Removed jumper on the BDP2 temperature read-back, bypassed BV2, and BCG2. This allows pumping through the diffussion pump. Cold box has been pumping for a few weeks. It's at 2.0 x 10-3
  19   Thu Apr 29 23:43:15 2004 Doug PreddyQuench RecoveryStarting flow to solenoid
Solenoid quenched due to power leads at 9:00 tonight. Restarted flow in 
parallel warm mode. 
  36   Wed Jul 28 22:01:31 2004 Doug PreddyQuench RecoveryStarting flow to solenoid
c1 28, c12 29, c6 11.6, c7 12.5 TI121 22.5
  162   Wed Oct 4 07:29:04 2006 Doug PreddyStart UpStarting cool down
Started compressor yesterday with flow through the purifier overnight.
Starting tubines this morning in parallel warm return
  5   Mon Apr 19 15:06:49 2004 Doug PreddyStart UpStarted turbines
Purifier run over the weekend. Both LN2 dewars empty, but purifier cold. 
Disabled CV211 and flowed all charge gas through purifier. Started turbine 
at 11:00am. set strip tool to monitor temps. Started M9BCP2. Pumped out 
beamline between BV% and BV7. Opened BV6.
  52   Mon Oct 4 09:16:29 2004 Doug PreddyStart UpStarted turbines
Started flow to the turbimes with the solenoid in parallel warm return.
  117   Tue May 23 08:44:28 2006 Doug PreddyStart UpStarted turbines
Solenoid and cold box were pumped out last Tuesday. Cold box roughing pump 
initially would not start. had to retighten all wires on the contactor in 
the Sulzer control panel (RP900).
Started compressor on Friday. ran purifier over the weekend.
Started flow in parallel warm return. Started flow to turbines.
Replacing dead monitor on other control computer.
  190   Tue Apr 24 10:21:49 2007 Doug PreddyStart UpStarted flow to turbines
Started flow to turbines at about 10:00 in parallel warm mode.
  121   Fri May 26 10:03:03 2006 Doug PreddyStart UpStarted flow to power leads
Started flow to power leads.
Liquid level 15%
  220   Mon Sep 24 15:08:41 2007 Doug PreddyStart UpStarted cool down this morning
The vacuum in the cold box was good on Friday 0.0 X 10-4. Let it pump over 
the weekend. Added two bottles of UHP helium to system Friday.

Started the turbines this morning. Cooling in paralell warm return. 

Solenoid vacuum is 3.9 x 10-6
  275   Mon Mar 29 09:25:52 2010 Kevin TrithardtStart UpStarted cool down this morning
  189   Mon Apr 23 16:31:02 2007 Doug PreddyStart UpStarted compressor
Startewd compressor and purifer today.
  274   Wed Mar 24 08:30:33 2010 Doug PreddyStart UpStart up

Compressor started yesterday afternoon. Pumping out solenoid overnight. Cold box vacuum good. circulating helium to solenoid in Parallel warm mode, CV142 open. Purifier running.

  160   Thu Aug 17 08:04:44 2006 Doug PreddyQuench RecoverySolenoid tripped at about 3:00 this morning
restarting system.
Compressor would not start. Idid not see anything that would not let it 
start. Pushed rest button upstairs, and reset contactor. Compressor started.
preparing to start flow to solenoid in // cold return.