112
Thu Nov 3 06:19:50 2005
Doug Preddy
Quench Recovery
restart system
Last night at about 7:00 there was a large
power bump that shut the
compressor and water pump off. I came in
113
Thu Nov 3 09:27:16 2005
Doug Preddy
Quench Recovery
started flow to solenoid
started flow to solenoid in parallel cold
return.
114
Tue Nov 15 07:25:10 2005
Doug Preddy
Normal Running
power supply problem
Saturday November 12 the solenoid power supply
ramped itself down to zero.
It would not ramp up unless the ramp rate
115
Mon Dec 12 09:24:29 2005
Doug Preddy
Quench Recovery
Solenoid Quench on power lead temp
The solenoid tripped on power lead temps at
about 9:15. I am recovering it
from it's all around mode into parallel cold
116
Thu Dec 22 15:10:44 2005
Doug Preddy
Shut Down
shut down system
shuttting down the system. isolated cold box
by closing v135. preparing to
pump out cold box.
117
Tue May 23 08:44:28 2006
Doug Preddy
Start Up
Started turbines
Solenoid and cold box were pumped out last
Tuesday. Cold box roughing pump
initially would not start. had to retighten
118
Thu May 25 07:34:18 2006
Doug Preddy
Start Up
switched to parallel cold return
switched to parallel cold return
119
Fri May 26 07:29:54 2006
Doug Preddy
Start Up
CV155 closed
CV155 closed.
Temps about 15K
120
Fri May 26 08:45:38 2006
Doug Preddy
Start Up
Changed to series mode
Changed to series mode.
Liquid level building.
121
Fri May 26 10:03:03 2006
Doug Preddy
Start Up
Started flow to power leads
Started flow to power leads.
Liquid level 15%
122
Fri May 26 14:34:24 2006
Doug Preddy
Start Up
Ready to run
Solenoid is cold, liquid level is 35%, power
leads cold and seem stable.
Added one bottle helium to buffer tank. Pressure
123
Tue May 30 14:36:41 2006
Doug Preddy
Normal Running
power leads warming
power leads warming. closed HIC 142, then
re-opened. temps falling
124
Tue Jun 13 10:54:27 2006
Doug Preddy
Quench Recovery
Compressor tripped off
No indicator lights etc as to why the compressor
tripped. The only light
was a cold box fault. I had to shut off the
129
Fri Jun 16 08:12:04 2006
Doug Preddy
Normal Running
LI108 indicates high on EPICS screen
LI108 indicates high on EPICS screen (150%)
Controller shows ~64%
130
Fri Jun 30 11:13:00 2006
Doug Preddy
Normal Running
E-log is back up and running
I would have made an entry when it died, but-----
131
Fri Jun 30 11:17:57 2006
Doug Preddy
Normal Running
Solenoid recovered
Tuesday night. June 27, 2006, the solenoid
quenched for no apparent
reason. It was restarted Wednesday morning.
132
Mon Jul 3 23:01:34 2006
Doug Preddy
Quench Recovery
power lead warming/ripples caused quench
power lead warming/ripples caused quench at
18:45 today. Compressor stayed
on, but flow to solenoid stopped.
133
Thu Jul 6 15:00:15 2006
Doug Preddy
Normal Running
test
134
Thu Jul 6 15:04:17 2006
Doug Preddy
Normal Running
test
test
135
Thu Jul 20 00:25:16 2006
Doug Preddy
Quench Recovery
Solenoid quenched @ 11:15
Solenoid Quenched on C1 temp. It went above
5K. Shut off compressor.
Restarting system @ 12:15