| ID | Date | Author | Operational Mode | Subject | Text |  | 
| 186 | Mon Mar  5 14:59:41 2007 | Doug Preddy | Maintenance | Password | On the weekend someone used this log to send spam messages. I have
 installed a password for the write privilege.
 |  | 
| 185 | Fri Jan 26 15:10:37 2007 | Doug Preddy | Maintenance | new control computer installed | I have installed a new control computer at the M9 solenoid control panel
 |  | 
| 184 | Tue Nov 28 10:32:04 2006 | Doug Preddy | Quench Recovery | recovering after looong power outage | The power has been out for about 24 hours. Recovering coldbox, solenoid and
 beamline vacuum.PLC is down for the vacuum
 |  | 
| 183 | Wed Nov 15 17:05:49 2006 | Willy Andersson | Quench Recovery | powerleads cold, energizing solenoid | Powerlead temperatures have bottomed out, energizing solenoid to 300
 amps, ramp rate 60.
 |  | 
| 182 | Wed Nov 15 13:56:00 2006 | Doug Preddy | Quench Recovery | switch to series mode |  |  | 
| 181 | Wed Nov 15 11:09:28 2006 | Doug Preddy | Quench Recovery | Compressor tripped off | Compressor tripped this morning at ~8:45. I restarted the compressor, but
 turbine 1 would not spin up. I let it warm
 |  | 
| 180 | Tue Nov  7 10:35:32 2006 | Doug Preddy | Normal Running | Maintenance day | I ramped down the solenoid power suplly and shut off flow to the power
 leads. I will allow them to warm up and (hopefully)
 |  | 
| 179 | Fri Nov  3 12:09:51 2006 | Doug Preddy | Normal Running | Fine tuning | The power lead temperatures have been climbing at a rate of ~ 0.02K very 8
 hours. The negative lead started to show
 |  | 
| 178 | Tue Oct 31 08:37:34 2006 | Doug Preddy | Quench Recovery | Energizing solenoid |  |  | 
| 177 | Tue Oct 31 05:36:14 2006 | Doug Preddy | Quench Recovery | switch to series mode |  |  | 
| 176 | Tue Oct 31 03:52:29 2006 | Doug Preddy | Quench Recovery | power lead quench at 2:30 this morning | Compressor still on. Starting flow to solenoid |  | 
| 175 | Thu Oct 26 14:14:57 2006 | Doug Preddy | Quench Recovery | power lead quench | solenoid tripped on power lead temp. I was here and caught it. System
 cooling.
 |  | 
| 174 | Thu Oct 26 11:45:22 2006 | Doug Preddy | Quench Recovery | Solenoid cold | power leads cold 50% liquid. Turning on power supply
 |  | 
| 173 | Thu Oct 26 08:18:10 2006 | Doug Preddy | Quench Recovery | switched to series mode | Switched to series mode at about 7:45. |  | 
| 172 | Thu Oct 26 04:34:15 2006 | Willy Andersson | Quench Recovery | Still in // cold return mode. | Leaving system cooling in // cold return mode, cooldown rates still good.
 Had flow on to powerleads since re-starting
 |  | 
| 171 | Thu Oct 26 04:27:31 2006 | Willy Andersson | Quench Recovery | Powerlead overtemp | 02:30 this morning got page from operations. Looked at system online, appears that there
 was a powerlead overtemp
 |  | 
| 170 | Sun Oct 15 07:09:41 2006 | Other | Normal Running | cryopump bcp2 trip | Donald Arseneau reporting 
 Not really solenoid realated, but FYI...
 |  | 
| 169 | Fri Oct 13 00:13:57 2006 | Experimentor | Quench Recovery | Energized solenoid. | Hi there, 
 This is Khashayar (experimenter). Lookin
 |  | 
| 168 | Thu Oct 12 19:54:10 2006 | Willy Andersson | Quench Recovery | Energized solenoid. | Switched to series mode about 6 hours ago. Cooling continued including
 power leads until LHe level at about 50%
 |  | 
| 167 | Thu Oct 12 11:53:04 2006 | Willy Andersson | Quench Recovery | Compressor trip, 09:50 this morning. | Main compressor tripped, refrigerator shutdown. Checked with experimentors
 and main control room, no indications of
 |  |