ID Date Author Operational Mode Subject Text Attachments
  186   Mon Mar 5 14:59:41 2007 Doug PreddyMaintenancePasswordOn the weekend someone used this log to send
spam messages. I have 
installed a password for the write privilege.
  
  185   Fri Jan 26 15:10:37 2007 Doug PreddyMaintenancenew control computer installedI have installed a new control computer at
the M9 solenoid control panel
  
  184   Tue Nov 28 10:32:04 2006 Doug PreddyQuench Recoveryrecovering after looong power outageThe power has been out for about 24 hours.
Recovering coldbox, solenoid and 
beamline vacuum.PLC is down for the vacuum
  
  183   Wed Nov 15 17:05:49 2006 Willy AnderssonQuench Recoverypowerleads cold, energizing solenoidPowerlead temperatures have bottomed out,
energizing solenoid to 300 
amps, ramp rate 60.
  
  182   Wed Nov 15 13:56:00 2006 Doug PreddyQuench Recoveryswitch to series mode    
  181   Wed Nov 15 11:09:28 2006 Doug PreddyQuench RecoveryCompressor tripped offCompressor tripped this morning at ~8:45.
I restarted the compressor, but 
turbine 1 would not spin up. I let it warm
  
  180   Tue Nov 7 10:35:32 2006 Doug PreddyNormal RunningMaintenance dayI ramped down the solenoid power suplly and
shut off flow to the power 
leads. I will allow them to warm up and (hopefully)
  
  179   Fri Nov 3 12:09:51 2006 Doug PreddyNormal RunningFine tuningThe power lead temperatures have been climbing
at a rate of ~ 0.02K very 8 
hours. The negative lead started to show
  
  178   Tue Oct 31 08:37:34 2006 Doug PreddyQuench RecoveryEnergizing solenoid    
  177   Tue Oct 31 05:36:14 2006 Doug PreddyQuench Recoveryswitch to series mode    
  176   Tue Oct 31 03:52:29 2006 Doug PreddyQuench Recoverypower lead quench at 2:30 this morningCompressor still on. Starting flow to solenoid   
  175   Thu Oct 26 14:14:57 2006 Doug PreddyQuench Recoverypower lead quenchsolenoid tripped on power lead temp. I was
here and caught it. System 
cooling.
  
  174   Thu Oct 26 11:45:22 2006 Doug PreddyQuench RecoverySolenoid coldpower leads cold 50% liquid.
Turning on power supply
  
  173   Thu Oct 26 08:18:10 2006 Doug PreddyQuench Recoveryswitched to series modeSwitched to series mode at about 7:45.   
  172   Thu Oct 26 04:34:15 2006 Willy AnderssonQuench RecoveryStill in // cold return mode.Leaving system cooling in // cold return mode,
cooldown rates still good. 
Had flow on to powerleads since re-starting
  
  171   Thu Oct 26 04:27:31 2006 Willy AnderssonQuench RecoveryPowerlead overtemp02:30 this morning got page from operations.
Looked at system online, appears that there
was a powerlead overtemp
  
  170   Sun Oct 15 07:09:41 2006 OtherNormal Runningcryopump bcp2 tripDonald Arseneau reporting

Not really solenoid realated, but FYI...
  
  169   Fri Oct 13 00:13:57 2006 ExperimentorQuench RecoveryEnergized solenoid.Hi there,

This is Khashayar (experimenter). Lookin
  
  168   Thu Oct 12 19:54:10 2006 Willy AnderssonQuench RecoveryEnergized solenoid.Switched to series mode about 6 hours ago.
Cooling continued including 
power leads until LHe level at about 50%
  
  167   Thu Oct 12 11:53:04 2006 Willy AnderssonQuench RecoveryCompressor trip, 09:50 this morning.Main compressor tripped, refrigerator shutdown.
Checked with experimentors 
and main control room, no indications of
  