3
|
Fri Dec 19 13:59:53 2003 |
Doug Preddy | Maintenance | To-Do list for M9SOL |
> I had distributed this list by email earlier this month. I have added
> a few items at the bottom. I am posting it on the M9SOL Elog under
> Maintenance. Some of the items may not be achievable or practical but
> I thought it would be a useful first attempt at a To-Do list.
A couple of other things to add:
Rebuild cooling water circulation pump with new seals and bearings.
Add air bleeds to the copper nonactive water lines.
Doug |
|
28
|
Fri May 21 09:27:39 2004 |
Doug Preddy | Maintenance | Idle mode |
Power supply was ramped down yesterday as there are no users.
The Sundance server was rebooted yesterday. This morning the displays were
all white screened, so I restarted the session. The computer was running
very slow and I could not get the strip tool to run at all. I rebooted the
local PC and everything seems to be okay now.
While rebooting the computer had an error message that the hard drive may
have some bad sectors. Did a scan disk and it seems okay, but the hard
drive might be dying in the local machine. |
|
51
|
Fri Oct 1 11:39:55 2004 |
Doug Preddy | Maintenance | Loading of compressor not working |
When the compressor was restarted on Monday, the remote load button would
not respond. The compressor was shut off. After two days of searching the
wiring etc. a bad connection was found on DR1 relay (upstairs in the
compressor control panel). The wire was the "compressor on" sensor.
Connection was cleaned and retightened.
The compressor was started this morning and is working.
I pressurized the cold box and started flow to the solenoid and through the
power leads. I started a continuous flow throughthe LN2 adsorber. This will
be run over the weekend. Cooling to start Monday morning. |
|
91
|
Fri Mar 11 09:18:30 2005 |
Doug Preddy | Maintenance | services on system |
The cold box was isolated and pumped out after it warmed up. It has been
refilled with helium.
Turbine 2 speed indicator failed at some time during the last run. After
the turbine inlet vavle was closed the turbine continued to read full
speed. I am removing the controller to look at it. The main power has been
turned of for this procedure. |
|
92
|
Mon Apr 18 09:52:11 2005 |
Doug Preddy | Maintenance | Relays replaced |
The following relays have been replaced in the Sulzer control panel:
K341, K379, K132, K324A, K314, K125, K127, K133, K252, K227, K233, K238,
K239, K245, K246.
These were the one that had indications of sever rusting inside the case.
There are five more spare relays in the gray cupboard. |
|
93
|
Thu May 19 15:05:31 2005 |
Doug Preddy | Maintenance | STA 120 replaced |
The turbine 2 speed indicator control box STA 120 has been replaced with a
newer style Jaquet controller. This is similar to the ones used in the
ISAC helium liquifier. See attached photos for set up parameters. These
were taken from the old controller, Sulzer documentation, and the ISAC 2
controllers. |
|
185
|
Fri Jan 26 15:10:37 2007 |
Doug Preddy | Maintenance | new control computer installed |
I have installed a new control computer at the M9 solenoid control panel |
|
186
|
Mon Mar 5 14:59:41 2007 |
Doug Preddy | Maintenance | Password |
On the weekend someone used this log to send spam messages. I have
installed a password for the write privilege. Please contact me at 6399
for the password.
Doug |
|
196
|
Wed May 16 10:02:35 2007 |
Doug Preddy | Maintenance | warming up power leads |
Power lead temps have been slowly increasing over the last wekk. I ramped
the solenoid down and am warming the power leads to >80k. |
|
201
|
Tue Jun 12 12:48:25 2007 |
Doug Preddy | Maintenance | added one bottle UHP helium to system |
|
|
214
|
Tue Aug 14 10:15:22 2007 |
Doug Preddy | Maintenance | Solenoid ramped down |
Solenoid ramped down and locked out for M9A slits work. |
|
216
|
Wed Sep 12 12:01:02 2007 |
Doug Preddy | Maintenance | regenerated adsorber |
regenerated adsorber
replaced MV902 "ON" switch (used nitrogen precooling on switch) |
|
218
|
Tue Sep 18 18:48:11 2007 |
Curtis Ballard | Maintenance | cold box leak check |
The leak check of the cold box diff pump fittings. Found a leak at the
base of the diff pump on the tee that is apparently used to fill the pump
with oil. The fitting was removed and a plug was removed that resembled a
dip stick (approx 1" long). Tbe dip stick was charred black as if it had
been hot. The seal between the dip stick and the tee was not an O-ring.
It looked like a high temperature glue? For leak checking purposes, an O-
ring was fitted to the dip stick and reinserted in situ. Only the top
portion of the tee was removed. The lower cap was slathered in vacuum
grease. Pumping back down was slow. After two hours the vacuum was down
to 2 torr. At 6:30 pm, the vacuum was 200 mtorr, so I turned on the
filament and tried again to leak check. I was able to get the throttle
valve open a little wider than previously, but we still had a big leak. I
checked at the tee and got a strong response, but I also checked at the
diff pump connections and also got a response, albeit much less. We may
have to blank off the entire diff pump to isolate the cold box so that we
can leak check the system. Dimo said he will be able to help tomorrow.
For now, I will leave the roughing pump on the cold box and isolte the
leak detector. See you tomorrow.
P.S. I left a voice mail re: welding at TITAN 8:30 am/blankets/work
permits/extractor. cb |
|
219
|
Wed Sep 19 20:53:18 2007 |
Doug Preddy | Maintenance | More diff pump news |
Cold box pumped out to ~ 190 mTorr overnight.Leak checked the pump again.
Still leaks at dip stick.Tried resetting the o-ring on the dip stick. It
still leaks baddd!!!
Machined the old seal off and installed a smaller o-ring onto dip stick.
Installed 60 ml of new pump oil into pump. Pumping down again.
After 4 hours the cold box is down to 160 mTorr. |
|
229
|
Wed Oct 24 12:04:06 2007 |
Doug Preddy | Maintenance | power lead and cryopump regeneration |
Regenerating M9B:CP2 as temperature was up to almost 20k.
Shut off flow to power leads to allow them to warm up. The positive and
negative temps were beginning to creep up.
Regeneraing the adsorber. |
|
233
|
Fri Nov 9 16:24:33 2007 |
Doug Preddy | Maintenance | 2APLC Died |
thumb wheel panel all lite up. 2APLC rebooted by Klara (Thanks). M9BMP2 not
pumping properly. Edi is looking at it. |
|
234
|
Fri Nov 9 18:56:20 2007 |
Doug Preddy | Maintenance | System Ready to Run |
M9BMP2 was replaced. Vacuum good. Solenoid is ramped up. :) |
|
258
|
Tue Sep 9 15:04:23 2008 |
Doug Preddy | Maintenance | Jumpered interlock for diff pump water flow |
| Jumpered interlock for diff pump water flow in Proteus box. |
|
259
|
Fri Sep 26 09:44:25 2008 |
Doug Preddy | Maintenance | Water flow interlock restored |
| yesterday |
|
266
|
Wed Dec 10 15:44:58 2008 |
Doug Preddy | Maintenance | Repairs to solenoid power supply |
| Klaus and company pulled the ramp rate card out of the solenoid power supply. Operational checks were done and the rate switch was cleaned. The card was reinserted and the ramp rate was checked. The power supply seems to ramp at a proper rate in all positions. ***Note*** The output voltage of the power supply is proportional to the ramp rate. The faster the ramp rate the higher the output voltage. |