Message ID: 3
Entry time: Sat Apr 9 14:57:30 2011
|
Author: |
Tom Davinson |
Type: |
|
Category: |
|
Subject: |
MIDAS DAQ configuration |
|
|
tuda2 (by TUDA chamber, SEBT1) was powered off on arrival
- this may have been due to a previous power outage between
March 18-April 4. Rebooted OK.
MIDAS DAQ operational
MIDAS DAQ reads out CAEN V560 Scaler #1 channels #0-7 (see
file VMEDataSources below) on a per event basis.
Data directory tuda2:/data2/data/S1287
S1287 requires 64+64+32+16=176 channels of ADCs/TDCs
i.e. 6x Silena 9418/6V ADCs and 2x CAEN V1190A TDCs.
Currently configured for 11x ADCs, 3x TDCs and 2x Scalers.
Note - VME crate #2 (top) displays a yellow 'overheat'
warning LED. This is probably bogus - the air and PSU
temperatures reported by the front panel look OK.
MIDAS DAQ works for mode 'MIDAS Data Acquisition' and
'TUDA'. For the latter, all scalers and the EPICS variables
(defined by tuda:/MIDAS/TUDA/tcl/EpicsDefn.tcl - see below)
are readout every 20s. |
|
SAC sac#1 vme 0x0400:0x000000:0x00000000:0:0 1 0 0 0 0 0 1 {} 1
{Silena S9418} s9418#1 vme 0x0000:0x000000:0x02000000:0:0 1 0 0 0 0 0 1 {} 1
{Silena S9418} s9418#2 vme 0x0000:0x000000:0x02010000:0:0 2 0 0 0 0 0 1 {} 2
{Silena S9418} s9418#3 vme 0x0000:0x000000:0x02020000:0:0 3 0 0 0 0 0 1 {} 3
{Silena S9418} s9418#4 vme 0x0000:0x000000:0x02030000:0:0 4 0 0 0 0 0 1 {} 4
{Silena S9418} s9418#5 vme 0x0000:0x000000:0x02040000:0:0 5 0 0 0 0 0 1 {} 5
{Silena S9418} s9418#6 vme 0x0000:0x000000:0x02050000:0:0 6 0 0 0 0 0 1 {} 6
{Silena S9418} s9418#7 vme 0x0000:0x000000:0x02060000:0:0 7 0 0 0 0 0 1 {} 7
{Silena S9418} s9418#8 vme 0x0000:0x000000:0x02070000:0:0 8 0 0 0 0 0 1 {} 8
{Silena S9418} s9418#9 vme 0x0000:0x000000:0x02080000:0:0 9 0 0 0 0 0 1 {} 9
{CAEN V1190} v1190#1 vme 0x0000:0x000000:0x01700000:0:0 1 0 0 0 0 0 1 {} 1
{CAEN V1190} v1190#2 vme 0x0000:0x000000:0x01710000:0:0 2 0 0 0 0 0 1 {} 2
{CAEN V1190} v1190#3 vme 0x0000:0x000000:0x01720000:0:0 3 0 0 0 0 0 1 {} 3
{CAEN V560} v560#1 vme 0x0000:0x000000:0x01000000:0:0 1 0 0 0 0 0 1 {} 1
{CAEN V560} v560#2 vme 0x0000:0x000000:0x01100000:0:0 2 0 0 0 0 0 1 {} 2
{Silena S9418} s9418#10 vme 0x0000:0x000000:0x02090000:0:0 10 0 0 0 0 0 1 {} 10
{Silena S9418} s9418#11 vme 0x0000:0x000000:0x020a0000:0:0 11 0 0 0 0 0 1 {} 11
|
|
|
## sys
v1190_GroupBase 20
Include.VME 1
Include.V1190 1
v1190.channels 2048
v879_GroupBase 32
v767.shift 0
v1290.offset 0
v879.offset 0
Include.V767 1
Histogram.V767 1
v767.offset 0
v878.offset 0
Include.V775 1
Include.V1290 1
s9418.shift 0
Histogram.V775 1
Histogram.V1190 1
s9418.offset 0
Rate.channels 4096
Stat.channels 4096
v785.shift 0
v878.channels 4096
s9418_GroupBase 1
vme.channels 8192
s9418.channels 4096
Include.S9418 1
Include.V785 1
v1290.channels 4096
v775.shift 0
Histogram.V785 1
v775.offset 0
v785.offset 0
v792_GroupBase 32
v775_GroupBase 48
v879.channels 4096
v1290_GroupBase 24
Include.V792 1
v879.shift 0
v775.channels 4096
v767.channels 2048
Histogram.V792 1
vme.offset 0
Histogram.V1290 1
Include.V878 1
Include.V879 1
v767_GroupBase 28
Histogram.V878 1
v792.channels 4096
Histogram.V879 1
vme.shift 3
Histogram.S9418 1
v878.shift 0
Histogram.VME 1
v785.channels 4096
v878_GroupBase 32
v792.offset 0
v785_GroupBase 32
v792.shift 0
v1190.shift 0
v1290.shift 4
v1190.offset 0
##
|
|
32 0x01000010 32 30 0
32 0x01000014 32 30 2
32 0x01000018 32 30 4
32 0x0100001c 32 30 6
32 0x01000020 32 30 8
32 0x01000024 32 30 10
32 0x01000028 32 30 12
32 0x0100002c 32 30 14
|
|
AcqMde02 1
Trig16 40 -20 8 4 1
Edge23 2
Res26 0
Dtime29 0
RdOut32 0
RdOut34 9
RdOut3c 7
Enble45 -1 -1 -1 -1 -1 -1 -1 -1
Off51 0 0
|
|
|
|
AcqMde02 1
Trig16 40 -20 8 4 1
Edge23 1
Res26 0
Dtime29 0
RdOut32 0
RdOut34 9
RdOut3c 7
Enble45 -1 -1 -1 -1 -1 -1 -1 -1
Off51 0 0
|
|
|
# format of an EPICS item
#
# EPICS parameter name
# first item number to use for data
# format of EPICS parameter 0 = 32 bit floating number
# 1 = 32 bit integer
# 2 = 16 bit integer
set EPICS {}
lappend EPICS {SEBT1:CHAN1CI:SCALECUR 0 0}
lappend EPICS {SEBT1:CHAN2CI:SCALECUR 2 0}
#lappend EPICS {SEBT1:CI1:CONV 4 1}
#lappend EPICS {SEBT1:CI1:RANGE 6 1}
#lappend EPICS {SEBT1:CI1:ALIVE 8 1}
lappend EPICS {SEBT1:CHAN2CI:FULLSCALE 4 1}
#lappend EPICS {SEBT1:CHAN3CI:SCALECUR 12 0}
#lappend EPICS {SEBT1:CHAN4CI:SCALECUR 14 0}
#lappend EPICS {SEBT1:CHAN5CI:SCALECUR 16 0}
#lappend EPICS {SEBT1:CHAN6CI:SCALECUR 18 0}
#lappend EPICS {SEBT1:CI2:CONV 20 0}
#lappend EPICS {SEBT1:CI2:RANGE 22 0}
#lappend EPICS {SEBT1:CI2:ALIVE 24 1}
#lappend EPICS {SEBT1:CHAN6CI:FULLSCALE 26 1}
#lappend EPICS {SEBT1:CHAN7CI:SCALECUR 28 0}
#lappend EPICS {SEBT1:CHAN8CI:SCALECUR 30 0}
#lappend EPICS {SEBT1:CHAN9CI:SCALECUR 32 0}
#lappend EPICS {SEBT1:CHAN10CI:SCALECUR 34 0}
#lappend EPICS {SEBT1:CI3:CONV 36 0}
#lappend EPICS {SEBT1:CI3:RANGE 38 0}
#lappend EPICS {SEBT1:CI3:ALIVE 40 1}
#lappend EPICS {SEBT1:CHAN10CI:FULLSCALE 42 1}
#lappend EPICS {SEBT1:BIASON 44 0}
lappend EPICS {TUDA2:PNG5:RDVAC 6 0}
lappend EPICS {CCS2ISAC:BL2ACURRENT 8 0}
set env(EPICS_CA_ADDR_LIST) 142.90.133.255
set env(EPICS_CA_AUTO_ADDR_LIST) YES
set env(EPICS_CA_REPEATER_PORT) 9003
set env(EPICS_CA_SERVER_PORT) 9004
set env(PATH) /MIDAS/epics/bin:$env(PATH)
set env(LD_LIBRARY_PATH) /MIDAS/epics/lib:/usr/sfw/lib:$env(LD_LIBRARY_PATH)
# run caget to conditionally launch caRepeater
set z [catch {exec caget CCS2ISAC:BL2ACURRENT} m]
puts "EPICS initialise returned code=$z and information \"$m\""
|
|
#device configuration information
# any line starting with a # is a comment and is ignored
stats ./stats
# devices available - list ends with a null line or a line starting with a &
# format class_name device_name driver_task_path_name
# class_name MUST be one of dlt, exabyte, scsitape, sink
dlt /dev/rmt/1 ./driver
exabyte /dev/rmt/0 ./driver
file /dev/file/0 ./driver
sink /dev/null/0 ./driver
&
#data link configuration information - list end with a null line
# format link_task_path_name
./linkTCP
&
#program options - list end with a null line
msg_reporting_level 0x0180fff8
# use 0x0080fff8 to enable msg logging
msg_logging_level 0xfff8
tapeserver_options 2
# default for following is 16Kbytes - both MUST be the same at present
data_buffer_size 32
tape_block_size 32
disc_file_size 2000
#
file_path_base /data2/data
&
#end of information
|