Subject:
> > tuda2 (by TUDA chamber, SEBT1) was powered off on arrival > - this may have been due to a previous power outage between > March 18-April 4. Rebooted OK. > > MIDAS DAQ operational > > MIDAS DAQ reads out CAEN V560 Scaler #1 channels #0-7 (see > file VMEDataSources below) on a per event basis. > > Data directory tuda2:/data2/data/S1287 > > S1287 requires 64+64+32+16=176 channels of ADCs/TDCs > i.e. 6x Silena 9418/6V ADCs and 2x CAEN V1190A TDCs. > Currently configured for 11x ADCs, 3x TDCs and 2x Scalers. > > Note - VME crate #2 (top) displays a yellow 'overheat' > warning LED. This is probably bogus - the air and PSU > temperatures reported by the front panel look OK. > > MIDAS DAQ works for mode 'MIDAS Data Acquisition' and > 'TUDA'. For the latter, all scalers and the EPICS variables > (defined by tuda:/MIDAS/TUDA/tcl/EpicsDefn.tcl - see below) > are readout every 20s.