Tuda logbooks General E1104 ISACII E1195 E1103 E1147 S1287 S1203 S1233 S1284 S1447 23Na(a,p) S1540 S1381  S1847 Catania
  S1287  ELOG logo
Message ID: 3     Entry time: Sat Apr 9 14:57:30 2011
Author: Tom Davinson 
Type:  
Category:  
Subject: MIDAS DAQ configuration 
tuda2 (by TUDA chamber, SEBT1) was powered off on arrival
- this may have been due to a previous power outage between
March 18-April 4. Rebooted OK.

MIDAS DAQ operational

MIDAS DAQ reads out CAEN V560 Scaler #1 channels #0-7 (see
file VMEDataSources below) on a per event basis.

Data directory tuda2:/data2/data/S1287

S1287 requires 64+64+32+16=176 channels of ADCs/TDCs
i.e. 6x Silena 9418/6V ADCs and 2x CAEN V1190A TDCs. 
Currently configured for 11x ADCs, 3x TDCs and 2x Scalers. 

Note - VME crate #2 (top) displays a yellow 'overheat'
warning LED. This is probably bogus - the air and PSU
temperatures reported by the front panel look OK.

MIDAS DAQ works for mode 'MIDAS Data Acquisition' and
'TUDA'. For the latter, all scalers and the EPICS variables
(defined by tuda:/MIDAS/TUDA/tcl/EpicsDefn.tcl - see below)
are readout every 20s.
Attachment 1: latest  1 kB  Uploaded Sat Apr 9 17:33:34 2011  | Hide | Hide all | Show all
SAC sac#1 vme 0x0400:0x000000:0x00000000:0:0 1 0 0 0 0 0 1 {} 1
{Silena S9418} s9418#1 vme 0x0000:0x000000:0x02000000:0:0 1 0 0 0 0 0 1 {} 1
{Silena S9418} s9418#2 vme 0x0000:0x000000:0x02010000:0:0 2 0 0 0 0 0 1 {} 2
{Silena S9418} s9418#3 vme 0x0000:0x000000:0x02020000:0:0 3 0 0 0 0 0 1 {} 3
{Silena S9418} s9418#4 vme 0x0000:0x000000:0x02030000:0:0 4 0 0 0 0 0 1 {} 4
{Silena S9418} s9418#5 vme 0x0000:0x000000:0x02040000:0:0 5 0 0 0 0 0 1 {} 5
{Silena S9418} s9418#6 vme 0x0000:0x000000:0x02050000:0:0 6 0 0 0 0 0 1 {} 6
{Silena S9418} s9418#7 vme 0x0000:0x000000:0x02060000:0:0 7 0 0 0 0 0 1 {} 7
{Silena S9418} s9418#8 vme 0x0000:0x000000:0x02070000:0:0 8 0 0 0 0 0 1 {} 8
{Silena S9418} s9418#9 vme 0x0000:0x000000:0x02080000:0:0 9 0 0 0 0 0 1 {} 9
{CAEN V1190} v1190#1 vme 0x0000:0x000000:0x01700000:0:0 1 0 0 0 0 0 1 {} 1
{CAEN V1190} v1190#2 vme 0x0000:0x000000:0x01710000:0:0 2 0 0 0 0 0 1 {} 2
{CAEN V1190} v1190#3 vme 0x0000:0x000000:0x01720000:0:0 3 0 0 0 0 0 1 {} 3
{CAEN V560} v560#1 vme 0x0000:0x000000:0x01000000:0:0 1 0 0 0 0 0 1 {} 1
{CAEN V560} v560#2 vme 0x0000:0x000000:0x01100000:0:0 2 0 0 0 0 0 1 {} 2
{Silena S9418} s9418#10 vme 0x0000:0x000000:0x02090000:0:0 10 0 0 0 0 0 1 {} 10
{Silena S9418} s9418#11 vme 0x0000:0x000000:0x020a0000:0:0 11 0 0 0 0 0 1 {} 11
Attachment 2: latest  9 kB  Uploaded Sat Apr 9 17:33:55 2011  | Show | Hide all | Show all
Attachment 3: MidasOptions  1002 Bytes  Uploaded Sat Apr 9 17:34:15 2011  | Hide | Hide all | Show all
## sys
v1190_GroupBase 20
Include.VME 1
Include.V1190 1
v1190.channels 2048
v879_GroupBase 32
v767.shift 0
v1290.offset 0
v879.offset 0
Include.V767 1
Histogram.V767 1
v767.offset 0
v878.offset 0
Include.V775 1
Include.V1290 1
s9418.shift 0
Histogram.V775 1
Histogram.V1190 1
s9418.offset 0
Rate.channels 4096
Stat.channels 4096
v785.shift 0
v878.channels 4096
s9418_GroupBase 1
vme.channels 8192
s9418.channels 4096
Include.S9418 1
Include.V785 1
v1290.channels 4096
v775.shift 0
Histogram.V785 1
v775.offset 0
v785.offset 0
v792_GroupBase 32
v775_GroupBase 48
v879.channels 4096
v1290_GroupBase 24
Include.V792 1
v879.shift 0
v775.channels 4096
v767.channels 2048
Histogram.V792 1
vme.offset 0
Histogram.V1290 1
Include.V878 1
Include.V879 1
v767_GroupBase 28
Histogram.V878 1
v792.channels 4096
Histogram.V879 1
vme.shift 3
Histogram.S9418 1
v878.shift 0
Histogram.VME 1
v785.channels 4096
v878_GroupBase 32
v792.offset 0
v785_GroupBase 32
v792.shift 0
v1190.shift 0
v1290.shift 4
v1190.offset 0
##
Attachment 4: VMEDataSources  179 Bytes  Uploaded Sat Apr 9 17:34:34 2011  | Show | Hide all | Show all
Attachment 5: v1190#1  148 Bytes  Uploaded Sat Apr 9 17:34:54 2011  | Hide | Hide all | Show all
AcqMde02 1 
Trig16 40 -20 8 4 1 
Edge23  2 
Res26  0 
Dtime29  0 
RdOut32  0 
RdOut34  9 
RdOut3c  7 
Enble45  -1 -1 -1 -1 -1 -1 -1 -1 
Off51  0 0 
Attachment 6: v1190#2  148 Bytes  Uploaded Sat Apr 9 17:35:11 2011  | Hide | Hide all | Show all
AcqMde02 1 
Trig16 40 -20 8 4 1 
Edge23  2 
Res26  0 
Dtime29  0 
RdOut32  0 
RdOut34  9 
RdOut3c  7 
Enble45  -1 -1 -1 -1 -1 -1 -1 -1 
Off51  0 0 
Attachment 7: v1190#3  148 Bytes  Uploaded Sat Apr 9 17:35:27 2011  | Hide | Hide all | Show all
AcqMde02 1 
Trig16 40 -20 8 4 1 
Edge23  2 
Res26  0 
Dtime29  0 
RdOut32  0 
RdOut34  9 
RdOut3c  7 
Enble45  -1 -1 -1 -1 -1 -1 -1 -1 
Off51  0 0 
Attachment 8: v1190#4  148 Bytes  Uploaded Sat Apr 9 17:35:42 2011  | Show | Hide all | Show all
Attachment 9: GroupDefn.tcl  39 Bytes  Uploaded Sat Apr 9 17:36:24 2011  | Hide | Hide all | Show all

set Scaler_Base 30
set EPICS_Base 31
Attachment 10: epicsDefn.tcl  1 kB  Uploaded Sat Apr 9 17:36:40 2011  | Hide | Hide all | Show all
#    format of an EPICS item
#
#       EPICS parameter name
#       first item number to use for data
#       format of EPICS parameter    0 = 32 bit floating number
#                                    1 = 32 bit integer
#                                    2 = 16 bit integer

set EPICS {}

lappend EPICS {SEBT1:CHAN1CI:SCALECUR 0 0}
lappend EPICS {SEBT1:CHAN2CI:SCALECUR 2 0}
#lappend EPICS {SEBT1:CI1:CONV 4 1}
#lappend EPICS {SEBT1:CI1:RANGE 6 1}
#lappend EPICS {SEBT1:CI1:ALIVE 8 1}
lappend EPICS {SEBT1:CHAN2CI:FULLSCALE 4 1}
#lappend EPICS {SEBT1:CHAN3CI:SCALECUR 12 0}
#lappend EPICS {SEBT1:CHAN4CI:SCALECUR 14 0}
#lappend EPICS {SEBT1:CHAN5CI:SCALECUR 16 0}
#lappend EPICS {SEBT1:CHAN6CI:SCALECUR 18 0}
#lappend EPICS {SEBT1:CI2:CONV 20 0}
#lappend EPICS {SEBT1:CI2:RANGE 22 0}
#lappend EPICS {SEBT1:CI2:ALIVE 24 1}
#lappend EPICS {SEBT1:CHAN6CI:FULLSCALE 26 1}
#lappend EPICS {SEBT1:CHAN7CI:SCALECUR 28 0}
#lappend EPICS {SEBT1:CHAN8CI:SCALECUR 30 0}
#lappend EPICS {SEBT1:CHAN9CI:SCALECUR 32 0}
#lappend EPICS {SEBT1:CHAN10CI:SCALECUR 34 0}
#lappend EPICS {SEBT1:CI3:CONV 36 0}
#lappend EPICS {SEBT1:CI3:RANGE 38 0}
#lappend EPICS {SEBT1:CI3:ALIVE 40 1}
#lappend EPICS {SEBT1:CHAN10CI:FULLSCALE 42 1}
#lappend EPICS {SEBT1:BIASON 44 0}
lappend EPICS {TUDA2:PNG5:RDVAC 6 0}
lappend EPICS {CCS2ISAC:BL2ACURRENT 8 0}

set env(EPICS_CA_ADDR_LIST) 142.90.133.255 
set env(EPICS_CA_AUTO_ADDR_LIST) YES
set env(EPICS_CA_REPEATER_PORT) 9003
set env(EPICS_CA_SERVER_PORT) 9004

set env(PATH) /MIDAS/epics/bin:$env(PATH)
set env(LD_LIBRARY_PATH) /MIDAS/epics/lib:/usr/sfw/lib:$env(LD_LIBRARY_PATH)

#    run caget to  conditionally launch caRepeater
set z [catch {exec caget CCS2ISAC:BL2ACURRENT} m]
puts "EPICS initialise returned code=$z and information \"$m\""

Attachment 11: TS_configuration  944 Bytes  Uploaded Tue Apr 12 08:50:55 2011  | Hide | Hide all | Show all
#device configuration information
#    any line starting with a # is a comment and is ignored
stats  ./stats
#    devices available   -   list ends with a null line or a line starting with a &
#    format       class_name     device_name    driver_task_path_name
#       class_name MUST be one of dlt, exabyte, scsitape, sink
dlt           /dev/rmt/1     ./driver
exabyte       /dev/rmt/0     ./driver
file          /dev/file/0    ./driver
sink          /dev/null/0    ./driver
&
#data link configuration information  - list end with a null line
#    format   link_task_path_name  
./linkTCP
&
#program options - list end with a null line
msg_reporting_level 0x0180fff8
#   use 0x0080fff8 to enable msg logging
msg_logging_level 0xfff8
tapeserver_options 2
#    default for following is 16Kbytes - both MUST be the same at present
data_buffer_size 32
tape_block_size 32
disc_file_size 2000
#
file_path_base /data2/data 
&
#end of information


ELOG V2.9.2-2455