Tuesday, December 02, 2024, 15:35 - Target loaded into the test stand
Thursday, December 03, 2024, 14:33 - as per elog entry 1033, test stand has been changed to CATHODE mode, and wiring within the Faraday cage has been altered appropriately. Additionally, a GeMov resistor has been added to potentially reduce the sparking problems that have previously tripped the TGHT power supply
Thursday, December 04, 2024, 10:00 - Roughing pump having trouble reaching suitable vacuum - possible replacement required.
Thursday, December 05, 2024, 13:37 - Turbo pumps on
Monday, December 09, 2024, 09:51 Vacuum at 3e-7 T
Monday, December 09, 2024, 12:10 Water cooling switched on, monitored vacuum. No change to vacuum.
Monday, December 09, 2024, 16:58: TGHT / TBHT ramping to 580A/230A at 1A/30.
Tuesday, December 10, 2024, 11:14 increasing COIL current 10A at a time, monitoring vacuum pressure
Tuesday, December 10, 2024, 18:42 COIL at 100A after allowing pressure to fall to reasonable levels throughout the day. allowing to stabilize for a few hours.
Tuesday, December 10, 2024, 19:00 vacuum plateauing now, awaiting decrease to 1/3 of peak before reducing COIL to zero
Tuesday, December 10, 2024, 19:37 - vacuum at 9.9e-7T. dropping COIL to 0A
Wednesday, December 11, 2024, 10:27 increasing BIAS/EE/EZL to 30/1.5/13 kV
Wednesday, December 11, 2024, 11:30 TGHT TBHT adjusted to 490/290A
Wednesday, December 11, 2024, 13:47 degaussing script not currently available. For now. will examine beam extracted at CAT=120V COIL=40A
Wednesday, December 11, 2024, 14:27 The CATHODE is completely shorted. cannot hold even 1 volt and is currently limited immediately upon switching on cathode bias power supply.
 
Wednesday, December 11, 2024, 15:02 dropped bias/ee/ezl to 0
Wednesday, December 11, 2024, 15:02 Must cool target to investigate Cathode short. Prior to installation, target isolation was confirmed. If the cathode dropped to contact the copper plate shielding the coil, this would be evidence to support keeping the retainer.
Thursday, December 12, 2024, 09:24 after cooling, cathode power supply was checked to see if shorted and it is indeed still shorted out.
Thursday, December 12, 2024, 09:25 turning off turbo pumps to spin down
Thursday, December 12, 2024, 11:30: venting chamber
Friday December 13: after venting the chamber and removing the target. The target appears to be unshorted and intact. However, upon inspection of the test stand, there appears to be a ~100Ω load between the cathode and ground. This load is able to sustain voltage from the cathode power supply, but draws a significant amount of current. It is too much for operation, and might be further aggravated by the presence of the target. Further, when the TGHT and TBHT terminals are shorted together, the resistance from cathode to ground drops to ~1Ω. Perhaps this could be associated with the installation of the TGHT power supply, as FEBIAD has not been run on this machine since it was installed. Wiring at the feedthroughs as been quadruple checked, I don't believe that the problem lies there.
With no target in place, the cathode was able to hold a voltage, despite drawing 100+mA current. However, I was unable to repeat further tests, as the power supply stopped responding. Despite ramping up to ~10V in epics, the readback is 0V/0A, unlike before.
|
Friday, December 20, 2024, 09:37 turned off all 4 turbopumps. winding down prior to evacuating.
Friday, December 20, 2024, 11:43 Purged and closed all water cooling lines.
Friday, December 20, 2024, 15:54 removed target and installed Uline.
roughing down test stand
Friday, December 20, 2024, 16:17 valves closed and roughing pump off.
Friday, December 20, 2024, 16:17 shut down test stand computer for holiday break |
Recent tests have shown that CATHODE voltage breaks down at ~130V, continually leaking current and progressively dropping voltage, making febiad CATHODE testing impossible.
Thursday, January 23, 2025, 15:18 disconnected cathode wire from TBHT feed through, simply suspended in cage.
voltages are held for ~1min prior to increase. current will be reported after 1min and stable.
100V - 0.000mA
120V - 0.000mA
130V - 0.000mA
140V - 0.000mA
150V - 0.000mA
170V - 0.000mA
200V - 0.000mA
- the Cathode wire and power supply have no leak problem
Thursday, January 23, 2025, 15:42
Reattached the CATHODE line to TBHT feedthrough. Removed the 100||100kΩ resistor pair that connects TBHT to HVC.
Same process as before:
100V - 0.000mA
120V - 0.000mA
125V - 0.000mA
130V - 0.000mA
135V - 0.000mA
140V - current limit ---> dropped to 52V
-- The TBHT resistor pair is not the problem
venting chamber to investigate lines.
Thursday, January 23, 2025, 17:05
removed piece of broken ceramic that held the coil terminal
still current leaking, but it's not as immediate or as shorted. more resistive.
__V__mA__
30 0
35 2.4
40 6.1
45 12.2
50 19.5
55 32
60 gradually ramped up to limited
Thursday, January 23, 2025, 17:10
Disconnected COIL terminal. Behaviour still similarly bad. Doesn't seem to have improved.
Disconnected all TBHT and TGHT feedthroughs. ALL are now isolated from ground (AS BEFORE). TBHT + and - are isolated from all others, TGHT + and - are isolated from all others but 100R apart due to resistor at feedthroughs (as expected).
Increasing CATHODE voltage on TBHT+ feedthrough to test isolation.
V ____ mA
120 ___ 0
135 ____ 0
150 ____ 0
200 ____ 0
- The TBHT+ terminal does not leak directly to ground
Connecting CATHODE cable to TBHT(-) feedthrough to test
V___mA
130 __ 0
150 __ 0
200 __ 0
- the TBHT- terminal does not leak directly to ground.
Connecting CATHODE to TGHT+ feedthrough , and disconnected the 100R between - and +
V ___ mA
100 __ 0
150 __ 0
200 __ 0
- the TGHT + terminal does not leak directly to ground
Connecting CATHODE to TGHT- feedthrough:
V __ mA
125 __ 0
150 __ 0
200 __ 0
- the TGHT- terminal does not leak directly to ground.
connected cathode cable to TBHT+ cable.
V __ mA
30 __ 0
40 __ 0
50 __ 12
60 __ 35
70 -> 60__ 80 current limited
Disconnected 100R path between TBHT and TGHT power supply
CATHODE cable still on TBHT+ cable
V __ mA
150 __ 0
200 __ 0
TBHT is not the problem
reconnected 100 path from TBHT to TGHT. REMOVED varistors that connect TGHT to ground
V __ mA
100 __ 0
200 __ 0
These are likely the issue.
Reconnected varistors to confirm
SHORTED at 50 V -- 10mA, 65V gives 80mA, current limited.
Removed varistors. WIll have to acquire appropriately specced ones |