ID |
Date |
Author |
Target/No.# |
Source |
Station |
Module |
Target Oven W.O# |
Heat Shield W.O# |
Status |
Subject |
|
169
|
Monday, August 13, 2012, 14:56 |
Pierre Bricault | | | | | | | In progress | FEBIAD test |
One thing that seems to be strange regarding this FEBIAD is the best results are without any current in the coil.
Normally for the neon ionization the on-line source wanted to be at around 50 to 80 Amp. Not this one.
Is there anything special about this coil?
|
|
168
|
Monday, August 13, 2012, 14:26 |
Pierre Bricault | | | | | | | In progress | FEBIAD test |
I am done with the FEBIAD test (anode biased)
We will swap the bias now to the cathode and change the polarity to negative. |
|
167
|
Monday, August 13, 2012, 10:59 |
Pierre Bricault | | | | | | | Problems/Solutions | Test FEBIAD |
This morning I try to contact PS group no one answer. Dan Louie is in vacation.
I went into the can diagnostics panel and reconfigure MB1 power supply.
This worked.
|
|
166
|
Monday, August 13, 2012, 08:39 |
Anders Mjos | | | | | | | Problems/Solutions | Started Archivers |
Archivers not running this morning. Started both ISACT and SOURCES |
|
165
|
Sunday, August 12, 2012, 22:07 |
Pierre Bricault | | | | | | | In progress | FBIAD test |
The reset restored the control.
But magnet supply does not put any amperage out.
Try turning it off and on. No chance.
I gave up. See tomorrow. |
|
164
|
Sunday, August 12, 2012, 21:50 |
Pierre Bricault | | | | | | | In progress | Test FEBIAD |
Lost communication with EPICS. Came in reset the ...
So far no luck.
|
|
163
|
Saturday, August 11, 2012, 16:18 |
Pierre Bricault | | | | | | | | Test FEBIAD anode biased at 120 V |
The ramp up finished.
I have 3.35E-6 Amp out of the source on FC0.
Vacuum 4.3 E-7 T increase Coil current to 40 A. No effect on vac.
Increase coil to 50 A. Vacuum still good.
I will continue from home.. |
|
162
|
Saturday, August 11, 2012, 13:34 |
Pierre Bricault | NA | LP-FEBIAD | | | NA | NA | In progress | Test FEBIAD anode bias 120 V |
|
|
161
|
Friday, August 10, 2012, 15:57 |
Pierre Bricault | | | | | | | | |
Test high voltage to 30 kV ok
Test gas valve open at 6.1 V
Test Ne leak ok
Will run test from home.
|
|
160
|
Friday, August 10, 2012, 15:29 |
Pierre Bricault | NA | LP-FEBIAD | | | NA | NA | | finishing the gas installation and its pump |
1524H Start small pump.
1529H stop the pump.
|
|
159
|
Friday, August 10, 2012, 14:08 |
Rick Maharaj | | | | | | | | Gas line pump |
Gas line vacuum pump reinstalled after fitting was modified to receive a standard swagelok connector. The vacuum line was plumbed using 1/4" copper tube (not Stainless) for the the ability to flex and manipulate the line for pump removal and and insertion. |
|
158
|
Friday, August 10, 2012, 13:48 |
Anders Mjos | | | | | | | Problems/Solutions | He line leak checked - leak fixed |
A major leak was found on the connection to the regulation valve on the He bottle. Ferrule replaced. System leak checked.
After He bottle: base 5E-6, no response to He spray
AfterSV1: base 5E-7, no response
After SV2: base 4E-7, no response
The system should now be operational. Gas lines were evacuated with the leak checker. |
|
157
|
Friday, August 10, 2012, 12:44 |
Pierre Bricault | NA | LP-FEBIAD | | | NA | NA | In progress | FEBIAD test |
Rick put the helium bottle back in the rack.
Waiting for status of the gas lines.
|
|
156
|
Friday, August 10, 2012, 11:29 |
Pierre Bricault | NA | | | | N A | NA | In progress | Test FEBIAD |
The helium bottle was found empty.
Rick will refill it and hook the leak detector on the system to find out
if we have a leak on the line.
|
|
155
|
Thursday, August 09, 2012, 13:08 |
Pierre Bricault | | | | | | | | Test FEBIAD |
Waiting Rick for the fitting on the vacuum pump. |
|
154
|
Thursday, July 19, 2012, 12:59 |
Anders Mjos | | | | | | | Problems/Solutions | Leak checking doors - no response |
Thursday, July 19, 2012, 12:57: Doors leak checked. No response. Base rate 1.6E-8 atm-cc/sec. TP pressuer 0.0E-4 Torr . TSGIG1 4.2E-6 Torr. TSG:CG1A 0.014 Torr. |
|
153
|
Thursday, July 19, 2012, 11:32 |
Anders Mjos | | | | | | | Done | Pumping down |
Thursday, July 19, 2012, 11:29: roughing
Thursday, July 19, 2012, 11:49: starting turbos
Thursday, July 19, 2012, 11:54: TP1B at speed
Thursday, July 19, 2012, 11:58: TP1A timed out
Thursday, July 19, 2012, 11:58 TP1A restarted. at speed.
Thursday, July 19, 2012, 11:59: both turbos at speed. Vacuum. 2E-5
|
|
152
|
Thursday, July 19, 2012, 11:31 |
Anders Mjos | | | | | | | Problems/Solutions | Wiped o-rings and sealing faces |
Anders Mjos wrote: |
Due to high vacuum a leak check will be done on the system.
Thursday, July 19, 2012, 09:00 : Base leak rate 4E-9 atm-cc/sec
Thursday, July 19, 2012, 09:07: Leak detected on side door. Mid 1E-8 range.
Thursday, July 19, 2012, 09:08: Leak detected on front door. Mid 1E-6 range. Main leak left or bottom side.
Thursday, July 19, 2012, 09:20: Leak on side door could not be reproduced. Previous response possibly due to "cross-talk" with front door.
|
Thursday, July 19, 2012, 11:29: Wiped both o-rings, grooves and sealing faces with kimwipes. Removed black and brown residues. |
|
151
|
Thursday, July 19, 2012, 09:35 |
Anders Mjos | | | | | | | Done | Venting to investigate leak on front door |
Thursday, July 19, 2012, 09:35: Switched off both turbos.
Thursday, July 19, 2012, 10:23: Venting
|
|
150
|
Thursday, July 19, 2012, 08:42 |
Anders Mjos | | | | | | | Problems/Solutions | Leak checking source chamber - leak found on front door |
Due to high vacuum a leak check will be done on the system.
Thursday, July 19, 2012, 09:00 : Base leak rate 4E-9 atm-cc/sec
Thursday, July 19, 2012, 09:07: Leak detected on side door. Mid 1E-8 range.
Thursday, July 19, 2012, 09:08: Leak detected on front door. Mid 1E-6 range. Main leak left or bottom side.
Thursday, July 19, 2012, 09:20: Leak on side door could not be reproduced. Previous response possibly due to "cross-talk" with front door. |