Message ID: 819     Entry time: Friday, November 06, 2020, 12:53
Author: Aaron Schmidt 
Target/No.#: Nb#9 
Source: LP-SIS 
Station: ITW 
Module: TM2 
Target Oven W.O#: 44508-2 
Heat Shield W.O#: 43515-4 
Status: In progress 
Subject: Conditioning Nb#9 w/ P/S Target & Bracket 

Attached end plate parasitic / symbiotic target & PEEK assembly bracket.

Friday, November 06, 2020, 12:24: Roughing.

Friday, November 06, 2020, 12:53: TP1B on.

Friday, November 06, 2020, 13:05: TP1A on. 

Friday, November 06, 2020, 13:16: IG1 on.


Monday, November 09, 2020, 06:26: Vacuum @ 1.25e-06T.

Monday, November 09, 2020, 06:28: HS waterline opened. Vacuum ok @ 1.25e-06T.

Monday, November 09, 2020, 06:36: Power/sources on.

Monday, November 09, 2020, 06:39: Auto-ramping TGHT / TBHT to 350 / 150 A @ 4 A per min.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 23.96 5.15
 TBHT 23.78

8.02

 COIL + EE 23.67 2.39
 TGHT 23.77 6.19

Monday, November 09, 2020, 06:4: Optics on.

Monday, November 09, 2020, 08:10: IV4 open.

Monday, November 09, 2020, 08:15: BIAS / EE / EZL @ 30 / -1.5 / 13 kV.

Monday, November 09, 2020, 08:19: Wire & Mass scans @ 350 / 150 A. Ok.

Monday, November 09, 2020, 08:27: Auto-ramping TGHT / TBHT to 580 / 230 A @ 2 A per min.

Monday, November 09, 2020, 11:40: Mass & wire scans complete.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 25.27 7.12
 TBHT 24.88

9.73

 COIL + EE 24.65 3.26
 TGHT 24.88 7.05

Monday, November 09, 2020, 13:46: E-rig PS off.


Tuesday, November 10, 2020, 04:47: Auto-cooling @ 2A per min. 

Tuesday, November 10, 2020, 06:21: HV @ 0 V. 

Tuesday, November 10, 2020, 06:58: TBHT @ 0 A. TGHT @ 317 A. Auto-cooling @ 3A per min. 

Tuesday, November 10, 2020, 09:06: IV4 Close. Sources off.

Tuesday, November 10, 2020, 09:07: IG1, power, & Turbo pumps off.

Tuesday, November 10, 2020, 10:15: Waterline purged. Venting.

Attachment 1: Nb#9_Conditioning_with_P-S_Mods.docx  1.522 MB  Uploaded Monday, November 16, 2020, 14:42
Attachment 2: ITW-TM2-Nb#9-LP-SIS_Test_Stand.docx  860 kB  Uploaded Monday, November 16, 2020, 14:43
Attachment 3: ER-580-230-30kV-m7.txt  3 kB  Uploaded Monday, November 16, 2020, 14:43  | Show | Hide all
Attachment 4: ER-580-230-30kV-m23.txt  3 kB  Uploaded Monday, November 16, 2020, 14:43  | Show | Hide all
Attachment 5: ER-580-230-30kV-m39.txt  3 kB  Uploaded Monday, November 16, 2020, 14:43  | Hide | Hide all
value (Voltage offset      40.0000)
     75.0000
      1.0000
     95.0000
  29978.0220
21
21
./emit.dat
 -0.1260 -0.0723 -0.0952 -0.0980 -0.0671 -0.0824 -0.0494 -0.0656 -0.0723 -0.0754 -0.0720 -0.0717 -0.0739 -0.0726 -0.0720 -0.0787 -0.0739 -0.0757 -0.0754 -0.0723 -0.0726
 -0.0940 -0.0674 -0.0928 -0.0946 -0.0662 -0.0870 -0.0662 -0.0473 -0.0058 -0.0684 -0.0671 -0.0690 -0.0677 -0.0665 -0.0684 -0.0732 -0.0677 -0.0696 -0.0711 -0.0635 -0.0677
 -0.0922 -0.0632 -0.0552 -0.0656 -0.0671 -0.0571 -0.0574 -0.0476 -0.0510 -0.0211 -0.0436 -0.0687 -0.0653 -0.0671 -0.0659 -0.0729 -0.0677 -0.0653 -0.0629 -0.0653 -0.0647
 -0.0900 -0.0671 -0.0665 -0.0610 -0.0665 -0.0644 -0.0656 -0.0662 -0.0668 0.0046 -0.0204 -0.0067 -0.0674 -0.0674 -0.0684 -0.0729 -0.0656 -0.0690 -0.0668 -0.0674 -0.0647
 -0.0894 -0.0668 -0.0665 -0.0641 -0.0656 -0.0671 -0.0659 -0.0656 -0.0671 -0.0653 -0.0742 -0.0162 0.0131 -0.0146 -0.0684 -0.0726 -0.0690 -0.0681 -0.0638 -0.0668 -0.0665
 -0.0900 -0.0656 -0.0674 -0.0638 -0.0653 -0.0665 -0.0659 -0.0620 -0.0693 -0.0665 -0.0650 -0.0653 -0.0415 0.0394 0.0021 -0.0739 -0.0699 -0.0638 -0.0711 -0.0641 -0.0690
 -0.0919 -0.0641 -0.0659 -0.0662 -0.0644 -0.0635 -0.0671 -0.0665 -0.0696 -0.0635 -0.0647 -0.0632 -0.0635 -0.0638 -0.0281 0.0723 -0.0082 -0.0677 -0.0644 -0.0665 -0.0647
 -0.0912 -0.0662 -0.0629 -0.0677 -0.0656 -0.0638 -0.0659 -0.0662 -0.0699 -0.0656 -0.0647 -0.0656 -0.0653 -0.0638 -0.0638 -0.0595 0.0208 0.0162 -0.0330 -0.0665 -0.0641
 -0.0894 -0.0656 -0.0644 -0.0760 -0.0650 -0.0659 -0.0632 -0.0644 -0.0677 -0.0653 -0.0693 -0.0662 -0.0671 -0.0635 -0.0662 -0.0720 -0.0635 -0.0397 0.0357 0.0061 -0.0629
 -0.0903 -0.0653 -0.0650 -0.0671 -0.0681 -0.0681 -0.0644 -0.0647 -0.0687 -0.0668 -0.0650 -0.0653 -0.0684 -0.0656 -0.0647 -0.0735 -0.0681 -0.0635 -0.0629 0.0140 -0.0177
 -0.0916 -0.0638 -0.0671 -0.0656 -0.0638 -0.0641 -0.0650 -0.0641 -0.0696 -0.0668 -0.0635 -0.0647 -0.0674 -0.0656 -0.0644 -0.0729 -0.0674 -0.0650 -0.0671 -0.0626 -0.0531
 -0.0906 -0.0653 -0.0647 -0.0671 -0.0641 -0.0629 -0.0668 -0.0644 -0.0690 -0.0635 -0.0668 -0.0659 -0.0659 -0.0665 -0.0650 -0.0711 -0.0677 -0.0650 -0.0638 -0.0650 -0.0629
 -0.0906 -0.0656 -0.0635 -0.0684 -0.0671 -0.0653 -0.0650 -0.0665 -0.0690 -0.0650 -0.0659 -0.0659 -0.0671 -0.0656 -0.0656 -0.0726 -0.0696 -0.0659 -0.0647 -0.0668 -0.0638
 -0.0891 -0.0668 -0.0656 -0.0696 -0.0659 -0.0665 -0.0638 -0.0644 -0.0693 -0.0644 -0.0647 -0.0653 -0.0684 -0.0665 -0.0662 -0.0732 -0.0647 -0.0653 -0.0653 -0.0668 -0.0653
 -0.0873 -0.0644 -0.0653 -0.0681 -0.0641 -0.0647 -0.0632 -0.0629 -0.0665 -0.0653 -0.0641 -0.0671 -0.0671 -0.0644 -0.0647 -0.0748 -0.0659 -0.0677 -0.0681 -0.0650 -0.0665
 -0.0909 -0.0635 -0.0638 -0.0711 -0.0644 -0.0626 -0.0644 -0.0623 -0.0668 -0.0647 -0.0699 -0.0635 -0.0665 -0.0656 -0.0653 -0.0726 -0.0711 -0.0626 -0.0641 -0.0647 -0.0659
 -0.0912 -0.0629 -0.0641 -0.0699 -0.0650 -0.0635 -0.0644 -0.0650 -0.0665 -0.0632 -0.0674 -0.0656 -0.0647 -0.0650 -0.0656 -0.0699 -0.0668 -0.0653 -0.0647 -0.0653 -0.0635
 -0.0903 -0.0653 -0.0644 -0.0650 -0.0656 -0.0647 -0.0638 -0.0653 -0.0687 -0.0653 -0.0641 -0.0644 -0.0659 -0.0629 -0.0629 -0.0723 -0.0674 -0.0659 -0.0641 -0.0665 -0.0650
 -0.0900 -0.0659 -0.0647 -0.0644 -0.0659 -0.0650 -0.0641 -0.0650 -0.0687 -0.0644 -0.0620 -0.0647 -0.0659 -0.0635 -0.0653 -0.0729 -0.0650 -0.0653 -0.0650 -0.0647 -0.0668
 -0.0879 -0.0638 -0.0653 -0.0635 -0.0659 -0.0641 -0.0629 -0.0641 -0.0668 -0.0656 -0.0644 -0.0647 -0.0674 -0.0644 -0.0647 -0.0723 -0.0653 -0.0626 -0.0653 -0.0641 -0.0665
 -0.0888 -0.0650 -0.0653 -0.0650 -0.0650 -0.0641 -0.0638 -0.0629 -0.0671 -0.0638 -0.0659 -0.0623 -0.0659 -0.0647 -0.0641 -0.0720 -0.0671 -0.0638 -0.0656 -0.0635 -0.0656
Attachment 6: ER-580-230-30kV-m83.txt  3 kB  Uploaded Monday, November 16, 2020, 14:43  | Show | Hide all
Attachment 7: ER-580-230-30kV-m133.txt  3 kB  Uploaded Monday, November 16, 2020, 14:43  | Show | Hide all
ELOG V2.9.2-2455