Previous short fixed. Grid clamp removed. Grid tack welded.

Friday, October 13, 2017, 15:49: Roughing.


By Anders

Friday, October 13, 2017, 16:10: TPs started

Friday, October 13, 2017, 16:20: Vacuum at 2.7E-7 T


Saturday, October 14, 2017, 06:37: Opened water supply to HS. 95/10 psi. Vacuum stable at 5.7E-7 T

Saturday, October 14, 2017, 06:42: Opened gas lines.

Saturday, October 14, 2017, 06:43: Started heater ramping to 580 / 290 A

Saturday, October 14, 2017, 06:45: BIAS/EZL/EE to 30 / 13.5 / 1.5 kV. EE current somewhat unstable around 500 uA - seems related to elevated BIAS current. Improving as BIAS current is gradually coming down.

Saturday, October 14, 2017, 15:19: TBHT to 310 A. Local readback 316 A


Sunday, October 15, 2017, 17:36 : Ramping down heaters.


Monday, October 16, 2017, 09:10: Sources off.

Monday, October 16, 2017, 09:13; IV4 Closed & Reset. FEBIAD Gas lines closed.

Monday, October 16, 2017, 09:15: RGA, Ion gauge, turbos off.

Monday, October 16, 2017, 10:40: Power off. Waterlines purged.

Monday, October 16, 2017, 10:43: Venting.

Attachment 1: SiC#37_(iii).docx  225 kB  Uploaded Tuesday, October 17, 2017, 12:26
Attachment 2: TS-2017-10-13-17-44.rga  415 kB  Uploaded Thursday, October 19, 2017, 12:04
Attachment 3: TS-2017-10-14-06-35.rga  1.548 MB  Uploaded Thursday, October 19, 2017, 12:04
Attachment 4: ER-580-310-30kV-m20.txt  3 kB  Uploaded Thursday, October 19, 2017, 12:04  | Show | Hide all
Attachment 5: ER-580-310-30kV-m23.txt  3 kB  Uploaded Thursday, October 19, 2017, 12:04  | Show | Hide all
Attachment 6: ER-580-310-30kV-m28.txt  3 kB  Uploaded Thursday, October 19, 2017, 12:04  | Hide | Hide all
value (Voltage offset       0.0000)
     76.0000
      0.9000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1187 -0.0696 -0.0842 -0.0742 -0.0809 -0.0940 -0.0714 -0.0360 0.0162 0.0665 0.1373 0.5869 1.1185 0.8875 0.5414 0.1840 0.0412 -0.0162 -0.0540 -0.0662 -0.0754
 -0.0595 -0.0696 -0.0818 -0.0671 -0.0717 -0.0870 -0.0842 -0.0754 -0.0403 0.0131 0.0674 0.2820 0.8087 0.7736 0.5081 0.1941 0.0494 -0.0217 -0.0558 -0.0607 -0.0705
 -0.0598 -0.0729 -0.0797 -0.0629 -0.0592 -0.0653 -0.0629 -0.0919 -0.0983 -0.0595 0.0137 0.0867 0.4562 0.5594 0.4379 0.1944 0.0574 -0.0235 -0.0562 -0.0644 -0.0589
 -0.0647 -0.0735 -0.0775 -0.0659 -0.0461 0.1990 0.1093 -0.0995 -0.1205 -0.1315 -0.0946 -0.0339 0.0912 0.2820 0.3189 0.1581 0.0604 -0.0229 -0.0626 -0.0671 -0.0632
 -0.0677 -0.0653 -0.0790 -0.0671 -0.0525 0.3851 2.2314 2.3416 -0.0034 -0.1788 -0.1575 -0.1410 -0.0906 0.0049 0.1236 0.0906 0.0482 -0.0162 -0.0613 -0.0681 -0.0677
 0.0015 -0.0677 -0.0839 -0.0659 -0.0623 -0.0375 2.6950 5.9198 4.4141 1.5039 -0.2081 -0.1733 -0.1437 -0.1343 -0.0751 -0.0055 0.0079 -0.0189 -0.0604 -0.0671 -0.0653
 1.3000 0.5222 -0.0537 -0.0626 -0.0696 -0.0665 0.3500 6.6342 9.9997 9.9997 6.5463 0.6955 -0.1389 -0.1205 -0.1266 -0.0842 -0.0458 -0.0369 -0.0616 -0.0681 -0.0665
 -0.0034 1.7517 1.2799 0.7275 0.1010 -0.0507 -0.0269 1.8335 9.9997 9.9997 9.9997 9.9997 3.0206 -0.0671 -0.0848 -0.0958 -0.0940 -0.0616 -0.0626 -0.0665 -0.0653
 -0.0598 -0.0211 1.6721 2.0697 1.9296 1.1948 0.5298 0.4449 7.5043 9.9997 9.9997 9.9997 9.9997 6.2766 1.6965 -0.0626 -0.0717 -0.0607 -0.0656 -0.0653 -0.0620
 -0.0601 -0.0604 -0.0446 0.3900 3.1848 2.8616 2.8793 3.1970 3.4647 9.9997 9.9997 9.9997 9.9997 8.4708 3.4933 2.9553 -0.0632 -0.0583 -0.0653 -0.0659 -0.0641
 -0.0571 -0.0690 -0.0735 -0.0558 0.0223 3.2553 4.3109 5.0281 6.7178 9.9997 9.9997 9.9997 9.9997 9.9997 5.2866 2.2714 1.7648 0.7895 -0.0616 -0.0681 -0.0647
 -0.0552 -0.0714 -0.0766 -0.0616 -0.0546 0.0031 1.1572 6.2912 6.8265 9.1522 9.9997 9.9997 9.9997 9.9997 9.4885 1.9513 1.3593 1.0706 0.9070 -0.0662 -0.0647
 -0.0583 -0.0641 -0.0766 -0.0616 -0.0528 -0.0525 -0.0238 0.1154 4.3103 9.9997 9.9997 9.9997 9.9997 9.9997 9.9997 2.5046 0.6808 0.9274 0.7187 0.4718 0.0467
 -0.0616 -0.0589 -0.0775 -0.0601 -0.0522 -0.0519 -0.0360 -0.0238 0.0760 1.6986 9.2844 9.9997 9.9997 9.9997 9.9997 6.6360 0.0238 0.0186 0.2628 0.3699 0.2271
 -0.0632 -0.0604 -0.0793 -0.0598 -0.0565 -0.0549 -0.0290 -0.0119 -0.0073 0.0375 0.2820 2.9593 9.9997 9.9997 9.9997 9.9997 3.8776 -0.0449 -0.0552 -0.0046 0.1202
 -0.0586 -0.0693 -0.0772 -0.0629 -0.0613 -0.0571 -0.0192 0.0146 0.0284 0.0238 0.0653 0.1486 0.5829 5.3879 9.9997 9.9997 6.1002 1.5350 -0.0574 -0.0616 -0.0595
 -0.0552 -0.0641 -0.0745 -0.0665 -0.0592 -0.0586 -0.0189 0.0372 0.0952 0.0946 0.0949 0.0949 0.1031 0.1544 0.3656 5.3406 6.2766 2.5500 -0.0174 -0.0601 -0.0607
 -0.0565 -0.0565 -0.0677 -0.0610 -0.0516 -0.0601 -0.0195 0.0439 0.1477 0.2090 0.1990 0.1373 0.0867 0.0418 0.0128 0.0638 0.1416 1.5460 0.3824 -0.0607 -0.0623
 -0.0580 -0.0522 -0.0635 -0.0522 -0.0433 -0.0516 -0.0177 0.0443 0.1752 0.3250 0.3522 0.2570 0.1184 0.0406 -0.0015 -0.0357 -0.0537 -0.0323 -0.0510 -0.0592 -0.0620
 -0.0580 -0.0531 -0.0650 -0.0449 -0.0363 -0.0345 -0.0098 0.0555 0.1871 0.3961 0.5533 0.4825 0.1907 0.0653 0.0006 -0.0372 -0.0580 -0.0558 -0.0623 -0.0708 -0.0659
 -0.0552 -0.0562 -0.0626 -0.0427 -0.0360 -0.0241 0.0079 0.0748 0.2063 0.4385 0.7339 0.7773 0.3360 0.1132 0.0156 -0.0320 -0.0555 -0.0558 -0.0607 -0.0690 -0.0632
Attachment 7: ER-580-310-30kV-m40.txt  3 kB  Uploaded Thursday, October 19, 2017, 12:04  | Show | Hide all
Attachment 8: SiC#37_(ii).docx  1.898 MB  Uploaded Thursday, October 19, 2017, 12:17
Attachment 9: OMEGA_OM-DAQPRO-5300_19-Oct-17_11.17.50.dlp  3.676 MB  Uploaded Thursday, October 19, 2017, 12:18
Attachment 10: ITE-TM4-SiC#37-HP-FEBIAD_Test_Stand.docx  692 kB  Uploaded Thursday, October 19, 2017, 13:03