Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  Test-Stand  Not logged in ELOG logo
Message ID: 547     Entry time: Tuesday, October 18, 2016, 13:29
Author: Aaron Schmidt 
Target/No.#: Ta#50 
Source: LP-SIS 
Station: ITE 
Module: TM4 
Target Oven W.O#: 41790-5 
Heat Shield W.O#: 41751-1 
Status: In progress 
Subject: Conditioning Ta#50 

Tuesday, October 18, 2016, 13:22: Roughing.

Tuesday, October 18, 2016, 13:46: Turbo pumps started.

Tuesday, October 18, 2016, 13:52: Ion gauge on.

Tuesday, October 18, 2016, 13:54: Both turbo pumps at speed.

Tuesday, October 18, 2016, 15:05: RGA on.


Wednesday, October 19, 2016, 08:46: Power/Sources on

Wednesday, October 19, 2016, 08:49: Waterline opened. No response on RGA. Vacuum stable @4.40E-7T

Wednesday, October 19, 2016, 08:57: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .

Wednesday, October 19, 2016, 08:58 : IV4 opened.

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 27.9 3.90
 TBHT 27.7 3.06
 COIL + EE 27.6 5.08
 TGHT  27.7 5.70

Wednesday, October 19, 2016, 09:01: Auto ramping TGHT / TBHT to 580 / 200 A @ 2A per min.

Wednesday, October 19, 2016, 11:50: TMS:FC0 moved out & TMS:RPM0B began scanning continuously on their own. Monitor.

Wednesday, October 19, 2016, 14:45: RGA Restarted.

Wednesday, October 19, 2016, 15:25: RGA Restarted (Unable to save 14:45 Startup)


Friday, October 21, 2016, 13:27: Auto Cooling @ 2A per min

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 37.3 3.92
 TBHT 28.2 3.07
 COIL + EE 27.9 5.07
 TGHT  29.8 5.62

Friday, October 21, 2016, 15:39: IV4 Closed


Monday, October 24, 2016, 09:10:

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 28.1 3.95
 TBHT 28.2 3.10
 COIL + EE 27.7 5.04
 TGHT  27.7 5.70

Monday, October 24, 2016, 09:10: Power/Sources off.

Monday, October 24, 2016, 09:20: Water lines purged.

Monday, October 24, 2016, 09:21: RGA off.

Monday, October 24, 2016, 09:23: Ion gauge & Turbo pumps off

Monday, October 24, 2016, 12:54: Venting.

Attachment 1: ER-580-200-30kV-m27.txt  3 kB  Uploaded Monday, October 24, 2016, 12:07  | Show | Hide all
Attachment 2: ER-580-200-30kV-m27(2).txt  3 kB  Uploaded Monday, October 24, 2016, 12:07  | Show | Hide all
Attachment 3: ER-580-200-30kV-m27(3).txt  3 kB  Uploaded Monday, October 24, 2016, 12:07  | Show | Hide all
Attachment 4: ER-580-200-30kV-m39.txt  3 kB  Uploaded Monday, October 24, 2016, 12:07  | Hide | Hide all
value (Voltage offset      40.0000)
     80.0000
      0.7000
     80.0000
  29978.0220
21
21
./emit.dat
 -0.1172 -0.0961 -0.0958 -0.0970 -0.0995 -0.0980 -0.0961 -0.0995 -0.0992 -0.0961 -0.0995 -0.0989 -0.0992 -0.0961 -0.0983 -0.0974 -0.0986 -0.0940 -0.1013 -0.0989 -0.0989
 -0.0906 -0.0903 -0.0906 -0.0909 -0.0922 -0.0928 -0.0931 -0.0986 -0.0946 -0.0946 -0.0943 -0.0961 -0.0937 -0.0958 -0.0922 -0.0916 -0.0922 -0.0940 -0.0986 -0.0955 -0.0955
 -0.0928 -0.0925 -0.0912 -0.0928 -0.0940 -0.0934 -0.0934 -0.0958 -0.0964 -0.0937 -0.0940 -0.0934 -0.0928 -0.0928 -0.0925 -0.0925 -0.0952 -0.0934 -0.0986 -0.0970 -0.0940
 -0.0940 -0.0931 -0.0937 -0.0970 -0.0897 -0.0925 -0.0958 -0.0931 -0.0919 -0.0909 -0.0909 -0.0931 -0.0897 -0.0916 -0.0940 -0.0940 -0.0934 -0.0928 -0.0958 -0.0967 -0.0900
 -0.0934 -0.0928 -0.0922 -0.0986 -0.0934 -0.0897 -0.0922 -0.0943 -0.0934 -0.0912 -0.0919 -0.0912 -0.0946 -0.0922 -0.0916 -0.0937 -0.0925 -0.0919 -0.0946 -0.0955 -0.0940
 -0.0912 -0.0906 -0.0925 -0.0958 -0.0952 -0.0955 -0.0961 -0.0928 -0.0952 -0.0916 -0.0928 -0.0949 -0.0937 -0.0952 -0.0937 -0.0952 -0.0912 -0.0931 -0.0946 -0.0946 -0.0894
 -0.0897 -0.0922 -0.0888 -0.0949 -0.0940 -0.0934 -0.0961 -0.0946 -0.0912 -0.0928 -0.0943 -0.0940 -0.0937 -0.0961 -0.0928 -0.0937 -0.0922 -0.0931 -0.0949 -0.0961 -0.0958
 -0.0919 -0.0906 -0.0912 -0.0970 -0.0906 -0.0912 -0.0958 -0.0925 -0.0809 -0.0858 -0.0897 -0.0937 -0.0903 -0.0919 -0.0906 -0.0934 -0.0937 -0.0928 -0.0974 -0.0964 -0.0916
 -0.0943 -0.0940 -0.0925 -0.0961 -0.0931 -0.0906 -0.0912 -0.0934 -0.0610 -0.0494 -0.0681 -0.0894 -0.0937 -0.0894 -0.0900 -0.0909 -0.0934 -0.0925 -0.0961 -0.0952 -0.0912
 -0.0928 -0.0903 -0.0934 -0.0998 -0.0937 -0.0919 -0.0940 -0.0934 -0.0641 -0.0165 -0.0314 -0.0763 -0.0903 -0.0925 -0.0928 -0.0952 -0.0925 -0.0922 -0.0955 -0.0943 -0.0909
 -0.0912 -0.0925 -0.0900 -0.0961 -0.0961 -0.0934 -0.0937 -0.0937 -0.0851 -0.0357 -0.0394 -0.0638 -0.0909 -0.0906 -0.0928 -0.0952 -0.0919 -0.0916 -0.0952 -0.0940 -0.0928
 -0.0912 -0.0900 -0.0909 -0.0974 -0.0934 -0.0925 -0.0977 -0.0949 -0.0903 -0.0745 -0.0598 -0.0702 -0.0876 -0.0900 -0.0839 -0.0900 -0.0916 -0.0925 -0.0952 -0.0961 -0.0922
 -0.0934 -0.0922 -0.0912 -0.0977 -0.0916 -0.0909 -0.0958 -0.0940 -0.0931 -0.0894 -0.0848 -0.0836 -0.0912 -0.0900 -0.0528 -0.0897 -0.0888 -0.0922 -0.0977 -0.0958 -0.0925
 -0.0931 -0.0906 -0.0940 -0.0989 -0.0909 -0.0916 -0.0952 -0.0919 -0.0925 -0.0916 -0.0888 -0.0912 -0.0885 -0.0891 -0.0864 -0.0897 -0.0897 -0.0912 -0.0922 -0.0943 -0.0900
 -0.0937 -0.0928 -0.0919 -0.0992 -0.0940 -0.0943 -0.0937 -0.0937 -0.0937 -0.0897 -0.0912 -0.0922 -0.0931 -0.0925 -0.0912 0.2444 -0.0891 -0.0900 -0.0952 -0.0925 -0.0885
 -0.0912 -0.0903 -0.0912 -0.0974 -0.0943 -0.0934 -0.0958 -0.0940 -0.0925 -0.0937 -0.0922 -0.0946 -0.0919 -0.0922 -0.0934 -0.0317 -0.0873 -0.0912 -0.0964 -0.0946 -0.0894
 -0.0922 -0.0906 -0.0903 -0.0958 -0.0952 -0.0937 -0.0955 -0.0955 -0.0949 -0.0916 -0.0931 -0.0934 -0.0937 -0.0943 -0.0919 -0.0595 -0.0928 -0.0922 -0.0967 -0.0955 -0.0952
 -0.0931 -0.0897 -0.0906 -0.0961 -0.0928 -0.0916 -0.0970 -0.0946 -0.0922 -0.0925 -0.0897 -0.0922 -0.0903 -0.0873 -0.0900 -0.0919 -0.0949 -0.0937 -0.0937 -0.0952 -0.0916
 -0.0940 -0.0928 -0.0925 -0.0974 -0.0922 -0.0912 -0.0946 -0.0937 -0.0922 -0.0891 -0.0916 -0.0916 -0.0916 -0.0900 -0.0916 -0.0922 -0.0919 -0.0922 -0.0949 -0.0943 -0.0931
 -0.0931 -0.0909 -0.0919 -0.0970 -0.0931 -0.0922 -0.0949 -0.0925 -0.0900 -0.0922 -0.0891 -0.0928 -0.0903 -0.0922 -0.0940 -0.0958 -0.0925 -0.0903 -0.0922 -0.0952 -0.0900
 -0.0919 -0.0900 -0.0888 -0.0958 -0.0940 -0.0922 -0.0949 -0.0937 -0.0934 -0.0922 -0.0928 -0.0937 -0.0943 -0.0949 -0.0928 -0.0937 -0.0903 -0.0906 -0.0955 -0.0937 -0.0925
Attachment 5: ITE-TM4-Ta#50-LP-SIS_Test_Stand.docx  724 kB  Uploaded Monday, October 24, 2016, 12:07
Attachment 6: Ta#50.docx  2.445 MB  Uploaded Monday, October 24, 2016, 12:08
Attachment 7: TS-2016-10-18-15-05.rga  655 kB  Uploaded Monday, October 24, 2016, 12:08
Attachment 8: TS-2016-10-19-14-42.rga  3.461 MB  Uploaded Monday, October 24, 2016, 12:08
Attachment 9: OMEGA_OM-DAQPRO-5300_24-Oct-16_11.03.10.dlp  1.211 MB  Uploaded Monday, October 24, 2016, 12:08
ELOG V2.9.2-2455