ID |
Date |
Author |
Target/No.# |
Source |
Station |
Module |
Target Oven W.O# |
Heat Shield W.O# |
Status |
Subject |
9
|
Friday, June 17, 2011, 11:58 |
John Wong | UCx (flakes) | LP-SIS | | | | | Problems/Solutions | Making UC2 flakes (stock# 2011_A & _B) |
Fri Jun 17 10:28:47 2011
First batch of UO2/C mixture of carbonization (Stock# 2011_A):
UO2 used = 50.120g (left overs + container = 411.23g)
C(grv) used = 10.653g (or 21.25% w/w)
Mass of organics = 4.842g (or 9.66% w/w)
Mass of dried cast = 47.420g
Waste collected in jar = 18.195g (including UO2, C and Org)
Waste accountable for UO2 = 12.57g
The carbonization process started from May 26 to June 4 (9 days).
See attachment #1 and #2 for the striptools during the UO2/C carbonization
Total Mass UC2 flakes = 35.310g
Fri Jun 17 11:56:15 2011
Attempted to make the second batch of UO2/C mixture of carbonization on June 13, but it failed. The milling jar gasket seal partially came out during the high speed milling. As a result, the UO2/C mixture squirted out and contaminated the planetary mill (See attachment #3 & #4). RPG and all other groups were informed. The planetary mill and the lab were properly cleaned up (see attachment #5), no activities were found after the cleaning. Attachments #6, #7, #8 are some of the discussion about the UO2 spill.
All the UO2 taken out from the container (50.23g) is accountable for waste.
June 28, 2011 - Re-making 2nd batch of UO2/C mixture was successful.
Stock#2011_B:
UO2 used = 50.29g
C(grv) used = 10.0g
Mass of Organics = 4.84g
Mass of Dried green = 71.96g (the cast was let dry over night <24hrs due to time restriction. It might not be completely dried as a result of such heavy mass remained. However, the UO2 waste was very minimum, <5.0g)
Total Mass UC2 flakes = 56.853g
|
Attachment 1: UO2_C_IG1_CG2_2011_A.jpg
|
|
Attachment 2: UO2_C_TGHT_IG1_2011_A.jpg
|
|
Attachment 3: UO2_mill_spill_01.png
|
|
Attachment 4: UO2_mill_spill_02.png
|
|
Attachment 5: UO2_spill_cleanup.png
|
|
Attachment 6: UO2_spill_discussion_01.png
|
|
Attachment 7: UO2_spill_discussion_02.png
|
|
Attachment 8: UO2_spill_discussion_03.png
|
|
3
|
Tuesday, March 29, 2011, 14:14 |
John Wong | Ta#35 | HP-SIS | ITE | TM4 | 38481-3 | 38303-4 | Spent | ITE-TM4-Ta#35-Hp-SIS |
ITE-TM4-Ta#35-Hp-SIS
Thu Apr 7 15:32:32 2011:
Target oven - w.o#38481-3.
Packs of Ta foils were previously prepared by Vicky H.
Marik loaded the target: 400 of 0.025mm foils (25 foils/pk) in 3.5cm length. (Recommended length 4.0-4.5cm; If too closely packed, wires get squeezed and snapped)
Average mass per discs = 0.105g
Total mass (Ta+wire) = 45.06g
Disc area = 2.12cm2
Ta thickness = 21.25g/cm2
Thu Apr 28 :
Target heat Shield (w.o 38303-4) cleaned
Target is now assembled by John and Rick
Megger Test Performed:
TA-->HS = 540V, OL
TA-->EE = 540V, OL
HS--> EE = 1049V, OL
Leak Check OK
*SIS restraint used in this target ion source
Mon May 9 10:32:40 2011: Standard salts added: Li, Na, Rb @ 10uL, K @ 5uL (but added 10uL), Cs @ 15uL. Target will be going to the test stand.
Fri May 20 16:57:22 2011:
Target was tested on the Test Stand but problems with the software (see Test stand elog).
Standard salts added: Li, Na, Rb @ 10uL, K @ 5uL (but added 10uL), Cs @ 15uL, and Be @ 10uL.
Monday, July 18, 2011, 11:15: Target ended on July 16. The target was used online for an extra 2 weeks and performed very well.
|
Attachment 1: Ta-HPSIS.jpg
|
|
4
|
Thursday, March 31, 2011, 15:38 |
John Wong | SiC#25 | HP-SIS | ITE | TM4 | 38481-1 | 38307-? | Spent | ITE-TM4-SiC#25-Hp-SIS |
Thu Mar 31 15:19:51 2011: SiC#25 was made sometime in December of 2010. It is now in the target module and ready to be conditioned.
Mon Apr 18 18:17:05 2011: It was found later that no salt was added to this target after it came out from the test stand. However, there were enough 40Ca and 39K for tuning (See attachment#1 for the big mass scan). Once the P+ is back, the yields will be measured.
Fri Apr 29 15:30:26 2011: Low 27Al was ionized from the surface for some reason. However, after the green laser was shone into the extraction tube, more 27Al was produced. Then, the laser was turned off, the target continued to surface ionizing 27Al. There could be a blockage from the rhenium (Re) ionizer foil. Jens L. will investigate this further and do a test with the lasers and the rhenium foil.
|
Attachment 1: big_mass_scan.jpg
|
|
21
|
Thursday, May 24, 2012, 11:04 |
John Wong | SiC#27 | HP-SIS | ITE | TM4 | 38817-3 | 39069-2 | Spent | HP-SIS SiC27 target description |
Thursday, May 24, 2012, 10:34
SiC#27- HPSIS target description:
Total SiC foils = 240 discs
Total SiC+ C-grv = 27.79g
Total length = 8.3cm
Average mass of SiC/disc (without C-grv) = 0.085g
Average thickness of SiC (without C-grv) = 0.24mm
No Rh foil added
Salt added = 10uL Li, 10uL Na
see attachment#1: SiC#27_density measurements
see link for more details: http://documents.triumf.ca/docushare/dsweb/View/Collection-8530
|
Attachment 1: SiC#27_specification.png
|
|
152
|
Friday, May 30, 2025, 14:23 |
Darwin Ortiz Rosales | C#5 | LP-SIS | | | 45220-6 | | Ready | C#5 conditioning |
Friday, May 30, 2025, 14:21 |
Attachment 1: C#5_Description.png
|
|
96
|
Tuesday, November 19, 2019, 12:20 |
Darwin Ortiz Rosales | UCx#30 | LP-SIS | | | 44062-3 | | Ready | UCx30-target_description |
|
Attachment 1: UCx#30.PNG
|
|
Attachment 2: UCx30-taret_description_revised.png
|
|
124
|
Tuesday, November 15, 2022, 08:41 |
Darwin Ortiz Rosales | UCx#41 | LP-SIS | | | 45220-13 | | Ready | |
|
Attachment 1: UCx#41_Description.png
|
|
131
|
Friday, May 26, 2023, 11:53 |
Darwin Ortiz Rosales | SiC#45 | HP-SIS | | | 44692-3 | | Ready | |
|
Attachment 1: SiC#45.png
|
|