ID |
Date |
Author |
Target Material |
Target Oven W.O# |
Source |
Status |
Subject |
173
|
Tuesday, November 14, 2017, 16:16 |
Laura Lambert | UO2/C | 40282-2 | TARGET BOAT | Done | Carbonizing UC2 discs for 1stepUCx in graphite sample container _#2 | Tuesday, November 14, 2017, 16:12
Circles from cast 1 were used, the same as the last round of 1stepUCx [see elog i.d. 164]
A 14mm punch was used to punch out circles from the cast.
35 circles were packed into a graphite sample container.
35 x UO2/C circles = 5.43g
Graphite tube is placed inside a target boat.
Boat previously coated with TaC and sintered.
Ta-target boat is wrapped with 3-layers of heat shield and installed in Evap1.
Chamber currently pumping down.
Wednesday, November 15, 2017, 08:53
IGP1 = 1.6e-6 Torr
Tapping up manually...
Wednesday, November 15, 2017, 11:59
IGP1 = 5.0e-5 Torr
Tapped up entire process manually to 520A in 3hrs!!!
TGHT = 520A, 4.98V
Vacuum coming down nicely indicating carbonization process complete
Will remain at max setpoint for 4hrs to ensure full carbonization before cooling down later this afternoon.
Wednesday, November 15, 2017, 15:16
IG1 = 8.65e-6 Torr
TGHT = 520A, 4.92V
Carbonization complete.
Initiating auto-cool at 2.0A/min
Monday, November 20, 2017, 13:28
Evap1 was vented in argon and sample removed/ transfered to glove box on friday.
Discs look good, slightly lighter grayer compared to production UCx discs after second cast but before final carbonization.
Material seems to adhere to graphite backing foil well, no flakes coming off from just moving discs around.
Weight of 35 carbonized discs = 4.387g
Weight loss = 5.43g - 4.387g = 1.043g
|
Attachment 1: striptool.JPG
|
|
164
|
Wednesday, August 23, 2017, 15:54 |
Laura Lambert | UO2/C | 40282-2 | TARGET BOAT | Done | Carbonizing UC2 discs for 1stepUCx in graphite sample container _#1 | Wednesday, August 23, 2017, 15:41
2 rounds of UO2/C were cast onto a graphite sheet in the fume hood and let dry.
For this round of samples, circles from cast 1 were used
Graphite powder used = 5.16g
A 14mm punch was used to punch out circles from a 16mm wide strip.
35 circles were packed into a graphite sample container.
graphite container [no endcap] = 3.37g
35 x UO2/C circles = 4.76g
graphite container+smpls = 8.13g
container+smpls+endcap = 10.08g
Graphite tube is placed inside a target boat.
Boat previously coated with TaC and sintered - 1st use of newly sintered boat.
Ta-target boat is wrapped with 3-layers of heat shield and installed in Evap1.
Chamber currently pumping down.
Friday, August 25, 2017, 09:59
IG1 = 3.45e-6 Torr.
Stepped up manually to 80A.
Vacuum stable.
IG1 = 3.0e-5 Torr.
Initiating auto-ramp.
Target boat will be ramped up to 520A at the rate of 0.1A/min.
Vacuum limit stop at 1.0e-4 Torr, backing off at 5.0e-4 Torr.
Monday, August 28, 2017, 08:09
IG1 = 6.3e-6 Torr
TGHT = 455A, 4.65V
Vacuum stable.
Increasing ramping rate to 0.2A/min.
Monday, August 28, 2017, 08:53
IG1 = 7.3e-6 Torr
TGHT = 465A, 4.75V
Vacuum stable.
Increasing ramping rate to 0.5A/min.
Monday, August 28, 2017, 13:17
IG1 = 1.32e-5 Torr
TGHT = 520A, 5.44V
Vacuum still fairly high and not dropping very fast indicating carbonization incomplete.
Increasing max setpoint to 530A and restarting auto-ramp.
Monday, August 28, 2017, 15:13
IG1 = 1.32e-5 Torr
TGHT = 530A, 5.58V
Vacuum still high, wavering up and down/ not dropping - carbonization remains incomplete.
Increasing max setpoint to 540A and restarting auto-ramp.
Tuesday, August 29, 2017, 09:28
IG1 = 3.72e-6 Torr
TGHT = 540A, 5.66V
Carbonization complete.
Initiating auto-cool at 2.0A/min
|
Attachment 1: IMG_1768.JPG
|
|
Attachment 2: IMG_1774.JPG
|
|
288
|
Thursday, January 19, 2023, 10:02 |
Darwin Ortiz Rosales | UO2/C | | LP TUBE | Done | C230105 Reduction | Thursday, January 19, 2023, 09:57
Cartridge #8 loaded with 198 discs
To be reduced inside a Ta container tube. Tube wrapped in 3 Ta foil heatshields.
Reduction for 4 hours and then 10 hours @ 425A
Good Vacuum: 5.0e-04 Torr
Bad Vacuum: 5.0e-02 Torr
Initial pressure is 1.41e-06 Torr
Thursday, January 19, 2023, 10:02 Auto ramping to 120A @ 1A/min
Thursday, January 19, 2023, 18:58
Autoramping to 300A @ 4/min (2A/30s)
IGP1=3.34e-06 Torr
Friday, January 20, 2023, 07:37
left at 300 A for 8 hours and pressure dropped. Restarted autoramp at 2A/30s and pressure was increasing until 425A where it peaked. Reduced autoramp to 1A/min and observed slow but definite pressure drop. Will go to 450A and leave there for 4 hours.
Currently 440A, 5.3V on TGHT and 1.28e-05 Torr on IGP1
Friday, January 20, 2023, 12:06
After 4 hours at 450A IGP1= 3.49e-06 Torr
TGHT = 450A 5.4V
Will drop to 425A and leave for 10 hours
Friday, January 20, 2023, 21:53
TGHT 425A, 5.2V
IGP1= 1.81e-06 Torr
autoramping to 0 A @ -2A/min |
286
|
Monday, November 28, 2022, 05:59 |
Darwin Ortiz Rosales | UO2/C | | LP-SIS | Done | C221125experiment | Monday, November 28, 2022, 05:58
Left weekend to pump down
current IGPI= 7.23e-7 Torr
Material will be reduced in LP-SIS wrapped with 3 Ta foils.
a cap was installed to collect material coming out from the ionizer tube
Initial weight of target material plus cartridge was recorded
All relevant data and findings will be reported and added to docushare as document :
Good vacuum = 5e-4 Torr
Bad vacuum = 5e-2 Torr
Power desired 1900W as approximate power used online.
Will autoramp @2A/min to 30A, then slow down ramp to 1A/min to 120A.
After first pressure bump cleared, will ramp up to 400 A @ 2A/20s to 425A on TGHT. TBHT will be topped off at 230A.
from 425A, TGHT will be ramped at 1A/min and monitored for pressure behaeviour.
sytem will be left at reduction current for 4 hours. Then, the current will be dropped to 425 A and allow to run for 8 hours.
The cup will be collected and the system will be run again for two hours. at the same current until no material visible on cap. Then it will be run at 450A for increments of two hours until no significant mass change of target material.
Monday, November 28, 2022, 06:05
started initial heating to 30A at 2A/min on both TGHT and TBHT.
Monday, November 28, 2022, 08:46
was ramping at 6A/min (2A/20s) on both tGHT and TBHT but aborted due to pressure spike. will allow pressure to stabilize.
TGHT = 162A, 1.35V
TBH 162A, 1.9V
Monday, November 28, 2022, 08:51
Restarting the heating at 6A/min (2A/20s)
Monday, November 28, 2022, 09:21
Pausing heating @ TGHT= 323 A 3.0V
Allow pressure to stabilize to better spot reduction current
Monday, November 28, 2022, 09:49
Started heating TGHT at 2A/min so reduction current can be better resolved
Monday, November 28, 2022, 10:12
heating aborted to allow pressure to stabilize
IGP1> 5.0e-4 Torr
Monday, November 28, 2022, 11:31
IGP1= 1.5e-4 Torr start autoramp TGHT @1A/min
Monday, November 28, 2022, 11:51
paused heating. Second peak has been passed. Will let pressure drop before starting heating again to try and better observe reduction peak.
Monday, November 28, 2022, 12:46
start autoramp at 2A/min from 389A on TGHT
IGP1- 6.4e-6 Torr
Monday, November 28, 2022, 14:21
Pressure starts to drop at around 535A, but large bump observed around 386A.
Will leave four hours.
TGHT= 539A, 4.98V
TBHT= 229A, 2.9V
Monday, November 28, 2022, 18:19
Ramping down at 4A/min to 425A
Currently IGP1= 6.2e-6 Torr
TGHT = 539A, 4.98 V
TBHT= 230A, 2.98V
Monday, November 28, 2022, 22:23
425 A on TGHT reached at 6:50 pm. Will ramp down at 2:50 am . -2A/min TGHT and -1A/min TBHT
Tuesday, November 29, 2022, 09:34
Final IGP1 when TGHT and TBHT = 0A is 5.0e-7 Torr
Tuesday, November 29, 2022, 14:08
Weighed materials and reloaded
Currently at 1.1 e-05 Torr
Will initiate heat ramp at 2A/min to 425 A
@241A increase ramp rate to 4A/min on TGHT
Wednesday, November 30, 2022, 12:44
heating B finished and cartridge loading for heating C
Current pressure = 2.09 E-05 Torr. Will ramp up to 450A at 4A/min and regulate as needed.
Will heat at 450A TGHT for 2 hours, TBHT max= 230A
Wednesday, November 30, 2022, 16:30
Starting rampdown @-2A/min TGHT and -1A/min TBHT
Final heating was @ 450A, 4.0V TGHT and 230A, 2.7V TBHT
Friday, December 02, 2022, 09:00
IGP1= 7.2 e-07 Torr
Starting ramp at 2A/min to 450A on TGHT and 230A on TBHT
Friday, December 02, 2022, 10:02
pressure stated dropping. Will increase ramp to 4A/min (2A/30s)
Friday, December 02, 2022, 15:35
Starting ramp down at -2A/min TGHT and -1A/min TBHT
IGP1= 2.6 e-06 Torr
TGHT 450A 4.2V
TBHT 230A 2.9V
|
Attachment 1: Data_summary.png
|
|
287
|
Monday, December 19, 2022, 10:45 |
Darwin Ortiz Rosales | UO2/C | | LP TUBE | Done | C-221207 Reduction | Monday, December 19, 2022, 10:38
Cartridge with 175 discs (including 5 analysis discs) loaded into a graphite cartridge.
To be reduced inside a Ta container tube. Tube wrapped in 3 Ta foil heatshields
Will keep at reducing current for 4 hours and then 10 hours @ 425A
Good Vacuum: 5.0e-04 Torr
Bad Vacuum: 5.0e-02 Torr
Initial pressure was 6.36 e-07 Torr
Autoramped to 56A @ 1A/30s
Outgassing bump observed starting at 50A and at 56A the maximum pressure was ~1.9e-05 Torr
Monday, December 19, 2022, 10:47
started autoramping at 1A/30s to 90A
Monday, December 19, 2022, 10:50
observed that pressure not increasing radically, therefore ramping up faster. 2A/30s
Monday, December 19, 2022, 10:58
Second pressure bump observed at 90A, 1.0V
leaving at this current until pressure drops
Monday, December 19, 2022, 17:15
ramped to 280A and left at 280 until pressure started dropping
ramping again to 500A @ 2A/30s
Monday, December 19, 2022, 17:26
Stopped at 330A due to pressure hike. Will allow pressure to drop then continue ramp.
Monday, December 19, 2022, 18:00
restarting ramp @ 2A/30s
Monday, December 19, 2022, 18:11
Pressure started dropping around 350A corresponding to additive gas-off. Will rampt to 500A @ 2A/30 seconds and then at 1A/30s to 530A
Final current was 548A, 6.7 V. Will test increasing current after 4 hours (11 pm)
Monday, December 19, 2022, 23:10
ramping down @ 2A/min to 425A . should be done by 12:00 am and cooling at9:00 am (includes 1 hour ramp down to 425).
Ramping down to zero @ -2A/min
Wednesday, December 21, 2022, 10:47
Final IGP1= 3.7e-07 Torr |
284
|
Friday, October 07, 2022, 08:14 |
Darwin Ortiz Rosales | UO2/C | | LP-SIS | Done | C-220928UCx conditioning experiments | Friday, October 07, 2022, 07:55
Cartridge-220928UCx was installed in the EVAP1 and pumped down yesterday.
Material will be reduced in LP-SIS wrapped with 3 Ta foils.
IGP1 = 2.1e-6 Torr
a cap was installed to collect material coming out from the ionizer tube
The initial weight of the cap 1 + small bag was recorded to be: 2.82 g Final weight of cap1 + small bag = 2.84g (see attached picture)
Good vacuum = 5e-4 Torr
Bad vacuum = 5e-2 Torr
Power desired 1900W as approximate power used online.
Will autoramp @2A/min to 30A, then slow down ramp to 1A/min to 120A.
After first pressure bump cleared, will ramp up to 400 A @ 2A/20s to 425A on TGHT. TBHT will be topped off at 230A.
from 425A, TGHT will be ramped at 1A/min and monitored for pressure behaeviour.
sytem will be left at reduction current for 4 hours. Then, the current will be dropped to 1900W and allow to run for 8 hours.
The cup will be collected and the system will be run again in increments of two hours.
*slight change from test plan to save time*
Friday, October 07, 2022, 21:55
A true pressure max was not observed. A max while current still rising. stopped at 560A on TGHT and 5.3V
will leave at the above conditions for 4 hours.
Suspect that the thinner Ta shields make a difference. Suggest to add a fourth shield and process the previous cartridge in a tube rather than a boat. Treat for four hours (pressure max) and then test increasing the current to see if there is a pressure bump. If no pressure bump then lower current to deposit 1900W for 8 additional hours.
After 4 hours at max current, the current was dropped to 450A (TGHT) and left for 8 hours.
Saturday, October 08, 2022, 09:26
start autoramp down @-2A/min for TGHT and -1A/min for TBHT
Cap+ bag = 2.82 g
final cap +bag = 2.84 g
Tuesday, October 11, 2022, 10:15
Today will switch end-cup. Will weigh new end-cup and place inside the evap1 glovebox. Then will remove used end-cap and install new end-cap. Used end-cup will be weighed on same balance as before. All mass measurements will be done with the same balance. Note, that a makeshift balance air isolation container was built to allow balance to not be affected by air currents.
Cap 2+bag initial weight = 2.80 g +/- 0.01 g
Cap 2+bag final weight = 2.81g +/- 0.01 g
No visible deposition on cap
Wednesday, October 12, 2022, 07:51
Start ramp up at 2A/min. TGHT to 450A, TBHT to 230A
IGP1= 6.7e-07Torr
Wednesday, October 12, 2022, 09:47
Ramping up @ 4A/min to same final settings: TGHT = 450A and TBHT = 230A
Wednesday, October 12, 2022, 11:15
@10:45am 450A, 4.15V were reached on TGHT
Will leave for 2 hours at 450A then ramp down, then switch cap.
Monday, October 17, 2022, 09:40
The second cap was taken out over the weekend (attachment number 2).
The final weight of this cap with bag = 2.81 +/- 0.01 g
Installed a nex cap (3) with initial weigh of cap +bag = 2.74 g
Allowed vacuum to settle over Sunday and will start ramping today
Will ramp to 470 A and hold for 2 hours. Ramp rate 2A /30s
IGP1= 4.6e-07 Torr
Monday, October 17, 2022, 11:39
ramping with no issues. will finish ramp around 12:03 pm and will start rampdown @ around 2:05 pm.
Monday, October 17, 2022, 12:03
TGHT= 470A, 4.5V
TBHT= 228A, 2.9V
IGP1= 5.6e-06 Torr
Monday, October 17, 2022, 14:05
Start rampdown at 2A/min TGHT and 1A/min TBHT to zero amp
current IGP1= 2.6e-06 Torr
TGHT = 470A, 4.5V
TBHT= 228.3A, 2.9V
Tuesday, October 18, 2022, 09:39
IGP1=4.4e-7 Torr
ready to dismount
Tuesday, October 18, 2022, 13:50
Cap 3 + bag initial weight = 2.74 g
Final weight of cap+bag = 2.74 g (no change)
no visible deposition on cap either |
Attachment 1: Cap_1_221011.png
|
|
Attachment 2: Cap_2.png
|
|
Attachment 3: cap3.png
|
|
285
|
Wednesday, October 19, 2022, 12:28 |
Darwin Ortiz Rosales | UC2 | 40253-5 | LP TUBE | Done | C-220510 further conditioning | Wednesday, October 19, 2022, 10:27
Cartridge-220510Cx was installed in the EVAP1 and pumped down yesterday.
Material will be further conditioned at 450A for 4 hours.
The material will be conditioned inside a container tube with both ends open at the ends. Two springs are inserted at the ends of the graphite cartridge to contain the material as graphite caps not inserted.
IGP1 = 7.8e-07 Torr
Good vacuum = 5e-4 Torr
Bad vacuum = 5e-2 Torr
Will autoramp @2A/30s (4A/min) to 40 A and then to 120A. Fast ramp rate because material had previously been conditioned
Wednesday, October 19, 2022, 12:23
TGHT 450A, 5.2 V
Will leave for 4 hours and then begin ramp-down.
Wednesday, October 19, 2022, 16:21
IGP1= 3.1e-06Torr
450A, 5.2V
Start rampdown @- 2A/min
Thursday, October 20, 2022, 09:23
IGP1= 4.6e-7 Torr
|
24
|
Thursday, June 30, 2011, 12:42 |
John Wong | UO2/C | UCx-heatboat#1 | LP-SIS | Done | Batch#2: UO2/C for carbonization | Before began the carbonization process, two cameras were setup in the hotlab for monitoring the evaporator and the power supply. On duty ISAC Ops also helps to monitor on a regular basis and to grab the striptools + cameras (twice/shift) and put this information in the ISAC Logbook with subject header: "UCx_2011_B 'date time'", for easy search.
P/S camera: http://isaccamera1, user/pw = admin
Evaporator camera: http://isac-cam01, user =admin, pw=yield
Thu Jun 30 12:24:31 2011
UO2 used = 50.29g
Green UO2/C = 71.96g, rolled up /w graphite sheet (15cm x 12cm), and put in the target boat, then wrapped with 3 Ta-heat shields. The target is installed in the evaporator and is being pumped down.
Thu Jun 30 12:45:39 2011: Currently Vac @ 3.75e-5, CG2 @ 0.045T.
Friday, July 01, 2011, 13:28: Vac @ 2.2-6 Torr. TGHT set at 10A yesterday.
Friday, July 01, 2011, 13:42: TGHT stepped up to 36A (panel 33A), 0.134V. No change in vacuum, 2.38e-6 Torr, CG @33mT.
Friday, July 01, 2011, 13:51: TGHT stepped up to 50A (panel 47A), 0.194V. Vac @ 5.7e-6 Torr, CG2 @ 33mT.
Friday, July 01, 2011, 14:09 : Vac was rising slowly. TGHT stepped up to 57A and the pressure started to increase faster. TGHT was brought back down to 30A, and vac is coming down gradually. (See attachment 1#2: Startup Vac/Current)
Friday, July 01, 2011, 14:16 : Auto ramping began. 0.1/min, max currently @ 525A. (see attachement #3: Parameter)
Monday, July 04, 2011, 12:33: TGTH @ 146.20A, Vac @ 4.74e-5 Torr. (see attachment#4: 3-day trend). Increase auto ramping rate to 0.5/min.
Monday, July 04, 2011, 13:01: TGHT @ 148A, Vac @ 5.20e-5 torr. Auto rampiing paused until vacuum goes down to 5.0e-5 torr. (see attachment#4)
Monday, July 04, 2011, 17:34: TGHT @ 220A, VAC started to come down, indicating the burning off organics was done. Increased heating rate to 1A/min.
Friday, July 08, 2011, 21:30: As per Peter Kunz, "I started the target cooldown sequence for IEVAP:TGHT1 at 20:35. The pressure came down to 4e-6 torr. The ramp down rate is 2A/min. Please check if the target heater current comes down to 0A. Then it can cool off over the weekend."
Monday, July 11, 2011, 12:17: Evap glove box flashed with Ar gas, set pressure ~100psi. Chamber also vented with Ar, set pressure @ ~5psi.
|
Attachment 1: startup_vac.jpg
|
|
Attachment 2: startup_current.jpg
|
|
Attachment 3: parameter.jpg
|
|
Attachment 4: UCx_2011_b_3days.png
|
|
Attachment 5: grab.S13888.png
|
|
23
|
Wednesday, May 25, 2011, 14:42 |
John Wong | UO2/C | UCx-heatboat#1 | LP-SIS | Done | Batch#1: UO2/C mixing & carbonization | Before began the carbonization process, two cameras were setup in the hotlab for monitoring the evaporator and the power supply. On duty ISAC Ops also helps to monitor on a regular basis and to grab the striptools + cameras (twice/shift) and put this information in the ISAC Logbook with subject header: "UCx_2011_A 'date time'", for easy search.
Wed May 25 14:39:31 2011: UO2/C cast was dried in the fumehood over the weekend. The cast was pilled off from the coated WD40 glass plate, rolled up and put in a target heating boat (/w TaC coating & sintered). The boat was wrapped with 3 Ta-heat shields, then installed in the evaporator which is currently being pumped down.
Thu May 26 11:40:06 2011: Vac @ 1.75e-6 torr, BP @ 0.032T. TGHT stepped up to 30A/0.092V (panel @ 27A), ; TGHT has been held for over 10 mins, no change in vacuum yet.
Thu May 26 11:46:48 2011: TGHT @ 40A/0.123V (panel @ 37A), Vac 2.3e-6 Torr, BP@0.032T.
Thu May 26 15:13:19 2011: TGHT has been at 40A/0.123V for sometime. Currently, Vac @ 4.39e-5 Torr, BP@ 0.27T. (See Attachement#1)
Thu May 26 17:28:36 2011: Auto ramping has begun. (see attachment#2). TGHT @ 40.60A, (panel @ 38A), vac @ 3.84 e-5 Torr.
Wed Jun 1 13:28:58 2011: TGHT currently at 395A/3.16V, vac @ 4.47e-5 Torr. Heating sequence changed for the increment from 4.5e-5 Torr to 5.0e-5 Torr.
Fri Jun 3 09:54:28 2011 : The organics burnoff and the carbonization process are complete (see Attachment#3 for the striptool). The cooling process will begin at a rate of 2A/min.
Tue Jun 14 21:24:41 2011 : The evaporator was about to be vented by argon on Jun 10 (Fri), but it was found that the argon cylinder was almost empty; a new one was replaced (see Attachment #4: Evaporator_Ar_flow). After the argon cylinder was replaced, the evaporator was purged with the gas and the vent valve was connected to the cylinder for venting of the evaporator. The target oven was removed and stored in the hermetic transfer vessel, then transferred to the glove box. (See attachment #5: UC2, Note: typo. Should read Ar gas, NOT N2). The flakes were stored in the jar labelled "Stock# 2011_A", with total weight of 35g. |
Attachment 1: Screen_shot_2011-05-26_at_3.14.00_PM.png
|
|
Attachment 2: auto_ramping.png
|
|
Attachment 3: UCx_2011_A_carbonization.png
|
|
Attachment 4: Evaporator_Ar_flow.png
|
|
Attachment 5: UC2.png
|
|
289
|
Wednesday, March 22, 2023, 22:29 |
Darwin Ortiz Rosales | UO2/C | | LP TUBE | Done | 230316 Reduction | Wednesday, March 22, 2023, 22:29
Cartridge #9 loaded with 213 plus 3 analysi discs
To be reduced inside a Ta container tube. Tube wrapped in 3 Ta foil heatshields.
Reduction for 4 hours and then 8 hours @ 450A
Good Vacuum: 5.0e-04 Torr
Bad Vacuum: 5.0e-02 Torr
Initial pressure is 3.71e-06 Torr
Wednesday, March 22, 2023, 22:33
Autoramping to 400A @ 1A/min
Thursday, March 23, 2023, 06:57
Starting heating at 2A/min
Currently at 400A and 4.5V
IGP1= 9.6e-06 Torr
At around 500A, the reduction was observed. Current manually increased to 508 A and left there for remainder of four hours. 4 hours will be done at 1210 pm, then will ramp down to 450 and 8 hours
Thursday, March 23, 2023, 12:01
IGP1= 7.6 e-06 Torr
508 A, 6.0 V TGHT
Start rampdown to 450 @ -2A/min
Thursday, March 23, 2023, 20:27
IGP1= 2.36e-06 Torr
TGHT = 452A, 5.3A
Starting ramp down @-2A/min
|
124
|
Thursday, December 10, 2015, 10:59 |
Laura Lambert | UO2/C | 40282-1 | TARGET BOAT | Done | 1stepUCx project - round2 | Thursday, December 10, 2015, 08:54
Filled graphite tube with 34 UO2/C discs cast onto graphite.
34 discs = 4.10g
graphite tube = 4.64g
IG1 = 2.8e-6 Torr.
Stepped up manually to 35A.
Waited ~1hr15, vacuum stable.
IG1 = 5.8e-6 Torr.
Initiating auto-ramp.
Target boat will be ramped up to 520A at the rate of 0.1A/min.
Vacuum limit stop at 5.0e-5 Torr, backing off at 6.0e-5 Torr.
Monday, December 14, 2015, 08:58
IG1 = 3.9e-6 Torr.
TGHT = 520A, 5.7
UO2/C reached 520A yesterday and has remained at max setpoint for ~12hrs.
Carbonization process is complete.
Initiating auto-cool at 2.0A/min.
Friday, December 18, 2015, 15:37
Vented in argon and removed from evap1
Graphite tube appears golden in colour [ see photo attached]
Materials spilled out of graphite tube, could be partially due to movement of target boat between evaporator and glove box and removal of heat shields.
Black graphite tube in weigh boat behind target is the graphite tube from the first round in evap1, shows a comparison in colour difference between the two rounds.
The difference between round 1 and round2:
-round1 contained 5 single discs packed together and a single 1cm/2cm square on the other end
-round2 contained 34 discs packed together consuming the whole graphite tube
All other parameters ie. max setpoints, were the same.
|
Attachment 1: image.jpeg
|
|
123
|
Friday, December 04, 2015, 15:06 |
Laura Lambert | UO2/C | 40282-1 | TARGET BOAT | Done | 1stepUCx project - round1 | Wednesday, December 02, 2015, 14:08
UO2/C was cast onto a graphite sheet in the fume hood and let dry.
Initial UO2 powder used = 25.11g (Inventory R-00835A, UO2 + bottle = 243.44g before use)
Graphite powder used = 5.16g
A 13mm punch was used to punch out discs from an 11mm wide strip.
5 discs were packed into one end of the graphite tube and a single 1cmx2cm strip was placed in the other end.
graphite tube = 4.57g
5 x UO2/C discs = 0.62g
2cm x 1cm strip = 0.21g
the graphite tube is placed inside a target boat.
The Ta-target boat is wrapped with 3-layers of heat shield and installed in Evap1.
Pumping down overnight.
Thursday, December 03, 2015, 09:23
IG1 = 5.7e-6 Torr.
Stepped up manually to 60A.
Waited ~1hr45, vacuum stable.
IG1 = 6.3e-6 Torr.
Initiating auto-ramp.
Target boat will be ramped up to 520A at the rate of 0.1A/min.
Vacuum limit stop at 5.0e-5 Torr, backing off at 6.0e-5 Torr.
Monday, December 07, 2015, 10:45
UO2/C reached 520A yesterday and has remained at max setpoint for 15hrs.
IG1 = 5.2e-6 Torr.
TGHT = 520A, 5.35V
Carbonization process is complete.
Initiating auto-cool at 2.0A/min.
Wednesday, December 09, 2015, 10:09
Vented in argon and took the materials out yesterday.
Currently sitting in glove box.
All five discs and strip had fallen out of graphite tube and were sitting directly on the target boat.
^not sure if due to removal of heat shields [moving target boat around shook materials out of tube] or if this happened during carbonization process.
materials stuck to graphite foil fairly well, able to handle foils with tweezers and inspect without materials flaking off.
lightly scraped top of discs with spatula - no materials flaked off.
minor cracks are visible on discs.
graphite tube shows no visible changes.
|
156
|
Thursday, July 13, 2017, 12:49 |
Laura Lambert | | | | Problems | *hot lab is HOT* | Thursday, July 13, 2017, 12:48
NOTE: hot lab reached 27degrees every day this week and went up to 28C today!
This is with the small portable fan running.
27C was reached even with both evaporators off.
When at 28C: evap2 = ~1500C and evap1 = 35A
Thursday, July 20, 2017, 15:58
Hot lab reached 30C today.
Small fan off - unsure if able to run due to contamination spread...
Both evap1 and evap2 were under vacuum but had not initiated ramping up yet.
|
Attachment 1: Capture.JPG
|
|
163
|
Thursday, July 27, 2017, 09:08 |
Laura Lambert | | | | Problems resolved | *hot lab is HOT* |
Laura Lambert wrote: |
Thursday, July 13, 2017, 12:48
NOTE: hot lab reached 27degrees every day this week and went up to 28C today!
This is with the small portable fan running.
27C was reached even with both evaporators off.
When at 28C: evap2 = ~1500C and evap1 = 35A
Thursday, July 20, 2017, 15:58
Hot lab reached 30C today.
Small fan off - unsure if able to run due to contamination spread...
Both evap1 and evap2 were under vacuum but had not initiated ramping up yet.
|
Thursday, July 27, 2017, 09:06
There was a broken valve motor affecting all ISAC rooms - temporary fix in place.
Ventilation in ISAC hot lab has been increased.
Temperature back to a reasonable level - currently sitting at 22C!
|
130
|
Tuesday, April 19, 2016, 10:48 |
Laura Lambert | UC2 | *** | | Done | ***ALL FUTURE UCx TARGETS*** | Tuesday, April 19, 2016, 09:47
Temperature measurements have been taken in evap2 on target containers.
Previously all UCx targets have been conditioned in evap1 at 520A. The respective target containers TaC coating was sintered in evap2 at a different current.
All future UCx targets will now be conditioned in evap1 at the same current required to reach 1900C in evap2 when sintering the TaC coating on the container to be used for the respective target.
ie. If a container is being sintered in evap2 at 450A [1900C], this same container is used for UCx - then that specific UCx target will be conditioned in evap1 at 450A.
Confirm full carbonization has been achieved by following with the vacuum on the strip tool and make sure the final vacuum comes down at the end of the whole process.
|
20
|
Wednesday, May 11, 2011, 16:23 |
John Wong | SiC | 38481-8 | HP-FEBIAD | Problems | Organics burnoff & Sintering | Wed May 11 16:20:21 2011: The HP-FEBIAD (w.o#38481-8) was filled with SiC (8.5mm, 555 discs). The target is wrapped with 3 Ta-heat shields and is in the evaporator, which is being pumped down.
Thu May 12 10:52:48 2011: The evaporator ion gauge, IEVAPIG1 filament burnt out. The gauge comes in two filaments, so it is now running on the second one. The chamber has been pumped down overnight, currently it's @ 2.29e-6 torr.
Thu May 12 11:15:59 2011: TGHT was stepped up to 80A, vac @ 2.47e-6 Torr. Begin auto heating at 1A/min to 820A and will increase the rate to 2A/min if the vacuum is stable.
Thu May 12 12:27:35 2011: TGHT went up to 101A, and the vacuum started to reach close to the trip limit (8.7e-6 Torr). TGHT was auto ramped down but the vacuum kept rising and tripped off the heater @ 94A. TGHT was turned back on after 30mins, vac @ 8.7e-6 Torr. TGHT was tapped up slowly to 95A. The ramping rate has also decreased to 0.5A/min. (See attachment#1: high_vacuum)
Thu May 12 12:58:51 2011: Lowered TGHT to 90A. Vac stayed at 7.8e-5 Torr.
Fri May 13 15:47:42 2011: TGHT @ 128A, incrementing rate @ 1A/min, Vac @ 5.4e-5 Torr after 30 hrs. (See Attachment#2: current_pressure_30hrs)
Fri May 13 16:22:54 2011: THGT and vacuum over the past 12 hrs. (See Attachment#3: current_pressure_past12hrs)
Mon May 16 11:03:54 2011: THGT reached to 820A and held for over 12 hrs. (See Attachment#4_Sintering process).
Tue May 17 16:26:34 2011
Problem Occurred:
- The end caps of the target got vaporized after the sintering process. (See Attachment#5: end_caps_vaporized)
- The caps were not coated with TaC and a D-shaped graphite foil were used.
Solutions:
- The caps were removed - No damage to the target oven. (See Attachment#6/#7: left_end/ right_end )
- New caps were used and coated with TaC. Two circled-graphite foils were used on the both ends and then sealed with the TaC coated caps. (See Attachment 8: TaC_coated_endcaps)
- Re-heating the target to 820A.
Tue May 17 16:35:13 2011: Target is wrapped with 3 Ta-heat shields. The target is in the evaporator which is being pumped down.
Tue May 17 19:34:55 2011: TGHT stepped up to 80A, Vac holding at 5.2e-6 Torr. Auto heating began after 25 mins later. Rate set at 2A/min to max 820A.
Tue May 17 22:02:36 2011: TGHT @ 382A, Vac @ 1.23 e-5 Torr. Increased ramping rate @ 3A/min. Cooling down process will begin at 04:00 of May18, 2011; ISAC Ops will help to start the auto cooling process.
Wed May 18 11:39:47 2011: The cooling down process ended; TGHT reached to zero. The camber will be vented.
|
Attachment 1: high_vacuum.jpg
|
|
Attachment 2: current_pressure_30hrs.jpg
|
|
Attachment 3: current_pressure_past12hrs.jpg
|
|
Attachment 4: Sintering_process.jpg
|
|
Attachment 5: end_caps_vaporized.jpg
|
|
Attachment 6: Left_end.jpg
|
|
Attachment 7: right_end.jpg
|
|
Attachment 8: TaC_coated_endcaps.jpg
|
|
Attachment 9: SiC_EDX.jpg
|
|
|